OPERATING PRECAUTIONS
GENERAL PRECAUTIONS
.
Federal Law restricts this device to sale by or on order of a physician.
.
Infant radiant warmers should be used only by properly trained personnel as directed by an appropriately qualified physician aware of currently known hazards and benefits.
.
The functional checkout procedure should be performed before each use and after disassembly for cleaning, servicing or maintenance. Refer to qualified service personnel if the unit does not perform as specified.
.
The Cart/Bassinet End and Side panels cannot be used for pushing or pulling the Cart/Bassinet and or
Warmer.
.
Do not leave the infant unattended in the Cart/Bassinet when the side panels orthefront panel arefolded down.
.
To avoid overheating or underheating, skin temperature must be continuously monitored and controlled either manually or automatically. Rectal temperature should never be used to control skin temperature.
.
To avoid overheating or underheating when operating in manual mode, observe the infant constantly and monitor the temperature using the temperature probe supplied with the equipment or another electronic thermometer.
.
The skin temperature sensing probe must be in direct contact with the skin to provide accurate monitoring of the infant’s skin temperature. Failure to maintain direct skin contact can result in overheating and possible burning. Check infant’s condition at least every fifteen minutes for correct Sensor attachment, reddened skin areas, and proper skin temperature.
.
The skin temperature sensing probe should be placed at least 1.5 inches from any transcutaneous monitor (TcPOz or TcPCOz) probe to prevent false temperature indications. The skin temperature probe should not be placed on an area previously used by a TcPOz or TcPC02 probe.
.
To avoid overheating the skin, the location of the skin temperature probe must be such that the skin around the Sensor is in direct line with the radiation from the warmer. Do not place anything between the radiant warmer and the infant that will interfere with the radiation from the warmer.
.
Radiant warming increases insensible water loss. Appropriate measures to maintain proper fluid balance should be considered.
.
For effective heating, make sure the Warmer Module is positioned on the center axis and locked with its detent.
.
This Warmer is intended for use with the Cart/Bassinet. If the Warmer is used to warm the mother immediately after delivery, all warnings and instructions should be followed with particular regard to the heater-to-mattress distance and skin probe location. Avoid placement of objects near the mother which can absorb heat and become hot to the touch.
.
Compressed gas cylinders, such as oxygen cylinders, can become hazardous projectiles if the gas is released rapidly due to damage or other causes. Cylinders must be securely fastened.
.
To avoid overheating the Warmer, do not place objects (equipment, blankets, clothing or sterile packs) on top of the Warmer.
COPYRIGHT@3 1998, HILL-ROM AIR-SHIELDS, INC.
OPERATING PRECAUTIONS (Continued)
l Avoid placement of objects between the infant and the Warmer Module that can block heat transfer or absorb heat. These objects may be heated directly by the heater and can cause skin burns.
a
Temperature uniformity (per IEC 601-2-21) across the mattress surface may not be maintained when the Bassinet is tilted in the 5- and 1 O-degree positions.
l
During calibration, inspect the secondary reflector directly under the warmer heater element for particles.
If particles are present, replace the heater element. The life expectancy of the heater element is 1000 hours of operation.
0
Evidence that the safety certification of the ACCESSORY has been performed in accordance with the appropriate IEC 601-l and/or IEC 601-l-l harmonized national standard.
RESUSCITATION PRECAUTIONS
A one-way valve is installed at the Patient Outlet connection. This valve opens when pressure in the hose delivering gas to the patient falls below -4 cm H20. Its purpose is to allow patient inspiration in the unlikely event of failure of the gas supply.
A humidifier, when used, must be placed between the Patient Outlet connection and the patient circuit.
DO NOT PLACE A HUMIDIFIER IN THE SUPPLY LINE.
Humidifier must have low series flow resistance and not be placed between exhalation valve and patient airway.
Gas going to the patient should not be super-saturated as evidenced by excessive condensation in the tubing. Condensation droplets on the inner walls of the tubing are normal.
Periodic blood gas measurements should be made to ensure proper levels of ventilation.
Always confirm the airway pressure relief valve setting before patient use.
Confirm that the oxygen/air blender control of the Blended Gas Supply Module is correctly set prior to use.
Confirm that the patient circuit contains all the parts needed prior to use.
Ensure that the patient breathing circuit connections are secure and free of obstructions.
The Aux Outlet circuit does not provide adjustable pressure limiting.
Auxiliary Gas Supply Pressure is internally limited to 160 cm circuit that includes appropriate pressure relief.
HzO. Always use a patient resuscitation
Always monitor Airway Pressure and/or provide appropriate pressure relief during infant resuscitation.
Confirm that the oxygen supply is turned off and that the equipment is disconnected from the oxygen supply when performing cleaning and maintenance procedures; a fire and explosion hazard exists when performing cleaning and/or maintenance procedures in an oxygen-enriched environment.
Gas supplies (02 and Air) should always be clean and dry. Water/Trap Filters should be used in the supply lines if necessary.
All hoses should be securely fastened to fittings. Hand-tighten to avoid damage to fittings.
Gas supplies should be maintained at 40 to 75 psi.
Flow restrictions (e.g., flowmeter, valve, etc.) must not be placed in the supply line.
If a compressor is used as the air source, steps should be taken to filter and dehumidify the room air before introducing it into the Gas Supply Module.
_ ii _
OPERATING PRECAUTIONS (Continued)
ELECTRICAL PRECAUTIONS a
An electrical shock hazard exists within the Warmer Module and Controller. Any substitution of components within the Controller or Warmer Module may impair the intrinsic safety of the unit. Service should be performed by qualified personnel.
0
Connect the power cord only to a properly grounded receptacle that is approved for hospital use and of the correct voltage. DO NOT USE EXTENSION CORDS. l
Use only with power cords supplied with the Warmer.
EXPLOSION PRECAUTIONS
l
Do not use in the presence of flammable anesthetics.
0
Confirm that the oxygen supply is turned off and that the equipment is disconnected from the oxygen supply when performing cleaning and maintenance procedures; a fire and explosion hazard exists when performing cleaning and/or maintenance procedures in an oxygen-enriched environment.
.
OXYGEN PRECAUTIONS
l
Improper use of supplemental oxygen may be associated with serious side effects including blindness, brain damage, and death. The risks vary with each infant. The method, the concentration, and the duration of oxygen administration should be prescribed by the attending physician.
If it is necessary to administer oxygen in an emergency, the attending physician should be notified immediately.
NOTE: See the current edition of “Guidelines for Perinatal Care”of the American Academy of Pediatrics/The
American College of Obstetricians and Gynecologists.
.
.
.
.
The oxygen concentration inspired by an infant does not predictably determine the partial pressure of oxygen (PO,) in the blood. When deemed advisable by the attending physician, blood PO2 should be measured by accepted clinical techniques.
Oxygen flow rates cannot be used as an accurate indication of oxygen concentrations. Oxygen concentrations should be measured with acalibrated oxygen analyzer at intervals directed by the attending physician.
Keep matches and all other sources of ignition out of the room in which the Warmer is located and keep spark-producing equipment or any sources of ignition out of the Bassinet. Wood, textiles, oils, and other combustibles are easily ignited and burn with great intensity in air enriched with oxygen.
Although oxygen compatible materials are used in the oxygen delivery system, special care must be taken when high pressure oxygen such as found in a medical oxygen cylinder is used. Violent ignition of oil, grease, greasy substances, small particles of dust, dirt or other particulate contaminants (even small particles of metal), can occur in the presence of high pressure oxygen if their ignition temperature is reached. An instantaneous increase in temperature can occur due to friction, particle acceleration, or adiabatic compression, if the oxygen cylinder valve is opened too rapidly. SERIOUS INJURY MAY RE-
SULT! Always observe the following precautions:
Oil, grease, greasy substances, dust, dirt and other particulate contaminants must be kept away from oxygen regulators, cylinder valves, tubing and all other oxygen equipment.
-
III -
OPERATING PRECAUTIONS (Continued)
-
Always open oxygen cylinder shut-off valves very slowly and carefully.
-
On high pressure oxygen cylinders use only pressure regulators or reducing valves approved for oxygen service. Do not use oxygen pressure regulators or reducing valves for air or gases other than oxygen as they may be hazardous. Operate such devices in strict accordance with the manufacturer’s recommendations.
When new or replacement oxygen cylinders are to be installed, they should have their outlet ports cleared by cracking the cylinder valve momentarily before attachment to the equipment.
LOW FLOW MICROBLENDER WARNINGS
l
If either the air or oxygen gas source pressure is reduced or increased creating a pressure differential of 30 psi, the microblender alarm will sound. This condition significantly alters the
FiO2 and flow output from the microblender.
l
Always operate the low flow microblender with clean/dry medical grade gases.
.
Air inlet water filters are recommended for use with the low flow microblender.
.
The level of oxygen and the partial pressure of oxygen within the patient’s blood should be monitored.
(Pa02)
- iv -
TABLE OF DEFINITIONS AND SYMBOLS
NOTE, IMPORTANT, PRECAUTION, CAUTION, AND WARNING
NOTE: A Note is inserted in text to point outprocedures or conditions which may otherwise be misinterpreted or overlooked.
A Note may also be used to clarify apparent/y contradictory or confusing situations.
IMPORTANT Similar to a Note but used when greater emphasis is required.
PRECAUTION: A Precaution is supplementalinformation equipment.
to assist the
user in the safe and effective use of the
CAUTION: A Caution is inserted in text to call attention to a procedure which, if not followed exactly, can lead to damage or destruction of the equipment.
WARNING: A Warning is inserted in text to call attention to dangerous or hazardous conditions inherent to the operation, cleaning, and maintenance of the equipment which may result In personal injury or death of the operator or patient.
SYMBOLS
Attention: consult accompanying documents.
Type B equipment with an F-type isolated (floating) applied part.
-
\
’
0
8
I
o-
’
/
A
8
Danger! High Voltage!
Examination Light
Examination Light Switch
Mode Control Key
Temperature Override Mode
Key
Keypad Lock Key
Disposable Suction Bottle
Reusable Suction Bottle
00
Set Temperature Keys
I Power On/Off Switch
Celsius/Fahrenheit Selection
Key
Patient
Heater Element
Silence/Reset Key
Suction Line Filter
Procedural Silence Indicator
w@
Apgar Timer Keys
Load Symbol
-V-
TABLE OF CONTENTS
SECTION
PAGE
1 GENERALINFORMATION..
.........................................................
1
1.1 INTRODUCTION..
......................................................................
1.2 DESCRIPTION
.........................................................................
1.3 SPECIFICATIONS
....................................................................... 1
1
1
1.4 EQUIPMENTPROVIDED
................................................................
1.5 FACTORY INSTALLED ACCESSORIES
3
................................................... 4
1.6 FIELD INSTALLED ACCESSORIES (Refer to Section 6 for Part Numbers)
.................... 4
2 INSTALLATION .....................................................................
6
2.1 UNPACKING
...........................................................................
2.2 ASSEMBLY (Refer to Figure 2.1)
6
.......................................................... 6
2.3 ASSEMBLY OF CART/BASSINET
......................................................... 6
3 FUNCTIONAL DESCRIPTION
........................................................
8
3.1 GENERAL ..............................................................................
8
3.2 FUNCTIONALDESCRIPTION
............................................................
8
3.2.1 WARMERMODULE..
..............................................................
8
3.2.2 CART/BASSINET
..................................................................
8
3.2.3 CONTROLLER
....................................................................8
3.2.4 BLENDER MODULE (OPTIONAL)
.................................................... 8
3.2.5 RESUSCITATION MODULE (OPTIONAL)
3.2.6 GAS SUPPLY MODULE (OPTIONAL)
............................................. 8
............................................... 12
3.2.7 ALARMS .........................................................................
3.2.8 APGARTIMER
13
...................................................................13
4 OPERATION
......................................................................
14
4.1 CONTROLS, INDICATORS AND CONNECTORS
.......................................... 14
4.2 OPERATIONAL CHECKOUT PROCEDURE
4.3 MECHANICALCHECKOUT
.............................................................
4.4 RESUSCITATION EQUIPMENT PRE-USE CHECKOUT AND SETUP (OPTIONAL)
4.4.1 SUPPLYPRESSURE
- CONTROLLER
............................... 24
..............................................................
25
............ 27
27
4.4.2 BLENDED GAS SUPPLY PRE-USE CHECKOUT AND SETUP (OPTIONAL)
............. 27
4.4.3 RESUSCITATION MODULE PRE-USE CHECKOUT AND SETUP (OPTIONAL)
.......... 28
4.5 CONTROLLEROPERATION
............................................................
29
4.5.1
PRE-WARMMODE ...............................................................
29
4.5.2 MANUALMODE
..................................................................
29
4.5.3 BABYMODE .....................................................................
30
4.5.4 EXAMINATION LIGHT .............................................................
31
4.6 CART/BASSINET ......................................................................
32
4.7 DOCKlNGANDUNDOCKlNGTHECART/BASSlNET
......................................
4.8 X-RAYPROCEDURES
.................................................................
32
32
5
CLEANINGANDMAINTENANCE
...................................................
33
5.1 GENERAL .............................................................................
33
5.2 CLEANING ............................................................................
33
5.3 DISASSEMBLYFORCLEANING
.........................................................
33
5.4 CLEANINGPROCEDURES
.............................................................
34
5.4.1 CLEANING AGENTS..
............................................................
5.4.2 PAINTEDSURFACES
.............................................................
54.3 CLEAR PLASTIC AND ACRYLIC SURFACES
34
34
........................................ 34
-vi-
TABLE OF CONTENTS (continued)
SECTION
5.4.4 METAL SURFACES ...............................................................
5.4.5 WOODEN SURFACES
............................................................
5.4.6 SKIN TEMPERATURE PROBE (REUSABLE) .........................................
5.5 STERILIZATION
.......................................................................
5.6 REASSEMBLY AFTER CLEANING
......................................................
5.7 CALIBRATION .........................................................................
5.8 TROUBLESHOOTING ..................................................................
6 PARTS LIST .......................................................................
6.1 GENERAL .............................................................................
PAGE
34
34
35
35
35
35
35
37
37
-vii-
RESUSCITAIREB B
IRTHING
R
OOM
W
ARMER
SECTION 1
GENERAL INFORMATION
1 .l INTRODUCTION
This manual provides instructions for installation, use, operator maintenance and troubleshooting of the equipment. Hill-Rom Air-Shields cannot be responsible for the performance of the equipment if the user does not operate the equipment in accordance with the instructions, fails to follow the maintenance recommendations in Section 5 of this manual or effects any repairs with unauthorized components. Calibration and repair should be performed only by qualified service personnel. Service manuals are available from Hill-Rom Air-Shields.
This manual should be read, thoroughly understood, and be readily accessible to all personnel who will be working with the equipment. The manual should be stored with the equipment when not in use. If there is anything you do not understand, please contact your Hill-Rom Air-Shields’ representative for further information.
1.2 DESCRIPTION
The ResuscitaireB Birthing Room Warmer is designed specifically for birthing room use. The
Re-
suscitaire8 Birthing Room Warmer consists of a detachable Cart/Bassinet, Warmer, and a Controller module which provides heat control, monitoring of skin temperature and Apgar timing. The
Resus-
citaire8 Birthing Room Warmer also includes an optional resuscitation package with suction and oxygen delivery. Other options include a reserve gas supply and blender.
1.3 SPECIFICATIONS
Specifications for the
ResoscitaireB Birthing
Room
Warmer are provided in Table 1 .l . All specifications are subject to change without notice.
-l-
RESUSCITAIREGI B
IRTHING
R
OOM
W
ARMER
TABLE 1.1 SPECIFICATIONS
POWER REQUIREMENTS ...........................................
12OV, 60 Hz, 750W
OVERLOAD PROTECTION
..................................... Dual 12A Circuit Breakers
CHASSIS LEAKAGE CURRENT ....................................... Less than 300 uA
EXAMINATION LIGHT.
..............................
>lOO Foot Candles (0.12 lumens/cm:z)
ALARMS
High Temperature
Check Patient
................................ Activates if Skin Temperature Probe is attached and the skin temperature sensor reaches 39.0 “C. Resets at 38.5 “(3.
................................ Activates in Manual Mode after 10 minutes.
Apgar Timer
Remains on with audible alarm every
30 seconds for 5 minutes; totalling 15 minutes. Then the heater is turned Off.
............................................Activates at the l-,
5-and lominute Apgar Time intervals.
Power Fail
Probe..
................................................... Activates when there is a
.......................................
loss of power.
Activates if Skin Temperature Probe fails
SystemFail ..........................................
(open or short).
Indicates system failure; refer unit to service immediately.
BabyTemp ..............................................
Electrical Module Audio Alarms
Activates if Baby Temperature fluctuates 1 “C above or below set poirit.
....................... Tone Frequency: 1.2 KHz maximum
Blender Module Pneumatic Audio Alarm
Three-stage sound level: 15 seconcls low, 15 seconds medium, then high.
..................................Vibrating Reed.
DISPLAYS
Skin Temperature Display
Accuracy..
Resolution
.................................
f 0.2 “C for 31 “C to 37 “C (88 “F to 98.6
.............................................................0.1 “C (0.5
“F)
“F)
Apgar Timer Display
Range ...............................................
Resolution .................................................................
Accuracy .............................................................
MANUAL HEAT CONTROL
0 to 59 minutes, 0 to 59 seconcls lsecond
Of0.5second
..................................Adjustable in 10% increments from zero to full power (100 %#)
DATA PORT .....................................................
2400 Bits/second fixed
Baud Rate. RS232C Compatible
DIMENSIONS AND WEIGHT-FREE STANDING WARMER
Height ...........................................................
Width (side to side) ................................................
Depth (front to back)
74 inches (188 cm)
28 inches (102 cm)
................................................
.40 inches (89 cm)
Weight (without tanks) ................................................. 170 Ibs (77 kg)
Warmer Head Rotation (right and left) .......................................
90 Degrees
-2-
RESUSCITAIREGO B
IRTHING
R
OOM
W
ARMER
TABLE 1 .I SPECIFICATIONS (Cont.)
DIMENSIONS AND WEIGHT - Cart/Bassinet
Height to Mattress ................................................
Width (to outside of bumper)
38.5 inches (98 cm)
......................................... 24 inches (61 cm)
Length (to outside bumper)
Weight
.......................................... 31 inches (79 cm)
...........................................................
c 160 Ibs (c73 Kg)
MattressTilt .................................
5 and 10 degrees Reverse Trendelenburg.
Infant’s head should be toward front of warmer.
TOTAL SYSTEM DIMENSIONS AND WEIGHT
Height .........................................................
Width .............................................................
Length ..........................................................
Weight ........................................................
..74inches(188cm)
28inches (71 cm)
48inches(122cm)
..c320Ibs(c152Kg )
ACCESSORY WEIGHT LIMITS
Infusion Pump/IV Pole ...................................................
Monitor Shelf
5 Ibs (2.2 Kg)
.......................................................... 10 Ibs (4.5 Kg)
ENVIRONMENTAL
Operating Temperature Range
Storage Temperature Range
.................................. 18
“C to 30 “C ambfent
..................................
Relative Humidity Operating Range
- 20 “C to
.................. 5% RH to 95% RH,
+45 “C ambient non-condensrng
RESUSCITATION
Wall Supply Pressure
...................................................
Cylinder Pressure .....................................................
Cylinder Diameter
............................................
4-5
40 to
2900
75 psi psi max inches( 1 O-l 2 cm)
Patient Gas Supply
Airway Pressure Limiting Operator-Adjustable
Fixed Pressure Relief, Factory Set .....................................
Suction Circuit
Adjustable Suction Intensity ................................
0 to 50 cm
.. . 60 cm
0 to 150
Hz0 (4.9 kPa) + 10%
Hz0 (5.9 kPa) 2 10% mmHg (0 to 20 kPa)
Auxiliary Supply
Auxiliary Flow Range .....................................................
Auxiliary Pressure Limit
.................................... 160 cm
0 to 15 LPM
Hz0 (16 kPa) -c 10%
.
1.4 EQUIPMENT PROVIDED
.
Bassinet
- The Bassinet provides maximum visibility and access to the infant. The Bassinet tilts up in the rear to 5 and 10 degrees and provides for X-ray Cassette tray (optional) insertion.
Warmer Module
- The Warmer Module houses a heating element and an Examination Light for special procedures.
.
.
Controller
- The Controller provides prewarm, manual heat control, automatic skin temperature servo-control and contains an
Apgar Timer, Skin Temperature monitor and probe.
Resuscitation Module (optional) -The Resuscitation Module contains a suction circuit, a patient oxygen delivery circuit with airway pressure relief and an auxiliary oxygen delivery circuit.
-3-
RESUSCITAIR~ B
IRTHING
R
OOM
W
ARMER
.
.
1.5 FACTORY INSTALLED OPTIONS
.
Resuscitation Module
Integrated Precision Blender
Gas Supply Module
-
02 Pipeline and Cylinder
-
Oz/Air Pipeline and Cylinder
.
.
.
.
.
.
1.6
FIELD INSTALLED ACCESSORIES
(Refer to Section 6 for Part Numbers)
X-ray Cassette Tray
Air Hose Assembly
Oxygen Hose Assembly
Monitor Shelf
I.V.
Pole
Drawer Organizer
-4-
RESUSCITAIREB B
IRTHING
R
OOM
W
ARMER
DRAWER
LOCK
PASS-THROUGH
DRAWER
WARMER MODULE
MONITOR SHELF
3c.
iT TILT
MtwiANISM
\
X-RAY TRAY 4
WALL 02/AIR
GAS HOSES
- GAS TANKS
(NOT SUPPLIED)
SUCTION BOTTLE
CAiTER LOCK
CASTER LOCK
CONTROLLER
/
BLENDED GAS
SUPPLY MODULE
RESUSCITATION
GAS SUPPLY MODULE
FIGURE 1
PASS-THROUGH DRAWER
.I EQUIPMENT PROVIDED WITH
FACTORY INSTALLED OPTIONS AND
FIELD INSTALLED ACCESSORIES
-5-
RESUSCITAIREGQ B
IRTHING
R
OOM
W
ARMER
2.1 UNPACKING
The
ResuscitaireB Birthing Room Warmer
is shipped in two cartons. One carton contains the following assemblies:
.
Upper Post Assembly
.
Lower Post Assembly
.
Warmer Module Assembly
.
Post Handle
.
Any user installed Accessories that were ordered
The other carton contains the Cart/Bassinet.
2.2
ASSEMBLY OF THE WARMER
(RE-
FER TO FIGURE 2.1)
When removing the equipment from the carton, use care not to scratch or otherwise damage unprotected surfaces; remove all packing material.
NOTE: If a Monitor Shelf is to be installed on the unit, refer to the instructions provided with the Monitor Shelf and install the weights provided with these accessories in the bottom of the Lower Post.
1.
REMOVE THE CONTROLLER (3) from the
Upper Post (2).
CAUTION: Two persons of sufficient strength
are required when performing Steps
3 through 6.
2.
3.
REST THE UPPER POST (2) on top of the
Lower Post opening.
CONNECT THE SUCTION HOSE (4) to the
Suction Hose (11). Release the joined Suction
Hoses from the Neat Clip inside the Lower Column.
4.
5.
6.
MOUNT THE UPPER POST (2) on the Lower
Post using four 10 - 32 x 3/8 inch screws (16) provided.
INSTALL TWO 10
- 32 X 3/8 INCH SCREWS
(6) IN THE UPPER HOLES OF THE UPPER
POST. Do not tighten the screws.
RAISE THE WARMER (7) above the open end of the Upper Post (2) and insert the Power
Cable (10) into the open end of the column.
SECTION 2
INSTALLATION
7.
8.
9.
10
11.
12.
13.
SLOWLY LOWER THE WARMER (7) onto the
Upper Post. Tighten the upper screws and install the remaining four 10 screws (6).
- 32 x 3/8 inch
THREAD THE WARMER POWER CABLE out through the Controller opening. Connect the Power Cable (10) to connector J4 on the
Controller (3).
REMOUNTTHE CONTROLLER on the Upper
Post.
MOUNT THE BACK COVER (1) on the Upper
Column using the six 8 - 32 x 3/4 painted screws (5).
MOUNT THE HANDLE (12) on the Post using the two 10 - 32 x 2.6 inch Phillips Head
Screws, No. 10 Lock Washers and Spacers
(13 1 (14 1 (15 ).
CONNECT THE LINE CORD to the Power
Connector on the rear of the Controller (refer to
Figure 4.2).
SECURE THE LINE CORD to the Back Cover
(1) using the Cable Clamp (8) and 8 - 32 x 3/8 screw (9).
CAUTION: Securing the Line Cord to the Back
Panel is required to prevent removal of the Controller chassis with the ac power applied.
14.
2.3
1.
2.
INSTALL ANY ACCESSORIES that were ordered using the installation instructions provided with the accessory.
ASSEMBLY OF THE CART/BASSINET
Refer to Paragraph 5.6, Step 4, and install the
Rear and Side Panels. Raise the Front Panel from its storage slot and lock it in place.
ENGAGE THE CART/BASSINET TO THE
WARMER by pushing the Cart/Bassinet straight onto the Warmer until it latches. Pull on the Cart/Bassinet to ensure that it is latched properly.
To disengage the Cart/Bassinet from the
Warmer, push and hold the release switch
(Figure 4.14) and pull the Cart/Bassinet away from the Warmer.
-6
RESUSCITAIREB B
IRTHING
R
OOM
W
ARMER
PARTS LIST
Screw, lo-32~3/8(QtylO) .
Screw, 10
...99041 36
- 32 x 2.6 (Qty 2) . . . . .99 045 30
Screw, 8 - 32 x
Screw, 8 -32 x
3/4 (Qty 6) . . . .
3/8 (Qty 1) . . . . .
.99 032 85
.99 031 38
No.iOLockWasher . . . . . . . . . . . . 9912416
Spacer (Qty 2) . . . . . . . . . . . . . . . .
Cable Clamp
Post Handle . : : : : 1: : : : : : : : 1: : .
.99 124 61
..1772564
.81 005 73
FIGURE 2.1 INSTALLATION
-7-
RESUSCITAIREB B
IRTHING
R
OOM
W
ARMER
SECTION 3
FUNCTIONAL DESCRIPTION
3.1 GENERAL
This section provides a functional description of the equipment.
3.2 FUNCTIONAL DESCRIPTION
3.2.1 WARMER MODULE
The Warmer is controlled by a Controller which provides Pre-Warm Mode, Manual Mode heater control, or Baby Mode (automatic skin temperature control). An Examination Light provides added illumination of the mattress area. A Warmer Head Pivot permits the Warmer to be pivoted to either side for
X-ray procedures. In addition, when the Warmer is pivoted, it can be powered and provide heat.
3.2.2 CART/BASSINET
The Bassinet is designed to provide maximum function and utility to aid in the care of the newborn. It is detachable so that the infant can be transported to the NICU, the general nursery or another area of the hospital. The side and front panels may be folded down to permit maximum access to the infant. The mattress may be tilted up in the rear at a 0, 5- or lo-degree angle. Openings are provided on each side of the Bassinet for the insertion of the optional
X-ray Cassette Tray.
3.2.3 CONTROLLER
At power-up, the microprocessor within the Controller performs a series of diagnostic tests to confirm the proper operation of the system. During this time, all displays and indicators are lighted (except
Power Fail) and an audible tone is sounded.
When powered up, the system initializes in
100% power and maintain that setting
Pre-
Warm Mode, the Controller will start the heater at forthree minutes, reduce to 60% for 12 minutes and then reduce the heater power to 30%.
When operating the Controller in the Manual Mode, the operator can adjust the heater power from 0 to full power. After 10 minutes of operation in the Manual Mode, a Chk Patient Alarm occurs.
Failure to acknowledge the Check Patient Alarm by pressing the Silence/Reset Key within the next 5 minutes will cause the heater to be shut down.
The Apgar Timer displays the elapsed time and sounds an audible dual tone to alert the operator that 1, 5, and 10 minutes have elapsed since the timer was activated.
When operated in the Baby Mode, the Controller utilizes a Skin Temperature Probe, connected between the Controller input and the infant, to automatically adjust the heater output of the Warmer
Module to maintain a digitally displayed preset Set
Temperature.
The Keypad Lock Key, when pressed, renders the
Up/Down Arrows and Select Mode Keys inactive or active.
Also included is a Procedural Silence Timer to block out Baby Temp audible Alarms during routine procedures.
3.2.4 BLENDER MODULE (OPTIONAL)
The Blender Module provides blended oxygen from
21% to 100% to the Patient Outlet on the Resuscitation Module.
3.2.5 RESUSCITATION MODULE (OPTIONAL)
The Resuscitation Module contains pneumatic circuitry necessary for infant resuscitation. Controls and displays for the module are located above and to the rear of the Bassinet.
The Resuscitation Module consists of the following components:
.
Suction
- The Suction Circuit (reusable or disposable) is driven by a gas powered venturi actuated vacuum generator which provides a negative pressure for suctioning the patient’s airway. The suction pressure is indicated on the Suction Gauge (Figure 3.1). Suction pres-
-8-
RESUSCITAIR~ B
IRTHING
R
OOM
W
ARMER
sure may be adjusted using the Suction trol or turned on and off using the On/Off
Con-
Switch. A fixed internal relief valve limits the maximum suction level to 150 mmHg.
Extension Tube
Suction Catheter
Disposable
Note: Disposable bottle has built-in filter
I
Reusable
Suction Catheter
I
‘Extension Tube
FIGURE 3.1 SUCTION FUNCTIONAL BLOCK DIAGRAM
-9-
RESUSCITAIR~ B
IRTHING
R
OOM
W
ARMER
.
FIGURE 3.2 RESUSCITATION MODULE WITH PATIENT GAS SUPPLY AND
AUXILIARY FLOW
Patient Gas Supply -
The Patient Gas Supply Circuit may be used to provide continuous gas flow to the patient. Controls are provided for Airway Pressure Relief (maximum pressure) and Flow Rate (LPM) (circuit flow delivering 100% oxygen or blended gas). The adjustable Airway Pressure Relief Control is always operative.
.
Auxiliary Flow
The Aux Flow circuit supplies 100% oxygen through the Aux Flow (LPM) Control to the Aux
Outlet Connector. It is intended for oxygen enrichment of a manual bag resuscitator, for supplemental delivery to the patient, the mother, or a second neonate, i.e., twins. The Aux Flow (LPM) Control adjusts the flow rate from 0 to 15 Ipm. An internal preset relief valve limits the AUX Outlet pressure to
160 cm HzO.
A fixed internal safety relief valve is also provided and is also always operable. This valve provides redundant maximum pressure relief at 50 + 10 cm
Hz0 and also allows the patient to inspire room air in the event of gas supply failure.
.
Airway Pressure
The Airway Pressure Gauge displays airway pressure when connected to patient circuits via external connection.
WARNING: Breathing room air through the
2-way relief valve requires extra effort. This condition, if it occurs, should be rectified as soon as possible.
.
Patient Breathing and Supply Circuits
The patient breathing circuit used in conjunction with the Resuscitation Module is illustrated in Figure
3.3. In addition, a patient supply circuit for Manual
Bagging (Figure 3.4) may also be used.
-10
RESUSCITAIREQXJ B
IRTHING
R
OOM
W
ARMER
l-CORRUGATED TUBING
2-EXHALATION VALVE HOUSING
3-EXHALATION VALVE CAP
4-EXHALATION VALVE DIAPHRAGM
5-EXHAUST PORT
6-PATIENT PORT
7-“T” FITTING
NOTE: EXHALATION VALVE DIAPHRAGM (ITEM 4)
MUST BE INSTALLED WITH DIAPHRAGM STEM
STEM
OF
ITEM
4
GAUGE
FIGURE 3.3 PATIENT BREATHING CIRCUIT FOR MANUAL VENTILATION
l-10/7.5 mm ID Adapter 2-10 mm Corrugated Tubing
3-15 mm/l0 mm Adapter
CONNECT BETWEEN PATIENT
OUTLET AND RESUSCITATION
FIGURE 3.4 PATIENT BREATHING CIRCUIT FOR MANUAL BAGGING
-ll-
RESUSCITAIR~ B
IRTHING
R
OOM
W
ARMER
3.2.6
GAS SUPPLY MODULE (Optional)
The Gas Supply Module includes an On/Off Switch which controls the pipeline and cylinder gas supply to the Resuscitation Module. An oxygen cylinder
I
Pressure Gauge is provided if the Resuscitation option is included. Oxygen and Air Pressure Gauges are provided on units equipped with the Blender
Module.
FIGURE 3.5 GAS SUPPLY MODULE
-12-
/?EsuscimmD
B
IRTHING
R
OOM
W
ARMER
3.2.7 ALARMS
HIGH TEMPERATURE. When the Skin Temperature Probe is attached to the infant and the skin temperature exceeds 39.0 “C, the heater is automatically turned off, the High Temp Indicator will flash and the audible alarm will sound continuously.
Press the Silence/Reset Key to silence the alarm for two minutes. After the alarm condition is corrected (a skin temperature of 38.5 “C or less), the alarm will automatically reset. If the Skin Probe should detect a temperature of 39.5 “C, a hardware shutdown will occur and a System Fail Alarm will be generated.
CHECK PATIENT. When in the Manual Mode after
10 minutes of operation, the Chk Patient Indicator will illuminate and the alarm will sound one time.
Thereafter, the Chk Patient Indicator will remain illuminated and the audible alarm will sound every 30 seconds for 5 minutes. If the alarm has not been acknowledged at the end of 5 minutes, the heater will be shut down and a continuous ramping audible alarm will sound. The Silence/Reset Key then must be pressed to reactivate the heater.
PROBE. If the Skin Temperature Probe fails (shortcircuited or open) or if the Controller is in Baby Mode and there is not a Skin Probe connected to the Controller, the Probe Indicator will flash and a ramping audible alarm will sound. Afterthe alarm condition is corrected (the Probe is replaced), the alarm will automatically reset.
BABY TEMPERATURE. When the temperature sensed by the Skin Temperature Probe is 1 “C above or 1 “C below the selected Set Temperature
Display setting (in Baby Mode), the Baby Temp Indicator will flash and an audible alarm will sound, first at a low level, then at a medium level, then at a high level. In addition, if the temperature is 0.2 “C above the selected Set Temperature, the heater will be turned off automatically. Press Silence/Re-
set to silence the alarm for 10 minutes.
POWER FAIL. When ac power to the unit is interrupted while the Controller is on, the Power Fail Indicator will flash and the audible alarm will beep.
When power is restored to the unit, the alarm will automatically reset and all the settings will be retained. The alarm may be silenced by turning off the power switch.
SYSTEM FAIL. If an internal malfunction is detected, the System Fail Indicator will flash and the audible alarm will beep. In addition, an Error Code
(eRO0 to eR025) will be displayed in the
BabyTem-
perature Display. This alarm is not resettable and the unit should be referred to qualified service personnel. A prolonged brown-out (five minutes or more) will also cause a System Fail alarm.
BLENDER DIFFERENTIAL BYPASS ALARM
(Optional). The Blender Module (factory installed option) will alarm and bypass whenever the pressure differential between the 02 and Air supplies exceeds 30 psi f2 psi. When this condition occurs, the blender will continue to supply whichever gas is the higher pressure: either 100% Air or 100% Oxygen.
This is an audible alarm only. There are no visual indicators.
3.2.8 APGAR TIMER
When the Apgar Timer is running, the Apgar Timer
Display will show elapsed minutes and seconds and the audible alarm will sound at the l-,
1 O-minute Apgar time intervals.
5- and
-13-
RESUSCITAIRE~ B
IRTHING
R
OOM
W
ARMER
SECTION 4
OPERATION
4.1 CONTROLS, INDICATORS AND
NECTORS
CON-
Controls, Indicators and Connectors for the Controller are presented in Figures 4.1 and 4.2 and Tables
4.1 and 4.2. Controls, Indicators and Connectors for
the Resuscitation Module are presented in Figure
4.3 and Table 4.3.
FIGURE 4.1 CONTROLLER FRONT PANEL: CONTROLS, INDICATORS AND
CONNECTORS
-14-
RESUSCITAIR~ B
IRTHING
R
OOM
W
ARMER
5
6
7
TABLE 4.1 CONTROLLER FRONT PANEL: CONTROLS, INDICATORS AND
CONNECTORS
ITEM
1
NAME
Mode
Pre-Warm Indicator
DESCRIPTION
Manual Indicator
Baby Indicator
Indicates that the Controller is operating in the
Warm Mode.
Pre-
Indicates that the Controller is operating in the Manual
Mode.
Indicates that the Controller is operating in the
Baby
Mode.
2
Press to select either Pre-Warm, Manual or Baby
Mode of operation.
Mode Select Key
/\
8
Skin Temp Probe
Connector
>37 “C Key
0
>37
“C Indicator
Accepts Skin Temperature Probe for monitoring infant skin temperature. When connected, the Baby Temperature Display indicates the temperature sensed by the probe. When probe isdisconnected, the BabyTemperature Display is blank. When disconnected in Baby
Mode, a Probe Alarm also occurs.
Press to place Set Temperature Display (refer to Item 9) in Temperature Override Mode, >37 “C (98.6 “F).
NOTE: This Key is inactive until the Set Temperature has been set to 37.0 “C.
Keypad Lock Key e
0
Exam Light Key
Lights to indicate that the Temperature Override Mode,
>37 “C (98.6 OF), has been selected.
Press to disable the ~37 “C, Up/Down Arrow and Mode
Select Keys (refer to Items 2, 4 and 8). Press again to enable the ~37 “C, Up/Down Arrow and Mode Select
Keys. Key lights to indicate that Keypad is locked.
Press to turn on or turn off the Examination Light located in the Warmer Module.
8
0
8 @ @ Display).
Manual Mode
Press the Up Arrow Key to raise heater power from 0% to
100% in 10% increments (refer to Item 11, Heater Power
Press the Down Arrow Key to lower relative heater power from 100% to 0% in 10% increments (refer to Item 11,
Heater Power Display).
-15-
RESUSCITAIR~ B
IRTHING
R
OOM
W
ARMER
TABLE 4.1 CONTROLLER FRONT PANEL: CONTROLS, INDICATORS AND
CONNECTORS (cont.)
ITEM
8
NAME
@@
DESCRIPTION
Baby Mode
Press the Up Arrow Key to raise the Set from 34.0 “C (93 “F) to 37.0 “C (98.6
Temperatu?e
“F). In the Ternperature Override Mode (refer to Items 4 and to raise the Set Temperature from 37.0
5), press
“C (98.6 “F) to
38.0 “C (102.2 “F).
Press one time to raise the Set Temperature in 0.1 degree increments. Press and hold to raise the Set Ternperature rapidly.
Press the Down Arrow Key to lower the Set turefrom 37.0 “C (98.6 “F) to 34.0 “C (93
Tempera-
“F). In the Teriperature Override Mode (refer to Items 4 and 5), press to lower the Set Temperature from 38.0 “C (102.2 “F) to
37.0 “C (98.6 “F).
Press one time to lower the Set Temperature in 0.1 degree increments, Press and hold to lower the Set Ternperature rapidly.
NOTE: The Up/Down Arrow Keys may be locked the Keypad Lock Key (refer to Item 6).
by
9
10
11
Set Temperature Display
Power Key
-
Heater Power Display
In Baby Mode, displays the Set Temperature as
selected by the Up/Down Arrow Keys (refer to Item
8) in “C or “F as selected by the “C/OF Key (refer to
Item 12). Display is blank in Pre-Warm and Manual
Modes.
Press to turn on or turn off the Controller and Warmer
Module.
12
Displays relative heater power in 10% increments from
0% to 100%.
Press to alternately select “C or “F for the Baby
perature and Set Temperature Displays.
Tem-
13
Baby Temperature Display
Digital display of infant temperature in “C or “F (referto
Item 12), whether in Manual, Pre-Warm or Baby
Mode. The display is blank if the Skin Temperature
Probe (refer to Item 3) is disconnected from the Controller.
-16-
RESUSCITAIREB B
IRTHING
R
OOM
W
ARMER
15
TABLE 4.1 CONTROLLER FRONT PANEL: CONTROLS, INDICATORS AND
CONNECTORS (cont.)
ITEM
14
16
NAME
Silence/Reset Key
w
@
Procedural Silence
Indicator
M
Alarms
Baby Temp
High Temp
Probe
DESCRIPTION
In Manual Mode
Press to silence the High Temp Alarm (refer to Item 16) for 2 minutes.
Resets Chk Patient Alarm (refer to Item 16), and silences Audible Alarm at any time after 10 minutes of
Manual Mode.
Resets Chk Patient Alarm (refer to Item 16), silences
Audible Alarm and restores heater power after the
15-minute Manual Mode period is complete.
In Baby Mode
Press to silence the High Temp Alarm (refer to Item 16) for 2 minutes.
Press to silence Baby Temp (refer to Item 16) Alarm for
10 minutes.
During non-alarm conditions, press to enter
Procedural Silence (refer to Item 15).
When illuminated, indicates thatthe unit is in Procedural
Silence. Procedural silence interval is 5 minutes. During
Procedural Silence, the Baby Temp Alarms are blocked.
The Baby Temp Indicator will flash with a three-level audible alarm to indicate that the baby’s skin temperature is more than 1 “C above or below the selected Set
Temperature (refer to Item 9). Press Silence/Reset
Key to silence alarm for 10 minutes.
The High Temp Indicator will flash, the audible alarm will sound continuously, and the heater will be turned off when the Skin Temperature Probe is attached to the infant and the skin temperature exceeds 39.0 “C. High
Temp (39.0 “C) Alarms can only be silenced for two minutes by the Silence/Reset Key.
Press the Silence/Reset Key to silence the audible
alarm for 2 minutes.
The Alarm condition is corrected when the temperature falls to 38.5 “C; the alarm will automatically reset.
When in Baby Mode, if the Skin Temperature Probe fails, the Probe Indicator will flash and a three-level
audible alarm will sound. After the Alarm condition is corrected (the Skin Temperature Probe is replaced), the alarm will automatically reset. Also refer to Table
5.1)
When in Baby Mode, if the Skin Temperature Probe
fails (shorted probe), the System Fail Indicator will light
and an audible alarm will sound. This Alarm cannot be silenced. The Power MUST BE TURNED OFF then
ON to reset the alarm condition.
- 17-
RESUSCITAIRE~ B
IRTHING
R
OOM
W
ARMER
TABLE
4.1 CONTROLLER FRONT PANEL: CONTROLS INDICATORS AND
CONNECTORS (cont.)
ITEM
16
NAME
Alarms (Cont.)
Chk Patient
System Fail
DESCRIPTION
When in the Manual Mode, if the Warmer is in operation for longer than 10 minutes, the Chk Patient Indicator will illuminate and the audible alarm will sound one time.
Thereafter, the Chk Patient Indicator will remain on and the audible alarm will sound every 30 seconds for the next
5 minutes. If at the end of 5 minutes (15 minutes total) the alarm has not been acknowledged by pressing the Silence/Reset Key (refer to Item 14), the heater will be shut down and a three level audible alarm will sound continuously.
When an internal malfunction is detected, the System
Fail Indicator will illuminate and the audible alarm wlill sound continuously. In addition, an Error Code (ErOO to
Er025) will be displayed in the Baby Temperature Display. When a malfunction is detected, the Controller wiill automatically perform the self-test function (refer to para. 4.2 Step 3) to determine if the fault has corrected itself. If the fault has not corrected itself, the error code will be displayed until corrected.
This alarm is not resettable and the unit should be referred to qualified service personnel.
NOTE: Error code 023 may be corrected by the Operator; refer to Table 5.1.
17
18
Power Fail
Apgar Timer
When power to the unit is interrupted while the Control-, ler is on, the Power Fail Indicator will flash and the audi-,
ble alarm will beep. When power is restored to the unit, the alarm will automatically reset. Push the Power Key to silence the Alarm.
When the Apgar Timer (refer to Item 18) is running, the
Apgar Timer displays elapsed minutes and seconds and the audible alarm will sound at the l-, 5- and
1 O-minute Apgar Time intervals.
Press to start or stop the Apgar Timer.
Stop/Start bm
Reset
0
//
When timer is runnlng, press to reset the timer to zero and restart the Apgar count.
When timer is stopped, press to turn timer off.
NOTE: The Timer can be reset at any time.
- 18-
RESUSCITAIR~ B
IRTHING
R
OOM
W
ARMER
FIGURE 4.2 CONTROLLER REAR PANEL: CONTROLS AND CONNECTORS
TABLE 4.2 CONTROLLER REAR PANEL: CONTROLS AND CONNECTORS
ITEM NAME
DESCRIPTION
1
2
3
CIRCUIT BREAKER
POWER
AUX PORT
Turns Controller on and off when switched by operator or the presence of excessive current drain is detected.
Accepts ac power cord.
Data port for connection to printer or host system.
-19-
RESUSCITAIREB B
IRTHING
R
OOM
W
ARMER
FIGURE 4.3A RESUSCITATION MODULE, INDICATORS AND
CONNECTORS
-2o-
12
13
14
15
RESUSCITAIREB B
IRTHING
R
OOM
W
ARMER
TABLE 4.3A RESUSCITATION MODULE: CONTROLS, INDICATORS AND
CONNECTORS
I
ITCM
I LI.,
I
NAMF
1.e-nm.mmm
I
I-bFSCRIPTlflN
. .
. .-..
5
6
7
Suction
Suction Gauge
Suction Min Max Control
On/Off Switch
Patient Supply
Airway Pressure Gauge
Airway Pressure Relief
Min Max Control
Flow Rate (LPM) Control
Patient Outlet Connector
Displays suction level from 0 to 200 mmHg of vacuum.
Adjusts suction level from 0 to 150 mmHg of vacuum.
Turns Suction on and off.
Displays airway pressure from -10 to +80 cm HzO.
Adjusts airway pressure relief setting from 0 to 50 cm HzO.
Adjusts patient gas flow from 0 to 15 LPM.
Accepts patient circuit outlet line.
16
17
Auxiliarv Flow
Aux Outlet 100% Oxygen
Connector
Aux Flow (LPM) Control
Accepts auxiliary gas delivery line.
Adjusts auxiliary patient gas flow from 0 to 15 LPM.
Connects Airway Pressure Gauge to Patient Circuit.
18 Airway Pressure Port
-2l-
RESUSCITAIREB B
IRTHING
R
OOM
W
ARMER
FIGURE 4.3B BLENDED GAS SUPPLY MODULE CONTROL
ITEM
1
2
TABLE 4.38 BLENDED GAS SUPPLY MODULE CONTROL
DESCRIPTION
NAME
Blended Gas Supply Module (Optional)
Blender % Oxygen Control
-
Blends air and oxygen mixture from 21 to 100% 02.
-22-
RESUSCITAIREB B
IRTHING
R
OOM
W
ARMER
FIGURE 4.3C GAS SUPPLY MODULE: CONTROLS AND INDICATORS
ITEM
TABLE 4.3C GAS SUPPLY MODULE: CONTROLS AND INDICATORS
NAME
DESCRIPTION
Gas SUPPIV (Optional )
19
20
21
Air Cylinder Gauge
Oxygen Cylinder Gauge
Gas Supply On/Off
Switch
Provides indication of air cylinder supply pressure 0
to 4000 psi.
Provides indication of oxygen cylinder supply pressure 0 to 4000 psi.
Turns gas supply to pneumatic system on and off.
-23-
RESUSCITAIREB B
IRTHING
R
OOM
W
ARMER
4.2 OPERATIONAL CHECKOUT PROCE-
DURE - CONTROLLER
WARNING: The Warmer should not be used if the Controller or other parts of the equip ment fail to function as described below. A potential shock hazard exists within the
Warmer Module. Service should be ferred to qualified personnel.
ra-
IMPORTANT:
Before attempting to perform this procedure, refer to Paragraph 4.1,
Controls, lndica tors and Connectors.
NOTE: The Operational Checkout procedure should be performed before each use and after disassembly for cleaning, servicing or maintenance. Refer to qualified service personnel if the unit does not perform as specified.
1.
CONNECT THE AC LINE CORD TO THE
POWER CONNECTOR on the Rear Panel.
WARNING: Connect the power cord only to a properly grounded wall receptacle that is approved for hospital-use and of the correct voltage. DO NOT use extension cords or an ac Receptacle Box for this device.
2.
CHECK THE SELF-TEST FUNCTION. Turn on the Power Switch, the Self-Test Function should be initiated and the following should occur:
Apgar Timer, Baby Temperature and
Set Temperature Digital Displays show all eights
All Alarm Indicators light (except Power
Fail)
All Mode Indicators light
The > 37 “C Indicator lights
All ten segments of the Heater Power Indicator light
The Procedural Silence Indicator lights
The Keypad Lock Switch lights
The audible alarm will sound a high pitch tone, a low pitch tone, then a beep-beepbeep
3.
4.
5.
6.
When the Self-Test Function is complete, the
Controller should begin operating in the
Warm Mode.
Pre-
CHECK THE PRE-WARM MODE. The
Pre-
Warm Indicator should be on and the Heater
Power Indicator should display 10 segments
(100%) for three minutes, reduce to 6 segments (60%) for 12 minutes, then reduce to 3 segments (30%).
CHECK THE MANUAL MODE. Select Manu-
al Mode by pressing the Mode Select Key. The
Manual Indicator should light.
Press the Up Arrow Key until all the Heater
Power Display segments are lit. Press the
Down Arrow Key until all the Heater Power Indicators are off. Connect the skin temperature probe to the Skin Temp Probe Connector, the
Baby Temperature Display should come on.
Attach the Skin Temperature Probe to the Mattress, with a Probe Cover, to prevent Error
Code 18.
Set the Heater Power Indicator to lOO%, all segments are lit. Wait 10 minutes. After 10 minutes have elapsed, the Chk Patient Indicator should come on and the audible alarm should sound one time. Wait an additional 5 minutes.
During this time the audible alarm should sound at 30-second intervals. At the end of 5 minutes (15 total), the heater should shut down and the Heater Power Indicators should go off.
Press the Silence/Reset Key, the Chk Patient
Indicator and audible alarm should go off and
the heater should come back on along with the
Heater Power Indicators.
CHECKTHE KEYPAD LOCK. Press the Keypad Lock Switch. The Keypad Lock Switch should light up. The Mode Key and Up/Down
Arrow Keys should be inoperative. Press the
Keypad Lock Switch. The Keypad Lock
Switch Light should go off and the Keypad should be enabled.
CHECK THE BABY MODE. Select Baby
Mode by pressing the Mode Select Key. The
Baby Indicator should light and the Set Tem-
perature Display should read 36.5 “C. In addition, the Baby Temp Indicator should flash and the audible alarm should sound. Press the Si-
- 24
RESUSCITAIREGII B
IRTHING
R
OOM
W
ARMER
lence/Reset Key, the audible alarm should go
off, the Baby Temp Indicator should remain on.
7.
CHECK THE TEMPERATURE OVERRIDE
MODE. Press the Up Arrow Key to raise the
Set Temperature to 37.0 “C. Press the >37 “C
Key, the >37 “C Indicator should come on.
Press the Up Arrow Key to raise the Set Temperature to 38.0 “C.
Press the Down Arrow Key to lower the Set
Temperature to below 37.0 “C. When the Set
Temperature falls below 37.0
Indicator should go off.
“C, the ~37 “C
8.
CHECK THE PROBE ALARM. Disconnect the skin temperature probe from the Skin
Temp Probe Connector. The Baby Temperature Display should go off, the Probe Indicator should flash and the audible alarm should sound. Connect the probe to the Skin Temp
Probe Connector.
9.
CHECK THE APGAR TIMER. Press the Start
/Stop Key, the Apgar Timer Display should come on and begin to count up from zero seconds. Press the Start/Stop Key, the Apgar
Timer count should stop. Press the Reset Key, the Apgar Timer Display should go off.
10.
CHECK THE EXAMINATION LIGHT. Press the Exam Light Switch. The Examination Light should come on. Press the Exam Light Switch, the Examination Light should go off.
11.
CHECK THE POWER FAILURE ALARM.
Turn off the CIRCUIT BREAKER on the Rear
Panel. Turn on the Power Switch on the Front
Panel. The Power Fail Indicator should come on and the Audible alarm should sound. Turn off the Power Switch and turn on the CIRCUIT
BREAKER. Disconnect the Skin Temperature
Probe from the Skin Temp Probe Connector and the Mattress.
NOTE: The unit must be connected to the ac line and power turned on for at least three minutes before the Power Fail circuitry becomes active.
4.3 MECHANICAL CHECKOUT
1.
CHECK THE BASSINET TILT CONTROL
(Figure 4.4) by pulling up on the lever located at the bottom rear of the Bassinet while supporting the rear lower edge of the Bassinet with the palm. Place the Bassinet in the 5-degree and then the lo-degree tilt position; return the
Bassinet to the level position.
FIGURE 4.4 CHECKING BASSINET
TILT CONTROL
2.
CHECK THE BASSINET SIDE PANELS (Figure 4.5) by raising each panel and pivoting it to hang straight down. Return the panel by reversing the procedure. Check that the panel is positively engaged to confine the infant.
3.
FIGURE 4.5 CHECKING THE
BASSINET SIDE PANELS
CHECK THE BASSINET FRONT PANEL
(Figures 4.6 and 4.7) by raising the panel and sliding it under the mattress. Return the panel by reversing the procedure. Check that the panel is positively engaged to confine the infant.
- 25 -
RESUSCITAIREB B
IRTHING
R
OOM
W
ARMER
FIGURE 4.6 CHECKING THE
BASSINET FRONT PANEL
FIGURE 4.8 CHECKING THE
THROUGH DRAWER
PASS-
5.
CHECK THE WARMER MODULE SWIVEL
OPERATION (Figure 4.9) by rotating the
Warmer Module 90 degrees to the left or right of center. Return to center position.
PRECAUTION: For effective heating, make sure the Warmer Module is positioned on the center axis and locked with its detent.
FIGURE 4.7 CHECKING THE
BASSINET FRONT PANEL
4. CHECK THE PASS-THROUGH DRAWER
(Figure 4.8) by sliding the drawer in and out on both sides of the Cart/Bassinet. Return to center position.
FIGURE 4.9 CHECKING THE WARMER
MODULE SWIVEL
6.
CHECK THE OPERATION OF THE
X-RAY
CASSETTE TRAY (ACCESSORY)
(Figure
4.10) by grasping the middle of a Side Panel and pulling the X-ray Cassette Tray out from under the Bassinet. Replace the X-ray Cassette Tray by reversing the procedure.
- 26 -
RESUSCITAIRED B
IRTHING
R
OOM
W
ARMER
FIGURE 4.10 CHECKING THE X-RAY
CASSETTE TRAY
7. CHECK THE WRITING SURFACE
(Figure
4.11) by pulling it out from under the Bassinet.
Replace the Writing Surface by reversing the procedure.
FIGURE 4.11 CHECKING THE
WRITING SURFACE
8. CHECK THE CART/BASSINET DOCKING
MECHANISM AND CASTER LOCKS. Refer to paragraph 4.7 and undock the Cart. Depress the Caster Locks (Figure 1.1) and check that they prevent the Can from moving. Release the Caster Locks and dock the Cart/Bassinet.
Pull on the Car-t/Bassinet to ensure that it is latched properly.
4.4 RESUSCITATION EQUIPMENT
USE CHECKOUT AND SETUP (Op-
PREtional)
4.4.1 SUPPLY PRESSURE
1.
Ensure that O2 (and AIR) pipeline(s) are securely attached to appropriate fittings on the rear of the unit and that gas supply present is
40 to 75 psi.
If using Reserve Gas Supply from cylinders:
2.
Ensure that cylinder(s) are properly secured in the mounting yokes on the rear of the Warmer and that the cylinder valve located on the top of the cylinder is open.
3. Examine the appropriate cylinder pressure gauges on the front of the upper column to ensure that sufficient reserve gas supply is present.
4. Set the handle of the
Gas Supply On/Off
Switch to the On position,
1.4.2
BLENDED GAS SUPPLY PRE-USE
CHECKOUT AND SETUP (Optional)
.
.
.
LOW FLOW MICROBLENDER WARNINGS:
.
If either the air or oxygen gas source pressure is reduced or increased, creating a pressure differential of 30 psi, the microblender alarm will sound. This condition significantly alters the FiO2 and flow output from the microblender.
Always operate the low flow microblender with clean/dry medical grade gases.
Air inlet water filters are recommended for use with the low flow microblender.
The concentration of oxygen provided and the partial pressure of oxygen within the patient’s blood (Pa02) should be monitored.
1.
If present, turn the precision blender knob to ensure free rotation.
2
Remove either the oxygen or air supply (pipeline or cylinder). The blender alarm should sound. Replace the supply.
3.
Set the blender to the desired % oxygen concentration.
-27-
RESUSCITAIREB B
IRTHING
R
OOM
W
ARMER
4.4.3 RESUSCITATION MODULE PRE-USE
CHECKOUT AND SETUP (Optlonal)
SUCTION
FIGURE 4.12 CHECKING THE SUCTION
NOTE: To obtain suction, the Gas Supply On/Off
Switch (Figure 4.3C) must be ON.
BOlTLE
tion Min Max Control while viewing the suction level on the Suction Gauge. Adjust the suction magnitude to the desired maximum suction pressure value.
1.
Check that a clean suction bottle (reusable or disposable, Figure 4.12) is installed and properly connected at the base of the Warmer column.
6.
Turn off the Suction On/Off Switch.
2.
Ensure that a bacterial filter is connected inline with the supply connection to the reusable suction bottle (a filter is built-in on the disposable bottle).
RESUSCITATION (Optional)
Manual Resuscitation
- Use with Patient Breathing Circuit - 10 mm tubing with thumb (finger) hole at patient end.
3.
Connect the desired extension tubing to the outlet of the suction bottle outlet port (refer to
Figure 3.1) and secure the free end of the extension tubing in either tubing retaining slot provided on the front and rear panels of the
Bassinet.
WARNING: Excessive airway pressure can cause damage to patient’s lungs. Always monitor airway pressure and/or provide appropriate pressure relief during resuscitation.
4.
Turn on the Suction On/Off Switch. There may be an initial reading of up to 30 mmHg on the Suction Gauge (refer to Figure 3.1) due to flow resistance of the hydrophobic filter and suction tubing.
WARNING: For prolonged ventilation, use of a heat and moisture exchanger is recommended.
NOTE: The filter and tubing resistance will not affect the desired maximum value that is set in Step 5 below. The pressure value on the
Suction Gauge matches the actual pressure value at the end of the catheter.
There are potential hazards associated with the delivery of supplemental oxygen. If it is necessary to administer oxygen, the attending physician should be notified immediately.
5.
Block the patient outlet of the suction bottle.
Adjust the suction magnitude using the Suc-
1.
Connect the Patient Circuit to the Patient Outlet (refer to Figures 3.2, 3.3 and 3.4).
- 28 -
RESUSCITAIREB B
IRTHING
R
OOM
W
ARMER
2.
Adjust the flow rate to the desired fresh gas flow rate using the Patient Supply Flow Rate
(LPM) Control.
3.
Set the Airway Pressure Relief control to the desired pressure limit according to the color coded bands on the Airway Pressure Gauge and Airway Pressure Relief Control. Alternately, a “T” Fitting with an airway pressure monitor port can be inserted into the Patient
Outlet Port and connected to the Airway
Pressure Fitting to indicate the breathing circuit pressure. Adjust the Airway Pressure Relief Control as necessary.
4.5 CONTROLLER OPERATION
WARNING: Connect the power cord only to a properly grounded wall receptacle that is approved for hospital-use and of the correct voltage. DO NOT use extension cords or an ac Receptacle Box for this device.
Connect the unit to the ac line. Turn on the CIRCUIT
BREAKER on the Rear Panel and the Power
Switch on the Front Panel. Observe the Functional
Test.
4.5.1 PRE-WARM MODE
After the Functional Test is complete, the Pre-
Warm Mode will activate. The Heater Power Indicator will be at 100% (all lights on) for three minutes, reduce to 60% (six lights on) for 12 minutes and then
be reduced to 30% (three lights on).
NOTE: Selection of Manual or Baby Mode and then returning to Pfe-Warm Mode during the three minutes of 700% or 12 minutes of
60% power will automatically reduce the
power to 30%.
During Pre-Warm Mode, the Chk Patient Alarm is disabled.
WARNING:
To avoid overheating or underheating, observe the infant constantly and monitor the temperature using the skin temperature probe supplied with the equipment or other electronic thermometer.
Inspect infant’s skin for reddened areas which may result from heating of materials in contact with the skin (plastic diapers, pins, etc.).
Avoid placement of objects (equipment, blankets, or clothing) between the infant and Warmer Module that can block heat transfer or absorb heat. These objects may be heated directly by the heater and can cause skin burns.
Radiant warming increases insensible water loss. Appropriate measures to maintain fluid balance should be considered.
4.5.2 MANUAL MODE
1.
Use the Mode Key to select Manual Mode.
2.
Use only for short-term warming with nursing personnel in constant attendance.
3.
Do not use the Warmer in Manual Mode if
Manual Indicator is not on.
4.
Set the Heater Power Indicator to the desired level. The heater power will be maintained for
10 minutes.
5.
After 10 minutes, the Chk Patient Alarm will activate. Press the Silence/Reset Key to initiate another 1 O-minute warming period.
6.
If the Chk Patient Alarm is not acknowledged, the heater will be automatically disabled after an additional 5 minutes of operation.
7.
Heater power output must be adjusted manually to maintain Baby Temperature within the desired range.
-29-
RESUSCITAIREB B
IRTHING
R
OOM
W
ARMER
8. Check infant’s temperature and condition at least every 15 minutes. When initially setting or when changing heater power output, check
Baby Temperature more frequently to be sure it
is maintained within the desired range.
CAUTION: A change in heater power output will not result in an immediate change in
Baby Temperature. Wait for results.
Large changes in heater power output will cause a more rapid change in Baby Temperature.
9. Use Skin Temperature Probe to continuously monitor Baby Temperature whenever possible.
Refer to paragraph 4.5.3 to attach the probe to the patient.
IMPORTANT: In Manual Mode, the Skin Temperature Probe monitors only -- it does not control heater output and the
Baby Temp alarms are not functional.
NOTE: It is not necessary that the Skin Temperature Probe be connected to the Controller for Manual Mode.
4.5.3 BABY MODE
1.
Use the Mode Key to select Baby Mode.
2. Plug Skin Temperature Probe into Controller
Skin Temp Probe Connector.
3. Attach the Skin Temperature Probe to the infant. The probe should be located on the infant’s abdomen, halfway between the xyphoid and the umbilicus (Figure 4.13). The metal side of the probe should be placed in direct contact with the skin.
WARNING:
Infant radiant warmers should be used only by properly trained personnel as directed by an appropriately qualified physician aware of currently known hazards and benefits.
To avoid hazards of overheating or underheating, the infant should not be left unattended. Use only with the Hill-Rom
Air-Shields Skin Temperature Probe supplied with the unit. Inspect the infant’s skin for reddened areas which may result from heated materials in contact with the skin (plastic diapers, pins, etc.).
Dressing the infant or swaddling with a blanket during radiant warming can lead to skin temperature variations which can cause Set Point Temperature alarms. Always uncover or undress the infant during radiant warming to prevent excessive skin temperature variations.
FIGURE 4.13 ATTACHING SKIN PROBE
-3o-
RESUSCITAIR~ B
IRTHING
R
OOM
W
ARMER
WARNING:
The location of the Skin Temperature
Probe must be such that the skin around the Sensor is in direct line with the heat from the Warmer Module. If the location is shadowed, for example, by the infant’s body, overheating and possible burning of the infant’s skin can result. Do not use a rectal probe. Use of a rectal probe can result in overheating of the infant.
The Skin Temperature Probe must be in direct contact with the skin to provide accurate monitoring of the infant’s skin temperature. Failure to maintain intimate skin contact can result in overheating and possible burning. Check infant’s condition at least every fifteen minutes for correct Sensor attachment and feel infant’s skin for signs of overheating.
The Skin Temperature Probe should be placed at least 1.5 inches from any transcutaneous monitor (TcPOz or TcPCOz) probe to prevent false temperature indications. The Skin Temperature Probe should not be placed on an area previously used by a TcP02 or TcPCO2 probe.
4.
When the infant is prone, the Skin Temperature
Probe should be located on the infant’s back.
5.
The skin area around the probe should be thoroughly cleansed and dried before the probe is placed on the skin.
6. To obtain an accurate reading of the infant’s skin temperature, place the probe in position and cover with a Critter Covers@ Disposable
Probe Cover, Care-For-Me Cover or tape the probe into position, cover it with a small piece of cotton just large enough to cover the tip of the probe, and then place a second piece of tape over the cotton. If it is desired to reduce tape contact on the infant’s skin, the cotton can be applied directly to the probe tip without the first piece of tape. To stabilize the attached probe, a third piece of tape may be placed overthe probe wire approximately three to four centimeters from the probe tip.To minimize the effect of direct radiation on the Skin Temperature Probe, in order to obtain a more accurate Baby Temper-
ature measurement, cover the Sensor with a
Critter Covers@ Disposable Probe Cover,
Care-For-Me Cover or an equivalent insulating foam cover with a reflective surface facing the
Warmer Module.
7.
Baby Mode should be used for long-term warming and when attending personnel cannot be in constant attendance.
8.
Set the Set Temperature Display to the prescribed temperature. A higher Set Temperature setting does not increase rapid warming.
9.
Verify that Baby Temperature Display reading
stabilizes within 0.2 “C of SetTemperature Dis-
play. Fluctuations in the Heater Power Indicators or the Baby Temperature Display reading can result from air currents, obstruction of radiation to the infant or the Skin Temperature
Probe not being in intimate contact with the skin.
10.
Baby Temp Alarms can be silenced for 10 minutes by pressing the Silence/Reset Key.
11.
Baby Temp, Probe and High Temp (39.0 “C)
Alarms are automatically reset after the alarm condition is corrected.
12.
The High Temp Alarm may be silenced for 2 minutes by pressing the Silence/Reset Key.
NOTE: In the event of a Probe Alarm,
Manual
Mode can be used temporarily until a replacement Skin Temperature Probe is available and only if nursing personnel are in constant attendance.
4.5.4 EXAMINATION LIGHT
The light is turned on and off by the Exam Light
Switch. Turn the light on only as required for optimum bulb life.
Protect infant’s eyes to minimize risk of injury.
I
-3l-
RESUSCITAIR~ B
IRTHING
R
OOM
W
ARMER
4.6
CART/BASSINET
CAUTION: Do not attempt to move the Cart/
Bassinet by using the acrylic End and Side Panels as a handle.
4.7
DOCKING AND
CART/BASSINET
UNDOCKING THE
1. The End and Side Panels raise and pivot to hang down (refer to Figures 4.5,4.6 and 4.7).
1.
The Car-t/Bassinet is attached to the Warmer by a latching mechanism located underneath the
Cart/Bassinet which attaches to a stud protruding from the Warmer center leg.
2.
The Bassinet Tilt Control (Figure 4.4) operates by pulling up on the lever located at the bottom rear of the Bassinet while supporting the rear lower edge of the Bassinet with the palm. Place the Bassinet in the 5-degree or the lo-degree tilt position.
I
WARNING:
This Warmer is intended for use with infants in the
Cart/Bassinet. If the Warmer is used to warm the mother immediately after delivery, all warnings and instructions should be followed with a particular regard to heater-to-mattress distance and skin probe location.
Avoid placement of objects near the mother which can absorb heat and become hot to the touch.
To engage the Cart/Bassinet to the Warmer, push the Cart/Bassinet straight onto the Warmer until it latches. To release the Cart/Bassinet, push and hold the Release Switch on the front of the Cart/Bassinet (Figure 4.14) and pull the
Cart/Bassinet away from the Warmer.
b
PRECAUTION
FOR PATENT SAFETY CART
~.__ ~~~
DOCKED
AND LATCHED TO ENSURE
CORRECT BASSINET
POSITIONI;Q$IoEFFECTiVE
When repositioning the mattress with an infant on the mattress, hold the infant securely to prevent the infant from moving during repositioning.
Do not leave infant unattended when the panels are down. When transporting the infant in the CatVBassinet, the Mattress should be in the horizontal position.
All panels must be erect and secure during transport.
/
PRESS AND HOLD
TO RELEASE CART.
FIGURE 4.14 CART/BASSINET
RELEASE SWITCH
- 32 -
4.8 X-RAY PROCEDURES
1.
Swing the Warmer Module (Figure 4.9) to the right or left of center as required to position the
X-ray machine.
2.
Lift the Left or Right Bassinet Side Panel up, slide the X-ray Cassette Tray out (Figure 4.10); place the X-ray Cassette on the tray and return the tray to the Bassinet. Align the tray as desired with the markings on the X-ray Cassette and relative markings on the inside of the Bassinet walls.
3.
When the X-ray is complete, remove the X-ray
Cassette and return the X-ray Cassette Tray.
Place the Warmer Module in its normal operating position.
RESUSCITAIR~ B
IRTHING
R
OOM
W
ARMER
SECTION 5
CLEANING AND MAINTENANCE
5.1 GENERAL
This section provides cleaning and maintenance instructions.
Where necessary, disassembly instructions are provided. Maintenance other than that provided in this section should be pet-formed only by qualified service personnel.
WARNING:
If oxygen is in use, make sure that the oxygen supply to the equipment is turned off and that it is disconnected from the oxygen supply when performing cleaning and maintenance procedures. A fire and explosion hazard exists when performing cleaning and/or maintenance procedures in an oxygen-enriched environment.
An electrical shock hazard exists when performing cleaning and maintenance procedures; make sure that the Power Cord is disconnected from the wall receptacle.
FIGURE 5.1 REMOVING BASSINET
SIDE PANELS
2.
Remove the Bassinet Back Panel (Figure 5.2) by raising it straight up until the bottom pins are adjacent to the slots in the corner brackets.
5.2 CLEANING
When an infant is discharged, or at least once a week, the equipment should be thoroughly disinfected. Cleaning can most effectively be accomplished by disassembling, then grouping the parts and/or assemblies in categories according to the method of cleaning required.
5.3 DISASSEMBLY FOR CLEANING
1.
Remove both Bassinet Side Panels (Figure
5.1) by pulling them straight up.
-33-
FIGURE 5.2 REMOVING BASSINET
BACK PANEL
3.
Remove the Bassinet Front Panel (Figure 5.4) by raising it and then swiveling it down. At the corners, press up on the release buttons (Figure 5.3) and pull the panel straight out.
RESUSCITAIR~ B
IRTHING
R
OOM
W
ARMER
RELEASEBUTTON
5.4 CLEANING PROCEDURES
When an infant is discharged, or at least once a week, the equipment should be thoroughly cleaned and disinfected.
FIGURE 5.3 BASSINET FRONT PANEL
RELEASE BUTTONS
5.4.1 CLEANING AGENTS.
An iodophor or quaternary disinfectant-detergent registered by the U.S. Environmental Protection
Agency should be used, but only when the equipment is not in use and disassembled as described elsewhere in this section. A cleanser such as Kleenaseptic@b Germicidal Surface Cleanser may be used. When using any cleaning agent, follow the manufacturer’s directions for use. Before cleaning, remove all solid wastes and contaminants from the disassembled parts.
5.4.2 PAINTED SURFACES
Use a disinfectant-detergent to clean all surfaces thoroughly; then dry with a clean cloth or paper towel.
FIGURE 5.4 REMOVING BASSINET
FRONT PANEL
4.
Remove the Mattress from the Bassinet.
5. Remove the X-ray Cassette Tray (Figure
4.10).
6.
Remove the Suction Bottle and Filter (Figure
4.12) from the side of the column.
-34-
5.4.3 CLEAR PLASTIC AND ACRYLIC SUR-
FACES
CAUTION: Alcohol can cause crazing of plastic and acrylic. Do not use alcohol, acetone, or any organic solvents for cleaning.
Do not expose plastic and acrylic to direct radiation from germicidal lamps. Ultraviolet radiation from these sources can cause cracking and crazing of clear plastic and acrylic.
Use a disinfectant-detergent to clean all surfaces thoroughly. Make sure to clean all holes, indentations, etc.; then dry with a clean cloth or papertowel.
5.4.4 METAL SURFACES
Use a disinfectant-detergent to thoroughly clean all surfaces; then dry with a clean cloth or paper towel.
5.4.5 WOODEN SURFACES
Use a disinfectant-detergent to thoroughly clean all surfaces; then dry with a clean cloth or paper towel.
RESUSCITAIR~ B
IRTHING
R
OOM
W
ARMER
5.4.6 SKIN TEMPERATURE PROBE, REUS-
ABLE
CAUTION: Do not pull on the tip of the skin temperature probe when cleaning or drying; damage to the probe may result.
Use a disinfectant-detergent to thoroughly clean all surfaces; then dry with a clean soft cloth or paper towel.
IMPORTANT
After cleaning, a complete operational checkout procedure should be performed before returning the unit to service.
5.5 STERILIZATION
CAUTION: DO NOT STEAM AUTOCLAVE. Gas sterilization temperature should not exceed 54.4 “C (130 “F).
Sterilization can be accomplished with the following:
A. COLD STERILIZATION
CAUTION: Do not expose plastic and acrylic to direct radiation from germicidal lamps. Ultraviolet radiation from these sources can cause cracking of gasket surfaces, fading of paint, and ultimately, crazing of plastic and acrylic.
B. GAS STERILIZATION (ETHYLENE OXIDE).
Prior to gas sterilization, the entire unit should be thoroughly cleaned as described elsewhere
in this section. Remove and discard all used disposable elements. New disposable elements should be installed after sterilization.
Standard gas sterilization procedures, as programmed by automatic equipment such as made by American Sterilizers and Wilmot
Castle, are satisfactory as these do not normally exceed 54.4 “C (130 “F).
IMPORTANTI
After sterilization, a complete func-
tional checkout procedure should be performed before returning the unit to service.
5.6 REASSEMBLY AFTER CLEANING
1.
Replace the Mattress on the Bassinet.
2
3.
4.
Replace the X-ray Cassette Tray (Figure
4.10).
Replace the Bassinet Back Panel by inserting the pins in the Corner Brackets (Figure 5.2).
Replace the Bassinet Side Panels by pushing them straight down into their slots (Figure 5.1).
5.
6.
Replace the Bassinet Front Panel by sliding it into the Front of the Bassinet (Figure 5.4) until the release tabs catch. Raise the Panel into position.
Install a new Suction Filter if a Reusable Bottle is in use (Figure 4.12). Replace the Suction
Bottle if a Disposable Suction Bottle is in use.
5.7 CALIBRATION
The equipment should be completely checked and calibrated at least once a year by qualified service personnel. Refer to the appropriate Service Manual
for details.
5.8 TROUBLESHOOTING
Troubleshooting for the operator of the equipment is presented in Table 5.1. If the fault cannot be localized from the chart, the unit should be removed from use and referred to factory trained or otherwise qualified service personnel.
- 35 -
RESUSCITAIREB B
IRTHING
R
OOM
W
ARMER
TABLE 5.1 TROUBLESHOOTING
SYMPTOM POSSIBLE CAUSE
CORRECTIVE ACTION
No Power and Power Fail Alarm a.
Circuit Breaker not set to On.
a.
Set Circuit Breaker to On.
is not activated
Power Fail Alarm activated a. Circuit Breaker tripped.
b. Power Cord unplugged.
c. Defective Power Cord.
a.
Reset Circuit Breaker (Figure 4.2).
b.
Connect Power Cord to
POWER connector (Figure
4.2) or wall socket.
c. Replace Power Cord.
System Fail Alarm activated a. Internal malfunction.
a. Refer to service.
Probe Alarm activated.
Possible Defective Skin
Probe(s).
a.
Check to ensure Skin Probe is in good contact with the skin.
b. Replace Skin Probe(s). If condition is not corrected, refer to service.
a. Refer to service.
Error Code ErOl through Er022
Er024 and Er025
Error Code Er023 a. Internal malfunction.
Ambient Temperature in excess of 32 “C (90 “F).
Verify ambient temperature with an external thermometer.
-36-
RESUSCITAIREB B
IRTHING
R
OOM
W
ARMER
SECTION 6
PARTS LIST
6.1 GENERAL
This section provides a listing of Operator replacement parts. Parts other than those listed here
should be replaced by qualified service personnel.
For a listing of accessories, refer to Figure 1.1 of this manual.
REPLACEMENT PARTS
PART NUMBER
Bassinet Side Panel
Bassinet Rear Panel
. . . . . . , , . . . . . . . . . . . . . . . . . . . . .
. . , . . . . . . . . . . . . . . , , , . . . . . . . . .
Bassinet Front Panel
PowerCord
. . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
,.......,,.....,....................81 900 00
,..........,,.......................81 90001
,...................................81 900 02
Skin Temperature Probe (Reusable)
Reusable Suction Bottle Kit (750 ml) (Bottle, Stopper, Tubing and Filter)
Reusable Suction Bottle Only
Filters (Box of 25)
Mattress/Cover (Mattress with Cover)
.................... 81 001 50
DISPOSABLES
Premi-Probe@ Skin Temperature Probe@ (Box of 10)
....................................81 300 08
Premi-Probe@ Skin Temperature Probe@ (10 Boxes of 10)
Critter Covers@ Probe Cover (Box of 100)
...............................81 300 09
.............................................
.68 209 46
Critter Covers@ Probe Cover (Box of 600)
.............................................
.68 209 45
Care-for-Me Probe Covers, 100 Large (10% discount when you order 5)
.................. 68 209 47
Care-for-Me Probe Covers, 100 Standard (10% discount when you order 5)
................ 68 209 48
Patient Supply Circuit (Box of 25)
Infant Breathing Circuit (Box of 25)
NeatClips-3/8”Diameter(BoxoflOO)
...............................................
1 .OO” Diameter (50/Case)
.6812053
................................................
.68 120 54
Disposable Suction Bottle and tubing, 800 ml (Box of 100)
Disposable Suction Bottle and tubing, 800 ml (Box of 20)
................................81 001 51
.................................81 001 49
OPTIONS
AirHoseAssembly,USA(lOft)Yelloww/WingNuts .....................................
7846410
OxygenHoseAssembly, USA(lOft)Greenw/WingNuts.. ...............................
X-rayCassetteTray ...............................................................
MonitorShelf .....................................................................
7846510
..811004 4
..820015 4
I .V. Pole...........................................................................820015
- 37 -
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 1 Operating Precautions
- 5 Table of Definitions & Symbols
- 6 Table of Contents
- 8 General Information
- 9 Specifications
- 13 Installation
- 15 Functional Description
- 20 Alarms
- 21 Operation
- 31 Operational Checkout Procedure
- 32 Mechanical Checkout
- 34 Resuscitation Pre-use Checkout
- 36 Controller Operation
- 39 Cart / Bassinet
- 40 Cleaning and Maintenance
- 43 Troubleshooting
- 44 Parts List