What are the approved operating parameters for ProTaper Ultimate™ files?
All ProTaper Ultimate™ files are approved for use at a speed of 400 rpm and a torque of 4 to 5.2 Ncm.
Dentsply Sirona ProTaper Ultimate Rotary Files are designed to help you achieve optimal results in your endodontic procedures. With their unique design and advanced features, these files offer a number of benefits that can make your work easier and more efficient.
ProTaper Ultimate Rotary Files are made from a high-quality nickel-titanium alloy that is both strong and flexible. This makes them resistant to breakage and allows them to follow the curvature of the root canal more easily. The files also feature a unique "variable taper" design that helps to create a smooth, tapered canal that is ideal for obturation.
In addition, ProTaper Ultimate Rotary Files are color-coded for easy identification. This makes it easy to quickly select the right file for the job. The files are also packaged in sterile blisters for added convenience.
Advertisement
Advertisement
...............................................................................................
...........................................................................................
...................................................................................................
............................................................................................
..............................................................................................
.........................................................................................
.................................................................................................
............................................................................................
....................................................................................................
.............................................................................................
............................................................................................
...................................................................................................
.................................................................................................
...........................................................................................
.............................................................................................
.............................................................................................
.............................................................................................
........................................................................................
....................................................................................................
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 1/151
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 2/151
ProTaper Ultimate™
FOR DENTAL USE ONLY
FOR PROFESSIONAL USE ONLY
STERILE – STERILIZED BY RADIATION
LIABLE FOR REPROCESSING
REUSABLE STERILE ENDODONTIC INSTRUMENTS
● The products are indicated for treatment of endodontic disease.
● Intended purpose: Engine-driven instrument intended for root canal preparation (glide path, shaping and debridement of the root canal).
● Endodontic instruments are to be used only in a clinical or hospital environment, following good dental practice, by qualified dental professionals such as general practitioners as well as Endo specialists (Endodontist) and Dental Assistants.
● Instruments shall be used in combination with an endodontic motor (manually in severe curvatures).
None.
ProTaper Ultimate™ instruments are made of three main components: the working part made of a nickel-titanium alloy, a colored silicone stopper (except for instrument SX) and a plated brass shank with a colored ring (except for instrument SX).
The files are manufactured with a process that results in a file that has a colored appearance.
Due to this proprietary processing, the files may appear slightly curved. This is not a manufacturing defect. While the file can be easily straightened using only your fingers, it is not necessary to straighten the file prior to use. Once inside the canal, the file will follow the root canal anatomy.
The ProTaper Ultimate™ range comprises following endodontic instruments:
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 3/151
Code xx-length y-file size z-quantity per blister
B ST PULR z
19 0SX
B ST PULR z xx 0SL
B ST PULR z xx 0SH
Product
ProTaper Ultimate™
SX
ProTaper Ultimate™
SLIDER
ProTaper Ultimate™
SHAPER
ISO XXX YYYv
XXX-ISO size
YYY-taper size v-variable taper
ISO 020.003v
ISO 016.002v
ISO 020.004v
Available lengths
19 mm
21/25/31 mm
21/25/31 mm
Packaging and content
Blister with 6 instruments (sterile)
Blister with 3 instruments (sterile)
Blister with 6 instruments (sterile)
Blister with 3 instruments (sterile)
Blister with 6 instruments (sterile)
Blister with 3 instruments (sterile)
Blister with 6 instruments (sterile)
B ST PULR z xx 0F y
B ST PULR 6 xx yyy
B ST PULR 5 xx SQE
B DD PULR
525 SQE
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
Sequence
ProTaper Ultimate™
Demo Pack
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
FX: ISO 035.012v
FXL: ISO 050.010v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
21/25/31 mm
21/25/31 mm
21/25/31 mm
25 mm
Blister with 3 instruments (sterile)
Blister with 6 instruments (sterile)
Blister with 1 instrument of each instrument (sterile)
Blister with 1 instrument of each instrument (sterile)
● Torque control devices ensure optimal usage.
● Use with endodontic motors in a constant rotation at a speed of 400 rpm.
● Set torque at: 4 to 5.2 Ncm.
● Lubricants such as NaOCl, EDTA, ProLube, Glyde™ shall be used.
● The use of radiographs in combination with an apex locator and a tool for adjusting the silicon stopper to the correct working length is the appropriate method of working length determination.
● Strictly follow this Instruction for Use and the General Processing Instructions For
Endodontic Products (see section 8) to minimize following risks to the device, the patient and/or the user:
○ Breakage of instrument.
○ Cross-contamination.
○ Heat generation due to insufficient lubrication and irrigation.
○ Swallowing of working part of the instrument.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 4/151
○ Toxic or allergic reactions caused by processing residues.
● Similar to all machine-driven root canal instruments, ProTaper Ultimate™ instruments should not be used in a root canal with an abrupt apical curvature due to heightened risk of separation. In this case, pre-curved hand files should be used in the apical region.
● Safety and effectiveness of use have not been established in pregnant or breastfeeding women or in children.
● Use a rubber dam system during the endodontic procedure.
● For your own safety, wear personal protective equipment (gloves, glasses, mask).
● Inspect the packaging before use and do not use the instruments if the packaging is damaged.
● Do not use the instruments after expiration date.
● Check the instrument before each use for signs of defects such as deformations (bent, unwound), breakage, corrosion, damaged cutting edges, loss of color coding or marking.
With these indications the devices are not able to fulfil the intended use with the required safety level, instruments should be discarded.
● Before using any instrument, make sure it is well connected to the contra-angle head.
● Check instrument and clean working part frequently during instrumentation, inspecting for signs of distortion, elongation or wear, such as uneven flutes, dull spots. With these indications that the devices are not able to fulfil the intended use with the required safety level, instruments should be discarded.
● Instruments should not be fully immersed in a sodium hypochlorite solution (NaOCl). Only the working part of the NiTi instrument, which is in contact with the patient should be immersed in a NaOCl solution concentrate at NOT more than 5%.
● Exercise caution in the apical area and in canals that divide, and/or exhibit abrupt curvatures or recurvatures.
● Irrigate abundantly and frequently the canal throughout the procedure and after each instrument used (according to good dental practices).
● Always use minimal apical pressure. Never force the files down the canal.
● When instrument does not easily progress, clean and inspect the cutting flutes, then irrigate, recapitulate with a manual file and reirrigate.
● For shaping extremely curved canals it is safer to use the file only to shape one canal in order to reduce the risk of breakage. Pay attention to the following good practices:
○ Use a new file and discard it after the canal was treated (single canal use).
○ Use manual instead of rotary files.
○ Use small size, flexible or/and NiTi files (this will help avoid canal transportation).
○ Visually inspect the working part for all the defects listed in the former paragraph during use (i.e after each wave).
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 5/151
○ Avoid the standard reaming continual rotational motion and instead use small angle motions (filing motion, watch winding oscillation motion, or balanced force technique) in order to limit the rotational bending fatigue on the instruments and improve their expected life.
In the present technical state, no adverse reaction has been reported so far.
Keep products in a dry and clean environment, away from sources of moisture and direct sunlight.
All ProTaper Ultimate™ files have been validated to be used at 400 rpm and 4 to 5.2 Ncm torque.
The ProTaper Ultimate™ instruments can be used with an outward brushing motion in all canals, especially those canals that exhibit an irregular cross-section, or with a light inward pecking motion to progressively advance toward the working length (WL).
Always cradle the handpiece in the webbing between the thumb and index finger. Avoid pushing; rather, let the files passively progress and follow the Slidepath.
1) Review different horizontally angulated radiographs to diagnostically determine the width, length, and curvature of any given root and its canal(s).
2) Prepare an access that enables the easy identification of each canal orifice.
3) The auxiliary shaping file, SX, may be used when there is restrictive space, or to preenlarge the body of a canal, or to relocate the coronal-most aspect of a canal away from an external root concavity.
4) In the presence of a lubricant, select the Slider and PASSIVELY follow the canal, in one or more passes, to its terminus. Determine WL using an electronic apex locator in combination with a radiographic image, then confirm patency.
5) If the Slider will not easily reach the canal terminus, select a small-sized manual Stainless
Steel (SS) hand file. In the presence of a lubricant catheterize this canal, establish working length, confirm patency, and verify the Slidepath. Now, repeat step #4 above.
6) In the presence of NaOCl, select the Shaper and advance along the Slidepath, in one or more passes, until the WL is reached.
7) Upon removing the Shaper, irrigate, EndoActivate to break up debris and move it into solution, then re-irrigate to liberate this debris.
8) Reconfirm WL, especially in more curved canals.
9) Select the FINISHER F1 (020.07) and passively follow the canal to the WL, in one or more passes. Remove and inspect its apical flutes. When loaded with dentinal debris, the preparation is finished.
10) If the FINISHER F1 is loose at length and its apical flutes are not loaded with debris, select the FINISHER F2 (025.08) and use this file in the same manner described above for
FINISHER F1.
11) If the FINISHER F2 is loose at length and its apical flutes are not loaded with debris, select the FINISHER F3 (030.09) and use this file in the same manner described above for
FINISHER F1 and F2.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 6/151
12) Upon removing any given file, clean and inspect its cutting flutes, irrigate, recapitulate with either a size 10 file or Endo Activator to break up debris, then re-irrigate.
13) Inspect the file’s cutting flutes upon removal for the presence of unwinding, straightening or stretching. If deformation is noted, discard and use a new ProTaper Ultimate™ file.
14) The preparation is finished when the apical extent of any Finisher is loaded with debris, and the correspondingly sized GPMC or Size Verifier fits at the WL.
Use the auxiliary Finishers in larger and straighter canals only, such as maxillary central incisors, some palatal or distal canals of molars, or when there is a pathologic or iatrogenic defect:
15) Select either a mechanically driven or manual auxiliary file, FINISHER FX (035.12), when working length is established and patency is confirmed. Use this stand-alone file directly and passively to follow the canal to the WL using the same technique as described above.
Remove and inspect its apical flutes. When loaded with dentinal debris, the preparation is finished.
16) If the FINISHER FX is loose at length and its apical flutes are not loaded with debris, select either a mechanically-driven or manual auxiliary FINISHER FXL (050.10) and use this file in the same manner described above for FINISHER FX.
17) The preparation is finished when the apical extent of any auxiliary Finisher is loaded with debris, and the correspondingly-sized GPMC or Size Verifier fits at the WL.
● Please follow the General Processing Instructions For Endodontic Products ( https:// www.dentsplysirona.com/en/service-contact/ifu.html
) for the disinfection, cleaning and sterilization steps.
● Do not re-use silicone stops. Remove and discard silicone stops after each use.
● Products shall be disposed according to local regulations for the safe disposal of sharp and contaminated devices.
● Any serious incident in relation to the product should be reported to the manufacturer and the competent authority according to local regulations.
● Sterility cannot be guaranteed if packaging is open, damaged or wet.
● To get a free printed copy of IFU please see section “Get a print copy of IFU” on website https://www.dentsplysirona.com/en/service-contact/ifu.html
.
● Explanation of non-harmonized symbols for IFUs and labels, see IFU Symbols ( https:// www.dentsplysirona.com/en/service-contact/ifu/ifu-symbols.html
).
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 7/151
2797
Authorised Representative
DENTSPLY DeTrey GmbH
De-Trey-Straße 1
D-78467 Konstanz
Germany
Manufacturer
Maillefer Instruments Holding Sàrl
Chemin du Verger 3
CH-1338 Ballaigues
Suisse www.dentsplysirona.com/ifu
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 8/151
ProTaper Ultimate™
RÉSERVÉ À UN USAGE DENTAIRE
RÉSERVÉ À UN USAGE PROFESSIONNEL
STÉRILE - STÉRILISÉ PAR IRRADIATION
RÉUTILISABLE
INSTRUMENTS ENDODONTIQUES STÉRILES RÉUTILISABLES
● Les produits sont indiqués dans le traitement des pathologies endodontiques.
● Utilisation prévue : instrument motorisé conçu pour la préparation du canal radiculaire (passage libre, mise en forme et débridement du canal radiculaire).
● Les instruments endodontiques doivent être utilisés uniquement en milieu clinique ou hospitalier, en suivant les bonnes pratiques dentaires, par des professionnels dentaires qualifiés tels que dentistes généralistes ainsi que spécialistes en endodontie (endodontistes) et assistants dentaires.
● Les instruments doivent être utilisés en association avec un moteur endodontique (manuellement pour les courbures importantes).
Aucune.
Les instruments ProTaper Ultimate™ se composent de trois éléments principaux : la partie travaillante en alliage nickel-titane, une butée en silicone de couleur (sauf pour l'instrument SX) et un manche en laiton chromé avec anneau de couleur (sauf pour l'instrument SX).
Le processus de fabrication des limes leur confère une apparence colorée.
En raison de ce traitement breveté, les limes peuvent présenter une légère courbure. Il ne s’agit pas d’un défaut de fabrication. Même si la lime peut être facilement redressée manuellement, il n’est pas nécessaire de le faire pour l’utiliser. Une fois dans le canal, la lime suivra l'anatomie du canal radiculaire.
La gamme ProTaper Ultimate™ comprend les instruments endodontiques suivants :
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 9/151
Code xx-longueur y-taille de la lime z-quantité par blister
B ST PULR z
19 0SX
B ST PULR z xx 0SL
B ST PULR z xx 0SH
Produit
ProTaper Ultimate™
SX
ProTaper Ultimate™
SLIDER
ProTaper Ultimate™
SHAPER
ISO XXX YYYv
XXX-ISO taille
YYY-taille de conicité v-conicité variable
ISO 020.003v
ISO 016.002v
ISO 020.004v
Longueurs disponibles
19 mm
21/25/31 mm
21/25/31 mm
Emballage et contenu
Blister de 6 instruments (stériles)
Blister de 3 instruments (stériles)
Blister de 6 instruments (stériles)
Blister de 3 instruments (stériles)
Blister de 6 instruments (stériles)
Blister de 3 instruments (stériles)
Blister de 6 instruments (stériles)
B ST PULR z xx 0F y
B ST PULR 6 xx yyy
B ST PULR 5 xx SQE
B DD PULR
525 SQE
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
Sequence
ProTaper Ultimate™
Demo Pack
F1 : ISO 020.007v
F2 : ISO 025.008v
F3 : ISO 030.009v
FX : ISO 035.012v
FXL : ISO 050.010v
SLIDER :
ISO 016.002v
SHAPER :
ISO 020.004v
F1 : ISO 020.007v
F2 : ISO 025.008v
F3 : ISO 030.009v
SLIDER :
ISO 016.002v
SHAPER :
ISO 020.004v
F1 : ISO 020.007v
F2 : ISO 025.008v
F3 : ISO 030.009v
21/25/31 mm
21/25/31 mm
21/25/31 mm
25 mm
Blister de 3 instruments (stériles)
Blister de 6 instruments (stériles)
Blister avec 1 instrument de chaque
(stérile)
Blister avec 1 instrument de chaque
(stérile)
● Les dispositifs de contrôle du couple permettent une utilisation optimale.
● Utilisez avec des moteurs endodontiques avec une rotation constante à une vitesse de
400 rpm.
● Définissez le couple à : 4 to 5.2 Ncm.
● Des lubrifiants tels que NaOCl, EDTA, ProLube, Glyde™ doivent être utilisés.
● Pour déterminer la longueur appropriée, il convient d'utiliser en parallèle les radiographies, un localisateur d'apex et un outil pour régler la butée en silicone à la bonne longueur de travail.
● Suivez scrupuleusement ces instructions d'utilisation et les instructions de traitement général pour produits endodontiques (voir section 8) afin de réduire les risques suivants pour l'instrument, le patient et/ou l'utilisateur :
○ Rupture de l'instrument.
○ Contamination croisée.
○ Production de chaleur due à une irrigation et une lubrification insuffisantes.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 10/151
○ Ingestion de la partie travaillante de l'instrument.
○ Réactions toxiques ou allergiques provoquées par des résidus de traitement.
● À l'instar des instruments endodontiques motorisés, les instruments ProTaper Ultimate™ ne doivent pas être utilisés dans un canal radiculaire avec une courbure apicale abrupte en raison du risque accru de séparation. Dans ce cas, des limes manuelles pré-courbées doivent être utilisées dans la région apicale.
● L’efficacité et la sécurité d'emploi n’ont été établies ni chez la femme enceinte ou allaitante ni chez l'enfant.
● Utilisez une digue dentaire lors de la procédure endodontique.
● Par mesure de sécurité, portez un équipement de protection individuelle (gants, lunettes, masque).
● Inspectez l’emballage avant utilisation et n'utilisez pas les instruments si l’emballage est endommagé.
● N'utilisez pas les instruments au-delà de leur date d’expiration.
● Contrôlez l'instrument avant chaque utilisation afin de vérifier qu'il n'y a pas de défaut tel qu’une déformation (tordu, déroulé), une rupture, de la corrosion, des bords tranchants endommagés, une perte du marquage ou du code couleur. Le cas échéant, les instruments ne peuvent pas fonctionner comme prévu avec le niveau de sécurité requis et doivent donc
être éliminés.
● Avant d’utiliser un instrument, assurez-vous qu'il est bien inséré dans la tête du contreangle.
● Vérifiez l'instrument et nettoyez fréquemment la partie travaillante pendant l’instrumentation
à la recherche d’éventuels signes de déformation, d’étirement ou d’usure, par exemple des spires ébréchées ou des zones émoussées. Le cas échéant, les instruments ne peuvent pas fonctionner comme prévu avec le niveau de sécurité requis et doivent donc être éliminés.
● Ces instruments ne doivent pas être immergés dans une solution d’hypochlorite de sodium
(NaOCl). Seule la partie travaillante de l’instrument en NiTi en contact avec le patient doit
être plongée dans une solution de NaOCl dont la concentration NE DOIT PAS excéder
5 %.
● Progressez avec précaution dans la zone apicale et dans les canaux présentant une division et/ou des courbures marquées ou multiples.
● Irriguez abondamment et fréquemment le canal tout au long de la procédure et après l'utilisation de chaque instrument (conformément aux bonnes pratiques dentaires).
● Utilisez toujours une pression apicale minimale. Ne forcez jamais le passage des limes dans le canal.
● Lorsque l'instrument ne progresse pas facilement, nettoyez et inspectez les spires, puis irriguez, recommencez avec une lime manuelle et irriguez à nouveau.
● En cas de mise en forme de canaux à courbure sévère, il est plus sûr d’utiliser une nouvelle lime pour chaque canal à mettre en forme afin de réduire le risque de rupture. Veuillez respecter les bonnes pratiques suivantes :
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 11/151
○ Utilisez une nouvelle lime et éliminez-la après le traitement du canal (utilisation unique par canal).
○ Préférez des limes manuelles aux limes rotatives.
○ Privilégiez des limes de petite taille, flexibles et/ou en NiTi (cela contribuera à prévenir le transport canalaire).
○ Pour détecter les éventuels défauts répertoriés dans le paragraphe précédent, contrôlez visuellement la partie travaillante de la lime durant son utilisation (c.-à-d. après chaque passage).
○ Pour élargir le canal, évitez les mouvements de rotation continue classique et privilégiez les petits mouvements angulaires (mouvement de limage, mouvement d’oscillation de remontoir de montre ou technique des forces équilibrées) afin de limiter la fatigue des instruments et d’améliorer leur durée de vie attendue.
Dans l’état technique actuel, aucun effet indésirable n’a été rapporté à ce jour.
Conservez les produits dans un environnement propre et sec, à l´écart de toute source d’humidité et à l’abri de la lumière directe du soleil.
Toutes les limes ProTaper Ultimate™ ont été validées pour une utilisation à 400 rpm et à un couple de 4 to 5.2 Ncm.
Les instruments ProTaper Ultimate™ peuvent être utilisés avec un mouvement de brossage vers l'extérieur dans tous les canaux, particulièrement les canaux qui présentent une section transversale irrégulière, ou avec un léger mouvement de va-et-vient pour avancer progressivement vers la longueur de travail.
Prenez toujours la pièce à main entre le pouce et l'index. Évitez de pousser les limes. Laissezles plutôt pénétrer passivement et suivre le passage libre.
1) Étudiez les radiographies de différentes angulations horizontales pour déterminer la largeur, la longueur et la courbure de toute racine et de son canal ou ses canaux.
2) Préparez un accès qui permet d'identifier facilement chacun des orifices canalaires.
3) La lime de mise en forme auxiliaire, SX, peut être utilisée en cas d'espace réduit ou pour pré-élargir le corps d'un canal, ou pour éloigner le point le plus coronaire d'un canal d'une concavité radiculaire externe.
4) En présence de lubrifiant, sélectionnez le Slider et suivez PASSIVEMENT le canal, en un ou plusieurs passages, jusqu'à son extrémité. Déterminez la longueur de travail à l'aide d'un localisateur d'apex électronique associé à une image radiographique, puis confirmez la perméabilité.
5) Si le Slider n'atteint pas facilement l'extrémité du canal, sélectionnez une lime manuelle en acier inoxydable de petite taille. En utilisant du lubrifiant, cathétérisez ce canal, définissez la longueur de travail, confirmez la perméabilité et vérifiez le passage libre. Répétez maintenant l'étape n° 4 ci-dessus.
6) En utilisant du NaOCl, sélectionnez le Shaper et avancez le long du passage libre, en un ou plusieurs passages, jusqu'à ce que la longueur de travail soit atteinte.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 12/151
7) Après avoir retiré le Shaper, irriguez avec un irrigateur de canal radiculaire EndoActivator pour décomposer les débris et les transporter dans une solution, puis irriguez à nouveau pour retirer les débris.
8) Reconfirmez la longueur de travail, particulièrement dans les canaux plus courbés.
9) Sélectionnez le FINISHER F1 (020.07) et suivez passivement le canal jusqu'à la longueur de travail, en un ou plusieurs passages. Retirez et inspectez ses spires apicales. Lorsqu'elles sont chargées de débris dentinaires, la préparation est terminée.
10) Si le FINISHER F1 a du jeu et que ses spires apicales ne sont pas chargées de débris, sélectionnez le FINISHER F2 (025.08) et utilisez cette lime de la même manière quelle celle décrite ci-dessus pour le FINISHER F1.
11) Si le FINISHER F2 a du jeu et que ses spires apicales ne sont pas chargées de débris, sélectionnez le FINISHER F3 (030.09) et utilisez cette lime de la même manière quelle celle décrite ci-dessus pour les FINISHER F1 et F2.
12) Qu'elle que soit la lime, après l'avoir retirée, nettoyez et inspectez ses spires, irriguez, répétez soit avec une lime de taille 10 ou un irrigateur de canal radiculaire EndoActivator pour décomposer les débris, puis irriguez à nouveau.
13) Après avoir retiré la lime, inspectez ses spires à la recherche de tout déroulement, redressement ou étirement. En cas de déformation, éliminez la lime et utilisez une lime ProTaper
Ultimate™ neuve.
14) La préparation est terminée lorsque l'extrémité apicale de tout Finisher est chargée de débris, et que le maître cône de gutta percha ou le vérificateur de taille de la taille correspondante est adapté à la longueur de travail.
Utilisez les Finishers auxiliaires exclusivement dans les canaux plus grands et plus droits, comme ceux des incisives centrales supérieures, certains canaux palatins ou distaux des molaires, ou en cas de défaut pathologique ou iatrogène :
15) Sélectionnez une lime auxiliaire manuelle ou motorisée, FINISHER FX (035.12), lorsque la longueur de travail est établie et la perméabilité confirmée. Utilisez cette lime autonome directement et passivement pour suivre le canal jusqu'à la longueur de travail, à l'aide de la même technique que celle décrite ci-dessus. Retirez et inspectez ses spires apicales.
Lorsqu'elles sont chargées de débris dentinaires, la préparation est terminée.
16) Si le FINISHER FX a du jeu et que ses spires apicales ne sont pas chargées de débris, sélectionnez un FINISHER FXL (050.10) auxiliaire manuel ou motorisé et utilisez cette lime de la même manière quel celle décrite ci-dessus pour le FINISHER FX.
17) La préparation est terminée lorsque l'extrémité apicale de tout Finisher auxiliaire est chargée de débris, et le maître cône de gutta percha ou le vérificateur de taille de la taille correspondante s'adapte à la longueur de travail.
● Veuillez suivre les instructions de traitement général pour produits endodontiques ( https:// www.dentsplysirona.com/en/service-contact/ifu.html
) en ce qui concerne les étapes de désinfection, nettoyage et stérilisation.
● Ne réutilisez pas les butées en silicone. Retirez et éliminez les butées en silicone après chaque utilisation.
● Les produits doivent être éliminés conformément aux réglementations locales relatives à l’élimination en toute sécurité des dispositifs tranchants et contaminés.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 13/151
● Tout incident grave en lien avec le produit doit être signalé au fabricant et à l'autorité compétente conformément aux réglementations locales.
● La stérilité n’est plus garantie si l'emballage est ouvert, endommagé ou humide.
● Pour obtenir une copie papier gratuite des instructions d'utilisation, veuillez consulter la section « Obtenir une copie papier gratuite des instructions d'utilisation » sur le site internet https://www.dentsplysirona.com/en/service-contact/ifu.html
.
● Signification des symboles non harmonisés pour les instructions d'utilisation et les étiquettes, voir les symboles des instructions d'utilisation ( https://www.dentsplysirona.com/en/ service-contact/ifu/ifu-symbols.html
).
2797
Représentant autorisé
DENTSPLY DeTrey GmbH
De-Trey-Straße 1
D-78467 Konstanz
Germany
Fabricant
Maillefer Instruments Holding Sàrl
Chemin du Verger 3
CH-1338 Ballaigues
Suisse www.dentsplysirona.com/ifu
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 14/151
ProTaper Ultimate™
SOLO PER USO ODONTOIATRICO
SOLO PER USO PROFESSIONALE
STERILE – STERILIZZATO MEDIANTE IRRADIAZIONE
RESPONSABILITÀ PER LA RISTERILIZZAZIONE
STRUMENTI ENDODONTICI STERILI RIUTILIZZABILI
● I prodotti sono indicati per il trattamento di patologie endodontiche.
● Scopo previsto: strumento motorizzato destinato alla preparazione del canale radicolare
(percorso di avanzamento, sagomatura e sbrigliamento del canale radicolare).
● Gli strumenti endodontici devono essere usati solo in ambiente clinico od ospedaliero, seguendo la buona pratica odontoiatrica, da odontoiatri specializzati, endodontisti e assistenti dentali.
● Gli strumenti devono essere usati in combinazione con un motore endodontico (manualmente in curvature accentuate).
Nessuna.
Gli strumenti ProTaper Ultimate™ sono costituiti da tre componenti principali: la parte in movimento in lega di nickel-titanio, un tappo di silicone colorato (tranne che per lo strumento SX) e uno stelo di ottone rivestito con un anello colorato (tranne che per lo strumento SX).
Le lime sono prodotte con un processo che conferisce loro un aspetto colorato.
A causa di questa lavorazione brevettata le lime possono apparire leggermente ricurve. Questo non è un difetto di fabbricazione. La lima può essere facilmente raddrizzata con le dita ma non
è necessario raddrizzarla prima dell'uso. Una volta all'interno del canale la lima seguirà l'anatomia del canale radicolare.
La linea ProTaper Ultimate™ comprende i seguenti strumenti endodontici:
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 15/151
Codice lunghezza xx dimensioni lima y quantità z per blister
B ST PULR z
19 0SX
B ST PULR z xx 0SL
B ST PULR z xx 0SH
Prodotto
ProTaper Ultimate™
SX
ProTaper Ultimate™
SLIDER
ProTaper Ultimate™
SHAPER
ISO XXX YYYv dimensioni ISO XXX dimensioni taper
YYY taper variabile v
ISO 020.003v
ISO 016.002v
ISO 020.004v
Lunghezze disponibili
19 mm
21/25/31 mm
21/25/31 mm
Confezione e contenuto
Blister con 6 strumenti (sterili)
Blister con 3 strumenti (sterili)
Blister con 6 strumenti (sterili)
Blister con 3 strumenti (sterili)
Blister con 6 strumenti (sterili)
Blister con 3 strumenti (sterili)
Blister con 6 strumenti (sterili)
B ST PULR z xx 0F y
B ST PULR 6 xx yyy
B ST PULR 5 xx SQE
B DD PULR
525 SQE
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
Sequence
ProTaper Ultimate™
Demo Pack
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
FX: ISO 035.012v
FXL: ISO 050.010v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
21/25/31 mm
21/25/31 mm
21/25/31 mm
25 mm
Blister con 3 strumenti (sterili)
Blister con 6 strumenti (sterili)
Blister con uno strumento di ciascun tipo (sterili)
Blister con uno strumento di ciascun tipo (sterili)
● I dispositivi di controllo della coppia assicurano un utilizzo ottimale.
● Usare con motori endodontici a rotazione costante con una velocità di 400 giri/min.
● Regolare la coppia su: 4 to 5.2 Ncm.
● Si devono usare lubrificanti del tipo NaOCl, EDTA, ProLube, Glyde™.
● L'uso di radiografie in combinazione con un localizzatore d'apice e uno strumento per regolare lo stopper di silicone alla corretta lunghezza di lavoro è il metodo appropriato per determinare la lunghezza di lavoro.
● Seguire rigorosamente queste istruzioni per l'uso e le istruzioni generali per il trattamento dei prodotti endodontici (vedere sezione 8) per ridurre al minimo i seguenti rischi per il dispositivo, il paziente e/o l'utente:
○ Rottura dello strumento.
○ Contaminazione incrociata.
○ Generazione di calore per insufficiente lubrificazione e irrigazione.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 16/151
○ Ingestione della parte in movimento dello strumento.
○ Reazioni tossiche o allergiche causate dai residui del trattamento.
● Analogamente a tutti gli strumenti automatici per il canale radicolare, gli strumenti ProTaper
Ultimate™ non devono essere usati in un canale radicolare con curvatura apicale improvvisa a causa del maggior rischio di rottura. In questo caso si devono usare lime manuali precurvate per la regione apicale.
● La sicurezza e l'efficacia d'uso non sono state determinate nelle donne in gravidanza o in allattamento e nei bambini.
● Utilizzare un telaio dentale in gomma durante la procedura endodontica.
● Per motivi di sicurezza consigliamo di indossare dispositivi di protezione personale (guanti, occhiali, maschera).
● Ispezionare la confezione prima dell'uso e non usare gli strumenti se la confezione è danneggiata.
● Non utilizzare gli strumenti dopo la data di scadenza.
● Prima di ogni utilizzo controllare lo strumento per verificare che non presenti eventuali difetti, come ad esempio deformazioni (piegature, svergolamento), rottura, corrosione, bordi di taglio rovinati, perdita del codice colorato o della marcatura. Se i dispositivi presentano una delle condizioni sopra descritte, non sono in grado di soddisfare l'uso previsto e garantire il livello di sicurezza richiesto, pertanto devono essere eliminati.
● Prima di utilizzare qualsiasi strumento assicurarsi che questo sia ben fissato alla testina del contrangolo.
● Controllare frequentemente lo strumento e pulire la parte in movimento durante l'utilizzo, cercando eventuali segni di distorsione, allungamento o usura, come ad esempio scanalature irregolari o zone opache. Se i dispositivi presentano una delle condizioni sopra descritte, non sono in grado di soddisfare l'uso previsto e garantire il livello di sicurezza richiesto, pertanto devono essere eliminati.
● Gli strumenti non devono essere immersi in soluzioni di ipoclorito di sodio (NaOCl). Solo la parte in movimento dello strumento NiTi a contatto con il paziente deve essere immersa in una soluzione di NaOCl con una concentrazione non superiore al 5%.
● Fare attenzione nelle aree apicali e nei canali che si dividono e/o presentano curvature accentuate o ricurvature.
● Durante la procedura, e dopo l'utilizzo di ogni strumento, irrorare abbondantemente e frequentemente il canale (secondo le buone pratiche odontoiatriche).
● Esercitare sempre una pressione apicale minima. Non forzare mai la lima nel canale.
● Se lo strumento non avanza con facilità, pulire e ispezionare le scanalature di taglio, quindi irrorare, ricominciare con una lima manuale e irrorare di nuovo.
● Per la sagomatura di canali estremamente curvi, è consigliabile utilizzare la lima per sagomare un solo canale onde ridurre il rischio di rottura. Prestare attenzione alle buone pratiche di seguito riportate:
○ Usare una nuova lima ed eliminarla in seguito al trattamento del canale in questione (monouso per singolo canale).
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 17/151
○ Utilizzare lime manuali anziché rotative.
○ Usare lime di piccole dimensioni, flessibili e/o in NiTi (ciò impedirà il trasporto canalare).
○ Ispezionare visivamente la parte in movimento per accertarsi che non presenti i difetti sopra elencati durante l'uso (ossia dopo ogni passaggio).
○ Evitare il movimento rotatorio continuo normalmente eseguito per allargare il canale ed effettuare, invece, piccoli movimenti angolari (movimento di limatura, prestando attenzione al movimento oscillatorio, oppure tecnica di bilanciamento delle forze) al fine di limitare la fatica flessionale esercitata sugli strumenti con il movimento rotatorio e quindi prolungarne la durata prevista.
Nell'attuale stato tecnico non sono state finora riferite reazioni indesiderate.
Conservare i prodotti in un ambiente asciutto e pulito, lontano da fonti di umidità e dalla luce solare diretta.
Tutte le lime ProTaper Ultimate™ sono state approvate per essere utilizzate ad una velocità di
400 giri/min e una coppia di 4 to 5.2 Ncm.
Gli strumenti ProTaper Ultimate™ possono essere usati con un movimento di brushing verso l'esterno, specialmente in quei canali che presentano una sezione trasversale irregolare o con un leggero movimento di pecking per avanzare gradualmente verso la lunghezza di lavoro
(WL).
Sostenere sempre il manipolo con la parte morbida tra il pollice e l'indice. Evitare di spingere. È invece opportuno lasciare che la lima avanzi passivamente seguendo il percorso di avanzamento.
1) Rivedere diverse radiografie orizzontali angolate per determinare in modo diagnostico la larghezza, la lunghezza e la curvatura di una data radice e dei relativi canali.
2) Preparare un accesso che consenta una facile identificazione di ciascun orifizio del canale.
3) La lima ausiliaria di sagomatura SX può essere usata in caso di spazi ristretti o per pre-allargare il corpo di un canale o riposizionare l'aspetto più coronale di un canale lontano da una concavità radicolare esterna.
4) Utilizzando un lubrificante, selezionare la Slider e seguire PASSIVAMENTE il canale, in uno o più passaggi, fino al suo terminus. Determinare la lunghezza di lavoro usando un localizzatore di apice elettronico in combinazione a un'immagine radiografica, quindi confermare la pervietà.
5) Se la Slider non raggiunge facilmente il terminus del canale, selezionare una lima manuale di piccole dimensioni in acciaio inossidabile. Utilizzando un lubrificante, cateterizzare questo canale, determinare la lunghezza di lavoro, confermare la pervietà e verificare il percorso di scorrimento. A questo punto, ripetere il passaggio n. 4 di cui sopra.
6) In presenza di NaOCl, selezionare lo Shaper e avanzare lungo il percorso di scorrimento, con uno o più passaggi, fino a raggiungere la lunghezza di lavoro.
7) Durante la rimozione dello Shaper, irrorare, eseguire l'endoattivazione per frammentare i detriti e trasferirli nella soluzione, quindi irrorare di nuovo per liberare questi detriti.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 18/151
8) Riconfermare la lunghezza di lavoro, specialmente nei canali più ricurvi.
9) Selezionare il FINISHER F1 (020.07) e seguire passivamente il canale fino alla lunghezza di lavoro con uno o più passaggi. Rimuovere e ispezionare le scanalature apicali. Quando sono cariche di fango dentinale, la preparazione può considerarsi terminata.
10) Se il FINISHER F1 è libero su tutta la lunghezza e le sue scanalature apicali non sono cariche di fango, selezionare il FINISHER F2 (025.08) e utilizzare questa lima nello stesso modo descritto per il FINISHER F1.
11) Se il FINISHER F2 è libero su tutta la lunghezza e le sue scanalature apicali non sono cariche di fango, selezionare il FINISHER F3 (030.09) e utilizzare questa lima nello stesso modo descritto per il FINISHER F1 e F2.
12) Durante la rimozione di qualsiasi lima pulire e ispezionare le scanalature di taglio, irrorare, ricominciare con una lima 10 o un endoattivatore per frammentare i detriti, quindi irrorare di nuovo.
13) Al momento della rimozione, ispezionare le scanalature di taglio della lima per l'eventuale presenza di svergolamento, raddrizzamento o allungamento. In caso si notino deformazioni, eliminare la lima e utilizzare una nuova lima ProTaper Ultimate™.
14) La preparazione può considerarsi conclusa quando la parte apicale di qualsiasi Finisher è colma di detriti e il corrispondente cono master di guttaperca o il Size Verifier si adattano alla lunghezza di lavoro.
Usare i Finisher ausiliari soltanto nei canali più ampi e più dritti, come ad esempio per gli incisivi centrali mascellari, alcuni canali palatali o distali di molari o qualora sia presente un difetto patologico o iatrogeno:
15) Selezionare una lima ausiliaria ad azionamento meccanico o manuale, del tipo FINISHER
FX (035.12), se la lunghezza di lavoro è nota e la pervietà confermata. Usare questa lima stand-alone in modo diretto e passivo per seguire il canale fino alla lunghezza di lavoro mediante la stessa tecnica descritta sopra. Rimuovere e ispezionare le scanalature apicali.
Quando sono cariche di fango dentinale la preparazione può considerarsi terminata.
16) Se il FINISHER FX è libero su tutta la lunghezza e le sue scanalature apicali non sono cariche di fango, selezionare un FINISHER FXL (050.10) ausiliario ad azionamento meccanico o manuale e utilizzare questa lima nello stesso modo descritto per il FINISHER FX.
17) La preparazione può considerarsi conclusa quando la parte apicale di qualsiasi Finisher è colma di detriti e il corrispondente cono master di guttaperca o il Size Verifier si adattano alla lunghezza di lavoro.
● Seguire le Istruzioni Generali per il Trattamento dei Prodotti per Endodonzia ( https:// www.dentsplysirona.com/en/service-contact/ifu.html
) per la disinfezione, la pulizia e la sterilizzazione.
● Non riutilizzare gli stop in silicone. Rimuovere ed eliminaregli stop di silicone dopo ogni utilizzo.
● I prodotti devono essere smaltiti conformemente alle disposizioni locali in materia di smaltimento sicuro dei dispositivi di taglio e contaminati.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 19/151
● Qualsiasi incidente correlato al prodotto deve essere segnalato al produttore e all'autorità competente in conformità con le disposizioni locali.
● La sterilità non può essere garantita se la confezione è aperta, danneggiata o bagnata.
● Per ottenere una copia cartacea delle istruzioni per l'uso, consultare la sezione “ Come ricevere una copia cartacea delle istruzioni per l'uso” sul sito web https:// www.dentsplysirona.com/en/service-contact/ifu.html
.
● Per la spiegazione dei simboli non armonizzati nelle istruzioni per l'uso e le etichette, consultare ( https://www.dentsplysirona.com/en/service-contact/ifu/ifu-symbols.html
).
2797
Rappresentante autorizzato
DENTSPLY DeTrey GmbH
De-Trey-Straße 1
D-78467 Konstanz
Germany
Produttore
Maillefer Instruments Holding Sàrl
Chemin du Verger 3
CH-1338 Ballaigues
Suisse www.dentsplysirona.com/ifu
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 20/151
ProTaper Ultimate™
NUR FÜR DEN DENTALEN GEBRAUCH
NUR FÜR DEN ZAHNÄRZTLICHEN GEBRAUCH
STERIL – STERILISIERT DURCH BESTRAHLUNG
AUFBEREITUNG ERFORDERLICH
WIEDERVERWENDBARE STERILE ENDODONTISCHE INSTRUMENTE
● Die Produkte sind zur Behandlung endodontischer Erkrankungen vorgesehen.
● Verwendungszweck: Maschinelle Instrumente zur Wurzelkanalaufbereitung (Gleitpfad, Aufbereitung und Reinigung des Wurzelkanals).
● Endodontische Instrumente dürfen nur in Zahnarztpraxen oder Zahnkliniken gemäß guter zahnärztlicher Praxis von entsprechend qualifiziertem Personal wie Allgemeinzahnärzten oder Endodontologen sowie von deren Assistenten verwendet werden.
● Die Instrumente sind zur Verwendung mit einem endodontischen Motor vorgesehen (bei starken Krümmungen auch manuell).
Keine.
ProTaper Ultimate™-Instrumente bestehen aus drei Hauptbestandteilen: dem Arbeitsteil aus einer Nickel-Titan-Legierung, einem farbcodierten Silikonstopper (außer Instrument SX) und einem vermessingten Schaft mit einem farbcodierten Ring (außer Instrument SX).
Die Feilen werden in einem Verfahren hergestellt, das ihnen ein verfärbtes Aussehen verleiht.
Dieses patentierte Verfahren kann die Feilen leicht gebogen erscheinen lassen. Dabei handelt es sich nicht um einen Herstellungsfehler. Die Feilen lassen sich mit den Fingern leicht geradebiegen, doch dies ist für den Gebrauch nicht notwendig. Im Kanal folgen die Feilen der Wurzelkanalanatomie.
Die ProTaper Ultimate™-Produktlinie umfasst die folgenden endodontischen Instrumente:
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 21/151
Code xx-Länge y-Feilengröße z-Menge pro
Blister
B ST PULR z
19 0SX
B ST PULR z xx 0SL
B ST PULR z xx 0SH
Produkt
ProTaper Ultimate™
SX
ProTaper Ultimate™
SLIDER
ProTaper Ultimate™
SHAPER
ISO XXX YYYv
XXX-ISO-Größe
YYY-Konizität v-variable Konizität
ISO 020.003v
ISO 016.002v
ISO 020.004v
Erhältliche
Längen
19 mm
21/25/31 mm
21/25/31 mm
Packungsform und Inhalt
Blister à 6 Instrumente (steril)
Blister à 3 Instrumente (steril)
Blister à 6 Instrumente (steril)
Blister à 3 Instrumente (steril)
Blister à 6 Instrumente (steril)
Blister à 3 Instrumente (steril)
Blister à 6 Instrumente (steril)
B ST PULR z xx 0F y
B ST PULR 6 xx yyy
B ST PULR 5 xx SQE
B DD PULR
525 SQE
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™-
Sequenz
ProTaper Ultimate™
Demo-Packung
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
FX: ISO 035.012v
FXL: ISO 050.010v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
21/25/31 mm
21/25/31 mm
21/25/31 mm
25 mm
Blister à 3 Instrumente (steril)
Blister à 6 Instrumente (steril)
Blister à 1 Instrument je Ausführung
(steril)
Blister à 1 Instrument je Ausführung
(steril)
● Geräte mit Drehmomentbegrenzung sorgen für eine optimale Anwendung.
● Betreiben Sie diese mit einem endodontischen Motor in kontinuierlicher Rotation bei einer
Drehzahl von 400 U/min.
● Stellen Sie das Drehmoment auf: 4 to 5.2 N cm.
● Gleitmittel wie z. B. NaClO, EDTA, ProLube, Glyde™ sollten verwendet werden.
● Bei der adäquaten Methode zur Festlegung der Arbeitslänge werden Röntgenbilder in Verbindung mit einem Wurzelkanallängenmessgerät verwendet sowie ein Werkzeug zum Einstellen des Silikonstoppers auf die korrekte Länge.
● Beachten Sie diese Gebrauchsanweisung und die Allgemeinen Aufbereitungsanweisungen für endodontische Produkte (siehe Abschnitt 8) genau , um folgende Risiken für das Produkt, den Patienten und/oder den Anwender zu minimieren:
○ Bruch des Instruments.
○ Kreuzkontamination.
○ Hitzeentwicklung wegen ungenügender Schmierung und Spülung.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 22/151
○ Verschlucken des Arbeitsteils des Instruments.
○ Toxische oder allergische Reaktionen durch Aufbereitungsrückstände.
● Wie alle maschinellen Wurzelkanalinstrumente sollten ProTaper Ultimate™-Instrumente aufgrund des erhöhten Bruchrisikos nicht in Kanälen mit abrupten apikalen Krümmungen verwendet werden. Hier sollte im Apikalbereich mit vorgebogenen Handfeilen gearbeitet werden.
● Die Sicherheit und Wirksamkeit der Behandlung wurden bei schwangeren oder stillenden
Frauen und bei Kindern noch nicht nachgewiesen.
● Verwenden Sie bei der endodontischen Behandlung einen Kofferdam.
● Zu Ihrer eigenen Sicherheit sollten Sie persönliche Schutzausrüstung (Handschuhe, Brille,
Maske) tragen.
● Überprüfen Sie die Verpackung vor dem Gebrauch und verwenden Sie die Instrumente nicht , wenn die Verpackung beschädigt ist.
● Verwenden Sie die Instrumente nach dem Verfallsdatum nicht mehr.
● Prüfen Sie die Instrumente vor jedem Gebrauch auf Anzeichen von Defekten wie z. B. Verformungen (verbogen, aufgedreht), Brüche, Korrosion, schadhafte Schneidekanten, fehlende Farbcodierung oder Markierung. Mangelhafte Instrumente können den Verwendungszweck nicht mit der nötigen Sicherheit erfüllen und sollten verworfen werden.
● Achten Sie vor der Verwendung eines Instruments stets darauf, dass es sicher am Winkelstückkopf befestigt ist.
● Kontrollieren Sie das Instrument und reinigen Sie das Arbeitsteil bei der Anwendung häufig , achten Sie dabei auf Anzeichen von Verformungen, Dehnungen oder Abnutzung, wie etwa unebene Schneiden oder stumpfe Stellen. Mit Instrumenten, die derartige Mängel aufweisen, kann der Verwendungszweck nicht mit der nötigen Sicherheit erfüllt werden und diese sollten verworfen werden.
● Die Instrumente dürfen nicht vollständig in Natriumhypochlorit-Lösung (NaClO) eingetaucht werden. Nur das Arbeitsteil des NiTi-Instruments, das mit dem Patienten in Kontakt kommt, sollte in NaClO-Lösung mit einer Konzentration von maximal 5 % eingetaucht werden.
● Gehen Sie im apikalen Bereich und in Kanälen, die sich teilen und/oder abrupte Krümmungen oder Gegenkrümmungen aufweisen, vorsichtig vor.
● Während der gesamten Behandlung und nach jeder Verwendung eines Instruments muss der Kanal reichlich und häufig gespült werden (gemäß guter zahnärztlicher Praxis).
● Üben Sie nach apikal stets nur minimalen Druck aus. Wenden Sie beim Einführen der Feilen in den Kanal niemals Kraft an.
● Wenn das Instrument nur noch schwer vordringt, sollten Sie die Schneiden reinigen und kontrollieren, danach spülen, mit einer Handfeile wiederholen und erneut spülen.
● Bei der Aufbereitung stark gekrümmter Kanäle ist es sicherer, die Feile nur zur Aufbereitung eines einzigen Kanals zu verwenden, um die Bruchgefahr zu verringern. Beachten Sie die folgenden guten Praktiken:
○ Verwenden Sie eine neue Feile und entsorgen diese nach der Aufbereitung des Kanals
(Verwendung für nur einen Kanal).
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 23/151
○ Verwenden Sie manuelle statt rotierender Feilen.
○ Verwenden Sie feine, flexible und/oder NiTi-Feilen (dies hilft dabei, eine Verlagerung des
Kanals zu vermeiden).
○ Prüfen Sie das Arbeitsteil während der Behandlung visuell auf alle oben aufgeführten Defekte hin (d.h. nach jeder Hin- und Herbewegung).
○ Vermeiden Sie die übliche reibende, kontinuierliche Rotationsbewegung , besser sind Bewegungen in kleinen Winkeln (feilend, oszillierend wie beim Uhraufziehen oder die Balanced-Force-Technik), um die Biegeermüdung der Instrumente durch Rotation zu begrenzen und ihre Lebensdauer zu verlängern.
Nach aktuellem technischen Stand liegen bislang keine Berichte zu unerwünschten Wirkungen vor.
Die Produkte in einer trockenen und sauberen Umgebung und vor Feuchtigkeit und direkter
Sonneneinstrahlung geschützt lagern.
Alle ProTaper Ultimate™-Feilen sind für den Betrieb bei einer Drehzahl von 400 U/min und einem Drehmoment von 4 to 5.2 N cm validiert.
Die ProTaper Ultimate™-Instrumente können in allen Kanälen und vor allem in Kanälen mit unregelmäßigem Querschnitt mit bürstenden Auswärts-Bewegungen oder kurzen tupfenden Einwärts-Bewegungen verwendet werden, um schrittweise die Arbeitslänge (AL) zu erreichen.
Halten Sie das Winkelstück stets in der Mulde zwischen Daumen und Zeigefinger. Vermeiden
Sie es, die Feilen einzuschieben; lassen Sie diese besser passiv entlang des Gleitpfades vordringen.
1) Fertigen Sie Röntgenbilder zur diagnostischen Bestimmung von Weite, Länge und Krümmung der Wurzelkanäle aus verschiedenen Horizontalwinkeln an.
2) Präparieren Sie den Zugang so , dass alle Kanaleingänge leicht erkennbar sind.
3) Das Zusatzinstrument Shaping File SX kann bei beengten Platzverhältnissen zur Vorerweiterung des Kanal-Hauptteils oder zur Verlagerung des koronalsten Kanalbereichs weg von einer externen Wurzelkonkavität verwendet werden.
4) Lassen Sie unter Verwendung eines Gleitmittels den Slider in einem oder mehreren Durchgängen dem Kanal bis zum Endpunkt PASSIV folgen. Bestimmen Sie die Arbeitslänge
(AL) mit einem elektronischen Wurzelkanallängenmessgerät in Kombination mit einem
Röntgenbild und prüfen Sie den Kanal danach auf Gängigkeit.
5) Wenn der Slider nur schwer bis zum Endpunkt vordringt, wählen Sie eine Handfeile aus
Edelstahl (SSt) in einer kleinen Größe aus. Sondieren Sie unter Verwendung eines Gleitmittels den Kanal, bestimmen Sie die AL, prüfen Sie die Gängigkeit und kontrollieren Sie den Gleitpfad. Wiederholen Sie nun Schritt 4.
6) Wählen Sie den Shaper aus und führen Sie diesen unter Verwendung von NaClO in einem oder mehreren Durchgängen entlang des Gleitpfades bis zur AL ein.
7) Spülen Sie nach dem Entfernen des Shapers mit Endo Activator, um das Debris zu zerkleinern und aufzulösen, danach spülen Sie erneut, um es zu entfernen.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 24/151
8) Kontrollieren Sie besonders bei stärker gekrümmten Kanälen die AL erneut die AL.
9) Wählen Sie den FINISHER F1 (020.07) aus und lassen diesen in einem oder mehreren
Durchgängen dem Kanal passiv bis zur AL folgen. Ziehen Sie diesen heraus und kontrollieren Sie den apikalen Schneidenbereich. Ist dieser mit Dentin (Debris) gefüllt, ist die Aufbereitung damit beendet.
10) Ist der FINISHER F1 bei erreichter AL locker und der apikale Schneidenbereich nicht mit
Dentin gefüllt, verwenden Sie den FINISHER F2 (025.08) wie oben zum FINISHER F1 beschrieben.
11) Ist der FINISHER F2 bei erreichter AL locker und der apikale Schneidenbereich nicht mit
Dentin gefüllt, verwenden Sie den FINISHER F3 (030.09) wie oben zu FINISHER F1 und
F2 beschrieben.
12) Reinigen und kontrollieren Sie bei jeder Feile nach dem Herausziehen den Schneidenbereich , spülen und verwenden Sie daraufhin eine Feile der Größe 10 oder den Endo Activator, um Debris zu zerkleinern, spülen Sie danach erneut.
13) Prüfen Sie den Schneidenbereich nach dem Herausziehen auf Aufdrehung, Begradigung oder Dehnung. Verwerfen Sie die Feile, wenn Verformungen vorliegen, und verwenden Sie eine neue ProTaper Ultimate™-Feile.
14) Die Aufbereitung ist abgeschlossen, wenn der apikale Schneidenbereich eines Finishers mit Debris gefüllt ist und der Guttapercha-Mastercone oder Size-Verifier in der entsprechenden Größe bei erreichter AL exakt passt.
Verwenden Sie die Zusatz-Finisher nur in relativ weiten und geraden Kanälen, wie z. B. bei mittleren Oberkiefer-Schneidezähnen, manchen palatinalen oder distalen
Kanälen von Molaren oder bei pathologischen oder iatrogenen Defekten.
15) Wählen Sie entweder das maschinelle oder das manuelle Zusatzinstrument FINISHER FX
(035.12) , wenn die AL bestimmt und der Kanal auf Gängigkeit geprüft worden ist. Lassen
Sie diese eigenständige Feile direkt und passiv in der oben beschriebenen Technik dem
Kanal bis zur AL folgen. Ziehen Sie diese heraus und kontrollieren Sie den apikalen
Schneidenbereich. Ist dieser mit Dentin (Debris) gefüllt, ist die Aufbereitung damit beendet.
16) Ist der FINISHER FX bei erreichter AL locker und der apikale Schneidenbereich nicht mit
Dentin gefüllt, verwenden Sie die maschinelle oder manuelle Zusatzfeile FINISHER FXL
(050.10) wie oben für den FINISHER FX beschrieben.
17) Die Aufbereitung ist abgeschlossen, wenn der apikale Schneidenbereich eines Zusatz-Finishers mit Debris gefüllt ist und der Guttapercha-Mastercone oder Size-Verifier in der entsprechenden Größe bei erreichter AL exakt passt.
● Bitte beachten Sie die Allgemeinen Aufbereitungsanweisungen für endodontische Produkte
( https://www.dentsplysirona.com/en/service-contact/ifu.html
) zur Reinigung, Desinfektion und Sterilisation.
● Die Silikonstopper dürfen nicht wiederverwendet werden. Entfernen und entsorgen Sie die
Silikonstopper nach jeder Verwendung.
● Die Produkte sind entsprechend den lokalen Vorschriften zur sicheren Entsorgung von scharfkantigen und kontaminierten Materialien zu entsorgen.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 25/151
● Jeder ernste Vorfall im Zusammenhang mit dem Produkt ist dem Hersteller und der zuständigen Behörde gemäß den lokalen Vorschriften zu melden.
● Die Sterilität kann nicht garantiert werden, wenn die Verpackung geöffnet, beschädigt oder durchnässt ist.
● Eine kostenlose gedruckte Kopie der Gebrauchsanweisung erhalten Sie auf unserer Website unter dem Punkt „Get a print copy of IFU“ https://www.dentsplysirona.com/en/servicecontact/ifu.html
.
● Zur Erläuterung der nicht harmonisierten Symbole für Gebrauchsanweisungen und Etiketten siehe „IFU Symbols“ ( https://www.dentsplysirona.com/en/service-contact/ifu/ifusymbols.html
).
2797
Autorisierter Vertreter
DENTSPLY DeTrey GmbH
De-Trey-Straße 1
D-78467 Konstanz
Germany
Hersteller
Maillefer Instruments Holding Sàrl
Chemin du Verger 3
CH-1338 Ballaigues
Suisse www.dentsplysirona.com/ifu
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 26/151
ProTaper Ultimate™
ÚNICAMENTE PARA USO DENTAL
ÚNICAMENTE PARA USO PROFESIONAL
ESTÉRIL - ESTERILIZADO POR RADIACIÓN
APTO PARA SER REPROCESADO
INSTRUMENTOS DE ENDODONCIA ESTÉRILES REUTILIZABLES
● Los productos están indicados para el tratamiento de la enfermedad endodóntica.
● Finalidad prevista: Instrumento accionado por motor indicado para la preparación del conducto radicular (permeabilización, conformado y desbridamiento del conducto radicular).
● Los instrumentos de endodoncia deben ser utilizados únicamente en un entorno clínico u hospitalario, siguiendo las buenas prácticas odontológicas, por profesionales dentales cualificados, como médicos generales, así como especialistas en endodoncia (endodoncistas) y asistentes dentales.
● Los instrumentos se utilizarán en combinación con un motor endodóntico (manualmente en curvaturas severas).
Ninguna.
Los instrumentos ProTaper Ultimate™ están formados por tres componentes principales: la pieza operante hecha de una aleación de níquel-titanio, un tapón de silicona de color (excepto para el instrumento SX) y un vástago de latón chapado con un anillo de color (excepto para el instrumento SX).
Las limas están fabricadas siguiendo un proceso que les confiere una apariencia coloreada.
Debido a este proceso exclusivo, es posible que el aspecto de las limas sea algo curvado. Esto no es un defecto de fabricación. Aunque la lima se puede enderezar fácilmente con los dedos, no es necesario enderezarla antes del uso. Cuando esté dentro del conducto, la lima se adaptará a la anatomía del conducto radicular.
La gama de ProTaper Ultimate™ comprende los siguientes instrumentos endodónticos:
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 27/151
Código longitud xx tamaño y de la lima cantidad z por envase blíster
Producto
ISO XXX YYYv tamaño XXX-ISO tamaño de punta
YYY punta variable-v
Longitudes disponibles
Embalaje y contenido
B ST PULR z
19 0SX
B ST PULR z xx 0SL
B ST PULR z xx 0SH
B ST PULR z xx 0F y
B ST PULR 6 xx yyy
B ST PULR 5 xx SQE
B DD PULR
525 SQE
ProTaper Ultimate™
SX
ProTaper Ultimate™
SLIDER
ProTaper Ultimate™
SHAPER
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
Sequence
ProTaper Ultimate™
Demo Pack
ISO 020.003v
ISO 016.002v
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
FX: ISO 035.012v
FXL: ISO 050.010v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
19 mm
21/25/31 mm
21/25/31 mm
21/25/31 mm
21/25/31 mm
21/25/31 mm
25 mm
Envase blíster con 6 instrumentos (estériles)
Envase blíster con 3 instrumentos (estériles)
Envase blíster con 6 instrumentos (estériles)
Envase blíster con 3 instrumentos (estériles)
Envase blíster con 6 instrumentos (estériles)
Envase blíster con 3 instrumentos (estériles)
Envase blíster con 6 instrumentos (estériles)
Envase blíster con 3 instrumentos (estériles)
Envase blíster con 6 instrumentos (estériles)
Envase blíster con 1 instrumento de cada (estéril)
Envase blíster con 1 instrumento de cada (estéril)
● Los dispositivos de control de par garantizan un uso óptimo.
● Utilización con motores de endodoncia en rotación constante a una velocidad de 400 rpm.
● Ajuste el par a: 4 to 5.2 Ncm.
● Deben utilizarse lubricantes como NaOCl, EDTA, ProLube, Glyde™.
● El uso de radiografías en combinación con un localizador de ápice y una herramienta para ajustar el tope de silicona a la longitud de trabajo correcta es el método adecuado para determinar la longitud de trabajo.
● Siga estrictamente estas instrucciones de uso y las instrucciones generales de procesamiento para productos endodónticos (consulte el apartado 8) para minimizar los siguientes riesgos para el dispositivo, el paciente o el usuario:
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 28/151
○ Rotura del instrumento.
○ Contaminación cruzada.
○ Generación de calor debido a una lubricación e irrigación insuficientes.
○ Ingestión de la pieza operante del instrumento.
○ Reacciones tóxicas o alérgicas causadas por los residuos del proceso.
● Al igual que todos los instrumentos para el conducto radicular, ProTaper Ultimate™ los instrumentos no deben utilizarse en conductos radiculares que presenten curvaturas apicales severas y abruptas debido al elevado riesgo de separación. En este caso, deben utilizarse limas manuales precurvadas en la región apical.
● No se han determinado la seguridad y efectividad de uso en mujeres embarazadas o en periodo de lactancia ni en niños.
● Utilice un sistema de dique de goma durante el procedimiento endodóntico.
● Por su propia seguridad, utilice equipos de protección individual (guantes, gafas y mascarilla).
● Inspeccione el embalaje antes del uso y, si está dañado no utilice los instrumentos.
● No use los instrumentos si se ha excedido la fecha de caducidad.
● Compruebe el instrumento antes de cada uso por si presenta indicios de defectos como deformaciones (doblado, desenrollado), rotura, corrosión, bordes de corte dañados, pérdida de la identificación por color o marcado. Con estas indicaciones los dispositivos no son capaces de cumplir con el uso previsto al nivel de seguridad requerido; los instrumentos deben ser desechados.
● Antes de usar cualquier instrumento asegúrese de que está bien conectado al cabezal del contra-ángulo.
● Compruebe el instrumento y limpie la pieza operante con frecuencia durante la instrumentación, comprobando si hay signos de distorsión, elongación o desgaste, como irregularidades en las espiras o zonas mate. Con estas indicaciones los dispositivos no son capaces de cumplir con el uso previsto al nivel de seguridad requerido; los instrumentos deben ser desechados.
● Los instrumentos no deben sumergirse completamente en soluciones de hipoclorito de sodio (NaOCl). Solamente la pieza de trabajo del instrumento de NiTi que está en contacto con el paciente deberá sumergirse en una solución de NaOCl con una concentración NO superior al 5 %.
● Actúe con precaución en la zona apical y en conductos que se dividan o muestren curvaturas o recurvaturas abruptas.
● Irrigue abundantemente y con frecuencia el conducto durante todo el procedimiento y después de cada instrumento utilizado (según las buenas prácticas odontológicas).
● Utilice siempre la mínima presión apical. No fuerce nunca las limas para introducirlas en el conducto.
● Cuando el instrumento no progresa fácilmente, limpie e inspeccione las espiras cortantes, luego irrigue, recapitule con una lima manual y vuelva a irrigar.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 29/151
● Para dar forma a conductos extremadamente curvados es más seguro usar la lima solo para dar forma a un conducto con el fin de reducir el riesgo de rotura. Preste atención a las buenas prácticas siguientes:
○ Use una lima nueva y deséchela una vez tratado el conducto (uso en un solo conducto).
○ Use limas manuales en lugar de rotatorias.
○ Use limas de NiTi o pequeñas y flexibles (para evitar la transportación del conducto).
○ Inspeccione visualmente la pieza operante para detectar los defectos enumerados en el párrafo anterior durante el uso (es decir, después de cada movimiento de vaivén).
○ Evite el movimiento rotativo continuo de ensanchado estándar y en lugar de ello haga pequeños movimientos en ángulo (movimiento de limado, movimiento de oscilación de dar cuerda a un reloj o técnica de fuerza equilibrada) para limitar la fatiga de flexión rotativa de los instrumentos y mejorar su previsión de vida útil.
En el estado técnico actual no se ha detectado ninguna reacción adversa por el momento.
Guarde los productos en un lugar seco y limpio alejado de fuentes de humedad y de la luz solar directa.
Todas las limas ProTaper Ultimate™ han sido validadas para ser usadas a 400 rpm y a un par de 4 to 5.2 Ncm.
Los instrumentos ProTaper Ultimate™ pueden utilizarse con un movimiento de cepillado hacia fuera en todos los conductos, especialmente en los que presentan una sección transversal irregular, o con un ligero movimiento de picoteo hacia dentro para avanzar progresivamente hacia la longitud de trabajo.
Apoye siempre la pieza de mano entre los dedos pulgar e índice. Evite empujar; más bien, deje que las limas progresen de forma pasiva y sigan el trayecto de deslizamiento.
1) Revise las radiografías tomadas desde diferentes ángulos del plano horizontal para determinar la anchura, la longitud y la curvatura de cualquier raíz o conducto(s).
2) Prepare un acceso que permita identificar fácilmente cada orificio del conducto.
3) La lima auxiliar de conformación, SX, puede utilizarse cuando hay un espacio limitado, o para agrandar previamente el cuerpo de un conducto o reubicar el aspecto más coronal de un conducto lejos de una concavidad radicular externa.
4) En presencia de un lubricante, seleccione el Slider y siga PASIVAMENTE el conducto, en una o más pasadas, hasta su terminación. Determine la longitud de trabajo utilizando un localizador de ápice electrónico en combinación con una imagen radiográfica, y a continuación confirme la permeabilidad.
5) Si el Slider no alcanza fácilmente la terminación del conducto, seleccione una lima manual de acero inoxidable (AI) de pequeño tamaño. En presencia de un lubricante, cateterice este conducto, establezca la longitud de trabajo, confirme la permeabilidad y verifique el trayecto de deslizamiento. Ahora, repita el paso n.º 4 anterior.
6) En presencia de NaOCl, seleccione el Shaper y avance a lo largo del trayecto de deslizamiento, en una o más pasadas, hasta alcanzar la longitud de trabajo.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 30/151
7) Al retirar el Shaper, irrigue, use EndoActivate para romper los restos y moverlos a la solución, luego vuelva a irrigar para liberar estos restos.
8) Reconfirme la longitud de trabajo, especialmente en los conductos más curvos.
9) Seleccione FINISHER F1 (020.07) y siga pasivamente el conducto hasta la longitud de trabajo en una o varias pasadas. Retire e inspeccione las espiras apicales. Cuando se haya cargado con restos de dentina, la preparación estará terminada.
10) Si la FINISHER F1 está suelta en toda su longitud y sus espiras apicales no están cargadas de restos, seleccione la FINISHER F2 (025.08) y utilice esta lima de la misma manera descrita anteriormente para la FINISHER F1.
11) Si la FINISHER F2 está suelta en toda su longitud y sus espiras apicales no están cargadas de restos, seleccione la FINISHER F3 (030.09) y utilice esta lima de la misma manera descrita anteriormente para la FINISHER F1 y F2.
12) Al retirar una lima determinada, limpie e inspeccione sus espiras cortantes, irrigue, recapitule con una lima de tamaño 10 o con Endo Activator para romper los restos y vuelva a irrigar.
13) Inspeccione las espiras cortantes de la lima al retirarla para detectar la presencia de desprendimiento, enderezamiento o estiramiento. Si observa alguna deformación, deseche la lima y use una nueva ProTaper Ultimate™.
14) La preparación estará terminada cuando la extensión apical de cualquier Finisher esté cargada de restos, y el cono maestro de gutapercha o el Size Verifier correspondiente encaje en la longitud de trabajo.
Utilice los Finishers auxiliares solo en los conductos más grandes y rectos, como los incisivos centrales maxilares, algunos conductos palatales o distales de molares, o cuando haya un defecto patológico o iatrogénico:
15) Seleccione la lima auxiliar de accionamiento mecánico o manual FINISHER FX (035.12) cuando se haya establecido la longitud de trabajo y esté confirmada la permeabilidad. Utilice esta lima auxiliar directamente y de forma pasiva para seguir el conducto hasta la longitud de trabajo utilizando la misma técnica descrita anteriormente. Retire e inspeccione las espiras apicales. Cuando se haya cargado con restos de dentina, la preparación estará terminada.
16) Si la FINISHER FX está suelta en toda su longitud y sus espiras apicales no están cargadas de restos, seleccione la FINISHER FXL (050.10) de accionamiento mecánico o auxiliar manual y utilice esta lima de la misma manera descrita anteriormente para la FI-
NISHER FX.
17) La preparación estará terminada cuando la extensión apical de cualquier Finisher auxiliar esté cargada de restos, y el cono maestro de gutapercha o el Size Verifier correspondiente encaje en la longitud de trabajo.
● Siga las instrucciones generales de procesamiento para productos endodónticos ( https:// www.dentsplysirona.com/en/service-contact/ifu.html
) para los pasos de desinfección, limpieza y esterilización.
● No reutilice los topes de silicona. Quite y descarte los topes de silicona después de cada uso.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 31/151
● Los productos se han de desechar de conformidad con la normativa local para la eliminación segura de productos punzantes y contaminados.
● Cualquier incidente grave relacionado con el producto debe notificarse al fabricante y a la autoridad competente según la normativa local.
● La esterilidad no está garantizada si el embalaje está abierto, dañado o mojado.
● Para obtener una copia impresa gratuita de las instrucciones de uso, consulte el apartado
«Obtener una copia impresa de las instrucciones de uso» en el sitio web https:// www.dentsplysirona.com/en/service-contact/ifu.html
.
● Explicación de los símbolos no armonizados para las instrucciones de uso y etiquetas, consulte Símbolos de las instrucciones de uso ( https://www.dentsplysirona.com/en/servicecontact/ifu/ifu-symbols.html
).
2797
Representante autorizado
DENTSPLY DeTrey GmbH
De-Trey-Straße 1
D-78467 Konstanz
Germany
Fabricante
Maillefer Instruments Holding Sàrl
Chemin du Verger 3
CH-1338 Ballaigues
Suisse www.dentsplysirona.com/ifu
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 32/151
ProTaper Ultimate™
APENAS PARA UTILIZAÇÃO EM MEDICINA DENTÁRIA
APENAS PARA UTILIZAÇÃO PROFISSIONAL
ESTÉRIL – ESTERILIZADO POR RADIAÇÃO
RESPONSÁVEL PELO REPROCESSAMENTO
INSTRUMENTOS ENDODÔNTICOS ESTÉREIS REUTILIZÁVEIS
● Os produtos são indicados para o tratamento de doença endodôntica.
● Finalidade prevista: instrumento motorizado concebido para a preparação do canal radicular (via de permeabilidade, moldagem e desbridamento do canal radicular).
● Os instrumentos endodônticos devem ser utilizados apenas num ambiente clínico ou hospitalar, seguindo uma boa prática dentária, por profissionais dentários qualificados, como clínicos gerais, bem como por especialistas de Endodontia (Endodontistas) e assistentes dentários.
● Os instrumentos devem ser utilizados em conjunto com um motor endodôntico (manualmente em curvaturas acentuadas).
Nenhuma.
Os instrumentos ProTaper Ultimate™ são feitos de três componentes principais: a peça de trabalho, feita de uma liga de níquel-titânio, um batente de silicone colorido (exceto para o instrumento SX) e uma haste de latão cromado com um anel colorido (exceto para o instrumento
SX).
As limas são fabricadas com um processo que resulta numa lima com um aspeto colorido.
Devido a este processamento proprietário, as limas podem parecer ligeiramente curvas. Não se trata de um defeito de fabrico. Embora a lima possa ser facilmente endireitada apenas com os dedos, não é necessário endireitá-la antes da utilização. Uma vez dentro do canal, a lima seguirá a anatomia do canal radicular.
A gama ProTaper Ultimate™ inclui os seguintes instrumentos endodônticos:
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 33/151
Código comprimento xx tamanho y da lima quantidade z por blister
B ST PULR z
19 0SX
B ST PULR z xx 0SL
B ST PULR z xx 0SH
Produto
ProTaper Ultimate™
SX
ProTaper Ultimate™
SLIDER
ProTaper Ultimate™
SHAPER
ISO XXX YYYv
Tamanho XXX-ISO
Tamanho YYY-cone cone variável v
ISO 020.003v
ISO 016.002v
ISO 020.004v
Comprimentos disponíveis
19 mm
21/25/31 mm
21/25/31 mm
Embalagem e conteúdo
Blister com 6 instrumentos (estéril)
Blister com 3 instrumentos (estéril)
Blister com 6 instrumentos (estéril)
Blister com 3 instrumentos (estéril)
Blister com 6 instrumentos (estéril)
Blister com 3 instrumentos (estéril)
Blister com 6 instrumentos (estéril)
B ST PULR z xx 0F y
B ST PULR 6 xx yyy
B ST PULR 5 xx SQE
B DD PULR
525 SQE
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
Sequence
ProTaper Ultimate™
Demo Pack
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
FX: ISO 035.012v
FXL: ISO 050.010v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
21/25/31 mm
21/25/31 mm
21/25/31 mm
25 mm
Blister com 3 instrumentos (estéril)
Blister com 6 instrumentos (estéril)
Blister com 1 instrumento de cada (estéril)
Blister com 1 instrumento de cada (estéril)
● Dispositivos de controlo do binário asseguram uma utilização ideal.
● Utilizar com motores endodônticos numa rotação constante, a uma velocidade de 400 rpm.
● Definir binário para: 4 to 5.2 Ncm.
● Devem ser utilizados lubrificantes como NaOCl, EDTA, ProLube, Glyde™.
● A utilização de radiografias em combinação com um localizador apical e uma ferramenta para ajustar o batente de silicone ao comprimento de trabalho correto é o método apropriado para a determinação do comprimento de trabalho.
● Seguir rigorosamente estas Instruções de Utilização e as Instruções Gerais de Processamento para Produtos Endodônticos (ver secção 8) para minimizar os seguintes riscos para o dispositivo, o paciente e/ou o utilizador:
○ Quebra do instrumento.
○ Contaminação cruzada.
○ Geração de calor devido a lubrificação e irrigação insuficientes.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 34/151
○ Deglutição da peça de trabalho do instrumento.
○ Reações tóxicas ou alérgicas causadas por resíduos de processamento.
● De modo similar a todos os instrumentos motorizados para o canal radicular, os instrumentos ProTaper Ultimate™ não devem ser usados num canal radicular com uma curvatura apical abrupta devido ao aumento do risco de separação. Neste caso, devem ser usadas limas manuais pré-curvadas na região apical.
● A segurança e a eficácia de utilização não foram determinadas para grávidas ou lactantes ou para crianças.
● Utilizar um sistema de dique de borracha durante o procedimento endodôntico.
● Para a sua própria segurança, utilizar equipamento de proteção individual (luvas, óculos, máscara).
● Inspecionar a embalagem antes da utilização e não utilizar os instrumentos se a embalagem estiver danificada.
● Não utilizar os instrumentos após a data de validade.
● Verificar o instrumento antes de cada utilização para detetar sinais de defeitos, como deformações (dobrado, desenrolado), quebra, corrosão, arestas de corte danificadas, perda da cor da codificação ou da marcação. Com estas indicações, os dispositivos não conseguem cumprir a utilização prevista com o nível de segurança exigido, pelo que devem ser descartados.
● Antes de usar qualquer instrumento, certifique-se de que o mesmo está bem ligado à cabe-
ça contra-ângulo.
● Durante a instrumentação, verificar os instrumentos e limpar frequentemente a peça de trabalho, procurando sinais de distorção, alongamento ou desgaste, como estrias desiguais ou manchas baças. Com estas indicações de que os dispositivos não conseguem cumprir a utilização prevista com o nível de segurança exigido, os instrumentos devem ser descartados.
● Os instrumentos não devem ser totalmente mergulhados numa solução de hipoclorito de sódio (NaOCl). Apenas a peça de trabalho do instrumento NiTi que está em contacto com o paciente deve ser imersa numa solução de NaOCl concentrada de NÃO mais de 5%.
● Proceder com cuidado na zona apical e em canais que se dividam e/ou apresentem curvaturas abruptas ou múltiplas.
● Irrigar abundante e frequentemente o canal durante todo o procedimento e após a utiliza-
ção de cada instrumento (de acordo com as boas práticas dentárias).
● Utilizar sempre uma pressão apical mínima. Nunca forçar as limas no interior do canal.
● Quando o instrumento não progredir facilmente, limpar e inspecionar as ranhuras de corte, depois irrigar, repetir com uma lima manual e irrigar novamente.
● Para a instrumentação de canais extremamente curvos, é mais seguro utilizar a lima apenas para a instrumentação de um canal, de modo a reduzir o risco de quebra. Prestar aten-
ção às boas práticas seguintes:
○ Utilizar uma lima nova e eliminá-la após o tratamento do canal (utilização num único canal).
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 35/151
○ Utilizar limas manuais ao invés de limas rotativas.
○ Utilizar limas pequenas, flexíveis e/ou de NiTi (isto ajudará a evitar o transporte do canal).
○ Inspecionar visualmente a peça de trabalho para detetar a presença dos defeitos indicados no parágrafo anterior durante a utilização (ou seja, após cada movimento).
○ Evitar o movimento rotativo contínuo de alargamento padrão e, em vez disso, executar pequenos movimentos em ângulo (movimento da lima, movimento de rotação para a direita ou técnica de força equilibrada) para limitar a fadiga de flexão rotativa dos instrumentos e aumentar a sua vida útil.
Não foram reportadas quaisquer reações adversas até à data.
Manter os produtos num ambiente seco e limpo, longe de fontes de humidade e da luz solar direta.
Todas as ProTaper Ultimate™ foram validadas para serem usadas a 400 rpm e a um binário de
4 to 5.2 Ncm.
Os instrumentos ProTaper Ultimate™ podem ser usados movimentando o escovilhão de dentro para fora em todos os canais, especialmente nos canais que exibem uma secção transversal irregular, ou com movimentos ligeiros de “picada”, de fora para dentro, para avançar progressivamente em direção ao comprimento de trabalho (CT).
Segurar na peça de mão sempre com cuidado, mantendo a cinta entre o polegar e o indicador.
Evitar empurrar; ao invés disso, permitir que as limas avancem sem resistência e sigam a via de permeabilidade.
1) Analisar diferentes radiografias com variações da angulação horizontal, para determinar a largura, o comprimento e a curvatura de qualquer raiz e do(s) seu(s) canal(ais).
2) Preparar um acesso que permita a fácil identificação de cada orifício do canal.
3) A lima de modelagem auxiliar SX pode ser utilizada quando há pouco espaço, ou para pré-alargar o corpo de um canal, ou para afastar o aspeto mais coronal de um canal de uma concavidade de raiz externa.
4) Na presença de um lubrificante, selecionar a Slider e seguir PASSIVAMENTE o canal, numa ou em mais passagens, até ao seu término. Determinar o CT com um localizador apical eletrónico em conjunto com uma imagem radiográfica, depois confirmar a patência.
5) Se a Slider não atingir facilmente o término do canal, selecionar uma lima manual pequena de aço inoxidável (AI). Na presença de um lubrificante, cateterizar este canal, estabelecer o comprimento de trabalho, confirmar a patência e verificar a via deslizante. Agora, repetir o passo n.º 4 acima.
6) Na presença de NaOCl, selecionar a Shaper e avançar ao longo da via deslizante, numa ou em mais passagens, até que o CT seja alcançado.
7) Ao remover a Shaper, irrigar, endoativar para quebrar os resíduos e transferi-los para a solução, e depois irrigar novamente para libertar esses resíduos.
8) Reconfirmar o CT, especialmente em canais mais curvos.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 36/151
9) Selecionar a FINISHER F1 (020.07) e seguir passivamente o canal até ao CT, numa ou em mais passagens. Remover e inspecionar as respetivas ranhuras apicais. Quando estiverem cheias de resíduos dentinários, a preparação está concluída.
10) Se a FINISHER F1 estiver folgada em comprimento e as ranhuras apicais não estiverem cheias de resíduos, selecionar a FINISHER F2 (025.08) e utilizá-la tal como descrito acima para a FINISHER F1.
11) Se a FINISHER F2 estiver folgada em comprimento e as ranhuras apicais não estiverem cheias de resíduos, selecionar a FINISHER F3 (030.09) e utilizá-la tal como descrito acima para a FINISHER F1 e F2.
12) Ao remover qualquer lima, limpar e inspecionar as suas ranhuras de corte, irrigar, repetir com uma lima de tamanho 10 ou Endo Activator para quebrar os resíduos, e depois irrigar novamente.
13) Inspecionar as ranhuras de corte da lima após a sua remoção para verificar a presença de desenrolamento, endireitamento ou alongamento. Se for detetada deformação, descartar e utilizar uma nova lima ProTaper Ultimate™.
14) A preparação está terminada quando a extensão apical de qualquer Finisher estiver carregada com resíduos e o Cone Gutta-Percha Master ou o verificador de tamanho correspondente couber no CT.
Utilizar os Finishers auxiliares apenas em canais maiores e mais retos, tais como incisivos centrais maxilares, alguns canais palatinos ou distais de molares, ou quando existir um defeito patológico ou iatrogénico:
15) Selecionar uma lima auxiliar manual ou mecânica FINISHER FX (035.12), quando se estabelecer o comprimento de trabalho e for possível confirmar a patência. Utilizar esta lima independente direta e passivamente para seguir o canal até ao CT, usando a mesma técnica descrita acima. Remover e inspecionar as respetivas ranhuras apicais. A preparação está concluída quando a lima estiver cheia de resíduos dentinários.
16) Se a FINISHER FX estiver folgada no comprimento e as ranhuras apicais não estiverem cheias de resíduos, selecionar uma lima auxiliar FINISHER FXL (050.10) manual ou motorizada e utilizá-la tal como descrito acima para a FINISHER FX.
17) A preparação está terminada quando a extensão apical de qualquer Finisher auxiliar estiver carregada com resíduos, e o Cone Gutta-Percha Master ou o verificador de tamanho correspondente couber no CT.
● Seguir as Instruções Gerais de Processamento de Produtos Endodônticos ( https:// www.dentsplysirona.com/en/service-contact/ifu.html
) para os passos de desinfeção, limpeza e esterilização.
● Não reutilizar os batentes de silicone. Remover e descartar os batentes de silicone após cada utilização.
● Os produtos devem ser eliminados de acordo com os regulamentos locais para a elimina-
ção segura de dispositivos afiados e contaminados.
● Qualquer incidente grave relacionado com o produto deve ser comunicado ao fabricante e
à autoridade competente, de acordo com os regulamentos locais.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 37/151
● A esterilidade não pode ser garantida caso a embalagem seja aberta, danificada ou molhada.
● Para obter uma cópia impressa das Instruções de utilização, ver a secção “Obter uma cópia impressa das Instruções de utilização” no website https://www.dentsplysirona.com/en/ service-contact/ifu.html
.
● Explicação dos símbolos não harmonizados para as instruções de utilização e etiquetas, ver símbolos das Instruções de Utilização ( https://www.dentsplysirona.com/en/servicecontact/ifu/ifu-symbols.html
).
2797
Representante autorizado
DENTSPLY DeTrey GmbH
De-Trey-Straße 1
D-78467 Konstanz
Germany
Fabricante
Maillefer Instruments Holding Sàrl
Chemin du Verger 3
CH-1338 Ballaigues
Suisse www.dentsplysirona.com/ifu
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 38/151
ProTaper Ultimate™
UITSLUITEND VOOR TANDHEELKUNDIG GEBRUIK
UITSLUITEND VOOR PROFESSIONEEL GEBRUIK
STERIEL - GESTERILISEERD DOOR MIDDEL VAN STRALING
VERANTWOORDELIJK VOOR HERVERWERKING
HERBRUIKBARE STERIELE ENDODONTISCHE INSTRUMENTEN
● De producten zijn bestemd voor de behandeling van endodontische aandoeningen.
● Beoogde doel: Motoraangedreven instrument bedoeld voor wortelkanaalpreparatie (glijpad, vormgeving en weefselverwijdering van het wortelkanaal).
● Endodontische instrumenten mogen alleen worden gebruikt in een klinische of ziekenhuisomgeving, volgens goede tandheelkundige praktijken, door gekwalificeerde tandheelkundigen zoals huisartsen, evenals endo-specialisten (endodontisten) en tandartsassistenten.
● Instrumenten moeten worden gebruikt in combinatie met een endodontische motor (handmatig bij ernstige krommingen).
Geen.
ProTaper Ultimate™ instrumenten zijn gemaakt van drie hoofdcomponenten: het werkende deel gemaakt van een nikkel-titanium legering, een gekleurde siliconen stop (met uitzondering van instrument SX) en een vergulde messing schacht met een gekleurde ring (met uitzondering van instrument SX).
De vijlen zijn zodanig vervaardigd dat ze een gekleurde look hebben.
Als gevolg van deze exclusieve verwerking kunnen vijlen iets gebogen zijn. Dit is geen fabricagefout. Hoewel de vijl met de vingers gemakkelijk rechtgebogen kan worden, is het niet noodzakelijk dit vóór gebruik te doen. Eenmaal in het kanaal volgt de vijl de anatomie van het wortelkanaal.
Het ProTaper Ultimate™ assortiment omvat de volgende endodontische instrumenten:
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 39/151
Code xx-lengte y-vijlmaat z-hoeveelheid per blister
B ST PULR z
19 0SX
B ST PULR z xx 0SL
B ST PULR z xx 0SH
Product
ProTaper Ultimate™
SX
ProTaper Ultimate™
SLIDER
ProTaper Ultimate™
SHAPER
ISO XXX YYYv
XXX-ISO-maat
YYY- conusmaat v-variabele conus
ISO 020.003v
ISO 016.002v
ISO 020.004v
Beschikbare lengtes
19 mm
21/25/31 mm
21/25/31 mm
Verpakking en inhoud
Blister met 6 instrumenten (steriel)
Blister met 3 instrumenten (steriel)
Blister met 6 instrumenten (steriel)
Blister met 3 instrumenten (steriel)
Blister met 6 instrumenten (steriel)
Blister met 3 instrumenten (steriel)
Blister met 6 instrumenten (steriel)
B ST PULR z xx 0F y
B ST PULR 6 xx yyy
B ST PULR 5 xx SQE
B DD PULR
525 SQE
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
Volgorde
ProTaper Ultimate™
Demopakket
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
FX: ISO 035.012v
FXL: ISO 050.010v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
21/25/31 mm
21/25/31 mm
21/25/31 mm
25 mm
Blister met 3 instrumenten (steriel)
Blister met 6 instrumenten (steriel)
Blisterverpakking met 1 instrument van elk instrument (steriel)
Blisterverpakking met 1 instrument van elk instrument (steriel)
● Instrumenten met koppelregelaars maken optimaal gebruik mogelijk.
● Gebruik met endodontische motoren in een constante rotatie met een snelheid van400 toeren per minuut.
● Stel koppel in op: 4 to 5.2 Ncm.
● Er moeten glijmiddelen zoals NaOCl, EDTA, ProLube, Glyde™ worden gebruikt.
● Het gebruik van röntgenfoto's in combinatie met een apexlocator en een hulpmiddel om de silicone stopper op de juiste werklengte af te stellen, is de geschikte methode voor het bepalen van de werklengte.
● Volg deze gebruiksaanwijzing en de algemene verwerkingsinstructies voor endodontische producten (zie deel 8) strikt op om de volgende risico's voor het hulpmiddel, de patiënt en/of de gebruiker tot een minimum te beperken:
○ Breuk van instrument.
○ Kruisbesmetting.
○ Warmteontwikkeling door onvoldoende smering en spoeling.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 40/151
○ Inslikken van het werkende deel van het instrument.
○ Giftige of allergische reacties veroorzaakt door verwerkingsresten.
● Net als alle machinaal aangedreven wortelkanaalinstrumenten, mogen ProTaper
Ultimate™-instrumenten niet worden gebruikt in een wortelkanaal met een abrupte apicale kromming vanwege het verhoogde risico op een breuk. In dit geval moeten voorgebogen handvijlen worden gebruikt in het apicaal gebied.
● De veiligheid en werkzaamheid zijn niet vastgesteld bij zwangere vrouwen, vrouwen die borstvoeding geven en kinderen.
● Gebruik een rubberdamsysteem tijdens de endodontische behandeling.
● Draag voor uw eigen veiligheid persoonlijke beschermingsmiddelen (handschoenen, bril, masker).
● Inspecteer de verpakking voor gebruik en gebruik de instrumenten niet als de verpakking beschadigd is.
● Gebruik de instrumenten niet na de vervaldatum.
● Controleer het instrument voor elk gebruik op tekenen van defecten zoals vervormingen
(verbogen, afgewikkeld), breuk, corrosie, beschadigde snijkanten, verdwenen kleurcodering of -markering. Met deze indicaties kunnen de apparaten niet voldoen aan het beoogde gebruik met het vereiste veiligheidsniveau, instrumenten moeten worden weggegooid.
● Voordat u een instrument gebruikt, moet u ervoor zorgen dat het goed is aangesloten op de contra-hoek kop.
● Controleer het instrument en reinig het werkzame deel regelmatig tijdens gebruik. Let hierbij op tekenen van vervorming, rek of slijtage, zoals ongelijkmatige windingen en doffe plekken. Indien de apparaten op basis van deze indicaties niet kunnen voldoen aan het beoogde gebruik met het vereiste veiligheidsniveau, moeten instrumenten worden weggegooid.
● Instrumenten mogen niet volledig worden ondergedompeld in een natriumhypochlorietoplossing (NaOCl). Alleen het werkende deel van het NiTi-instrument dat in contact met de patiënt komt, mag ondergedompeld worden in een oplossing met NIET meer dan 5% NaO-
Cl.
● Ga bijzonder voorzichtig te werk in het apicaal gebied en in kanalen die zich splitsen, en/of abrupte krommingen of nieuwe krommingen vertonen.
● Spoel het kanaal overvloedig en regelmatig tijdens de behandeling en na elk gebruikt instrument (volgens goede tandheelkundige praktijken).
● Oefen te allen tijde minimale apicale druk uit. De vijlen mogen nooit met kracht in het wortelkanaal worden ingebracht.
● Wanneer het instrument niet gemakkelijk vordert, reinigt en inspecteert u de snijkanten, spoelt u, herhaalt u met een handmatige vijl en spoelt u opnieuw.
● Bij het vormgeven van extreem gebogen kanalen is het veiliger om de vijl te gebruiken voor het vormgeven van slechts één kanaal, om zo het risico op breuk te verminderen. Neem het volgende in acht:
○ Gebruik een nieuwe vijl en gooi deze weg nadat het kanaal is behandeld (één vijl per kanaal).
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 41/151
○ Gebruik handvijlen in plaats van roterende vijlen.
○ Gebruik kleine, flexibele en/of NiTi-vijlen (dit helpt bij het voorkomen van kanaaltransport).
○ Controleer tijdens het gebruik van het product het werkende deel visueel op alle bij het vorige punt genoemde defecten (d.w.z. na elke in- en uitgaande beweging).
○ Vermijd de standaard ruimende, continu draaiende beweging en maak in plaats daarvan kleine hoekbewegingen (vijlbeweging, roterende oscillerende beweging of 'balanced force'-methode) om de roterende buigvermoeidheid van de instrumenten te beperken en hun verwachte levensduur te verbeteren.
In de huidige technische staat zijn er tot dusver geen ongewenste reacties gemeld.
Houd producten in een droge en schone omgeving en uit de buurt van vocht en direct zonlicht.
Alle ProTaper Ultimate™ vijlen zijn gevalideerd voor gebruik op een 400 in toeren per minuut en
4 to 5.2 Ncm torsie.
De ProTaper Ultimate™ instrumenten kunnen worden gebruikt met een naar buiten gerichte borstelbeweging in alle kanalen, vooral die kanalen met een onregelmatige dwarsdoorsnede, of met een lichte naar binnen gerichte op- en neerwaartse beweging om progressief naar de werklengte te gaan.
Wieg het handstuk altijd in de ruimte tussen duim en wijsvinger. Vermijd duwen; laat de vijlen liever passief vorderen en volg het glijpad.
1) Bestudeer verschillende horizontaal georiënteerde röntgenopnamen teneinde diagnostisch de breedte, lengte en kromming te bepalen van een gegeven wortel en kana(a)l(en).
2) Bereid een toegang voor die een gemakkelijke identificatie van elke kanaalopening mogelijk maakt.
3) De hulpvormgevingsvijl, SX, kan worden gebruikt wanneer er beperkte ruimte is, of om de ruimte van een kanaal vooraf te vergroten, of om het meest coronale deel van een kanaal van een externe wortelconcaviteit vandaan te plaatsen.
4) Selecteer, met gebruikmaking van een glijmiddel, de Slider en volg het kanaal PASSIEF, in een of meer bewegingen, tot het eindpunt. Bepaal de werklengte met behulp van een elektronische apexlocator in combinatie met een radiografisch beeld en bevestig vervolgens de doorlatendheid.
5) Als de Slider het eindpunt van het kanaal niet gemakkelijk kan bereiken, kiest u een kleine handmatige roestvrij stalen (RVS) handvijl. Katheteriseer met gebruikmaking van een glijmiddel dit kanaal, bepaal de werklengte, controleer de doorlatendheid en verifieer het glijpad. Herhaal nu stap #4 hierboven.
6) Selecteermet gebruikmaking van NaOCl de Shaper en ga verder langs het glijpad, in een of meer bewegingen, totdat de werklengte is bereikt.
7) Na het verwijderen van de Shaper spoelt u en gebruikt u de Endo Activator om resten los te breken en in de oplossing te brengen, vervolgens spoelt u opnieuw om de resten te verwijderen.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 42/151
8) Herbevestig de werklengte, vooral in meer gebogen kanalen.
9) Selecteer de FINISHER F1 (020.07) en volg passief het kanaal naar de werklengte, in één of meerdere bewegingen. Verwijder en inspecteer de apicale windingen. Indien deze vol zitten met dentale resten, is de voorbereiding voltooid.
10) Als de FINISHER F1 in de lengterichting los zit en de apicale windingen niet vol resten zitten, selecteer dan de FINISHER F2 (025.08) en gebruik deze vijl op dezelfde manier als hierboven beschreven voor de FINISHER F1.
11) Als de FINISHER F2 in de lengterichting los zit en de apicale windingen niet vol resten zitten, selecteer dan de FINISHER F3 (030.09) en gebruik deze vijl op dezelfde manier als hierboven beschreven voor de FINISHER F1 en F2.
12) Na het verwijderen van een bepaalde vijl reinigt en inspecteert u de windingen, spoelt u, recapituleert u met een vijl van maat 10 of Endo Activator om vuil af te breken, en spoelt u vervolgens opnieuw.
13) Inspecteer de windingen van de vijl bij verwijdering op de aanwezigheid van beschadigingen, vervorming of uitrekking. Gooi de vijl weg als vervormingen worden geconstateerd en gebruik een nieuwe ProTaper Ultimate™ vijl.
14) De voorbereiding is voltooid wanneer de apicale omvang van een Finisher vol zit met resten en de overeenkomstige maat GPMC of Size Verifier op de werklengte past.
Gebruik de hulp Finishers alleen in grotere en rechtere kanalen, zoals maxillaire centrale incisieven, bepaalde palatinale of distale kanalen van kiezen, of wanneer er een pathologisch of iatrogeen gebrek is:
15) Selecteer een mechanisch aangedreven of handmatige hulpvijl, FINISHER FX (035.12), wanneer de werklengte is vastgesteld en de doorlatendheid is bevestigd. Gebruik deze op zelfstandige vijl direct en passief om het kanaal naar de werklengte te volgen met dezelfde techniek als hierboven beschreven. Verwijder en inspecteer de apicale windingen. Indien deze vol zitten met dentale resten, is de voorbereiding voltooid.
16) Als de FINISHER FX in de lengterichting los zit en de apicale windingen niet vol resten zitten, selecteer dan de een mechanisch aangedreven of handmatige hulpvijl FINISHER FXL
(050.10) en gebruik deze vijl op dezelfde manier als hierboven beschreven voor de FINIS-
HER FX.
17) De voorbereiding is voltooid wanneer de apicale omvang van een Finisher hulpvijl vol zit met resten en de overeenkomstige - maat GPMC of Size Verifier op de werklengte past.
● Volg de Algemene verwerkingsinstructies voor endodontische producten ( https:// www.dentsplysirona.com/en/service-contact/ifu.html
) voor de desinfectie-, reinigings- en sterilisatiestappen.
● Gebruik siliconen stoppen niet opnieuw. Verwijder en gooi siliconen stoppen na elk gebruik weg.
● De producten moeten worden afgevoerd volgens de plaatselijke voorschriften betreffende de veilige afvoer van scherpe en vervuilde materialen.
● Elk ernstig incident met betrekking tot het product moet worden gemeld aan de fabrikant en de bevoegde autoriteit in overeenstemming met de lokale regelgeving.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 43/151
● Steriliteit kan niet gegarandeerd worden als de verpakking open, beschadigd of vochtig is.
● Om een gratis gedrukt exemplaar van de IFU te krijgen, raadpleegt u het deel "Een gedrukt exemplaar van de IFU verkrijgen" op de website https://www.dentsplysirona.com/en/ service-contact/ifu.html
.
● Betekenis van niet-geharmoniseerde symbolen voor gebruiksaanwijzingen en labels, zie
IFU-symbolen ( https://www.dentsplysirona.com/en/service-contact/ifu/ifu-symbols.html
).
2797
Geautoriseerd vertegenwoordiger
DENTSPLY DeTrey GmbH
De-Trey-Straße 1
D-78467 Konstanz
Germany
Fabrikant
Maillefer Instruments Holding Sàrl
Chemin du Verger 3
CH-1338 Ballaigues
Suisse www.dentsplysirona.com/ifu
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 44/151
ProTaper Ultimate™
ENDAST FÖR DENTAL ANVÄNDNING
ENDAST FÖR PROFESSIONELL ANVÄNDNING
STERIL – STERILISERAD GENOM STRÅLNING
KAN ÅTERANVÄNDAS
FLERGÅNGS STERILA ENDODONTISKA INSTRUMENT
● Produkterna är avsedda för behandling av endodontiska sjukdomar.
● Avsedd användning: Mekaniskt drivet instrument avsedd för förberedelse av rotkanalen
(glidväg, formning och debridering av rotkanalen).
● Endodontiska instrument får endast användas i kliniska eller sjukhusmiljöer, enligt beprövad dental rutin, av kvalificerade tandvårdspecialister såsom både tandläkare och endodontologer och tandvårdsassistenter.
● Instrumenten ska användas i kombination med en endodontisk mekanisk drivning (manuellt i svåra kurvaturer).
Inga.
ProTaper Ultimate™ instrument består av tre huvudkomponenter: arbetsdelen i en nickel-titaniumlegering, ett färgat silikonstopp (undantag för instrument SX) och ett pläterat mässingskaft med en färgad ring (undantag för instrument SX).
Filerna tillverkas i en process som resulterar i en fil som har ett färgat utseende.
På grund av denna särskilda bearbetning, kan filerna se något böjda ut. Detta är inte ett fabrikationsfel. Trots att det är lätt att räta ut filen med bara fingrarna, är det inte nödvändigt att räta ut filen före användning. När den är inuti kanalen, följer filen rotkanalens anatomi.
ProTaper Ultimate™ sortimentet omfattar de följande endodontiska instrument:
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 45/151
Kod xx-längd
Y-fil storlek
Z-antal per blister
Produkt
ISO XXX YYYv
XXX-ISO storlek
YYY-avsmalnande storlek
V-variabel avsmalning
Tillgängliga längder
Förpackning och innehåll
B ST PULR z
19 0SX
B ST PULR z xx 0SL
B ST PULR z xx 0SH
ProTaper Ultimate™
SX
ProTaper Ultimate™
SLIDER
ProTaper Ultimate™
SHAPER
ISO 020.003v
ISO 016.002v
ISO 020.004v
19 mm
21/25/31 mm
21/25/31 mm
Blister med 6 instrument (steril)
Blister med 3 instrument (steril)
Blister med 6 instrument (steril)
Blister med 3 instrument (steril)
Blister med 6 instrument (steril)
Blister med 3 instrument (steril)
Blister med 6 instrument (steril)
B ST PULR z xx 0F y
B ST PULR 6 xx yyy
B ST PULR 5 xx SQE
B DD PULR
525 SQE
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™ sekvens
ProTaper Ultimate™
Demopack
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
FX: ISO 035.012v
FXL: ISO 050.010v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
21/25/31 mm
21/25/31 mm
21/25/31 mm
25 mm
Blister med 3 instrument (steril)
Blister med 6 instrument (steril)
Blister med 1 instrument av varje (steril)
Blister med 1 instrument av varje (steril)
● Vridmomentkontroll säkerställer optimal användning.
● Används med endodontiska mekaniska drivningar med en konstant rotation på 400 vpm.
● Ställ vridmomentet in på: 4 to 5.2 Ncm.
● Smörjning såsom NaOCl, EDTA, ProLube, Glyde™ ska användas.
● Användning av röntgenbilder i kombination med en APEX-lokaliserare och ett verktyg för att justera silikonstoppet till den korrekta arbetslängden är den lämpliga metoden för att bestämma arbetslängden.
● Bruksanvisningarna och de allmänna hanteringsanvisningarna för endodontiska produkter
(se avsnitt 8) skall följas exakt för att minimera riskerna för patienten och/eller brukaren samt apparaten:
○ brott på instrumenten.
○ korskontaminering.
○ uppkomst av värme på grund av otillräcklig smörjning och spolning.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 46/151
○ att svälja arbetsdelen av instrumentet.
○ toxiska eller allergiska reaktioner förorsakade av bearbetningsrester.
● Som alla andra mekaniskt drivna rotkanalsinstrument ska ProTaper Ultimate™ instrument inte användas i rotkanaler med kraftig och abrupt apikal kurvatur p.g.a. av hög risk för separation. I dessa fall bör böjda handfiler användas i det apikala området.
● Säkerheten och effektiviteten för gravida kvinnor och barn har inte fastställts.
● Använd en gummiduk under den endodontiska proceduren.
● För din egen säkerhet, bär personlig skyddsutrustning (handskar, skyddsglasögon, visir).
● Inspektera förpackningen före användning och använd inte instrumenten om förpackningen
är skadad.
● Använd inte instrument med utgånget bäst-före-datum.
● Inspektera instrumentet före varje användning på tecken för defekter såsom deformationer
(böjd, avrullad), brott, korrosion, skadade skärytor, lossnande färgkodning eller märkning.
Vid dessa indikationer är anordningen inte lämplig för den avsedda användningen vid den erforderliga säkerhetsnivån, instrumenten bör kasseras.
● Innan du använder något instrument, se till att det är väl anslutet till motvinkelhuvudet.
● Inspektera och rengör arbetsdelen ofta under instrumentering, kontrollera för tecken på snedvridning, töjning eller slitage, såsom ojämna ytor, matta fläckar. Vid dessa indikationer
är anordningen inte lämplig för den avsedda användningen vid den erforderliga säkerhetsnivån, instrumenten bör kasseras.
● Instrumenten ska inte sänkas ner helt i en lösning av natriumhypoklorit (NaOCl). Endast arbetsdelen av NiTi-instrumentet, som är i kontakt med patienten ska vara nedsänkt i en Na-
OCl-lösning koncentrat på högst 5 %.
● Var försiktig i apikala områden och i kanaler som delar sig och/eller uppvisar skarpa krökningar.
● Spola kanalen ofta och rikligt under arbetet och efter varje insatt instrument (enligt god dental rutin).
● Använd alltid minsta möjliga apikalt tryck. Tryck aldrig ner filerna i kanalen.
● När ett instrument inte lätt går fram, ska instrumentet rengöras och inspekteras, sedan sköljas, avsluta med en manuell fil och spola sedan igen.
● För att forma extremt böjda kanaler är det säkrare att använda filen bara för att forma en kanal så att man minskar risken för brott. Var uppmärksam på följande goda praxis:
○ Använd en ny fil och släng den efter att ha kanalen har behandlats (användning enbart till en kanal).
○ Använd manuella i stället för roterande filar.
○ Använd små, flexibla eller/och NiTi-filar (detta hjälper till att undvika kanaltransport).
○ Inspektera arbetsdelen visuellt med tanke på alla defekter som listas i föregående stycke under användningen (d.v.s. efter varje omgång).
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 47/151
○ Undvik standardrullning med kontinuerlig rotationsrörelse och använd istället små vinkelrörelser (filningsrörelse, den svängande rörelsen för att dra upp en klocka eller en teknik med balanserad kraft) för att begränsa rotationsböjningsutmattning på instrumentet samt
öka deras förväntade livslängd.
I nuvarande tekniska status, har ingen biverkning rapporterats hittills.
Produkterna lagras i torr, ren miljö, skyddade mot fuktkällor och direkt solljus.
Alla ProTaper Ultimate™ filer har utvärderats för användning vid 400 vpm och 4 to 5.2 Ncm vridmoment.
ProTaper Ultimate™ instrument kan användas i alla kanaler med en utåt gående borstande rörelse, i synnerhet i de kanalerna som uppvisar ett oregelbundet tvärsnitt, eller med en lätt inåt riktade rörelse för att successivt närma sig arbetslängden (working length - WL).
Håll alltid handtaget i bandet mellan tummen och pekfingret. Tryck inte på, låt filerna mata sig fram passivt och följa glidvägen.
1) Granska olika röntgenbilder tagna med olika horisontella vinklar för att fastställa den aktuella rotens och dess kanalers bredd, längd och kurvatur.
2) Förbered en ingång som ger en enkel identifikation av varje kanalutrymme.
3) Tillsats-Shaper-filen, SX, kan användas när det är ont om plats eller för att utvidga en kanals kropp eller för att förskjuta den mest koronala delen av en kanal bort från en yttre rotkonkavitet.
4) Välj, om smörjning finns, en Slider och följ kanalen PASSIVT in en eller flera omgångar till den tar slut. Fastställ arbetslängden med hjälp av en APEX-lokaliserare i kombination med röntgenbilder och bekräfta sedan öppenheten.
5) Om Slider inte enkelt når kanalens slut, väljer du en liten, manuell handfil i rostfritt stål
(SS). Med smörjning ska kanalen kateteriseras, arbetslängden fastställas, öppenheten bekräftas och glidvägen verifieras. Upprepa sedan steg #4 ovan.
6) Med NaOCl väljer du en Shaper och glider längsmed glidvägen i en eller flera omgångar tills du når arbetslängden.
7) Skölj, när Shapern har dragits ut, EndoActivatorn för att få skräpet att lossna, spola sedan igen för att frigöra skräpet.
8) Bekräfta arbetslängden igen, i synnerhet i svängda kanaler.
9) Välj FINISHER F1 (020.07) och följ kanalen passivt till arbetslängden i en eller flera omgångar. Ta bort och inspektera dess apikala räfflor. När de är fyllda med tandskräp, är förberedelserna klara.
10) Om FINISHER F1 sitter löst på längden och dess apikala räfflor inte är fyllda med skräp, väljer du en FINISHER F2 (025.08) och använder den filen på samma sätt som i den ovanstående beskrivningen för FINISHER F1.
11) Om FINISHER F2 sitter löst på längden och dess apikala räfflor inte är fyllda med skräp, väljer du en FINISHER F3 (030.09) och använder den filen på samma sätt som i den ovanstående beskrivningen för FINISHER F1 och F2.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 48/151
12) När en fil tas ut, ska dess skärande räfflor rengöras och inspekteras, sköljas, bearbetas igen med antingen en storleks 10 fil eller EndoActivatorn för att få skräpet att lossna och sedan sköljas igen.
13) Inspektera filens skärande räfflor när filen tas ut för att upptäcka eventuella avlindningar, rätningar eller sträckningar. Om deformationer upptäcks, ska den kasseras och använd en ny ProTaper Ultimate™ fil ska användas.
14) Förberedelserna är avslutade när det apikala utrymmet på varje Finisher är fylld med skräp och en GPMS med motsvarande storlek eller en Size Verifier passar in i WL.
Tillsats-Finisher ska endast användas i större och rakare kanaler, såsom maxilara, centrala tänder, vissa palatala eller distala kanaler på molarer eller vid patologiska eller iatrogeniska defekter:
15) Välj antingen en mekaniskt driven eller en manuell tillsats-fil, FINISHER FX (035.12), när arbetslängden är fastställd och öppenheten har bekräftats. Använd denna fristående fil direkt och passivt med samma teknik som i beskrivningen ovan för att följa kanalen till WL.
Ta bort och inspektera dess apikala räfflor. När de är fyllda med tandskräp, är förberedelserna klara.
16) Om FINISHER FX sitter löst på längden och dess apikala räfflor inte är fyllda med skräp, väljer du antingen en mekaniskt driven eller en manuell tillsats-FINISHER FXL (050.10) och använder den filen på samma sätt som i den ovanstående beskrivningen för FINIS-
HER FX.
17) Förberedelserna är avslutade när det apikala utrymmet på varje tillsats-Finisher är fylld med skräp och en GPMS med motsvarande storlek eller en Size Verifier passar in i WL.
● Följ de allmänna råden för hantering av endodontiska produkter ( https:// www.dentsplysirona.com/en/service-contact/ifu.html
) för desinfektionen, rengöringen och steriliseringen.
● Återanvänd inte silikonstoppen. Silikonstoppen ska tas bort och slängas efter varje användning.
● Produkterna ska omhändertas enligt lokala föreskrifter för säker avfallshantering av vassa och kontaminerade ämnen.
● Alla allvarliga händelser i samband med produkten bör rapporteras till tillverkaren och de enligt de lokala föreskrifterna ansvariga myndigheter.
● Sterilitet kan inte garanteras om förpackningen är öppen, skadad eller våt.
● Du beställer en gratis tryckt version av IFU i avsnittet “Beställ en gratis utskrift av IFU” på hemsidan https://www.dentsplysirona.com/en/service-contact/ifu.html
.
● Förklaring av icke-harmoniserade symboler för IFU:er och etiketter, IFU ( https:// www.dentsplysirona.com/en/service-contact/ifu/ifu-symbols.html
).
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 49/151
2797
Auktoriserade återförsäljare
DENTSPLY DeTrey GmbH
De-Trey-Straße 1
D-78467 Konstanz
Germany
Tillverkare
Maillefer Instruments Holding Sàrl
Chemin du Verger 3
CH-1338 Ballaigues
Suisse www.dentsplysirona.com/ifu
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 50/151
ProTaper Ultimate™
KUN TIL ODONTOLOGISK ANVENDELSE
KUN TIL PROFESSIONEL ANVENDELSE
STERIL – STERILISERET VED BESTRÅLING
SKAL RENGØRES OG STERILISERES
GENANVENDELIGE STERILE RODBEHANDLINGSINSTRUMENTER
● Produkter er indiceret til rodbehandling.
● Tilsigtet anvendelse: Motordrevet instrument beregnet til præparering af rodkanalen (adgangsvej, shaping og debridement af rodkanalen).
● Rodbehandlingsinstrumenter må kun anvendes i et klinik- eller hospitalsmiljø, hvor god odontologisk praksis følges, af kvalificerede tandlæger, speciallæger i rodbehandling og klinikassistenter.
● Instrumenterne skal anvendes sammen med en rodbehandlingsmotor (manuelt i kraftige buer).
Ingen.
ProTaper Ultimate™-instrumenter er fremstillet af tre hovedkomponenter: Den arbejdende del, som er fremstillet af en nikkeltitaniumlegering, en farvet silikoneprop (med undtagelse af SX-instrumentet) og et pletteret skaft i messing med en farvet ring (med undtagelse af SX-instrumentet)
Filene er fremstillet ved hjælp af en proces, der giver filen et farvet udseende.
På grund af denne varemærkebeskyttede behandling kan filene være let buede. Dette er ikke en fejl i fremstillingen. Filen kan nemt rettes ud med fingrene, men det er ikke nødvendigt at rette den ud, før den bruges. Når filen er inde i rodkanalen, følger den anatomien.
ProTaper Ultimate™-serien består af følgende rodbehandlingsinstrumenter:
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 51/151
Kode xx-længde y-filstørrelse z-antal pr. blisterpakning
Produkt
ISO XXX YYYv
XXX-ISO-størrelse
YYY-konusstørrelse v-variabel konus
Tilgængelige længder
Emballage og indhold
B ST PULR z
19 0SX
B ST PULR z xx 0SL
B ST PULR z xx 0SH
B ST PULR z xx 0F y
B ST PULR 6 xx yyy
B ST PULR 5 xx SQE
B DD PULR
525 SQE
ProTaper Ultimate™
SX
ProTaper Ultimate™
SLIDER
ProTaper Ultimate™
SHAPER
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
Sequence
ProTaper Ultimate™
Demo Pack
ISO 020.003v
ISO 016.002v
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
FX: ISO 035.012v
FXL: ISO 050.010v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
19 mm
21/25/31 mm
21/25/31 mm
21/25/31 mm
21/25/31 mm
21/25/31 mm
25 mm
Blisterpakning med 6 instrumenter (sterile)
Blisterpakning med 3 instrumenter (sterile)
Blisterpakning med 6 instrumenter (sterile)
Blisterpakning med 3 instrumenter (sterile)
Blisterpakning med 6 instrumenter (sterile)
Blisterpakning med 3 instrumenter (sterile)
Blisterpakning med 6 instrumenter (sterile)
Blisterpakning med 3 instrumenter (sterile)
Blisterpakning med 6 instrumenter (sterile)
Blisterpakning med 1 stk. af hvert instrument (steril)
Blisterpakning med 1 stk. af hvert instrument (steril)
● Momentstyringsanordninger sikrer optimal brug.
● Brug med rodbehandlingsmotorer i en konstant rotation ved en hastighed på 400 rpm.
● Indstil moment til: 4 to 5.2 Ncm.
● Der skal anvendes smøremidler som NaOCl, EDTA, ProLube, Glyde™.
● Brugen af røntgenbilleder sammen med en tandrodssøger og et værktøj til justering af silikoneproppen til den korrekte arbejdslængde er den foretrukne metode til bestemmelse af arbejdslængden.
● Følg denne brugsanvisning og den generelle vejledning i rengøring og sterilisering af rodbehandlingsinstrumenter til punkt og prikke (se afsnit 8) for at minimere følgende risici for udstyret, patienten og/eller brugeren:
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 52/151
○ Brud på instrumentet.
○ Krydsinfektion.
○ Varmedannelse pga. utilstrækkelig smøring og skylning.
○ Indtagelse af den arbejdende del af instrumentet.
○ Giftvirkninger eller allergiske reaktioner forårsaget af restprodukter efter rengøring og sterilisering.
● Som med alle mekanisk drevne rodkanalinstrumenter bør ProTaper Ultimate™-instrumenter ikke anvendes i en rodkanal med meget kraftigt buet rodspids, da dette medfører en
øget risiko for separation. I dette tilfælde bør der anvendes forbuede håndfiler i rodspidsområdet.
● Sikkerheden og effektiviteten er ikke blevet afprøvet på gravide og ammende kvinder samt børn.
● Anvend et kofferdamsystem under rodbehandlingen.
● Af hensyn til din sikkerhed skal du bære personlige værnemidler (handsker, beskyttelsesbriller, maske).
● Emballagen skal kontrolleres før brug, og instrumenterne må ikke anvendes, hvis emballagen er beskadiget.
● Instrumenterne må ikke anvendes efter udløbsdatoen.
● Kontrollér instrumentet før hver ibrugtagning for tegn på defekter som deformationer (bøjning, afhaspning), brud, korrosion, beskadigede skærekanter, manglende farvekodning eller mærkning. Med disse indikationer for, at anordningerne ikke kan opfylde den tilsigtede anvendelse med det påkrævede sikkerhedsniveau, skal instrumenterne kasseres.
● Før ethvert instrument tages i brug, skal det kontrolleres, er det er korrekt monteret på vinkelstykkehovedet.
● Efterse instrumentet og rengør den arbejdende del hyppigt under brug af instrumenterne, og kontrollér for tegn på forvridning, forlængelse eller slitage, som uregelmæssige skær og sløve punkter. Med disse indikationer for, at anordningerne ikke kan opfylde den tilsigtede anvendelse med det påkrævede sikkerhedsniveau, skal instrumenterne kasseres.
● Instrumenterne må ikke nedsænkes i en opløsning af natriumhypochlorit (NaOCl). Kun den arbejdende del af NiTi-instrumentet, som er i kontakt med patienten, må nedsænkes i en
NaOCl-opløsning på HØJST 5 %.
● Udvis forsigtighed i rodspidsområdet og ved behandling af delte og/eller kraftigt buede rodkanaler.
● Skyl rodkanalen grundigt og ofte under hele proceduren og efter hvert anvendt instrument
(iht. god odontologisk praksis).
● Der skal altid påføres et minimalt rodspidstryk. Filene må ikke tvinges ned i rodkanalen.
● Skærene skal renses og efterses, hvis instrumentet ikke er nemt at fremføre, hvorefter det skal skylles og slibes med en manuel fil og skylles igen.
● Til præparering af meget buede rodkanaler er det mere sikkert kun at bruge filen til at præparere én rodkanal, så risikoen for brud reduceres. Overhold følgende retningslinjer for god praksis:
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 53/151
○ Brug en ny fil, og kassér den efter behandlingen af rodkanalen (må kun bruges til én rodkanal).
○ Brug manuelle file i stedet for roterende file.
○ Brug små eller fleksible file og/eller NiTi-file (bidrager til at undgå rodkanaltransport).
○ Kontrollér den arbejdende del visuelt for alle de defekter, der er angivet i ovenstående afsnit, under brugen (dvs. efter hver frem- og tilbagebevægelse).
○ Undgå den kontinuerlige bevægelse ved almindelig udvidelse og brug i stedet små vinklede bevægelser (filebevægelse, oscillerende uroptrækningsbevægelse eller balanceret trykkende teknik) for at begrænse rotationsbøjningstræthed i instrumenterne og forlænge deres forventede levetid.
I den nuværende tekniske tilstand er der indtil nu ikke rapporteret nogen bivirkninger.
Produkterne skal opbevares i et tørt og rent miljø, hvor de ikke udsættes for fugtkilder og direkte sollys.
Alle ProTaper Ultimate™-filer er godkendt til brug ved 400 rpm og 4 to 5.2 Ncm moment.
ProTaper Ultimate™-instrumenterbe kan bruges med en udadgående børstende bevægelse i alle rodkanaler, især de rodkanaler, der udviser et uregelmæssigt tværsnit, eller med en let indadgående hakkebevægelse for gradvist at komme frem mod arbejdslængden.
Hold altid håndstykket i båndet mellem tommelfingeren og pegefingeren. Undgå tvang, før filerne passivt fremad, og følge adgangsvejen.
1) Anvend forskellige horisontalt vinklede røntgenbilleder til at foretage en diagnostisk vurdering af bredde, længde og kurvatur af de(n) behandlingskrævende rod og rodkanal(er).
2) Præparér adgangsstedet, hvilket giver mulighed for nem identifikation af hver rodkanalåbning.
3) SX-præpareringsfilen (tilbehør) kan anvendes, når der er begrænset plads, eller til at forstørre rodkanalrummet, eller til at flytte den koronale del af rodkanalen væk fra en ekstern rodkonkavitet.
4) Sørg for påføring af et smøremiddel, og brug slideren til at følge rodkanalen PASSIVT i et eller flere strøg, indtil den når rodkanalens endepunkt. Fastslå arbejdslængden ved hjælp af en elektronisk tandrodssøger kombineret med et røntgenbillede, og bekræft derefter synligheden.
5) Hvis slideren ikke har nemt ved at nå rodkanalens endepunkt, skal du vælge en lille, manuel håndfil i rustfrit stål. Sørg for påføring af et smøremiddel, og kateterisér denne rodkanal, fastslå arbejdslængden, bekræft synligheden og verificér adgangsvejen. Gentag derefter trin 4 ovenfor.
6) Sørg for påføring af NaOCI, og brug shaperen til at fremføre langs med adgangsvejen i et eller flere strøg, indtil arbejdslængden er nået.
7) Idet shaperen fjernes, skylles og endoaktiveres for at løsne debris og fjerne det i én bevægelse, hvorefter der skylles efter for at frigøre debriset.
8) Bekræft arbejdslængden på ny, især i meget buede rodkanaler.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 54/151
9) Brug FINISHER F1 (020.07) til at følge rodkanalen passivt frem til arbejdslængden i et eller flere strøg. Fjern og efterse rodkanalens rodspidsskær. Når de er fyldt med dentinmateriale, er præpareringen afsluttet.
10) Hvis FINISHER F1 er løs i længderetningen og filens rodspidsskær ikke er fyldt med debris, skal du vælge FINISHER F2 (025.08) og bruge denne fil på den samme måde som beskrevet ovenfor for FINISHER F1.
11) Hvis FINISHER F2 er løs i længderetningen og filens rodspidsskær ikke er fyldt med debris, skal du vælge FINISHER F3 (030.09) og bruge denne fil på den samme måde som beskrevet ovenfor for FINISHER F1 og F2.
12) Ved fjernelse af en vilkårlig fil skal dens skær renses og efterses, skylles og slibes med enten en størrelse 10 fil eller endoaktivator for at fjerne debris, hvorefter den skal skylles efter.
13) Efterse filens skær ved fjernelse for tilstedeværelse af afhaspning, udretning eller strækning. Hvis deformation konstateres, skal filen kasseres, og der anvendes en ny ProTaper
Ultimate™-fil.
14) Præpareringen er fuldført, når rodspidsen på en vilkårlig finisher er fyldt med debris og GP-
MC'en eller størrelsesbekræftelsesanordningen i en tilsvarende størrelse passer med arbejdslængden.
Brug kun de ekstra finishere i større eller mere regelrette rodkanaler som maksillære centrale fortænder, visse palatale eller distale rodkanaler i molarer, eller når der er en patologisk eller iatrogen defekt:
15) Vælg enten en mekanisk drevet eller manuel hjælpefil FINISHER FX (035.12), når arbejdslængden er etableret og synligheden er bekræftet. Brug denne særskilte fil direkte og passivt til at følge rodkanalen til arbejdslængden ved hjælp af den samme teknik som beskrevet ovenfor. Fjern og efterse rodkanalens rodspidsskær. Når de er fyldt med dentinmateriale, er præpareringen afsluttet.
16) Hvis FINISHER FX er løs i længderetningen og filens rodspidsskær ikke er fyldt med debris, skal du vælge enten en mekanisk drevet eller manuel hjælpefil FINISHER FXL
(050.10) og bruge denne fil på den samme måde som beskrevet ovenfor for FINISHER
FX.
17) Præpareringen er fuldført, når rodspidsen på en vilkårlig ekstra finisher er fyldt med debris og GPMC'en eller størrelsesbekræftelsesanordningen i en tilsvarende størrelse passer med arbejdslængden.
● Følg den generelle vejledning i rengøring og sterilisering af rodbehandlingsinstrumenter
( https://www.dentsplysirona.com/en/service-contact/ifu.html
) for anvisninger i desinfektion, rengøring og sterilisering.
● Silikonestop må ikke genanvendes. Fjern og kassér silikonestoppene efter hver brug.
● Produkterne skal bortskaffes overensstemmelse med de lokale bestemmelser for sikker bortskaffelse af skarpe og kontaminerede enheder.
● Enhver alvorlig hændelse relateret til brug af produktet skal indberettes til producenten og den kompetente myndighed iht. lokale bestemmelser.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 55/151
● Steriliteten kan ikke garanteres, hvis emballagen er åben, beskadiget eller våd.
● Se afsnittet “Modtag en papirkopi af brugsanvisningen” på webstedet https:// www.dentsplysirona.com/en/service-contact/ifu.html
for at modtage en gratis papirkopi af brugsanvisningen.
● Se afsnittet "Symboler i brugsanvisninger" på webstedet ( https://www.dentsplysirona.com/ en/service-contact/ifu/ifu-symbols.html
) for en beskrivelse af ikke-harmoniserede symboler i brugsanvisninger og på mærkning.
2797
Autoriseret repræsentant
DENTSPLY DeTrey GmbH
De-Trey-Straße 1
D-78467 Konstanz
Germany
Producent
Maillefer Instruments Holding Sàrl
Chemin du Verger 3
CH-1338 Ballaigues
Suisse www.dentsplysirona.com/ifu
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 56/151
ProTaper Ultimate™
BARE FOR BRUK TIL TANNBEHANDLING
KUN TIL PROFESJONELL BRUK
STERIL – STERILISERT MED STRÅLING
ANSVARLIG FOR REPROSESSERING
GJENBRUKBARE STERILE ENDODONTISKE INSTRUMENTER
● Disse produktene er indikert for behandling av endodontisk sykdom.
● Tiltenkt formål: Motordrevent instrument som er beregnet til rotkanalklargjøring (glidebane, forming og debridement av rotkanal).
● Endodontiske instrumenter skal kun brukes kun i et klinisk miljø eller sykehusmiljø med god tannlegepraksis, av kvalifiserte tannlegefagpersoner, slik som allmenntannleger og endospesialister (endodontist) og tannlegeassistenter.
● Instrumenter skal brukes i kombinasjon med en endodontisk motor (manuelt i alvorlige kurvaturer).
Ingen.
ProTaper Ultimate™ instrumenter er produsert i tre hovedkomponenter: arbeidsdelen er laget i nikkel-titan-legering, en farget silikonstopper (unntatt instrument SX) og et belagt messingskaft med en farget ring (unntatt instrument SX).
Filene er framstilt med en prosess som resulterer i en fil som har et farget utseende.
På grunn av denne egenutviklede prosesseringen kan filene virke lett buet. Dette er ikke en produksjonsdefekt. Selv om filen enkelt kan rettes ut ved hjelp av fingrene, er det ikke nødvendig å rette den ut før bruk. Inne i kanalen vil filen følge rotkanalanatomien.
ProTaper Ultimate™-utvalget omfatter følgende endodontiske instrumenter:
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 57/151
Kode xx-lengde y-filstørrelse z-antall per blister
B ST PULR z
19 0SX
B ST PULR z xx 0SL
B ST PULR z xx 0SH
Produkt
ProTaper Ultimate™
SX
ProTaper Ultimate™
SLIDER
ProTaper Ultimate™
SHAPER
ISO XXX YYYv
XXX-ISO-størrelse
YYY-konusstørrelse v-variabel konus
ISO 020.003v
ISO 016.002v
ISO 020.004v
Tilgjengelige lengder
19 mm
21/25/31 mm
21/25/31 mm
Pakning og innhold
Blister med 6 instrumenter (sterile)
Blister med 3 instrumenter (sterile)
Blister med 6 instrumenter (sterile)
Blister med 3 instrumenter (sterile)
Blister med 6 instrumenter (sterile)
Blister med 3 instrumenter (sterile)
Blister med 6 instrumenter (sterile)
B ST PULR z xx 0F y
B ST PULR 6 xx yyy
B ST PULR 5 xx SQE
B DD PULR
525 SQE
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
Sequence
ProTaper Ultimate™
Demo Pack
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
FX: ISO 035.012v
FXL: ISO 050.010v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
21/25/31 mm
21/25/31 mm
21/25/31 mm
25 mm
Blister med 3 instrumenter (sterile)
Blister med 6 instrumenter (sterile)
Blister med 1 instrument av hvert instrument (sterilt)
Blister med 1 instrument av hvert instrument (sterilt)
● Momentkontrollenheter sikrer optimal bruk.
● Brukes med endodontiske motorer i en konstant rotasjon ved en hastighet på 400 o/min.
● Still inn moment på: 4 to 5.2 Ncm.
● Smøremidler, slik som NaOCl, EDTA, ProLube, Glyde™ skal brukes.
● Bruken av radiografer i kombinasjon med en apekslokator og et verktøy for å justere silikonstopperen til riktig arbeidslengde er egnet metode for arbeidslengdebestemmelse.
● Følg denne bruksanvisningen og de generelle behandlingsanvisningene for endodontiske produkter (se avsnitt 8) for å minimere følgende risikoer for enheten, pasienten og/eller brukeren:
○ Instrumentbrudd.
○ Krysskontaminering.
○ Varmegenerering på grunn av utilstrekkelig smøring og skylling.
○ Svelging av virksom del av instrumentet.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 58/151
○ Toksiske eller allergiske reaksjoner forårsaket av behandlingsrester.
● Lignende med alle maskindrevne rotkanalinstrumenter skal ProTaper Ultimate™-instrumenter ikke brukes i en rotkanal med en brå apikal kurvatur på grunn av den økte faren for separasjon. I dette tilfellet skal forhåndskurvede håndfiler brukes i den apikale regionen.
● Sikkerhet og effektivitet ved bruk er ikke fastslått hos gravide eller ammende kvinner eller hos barn.
● Bruk et kofferdamsystem under den endodontiske prosedyren.
● For din egen sikkerhet må du bruke personlig verneutstyr (hansker, briller, maske).
● Inspiser forpakningen før bruk og ikke bruk instrumentene hvis forpakningen er skadet.
● Ikke bruk instrumentene etter utløpsdatoen.
● Kontroller instrumentet før hver bruk for tegn på defekter som deformeringer (bøyninger, ikke avviklet), brudd, korrosjon, skadede kuttekanter, tap av fargekoding eller merking. Med disse indikasjonene er enhetene ikke i stand til å oppfylle tiltenkt bruk med nødvendig sikkerhetsnivå, instrumentene skal kasseres.
● Før bruk av noe instrument må du kontrollere at det er godt festet til kontravinkelstykkehodet.
● Kontroller instrumentet og rengjør den virksomme delen hyppig under instrumentering, og se etter om det er tegn på skjevhet, forlengelse eller slitasje, som ujevne renner, matte flekker. Med disse indikasjonene er enhetene ikke i stand til å oppfylle tiltenkt bruk med nødvendig sikkerhetsnivå, instrumentene skal kasseres.
● Instrumentene må ikke senkes helt i en natriumhypoklorittløsning (NaOCl). Bare den virksomme delen av NiTi-instrumentet, som er i kontakt med pasienten, skal senkes ned i en
NaOCl-løsning med en konsentrasjon på MAKS. 5 %.
● Utvis varsomhet i det apikale området og i kanaler som deler seg og/eller har bratte kurvaturer eller tilbakekurvaturer.
● Skyll kanalen grundig og hyppig i løpet av prosedyren og etter hvert instrument er brukt (i henhold til god tannlegepraksis).
● Bruk minimalt apikalt trykk. Press aldri filene ned kanalen.
● Når instrumentet ikke går enkelt videre, rengjør og inspiser skjærefurene, irriger, rekapituler med en manuell fil og irriger på nytt.
● Ved forming av ekstremt buede kanaler er det tryggest å bare bruke filen til å forme én kanal, for å redusere faren for brudd. Følgende gode praksis anbefales:
○ Bruk en ny fil og kast den etter at kanalen er behandlet (én kanals bruk).
○ Bruk manuelle i stedet for roterende filer.
○ Bruk små, fleksible filer og/eller NiTi-filer (dette bidrar til å unngå kanaltransport).
○ Inspiser den virksomme delen visuelt under bruk for alle defektene i listen i forrige avsnitt
(dvs. etter hver kurve).
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 59/151
○ Unngå standard kontinuerlig roterende borebevegelse, og bruk i stedet små vinkelbevegelser (filebevegelse, oscillerende klokkeopptrekkingsbevegelse eller balansert kraftteknikk) for å begrense rotasjonsbøyingstrettheten på instrumentene og forlenge forventet levetid.
I den nåværende tekniske tilstanden er det ikke rapportert noen bivirkninger.
Oppbevar produkter i tørt og rent miljø, atskilt fra fuktkilder og direkte sollys.
Alle ProTaper Ultimate™ -filer har blitt validert til bruk ved 400 o/min og4 to 5.2 Ncm moment.
ProTaper Ultimate™instrumenter kan brukes med en utovergående børste i alle kanaler, spesielt de kanalene som fremviser et uregelmessig tverrsnitt, eller med en lett innovervendt hakkebevegelse for å progressivt føre frem mot arbeidslengden (WL).
Legg alltid inn håndstykket inn i nettingen mellom tommelen og pekefingeren. Unngå å skyve; la heller filene føres passivt frem og følge glidebanen.
1) Se over forskjellige horisontalt vinklede røntgenbilder for å diagnostisk bestemme bredde, lengde og kurvatur til en gitt rot og kanal(er).
2) Klargjør en tilgang som gjør det mulig med enkel identifisering av hver kanalåpning.
3) Hjelpeformfilen, SX, kan brukes når det er begrenset plass, eller til å forhåndsforstørre kroppen til en kanal eller å bevisst forskyve det mest koronale aspektet av kanalen bort fra en utvendig rotkonkavitet.
4) I nærvær av et smøremiddel, velg skyver og følg kanalen PASSIVT, i ett eller flere pass, til terminus. Bestem WL ved bruk av en elektronisk apekslokator kombinert med et røntgenbilde, bekreft deretter åpning.
5) Hvis Slider ikke enkelt når kanalterminus, velg en liten manuell håndfil i rustfritt stål (SS).
Ved tilstedeværelse av smøremiddel kateteriser denne kanalen, etabler arbeidslengde, bekreft åpningen og verifiser glidebanen. Gjenta nå trinn nr. 4 ovenfor.
6) I nærvær av NaOCl, velg Shaper og før frem langs glidebanen, i ett eller flere pass, inntil
WL nås.
7) Ved fjerning av Shaper, skyll, EndoActivate for å bryte opp avfall og flytte det inn i løsningen, skyll på nytt for å frigi dette avfallet.
8) Kontroller arbeidslengden på nytt, spesielt i mer kurvede kanaler.
9) Velg FINISHER F1 (020.07) og følg kanalen passivt til WL, i ett eller flere pass. Fjern og inspiser apikale furer. Hvis de er fulle av dentinspon, er klargjøringen ferdig.
10) Hvis FINISHER F1 er løs i lengden og apikale furer ikke er lastet med avfall, velg
FINISHER F2 (025.08) og bruk denne filen på samme måte som beskrevet ovenfor for
FINISHER F1.
11) Hvis FINISHER F2 er løs i lengden og apikale furer ikke er lastet med avfall, velg
FINISHER F3 (030.09) og bruk denne filen på samme måte som beskrevet ovenfor for
FINISHER F1 og F2.
12) Ved fjerning av enhver gitt fil, rengjør og inspiser skjærefurene, irriger, rekapituler med enten en fil i størrelse 10 eller Endo Activator for å bryte opp avfall, og irriger på nytt.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 60/151
13) Inspiser filens skjærefurer ved fjerning for nærvær av utstrekking, utretting eller strekking.
Hvis du oppdager deformasjon, må du kaste filen og bruke en ny ProTaper Ultimate™-fil.
14) Klargjøringen er fullført når den apikale utstrekningen av enhver Finisher og GPMC i tilsvarende størrelse eller Size Verifier passer ved WL.
Bruk ekstra Finishers kun i større og rettere kanaler, slik som maksillære sentrale innsnittsenheter, noen palatale eller distale molære kanaler eller når det er en patologisk eller iatrogen defekt:
15) Velg enten en mekanisk dreven eller den manuelle hjelpefilen, FINISHER FX (035.12) når arbeidslengden er etablert og patenten er bekreftet. Bruk denne enkeltstående filen direkte og passivt for å følge kanalen til arbeidslengden ved bruk av samme teknikk som beskrevet ovenfor. Fjern og inspiser apikale furer. Hvis de er fulle av dentinspon, er klargjøringen ferdig.
16) Hvis FINISHER FX er løs i lengden og apikale furer ikke er lastet med avfall, velg enten en mekanisk dreven eller manuell hjelpe-FINISHER FXL (050.10) og bruk denne filen på samme måte som beskrevet ovenfor for FINISHER FX.
17) Klargjøringen er fullført når den apikale utstrekningen av enhver ekstra Finisher og GPMC i tilsvarende størrelse eller Size Verifier passer ved WL.
● Følg de generelle behandlingsinstruksjonene for endodontiske produkter ( https:// www.dentsplysirona.com/en/service-contact/ifu.html
) for desinfiserings-, rengjørings- og steriliseringstrinn.
● Silikonplugger må ikke brukes flere ganger. Fjern og kast og silikonstopperne etter hver bruk.
● Produkter skal avhendes i samsvar med lokale forskrifter for sikker avhending av skarpe og kontaminerte enheter.
● Enhver alvorlig hendelse i forhold til produktet skal rapporteres til produsenten og den kompetente myndigheten etter lokale forskrifter.
● Sterilitet kan ikke garanteres dersom forpakningen er åpnet, skadd eller våt.
● For å få en gratis papirkopi av bruksanvisningen se avsnittet "Få en papirkopi av bruksanvisningen" på nettsidene https://www.dentsplysirona.com/en/service-contact/ifu.html
.
● Forklaring av ikke-harmoniserte symboler for bruksanvisninger og etiketter, se bruksanvisningssymboler ( https://www.dentsplysirona.com/en/service-contact/ifu/ifu-symbols.html
).
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 61/151
2797
Autorisert representant
DENTSPLY DeTrey GmbH
De-Trey-Straße 1
D-78467 Konstanz
Germany
Produsent
Maillefer Instruments Holding Sàrl
Chemin du Verger 3
CH-1338 Ballaigues
Suisse www.dentsplysirona.com/ifu
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 62/151
ProTaper Ultimate™
TYLKO DO UŻYTKU STOMATOLOGICZNEGO
TYLKO DO UŻYTKU PROFESJONALNEGO
PRODUKT JAŁOWY — WYSTERYLIZOWANY PROMIENIOWANIEM
PRODUKT PRZEZNACZONY DO DEKONTAMINACJI
WIELORAZOWE JAŁOWE INSTRUMENTY ENDODONTYCZNE
● Produkty są wskazane do stosowania w leczeniu chorób endodontycznych.
● Przeznaczenie: napędzany silnikiem instrument przeznaczony do opracowywania kanałów korzeniowych (tor wprowadzania, kształtowanie i oczyszczanie kanału korzeniowego).
● Instrumenty endodontyczne są przeznaczone do stosowania wyłącznie w środowisku klinicznym lub szpitalnym, zgodnie z dobrą praktyką stomatologiczną, przez wykwalifikowanych lekarzy stomatologów, takich jak stomatolodzy ogólni, a także lekarze specjalizujący się w endodoncji (endodonci) i asystentki stomatologiczne.
● Instrumenty należy stosować w połączeniu z silnikiem endodontycznym (ręcznie w przypadku dużych krzywizn).
Brak.
Instrumenty ProTaper Ultimate™ są złożone z trzech głównych elementów: części roboczej wykonanej ze stopu niklowo-tytanowego, ogranicznika wykonanego z kolorowego silikonu (z wyjątkiem instrumentu SX) oraz chwytu platerowanego mosiądzem z kolorowym pierścieniem (z wyjątkiem instrumentu SX).
Pilniki są produkowane w procesie, w wyniku którego powstaje kolorowy pilnik.
Ten zastrzeżony proces obróbki może sprawić, że pilniki mogą wydawać się lekko zakrzywione.
Nie jest to wada produkcyjna. Chociaż do wyprostowania pilnika wystarczy użyć palców, nie należy tego robić przed użyciem narzędzia. Po wprowadzeniu do kanału pilnik będzie dostosuje się do anatomii kanału korzeniowego.
Gama ProTaper Ultimate™ obejmuje następujące instrumenty endodontyczne:
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 63/151
Kod xx = długość y = rozmiar pilnika
Z = ilość w blistrze
Produkt
ISO XXX YYYv
XXX = rozmiar ISO
YYY = rozmiar stożka v = zmienny stożek
Dostępne długości
Opakowanie i zawartość
B ST PULR z
19 0SX
B ST PULR z xx 0SL
B ST PULR z xx 0SH
B ST PULR z xx 0F y
B ST PULR 6 xx yyy
B ST PULR 5 xx SQE
B DD PULR
525 SQE
ProTaper Ultimate™
SX
SLIDER ProTaper
Ultimate™
SHAPER ProTaper
Ultimate™
FINISHER ProTaper
Ultimate™
FINISHER ProTaper
Ultimate™
Sekwencja ProTaper
Ultimate™
Pakiet demonstracyjny ProTaper
Ultimate™
ISO 020.003v
ISO 016.002v
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
FX: ISO 035.012v
FXL: ISO 050.010v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
19 mm
21/25/31 mm
21/25/31 mm
21/25/31 mm
21/25/31 mm
21/25/31 mm
25 mm
Blister zawierający 6 instrumentów (sterylny)
Blister zawierający 3 instrumenty (sterylny)
Blister zawierający 6 instrumentów (sterylny)
Blister zawierający 3 instrumenty (sterylny)
Blister zawierający 6 instrumentów (sterylny)
Blister zawierający 3 instrumenty (sterylny)
Blister zawierający 6 instrumentów (sterylny)
Blister zawierający 3 instrumenty (sterylny)
Blister zawierający 6 instrumentów (sterylny)
Blister z jednym instrumentem każdego rodzaju (sterylny)
Blister z jednym instrumentem każdego rodzaju (sterylny)
● Urządzenia kontrolujące moment obrotowy zapewniają optymalne wykorzystanie.
● Stosować z silnikami endodontycznymi o stałej rotacji przy prędkości 400 obr./min.
● Ustawić moment obrotowy na: 4 to 5.2 Ncm.
● Należy stosować takie środki poślizgowe, jak NaOCl, EDTA, ProLube, Glyde™.
● Zastosowanie radiogramów w połączeniu z lokalizatorem wierzchołków i narzędziem do dopasowania silikonowego ogranicznika do właściwej długości roboczej jest właściwą metodą ustalania długości roboczej.
● Należy ściśle przestrzegać niniejszej instrukcji użycia oraz ogólnych instrukcji dotyczących przetwarzania produktów endodontycznych (patrz punkt 8) w celu zminimalizowania następujących zagrożeń dla urządzenia, pacjenta i/lub użytkownika:
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 64/151
○ Złamanie instrumentu.
○ Przeniesienie zakażenia.
○ Wytwarzanie ciepła z powodu niewystarczającego smarowania i irygacji.
○ Połknięcie części roboczej instrumentu.
○ Reakcje toksyczne lub alergiczne spowodowane pozostałościami procesowymi.
● Podobnie jak w przypadku wszystkich napędzanych silnikiem instrumentów do kanałów korzeniowych, instrumentów ProTaper Ultimate™ nie należy używać w kanałach korzeniowych z ostrą krzywizną dowierzchołkową ze względu na zwiększone ryzyko separacji. W takim przypadku w obszarze wierzchołkowym należy użyć wstępnie zakrzywionych pilników ręcznych.
● Nie określono bezpieczeństwa i skuteczności w grupie kobiet ciężarnych i karmiących piersią oraz dzieci.
● Podczas zabiegu endodontycznego należy stosować koferdam.
● Dla własnego bezpieczeństwa użytkownik powinien stosować środki ochrony indywidualnej
(rękawiczki, okulary, maska).
● Przed użyciem należy sprawdzić opakowanie. Nie używać instrumentów, jeśli opakowanie jest uszkodzone.
● Nie używać instrumentów po upływie terminu ważności.
● Przed każdym użyciem należy sprawdzić instrument pod kątem uszkodzeń, takich jak zniekształcenia (wygięcie, odwinięcie), złamanie, korozja, uszkodzone krawędzie tnące, utrata kodu barwnego lub oznakowania. W przypadku tych wskazań urządzenia nie mogą działać zgodnie z przeznaczeniem przy wymaganym poziomie bezpieczeństwa. Instrumenty należy wyrzucić.
● Przed użyciem jakiegokolwiek instrumentu należy upewnić się, że jest on dobrze połączony z główką kątnicy.
● Należy często sprawdzać instrument i czyścić część roboczą podczas jego stosowania, sprawdzając pod kątem zniekształceń, wydłużenia lub oznak zużycia, takich jak nierówne rowki, matowe plamy. W przypadku tych wskazań urządzenia nie mogą działać zgodnie z przeznaczeniem przy wymaganym poziomie bezpieczeństwa. Instrumenty należy wyrzucić.
● Instrumenty nie powinny być w całości zanurzane w roztworze podchlorynu sodu (NaOCl).
Tylko część robocza instrumentu NiTi, która styka się z pacjentem, powinna być zanurzona w stężonym roztworze NaOCl o stężeniu NIEPRZEKRACZAJĄCYM 5%.
● Należy zachować ostrożność w obszarze wierzchołkowym oraz w kanałach rozdzielających się i/lub wykazujących nagłe krzywizny lub krzywizny wsteczne.
● Obficie i często przepłukiwać kanał w trakcie zabiegu oraz po każdym użyciu instrumentu
(zgodnie z dobrą praktyką stomatologiczną).
● W okolicy wierzchołka korzenia należy zawsze stosować minimalny nacisk. W trakcie stosowania pilników w kanale nie wolno używać zbyt dużej siły.
● Gdy instrument nie porusza się łatwo, należy wyczyścić i skontrolować rowki tnące, a następnie przepłukać, powtórzyć rekapitulację za pomocą pilnika ręcznego i ponownie przepłukać.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 65/151
● Do formowania ekstremalnie zakrzywionych kanałów bezpieczniej jest używać pilnika tylko do kształtowania jednego kanału, aby zmniejszyć ryzyko złamania. Należy zwrócić uwagę na następujące dobre praktyki:
○ Należy użyć nowego pilnika i wyrzucić go po wyleczeniu kanału (użycie w jednym kanale).
○ Używać pilników ręcznych zamiast pilników obrotowych.
○ Używać pilników w małych rozmiarach, elastycznych lub/i NiTi (pomoże to uniknąć transportacji kanału).
○ Podczas stosowania (tj. po każdej fali) należy oglądać część roboczą pod kątem wszystkich wad wymienionych w poprzednim akapicie.
○ Unikać standardowego ciągłego obrotowego ruchu rozwiercającego. W zamian należy wykonywać ruchy pod małym kątem (ruch piłowania, ruch oscylacyjny lub technika zrównoważonej siły), aby ograniczyć zmęczenie instrumentów zginaniem rotacyjnym i wydłużyć ich oczekiwaną żywotność.
Przy obecnym stanie technicznym dotychczas nie zgłoszono działań niepożądanych.
Produkty należy przechowywać w suchym i czystym środowisku, z dala od źródeł wilgoci i bezpośredniego nasłonecznienia.
Wszystkie pilniki ProTaper Ultimate™ zostały zatwierdzone do użytku z prędkością 400 obr./min i momentem obrotowym 4 to 5.2 Ncm.
Instrumenty ProTaper Ultimate™ mogą być używane z ruchem szczotkującym na zewnątrz we wszystkich kanałach, zwłaszcza tych o nieregularnym przekroju, lub z lekkim ruchem dziobania do wewnątrz, aby stopniowo przesuwać się w kierunku długości roboczej (WL).
Zawsze należy trzymać prostnicę w zagłębieniu pomiędzy kciukiem a palcem wskazującym. Należy unikać pchania. Zamiast tego należy pozwolić pilnikom pasywnie postępować i podążać torem wprowadzania.
1) Należy przejrzeć różne radiogramy ustawione pod kątem poziomym, aby diagnostycznie określić szerokość, długość i krzywiznę danego korzenia i jego kanału (kanałów).
2) Przygotować dostęp umożliwiający łatwą identyfikację każdego ujścia kanału.
3) Pomocniczy pilnik kształtujący SX może być stosowany w przypadku ograniczonej przestrzeni, do wstępnego powiększenia głównej cześć kanału lub do przeniesienia najbardziej dokoronowego aspektu kanału z zewnętrznego zagłębienia korzenia.
4) W obecności środka poślizgowego należy wybrać slider i PASYWNIE podążać wzdłuż kanału, wykonując jedno lub kilka przejść, aż do jego końca. Określić długość roboczą za pomocą elektronicznego lokalizatora wierzchołka w połączeniu z obrazem radiograficznym, a następnie potwierdzić drożność.
5) Jeśli slider nie dotrze do końca kanału, wybrać niewielki pilnik ręczny ze stali nierdzewnej
(SS). W obecności środka poślizgowego przeprowadzić cewnikowanie tego kanał, ustalić długość roboczą, potwierdzić drożność i zweryfikować tor wprowadzania. Powtórzyć czynność 4 powyżej.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 66/151
6) W obecności NaOCl wybrać shaper i wprowadzać go wzdłuż toru wprowadzania, w jednym lub kilku przejściach, aż do osiągnięcia długości roboczej.
7) Po wyjęciu shapera należy przeprowadzić irygację i użyć urządzenia EndoActivate, aby rozbić zanieczyszczenia i przenieść je do roztworu, a następnie powtórzyć irygację w celu usunięcia zanieczyszczeń.
8) Ponownie potwierdzić długość roboczą, zwłaszcza w bardziej zakrzywionych kanałach.
9) Wybrać FINISHER F1 (020.07) i pasywnie wprowadzać go przez kanał do osiągnięcia długości roboczej, wykonując jedno lub kilku przejść. Wyjąć i sprawdzić rowki wierzchołkowe instrumentu. Po załadowaniu resztkami zębiny preparacja jest zakończona.
10) Jeśli FINISHER F1 jest luźny na długości, a jego rowki wierzchołkowe nie są załadowane zanieczyszczeniami, należy wybrać FINISHER F2 (025.08) i użyć tego pilnika w taki sam sposób, jak opisano powyżej dla FINISHERA F1.
11) Jeśli FINISHER F2 jest luźny na długości, a jego rowki wierzchołkowe nie są załadowane zanieczyszczeniami, należy wybrać FINISHER F3 (030.09) i użyć tego pilnika w taki sam sposób, jak opisano powyżej dla FINISHERA F1 i F2.
12) Po wyjęciu każdego pilnika należy oczyścić i sprawdzić jego rowki tnące, wykonać irygację, przeprowadzić rekapitulację pilnikiem w rozmiarze 10 lub urządzeniem EndoActivator, aby rozbić zanieczyszczenia, a następnie powtórzyć irygację.
13) Po wyjęciu sprawdzić rowki tnące pilnika pod kątem odwijania, prostowania lub rozciągania. W przypadku wykrycia deformacji usunąć pilnik i użyć nowego pilnika ProTaper
Ultimate™.
14) Opracowanie jest zakończone, gdy wierzchołkowy obszar Finishera zostanie załadowany zanieczyszczeniami, a GPMC lub narzędzie potwierdzające rozmiar w odpowiednim rozmiarze pasuje do długości roboczej.
Finisherów pomocniczych należy używać tylko w większych i prostszych kanałach, takich jak centralne siekacze szczęki, niektóre podniebienne lub dystalne kanały zębów trzonowych, lub gdy występuje ubytek patologiczny lub jatrogenny:
15) Po ustaleniu długości roboczej i potwierdzeniu drożności należy wybrać pilnik pomocniczy napędzany mechanicznie lub ręczny FINISHER FX (035.12). Tego niezależnego pilnika należy użyć bezpośrednio i pasywnie, aby podążać kanałem do długości roboczej przy użyciu tej samej techniki, jak opisano powyżej. Wyjąć i sprawdzić rowki wierzchołkowe instrumentu. Po załadowaniu resztkami zębiny preparacja jest zakończona.
16) Jeśli FINISHER FX jest luźny na długości, a jego rowki wierzchołkowe nie są załadowane zanieczyszczeniami, należy wybrać napędzany mechanicznie lub ręczny, pomocniczy FI-
NISHER FXL (050.10) i użyć tego pilnika w taki sam sposób, jak opisano powyżej dla FINI-
SHERA FX
17) Opracowanie jest zakończone, gdy wierzchołkowy obszar Finishera pomocniczego zostanie załadowany zanieczyszczeniami, a GPMC lub narzędzie potwierdzające rozmiar w odpowiednim rozmiarze pasuje do długości roboczej.
● Należy postępować zgodnie z ogólnymi instrukcjami przetwarzania produktów endodontycznych ( https://www.dentsplysirona.com/en/service-contact/ifu.html
) w zakresie etapów dezynfekcji, czyszczenia i sterylizacji.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 67/151
● Ograniczników silikonowych nie wolno używać ponownie. Silikonowe ograniczniki należy usunąć i wyrzucić po każdym użyciu.
● Produkty należy usuwać zgodnie z lokalnymi przepisami dotyczącymi bezpiecznego usuwania ostrych i skażonych urządzeń.
● Wszelkie poważne incydenty związane z produktem należy zgłaszać producentowi i właściwym władzom zgodnie z lokalnymi przepisami.
● Nie można zapewnić sterylności, jeżeli opakowanie jest otwarte, uszkodzone lub mokre.
● Aby uzyskać bezpłatny drukowany egzemplarz instrukcji użycia, należy zapoznaj się z częścią „Get a print copy of IFU” na stronie internetowej https://www.dentsplysirona.com/en/ service-contact/ifu.html
.
● Wyjaśnienie niezharmonizowanych symboli zastosowanych w instrukcji użycia i na oznakowaniu, patrz Symbole w instrukcji użytkowania ( https://www.dentsplysirona.com/en/servicecontact/ifu/ifu-symbols.html
).
2797
Autoryzowany przedstawiciel
DENTSPLY DeTrey GmbH
De-Trey-Straße 1
D-78467 Konstanz
Germany
Producent
Maillefer Instruments Holding Sàrl
Chemin du Verger 3
CH-1338 Ballaigues
Suisse www.dentsplysirona.com/ifu
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 68/151
ProTaper Ultimate™
NAUDOTI TIK DANTIMS GYDYTI
NAUDOTI TIK PROFESIONALAMS
STERILUS – STERILIZUOTAS NAUDOJANT SPINDULIAVIMĄ
BŪTINA PAKARTOTINAI APDOROTI
DAUGKARTINIAI STERILŪS ENDODONTINIAI INSTRUMENTAI
● Gaminiai, skirti endodontinėms ligoms gydyti.
● Numatytoji paskirtis: Varomas variklio instrumentas šaknies kanalui paruošti (slydimo kelias, formavimas ir šaknies kanalo valymas).
● Endodontiniai instrumentai turi būti naudojami tik klinikinėje ar ligoninės aplinkoje, laikantis gerosios odontologinės praktikos. Juos naudoti gali tik kvalifikuoti odontologijos specialistai, pavyzdžiui, bendrosios praktikos gydytojai, endodontologai ir odontologų asistentai.
● Instrumentai bus naudojami kartu su endodontiniu varikliu (rankiniu būdu staigiuose linkiuose).
Nėra.
ProTaper Ultimate™ instrumentai pagaminti iš trijų komponentų: darbinė dalis iš nikelio ir titano lydinio, spalvotas silikono kamštis (išskyrus instrumentus SX) ir žalvariu dengta rankenėlė su spalvotu žiedu (išskyrus instrumentą SX).
Dėl gaminimo proceso ypatybių dildės yra spalvotos.
Dėl šio patentuoto apdorojimo dildės gali atrodyti šiek tiek išlenktos. Tai nėra gamybos defektas. Nors dildę galima nesunkiai ištiesinti pirštais, prieš naudojant to daryti nebūtina. Kanalo viduje dildė atkartoja šaknies kanalo anatomiją.
ProTaper Ultimate™ asortimentą sudaro šie endodontiniai instrumentai:
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 69/151
Kodas xx – ilgis y – dildės dydis z – vienetų lizdinėje pakuotėje
B ST PULR z
19 0SX
B ST PULR z xx 0SL
B ST PULR z xx 0SH
Gaminys
ProTaper Ultimate™
SX
ProTaper Ultimate™
SLIDER
ProTaper Ultimate™
SHAPER
ISO XXX YYYv
XXX-ISO dydis
YYY – kūgiškumas v – kintamasis kūgiškumas
ISO 020.003v
ISO 016.002v
ISO 020.004v
Galimi ilgiai
19 mm
21/25/31 mm
21/25/31 mm
Pakuotė ir turinys
6 instrumentų lizdinė pakuotė (sterili)
3 instrumentų lizdinė pakuotė (sterili)
6 instrumentų lizdinė pakuotė (sterili)
3 instrumentų lizdinė pakuotė (sterili)
6 instrumentų lizdinė pakuotė (sterili)
3 instrumentų lizdinė pakuotė (sterili)
6 instrumentų lizdinė pakuotė (sterili)
B ST PULR z xx 0F y
B ST PULR 6 xx yyy
B ST PULR 5 xx SQE
B DD PULR
525 SQE
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
Sequence
ProTaper Ultimate™
Demo Pack
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
FX: ISO 035.012v
FXL: ISO 050.010v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
21/25/31 mm
21/25/31 mm
21/25/31 mm
25 mm
3 instrumentų lizdinė pakuotė (sterili)
6 instrumentų lizdinė pakuotė (sterili)
1 instrumento lizdinė pakuotė kiekvienam instrumentui (sterili)
1 instrumento lizdinė pakuotė kiekvienam instrumentui (sterili)
● Sukimo momento kontrolės prietaisas užtikrina optimalų naudojimą.
● Naudokite su endodontiniais varikliais, kurie nuolat sukasi 400 aps./min. greičiu.
● Nustatykite sukimo momentą: 4 to 5.2 Ncm.
● Reikia naudoti tepalus, pavyzdžiui, NaOCl, EDTA, ProLube, Glyde™.
● Rentgeno nuotraukų naudojimas kartu su viršūnės lokatoriumi ir įrankiu, skirtu reguliuoti silikono kamščiui iki reikiamo darbinio gylio, yra tinkamas darbinio gylio nustatymo metodas.
● Griežtai laikykitės bendrojo endodontinių gaminių apdorojimo instrukcijos nurodymų (žr. 8 skyriuje), kad sumažintumėte galimas rizikas įrenginiui, pacientui ir (arba) operatoriui.
○ Instrumento lūžimas.
○ Kryžminė tarša.
○ Šilumos išsiskyrimas, nes nepakankamai sutepta ir išplauta.
○ Instrumento darbinės dalies prarijimas.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 70/151
○ Toksinės ar alerginės reakcijos dėl proceso likučių.
● Kaip ir visų mašininių šaknies kanalo instrumentų, ProTaper Ultimate™ instrumentų negalima naudoti, jei danties šaknies viršūnė labai lenkta, nes padidėja atsisluoksniavimo rizika.
Šiuo atveju viršūnės srityje reikia naudoti lenktas rankines dildes.
● Sauga ir veiksmingumas nėščioms arba krūtimi maitinančioms moterims ir vaikams nenustatyti.
● Atlikdami endodontinę procedūrą naudokite koferdamo sistemą.
● Dėl pačių saugumo naudokite asmenines apsaugos priemones (pirštines, akinius, kaukę).
● Prieš naudodami patikrinkite pakuotę ir nenaudokite instrumentų, jei pakuotė pažeista.
● Nenaudokite instrumentų pasibaigus jų galiojimo laikui.
● Kaskart prieš naudodami patikrinkite instrumentą, ar nėra defektų, pavyzdžiui, deformacijų
(sulinkimų, išsipynimų), lūžimų, rūdžių, ašmenų pažeidimų, ar nenusitrynė spalvinis kodas arba žymės. Jei yra tokių požymių, instrumentu negalima saugiai atlikti numatytų užduočių, todėl jį reikia utilizuoti.
● Prieš naudodami bet kurį instrumentą, įsitikinkite, kad jis tinkamai sujungtas su kampinio elemento galvute.
● Patikrinkite instrumentą ir naudodami dažnai valykite darbinę dalį, patikrinkite, ar nėra deformacijos arba nusidėvėjimo požymių, pavyzdžiui, nevienodų griovelių, pilkšvų dėmių. Jei yra tokių požymių, instrumentu negalima saugiai atlikti numatytų užduočių, todėl jį reikia utilizuoti.
● Instrumentų negalima merkti į natrio hipochlorito tirpalą (NaOCl). Į NE didesnės nei 5 % koncentracijos NaOCl tirpalą reikėtų merkti tik darbinę „NiTi“ instrumento dalį, kuri liečiasi prie paciento.
● Išsišakojančiuose ir (arba) staigiai užsilenkiančiuose ar atsilenkiančiuose kanaluose elkitės atsargiai.
● Gausiai ir dažnai plaukite kanalą per visą procedūrą ir po kiekvieno naudoto instrumento
(pagal gerąsias odontologijos praktikas).
● Visuomet naudokite minimalų apikalinį spaudimą. Niekuomet nestumkite dildžių į kanalą per jėgą.
● Kai instrumento negalite lengvai stumti, nuvalykite ir apžiūrėkite kertamąsias apvijas, tada išplaukite kanalą, trumpai pakartokite naudodami rankinę dildę ir vėl išplaukite.
● Formuojant labai riestus kanalus saugiau naudotis dilde tik vienam kanalui suformuoti, kad nelūžtų. Laikykitės toliau pateiktų gerųjų praktikų.
○ Naudokitės nauja dilde ir utilizuokite ją suformavę kanalą (naudokite vienam kanalui).
○ Naudokite rankines, o ne sukamąsias dildes.
○ Naudokite mažas, lanksčias arba (ir) „NiTi“ dildes (tai padės išvengti kanalo perkėlimo).
○ Naudodami apžiūrėkite darbinę dalį, ar ji neturi defektų, nurodytų ankstesnės pastraipos sąraše (t. y. po kiekvienos bangos).
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 71/151
○ Nenaudokite standartinio nepertraukiamo gręžiamojo judesio, geriau naudokite trumpus kampinius judesius (šlifavimo judesį, rankinio laikrodžio prisukimo judesį arba subalansuotos jėgos techniką), kad instrumentams netektų per didelė sukamojo lenkimo apkrova ir juos būtų galima ilgiau naudoti.
Iki šiol pranešimų apie nepageidaujamas reakcijas negauta.
Laikykite gaminius sterilizavimo pakuotėje švarioje aplinkoje atokiau nuo drėgmės šaltinių ir tiesioginių saulės spindulių.
Visos ProTaper Ultimate™ dildės patvirtintos naudoti 400 aps./min. greičiui ir 4 to 5.2 Ncm sukimo momentui.
ProTaper Ultimate™ instrumentus galima naudoti braukiant į išorę visuose kanaluose, ypač tuose, kuriuose yra netaisyklingų skersinių sričių arba švelniais trumpos amplitudės stūmiais gilinant link darbinio gylio (WL).
Atsargiai laikykite rankinį įtaisą tarp nykščio ir rodomojo piršto. Stenkitės nestumti, bet leiskite dildei judėti pasyviai slydimo keliu.
1) Kad diagnostiškai nustatytumėte bet kurios konkrečios šaknies ir kanalo (-ų) plotį, ilgį ir lenktumą, peržiūrėkite skirtingas horizontalaus kampo rentgeno nuotraukas.
2) Paruoškite prieigą taip, kad galėtumėte lengvai identifikuoti kiekvieną kanalo angą.
3) Pagalbines formavimo dildes XF galima naudoti, kai ertmės sunkiau prieinamos arba norint padidinti kanalą, arba norint perkelti kanalo vainikinę kryptį nuo išorinių kanalo įgaubų.
4) Naudodami tepalą, pasirinkite dildę SLIDER ir PASYVIAI pereikite kanalą vienu ar keliais etapais iki kanalo viršūnės. Nustatykite darbinį gylį, naudodami elektroninį viršūnės lokatorių su rentgeno nuotrauka, tuomet patvirtinkite praeinamumą.
5) Jei dildė SLIDER lengvai nepasiekia kanalo viršūnės, pasirinkite mažesnio dydžio nerūdijančiojo plieno rankinę dildę. Naudodami tepalą, kateterizuokite kanalą, pasiekite darbinį gylį, patvirtinkite praeinamumą ir patikrinkite slydimo kelią. Dabar pakartokite 4 veiksmą.
6) Naudodami NaOCl, pasirinkite dildę SHAPER ir judėkite slydimo keliu vienu ar keliais etapais, kol pasieksite darbinį gylį.
7) Išėmę dildę SHAPER, išplaukite, endoaktyvatoriumi suardykite drožles ir perkelkite į tirpalą, tuomet dar kartą išplaukite, kad nešvarumai atsilaisvintų.
8) Dar kartą patikrinkite darbinį gylį, ypač labiau lenktuose kanaluose.
9) Pasirinkite dildę FINISHER F1 (020.07) ir pasyviai pereikite kanalą iki darbinio gylio vienu ar keliais etapais. Pašalinkite ir patikrinkite apkalinius griovelius. Jei juose yra dentino drožlių, paruošta.
10) Jei dildė FINISHER F1 yra laisva ilgiu ir jos apikaliniuose grioveliuose nėra nešvarumų, pasirinkite dildę FINISHER F2 (025.08) ir naudokite ją taip pat kaip dildę FINISHER F1.
11) Jei dildė FINISHER F2 yra laisva ilgiu ir jos apikaliniuose grioveliuose nėra nešvarumų, pasirinkite dildę FINISHER F3 (030.09) ir naudokite ją taip pat kaip dildę FINISHER F1 ir F2.
12) Išėmę bet kurią dildę, išvalykite ir patikrinkite kertamąsias apvijas, trumpai pakartokite naudodami 10 dydžio dildę arba endoaktyvatorių drožlėms suardyti ir vėl išplaukite.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 72/151
13) Išėmę dildę, patikrinkite dildės kertamąsias apvijas, ar nėra išsipynimų, išsitiesinimų ar įsitempimų. Pastebėję deformaciją, utilizuokite ir naudokite naują ProTaper Ultimate™ dildę.
14) Paruošta, kai dildės FINISHER apikalinėje dalyje yra nešvarumų ir atitinkamo dydžio
GPMC (gutaperčios konuso) ar dydžio nustatymo instrumentas telpa darbiniame gylyje.
Naudokite pagalbines dildes FINISHER tik didesniems ir tiesesniems kanalams, pavyzdžiui, viršutinio žandikaulio priekiniams kandžiams, gomuriniams ar distaliniams krūminių dantų kanalams arba kai yra patologija ar jatrogeninis defektas.
15) Pasirinkite rankinę pagalbinę dildę FINISHER FX (035.12), nustatę darbinį gylį ir patvirtinę praeinamumą. Naudokite šią autonominę dildę, kad galėtumėte tiesiogiai ir pasyviai pereiti kanalą iki darbinio gylio ta pačia technika, kuri aprašyta pirmiau. Pašalinkite ir patikrinkite apkalinius griovelius. Jei juose yra dentino drožlių, paruošta.
16) Jei dildė FINISHER FX yra laisva ilgiu ir jos apikaliniuose grioveliuose nėra nešvarumų, pasirinkite dildę FINISHER FXL (050.10) ir naudokite ją taip pat kaip dildę FINISHER FX.
17) Paruošta, kai papildomos dildės FINISHER apikalinėje dalyje yra nešvarumų ir atitinkamo dydžio GPMC (gutaperčios konuso) ar dydžio nustatymo instrumentas telpa darbiniame gylyje.
● Vadovaukitės bendrojo endodontinių gaminių apdorojimo instrukcijos ( https:// www.dentsplysirona.com/en/service-contact/ifu.html
) reikalavimais, norėdami dezinfekuoti, valyti ir sterilizuoti.
● Nenaudokite silikoninių antgalių pakartotinai. Kaskart panaudoję silikoniniu antgalius nuimkite ir išmeskite.
● Gaminiai utilizuojami laikantis vietinių saugaus aštrių ir užterštų priemonių utilizavimo reikalavimų.
● Apie bet kokį rimtą incidentą, susijusį su gaminiu, reikia pranešti gamintojui ir kompetentingai įstaigai pagal vietos teisės aktų reikalavimus.
● Sterilumo negalima garantuoti, jei pakuotė atidaryta, pažeista arba drėgna.
● Norėdami gauti nemokamą spausdintinę naudojimo instrukcijos kopiją, žr. skyriuje „Gauti spausdintinę IFU (naudojimo instrukcijos) kopiją“ interneto svetainėje https:// www.dentsplysirona.com/en/service-contact/ifu.html
.
● Nesuderintųjų simbolių, skirtų naudojimo instrukcijai ir etiketėms, paaiškinimų ieškokite
„Naudojimo instrukcijos simboliai“ ( https://www.dentsplysirona.com/en/service-contact/ifu/ ifu-symbols.html
).
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 73/151
2797
Įgaliotasis atstovas
DENTSPLY DeTrey GmbH
De-Trey-Straße 1
D-78467 Konstanz
Germany
Gamintojas
Maillefer Instruments Holding Sàrl
Chemin du Verger 3
CH-1338 Ballaigues
Suisse www.dentsplysirona.com/ifu
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 74/151
ProTaper Ultimate™
TIKAI IZMANTOŠANAI ZOBĀRSTNIECĪBĀ
TIKAI PROFESIONĀLAI IZMANTOŠANAI
STERILI – STERILIZĒTI AR STAROJUMU
DERĪGI PĀRSTRĀDEI
ATKĀRTOTI LIETOJAMI STERILI ENDODONTIJAS INSTRUMENTI
● Produkti ir paredzēti endodontijas slimību ārstēšanai.
● Paredzētais nolūks: Ar motoru darbināms instruments, kas paredzēts sakņu kanālu sagatavošanai (sakņu kanāla caurejamības un formas veidošana un iztīrīšana).
● Endodontijas instrumentus drīkst lietot kvalificēti zobārstniecības speciālisti, piemēram, vispārējās prakses zobārsti, kā arī endodontijas speciālisti (endodontisti) un zobārstniecības palīgi, tikai klīniskajā vai slimnīcas vidē, ievērojot labu zobārstniecības praksi.
● Instrumenti jāizmanto kopā ar endodontisko motoru (manuāli, ja ir ļoti izliekti kanāli).
Nav.
ProTaper Ultimate™ instrumenti ir izgatavoti no trim galvenajām sastāvdaļām: darba daļa, kas izgatavota no niķeļa-titāna sakausējuma, krāsains silikona aizturis (izņemot instrumentu SX) un kāts ar misiņa pārklājumu un krāsainu gredzenu (izņemot instrumentu SX).
Faiļu ražošanas procesa dēļ tās izskatās kā nokrāsotas.
Šīs patentētās apstrādes dēļ failes var izskatīties nedaudz izliektas. Tas nav ražošanas defekts.
Lai gan faili var vienkārši iztaisnot ar pirkstiem, pirms lietošanas faile nav jāiztaisno. Iekļūstot kanālā, faile pielāgosies saknes kanāla anatomijai.
ProTaper Ultimate™ klāsts ietver šādus endodontijas instrumentus:
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 75/151
Kods xx-garums
Y-failes izmērs z-daudzums blisterī
B ST PULR z
19 0SX
B ST PULR z xx 0SL
B ST PULR z xx 0SH
Produkts
ProTaper Ultimate™
SX
ProTaper Ultimate™
SLIDER
ProTaper Ultimate™
SHAPER
ISO XXX YYYv
XXX-ISO izmērs
YYY-konusa izmērs v-mainīgais konuss
ISO 020.003v
ISO 016.002v
ISO 020.004v
Pieejamie garumi
19 mm
21/25/31 mm
21/25/31 mm
Iepakojums un saturs
Blisteris ar 6 instrumentiem (sterils)
Blisteris ar 3 instrumentiem (sterils)
Blisteris ar 6 instrumentiem (sterils)
Blisteris ar 3 instrumentiem (sterils)
Blisteris ar 6 instrumentiem (sterils)
Blisteris ar 3 instrumentiem (sterils)
Blisteris ar 6 instrumentiem (sterils)
B ST PULR z xx 0F y
B ST PULR 6 xx yyy
B ST PULR 5 xx SQE
B DD PULR
525 SQE
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
Sequence
ProTaper Ultimate™
Demo Pack
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
FX: ISO 035.012v
FXL: ISO 050.010v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
21/25/31 mm
21/25/31 mm
21/25/31 mm
25 mm
Blisteris ar 3 instrumentiem (sterils)
Blisteris ar 6 instrumentiem (sterils)
Blisteris ar 1 instrumentu no katra instrumentu veida (sterils)
Blisteris ar 1 instrumentu no katra instrumentu veida (sterils)
● Griezes kontroles ierīces nodrošina optimālu lietošanu.
● Izmantojiet ar endodontiskajiem motoriem ar konstantu rotācijas ātrumu 400 apgr./min.
● Iestatiet griezi uz: 4 to 5.2 Ncm.
● Jāizmanto tādi lubrikanti kā NaOCl, EDTA, ProLube, Glyde™.
● Rentgenuzņēmumu izmantošana kopā ar apeksa lokatoru un instrumentu silikona aiztura pielāgošanai pareizajam darba garumam ir piemērota darba garuma noteikšanas metode.
● Stingri ievērojiet šo lietošanas instrukciju un Endodontisko izstrādājumu vispārējās apstrādes instrukcijas (sk. 8. sadaļu), lai mazinātu sekojošus riskus ierīcei, pacientam un/vai lietotājam:
○ Instrumenta salūšana.
○ Savstarpēja inficēšanās.
○ Siltuma veidošanās nepietiekamas lubrikācijas un skalošanas dēļ.
○ Instrumenta darba daļas norīšana.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 76/151
○ Toksiskas vai alerģiskas reakcijas, ko izraisa apstrādes atliekas.
● Tāpat kā visus mehāniski darbināmos saknes kanāla instrumentus, ProTaper Ultimate™ instrumentus nedrīkst izmantot izteiktu sakņu kanālu apeksu izliekumu gadījumos, jo pastāv paaugstināts atdalīšanās risks. Šajā gadījumā apikālajā zonā jāizmanto iepriekš izliektas rokas failes.
● Lietošanas drošums un efektivitāte, lietojot grūtniecēm vai sievietēm, kuras baro bērnu ar krūti, un bērniem, nav novērtēta.
● Endodontijas procedūras laikā lietot koferdama sistēmu.
● Drošības nolūkā, lūdzu, lietojiet personisko aizsargaprīkojumu (cimdus, brilles, masku).
● Pirms lietošanas pārbaudiet iepakojumu un nelietojiet instrumentus, ja iepakojums ir bojāts.
● Nelietojiet instrumentus pēc derīguma termiņa beigām.
● Pirms katras lietošanas pārbaudiet, vai instrumentam nav defektu, piemēram, deformāciju
(saliekts, atritināts), lūzumu, korozijas, bojātu griešanas malu, vai nav nodzisis krāsas kods vai marķējums. Ar šādām pazīmēm ierīces nespēj izpildīt paredzēto lietojumu ar nepieciešamo drošības līmeni, un instrumenti jāiznīcina.
● Pirms jebkura instrumenta lietošanas pārliecinieties, ka tas ir labi savienots ar pretleņķa galvu.
● Instrumentācijas laikā regulāri pārbaudiet instrumentu, notīriet darba daļu un pārbaudiet, vai nav redzamas deformācijas, pagarināšanās vai nodiluma pazīmes, piemēram, nevienmērīgas rievas, notrulinājušās vietas. Ar šādām pazīmēm ierīces nespēj izpildīt paredzēto lietojumu ar nepieciešamo drošības līmeni, un instrumenti jāiznīcina.
● Instrumentus nedrīkst pilnībā iegremdēt nātrija hipohlorīta šķīdumā (NaOCl). Tikai NiTi instrumenta darba daļu, kas ir saskarē ar pacientu, drīkst iegremdēt NaOCl šķīduma koncentrātā, kura koncentrācija NEPĀRSNIEDZ 5%.
● Rīkojieties piesardzīgi apikālajā zonā un kanālos, kas dalās un/vai kuros ir strauji izliekumi vai ieliekumi.
● Visas procedūras laikā un pēc katra izmantotā instrumenta (saskaņā ar labu zobārstniecības praksi) bagātīgi un bieži skalojiet kanālu.
● Vienmēr izmantojiet minimālu spiedienu uz apeksu. Nekādā gadījumā nebīdiet faili kanālā ar spēku.
● Ja instruments nevirzās viegli, notīriet un pārbaudiet griešanas rievas, pēc tam skalojiet, atkārtojiet darbību ar manuālo faili un skalojiet vēlreiz.
● Ārkārtīgi izliektu kanālu formas veidošanai drošāk ir izmantot faili, lai veidotu tikai viena kanāla formu nolūkā samazināt bojājumu risku. Ievērojiet tālāk norādītos labas prakses norādījumus:
○ Izmantojiet jaunu faili un izmetiet to pēc kanāla apstrādes (ja tiek lietots vienam kanālam).
○ Izmantojiet manuālās, nevis rotējošās failes.
○ Izmantojiet maza izmēra elastīgas un/vai NiTi failes (šādi var izvairīties no kanāla transportēšanas).
○ Lai noteiktu iepriekšējā punktā uzskaitītos bojājumus, vizuāli pārbaudiet darba daļu lietošanas laikā (t. i. pēc katras kārtas).
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 77/151
○ Izvairieties no standarta izrīvēšanas nepārtrauktās rotējošās kustības, bet tā vietā izmantojiet neliela leņķa kustības (vīlēšanas kustību, pulksteņa uzvilkšanas svārstību kustību vai līdzsvarota spēka tehniku), lai ierobežotu instrumentu rotējošās lieces nogurumu un pagarinātu to kalpošanas laiku.
Pašreizējā tehniskajā stāvoklī par nelabvēlīgas reakcijas gadījumiem līdz šim nav ziņots.
Glabājiet izstrādājumus sausā un tīrā vidē, kas nav pakļauta mitruma vai tiešai saules gaismas iedarbībai.
Visas ProTaper Ultimate™ failes ir apstiprinātas lietošanai ar ātrumu 400 apgr./min un 4 to
5.2 Ncm griezi.
ProTaper Ultimate™ instrumentus var izmantot ar uz āru vērstu slaukošu kustību visos kanālos, it īpaši tajos kanālos, kuru šķērsgriezums ir neregulārs, vai ar vieglu uz iekšu vērstu knābājošu kustību, lai pakāpeniski virzītos uz darba garumu (WL).
Vienmēr turiet rokturi starp īkšķi un rādītājpirkstu. Izvairieties no spiešanas; drīzāk ļaujiet failēm pasīvi virzīties un sekot pieejas ceļam.
1) Apskatiet dažādus horizontālā leņķī uzņemtus rentgenuzņēmumus, lai noteiktu konkrētās saknes un tās kanāla(-u) platumu, garumu un izliekumu.
2) Sagatavojiet piekļuvi, kas ļauj viegli identificēt katru kanāla atveri.
3) Papildu formēšanas faili SX var izmantot, ja ir ierobežota telpa vai lai iepriekš palielinātu kanāla atveri, vai arī pārvietotu kanāla koronālo – lielāko daļu – prom no ārējā saknes ieliekuma.
4) Izmantojot lubrikantu, izvēlieties Slider un PASĪVI sekojiet kanālam vienā vai vairākos kanālos līdz to galam. Nosakiet darba garumu (WL), izmantojot elektronisko apeksa lokatoru kopā ar rentgenuzņēmumu, pēc tam pārliecinieties par caurejamību.
5) Ja Slider nesasniedz kanāla galu viegli, izvēlieties maza izmēra manuālo nerūsējošā tērauda (SS) rokas faili. Izmantojot lubrikantu, izejiet šo kanālu, nosakiet darba garumu, pārbaudiet caurejamību un pārbaudiet Slidepath. Atkārtojiet iepriekšminēto soli #4.
6) Izmantojot NaOCl, izvēlieties Shaper un virziet gar Slidepath vienā vai vairākos kanālos, līdz tiek sasniegts WL.
7) Izņemot Shaper, skalojiet kanālu, lietojiet EndoActivate, lai izšķīdinātu skaidiņu slāni, pēc tam atkārtoti skalojiet, lai izskalotu skaidiņas ārā.
8) Pārbaudiet WL, it īpaši vairāk izliektos kanālos.
9) Izvēlieties FINISHER F1 (020.07) un pasīvi sekojiet kanālam līdz WL vienā vai vairākos kanālos. Izņemiet un pārbaudiet apikālās rievas. Ja tās ir aizpildītas ar dentīna skaidiņām, sagatavošanās ir pabeigta.
10) Ja FINISHER F1 ir vaļīgs visā garumā un tā apikālajās rievās nav skaidiņu, atlasiet FI-
NISHER F2 (025.08) un izmantojiet šo faili tādā pašā veidā, kā aprakstīts iepriekš FI-
NISHER F1.
11) Ja FINISHER F2 ir vaļīgs visā garumā un tā apikālajās rievās nav skaidiņu, atlasiet FI-
NISHER F3 (030.09) un izmantojiet šo faili tādā pašā veidā, kā aprakstīts iepriekš FI-
NISHER F1 un F2.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 78/151
12) Izņemot jebkuru norādīto faili, notīriet un pārbaudiet tās griešanas rievas, izskalojiet, atkārtojiet ar 10 izmēra faili vai Endo Activator, lai atdalītu skaidiņas, pēc tam atkal izskalojiet.
13) Pēc izņemšanas pārbaudiet, vai failes griešanas rievas nav atritinātas, iztaisnotas vai izstieptas. Ja tiek konstatēta deformācija, izmetiet un izmantojiet jaunu ProTaper Ultimate™ faili.
14) Sagatavošana ir pabeigta, kad jebkura Finisher apikālais pagarinājums ir pilns ar skaidiņām un atbilstoša izmēra GPMC (gutaperčas Master konusu) vai Size Verifier der WL.
Papildu Finisher izmantojiet tikai lielākos un taisnākos kanālos, piemēram, augšžokļa centrālajos priekšzobos, dažos dzerokļu palatinālajos vai distālajos kanālos vai ja ir patoloģisks vai jatrogēns defekts:
15) Kad darba garums ir noteikts un caurejamība ir apstiprināta, izvēlieties mehāniski darbināmu vai manuālo palīgfaili FINISHER FX (035.12). Izmantojiet šo atsevišķo faili, lai tieši un pasīvi sekotu kanālam līdz WL, izmantojot to pašu tehniku, kas aprakstīta iepriekš.
Izņemiet un pārbaudiet apikālās rievas. Ja tās ir aizpildītas ar dentīna skaidiņām, sagatavošanās ir pabeigta.
16) Ja FINISHER FX ir vaļīgs visā garumā un tā apikālajās rievās nav skaidiņu, atlasiet mehāniski darbināmu vai manuālo palīgfaili FINISHER FXL (050.10) un izmantojiet šo faili tādā pašā veidā, kā aprakstīts iepriekš FINISHER FX.
17) Sagatavošana ir pabeigta, kad jebkura papildu Finisher apikālais pagarinājums ir pilns ar skaidiņām un atbilstoša izmēra GPMC vai Size Verifier der WL.
● Dezinfekcijas, tīrīšanas un sterilizācijas soļos, lūdzu, ievērojiet endodontijas izstrādājumu vispārīgās apstrādes instrukcijas (GPI) ( https://www.dentsplysirona.com/en/service-contact/ ifu.html
).
● Neizmantojiet silikona aizturus atkārtoti. Noņemiet un izmetiet silikona aizturus pēc katras lietošanas reizes.
● Izstrādājumi ir jāizmet saskaņā ar vietējiem noteikumiem par drošu asu un piesārņotu ierīču iznīcināšanu.
● Par visiem nopietniem incidentiem saistībā ar izstrādājumu jāziņo ražotājam un kompetentajai iestādei saskaņā ar vietējiem noteikumiem.
● Sterilitāti nevar garantēt, ja iepakojums ir atvērts, bojāts vai samircis.
● Lai saņemtu IFU bezmaksas drukātu kopiju, lūdzu, ejiet uz sadaļu “Saņemt drukātu IFU kopiju” vietnē https://www.dentsplysirona.com/en/service-contact/ifu.html
.
● Neharmonizēto simbolu skaidrojumus lietošanas instrukcijām un etiķetēm skatīt IFU Simboli
( https://www.dentsplysirona.com/en/service-contact/ifu/ifu-symbols.html
).
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 79/151
2797
Pilnvarotais pārstāvis
DENTSPLY DeTrey GmbH
De-Trey-Straße 1
D-78467 Konstanz
Germany
Ražotājs
Maillefer Instruments Holding Sàrl
Chemin du Verger 3
CH-1338 Ballaigues
Suisse www.dentsplysirona.com/ifu
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 80/151
ProTaper Ultimate™
KASUTAMISEKS AINULT HAMBARAVIS
AINULT PROFESSIONAALSEKS KASUTAMISEKS
STERIILNE – STERILISEERITUD KIIRGUSE ABIL
TÖÖTLEMISE EEST VASTUTAV
KORDUVKASUTATAVAD STERIILSED ENDODONTILISED INSTRUMENDID
● Tooted on ette nähtud endodontiliste haiguste raviks.
● Sihtotstarve: Juurekanali ettevalmistamiseks (juurekanali libisemistee loomine, kujundamine ja puhastamine) mõeldud mootoriga käitatav instrument.
● Endodontilised instrumendid on ette nähtud kasutamiseks kliinikus või haiglakeskkonnas hambaarstide, endodontia spetsialistide (endodontistide) ja hambaravi assistentide poolt, kes järgivad stomatoloogias heakskiidetud praktikat.
● Instrumente kasutatakse kombinatsioonis endodontilise mootoriga (tugevate kõveruste korral käsitsi).
Puuduvad.
ProTaper Ultimate™ instrumendid on valmistatud kolmest põhikomponendist: nikli-titaani sulamist tööosa, värviline silikoonstopper (va instrument SX) ja kaetud messingvarras värvilise rõngaga (va instrument SX).
Viilide tootmisel on kasutatud protsessi, mille tulemusel on viili välimus värviline.
Selle patenditud protsessi tõttu võivad viilid olla veidi kõverad. See pole tootmisdefekt. Kuigi viili saab sõrmega kergelt sirgeks painutada, pole see enne viili kasutamist vajalik. Kanalis liikudes järgib viil juurekanali anatoomiat.
ProTaper Ultimate™ tootevalikusse kuuluvad järgmised endodontilised instrumendid:
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 81/151
Kood xx-pikkus y-viili suurus z-kogus blisterpakendi kohta
Toode
ISO XXX YYYv
XXX-ISO suurus
YYY-taper tihvti suurus v-varieeruv taper tihvt
Saadaolevad pikkused
Pakend ja sisu
B ST PULR z
19 0SX
B ST PULR z xx 0SL
B ST PULR z xx 0SH
B ST PULR z xx 0F y
B ST PULR 6 xx yyy
B ST PULR 5 xx SQE
B DD PULR
525 SQE
ProTaper Ultimate™
SX
ProTaper Ultimate™
SLIDER
ProTaper Ultimate™
SHAPER
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
Sequence
ProTaper Ultimate™
Demo Pack
ISO 020.003v
ISO 016.002v
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
FX: ISO 035.012v
FXL: ISO 050.010v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
19 mm
21/25/31 mm
21/25/31 mm
21/25/31 mm
21/25/31 mm
21/25/31 mm
25 mm
Blisterpakend 6 instrumendiga (steriilsed)
Blisterpakend 3 instrumendiga (steriilsed)
Blisterpakend 6 instrumendiga (steriilsed)
Blisterpakend 3 instrumendiga (steriilsed)
Blisterpakend 6 instrumendiga (steriilsed)
Blisterpakend 3 instrumendiga (steriilsed)
Blisterpakend 6 instrumendiga (steriilsed)
Blisterpakend 3 instrumendiga (steriilsed)
Blisterpakend 6 instrumendiga (steriilsed)
Blisterpakend ühega igast instrumendist
(steriilsed)
Blisterpakend ühega (1) igast instrumendist (steriilsed)
● Pöördemomendi reguleerimisseadmed tagavad optimaalse kasutamise.
● Kasutamine endodontiliste mootoritega ühtlasel pöörlemissagedusel kiirusega 400 p/min.
● Pöördemomendi reguleerimine sagedusele: 4 to 5.2 Ncm.
● Kasutada tuleb järgmisi libesteid: NaOCl, EDTA, ProLube, Glyde™.
● Tööpikkuse määramise sobiv meetod on röntgenpiltide kasutamine koos tipuleidjaga ja silikoonstopperi õigele tööpikkusele reguleerimise tööriistaga.
● Järgige rangelt käesolevat kasutusjuhendit ja endodontiliste toodete üldisi töötlemisjuhiseid
(vt punkt 8), et minimeerida järgmisi ohte seadmele, patsiendile ja/või kasutajale:
○ instrumendi purunemine.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 82/151
○ ristsaastumine.
○ soojuse teke ebapiisava libestite kasutamise ja loputamise tõttu.
○ instrumendi tööosa allaneelamine.
○ töötlemisjääkidest tingitud toksilised või allergilised reaktsioonid.
● Sarnaselt kõigi masinaga käitatavate juurekanali instrumentidega, ei tohi ProTaper
Ultimate™ instrumente kasutada kõrgenenud murdumisriski tõttu järskude apikaalsete kõverustega juurekanalis. Sellisel juhul tuleb apikaalses piirkonnas kasutada eelkõverdatud käsiviile.
● Kasutamise ohutust ja toimivust rasedatel või imetavatel naistel või lastel pole tõestatud.
● Endodontilise protseduuri ajal kasutage kofferdamisüsteemi.
● Ohutuse tagamiseks kasutage isikukaitsevahendeid (kindad, prillid, mask).
● Enne kasutamist kontrollige pakendit ja ärge kasutage instrumente, kui pakend on kahjustatud.
● Ärge kasutage instrumente pärast aegumiskuupäeva.
● Enne iga kasutamist kontrollige instrumendil kahjustuste ilmingud, näiteks kujumuutused
(kõverdumine, lahtikeerdumine), purunemine, korrosioon, kahjustatud lõikeservad, tunnusvärvi või -märgi kulumine. Need on märgiks, et seadmetega pole võimalik teha vajalikku toimingut nõutava ohutustasemega ning need tuleb kasutusest kõrvaldada.
● Enne mis tahes instrumendi kasutamist veenduge, et see on korralikult nurkotsiku pea külge kinnitatud.
● Instrumentide kasutamise ajal kontrollige ja puhastage sageli tööosi, otsides kujumuutusi, pikenemis- ja kulumistunnuseid nagu ebatasased sooned, nürid kohad. Need on märgiks, et seadmetega pole võimalik teha vajalikku toimingut nõutava ohutustasemega ning need tuleb kasutusest kõrvaldada.
● Instrumente ei tohi täielikult kasta naatriumhüpokloriti (NaOCl) lahusesse. NaOCl lahusesse, mille kontsentratsioon EI ÜLETA 5%, võib kasta ainult patsiendiga kokku puutuvaid NiTi instrumentide tööosi.
● Olge ettevaatlik apikaalses piirkonnas ja kanalites, mis jagavad ja/või millel on järsud kõverused või vastukõverused.
● Loputage kanalit rikkalikult ja sageli kogu protseduuri vältel ja pärast iga instrumendi kasutamist (stomatoloogias heakskiidetud praktika kohaselt).
● Rakendage alati minimaalset apikaalset survet. Ärge kunagi kasutage viili juurekanalisse lükkamiseks jõudu.
● Kui instrument ei liigu kergelt edasi, siis puhastage ja kontrollige lõikesooni, seejärel loputage, korrake käsiviiliga ja loputage uuesti.
● Väga kõverate kanalite vormimisel on ohutum kasutada viili ainult ühe kanali kujundamiseks, kuna see vähendab purunemisohtu. Järgige järgmisi häid tavasid:
○ Kasutage uut viili ning kõrvaldage see pärast kanali ravi (ühes kanalis kasutamine).
○ Kasutage pöördviilide asemel käsiviile.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 83/151
○ Kasutage väikese suurusega, painduvaid või/ja NiTi viile (aitab vältida kanali transportimist).
○ Kontrollige kasutamise ajal iga tööosa visuaalselt eelmises lõigus loetletud defektide suhtes (st pärast iga etappi).
○ Vältige standardset pidevalt pöörlevat hõõritsusliikumist ning kasutage selle asemel väikese nurga all liigutusi (viilimisliigutus, kellakeeramise võnkuv liigutus või tasakaalustatud jõutehnika), et piirata seadmete pöördpaindest tulenevat kulumist ja pikendada nende eeldatavat tööiga.
Praeguse tehnilise seisundi juures pole mingeid kõrvaltoimeid täheldatud.
Säilitage tooteid kuivas ja puhtas keskkonnas, eemal niiskusallikatest ja otsest päikesevalgusest.
Kõik ProTaper Ultimate™ viilid on kinnitatud kasutamiseks kiirusel 400 p/min ja pöördemomendiga 4 to 5.2 Ncm.
ProTaper Ultimate™ instrumente võib väljatõmbavate harjamisliigutustega kasutada kõigis kanalites, eelkõige korrapäratu ristlõikega kanalites, või kergete sissepoole suunatud liigutustega, et tööpikkuse (WL) suunas järk-järgult edasi liikuda.
Hoidke käsiinstrumenti alati pöidla ja nimetissõrme vahelise naha vastas. Vältige surumist, laske pigem viilil passiivselt edasi liikuda ja järgida libisemisteed.
1) Vaadake üle erinevad horisontaalse nurga all tehtud röntgenpildid, et hinnata diagnostiliselt antud juure ja selle kanali(te) laiust, pikkust ja kõverust.
2) Prepareerige juurdepääs, mis võimaldab iga kanali suudme hõlpsasti tuvastada.
3) Kujundamise abiviili SX võib kasutada takistuseni jõudes või kanali korpuse eelsuurendamiseks või kanali koronaaltasapinna eemale juhtimiseks välisest juure süvendist.
4) Kasutage libesti juuresolekul viili Slider, et kanal PASSIIVSELT ühe või mitme viilimiskäiguga läbida, kuni jõuate juure tipuni. Tööpikkuse määramiseks kasutage elektroonilist tipuleidjat koos röntgenpiltidega, siis kinnitage läbitavus.
5) Kui Slider ei liigu kergelt kanali tipuni, siis valige väiksem roostevabast terasest (SS) käsiviil. Katetriseerige libesti juuresolekul see kanal, määrake tööpikkus, kinnitage läbitavus ja kontrollige libisemistee olemasolu. Nüüd korrake ülaltoodud sammu nr 4.
6) Valige viil Shaper ja liikuge NaOCl keskkonnas mööda libisemisteed ühe või mitme viilimiskäiguga kuni on saavutatud tööpikkus.
7) Pärast viili Shaper eemaldamist loputage, EndoActivate dentiinijääkide purustamiseks ja lahusesse viimiseks, seejärel loputage uuesti dentiinijääkide eemaldamiseks.
8) Kontrollige uuesti tööpikkust, eelkõige kõverate kanalite korral.
9) Valige viil FINISHER F1 (020.07) ja läbige kanal passiivselt kogu tööpikkuses ühe või mitme viilimiskäiguga. Eemaldage ja kontrollige selle apikaalseid sooni. Kui need on täis dentiinijääke, siis on prepareerimine lõpetatud.
10) Kui FINISHER F1 on pikkuses vaba ja selle apikaalsetes soontes pole dentiinijääke, valige
FINISHER F2 (025.08) ja kasutage seda viili sarnaselt ülalkirjeldatud viilile FINISHER F1.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 84/151
11) Kui FINISHER F2 on pikkuses vaba ja selle apikaalsetes soontes pole dentiinijääke, valige
FINISHER F3 (030.09) ja kasutage seda viili sarnaselt ülalkirjeldatud viilidele FINISHER
F1 ja F2.
12) Pärast iga viili eemaldamist puhastage ja kontrollige selle lõikesooni, loputage, korrake käsiviiliga suurus 10 või ja Endo Activatoriga dentiinijääkide purustamiseks, siis loputage uuesti.
13) Pärast eemaldamist kontrollige viili lõikesooni, et need ei ole lahti keerdunud, sirgenenud või veninud. Deformeerumise korral kõrvaldage ja kasutage uut ProTaper Ultimate™ viili.
14) Prepareerimine on lõpetatud, kui viili Finisher apikaalne ulatus on täis dentiinijääke ja vastava suurusega gutapertšist peatihvt või Size Verifier sobib tööpikkusega.
Kasutage lisaviile Finisher ainult suuremates ja sirgemates kanalites, nt ülalõua tsenraalsed intsisiivid, mõned molaaride palataalsed või distaalsed kanalid, või kui esineb patoloogiline või iatrogeenne defekt:
15) Kui tööpikkus on määratud ja läbitavus kinnitatud, valige kas mehaaniliselt või käsitsi käitatav lisaviil FINISHER FX (035.12). Kasutage seda eraldiseisvat viili kanali tööpikkuse otseseks ja passiivseks läbimiseks, kasutades ülalkirjeldatud tehnikat. Eemaldage ja kontrollige selle apikaalseid sooni. Kui need on täis dentiinijääke, siis on prepareerimine lõpetatud.
16) Kui FINISHER FX on pikkuses vaba ja selle apikaalsetes soontes pole dentiinijääke, valige kas mehaaniliselt või käsitsi käitatav lisaviil FINISHER FXL (050.10) ja kasutage seda viili sarnaselt ülalkirjeldatud viilile FINISHER FX.
17) Prepareerimine on lõpetatud, kui lisaviili Finisher apikaalne ulatus on täis dentiinijääke ja vastava suurusega gutapertšist peatihvt või Size Verifier sobib tööpikkusega.
● Desinfitseerimise, puhastamise ja steriliseerimise ajal järgige endodontiliste toodete üldisi töötlemisjuhised ( https://www.dentsplysirona.com/en/service-contact/ifu.html
).
● Ärge kasutage silikoonstoppereid korduvalt. Eemaldage ja visake silikoonstopperid ära.
● Tooted tuleb kõrvaldada vastavalt teravate ja saastunud seadmete ohutut kõrvaldamist reguleerivatele kohalikele õigusaktidele.
● Kõikidest tootega seotud tõsistest juhtumitest tuleb vastavalt kohalikele eeskirjadele teatada tootjale ja pädevale asutusele.
● Avatud, kahjustatud või niiske pakendi korral pole steriilsus tagatud.
● Kasutusjuhendi tasuta väljatrükiks vt jaotist “Get a print copy of IFU” veebilehel https:// www.dentsplysirona.com/en/service-contact/ifu.html
.
● Kasutusjuhendite ja siltide ühtlustamata tähiste selgitus, vt kasutusjuhendi sümbolid
( https://www.dentsplysirona.com/en/service-contact/ifu/ifu-symbols.html
).
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 85/151
2797
Volitatud esindaja
DENTSPLY DeTrey GmbH
De-Trey-Straße 1
D-78467 Konstanz
Germany
Tootja
Maillefer Instruments Holding Sàrl
Chemin du Verger 3
CH-1338 Ballaigues
Suisse www.dentsplysirona.com/ifu
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 86/151
ProTaper Ultimate™
POUZE PRO POUŽITÍ V ZUBNÍM LÉKAŘSTVÍ
POUZE PRO PROFESIONÁLNÍ POUŽITÍ
STERILNÍ - STERILIZOVANÉ ZÁŘENÍM
VHODNÉ NA OPĚTOVNÉ ZPRACOVÁNÍ
OPAKOVANĚ POUŽITELNÉ STERILNÍ ENDODONTICKÉ NÁSTROJE
● Výrobky jsou indikovány k léčbě endodontických onemocnění.
● Určený účel: Nástroj poháněný motorem určený k preparaci kořenových kanálků (klouzání, tvarování a odstraňování nečistot z kořenového kanálku).
● Endodontické nástroje mají být používány pouze v klinickém nebo nemocničním prostředí, při dodržení správné stomatologické praxe, kvalifikovanými zubními odborníky, jako jsou praktičtí lékaři a endo-specialisté (endodontisté) a zubní asistenti.
● Nástroje se používají v kombinaci s endodontickým motorem (manuálně při silných zakřiveních).
Žádné.
ProTaper Ultimate™ nástroje jsou vyrobeny ze tří hlavních komponent: pracovní část je vyrobena ze slitiny niklu a titanu, barevné silikonové zarážky (kromě nástroje SX) a barevného mosazného dříku s barevným kroužkem (kromě nástroje SX).
Pilníky jsou vyráběny procesem, jehož výsledkem je pilník, který má barevný vzhled.
Díky tomuto speciálnímu zpracování mohou pilníky vypadat mírně zakřivené. Nejedná se o výrobní vadu. Pilník lze sice snadno narovnat pouze pomocí prstů, ale před použitím není nutné pilník narovnávat. Po vstupu do kanálku bude pilník kopírovat anatomii kořenového kanálku.
ProTaper Ultimate™ Produktová řada zahrnuje následující endodontické nástroje:
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 87/151
Kód xx - délka y - velikost pilníku z - množství v blistru
B ST PULR z
19 0SX
B ST PULR z xx 0SL
B ST PULR z xx 0SH
Produkt
NÁSTROJProTaper
Ultimate™ SX
NÁSTROJProTaper
Ultimate™ SLIDER
NástrojProTaper
Ultimate™ SHAPER
ISO XXX RRRRv
XXX - velikost ISO
YYY - velikost kuželové části nástroje v - variabilní zúžení
ISO 020.003v
ISO 016.002v
ISO 020.004v
Dostupné délky
19 mm
21/25/31 mm
21/25/31 mm
Balení a obsah
Blistr se 6 nástroji (sterilní)
Blistr se 3 nástroji (sterilní)
Blistr se 6 nástroji (sterilní)
Blistr se 3 nástroji (sterilní)
Blistr se 6 nástroji (sterilní)
Blistr se 3 nástroji (sterilní)
Blistr se 6 nástroji (sterilní)
B ST PULR z xx 0F y
B ST PULR 6 xx yyy
B ST PULR 5 xx SQE
B DD PULR
525 SQE
NástrojProTaper
Ultimate™ FINISHER
NástrojProTaper
Ultimate™ FINISHER
ProTaper Ultimate™
Série
ProTaper Ultimate™
Demo Pack
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
FX: ISO 035.012v
FXL: ISO 050.010v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
21/25/31 mm
21/25/31 mm
21/25/31 mm
25 mm
Blistr se 3 nástroji (sterilní)
Blistr se 6 nástroji (sterilní)
Blistr s 1 kusem každého nástroje (sterilní)
Blistr s 1 kusem každého nástroje (sterilní)
● Zařízení pro regulaci točivého momentu zajišťují optimální využití.
● Používejte s endodontickými motory s konstantní rychlostí otáčení. 400 rpm.
● Nastavte točivý moment na: Točivý moment4 to 5.2 Ncm.
● Používají se lubrikanty, jako například NaOCl, EDTA, ProLube, Glyde™.
● Vhodnou metodou určení pracovní délky je použití rentgenových snímků v kombinaci s apexlokátorem a nástrojem pro nastavení silikonové zátky na správnou pracovní délku.
● Přísně dodržujte tento návod k použití a Všeobecné pokyny pro zpracování endodontických výrobků (viz část 8), abyste minimalizovali následující rizika pro zařízení, pacienta a/nebo uživatele:
○ Zlomení nástroje.
○ Křížová kontaminace.
○ Tvorba tepla v důsledku nedostatečné lubrikace a irigace.
○ Spolknutí pracovní části nástroje.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 88/151
○ Toxické nebo alergické reakce způsobené zbytky po zpracování.
● Podobně jako u všech strojově poháněných nástrojů na kořenové kanálky, ProTaper
Ultimate™ nástroje by se neměly používat v kořenovém kanálku s náhlým apikálním zakřivením z důvodu zvýšeného rizika separace. V tomto případě by měly být v apikální oblasti použity předem zakřivené ruční pilníky.
● Bezpečnost a účinnost použití u těhotných a kojících žen nebo u dětí nebyla stanovena.
● Při endodontickém zákroku používejte kofferdamový systém.
● Pro vlastní bezpečnost používejte osobní ochranné pomůcky (rukavice, brýle, masku).
● Před použitím zkontrolujte obal a nepoužívejte nástroje, pokud je obal poškozen.
● Nepoužívejte nástroje po uplynutí doby použitelnosti.
● Před každým použitím nástroje zkontrolujte, zda nevykazují známky závad, jako jsou deformace (ohnutí, rozvinutí), zlomení, koroze, poškozené řezné hrany, ztráta barvy nebo označení. Při těchto příznacích přístroje nejsou schopny plnit určené použití s požadovanou
úrovní bezpečnosti, přístroje by měly být vyřazeny.
● Před použitím jakéhokoli nástroje se ujistěte, že je dobře připojen k protiúhlové hlavě.
● Během práce s nástrojem často kontrolujte nástroj a čistěte pracovní část. Zkontrolujte, zda nevykazuje známky deformace, prodloužení nebo opotřebení, jako jsou nerovnoměrné drážky, matná místa. Při těchto příznacích přístroje nejsou schopny plnit určené použití s požadovanou úrovní bezpečnosti, přístroje by měly být vyřazeny.
● Přístroje se nesmí zcela ponořit do roztoku chlornanu sodného (NaOCl). Pouze pracovní část nástroje z NiTi, která je v kontaktu s pacientem, by měla být ponořena do roztoku
NaOCl o koncentraci NE vyšší než 5 %.
● Buďte opatrní v apikální oblasti a v kanálech, které se dělí a/nebo vykazují náhlé zakřivení nebo opětovné zakřivení.
● Hojně a často irigujte kanálek během celého zákroku a po každém použitém nástroji (podle správné stomatologické praxe).
● Vždy používejte minimální apikální tlak. Nikdy netlačte pilníky do kanálku násilím.
● Pokud nástroj nepostupuje snadno, vyčistěte a zkontrolujte řezné drážky, poté irigujte, rekapitulujte ručním pilníkem a znovu irigujte.
● Při tvarování extrémně zakřivených kanálků je bezpečnější použít pilník pouze k tvarování jednoho kanálku, aby se snížilo riziko zlomení. Dbejte na následující správné postupy:
○ Použijte nový pilník a po úpravě kanálku ho zlikvidujte (použití jen pro jeden kanálek).
○ Používejte ruční pilníky místo rotačních.
○ Používejte pilníky malých rozměrů, ohebné nebo/i NiTi pilníky (tím se můžete vyhnout přesunu kanálku).
○ Během používání (tj. po každé vlně) vizuálně zkontrolujte pracovní část na všechny závady uvedené v předchozím odstavci.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 89/151
○ Vyhněte se standardnímu kontinuálnímu rotačnímu pohybu a místo toho používejte pohyby pod malými úhly (pohyb pilování, oscilační pohyb při navíjení hodinek nebo techniku vyvážené síly), abyste omezili únavu nástrojů z rotačního ohybu a zvýšili jejich očekávanou životnost.
Za současného technického stavu nebyly dosud hlášeny žádné nežádoucí reakce.
Výrobky skladujte v suchém a čistém prostředí, mimo zdroj vlhkosti a přímého slunečního záření.
Všechny ProTaper Ultimate™ pilníky byly ověřeny pro použití při 400 ot/min a otáčivém momentu4 to 5.2 Ncm torque.
ProTaper Ultimate™ Nástroje lze používat s kartáčováním směrem ven ve všech kanálcích, zejména v těch, které mají nepravidelný průřez, nebo s lehkým klovacím pohybem směrem dovnitř, aby se postupně přiblížily k pracovní délce (WL).
Rukojeť vždy držte v části mezi palcem a ukazovákem. Vyhněte se tlačení; raději nechte pilníky pasivně postupovat a sledujte dráhu posuvu.
1) Prohlédněte si různé rentgenové snímky nasnímané pod různým horizontálním úhlem, abyste diagnosticky stanovili šířku, délku a zakřivení daného kořene a jeho kanálku (kanálků).
2) Připravte si přístup, který umožní snadnou identifikaci každého ústí kanálku.
3) Pomocný tvarovací pilník SX lze použít v případě, že je omezený prostor, nebo k předběžnému zvětšení těla kanálku, případně k přemístění nejkoronálnější části kanálku od vnější konkávní části kořene.
4) Za přítomnosti lubrikantu zvolte Slider a jedním nebo více průchody PASIVNĚ postupujte kanálkem až k jeho konci. Určete pracovní délku pomocí elektronického apexlokátoru v kombinaci s radiografickým snímkem a poté potvrďte průchodnost.
5) Pokud Slider nedosáhne snadno konec kanálku, zvolte ruční pilník z nerezové oceli (SS) malých rozměrů. Za přítomnosti lubrikantu katetrizujte tento kanálek, stanovte pracovní délku, potvrďte průchodnost a ověřte dráhu posouvání. Nyní zopakujte výše uvedený krok č. 4.
6) Za přítomnosti NaOCl zvolte Shaper a postupujte podél posuvné dráhy jedním nebo více průchody, dokud nedosáhnete pracovní délky.
7) Po vyjmutí Shapera, irigujte, pomocí EndoActivate, abyste rozbili zbytky a přesunuli je do roztoku, a poté znovu proveďte irigaci, abyste tyto zbytky uvolnili.
8) Znovu ověřte pracovní délku, zejména v zakřivenějších kanálcích.
9) Zvolte FINISHER F1 (020.07) a pasivně sledujte kanálek až po pracovní délku, a to jedním nebo více průchody. Vyjměte a zkontrolujte jeho apikální drážky. Při naplnění dentinovými zbytky je preparace dokončena.
10) Pokud je FINISHER F1 volný na pracovní délku a jeho apikální drážky nejsou naplněny nečistotami, zvolte FINISHER F2 (025.08) a použijte tento pilník stejným způsobem, jaký byl popsán výše pro FINISHER F1.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 90/151
11) Pokud je FINISHER F2 na pracovní délku volný a jeho apikální drážky nejsou naplněny nečistotami, vyberte FINISHER F3 (030.09) a použijte tento pilník stejným způsobem, jaký je popsán výše pro FINISHER F1 a F2.
12) Po vyjmutí kteréhokoli z uvedených pilníků očistěte a zkontrolujte jeho řezné drážky, irigujte, zopakujte postup buď pilníkem velikosti 10, nebo přípravkem Endo Activator, abyste rozdrobili nečistoty, a poté znovu irigujte.
13) Po vyjmutí pilníku zkontrolujte jeho řezné drážky, zda se nerozvinuly, nenarovnaly nebo neroztáhly. Pokud si všimnete deformací, pilník zlikvidujte a použijte nový pilníkProTaper
Ultimate™.
14) Preparace je dokončena, když je apikální část kteréhokoli Finishera naplněna nečistotami a na pracovní délku je možné umístit GPMC příslušné velikosti nebo nástroj Size Verifier.
Pomocné Finishery používejte pouze u větších a rovnějších kanálků, jako jsou horní střední řezáky, některé palatální nebo distální kanálky stoliček nebo při patologickém nebo iatrogenním defektu:
15) Po zjištění pracovní délky a potvrzení průchodnosti zvolte buď mechanicky poháněný nebo ruční pomocný pilník FINISHER FX (035.12). Tento samostatný pilník použijte přímo a pasivně ke sledování kanálku do pracovní délky pomocí stejné techniky, jak je popsáno výše.
Vyjměte a zkontrolujte jeho apikální drážky. Po naplnění dentinovými zbytky je preparace dokončena.
16) Pokud je FINISHER FX v pracovní délce volný a jeho apikální drážky nejsou zatíženy nečistotami, zvolte buď mechanicky poháněný, nebo ruční pomocný FINISHER FXL (050.10) a použijte tento pilník stejným způsobem, jaký je popsán výše pro FINISHER FX.
17) Preparace je dokončena, když je apikální rozsah libovolného pomocného Finishera naplněn nečistotami a GPMC nebo nástroj Size Verifier se vejde o pracovní délky.
● Při dezinfekci, čištění a sterilizaci postupujte podle obecných pokynů pro zpracování endodontických výrobků https://www.dentsplysirona.com/en/service-contact/ifu.html
.
● Silikonové zarážky nepoužívejte opakovaně. Po každém použití silikonové zarážky odstraňte a zlikvidujte.
● Výrobky se likvidují v souladu s místními předpisy pro bezpečnou likvidaci ostrých a kontaminovaných zařízení.
● Jakýkoli závažný incident v souvislosti s výrobkem by měl být nahlášen výrobci a příslušnému orgánu v souladu s místními předpisy.
● Sterilitu nelze zaručit, pokud je obal otevřený, poškozený nebo vlhký.
● Chcete-li získat bezplatnou tištěnou kopii uživatelské příručky, přečtěte si část "Získání tištěné kopie uživatelské příručky" na webových stránkách https://www.dentsplysirona.com/ en/service-contact/ifu.html
.
● Vysvětlení neharmonizovaných symbolů v návodu k použití a na štítcích naleznete v části
Symboly v návodu k použití ( https://www.dentsplysirona.com/en/service-contact/ifu/ifusymbols.html
).
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 91/151
2797
Autorizovaný zástupce
DENTSPLY DeTrey GmbH
De-Trey-Straße 1
D-78467 Konstanz
Germany
Výrobce
Maillefer Instruments Holding Sàrl
Chemin du Verger 3
CH-1338 Ballaigues
Suisse www.dentsplysirona.com/ifu
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 92/151
ProTaper Ultimate™
LEN NA DENTÁLNE POUŽITIE
LEN NA PROFESIONÁLNE POUŽITIE
STERILNÉ – STERILIZOVANÉ OŽIARENÍM
VHODNÉ NA OPÄTOVNÉ SPRACOVANIE
OPAKOVANE POUŽITEĽNÉ STERILNÉ ENDODONTICKÉ NÁSTROJE
● Produkty sú určené na liečbu ochorení zubnej drene a tkanív zubných koreňov.
● Určený účel: Motorom poháňaný nástroj určený na preparáciu koreňového kanálika (dráha posúvania, tvarovanie a čistenie koreňového kanálika).
● Endodontické nástroje sa smú používať iba v klinickom alebo nemocničnom prostredí v súlade s osvedčenými zubárskymi postupmi, a to kvalifikovanými stomatologickými odborníkmi, ako sú praktickí lekári a odborníci v oblasti endodoncie (endodontisti) a zubní asistenti.
● Nástroje sa musia používať v kombinácii s endodontickým motorom (manuálne v prípade výrazných zakrivení).
Žiadne.
Nástroje ProTaper Ultimate™ sú vyrobené z troch hlavných komponentov: pracovná časť je vyrobená zo zliatiny niklu a titánu, farebnej silikónovej zarážky (s výnimkou nástroja SX) a pokovaného mosadzného drieku s farebným krúžkom (s výnimkou nástroja SX).
Pilníky sa vyrábajú procesom, ktorého výsledkom je pilník s farebným vzhľadom.
Z dôvodu tohto špeciálneho spracovania môžu byť pilníky mierne zakrivené. Nejde o výrobnú chybu. Aj keď pilník môžete ľahko narovnať prstami, pilník sa pred použitím nemusí narovnať.
Po vložení do kanálika bude pilník sledovať anatómiu koreňového kanálika.
Produktový rad ProTaper Ultimate™ pozostáva z nasledujúcich endodontických nástrojov:
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 93/151
Kód xx – dĺžka y – veľkosť pilníka z – množstvo v blistrovom balení
Produkt
ISO XXX YYYv
XXX – veľkosť ISO
YYY – veľkosť kužeľovitej časti nástroja v – variabilná veľkosť kužeľovitej časti nástroja
Dostupné dĺžky
Balenie a obsah
B ST PULR z
19 0SX
B ST PULR z xx 0SL
B ST PULR z xx 0SH
Nástroj ProTaper
Ultimate™ SX
Nástroj ProTaper
Ultimate™ SLIDER
Nástroj ProTaper
Ultimate™ SHAPER
ISO 020.003v
ISO 016.002v
ISO 020.004v
19 mm
21/25/31 mm
21/25/31 mm
Blistrové balenie so 6 nástrojmi (sterilné)
Blistrové balenie s 3 nástrojmi (sterilné)
Blistrové balenie so 6 nástrojmi (sterilné)
Blistrové balenie s 3 nástrojmi (sterilné)
Blistrové balenie so 6 nástrojmi (sterilné)
Blistrové balenie s 3 nástrojmi (sterilné)
Blistrové balenie so 6 nástrojmi (sterilné)
Blistrové balenie s 3 nástrojmi (sterilné)
B ST PULR z xx 0F y
B ST PULR 6 xx yyy
B ST PULR 5 xx SQE
B DD PULR
525 SQE
Nástroj ProTaper
Ultimate™ FINISHER
Nástroj ProTaper
Ultimate™ FINISHER
Séria nástrojov
ProTaper Ultimate™
Demonštračná súprava nástrojov ProTaper
Ultimate™
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
FX: ISO 035.012v
FXL: ISO 050.010v
Nástroj SLIDER:
ISO 016.002v
Nástroj SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
Nástroj SLIDER:
ISO 016.002v
Nástroj SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
21/25/31 mm
21/25/31 mm
21/25/31 mm
25 mm
Blistrové balenie so 6 nástrojmi (sterilné)
Blistrové balenie s 1 nástrojom každého typu (sterilné)
Blistrové balenie s 1 nástrojom každého typu (sterilné)
● Optimálne použitie zabezpečia zariadenia na reguláciu krútiaceho momentu.
● Používajte s endodontickými motormi pri konštantnom otáčaní rýchlosťou 400 ot./min.
● Nastavte krútiaci moment na: 4 to 5.2 Ncm.
● Používajú sa mazivá, ako napríklad NaOCl, EDTA, ProLube a Glyde™.
● Vhodnou metódou na stanovenie pracovnej dĺžky je použitie röntgenových snímok v kombinácii s apexlokátorom a nástrojom na nastavenie silikónovej zarážky na správnu pracovnú dĺžku.
● Dôsledne dodržujte tento návod na použitie a všeobecné pokyny na spracovanie pre endodontické výrobky (pozri časť 8), aby sa minimalizovali nasledujúce riziká týkajúce sa pomôcky, pacienta a/alebo používateľa:
○ Zlomenie nástroja.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 94/151
○ Krížová kontaminácia.
○ Tvorba tepla v dôsledku nedostatočného mazania a irigácie.
○ Prehltnutie pracovnej časti nástroja.
○ Toxické alebo alergické reakcie spôsobené zvyškami po spracovaní.
● Podobne ako všetky strojovo poháňané nástroje na koreňové kanáliky, nástroje ProTaper
Ultimate™ sa nesmú použiť v koreňovom kanáliku s náhlym apikálnym zakrivením z dôvodu zvýšeného rizika oddelenia. V tomto prípade sa v apikálnej oblasti musia použiť vopred zakrivené ručné pilníky.
● Bezpečnosť a účinnosť používania nebola stanovená pre tehotné alebo dojčiace ženy ani pre deti.
● Počas endodontického zákroku používajte kofferdamový systém.
● Z dôvodu vlastnej bezpečnosti používajte osobné ochranné prostriedky (rukavice, okuliare, masku).
● Pred použitím skontrolujte obal a ak je poškodený, nástroje nepoužívajte.
● Nástroje nepoužívajte po dátume exspirácie.
● Nástroj pred každým použitím skontrolujte, či nevykazuje známky poškodenia, ako sú napríklad deformácie (ohnutie, odvinutie), zlomenie, poškodené rezné hrany, zmena farby alebo strata označenia. Ak zistíte tieto príznaky, pomôcky nie sú schopné plniť svoj zamýšľaný účel s požadovanou úrovňou bezpečnosti, takže sa musia zlikvidovať.
● Pred použitím akéhokoľvek nástroja sa uistite, že je dobre pripojený k hlave nadstavca.
● Počas prípravy nástrojov nástroj často kontrolujte a čistite pracovnú časť, skontrolujte, či nástroj nevykazuje známky deformácie, vychýlenia alebo opotrebenia, ako sú rozdielne drážky a matné škvrny. Ak zistíte tieto príznaky, pomôcky nie sú schopné plniť svoj zamýšľaný účel s požadovanou úrovňou bezpečnosti, takže sa musia zlikvidovať.
● Nástroje sa nesmú úplne ponárať do roztoku chlórnanu sodného (NaOCl). Iba pracovná časť nástroja zo zliatiny niklu a titánu (Ni-Ti), ktorá je v kontakte s pacientom, sa smie ponoriť do koncentrátu roztoku NaOCl v koncentrácii NIE VIAC ako 5 %.
● Postupujte opatrne v apikálnej oblasti a kanálikoch, ktoré rozdeľujú a/alebo vykazujú náhle zakrivenia alebo opätovné zakrivenia.
● Počas celého zákroku a po každom použití nástroja kanálik často a dôkladne irigujte (podľa osvedčených zubárskych postupov).
● Vždy používajte minimálny apikálny tlak. Pilníky nikdy netlačte dolu kanálikom.
● Ak nástroj nepostupuje ľahko, vyčistite a skontrolujte rezné drážky, následne irigujte, zopakujte postup pomocou ručného pilníka a následne znovu irigujte.
● Na tvarovanie extrémne zakrivených kanálikov je bezpečnejšie použiť pilník iba na tvarovanie jedného kanálika, aby sa znížilo riziko zlomenia. Venujte pozornosť nasledujúcim osvedčeným postupom:
○ Použite nový pilník a po ošetrení kanálika ho zlikvidujte (len na použitie v jednom kanáliku).
○ Používajte ručné, nie rotačné pilníky.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 95/151
○ Použite pilníky malej veľkosti, pružné pilníky a/alebo pilníky NiTi (tým sa pomôže zabrániť presunu kanálika).
○ Počas používania vizuálne kontrolujte pracovnú časť, či nevykazuje akékoľvek chyby uvedené v predchádzajúcom odseku (t. j. po každej vlne).
○ Vyhýbajte sa štandardnému vytrvalému otáčavému pohybu a namiesto toho používajte pohyby v malých uhloch (pohyb pri pilníkovaní, otáčavý pohyb tam a späť alebo technika vyrovnanej sily), aby ste obmedzili únavu z ohybu na nástrojoch a zlepšili ich očakávanú životnosť.
Na základe súčasného technického stavu neboli hlásené žiadne nežiaduce reakcie.
Produkty skladujte v suchom a čistom prostredí, mimo zdrojov vlhkosti a priameho slnečného žiarenia.
Všetky pilníky ProTaper Ultimate™ boli validované na použitie pri 400 ot./min a krútiacom momente 4 to 5.2 Ncm.
Nástroje ProTaper Ultimate™ sa môžu používať otieraním smerom von vo všetkých kanálikoch, najmä v tých, ktoré majú nepravidelný prierez, alebo jemným dlabaním smerom dovnútra, aby sa postupne posúvali po pracovnej dĺžke.
Rukoväť vždy držte v časti medzi palcom a ukazovákom. Vyhnite sa tlačeniu, namiesto toho nechajte pilníky pasívne postupovať a sledovať dráhu posúvania.
1) Pozrite si röntgenové snímky nasnímané pod rôznym horizontálnym uhlom, aby ste diagnosticky stanovili šírku, dĺžku a zakrivenie koreňa a jeho kanálika(-ov).
2) Pripravte prístup, ktorý umožňuje ľahkú identifikáciu každého ústia kanálika.
3) Pomocný tvarovací pilník SX sa môže použiť v prípade obmedzeného miesta alebo na predbežné zväčšenie tela kanálika, prípadne na premiestenie najkoronálnejšej časti kanálika od vonkajšej konkávnej časti koreňa.
4) Ak je prítomné mazivo, vyberte nástroj Slider a PASÍVNE sledujte kanálik jedným alebo viacerými prechodmi nástroja až po koniec kanálika. Stanovte pracovnú dĺžku pomocou elektronického apexlokátora v kombinácii s röntgenovou snímkou a následne potvrďte priechodnosť.
5) Ak nástroj Slider ľahko nedosiahne koniec kanálika, vyberte malý ručný pilník z nehrdzavejúcej ocele (SS). Za prítomnosti maziva katetrizujte tento kanálik, stanovte pracovnú dĺžku, potvrďte priechodnosť a overte dráhu posúvania. Teraz zopakujte vyššie uvedený krok č.
4.
6) Za prítomnosti NaOCl vyberte nástroj Shaper a posúvajte ho po dráhe posúvania jedným alebo viacerými prechodmi, kým sa nedosiahne pracovná dĺžka.
7) Po odstránení nástroja Shaper irigujte, pomocou EndoActivate rozbite zvyšky a presuňte ich do roztoku, následne znovu irigujte, aby sa tieto zvyšky uvoľnili.
8) Znovu potvrďte pracovnú dĺžku, najmä vo viac zakrivených kanálikoch.
9) Vyberte nástroj FINISHER F1 (020.07) a pasívne sledujte kanálik až po pracovnú dĺžku jedným alebo viacerými prechodmi nástroja. Vyberte nástroj a skontrolujte jeho apikálne drážky. Ak sú naplnené zvyškami dentínu, preparácia je dokončená.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 96/151
10) Ak je nástroj FINISHER F1 v pracovnej dĺžke voľný a jeho apikálne drážky nie sú naplnené zvyškami, vyberte nástroj FINISHER F2 (025.08) a tento pilník použite rovnakým spôsobom, ako je to opísané vyššie pre FINISHER F1.
11) Ak je nástroj FINISHER F2 v pracovnej dĺžke voľný a jeho apikálne drážky nie sú naplnené zvyškami, vyberte nástroj FINISHER F3 (030.09) a tento pilník použite rovnakým spôsobom, ako je to opísané vyššie pre nástroj FINISHER F1 a F2.
12) Po odstránení ktoréhokoľvek z uvedených pilníkov vyčistite a skontrolujte jeho rezné drážky, irigujte, zopakujte postup buď pomocou pilníka veľkosti 10, alebo Endo Activator, aby ste rozdrobili zvyšky. Následne znovu irigujte.
13) Po vybratí skontrolujte rezné drážky pilníka, či nie sú odvinuté, narovnané alebo natiahnuté. Ak si všimnete deformáciu, pilník zlikvidujte a použite nový pilník ProTaper
Ultimate™.
14) Preparácia je dokončená vtedy, keď je apikálna časť ktoréhokoľvek nástroja Finisher naplnená zvyškami a do pracovnej dĺžky je možné umiestniť hlavný gutaperčový čap (GPMC) príslušnej veľkosti alebo nástroj Size Verifier.
Pomocné nástroje Finisher používajte iba vo väčších a rovnejších kanálikoch, ako sú horné stredné rezáky, niektoré palatálne alebo distálne kanáliky stoličiek alebo v prípade patologického alebo iatrogénneho defektu:
15) Po stanovení pracovnej dĺžky a potvrdení priechodnosti vyberte buď mechanicky poháňaný, alebo ručný pomocný pilník, FINISHER FX (035.12). Tento samostatný pilník použite priamo a pasívne sledujte kanálik po pracovnú dĺžku pomocou rovnakej techniky, ako je to opísané vyššie. Vyberte nástroj a skontrolujte jeho apikálne drážky. Ak sú naplnené zvyškami dentínu, preparácia je dokončená.
16) Ak je nástroj FINISHER FX v pracovnej dĺžke voľný a jeho apikálne drážky nie sú naplnené zvyškami, vyberte buď mechanicky poháňaný, alebo ručný pomocný nástroj FINISHER FX
(050.10) a tento pilník použite rovnakým spôsobom, ako je to opísané vyššie pre FINIS-
HER FX.
17) Preparácia je dokončená vtedy, keď je apikálna časť ktoréhokoľvek pomocného nástroja
Finisher naplnená zvyškami a do pracovnej dĺžky je možné umiestniť hlavný gutaperčový čap (GPMC) príslušnej veľkosti alebo nástroj Size Verifier.
● Pri dezinfekcii, čistení a sterilizácii postupujte podľa všeobecných pokynov na spracovanie pre endodontické výrobky ( https://www.dentsplysirona.com/en/service-contact/ifu.html
).
● Silikónové zarážky nepoužívajte opakovane. Silikónové zarážky odstráňte a zlikvidujte po každom použití.
● Produkty je potrebné likvidovať v súlade s miestnymi nariadeniami na bezpečnú likvidáciu ostrých alebo kontaminovaných pomôcok.
● Akýkoľvek závažný incident súvisiaci s produktom sa musí nahlásiť výrobcovi a príslušnému orgánu v súlade s miestnymi predpismi.
● Sterilitu nemožno zaručiť, ak je obal otvorený, poškodený alebo vlhký.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 97/151
● Ak chcete získať bezplatnú tlačenú kópiu návodu na použitie, prečítajte si časť „Získanie tlačenej kópie návodu na použitie“ na webovej stránke https://www.dentsplysirona.com/en/ service-contact/ifu.html
.
● Vysvetlenie neharmonizovaných symbolov v návodoch na použitie a na štítkoch nájdete v časti Symboly v návode na použitie ( https://www.dentsplysirona.com/en/service-contact/ifu/ ifu-symbols.html
).
2797
Autorizovaný zástupca
DENTSPLY DeTrey GmbH
De-Trey-Straße 1
D-78467 Konstanz
Germany
Výrobca
Maillefer Instruments Holding Sàrl
Chemin du Verger 3
CH-1338 Ballaigues
Suisse www.dentsplysirona.com/ifu
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 98/151
ProTaper Ultimate™
KIZÁRÓLAG FOGÁSZATI FELHASZNÁLÁSRA
KIZÁRÓLAG PROFESSZIONÁLIS FELHASZNÁLÁSRA
STERIL – BESUGÁRZÁSSAL STERILIZÁLT
REGENERÁLÁSRA ALKALMAS
ÚJRAFELHASZNÁLHATÓ STERIL GYÖKÉRKEZELÉSI MŰSZEREK
● A termékek foggyökér betegségek kezelésére javallottak.
● Tervezett használat: Motoros műszer a gyökércsatorna előkészítéséhez (vezetőcsatorna, a gyökércsatorna formázása és sebtisztítása).
● A gyökérkezelési műszereket kizárólag szakképzett fogászati szakemberek, azaz általános fogorvosok, valamint endo-specialisták (endodontológusok) és fogászati szakasszisztensek használhatják klinikai vagy kórházi környezetben a helyes fogászati gyakorlat követésével.
● A műszereket gyökérkezelési motorral együtt kell használni (manuálisan az erős görbületekben)
Nincs.
A ProTaper Ultimate™ műszerek három fő komponensből állnak: a nikkel-titán ötvözetből készült működő rész, színes szilikon záródugó (az SX műszer kivételével) és színes gyűrűvel ellátott sárgarézzel bevont szár (az SX műszer kivételével).
A reszelő gyártása olyan technológiával történik, amelynek eredményeként a reszelő színes külsőt kap.
Ennek a védett eljárásnak köszönhetően reszelők enyhén görbének tűnhetnek. Ez nem gyártási hiba. Ujjainak használatával ugyan könnyen kiegyenesítheti a reszelőt, nem szükséges használat előtt kiegyenesíteni. Amint a csatornába jutott, a reszelő követni fogja a gyökércsatorna anatómiáját.
A ProTaper Ultimate™ termékcsoport a következő gyökérkezelési műszerekből áll:
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 99/151
Kód xx-hossz y-reszelő mérete z-mennyiség buborékcsomagolásonként
B ST PULR z
19 0SX
B ST PULR z xx 0SL
B ST PULR z xx 0SH
Termék
ProTaper Ultimate™
SX
ProTaper Ultimate™
SLIDER
ProTaper Ultimate™
SHAPER
ISO XXX YYYv
XXX-ISO méret
YYY-kúpméret v-változó kúp
ISO 020.003v
ISO 016.002v
ISO 020.004v
Elérhető hosszak
19 mm
21/25/31 mm
21/25/31 mm
Csomagolás és tartalom
Buborékcsomagolás 6 műszerrel (steril)
Buborékcsomagolás 3 műszerrel (steril)
Buborékcsomagolás 6 műszerrel (steril)
Buborékcsomagolás 3 műszerrel (steril)
Buborékcsomagolás 6 műszerrel (steril)
Buborékcsomagolás 3 műszerrel (steril)
Buborékcsomagolás 6 műszerrel (steril)
B ST PULR z xx 0F y
B ST PULR 6 xx yyy
B ST PULR 5 xx SQE
B DD PULR
525 SQE
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
Sequence
ProTaper Ultimate™
Demo Pack
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
FX: ISO 035.012v
FXL: ISO 050.010v
FOGTÜKÖR:
ISO 016.002v
FORMÁZÓ:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
FOGTÜKÖR:
ISO 016.002v
FORMÁZÓ:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
21/25/31 mm
21/25/31 mm
21/25/31 mm
25 mm
Buborékcsomagolás 3 műszerrel (steril)
Buborékcsomagolás 6 műszerrel (steril)
Buborékcsomagolás műszerenként 1 darabbal (steril)
Buborékcsomagolás műszerenként 1 darabbal (steril)
● A nyomatékszabályozó eszközök optimális használatot biztosítanak.
● Állandó, 400 rpm fordulatszámon használja a gyökérkezelési motorokat.
● Nyomaték beállítása: 4 to 5.2 Ncm.
● Kenőanyag, például NaOCl, EDTA, ProLube, Glyde™ használata szükséges.
● A munkahossz meghatározásának megfelelő módszere az, ha radiográfiai feltételeket használ apex lokátorral és a szilikon záródugó megfelelő munkahossz szerinti beállítására szolgáló eszközzel együtt.
● Szigorúan tartsa be a Használati utasítás és az Általános előkészítési utasítások gyökérkezelési termékek esetében dokumentumokat (lásd a 8. bekezdés), hogy minimálisra csökkentse az eszközt, a beteget és/vagy a felhasználót érő kockázatokat:
○ Műszer törése.
○ Keresztszennyeződés.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 100/151
○ Hőfejlődés elégtelen kenés és öblögetés miatt.
○ A műszer munkavégzési részének lenyelése.
○ Előkészítési maradványok által okozott toxikus vagy allergiás reakciók.
● Más mechanikus hajtású gyökércsatorna-kezelő műszerekhez hasonlóan a ProTaper
Ultimate™ műszereket sem szabad használni olyan esetekben, amikor a foggyökérben hirtelen apikális görbület található, hiszen ilyenkor fokozottan fennáll a leválás kockázata. Ebben az esetben előgörbített kézi reszelőket kell használni az apikális régióban.
● A felhasználás biztonságosságát és hatásosságát nem tesztelték sem várandós nőkön, sem szoptató nőkön, sem gyermekeken.
● Használjon nyálrekesz rendszert a gyökérkezelési eljárás alatt.
● Saját biztonsága érdekében viseljen egyéni védőfelszereléseket (védőkesztyű, védőszem-
üveg, maszk).
● Használat előtt ellenőrizze a csomagolást, és ne használja a műszereket, ha a csomagolás sérült.
● Ne használja a műszereket a lejárati dátum után.
● Minden egyes használat előtt ellenőrizze a műszeren a hibák jeleit, például a deformálódást
(meghajlás, megtekeredés), törést, korróziót, sérült vágóéleket, színkódolás vagy jelölés eltűnését. Ha ilyen jelek láthatók, az eszközök nem használhatók a rendeltetési céljuknak megfelelően a szükséges biztonsági szinten, és a műszereket ki kell dobni.
● Mielőtt bármilyen műszert használ, győződjön meg róla, hogy megfelelően csatlakozik-e a kézidarab fejhez.
● Gyakran ellenőrizze a műszert, és tisztítsa meg a munkavégzési részt; eközben vizsgálja meg a torzulás, megnyúlás vagy kopás jeleit, például az egyenetlen hornyokat, életlen pontokat. Ha ilyen jelek láthatók, az eszközök nem használhatók a rendeltetési céljuknak megfelelően a szükséges biztonsági szinten, és a műszereket ki kell dobni.
● A műszereket tilos nátrium-hipoklorit oldatba (NaOCl) teljesen alámeríteni. Az NiTi műszernek csak a beteggel érintkező munkavégzési részét kell NaOCl oldatba meríteni, aminek a koncentrációja NEM lehet 5%-nál nagyobb.
● Óvatosan járjon el az apikális területen és azokban a csatornákban, amelyek szétválnak,
és/vagy amelyekben hirtelen görbületek vagy visszagörbülések vannak.
● Bőségesen és gyakran öblítse a csatornát az eljárás alatt és az egyes műszerhasználatok után (a helyes fogászati gyakorlat szerint).
● Mindig csak minimális apikális nyomást fejtsen ki. Soha ne próbálja erővel letolni a reszelőket a csatornában.
● Amikor a műszer nem halad könnyedén előre, tisztítsa meg és ellenőrizze a maróhornyokat, majd öblítsen, rekapituláljon kézi reszelővel, és végezzen újra öblítést.
● Extrém mértékben ívelt csatornák alakításánál a biztonságos megoldás az, ha egy reszelőt csak egyetlen csatornához használ, mivel ez csökkenti a törés kockázatát. Ügyeljen az alábbi jó gyakorlatok követésére:
○ Használjon új reszelőt, és a csatorna kezelése után dobja ki (egycsatornás használat).
○ Forgó reszelők helyett kézi reszelőt használjon.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 101/151
○ Kis méretű, rugalmas és/vagy NiTi reszelőket használjon (ez segít megelőzni a csatorna eltolását).
○ Használat közben szemrevételezéssel ellenőrizze a munkavégzési részt a fenti bekezdésben felsorolt valamennyi sérülés tekintetében (tehát minden egyes hullám után).
○ Kerülje a folyamatos tágításhoz vezető körkörös mozgatást, ehelyett a mozgatás során kis íveket írjon le (töltő mozgás, órafelhúzásszerű szakaszos mozgás, kiegyensúlyozott erőt alkalmazó technika) annak érdekében, hogy a műszer forgás miatti hajlító fáradása minimálisra csökkenjen és ezzel az élettartama megnőjön.
A technológia jelen állása szerint eddig senki nem jelentett nemkívánatos reakciót.
Tiszta és száraz környezetben, nedvességforrásoktól és közvetlen napsugárzástól védetten tárolja a termékeket.
Valamennyi ProTaper Ultimate™ reszelő érvényesítve lett 400 rpm fordulatszámon és 4 to
5.2 Ncm nyomatékon történő használathoz.
A ProTaper Ultimate™ műszerek kifelé végzett kefélési mozgással használhatók az összes csatornában, kivéve azokat a csatornákat, amelyek keresztmetszete szabálytalan, illetve enyhe, befelé végzett petyegtető (‘tupfoló’) mozdulattal, amely révén progresszíven haladhat előre a munkavégzési hosszúságban (WL).
Mindig a hüvelykujja és a mutatóujjak között tartsa a kézidarabot. Kerülje a tolást; ehelyett hagyja passzívan haladni a reszelőket, és kövesse a bevezetési útvonalat.
1) A különböző, vízszintes szögben készített radiográfiai felvételek áttekintésével diagnosztikailag határozza meg az adott gyökér és csatorna (csatornák) szélességét, hosszúságát
és görbületét.
2) Olyan hozzáférést biztosítson, ami lehetővé teszi az egyes csatornanyílások egyszerű azonosítását.
3) Az SX kiegészítő formázó reszelő használható, ha szűk a hely, vagy segítségével a csatornatest előtágítása végezhető, vagy a külső gyökér konkavitásától távol álló csatorna szélső koronális helyzete módosítható.
4) Kenőanyag jelenlétében válassza ki a fogtükröt, és PASSZÍVAN kövesse a csatornát egy vagy több menetben a terminusig. Elektronikus apex lokátor és radiográfiás feltétel együttes használatával állítsa meg a munkavégzési hosszúságot, majd győződjön meg az átjárhatóságról.
5) Ha a fogtükörrel nem éri el a csatorna terminusát, válasszon kisebb méretű kézi rozsdamentes acél reszelőt. Kenőanyag jelenlétében katéterezze be ezt a csatornát, állapítsa meg a munkavégzési hosszúságot, és ellenőrizze a bevezetési útvonalat. Ezután ismételje meg a fenti 4. lépést.
6) NaOCl jelenlétében válassza ki a formázót, és tolja előre a bevezetési útvonalon egy vagy több menetben a munkavégzési hossz elérésig.
7) A formázó eltávolítása után öblítsen, EndoActivate használatával lazítsa fel a törmeléket, hogy az oldatba kerüljön, majd öblítsen újra a törmelék felszabadításához.
8) Ellenőrizze újra a munkavégzési hosszúságot, különösen a görbe csatornákban.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 102/151
9) Válassza ki az F1 FINÍROZÓ (020.07) eszközt, és passzívan kövesse a csatornát a munkavégzési hosszig egy vagy több menetben. Távolítsa el, és ellenőrizze az apikális hornyokat. Ha fogbéltörmelékkel van eltelítve, az előkészítés befejeződött.
10) Ha az F1 FINÍROZÓ végül kiszabadul, és az apikális hornyai nem telítődtek törmelékkel, válassza ki az F2 FINÍROZÓ (025.08) eszközt, és az F1 FINÍROZÓ esetében fentiekben leírt módon használja ezt a reszelőt is.
11) Ha az F2 FINÍROZÓ végül kiszabadul, és az apikális hornyai nem telítődtek törmelékkel, válassza ki az F3 FINÍROZÓ (030-09) eszközt, és az F1 és F2 FINÍROZÓ esetében fentiekben leírt módon használja ezt a reszelőt is.
12) Bármelyik reszelő eltávolításakor tisztítsa meg és ellenőrizze a maró hornyokat, öblítsen, rekapituláljon 10-es méretű reszelővel vagy Endo Activator eszközzel a törmelék feloldásához, majd öblítsen újra.
13) Eltávolítás után ellenőrizze a reszelő maró hornyain a megtekeredés, kiegyenesedés vagy megnyúlás jelenlétét. Ha deformálódást észlel, dobja ki, és használjon új ProTaper
Ultimate™ reszelőt.
14) Az előkészítés akkor fejeződik be, amikor az adott finírozó apikális hossza megtelik törmelékkel, és a megfelelően méretezett mester guttapercha csúcs vagy méretellenőrző illeszkedik a munkavégzési hosszúságba.
A kiegészítő finírozók kizárólag nagyobb és egyenesebb csatornákban, például a felső nagymetszőfogak, az őrlőfogak egyes palatális vagy disztális csatornái esetében vagy akkor használhatók, ha kóros vagy iatrogén rendellenesség áll fenn.
15) Válasszon FX FINÍROZÓ mechanikus meghajtású vagy kézi kiegészítő reszelőt (035.12) amikor a munkavégzési hosszúságot létrehozta, és meggyőződött az átjárhatóságról. Ennek a különálló reszelőnek a használatával közvetlenül és passzívan kövesse a csatornát a munkavégzési hosszúságon a fentiekben leírt technika alkalmazásával. Távolítsa el, és ellenőrizze az apikális hornyokat. Ha fogbéltörmelékkel van eltelítve, az előkészítés befejeződött.
16) Ha az FX FINÍROZÓ végül kiszabadul, és az apikális hornyai nem telítődtek törmelékkel, válasszon mechanikus meghajtású vagy kézi kiegészítő FXL FINÍROZÓ (050.10) eszközt,
és az FX FINÍROZÓ esetében fentiekben leírt módon használja ezt a reszelőt is.
17) Az előkészítés akkor fejeződik be, amikor bármelyik kiegészítő finírozó apikális hossza megtelik törmelékkel, és a megfelelően méretezett mester guttapercha csúcs vagy méretellenőrző illeszkedik a munkavégzési hosszúságba.
● Kövesse az Általános előkészítési utasítások gyökérkezelési termékek esetében ( https:// www.dentsplysirona.com/en/service-contact/ifu.html
) dokumentumot a fertőtlenítési, tisztítási és sterilizálási lépésekre vonatkozóan.
● Ne használja fel újra a szilikon záródugókat. Minden egyes használat után vegye ki és dobja ki a szilikon záródugókat.
● A termékeket az éles és fertőzött eszközök biztonságos selejtezésére vonatkozó helyi szabályozásnak megfelelően kell selejtezni.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 103/151
● A termékkel összefüggésben előforduló minden súlyos esetet jelenteni kell a gyártó és az illetékes hatóság felé a helyi előírások szerint.
● A sterilitás nincs garantálva, ha csomagolás nyitott, sérült vagy nedves.
● Ha a Használati utasítás ingyenes, nyomtatott példányára van szüksége, lásd a https:// www.dentsplysirona.com/en/service-contact/ifu.html
weboldaon a „Get a print copy of IFU” részt.
● Nem harmonizált szimbólumok magyarázata használati utasításokhoz és címkékhez kapcsolódóan, lásd a használati utasítás szimbólumokat ( https://www.dentsplysirona.com/en/ service-contact/ifu/ifu-symbols.html
).
2797
Megbízott képviselő
DENTSPLY DeTrey GmbH
De-Trey-Straße 1
D-78467 Konstanz
Germany
Gyártó
Maillefer Instruments Holding Sàrl
Chemin du Verger 3
CH-1338 Ballaigues
Suisse www.dentsplysirona.com/ifu
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 104/151
ProTaper Ultimate™
EXCLUSIV PENTRU UZ DENTAR
EXCLUSIV PENTRU UZ PROFESIONAL
STERIL - STERILIZAT CU RADIAŢII
SUSCEPTIBIL DE REPROCESARE
INSTRUMENTE ENDODONTICE STERILE REUTILIZABILE
● Produsele sunt indicate pentru tratarea bolii endodontice.
● Utilizarea prevăzută: Instrument acţionat mecanic destinat pregătirii canalului radicular (traiectoria de coborâre, modelarea şi debridarea canalului radicular).
● Instrumentele endodontice trebuie utilizate exclusiv într-un mediu clinic sau spitalicesc, cu respectarea bunelor practici stomatologice, de către profesionişti calificaţi cum ar fi generaliştii şi specialiştii în endodonţie (endodontist) şi asistenţii stomatologi.
● Instrumentele se vor utiliza în combinaţie cu un motor endodontic (manual în cazul curburilor severe).
Niciuna.
Instrumentele ProTaper Ultimate™ se compun din trei componente principale: piesa de lucru fabricată dintr-un aliaj de nichel şi titan, un opritor din silicon colorat (cu excepţia instrumentului
SX) şi o coadă din alamă placată cu un inel colorat (cu excepţia instrumentului SX).
Acele sunt fabricate printr-un proces care conferă acului un aspect colorat.
Datorită acestei prelucrări brevetate, acele pot părea uşor curbate. Acesta nu este un defect de fabricaţie. Cu toate că acul poate fi îndreptat uşor cu degetele, nu este necesară îndreptarea acului înainte de utilizare. Odată introdus în canal, acul va urma anatomia canalului radicular.
Gama ProTaper Ultimate™ cuprinde următoarele instrumente endodontice:
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 105/151
Cod xx-lungime y-mărime ac z-cantitate în blister
B ST PULR z
19 0SX
B ST PULR z xx 0SL
B ST PULR z xx 0SH
Produs
ProTaper Ultimate™
SX
ProTaper Ultimate™
SLIDER
ProTaper Ultimate™
SHAPER
ISO XXX YYYv
XXX-mărime ISO
YYY-conicitate v-conicitate variabilă
ISO 020.003v
ISO 016.002v
ISO 020.004v
Lungimi disponibile
19 mm
21/25/31 mm
21/25/31 mm
Ambalaj şi conţinut
Blister cu 6 instrumente (sterile)
Blister cu 3 instrumente (sterile)
Blister cu 6 instrumente (sterile)
Blister cu 3 instrumente (sterile)
Blister cu 6 instrumente (sterile)
Blister cu 3 instrumente (sterile)
Blister cu 6 instrumente (sterile)
B ST PULR z xx 0F y
B ST PULR 6 xx yyy
B ST PULR 5 xx SQE
B DD PULR
525 SQE
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
Sequence
ProTaper Ultimate™
Demo Pack
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
FX: ISO 035.012v
FXL: ISO 050.010v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
21/25/31 mm
21/25/31 mm
21/25/31 mm
25 mm
Blister cu 3 instrumente (sterile)
Blister cu 6 instrumente (sterile)
Blister cu 1 instrument din fiecare tip de instrument (steril)
Blister cu 1 instrument din fiecare tip de instrument (steril)
● Dispozitivele cu opţiune de comandă a cuplului asigură o utilizare optimă.
● A se utiliza cu motoare endodontice cu rotaţie constantă la o turaţie de 400 rpm.
● Stabiliţi cuplul la: 4 to 5.2 Ncm.
● Se vor utiliza lubrifianţi cum ar fi NaOCl, EDTA, ProLube, Glyde™.
● Metoda adecvată pentru stabilirea lungimii de lucru constă în utilizarea radiografiilor în combinaţie cu un dispozitiv pentru localizarea apexului şi un instrument pentru ajustarea opritorului din silicon la lungimea de lucru corectă.
● Urmaţi cu stricteţe prezentele Instrucţiuni de utilizare şi Instrucţiunile generale de prelucrare a produselor endodontice (a se vedea secţiunea 8) pentru a se reduce la minimum riscurile pentru dispozitiv, pacient şi/sau utilizator:
○ Ruperea instrumentului.
○ Contaminarea încrucişată.
○ Generarea de căldură din cauza lubrifierii şi irigării insuficiente.
○ Înghiţirea piesei de lucru a instrumentului.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 106/151
○ Reacţii toxice sau alergice provocate de reziduurile rezultate în urma prelucrării.
● La fel ca toate instrumentele de tratare a canalelor radiculare, acţionate mecanic, instrumentele ProTaper Ultimate™ nu trebuie utilizate într-un canal radicular cu o curbură apicală pronunţată din cauza riscului ridicat de separare. În acest caz, în zona apicală trebuie utilizate ace manuale precurbate.
● Siguranţa şi eficienţa utilizării nu au fost stabilite în cazul femeilor însărcinate sau care alăptează, ori în cazul copiilor.
● Utilizaţi o digă din cauciuc în timpul procedurii endodontice.
● Pentru siguranţa dumneavoastră, purtaţi echipament individual de protecţie (mănuşi, ochelari, mască).
● Inspectaţi ambalajul înainte de utilizare şi nu utilizaţi instrumentele dacă ambalajul este deteriorat.
● Nu utilizaţi instrumente al căror termen de valabilitate a expirat.
● Înainte de fiecare utilizare, verificaţi dacă instrumentul nu prezintă semne de deteriorare ca deformări (îndoire, desprindere), rupturi, urme de coroziune, muchii ascuţite deteriorate, pierderea codului de culoare sau a marcajelor. Dacă prezintă aceste semne, instrumentele nu îşi mai pot îndeplini funcţia prevăzută la nivelul de siguranţă necesar şi trebuie eliminate la deşeuri.
● Înainte de utilizarea oricărui instrument, asiguraţi-vă că acesta este bine conectat la capul piesei contraunghi.
● Verificaţi instrumentul şi curăţaţi piesa de lucru în timpul instrumentării, verificând dacă există semne de deformare, alungire sau uzură, cum ar fi caneluri neregulate, puncte tocite.
Dacă prezintă aceste semne, instrumentele nu îşi mai pot îndeplini funcţia prevăzută la nivelul de siguranţă necesar şi trebuie eliminate la deşeuri.
● Aceste instrumente nu trebuie scufundate complet într-o soluţie de hipoclorit de sodiu
(NaOCl). Numai piesa de lucru a instrumentului din NiTi, care intră în contact cu pacientul, trebuie scufundată într-o soluţie NaOCl cu o concentraţie de MAXIM 5%.
● Acordaţi o atenţie deosebită zonei apicale şi canalelor care separă şi/sau care prezintă canale curbate sau recurbate brusc.
● Irigaţi abundent şi frecvent canalul, pe parcursul procedurii şi după utilizarea fiecărui instrument (conform bunelor practici stomatologice).
● Utilizaţi întotdeauna presiunea apicală minimă. Nu forţaţi acele în jos pe canal.
● Dacă instrumentul nu avansează cu uşurinţă, curăţaţi şi inspectaţi canelurile ascuţite, după aceea irigaţi, reluaţi operaţia cu un ac manual şi reirigaţi.
● Pentru modelarea canalelor extrem de curbate, se recomandă să utilizaţi acul numai pentru a modela un singur canal în vederea reducerii riscului de rupere a acestuia. Aveţi grijă să respectaţi bunele practici:
○ Utilizaţi un ac nou şi eliminaţi-l după tratarea canalului (utilizare pe un singur canal).
○ Utilizaţi ace manuale în locul celor cu mişcare rotativă.
○ Utilizaţi ace de dimensiune mică, flexibile şi/sau din NiTi (acest lucru contribuie la evitarea deplasării canalului).
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 107/151
○ Inspectaţi vizual piesa de lucru pentru a detecta dacă în timpul utilizării a apărut vreuna dintre defecţiunile indicate în paragraful anterior (adică după fiecare mişcare de prelucrare a canalului).
○ Evitaţi mişcarea de rotaţie continuă standard pentru lărgire şi optaţi pentru mişcări mici în unghi (mişcare de umplere, mişcare oscilatorie în sens orar sau tehnică de forţare compensată) pentru a limita solicitarea instrumentelor prin îndoire din cauza mişcării rotative a acestora şi a optimiza durata lor de viaţă preconizată.
În starea tehnică actuală, nu s-a raportat nicio reacţie adversă.
Păstraţi produsele într-un mediu uscat şi curat, ferite de surse de umezeală şi de lumina directă a soarelui.
Toate acele ProTaper Ultimate™ au fost aprobate pentru utilizare la 400 rpm şi cuplu de 4 to
5.2 Ncm.
Instrumentele ProTaper Ultimate™ se pot utiliza cu o mişcare de periere spre exterior în toate canalele, în special în canalele care prezintă o secţiune transversală neregulată sau cu o mişcare de scobire uşoară spre interior pentru avansarea treptată până la lungimea de lucru.
Sprijiniţi întotdeauna mânerul în porţiunea dintre degetul mare şi index. Evitaţi împingerea; în schimb, lăsaţi acul să avanseze pasiv şi urmaţi traiectoria de alunecare.
1) Analizaţi diferite radiografii pe o direcţie unghiulară orizontală pentru a determina prin diagnosticare lăţimea, lungimea şi curbura rădăcinii şi a canalului(-elor) sale.
2) Pregătiţi un punct de acces care să permită identificarea uşoară a fiecărui orificiu de canal.
3) Dacă spaţiul este restrâns sau pentru lărgirea preliminară a corpului canalului, pentru repoziţionarea părţii celei mai coronale a unui canal departe de o concavitate radiculară externă, se poate utiliza acul de modelare auxiliar, SX.
4) În prezenţa unui lubrifiant, selectaţi instrumentul Slider şi urmaţi PASIV canalul în una sau mai multe treceri până la terminaţie. Stabiliţi lungimea de lucru utilizând un dispozitiv electronic pentru localizarea apexului în combinaţie cu o radiografie şi după aceea confirmaţi vizibilitatea.
5) Dacă instrumentul Slider nu atinge uşor terminaţia canalului, alegeţi un ac manual din oţel inoxidabil de mici dimensiuni. În prezenţa unui lubrifiant, cateterizaţi acest canal, stabiliţi lungimea de lucru, confirmaţi vizibilitatea şi verificaţi traseul de alunecare. Acum, repetaţi pasul 4 de mai sus.
6) În prezenţa NaOCl, selectaţi instrumentul Shaper şi avansaţi pe traseul de alunecare, în una sau mai multe treceri, până se atinge lungimea de lucru.
7) La scoaterea instrumentului Shaper, irigaţi, utilizaţi instrumentul Endo Activator pentru a dispersa reziduurile şi a le muta în soluţie şi după aceea reirigaţi pentru degajarea reziduurilor.
8) Reverificaţi lungimea de lucru, mai ales în canalele mai curbate.
9) Selectaţi instrumentul FINISHER F1 (020.07) şi urmaţi pasiv canalul până la lungimea de lucru în una sau mai multe treceri. Scoateţi şi inspectaţi canelurile apicale. Dacă sunt încărcate cu reziduuri de dentină, pregătirea este finalizată.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 108/151
10) În cazul în care instrumentul FINISHER F1 nu ajunge pe toată lungimea şi canelurile sale apicale nu sunt încărcate cu reziduuri, selectaţi instrumentul FINISHER F2 (025.08) şi utilizaţi acest ac în modul descris mai sus pentru FINISHER F1.
11) Dacă instrumentul FINISHER F1 nu ajunge pe toată lungimea şi canelurile sale apicale nu sunt încărcate cu reziduuri, selectaţi instrumentul FINISHER F3 (030.09) şi utilizaţi acest ac în modul descris mai sus pentru FINISHER F1 şi F2.
12) La scoaterea oricărui ac, curăţaţi şi inspectaţi canelurile acestuia, irigaţi, reluaţi operaţia cu un ac de mărimea 10 sau un instrument Endo Activator pentru dispersarea reziduurilor şi după aceea reirigaţi.
13) Inspectaţi canelurile acului la scoatere pentru a depista prezenţa destinderii, îndreptării sau
întinderii. Dacă se observă o deformare, eliminaţi şi utilizaţi un ac ProTaper Ultimate™ nou.
14) Pregătirea este finalizată când prelungirea apicală a oricărui instrument Finisher este încărcată cu reziduuri şi conul de ghidaj din gutapercă de mărime corespunzătoare sau Size
Verifier se potriveşte la lungimea de lucru.
Utilizaţi instrumentele Finisher auxiliare doar în canale mai largi şi mai drepte, cum ar fi cele de la incisivii centrali maxilari, unele canale palatale sau distale ale molarilor sau dacă există un defect patologic sau iatrogenic:
15) Selectaţi un ac auxiliar acţionat mecanic sau manual, FINISHER FX (035.12), când este stabilită lungimea de lucru şi este confirmată vizibilitatea. Utilizaţi acest ac independent direct şi pasiv pentru urmărirea canalului până la lungimea de lucru utilizând aceeaşi tehnică descrisă mai sus. Scoateţi şi inspectaţi canelurile apicale. Dacă sunt încărcate cu reziduuri de dentină, pregătirea este finalizată.
16) Dacă instrumentul FINISHER FX nu ajunge pe toată lungimea şi canelurile sale apicale nu sunt încărcate cu reziduuri, selectaţi un instrument auxiliar acţionat mecanic sau manual
FINISHER FXL (050.10) şi utilizaţi acest ac în modul descris mai sus pentru FINISHER
FX.
17) Pregătirea este finalizată când prelungirea apicală a oricărui instrument auxiliar Finisher este încărcată cu reziduuri şi conul de ghidaj din gutapercă de mărime corespunzătoare sau Size Verifier se potriveşte la lungimea de lucru.
● Respectaţi Instrucţiuni generale de prelucrare a produselor endodontice ( https:// www.dentsplysirona.com/en/service-contact/ifu.html
) pentru etapele de dezinfectare, curăţare şi sterilizare.
● Nu reutilizaţi opritoarele din silicon. Scoateţi şi eliminaţi opritoarele din silicon după fiecare utilizare.
● Produsele vor fi eliminate în conformitate cu reglementările locale privind eliminarea în siguranţă a dispozitivelor ascuţite şi contaminate.
● Orice incident grav referitor la produs trebuie raportat producătorului şi autorităţii competente, conform reglementărilor locale.
● Sterilitatea nu poate fi garantată dacă ambalajul este deschis, deteriorat sau umed.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 109/151
● Pentru a obţine un exemplar gratuit al Instrucţiunilor de utilizare, consultaţi secţiunea „Obţinerea unui exemplar al Instrucţiunilor de utilizare” pe site-ul Web https:// www.dentsplysirona.com/en/service-contact/ifu.html
.
● Pentru explicarea simbolurilor nearmonizate pentru instrucţiunile de utilizare şi etichete, consultaţi secţiunea Simboluri din Instrucţiunile de utilizare ( https:// www.dentsplysirona.com/en/service-contact/ifu/ifu-symbols.html
).
2797
Reprezentant autorizat
DENTSPLY DeTrey GmbH
De-Trey-Straße 1
D-78467 Konstanz
Germany
Producător
Maillefer Instruments Holding Sàrl
Chemin du Verger 3
CH-1338 Ballaigues
Suisse www.dentsplysirona.com/ifu
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 110/151
ProTaper Ultimate™
SAMO ZA STOMATOLOŠKU UPOTREBU
SAMO ZA PROFESIONALNU UPOTREBU
STERILNO - STERILIZOVANO RADIJACIJOM
POGODNO ZA PONOVNU OBRADU
VIŠEKRATNI STERILNI ENDODONTSKI INSTRUMENTI
● Proizvodi su indikovani za lečenje endodontske bolesti.
● Namena: Motorni instrument namenjen preparaciji korenskog kanala (klizna putanja, oblikovanje i čišćenje korenskog kanala).
● Endodontske instrumente treba da koriste kvalifikovani stomatološki stručnjaci kao što su lekari opšte prakse, kao i specijalisti endodoncije (endodontisti) i stomatološki asistenti, samo u kliničkom ili bolničkom okruženju, prateći dobru stomatološku praksu.
● Treba koristiti instrumente u kombinaciji sa endodontskim motorom (ručno kod izraženih zakrivljenosti).
Nema.
ProTaper Ultimate™ instrumenti su napravljeni od tri glavne komponente: radnog dela izrađenog od nikl-titanijumske legure, obojenog silikonskog stopera (osim instrumenta SX) i držača obloženog mesingom sa obojenim prstenom (osim instrumenta SX).
Turpijice su proizvedene postupkom koji rezultira turpijicom koja je u boji.
Zbog takve vlasničke obrade instrumenti mogu da izgledaju neznatno zakrivljeni. Ovo nije proizvodna greška. Dok se turpijica može lako ispraviti samo prstima, turpijicu nije potrebno ispraviti pre upotrebe. Kad uđe u kanal, turpijica će pratiti anatomiju korenskog kanala.
Opseg ProTaper Ultimate™ proizvoda sačinjavaju sledeći endodontski instrumenti:
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 111/151
Kôd: xx-dužina
Veličina y-turpijice z-kvantitet po blisteru
B ST PULR z
19 0SX
B ST PULR z xx 0SL
B ST PULR z xx 0SH
Proizvod
ProTaper Ultimate™
SX
ProTaper Ultimate™
SLIDER
ProTaper Ultimate™
SHAPER
ISO XXX YYYv
XXX-ISO veličina
YYY-konusna veličina v-promenljivi konus
ISO 020.003v
ISO 016.002v
ISO 020.004v
Dostupne dužine
19 mm
21/25/31 mm
21/25/31 mm
Pakovanje i sadržina
Blister sa 6 instrumenata (sterilno)
Blister sa 3 instrumenta (sterilno)
Blister sa 6 instrumenata (sterilno)
Blister sa 3 instrumenta (sterilno)
Blister sa 6 instrumenata (sterilno)
Blister sa 3 instrumenta (sterilno)
Blister sa 6 instrumenata (sterilno)
B ST PULR z xx 0F y
B ST PULR 6 xx yyy
B ST PULR 5 xx SQE
B DD PULR
525 SQE
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
Sequence
ProTaper Ultimate™
Demo Pack
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
FX: ISO 035.012v
FXL: ISO 050.010v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
21/25/31 mm
21/25/31 mm
21/25/31 mm
25 mm
Blister sa 3 instrumenta (sterilno)
Blister sa 6 instrumenata (sterilno)
Blister sa 1 instrumentom od svakog instrumenta (sterilno)
Blister sa 1 instrumentom od svakog instrumenta (sterilno)
● Uređaji za kontrolu obrtnog momenta osiguravaju optimalno korišćenje.
● Koristite sa endodontskim motorima u konstantnoj rotaciji pri brzini od 400 rpm.
● Podesite obrtni momenat na: 4 to 5.2 Ncm.
● Lubrikanti kao što je NaOCl, EDTA, ProLube, Glyde™ treba da budu korišćeni.
● Upotreba radiografskih snimaka u kombinaciji sa apeks lokatorom i alatom za prilagođavanje silikonskog graničnika tačnoj radnoj dužini je odgovarajući metod za određivanje radne dužine.
● Strogo se pridržavajte ovog Uputstava za upotrebu i Opštih uputstava za obradu endodontskih proizvoda (pogledajte odeljak 8) da biste umanjili sledeće rizike po uređaj, pacijenta i/ili korisnika:
○ Kvar instrumenta.
○ Unakrsna kontaminacija.
○ Stvaranje toplote usled nedovoljne lubrikacije i irigacije.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 112/151
○ Gutanje radnog dela instrumenta.
○ Toksične ili alergijske reakcije izazvane ostacima od prerade.
● Kao što je slučaj i sa svim instrumentima za obradu korenskih kanala na mašinski pogon,
ProTaper Ultimate™ instrumenti se ne smeju koristiti ako postoji nagla apikalna zakrivljenost zbog povećanog rizika od separacije. U ovom slučaju, prethodno zakrivljene ručne turpije treba koristiti u apikalnom području.
● Bezbednost i delotvornost upotrebe nisu utvrđene kod trudnica ili dojilja ili kod dece.
● Tokom endodontskog postupka koristite gumeni sistem zaštite.
● Radi sopstvene sigurnosti nosite ličnu zaštitnu opremu (rukavice, naočare, masku).
● Pre upotrebe pregledajte pakovanje i nemojte koristiti instrumente ako je ono oštećeno.
● Ne koristite instrumente nakon isteka roka upotrebe.
● Pre svake upotrebe proverite instrument da li ima znakova oštećenja poput deformacija (savijenih, odmotanih), kvarova, korozije, oštećenih reznih ivica, gubitak koda boje ili oznake.
Uz ove indikacije uređaji ne mogu da ispune svoju predviđenu namenu sa potrebnim nivoom bezbednosti, pa instrumente treba odložiti u otpad.
● Pre korišćenja bilo kog instrumenta, proverite da li je on dobro povezan za glavu kontraugla.
● Proveravajte instrument i tokom upotrebe instrumenta redovno čistite radni deo tako što ćete proveravati da li su se javili znaci izobličenja, izduženja ili habanja, poput neujednačenih brazda ili tamnih mrlja. Uz ove indikacije da uređaji ne mogu da ispune svoju predviđenu namenu sa potrebnim nivoom bezbednosti, instrumente treba odložiti u otpad.
● Instrumenti se ne smeju celi potapati u rastvor natrijum hipohlorita (NaOCl). Samo radni deo NiTi instrumenta, koji je u kontaktu sa pacijentom, treba da bude uronjen u koncentrat
NaOCl rastvora, i to NE veći od 5%.
● Budite oprezni u apikalnom području i u kanalima koji se račvaju i/ili ispoljavaju nagle zakrivljenosti ili višestruke zakrivljenosti.
● Obilno i često irigirati kanal tokom postupka i nakon svakog korišćenog instrumenta (u skladu sa dobrom stomatološkom praksom).
● Uvek primenjujte minimalni apikalni pritisak. Nikad ne uvodite turpijicu kroz kanal na silu.
● Kada instrument ne napreduje s lakoćom, očistite i pregledajte rezne brazde, zatim irigirajte, rekapitulirajte ručnom turpijicom i ponovo irigirajte.
● Za oblikovanje izuzetno zakrivljenih kanala, bezbednije je koristiti turpijicu za oblikovanje samo jednog kanala kako bi se smanjila opasnost od lomljenja. Obratite pažnju na sledeće dobre prakse:
○ Koristite novu turpijicu i odložite je kada se kanal obradi (upotreba za samo jedan kanal).
○ Koristite ručne umesto rotacionih turpijica.
○ Koristite turpijice male veličine, fleksibilne i/ili NiTi turpijice (ovo pomaže da se izbegne kanalna transportacija).
○ Vizuelno proverite radni deo zbog svih oštećenja navedenih u prethodnom pasusu tokom korišćenja (tj. nakon svakog zamaha).
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 113/151
○ Izbegavajte standardni kontinuirani rotacioni pokret za proširivanje i umesto toga koristite pokrete sa malim uglom (pokret turpijanja, oscilirajući pokret uvijanja sata ili tehniku balansirane sile) kako bi se ograničilo opterećenje instrumenata rotacionim savijanjem i poboljšao njihov očekivani radni vek.
Za trenutno tehničko stanje do sada nije prijavljena nijedna neželjena reakcija.
Čuvajte proizvode u suvom i čistom okruženju, dalje od izvora vlage i direktne sunčeve svetlosti.
Sve ProTaper Ultimate™ turpijice su odobrene za korišćenje pri brzini rotacije od 400 rpm i sa obrtnim momentom od 4 to 5.2 Ncm.
Instrumenti ProTaper Ultimate™ se mogu koristiti sa pokretima četkanja prema spolja u svim kanalima, posebno onim kanalima koji imaju nepravilan poprečni presek, ili sa laganim kretanjem prema unutra, kako bi se progresivno napredovalo ka radnoj dužini (WL).
Uvek postavljajte brusilicu u pregib između palca i kažiprsta. Izbegavajte guranje, već pustite da turpijice pasivno napreduju i prate kliznu putanju.
1) Pregledajte različite vodoravno angulirane radiografe da biste dijagnostikovali širinu, dužinu i zakrivljenost svih datih korena i njegovog/njegovih kanala.
2) Pripremite pristup koji omogućava lako prepoznavanje svakog otvora kanala.
3) Pomoćna turpijica za oblikovanje, SX, može se koristiti kada postoji ograničen prostor ili za prethodno uvećanje tela kanala ili za relokaciju kruničnog aspekta kanala dalje od spoljnog konkaviteta korena.
4) U prisustvu lubrikanta, odaberite Slider i PASIVNO pratite kanal, u jednom ili više prolaza, do njegove završnice. Odredite WL (radnu dužinu) pomoću elektronskog apeks lokatora u kombinaciji sa radiografskim snimkom, a zatim potvrdite prohodnost.
5) Ako Slider ne dolazi lako do završnice kanala, odaberite manuelnu ručnu turpiju od nerđajućeg čelika (SS) manje veličine. U prisustvu lubrikanta kateterizujte ovaj kanal, utvrdite radnu dužinu, potvrdite prohodnost i verifikujte kliznu putanju. Sada ponovite gornji korak #
4.
6) U prisustvu NaOCl, izaberite Shaper i napredujte duž klizne putanje, u jednom ili više prolaza, dok se ne dostigne radna dužina (WL).
7) Nakon što je uklonjen Shaper, irigirajte, EndoActivate da razbijete debris i premestite ga u rastvor, a zatim ponovo irigirajte da biste oslobodili debris.
8) Ponovo potvrdite radnu dužinu (WL), posebno u zakrivljenijim kanalima.
9) Izaberite FINISHER F1 (020.07) i pasivno pratite kanal do radne dužine (WL), u jednom ili više prolaza. Uklonite i pregledajte njene apikalne brazde. Kada se napune debrisom od dentina, priprema je gotova.
10) Ako je FINISHER F1 labav po dužini i njegove apikalne brazde nisu ispunjene debrisom, odaberite FINISHER F2 (025.08) i koristite ovu turpijicu na isti način kao što je prethodno opisano za FINISHER F1.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 114/151
11) Ako je FINISHER F2 labav po dužini i njegove apikalne brazde nisu ispunjene debrisom, odaberite ručni pomoćni FINISHER F3 (030.09) i koristite ovu turpijicu na isti način kao što je prethodno opisano za FINISHER F1 i F2.
12) Nakon uklanjanja bilo koje date turpijice, očistite i pregledajte njene rezne brazde, irigirajte, rekapitulirajte turpijicom veličine 10 ili Endo Activatorom da biste razbili debris, a zatim ponovo irigirajte.
13) Nakon uklanjanja pregledajte rezne brazde turpijice, da li se odmotavaju, ispravljaju ili istežu. Ako primetite deformaciju odložite u otpad i uzmite novu ProTaper Ultimate™ turpijicu.
14) Priprema se završava kada se apikalni opseg bilo kog Finishera napuni debrisom, a odgovarajuća veličina glavnog gutaperka kočića (GPMC) ili Size Verifier odgovara radnoj dužini
(WL).
Koristite pomoćne Finishere samo u većim i ravnijim kanalima, kao što su maksilarni centralni sekutići, neki nepčani ili distalni kanali molara ili kada postoji patološki ili jatrogeni defekt:
15) Izaberite ili mehaničku ili ručnu pomoćnu turpijicu FINISHER FX (035.12) kada se utvrdi radna dužina i potvrdi prohodnost. Koristite ovu samostalnu turpijicu direktno i pasivno da biste pratili kanal do WL (radne dužine) koristeći istu tehniku kao što je gore opisano. Uklonite i pregledajte njene apikalne brazde. Kada se napune debrisom od dentina, priprema je gotova.
16) Ako je FINISHER FX labav po dužini i njegove apikalne brazde nisu opterećene debrisom, odaberite pomoćni FINISHER FXL (050.10) ili na mašinski pogon ili ručni i koristite ovu turpijicu na isti način kao što je prethodno opisano za FINISHER FX.
17) Priprema se završava kada se apikalni opseg bilo kog pomoćnog Finishera napuni debrisom, a odgovarajuća veličina glavnog gutaperka kočića (GPMC) ili Size Verifier odgovara radnoj dužini (WL).
● Pridržavajte se Opštih uputstava za obradu za endodontske proizvode ( https:// www.dentsplysirona.com/en/service-contact/ifu.html
) za korake za dezinfekciju, čišćenje i sterilizaciju.
● Ne koristiti ponovo silikonske stopere. Uklonite i odložite silikonske stopere nakon svake upotrebe.
● Proizvode treba odlagati u otpad u skladu sa lokalnim propisima za bezbedno odlaganje oštrih i kontaminiranih uređaja.
● Svaki ozbiljni incident u vezi sa proizvodom treba prijaviti proizvođaču i nadležnom organu u skladu sa lokalnim propisima.
● Sterilnost se ne može garantovati ako je ambalaža otvorena, oštećena ili vlažna.
● Da biste dobili besplatnu štampanu kopiju Uputstava za upotrebu, pogledajte odeljak „Nabavite štampanu kopiju Uputstava za upotrebu“ na veb stranici https:// www.dentsplysirona.com/en/service-contact/ifu.html
.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 115/151
● Objašnjenje neharmonizovanih simbola za Uputstva za upotrebu i nalepnice, pogledajte
Simbole za Uputstva za upotrebu ( https://www.dentsplysirona.com/en/service-contact/ifu/ ifu-symbols.html
).
2797
Ovlašćeni predstavnik
DENTSPLY DeTrey GmbH
De-Trey-Straße 1
D-78467 Konstanz
Germany
Proizvođač
Maillefer Instruments Holding Sàrl
Chemin du Verger 3
CH-1338 Ballaigues
Suisse www.dentsplysirona.com/ifu
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 116/151
ProTaper Ultimate™
SAMO ZA STOMATOLOŠKU UPOTREBU
SAMO ZA PROFESIONALNU UPOTREBU
STERILNO - STERILIZOVANO RADIJACIJOM
POGODNO ZA PONOVNU OBRADU
VIŠEKRATNI STERILNI ENDODONTSKI INSTRUMENTI
● Proizvodi su indikovani za liječenje endodontske bolesti.
● Namjena: Motorni instrument namijenjen preparaciji korijenskog kanala (klizna putanja, oblikovanje i čišćenje korijenskog kanala).
● Endodontske instrumente treba da koriste kvalifikovani stomatološki stručnjaci kao što su ljekari opšte prakse, kao i specijalisti endodoncije (endodontisti) i stomatološki asistenti, samo u kliničkom ili bolničkom okruženju, prateći dobru stomatološku praksu.
● Treba koristiti instrumente u kombinaciji s endodontskim motorom (ručno kod izraženih zakrivljenosti).
Nema.
ProTaper Ultimate™ instrumenti su napravljeni od tri glavne komponente: radnog dijela izrađenog od nikl-titanijumske legure, obojenog silikonskog stopera (osim instrumenta SX) i držača obloženog mesingom s obojenim prstenom (osim instrumenta SX).
Turpijice su proizvedene postupkom koji rezultira turpijicom u boji.
Zbog takve vlasničke obrade instrumenti mogu izgledati blago zakrivljeni. Ovo nije proizvodna greška. Dok se turpijica može lako ispraviti samo prstima, turpijicu nije potrebno ispraviti prije upotrebe. Kad uđe u kanal, turpijica će pratiti anatomiju korijenskog kanala.
Opseg proizvoda ProTaper Ultimate™ sačinjavaju sljedeći endodontski instrumenti:
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 117/151
Šifra xx-dužina
Veličina yturpijice z-količina po blisteru
B ST PULR z
19 0SX
B ST PULR z xx 0SL
B ST PULR z xx 0SH
Proizvod
ProTaper Ultimate™
SX
ProTaper Ultimate™
SLIDER
ProTaper Ultimate™
SHAPER
ISO XXX YYYv
XXX-ISO veličina
YYY-konusna veličina v-promjenjivi konus
ISO 020.003v
ISO 016.002v
ISO 020.004v
Dostupne dužine
19 mm
21/25/31 mm
21/25/31 mm
Pakovanje i sadržaj
Blister sa 6 instrumenata (sterilno)
Blister s 3 instrumenta (sterilno)
Blister sa 6 instrumenata (sterilno)
Blister s 3 instrumenta (sterilno)
Blister sa 6 instrumenata (sterilno)
Blister s 3 instrumenta (sterilno)
Blister sa 6 instrumenata (sterilno)
B ST PULR z xx 0F y
B ST PULR 6 xx yyy
B ST PULR 5 xx SQE
B DD PULR
525 SQE
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
Sequence
ProTaper Ultimate™
Demo Pack
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
FX: ISO 035.012v
FXL: ISO 050.010v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
21/25/31 mm
21/25/31 mm
21/25/31 mm
25 mm
Blister s 3 instrumenta (sterilno)
Blister sa 6 instrumenata (sterilno)
Blister s 1 instrumentom od svakog instrumenta (sterilno)
Blister s 1 instrumentom od svakog instrumenta (sterilno)
● Uređaji za kontrolu obrtnog momenta omogućavaju optimalno korištenje.
● Koristite s endodontskim motorima u konstantnoj rotaciji pri brzini od 400 rpm.
● Podesite obrtni momenat na: 4 to 5.2 Ncm.
● Treba da se koriste lubrikanti kao što je NaOCl, EDTA, ProLube, Glyde™.
● Upotreba radiografskih snimaka u kombinaciji s apeks lokatorom i alatom za prilagođavanje silikonskog graničnika tačnoj radnoj dužini je odgovarajući način određivanja radne dužine.
● Strogo se pridržavajte ovog Uputstava za upotrebu i Opštih uputstava za obradu endodontskih proizvoda (pogledajte odjeljak 8) da biste umanjili sekundarne rizike po uređaj, pacijenta i/ili korisnika:
○ Kvar instrumenta.
○ Unakrsna kontaminacija.
○ Stvaranje toplote uslijed nedovoljne lubrikacije i irigacije.
○ Gutanje radnog dijela instrumenta.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 118/151
○ Toksične ili alergijske reakcije izazvane ostacima od prerade.
● Kao što je slučaj i sa svim instrumentima za obradu korijenskih kanala na mašinski pogon,
ProTaper Ultimate™ instrumenti se ne smiju koristiti ako postoji nagla apikalna zakrivljenost zbog povećanog rizika od separacije. U ovom slučaju, prethodno zakrivljene ručne turpije treba koristiti u apikalnom području.
● Sigurnost i efikasnost upotrebe nisu utvrđene kod trudnica, dojilja ni kod djece.
● Tokom endodontskog postupka koristite gumeni sistem zaštite.
● Radi svoje sigurnosti nosite ličnu zaštitnu opremu (rukavice, naočale, masku).
● Prije upotrebe pregledajte ambalažu i nemojte koristiti instrumente ako je oštećena.
● Ne koristite instrumente nakon isteka roka upotrebe.
● Prije svake upotrebe provjerite da li instrument ima znakova oštećenja poput deformacija
(savijenih, odmotanih), kvarova, korozije, oštećenih reznih rubova, gubitak boje ili oznake
šifriranja. Uz ove indikacije uređaji ne mogu da ispune svoju predviđenu namjenu s potrebnim nivoom sigurnosti, pa instrumente treba odložiti u otpad.
● Prije korištenja bilo kog instrumenta, provjerite da li je čvrsto spojen za glavu suprotnog ugla.
● Provjeravajte instrument i tokom upotrebe instrumenta redovno čistite radni dio tako što ćete provjeravati da li su se pojavili znaci deformacije, izduženja ili habanja, poput neujednačenih brazda ili tamnih mrlja. Uz ove indikacije da uređaji ne mogu da ispune svoju predviđenu namjenu s potrebnim nivoom sigurnosti, instrumente treba odložiti u otpad.
● Instrumenti se ne smiju kompletni potapati u rastvor natrijum hipohlorita (NaOCl). Samo radni dio NiTi instrumenta, koji je u kontaktu s pacijentom, treba da bude uronjen u koncentrat NaOCl rastvora, i to NE veći od 5%.
● Budite oprezni u apikalnom području i kanalima koji se račvaju i/ili ispoljavaju nagle zakrivljenosti ili višestruke zakrivljenosti.
● Obilno i često irigirati kanal tokom postupka i nakon svakog korištenog instrumenta (u skladu s dobrom stomatološkom praksom).
● Uvijek primjenjujte minimalni apikalni pritisak. Nikad ne uvodite turpijicu kroz kanal na silu.
● Kada instrument ne napreduje s lakoćom, očistite i pregledajte rezne brazde, zatim irigirajte, rekapitulirajte ručnom turpijicom i ponovo irigirajte.
● Za oblikovanje izuzetno zakrivljenih kanala, sigurnije je koristiti turpijicu za oblikovanje samo jednog kanala kako bi se smanjila opasnost od lomljenja. Obratite pažnju na sljedeće dobre prakse:
○ Koristite novu turpijicu i odložite je kada se kanal obradi (upotreba za samo jedan kanal).
○ Koristite ručne umjesto rotacionih turpijica.
○ Koristite turpijice male veličine, fleksibilne i/ili NiTi turpijice (to pomaže da se izbjegne kanalna transportacija).
○ Vizuelno provjerite radni dio zbog svih oštećenja navedenih u prethodnom paragrafu tokom korištenja (tj. nakon svakog zamaha).
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 119/151
○ Izbjegavajte standardni kontinuirani rotacioni pokret za proširivanje i umjesto toga koristite pokrete s malim uglom (pokret turpijanja, oscilirajući pokret uvijanja sata ili tehniku balansirane sile) kako bi se ograničilo opterećenje instrumenata rotacionim savijanjem i poboljšao njihov očekivani radni vijek.
Za trenutno tehničko stanje do sada nije prijavljena nijedna neželjena reakcija.
Čuvajte proizvode u suhom i čistom okruženju, dalje od izvora vlage i direktne sunčeve svjetlosti.
Sve ProTaper Ultimate™ turpijice su odobrene za korištenje pri brzini rotacije od 400 rpm i s obrtnim momentom od 4 to 5.2 Ncm.
Instrumenti ProTaper Ultimate™ mogu se koristiti s pokretima četkanja prema van u svim kanalima, posebno onim kanalima koji imaju nepravilan poprečni presjek, ili s laganim kretnjama prema unutra, kako bi se progresivno napredovalo k radnoj dužini (WL).
Uvijek postavljajte brusilicu u pregib između palca i kažiprsta. Izbjegavajte guranje, nego pustite da turpijice pasivno napreduju i prate kliznu putanju.
1) Pregledajte različite horizontalno angulirane radiografe da biste dijagnostikovali širinu, dužinu i zakrivljenost svih datih korijena i njegovog/njegovih kanala.
2) Pripremite pristup koji omogućava lako prepoznavanje svakog otvora kanala.
3) Pomoćna turpijica za oblikovanje, SX, može se koristiti kada postoji ograničen prostor ili za prethodno uvećanje tijela kanala ili za relokaciju kruničnog aspekta kanala dalje od spoljnog konkaviteta korijena.
4) U prisustvu lubrikanta, odaberite Slider i PASIVNO pratite kanal, u jednom ili više prolaza, do njegove završnice. Odredite WL (radnu dužinu) pomoću elektronskog apeks lokatora u kombinaciji s radiografskim snimkom, a zatim potvrdite prohodnost.
5) Ako Slider ne dolazi lako do završnice kanala, odaberite manuelnu ručnu turpiju od nerđajućeg čelika (SS) manje veličine. U prisustvu lubrikanta kateterizujte ovaj kanal, utvrdite radnu dužinu, potvrdite prohodnost i verifikujte kliznu putanju. Sada ponovite gornji korak # 4.
6) U prisustvu NaOCl, izaberite Shaper i napredujte duž klizne putanje, u jednom ili više prolaza, dok se ne dostigne radna dužina (WL).
7) Nakon što je uklonjen Shaper, irigirajte, EndoActivate da razbijete debris i premjestite ga u rastvor, a zatim ponovo irigirajte da biste oslobodili debris.
8) Ponovo potvrdite radnu dužinu (WL), posebno u zakrivljenijim kanalima.
9) Izaberite FINISHER F1 (020.07) i pasivno pratite kanal do radne dužine (WL), u jednom ili više prolaza. Uklonite i pregledajte njene apikalne brazde. Kada se napune debrisom od dentina, priprema je gotova.
10) Ako je FINISHER F1 labav po dužini i njegove apikalne brazde nisu ispunjene debrisom, odaberite FINISHER F2 (025.08) i koristite ovu turpijicu na isti način kao što je opisano za
FINISHER F1.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 120/151
11) Ako je FINISHER F2 labav po dužini i njegove apikalne brazde nisu ispunjene debrisom, odaberite ručni pomoćni FINISHER F3 (030.09) i koristite ovu turpijicu na isti način kao što je opisano za FINISHER F1 i F2.
12) Nakon uklanjanja bilo koje date turpijice, očistite i pregledajte njene rezne brazde, irigirajte, rekapitulirajte turpijicom veličine 10 ili Endo Activatorom da biste razbili debris, a zatim ponovo irigirajte.
13) Nakon uklanjanja pregledajte rezne brazde turpijice, da li se odmotavaju, ispravljaju ili istežu. Ako primijetite deformaciju, odložite u otpad i uzmite novu ProTaper Ultimate™ turpijicu.
14) Priprema se završava kada se apikalni opseg svakog Finishera napuni debrisom, a odgovarajuća veličina glavnog gutaperka kočića (GPMC) ili Size Verifier odgovara radnoj dužini (WL).
Koristite pomoćne finišere samo u većim i ravnijim kanalima, kao što su maksilarni centralni sjekutići, neki nepčani ili distalni kanali molara ili kada postoji patološki ili jatrogeni defekt:
15) Izaberite ili mehaničku ili ručnu pomoćnu turpijicu FINISHER FX (035.12) kada se utvrdi radna dužina i potvrdi prohodnost. Koristite ovu samostalnu turpijicu direktno i pasivno da biste pratili kanal do WL (radne dužine) koristeći istu tehniku koja je prethodno opisana.
Uklonite i pregledajte njene apikalne brazde. Kada se napune debrisom od dentina, priprema je gotova.
16) Ako je FINISHER FX labav po dužini i njegove apikalne brazde nisu opterećene debrisom, odaberite pomoćni FINISHER FXL (050.10) ili na mašinski pogon ili ručni i koristite ovu turpijicu na isti način kao što je opisano za FINISHER FX.
17) Priprema se završava kada se apikalni opseg svakog pomoćnog Finishera napuni debrisom, a odgovarajuća veličina glavnog gutaperka kočića (GPMC) ili Size Verifier odgovara radnoj dužini (WL).
● Pridržavajte se Opštih uputstava za obradu za endodontske proizvode ( https:// www.dentsplysirona.com/en/service-contact/ifu.html
) za korake za dezinfekciju, čišćenje i sterilizaciju.
● Ne koristiti ponovo silikonske stopere. Uklonite i odložite silikonske stopere nakon svake upotrebe.
● Proizvode treba odlagati u otpad u skladu s lokalnim propisima za sigurno odlaganje oštrih i kontaminiranih uređaja.
● Svaki ozbiljni incident u vezi s proizvodom treba prijaviti proizvođaču i nadležnom organu u skladu s lokalnim propisima.
● Sterilnost se ne može garantovati ako je ambalaža otvorena, oštećena ili vlažna.
● Da biste dobili besplatnu štampanu kopiju Uputstava za upotrebu, pogledajte odjeljak
„Nabavite štampanu kopiju Uputstava za upotrebu“ na veb stranici https:// www.dentsplysirona.com/en/service-contact/ifu.html
.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 121/151
● Objašnjenje neharmonizovanih simbola za Uputstva za upotrebu i naljepnice, pogledajte
Simbole za Uputstva za upotrebu ( https://www.dentsplysirona.com/en/service-contact/ifu/ ifu-symbols.html
).
2797
Ovlašteni zastupnik
DENTSPLY DeTrey GmbH
De-Trey-Straße 1
D-78467 Konstanz
Germany
Proizvođač
Maillefer Instruments Holding Sàrl
Chemin du Verger 3
CH-1338 Ballaigues
Suisse www.dentsplysirona.com/ifu
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 122/151
ProTaper Ultimate™
SAMO ZA STOMATOLOŠKU UPOTREBU
SAMO ZA PROFESIONALNU UPOTREBU
STERILNO - STERILIZOVANO RADIJACIJOM
POGODNO ZA PONOVNU OBRADU
VIŠEKRATNI STERILNI ENDODONTSKI INSTRUMENTI
● Proizvodi su indikovani za liječenje endodontske bolesti.
● Namjena: Motorni instrument namijenjen preparaciji korijenskog kanala (klizna putanja, oblikovanje i čišćenje korijenskog kanala).
● Endodontske instrumente treba da koriste kvalifikovani stomatološki stručnjaci kao što su ljekari opšte prakse, kao i specijalisti endodoncije (endodontisti) i stomatološki asistenti, samo u kliničkom ili bolničkom okruženju, prateći dobru stomatološku praksu.
● Treba koristiti instrumente u kombinaciji sa endodontskim motorom (ručno kod izraženih zakrivljenosti).
Nema.
ProTaper Ultimate™ instrumenti su napravljeni od tri glavne komponente: radnog dijela izrađenog od nikl-titanijumske legure, obojenog silikonskog stopera (osim instrumenta SX) i držača obloženog mesingom sa obojenim prstenom (osim instrumenta SX).
Turpijice su proizvedene postupkom koji rezultira turpijicom u boji.
Zbog takve vlasničke obrade instrumenti mogu izgledati blago zakrivljeni. Ovo nije proizvodna greška. Dok se turpijica može lako ispraviti samo prstima, turpijicu nije potrebno ispraviti prije upotrebe. Kad uđe u kanal, turpijica će pratiti anatomiju korijenskog kanala.
Opseg proizvoda ProTaper Ultimate™ sačinjavaju sljedeći endodontski instrumenti:
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 123/151
Šifra xx-dužina
Veličina yturpijice z-količina po blisteru
B ST PULR z
19 0SX
B ST PULR z xx 0SL
B ST PULR z xx 0SH
Proizvod
ProTaper Ultimate™
SX
ProTaper Ultimate™
SLIDER
ProTaper Ultimate™
SHAPER
ISO XXX YYYv
XXX-ISO veličina
YYY-konusna veličina v-promjenjivi konus
ISO 020.003v
ISO 016.002v
ISO 020.004v
Dostupne dužine
19 mm
21/25/31 mm
21/25/31 mm
Pakovanje i sadržina
Blister sa 6 instrumenata (sterilno)
Blister s 3 instrumenta (sterilno)
Blister sa 6 instrumenata (sterilno)
Blister s 3 instrumenta (sterilno)
Blister sa 6 instrumenata (sterilno)
Blister s 3 instrumenta (sterilno)
Blister sa 6 instrumenata (sterilno)
B ST PULR z xx 0F y
B ST PULR 6 xx yyy
B ST PULR 5 xx SQE
B DD PULR
525 SQE
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
Sequence
ProTaper Ultimate™
Demo Pack
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
FX: ISO 035.012v
FXL: ISO 050.010v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
21/25/31 mm
21/25/31 mm
21/25/31 mm
25 mm
Blister s 3 instrumenta (sterilno)
Blister sa 6 instrumenata (sterilno)
Blister sa 1 instrumentom od svakog instrumenta (sterilno)
Blister sa 1 instrumentom od svakog instrumenta (sterilno)
● Uređaji za kontrolu obrtnog momenta omogućavaju optimalno korišćenje.
● Koristite sa endodontskim motorima u konstantnoj rotaciji pri brzini od 400 rpm.
● Podesite obrtni momenat na: 4 to 5.2 Ncm.
● Treba da se koriste lubrikanti kao što je NaOCl, EDTA, ProLube, Glyde™.
● Upotreba radiografskih snimaka u kombinaciji sa apeks lokatorom i alatom za prilagođavanje silikonskog graničnika tačnoj radnoj dužini je odgovarajući način određivanja radne dužine.
● Strogo se pridržavajte ovog Uputstava za upotrebu i Opštih uputstava za obradu endodontskih proizvoda (pogledajte odjeljak 8) da biste umanjili sekundarne rizike po uređaj, pacijenta i/ili korisnika:
○ Kvar instrumenta.
○ Unakrsna kontaminacija.
○ Stvaranje toplote uslijed nedovoljne lubrikacije i irigacije.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 124/151
○ Gutanje radnog dijela instrumenta.
○ Toksične ili alergijske reakcije izazvane ostacima od prerade.
● Kao što je slučaj i sa svim instrumentima za obradu korijenskih kanala na mašinski pogon,
ProTaper Ultimate™ instrumenti se ne smiju koristiti ako postoji nagla apikalna zakrivljenost zbog povećanog rizika od separacije. U ovom slučaju, prethodno zakrivljene ručne turpije treba koristiti u apikalnom području.
● Sigurnost i efikasnost upotrebe nisu utvrđene kod trudnica, dojilja ni kod djece.
● Tokom endodontskog postupka koristite gumeni sistem zaštite.
● Radi svoje sigurnosti nosite ličnu zaštitnu opremu (rukavice, naočare, masku).
● Prije upotrebe pregledajte ambalažu i nemojte koristiti instrumente ako je oštećena.
● Ne koristite instrumente nakon isteka roka upotrebe.
● Prije svake upotrebe provjerite da li instrument ima znakova oštećenja poput deformacija
(savijenih, odmotanih), kvarova, korozije, oštećenih reznih rubova, gubitak boje ili oznake
šifriranja. Uz ove indikacije uređaji ne mogu da ispune svoju predviđenu namjenu sa potrebnim nivoom sigurnosti, pa instrumente treba odložiti u otpad.
● Prije korišćenja bilo kog instrumenta, provjerite da li je čvrsto spojen za glavu suprotnog ugla.
● Provjeravajte instrument i tokom upotrebe instrumenta redovno čistite radni dio tako što ćete provjeravati da li su se pojavili znaci deformacije, izduženja ili habanja, poput neujednačenih brazda ili tamnih mrlja. Uz ove indikacije da uređaji ne mogu da ispune svoju predviđenu namjenu sa potrebnim nivoom sigurnosti, instrumente treba odložiti u otpad.
● Instrumenti se ne smiju kompletni potapati u rastvor natrijum hipohlorita (NaOCl). Samo radni dio NiTi instrumenta, koji je u kontaktu sa pacijentom, treba da bude uronjen u koncentrat NaOCl rastvora, i to NE veći od 5%.
● Budite oprezni u apikalnom području i kanalima koji se račvaju i/ili ispoljavaju nagle zakrivljenosti ili višestruke zakrivljenosti.
● Obilno i često irigirati kanal tokom postupka i nakon svakog korištenog instrumenta (u skladu sa dobrom stomatološkom praksom).
● Uvijek primjenjujte minimalni apikalni pritisak. Nikad ne uvodite turpijicu kroz kanal na silu.
● Kada instrument ne napreduje neometano, očistite i pregledajte rezne brazde, zatim irigirajte, rekapitulirajte ručnom turpijicom i ponovo irigirajte.
● Za oblikovanje izuzetno zakrivljenih kanala, sigurnije je koristiti turpijicu za oblikovanje samo jednog kanala kako bi se smanjila opasnost od lomljenja. Obratite pažnju na sljedeće dobre prakse:
○ Koristite novu turpijicu i odložite je kada se kanal obradi (upotreba za samo jedan kanal).
○ Koristite ručne umjesto rotacionih turpijica.
○ Koristite turpijice male veličine, fleksibilne i/ili NiTi turpijice (to pomaže da se izbjegne kanalna transportacija).
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 125/151
○ Vizuelno provjerite radni dio zbog svih oštećenja navedenih u prethodnom paragrafu tokom korišćenja (tj. nakon svakog zamaha).
○ Izbjegavajte standardni kontinuirani rotacioni pokret za proširivanje i umjesto toga koristite pokrete pod malim uglom (pokret turpijanja, oscilirajući pokret uvijanja sata ili tehniku balansirane sile) kako bi se ograničilo opterećenje instrumenata rotacionim savijanjem i poboljšao njihov očekivani radni vijek.
Za trenutno tehničko stanje do sada nije prijavljena nijedna neželjena reakcija.
Čuvajte proizvode u suvom i čistom okruženju, dalje od izvora vlage i direktne sunčeve svjetlosti.
Sve ProTaper Ultimate™ turpijice su odobrene za korišćenje pri brzini rotacije od 400 rpm i sa obrtnim momentom od 4 to 5.2 Ncm.
Instrumenti ProTaper Ultimate™ mogu se koristiti sa pokretima četkanja prema van u svim kanalima, posebno onim kanalima koji imaju nepravilan poprečni presjek, ili sa laganim kretnjama prema unutra, kako bi se progresivno napredovalo ka radnoj dužini (WL).
Uvijek postavljajte brusilicu u pregib između palca i kažiprsta. Izbjegavajte guranje, nego pustite da turpijice pasivno napreduju i prate kliznu putanju.
1) Pregledajte različite horizontalno angulirane radiografe da biste dijagnostikovali širinu, dužinu i zakrivljenost svih datih korijena i njegovog/njegovih kanala.
2) Pripremite pristup koji omogućava lako prepoznavanje svakog otvora kanala.
3) Pomoćna turpijica za oblikovanje, SX, može se koristiti kada postoji ograničen prostor ili za prethodno uvećanje tijela kanala ili za relokaciju kruničnog aspekta kanala dalje od spoljnog konkaviteta korijena.
4) U prisustvu lubrikanta, odaberite Slider i PASIVNO pratite kanal, u jednom ili više prolaza, do njegove završnice. Odredite WL (radnu dužinu) pomoću elektronskog apeks lokatora u kombinaciji sa radiografskim snimkom, a zatim potvrdite prohodnost.
5) Ako Slider ne dolazi lako do završnice kanala, odaberite manuelnu ručnu turpiju od nerđajućeg čelika (SS) manje veličine. U prisustvu lubrikanta kateterizujte ovaj kanal, utvrdite radnu dužinu, potvrdite prohodnost i verifikujte kliznu putanju. Sada ponovite gornji korak # 4.
6) U prisustvu NaOCl, izaberite Shaper i napredujte duž klizne putanje, u jednom ili više prolaza, dok se ne dostigne radna dužina (WL).
7) Nakon što je uklonjen Shaper, irigirajte, EndoActivate da razbijete debris i premjestite ga u rastvor, a zatim ponovo irigirajte da biste oslobodili debris.
8) Ponovo potvrdite radnu dužinu (WL), posebno u zakrivljenijim kanalima.
9) Izaberite FINISHER F1 (020.07) i pasivno pratite kanal do radne dužine (WL), u jednom ili više prolaza. Uklonite i pregledajte njene apikalne brazde. Kada se napune debrisom od dentina, priprema je gotova.
10) Ako je FINISHER F1 labav po dužini i njegove apikalne brazde nisu ispunjene debrisom, odaberite FINISHER F2 (025.08) i koristite ovu turpijicu na isti način kao što je opisano za
FINISHER F1.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 126/151
11) Ako je FINISHER F2 labav po dužini i njegove apikalne brazde nisu ispunjene debrisom, odaberite ručni pomoćni FINISHER F3 (030.09) i koristite ovu turpijicu na isti način kao što je opisano za FINISHER F1 i F2.
12) Nakon uklanjanja bilo koje date turpijice, očistite i pregledajte njene rezne brazde, irigirajte, rekapitulirajte turpijicom veličine 10 ili Endo Activatorom da biste razbili debris, a zatim ponovo irigirajte.
13) Nakon uklanjanja pregledajte rezne brazde turpijice, da li se odmotavaju, ispravljaju ili istežu. Ako primijetite deformaciju, odložite u otpad i uzmite novu ProTaper Ultimate™ turpijicu.
14) Priprema se završava kada se apikalni opseg bilo kog Finishera napuni debrisom, a odgovarajuća veličina glavnog gutaperka kočića (GPMC) ili Size Verifier odgovara radnoj dužini (WL).
Koristite pomoćne Finishere samo u većim i ravnijim kanalima, kao što su maksilarni centralni sjekutići, neki nepčani ili distalni kanali molara ili kada postoji patološki ili jatrogeni defekt:
15) Izaberite ili mehaničku ili ručnu pomoćnu turpijicu FINISHER FX (035.12) kada se utvrdi radna dužina i potvrdi prohodnost. Koristite ovu samostalnu turpijicu direktno i pasivno da biste pratili kanal do WL (radne dužine) koristeći istu tehniku koja je prethodno opisana.
Uklonite i pregledajte njene apikalne brazde. Kada se napune debrisom od dentina, priprema je gotova.
16) Ako je FINISHER FX labav po dužini i njegove apikalne brazde nisu opterećene debrisom, odaberite pomoćni FINISHER FXL (050.10) ili na mašinski pogon ili ručni i koristite ovu turpijicu na isti način kao što je opisano za FINISHER FX.
17) Priprema se završava kada se apikalni opseg bilo kog pomoćnog Finishera napuni debrisom, a odgovarajuća veličina glavnog gutaperka kočića (GPMC) ili Size Verifier odgovara radnoj dužini (WL).
● Pridržavajte se Opštih uputstava za obradu za endodontske proizvode ( https:// www.dentsplysirona.com/en/service-contact/ifu.html
) za korake za dezinfekciju, čišćenje i sterilizaciju.
● Ne koristiti ponovo silikonske stopere. Uklonite i odložite silikonske stopere nakon svake upotrebe.
● Proizvode treba odlagati u otpad u skladu sa lokalnim propisima za sigurno odlaganje oštrih i kontaminiranih uređaja.
● Svaki ozbiljni incident u vezi sa proizvodom treba prijaviti proizvođaču i nadležnom organu u skladu sa lokalnim propisima.
● Sterilnost se ne može garantovati ako je ambalaža otvorena, oštećena ili vlažna.
● Da biste dobili besplatnu štampanu kopiju Uputstava za upotrebu, pogledajte odjeljak
„Nabavite štampanu kopiju Uputstava za upotrebu“ na veb stranici https:// www.dentsplysirona.com/en/service-contact/ifu.html
.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 127/151
● Objašnjenje neharmonizovanih simbola za Uputstva za upotrebu i naljepnice, pogledajte
Simbole za Uputstva za upotrebu ( https://www.dentsplysirona.com/en/service-contact/ifu/ ifu-symbols.html
).
2797
Ovlašćeni predstavnik
DENTSPLY DeTrey GmbH
De-Trey-Straße 1
D-78467 Konstanz
Germany
Proizvođač
Maillefer Instruments Holding Sàrl
Chemin du Verger 3
CH-1338 Ballaigues
Suisse www.dentsplysirona.com/ifu
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 128/151
ProTaper Ultimate™
VETËM PËR PËRDORIM DENTAR
VETËM PËR PËRDORIM PROFESIONAL
STERIL – I STERILIZUAR ME RREZATIM
PËRGJEGJËS PËR RIPËRPUNIMIN
INSTRUMENTE ENDODONTIKE STERILE TË RIPËRDORSHME
● Produktet tregohen për trajtimin e sëmundjes endodontike.
● Qëllimi i synuar: Instrument me operim motorik me qëllim përgatitjen e kanalit të rrënjës
(rruga e kalimit, formësimi dhe prishja e kanalit të rrënjës).
● Instrumentet endodontike duhet të përdoren vetëm në një mjedis klinik ose spitalor, duke ndjekur praktikën e mirë dentare nga profesionistët e kualifikuar dentarë siç janë mjekët e përgjithshëm si dhe specialistët endodontikë (endodontistët) dhe ndihmësit dentarë.
● Instrumentet duhet të përdoren në kombinim e një motor endodontik (manualisht në përkuljet e ashpra).
Nuk ka.
Instrumentet ProTaper Ultimate™ janë prej tre komponentësh kryesorë: pjesa operuese prej aliazhi nikel-titani, një tapë silikoni me ngjyrë (përveç instrumentit SX) dhe një fyell tunxhi i praruar me unazë me ngjyrë (përveç instrumentit SX).
Limat prodhohen nëpërmjet një procesi që i shkakton limës një pamje me ngjyrë.
Për shkak të këtij përpunimi të patentuar, limat mund të shfaqen disi të lakuara. Ky nuk është një defekt prodhimi. Ndërkohë që lima mund të drejtohet lehtësisht duke përdorur vetëm gishtat, nuk është e nevojshme që ajo të drejtohet para përdorimit. Pasi të hyjë në kanal, lima do të ndjekë anatominë e kanalit të rrënjës.
Gama e ProTaper Ultimate™ përfshin këto instrumente endodontike:
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 129/151
Kodi
Gjatësia-xx
Madhësia e limës-y
Sasia-z për blister
B ST PULR z
19 0SX
B ST PULR z xx 0SL
B ST PULR z xx 0SH
Produkti
ProTaper Ultimate™
SX
ProTaper Ultimate™
SLIDER
ProTaper Ultimate™
SHAPER
ISO XXX YYYv
Madhësia ISO-XXX
Madhësia e turjelës-
YYY
Turjela e ndryshueshme-v
ISO 020.003v
ISO 016.002v
ISO 020.004v
Gjatësia e disponueshm e
19 mm
21/25/31 mm
21/25/31 mm
Paketimi dhe përmbajtja
Blister me 6 instrumente (sterile)
Blister me 3 instrumente (sterile)
Blister me 6 instrumente (sterile)
Blister me 3 instrumente (sterile)
Blister me 6 instrumente (sterile)
Blister me 3 instrumente (sterile)
Blister me 6 instrumente (sterile)
B ST PULR z xx 0F y
B ST PULR 6 xx yyy
B ST PULR 5 xx SQE
B DD PULR
525 SQE
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
Sequence
ProTaper Ultimate™
Demo Pack
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
FX: ISO 035.012v
FXL: ISO 050.010v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
21/25/31 mm
21/25/31 mm
21/25/31 mm
25 mm
Blister me 3 instrumente (sterile)
Blister me 6 instrumente (sterile)
Blister me 1 instrument nga secili instrument (steril)
Blister me 1 instrument nga secili instrument (steril)
● Pajisjet e kontrollit të çiftit rrotullues garantojnë përdorim optimal.
● Përdoreni me motorë endodontikë në një rrotullim konstante me një shpejtësi prej 400 rrpm.
● Vendosni çiftin rrotullues: 4 to 5.2 Ncm.
● Do të përdoren lubrifikantë si NaOCl, EDTA, ProLube, Glyde™.
● Përdorimi i radiografive në kombinim me një lokalizues kulmi dhe një mjet për përshtatjen e tapës së silikonit me gjatësinë e saktë funksionale është mënyra e duhur e përcaktimit të gjatësisë funksionale.
● Ndiqni në mënyrë rigoroze këtë udhëzim për përdorim dhe udhëzimet e përgjithshme të përpunimit për produktet endodontike (shih seksionin 8) për të minimizuar rreziqet e mëposhtme ndaj pajisjes, pacientit dhe/ose përdoruesit:
○ Thyerja e instrumentit.
○ Kontaminim i kryqëzuar.
○ Gjenerim nxehtësie për shkak të lubrifikimit të pamjaftueshëm dhe shpëlarjes.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 130/151
○ Gëlltitja e një pjese operuese të instrumentit.
○ Reaksione toksike ose alergjike të shkaktuara nga mbetjet përpunuese.
● Njësoj si të gjitha instrumentet e kanalit të rrënjës me operim makinerie, instrumentet
ProTaper Ultimate™ nuk duhen përdorur në një kanal rrënje me një përkulje apikale të prerë për shkak të rritjes së rrezikut të ndarjes. Në këtë rast, limat e dorës të lakuara paraprakisht duhen përdorur në regjionin apikal.
● Siguria dhe efikasiteti i përdorimit nuk janë përcaktuar te gratë shtatzëna ose që ushqejnë me gji apo te fëmijët.
● Përdorni një sistem dige gome gjatë procedurës endodontike.
● Për sigurinë tuaj, mbani pajisje personale mbrojtëse (doreza, syze, maskë).
● Kontrolloni paketimin para përdorimit dhe mos i përdorni instrumentet nëse është i dëmtuar paketimi.
● Mos i përdorni instrumentet pas datës së skadimit.
● Kontrolloni instrumentin para çdo përdorimi për shenja defektesh si deformime (shtypje, shtendosje), thyerje, gërryerje, buzë prerëse të dëmtuara, humbja e kodimit ose shënimit me ngjyrë. Me këto indikacione pajisjet nuk mund të përmbushin përdorimin e synuar me nivelin e kërkuar të sigurisë, instrumentet duhen hedhur.
● Para përdorimit të ndonjë instrumenti, sigurohuni që është i lidhur mirë me kokën e kundërkëndit.
● Kontrolloni instrumentin dhe pastrojeni shpesh pjesën operuese gjatë përdorimit të instrumenteve, duke kontrolluar për shenja deformimi, zgjatjeje ose konsumimi, siç janë hulli të pabarabarta, pikat e turbullta. Me këto indikacione pajisjet nuk mund të përmbushin përdorimin e synuar me nivelin e kërkuar të sigurisë, instrumentet duhen hedhur.
● Instrumentet nuk duhen zhytur plotësisht në një solucion me hipoklorit natriumi (NaOCl).
Vetëm pjesa operuese e instrumentit NiTi, e cila është në kontakt me pacientin duhet zhytur në një koncentrat solucioni NaOCI në një nivel JO më të madh se 5%.
● Bëni kujdes në zonën apikale dhe në kanalet që ndajnë dhe/ose shfaqin përkulje të prera ose ripërkulje.
● Shpëlani bollshëm dhe shpesh kanalin gjatë gjithë procedurës dhe pas përdorimit të çdo instrumenti (sipas praktikave të mira dentare).
● Përdorni gjithmonë presionin minimal apikal. Mos i fusni kurrë me zor limat në kanal.
● Kur instrumenti nuk ecën përpara lehtësisht, pastroni dhe kontrolloni hullitë prerëse, pastaj shpëlani, rinisni me një limë manuale dhe shpëlani sërish.
● Për formësimin e kanaleve tepër të harkuar, është më e sigurt ta përdorni limën vetëm për t'i dhënë formë një kanali, për të reduktuar rrezikun e thyerjes. Ndiqni me vëmendje praktikat e mira të mëposhtme:
○ Përdorni një limë të re dhe hidheni pasi të jetë trajtuar kanali (vetëm një përdorim për kanal).
○ Përdorni lima manuale në vend të atyre rrotulluese.
○ Përdorni lima të vogla, fleksibël ose/dhe prej nikel-titaniumi NiTi (kjo do të shmangë transportimin e kanalit).
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 131/151
○ Gjatë përdorimit, inspektoni me sy pjesën operuese për gjithë defektet e listuara në paragrafin e mëparshëm (p.sh. pas çdo vale).
○ Shmangni lëvizjen e vazhdueshme rrotulluese shpuese dhe përdorni më mirë lëvizje të vogla këndore (lëvizje limuese, lëvizje lëkundëse gjarpëruese ose teknikë me fuqi të balancuar) për të kufizuar sforcimin e përkuljes rrotulluese në instrumente dhe për të përmirësuar jetëgjatësinë e tyre.
Në gjendjen aktuale teknike, nuk është raportuar asnjë reaksion negativ deri tani.
Mbajini produktet në një mjedis të thatë e të pastër, larg burimeve të lagështisë dhe dritës së drejtpërdrejtë të diellit.
Të gjitha limat ProTaper Ultimate™ janë vërtetuar për t'u përdorur me 400 rrpm dhe çift rrotullues 4 to 5.2 Ncm.
Instrumentet ProTaper Ultimate™ mund të përdoren me një lëvizje fshirjeje në pjesën e jashtme të të gjitha kanaleve, veçanërisht në ato kanale që shfaqin një prerje të çrregullt tërthore ose me një lëvizje të lehtë çukitëse në brendësi për të avancuar në mënyrë progresive drejt gjatësisë operuese (WL).
Mbajeni gjithmonë instrumentin dentar mes gishtit të madh të dorës dhe gishtit tregues.
Shmangni shtyrjen, kështu që më mirë lini limat të avancojnë në mënyrë progresive dhe të ndjekin Slidepath.
1) Rishikoni radiografi të ndryshme me kënde horizontale për të përcaktuar nga ana diagnostike gjerësinë, gjatësinë dhe përkuljen e çdo rrënje të dhënë dhe kanalit (kanaleve) të saj.
2) Përgatisni një hyrje që mundëson identifikimin e lehtë të secilës grykë të kanalit.
3) Lima ndihmëse e formësimit, SX, mund të përdoret kur ka një hapësirë kufizuese se për të zmadhuar trupin e kanalit ose për të zhvendosur aspektin më kurorëzor të një kanali larg një lugëtie të jashtme të rrënjës.
4) Në praninë e një lubrifikanti, zgjidhni Slider dhe ndiqni PASIVISHT kanalin në një ose disa kalime për te kufiri fundor. Përcaktoni WL duke përdorur një lokalizues elektronik kulmi në kombinim me një imazh radiografik, pastaj konfirmoni kalueshmërinë.
5) Nëse Slider nuk do të arrijë lehtësisht kufirin fundor të kanalit, zgjidhni një limë dore manuale të vogël prej inoksi (SS). Në praninë e një lubrifikanti kateterizoni këtë kanal, vendosni gjatësinë operuese, konfirmoni kalueshmërinë dhe verifikoni Slidepath. Tani, përsëritni hapin nr. 4 të mësipërm.
6) Në praninë e NaOCI, zgjidhni Shaper dhe avanconi përgjatë Slidepath, në një ose shumë kalime, derisa të arrihet WL.
7) Pas heqjes së Shaper, shpëlani, EndoActivate për të shpërndarë mbetjet dhe kaloni te solucioni, pastaj shpëlani sërish për të nxjerrë mbetjet.
8) Rikonfirmoni WL, veçanërisht në kanalet më të lakuara.
9) Zgjidhni FINISHER F1 (020.07) dhe ndiqni pasivisht kanalin në WL, në një ose më shumë kalime. Hiqni dhe kontrolloni hullitë e tij apikale. Kur është e ngarkuar me mbetjet dentinale, përgatitja ka përfunduar.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 132/151
10) Nëse FINISHER F1 është i lirë në gjatësi dhe hullitë apikale nuk janë të ngarkuara me mbetje, zgjidhni FINISHER F2 (025.08) dhe përdorni këtë limë në të njëjtën mënyrë që përshkruhet më sipër për FINISHER F1.
11) Nëse FINISHER F2 është i lirë në gjatësi dhe hullitë apikale nuk janë të ngarkuara me mbetje, zgjidhni FINISHER F3 (030.09) dhe përdorni këtë limë në të njëjtën mënyrë që përshkruhet më sipër për FINISHER F1 dhe F2.
12) Pas heqjes së çdo lime, pastroni dhe kontrolloni hullitë prerëse, shpëlani, rinisni qoftë me një limë të madhësisë 10 ose Endo Activator për të shpërndarë mbetjet, pastaj shpëlani sërish.
13) Kontrolloni hullitë prerëse të limës pas heqjes për praninë e shkurdisjes, drejtimit ose tendosjes. Nëse vihet re deformim, hidhni dhe përdorni një limë të re ProTaper Ultimate™.
14) Përgatitja përfundon kur shtrirja apikale e çdo Finisher ngarkohet me mbetje dhe GMPC
(Koni Mjeshtër Gutta-Percha) ose Size Verifier me madhësi të barabartë përputhet në WL.
Përdorni Finishers ndihmës vetëm në kanalet më të mëdha dhe më të drejtë, siç janë dhëmbët prerës qendrorë maksilarë, disa kanale palatale ose distale të molarëve ose kur ka një defekt patologjik ose jatrogjenik:
15) Zgjidhni një limë ndihmëse me operim mekanik ose manual, FINISHER FX (035.12), kur vendoset gjatësia funksionale dhe konfirmohet kalueshmëria. Përdorni këtë limë të pavarur direkt dhe pasivisht për të ndjekur kanalin në WL duke përdorur të njëjtën teknikë siç përshkruhet më sipër. Hiqni dhe kontrolloni hullitë e tij apikale. Kur është e ngarkuar me mbetjet dentinale, përgatitja ka përfunduar.
16) Nëse FINISHER FX është i lirë në gjatësi dhe flautet apikale nuk janë të ngarkuara me mbetje, zgjidhni FINISHER FXL (050.10) ndihmës me operim mekanik ose manual dhe përdorni këtë limë në të njëjtën mënyrë që përshkruhet më sipër për FINISHER FX.
17) Përgatitja përfundon kur shtrirja apikale e çdo Finisher ndihmës ngarkohet me mbetje dhe
GMPC (Koni Mjeshtër Gutta-Percha) ose Size Verifier me madhësi të barabartë përputhet në WL.
● Ndiqni Udhëzimet e Përgjithshme të Përpunimit për Produktet Endodontike ( https:// www.dentsplysirona.com/en/service-contact/ifu.html
) për dezinfektimin, pastrimin dhe hapat e sterilizimit.
● Mos përdorni sërish ndalesa silikoni. Hiqni dhe hidhni ndalesat e silikonit pas çdo përdorimi.
● Produktet duhet të asgjësohen në përputhje me rregulloret lokale për asgjësimin e sigurt të pajisjeve të mprehta dhe të ndotura.
● Çdo incident i rëndë në lidhje me produktin, duhet t'i raportohet prodhuesit dhe autoritetit kompetent përkatës sipas rregullave vendore.
● Steriliteti nuk mund të garantohet nëse paketimi është i hapur, i dëmtuar ose i lagur.
● Për të marrë një kopje të shtypur falas të Udhëzimit për Përdorim (IFU), shikoni pjesën
"Merrni një kopje të shtypur të IFU" në faqen e internetit https://www.dentsplysirona.com/en/ service-contact/ifu.html
.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 133/151
● Shpjegimi i simboleve jo të harmonizuara për IFU-të dhe etiketat, shih Simbolet IFU ( https:// www.dentsplysirona.com/en/service-contact/ifu/ifu-symbols.html
).
2797
Përfaqësuesi i autorizuar
DENTSPLY DeTrey GmbH
De-Trey-Straße 1
D-78467 Konstanz
Germany
Prodhuesi
Maillefer Instruments Holding Sàrl
Chemin du Verger 3
CH-1338 Ballaigues
Suisse www.dentsplysirona.com/ifu
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 134/151
ProTaper Ultimate™
ΜΌΝΟ ΓΙΑ ΟΔΟΝΤΙΑΤΡΙΚΉ ΧΡΉΣΗ
ΜΌΝΟ ΓΙΑ ΕΠΑΓΓΕΛΜΑΤΙΚΉ ΧΡΉΣΗ
ΑΠΟΣΤΕΙΡΩΜΈΝΟ - ΑΠΟΣΤΕΙΡΏΝΕΤΑΙ ΜΕ ΑΚΤΙΝΟΒΟΛΊΑ
ΜΕ ΔΥΝΑΤΌΤΗΤΑ ΕΠΑΝΕΠΕΞΕΡΓΑΣΊΑΣ
ΕΠΑΝΑΧΡΗΣΙΜΟΠΟΙΉΣΙΜΑ ΑΠΟΣΤΕΙΡΩΜΈΝΑ ΕΝΔΟΔΟΝΤΙΚΆ ΕΡΓΑΛΕΊΑ
● Τα προϊόντα προορίζονται για τη θεραπεία ενδοδοντικών παθήσεων.
● Προβλεπόμενος σκοπός: Μηχανοκίνητο εργαλείο που προορίζεται για την προετοιμασία
του ριζικού σωλήνα (οδός πρόσβασης, διαμόρφωση και χειρουργικός καθαρισμός του
ριζικού σωλήνα).
● Τα ενδοδοντικά εργαλεία πρέπει να χρησιμοποιούνται μόνο σε κλινικό ή νοσοκομειακό
περιβάλλον τηρώντας την ορθή οδοντιατρική πρακτική από εξειδικευμένους επαγγελματίες
οδοντιατρικής, όπως οδοντίατροι, ειδικοί ενδοδοντολογίας (ενδοδοντολόγοι) και βοηθοί
οδοντιάτρων.
● Τα εργαλεία πρέπει να χρησιμοποιούνται σε συνδυασμό με μοτέρ ενδοδοντίας (χειροκίνητα
σε περιπτώσεις έντονης καμπυλότητας).
Καμία.
Τα εργαλεία ProTaper Ultimate™ αποτελούνται από τρία βασικά μέρη: το μέρος εργασίας που
είναι κατασκευασμένο από κράμα νικελίου-τιτανίου, έναν χρωματιστό αναστολέα σιλικόνης
(εξαιρουμένου του οργάνου SX) και ένα στέλεχος από ορείχαλκο με επίστρωση με χρωματιστό
δακτύλιο (εξαιρουμένου του οργάνου SX).
Οι ρίνες κατασκευάζονται με μια διαδικασία, έτσι ώστε η ρίνη να έχει χρωματιστή όψη.
Λόγω αυτής της αποκλειστικής επεξεργασίας, οι ρίνες ενδέχεται να φαίνονται ελαφρώς κυρτές.
Δεν πρόκειται για κατασκευαστικό ελάττωμα. Αν και η ρίνη μπορεί εύκολα να ισιώσει μόνο με τα
δάχτυλά σας, δεν χρειάζεται να την ισιώσετε πριν την χρησιμοποιήσετε. Μόλις εισέλθει στον
ριζικό σωλήνα, η ρίνη θα ακολουθήσει την ανατομία του σωλήνα.
Η σειρά ProTaper Ultimate™ περιλαμβάνει τα ακόλουθα ενδοδοντικά εργαλεία:
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 135/151
Κωδικός xx-μήκος y-μέγεθος
ρίνης z-ποσότητα
ανά κυψέλη
(blister)
B ST PULR z
19 0SX
B ST PULR z xx 0SL
B ST PULR z xx 0SH
Προϊόν
ProTaper Ultimate™
SX
ProTaper Ultimate™
SLIDER
ProTaper Ultimate™
SHAPER
ISO XXX YYYv
XXX-ISO μέγεθος
YYY-μέγεθος κώνου v-μεταβλητός κώνος
ISO 020.003v
ISO 016.002v
ISO 020.004v
Διαθέσιμα
μήκη
19 mm
21/25/31 mm
21/25/31 mm
Συσκευασία και περιεχόμενα
Κυψέλη με 6 εργαλεία (αποστειρωμένα)
Κυψέλη με 3 εργαλεία (αποστειρωμένα)
Κυψέλη με 6 εργαλεία (αποστειρωμένα)
Κυψέλη με 3 εργαλεία (αποστειρωμένα)
Κυψέλη με 6 εργαλεία (αποστειρωμένα)
Κυψέλη με 3 εργαλεία (αποστειρωμένα)
Κυψέλη με 6 εργαλεία (αποστειρωμένα)
B ST PULR z xx 0F y
B ST PULR 6 xx yyy
B ST PULR 5 xx SQE
B DD PULR
525 SQE
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
Sequence
ProTaper Ultimate™
Demo Pack
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
FX: ISO 035.012v
FXL: ISO 050.010v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
21/25/31 mm
21/25/31 mm
21/25/31 mm
25 mm
Κυψέλη με 3 εργαλεία (αποστειρωμένα)
Κυψέλη με 6 εργαλεία (αποστειρωμένα)
Κυψέλη με 1 εργαλείο από κάθε είδος
εργαλείου (αποστειρωμένα)
Κυψέλη με 1 εργαλείο από κάθε είδος
εργαλείου (αποστειρωμένα)
● Οι συσκευές ελέγχου της ροπής διασφαλίζουν τη βέλτιστη χρήση.
● Χρήση με μοτέρ ενδοδοντίας με συνεχή περιστροφή με ταχύτητα 400 rpm.
● Ρύθμιση ροπής σε: 4 to 5.2 Ncm.
● Πρέπει να χρησιμοποιούνται λιπαντικά όπως NaOCl, EDTA, ProLube, Glyde™.
● Η χρήση ακτινογραφιών σε συνδυασμό με τον εντοπιστή ακρορριζίου και ένα εργαλείο
προσαρμογής του αναστολέα σιλικόνης στο σωστό μήκος εργασίας αποτελεί την κατάλληλη
μέθοδο καθορισμού του μήκους εργασίας.
● Να τηρείτε αυστηρά τις παρούσες Οδηγίες χρήσης και τις Γενικές οδηγίες προετοιμασίας για
προϊόντα ενδοδοντίας (ανατρέξτε στην ενότητα 8) για ελαχιστοποίηση των ακόλουθων
κινδύνων για τη συσκευή, τον ασθενή ή/και τον χρήστη:
○ Θραύση του εργαλείου.
○ Διασταυρούμενη μόλυνση.
○ Παραγωγή θερμότητας λόγω ανεπαρκούς λίπανσης και έκπλυσης.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 136/151
○ Κατάποση του μέρους εργασίας του εργαλείου.
○ Τοξικές ή αλλεργικές αντιδράσεις που προκαλούνται από την επεξεργασία των
υπολειμμάτων.
● Παρόμοια με όλα τα μηχανοκίνητα εργαλεία ριζικού σωλήνα, τα εργαλεία ProTaper
Ultimate™ δεν θα πρέπει να χρησιμοποιούνται σε ριζικό σωλήνα με έντονη καμπυλότητα
του ακρορριζίου λόγω του αυξημένου κινδύνου θραύσης. Σε αυτή την περίπτωση θα πρέπει
να χρησιμοποιούνται ρίνες χειρός με προσχηματισμένη καμπυλότητα στην ακρορριζική
περιοχή.
● Η ασφάλεια και η αποτελεσματικότητα χρήσης δεν έχουν αποδειχθεί σε εγκύους ή γυναίκες
που θηλάζουν ή σε παιδιά.
● Χρησιμοποιήστε σύστημα ελαστικού απομονωτήρα κατά τη διάρκεια της ενδοδοντικής
επέμβασης.
● Για τη δική σας ασφάλεια, χρησιμοποιείτε μέσα ατομικής προστασίας (γάντια, γυαλιά,
μάσκα).
● Επιθεωρήστε τη συσκευασία πριν από τη χρήση και μην χρησιμοποιήσετε τα εργαλεία εάν
η συσκευασία είναι κατεστραμμένη.
● Μην χρησιμοποιείτε τα εργαλεία μετά την ημερομηνία λήξης.
● Ελέγξτε το εργαλείο πριν από κάθε χρήση για τυχόν ελαττώματα, όπως παραμορφώσεις
(κάμψη, εκτύλιξη), θραύση, διάβρωση, κατεστραμμένα κοπτικά άκρα, απώλεια της
χρωματικής κωδικοποίησης ή της σήμανσης. Εάν ισχύει κάτι από αυτά, οι συσκευές δεν
μπορούν να καλύπτουν τη χρήση για την οποία προορίζονται με το απαιτούμενο επίπεδο
ασφάλειας, τα εργαλεία θα πρέπει να απορρίπτονται.
● Πριν από τη χρήση οποιουδήποτε οργάνου, βεβαιωθείτε ότι έχει συνδεθεί σωστά στην
κεφαλή της γωνιακής χειρολαβής.
● Πρέπει να ελέγχετε το εργαλείο και να καθαρίζετε το μέρος εργασίας συχνά κατά τη
διάρκεια της προετοιμασίας των εργαλείων, επιθεωρώντας για σημεία παραμόρφωσης,
επιμήκυνσης ή φθοράς, όπως ανομοιόμορφες αύλακες, θαμπά σημεία. Εάν ισχύει κάτι από
αυτά, οι συσκευές δεν μπορούν να καλύπτουν τη χρήση για την οποία προορίζονται με το
απαιτούμενο επίπεδο ασφάλειας, τα εργαλεία θα πρέπει να απορρίπτονται.
● Τα εργαλεία δεν θα πρέπει να εμβαπτίζονται πλήρως σε διάλυμα υποχλωριώδους νατρίου
(NaOCl). Μόνο το τμήμα εργασίας του εργαλείου NiTi, το οποίο έρχεται σε επαφή με τον
ασθενή, θα πρέπει να εμβαπτίζεται σε πυκνό διάλυμα NaOCl συγκέντρωσης ΟΧΙ πάνω
από 5%.
● Προσέξτε στην περιοχή του ακρορριζίου και στους ριζικούς σωλήνες που διαχωρίζονται ή/
και παρουσιάζουν έντονες καμπυλότητες ή επαναλαμβανόμενες καμπυλότητες.
● Κάντε άφθονους και συχνούς διακλυσμούς του σωλήνα σε όλη τη διάρκεια της επέμβασης
και μετά τη χρήση κάθε εργαλείου (σύμφωνα με τις ορθές οδοντιατρικές πρακτικές).
● Χρησιμοποιείτε πάντοτε ελάχιστη ακρορριζική πίεση. Μην ωθείτε ποτέ τις ρίνες με δύναμη
μέσα στον ριζικό σωλήνα.
● Όταν ένα εργαλείο δεν προωθείται εύκολα, καθαρίστε και επιθεωρήστε τις αύλακες κοπής,
κατόπιν πραγματοποιήστε διακλυσμό, ελέγξτε με μια ρίνη χειρός και επαναλάβετε τον
διακλυσμό.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 137/151
● Κατά τη διαμόρφωση ριζικών σωλήνων εξαιρετικά μεγάλης καμπυλότητας, είναι πιο
ασφαλές να χρησιμοποιείται η ρίνη μόνο για τη διαμόρφωση ενός ριζικού σωλήνα, έτσι ώστε να μειώνεται ο κίνδυνος θραύσης. Προσέξτε τις εξής ορθές πρακτικές:
○ Χρησιμοποιήστε καινούργια ρίνη και απορρίψτε τη μετά την υποβολή του ριζικού σωλήνα
σε θεραπεία (χρήση μόνο σε έναν ριζικό σωλήνα).
○ Χρησιμοποιήστε χειροκίνητες αντί για περιστροφικές ρίνες.
○ Χρησιμοποιήστε εύκαμπτες ρίνες μικρού μεγέθους ή/και ρίνες NiTi (με αυτόν τον τρόπο
αποτρέπεται η μετατόπιση του ριζικού σωλήνα).
○ Επιθεωρήστε οπτικά το τμήμα εργασίας για κάθε ελάττωμα που αναφέρθηκε στην
προηγούμενη παράγραφο κατά τη διάρκεια της χρήσης (δηλ. ύστερα από κάθε κίνηση
μέσα-έξω).
○ Αποφύγετε τη συνεχή περιστροφική κίνηση της τυπικής διάνοιξης και, αντ' αυτής,
χρησιμοποιήστε κινήσεις υπό μικρή γωνία (παλινδρομική κίνηση παρόμοια με το
κούρδισμα ρολογιού ή τεχνική ισορροπημένης δύναμης), προκειμένου να περιοριστεί η
κόπωση περιστροφικής κάμψης των εργαλείων και να παραταθεί η αναμενόμενη διάρκεια
ζωής τους.
Στην παρούσα τεχνική κατάσταση, δεν έχει αναφερθεί μέχρι σήμερα καμία ανεπιθύμητη
αντίδραση.
Διατηρείτε τα προϊόντα σε στεγνό και καθαρό περιβάλλον, μακριά από πηγές υγρασίας και το άμεσο ηλιακό φως.
Οι ρίνες ProTaper Ultimate™ έχουν εγκριθεί για χρήση με 400 rpm και ροπή 4 to 5.2 Ncm.
Τα εργαλεία ProTaper Ultimate™ μπορούν να χρησιμοποιούνται με κίνηση βουρτσίσματος
προς τα έξω σε όλους τους ριζικούς σωλήνες, ειδικά σε σωλήνες με ακανόνιστες διατομές, ή με ήπια κίνηση «ραμφίσματος» προς τα μέσα για να προωθούνται σταδιακά προς το μήκος
εργασίας (WL).
Πάντα να στηρίζετε τη χειρολαβή στη μεμβράνη μεταξύ αντίχειρα και δείκτη. Αποφεύγετε την ώθηση και προτιμήστε να αφήνετε τις ρίνες να προωθούνται παθητικά ακολουθώντας το
Slidepath.
1) Εξετάστε ακτινογραφίες που έχουν ληφθεί υπό διαφορετικές οριζόντιες γωνίες, για να
προσδιορίσετε διαγνωστικά το πλάτος, το μήκος και την καμπυλότητα κάθε ρίζας και του
ριζικού σωλήνα(-ων) της.
2) Προετοιμάστε ένα σημείο πρόσβασης που επιτρέπει την εύκολη αναγνώριση κάθε στομίου
του ριζικού σωλήνα.
3) Η βοηθητική ρίνη διαμόρφωσης, SX, μπορεί να χρησιμοποιηθεί όταν υπάρχει
περιορισμένος χώρος ή για να γίνει εκ των προτέρων μεγέθυνση του σώματος του σωλήνα ή για αλλαγή του ακραίου μυλικού τμήματος ενός σωλήνα μακριά από ΄την εξωτερική
κοιλότητα της ρίζας.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 138/151
4) Όταν υπάρχει λιπαντικό, επιλέξτε το Slider και ακολουθήστε ΠΑΘΗΤΙΚΑ τον σωλήνα με ένα ή περισσότερα περάσματα μέχρι το άκρο. Καθορίστε το μήκος εργασίας (WL)
χρησιμοποιώντας έναν ηλεκτρονικό εντοπιστή ακρορριζίου σε συνδυασμό με την
ακτινογραφία και κατόπιν επιβεβαιώστε τη βατότητα.
5) Εάν δεν είναι εύκολη η προσέγγιση του Slider στο άκρο του ριζικού σωλήνα, επιλέξτε μια
χειροκίνητη ρίνη χειρός μικρού μεγέθους από ανοξείδωτο ατσάλι (SS). Με τη χρήση
λιπαντικού, καθετηριάστε αυτόν τον σωλήνα, καθορίστε το μήκος εργασίας, επιβεβαιώστε
τη βατότητα και επαληθεύστε το Slidepath. Τώρα επαναλάβετε το βήμα 4 παραπάνω.
6) Εάν υπάρχει NaOCl, επιλέξτε το Shaper και προωθήστε το κατά μήκος του Slidepath με ένα ή περισσότερα περάσματα μέχρι να επιτευχθεί το WL.
7) Κατά την αφαίρεση του Shaper, πραγματοποιήστε διακλυσμό, το EndoActivate θα
διασπάσει τα υπολείμματα, και μετακινήστε το στο διάλυμα, κατόπιν πραγματοποιήστε εκ
νέου διακλυσμό για να απελευθερωθούν τα υπολείμματα.
8) Επιβεβαιώστε ξανά το WL, ειδικά σε σωλήνες με μεγαλύτερη κυρτότητα.
9) Επιλέξτε τη FINISHER F1 (020.07) και ακολουθήστε παθητικά το κανάλι έως το WL με ένα ή περισσότερα περάσματα. Αφαιρέστε και επιθεωρήστε τις ακρορριζικές αύλακες. Όταν
είναι γεμάτες οδοντινικά υπολείμματα, η προετοιμασία έχει ολοκληρωθεί.
10) Εάν η FINISHER F1 είναι χαλαρή σε όλο το μήκος και οι ακρορριζικές αύλακες δεν είναι
γεμάτες με υπολείμματα, επιλέξτε τη FINISHER F2 (025.08) και χρησιμοποιήστε αυτή τη
ρίνη κατά τον ίδιο τρόπο που περιγράφεται παραπάνω για τη FINISHER F1.
11) Εάν η FINISHER F2 είναι χαλαρή σε όλο το μήκος και οι ακρορριζικές αύλακες δεν είναι
γεμάτες με υπολείμματα, επιλέξτε τη FINISHER F3 (030.09) και χρησιμοποιήστε τη
συγκεκριμένη ρίνη κατά τον ίδιο τρόπο που περιγράφεται παραπάνω για τη FINISHER F1
και F2.
12) Κατά την αφαίρεση οποιασδήποτε ρίνης, καθαρίστε και επιθεωρήστε τις αύλακες κοπής,
πραγματοποιήστε διακλυσμό, ελέγξτε είτε με μια ρίνη μεγέθους 10 ή με το Endo Activator
για διάσπαση των υπολειμμάτων και κατόπιν πραγματοποιήστε εκ νέου διακλυσμό.
13) Κατά την αφαίρεση της ρίνης επιθεωρήστε τις αύλακες κοπής της για τυχόν εκτύλιξη, ίσιωμα ή επιμήκυνση. Αν παρατηρήσετε παραμόρφωση, απορρίψτε την και
χρησιμοποιήστε καινούργια ρίνη ProTaper Ultimate™.
14) Η προετοιμασία ολοκληρώνεται όταν η επέκταση ακρορριζίου οποιασδήποτε Finisher έχει
γεμίσει με υπολείμματα και ο κώνος γουταπέρκας γενικής χρήσης ή το Size Verifier
αντίστοιχου μεγέθους ταιριάζει στο WL.
Χρησιμοποιήστε τις βοηθητικές Finisher σε μεγαλύτερους και πιο ίσιους ριζικούς
σωλήνες μόνο, όπως των πρόσθιων κεντρικών τομέων της άνω γνάθου, σε
ορισμένους υπερώιους ή άπω ριζικούς σωλήνες των γομφίων ή όταν υπάρχει
παθολογική ή ιατρογενής βλάβη:
15) Επιλέξτε είτε μηχανοκίνητη ή χειροκίνητη βοηθητική ρίνη, FINISHER FX (035.12), όταν έχει
προσδιοριστεί το μήκος εργασίας και έχει επιβεβαιωθεί η βατότητα. Χρησιμοποιείστε αυτή
την αυτόνομη ρίνη απευθείας και παθητικά για να ακολουθήσετε τον σωλήνα έως το WL
χρησιμοποιώντας την ίδια τεχνική που περιγράφεται παραπάνω. Αφαιρέστε και
επιθεωρήστε τις ακρορριζικές αύλακες. Όταν είναι γεμάτες οδοντινικά υπολείμματα, η
προετοιμασία έχει ολοκληρωθεί.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 139/151
16) Εάν η FINISHER FX είναι χαλαρή σε όλο το μήκος και οι ακρορριζικές αύλακες δεν είναι
γεμάτες με υπολείμματα, επιλέξτε είτε μια μηχανοκίνητη ή μια χειροκίνητη βοηθητική
FINISHER FXL (050.10) και χρησιμοποιήστε τη συγκεκριμένη ρίνη κατά τον ίδιο τρόπο
που περιγράφεται παραπάνω για τη FINISHER FX.
17) Η προετοιμασία ολοκληρώνεται όταν η επέκταση ακρορριζίου οποιασδήποτε βοηθητικής
Finisher έχει γεμίσει με υπολείμματα και ο κώνος γουταπέρκας γενικής χρήσης ή το Size
Verifier αντίστοιχου μεγέθους ταιριάζει στο WL.
● Θα πρέπει να τηρείτε τις Γενικές οδηγίες προετοιμασίας για προϊόντα ενδοδοντίας ( https:// www.dentsplysirona.com/en/service-contact/ifu.html
) για τα βήματα απολύμανσης,
καθαρισμού και αποστείρωσης.
● Μην επαναχρησιμοποιείτε τα στοπ σιλικόνης. Αφαιρέστε και απορρίψτε τα στοπ σιλικόνης
μετά από κάθε χρήση.
● Τα προϊόντα πρέπει να απορρίπτονται σύμφωνα με τους τοπικούς κανονισμούς σχετικά με
την ασφαλή απόρριψη αιχμηρών και μολυσμένων συσκευών.
● Κάθε σοβαρό περιστατικό σε σχέση με το προϊόν θα πρέπει να αναφέρεται στον
κατασκευαστή και την αρμόδια αρχή σύμφωνα με τους τοπικούς κανονισμούς.
● Δεν παρέχονται εγγυήσεις για την αποστείρωση σε περίπτωση που η συσκευασία είναι
ανοιγμένη, κατεστραμμένη ή υγρή.
● Για ένα δωρεάν έντυπο αντίγραφο των Οδηγιών χρήσης, ανατρέξτε στην ενότητα «Λήψη έντυπου αντιγράφου των Οδηγιών χρήσης» στην ιστοσελίδα https:// www.dentsplysirona.com/en/service-contact/ifu.html
.
● Επεξήγηση των μη εναρμονισμένων συμβόλων για οδηγίες χρήσης και ετικέτες, ανατρέξτε
στα Σύμβολα των Οδηγιών χρήσης ( https://www.dentsplysirona.com/en/service-contact/ifu/ ifu-symbols.html
).
2797
Εξουσιοδοτημένος αντιπρόσωπος
DENTSPLY DeTrey GmbH
De-Trey-Straße 1
D-78467 Konstanz
Germany
Κατασκευαστής
Maillefer Instruments Holding Sàrl
Chemin du Verger 3
CH-1338 Ballaigues
Suisse www.dentsplysirona.com/ifu
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 140/151
ProTaper Ultimate™
EXCLUSIVAMENTE PARA USO DENTÁRIO
APENAS PARA USO PROFISSIONAL
ESTÉRIL – ESTERILIZADO POR RADIAÇÃO
RESPONSÁVEL PELO REPROCESSAMENTO
INSTRUMENTOS ENDODÔNTICOS ESTÉREIS REUTILIZÁVEIS
● Os produtos são indicados para tratamento de doença endodôntica.
● Finalidade prevista: instrumento motorizado concebido para a preparação do canal radicular (via de permeabilidade, modelagem e desbridamento do canal radicular).
● Os instrumentos endodônticos devem ser utilizados somente em um ambiente clínico ou hospitalar, seguindo uma boa prática dentária, por profissionais dentários qualificados, como clínicos gerais, bem como por especialistas de Endodontia (Endodontistas) e assistentes dentários.
● Os instrumentos devem ser utilizados em conjunto com um motor endodôntico (manualmente em curvaturas acentuadas).
Nenhuma.
Os instrumentos ProTaper Ultimate™ são feitos de três componentes principais: a peça de trabalho, feita de uma liga de níquel-titânio, um stop de silicone colorido (exceto para o instrumento SX) e uma haste de latão cromado com um anel colorido (exceto para o instrumento SX).
As limas são fabricadas com um processo que resulta em uma lima com um aspeto colorido.
Devido a este processamento proprietário, as limas podem parecer ligeiramente curvas. Não é um defeito de fabricação. Embora as limas possam ser facilmente endireitadas apenas com os dedos do usuário, não é necessário endireitá-las antes de seu uso. Uma vez dentro do canal, a lima seguirá a anatomia do canal radicular.
A linha ProTaper Ultimate™ inclui os seguintes instrumentos endodônticos:
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 141/151
Código comprimento xx dimensões da lima y quantidade z por blister
B ST PULR z
19 0SX
B ST PULR z xx 0SL
B ST PULR z xx 0SH
Produto
ProTaper Ultimate™
SX
ProTaper Ultimate™
SLIDER
ProTaper Ultimate™
SHAPER
ISO XXX YYYv
Tamanho XXX-ISO
Tamanho YYY-cone cone variável v
ISO 020.003v
ISO 016.002v
ISO 020.004v
Comprimentos disponíveis
19 mm
21/25/31 mm
21/25/31 mm
Embalagem e conteúdo
Blister com 6 instrumentos (estéril)
Blister com 3 instrumentos (estéril)
Blister com 6 instrumentos (estéril)
Blister com 3 instrumentos (estéril)
Blister com 6 instrumentos (estéril)
Blister com 3 instrumentos (estéril)
Blister com 6 instrumentos (estéril)
B ST PULR z xx 0F y
B ST PULR 6 xx yyy
B ST PULR 5 xx SQE
B DD PULR
525 SQE
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
Sequence
ProTaper Ultimate™
Demo Pack
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
FX: ISO 035.012v
FXL: ISO 050.010v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
SLIDER:
ISO 016.002v
SHAPER:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
21/25/31 mm
21/25/31 mm
21/25/31 mm
25 mm
Blister com 3 instrumentos (estéril)
Blister com 6 instrumentos (estéril)
Blister com 1 instrumento de cada (estéril)
Blister com 1 instrumento de cada (estéril)
● Dispositivos de controle do torque asseguram um uso ideal.
● Utilizar com motores endodônticos em uma rotação constante, a uma velocidade de
400 rpm..
● Definir torque para: 4 to 5.2 Ncm.
● Deve-se utilizar lubrificantes como NaOCl, EDTA, ProLube, Glyde™.
● O uso de radiografias em combinação com um localizador apical e uma ferramenta para ajustar o stop de silicone ao comprimento de trabalho correto é o método apropriado para a determinação do comprimento de trabalho.
● Seguir rigorosamente estas Instruções de Uso e as Instruções Gerais de Processamento para Produtos Endodônticos (ver secção 8) para minimizar os seguintes riscos para o dispositivo, o paciente e/ou o usuário:
○ Quebra do instrumento.
○ Contaminação cruzada.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 142/151
○ Geração de calor devido a lubrificação e irrigação insuficientes.
○ Deglutição da peça de trabalho do instrumento.
○ Reações tóxicas ou alérgicas causadas por resíduos de processamento.
● De modo similar a todos os instrumentos motorizados para o canal radicular, os instrumentos ProTaper Ultimate™ não devem ser usados em um canal radicular com uma curvatura apical abrupta devido ao aumento do risco de separação. Neste caso, devem ser usadas limas manuais pré-curvadas na região apical.
● A segurança e a eficácia do uso não foram estabelecidas em mulheres grávidas, lactantes ou em crianças.
● Usar um sistema de dique de borracha durante o procedimento endodôntico.
● Para sua própria segurança, use equipamento de proteção individual (luvas, óculos, máscara).
● Inspecione a embalagem antes do uso e não use os instrumentos se ela estiver danificada.
● Não utilizar os instrumentos após a data de validade.
● Verificar o instrumento antes de cada uso para detectar sinais de defeitos, como deforma-
ções (dobrado, desenrolado), quebra, corrosão, arestas de corte danificadas, perda da cor da codificação ou da marcação. Com estas indicações, os dispositivos não conseguem cumprir o uso previsto com o nível de segurança exigido. Os instrumentos devem ser descartados.
● Antes de usar qualquer instrumento, certifique-se de que o mesmo está bem ligado à cabe-
ça contra-ângulo.
● Durante a instrumentação, verificar os instrumentos e limpar frequentemente a peça de trabalho, procurando sinais de distorção, alongamento ou desgaste, como estrias desiguais ou manchas baças. Com estas indicações de que os dispositivos não conseguem cumprir o uso previsto com o nível de segurança exigido, os instrumentos devem ser descartados.
● Os instrumentos não devem ser totalmente mergulhados em uma solução de hipoclorito de sódio (NaOCl). Apenas a peça de trabalho do instrumento NiTi que está em contato com o paciente deve ser imersa em uma solução de NaOCl concentrada de NÃO mais de 5%.
● Proceder com cuidado na zona apical e em canais que se dividam e/ou apresentem curvaturas abruptas ou múltiplas.
● Irrigar abundante e frequentemente o canal durante todo o procedimento e após o uso de cada instrumento (de acordo com as boas práticas dentárias).
● Exerça uma pressão apical mínima. Nunca force as limas no interior do canal.
● Quando o instrumento não progredir facilmente, limpar e inspecionar as ranhuras de corte, depois irrigar, repetir com uma lima manual e irrigar novamente.
● Para a instrumentação de canais extremamente curvos, é mais seguro usar a lima apenas para a instrumentação de um canal para reduzir o risco de fratura. Preste atenção às seguintes boas práticas:
○ Use uma lima nova e descarte-a após o tratamento do canal (uso em canal único).
○ Usar limas manuais ao invés de limas rotativas.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 143/151
○ Use limas pequenas, flexíveis e/ou de NiTi (isso ajudará a evitar o transporte do canal).
○ Inspecionar visualmente a peça de trabalho para detectar a presença dos defeitos indicados no parágrafo anterior durante o uso (ou seja, após cada movimento).
○ Evitar o movimento rotativo contínuo de alargamento padrão e, em vez disso, executar pequenos movimentos em ângulo (movimento da lima, movimento oscilatório horário e anti-horário) para limitar a fadiga de flexão rotativa dos instrumentos e aumentar a sua vida útil.
5)
6)
7)
1)
2)
3)
4)
5)
6)
7)
8)
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 144/151
9) Selecionar a FINISHER F1 (020.07) e seguir passivamente o canal até ao CT, em uma ou em mais passagens. Remover e inspecionar as respectivas ranhuras apicais. A preparação está concluída quando a lima estiver cheia de resíduos dentinários.
10) Se a FINISHER F1 estiver folgada em comprimento e as ranhuras apicais não estiverem cheias de resíduos, selecionar a FINISHER F2 (025.08) e utilizá-la tal como descrito acima para a FINISHER F1.
11) Se a FINISHER F2 estiver folgada em comprimento e as ranhuras apicais não estiverem cheias de resíduos, selecionar a FINISHER F3 (030.09) e utilizá-la tal como descrito acima para a FINISHER F1 e F2.
12) Ao remover qualquer lima, limpar e inspecionar as suas ranhuras de corte, irrigar, repetir com uma lima de tamanho 10 ou Endo Activator para quebrar os resíduos. Em seguida, irrigar novamente.
13) Inspecionar as ranhuras de corte da lima após a sua remoção para verificar a presença de desenrolamento, endireitamento ou alongamento. Se for observada deformação, descarte e use uma nova lima ProTaper Ultimate™.
14) A preparação está terminada quando a extensão apical de qualquer Finisher estiver carregada com resíduos e o Cone Gutta-Percha Master ou o verificador de tamanho correspondente couber no CT.
Utilizar os Finishers auxiliares apenas em canais maiores e mais retos, tais como incisivos centrais maxilares, alguns canais palatinos ou distais de molares, ou quando houver um defeito patológico ou iatrogênico:
15) Selecionar uma lima auxiliar manual ou mecânica FINISHER FX (035.12), quando se estabelecer o comprimento de trabalho e for possível confirmar a patência. Utilizar esta lima independente direta e passivamente para seguir o canal até ao CT, usando a mesma técnica descrita acima. Remover e inspecionar as respectivas ranhuras apicais. A preparação está concluída quando a lima estiver cheia de resíduos dentinários.
16) Se a FINISHER FX estiver solta em comprimento e as ranhuras apicais não estiverem cheias de resíduos, selecionar uma lima auxiliar FINISHER FXL (050.10) manual ou motorizada e utilizá-la tal como descrito acima para a FINISHER FX.
17) A preparação está terminada quando a extensão apical de qualquer Finisher auxiliar estiver carregada com resíduos, e o Cone Gutta-Percha Master ou o verificador de tamanho correspondente couber no CT.
● Seguir as Instruções Gerais de Processamento de Produtos Endodônticos ( https:// www.dentsplysirona.com/en/service-contact/ifu.html
) para os passos de desinfeção, limpeza e esterilização.
● Não reutilizar os stops de silicone. Remover e descartar os stops de silicone após cada uso.
● Os produtos devem ser eliminados de acordo com os regulamentos locais para a eliminação segura de dispositivos afiados e contaminados.
● Qualquer incidente grave relacionado com o produto deve ser comunicado ao fabricante e
à autoridade competente, de acordo com os regulamentos locais.
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 1
45/151
● A esterilidade não pode ser garantida caso a embalagem seja aberta, danificada ou molhada.
● Para obter uma cópia impressa das Instruções de Uso, ver a secção “Obter uma cópia impressa das Instruções de Uso” no website https://www.dentsplysirona.com/en/servicecontact/ifu.html
.
● Explicação dos símbolos não harmonizados para as instruções de uso e etiquetas, ver símbolos das Instruções de Uso ( https://www.dentsplysirona.com/en/service-contact/ifu/ifusymbols.html
).
2797
Representante autorizado
DENTSPLY DeTrey GmbH
De-Trey-Straße 1
D-78467 Konstanz
Germany
Fabricante
Maillefer Instruments Holding Sàrl
Chemin du Verger 3
CH-1338 Ballaigues
Suisse www.dentsplysirona.com/ifu
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 1
46-151
ProTaper Ultimate™
-
사용 방법
재사용 멸균 근관 치료기
제품 용도
● 해당 제품은 치근단 질환 치료에 쓰입니다 .
● 설계 용도 : 치근관 형성에 쓰이는 엔진 구동 기구 ( 글라이드 패스 , 치근관 성형 및 괴사 조직
제거 ).
● 근관 치료기는 임상 혹은 병원 환경에서 풍부한 진료 경험이 있는 근관 치료 전문의
(Endodontist) 및 치위생사를 비롯하여 일반의와 같은 자격을 갖춘 치과 전문가만이 사용해
야 합니다 .
● 기구는 근관 치료용 모터와의 조합으로 사용합니다 ( 매우 심한 곡률에서는 수동으로 ).
금기 사항
없음 .
내용물
구성 및 호환 장치
ProTaper Ultimate™ 기구는 세 가지 주요 부품인 니켈 티타늄 합금으로 만들어진 작동 부분 , 유
색의 실리콘 스토퍼 (SX 기구는 제외 ), 황동 생크를 도금한 유색의 고리로 (SX 기구는 제외 ) 구성
됩니다 .
파일의 제조 공정에는 유색의 외형을 만드는 공정이 포함되어 있습니다 .
이러한 전용 공정으로 인해 파일이 약간 구부러진 것처럼 보일 수 있습니다 . 이는 제조상 결함이
아닙니다 . 손가락으로 쉽게 파일을 곧게 만들 수 있지만 , 사용 전 곧게 만들 필요는 없습니다 . 일
단 파일이 근관 내로 진입하면 근관 구조를 따라갑니다 .
ProTaper Ultimate™ 세트는 다음 근관 치료기로 구성되어 있습니다 .
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 147/151
코드 xx길이 y파일 크기
블리스터 당 z-
양
B ST PULR z
19 0SX
B ST PULR z xx 0SL
B ST PULR z xx 0SH
제품
ProTaper Ultimate™
SX
ProTaper Ultimate™
SLIDER
ProTaper Ultimate™
Shaper
ISO XXX YYYv
XXX-ISO 크기
YYY테이퍼 크기 v가변 테이퍼
ISO 020.003v
ISO 016.002v
ISO 020.004v
이용 가능한 길
이
19 mm
21/25/31 mm
21/25/31 mm
포장 및 내용물
기구 6 개입 블리스터 포장 ( 멸균 )
기구 3 개입 블리스터 포장 ( 멸균 )
기구 6 개입 블리스터 포장 ( 멸균 )
기구 3 개입 블리스터 포장 ( 멸균 )
기구 6 개입 블리스터 포장 ( 멸균 )
기구 3 개입 블리스터 포장 ( 멸균 )
기구 6 개입 블리스터 포장 ( 멸균 )
B ST PULR z xx 0F y
B ST PULR 6 xx yyy
B ST PULR 5 xx SQE
B DD PULR
525 SQE
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
FINISHER
ProTaper Ultimate™
시퀀스
ProTaper Ultimate™
Demo Pack
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
FX: ISO 035.012v
FXL: ISO 050.010v
SLIDER:
ISO 016.002v
Shaper:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
SLIDER:
ISO 016.002v
Shaper:
ISO 020.004v
F1: ISO 020.007v
F2: ISO 025.008v
F3: ISO 030.009v
21/25/31 mm
21/25/31 mm
21/25/31 mm
25 mm
기구 3 개입 블리스터 포장 ( 멸균 )
기구 6 개입 블리스터 포장 ( 멸균 )
각 기구 개별 블리스터 포장 ( 멸균 )
각 기구 개별 블리스터 포장 ( 멸균 )
● 토크 제어 장치는 최적의 사용을 보장합니다 .
● 400 rpm 의 일정한 회전 속도를 내는 근관 치료용 모터와 함께 사용합니다 .
● 토크 설정 : 4 to 5.2 Ncm.
● 차아염소산나트륨 (NaOCl) 과 같은 윤활제 , EDTA, ProLube, Glyde™ 를 사용해야 합니다 .
● 근관 길이 측정기와 알맞은 치료 길이에 맞춰 실리콘 스토퍼를 조정하는 도구와 조합한 방사
선 사진의 사용은 치료 길이를 결정하는 올바른 방법입니다 .
경고
● 장치 , 환자 또는 사용자에게 다음 위험을 줄이기 위하여 본 근관 치료 제품 사용과 일반 처리
지침 (8 섹션 참조 ) 을 엄격하게 따르십시오 .
○ 기구의 파손 .
○ 교차 오염 .
○ 부족한 윤활제 및 불충분한 세척으로 인한 열 발생 .
○ 기구 중 작동 부분 삼킴 .
○ 잔여물 처리 과정에서 유발되는 독성 또는 알러지 반응 .
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 148/151
● 기계식으로 구동되는 모든 근관 기구와 마찬가지로 ProTaper Ultimate™ 기구 역시 분리 위
험이 높기 때문에 급격한 치근첨단 곡률이 있는 근관에 사용해서는 안 됩니다 . 이 경우에는
치근첨단 구역에 구부러진 핸드 파일을 사용해야 합니다 .
주의 사항
● 임산부 또는 수유 중인 여성 , 어린이에 대해 안전성과 효과는 입증되지 않았습니다 .
● 근관 치료 과정에서 러버댐 시스템을 사용하십시오 .
● 안전을 위해 개인 보호 장구 ( 장갑 , 보호경 , 마스크 ) 를 착용하십시오 .
● 사용 전 포장을 점검하고 , 포장에 손상이 있는 경우에는 기구를 사용하지 마십시오 .
● 사용 유효 기간이 지났다면 기구를 사용하지 마십시오 .
● 사용 전 기구의 변형 ( 구부러짐 , 풀림 ), 파손 , 부식 , 절단부 손상 , 색상 코드 또는 마크 손실과
같은 결함을 확인하십시오 . 앞선 예와 같은 현상이 보인다면 해당 기구는 필요한 안전 수준에
서 의도한 용도를 수행할 수 없으므로 폐기해야 합니다 .
● 모든 기구 사용 전에는 콘트라앵글 헤드에 파일이 잘 연결되었는지 확인합니다 .
● 고르지 않은 플루트 , 흐린 얼룩과 같은 변형 , 늘어짐 또는 마모를 확인하여 기구를 점검하고 ,
사용 중에는 시술부를 자주 세척합니다 . 앞선 예와 같은 현상이 보인다면 해당 기구는 필요한
안전 수준에서 의도한 용도를 수행할 수 없으므로 폐기해야 합니다 .
● 차아염소산나트륨 (NaOCl) 용액에 기구를 완전히 담그지 마십시오 . 환자와 접촉하는 NiTi 기
구의 작동 부분만 5% 의 농도를 넘지 않는 NaOCl 용액에 담가야 합니다 .
● 치근첨단부 및 근관 내 분할 또는 갑작스러운 곡률 또는 전향을 보이는 근관에서는 주의를 기
울입니다 .
● 과정 내내 및 각 기구를 사용한 후 ( 이상적인 치과 시술에 따라 ) 에는 충분히 그리고 빈번하게
근관을 물로 세척해 주어야 합니다 .
● 항상 최소의 치근첨단 압력을 사용하십시오 . 절대로 파일을 근관으로 밀어 넣지 마십시오 .
● 기구가 수월하게 진입이 되지 않을 때는 절단 플루트를 세척 및 점검하고 , 물로 세척 후 수동
파일로 진행하며 재세척하는 작업을 반복합니다 .
● 휘어진 각도가 매우 큰 근관 성형 시에는 파손의 위험을 줄이기 위해 파일만을 사용하여 한
개의 근관을 형성하는 것이 안전합니다 . 다음과 같은 올바른 관행에 주의를 기울이십시오 .
○ 새로운 파일을 사용하고 근관을 치료한 후에는 파일을 폐기하십시오 ( 근관 일회성 사용 ).
○ 회전 파일 대신 수동 파일을 사용합니다 .
○ 작은 크기의 플렉시블 또는 나이타이 (NiTi) 파일 ( 근관 변위를 막는 데 도움이 됨 ) 을 사용하
십시오 .
○ 사용 중에는 작동 부분에 대해 이전 단락에서 명시한 모든 결함을 육안으로 검사하십시오
( 예 : 각 웨이브 (wave) 후 ).
○ 기구의 회전굽힘 피로를 제한하고 기대 수명을 높이기 위해서는 표준 리밍 연속 회전 동작
을 피하고 대신 각도가 작은 동작 ( 충전 동작 , 시계 태엽감기 진동 동작 또는 BFT(balanced force technique)) 을 사용하십시오 .
부작용
현재의 기술 수준에서는 아직 보고된 부작용이 없습니다 .
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 149/151
보관 조건
제품은 습기가 없고 직사광선이 비치지 않는 청결한 장소에 보관하십시오 .
단계별 사용 지침
모든 ProTaper Ultimate™ 파일은 400 rpm 및 4 to 5.2 Ncm 토크에서 사용할 수 있는 검증을 마쳤
습니다 .
ProTaper Ultimate™ 기구는 작업 길이 (WL) 를 향해 점차 나아갈 수 있도록 모든 근관 , 특별히 불
규칙한 절단면에서 바깥 방향으로 빗질하는 동작 혹은 내부로 가벼운 상하 움직임을 주어 사용합
니다 .
핸드피스는 항상 엄지와 집게손가락 사이의 마디로 부드럽게 잡습니다 . 파일을 밀어넣기 보다는
슬라이드패스를 따라 조심스럽게 진행할 수 있도록 수동으로 진입합니다 .
1) 다양한 수평 각도의 방사선 사진을 검토하여 해당 근관의 너비 , 길이 및 곡률을 결정합니다 .
2) 각 근관 입구의 쉬운 식별을 가능하게 해주는 경로를 준비해 놓습니다 .
3) 보조 성형 파일 SX 는 공간이 제한적인 상황 또는 근관의 몸체를 사전에 확장하거나 근관의
관상 측을 외부 치근 함몰에서 멀리 재배치하는데 사용할 수 있습니다 .
4) 윤활제가 있는 상황에서 SLIDER 를 선택하고 , 하나 혹은 그 이상의 경로에서 근관의 말단까
지 조심스럽게 따라갑니다 . 방사선 이미지와 전자 극관 길이 측정기로 치료 길이 (WL) 를 결
정한 다음 개방성을 확인합니다 .
5) SLIDER 가 근관 말단부에 쉽게 닿지 않는다면 , 작은 크기의 스테인리스 강 (SS) 핸드 파일을
선택합니다 . 윤활제가 있는 상황에서 치료 길이의 확립 , 개방성 확인 , 슬라이드패스를 점검
합니다 . 이제 위의 4 단계를 반복합니다 .
6) 차아염소산나트륨이 있는 상황에서 Shaper 를 선택하고 , 하나 혹은 여러 경로에서 치료 길
이에 도달할 때까지 슬라이드패스를 따라 진행합니다 .
7) Shaper 를 제거하고 세척 후 EndoActivator 로 잔여물을 부수고 용액으로 옮겨 재새척하여 해
당 잔여물을 제거합니다 .
8) 특히나 근관이 더욱 만곡되어 있다면 치료 길이 (WL) 를 재확인합니다 .
9) FINISHER F1(020.07) 을 선택하고 하나 혹은 그 이상의 경로에서 치료 길이까지 조심스럽
게 근관을 따라갑니다 . 치근첨단 플루트를 제거하고 점검합니다 . 상아질 잔여물이 확보되었
다면 준비가 완료된 것입니다 .
10) FINISHER F1 의 길이에 여유가 있고 치근첨단 플루트에 잔여물이 적재되어 있다면 ,
FINISHER F2(025.08) 를 선택하여 해당 파일을 위의 FINISHER F1 을 설명한 것과 같은 방
식으로 적용합니다 .
11) FINISHER F2 의 길이에 여유가 있고 치근첨단 플루트에 잔여물이 적재되어 있다면
FINISHER F3(030.09) 를 선택하여 해당 파일을 위의 FINISHER F1 및 F2 를 설명한 것과 같
은 방식으로 적용합니다 .
12) 모든 파일을 제거한 후 곧바로 10 번 파일 또는 Endo Activator 로 잔여물을 제거하고 절단 플
루트 세척 , 물세척을 반복합니다 .
13) 제거 시에 풀림 , 펴짐 또는 늘어남이 있는지 파일의 절단 플루트를 점검합니다 . 변형된 부분
이 있는 경우 , 기존 파일을 폐기하고 새로운 ProTaper Ultimate™ 파일을 사용합니다 .
14) 치근첨단 외부의 어떤 FINISHER 라도 잔여물이 적재되어 있고 , 그에 맞는 크기의 구타 페르
카 마스터 콘 또는 Size Verifier 가 치료 길이와 맞는다면 준비가 완료된 것입니다 .
상악골 중앙 치아 , 일부 구개측 또는 어금니 말단 근관과 같은 더 크고 곧은 근관의 전용 또
는 증상이나 의사가 원인인 결함이 있을 때 보조 FINISHER 를 사용합니다 .
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 150/151
15) 치료 길이가 확립되고 개방성이 확인되면 기계식 구동 파일 혹은 수동 보조 파일 FINISHER
FX(035.12) 를 선택합니다 . 위의 설명과 같은 기법으로 해당 독립형 파일을 직접적으로 조심
스럽게 사용하여 치료 길이까지의 근관을 따라갑니다 . 치근첨단 플루트를 제거하고 점검합
니다 . 상아질 잔여물이 확보되었다면 준비가 완료된 것입니다 .
16) FINISHER FX 의 길이에 여유가 있고 치근첨단 플루트에 잔여물이 적재되어 있다면 기계식
구동 파일 혹은 수동 보조 파일 FINISHER FXL(050.10) 를 선택하여 해당 파일을 위의
FINISHER FX 를 설명한 것과 같은 방식으로 적용합니다 .
17) 치근첨단 외부의 어떤 보조 FINISHER 라도 잔여물이 적재되어 있고 , 그에 - 맞는 크기의 구
타 페르카 마스터 콘 또는 Size Verifier 가 치료 길이와 맞는다면 준비가 완료된 것입니다 .
위생
소독
세척 및 멸균
● 소독 , 세척 및 멸균을 위한 다음의 근관치료 제품 일반 처리 사용 지침 ( https:// www.dentsplysirona.com/en/service-contact/ifu.html
) 을 따라주십시오 .
● 실리콘 스톱은 재사용하지 마십시오 . 사용 후에는 실리콘 스톱을 제거하고 버리십시오 .
● 날카롭고 오염된 기기의 안전한 폐기를 위해 해당 지역 규정에 따라 제품을 폐기하십시오 .
추가 정보
● 제품과 관련된 모든 사고는 현지 규정에 따라 제조사 및 담당 기관에 보고되어야 합니다 .
● 포장재가 열려 있거나 , 손상되었거나 , 젖어 있으면 멸균을 보장할 수 없습니다 .
● IFU 의 무료 인쇄 사본을 얻으러면 웹사이트의 “IFU 의 인쇄 사본 얻기 ” 섹션을 참조하십시오
( https://www.dentsplysirona.com/en/service-contact/ifu.html
).
● IFU 용 불균형 기호 및 라벨에 대한 설명은 IFU 기호를 참조하십시오 ( https:// www.dentsplysirona.com/en/service-contact/ifu/ifu-symbols.html
).
2797
공인 대표
DENTSPLY DeTrey GmbH
De-Trey-Straße 1
D-78467 Konstanz
Germany
제조사
Maillefer Instruments Holding Sàrl
Chemin du Verger 3
CH-1338 Ballaigues
Suisse www.dentsplysirona.com/ifu
END-IFU-PROTAPER-ULTIMATE-ROTARY-REUSABLE-STERILE-V00-WEB-ROW-Multilingual-2021-04 151/151
Advertisement
Unique "variable taper" design creates a smooth, tapered canal that is ideal for obturation.
Color-coded for easy identification.
Packaged in sterile blisters for added convenience.
ProTaper Ultimate Rotary Files are made from a high-quality nickel-titanium alloy that is both strong and flexible.
The unique feature of ProTaper Ultimate Rotary Files is their "variable taper" design, which helps to create a smooth, tapered canal that is ideal for obturation.
Yes, ProTaper Ultimate Rotary Files are color-coded for easy identification.
What are the approved operating parameters for ProTaper Ultimate™ files?
All ProTaper Ultimate™ files are approved for use at a speed of 400 rpm and a torque of 4 to 5.2 Ncm.
What techniques are suitable for use with ProTaper Ultimate™ instruments?
ProTaper Ultimate™ instruments can be used with an outward brushing motion in all canals, especially those with irregular cross-sections, or with a light inward pecking motion to gradually move towards the working length (WL).
What precautions should be taken when using ProTaper Ultimate™ instruments to minimize the risk of complications?
Adhere strictly to the instructions for use and processing endodontic products to minimize the risk of instrument fracture, cross-contamination, and other complications. Avoid pushing the instrument, allowing it to progress passively.
Login to continue