advertisement
Sm
o
ker
lyzer ®
For use with Micro+ basic
F ormer piCO+
0086
Scientific contributions to health.
www.covita.net | (800)707-5751
Contents
Intended Use ...................................................................................................................................... 2
Introduction ........................................................................................................................................ 2
Taking a breath test ............................................................................................................................ 2
Converting %COHb to %FCOHb ................
...................................................................................... 3
Reviewing history ..................................................................................................................
.
.... 3
Changing date and time ..................................................................................................................... 3
Technical Specification .......................................................................................................
.
..........
4
Safety Information and Device Symbols ............................................................................................. 4
Environment ....................................................................................................................................... 5
Instrument Layout .............................................................................................................................. 5
Display and Device Symbols ............................................................................................................... 6
Warnings and Maintenance ............................................................................................................... 8
Cleaning .............................................................................................................................................. 8
Routine Maintenance ......................................................................................................................... 9
Troubleshooting ................................................................................................................................. 9
Calibration Procedure ...................................................................................................................... 10
COdata
+
™ Software – connecting to the PC ..................................................................................... 10
Returns Procedure ............................................................................................................................ 11
Spares ............................................................................................................................................... 11
Warranty .......................................................................................................................................... 11
www.covita.net | (800)707-5751
1
Intended Use
The M icro
+ basic
™ Smokerlyzer® is a breath carbon monoxide monitor intended for multi-‐patient use by healthcare professionals in smoking cessation programmes and research. They can also be us ed as an indicator of carbon monoxide poisoning in a healthcare environment.
Introduction
Carbon monoxide (CO) is a toxic, odourless, colourless and tasteless gas. It is formed by the incomplete combustion of organic material at high temperatures, with an insufficient oxygen s u pply. When inhaled, CO displaces oxygen in the bloodstream to form carboxyhaemoglobin (COHb). This starves body tissue of oxygen restricting the body’s perfor mance and causing health i m plications. CO can remain in the bloodstream for up to 24 hours, depending on factors such as physical activity, sex, and inhalation intensity. The half-‐life is about 5 hours.
CO (ppm), carboxyhaemoglobin (%COHb) and foetal carboxyhaemoglobin (%fCOHb) correlation:
Breath carbon monoxide is measured in parts per million (ppm) and blood carboxyhaemoglobin is measured in percentages (%COHb and %fCOHb). The three are compatible and convertible: the ppm
CO reading relates to gas in the lungs and on the breath (i.e. the amount of poisonous CO that has been inhaled), and the %COHb and %fCOHb readings relate to the percentage of vital oxygen that has been replaced in the bloodstream.
The Micro
+ basic
™ Smokerlyzer® displays CO ppm, %COHb and %FCOHb. However all devices on ly d ire ctly measure CO ppm, the %COHb and %fCOHb is a calculation based on clinical evidence. Clinical research has demonstrated that a useful relationship between carbon monoxide and carboxyhaemoglobin is obtained after a short period of breath-‐holding.
The cut-‐off point between smoker and non-‐smoker differs based on local regulation,
Bedfont have created cut-‐off levels as a guide only, see ‘Traffic Light Pre-‐set Cut off Levels’, these can be alt ered
in-‐line with your local regulations from the settings screen.
Taking a breath test
1.
Attach a breath sampling D-‐piece and new mouthpiece.
2.
Turn on the monitor by pressing the power button once.
3.
Press ‘Breath Test’ symbol on screen,
4.
To cancel the breath test, press . or .
5.
Inhale and hold breath for the pre-‐set 15 second countdown.
6.
A beep will sound during the last three seconds of the countdown.
7.
Blow slowly into mouthpiece, aiming to empty lungs completely.
8.
The ppm and equivalent %COHb levels will rise and hold on-‐screen.
2
9.
When the test is finished will appear at the bottom of the screen.
10.
If a high reading has been recorded, you can mute the sound by pressing .
11.
To repeat breath test, press once to return to the home screen and repeat steps 3-‐8.
12.
To save the reading press and select the relevant patient profile.
13.
Remove the D-‐piece between tests to purge sensor with fresh air.
14.
To switch off, press and hold the power button for 3 seconds, unit will also power off after 2 minutes of inactivity to save power.
Converting %COHb to %FCOHb –
Refer to the included Maternity Chart to convert the pregant mother's %COHb to
%FCOHb.
Reviewing history
The Micro
+ basic
™ will record every reading taken in its history up to 150 readings, to access this press
and then the history will then be shown.
Changing date and time
Press and then you can then select either d-‐m-‐y or m-‐d-‐y for the date format and 12h or
24h for the time format. Dates and times are then adjusted by selecting the number you wish to
change and pressing or to increase and decrease. Press to save your settings.
3
Traffic Light Pre-‐set Cut-‐Off Levels
Traffic light col o r Description
Green
Amber
1 Red
2 Red
3 Red
3 Red Flashing
Non-‐Smoker
Borderline
Reading (ppm)
0-‐6
7-‐9
Smoker – Low addicted 10-‐15
Smoker – Moderately addicted 16-‐25
Smoker – Heavily addicted 26-‐35
Smoker – Very heavy addicted 36+
Pregnant Smokers
Reading (ppm)*
0-3
4-6
7-10
11-15
16-25
26+
To adjust the pre-set cut off levels, press from the settings menu and drag the cut off levels up
and down to the desired ppm level. These settings will remain until adjusted again.
Technical Specification
Concentration Range (CO)
Display
Detection Principle
Repeatability
H
2
Cross Interference
Power
3 x AA (LR6 or Equivalent)
Approx. 3 x AA Battery Life
Micro+ ™ basic
0-‐ 15 0ppm
Full colo r touchscreen
Electrochemical sensor
<±5% on consecutive readings
Approx. 1 x CR2032 Battery Life
T
90
Response Time
Operating Temperature Range
Operating/Transport/Storage Pressure
Operating Humidity
Sensor Operating Life
Sensor Sensitivity
Sensor Drift
Dimensions
Weight
Monitor Construction
1 x CR2032 Lithium coin cell
1,000 uses on new alkaline batteries (approx.
3 min per use)
175,000 hours
≤30 seconds
15 -‐40°C (storage 0-‐50°C)
800-‐1200mbar
1 5 -‐90%non-‐condensing (storage 0-‐95%)
5 years – 5 year warranty
1ppm
<5% per annum
Approx. 34 x 75 x 140 mm
Approx. 200g (including batteries)
Case: Polycarbonate/ABS blend
SteriTouch anti-‐Microbial additive
D-‐Piece: Polypropylene
Safety Information and Device Symbols
Degree of protection against electric shock
Type of protection against electric shock
Degree of protection against ingress of liquid
Degree of safety application in the presence of a flammable anaesthetic mixture with air, oxygen or nitrous oxide
Refer to Manual
Type BF applied part
Internally powered equipment
IPXO -‐ not protected against water ingress
Equipment not suitable for use in the presence of flammable mixtures.
Direct Current
CE Mark
4
Type BF applied part
Dispose of according to WEEE
Serial Number
Manufacture by and Date
Bedfont Logo
Environment
The Micro
+ basic
™ Smokerlyzer® complies with the directive EN60601-‐1-‐2 electromagnetic compatibility but can be affected by cellular phones and by electromagnetic interference exceeding the levels specified in EN50082:1. This equipment should be moved if n
eces sary to avoid interference.
Instrument Layout
1.
Display
2.
Power button
3.
Breath sampling D-‐Piece™
4.
D-‐Piece™ aperture
5.
Exhaust port for breath sample
6.
USB connector (for use with COdata Software)
7.
Single-‐use SteriBreath™ mouthpiece
8.
Battery compartment
9.
Battery compartment clip
10.
Reset button
5
Display and Device Symbols
Description
Battery condition: Full
Battery condition: Low
Battery condition: Empty
Breath Test
Environmental Test
Settings
Patient Profiles
View patient results
Inhale
Hold Breath
Countdown Timer
Exhale ppm Reading
%COHb Reading
Show %fCOHb
%fCOHb Reading
Micro+ basic ™
6
Description
Ambient CO reading
Home
Change D-‐Piece
Change Traffic Light Cut-‐off Levels
Save
Change Breath Hold Time
History
Mute
Change date/time
Next Step
Countdown to Sensor Change
Change Sensor
Calibrate Device
Attach Flow Meter to Gas
Canister
Micro+ basic ™
7
Description
Attach Calibration Adaptor to D-‐Piece
Micro+ basic ™
Attach D-‐Piece to monitor and Turn on Gas Flow
Unit Calibrating
Calibration Successful
Calibration Failed
Retry Calibration
Firmware Version
6 Month Calibration Reminder
N/A = Not Available on this device
Warnings and Maintenance
1.
Mouthpieces should be replaced after every use.
2.
Hands should be washed regularly in accordance with infection control practice.
3.
Please do not attempt to modify the equipment in any way or use accessories not specified by the manufacturer. Any attempt to do so will invalidate the warranty and may compromise the safety of the device.
4.
Bedfont will make available upon request service training to appropriately qualified persons.
5.
Holding the reset button down for 30 seconds will perform complete device reset, this will
Cleaning clear any saved data and revert all settings to the factory defaults . After performing a reset the device will need to have the date/time set and be calibrated before it can be used.
1.
The Smokerlyzer® products are moulded with SteriTouch® technology for optimum infection control and bacterial efficacy. However Bedfont recommend wiping the instrument and D-‐ piece external surfaces with a product specifically developed for this purpose. Bedfont provides instrument cleaning wipes. The D-‐Piece cannot be sterilised.
2.
NEVER use alcohol or cleaning agents containing alcohol or other organic solvents as long term exposure to these vapours will damage the CO sensor inside.
3.
Under no circumstances should the instrument be immersed in liquid or splashed with liquid.
8
Routine Maintenance
1.
Replace batteries when indicated by the empty symbol .
2.
Bedfont recommend removal of the batteries when the device is not used for prolonged periods of time to prevent leakage.
3.
Replace breath sampling D-‐piece every 30 days or if visibly soiled or contaminated. The
Micro
+ basic
™ Smokerlyzer® will give a reminder during start-‐up when the D-‐piece should be replaced, see ‘Change D-‐Piece’ symbol.
4.
The sensor should be replaced every 5 years, 60 days prior to the sensor change ‘Countdown to sensor change’ symbol will be shown with the date on which the sensor should be changed. This can be ignored by pressing until the date at which the sensor should be changed arrives, at this point the ‘Change Sensor’ symbol will be shown. Change sensor if trained to by an approved Bedfont engineer or send to Bedfont or the local representative.
5.
Additional technical information can be made available on request; please contact Bedfont or its distributor.
6.
The Micro
+ basic
™ Smokerlyzer® requires calibration every 6 months this will be indicated by press the
Troubleshooting
and follow instructions to calibrate the device or press to ignore.
The unit fails to turn on:
If the unit fails to turn on, replace the batteries. Ensure that the batteries are inserted the correct way around, matching the symbols moulded into the plastic.
High readings recorded for reported non-‐smoker:
If a breath test on a non-‐smoker gives an indication of CO being present, it may indicate the following:
1.
A high ambient level of CO (possibly due to a faulty combustion appliance or exhaust fumes).
2.
The effects of passive smoking.
3.
The device is contaminated by alcohol or another organic substance.
4.
The patient may have a very high level of hydrogen on their breath, which can be generated in the human digestive system.
5.
The sensor may have drifted out of specification (see below).
The sensor has drifted out of specification:
Smokerlyzer® products are calibrated before leaving coVita , however we recommend if you suspect the instrument is reading incorrectly, try the test again with another device if available to get a comparison. Alternatively you check the function using coVita check gas or send back to coVita . The check gas required is coVita 20ppm carbon monoxide in air, procedure below.
1.
Ensure the fine control valve is in the off position.
2.
Screw the fine control valve and flow indicator assembly to the gas can. This is best done by screwing the gas can into the valve.
3.
Press ‘Breath Test’ symbol on screen, or , to start a breath test.
9
4.
When prompted by the ‘Exhale’ symbol, or regulator and allow the gas to flow at 1 litre per minute.
, open the fine control valve or
5.
Allow the gas to flow through the instrument for the duration of the test, again monitoring the rate of flow.
6.
If the final displayed value is less than 18ppm or higher than 22ppm, stop the test and perform a calibration or send back to coVita for a n evaluation.
Calibration Procedure
1.
Tu r n o n t h e m o n ti o r b y p r e s s i n g t h e p o w e r b u t t o n o n c e
2.
Press
3.
4.
St ar t t h e pr oc e e d c al i b r at i o n p r o c es s , p r e s s i ng th e sy m b o l 5 it m e s an d t h en onc e to
If t h e t e m p e r a t u r e o f t h e m o n it o r is t o o lo w t o c a lib r a t e , t h e b lu e t h e r m o m e t e r w ill b e
5.
sh o w n co n t a ct B e d f o
.
n
Tr y t ® ag o ai n
r yo u r l at e l o ca
r o n c e
l su p p t h e u n i il e r .
t h as w ar m e d u p – if t h is p r o b le m p e r s is t s p le a
If t h e t e m p e r a t u r e o f t h e m o n it o r is t o o h ig h t o c a lib r a t e , t h e r e d t h e r m o m e t e r w ill b e s e sh o w n . T r y a g a in la t e r o n c e t h e u n it h a s c o o le d dow n – if t h is p r o b le m p e r s is t s p le a s e co n t a ct B e d f o n t ® o r yo u r l o ca l su p p il e r .
6.
Once this has been successfully carried out the device will perform a zero; do not apply gas
7.
at this stage
Th e f ri s t st e p o f t h e ca il b r a it o n p r o ce ss w li l b e sh o w n o n scr e e n ; in s t r u c t io n s t o f in is h t h e c a lib r a t io n p r o c e s s f ol l ow th e o n s c re e n
8.
Gas flow should be set to 1.0 litres per minute
9.
A successful calibration will be indicated by the , press to return to the home screen
10.
A failed calibration will be indicated by , press to attempt calibration again – if the problem persists see ‘troubleshooting’ or call your local supplier of Bedfont® products
11.
Return to the home screen by pressing
COdata
+
™ Software – connecting to the PC
Place one end of the connection lead into the USB socket on the top of the Smokerlyzer® product, connecting the other end to the USB port on the PC. Before starting the software, ensure that the
Smokerlyzer® product is connected to the PC and switched on. Double click the COdata
+
™ icon on the PC to start the programme. Refer to the supplied documentation for how to operate COdata
+
™.
10
Returns Procedure
If your equipment requires servicing, please contact coVita Customer Service
Department, before ret urning any goods.
1.
When you have supplied the Customer Repairs Department with the monitor serial number and description of the fault, you will be issued with a returns/ ticket number. Please state the returns/ticket number on the outside of the box when returning the monitor, and ensure that your full details, telephone, fax numbers and return address are clearly stated.
2.
The product must also be decontaminated before it is returned according to your local regulations. Bedfont can provide you with a decontamination certificate to complete, which also needs to be attached to the outside of the box. Failure to do so will result in the product being subject to Bedfont decontamination procedure and will delay your service/repair.
3.
coVita advise that you use a courier service when returning monitors. This enables you to insure goods for loss or damage in transit. When your goods are received, you will be se nt an email stating so.
4.
If the device has been returned for repair it will then be examined and you will be sent an
‘Engineer’s Report’ and a quotation for the repair, which will include an authorisation form.
Complete the authorisation form, and ensure that you include the ‘Official Purchase Order
Number’. Please contact the Customer Repairs Department if you are unable to supply an
‘Official Purchase Order Number’.
5.
If your monitor is still in warranty and the fault is covered by warranty, see ‘Warranty’ section of this manual, coVita will repair it and return it to you with an ‘En ginee r’s Report’, free of charge plus shipping and handling.
.
6.
If you choose not to proceed with the repair, a handling fee will be charged. Ensure that you return the completed authori z ation form with an ‘Official Purchase Order Number’.
7.
The equipment will be returned to you as soon as coVita has received all relevant paperwork. A carriage fee will be charged if the monitor is no longer in warranty.
Spares
D-‐PIECE Box of 12 D-‐Pieces (non-‐sterile)
STERIBREATH Bo x of 250 SteriBreath Mouthpieces (non-‐sterile)
WIPE S Pack of 50 Cleaning Wipes
USB-3F
Warranty
USB Cable
coVita warrants the Micro
+ ba sic
™ Smokerlyzer® (excluding batteries) to be free of defects in materials and workmanship for a period of five years from the date of shipment. coVita ’s sole obligation under this warranty is limited to repairing or replacing, at its choice, any item covered under this warranty when such an item is returned, intact and prepaid, to coVita . These warranties are automatically invalidated if the products are repaired; altered, void labels removed or otherwise tampered with by unauthorised personnel, or have been subject to misuse, neglect or accident.
Never dispose of any electronic instrument or batteries in domestic waste. At the end of the product’s life, contact coVita for trade in credit and/or disposal instructions.
11
www.covita.net | (800)707-5751
Scientific contributions to health.
coVita
212 Cottage Grove Ave Suite C.
Santa Barbara, CA 93101
Tel:
(800) 707-5751
Fax: (800) 721-2377
Email: [email protected]
www.covita.net
© Bedfont Scientific Limited 2012
Issue 3 - March 201 7 , Part No: LAB679_USA
Bedfont Scientific Limited reserves the right to change or update this literature without prior notice.
Registered in: England and Wales. Registered No: 1289798
ISO 9001:2008
Cert No. FM 31664
ISO 13485:2003
Cert No. MD 502905
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement