BW/BWS-E series


Add to my manuals
24 Pages

advertisement

BW/BWS-E series | Manualzz

BW/BWS-E series

Indicator Scales

User’s guide

BW/BWS-

E v1.03

BW/BWS-E serial Indicator user’s manual

Table of Contents

SECTION 1

INTRODUCTION…………………………………………………….………1

SECTION 2 SPECIFICATIONS………………………………………………….…..……2

SECTION 3 INSTALLATION……………………………………………………………….3

3. 1 General installation

…………………………………………………………….………3

3. 2 Installation of BW/BWS-E series

…………………………………………….3

3. 3 Port connect

……………………………………………………………….……………4

3.3.1 Port location

……………………………………………………………….…………..4

3.3.2 Load cell port

…………………………………………………………….….…………5

3.3.3 RS-232

…………………………………………………………………………………5

3.3.4 Analog output and check weighing output

………………………………………….5

3.3.5 Mercury level bubble

…………………………………………………………………6

SECTION 4 KEY DESCRIPTIONS………………………………………….……………7

SECTION 5 DISPLAYS…………………………………………………………………….8

SECTION 6 OPERATION…………………….……………………………….……….…..9

6.1 Zeroing the display…………………………………………………………….……….9

6.2 Taring……………………………………………………………………………....…….9

6.3 Weighing a sample…….………………………………………………….…………….9

6.4 Check-weighi ng………………………………………………………………………….9

6.4.1 About check weighing

……………………………………………..………………….9

6.4.2 Set limit

………………………………………………………………..……………10

6.4.3 Set check weighing mode

…………………………………………………………10

6.4.4 Note

………………………………………………………………………….………..10

6.5 Accumulated total……………………………………………………………………10

6.5.1Note

……………………………………………………………………………………10

6.5.2 Accumulate operate

………………………………………………………..………10

6.5.3 Memory recall

……………………………………………………………..…………11

6.5.4 Memory clear

……………………………………………………………..…………11

6.5.5 Automatically accumulate

…………………………………………………………..11

6.6 animal

…………………………………………………………………………….11

6.7 keyboard lock

…………………………………………………………………………..11

6.8 set auto power off ……………………………………………………………………..11

SECTION 7 PARAMETERS………………………………………………………………12

SECTION 8 BATTERY OPERATION…………………………………………..……..…16

SECTION 9 RS-

232 OUTPUT…………………………………………………..………..17

9.1 basic information

……………………………………………………………….………17

9.2 normal output

……………………………………………………………………..……17

9.3 continuously output protocol

…………………………………………………….……17

SECTION 1

0 CALIBRATION……………………………………..……………….……..18

SECTION 11 ERROR CODE…………………………………….………………………20

- I -

BW/BWS-E serial Indicator user’s manual

SECTION 1 INTRODUCTION

The BW/BWS-E series of indicator provides an accurate, fast and versatile series of general purpose weighing scale with counting and check-weighing functions.

There are 3 series s within the range, the platform size from 350mm x 450mm to

600mm x 800mm, the capacity range from 30kg to 1000kg

All the displays are large easy to read liquid crystal type displays (LED)

All units include automatic zero tracking, audible alarm for pre-set weights, and an accumulation facility that allows the individual weights to be stored and recalled as an accumulated total.

- 1 -

BW/BWS-E serial Indicator user’s manual

SECTION 2 SPECIFICATIONS

Model SBW/BWS-E MBW/BWS-E LBW/BWS-E

Platform size 350mm x 450mm 420mm x 520mm 600mm x 800mm

Capacity 30kg/60kg/150kg 60kg/150kg/300kg 300kg/600kg/1000kg

Resolution

Interface

1:15.000

RS-232 Output Optional

Analog Output Optional

Stabilisation Time 1 Seconds typical

Operating

Temperature

Power supply

0°C - 40°C / 32°F - 104°F

External AC adapter, 9V 800mA

AC powered, 115V/230V

Calibration Automatic External

Display

1.2’’ LED

Balance Housing Indicator ABS Plastic / SST

Load cell drive voltage

Load cells

Max 5V/150mA

Up to four 350 ohms cells

- 2 -

BW/BWS-E serial Indicator user’s manual

SECTION 3 INSTALLATION

3. 1 GENERAL INSTALLATION

The s should be sited in a location that will not degrade the accuracy.

Avoid extremes of temperature. Do not place in direct sunlight or near air conditioning vents.

Avoid unsuitable tables. The tables or floor must be rigid and not vibrate. Do not place near vibrating machinery.

Avoid unstable power sources. Do not use near large users of electricity such as welding equipment or large motors.

Avoid high humidity that might cause condensation. Avoid direct contact with water.

Do not spray or immerse the s in water.

Avoid air movement such as from fans or opening doors. Do not place near open windows.

Keep the s clean.

Do not stack material on the s when they are not in use.

3. 2 INSTALLATION OF BW/BWS-E SERIES

The pillar is attached to the base using a bracket that must first be attached to the base frame using the 4 bolts supplied. The Pillar is secured to the bracket using 2 set screws. The cable from the base to the indicator module is run through the tube, out through the plastic support at the top. Excess cable can be stored within the tube.

The BW/BWS-E Series comes with a stainless steel platform packed separately.

Place the platform in the base.

Level the scale by adjusting the four feet. The scale should be adjusted such that the bubble in the spirit level is in the center of the level and the scale is supported by all four feet. If the scale rocks read just the feet.

Attach the indicator module to the pillar by sliding it over the bracket with the flanges engaged in the groves on the base. Attach the cable from the base to the connector on the rear of the indicator.

Attach the AC power adapter to the connector on the back of the indicator.

- 3 -

BW/BWS-E serial Indicator user’s manual

3. 3 PORT CONNECTION

3.3.1 port locations

BW/BWS-E (rear side)

Load cell

Analog output

RS-232

AC adapter

AC power socket

BWS (

Under Side

)

AC adapter

Load cell

- 4 -

RS-232

For pole neck

Analog output

BW/BWS-E serial Indicator user’s manual

3.3.2 load cell port

Load cell connect as below( 5pin air connecter)

EXC+

SEN+

SIG+

LOAD CELL

SIG-

SEN-

EXC-

3.3.3 RS-232

RS-232 (Air connector / D type connector)

RXD Input

TXD

GND

Output

3.3.4 Analog output and check weighing output

9 pin air connector pin1~pin6: check weighing output pin 1 hi (output) pin 2 ok (output) pin 3 low (output) pin 4 beep (output) pin 5 vcc (5V)(output) pin 6 com (gnd)

SHIELD

Receiving data

Transmission data

Signal ground

5

5

3

2

4

4

1

- 5 -

BW/BWS-E serial Indicator user’s manual pin6~pin9 analog output pin 6 com (gnd) pin 7 analog power input (10~32VDC, +) (input) pin 8 T+ (0~10V/4~20mA) (output) pin9 T- (0~10V/4~20mA) (output)

Check weighing output application sample (you need a additional output control

PCBA)

BW/BWS-E indicator

Output control PCBA

J1

N

AC power

L

3.3.5 Mercury level bubble

If you want to scale can

’t work when not level, you can add one mercury level bubble, connect 2 wire to K1, when scale not level, mercury level bubble short, indicator will show

“ err l ”, when this message appear, please adjust scale level.

If you don

’t need this function, just leave K1 unused.

- 6 -

BW/BWS-E serial Indicator user’s manual

SECTION 4 KEY DESCRIPTIONS

ON/ OFF

or

Turn on or off the power.

Zero

Set the zero point for all subsequent weighing. The display shows zero.

A secondary function of "Enter" key when set parameters or other functions.

Tare

Stores the current weight in memory as a tare value, subtracts the tare value from the weight and shows the results.

A secondary function incrementing the active digit when set a value for parameters or other functions.

Press

Zero

and

Tare

key together in normal weighing mode will turn on/off animal function. (When enter/escape animal mode, you will hear beeper on twice)

MR

Memory recall key, show total accumulate weight in memory.

In setting mode, this key used to move active digits right.

M+

Accumulate key, store current weight in memory.

In setting mode, this key used to move active digits left.

Press

M+

and

MR

key for MC (memory clear)

Print

For print the results to a PC or printer using the optional RS-232 interface.

Press Print during self checking will enter setting mode.

G/N

Gross weight/net weight shift key, Hold this key for 3 seconds to shift the unit.

Secondary function (ESC) , is to return to normal operation when the mode is in a parameter setting mode.

Press

G/N

and

Print

key together for setting check Weighing limit.

- 7 -

- 8 -

BW/BWS-E serial Indicator user’s manual

SECTION 5 DISPLAYS

The LED display will show a value. In addition there are LED lights for Zero, Stable,

Tare, Batt, Kg, U

2

.

BW/BWS-E serial Indicator user’s manual

SECTION 6 OPERATION

6.1 Zeroing The Display

You can press the

ZERO

key at any time to set the zero point from which all other weighing and counting is measured, within 4% of power up zero. This will usually only be necessary when the platform is empty. When the zero point is obtained the display will show the indicator for zero.

The scale has an automatic rezeroing function to account for minor drifting or accumulation of material on the platform. However you may need to press the

ZERO

key to rezero the scale if small amounts of weight are shown when the platform is empty.

6.2 Taring

Zero the scale by pressing the

ZERO

key if necessary. The zero indication will be on.

Place a container on the platform, a value for its weight will be displayed.

Press the

TARE

key to tare the scale. The weight that was displayed is stored as the tare value and that value is subtracted from the display, leaving zero on the display.

The "NET" indicator will be on. As product is added only the weight of the product will be shown. The scale could be tared a second time if another type of product was to be added to the first one. Again only the weight that is added after taring will be displayed.

When the container is removed a negative value will be shown. If the scale was tared just before removing the container this value is the gross weight of the container plus all product that was removed. The zero indicator will also be on because the platform is back to the same condition it was when the

ZERO

key was last pressed.

6.3 Weighing a sample

To determine the weight of a sample first tare the empty container then place the sample in the container. the display will show the weight and the units of weight currently in use.

6.4 Check-Weighing

6.4.1 About check-weighing

Check-weighing is a procedure to cause an alarm to sound when the weight on the meets or exceeds values stored in memory. The memory holds values for a high limit and a low limit.

- 9 -

BW/BWS-E serial Indicator user’s manual

Check mode OK:

When check range, the display will show OK and the beeper will sound when the weight is between the limits.

Check mode NG:

When check range, the display will show OK and the beeper will sound when the weight is out of the limits.

6.4.2 Set limits

Press

G/N

and

PRINT

key together, it will display “ set h ”, use

TARE

key to select

SET H

” or “

SET L

”, press

ZERO

key to enter, use

M+

and MR key to move active digit, use

TARE

key to change value, use

Print

key to clear value. After you enter the value, press

ZERO

key to sure, press

G/N

key to escape.

6.4.3 Set check weighing mode

Press G/N and print key together, it will display “ set h

”, use

TARE

key to select beep

, press

ZERO

key to enter, press

TARE

key to select ok

(check mode OK), ng

(check mode NG), no

(no beep), press

ZERO

key to sure, press

G/N

key to escape.

6.4.4 NOTE

Check weighing function only available when weight more than 20d (min).

To disable the Check-Weighing function enter zero into both limits by pressing the

PRINT

key when the current limits are shown then pressing ZERO/ENTER to store the zero values.

6.5 Accumulated Total

6.5.1 Note

The scale can be set to accumulate manually by pressing the

M+

key. See the

PARAMETERS Section for details of selecting the method using function "

P2 com

".

Please note before every accumulate operate, scale need return to zero, and only press

M+

key when stable, when weight less than 20d, accumulate operate will be invalid.

6.5.2 Accumulate operate

The weight displayed will be stored in memory when the

M+

key is pressed and the weight is stable.

The display will show "

ACC 1

" and then the total in memory for 2 seconds before returning to normal. (after do accumulate operate,

“M+” indicator will turn on) If the optional RS-232 interface is installed the weight will be output to a printer or PC.

Remove the weight, allowing the scale to return to zero and put a second weight on.

Press the

M+

key, the display will show "

ACC 2

" and then the new total.

- 10 -

BW/BWS-E serial Indicator user’s manual

Continue until all weights have been added.

6.5.3 memory recall

To view the totals in memory press

MR

key.

6.5.4 memory clear

To clear the memory, just press

M+

and

MR

together

6.5.5 automatically accumulate

Press Print key during self checking, enter setting mode, press Tare key until display show

P32COM

, press

ZERO

key to enter, press

TARE

key to select

“ mode ”, press

ZERO

key to sure, display will show current RS-232 mode, press TARE key to select

Auto

. After you set, AUTO indicator on.

Press weight on platform, after stable, you will hear beep on twice, you can add or remote weight now, scale will beep on again after stable, at last, remove all weight on platform, the last weight value will store in memory

6.6 Animal s

Let the animal on the platform, after some second, if reading data change not a lot, you can hear beep sound and reading data will be locked.

In reading data lock mode, if you add/remove big weight, display will still update and lock new reading data.

You can press

ZERO TARE

together to enter/escape animal weighing mode.

6.7 Keyboard lock

You can use keyboard lock when this function enable (see detail setting process in section 7), after keyboard haven

’t use for 10 minutes, keyboard will be lock, after enter lock status, if you press any key, display will show

K-LCK

”.

If you want to escape lock mode and return work mode, hold

PRINT

,

MR

,

ZERO

key 2 seconds, display will show

“ ULCK ”, return normal mode.

6.8 set auto power off

Hold

ZERO

key 3 second, display will show

” setbl “, press

ZERO

key to enter. press

TARE

key, display show

“ setoff ”, press

ZERO

key to enter auto power off setting, press

TARE

key to change auto power off time:0ff/on/3/5/15/30. ( of ON

: always on, of off

: always off, of xx

: auto power off after standby xx minutes), press

ZERO

key to sure, press

G/N

key to escape.

- 11 -

BW/BWS-E serial Indicator user’s manual

SECTION 7 PARAMETERS

The scale has 5 parameters that can be set by the user plus a method of entering the calibration section.

To set parameters press the

PRINT

key during self checking, display will show pn

(ask password), press

M+

,

G/N

,

TARE

to enter.

The display will show the first function, " p0 chk

".

Pressing the

TARE

key will cycle through the other functions.

Pressing

ZERO

will allow you to set the function. It may be necessary to either use

TARE

or set a value using the

M+ MR

key to move the active digit and then using the

TARE

key to increment a digit, followed by the

ZERO

key to enter the value. Use the

G/N

key to leave a parameter unchanged.

For example when the display shows “ p0 chk

” press the

ZERO

key to begin.

The display will show “ Set Lo ”, press the

ZERO

key to set the low limit, or press the

TARE

to skip to the next parameter, “

Set Hi

” for setting the high limit.

After pressing the

ZERO

key to set a limit, use the

M+ MR

keys to change the flashing digit, then use the

TARE

key to increment the flashing digit. Continue to the next digit and set it as needed.

When all digits have been set press the

ZERO

key to store the value. The display will go back to the param eter just set, i.e. “ Set Lo ”. Advance to another parameter if needed or press the

G/N

key to return to weighing.

Parameter setting table functio n

P0

CHK

Sub function

SET H describe

Set high limit, press

M+ MR

key to move active digit, press

TARE

to change value, press

PRINT

key to clear data, press

ZERO

key to sure.

SET LO

Set high limit, press

M+ MR

key to move active digit, press

TARE

to change value, press

PRINT

key to clear data, press

ZERO

key to sure.

BEEP

P1

REF

AZN 0

0AUTO

Set beep mode, no

: no beep for check Weighing

OK

: beep when weight between hi and low (OK) ng

: beep when weight out of hi-low range (NG)

This option is used to select the auto zero tracking range

Options : 0.5d, 1d, 2d, 4d

This option is used to select the auto zero range when turn the indicator.

Options : 0%, 2%, 5%, 10%, 20%, 50%, 100%

- 12 -

P2

COM

P3

Lin

P4

0RAGE

0-TARE

SPEED

ZERO

MODE

BAUD pr ptype lab prt

BW/BWS-E serial Indicator user’s manual

This option is used to select the manual zero range when press the

ZERO

key.

Options: 0%, 2%, 4%, 10%, 20%, 50%, 100%

This option is used to set whether Bench Scales will do auto zero tracking in net mode (after do tare operate, net weight is zero) Options: ON/OFF

Set the ADC speed

7.5/15/30/60 times/second

Set new zero point, after set this value, when in zero point, the reading will be

–xxx.xxx

This option is used to set RS-232 communication mode

Options:

CONT

(continuously send)

ST1

: send one frame data after stable

STC

: send data continuously when stable

PR1

: when press print key, send one frame data (printer mode)

PR2

: when press M+ key, do print data and M+ at the same time

AUTO

: auto accumulate (auto print) mode, when weight stable and then return to zero, indicator will do accumulate and print operate automatically

ASK

: ask mode, bi-direction,

Command R: read data

Command T: tare

Command Z: zero

Wireless

: wireless mode

Note: if you have selected the wireless model, the communication mode has to be set to wireless.

This option is used to set baud rate

Options: 600/1200/2400/4800/9600

This option is used to set parity verify

Optional: 7E1/7O1/8N1 tpup

: set the printer as tpup model. lp-50

: set the printer as LP-50 model.

“ Lab x ”, set gross/acc print format

“ prt

, set the date/time print format

DECI

Dual

ON

/

OFF

CAL

This option is used to select the decimal

Options : 0, 0.0, 0.00, 0.000

INC

This option is used to select the division

Options : 1, 2, 5, 10, 20, 50, 100, 200

CAP

This display will show xxxxxx for setting the capacity.

Nonlin

Non- linearity Calibrate

Liner

Linearity Calibrate

COUNT

LOCK

This display will show xxxxxx for indicating the internal counts.

Enable /disable key lock

- 13 -

OTH

P5 unt

ANM

Kg

/

Lb

/

L2

/ tj

/ hj

BW/BWS-E serial Indicator user’s manual

ON (animal )

OFF(normal mode)

Set unit, when the unit is set for U2, then this unit could be active to U2. Note: lb, oz, Tj and Hj could not be used at the same time. If you select U2, then will get kg and which you select secondary unit.

External calibration.

P6

Cal

Form one

LAB

0 1 2 3

TYPE

PRT tpup tpup tpup tpup

0

GS: 0.888kg

NT: 0.666kg

TW: 0.222kg

GW: 0.888kg

GS: 0.222kg

TOTAL: 0.222kg

1

2

3

4

5

6

7

DATE: 04/06/06

GS: 0.888kg

TIME: 11/11/11

GS: 0.888kg

DATE: 04/06/06

TIME: 11/11/11

GS: 0.888kg

NO.: 4

GS: 0.888kg

DATE: 04/06/06

NO.: 5

GS: 0.888kg

TIME: 11/11/11

NO.: 6

GS: 0.888kg

DATE: 04/06/06

TIME: 11/11/11

NO.: 7

GS: 0.888kg

DATE: 04/06/06

NT: 0.666kg

TW: 0.222Kg

GW: 0.888kg

TIME: 11/11/11

NT: 0.666kg

TW: 0.222kg

GW: 0.888kg

DATE: 04/06/06

TIME: 11/11/11

NT: 0.666kg

TW: 0.222kg

GW: 0.888kg

NO. : 4

NT : 0.666kg

TW: 0.222kg

GW: 0.888kg

DATE: 04/06/06

NO.: 5

NT: 0.666kg

TW: 0.222kg

GW: 0.888kg

TIME: 11/11/11

NO.: 6

NT: 0.666kg

TW: 0.222kg

GW: 0.888kg

DATE: 04/06/06

TIME: 11/11/11

NO.: 7

NT: 0.666kg

TW: 0.222kg

GW: 0.888kg

DATE: 04/06/06

GS: 0.222kg

TOTAL: 0.444kg

TIME: 11/11/11

GS: 0.222kg

TOTAL: 0.666kg

DATE: 04/06/06

TIME: 11/11/11

GS: 0.222kg

TOTAL: 0.888kg

NO.: 4

GS: 0.222kg

TOTAL: 1.000kg

DATE: 04/06/06

NO.: 5

GS: 0.222kg

TOTAL: 1.222kg

TIME: 11/11/11

NO.: 6

GS: 0.222kg

TOTAL: 1.444kg

DATE: 04/06/06

TIME: 11/11/11

NO.: 7

GS: 0.222kg

TOTAL: 1.666kg

NT: 0.222kg

TW: 0.666kg

GW: 0.888kg

TOTAL: 0.222kg

DATE: 04/06/06

NT: 0.222kg

TW: 0.666kg

GW: 0.888kg

TOTAL: 0.444kg

TIME: 11/11/11

NT: 0.222kg

TW: 0.666kg

GW: 0.888kg

TOTAL: 0.666kg

DATE

04/06/06

TIME: 11/11/11

NT: 0.222kg

TW: 0.666kg

GW: 0.888kg

TOTAL: 0.888kg

No.: 4

NT: 0.222kg

TW: 0.666kg

GW: 0.888kg

TOTAL: 1.000kg

DATE: 04/06/06

No.: 5

NT: 0.222kg

TW: 0.666kg

GW: 0.888kg

TOTAL: 1.222kg

TIME: 11/11/11

No.: 6

NT: 0.222kg

TW: 0.666kg

GW: 0.888kg

TOTAL: 1.444kg

DATE: 04/06/06

TIME: 11/11/11

No.: 7

NT: 0.222kg

TW:: 0.666kg

GW: 0.888kg

TOTAL: 1.666kg

- 14 -

BW/BWS-E serial Indicator user’s manual

7

4

5

6

0

1

3

2

Form two

LAB

PRT TYPE

0

LP-50

2000/00/00 00:00

S/N 1

GW 0.888kg

DATE: 2000/00/00

TIME: 00:00

GW: 0.888kg

DATE:

TIME: 00:00

S./NO.: 2

GROSS WT: 0.888kg

2000/00/00 00:00

S/N 0003

GW 0.888kg

2000/00/00 00:00

S/N 4

GW 0.888kg

DATE: 2000/00/00

TIME: 00:00

GW: 0.888kg

DATE:

TIME: 00:00

S./NO.: 6

GROSS WT: 0.888kg

2000/00/00 00:00

S/N 7

GW 0.888kg

1

LP-50

As left

As left

As left

As left

As left

As left

As left

As left

2

LP-50

As left

As left

As left

As left

As left

As left

As left

As left

3

LP-50

As left

As left

As left

As left

As left

As left

As left

As left

- 15 -

BW/BWS-E serial Indicator user’s manual

SECTION 8 BATTERY OPERATIONS

The indicator can be operated from the battery if desired. The battery life is approximately 70 hours.

When the battery needs charging a symbol on the weight display will turn on. The battery should be charged when the symbol is on. The scale will still operate for about 10 hours after which it will automatically switch off to protect the battery.

To charge the battery simply plug into the mains power. The scale does not need to be turned on.

The battery should be charged for 12 hours for full capacity.

Just left side of the display is an LED to indicate the status of battery charging.

When the scale is plugged into the mains power the internal battery will be charged.

If the LED is green the battery has a full charge. If it is Red the battery is nearly discharged and yellow indicates the battery is being charged.

As the battery is used it may fail to hold a full charge. If the battery life becomes unacceptable then contact your distributor.

- 16 -

BW/BWS-E serial Indicator user’s manual

SECTION 9 RS-232 OUTPUT

The BW/BWS-E Series of s can be ordered with an optional RS-232 output.

9. 1 basic information

Specifications:

RS-232 output of weighing data

ASCII code

7/8 data bits

Parity stable

Baud rate from 600bps to 9600bps

Connector: 9 pin d-subminiature socket / 9 pin air connector

Pin 2: Input,

Pin 3: Output

Pin 5: Signal Ground

9. 2 normal print out

Data Format for normal weighing operations, parts counting or recalling of totals from memory will all be different. Examples follow:

Normal Output

S/N The number increments every time a new value is stored in memory

GW GW for gross weight, NT for net weight and a unit of weight

<If>

<If> Includes 2 line feeds

When recalling the Total weight stored in the accumulation memory the output format is:

*************** A line of stars is shown

<lf> Includes 1 line feed

Total No: 3 Times of the accumulation memory

Total wt.: 0.447KG Weight of the accumulation memory

***************

9. 3 continuously output protocol con1: weighing mode k g CR LF

HEADER1: ST=STABLE US=UNSTABLE

HEADER2: NT=NET GS=GROSS

- 17 -

BW/BWS-E serial Indicator user’s manual

SECTION 10 CALIBRATION

Turn the power off.

Turn the power back on, during the counting from 9 to 0 press the

PRINT

key.

The display will show " pn

" (ask password), press

M+

,

G/N

,

TARE

to enter.

The display will show the first function, " p0chk

". , press

TARE

until display show

P3 Lin

, press zero to enter, press Tare to select

CAL

, press

ZERO

key to enter calibrate. The display will show "

Nonlin

".

1. Normal calibrate press

ZERO

key to enter calibrate. The display will show " unLd

".

Remove any weight from the platform. After stable indicator on, press the

ZERO

key.

Then the display will show the last calibration weight used. If this is correct you can continue by pressing the

ZERO

key. If it is not correct use the

M+, MR

,

TARE

keys to change the calibration weight value. When will correct, press the

ZERO

key.

Then display will show "

LoAd

". Place the calibration weight on the scale. After stable, press the

ZERO

key.

2. Linearity Calibrate

Press

ZERO

key to enter calibrate. Press

TARE

key. Then the display will show "

Liner

".

Press

ZERO

key to enter calibrate. Then the display will show

”PIN”, (ask password), press

G/N

,

M+

,

MR

to enter.

The display will show "

Load0

".

Remove any weight from the platform. After stable indicator on, press the

ZERO

key.

Then the display will show "

Load1

", apply the reference weight of the full capacity of the scale just in the center of the platform. The span adjustment will automatically be achieved.(Calibrate weight request: the front digits if the full capacity. For example, the full capacity is 300kg, the 30kg/40kg/

… to 290kg (integer value) ) will be accepted. then press the

ZERO

key.

- 18 -

BW/BWS-E serial Indicator user’s manual

Then the display will show "

Load2

", apply the reference weight of the full capacity of the scale just in the center of the platform. The rule is the same as "

Load1

". then press the

ZERO

key.

Then the display will show "

Load3

", apply the weight of the full capacity, the full capacity will automatically be achieved. then press the

ZERO

key.

If the calibration is acceptable the display will return to normal. If an error message is shown try calibration again as a disturbance may have prevented a successful calibration.

3. External Calibrate press

TARE

until display show

P6 Cal

, then press

ZERO

key to enter external calibrate. The display will show " unLd

".

Remove any weight from the platform. After stable indicator on, press the

ZERO

key.

Then the display will show the last calibration weight used. If this is correct you can continue by pressing the

ZERO

key. If it is not correct use the

M+, MR

,

TARE

keys to change the calibration weight value. When will correct, press the

ZERO

key.

Then display will show "

LoAd

". Place the calibration weight on the scale. After stable, press the

ZERO

key.

Note, only after the linearity calibrate has been completed, then the external calibrate could be set.

If the problem persists then contact your dealer.

After calibration, it should be checked to verify the calibration and linearity is correct.

If necessary repeat calibration, especially be certain the scale is stable before accepting any weight.

- 19 -

BW/BWS-E serial Indicator user’s manual

SECTION 11 ERROR CODES

Error Message

-----

Err 1

Err 2

Err 4

Err 5

Err 6

Err 9

Err 17

--ol--

Description

Maximum load exceeded

Incorrect date

Incorrect time

Zero setting error

Key board error

A/D value out of range

Unstable Reading

Tare out of range

Over range

Solution

Unload or reduce weight

Enter the date by using format

“yy;mm:dd”

Enter the time by using format

“hh:mm:ss”

Zero setting range exceeded due to switching on.(4%max)

Make sure platform empty.

Check the keys and connecter.

Make sure platform empty and check the pan is installed proper. Check the load cell connectors.

Check any air variation, vibration, RF noise and touching some where.

Check the load cell and connecters.

Remove the load and restart scale again.

Remove the load.

Re calibrate

Re calibrate

Fai l h

/

fai l l

Err p

Ba lo

/

lo ba

Calibration Error

Printer error

Battery low

Check the printer and settings

Re charge battery, check the voltages.

- 20 -