BTEK WorldWEIGH BWS Series Technical Manual
Add to My manuals26 Pages
BTEK WorldWEIGH BWS Series is a versatile weighing indicator, suitable for general weighing or more specialized applications such as check weighing, animal weighing, and accumulation applications. It features a large LCD with white LED backlight, 7-key operation, and can be connected to a printer, PC, or scoreboard. With automatic external calibration, 1 RS-232 output, and the ability to drive up to 4 load cells, the BWS Series is a reliable and accurate weighing solution for a variety of applications.
advertisement
![BTEK WorldWEIGH BWS Series Technical Manual | Manualzz BTEK WorldWEIGH BWS Series Technical Manual | Manualzz](http://s2.manualzz.com/store/data/066401810_1-45a5507f9646d227be1ac04afe2121f6-360x466.png)
BWS
Series
Indicator
Technical Manual
800-000316
V1.09p
CONTENTS
1. SPECIFICATION... 2
2. INTRODUCTION... 3
Charging...3
3. INSTALLATION. 4
Installation.. 4
Wiring ....
4
4. KEYS DESCRIPTION... 6
Key Board... 6
Secondary functions of the keys.8
5. OPERATION 8
5.1. Basic Operation... ...8
1. Power On /Off..8
2. Zero...8
3. Tare...8
4. Select Unit8
5.2. Check Weighing..9
5.2.1. Set Limits......9
5.2.2. Set Check Weighing...9
5.3. Accumulation10
5.3.1. Memory Recall.....10
5.3.2. Memory Clear...10
5.3.3. Automatic Accumulation...11
5.4. Parts Counting.....11
5.5. Animal Weighing..12
5.6. Key Board Lock12
5.7. Set Auto Power Off..13
5.8. Set Back Light..13
6. PARAMETERS.14
7. CALIBRATION ..
18
8. RS-232 OUTPUT19
8.1. Specifications.19
8.2. Connecter19
RS-232 Output19
8.3. Continuous output protocol.19
9. DRAWING20
9.1. BWS Drawing20
9.2. BWS Parts List..21
10. ERROR CODES..22
BWS Series Indicator Technical Manual
PRECAUTIONS
WARNING
DISCONNECT ALL POWER TO THIS UNIT
BEFORE INSTALLING, CLEANING, OR
SERVICING. FAILURE TO DO SO COULD
RESULT IN BODILY HARM OR DAMAGE TO
THE UNIT.
CAUTION
• Only qualified Service personal to service this instrument
• Before connecting or disconnecting any components, Disconnect power.
• Failure to observe these precautions could cause bodily harm or damage or destruction of the equipment.
• The BWS weighing indicator is a precision electronic instrument, handle with care.
• Do not install in direct sunlight.
• Verify the local voltage and receptacle type are correct for
the scale.
• Pluggable equipment must be installed near an easily accessible socket outlet.
• Avoid unstable power sources. Do not use near large consumers of electricity such as welding equipment or large motors.
• Avoid sudden temperature changes, vibration, wind and water.
• Avoid heavy RF noise.
• Avoid high stactic areas.
• Keep the indicator clean.
- 1 -
BWS Series Indicator Technical Manual
1. SPECIFICATION
Model
Resolution
Key Pad
Power
BWS
Display
Housing
52mm LCD/ 2” LCD
Stainless Steel
Operating Temperature 0°C - 40°C / 14°F - 104°F
1/30,000 (Non Approved) 1/6000 (NTEP Approved)
7 Keys
Calibration
Interface
Load cell drive Voltage
Load Cells
ADC
ADC Update
Stabilization Time
Battery (6V/4Ah) estimated run time on fully charged battery 50 hours
Automatic External
1 RS-232 Output
Max: 5V/150mA
Up to 4 load cell
Sigma Delta
≤1/10 second
One second typically
- 2 -
BWS Series Indicator Technical Manual
2. INTRODUCTION
The BWS series weighing indicator amplifies signals from a load cell, then converts it to digital data and displays it as a mass value number
It is suitable for general weighing or more specialized applications such as check weighing, animal weighing and accumulation applications
Can be connected to a printer, PC, or scoreboard
Large LCD with white LED back light displays
Charging
• To charge the battery, plug cord into power outlet. The scale does not need to be turned on, to charge.
• The battery estimated charge time to full capacity is 12 hours.
• An LED to indicate the status of battery charge is located on the left side of the Display. When the scale is plugged into a power source the internal battery will be recharging. If the LED is green, the battery has a full charge.
If it is red, the battery is nearly discharged and if yellow, the battery is being charged.
• Verify that the AC power outlet is properly protected.
Note: Please charge the battery before using the scale for the first time .
- 3 -
BWS Series Indicator Technical Manual
Installation
3. INSTALLATION
• Place the Indicator on a table or connect to the supplied mounting bracket.
• Connect the load cell cable to the indicator Bd9 load cell connecter.
• Plug equipment into an easily accessible outlet with a protective ground/ earth contact.
• Turn BWS on with use of the On/Off key. If you want to turn off, press this key again.
• Display will start self checking/ followed by number of calibrations/ number of option changes/ then error 4/ Then enter into main weighing screen.
• Warm-up time of 15 minutes helps stabilizes the measured values after switching on.
• Calibrate with exact calibration weights, a minimum 1/3 of the scale capacity should be used for calibration. For calibration instructions see details under parameters.
Wiring
Supplied 18 foot cable (to load cells) Self made cable (from BWS)
SIG- Green Pins to DB9
SIG+ White 1 SIG-
EX- Black
2 SIG+
EX+ Blue
3
4
EX+
EX-
SEN- Yellow
5 SEN+
SEN+ Red
6 Connect to 7
Shield Grey
7 Connect to 6
SEN and EX needs to be tied
Together if no sense wires..
8 SEN-
9 Shield
Terminal strip and Cut pigtail wiring on the next page
- 4 -
Wiring
Cut Pigtail
Pins to DB9
Green
White
Brown
Black
Grey
Sheild
Blue
Yellow
Red
Wiring
Terminal
Strip
BWS Series Indicator Technical Manual
SIG-
SIG+
Shield
EX-
SEN-
Sen+
Ex+
I/O gnd (Must connect to red)
I/O (must be connected to Yellow)
- 5 -
BWS Series Indicator Technical Manual
Keyboard
4. KEYS DESCRIPTION
Keys Description
Power turn ON/OFF
Zero Display
To perform a tare function, subtracts weight
Accumulator key, current values will store to the memory, to send the data to printer or PC
Shift between Gross / Net Weight
Counting
To change the unit of measure
- 6 -
BWS Series Indicator Technical Manual
Secondary functions of the keys (Setup)
(See Parameter section 6)
Secondary Functions
To confirm the selected menu
Keys
To change the menu and active digit
To move the active digit to right
To move the active digit to left
To enter in to the menu
Escape from the menu to normal operation
- 7 -
BWS Series Indicator Technical Manual
5. OPERATION
Initial Start-up
Warm-up time of 15 minutes helps stabilizes the measured values after switching on.
5.1. Basic Operation
1. Power On/Off:
Switch on the BWS by pressing the key.
The display is now switched on and the start-up test has started, if wanting to switch off, press the
2. Zero:
key again.
Environmental conditions can lead to the BWS not being exactly zero even if the weighing platform is clear of a weight. However, you can set the display of your BWS to zero any time by pressing the therefore ensure that weighing starts at zero.
3. Tare:
key and
The weight of any container can be tared by pressing button so that with subsequent weighing the net weight of the object being weighed is always displayed.
• Load weight on the platform.
• Press key. Zero is displayed, and tare is subtracted.
• Remove weight from the platform. Tared weight is displayed. The BWS can only store one tare value. (see parameter P4) It will be shown with a minus value.
• Press to change between gross weight and net weight.
• To clear the tare value, remove the load and press displayed, tare weight is cleared. key. Zero is
4. Select Unit: (Units not being used can be disabled see parameter 6)
Press the key, to change units kg/g/lb/oz.
- 8 -
BWS Series Indicator Technical Manual
5.2. Check Weighing
The BWS can set an upper or lower limit when weighing with a limited range.
During the limit controls the BWS will indicate whether the upper or lower limits value has been met and an alarm will sound.
5.2.1. Setting Limits: and key together, display will show set h.
• Press
• Press
• Press blink.
key to select set h or set l key to confirm, display will show 00000 and the last digit will
• Enter the high limit value by using active digits and press
and keys to change the
key to increase the value.
• Press key to confirm, display will show set l
• Enter the Low limit value by using active digits and press
and keys to change the
key to increase the value.
• Press key to confirm.
To escape from the settings press
5.2.2. Set Check Weighing:
• Press and key.
key together, display will show set h.
• Press key to select display beep.
• Press key to confirm, display will show none or ok or ng
• Check mode none : No beep sound in the limits. Function turned off.
• Check mode ok : When the weight is between the limits . OK will show
and beeper will be sounded.
- 9 -
BWS Series Indicator Technical Manual
• Check mode ng : When the weight is out of the limits, the beeper will be
sounded and OK will show.
Note: Check weighing available only when weight more than 20d
5.3. Accumulation
The scale can be set to accumulate manually by pressing see the parameter p 1 Com » mode » pr 2
key. For settings,
Before operation scale should be stable and returned to zero, accumulation available only when weight is more than 20d.
Accumulation Operation
• Place the load on the platform.
• Press key, when STABLE is displayed on the screen.
• Display will be show acc 1 then will show the total saved value. These displays will only be shown for three seconds.
• Remove the weight from the platform.
• When display get zeroed and stable then place the second weight on the platform.
• This will continue until the memory gets full or 99 items.
5.3.1 Memory Recall:
To recall the memory press key.
Display will be show acc X (X: Total number of accumulations) then will show the total saved value. These displays will be shown for only three seconds.
5.3.2. Memory Clear:
To clear the memory, press and keys together.
Display will show Acc 0 , all accumulation memory will be cleared.
- 10 -
BWS Series Indicator Technical Manual
5.3.3. Automatically accumulation:
The scale can be set to accumulate automatically. For settings, see the parameter p 1 Com » mode » auto
Automatic Accumulation Operation
• Place the load on the platform.
• When display shows a STABLE indication, display will be show acc 1 then will show the total saved value. These displays will be shown only for three seconds.
• Remove the weight from the platform
• When display get zeroed and stable then place the second weight on the platform.
• This can continue until the memory gets full or 99 items.
5.4. Parts Counting
Parts Counting Operation
• Place the load on the platform
• To enter parts counting, press key, display will then show p 10
Press to change the parts quantity.
Options: p 10 / p 20 / p 50 / p100 / p 200
• Press key to confirm, display will show quantity of the item on the platform.
---- then will show the
• Then items can be added to the platform, display will update the parts quantity automatically.
key to go back to the weighing mode. Press
- 11 -
BWS Series Indicator Technical Manual
5.5. Animal Weighing
BWS can be used for unstable loads. This function is called animal weighing. For settings, see parameter p 3 oth » anm
When load is applied on the platform, after a few seconds of a stable weight, the reading will be locked.
Adding or removing loads, updates the weighing locked values.
To enter or exit animal weighing mode, press and release and together at the same time.
When in animal weighing mode HOLD will be indicated on the display. key
5.6. Keyboard Lock
To set key board lock, see parameter p 3 oth » lock
When the keys are not used within 10 minutes, the keys will be locked automatically.
After entering into the lock function, when a key is pressed display will show klok. Then will return to the normal display.
To unlock and use the keys again, press and hold , and keys for three seconds. Display will show u lck Then the BWS returns to the normal display.
- 12 -
BWS Series Indicator Technical Manual
5.7. Set auto power off
To set auto power off of the BWS
• Hold
• Press key three seconds display will show
key to change Set of and press setbl key to confirm.
• Press
Set of off
Of 5 key to change the options.
To set auto off function off, for scale always on
Set to turn off five minutes later
Of 15 Set to turn off fifteen minutes later
• After selecting the auto off option press key to escape from the settings.
5.8. Set Back Light
To set back light. key to confirm and press
• Hold key for three seconds display will show setbl
• Press setbl
au
on key to confirm.
off
To set auto option. When in use back light will be on, when use stops back light also will off.
To set always on. Back light will always be on.
To set back light off. Back light will be always off key to confirm and press • After selecting the back light option press key to escape from the settings.
- 13 -
BWS Series Indicator Technical Manual
6. PARAMETERS
For Key function in Setup menu see page 7
To setup parameters, turn on the scale.
• Press and Release
• Display will be show pn
key during the self checking countdown
• Press and release, then press and release, then release keys to enter setup, display will show p O CH-
and
• Inside the menu, Press key to confirm selection
Menu
P 0 CH -
CHECK
WEIGHING
P 1 CC om
COM PORT
SETTINGS
key to change selection and press
Sub Menu
Set H
Set lo beep
Mode
( Type of print Function)
Description
Set high limit for check weighing
Set low limit for check weighing
None No beep for check weighing
Ok Beep, when check weighing is between the limits ng Beep, when check weighing is out of the limits
Options:
Cont : Send data continuously
St 1 : Send data one time, when stable.
St c : Send data continuously, when stable
P r1 : Send data one time, when press Key (weight needs to return to zero)
Pr 2 : Send data to print and accumulation, when press .key
Auto : Auto accumulate and auto print mode when weight is stable and or return to zero.
Ask :NOT Used
Wireles :NOT Used
- 14 -
BWS Series Indicator Technical Manual
Baud
(Baud Rate)
Pr
(Parity)
Ptype
(Printer Type)
Lab
(Print Formats)
(these are not customizable)
To set the baud rate
Options:
600 / 1200 / 2400 / 4800 /
9600
To set the parity
Options:
7 e1 / 7 o1 / 8 n1
To set printer model
Options:
Tm295 : Epson tm295
Tm220 : Epson tm220
Lp50 : WorldWEIGH LP50
To set print format 4 choices
Options:
LAB 0
(Accumulation)
NO.
NET: 0.666kg
TARE: 0.222kg
GROSS: 0.888kg
TOTAL: 0.888kg
LAB 1
(G/T/N)
NET: 0.666kg
TARE: 0.222kg
GROSS: 0.888kg
LAB 2
(What is on the screen)
NET: 0.222kg or GROSS: 0.222KG
LAB 3
(Piece count ONLY)
U.W.: 0.00998 lb
NET : 1.996 lb
TOTAL: 200Pcs
LAB 4
(Piece count and normal weighing mode)
U.W.: 0.01000 lb
NET : 2.002 lb
TARE : 1.000 lb
TOTAL: 200Pcs <piece count
U.W.: 0.00000 lb
NET : 0.996 lb
TARE : 1.018 lb
TOTAL: 1.199 lb <gross
- 15 -
BWS Series Indicator Technical Manual
P 2 mod
MODE
(K3 to enter)
P3 ot
OTHER
H
Si g r
Dual 1
Dual 2
Lock anm
To select single range operation
Count To check internal counts
Deci
Div
Cap
To set decimal points
0 / 0.0 / 0.00 / 0.000
To set increment
1 / 2 / 5 / 10 / 20 / 50
Set Capacity
Cal gra
Calibration
Gravity
To select dual range - mode 1
Note: Once active second interval (div 2), Then second interval will work until display return to zero
Count To check internal counts
Deci
Div
To set decimal points
Di v 1 To select first division
Cap
Di v 2
Cap 1
Cap 2
To select second division
To select first capacity
To select second capacity
Cal Calibration gra Gravity (Do not Change)
To select dual interval - mode 2
Note: First interval will active in CAP 1
Second interval will be active in CAP 2
Count To check internal counts
Deci
Div
To set decimal points
Di v 1 To select first division
Di v 2 To select second division
Cap Cap 1 To select first capacity
Cap 2 To select second capacity
Cal gra
Calibration
Gravity
To set keypad lock
Options: on / off
To set animal mode.
Options: on / off
- 16 -
lb oz
BWS Series Indicator Technical Manual
P4 S t
TARE
PP 55 CLR
RESET
P6 Unt
UNIT
St ON
St OFF
CLr CAL
CLr oPt kg g
Multi tare operation turn on
Multi tare operation turn off
Clear all calibration data
Clear ALL parameters
To set “kg” unit Options: on / off
To set “g” unit Options: on / off
(lb or kg must be set to .000 to be able to turn on grams) once selected to the main weight screen change unit to g then enter back into P6 and you can turn off kg or lb if needed
To set “lb” unit Options: on / off
To set “oz” unit Options: on / off
- 17 -
BWS Series Indicator Technical Manual
7. CALIBRATION
Jumper K2 comes shorted Allowing Front Panel access to Calibration
if k2 is not shorted press K3 button inside the unit to enter calibration
To calibrate, turn on the scale.
• Press and release during the self checking countdown
• Display will be show Pn (Password)
• Press and release, then press and release, then release, to enter setup, display will show p O CH-
• Tap the Key until display will show p 2 mod .
and
• Press the key to confirm and press to select duaL 1 duaL 2 (single or Dual range)
• Press the
CAL key to confirm and press the key till display shows
• Press key to confirm.
Calibration
Ca
• Press the key to enter calibration, display will show
• Remove all the weight from the platform. unL dd
• When indicator is stable, press the key to confirm.
• The Display will show the last calibration weight. To change the calibration weight value, press and keys to select the active digits then the key to increase the value.
• When the calibration value is correct, press key to confirm.
• Display will show
L oad
• Place the calibration weight on the platform. (suggested 1/3 capacity)
• When indicator is stable, press key to confirm.
• Display will show pass
and continue to normal weighing mode
- 18 -
BWS Series Indicator Technical Manual
8. RS-232 OUTPUT
8.1. Specifications: (1 Com port)
RS-232 output of weighing data
Code : ASCII
Data bits : 8 data bits
Parity : Parity selectable
Baud rate : 600bps to 9600bps selectable
8.2. RS-232 (Db9 type connector)
Pin 2 RXD Input White
Pin 3 TXD Output
Blue
Receiving data
Transmission data
Pin 5 GND ― Green Signal ground
Db9 Connecter to computer/printer (null modem)
Indicator Computer
Pin 2: Pin 3
Pin 3: Pin 2
Pin 5: Pin 5
8. 3. Continuous output protocol
Weighing mode
<STX> <POL> <WWWWWWWW> <UNIT> <G/N> <S> <TERM>
ASCII 02 decimal
Polarity:
<space>=Positive
<->=Negative
Weight:7 digits, right-justified, dummy zeroes, decimal point with no leading zeroes except for leading zero immediately preceding the decimal point. leading zeroes transmitted as spaces.
G=Gross
N=Net
L= pounds
K= kilograms
O= ounces
G= grams
<CR><LF> or<CR>
Status:
<space>=Valid
I=Invalid
M=Motion
O=Over/under range
- 19 -
BWS Series Indicator Technical Manual
9.1. BWS Drawing
9. DRAWING
- 20 -
BWS Series Indicator Technical Manual
9.2. BWS Parts List
24
25
26
27
28
29
17
18
19
20
21
22
23
12
13
14
15
16
3
4
5
6
7
No Parts
1
2
Key Panel
Front Cover
8
9
10
11
Display Protection Plate
Nut
Main Board
Washer
Star (+) Self Thread screw
Water Proof Rubber Bar
Star (+) Screw
Washer
Battery Clamp
Washer
Star (+) Big head Screw
Bracket
Bracket Screw
Water Proof Adaptor jack
Interface Module
Air connecter
Plug
Rubber Spacer
Air Connecter
Back Cover
Air Connecter Water Proof
Nut
Battery
Nut
Main Serial board
Spacer
Star (+) Screw
Micro Switch Cap
Homerun 15Ft with DB9
Rs232 Board
1
1
1
1
1
7
1
1
1
1
1
1
1
3
6
6
1
2
1
1
6
1
6
6
1
2
2
1
Qty PN
1
1
899-300086
899-300085
899-300022
899-300041
899-300087
M4
M4x12
5Pin
7Pin
Spec
M3*6
8x3.1x1.5
M3x8
M4x10
M4
6V/4Ah
M3x6
3Mx20
- 21 -
BWS Series Indicator Technical Manual
Error Message
10. ERROR CODES
Description Solution
Maximum load exceeded Unload or reduce weight.
Low reading Re-check wiring
Zero setting error
Key board error
Zero setting range exceeded due to switching on.(4%max)
Make sure platform is empty.
Check the keys and connecter.
A/D value out of range Make sure platform is empty.
Check the load cell connectors.
Unstable Reading
Tare out of range
Over range
Calibration Error
Check any air variation, vibration, RF noise and binding somewhere.
Check the load cell and connecters.
Remove the load and restart scale again.
Remove the load.
Recalibrate
Recalibrate
Err p
Printer error flashing backlight Battery low
Ba LO
Battery low
Check the printer and settings.
Recharge battery, check the voltages.
Recharge battery, check the voltages.
- 22 -
BWS Series Indicator Technical Manual