Honeywell ToxiRAE Pro CO2 Quick Start Guide

Add to My manuals
2 Pages

advertisement

Honeywell ToxiRAE Pro CO2 Quick Start Guide | Manualzz

ToxiRAE Pro CO

2

QuickStart Guide

RAE Systems, Inc.

3775 N. First St.

San Jose, CA 95134-1708 USA

Phone: 408-952-8200 Fax: 408-952-8480

Email: [email protected] www.raesystems.com

Charging The ToxiRAE Pro CO

2

Always fully charge the battery before use. Contacts on the bottom of the ToxiRAE Pro CO

2

meet the cradle’s contact pins, transferring power. While charging, the

LED on the cradle glows red. When the battery is fully charged, the LED glows green.

A Place ToxiRAE

Pro CO 2 in charging cradle

B Connect AC adapter to charging cradle and plug it into AC

Raised lip locks into cradle’s latch power source

WARNINGS

Read Before Operating

The ToxiRAE Pro CO

2

User’s Guide must be carefully read by all individuals who have or will have the responsibility of using, maintaining, or servicing this product. The product will perform as designed only if it is used, maintained, and serviced in accordance with the manufacturer’s instructions.

CAUTION!

Never operate the monitor when the cover is removed. Remove monitor top cap, rear cover, and battery only in an area known to be non-hazardous.

2.

User Interface

The ToxiRAE Pro CO

2

’s user interface consists of the display, four LED lights, an alarm buzzer, and two keys, [MODE] and [Y/+]. The LCD displays real-time reading in specified measurement units, sensor type, alarm type (when in alarm, including cal. or bump overdue), Man Down status, battery status, datalog (if on), and radio and connection quality (if available).

Gas inlet

Belt clip (on back)

LEDs

LED

Alarm

Buzzer

Display

[MODE] key

[Y/+] key

3.

Turning The ToxiRAE Pro CO

2

On

Press and hold [MODE] for 3 seconds. During startup, the battery, buzzer, vibration alarm, and

LEDs are tested, and then the ToxiRAE Pro CO

2 performs self-testing of its functions. When the main measurement screen appears, the ToxiRAE

Pro CO

2

is ready for calibration or use.

Note: If the battery is almost fully discharged, the message “Battery too low! Needs charging.

Powering off!” shows in the display and the

ToxiRAE Pro CO

2

shuts off. You must charge the battery before you can use the instrument.

Turning The ToxiRAE Pro CO

2

Off

Press and hold [MODE]. A 5-second countdown to shutoff begins. You must continue pressing on the key for the entire shutoff process. If you remove your finger from the key during the countdown, the shutoff operation is canceled and the ToxiRAE Pro

CO

2

continues normal operation.

When the countdown beeps stop and you see “Unit

Off,” release your finger from the [MODE] key.

The ToxiRAE Pro CO

2

is now off.

Latch release

Charging status LED

4. 5. 6.

Calibration

The ToxiRAE Pro CO

2

can be automatically bump tested and calibrated using the AutoRAE 2 Test and Calibration

Station (refer to its User’s Guide for instructions).

Manually calibrate using a fixed-flow regulator (flow rate between 0.5 and 1.0 liters per minute) and the supplied special calibration adapter that covers the gas inlet:

1.

To begin calibration, connect the calibration gas cylinder, flow regulator, and calibration adapter to the

ToxiRAE Pro CO

2

.

2.

Enter Programming Mode by pressing and holding

[MODE] and [Y/+] simultaneously until the password screen appears.

3.

Input the 4-digit password. (The default password is

“0000.” If you do not know the password, highlight and select “OK.”) Then follow the instructions for zero calibration and span calibration.

Important! After a bump test or calibration, remove the calibration adapter to ensure correct readings.

7.

Bump (Functional) Testing

RAE Systems recommends periodic bump testing to confirm that the sensor is working and that the alarms are functional.

With the ToxiRAE Pro CO

2

in Normal Mode:

1.

Connect the calibration gas cylinder, flow regulator, and calibration adapter to the

ToxiRAE Pro CO

2

and start the gas flow.

2.

Make sure the unit goes into alarm and that the buzzer produces at least two beeps per second, the LED lights flash on and off, and the vibration alarm functions. The display backlight should illuminate and an alarm message should be shown in the display.

3.

Turn off the gas flow.

4.

Remove the calibration adapter.

10.

Zero (Fresh Air) Calibration

The ToxiRAE Pro CO

2

should be zero-calibrated in

100% nitrogen. With the ToxiRAE Pro CO

2

in

Programming Mode and “Zero Calib” highlighted:

1.

Press [Y/+]. “Apply zero gas…” is displayed.

2.

Start the flow of the zero gas, if used, and press

[Y/+] to start the fresh air or zero calibration.

3.

The screen says, “Zeroing…” and shows the countdown.

4.

When done, it says, “Zero is done! Reading = X” (this reading should be 0.0 ppm or very close to it).

5.

“Span Calib” is highlighted.

6.

Shut off the flow of nitrogen.

7. Remove the calibration adapter.

Pressing [MODE] repeatedly allows you to step through the screens as shown here.

Note: The gray boxes indicate the datalog functions when datalogging is in Manual mode. When datalogging is in Automatic mode, these screens do not appear. Refer to the User’s

Guide for information on selecting Automatic or

Manual datalogging.

Manual Span Calibration

Note: The ToxiRAE Pro CO

2

should be calibrated with

5000 ppm CO

2

.

In Programming Mode, and with “Span Calib” highlighted:

1.

Press [Y/+]. The screen displays the current (or default) calibration gas and its concentration.

· Press [MODE] if you do not want to change it.

· Press [Y/+] if you do want to change it.

2.

Attach the calibration adapter, and connect the calibration gas cylinder and flow regulator to the

ToxiRAE Pro CO

2

, and start the gas flow.

3.

Press [Y/+] to start calibration.

4.

Upon completion, "Span is done!" and the reading are shown. Make sure it is within ±10% of the span gas value.

5. Remove the calibration adapter.

8.

Menu Navigation In Normal Mode

9.

11.

P/N: G02-4013-000 Rev A 201303

advertisement

Key Features

  • CO2 concentration monitoring in real-time
  • Audible, visual, and vibration alarms for quick response
  • Datalogging for up to 120 hours of data storage
  • Rugged design for harsh environments
  • Easy-to-use interface with intuitive navigation
  • Low maintenance and calibration needs
  • Compact and lightweight for convenient portability
  • Long battery life for extended operation
  • Bump testing capability for quick sensor verification

Related manuals

Frequently Answers and Questions

How do I turn on the ToxiRAE Pro CO2?
Press and hold the [MODE] key for 3 seconds.
How do I calibrate the ToxiRAE Pro CO2?
Connect the calibration gas cylinder, flow regulator, and calibration adapter to the ToxiRAE Pro CO2. Enter Programming Mode by pressing and holding [MODE] and [Y/+]. Input the 4-digit password and follow the instructions for zero and span calibration.
How do I perform a bump test on the ToxiRAE Pro CO2?
Connect the calibration gas cylinder, flow regulator, and calibration adapter to the ToxiRAE Pro CO2. Make sure the unit goes into alarm and that the buzzer produces at least two beeps per second, the LED lights flash on and off, and the vibration alarm functions.