AGS Merlin 1000S Installation & Operation Manual
Add to My manuals12 Pages
AGS Merlin 1000S is a gas isolation controller designed for educational establishments and laboratories to provide full control over incoming gas supply. It features a lockable main keyswitch, timeout facility for automatic gas solenoid valve shut-off, and compatibility with various gas sensors like LPG, NG, CO, and Hydrogen.
advertisement
Installation & Operation Manual Merlin 1000S
Merlin 1000S
Gas Isolation Controller
Installation & Operation Manual
Please read this manual carefully and retain for future use.
The Merlin 1000S gas pressure proving system is designed specifically for use in educational establishments and laboratories. The system is designed to give full control over incoming gas supply with the lockable main keyswitch.
The information contained within this manual should be referenced for typical installation and operation only.
For specific requirements that may deviate from the information in this guide – contact your supplier.
Rev: 10 02-21 1
Rev: 10 02-21
Installation & Operation Manual Merlin 1000S
Contents
IMPORTANT WARNING STATEMENTS
...................................................... 3
Installation
Typical Installation Arrangement
................................................................................. 5
................................................................................................. 5
.................................................................................................. 8
Operation
........................................................................................ 9
Maintenance
.............................................................................................. 11
Bump Test (Gas Response Check)
........................................................................... 11
2
Installation & Operation Manual Merlin 1000S
IMPORTANT WARNING STATEMENTS
Please take the time to thoroughly read this user’s guide which should be retained for future reference.
This device requires a continual supply of electrical power – it will not work without power.
This device should not be used to substitute proper installation, use and/or maintenance of fuel burning appliances including appropriate ventilation and exhaust systems.
This device does not prevent dangerous gasses from occurring or accumulating.
Non-continuous (intermittent) operation will not affect performance.
Actuation of alarms indicates the presence of a danger.
Your product should reach you in perfect condition, if you suspect it is damaged, contact your supplier .
Manufacturer’s Warranty
Warranty coverage: The manufacturer warrants to the original consumer purchaser, that this product will be free of defects in material and workmanship for a period of three (3) years from date of purchase.
The manufacturer’s liability hereunder is limited to replacement of the product with repaired product at the discretion of the manufacture. This warranty is void if the product has been damaged by accident, unreasonable use, neglect, tampering or other causes not arising from defects in material or workmanship. This warranty extends to the original consumer purchaser of the product only.
Warranty disclaimers: Any implied warranties arising out of this sale, including but not limited to the implied warranties of description, merchantability and intended operational purpose, are limited in duration to the above warranty period.
In no event shall the manufacturer be liable for loss of use of this product or for any indirect, special, incidental or consequential damages, or costs, or expenses incurred by the consumer or any other user of this product, whether due to a breach of contract, negligence, strict liability in tort or otherwise. The manufacturer shall have no liability for any personal injury, property damage or any special, incidental, contingent or consequential damage of any kind resulting from gas leakage, fire or explosion. This warranty does not affect your statutory rights.
Warranty Performance: During the above warranty period, your product will be replaced with a comparable product if the defective product is returned together with proof of purchase date. The replacement product will be in warranty for the remainder of the original warranty period or for six months – whichever is the greatest.
Information on waste disposal for consumers of electrical & electronic equipment.
When this product has reached the end of its life, it must be treated as Waste Electrical & Electronics Equipment
(WEEE). Any WEEE marked products must not be mixed with general household waste, but kept separate for the treatment, recovery and recycling of the materials used. Please contact your supplier or local authority for details of recycling schemes in your area.
Alternatively, all AGS products can be securely packaged and returned clearly marked for disposal.
Rev: 10 02-21 3
Installation & Operation Manual Merlin 1000S
Installation
Planning
The Merlin 1000S gas pressure proving system is designed specifically for use in educational establishments and laboratories.
This panel is to be used to carry out a gas-proving test on the pipe work in order to highlight if there is a gas appliance open or a gas leak in a laboratory. The Merlin 1000S is designed to give the teacher full control over the incoming gas supply with the lockable main key-switch.
The Merlin 1000S can work in conjunction with carbon dioxide, natural gas, carbon monoxide and
LPG sensors. The Merlin 1000S also has a built in “timeout” facility which will automatically shut off the gas solenoid valve at the end of a specific time period, this time period can be adjusted to 2, 5, 8 hours or can be overridden if required.
Please refer to your detector manual for important information regarding coverage, location and positioning including areas and conditions to avoid.
Fixing - Mounting
Place the panel 48-60 inches above finished floor level.
Unpack all the parts!
Designed for surface mounting, it must be installed by a licensed, insured contractor.
1.
Carefully remove the front cover from the unit by unscrewing the four bolts located at each corner.
To do this – use the socket wrench provided.
2.
Mark the four screw holes located on the back of the enclosure to the wall. Ensure the wall surface is flat to prevent base distortion.
3.
After executing the mounting and the connections – replace the front cover and insert the security caps over the four bolts.
Take care when making connections to high voltage connectors!
Any damage attempting to remove the circuit board may void any warranty!
All Class 2 wiring is to be installed within flexible tubing to maintain segregation between circuits!
Wiring of different circuits shall be separated by means of routing, clamping or barrier!
A flush mount kit is available, comprising of a mounting bracket and decorative surround strip. Contact your supplier for more information.
Rev: 10 02-21 4
Installation & Operation Manual Merlin 1000S
Typical Installation Arrangement
Terminal Connections
1.
POWER / LINE IN
110-120V AC Power should be supplied to the [POWER / LINE IN] terminal and fused at 3A.
2.
GAS SOLENOID VALVE OUTPUT
110-120V AC electrical power supplied from the [VALVE OUT] connector using a 3 core cable can be connected to a gas solenoid valve which can shut the gas supply on alarm status.
Pressure sensors should be screwed to the downstream port of the valve.
3.
BMS OUTPUT
Connections are available on the board for Building Management Systems.
[NO Normally Open] [COM Common] [NC Normally Closed] These are volt free connections.
This is a relay that changes state in alarm or when the gas is on/off and can be used in conjunction with the 12V DC output and other external relays that affect other devices and controls such as purge fans and audible alarms etc. See section; Switch Settings - for BMS options
Rev: 10 02-21 5
Installation & Operation Manual Merlin 1000S
4.
PRESSURE SENSOR
The pressure sensor is wired to the [PRESSURE SENSOR] connector and screwed into the downstream port of the gas solenoid valve. Connect the pressure sensor as shown:
Wiring: Red [+] Black [-] Blue [ IN ]
The sensor will monitor the gas supply pressure and if pressure drops below 0.17psi – the gas valve will close as this could mean a gas leak is present.
The pressure sensor operating pressure is: 0 – 1.45psi.
5.
EM REMOTE
Connections for remote emergency shut-off or stop buttons is detailed on the circuit board as [EM
REMOTE]. This is linked out as a factory setting. Remote emergency shut-off buttons should be volt free and wired to the Merlin 1000S using a plenum security cable, white, 18/2 (18AWG 2 conductor), stranded, CMP or similar.
6.
GAS DETECTOR
Connections can be made for a Merlin gas detector (LPG, NG,
CO or Hydrogen).
If no detector is being used leave the factory fitted link in.
A wiring example is shown.
Refer to your gas detector manual for further information.
7.
FAN SWITCHES
This terminal switches when the key is turned on and off.
This can be connected to a fan switch (supplied separately) which can provide power to fans when the control panel is switched on.
8.
CO2 MONITOR
This terminal can be used to connect a Merlin CO
2
monitor to shut off the system in the event of a high concentration of CO
2
.
If no CO2 monitor is connected, the panel will ‘beep’ upon power up and the CO2 LED will flash 3 times to indicate that this terminal has been disabled. An external power supply is required.
Refer to your CO
2
monitor manual for further information.
9.
12V DC
This is a power output for external auxiliary devices when there is power at the panel and can be used to create a relay switch with the BMS relay output.
Max output: 50mA
Rev: 10 02-21 6
Installation & Operation Manual Merlin 1000S
Switch Settings
Auto Reset
OFF
When power is restored after the power cut/ loss, the panel has to be restarted manually. (Default)
ON
This will instruct the system to restart automatically when power is restored after power cut/loss.
Building Management System Integration
[BMS SEL].
The panel can be integrated with a BMS to make or break a circuit when the gas is either on or off, (valve open or valve closed).
This will tell the BMS whether or not power is being sent to the solenoid valve.
OFF Signals the BMS when gas is on or gas is off. (Default)
ON
OFF
Signals the BMS on a fault i.e. high gas levels detected, emergency shut-off activated etc.
Fan Switch Integration
[EM SEL].There is the facility to connect a Fan Switch.
A fan switch (Merlin FS1 or FS2 - sold separately) provides the facility to turn on fan(s) when the 1000S panel is switched on and removes power to fan(s) when the switched off.
Instructs the system to shut down fan(s) and gas supply upon activation of emergency shut off button(s). (Default)
ON
Instructs the system to leave the fans on and shut off the gas supply only upon activation of emergency shut off button(s).
This option is not available if a Fan Switch is not installed.
Gas Fill Time and Prove Time
There are two switches located on the circuit board labelled [FILL TIME] and [PROVE TIME].
FILL TIME: Amount of time the gas valve opens to fill the gas line on power up or reset.
PROVE TIME: Amount of time the system tests the gas line for leaks on power up or reset.
FILL TIME PROVE TIME
OFF
ON
5 Seconds (Default)
10 Seconds
OFF
ON
30 Seconds (Default)
50 Seconds
Once the settings have been changed - remove power for 10 seconds.
Rev: 10 02-21 7
Installation & Operation Manual Merlin 1000S
Fire Panel Integration
The Merlin 1000S can be integrated with a fire alarm to close the gas supply automatically in the event of a fire. The volt free fire alarm signal can be wired in series with any remote emergency shut off buttons. If there are no remote emergency shut off buttons being installed, wire this directly to the terminal marked [EM REMOTE].
Auto Shut-Down - Timeout Period
The system has an auto-shut down feature after a selected time.
There are two switches located on the circuit board labelled [TIME1] & [TIME2].
These can be configured to select the required timeout/ shut-down period.
TIME 1 TIME 2 Timeout Period
OFF OFF 2 Hours (Default)
ON
OFF
OFF
ON
4 Hours
8 Hours
ON ON Disabled
Upon timeout the gas supply will be turned off unless other utilities are configured.
General Specification
Model:
Visual Indication
Mains Electrical Power Input
Gas Solenoid Valve Output
Current Consumption
Internal Fuse
Operating Temperature
Audible Alarm Buzzer dB
Pressure Sensor Operating Pressure
Housing Material
Flame Rating
UL Approval Reference
O/All Dimensions (H x W x D) mm / inch
1000S
LED
110-120VAC
110-120VAC
12W max (50mA) @ 120VAC
3.15A
32 – 104°F 0-95%RH Non-Condensing
65 dB (300mm distance in quiet conditions)
0 -1.45psi
Polylac PA-765
UL 94
E464760
7.08 x 10 x 3.03”
Rev: 10 02-21 8
Installation & Operation Manual Merlin 1000S
Operation
First Power Up
On connecting mains power, an LED on the front of the panel (AGS Logo) will illuminate red.
Turn the key switch to on position. Turn the key switch to the off position to turn the panel off.
Emergency Shut Off Button
The emergency shut off button is located on the front of the panel.
There is also a facility for remote shut off buttons to be wired in series on the circuit board.
The emergency shut off button(s) will cut off the gas supply and electrics only when activated.
To reinstate the system, the emergency shut off button(s) will need to be reset and the panel restarted.
LED Indicators
Gas On
When the key switch is turned on, the Merlin 1000S will check the installation for gas leaks. If gas proving is successful, the gas valve will open and the ‘Gas On’ LED will illuminate.
ON = Gas On
OFF = Gas Off
Testing
This LED will illuminate GREEN for approximately 30 seconds when the panel is checking the integrity of the gas installation upon start up.
ON = proving the gas line, do NOT operate any appliances.
Test Fail
Under normal working conditions this LED is off.
When the panel detects a gas leak on start-up, the LED will illuminate AMBER.
The gas valve will remain closed.
OFF = OK
ON = gas proving failed.
Pressure Low
Under normal working conditions the LED is off.
The LED will illuminate AMBER when pressure of the gas supply drops below 0.17psi for 10 secs. The gas valve will close.
OFF = OK
ON = gas supply pressure low.
Rev: 10 02-21 9
Installation & Operation Manual Merlin 1000S
Timeout
Under normal working conditions this LED is off.
This LED will illuminate AMBER when auto-shut down has occurred.
OFF = OK
ON = Auto-shut down activated.
EM Stop
If an emergency shut off button (either remote or on the panel) is pressed, the LED will illuminate
AMBER and the gas will be turned off.
The EM Stop button must be reset before restarting the system.
OFF = OK
ON = Emergency Shut-Off button activated.
Gas Detected
Under normal working conditions this LED is off.
If the external Merlin detector connected detects gas this will show RED and the gas valve will turn off.
OFF = OK
ON = Gas detected.
CO
2
High
Under normal working conditions this LED is off.
If the concentration of CO2 in the air is at alarm level (relevant detector required), the LED will show RED and the Gas valve will turn off.
OFF = OK
ON = the concentration of CO2 is at alarm level.
Rev: 10 02-21 10
Installation & Operation Manual Merlin 1000S
Maintenance
Cleaning
To keep your control panel in good working order - follow these basic principles;
Remove any dust/debris from the outer enclosure regularly using a slightly damp cloth.
Never spray air fresheners, hair spray, paint or other aerosols near gas detecting devices.
Never paint the device.
It is recommended that any detectors connected to the system are inspected and serviced at least annually from the date of installation for optimum performance and protection.
Bump Test (Gas Response Check)
What is a Gas Response Check?
Response checks are often referred to as a ‘BUMP TEST’. Bump tests are important to make sure a gas detection device is able to detect a release of gas as early as possible.
The aim of the bump test is to make sure a gas detector is working at its optimum by briefly exposing the unit to a known concentration of the target gas that exceeds the highest alarm point. If the detector goes into alarm and all system outputs/relays activate, then it is working safely.
If the system fails to operate as intended in an alarm state, the gas detector must not be used until a full inspection and service has been conducted.
Why is it important?
A detector may visually appear in good working order, but its sensitivity can be inhibited by external factors. Dust, humidity, temperature fluctuations, cleaning products, contaminants or sensor drift
(ageing) can cause a decline in sensitivity and eventual failure.
Regular bump tests are important to make sure the detector is able to detect a release of gas as early as possible.
How often should I Bump Test a detector?
Regular bump tests are important to make sure the detector is able to detect a release of gas as early as possible. A bump test usually takes seconds (gas type dependant) and is often completed alongside a scheduled fire alarm test, however the frequency should be determined following a risk assessment by the end user. Remember, bump testing does not remove the need to have gas detectors inspected, calibrated and serviced periodically by a trained personnel.
What equipment do I need to perform a Bump Test?
Contact your AGS representative for details of suitable bump testing kits and gases.
Kits usually consist of a certified gas cylinder; flow control regulator, tube pipe and applicator cone.
Rev: 10 02-21 11
Installation & Operation Manual Merlin 1000S
Installation Details
Please pass this manual to the system owner or system user.
Date of Installation:
Installation Location:
Organisation:
Stamp/ Signature of the installer:
American Gas Safety LLC
Head office:
6304 Benjamin Road, Suite 502, Tampa, FL 33634
Tel: (727) 608-4375 [email protected]
American Gas Safety LLC is the owner of this document and reserves all rights of modification without prior notice.
Rev: 10 02-21 12