Mindray BeneVision N Series Operator’s Manual

Add to My manuals
452 Pages

advertisement

Mindray BeneVision N Series Operator’s Manual | Manualzz

BeneVision N22/BeneVision N19

BeneVision N17/BeneVision N15/BeneVision N12

Patient Monitor

Operator’s Manual

© Copyright 2017-202

2

Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights reserved.

Release time:

April

202

2

Revision: 1

7

.0

WARNING

Federal Law (USA) restricts this device to sale by or on the order of a physician or other practitioner licensed by U.S. state law to use or order the use of this device.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual I

II

Intellectual Property Statement

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) owns the intellectual property rights to this Mindray product and this manual. This manual may refer to information protected by copyright or patents and does not convey any license under the patent rights or copyright of Mindray, or of others.

Mindray intends to maintain the contents of this manual as confidential information. Disclosure of the information in this manual in any manner whatsoever without the written permission of Mindray is strictly forbidden.

Release, amendment, reproduction, distribution, rental, adaptation, translation or any other derivative work of this manual in any manner whatsoever without the written permission of Mindray is strictly forbidden.

is the trademark, registered or otherwise, of Mindray in China and other countries. All other trademarks that appear in this manual are used only for informational or editorial purposes. They are the property of their respective owners.

This posting serves as notice under 35 U.S.C.§287(a) for Mindray patents: http://www.mindrayna.com/patents.

Manufacturer’s Responsibility

Contents of this manual are subject to changes without prior notice.

Mindray is responsible for the effects on safety, reliability and performance of this product, only if:

■ all installation operations, expansions, changes, modifications and repairs of this product are conducted by

Mindray authorized personnel; the electrical installation of the relevant room complies with the applicable national and local requirements; the product is used in accordance with the instructions for use.

WARNING

Only skilled/trained clinical professionals should operate this equipment.

It is important for the hospital or organization that uses this equipment to perform a recommended service/maintenance plan. Neglect of this may result in machine breakdown or personal injury.

Service

Mindray maintains a network of service representatives and factory-trained distributors. Prior to requesting service, perform a complete operational check of the instrument to verify proper control settings. If operational problems continue to exist, contact Mindray service.

In North America contact the Service Department at (800) 288-2121, ext: 8116 for Technical Support or (201)

995-8000 for assistance in determining the nearest field service location.

Please include the instrument model number, the serial number, and a description of the problem with all requests for service.

Any questions regarding the warranty should be directed to the local sales or service representative.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Company Contact

Manufacturer:

Address:

Website:

E-mail Address:

Tel:

Fax:

Distributor:

Address:

Tel:

Website:

Shenzhen Mindray Bio-Medical Electronics Co., Ltd.

Mindray Building, Keji 12th Road South, High-tech Industrial Park, Nanshan,

Shenzhen 518057, P.R.China

www.mindray.com

[email protected]

+86 755 81888998

+86 755 26582680

Mindray DS USA, Inc.

800 MacArthur Boulevard, Mahwah, New Jersey, 07430 USA

1.800.288.2121, 1.201.995.8000

www.mindray.com

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual III

Preface

Manual Purpose

This manual contains the instructions necessary to operate the product safely and in accordance with its function and intended use. Observance of this manual is a prerequisite for proper product performance and correct operation and ensures patient and operator safety.

This manual is based on the maximum configuration and therefore some contents may not apply to the product.

If you have any question, please contact Mindray.

This manual is an integral part of the product. It should always be kept close to the equipment so that it can be obtained conveniently when needed.

Intended Audience

This manual is geared for clinical professionals who are expected to have a working knowledge of medical procedures, practices and terminology as required for monitoring of patients.

Illustrations

All illustrations in this manual serve as examples only. They may not necessarily reflect the setup or data displayed on the patient monitor.

Conventions

Italic text is used in this manual to quote the referenced manuals, chapters, sections and formulas.

Bold text is used to indicate the screen texts and names of hard keys.

→ is used to indicate operational procedures.

IV BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Contents

1 Safety .................................................................................................................................................................................. 1 - 1

1.1 Safety Information ..........................................................................................................................................................................................1 - 1

1.1.1 Warnings ..............................................................................................................................................................................................1 - 1

1.1.2 Cautions ................................................................................................................................................................................................1 - 2

1.1.3 Notes ......................................................................................................................................................................................................1 - 2

1.2 Equipment Symbols .......................................................................................................................................................................................1 - 3

2 Equipment Introduction ................................................................................................................................................... 2 - 1

2.1 Indications for Use ..........................................................................................................................................................................................2 - 1

2.2 Applied Parts .....................................................................................................................................................................................................2 - 1

2.3 System Components ......................................................................................................................................................................................2 - 2

2.3.1 N22, N19

Main Unit .........................................................................................................................................................................2 - 2

2.3.2 N22, N19 Displays ..............................................................................................................................................................................2 - 4

2.3.3 N17, N15, N12 Main Unit .................................................................................................................................................................2 - 7

2.3.4 Satellite Module Rack (SMR) ....................................................................................................................................................... 2 - 12

2.3.5 External Modules ............................................................................................................................................................................ 2 - 12

2.3.6 Cable Management Kit ................................................................................................................................................................. 2 - 15

2.3.7 Input Devices ................................................................................................................................................................................... 2 - 15

2.3.8 Printing Devices .............................................................................................................................................................................. 2 - 16

3 Getting Started .................................................................................................................................................................. 3 - 1

3.1 Equipment Preparation Safety Information ..........................................................................................................................................3 - 1

3.2 Monitor Installation ........................................................................................................................................................................................3 - 1

3.2.1 Unpacking and Checking ...............................................................................................................................................................3 - 2

3.2.2 Environmental Requirements .......................................................................................................................................................3 - 2

3.3 Setting Up the Equipment ...........................................................................................................................................................................3 - 2

3.3.1 Connecting the AC Mains ...............................................................................................................................................................3 - 2

3.3.2 Connecting the Input Devices ......................................................................................................................................................3 - 3

3.3.3 Connecting the SMR ........................................................................................................................................................................3 - 3

3.3.4 Connecting Modules to the SMR .................................................................................................................................................3 - 3

3.3.5 Removing Modules from the SMR ..............................................................................................................................................3 - 3

3.4 Turning on the Monitor .................................................................................................................................................................................3 - 4

3.5 Operation and Navigation ...........................................................................................................................................................................3 - 4

3.5.1 Using the Touchscreen ....................................................................................................................................................................3 - 4

3.5.2 Using the On-Screen Keyboard ....................................................................................................................................................3 - 5

3.5.3 Using the Navigation Knob (for N22/N19) ...............................................................................................................................3 - 5

3.5.4 Using the Barcode Reader ..............................................................................................................................................................3 - 5

3.5.5 Using the Remote Controller ........................................................................................................................................................3 - 5

3.6 Screen Display ..................................................................................................................................................................................................3 - 5

3.6.1 On-screen Symbols ...........................................................................................................................................................................3 - 6

3.6.2 Dialogs ...................................................................................................................................................................................................3 - 7

3.6.3 Quick Keys ............................................................................................................................................................................................3 - 8

3.7 Operating Modes .......................................................................................................................................................................................... 3 - 10

3.7.1 Monitoring Mode ........................................................................................................................................................................... 3 - 10

3.7.2 Privacy Mode .................................................................................................................................................................................... 3 - 10

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 1

3.7.3 Night Mode ...................................................................................................................................................................................... 3 - 11

3.7.4 Standby Mode ................................................................................................................................................................................. 3 - 12

3.7.5 Discharge Mode ............................................................................................................................................................................. 3 - 12

3.8 Configuring the Monitor ........................................................................................................................................................................... 3 - 12

3.8.1 Setting the Screen Orientation (for N22 and N19) ............................................................................................................. 3 - 13

3.8.2 Setting the Date and Time .......................................................................................................................................................... 3 - 13

3.8.3 Adjusting the Screen Brightness .............................................................................................................................................. 3 - 13

3.8.4 Adjusting the Volume .................................................................................................................................................................. 3 - 13

3.9 General Operation ....................................................................................................................................................................................... 3 - 14

3.9.1 Switching On or Off a Parameter ............................................................................................................................................. 3 - 14

3.9.2 Displaying Parameter Numerics and Waveforms .............................................................................................................. 3 - 14

3.9.3 Displaying the Parameter List ................................................................................................................................................... 3 - 14

3.9.4 Accessing Parameter Setup Dialogs ....................................................................................................................................... 3 - 14

3.9.5 Changing Measurement Colors ................................................................................................................................................ 3 - 15

3.10 Initiating a Manual Event ........................................................................................................................................................................ 3 - 15

3.11 Freezing Waveforms ................................................................................................................................................................................. 3 - 15

3.11.1 Freezing Waveforms ................................................................................................................................................................... 3 - 15

3.11.2 Viewing Frozen Waveforms .................................................................................................................................................... 3 - 15

3.11.3 Unfreezing Waveforms .............................................................................................................................................................. 3 - 16

3.11.4 Printing Frozen Waveforms ..................................................................................................................................................... 3 - 16

3.12 N22 and N19 Secondary Display .......................................................................................................................................................... 3 - 16

3.12.1 Connecting the Secondary Display Power Supply (for N22/N19) ............................................................................. 3 - 16

3.12.2 Setting the ECG Waveform Size for the Independent External Display (for N17/N15/N12) ............................ 3 - 16

3.13 Using the iView System ........................................................................................................................................................................... 3 - 17

3.14 Capturing the Screen ............................................................................................................................................................................... 3 - 17

3.15 Connecting the CMS ................................................................................................................................................................................. 3 - 18

3.16 Connecting the eGateway ...................................................................................................................................................................... 3 - 18

3.17 Connecting the Wireless Network ....................................................................................................................................................... 3 - 18

3.18 Disconnecting the Wireless Network ................................................................................................................................................. 3 - 18

3.19 Turning Off the Monitor .......................................................................................................................................................................... 3 - 19

4 User Screens ....................................................................................................................................................................... 4 - 1

4.1 Choosing a Screen ..........................................................................................................................................................................................4 - 1

4.2 Normal Screen ..................................................................................................................................................................................................4 - 1

4.2.1 Entering the Normal Screen ..........................................................................................................................................................4 - 1

4.2.2 Configuring the Normal Screen ...................................................................................................................................................4 - 1

4.3 The Big Numerics Screen ..............................................................................................................................................................................4 - 1

4.3.1 Entering the Big Numerics Screen ..............................................................................................................................................4 - 1

4.3.2 Configuring the Big Numerics Screen .......................................................................................................................................4 - 2

4.4 Minitrends Screen ...........................................................................................................................................................................................4 - 2

4.4.1 Entering the Minitrends Screen ...................................................................................................................................................4 - 2

4.4.2 The Display of Minitrends Screen ...............................................................................................................................................4 - 3

4.4.3 Viewing the Long Trends ...............................................................................................................................................................4 - 3

4.4.4 Setting Minitrends Parameters ....................................................................................................................................................4 - 3

4.4.5 Setting the Minitrend Length .......................................................................................................................................................4 - 3

4.4.6 Setting the Alarm Statistics Switch .............................................................................................................................................4 - 4

4.4.7 Setting the Alarm Statistics Duration ........................................................................................................................................4 - 4

2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

4.4.8 Routine Vital/Baseline ......................................................................................................................................................................4 - 4

4.5 The OxyCRG Screen ........................................................................................................................................................................................4 - 4

4.5.1 Entering the OxyCRG Screen .........................................................................................................................................................4 - 4

4.5.2 The Display of the OxyCRG Screen .............................................................................................................................................4 - 5

4.5.3 OxyCRG Events ...................................................................................................................................................................................4 - 5

4.5.4 The Display of the ABD Event Area .............................................................................................................................................4 - 6

4.5.5 Setting OxyCRG Parameters ..........................................................................................................................................................4 - 6

4.5.6 Setting the Threshold of ABD Events .........................................................................................................................................4 - 6

4.5.7 Editing ABD Events ...........................................................................................................................................................................4 - 6

4.6 The Targeted Goal Screen ............................................................................................................................................................................4 - 6

4.6.1 Entering the Targeted Goal Screen .............................................................................................................................................4 - 6

4.6.2 The Display of the Targeted Goal Screen .................................................................................................................................4 - 6

4.6.3 Configuring the Targeted Goal Screen Layout .......................................................................................................................4 - 7

4.6.4 Setting Parameter Statistics ..........................................................................................................................................................4 - 8

4.6.5 Selecting the Range of each Parameter Section and the Target Section .....................................................................4 - 9

4.6.6 Selecting the Statistics Duration of the Target Parameter .................................................................................................4 - 9

4.7 Remote View .....................................................................................................................................................................................................4 - 9

4.7.1 Entering the Remote View Screen ..............................................................................................................................................4 - 9

4.7.2 About the Remote View ..................................................................................................................................................................4 - 9

4.7.3 Adding a Bed .................................................................................................................................................................................... 4 - 10

4.7.4 Removing a Bed .............................................................................................................................................................................. 4 - 11

4.7.5 Displaying the Main Bed ............................................................................................................................................................. 4 - 11

4.7.6 Saving a Manual Event ................................................................................................................................................................. 4 - 11

4.7.7 Managing Alarms from Remote Devices ............................................................................................................................... 4 - 11

4.7.8 Resetting Alarms for Remote Devices ..................................................................................................................................... 4 - 11

4.7.9 Selecting Beds By Care Group ................................................................................................................................................... 4 - 11

4.7.10 Alarm Watch .................................................................................................................................................................................. 4 - 12

4.7.11 Auto Displaying the Alarm Bed .............................................................................................................................................. 4 - 12

5 Managing Patients ............................................................................................................................................................ 5 - 1

5.1 Starting Monitoring a Patient .....................................................................................................................................................................5 - 1

5.2 Admitting a Patient ........................................................................................................................................................................................5 - 1

5.2.1 Entering the Patient Management Dialog ...............................................................................................................................5 - 1

5.2.2 Editing Patient Information ...........................................................................................................................................................5 - 1

5.2.3 Loading Patient Information from the CMS ............................................................................................................................5 - 2

5.2.4 Loading Patient Information from the ADT Server ...............................................................................................................5 - 2

5.3 Transferring Patient ........................................................................................................................................................................................5 - 3

5.3.1 Data Storage Introduction .............................................................................................................................................................5 - 3

5.3.2 Transferring Patient Data ...............................................................................................................................................................5 - 4

5.4 Exporting Patient Data ..................................................................................................................................................................................5 - 4

5.5 Deleting Patient Data .....................................................................................................................................................................................5 - 4

5.6 Stopping a Parameter Measurement .......................................................................................................................................................5 - 5

5.7 Discharging a Patient .....................................................................................................................................................................................5 - 5

6 Interfacing with External Devices .................................................................................................................................... 6 - 1

6.1 BeneLink Introduction ...................................................................................................................................................................................6 - 1

6.2 BeneLink Safety Information .......................................................................................................................................................................6 - 1

6.3 Differences in Displayed Values .................................................................................................................................................................6 - 2

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 3

6.4 Connecting an External Device ..................................................................................................................................................................6 - 2

6.5 Accessing the Integrated Devices Screen ..............................................................................................................................................6 - 3

6.6 Displaying

Parameters from External Devices ....................................................................................................................................6 - 4

6.6.1 Setting Waveform Properties for Parameters from External Devices ............................................................................6 - 4

6.6.2 Setting Alarms from External Devices .......................................................................................................................................6 - 5

6.6.3 Setting Parameters from External Devices for Display ........................................................................................................6 - 5

6.6.4 Displaying

Data from External Devices on the Main Screen ............................................................................................6 - 5

6.6.5 Setting Units for Parameters from External Devices ............................................................................................................6 - 6

6.6.6 Accessing the Loops Screen ..........................................................................................................................................................6 - 6

6.7 Viewing Alarms from External Devices ...................................................................................................................................................6 - 6

6.8 Viewing Parameter Trends from External Devices ..............................................................................................................................6 - 6

6.9 Recording and Printing Parameter Trends from External Devices ................................................................................................6 - 6

7 Using with the Telemetry Monitor ................................................................................................................................... 7 - 1

7.1 Introduction ......................................................................................................................................................................................................7 - 1

7.1.1 Pairing Procedure .............................................................................................................................................................................7 - 1

7.2 Binding the Telemetry with the Monitor ................................................................................................................................................7 - 2

7.3 Unpairing the Telemetry and the Monitor .............................................................................................................................................7 - 2

7.3.1 Unpairing the Telemetry and the Monitor via the Monitor ...............................................................................................7 - 2

7.3.2 Unpairing the Telemetry and the Monitor via the telemetry ...........................................................................................7 - 2

7.4 Troubleshooting ..............................................................................................................................................................................................7 - 3

8 Alarms ................................................................................................................................................................................. 8 - 1

8.1 Alarm Introduction .........................................................................................................................................................................................8 - 1

8.2 Alarm Safety Information .............................................................................................................................................................................8 - 1

8.3 Understanding the Alarms ..........................................................................................................................................................................8 - 1

8.3.1 Alarm Categories ...............................................................................................................................................................................8 - 1

8.3.2 Alarm Priorities ..................................................................................................................................................................................8 - 1

8.3.3 Alarm Indicators ................................................................................................................................................................................8 - 2

8.3.4 Alarm Status Symbols ......................................................................................................................................................................8 - 3

8.4 Accessing On-screen Help for Technical Alarms (AlarmSight) .......................................................................................................8 - 3

8.5 Checking the Physiological Alarms List ..................................................................................................................................................8 - 3

8.6 Changing Alarm Settings .............................................................................................................................................................................8 - 3

8.6.1 Setting Parameter Alarm Properties ..........................................................................................................................................8 - 3

8.6.2 Changing the Alarm Volume ........................................................................................................................................................8 - 4

8.6.3 Setting the Auto Limits for New Patient Switch ....................................................................................................................8 - 5

8.6.4 Initiating Auto Alarm Limits ..........................................................................................................................................................8 - 5

8.6.5 Setting the Apnea Delay Time .....................................................................................................................................................8 - 8

8.6.6 Switching On or Off V-Tach Latching ........................................................................................................................................8 - 8

8.6.7 Restoring the Default Alarm Settings ........................................................................................................................................8 - 8

8.6.8 Setting the Length of Printed Waveforms ...............................................................................................................................8 - 8

8.7 Pausing Alarms/Pausing Alarm Tones .....................................................................................................................................................8 - 8

8.7.1 Pausing Alarms ..................................................................................................................................................................................8 - 8

8.7.2 Pausing Alarm Sound ......................................................................................................................................................................8 - 9

8.8 Resetting Alarms .......................................................................................................................................................................................... 8 - 10

8.8.1 Resetting Physiological Alarms ................................................................................................................................................. 8 - 10

8.8.2 Resetting Technical Alarms ........................................................................................................................................................ 8 - 10

4 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

8.9 Latching Alarms ............................................................................................................................................................................................ 8 - 10

8.10 Nurse Call ...................................................................................................................................................................................................... 8 - 11

8.11 Calling for Help ........................................................................................................................................................................................... 8 - 11

8.11.1 Issuing the Call Help Signal ...................................................................................................................................................... 8 - 11

8.12 CPB Mode ...................................................................................................................................................................................................... 8 - 11

8.12.1 Entering the CPB Mode .............................................................................................................................................................. 8 - 11

8.12.2 Exiting the CPB Mode ................................................................................................................................................................. 8 - 12

8.13 Intubation Mode ........................................................................................................................................................................................ 8 - 12

8.13.1 Entering the Intubation Mode ................................................................................................................................................ 8 - 12

8.13.2 Exiting the Intubation Mode .................................................................................................................................................... 8 - 12

8.14 Testing Alarms ............................................................................................................................................................................................ 8 - 12

9 Monitoring ECG, Arrhythmia, ST and QT ......................................................................................................................... 9 - 1

9.1 ECG Introduction .............................................................................................................................................................................................9 - 1

9.2 ECG Safety Information .................................................................................................................................................................................9 - 1

9.3 ECG Display ........................................................................................................................................................................................................9 - 2

9.4 Preparing for ECG Monitoring ....................................................................................................................................................................9 - 2

9.4.1 Preparing the Patient Skin .............................................................................................................................................................9 - 2

9.4.2 Applying Electrodes and Connecting the Patient .................................................................................................................9 - 2

9.4.3 Lead Wire Color Code ......................................................................................................................................................................9 - 3

9.4.4 ECG Electrode Placements .............................................................................................................................................................9 - 3

9.4.5 Choosing the ECG Lead Type ........................................................................................................................................................9 - 6

9.4.6 Checking Paced Status ....................................................................................................................................................................9 - 6

9.4.7 Enabling Pacer Rejection ................................................................................................................................................................9 - 6

9.5 Changing ECG Settings .................................................................................................................................................................................9 - 7

9.5.1 Choosing an ECG Screen ................................................................................................................................................................9 - 7

9.5.2 Setting ECG Alarm Properties .......................................................................................................................................................9 - 7

9.5.3 Changing ECG Wave Settings .......................................................................................................................................................9 - 7

9.5.4 Disabling the Smart Lead Off Function .....................................................................................................................................9 - 9

9.5.5 Adjusting the QRS Volume ............................................................................................................................................................9 - 9

9.5.6 Adjusting the Minimum QRS Detection Threshold ..............................................................................................................9 - 9

9.6 Monitoring Arrhythmia .............................................................................................................................................................................. 9 - 10

9.6.1 Arrhythmia Safety Information ................................................................................................................................................. 9 - 10

9.6.2 Arrhythmia Events ......................................................................................................................................................................... 9 - 10

9.6.3 Displaying Arrhythmia Information ......................................................................................................................................... 9 - 11

9.6.4 Changing Arrhythmia Settings .................................................................................................................................................. 9 - 11

9.6.5 Intelligent Arrhythmia Alarm ..................................................................................................................................................... 9 - 13

9.7 ST Segment Monitoring ............................................................................................................................................................................. 9 - 15

9.7.1 ST Safety Information ................................................................................................................................................................... 9 - 15

9.7.2 Enabling ST Monitoring ............................................................................................................................................................... 9 - 15

9.7.3 Displaying ST Deviation Numerics ........................................................................................................................................... 9 - 15

9.7.4 Displaying ST Segments in the Waveform Area .................................................................................................................. 9 - 16

9.7.5 Entering the ST View ..................................................................................................................................................................... 9 - 17

9.7.6 Saving the Current ST Deviation and Segment as Baseline ............................................................................................ 9 - 17

9.7.7 Entering the ST Graphic ............................................................................................................................................................... 9 - 18

9.7.8 Changing ST Settings .................................................................................................................................................................... 9 - 18

9.7.9 Adjusting ST Measurement Points ........................................................................................................................................... 9 - 19

9.8 QT/QTc Interval Monitoring ..................................................................................................................................................................... 9 - 20

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 5

9.8.1 QT/QTc Monitoring Limitations ................................................................................................................................................ 9 - 20

9.8.2 Enabling QT/QTc Monitoring .................................................................................................................................................... 9 - 20

9.8.3 Displaying QT Numerics and Segments ................................................................................................................................ 9 - 21

9.8.4 Entering the QT View .................................................................................................................................................................... 9 - 21

9.8.5 Saving the Current QTc as Baseline ......................................................................................................................................... 9 - 22

9.8.6 Changing QT Settings .................................................................................................................................................................. 9 - 22

9.9 ECG Relearning ............................................................................................................................................................................................. 9 - 22

9.9.1 Auto ECG Relearning .................................................................................................................................................................... 9 - 22

9.9.2 Initiating an ECG Relearning Manually ................................................................................................................................... 9 - 23

9.10 Defibrillation Synchronization Pulse Output ................................................................................................................................... 9 - 23

9.11 ECG Troubleshooting ............................................................................................................................................................................... 9 - 23

10 Resting 12-Lead ECG Analysis ......................................................................................................................................10 - 1

10.1 Resting 12-Lead ECG Analysis Introduction ..................................................................................................................................... 10 - 1

10.2 Entering the 12-Lead Screen ................................................................................................................................................................. 10 - 1

10.3 Initiating Resting 12-Lead ECG Analysis ............................................................................................................................................ 10 - 1

10.4 Changing 12-Lead ECG Analysis Settings ......................................................................................................................................... 10 - 1

10.4.1 Setting the High Frequency Filter ......................................................................................................................................... 10 - 1

10.4.2 Setting the Baseline Drift Removal Filter ............................................................................................................................ 10 - 1

10.4.3 Setting the 12-Lead Waveform Layout ................................................................................................................................ 10 - 2

10.4.4 Editing Patient Information ..................................................................................................................................................... 10 - 2

10.4.5 Setting Tachycardia and Bradycardia Thresholds ........................................................................................................... 10 - 2

10.4.6 Setting the 12-Lead Interpretation Report ....................................................................................................................... 10 - 3

10.5 Saving the 12-Lead Interpretation Report ........................................................................................................................................ 10 - 3

10.6 Printing the 12-Lead Interpretation Report ..................................................................................................................................... 10 - 3

10.7 Exiting the ECG 12-Lead Screen ........................................................................................................................................................... 10 - 3

11 Monitoring Impedance Respiration (Resp) .................................................................................................................11 - 1

11.1 Resp Introduction ...................................................................................................................................................................................... 11 - 1

11.2 Impedance Resp Safety Information .................................................................................................................................................. 11 - 1

11.3 Resp Display ................................................................................................................................................................................................ 11 - 2

11.4 Preparing for Impedance Resp Monitoring ..................................................................................................................................... 11 - 2

11.4.1 Preparing the Patient ................................................................................................................................................................. 11 - 2

11.4.2 Placing the Electrodes ............................................................................................................................................................... 11 - 3

11.5 Changing Resp Settings .......................................................................................................................................................................... 11 - 4

11.5.1 Setting the Resp Alarm Properties ........................................................................................................................................ 11 - 4

11.5.2 Setting the RR Source ............................................................................................................................................................... 11 - 4

11.5.3 Choosing the Respiration Lead .............................................................................................................................................. 11 - 4

11.5.4 Setting the Resp Waveform Size ............................................................................................................................................ 11 - 4

11.5.5 Setting the Resp Waveform Speed ....................................................................................................................................... 11 - 4

11.5.6 Setting the Auto Detection Switch ....................................................................................................................................... 11 - 5

11.5.7 Adjusting the Resp Waveform Detection Threshold ...................................................................................................... 11 - 5

12 Monitoring Pulse Oxygen Saturation (SpO

2

) ..............................................................................................................12 - 1

12.1 SpO

2

Introduction ..................................................................................................................................................................................... 12 - 1

12.2 SpO

2

Safety Information ......................................................................................................................................................................... 12 - 2

12.3 SpO

2

Measurement Limitations ........................................................................................................................................................... 12 - 3

12.4 SpO

2

Display ................................................................................................................................................................................................ 12 - 4

6 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

12.5 Preparing for SpO

2

Monitoring ............................................................................................................................................................. 12 - 4

12.6 Changing the SpO

2

Settings .................................................................................................................................................................. 12 - 5

12.6.1 Changing the SpO

2

Alarm Settings ....................................................................................................................................... 12 - 5

12.6.2 Nellcor SatSeconds

TM

Alarm Management ....................................................................................................................... 12 - 5

12.6.3 Setting SpO

2

Sensitivity (for Masimo SpO

2

) ....................................................................................................................... 12 - 6

12.6.4 Enabling FastSAT (for Masimo SpO

2

) .................................................................................................................................... 12 - 7

12.6.5 Displaying SIQ (for Masimo SpO

2

) ......................................................................................................................................... 12 - 7

12.6.6 Changing Averaging Time (for Masimo SpO

2

) .................................................................................................................. 12 - 7

12.6.7 Monitoring SpO

2

and NIBP Simultaneously ....................................................................................................................... 12 - 8

12.6.8 Changing the Sweep Speed of the Pleth Waveform ...................................................................................................... 12 - 8

12.7 Changing the PR Settings ....................................................................................................................................................................... 12 - 8

12.7.1 Changing the PR Alarm Settings ............................................................................................................................................ 12 - 8

12.7.2 Changing the QRS Volume ....................................................................................................................................................... 12 - 8

12.7.3 Setting the PR Source ................................................................................................................................................................. 12 - 8

12.7.4 Showing/Hiding PR ..................................................................................................................................................................... 12 - 9

12.8 SpO

2

Troubleshooting ............................................................................................................................................................................. 12 - 9

12.9 Nellcor Information ................................................................................................................................................................................12 - 10

12.10 Masimo Information .............................................................................................................................................................................12 - 10

12.11 Masimo End-User License Agreement ...........................................................................................................................................12 - 10

13 Monitoring Temperature (Temp) ..................................................................................................................................13 - 1

13.1 Temperature Introduction ...................................................................................................................................................................... 13 - 1

13.2 Temperature Safety Information .......................................................................................................................................................... 13 - 1

13.3 Temperature Display ................................................................................................................................................................................ 13 - 1

13.4 Preparing for Temperature Monitoring ............................................................................................................................................. 13 - 2

13.5 Changing Temperature Settings .......................................................................................................................................................... 13 - 2

13.5.1 Setting the Temperature Alarm Properties ........................................................................................................................ 13 - 2

13.5.2 Selecting the Temperature Label ........................................................................................................................................... 13 - 2

13.5.3 Displaying the Temperature Difference .............................................................................................................................. 13 - 2

13.6 Temperature Troubleshooting ............................................................................................................................................................. 13 - 3

14 Monitoring Noninvasive Blood Pressure (NIBP) .........................................................................................................14 - 1

14.1 NIBP Introduction ...................................................................................................................................................................................... 14 - 1

14.2 NIBP Safety Information .......................................................................................................................................................................... 14 - 1

14.3 NIBP Measurement Limitations ............................................................................................................................................................ 14 - 2

14.4 Measurement Modes ................................................................................................................................................................................ 14 - 2

14.5 NIBP Display ................................................................................................................................................................................................. 14 - 3

14.6 Preparing for NIBP Measurements ...................................................................................................................................................... 14 - 3

14.6.1 Preparing the Patient for NIBP Measurements ................................................................................................................. 14 - 3

14.6.2 Placing the NIBP Cuff .................................................................................................................................................................. 14 - 4

14.7 Starting and Stopping NIBP Measurements .................................................................................................................................... 14 - 4

14.8 Changing NIBP Settings ........................................................................................................................................................................... 14 - 5

14.8.1 Setting the NIBP Alarm Properties ......................................................................................................................................... 14 - 5

14.8.2 Setting the Initial Cuff Inflation Pressure ............................................................................................................................ 14 - 5

14.8.3 Setting the NIBP Interval ........................................................................................................................................................... 14 - 5

14.8.4 Selecting NIBP Start Mode ........................................................................................................................................................ 14 - 5

14.8.5 Enabling the NIBP End Tone .................................................................................................................................................... 14 - 6

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 7

14.8.6 Setting NIBP Sequence .............................................................................................................................................................. 14 - 6

14.8.7 Setting the NIBP Display Format ............................................................................................................................................ 14 - 6

14.8.8 Setting the NIBP Alarm Limits Display Switch .................................................................................................................. 14 - 6

14.8.9 Showing/Hiding PR ..................................................................................................................................................................... 14 - 6

14.8.10 Correcting the NIBP Measurements ................................................................................................................................... 14 - 6

14.9 Assisting Venous Puncture ..................................................................................................................................................................... 14 - 6

14.10 NIBP Troubleshooting ........................................................................................................................................................................... 14 - 7

15 Monitoring Invasive Blood Pressure (IBP) ...................................................................................................................15 - 1

15.1 IBP Introduction ......................................................................................................................................................................................... 15 - 1

15.2 IBP Safety Information ............................................................................................................................................................................. 15 - 1

15.3 Preparing for IBP Monitoring ................................................................................................................................................................ 15 - 2

15.3.1 IBP Equipment to Patient Connection ................................................................................................................................. 15 - 2

15.3.2 Measuring an Invasive Blood Pressure ................................................................................................................................ 15 - 2

15.3.3 Zeroing the IBP transducer ...................................................................................................................................................... 15 - 3

15.4 Measuring ICP Using the Codman ICP Transducer ....................................................................................................................... 15 - 3

15.4.1 Zeroing the Codman ICP transducer .................................................................................................................................... 15 - 3

15.4.2 Measuring ICP ............................................................................................................................................................................... 15 - 4

15.5 IBP Display .................................................................................................................................................................................................... 15 - 4

15.5.1 Overlapping IBP Waveforms ................................................................................................................................................... 15 - 5

15.6 Changing IBP Settings ............................................................................................................................................................................. 15 - 5

15.6.1 Changing the IBP Alarm Settings .......................................................................................................................................... 15 - 5

15.6.2 Changing the Pressure Label .................................................................................................................................................. 15 - 5

15.6.3 Setting the Pressure Type for Display .................................................................................................................................. 15 - 6

15.6.4 Changing the Sensitivity ........................................................................................................................................................... 15 - 6

15.6.5 Setting the IBP Waveform ........................................................................................................................................................ 15 - 6

15.6.6 Setting the Display Format of Artery Pressure .................................................................................................................. 15 - 7

15.6.7 Showing/Hiding the Alarm Limits of Artery Pressure .................................................................................................... 15 - 7

15.6.8 Setting the Use PA-D as PAWP Switch ................................................................................................................................. 15 - 7

15.6.9 Enabling PPV Measurement .................................................................................................................................................... 15 - 7

15.7 PAWP .............................................................................................................................................................................................................. 15 - 8

15.7.1 PAWP Equipment to Patient Connection ........................................................................................................................... 15 - 8

15.7.2 Preparing to Measure PAWP ................................................................................................................................................... 15 - 8

15.7.3 Measuring PAWP ......................................................................................................................................................................... 15 - 9

15.7.4 Setting the Waveforms of the PAWP Screen ................................................................................................................... 15 - 10

15.7.5 Performing Hemodynamic Calculation ............................................................................................................................. 15 - 10

15.8 Connecting a Camino Device .............................................................................................................................................................. 15 - 10

15.9 IBP Troubleshooting ...............................................................................................................................................................................15 - 11

16 Monitoring Cardiac Output (C.O.) ................................................................................................................................16 - 1

16.1 C.O. Introduction ....................................................................................................................................................................................... 16 - 1

16.2 C.O. Safety Information ........................................................................................................................................................................... 16 - 1

16.3 C.O. Measurement Limitations ............................................................................................................................................................. 16 - 2

16.4 C.O. Display .................................................................................................................................................................................................. 16 - 2

16.5 C.O. Equipment to Patient Connection ............................................................................................................................................. 16 - 3

16.6 Performing C.O. Measurement ............................................................................................................................................................. 16 - 3

16.6.1 Preparing for C.O. Measurement ........................................................................................................................................... 16 - 3

16.6.2 Setting C.O. Measurement ....................................................................................................................................................... 16 - 3

8 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

16.6.3 Performing C.O. Measurement ............................................................................................................................................... 16 - 4

16.7 Changing C.O. Settings ............................................................................................................................................................................ 16 - 5

16.7.1 Setting C.O. Alarm Properties .................................................................................................................................................. 16 - 5

16.8 C.O. Troubleshooting ............................................................................................................................................................................... 16 - 5

17 Monitoring Central Venous Oxygen Saturation (ScvO

2

) ............................................................................................17 - 1

17.1 ScvO

2

Introduction .................................................................................................................................................................................... 17 - 1

17.2 ScvO

2

Safety Information ........................................................................................................................................................................ 17 - 1

17.3 ScvO

2

Display .............................................................................................................................................................................................. 17 - 1

17.4 Accessing the On-screen ScvO

2

Guide ............................................................................................................................................... 17 - 2

17.5 ScvO

2

Equipment to Patient Connection .......................................................................................................................................... 17 - 2

17.6 Measuring ScvO

2 ..........................................................................................................................................................................................................................................17 - 2

17.7 ScvO

2

Calibration ....................................................................................................................................................................................... 17 - 3

17.8 Accessing the HemoSight Dialog ......................................................................................................................................................... 17 - 3

17.9 Changing ScvO

2

Settings ........................................................................................................................................................................ 17 - 3

17.9.1 Changing ScvO

2

Alarm Settings ............................................................................................................................................. 17 - 3

17.9.2 Setting Hb or Hct ......................................................................................................................................................................... 17 - 3

17.9.3 Inputting the SaO

2

Value .......................................................................................................................................................... 17 - 4

18 Monitoring CCO/SvO

2 ...........................................................................................................................................................................................18 - 1

18.1 CCO/SvO

2

Introduction ............................................................................................................................................................................ 18 - 1

18.2 CCO/SvO

2

Safety Information ............................................................................................................................................................... 18 - 2

18.3 Connecting the Device ............................................................................................................................................................................ 18 - 2

18.3.1 CCO/SvO2 Cable ........................................................................................................................................................................... 18 - 2

18.3.2 Connecting to the Vigilance II Monitor ................................................................................................................................ 18 - 2

18.3.3 Connecting to the Vigileo Monitor ........................................................................................................................................ 18 - 4

18.3.4 Connecting the EV1000 Monitor ............................................................................................................................................ 18 - 5

18.3.5 Connecting the HemoSphere Monitor ................................................................................................................................ 18 - 5

18.4 Accessing the HemoSight dialog ......................................................................................................................................................... 18 - 6

18.5 Changing CCO Settings ........................................................................................................................................................................... 18 - 6

18.5.1 Changing the CCO Alarm Settings ........................................................................................................................................ 18 - 6

18.5.2 Setting Parameters for Display ............................................................................................................................................... 18 - 6

18.5.3 Setting the CCO Analog Output Signal ............................................................................................................................... 18 - 7

18.6 Changing SvO

2

/ScvO

2

Settings ............................................................................................................................................................ 18 - 7

18.6.1 Changing the SvO

2

/ScvO

2

Alarm Settings ......................................................................................................................... 18 - 7

18.6.2 Setting the SvO

2

/ScvO

2

Analog Output Signal ................................................................................................................. 18 - 8

18.7 CCO/SvO

2

Troubleshooting ................................................................................................................................................................... 18 - 9

19 Monitoring Carbon Dioxide (CO

2

) ................................................................................................................................19 - 1

19.1 CO

2

Introduction ........................................................................................................................................................................................ 19 - 1

19.2 Identifying CO

2

Modules ......................................................................................................................................................................... 19 - 1

19.3 CO

2

Safety Information ............................................................................................................................................................................ 19 - 2

19.4 CO

2

Measurement Limitations .............................................................................................................................................................. 19 - 2

19.5 CO

2

Display .................................................................................................................................................................................................. 19 - 2

19.6 Measuring CO

2

Using the Sidestream/Microstream CO

2

Module ........................................................................................... 19 - 3

19.6.1 Preparing to Measure CO

2

Using the Sidestream CO

2

Module .................................................................................. 19 - 3

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 9

19.6.2 Preparing to Measure CO

2

Using the Microstream CO

2

Module ............................................................................... 19 - 4

19.6.3 Zeroing the Sidestream/Microstream CO

2

Module ........................................................................................................ 19 - 5

19.7 Changing Settings for All CO

2

Modules ............................................................................................................................................ 19 - 5

19.7.1 Changing CO

2

Alarm Settings ................................................................................................................................................ 19 - 5

19.7.2 Setting the CO

2

Waveform ....................................................................................................................................................... 19 - 5

19.7.3 Setting the RR Source ................................................................................................................................................................ 19 - 6

19.7.4 Entering the Standby Mode .................................................................................................................................................... 19 - 6

19.7.5 Entering the Intubation Mode ................................................................................................................................................ 19 - 6

19.7.6 Setting the Auto Standby ......................................................................................................................................................... 19 - 6

19.7.7 Setting Humidity Compensation ........................................................................................................................................... 19 - 7

19.8 Setting the Gas Compensation ............................................................................................................................................................. 19 - 7

19.9 Choosing a Time Interval for Peak-Picking ....................................................................................................................................... 19 - 7

19.10 Performing the Leakage Test .............................................................................................................................................................. 19 - 7

19.11 CO

2

Calibration ........................................................................................................................................................................................ 19 - 8

19.12 Test Method Used to Determine the Respiration Rate Measuring Range ......................................................................... 19 - 8

19.13 CO

2

Troubleshooting ............................................................................................................................................................................. 19 - 8

19.13.1 Troubleshooting the Sidestream/Microstream CO

2

Module .................................................................................... 19 - 9

19.14 Oridion Information ............................................................................................................................................................................... 19 - 9

20 Monitoring Anesthetic Gas (AG) ...................................................................................................................................20 - 1

20.1 AG Introduction ......................................................................................................................................................................................... 20 - 1

20.2 AG Safety Information ............................................................................................................................................................................. 20 - 2

20.3 AG Measurement Limitations ............................................................................................................................................................... 20 - 2

20.4 AG Display .................................................................................................................................................................................................... 20 - 3

20.5 AG Equipment to Patient Connection ............................................................................................................................................... 20 - 3

20.6 Preparing for AG Monitoring ................................................................................................................................................................. 20 - 4

20.7 Zeroing the AG Module ........................................................................................................................................................................... 20 - 4

20.8 MAC Values .................................................................................................................................................................................................. 20 - 5

20.9 Changing AG Settings .............................................................................................................................................................................. 20 - 6

20.9.1 Changing AG Alarm Settings .................................................................................................................................................. 20 - 6

20.9.2 Setting the O

2

Compensation ................................................................................................................................................. 20 - 6

20.9.3 Entering the Standby Mode .................................................................................................................................................... 20 - 6

20.9.4 Setting Auto Standby ................................................................................................................................................................. 20 - 6

20.9.5 Setting the Gas Waveform ....................................................................................................................................................... 20 - 7

20.9.6 Setting the RR Source ................................................................................................................................................................ 20 - 7

20.9.7 Entering the Intubation Mode ................................................................................................................................................ 20 - 7

20.9.8 Enabling or Disabling MAC Display ...................................................................................................................................... 20 - 7

20.10 Changing the Anesthetic Agent ........................................................................................................................................................ 20 - 7

20.11 Performing AG Leakage Test .............................................................................................................................................................. 20 - 8

20.12 Calibrating the AG Module .................................................................................................................................................................. 20 - 8

20.13 AG Troubleshooting ............................................................................................................................................................................... 20 - 8

21 Monitoring Electroencephalogram (EEG) ...................................................................................................................21 - 1

21.1 EEG Introduction ........................................................................................................................................................................................ 21 - 1

21.2 EEG Safety Information ............................................................................................................................................................................ 21 - 1

21.3 EEG Parameters .......................................................................................................................................................................................... 21 - 2

10 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

21.4 EEG Display ................................................................................................................................................................................................... 21 - 2

21.5 Accessing the On-screen EEG Guide ................................................................................................................................................... 21 - 3

21.6 Prepare for EEG Monitoring ................................................................................................................................................................... 21 - 3

21.6.1 EEG Equipment to Patient Connection ................................................................................................................................ 21 - 3

21.6.2 EEG Electrode Montages ........................................................................................................................................................... 21 - 4

21.6.3 Attaching EEG Electrodes ......................................................................................................................................................... 21 - 6

21.7 Performing EEG Sensor Check ............................................................................................................................................................... 21 - 6

21.7.1 Setting the Interval of Auto Sensor Check .......................................................................................................................... 21 - 6

21.7.2 Displaying/Hiding Impedance Value .................................................................................................................................... 21 - 7

21.7.3 Manually Starting a Sensor Check ......................................................................................................................................... 21 - 7

21.7.4 EEG Electrode Status ................................................................................................................................................................... 21 - 7

21.7.5 Stopping Sensor Check ............................................................................................................................................................. 21 - 7

21.8 Changing EEG Settings ............................................................................................................................................................................ 21 - 8

21.8.1 Changing the EEG Scale ............................................................................................................................................................ 21 - 8

21.8.2 Changing the EEG Sweep Speed ............................................................................................................................................ 21 - 8

21.8.3 Changing the High/Low Filter ................................................................................................................................................. 21 - 8

21.8.4 Switching Off the Notch Filter ................................................................................................................................................. 21 - 8

21.8.5 Changing Displayed EEG Parameters ................................................................................................................................... 21 - 8

21.9 Displaying the EEG Expand View ......................................................................................................................................................... 21 - 8

21.9.1 CSA View ......................................................................................................................................................................................... 21 - 9

21.9.2 DSA View ......................................................................................................................................................................................... 21 - 9

21.10 EEG Troubleshooting ...........................................................................................................................................................................21 - 11

22 Monitoring Bispectral Index (BIS) ................................................................................................................................22 - 1

22.1 BIS Introduction .......................................................................................................................................................................................... 22 - 1

22.2 BIS Safety Information .............................................................................................................................................................................. 22 - 2

22.3 BIS Parameters ............................................................................................................................................................................................ 22 - 3

22.4 BIS Display .................................................................................................................................................................................................... 22 - 4

22.5 Accessing the On-screen BIS Guide ..................................................................................................................................................... 22 - 4

22.6 Preparing for BIS Monitoring ................................................................................................................................................................. 22 - 5

22.7 Changing BIS Settings .............................................................................................................................................................................. 22 - 6

22.7.1 Setting BIS Alarm Properties .................................................................................................................................................... 22 - 6

22.7.2 Choosing the BIS Smoothing Rate ........................................................................................................................................ 22 - 6

22.7.3 Setting the Display of BIS Waveform Area .......................................................................................................................... 22 - 6

22.7.4 Switching the Filter On or Off .................................................................................................................................................. 22 - 6

22.7.5 Setting the Displayed BIS Parameters .................................................................................................................................. 22 - 6

22.8 Sensor Check ............................................................................................................................................................................................... 22 - 7

22.8.1 Automatic Sensor Check ........................................................................................................................................................... 22 - 7

22.8.2 Disabling automatic sensor check ......................................................................................................................................... 22 - 7

22.8.3 Manual Sensor Check ................................................................................................................................................................. 22 - 7

22.8.4 BIS Sensor Status .......................................................................................................................................................................... 22 - 8

22.9 Monitoring Bilateral BIS ........................................................................................................................................................................... 22 - 8

22.9.1 Entering the BIS Expand View ................................................................................................................................................. 22 - 8

22.9.2 Selecting BIS Expand View Display ........................................................................................................................................ 22 - 8

22.10 BIS Troubleshooting .............................................................................................................................................................................22 - 10

22.11 Covidien Information ...........................................................................................................................................................................22 - 10

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 11

23 Monitoring Neuromuscular Transmission (NMT) .......................................................................................................23 - 1

23.1 NMT Introduction ...................................................................................................................................................................................... 23 - 1

23.2 NMT Safety Information .......................................................................................................................................................................... 23 - 1

23.3 Stimulation Modes .................................................................................................................................................................................... 23 - 2

23.3.1 Train-Of-Four (TOF) ..................................................................................................................................................................... 23 - 2

23.3.2 Single Twitch (ST) ........................................................................................................................................................................ 23 - 2

23.3.3 Post-Tetanic Count (PTC) .......................................................................................................................................................... 23 - 2

23.3.4 Double Burst Stimulation (DBS) ............................................................................................................................................. 23 - 3

23.4 NMT Parameters ......................................................................................................................................................................................... 23 - 3

23.5 NMT Display ................................................................................................................................................................................................. 23 - 3

23.6 Accessing the On-screen NMT Guide ................................................................................................................................................. 23 - 4

23.7 Preparing for NMT Monitoring ............................................................................................................................................................. 23 - 4

23.7.1 NMT Equipment to Patient Connection .............................................................................................................................. 23 - 4

23.7.2 Preparing the Skin ....................................................................................................................................................................... 23 - 5

23.7.3 Placing the Electrodes and Sensor ........................................................................................................................................ 23 - 5

23.8 Calibrating the NMT Measurement ..................................................................................................................................................... 23 - 6

23.8.1 Setting the Calibration Current .............................................................................................................................................. 23 - 6

23.8.2 Starting NMT Calibration .......................................................................................................................................................... 23 - 6

23.9 Starting NMT Measurements ................................................................................................................................................................. 23 - 7

23.10 Stopping NMT Measurements ........................................................................................................................................................... 23 - 7

23.11 Changing NMT Measurement Settings ........................................................................................................................................... 23 - 7

23.11.1 Selecting the NMT Measurement Mode ........................................................................................................................... 23 - 7

23.11.2 Changing the Stimulation Current ..................................................................................................................................... 23 - 7

23.11.3 Changing the Pulse Width ..................................................................................................................................................... 23 - 7

23.11.4 Enabling Block Recovery Notification ............................................................................................................................... 23 - 8

23.11.5 Adjusting NMT Stimulation Beep Volume ....................................................................................................................... 23 - 8

23.12 Recalling Calibration Information ..................................................................................................................................................... 23 - 8

23.13 NMT Troubleshooting ........................................................................................................................................................................... 23 - 8

24 Monitoring Regional Oxygen Saturation (rSO

2

) .........................................................................................................24 - 1

24.1 rSO

2

Introduction ...................................................................................................................................................................................... 24 - 1

24.2 rSO

2 Safety Information .......................................................................................................................................................................... 24 - 2

24.3 rSO

2

Measurement Limitations ............................................................................................................................................................ 24 - 2

24.4 rSO

2

Display ................................................................................................................................................................................................. 24 - 2

24.5 Accessing the On-screen rSO

2

Guide ................................................................................................................................................. 24 - 3

24.6 Preparing for rSO

2

Monitoring ............................................................................................................................................................. 24 - 3

24.6.1 rSO

2

Sensor Site Selection ........................................................................................................................................................ 24 - 3

24.6.2 Preparing the Skin ....................................................................................................................................................................... 24 - 3

24.6.3 Applying the rSO

2

Sensor ......................................................................................................................................................... 24 - 4

24.6.4 Connecting the rSO

2

Parts ....................................................................................................................................................... 24 - 4

24.7 Changing rSO

2

Settings .......................................................................................................................................................................... 24 - 5

24.7.1 Changing rSO

2

Alarm Settings ............................................................................................................................................... 24 - 5

24.7.2 Setting the rSO

2

Auto Low Limit Switch ............................................................................................................................ 24 - 5

24.7.3 Setting the rSO

2

Label ............................................................................................................................................................... 24 - 5

24.7.4 Setting the AUC Mode ............................................................................................................................................................... 24 - 5

24.7.5 Setting the Baseline .................................................................................................................................................................... 24 - 5

12 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

24.7.6 Selecting rSO

2

Parameters for Display ................................................................................................................................. 24 - 6

24.8 Covidien Information ............................................................................................................................................................................... 24 - 6

25 Review ............................................................................................................................................................................25 - 1

25.1 Review Overview ........................................................................................................................................................................................ 25 - 1

25.2 Review Dialog .............................................................................................................................................................................................. 25 - 1

25.2.1 Accessing the Review Dialog ................................................................................................................................................... 25 - 1

25.2.2 Example Review Dialog ............................................................................................................................................................. 25 - 1

25.2.3 Symbols on Review Pages ........................................................................................................................................................ 25 - 2

25.2.4 Common Operations .................................................................................................................................................................. 25 - 2

25.2.5 Tabular Trends Review Page .................................................................................................................................................... 25 - 3

25.2.6 Graphics Trends Review Page ................................................................................................................................................. 25 - 4

25.2.7 Events Review Page .................................................................................................................................................................... 25 - 5

25.2.8 Full Disclosure Review Page .................................................................................................................................................... 25 - 7

25.2.9 Arrhy Statistics Review Page .................................................................................................................................................... 25 - 9

25.2.10 OxyCRG Review Page ............................................................................................................................................................... 25 - 9

25.2.11 12-Lead ECG Review Page ....................................................................................................................................................25 - 10

25.2.12 ST Review Page ........................................................................................................................................................................25 - 11

25.3 Reviewing Discharged Patients ..........................................................................................................................................................25 - 12

25.3.1 Checking the Data of a Discharged Patient .....................................................................................................................25 - 12

25.3.2 Checking the Information of a Discharged Patient .......................................................................................................25 - 12

26 Clinical Assistive Applications (CAA) ...........................................................................................................................26 - 1

26.1 Checking Software Licenses ................................................................................................................................................................... 26 - 1

26.2 Early Warning Score (EWS) ...................................................................................................................................................................... 26 - 1

26.2.1 Displaying the EWS Numerics Area ....................................................................................................................................... 26 - 2

26.2.2 Accessing the EWS Screen ........................................................................................................................................................ 26 - 2

26.2.3 Performing EWS Scoring ........................................................................................................................................................... 26 - 3

26.2.4 Changing EWS Settings ............................................................................................................................................................. 26 - 4

26.2.5 Viewing Historical Scores .......................................................................................................................................................... 26 - 5

26.2.6 Viewing Parameter Trends ....................................................................................................................................................... 26 - 5

26.3 Glasgow Coma Scale (GCS) .................................................................................................................................................................... 26 - 5

26.3.1 Displaying the GCS Parameter Area ...................................................................................................................................... 26 - 6

26.3.2 Accessing the GCS Dialog ......................................................................................................................................................... 26 - 6

26.3.3 Performing GCS Scoring ............................................................................................................................................................ 26 - 7

26.3.4 Setting GCS Scoring Interval .................................................................................................................................................... 26 - 7

26.3.5 Reviewing GCS Trend Data ...................................................................................................................................................... 26 - 7

26.4 SepsisSight

TM ..................................................................................................................................................................................................................................................26 - 7

26.4.1 Accessing the SepsisSight Dialog .......................................................................................................................................... 26 - 8

26.4.2 Screening ........................................................................................................................................................................................ 26 - 8

26.4.3 Recommendations ...................................................................................................................................................................... 26 - 9

26.4.4 Reviewing SepsisSight Trend Data ........................................................................................................................................ 26 - 9

26.5 ECG 24h Summary ..................................................................................................................................................................................... 26 - 9

26.5.1 Viewing the ECG 24h Summary ............................................................................................................................................26 - 10

26.5.2 Selecting Typical ECG Strips ...................................................................................................................................................26 - 10

26.5.3 Reviewing ECG Summary ........................................................................................................................................................26 - 10

26.6 HemoSight

TM ............................................................................................................................................................................................................................................... 26 - 10

26.6.1 Accessing the HemoSight Dialog ........................................................................................................................................26 - 11

26.6.2 Viewing Hemodynamic Parameters ...................................................................................................................................26 - 11

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 13

26.6.3 Physiology Graphics ................................................................................................................................................................. 26 - 13

26.6.4 Physiology Relationship .......................................................................................................................................................... 26 - 15

26.6.5 Hemodynamic Test ................................................................................................................................................................... 26 - 16

26.6.6 Following Up the Patient Hemodynamic Status ............................................................................................................ 26 - 17

26.6.7 Evaluating the Hemodynamic Parameters ...................................................................................................................... 26 - 18

26.6.8 Setting Hemodynamic Parameter Ranges ....................................................................................................................... 26 - 20

26.6.9 Restoring Default HemoSight Range Values ................................................................................................................... 26 - 20

26.6.10 Setting the Pattern ................................................................................................................................................................. 26 - 21

27 Calculation .....................................................................................................................................................................27 - 1

27.1 Calculation Overview ............................................................................................................................................................................... 27 - 1

27.2 Calculation Safety Information ............................................................................................................................................................. 27 - 1

27.3 Drug Calculations ...................................................................................................................................................................................... 27 - 1

27.3.1 Performing Drug Calculations ................................................................................................................................................ 27 - 1

27.3.2 Checking the Titration Table ................................................................................................................................................... 27 - 2

27.3.3 Drug Calculation Formula ........................................................................................................................................................ 27 - 2

27.3.4 Titration Table Calculation Formula ..................................................................................................................................... 27 - 2

27.4 Hemodynamic Calculations ................................................................................................................................................................... 27 - 3

27.4.1 Performing Hemodynamic Calculations ............................................................................................................................. 27 - 3

27.4.2 Input Parameters for Hemodynamic Calculations .......................................................................................................... 27 - 3

27.4.3 Calculated Parameters and Formulas for Hemodynamic Calculations ................................................................... 27 - 3

27.5 Oxygenation Calculations ...................................................................................................................................................................... 27 - 4

27.5.1 Performing Oxygenation Calculations ................................................................................................................................ 27 - 4

27.5.2 Input Parameters for Oxygenation Calculations .............................................................................................................. 27 - 5

27.5.3 Calculated Parameters and Formulas for Oxygenation Calculations ....................................................................... 27 - 5

27.6 Ventilation Calculations .......................................................................................................................................................................... 27 - 6

27.6.1 Performing Ventilation Calculations .................................................................................................................................... 27 - 6

27.6.2 Input Parameters for Ventilation Calculations .................................................................................................................. 27 - 6

27.6.3 Calculated Parameters and Formulas for Ventilation Calculations ........................................................................... 27 - 6

27.7 Renal Calculations ..................................................................................................................................................................................... 27 - 7

27.7.1 Performing Renal Calculations ............................................................................................................................................... 27 - 7

27.7.2 Calculated Parameters and Formulas for Renal Calculations ...................................................................................... 27 - 7

27.7.3 Calculated Parameters and Formulas for Renal Calculations ...................................................................................... 27 - 8

28 Recording .......................................................................................................................................................................28 - 1

28.1 Recorder ........................................................................................................................................................................................................ 28 - 1

28.2 Starting Recordings .................................................................................................................................................................................. 28 - 1

28.2.1 Manually Starting Recordings ................................................................................................................................................. 28 - 1

28.2.2 Automatic Recordings ............................................................................................................................................................... 28 - 1

28.3 Stopping Recordings ................................................................................................................................................................................ 28 - 2

28.3.1 Stopping Recordings Manually .............................................................................................................................................. 28 - 2

28.3.2 Stopping Recordings Automatically .................................................................................................................................... 28 - 2

14 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

28.4 Recording Related Flags .......................................................................................................................................................................... 28 - 2

28.5 Setting up the Recorder .......................................................................................................................................................................... 28 - 2

28.6 Clearing Recording Tasks ........................................................................................................................................................................ 28 - 2

28.7 Loading Paper ............................................................................................................................................................................................. 28 - 3

28.8 Removing Paper Jam ................................................................................................................................................................................ 28 - 3

28.9 List of Recording Reports ........................................................................................................................................................................ 28 - 3

29 Printing ..........................................................................................................................................................................29 - 1

29.1 Supported Printer ...................................................................................................................................................................................... 29 - 1

29.2 End Case Reports ....................................................................................................................................................................................... 29 - 1

29.2.1 Printing the End Case Report .................................................................................................................................................. 29 - 1

29.2.2 Setting a Report as An End Case Report .............................................................................................................................. 29 - 1

29.2.3 Configuring the End Case Reports ........................................................................................................................................ 29 - 1

29.2.4 Setting the End Case Report Period ...................................................................................................................................... 29 - 2

29.3 Manually Starting a Printing Task ........................................................................................................................................................ 29 - 2

29.3.1 Printing from the Screen ........................................................................................................................................................... 29 - 2

29.3.2 Printing Realtime Reports ......................................................................................................................................................... 29 - 2

29.3.3 Printing Most Common Reports ............................................................................................................................................. 29 - 2

29.4 Automatically Printing Reports ............................................................................................................................................................. 29 - 2

29.5 Stopping a Printing Task ......................................................................................................................................................................... 29 - 3

29.6 Configuring Reports .................................................................................................................................................................................. 29 - 3

29.6.1 Configuring ECG Reports .......................................................................................................................................................... 29 - 3

29.6.2 Configuring Realtime Reports ................................................................................................................................................. 29 - 4

29.6.3 Configuring Tabular Trends Reports .................................................................................................................................... 29 - 4

29.6.4 Configuring Graphic Trends Reports .................................................................................................................................... 29 - 5

29.7 Viewing Printer Status .............................................................................................................................................................................. 29 - 5

29.8 Printer Out of Paper .................................................................................................................................................................................. 29 - 5

29.9 List of Reports .............................................................................................................................................................................................. 29 - 5

30 Using the On-Screen Timers .........................................................................................................................................30 - 1

30.1 Displaying Timers ...................................................................................................................................................................................... 30 - 1

30.2 Controlling the Timer ............................................................................................................................................................................... 30 - 1

30.3 Setting the Timer ....................................................................................................................................................................................... 30 - 1

31 Managing Configurations .............................................................................................................................................31 - 1

31.1 Configuration Introduction .................................................................................................................................................................... 31 - 1

31.2 Changing the Department ..................................................................................................................................................................... 31 - 1

31.3 Setting Default Patient Category ......................................................................................................................................................... 31 - 1

31.4 Setting Default Configuration ............................................................................................................................................................... 31 - 1

31.5 .......................................................................................................................................................................... Defining Age Segments 31 - 2

31.6 Saving Current Settings ........................................................................................................................................................................... 31 - 2

31.7 Deleting a Configuration ......................................................................................................................................................................... 31 - 2

31.8 Transferring a Configuration ................................................................................................................................................................. 31 - 3

31.8.1 Exporting a Configuration ........................................................................................................................................................ 31 - 3

31.8.2 Importing a Configuration ....................................................................................................................................................... 31 - 3

31.8.3 Loading a Configuration ........................................................................................................................................................... 31 - 3

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 15

31.9 Printing Configurations ........................................................................................................................................................................... 31 - 3

31.10 Modifying Configuration Password .................................................................................................................................................. 31 - 4

32 Password-protected User Settings ..............................................................................................................................32 - 1

32.1 Setting the Device Location .................................................................................................................................................................. 32 - 1

32.1.1 Setting Monitor Information ................................................................................................................................................... 32 - 1

32.1.2 Setting Monitor Location .......................................................................................................................................................... 32 - 1

32.1.3 Enabling the Auto Obtain Bed Number Function ........................................................................................................... 32 - 1

32.2 Changing Patient Management Settings ......................................................................................................................................... 32 - 2

32.2.1 Selecting Displayed Patient Information ............................................................................................................................ 32 - 2

32.2.2 Setting the Range of Finding Patient ................................................................................................................................... 32 - 2

32.2.3 Setting the Length of Data Transferred by N1 or T1 ....................................................................................................... 32 - 2

32.2.4 Setting the Strategy of Transferring Patient Information ............................................................................................. 32 - 2

32.2.5 Transferring the N1 or T1 Settings ........................................................................................................................................ 32 - 3

32.2.6 Auto Discharging a Patient after Monitor Power Off ..................................................................................................... 32 - 3

32.2.7 Configuring Alarms for Auto Deleting Discharged Patients ....................................................................................... 32 - 3

32.2.8 Configuring Discharged Patients Data ................................................................................................................................ 32 - 3

32.2.9 Clearing All Patient Data ........................................................................................................................................................... 32 - 3

32.2.10 Selecting Query Criteria for Searching Patients in ADT Server ................................................................................. 32 - 4

32.2.11 Setting Patient Location Options for the Discharge Screen ...................................................................................... 32 - 4

32.2.12 Hiding Patient Name ............................................................................................................................................................... 32 - 4

32.3 Changing the Alarm Settings ................................................................................................................................................................ 32 - 4

32.3.1 Setting Alarm Tone Properties ............................................................................................................................................... 32 - 4

32.3.2 Setting Alarm Pausing and Resetting Properties ............................................................................................................. 32 - 5

32.3.3 Defining the Pause Function ................................................................................................................................................... 32 - 5

32.3.4 Latching Physiological Alarms ................................................................................................................................................ 32 - 7

32.3.5 Changing Remote Alarm Settings ......................................................................................................................................... 32 - 7

32.3.6 Setting Nurse Call Properties .................................................................................................................................................. 32 - 8

32.3.7 Changing Alarm Priority Settings .......................................................................................................................................... 32 - 9

32.3.8 Changing Alarm Delay Settings ........................................................................................................................................... 32 - 11

32.3.9 Adjusting the Alarm Light Brightness ................................................................................................................................ 32 - 12

32.3.10 Changing Other Alarm Settings ........................................................................................................................................ 32 - 12

32.4 Changing CAA Settings ......................................................................................................................................................................... 32 - 13

32.4.1 Enabling Inputing the Clinician ID ...................................................................................................................................... 32 - 13

32.4.2 Selecting the Default Scoring Tool ..................................................................................................................................... 32 - 13

32.4.3 Managing the Scoring Tools ................................................................................................................................................. 32 - 13

32.4.4 Setting GCS Threshold for Each Consciousness Level .................................................................................................32 - 14

32.4.5 Changing SepsisSight Settings ............................................................................................................................................. 32 - 14

32.5 Changing Module Settings .................................................................................................................................................................. 32 - 14

32.5.1 Selecting the ECG Standard ................................................................................................................................................... 32 - 14

32.5.2 Changing the QTc Formula .................................................................................................................................................... 32 - 14

32.5.3 Setting the 12-Lead Order Switch ....................................................................................................................................... 32 - 15

32.5.4 Calibrating ECG .......................................................................................................................................................................... 32 - 15

32.5.5 Manually Zeroing the CO

2

Module ..................................................................................................................................... 32 - 15

32.5.6 Hiding the Invalid Display after Zeroing the CO

2

Module ..........................................................................................32 - 15

32.5.7 Manually Zeroing the AG Module ....................................................................................................................................... 32 - 15

32.5.8 Hiding the Invalid Display after Zeroing the AG Module ............................................................................................32 - 15

32.5.9 Selecting the Gas Type for Measurement ........................................................................................................................ 32 - 16

32.5.10 Setting Parameter Measurement Timeout .................................................................................................................... 32 - 16

32.5.11 Setting Flow Rate (For Sidestream CO

2

Module Without O

2

) .................................................................................32 - 16

16 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

32.6 Changing the Review Settings ............................................................................................................................................................32 - 16

32.6.1 Hiding Undesired Review Tabs .............................................................................................................................................32 - 16

32.6.2 Renaming Events .......................................................................................................................................................................32 - 16

32.6.3 Exporting Patient Data .............................................................................................................................................................32 - 17

32.7 Setting the Printer ...................................................................................................................................................................................32 - 17

32.7.1 Setting the Printer Properties ................................................................................................................................................32 - 17

32.7.2 Setting the Report Type ..........................................................................................................................................................32 - 17

32.7.3 Setting the Report Layout ......................................................................................................................................................32 - 18

32.7.4 Setting Patient Information on 12-lead ECG Reports ...................................................................................................32 - 18

32.7.5 Configuring the Name of PDF Files .....................................................................................................................................32 - 18

32.7.6 Showing the Second Mark .....................................................................................................................................................32 - 18

32.7.7 Setting the Arrhythmia Information to Output by the Recorder .............................................................................32 - 18

32.7.8 Setting Parameter Unit ............................................................................................................................................................32 - 19

32.8 Setting the Time .......................................................................................................................................................................................32 - 19

32.8.1 Setting the Time Synchronization .......................................................................................................................................32 - 19

32.8.2 Enabling Auto Daylight Savings Time ................................................................................................................................32 - 20

32.9 Viewing Version Information ...............................................................................................................................................................32 - 20

32.10 Checking Battery Information ...........................................................................................................................................................32 - 20

32.11 Changing Secondary Display Settings (for N22/N19) ..............................................................................................................32 - 20

32.11.1 Configuring Secondary Display Screen Contents (for N22/N19/N17) .................................................................32 - 20

32.11.2 Enabling Secondary Display Alarm Indications (for N22/N19) ...............................................................................32 - 21

32.12 Changing Scanner Settings ...............................................................................................................................................................32 - 21

32.12.1 Establish the Relationship between Patient Demographics and 2D Barcode Fields (for the Mindray Custom

2D Barcode Reader) ..................................................................................................................................................................32 - 21

32.12.2 Setting the Barcode Reader Information ........................................................................................................................32 - 21

32.12.3 Identifying the Barcode Reader (for the non-Mindray Custom 2D Barcode Reader) .....................................32 - 22

32.12.4 Selecting Patient Information Read by the Barcode Reader (for the Mindray Custom 2D Barcode Reader) 32

- 22

32.13 Configuring the Network ....................................................................................................................................................................32 - 22

32.13.1 Network Safety Information ................................................................................................................................................32 - 22

32.13.2 Selecting a Network Type .....................................................................................................................................................32 - 23

32.13.3 Setting the Wired Network ..................................................................................................................................................32 - 23

32.13.4 Setting the Wireless Network .............................................................................................................................................32 - 23

32.13.5 Managing Certifications ........................................................................................................................................................32 - 24

32.13.6 Enabling Selecting a CMS .....................................................................................................................................................32 - 24

32.13.7 Adding CMSs .............................................................................................................................................................................32 - 24

32.13.8 Setting Multicast Parameters ..............................................................................................................................................32 - 24

32.13.9 Setting the Master Server Address ....................................................................................................................................32 - 25

32.13.10 Setting the Network Service Quality Level ..................................................................................................................32 - 25

32.13.11 Using the ADT Gateway ......................................................................................................................................................32 - 25

32.13.12 Sending Realtime Data, Waveforms and Alarms via HL7 Protocol .....................................................................32 - 25

32.13.13 Selecting Data Encryption Type ......................................................................................................................................32 - 26

32.13.14 Disabling the Broadcasting Patient Demographics Function ..............................................................................32 - 26

32.14 MLDAP .......................................................................................................................................................................................................32 - 26

32.14.1 Setting MLDAP .........................................................................................................................................................................32 - 26

32.14.2 Testing MLDAP Server Connection ...................................................................................................................................32 - 26

32.14.3 Selecting Password for User Authentication .................................................................................................................32 - 27

32.14.4 Setting the Password Timeout Period .............................................................................................................................32 - 27

32.15 ......................................................................................................................................................................................................................32 - 27

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 17

32.16.1 Setting Notch Filter Frequency .......................................................................................................................................... 32 - 27

32.16.2 Setting Mouse Sensitivity ..................................................................................................................................................... 32 - 27

32.16.3 Setting the Manual Event Edit Switch ............................................................................................................................. 32 - 28

32.16.4 Setting the SpO

2

Tone Mode .............................................................................................................................................. 32 - 28

32.16.5 Selecting the Language ........................................................................................................................................................ 32 - 28

32.16.6 Switching On or Off a Parameter ....................................................................................................................................... 32 - 29

32.16.7 Setting If Setting Parameter Switches is Protected .................................................................................................... 32 - 29

32.16.8 Setting If Parameter Switch is Influenced by Configuration ...................................................................................32 - 29

32.16.9 Setting Whether the Monitor Stands by if MPM module is Removed .................................................................32 - 29

32.16.10 Browsing System Log .......................................................................................................................................................... 32 - 29

32.16.11 Exporting System Log ......................................................................................................................................................... 32 - 30

33 Battery ............................................................................................................................................................................33 - 1

33.1 Battery Introduction ................................................................................................................................................................................. 33 - 1

33.2 Battery Safety Information ..................................................................................................................................................................... 33 - 1

33.3 Installing or Replacing the Battery ..................................................................................................................................................... 33 - 1

33.4 Battery Indications .................................................................................................................................................................................... 33 - 2

33.4.1 Battery LED .................................................................................................................................................................................... 33 - 2

33.4.2 Battery symbols ............................................................................................................................................................................ 33 - 3

33.4.3 Battery-related Alarms ............................................................................................................................................................... 33 - 3

33.5 Charging the Battery ................................................................................................................................................................................ 33 - 3

33.6 Maintaining the Battery .......................................................................................................................................................................... 33 - 3

33.6.1 Conditioning the Battery .......................................................................................................................................................... 33 - 3

33.6.2 Checking Battery Performance ............................................................................................................................................... 33 - 4

33.7 Storing Batteries ......................................................................................................................................................................................... 33 - 4

33.8 Recycling Batteries .................................................................................................................................................................................... 33 - 4

34 Care and Cleaning .........................................................................................................................................................34 - 1

34.1 Care and Cleaning Introduction ........................................................................................................................................................... 34 - 1

34.2 Care and Cleaning Safety Information ............................................................................................................................................... 34 - 1

34.3 Cleaning and Disinfecting the Equipment and Mounting Kits ................................................................................................. 34 - 2

34.3.1 Approved Cleaning and Disinfecting Agents .................................................................................................................. 34 - 2

34.3.2 Cleaning the Equipment and Mounting Kits ..................................................................................................................... 34 - 3

34.3.3 Disinfecting the Equipment and Mounting Kits .............................................................................................................. 34 - 3

34.4 Cleaning and Disinfecting the Accessories ...................................................................................................................................... 34 - 3

34.4.1 Approved Accessories Cleaning and Disinfecting Agents ........................................................................................... 34 - 4

34.4.2 Cleaning the Accessories .......................................................................................................................................................... 34 - 6

34.4.3 Disinfecting the Accessories .................................................................................................................................................... 34 - 6

34.5 Sterilization .................................................................................................................................................................................................. 34 - 6

35 Maintenance ..................................................................................................................................................................35 - 1

35.1 Maintenance Introduction ..................................................................................................................................................................... 35 - 1

35.2 Maintenance Safety Information ......................................................................................................................................................... 35 - 1

35.3 Maintenance and Testing Schedule ................................................................................................................................................... 35 - 2

35.4 Checking System Software Version .................................................................................................................................................... 35 - 2

35.5 Testing Methods and Procedures ........................................................................................................................................................ 35 - 2

35.5.1 Performing Visual Inspection .................................................................................................................................................. 35 - 3

18 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

35.5.2 Performing Power-on Test ....................................................................................................................................................... 35 - 3

35.5.3 Checking the NMT Sensor ........................................................................................................................................................ 35 - 3

35.5.4 Testing the Recorder .................................................................................................................................................................. 35 - 3

35.5.5 Testing the Network Printer .................................................................................................................................................... 35 - 3

35.5.6 Checking the Battery .................................................................................................................................................................. 35 - 4

35.6 NIBP Maintenance ..................................................................................................................................................................................... 35 - 4

35.6.1 NIBP Leakage Test ....................................................................................................................................................................... 35 - 4

35.6.2 NIBP Accuracy Test ...................................................................................................................................................................... 35 - 4

35.7 Disposing of the Monitor ........................................................................................................................................................................ 35 - 4

36 Accessories .....................................................................................................................................................................36 - 1

36.1 ECG Accessories .......................................................................................................................................................................................... 36 - 1

36.1.1 ECG Electrodes Available for Purchase ............................................................................................................................... 36 - 1

36.1.2 ECG Electrodes Also Compatible ........................................................................................................................................... 36 - 1

36.1.3 12-Pin Trunk Cables Available for Purchase ....................................................................................................................... 36 - 2

36.1.4 12-Pin Trunk Cables Also Compatible .................................................................................................................................. 36 - 2

36.1.5 3-lead ECG Leadwires Available for Purchase ................................................................................................................... 36 - 2

36.1.6 3-lead ECG Leadwires Also Compatible ............................................................................................................................... 36 - 3

36.1.7 5-lead ECG Leadwires Available for Purchase ................................................................................................................... 36 - 3

36.1.8 5-lead ECG Leadwires Also Compatible ............................................................................................................................... 36 - 3

36.1.9 6-lead ECG Leadwires Available for Purchase ................................................................................................................... 36 - 4

36.1.10 12-lead ECG Chest Leadwires Available for Purchase .................................................................................................. 36 - 4

36.1.11 12-lead ECG Leadwires Also Compatible .......................................................................................................................... 36 - 4

36.2 SpO

2

Accessories ........................................................................................................................................................................................ 36 - 4

36.2.1 Extension Cables and Adapter Cables Available for Purchase .................................................................................... 36 - 4

36.2.2 Extension Cables and Adapter Cables Also Compatible ............................................................................................... 36 - 4

36.2.3 Masimo SpO

2

RD Set Sensors Available for Purchase .................................................................................................... 36 - 5

36.2.4 Masimo SpO

2

LNCS Sensors Also Compatible .................................................................................................................. 36 - 5

36.2.5 Nellcor SpO

2

Sensors Also Compatible ................................................................................................................................ 36 - 5

36.3 Temp Accessories ..................................................................................................................................................................................... 36 - 6

36.3.1 Temp Cable Available for Purchase ....................................................................................................................................... 36 - 6

36.3.2 Temp Cable Also Compatible .................................................................................................................................................. 36 - 6

36.3.3 Temp Probes Available for Purchase .................................................................................................................................... 36 - 6

36.3.4 Temp Probes Also Compatible ............................................................................................................................................... 36 - 6

36.4 NIBP Accessories ........................................................................................................................................................................................ 36 - 7

36.4.1 NIBP Hoses Available for Purchase ....................................................................................................................................... 36 - 7

36.4.2 Cuffs Available for Purchase .................................................................................................................................................... 36 - 7

36.4.3 Cuffs Also Compatible ................................................................................................................................................................ 36 - 7

36.5 IBP Accessories ............................................................................................................................................................................................ 36 - 8

36.5.1 IBP Accessories Available for Purchase ................................................................................................................................ 36 - 8

36.5.2 ICP Accessories Available for Purchase ................................................................................................................................ 36 - 8

36.5.3 ICP Accessories Also Compatible ........................................................................................................................................... 36 - 8

36.6 C.O. Accessories Available for Purchase ............................................................................................................................................ 36 - 8

36.7 ScvO2 Accessories Available for Purchase ........................................................................................................................................ 36 - 9

36.8 CO

2

Accessories .......................................................................................................................................................................................... 36 - 9

36.8.1 Sidestream CO

2

Accessories Available for Purchase ....................................................................................................... 36 - 9

36.8.2 Microstream CO

2

Accessories Also Compatible ............................................................................................................... 36 - 9

36.9 AG Accessories ..........................................................................................................................................................................................36 - 10

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 19

36.9.1 AG Accessories Available for Purchase .............................................................................................................................. 36 - 10

36.9.2 AG Accessories Also Compatible ......................................................................................................................................... 36 - 10

36.10 EEG Accessories Available for Purchase ........................................................................................................................................ 36 - 10

36.11 BIS Accessories .......................................................................................................................................................................................36 - 11

36.11.1 BIS Accessories Available for Purchase ........................................................................................................................... 36 - 11

36.11.2 BIS Accessories Also Compatible ....................................................................................................................................... 36 - 11

36.12 NMT Accessories Available for Purchase ...................................................................................................................................... 36 - 11

36.13 rSO

2

Accessories ....................................................................................................................................................................................36 - 11

36.13.1 rSO

2

Accessories Available for Purchase ........................................................................................................................ 36 - 11

36.13.2 rSO

2

Accessories Also Compatible ................................................................................................................................... 36 - 11

36.14 CCO/SvO2 Accessories Available for Purchase .......................................................................................................................... 36 - 12

36.15 BeneLink Accessories ........................................................................................................................................................................... 36 - 12

36.16 External Modules ...................................................................................................................................................................................36 - 12

36.17 Mount and Mounting Accessories .................................................................................................................................................. 36 - 13

36.18 Miscellaneous Accessories ................................................................................................................................................................. 36 - 14

A Product Specifications ..................................................................................................................................................... A - 1

A.1 Monitor Safety Specifications ................................................................................................................................................................... A - 1

A.2 Physical Specifications ................................................................................................................................................................................. A - 1

A.2.1 BeneVision N22/N19 ...................................................................................................................................................................... A - 1

A.2.2 BeneVision N17/N15/N12 ............................................................................................................................................................. A - 1

A.2.3 SMR and Modules ............................................................................................................................................................................ A - 2

A.3 Environmental Specifications ................................................................................................................................................................... A - 2

A.4 Power Supply Specifications ..................................................................................................................................................................... A - 3

A.4.1 External Power Supply Specifications ...................................................................................................................................... A - 3

A.4.2 Battery Specifications .................................................................................................................................................................... A - 3

A.4.3 N17/N15/N12 Battery Specifications ........................................................................................................................................ A - 4

A.5 Display Specifications .................................................................................................................................................................................. A - 4

A.6 Recorder Specifications ............................................................................................................................................................................... A - 4

A.7 LEDs .................................................................................................................................................................................................................... A - 4

A.8 Audio Indicator ............................................................................................................................................................................................... A - 5

A.9 Monitor Interface Specifications .............................................................................................................................................................. A - 5

A.9.1 Interface Specifications of the N22/N19 Main Unit ............................................................................................................. A - 5

A.9.2 Interface Specifications of the N22/N19 Separate Primary Display .............................................................................. A - 5

A.9.3 Interface Specifications of the N22/N19 Integrated Primary Display ........................................................................... A - 5

A.9.4 Interface Specifications of the N22/N19 Secondary Display ........................................................................................... A - 5

A.9.5 Interface Specifications of the N17/N15/N12 ........................................................................................................................ A - 5

A.10 Signal Outputs Specifications ................................................................................................................................................................. A - 6

A.11 Data Storage ................................................................................................................................................................................................. A - 7

A.12 Wi-Fi Specifications .................................................................................................................................................................................... A - 7

A.12.1 Wi-Fi Technical Specifications (MSD45N) ............................................................................................................................. A - 7

A.12.2 Wi-Fi Technical Specifications (SX-SDMAC-2832S+) ........................................................................................................ A - 7

A.12.3 Wi-Fi Performance Specifications ........................................................................................................................................... A - 8

A.13 Measurement Specifications ................................................................................................................................................................... A - 9

A.13.1 ECG Specifications ...................................................................................................................................................................... A - 10

A.13.2 Resp Specifications ..................................................................................................................................................................... A - 12

20 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

A.13.3 SpO

2

Specifications ....................................................................................................................................................................A - 13

A.13.4 PR Specifications ..........................................................................................................................................................................A - 14

A.13.5 Temp Specifications ...................................................................................................................................................................A - 15

A.13.6 NIBP Specifications .....................................................................................................................................................................A - 15

A.13.7 IBP Specifications ........................................................................................................................................................................A - 17

A.13.8 C.O. Specifications .......................................................................................................................................................................A - 18

A.13.9 ScvO2/SvO2 Specifications from the Edwards Monitor ................................................................................................A - 18

A.13.10 CCO Specifications from the Edwards Monitor .............................................................................................................A - 19

A.13.11 ScvO

2

Specifications ................................................................................................................................................................A - 20

A.13.12 CO

2

Specifications ....................................................................................................................................................................A - 20

A.13.14 EEG Specifications ....................................................................................................................................................................A - 25

A.13.15 BIS Specifications ......................................................................................................................................................................A - 26

A.13.16 NMT Specifications (NMT from Mindray NMT module) ..............................................................................................A - 26

A.13.17 rSO

2

Specifications ...................................................................................................................................................................A - 27

A.14 Software Specifications ...........................................................................................................................................................................A - 27

B EMC and Radio Regulatory Compliance ..........................................................................................................................B - 1

B.1 EMC ...................................................................................................................................................................................................................... B - 1

B.2 Radio Regulatory Compliance ................................................................................................................................................................... B - 4

C Default Settings .................................................................................................................................................................C - 1

C.1 ECG, Arrhythmia, ST and QT Default Settings ...................................................................................................................................... C - 1

C.1.1 ECG Default Settings ....................................................................................................................................................................... C - 1

C.1.2 Arrhythmia Default Settings ........................................................................................................................................................ C - 2

C.1.3 ST Default Settings .......................................................................................................................................................................... C - 3

C.1.4 QT Default Settings ......................................................................................................................................................................... C - 4

C.1.5 Glasgow 12-lead ECG Algorithm Default Settings ............................................................................................................... C - 4

C.2 Respiration Default Settings ...................................................................................................................................................................... C - 4

C.3 SpO

2

/SpO

2 b Default Settings .................................................................................................................................................................... C - 5

C.4 Temperature Default Settings ................................................................................................................................................................... C - 6

C.5 NIBP Default Settings .................................................................................................................................................................................... C - 6

C.6 IBP Default Settings ....................................................................................................................................................................................... C - 8

C.7 C.O. Default Settings ...................................................................................................................................................................................C - 11

C.8 ScvO

2

Default Settings ...............................................................................................................................................................................C - 12

C.9 CCO Default Settings(Vigilance II/Vigileo/EV1000/HemoSphere) .............................................................................................C - 12

C.10 ScvO

2

/SvO

2

Default Settings(Vigilance II/Vigileo/EV1000/HemoSphere) ............................................................................C - 12

C.11 CO

2

Default Settings .................................................................................................................................................................................C - 13

C.11.1 General Settings ...........................................................................................................................................................................C - 13

C.11.2 Sidestream CO2 Default Settings ..........................................................................................................................................C - 13

C.11.3 Microstream CO2 Default Settings ........................................................................................................................................C - 13

C.12 Gas Default Settings .................................................................................................................................................................................C - 14

C.13 EEG Default Settings .................................................................................................................................................................................C - 16

C.14 BIS Default Settings ...................................................................................................................................................................................C - 17

C.15 NMT Default Settings ...............................................................................................................................................................................C - 17

C.16 rSO

2

Default Settings ...............................................................................................................................................................................C - 17

C.17 Alarm Default Settings .............................................................................................................................................................................C - 18

C.18 Display Default Settings ..........................................................................................................................................................................C - 18

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 21

D Alarm Messages ............................................................................................................................................................... D - 1

D.1 Physiological Alarm Messages .................................................................................................................................................................. D - 1

D.1.1 General Physiological Alarm Messages ................................................................................................................................... D - 1

D.1.2 Arrhythmia Alarm Messages ....................................................................................................................................................... D - 1

D.1.3 Resp Physiological Alarm Messages ......................................................................................................................................... D - 2

D.1.4 SpO

2

Physiological Alarm Messages ........................................................................................................................................ D - 2

D.1.5 PR

Physiological Alarm Messages ............................................................................................................................................. D - 2

D.1.6 NIBP Physiological Alarm Messages ......................................................................................................................................... D - 2

D.1.7 IBP Physiological Alarm Messages ............................................................................................................................................ D - 3

D.1.8 CO

2

Physiological Alarm Messages .......................................................................................................................................... D - 3

D.1.9 AG Physiological Alarm Messages ............................................................................................................................................ D - 3

D.2 Technical Alarm Messages ......................................................................................................................................................................... D - 3

D.2.1 General Technical Alarm Messages .......................................................................................................................................... D - 3

D.2.2 ECG Technical Alarm Messages ................................................................................................................................................. D - 4

D.2.3 Resp Technical Alarm Messages ................................................................................................................................................ D - 4

D.2.4 SpO

2

Technical Alarm Messages ................................................................................................................................................ D - 4

D.2.5 Temp Technical Alarm Messages .............................................................................................................................................. D - 5

D.2.6 NIBP Technical Alarm Messages ................................................................................................................................................ D - 5

D.2.7 IBP Technical Alarm Messages ................................................................................................................................................... D - 6

D.2.8 C.O. Technical Alarm Messages .................................................................................................................................................. D - 6

D.2.9 ScvO

2

Technical Alarm Messages .............................................................................................................................................. D - 7

D.2.10 CO

2

Technical Alarm Messages ............................................................................................................................................... D - 7

D.2.11 AG Technical Alarm Messages ................................................................................................................................................. D - 8

D.2.12 EEG Technical Alarm Messages ............................................................................................................................................... D - 9

D.2.13 BIS Technical Alarm Messages ................................................................................................................................................. D - 9

D.2.14 NMT Technical Alarm Messages ............................................................................................................................................D - 10

D.2.15 rSO

2

Technical Alarm Messages ............................................................................................................................................D - 11

D.2.16 Power Supply

Technical Alarm Messages ..........................................................................................................................D - 11

D.2.17 Recorder Technical Alarm Messages ...................................................................................................................................D - 12

D.2.18 Printer Technical Alarm Messages ........................................................................................................................................D - 12

D.2.19 Technical Alarm Messages Related to Networked Monitoring ..................................................................................D - 13

D.2.20 Technical Alarm Messages Related to Telemetry Monitors ........................................................................................D - 13

D.2.21 Other System Technical Alarm Messages ..........................................................................................................................D - 14

E SpO

2

Sensor Accuracy ....................................................................................................................................................... E - 1

E.1 The Accuracy of Masimo SpO

2

Sensors ................................................................................................................................................... E - 1

E.2 The Accuracy of Nellcor SpO

2

Sensors .................................................................................................................................................... E - 3

F Units, Symbols and Abbreviations ................................................................................................................................... F - 1

F.1 Units ..................................................................................................................................................................................................................... F - 1

F.2 Symbols .............................................................................................................................................................................................................. F - 2

F.3 Abbreviations ................................................................................................................................................................................................... F - 3

22 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

1

Safety

1.1

1.1.1

Safety Information

WARNING

Indicates a potential hazard or unsafe practice that, if not avoided, could result in death or serious injury.

CAUTION

Indicates a potential hazard or unsafe practice that, if not avoided, could result in minor personal injury or product/property damage.

NOTE

Provides application tips or other useful information that does not involve patient or user risk.

Warnings

WARNING

Patient with a pacemaker – on ventricular paced patients, episodes of Ventricular Tachycardia may not always be detected. Do not rely entirely upon the system’s automated arrhythmia detection algorithm. Keep pacemaker patients under close surveillance.

This equipment is used for a single patient at a time.

To avoid explosion hazards, do not use the equipment in the presence of oxygen-rich atmospheres, flammable anesthetics, or other flammable agents.

Do not touch the equipment’s metal parts or connectors when in contact with the patient; otherwise patient injury may result.

Never mix patient electrode types or brands. Dissimilar metals or other incompatibilities may cause considerable baseline drift and may increase trace recovery time after defibrillation.

To reduce the hazard of burns during high-frequency surgical procedure, ensure that the monitor’s cables and transducers never come into contact with the electro-surgery unit (ESU).

The neutral electrode of the electro-surgery unit (ESU) shall properly contact the patient. Otherwise, burns may result.

MR unsafe: the N Series monitor is not intended to be used within the Magnetic Resonance (MR) environment.

Before connecting the equipment to the mains power, ensure that the voltage and frequency ratings of the power line are the same as those indicated on the equipment’s label or in this manual.

Before putting the system into operation, the operator must verify that the equipment, connecting cables, and accessories are in correct working order and operating condition.

To avoid risk of electric shock, the equipment must only be connected to mains power with protective earth, or operated on battery power.

Do not touch the patient and live parts simultaneously. Otherwise patient injury may result.

Do not touch the patient or metal parts in contact with the patient during defibrillation. Otherwise serious injury or death could result.

Do not open the equipment housings. All servicing and future upgrades must be carried out by trained and authorized personnel.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 1 - 1

Do not rely exclusively on the audible alarm system for patient monitoring. Turning the alarm volume to a low level or off may result in a hazard to the patient. Always keep the patient under close surveillance.

Alarm settings should be customized according to patient situations.

Do not place the equipment or accessories in any position that might cause it to fall on the patient.

Do not start or operate the equipment unless the setup was verified to be correct.

To avoid inadvertent disconnection, route all cables in a way to prevent a stumbling hazard. Wrap and secure excess cabling to reduce risk of entanglement by patients or personnel.

The physiological data and alarm messages displayed on the equipment are for reference only and cannot be directly used for diagnostic interpretation. The equipment should not be used as the sole basis for medical decisions. It must be used in conjunction with clinical signs and symptom. If any measurement seems questionable, first check the patient’s vital signs by alternate means and then check the equipment for proper functioning.

1.1.2

Cautions

CAUTION

Use only parts and accessories specified in this manual.

Disposable accessories are not designed to be reused. Reuse may cause a risk of contamination and affect the measurement accuracy.

Ensure that the equipment is supplied with continuous electric power during operation from the mains power or batteries. Sudden power failure may cause data loss.

When no battery is installed, ensure that the monitor is supplied with continuous electric power during operation. Sudden power failure may lead to data loss.

Store and use the equipment in specified environmental condition. The monitor and accessories may not meet the performance specification due to aging, stored or used outside the specified temperature and humidity range.

Magnetic and electrical fields are capable of interfering with the proper performance of the equipment. For this reason make sure that all external devices operated in the vicinity of the equipment comply with the relevant EMC requirements. Mobile phones, X-ray equipment, and MRI devices are possible sources of interference as they may emit higher levels of electromagnetic radiation.

Always install or carry the equipment properly to avoid damage caused by drops, impacts, strong vibrations or other mechanical forces.

Dry the equipment immediately in case of rain or water spray.

Some settings are password protected and can only be changed by authorized personnel. Contact your department manager or biomedical engineering department for the passwords used at your facility.

Do not loop the patient cabling into a tight coil or wrap around the device, as this can damage the patient cabling.

Dispose of the package material as per the applicable waste control regulations. Keep it out of children’s reach.

At the end of its service life, the equipment, as well as its accessories, must be disposed of in compliance with the guidelines regulating the disposal of such products. If you have any questions concerning disposal of the equipment, please contact Mindray.

1.1.3

Notes

NOTE

The equipment software copyright is solely owned by Mindray. No organization or individual shall resort to modifying, copying, or exchanging it or to any other infringement on it in any form or by any means without due permission.

Put the equipment in a location where it can be easily viewed and operated.

1 - 2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

1.2

The equipment uses a mains plug as isolation means to the mains power. Do not locate the equipment in a place difficult to access the mains plug.

During normal use, the operator is expected to face the front of the equipment.

Remove the battery before shipping the monitor or if it will not be used for an extended period of time.

Remove the battery before transporting the equipment or if the equipment will not be used for a long time.

The software was developed in compliance with IEC62304, to minimize the possibility of hazards arising from software errors.

This manual is based on the maximum configuration and therefore some contents may not apply to the product. If you have any question, please contact Mindray.

Observance of this manual is a prerequisite for proper product performance and correct operation and ensures patient and operator safety.

Keep this manual in the vicinity of the equipment so that it can be referenced when needed.

Equipment Symbols

Symbol Description

General warning sign

Symbol Description

Refer to instruction manual/booklet

Serial number

Date of manufacture

USB connector

Battery indicator

Equipotentiality

Catalogue number

Manufacturer

Protected against vertically falling water drops per IEC 60529

Computer network

Alternating current

Defibrillation-proof type CF applied part

Stop USB

Input/output

Defibrillation-proof type BF applied part

Output

IBP zero key

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 1 - 3

Symbol Description

NIBP start/stop

Start

Set baseline

Stand-by

Check sensor

Gas outlet

Pushing prohibited (wheels locked, no pushing)

Video output

Dispose of in accordance to local requirements

Symbol Description

Calibration

Stop

Batch code

Menu

Graphical record

Gas inlet

Unlocking

Non-ionizing electromagnetic radiation

Plastic identification symbol

Conforms to ANSI Std. ES60601-1, IEC Std. 60601-1-8, IEC Std. 60601-2-25, IEC Std. 60601-2-26, IEC Std. 60601-

2-27, IEC Std. 80601-2-30, IEC Std. 60601-2-34, IEC Std. 60601-2-49, ISO Std. 0601-2-55, ISO Std. 80601-2-56, and ISO Std. 80601-2-61

1 - 4 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

2

Equipment Introduction

2.1

2.2

Indications for Use

The BeneVision N12/N15/N17/N19/N22 patient monitors are intended for monitoring, displaying, reviewing, storing, alarming, and transferring of multiple physiological parameters including ECG (3-lead, 5-lead, 6-lead or

12-lead selectable, Arrhythmia Detection, ST Segment Analysis, QT Analysis, and Heart Rate (HR)), Respiration

Rate (Resp), Temperature (Temp), Pulse Oxygen Saturation (SpO

2

), Pulse Rate (PR), Non-invasive Blood Pressure

(NIBP), Invasive Blood Pressure (IBP), Pulmonary Artery Wedge Pressure (PAWP), Cardiac Output (C.O.),

Continuous Cardiac Output (CCO), Mixed/Central Venous Oxygen Saturation (SvO

2

/ScvO

2

), Carbon Dioxide

(CO

2

), Oxygen (O

2

), Anesthetic Gas (AG), Impedance Cardiograph (ICG), BispectralIndex (BIS), Respiration

Mechanics (RM), Neuromuscular Transmission Monitoring (NMT), Electroencephalograph (EEG), and Regional

Oxygen Saturation (rSO

2

). The system also provides an interpretation of resting 12-lead ECG.

All the parameters can be monitored on single adult, pediatric, and neonatal patients except for the following:

■ The BIS, RM, CCO, SvO

2

/ScvO

2

, PAWP, NMT monitoring, PNP, and PNC are intended for adult and pediatric patients only.

C.O. monitoring and A-Fib are intended for adult patients only. ■

■ ICG monitoring is intended for only adult patients who meet the following requirements: height: 122 to

229cm, weight: 30 to 155kg.

■ rSO

2

monitoring is intended for use in individuals greater than 2.5 kg.

The monitors are to be used in healthcare facilities by clinical professionals or under their guidance. They should only be used by persons who have received adequate training in their use. The BeneVision N12/N15/N17/N19/

N22 monitors are not intended for helicopter transport, hospital ambulance, or home use.

NOTE

This manual is based on the maximum configuration and therefore some contents may not apply to the product. If you have any question, please contact Mindray.

Applied Parts

The applied parts of the monitor are:

ECG electrode and leadwire

SpO

2

sensor

Temp probe

NIBP cuff

IBP transducer

C.O. sensor

CCO sensor

ScvO

2

sensor

CO

2

sampling line/nasal sampling cannula, water trap, and mask

AG sampling line, water trap, airway adapter, and mask

EEG electrode

BIS sensor

NMT sensor and electrode rSO

2

sensor

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 2 - 1

2.3

System Components

The N22 and N19 monitors consist of the main unit, primary display, secondary displays, external modules, satellite module rack (SMR), input devices, and output devices.

The N17, N15, N12 monitors consist of the main unit, external display, external modules, input devices, and output devices. The N17 and N15 can also configured with the satellite module rack (SMR).

NOTE

The monitor may not include all these components. Contact the local service personnel for available components.

2.3.1

N22, N19 Main Unit

The main unit processes data from modules and sends information to displays and, optionally, among networked devices.

2.3.1.1

N22, N19 Main Unit for Integrated Installation

If the main unit and the primary display are installed together. It provides the following connectors:

(1) (2) (3) (4)

(10)

(5) (6) (7) (8) (9)

N22, N19 Main Unit Bottom View

(11)

13

2 - 2

12

(1) Cable retainer

N22, N19 Main Unit Side View

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

(2)

(3)

(4)

AC Power input

Nurse call connector (NC)

It is a BNC connector. It connects the monitor to the hospital’s nurse call system through the nurse call cable (PN:

8000-21-10361). Alarms from the monitor are sent to the nurse station through the nurse call system, if configured to do so.

Satellite module rack connector (SMR1, 2, 3): connects the SMR and N1/T1 docking station.

(6)

(7)

(8)

(9)

(10)

(11)

(12)

(13)

When using the monitor together with other devices, connect their equipotential grounding terminals together to eliminate the potential difference between them.

Video output connector (VP1): connects the secondary display.

Network Connector (LAN1)

It is a standard RJ45 connector which connects the monitor to the central monitoring system (CMS) or other network devices.

Network Connector (LAN 2)

Reserved for future use.

Serial bus connectors (MSB1, 2, 3): connect approved USB devices, for example the keyboard, mouse, and barcode reader. If an independent secondary display is connected, the MSB1 connector is connected to the SBH connector at the rear of the secondary display to activate the MSB connector connecting the keyboard and mouse for the independent secondary display.

Network Connector (LAN3)

It is a standard RJ45 connector which connects the iView system to the external network.

Video output connector (VP2): connects the iView display.

USB connectors: available only when iView module is configured. They connect approved USB devices to the iView, for example keyboard and mouse.

UDI label

CAUTION

Main unit installation and troubleshooting should be executed by Mindray service personnel or authorized technicians.

NOTE

The UDI label on the main unit should be consistent with that on the display.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 2 - 3

2.3.1.2

N22, N19 Main Unit for Separate Installation

If the main unit and the primary display are separately installed, besides the connectors described in 2.3.1.1 N22,

N19 Main Unit for Integrated Installation , the main unit also provides connectors at the rear.

(1) (2) (3)

(4)

2.3.2

(1) Video output connector (VP3): connects the VP connector on the separated primary display.

(2) Serial bus connector (MSB4): connects the serial bus hub connector (SBH) on the separated primary display.

(3) Signal input/output connector (SIG): connects the SIG1 connector on the separated primary display.

(4) DC-output connector: connects the DC-in connector on the separated primary display.

N22, N19 Displays

The displays are used to display system information, alarm messages, parameter numerics, waveforms, and so on. The displays integrate audible and visual alarms and provide USB connectivity. There are two display sizes available: the 22-inch display and the 19-inch display. The monitor supports a primary display, a secondary display, and a iView display.

2 - 4 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

2.3.2.1

N22, N19 Integrated Primary Display

The primary display can be attached to or separately from the main unit, either horizontally or vertically.

The following picture shows the indicators and connectors on the primary display when it is installed with the main unit.

(1)

(5)

(4)

(2)(3)

(1)

(6)

(7)

(5)

(6)

(7)

(3)

(2)

(4)

(5)

(1) Alarm lamp:

When a physiological alarm or technical alarm occurs, this lamp lights and flashes corresponding with the alarm priority:

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 2 - 5

High priority alarms: the lamp quickly flashes red.

Medium priority alarms: the lamp slowly flashes yellow.

Low priority alarms: the lamp lights in cyan without flashing.

(2) Battery indicator:

Yellow: the battery is being charged.

Green: the battery is fully charged.

Flashing green: the monitor operates on battery power.

Off: no battery is installed, or the battery is malfunctioning, or the monitor is powered off and no AC power is connected.

(3) AC power indicator

On: when AC power is connected.

Off: when AC power is not connected.

(4) Power switch

Pressing this switch turns on the monitor.

When the monitor is on, pressing and holding this switch turns off the monitor.

(5) Serial bus connectors (MSB): connect approved USB devices, for example keyboard, mouse, and barcode reader.

Note that every connector may not be visible if the handle (7) is installed.

(6) Navigation knob

(7) Handle

2.3.2.2

N22, N19 Separated Primary Display and Secondary Display

Besides the indicators and connectors, which are the same as those of the integrated primary display, the separated primary display and the secondary display have connectors at the rear.

(6)

(1) (2) (3) (4) (5)

2 - 6

(1) Serial bus connector (MSB1, MSB2): connect approved USB devices, for example keyboard, mouse, and barcode reader.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

(2)

(3)

(4)

(5)

Signal input/output connector (SIG1): for the separate primary display, it connects the SIG connector at the rear of the main unit; for the secondary display, it is reserved for future use.

Serial bus hub connector (SBH): connects the MSB1 connector at the bottom of the main unit to activate MSB connectors for the independent secondary display.

DC-in connector: connects the DC adapter to run the secondary display.

Video connector (VP): connects only to the VP3 connector on the back of the main unit when used as a separate primary display.

UDI label (6)

NOTE

The UDI label on the main unit must be consistent with that on the display

VP of the primary display must connect to VP3 on the main unit to have a supported configuration.

2.3.3

N17, N15, N12 Main Unit

The main unit processes and displays data from modules, and optionally, among networked devices.

2.3.3.1

N17, N15, N12 Front View

(1)

(2)

(3)

(1)

(2)

(5) (4)

Alarm lamp:

When a physiological alarm or technical alarm occurs, this lamp lights and flashes corresponding with the alarm priority:

◆ High priority alarms: the lamp quickly flashes red.

◆ Medium priority alarms: the lamp slowly flashes yellow.

◆ Low priority alarms: the lamp lights in cyan without flashing.

Display

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 2 - 7

(3) AC power indicator

◆ On: when AC power is connected.

◆ Off: when AC power is not connected.

◆ Yellow: the battery is being charged.

◆ Green: the battery is fully charged.

Flashing green: the monitor operates on battery power.

◆ Off: no battery is installed, or the monitor is powered off and no AC power is connected.

Pressing this switch turns on the monitor.

◆ When the monitor is on, pressing and holding this switch turns off the monitor.

2.3.3.2

N17, N15, N12 Left View

(1)

(1)

(2)

(2)

(2)

(2)

N17/N15

(1) Handle

(2) Module racks

N12

2 - 8 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

2.3.3.3 N17, N15, N12 Right View

(1)

(2)

(1) Handle

(2) Recorder

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 2 - 9

2.3.3.4 N15, N12 Rear View

(1)

(1)

(2)

(3)

(4)

(5)

(6)

(7)

(1)

(2)

(3)

(4)

(5)

(6)

USB connectors

Connect approved USB devices, for example the keyboard, mouse, and barcode reader.

Equipotential Grounding Terminal

When using the monitor together with other devices, connect their equipotential grounding terminals together to eliminate the potential difference between them.

AC Power input

Satellite module rack connector

For N15, connects the SMR and N1/ T1 docking station.

For N12, connects the N1/ T1 docking station.

Nurse call connector

It is a BNC connector. It connects the monitor to the hospital’s nurse call system through the nurse call cable (PN:

8000-21-10361). Alarms from the monitor are sent to the nurse station through the nurse call system, if configured to do so.

Digital video connector: connects external display.

It is a standard RJ45 connector which connects the monitor to the central monitoring system (CMS) or other network devices.

2 - 10 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

2.3.3.5

N17 Rear View

(1) (3) (4)

(5)

(9)

(10)

(6)

(2)

(3)

(4)

(5)

(6)

(1)

(2) (7) (8)

Equipotential Grounding Terminal

When using the monitor together with other devices, connect their equipotential grounding terminals together to eliminate the potential difference between them.

AC Power input

USB connectors

Connect approved USB devices, for example the keyboard, mouse, and barcode reader.

USB connectors: available only when iView module is configured. They connect approved USB devices to the iView system, for example keyboard and mouse.

Digital video connector: connects a mirrored external display.

Digital video connector:

Connects the iView display if the iView module is configured.

Connects an independent external display if the N17 does not have an iView.

(8)

(9)

(10)

It is a standard RJ45 connector which connects the monitor to the central monitoring system (CMS) or other network devices.

Network Connector: available only when iView module is configured. It is a standard RJ45 connector which connects the iView system to an external network.

Satellite module rack (SMR) connector: connects the SMR or N1/T1 docking station.

Nurse call connector

It is a BNC connector. It connects the monitor to the hospital’s nurse call system through the nurse call cable (PN:

8000-21-10361). Alarms from the monitor are sent to the nurse station through the nurse call system, if configured to do so.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 2 - 11

2.3.4

Satellite Module Rack (SMR)

The SMR provides an interface between the monitor and external modules. The SMR has eight module slots. It is connected to the main unit through the SMR connector.

The following pictures show the indicator and connectors on the SMR.

(1)

(2) (3)

(1) SMR status indicator: illuminates when the SMR is powered on.

(2) Monitor connector: connects to the Passport 17m/12m monitor.

(3) Monitor connector: connects to the BeneVision monitor.

2.3.5

External Modules

The external modules are used to monitor the patient’s physiological parameters, record patient information and data, and connect external devices. The monitor provides the following modules:

Parameter modules: acquires and processes the patient’s data and sends the data to the main unit.

Recorder module: prints patient information, parameter measurements and waveforms.

BeneLink module: connects external devices. The monitor outputs data from external devices through the

BeneLink module.

2.3.5.1

Available Modules

Refer to 36.16 External Modules for available modules.

The monitor can simultaneously use maximum of three IBP modules and three Temp modules (besides the

Temp and IBP of the MPM module), and two rSO

2 modules. The other modules can only be used one at a time.

Otherwise, the monitor will issue a module conflict prompt. Using MPM module, T1, or N1 at the same time will also result in module conflict.

For example, if a CO

2 module is already loaded and then another CO

2

module is inserted, the monitor will then prompt module conflict. To solve the problem of module conflict, just remove a module.

2 - 12 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

2.3.5.2

Example Module

The parameter modules have similar structure:

The parameter label is marked at the upper left corner.

Hard keys are located on the upper part.

Patient cable connectors are located at the lower part.

Take the MPM module as an example:

(1) (2) (3)

(5)

(9)

(4)

(5)

(6)

(7)

(8)

(1)

(2)

(3)

(4)

(5)

(6)

(7)

(8)

(9)

NIBP Start/Stop key: starts/stops NIBP measurement.

Zero hard key: enters the Zero IBP dialog.

Analog out connector: outputs defibrillation synchronization pulse, ECG, and IBP analog signal.

Module status indicator

On: the module works properly.

Flashing: the module is initializing.

Off: the module is not connected or the module fails.

Connector for IBP cable

Connector for Temp probe

Connector for NIBP cuff

Connector for SpO

2

cable

Connector for ECG cable

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 2 - 13

2.3.5.3

BeneVision N1

The BeneVision N1 is used as an MPM module when being connected to the monitor through the module rack

(for N17, N15, N12), SMR or through the Dock.

(1) (2) (3)

(4)

(3)

(9)

(8)

(5)

(6)

(7)

(1) Power switch

(2) Sample line connector for sidestream

CO

2

(3) Temperature probe connector

(4) Gas outlet

(5) ECG cable connector

(6) SpO

2

sensor connector

(7) NIBP cuff connector

(8) Multifunctional connector: outputs analog ECG, IBP and defib synchronization signals.

(9) IBP cable connector

2.3.5.4

T1

The T1 is used as an MPM module when being connected to the monitor through the module rack (for N17, N15,

N12), SMR or through the T1 docking station.

(1) (1)

(6)

(5)

(2)

(3)

(4)

(1)

(2)

(3)

(4)

(5)

(6)

Temperature probe connector

NIBP cuff connector

ECG cable connector

SpO

2

sensor connector

Multifunctional connector: outputs analog

ECG, IBP, and defibrillation synchronization signals.

IBP cable connector

2 - 14 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

2.3.6

Cable Management Kit

The cable management kit is installed at the bottom of the SMR.

(1)

(2)

2.3.7

(1) Handle: can hold the NIBP cuff

(2) Cable hooks: can hold the cables and leadwires

Input Devices

The monitor allows data entry through touchscreen, keyboard, mouse, navigation knob, remote controller, and barcode reader.

Only Mindray specified input devices should be used. For N22 and N19, the primary display and the secondary display can have independent mice and keyboards. When the secondary display is used as an extended display, one mouse and one keyboard can control both the primary display and the secondary display.

For N22 and N19, you can use the navigation knob to operate some features of the monitor, such as scrolling through lists. The navigation knob is installed at the bottom of the display. The primary display and the secondary display can have independent navigation knobs. When the secondary display is configured as an extended display, you can use one navigation knob to control both the primary display and the secondary display.

(1) (2)

(3)

(1)

(2)

(3)

(4)

(5)

(5) (4)

Alarm Reset hard key: resets the alarm system

Main Menu hard key: enters the main menu.

Navigation knob

NIBP Start/Stop hard key: starts an NIBP measurement or stops the current NIBP measurement

Alarm Pause hard key: pauses physiological alarm system

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 2 - 15

2.3.8

Printing Devices

The monitor can use Mindray specified printer and/or recorder to output patient information and data.

N17, N15, N12 can be configured with a built-in recorder. If the built-in recorder is not available, an optional external recorder module is available.

For N22 and N19, to use the recorder, insert the recorder module into the SMR.

The printer can be connected to the monitor through the network to output patient reports.

2 - 16 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

3

Getting Started

3.1

3.2

Equipment Preparation Safety Information

WARNING

Use only installation accessories specified by Mindray.

Connect only approved devices to this equipment. Devices connected to the equipment must meet the requirements of the applicable IEC standards (e.g. IEC 60950 safety standards for information technology equipment and IEC 60601-1 safety standards for medical electrical equipment). The system configuration must meet the requirements of the IEC 60601-1 medical electrical systems standard. Any personnel who connect devices to the equipment’s signal input/output port are responsible for providing evidence that the safety certification of the devices has been performed in accordance to the IEC 60601-1. If you have any questions, please contact Mindray.

The monitor and parameter monitoring accessories are suitable for use within the patient environment. For other equipment and accessories connected to the monitor, consult corresponding manufacturers for the suitability within the patient environment.

If it is not evident from the equipment specifications whether a particular combination with other devices is hazardous, for example, due to summation of leakage currents, please consult the manufacturer or an expert in the field. A determination must be made that the proposed combination will not negatively affect the devices themselves or the patient's safety.

CAUTION

The equipment should be installed by Mindray authorized personnel.

When disposing of the packaging material, be sure to observe the applicable waste control regulations and keep it out of children’s reach.

Before use, verify whether the packages are intact, especially the packages of single use accessories.

In case of any damage, do not apply it to the patient.

Make sure that the equipment operating environment meets the specific requirements. Otherwise unexpected consequences, e.g. damage to the equipment, could result.

Avoid rough handling during transport.

NOTE

This manual is based on the maximum configuration and therefore some contents may not apply to the product. If you have any question, please contact Mindray.

Put the equipment in a location where it can be easily viewed and operated.

Keep this manual in the vicinity of the equipment so that it can be referenced when needed.

Save the packing case and packaging material as they can be used if the equipment must be reshipped.

Many settings in the patient monitor are password protected. It is recommended to change the default passwords and keep the passwords safe. Passwords should only be changed by authorized personnel. Contact the department manager or biomedical engineering department for the passwords used at the facility.

Monitor Installation

The monitor can be installed in various ways as required.

Wall mount

Installed on the medical supply unit

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 3 - 1

3.2.1

3.2.2

■ Installed on the anesthesia machine

Unpacking and Checking

Before unpacking, examine the packaging carefully for signs of damage. If any damage is detected, contact the carrier, distributor, or Mindray.

If the packing case is intact, open the package and remove the equipment and accessories carefully. Check all materials against the packing list and check for any mechanical damage. Contact Mindray in case of any problems.

Environmental Requirements

The operating environment of the equipment must meet the requirements specified in this manual.

The environment where the equipment is used shall be reasonably free from noises, vibration, dust, corrosive, flammable and explosive substances. Moreover, to maintain good ventilation, the equipment shall be at least 2 inches (5cm) away from around the cabinet.

When the equipment is moved from one place to another, condensation may occur as a result of temperature or humidity difference. In this case, never start the system before the condensation evaporates.

CAUTION

Make sure that the equipment operating environment meets the specific requirements. Otherwise unexpected consequences, e.g. damage to the equipment, could result.

3.3

3.3.1

Setting Up the Equipment

Observance of this manual is a prerequisite for proper product performance and correct operation. It ensures patient and operator safety.

Connecting the AC Mains

The monitor is powered by an AC power supply. Before connecting the equipment to the AC mains, check that the voltage and frequency ratings of the power line are the same as those indicated beside the AC power input.

To use the AC power source, follow this procedure:

1.

Connect the female end of the power cord with the AC power input.

2.

Connect the male end of the power cord with a wall AC outlet.

3.

Check that the AC indicator is on.

The AC indicator is off if the AC mains is not connected. When AC mains is connected, the AC indicator is illuminated in green.

WARNING

Operate the equipment on battery power if the integrity of the protective earth conductor or the protective earthing system in the installation is in doubt.

CAUTION

Always use the accompanying power cord delivered with the monitor.

Before connecting the equipment to the AC mains, ensure that the voltage and frequency ratings of the power line are the same as those indicated beside the AC power input.

NOTE

Use the cable retainer to secure the power cord to prevent it from falling off.

3 - 2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

3.3.2

3.3.3

3.3.4

Connecting the Input Devices

Connect the mouse, keyboard, navigation knob, and/or barcode reader to the USB ports if necessary.

Connecting the SMR

To connect the SMR, use the SMR cable (PN: 009-005121-00 or 009-005122-00) to connect the monitor connector on the rear of SMR to the SMR connector on the main unit.

Connecting Modules to the SMR

3.3.5

To connect a module to the SMR, follow this procedure:

1.

With the module properly oriented, align the module insertion guide slot with the SMR insertion guide.

Push the module into the SMR until there is a click.

2.

Push the lock at the bottom of the module inwards to lock the module.

Removing Modules from the SMR

To remove a module from the SMR, follow this procedure:

1.

Pull outwards on the lock at the bottom of the module to release the module.

2.

Lift the latches at the bottom of the module and slide the module out of the SMR. Hold on the module securely to make sure it does not drop when it comes out.

(1)

(2)

(1)

(2)

Latch

Lock

CAUTION

When removing modules, be careful not to drop them. Modules may be heavier than expected, so securely supporting a module with two hands during removal is recommended.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 3 - 3

3.4

Turning on the Monitor

Before beginning measurements, turn on the monitor, perform the following inspections:

1.

Check the monitor, SMR and modules for any mechanical damage. Make sure that all external cables, plugins and accessories are properly connected.

2.

Connect the power cord to the AC power source.

3.

Press the power switch to turn on the monitor. Secondary displays must be turned on separately.

When the monitor is turned on, the alarms are paused for two minutes. Then the alarm system is activated.

CAUTION

Check that visual and auditory alarm signals are presented correctly when the equipment is powered on.

Do not use the monitor for any monitoring procedure on a patient if you suspect it is not working properly, or if it is mechanically damaged. Contact the service personnel or Mindray.

3.5

3.5.1

Operation and Navigation

Everything needed to operate the monitor is on its screen. Almost every element on the screen is interactive.

Screen elements include parameter values, waveforms, quick keys, information fields, alarms fields, dialogs, and menus. Often you can access the same element in different ways. For example, you can access a parameter dialog by selecting the corresponding numeric area or waveform area, through the Menu hard key on the parameter module, or through the Parameters Setup quick key.

Using the Touchscreen

You can use the touchscreen to select a screen element by pressing directly on the monitor’s screen.

3.5.1.1

Tapping or Swiping across the Screen

■ Tapping the screen

To select an item from menus or lists or select a quick key, tap on it with your finger.

To enter a parameter menu, tap the corresponding numeric area or waveform area. For example, select the ECG numeric area or waveform area to enter the ECG menu.

Swiping the screen with a single finger

◆ To scroll through a list or a menu, swipe up or down.

◆ To show or expand the Minitrends screen or the EWS screen, swipe right across the corresponding screen.

To contract or hide the Minitrends screen or the EWS screen, swipe left across the corresponding screen.

Swiping across the screen with two fingers

◆ To switch to another screen, swipe left or right across the screen. For example, on the Normal screen, swipe with two fingers from left to right to switch to the Minitrends screen.

To discharge a patient, swipe from top to bottom.

3.5.1.2

Locking the Touchscreen

To avoid accidental use, you can temporarily disable the touchscreen by holding and pressing the Main Menu quick key and sliding as directed by the arrow. A padlock symbol key if the touchscreen is disabled.

displays at the top of the main menu quick

The touchscreen lock period is configurable. To do so, follow this procedure:

1.

Access Display in either of the following ways:

◆ Select the Screen Setup quick key → select the Display tab.

◆ Select the Main Menu quick key → from the Display column select Display .

2.

Set Screen Lock Duration .

3 - 4 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

The touchscreen is enabled when the preset time is reached. To manually enable the touchsceen, hold and press the Main Menu quick key and slide as directed by the arrow.

CAUTION

Check that the touchscreen is not damaged.broken, or loose. If there is any sign of damage, stop using the monitor and contact the service personnel.

3.5.2

3.5.3

3.5.4

Using the On-Screen Keyboard

The on-screen keyboard enables information entry:

Enter the information by selecting one character after another.

Select the Backspace key

Select the Caps Lock key

Select the Enter key

to delete single characters or select

to access uppercase letters.

to delete the entire entry.

to confirm the entry and close the on-screen keyboard.

If a conventional keyboard is connected to the monitor, you can use it instead of, or in combination with, the onscreen keyboard.

Using the Navigation Knob (for N22/N19)

You can use the navigation knob and associated buttons to access the main menu, pause alarms, reset alarms, and start/stop NIBP measurements.

Using the Barcode Reader

The monitor supports both a linear (1D) barcode reader and two-dimensional (2D) barcode reader. The barcode reader is connected to the monitor’s MSB connector (for N22/N19) or the USB connector (N17/N15/N12).

NOTE

Use the Mindray custom barcode reader to obtain all the available patient information from both the 2D and 1D barcodes. Using other barcode readers can only output the patient’s medical record number (MRN) and visit number.

3.5.4.1

Clearing Old Data Formats (for the Mindray Custom 2D Barcode Reader)

If you are using the Mindray custom 2D barcode reader (Model HS-1R or HS-1M), before using the it for the first time, clear old data formats and configure the barcode reader. To do so, follow this procedure:

1.

Scan the engineering barcode to clear the previous data format.

2.

Scan the 2D engineering barcode which contains the hospital’s data format.

NOTE

Contact the scanner manufacturer or Mindray to obtain the barcodes for clearing data formats and programming the hospital’s data format.

3.5.5

3.6

Using the Remote Controller

The remote controller can control the monitor by connecting the receiver of the remote controller to the monitor’s MSB connector (for N22/N19) or the USB connector (N17/N15/N12). For more information on how to use the remote controller, see the Instructions for Use delivered with the remote controller.

Screen Display

The following figure shows the normal screen:

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 3 - 5

(1) (2) (3) (4) (5)

(6)

(7)

(8)

3.6.1

(9)

(1) Patient information area: displays patient information, including patient category, gender, department, room number, bed number, and so on. The displayed patient information is configurable. Selecting this area enters the

Patient Management dialog. For more information, see

5.2.1 Entering the Patient Management Dialog .

(2) The current configuration: displays the name of the configuration that can be loaded from the main menu.

(3) Technical alarm information area: displays prompt messages at the top; displays technical alarm messages at the bottom.

(4) Physiological alarm information area: displays high priority physiological alarms at the top; displays medium and low priority physiological alarms at the bottom.

(5) System status information area: displays alarm symbol, battery status, network status, currently connected CMS,

storage device status, and system time. For more information, see 3.6.1 On-screen Symbols .

(6) Parameter waveform area: displays parameter waveforms, measurement ranges, and magnification. Selecting a waveform enters the corresponding parameter dialog.

(7) Parameter numerics area: displays parameter values, alarm limits, and alarm status. This area can also display the parameter list. For more information, see

3.9.3 Displaying the Parameter List

. Selecting a parameter numeric block enters the corresponding parameter dialog. Selecting the parameter list enters Tabular Trends review.

(8) Parameter waveform/numerics area: unique area of the screen that can be configured to display either parameter numerics or parameter waveforms. Depending on the configuration, the tiles take on the characteristics of either area (6) or area (7).

(9) Quick key area: displays quick keys.

On-screen Symbols

The following table lists the on-screen symbols displayed on the system status information area:

Symbol Description

Adult, male

Symbol Description

Adult, female

Pediatric, male Pediatric, female

3 - 6 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Symbol Description

Neonate, male

Wireless network is connected. The solid part indicates network signal strength.

Wired network is connected.

Symbol Description

Neonate, female

Wireless network is not connected.

Wired network is not connected.

All the alarms are paused.

Audible alarm tones are paused.

The alarm system is reset.

The battery has low power and needs to be charged.

The battery is being charged.

Individual physiological alarms are turned off or the monitor is in the alarm off status.

Audible alarm tones are turned off.

The battery is working correctly. The green portion represents the remaining charge.

The battery has critically low charge and needs to be charged immediately. Otherwise, the monitor will automatically shut down soon.

No battery is installed.

3.6.2

Dialogs

All dialogs have similar style and structure, see the figure below:

(1)

(2)

(2)

(4)

(5)

(6)

(3)

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 3 - 7

(4) Exit button: closes the current dialog.

(5) Main body area: includes dialog items and options.

(6) Switch:

Green: the switch is on.

Gray: the switch is off.

3.6.3

Quick Keys

The monitor provides quick keys for you to quickly access some functions. The quick key area is located at the bottom of the screen. Normally the quick key area displays 14 quick keys (12 quicks keys for N12). The Main

Menu key is permanently located the right, and the More key is permanently located at the left. Selecting the

More quick key shows more quick keys. The quick keys displayed on the screen are configurable.

3.6.3.1

Available Quick Keys

The following table shows available quick keys.

Symbol Label

Main Menu

Function

Enters the main menu.

Symbol Label

Alarm Reset

Function

Resets the alarm system.

Alarm Pause

Screen Setup

Pauses the physiological alarm system (Availability depends on configuration of Pause in Maintenance

→ Alarm → Pause/Reset dialog).

Enters the Screen Setup dialog.

Manual Event Manually saves an event into the Events list.

NIBP Start/

Stop

NIBP STAT

Alarm Setup

Starts an NIBP measurement or stops the current NIBP measurement.

Starts a five-minute continuous NIBP measurement.

Enters the Alarm dialog.

Zero IBP Enters the Zero IBP dialog.

Audio Pause Pauses alarm tone

(Availability depends on configuration of Pause in

Maintenance → Alarm →

Pause/Reset dialog).

Standby Enters the standby mode.

Print Starts printing a real-time report.

NIBP Measure Opens the NIBP Measure interval options.

NIBP Stop All Stops all NIBP measurements.

More Shows more quick keys.

C.O. Measure Opens the C.O.

Measure screen.

3 - 8 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Symbol Label

PAWP

Function

Opens the PAWP screen.

Symbol Label

Loops

Venipuncture Inflates the NIBP cuff to help venous puncture.

Parameters

Setup

Enters the Parameters

Setup dialog.

Review

OxyCRG

Enters the Review dialog to review historical data.

Opens the OxyCRG screen.

Function

Opens the Loops screen.

Start TOF Starts/stops TOF measurement.

Remote View Opens the Remote View screen to view other patients.

Minitrends Opens the Minitrends screen.

ECG Full-

Screen

Opens the ECG full screen.

Night Mode Enters the night mode.

Privacy Mode Enters the privacy mode.

CPB Mode Enters the CPB mode.

Volume Enters dialog.

Intubation

Mode

Enters the intubation mode.

Freeze Freezes waveforms.

Calculations Enters the Calculations dialog.

Patient

Management

Enters the Patient

Management dialog.

Integrated

Devices

Opens the Integrated

Devices screen

ECG Lead/Gain Enters the ECG Lead/Gain dialog.

HemoSight Enters the HemoSight dialog.

GCS Enters the GCS dialog.

Load Config Enters the Load Config dialog.

Record Starts/Stops a recording.

Rotate Screen

(N22 and N19 only)

Changes the screen orientation.

Call Help Sends the Call Help signal over the network.

EWS Opens the EWS screen.

SepsisSight Enters the SepsisSight dialog.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 3 - 9

Symbol Label

Discharge

Patient

ECG 24h Sum

Function

Enters the Discharge

Patient dialog to discharge the patient and restore pre-configured default settings.

Views the 24-hour ECG summary.

Symbol Label

Discharged

Patients

End Case

Report

Function

Enters the Discharged

Patients dialog and displays previously discharged patients.

Prints the selected end case reports

Bedside

Devices

Enters the Bedside

Devices dialog.

Targeted Goal Opens the Targeted Goal screen.

3.6.3.2

Configuring the Displayed Quick Keys

To select the quick keys for display, follow this procedure:

1.

Access Quick Keys in either of the following ways:

Select the Screen Setup quick key

→ the Select Quick Keys tab.

Select the Main Menu quick key

→ from the Display column select Quick Keys .

2.

Select the Current tab to configure the quick keys for display on the screen: From the top of this page, select a block location to show a certain quick key, and then select the quick key from the quick key list. For example, to show the Screen Setup quick key at the first block, select the first block, and then select

Screen Setup from the list.

3.

Select the More tab to configure the quick keys for display when the More quick key is selected.

3.7

Operating Modes

The monitor provides different operating modes. This section describes the different monitoring modes and standby modes.

3.7.1

Monitoring Mode

The monitoring mode is the most frequently used clinical mode for patient monitoring. When the monitor is turned on, it automatically enters the monitoring mode.

3.7.2

Privacy Mode

The privacy mode is a special clinical monitoring mode. In the privacy mode, the monitor does not display patient information and monitoring data. This provides controlled access to patient data and ensures confidentiality.

The privacy mode is only available when the patient admitted by the monitor is also monitored by the CMS. The monitor continues monitoring the patient, but patient data is only visible at the CMS.

3.7.2.1

Entering the Privacy Mode

To enter the privacy mode, choose either of the following ways:

Select the Privacy Mode quick key

select Ok .

Select the Main Menu quick key

from the Display column select Privacy Mode

select Ok .

The monitor has the following features after entering the privacy mode:

■ The screen turns blank.

3 - 10 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Except for the low battery alarm, the monitor inactivates alarm tones and alarm lights for all other alarms.

The monitor suppresses all system sounds, including heart beat tone, pulse tone, and prompt tone.

WARNING

. •

In the privacy mode, all audible alarms are suppressed and the alarm light is deactivated at the monitor. Alarms are presented only at the CMS. Pay attention to potential risk.

NOTE

The privacy mode is not available if the Department is set to OR.

You cannot enter the privacy mode if a low battery alarm is active.

3.7.2.2

Exiting the Privacy Mode

The monitor automatically exit the privacy mode in any of the following situations:

The monitor disconnects from the CMS.

The low battery alarm occurs.

You can also operate the touchscreen, mouse, or keyboard to manually exit the privacy mode.

3.7.3

Night Mode

The night mode is a special clinical monitoring mode. The night mode reduces screen brightness, sound volumes, and pauses NIBP measurements. To avoid disturbing the patient, you can use the night mode.

3.7.3.1

Entering the Night Mode

To enter the night mode, follow this procedure:

1.

Select the Night Mode quick key, or select the Main Menu quick key → from the Display column select

Night Mode .

2.

Change the night mode settings if necessary.

3.

Select Enter Night Mode .

The night mode settings are as follows by default:

Brightness

Alarm Volume

QRS Volume

Key Volume

NIBP End Tone

Stop NIBP :

: 1

:

:

0

1

Off

:

:

2

Off

CAUTION

Verify the night mode settings before entering the night mode. Pay attention to the potential risk if the setting value is low.

3.7.3.2

Exiting the Night Mode

To cancel the night mode, follow this procedure:

1.

Select the Night Mode quick key, or select the Main Menu quick key

from the Display column select

Exit Night Mode .

2.

Select Ok .

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 3 - 11

NOTE

If the monitor is connected to the CMS, it automatically exits the night mode if disconnected from the CMS.

The monitor resumes the previous settings after exiting the night mode.

3.7.4

Standby Mode

You can temperately stops patient monitoring without switching off the monitor by entering the standby mode.

3.7.4.1

Entering the Standby Mode

1.

Select the Standby quick key, or select the Main Menu quick key → from the Patient Management column select Standby .

2.

Define where the patient is by selecting a location in the drop down list when the monitor enters the standby mode.

3.

Select Ok .

The monitor behaves as follows after entering the standby mode:

Stops all parameter measurements.

Disables all the alarms and prompt messages, except for the battery low alarm.

Turns screen brightness to the dimmest after entering the standby mode for 30 seconds.

WARNING

Pay attention to the potential risk of placing the monitor to standby. In the standby mode, the monitor stops all parameter measurements and disable all the alarm indications, except for the battery low alarm.

3.7.4.2

Changing the Patient Location at Standby

To change the patient’s location, select patient location from the Standby screen.

3.7.4.3

Exiting the Standby Mode

To exit the standby mode, choose any of the following ways:

Select

Select

Resume Monitor to exit the standby mode and resume monitoring the current patient.

Discharge Patient to discharge the current patient.

If the monitor automatically enters the standby mode after a patient is discharged, choose any of the following ways to exit the standby mode:

Select

Select

Monitor to exit the standby mode and admit a new patient.

Patient Management to enter the patient information for preparing to admit a new patient.

When the monitor exits Standby mode and resumes monitoring, the alarms are paused for two minutes, after which the alarm system is activated.

3.7.5

3.8

Discharge Mode

The discharge mode is a special standby mode after a patient is discharged.

Configuring the Monitor

Configure the monitor before putting it in use.

3 - 12 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

3.8.1

3.8.2

3.8.3

3.8.4

Setting the Screen Orientation (for N22 and N19)

Both the primary display and the secondary display can be installed vertically or horizontally. Set the screen orientation accordingly. To do so, follow this procedure:

1.

Access Display in either of the following ways:

◆ Select the Screen Setup quick key → select the Display tab.

◆ Select the Main Menu quick key → from the Display column select Display .

2.

Select Screen Orientation to set the screen orientation of the corresponding display.

◆ Portrait : if the display is vertically installed, set Screen Orientation to Portrait .

◆ Landscape : if the display is horizontally installed, set Screen Orientation to Landscape .

You can select the Rotate Screen quick key to quickly switch the screen orientation.

Setting the Date and Time

To set the system time, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Time .

2.

Set Date and Time .

3.

Set Date Format .

4.

If you want to use the 12-hour mode, switch off 24 Hour Time .

5.

If you want to use daylight saving time, switch on Daylight Saving Time . You can manually switch on or off the daylight saving time only when the auto daylight saving time function is disabled. For more

information, see 32.8.2 Enabling Auto Daylight Savings Time for details.

If the monitor is connected to a central monitoring system (CMS) or hospital clinical system (HIS), the date and time are automatically taken from the CMS. In this case, you cannot change the date and time on the monitor.

CAUTION

Changing the date and time affects the storage of trends and events and may result in loss of data.

Adjusting the Screen Brightness

To adjust the screen brightness, follow this procedure:

1.

Access Display in either of the following ways:

◆ Select the Screen Setup quick key → select the Display tab.

◆ Select the Main Menu quick key → from the Display column select Display .

2.

Set Brightness . If Brightness is set to Auto , the monitor automatically adjust the screen brightness according to the ambient light.

NOTE

The brightness of the primary screen and secondary screen can be adjusted separately.

Adjusting the Volume

■ Set Alarm Volume in either of the following ways:

◆ Select the Volume quick key.

◆ Select the Main Menu quick key → from the Alarm column select Setup .

Set Reminder Volume by selecting the Main Menu quick key → from the Alarm column select Setup .

■ Set QRS Volume in any of the following ways:

◆ Select the Volume quick key.

◆ From the ECG dialog select Setup .

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 3 - 13

3.9

3.9.1

◆ From the SpO

2

dialog select PR Setup .

Select the Volume quick key to set Key Volume .

General Operation

This section describes the operations that are generally used when monitoring a patient.

Switching On or Off a Parameter

You can manually switch on or off a parameter when its module is connected. If setting parameter switches is not password protected, follow this procedure to set parameter switches:

1.

Access Parameters On/Off in either of the following ways:

◆ Select the Screen Setup quick key

→ select the Parameters On/Off tab.

◆ Select the Main Menu quick key

from the Parameters column select Parameters On/Off .

2.

Switch on or off desired parameters.

When a parameter is switched off, the monitor stops data acquisition and alarming for this measurement.

NOTE

When a parameter is manually switched off, you cannot monitor this parameter even if corresponding parameter module is plugged in and related accessories of this parameter are connected.

3.9.2

3.9.3

3.9.4

Displaying Parameter Numerics and Waveforms

You can configure the parameter numerics, waveforms, and their sequence displayed on the normal screen. To do so, follow this procedure:

1.

Access Tile Layout in either of the following ways:

◆ Select the Screen Setup quick key

select the Tile Layout tab.

◆ Select the Main Menu quick key

→ from the Display column select Tile Layout .

2.

Select a parameter numeric area or waveform area, and then from the popup list select an element you want to display in this area. The parameters and waveforms you did not select will not displayed.

If you are using the independent secondary display, you can select parameter and waveform locations on the secondary display.

Displaying the Parameter List

You can display the trends of HR, SpO

2

, RR, and NIBP/IBP in the parameter numerics area. To do so, follow this procedure:

1.

Access Tile Layout in either of the following ways:

◆ Select the Screen Setup quick key → select the Tile Layout tab.

◆ Select the Main Menu quick key → from the Display column select Tile Layout .

2.

Select the parameter numerics area where you want to display the parameter list, and then from the popup list select Parameter List .

Accessing Parameter Setup Dialogs

Each parameter has a setup dialog in which you can adjust the alarm and parameter settings. You can enter a parameter setup dialog by using any of the following methods:

Select the parameter numeric area or waveform area.

Press the setup hard key

Select the Parameters Setup

on the module front.

quick key , and then select the desired parameter .

3 - 14 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

■ Select the Main Menu quick key → from the Parameters column select Setup → select the desired parameter .

NOTE

In this manual, we always use the first method to enter the setup dialog. But you can use any method you prefer.

3.9.5

Changing Measurement Colors

You can set the color of measurement values and waveforms for each parameter. To do so, follow this procedure:

1.

Select Main Menu quick key

→ from the Parameters column select Parameter Color .

2.

Select the Current tab and set the colors of the currently monitoring measurement values and waveforms.

3.

Select the All tab and set the colors of measurement values and waveforms for all parameters.

3.10

Initiating a Manual Event

To save a manual event, follow this procedure:

1.

Select the Manual Event quick key to enter the Manual Event menu.

2.

Select a name for this event, for example Intubated , or input a name.

3.

Select OK .

To edit the name of preset event names, select to enter the Manual Event Setup menu.

Selecting or editing manual event name functionality is available only if the Manual Event Edit switch is turned

on. For more information, see 32.16.3 Setting the Manual Event Edit Switch .

You can review the manual events. For more information, see 25.2.7 Events Review Page .

3.11

Freezing Waveforms

During patient monitoring, the freeze feature allows you to freeze the currently displayed waveforms on the screen so that you can have a close examination of the patient’s status. Additionally, you can select any frozen waveform for recording.

3.11.1

Freezing Waveforms

To freeze waveforms, select the Freeze quick key. Except waveforms of the following screens, all displayed waveforms stop refreshing and scrolling after you select the Freeze quick key:

Minitrends

OxyCRG

EWS screen screen screen

Remote View screen

3.11.2

Viewing Frozen Waveforms

To view the frozen waveforms, follow this procedure:

Select the or button in the Freeze screen.

Slide the frozen waveform leftward or rightward.

At the lower right corner of the bottom-most waveform displays the freeze time. The initial frozen time is 0 s .

With the waveforms scrolling, the freeze time changes at an interval of 1 second. For example, -2 s means two seconds before the frozen time. This change will be applied for all waveforms on the screen.

NOTE

You can view the frozen waveforms of up to 120 seconds.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 3 - 15

3.11.3

Unfreezing Waveforms

To unfreeze the frozen waveforms, select the

Numerics.

button at the upper right corner of the Freeze Big

3.11.4

Printing Frozen Waveforms

To print the frozen waveforms, select the button at the upper left corner of the Freeze Big Numerics.

3.12

N22 and N19 Secondary Display

The secondary display of N22/N19 has the following features:

Be able to connect independent input devices.

Provides visual and audible alarm indications.

The secondary display runs on independent AC mains through a DC adapter. To use the secondary display, turn it on and verify that it is connected before turning on the monitor. The secondary display does not support hot swapping. If the secondary display is disconnected from the main unit, the primary display will announce an alarm.

3.12.1

Connecting the Secondary Display Power Supply (for N22/N19)

You need a power adapter to convert the AC mains to DC so as to power the secondary display. Before connecting the power adapter, check that the power adapter meets the specification.

To connect the power supply, follow this procedure:

1.

Connect one end of the power adapter to the DC-IN connector on the secondary display.

2.

Connect the other end of the power adapter to the AC mains.

3.

Check that the AC indicator on the secondary display is on.

CAUTION

Use only Mindray-specified power adapters for powering Mindray displays.

3.12.1.1

Setting Secondary Display Orientation (for N22/N19)

If the secondary display is an independent display or an extended display, you can set its orientation if necessary.

See

3.8.1 Setting the Screen Orientation (for N22 and N19) for details.

NOTE

If the secondary display is a mirror display, its orientation is the same with the primary display and cannot be set separately.

3.12.1.2

Setting Secondary Display Brightness (for N22/N19)

See

3.8.3 Adjusting the Screen Brightness for details.

3.12.2

Setting the ECG Waveform Size for the Independent External Display (for N17/

N15/N12)

For N17/N15/N12 monitors, if the independent external display is connected, you can set the speed and amplitude scales of the ECG waveforms for displays of different dimensions to achieve the best display effect. To do so, follow this procedure:

1.

From the independent external display screen, select the Main Menu quick key

from the System column select Maintenance → input the required password → select .

2.

Select Display → select the Screen Size tab.

3.

Select the screen size.

3 - 16 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

4.

Adjust the speed and amplitude of the ECG waveform by setting the scale so that 1 cm on the scale is actually equal to one centimeter.

5.

Restart the monitor.

(2)

(1)

(1) Adjust the amplitude of the ECG waveform

(2) Adjust the speed of the ECG waveform

NOTE

The setting of Screen Size for the independent external display takes effect only after the monitor restarts.

3.13

Using the iView System

The iView system provides a means of running clinical applications on a monitor for obtaining other patient data. The application data from the iView can show on the monitor’s display or on the iView display.

For more information on iView, see iView System Operator’s Manual (PN: 046-012637-00).

WARNING

Some clinical applications may show data from another patient. Be aware that some of the data on the patient monitor display may not always be from a single patient.

Applications running on the iView cannot act as a primary alarming device and cannot be relied upon for alarm notification. There may be no audible or visible indications apart from what is shown on the screen, and any data shown may be delayed.

CAUTION

Always use AC mains to run the monitor if iView is in use.

Ensure that any software installed on the iView complies with all relevant local regulations.

• iView is an independent PC platform that enables running third-party applications. The monitor runs independently of the iView, and vice versa.It is the user's responsibility to maintain iView security patches in compliance with hospital PC policy. The user’s IT administrator should determine what security patches and policies needed after third-party applications are installed.

Contact Microsoft service personnel if iView Windows problems occur.

3.14

Capturing the Screen

The monitor provides the function of screen capture. To capture the current screen display, follow this procedure:

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 3 - 17

1.

Connect the USB drive to the monitor’s MSB connector (for N22/N19) or the USB connector (N17/N15/N12).

2.

Press and hold the More quick key. Wait until it turns from blue to grey.

The captured pictures are automatically saved in the USB drive.

The screen capture function is disabled by default. If you need this function, contact the service personnel.

3.15

Connecting the CMS

You can connect the monitor to the BeneVision CMS through wired LAN or wireless LAN. When connected to the

CMS, the system provides the following function.

The monitor can transmit parameter values, waveforms, alarms, and events to the CMS. From the CMS, you can check the patient’s monitoring data and alarms.

The monitor can transmit parameter values and alarms from the connected external devices to the CMS.

From the CMS you can check the patient’s monitoring data and alarms obtained from the connected external devices.

Patient information, alarm settings, and alarm status can be synchronized between the monitor and the

CMS.

You can start or stop NIBP measurements from the CMS.

In case of network disconnection, the monitor can transmit the offline data to the CMS when the network is reconnected.

For more information on the CMS, see the operator’s manual of corresponding central monitoring system.

To select a CMS, select the system status information area at the top right corner of the main screen. Select the desired CMS from the popup CMS list.

NOTE

You can select CMS only when the Select CMS switch is on. For more information, see

32.13.6 Enabling Selecting a CMS

.

3.16

Connecting the eGateway

You can connect the monitor to the eGateway through wired LAN or wireless LAN to implement interaction between the monitor and external devices. When connected to the eGateway, the system provides the following functions:

The monitor can transmit parameter values, waveforms, alarm settings, and events to the eGateway.

The monitor can transmit parameter values and alarm settings received from the BeneLink-connected external devices to the eGateway.

Clock can be synchronized between the monitor and the eGateway.

3.17

Connecting the Wireless Network

You can add up to five wireless networks for the monitor. For more information, see

32.13.4 Setting the Wireless

Network

.

If connecting the current wireless network fails, the monitor automatically connects other wireless networks in the order when they were added.

To manually switch the wireless network, from the system status information area on the top right conner of the screen select , and select the desired wireless network.

3.18

Disconnecting the Wireless Network

To disconnect the wireless network manually, follow this procedure:

1.

Swipe the screen from top down with a single finger.

2.

Select .

To reconnect the wireless network after it is disconnected manually, follow this procedure:

3 - 18 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

1.

Swipe the screen from top down with a single finger.

2.

Select .

3.19

Turning Off the Monitor

Before turning off the monitor, perform the following check:

1.

Ensure that monitoring of the patient has been completed.

2.

Disconnect the cables and sensors from the patient.

3.

Save or clear the patient data as required.

To turn off the monitor, press and hold the power switch for 3 seconds.

CAUTION

Press and hold the power switch for 15 seconds (for N22/N19) or 10 seconds (for N17/N15/N12) to forcibly shut down the monitor if it could not be shut down normally. This may cause loss of patient data.

NOTE

Turning off the monitor does not disconnect the monitor from the AC mains. To completely disconnect the power supply, unplug the power cord.

If a monitor is on during a power loss, then the monitor is automatically switched on when the power is restored.

In case of a temporary power failure, if the power is restored within 30 minutes, monitoring will resume with all active settings unchanged. If the monitor is without power for more than 30 minutes, it behaves the same as if it were normally turned off.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 3 - 19

This page intentionally left blank.

3 - 20 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

4

User Screens

4.1

4.2

4.2.1

4.2.2

4.3

4.3.1

The monitor provides different user screens to facilitate patient monitoring in different departments and clinical applications.

Choosing a Screen

The monitor enters the normal screen after it is powered on. The normal screen is most frequently used for patient monitoring. To select other screens, follow this procedure:

1.

Access the Choose Screen tab in either of the following ways:

Select the Screen Setup quick key.

Select the Main Menu quick key

→ from the Display column select Choose screen .

2.

Select the desired screen.

Normal Screen

The normal screen is most frequently used for patient monitoring. For general department, ICU, and CCU, normal screen is used by default.

Entering the Normal Screen

To enter the normal screen, choose any of the following ways:

Swipe left or right on the touchscreen with two fingers to switch to the normal screen.

Select the Screen Setup quick key

select the Choose Screen tab

select Normal Screen .

Select the Main Menu quick key

from the Display column select Choose Screen

select Normal

Screen .

Configuring the Normal Screen

You can configure the parameter numerics, waveforms, and their sequence displayed on the normal screen. To do so, follow this procedure:

1.

Access Tile Layout in either of the following ways:

Select the Screen Setup quick key.

Select the Main Menu quick key

→ from the Display column select Tile Layout .

2.

Select a parameter numeric area or waveform area, and then from the popup list select an element you want to display in this area. The parameters and waveforms you did not select will not displayed.

The Big Numerics Screen

The Big Numerics screen displays parameter numerics in big font size. The Big Numerics screen displays measurement values and waveforms of up to six parameters. You can configure the parameters and their layout on the big numeric screen.

Entering the Big Numerics Screen

To enter the big numerics screen, choose any of the following ways:

Swipe left or right on the touchscreen with two fingers to switch to the big numerics screen.

Select the Screen Setup quick key → select the Choose Screen tab → select Big Numerics .

Select the Main Menu quick key → from the Display column select Choose Screen → select Big

Numerics .

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 4 - 1

4.3.2

4.4

4.4.1

Configuring the Big Numerics Screen

To configure the big numerics screen, follow this procedure:

1.

Access Choose Screen in either of the following ways:

◆ Select the Screen Setup quick key .

Select the Main Menu quick key

→ from the Display column select Choose Screen .

2.

Select the Big Numerics tab

3.

Select a parameter numeric area or waveform area, and then from the popup list select an element to display in this area.

Minitrends Screen

The Minitrends screen shows the recent graphic trends of parameters.

Entering the Minitrends Screen

Choose one of the following methods to enter the Minitrends screen:

Swipe left or right on the touchscreen with two fingers to switch among the Minitrends screen, normal screen, and the big numerics screen.

Select the

Select the

Select the

Minitrends quick key.

Screen Setup

Main Menu quick key

quick key

Select the from the

Choose Screen

Display

tab

column select

For adult and pediatric patients, when the Minitrends screen is hidden as following methods to quickly enter the Minitrends screen.

Choose Screen

, select Minitrends

.

select Minitrends you can also choose one of the

.

Swipe left or right across the touchscreen with two fingers until you switch to the Minitrends screen.

Swipe right across the touchscreen with a single finger.

Select the button .

4 - 2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

4.4.2

The Display of Minitrends Screen

The following figure shows the minitrends screen. Your display may be configured to look slightly different.

(3)

(1)

(4)

4.4.3

4.4.4

4.4.5

(6)

(2)

(1)

(2)

(3)

(4)

(5)

(6)

Scale

Routine Vital button. If the department is set to OR , then the Baseline button is displayed.

Routine Vital/Baseline

Select this button to view the long trends, or contract the long trends screen to the Minitrends screen.

Time line

Alarm statistic area

Viewing the Long Trends

To expand the Minitrends screen to view the long trends, choose either of the following ways:

Select the button.

Swipe right across the Minitrends screen with a finger.

Setting Minitrends Parameters

To set parameters, follow this procedure:

1.

Enter the Minitrends screen.

2.

Select the Setup button.

3.

Set parameters. If you want to use the default parameters, select Default Parameter .

Setting the Minitrend Length

To set the Minitrend length, follow this procedure:

1.

Enter the Minitrends screen.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

(5)

4 - 3

4.4.6

4.4.7

2.

Select the Setup button.

3.

Set the Minitrend Length .

Setting the Alarm Statistics Switch

The Minitrends screen can be configured to display the statistic number of physiological alarms in its lower half screen. To set the alarm statistics switch, follow this procedure:

1.

Enter the Minitrends screen.

2.

Select the Setup button.

3.

Switch on or off the Alarm Statistics switch.

Setting the Alarm Statistics Duration

The time length within which the alarms statistics are made is configurable. To set the alarm statistics length, follow this procedure:

1.

Enter the Minitrends screen.

2.

Select the Setup button.

3.

Set Alarm Statistics Duration .

4.4.8

Routine Vital/Baseline

The Routine vital/Baseline function is used for marking the parameter measurements of certain moment for later reference. If the department is set to OR , then the Baseline button is available. For other departments, the

Routine Vital button is available.

4.4.8.1

Manually Marking the Routine Vital/Baseline

To manually mark the Routine Vital/Baseline, follow this procedure:

1.

Enter the Minitrends screen.

2.

Select the Routine Vital button or Baseline button.

NOTE

If you do not see the Baseline button or Routine Vital button in the Minitrends screen, you can select the Setup button and switch on the Baseline switch, or set the Routine Vital to Manual or Auto.

4.4.8.2

Configuring Automatic Routine Vital Settings

The monitor can automatically mark the routine vital sign values. To enable this function, follow this procedure:

1.

Enter the Minitrends screen.

2.

Select the Setup button.

3.

Select Auto from the dropdown list of Routine Vital .

4.

Select Time to set the time for marking the first routine vital sign values.

5.

Select Interval to set the interval for marking the routine vital sign values.

4.5

4.5.1

The OxyCRG Screen

The monitor displays the OxyCRG screen by default when the neonatology department is selected. The OxyCRG screen is available in any department setting, but only when the patient size is set to Neo . This screen displays 6minute HR, SpO

2

trends, CO

2

/Resp compressed waveform, ABD parameters, and the latest ABD events.

The OxyCRG function is intended for neonatal patents only.

Entering the OxyCRG Screen

To enter the OxyCRG screen, choose any of the following ways:

4 - 4 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

4.5.2

Swipe left or right on the touchscreen with two fingers to switch to the OxyCRG screen.

Select the OxyCRG quick key.

Select the Screen Setup quick key → select the Choose Screen tab → select OxyCRG .

Select the Main Menu quick key → from the Display column select Choose Screen → select OxyCRG .

The Display of the OxyCRG Screen

The following figure shows the OxyCRG screen. Your display may be configured to look slightly different.

(8) (7)

(1)

(6)

(2)

4.5.3

(1)

(2)

(3)

(4)

(5)

(6)

(7)

(8)

(3) (4) (5)

HR, SpO

2

/SpO

2 b trend

Resp/CO

2

compressed waveform

Event time

Event type

Parameter values of ABD events

ABD event list: displays the latest red ABD events. Selecting the ABD event list area enters the OxyCRG review page.

Mark button: opens the Mark dialog to edit the events.

ABD event prompt area: displays parameter values of currently active OxyCRG events.

OxyCRG Events

The following table lists the ABD events and their criteria:

B

D

BD

AB

AD

ABD

Event type Description

A Apnea event: the apnea duration exceeds the threshold.

• A20: the apnea duration is greater or equal to 20 seconds.

• A15: the apnea duration is between 15 to 20 seconds (excluding 20 seconds).

• A10: the apnea duration is between 10 to 15 seconds (excluding 15 seconds).

Low Heart Rate event: the duration of low heart rate exceeds the threshold.

Low SpO2 event: the duration of SpO2 Desat exceeds the threshold.

Low Heart Rate and low SpO2 happen at the same time.

Apnea and Low Heart Rate happens at the same time.

Apnea and low SpO2 happen at the same time.

Apnea, Low Heart Rate, and low SpO2 happen at the same time.

Remarks

A20 is a red event

/

/

/

Red event

Red event

Red event

NOTE

The monitor records all ABD events for OxyCRG review, but only red events displays in the ABD list of the OxyCRG screen.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 4 - 5

4.5.4

4.5.5

4.5.6

4.5.7

4.6

4.6.1

4.6.2

The Display of the ABD Event Area

The ABD event area displays parameter values of currently active OxyCRG events and lists the latest red ABD events.

Setting OxyCRG Parameters

Select parameter trends or compressed waveform to set parameters and the compressed waveform you want to display. The selected parameters will be used for ABD event calculation.

Setting the Threshold of ABD Events

Select any parameter trend or the compressed waveform to perform the following setup:

■ Set the threshold of ABD events.

■ Set Event Storage Format :

◆ 1 min+3 min : stores data one minute before and three minutes after the event.

◆ 3 min+1 min : stores data three minutes before and one minute after the event.

◆ 2 min+2 min : stores data two minutes before and two minutes after the event.

The stored data includes the trends of the OxyCRG parameters, compressed waveform, alarm thresholds, NIBP, and Temp measurements.

Editing ABD Events

To edit ABD events, follow this procedure:

1.

Select the Mark button to enter the Mark dialog box.

2.

Drag the event list upwards and downwards to select the desired event.

3.

Select the patient’s status when the event happened.

4.

Select Save .

The Targeted Goal Screen

If you are concerned with specific parameters and their trends, you can use the Targeted Goal screen. The

Targeted Goal screen focuses on the target parameter and displays parameter measurements in big numerics.

You can easily identify whether parameter target is reached via a dashboard and review the statistics of the target parameter by sections.

The Targeted Goal screen displays parameter measurements and waveforms of ECG, SpO

2

, IBP, PI, PR, CO

2

, Resp,

NIBP, and Temp. You can define the target parameter and secondary parameters. The measurements of these parameters displays in big numerics.

Entering the Targeted Goal Screen

To enter the Targeted Goal screen, choose any of the following ways:

Select the Screen Setup quick key

select the Choose Screen tab

select Targeted Goal screen.

Select the Main Menu quick key

from the Display column select Choose Screen

select Targeted

Goal Screen.

■ If the Patient Category is set to Neo , swipe left or right on the touchscreen with two fingers to switch to the Targeted Goal screen.

The Display of the Targeted Goal Screen

The following figure shows the Targeted Goal screen. Your display may be configured to look slightly different.

4 - 6 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

(5)

(1)

(6)

4.6.3

(2)

(3)

(4)

(7)

(1)

(2)

(3)

(4)

(5)

(6)

(7)

Parameter trends area: displays trends of the target parameter and secondary parameters. If the target parameter is Art, this area only lists the trend of arterial pressure. Selecting this area enters the Tabular Trends review page.

Target parameter statistics area: displays the statistics of the target parameter by sections.

Target parameter trends area: displays the graphic trends of the target parameter. If this area is not configured to display the trends of the target area, other selected parameter is displayed.

Other parameter area: displays parameter measurements and alarm limits of parameters other than the target parameter and secondary parameters.

Target parameter area: displays the measurement of the target parameter in big numerics, as well as its target range, and alarm limits.

• If the target parameter is Resp or PR, parameter source is also displayed.

• The dashboard shows the target range in green.

• The

△ pointer below the dashboard indicates the current measurement value.

• Selecting this area enters the corresponding parameter setup menu.

Secondary parameters area: displays parameter measurement of secondary parameters in big numerics, as well as waveforms and alarm limits. If secondary parameters are Resp and PR, parameter sources are also displayed.

Target parameter waveform area: displays the waveform of the target parameter.

• If the target parameter is Resp or PR, the waveform of the source parameter is displayed.

• If the target parameter is ECG, the fist ECG waveform is displayed by default.

Configuring the Targeted Goal Screen Layout

To configure the parameter numerics, waveforms, and their sequence displayed on the Targeted Goal screen, follow this procedure:

1.

Access the Targeted Goal screen in either of the following ways:

◆ Select the Screen Setup quick key → select the Choose Screen tab → select Targeted Goal screen.

◆ Select the Main Menu quick key → from the Display column select Choose Screen → select

Targeted Goal Screen.

2.

Select a parameter numeric area or waveform area, and then from the popup list select an element to display in this area. The parameters and waveforms not selected will not displayed.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 4 - 7

(6)

(1)

4.6.4

(2)

(5)

(3)

(4)

(1) Select this area to define items to be displayed for the target area:

Statistics: this area displays the statistics of the target parameter by sections.

Statistics & Trend: this area displays the statistics of the target parameter by sections and the area below displays the graphic trends of the target parameter.

(2) If the graphic trends of the target parameter is not displayed, select this area to define other parameter to be displayed.

(3) Select this area to define other parameter (other than the target parameter and the secondary parameter) to be displayed.

(4) Select this area to define the target parameter.

(5) Select this area to define the secondary parameter.

(6) Select this area to define the secondary parameter.

Setting Parameter Statistics

You can show the statistics of the target parameter for a defined period of time. To do so, follow this procedure:

1.

Select from the target parameter statistics area to enter the parameter statistics menu.

2.

Select the range of each section: from the To column select the SpO2 value at which corresponding section ends.

3.

From the Target column select the target section. The target section is highlighted in green in the SpO

2 statistics area.

4.

From the target parameter statistics area, select the duration to redefine the statistics duration.

The following figure shows the target parameter statistics area when SpO

2 is set as the target parameter:

1

2

3

4

(1) Statistics duration

(2) Statistics setup icon

(3) Statistics results

(4) Sections for statistics

4 - 8 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

4.6.5

4.6.6

4.7

Selecting the Range of each Parameter Section and the Target Section

For the target parameter, to define the range of each section, follow this procedure:

1.

Select from the target parameter statistics area.

2.

From the To column select the parameter value at which corresponding section ends.

3.

From the Target column select the target section. The target section is highlighted in green in the statistics area.

Selecting the Statistics Duration of the Target Parameter

The statistics duration of the target parameter is configurable. From the target parameter statistics area, select the duration to redefine its duration.

Remote View

On the monitor, you can observe alarm conditions and view real time physiological data from patients on other compatible remote networked monitoring devices, such as a bedside monitor or a telemetry device.

A device from a remote site is called a remote device or bed. N22/N19/N17 can simultaneously watch up to 18 remote devices. N15/N12 can simultaneously watch up to 12 remote devices. The monitor can show waveforms of one remote device.

You can watch the remote devices in the Remote View screen, or the alarm watch tiles on the main screen.

From the Remote View screen you can watch the following information:

The alarm status and alarm messages of up to 18 remote devices for N22/N19/N17 or 12 remote devices for

N15/N12.

The realtime parameter values and waveforms from the main bed.

If an anesthesia system or ventilator is connected to the main bed, you can watch realtime parameter values and waveforms from the anesthesia system or the ventilator.

If pumps are connected to the main bed, you can view the pump alarm status.

NOTE

A particular monitor (for example Monitor #1) can be viewed by at most 32 remote devices at the same time, of which eight of those remote devices can watch Monitor #1’s waveforms.

4.7.1

4.7.2

Entering the Remote View Screen

To enter the Remote View screen, choose one of the following ways:

Select the Remote View quick key.

Select the bed at the alarm watch tile on the main screen. For more information, see 4.7.10.2 Displaying the

Alarm Watch Tile on the Main Screen .

Select the Screen Setup quick key → select the Choose Screen tab → select Remote View .

About the Remote View

The Remote View screen displays parameter measurements and waveforms of the remote device. If an anesthesia system is connected to the main bed, swipe left or right on the screen to view more information, including parameters and waveforms from the anesthesia system.

The following figure shows the Remote View screen.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 4 - 9

(1)

(2)

(3)

(1)

(2)

(3)

Alarm watch area

Displays all the monitored remote beds. Each bed displays the room number, bed number, connection status and alarm status. The background color indicates the alarm status on the corresponding bed.

Background Color Description

Green

Red

Yellow

Cyan

Grey

Main area

No alarm is occurring at the bed.

The remote device is disconnected or a high priority alarm is currently the highest alarm level on the bed. If the remote device is disconnected, the icon is displayed.

The medium priority alarm is currently the highest alarm level on the bed.

The low priority alarm is currently the highest alarm level on the bed.

The bed is in the standby mode.

Displays the patient’s information, alarm status and messages, waveforms, measurements, etc. of the selected bed.

This bed is called main bed. Scrolling up and down can view more parameters and waveforms. If anesthesia system is connected to the main bed, swipe left and right on the screen to view parameters and waveforms from the anesthesia system.

Remote View setup button: select it to enter the Remote View setup menu.

WARNING

The data presented in the Remote View screen has a delay. Do not rely on this screen for real time data.

4.7.3

Adding a Bed

After adding the desired remote devices, then the alarms from these devices can be viewed on the monitor. To add a remote device, follow this procedure:

1.

Enter the Select Bed dialog. To do so, choose either of the following ways:

◆ In the Remote View screen, select Select Bed . For more information, see

4.7.1 Entering the Remote

View Screen

.

◆ Select the icon at the alarm watch tile if the tile is configured to display on the main screen.

2.

In the Select Bed dialog, select a desired department. All the beds under this department will be listed.

4 - 10 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

4.7.4

4.7.5

4.7.6

4.7.7

4.7.8

4.7.9

3.

Select a desired tile at the A-W1, A-W2 or A-W3 areas and then select a bed from the bed list. The selected bed will appear in the tile.

NOTE

The added bed is indicated by a check mark (√) at the left of the bed list.

Removing a Bed

To remove a remote device from remote monitoring, follow this procedure:

1.

Enter the Select Bed dialog. Choose either of the following ways:

In the Remote View screen, select Select Bed

. For more information, see 4.7.1 Entering the Remote

View Screen

.

◆ Select the icon in the alarm watch tile if the tile is configured to display on the main screen.

2.

In the Select Bed dialog, select a bed at the A-W1, A-W2 or A-W3 areas, and then select Clear Bed . If you want remove all beds, select Clear All Beds .

Displaying the Main Bed

In the Remote View screen, you can select a bed in the alarm watch area, then the main body of the Remote

View screen will display the real time monitoring screen of the device.

Saving a Manual Event

You can initiate a manual event on the remote monitor by selecting Manual Event in the Remote View screen.

Managing Alarms from Remote Devices

You can view and manage the alarms coming from remote devices.

If the monitor and remote devices have alarms at the same time, the monitor presents alarms according to the following rules:

If the alarm tone pattern for the monitor is same as that for the remote devices, the monitor prioritizes all the alarms and sounds the highest alarm tone.

If the alarm tone pattern for the monitor is different from that for the remote devices, the monitor sounds the alarms in their corresponding tone.

The monitor provides the same alarm tone pattern for the remote device alarms as those for the monitor alarms by default. You can change the alarm tone pattern of the remote monitor; see

32.3.5.5 Setting the Tone Pattern for

Alarms from Remote Devices

.

Resetting Alarms for Remote Devices

You can reset the alarms on the remote devices by selecting Alarm Reset in the Remote View screen. This

function needs to be enabled. For more information, see 32.3.5.1 Resetting Alarms for Remote Devices .

Selecting Beds By Care Group

If configured, the monitor automatically selects beds in the same care group during the shift of care groups in the CMS. To enable this function, follow this procedure:

1.

Enter the Select Bed window. Choose either of the following ways:

◆ In the Remote View window, select Select Bed . For more information, see

4.7.1 Entering the Remote

View Screen

for entering the Remote View window.

Select the icon in the alarm watch tile if the tile is configured to display on the main screen.

2.

In lower left corner of the Select Bed window, select Select Beds By Care Group .

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 4 - 11

4.7.10

Alarm Watch

The alarm watch function provides the alarm notification by color and sound.

The monitor sounds the highest priority alarm tone from all the monitored remote devices.

The monitor displays the highest priority alarm in corresponding background color for each bed at the following areas:

◆ At the top of the Remote View . For more information, see

4.7.2 About the Remote View

for details.

◆ On the main screen. For more information, see

4.7.10.1 About Alarm Watch Tile for details.

4.7.10.1

About Alarm Watch Tile

The main screen can display up to three alarm watch tiles, namely A-W1, A-W2 and A-W3. Each tile can accommodate up to six beds.

The following figure shows the alarm watch tiles.

(1) (2)

(5) (4) (3)

(1)

(2)

(3)

(4)

Alarm watch tile label

Disconnection icon: when the remote device is disconnected, this icon displays at the tile, and the tile background color is red.

Bed area (multiple beds): if more than one bed is assigned to an alarm watch tile, each bed cyclically displays the bed number, room number, and the alarm of the highest priority. The background color of each bed indicated the status of this bed.

Bed area (one bed): if only one bed is assigned to an alarm watch area, this area displays the bed number, room number, parameter value, and alarm message from this bed, etc.

Select bed icon: select it to enter the Select Bed dialog.

(5)

The alarm watch tile in the main screen is similar to the alarm watch area in the Remote View.

For more

information, see 4.7.2 About the Remote View .

4.7.10.2

Displaying the Alarm Watch Tile on the Main Screen

To configure the alarm watch tile to be displayed on the monitor’s main screen, follow this procedure:

1.

Select the Main Menu quick key

from the Display column select Choose Screen to enter the Screen

Setup dialog.

2.

Select the Tile Layout tab.

3.

Select the numeric area where you want to display the alarm watch tile, and then in the drop-down list, select Alarm Watch

A-W1, A-W2 , or A-W3 .

4.7.11

Auto Displaying the Alarm Bed

The monitor provides the function of automatically displaying the remote alarm bed. If this function is enabled, when a remote bed issues an alarm, the monitor automatically displays this bed as the main bed on the Remote

View screen.

If multiple remote beds issue alarms, the monitor cyclically displays the alarm beds as per the preset interval and in the order of alarm time.

The auto displaying alarm bed function is disabled by default. To enable this function, follow this procedure:

1.

From the Remote View screen, select to enter the Remote View setup menu.

4 - 12 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

2.

Switch on Rollup Alarm Beds .

3.

Set Rollup Interval :

◆ Off : do not cyclically display the remote alarm beds. Once a new alarm is issued, the monitor automatically switches to the new alarm bed.

◆ 10 sec , 20 sec , or 30 sec : If multiple remote beds issue alarms, the monitor cyclically displays the alarm beds as per the preset interval and alarm priority in the order of alarm time.

4.

Set Alarm Priority:

◆ High Only: Only when a high priority alarm is issued, the monitor automatically switches to the alarm bed.

High & Med: If Rollup Interval is set to Off and when a high priority alarm or medium priority alarm is issued, the monitor automatically switches to the alarm bed. If Rollup Interval is set to 10 sec , 20 sec , or 30 sec and multiple remote beds issue alarms, the monitor cyclically displays the alarm beds with higher priority in the order of alarm time. For example, if both high priority alarms or medium priority alarm are issued, only beds with high priority alarms are cyclically displayed.

5.

Set Switch Bed Prompt Voice. If this function is enabled, the monitor issues a reminder sound each time the main bed switches.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 4 - 13

This page intentionally left blank.

4 - 14 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

5

Managing Patients

5.1

5.2

Starting Monitoring a Patient

After turning on the monitor, follow this procedure to monitor a patient:

1.

Admit the patient.

2.

Check patient settings. Make sure that alarm limits, patient category and paced status, and so on, are appropriate for the patient. Change them if necessary.

3.

Perform desired measurements. For more information, see corresponding measurement chapters.

Admitting a Patient

The monitor admits a new patient in the following situations:

After a patient is manually discharged, the monitor automatically admits a new patient.

After being switched off for the selected time period, the monitor automatically discharges the previous patient and admits a new patient at startup.

If the monitor has not detected certain patient vital signs (ECG, SpO

2

, PR, RR, NIBP) for 30 minutes, it will prompt whether to start monitoring a new patient if any of the above vital signs are detected again.

Always input patient information as soon as the patient is admitted. For more information, see

5.2.2 Editing

Patient Information for details.

WARNING

The default Patient Category setting is Adult, and Paced setting is Unspecified. Set Paced and check if the Patient Category setting is correct for the patient.

For paced patients, set Paced to Yes. If it is incorrectly set to No, the monitor could mistake a pace pulse for a QRS and fail to alarm when the ECG signal is too weak.

For non-paced patients, set Paced to No.

5.2.1

5.2.2

Entering the Patient Management Dialog

Use any of the following methods to enter the Patient Management dialog:

Select the patient information area at the top left corner of the screen.

Select the Patient Management quick key.

Select the Main Menu quick key

from the Patient Management column select Patient Management .

Editing Patient Information

Edit patient information after a patient has been admitted, or when patient information is incomplete, or when it is necessary to change patient information:

To edit patient information, follow this procedure:

1.

Enter the Patient Management

dialog. For more information, see 5.2.1 Entering the Patient Management

Dialog .

2.

Edit patient information as required.

If a barcode reader is connected to the monitor, scanning the patient’s barcode will enter the patient’s information.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 5 - 1

5.2.3

5.2.4

NOTE

The monitor will reload the configuration if you change the patient category.

Loading Patient Information from the CMS

If the monitor is connected to the central monitoring system (CMS). You can load patient information from the

CMS to the monitor. To do so, follow this procedure:

1.

Enter the Find Patient dialog in either of the following ways:

◆ Select the Main Menu quick key → from the Patient Management column select Find Patient .

◆ From the Patient Management dialog select Find Patient .

2.

Input query criteria.

3.

Select Search . Then a list pops up, including all the patients that meet the query criteria.

4.

Select a patient from the patient list, and then select Admit . Corresponding patient information in the monitor will be updated.

Patients that can be searched for are configurable based on location. For more information, see 32.2.2 Setting the

Range of Finding Patient

.

NOTE

If ADT Query is switched on, input query criteria from the Discharged Patients page.

Loading Patient Information from the ADT Server

If the monitor is connected with the Admit-Discharge-Transfer (ADT) server through the eGateway. You can load patient information from ADT server to the monitor.

To do so, follow this procedure:

1.

Enter the Find Patient dialog in either of the following ways:

◆ Select the Main Menu quick key → from the Patient Management column select Find Patient .

◆ Select Find Patient from the Patient Management dialog.

2.

Input query criteria.

3.

Select Search . Then a list pops up, including all the patients that meet the query criteria.

4.

Select a patient from the patient list, and then select Admit . Corresponding patient information in the monitor will be updated.

The patient where you can search is configurable. For more information, see

32.2.2 Setting the Range of Finding

Patient .

NOTE

You can load patient information from the ADT server only when ADT Query is enabled. For more

information, see 32.13.11 Using the ADT Gateway .

The monitor can automatically upload the patient information from the ADT server when:

The configured keywords of the monitor and the eGateway are the same.

The IP address of the monitor is added to the eGateway configuration dialog.

Loading patient information from the ADT server updates only patient information in the monitor.

The patient’s monitoring data is not changed and the patient is not discharged.

5 - 2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

5.3

Transferring Patient

Patient settings and data can be transferred from a T1, BeneVision N1, or MPM module into this monitor. The patient settings and data includes patient information, trend data, alarm events, and parameter alarm limits. The

N1 also allows transfer of full disclosure data.

The MPM allows transfer of patient information, such as patient ID, height, and weight, etc.

WARNING

Do not discharge a patient on the bedside monitor before patient information is completely transferred. Discharging the patient during transfer will result in incomplete patient information.

NOTE

The system automatically switches on the HR/PR (high, low) alarms and lethal arrhythmia alarms after transferring the patient data.

5.3.1

Data Storage Introduction

Familiarizing with the data respectively stored in the patient monitor, T1, N1, or MPM helps you understand the effects incurred by transferring patients with an T1, N1, or MPM.

Type of storage

Data

Settings

Patient demographics

Trend data

Calculation data

Event data

Full disclosure waveforms

Monitor settings

(Alarm pause, alarm volume, etc.)

Parameter settings

(Alarm limits, measurement setting, etc.)

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Can be stored in the monitor?

Can be transferred via MPM?

Can be transferred via T1?

Yes

Yes

No

No

No

No

Yes

Yes

No

Yes

No

No

Can be transferred via N1

Yes

Yes

Yes

Yes

Yes

No

Yes Yes Yes

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 5 - 3

5.3.2

Transferring Patient Data

To transfer the patient data via N1/ T1/MPM, insert it into the SMR.

If the patient demographics in the monitor are consistent with those of in the N1/T1/MPM, N1/T1/MPM automatically uploads the data to the monitor.

If the patient demographics in the monitor are not consistent with those of in the N1/T1/MPM, and Data

Transfer Strategy is set to Always Ask (for more information, see

32.2.4 Setting the Strategy of Transferring

Patient Information ), the monitor prompts the

Select Patient dialog automatically. In this case, you need to select an operation (see the following table) according to the actual situation. If the Data Transfer Strategy is not set to Always Ask then the monitor will follow the strategy setting without prompting the user.

5.4

5.5

Operations

Continue Patient in Monitor

Continue Patient in Module

New Patient

Same Patient

Operation Description Examples of applications

Continue to use the patient data in the monitor. This deletes all patient data in the

N1/T1/MPM and copies all data in the monitor to the N1/T1/MPM.

Continue to use the patient data in N1/T1/

MPM. The monitor discharges the patient, and automatically admits a new patient and copies all data from N1/T1/MPM.

Select this option if you will use neither the information in the monitor nor that in the

N1/T1/MPM. This deletes all data in the monitor and N1/T1/MPM and lets you admit a new patient on the monitor. In this case, you need to re-enter the patient demographics. The monitor will restore the settings according to the patient category.

Select this option if the patient information in the monitor and N1/T1/MPM are different but you are sure that it is the same patient. This merges the patient’s trend data in the monitor and N1/T1/MPM and copies the settings in N1/T1/MPM to the monitor as well.

1. Replace N1/T1/MPM during patient monitoring.

2. After the patient is admitted, connect the

N1/T1/MPM.

You are monitoring a patient using N1/T1/

MPM, and you need to transfer the patient, e.g. from a ward (original monitor) to the operating room (destination monitor).

Connect the N1/T1/MPM before admitting a new patient. However, the monitor and/ or N1/T1/MPM have stored the previous patient’s data and settings.

A patient monitored with the N1/T1/MPM is moved to another department and again moved back.

However, the patient information stored in the N1/T1/MPM was altered before connected to the original monitor.

NOTE

Selecting Apply Module Settings will transfer the N1/T1/MPM settings to the monitor along with the patient data.

Exporting Patient Data

You can export the demographic information and monitoring data of the current and discharged patients via a

USB drive. For more information, see 32.6.3 Exporting Patient Data .

Deleting Patient Data

To delete the data of discharged patients, follow this procedure:

1.

Access the Discharged Patients dialog box by either of the following ways:

◆ Select the Discharged Patients quick key.

◆ Select the Main Menu quick key

→ from the Patient Management column select Discharged

Patients .

2.

From the patient list select desired patients.

3.

Select Delete .

5 - 4 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

5.6

5.7

Stopping a Parameter Measurement

To stop monitoring a parameter, follow this procedure:

1.

Remove corresponding sensor from the patient.

2.

Disconnect the sensor from the patient cable.

3.

Disconnect the patient cable from the parameter module.

Discharging a Patient

Before monitoring a new patient, discharge the previous patient. The technical alarms are reset, and monitor

settings return to their defaults. For more information, see 31.3 Setting Default Patient Category

.

After a patient is discharged, the monitor automatically admits a new patient.

WARNING

Always discharge the previous patient before starting to monitor a new patient. Failure to do so can lead to data being associated with the wrong patient.

To manually discharge a patient, use any of the following methods to access the Discharge Patient dialog:

Swipe down the touchscreen with two fingers.

Select the Discharge Patient quick key.

Select the patient information area at the top left corner of the screen → Discharge Patient .

Select the Patient Management quick key

Discharge Patient .

Select the Main Menu quick key → from the Patient Management column select Discharge .

Select a button in the Discharge Patient dialog:

Print End Case Report : prints the end case report when the patient is discharged.

Discharge : clears the waveform data of the current patient. The monitor loads the default configuration and goes to the standby mode. The current patient becomes a discharged patient.

Clear Patient Data : discharges the current patient and clears the waveform data. The monitor loads the default configuration and does not go to the standby mode. The current patient becomes a discharged patient.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 5 - 5

This page intentionally left blank.

5 - 6 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

6

Interfacing with External Devices

6.1

BeneLink Introduction

BeneLink module is intended for connecting external devices, such as ventilators and anesthesia machines, to the monitor. It allows information (patient data, alarms, etc.) from external devices to be displayed, saved, recorded, or printed through the monitor. If the monitor is connected with the CMS or eGateway, information from external devices can also be transmitted to the CMS or eGateway.

For more information on connecting external devices via the BeneLink module, see BeneLink Module Operator’s

Manual (PN: 046-011948-00).

6.2

BeneLink Safety Information

WARNING

Devices of the same category cannot be connected to the BeneLink module simultaneously.

The parameter labels used on the patient monitor may be different from those used on the external device.

The alarms from external devices may be delayed before transmission to the patient monitor.

There can be differences between the alarm priorities displayed on the monitors and the priorities displayed on external devices interfaced through BeneLink.

CAUTION

Check the compatibility of the external devices and ID adapters before connection. Otherwise, unpredictable system failure may result.

NOTE

The alarm messages from external devices are derived from the open protocol of the corresponding external device. For more information about these alarms, please see the operator’s manual for the corresponding devices.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 6 - 1

6.3

6.4

Differences in Displayed Values

In certain cases, there may be differences between the numerics displayed on the monitor and those on external devices. The table below lists some situations and possible reasons.

Situation

Some parameter values are displayed as invalid values on the monitor.

The monitor and external device display parameter values with different numbers of places of decimals.

Non-continuously measured values and continuously measured values have the same displaying mode in the patient monitor.

Parameter values displayed on the patient monitor and those displayed in the external device are slightly different.

Possible Reasons

The patient monitor and the external device may have different parameter configuration or displaying range of values. If the patient monitor displays a parameter not configured in the external device, or a parameter value from the external device exceeds the displaying range of the monitor, then the corresponding parameter value is displayed on the monitor as an invalid value.

The monitor displays parameter values from the external device based on the monitor displaying rules. Same parameter value is displayed differently when the monitor and the external device display numbers with different precision.

For non-continuously measured values, the monitor displays the latest measured values until a new measurement is transmitted by the external device.

Some parameter values are converted to different units when transmitted to the monitor. Sometimes, values from the external device may be delayed before transmission to the patient monitor.

NOTE

When the pressure units are converted among cmH

2

O, hPa and mbar, the parameter values remain unchanged, for example, 1cmH

2

O=1hPa=1mbar, which may differ from some external devices.

Connecting an External Device

An external device is connected to the BeneLink module through an ID adapter. The ID adapter supports only its matching device.

(4)

(1)

(2)

(3)

(5)

6 - 2

(1) BeneLink Module

(3) RJ45 connecting cable

(5) External device

(6)

(2) Label

(4) ID Adapter

(6) Serial port adapting cable (optional)

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

6.5

To connect an external device, follow this procedure:

1.

Insert the BeneLink module into the SMR.

2.

Connect the ID adapter that matches the external device to the BeneLink module with an RJ45 connecting cable.

3.

Plug the ID adapter into the RS232 port on the external device. Some external devices may have ports incompatible with the ID adapter. In this case, a serial port adapting cable is required.

4.

Stick a device name label to the RJ45 connecting cable at the end close to the BeneLink module. When the

BeneLink module is connected to several external devices, you can tell devices easily with these labels.

Switch on the external device.

After the external device is connected to the monitor, the indicators on both the ID adapter and the BeneLink module illuminate to show that the monitor successfully communicates with the external device.

CAUTION

First installation and debugging should be executed by Mindray service personnel or an authorized technician.

Please check the compatibility of the external device and the ID adapter before connection.

Otherwise, unpredictable system failure may result.

Ports on the BeneLink module are not conventional network connectors. They are intended for connecting with the serial port of designated devices only. Do not connect them to network interfaces.

Accessing the Integrated Devices Screen

You can view the information of external devices in the Integrated Devices screen of the monitor. The

Integrated Devices screen provides information of both individual device and multi devices. To access the

Integrated Devices screen, follow this procedure:

Select the Integrated Devices quick key.

Select the Screen Setup quick key → select Integrated Devices .

Select the Main Menu quick key → from the Display column select Choose Screen → select Integrated

Devices .

Select the numeric area or waveform area of any parameter from the external device

→ select Integrated

Devices button.

The Integrated Devices screen has the following features:

For the parameters measured by the external device, the measurements display directly after the parameter labels.

For the parameters input on the external device, the settings are enclosed in parenthesis after parameter labels.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 6 - 3

6.6

■ For the measured parameters and input parameters that have the same label, parameter measurements and settings are displayed after parameter labels with the settings are enclosed in parenthesis. For example, PEEP 18 (20), in which PEEP is parameter label, 18 is the measurement, and (20) is the setting.

NOTE

Parameters in the Integrated Devices screen are displayed in the order as shown in the selection dialog. If the screen cannot display all the selected parameters, the parameter with higher positions

(from top to bottom and from left to right) in the Select Parameter dialogue are displayed.

Displaying Parameters from External Devices

This monitor can display parameters from external devices in the main screen:

Display waveforms from external devices in the waveform area.

Display labels and measurements of parameters from external devices in the numeric area.

Display respiratory loops of parameters from external deivices in the Loops screen.

The following is an example of parameters (from TOF-Watch SX monitor) displayed in the numeric area of the main screen.

(8) (7)

(6)

(1)

(5)

(2) (3) (4)

(1) Parameter unit

(3) Parameter measurement

(5) Skin temperature

(7) Measurement countdown

(2) Parameter label

(4) Response amplitude of stimulation

(6) Measurement interval

(8) Time of last measurement

NOTE

When displayed in the monitor main screen, parameter labels of external devices are prefixed with the plug sign “+”. For example, if SpO

2

is from an external device, its label is displayed as “+SpO

2

”, and its waveform label is displayed as “+Pleth”.

If a parameter can be obtained either from the monitor or an external device, the measured value, waveform or loops coming from the monitor will be displayed preferentially.

6.6.1

Setting Waveform Properties for Parameters from External Devices

To set the waveform properties for parameters from external devices, follow this procedure:

1.

Access the parameter setup dialog by selecting its waveform area or numeric area.

2.

Set Speed or Scale .

6 - 4 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

6.6.2

6.6.3

6.6.4

Setting Alarms from External Devices

To enable or disable the storage, display, and sound of the external device alarms of a certain priority and category, follow this procedure:

1.

From the Integrated Devices screen select Setup .

2.

Set switches as desired.

If the storage, display, or audio settings of a specific alarm are different from its category or priority, set them individually by adding the Alarm ID to the alarm list. To do so, follow this procedure:

1.

From the Integrated Devices screen select Setup .

2.

Input the Alarm ID for this alarm, and select Add .

To delete a specific external device alarm, select the desired Alarm ID, and select Delete .

Refer to the BeneLink Module Operator’s Manual (PN: 046-011948-00) for the list of supported alarms and

Alarm IDs.

Setting Parameters from External Devices for Display

To select parameters displayed in the Integrated Devices screen, follow this procedure:

1.

From the Integrated Devices screen select Select Parameter .

2.

Select desired parameters.

Displaying Data from External Devices on the Main Screen

This monitor can display data from external devices on the main screen:

Display waveforms from external devices in the waveform area.

Display labels and measurements of parameters from external devices in the numeric area.

If an anesthesia system is connected, display its set parameters (+Anes set) in the numeric area

If pumps are connected, display their alarm status in the numeric area

To display data from external devices on the main screen, follow this procedure:

1.

Access Tile Layout in either of the following ways:

Select the Screen Setup quick key

select the Tile Layout tab.

Select the Main Menu quick key

→ from the Display column select Tile Layout .

2.

Select a parameter numeric area or waveform area, then select the BeneLink element, and then from the popup list select the external device element you want to display in this area.

6.6.4.1

Setting Waveform Properties for Parameters from External Devices

To set the waveform properties for parameters from external devices, follow this procedure:

1.

On the main screen, select the waveform area or numeric area for the external device to enter the parameter setup menu.

2.

Set Speed or Scale .

6.6.4.2

Selecting Measured Parameters from the Anesthesia System for Display

On the main screen, the numeric area of some parameters from an external device, for example the +Paw parameter, can display multiple parameters. To select parameters for display, follow this procedure:

1.

On the main screen select the numeric area of the parameter from the anesthesia system.

2.

Select the Select Parameter tab.

3.

Select a parameter tile, and then select a parameter to display from the popup parameter.

4.

Repeat Step 3 to select additional desired parameters.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 6 - 5

6.6.5

6.6.6

Setting Units for Parameters from External Devices

To set units for parameters from external devices, follow this procedure:

1.

Select the numeric area or waveform area of any parameter from the external device.

2.

Select the Unit tab.

3.

Set the unit for O2 , CO2 , Pressure or Temp .

Accessing the Loops Screen

To enter the Loops screen, choose any of the following ways:

Select the Loops quick key.

Select the Screen Setup quick key → Select the Choose Screen tab → select Respiratory Loops .

Select the Main Menu quick key → from the Display column select Choose Screen → select Respiratory

Loops .

From the main screen, select the numeric area for the anesthesia system or ventilator, select Loops .

NOTE

The monitor only displays real-time loops of the external device, and these loops cannot be displayed or saved as reference loops.

6.7

6.8

6.9

Viewing Alarms from External Devices

If the patient monitor is configured to do so, the monitor displays alarms from external devices in the physiological and technical alarm information areas. A plus sign “+” is added before each alarm message from external devices.

Viewing Parameter Trends from External Devices

If the patient monitor is configured to do so, the monitor saves parameters trends and alarm events from external devices. You can review these data in the Tabular Trends , Graphic Trends , Events and Full Disclosure pages in the Review screen.The monitor adds a “+” before the parameter label of external devices.

For more information, see

25 Review

.

NOTE

Parameters from external devices are saved and displayed according to the time on the monitor.

Recording and Printing Parameter Trends from External Devices

You can record or print parameter trends from external devices. For more information, see

25 Review

,

28 Recording , and

29 Printing

.

6 - 6 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

7

Using with the Telemetry Monitor

7.1

Introduction

You can connect a TM80 or TM70 telemetry monitor (hereinafter called the telemetry) with the N Series monitor

(hereinafter called the monitor) to measure the ECG, Resp, and SpO

2

of ambulatory adult and pediatric patients.

If the telemetry is connected to the BP10 NIBP module, the NIBP measurements can also be transferred to the N

Series monitor via the telemetry.

The telemetry is connected with the monitor via the wireless network. The process of connecting the telemetry and the monitor is called "pairing”. After the telemetry is paired with the monitor, you can view the measurement data from the telemetry on the monitor’s screen.

NOTE

To connect to the TM70 telemetry monitor, N Series monitor system software V01.11.00 and above is required.

7.1.1

Pairing Procedure

Before pairing the telemetry and the monitor, verify the network settings of both the telemetry and the monitor.

Ensure that the telemetry and the monitor are in the same network.

To pair the telemetry with the monitor, follow this procedure:

1.

Access the Bedside Devices dialog in either of the following ways:

Select the Bedside Devices quick key.

Select the Main Menu quick key

→ from the Parameters column select Bedside Devices .

2.

Select + to enter the Add Device dialog. Select Refresh if you want to refresh the unpaired telemetries list.

From the All Unpaired tab or the Unpaired in Current Dept. tab, select the telemetry you want to pair.

◆ In the search box, input the device name, the patient’s MRN, department, or bed number, and then select Search . Select the desired telemetry from the search results.

3.

Select Pair and follow the on-screen instructions: a Select Confirm to start pairing the selected telemetry with the monitor.

b Select patient.

Use Patient in Monitor : uses the patient data in the monitor and deletes patient data in the telemetry.

◆ Use Patient in Telemetry : uses the patient data in the telemetry and deletes patient data in the monitor.

New Patient : discharges the current patient and admits a new patient. The new patient will be monitored.

c Check that the message “This telemetry is pairing with patient monitor XX, are you sure?” is displayed on the desired telemetry. Select Yes .

4.

At the completion of pairing, the message “Paired successfully” is displayed on the monitor. Close the

Pairing and the Bedside Devices dialogs.

After pairing with the telemetry, on the monitor, labels of parameters from the telemetry are followed by “-T”, for example ECG-T, indicating that these parameters come from the telemetry.

CAUTION

Make sure that correct telemetry is selected for pairing with the monitor.

If the wireless network signal is poor, the monitor may have the risk of data loss.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 7 - 1

7.2

NOTE

The telemetry is not intended for monitoring neonatal patients.

By default the device name of the telemetry is its serial number. If you need to change the name of the telemetry, refer to the operator’s manual of the corresponding telemetry.

Binding the Telemetry with the Monitor

After the telemetry is paired with the monitor, discharging the patient from either device automatically unpairs the telemetry and the monitor. If you need the monitor and the telemetry to remain paired after the patient is discharged, bind the two devices. To do so, enter the Bedside Devices dialog and select the unlock symbol on the top left corner. After the two devices are bound together, the lock symbol turns to be locked .

To release the binding of the monitor and the telemetry, select the lock symbol . Then the two devices are unbound and the lock symbol changes to be unlocked .

7.3

7.3.1

7.3.2

Unpairing the Telemetry and the Monitor

If you do not need to connect the monitor and the telemetry to monitor the patient, you can unpair them. After the monitor and the telemetry are unpaired, the CMS receives patient data from the monitor or the telemetry if connected.

Unpairing the Telemetry and the Monitor via the Monitor

To unpair the telemetry from the monitor, follow this procedure:

1.

Access the Bedside Devices dialog in either of the following ways:

◆ Select the Bedside Devices quick key.

◆ Select the Main Menu quick key → from the Parameters column select Bedside Devices .

2.

Select Unpair .

3.

Select which device is used to continue monitoring the patient.

◆ Monitor : the monitor is used to continue monitoring the patient.The telemetry enters standby.

◆ Telemetry : the telemetry is used to continue monitoring the patient. The monitor enters standby.

4.

Close the Unpair and the Bedside Devices dialogs.

Unpairing the Telemetry and the Monitor via the telemetry

To unpair the monitor from the telemetry, follow this procedure:

1.

Press the main menu button on the front panel of the telemetry.

2.

If prompted, input the password for the screen lock dialog. When Screen Lock is set to Off , skip this step.

3.

Select Unpair .

7 - 2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

7.4

4.

Select which device is used to continue monitoring the patient.

◆ Telemetry : the telemetry is used to continue monitoring the patient. The monitor enters standby.

◆ Monitor : the monitor is used to continue monitoring the patient. The telemetry enters standby.

◆ Cancel : do not unpair the telemetry and the monitor.

Troubleshooting

Problem Possible Cause

Telemetry Disconnected The telemetry is out of the wireless network coverage area.

The telemetry is powered off.

The monitor is not connected to the network or network settings are incorrect.

The telemetry is not connected to the wireless network or wireless network settings are incorrect.

The exchanger network connected by the telemetry and the monitor does not support multicast data transfer.

Wireless signal interference exists.

After the telemetry is paired with the monitor, ECG and

Pleth waveforms from the telemetry are not properly displayed on the monitor screen.

Wireless signals are weak.

Corrective Action

Put the telemetry in the Wi-Fi coverage area.

Power on the telemetry.

Verify that the monitor’s network settings are correct and the monitor is connected to the network.

Verify that the telemetry’s network settings are correct and the telemetry is connected to the wireless network.

Contact the service personnel.

Put the telemetry in the wireless coverage area.

Check the interference source and reduce or eliminate interference.

Put the telemetry in the wireless network coverage area with strong signals.

Remove any metallic obstructions that stand between the telemetry and the monitor.

Contact the service personnel.

The telemetry occasionally goes offline.

Insufficient network bandwidth or greater network delay leads to data transfer delay.

Wireless signal interference exists.

Wireless signals are weak.

Put the telemetry in the wireless network coverage area.

Check the interference source and reduce or eliminate interference.

Put the telemetry in the wireless network coverage area with strong signals strength.

Remove any metallic obstructions between the telemetry and the monitor.

Contact the service personnel.

Telemetry Low Battery or

Telemetry Battery Depleted

Insufficient network bandwidth or greater network delay leads to data transfer delay.

The telemetry battery charge is low or depleted.

Replace the battery with a known good one.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 7 - 3

This page intentionally left blank.

7 - 4 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

8

Alarms

8.1

8.2

Alarm Introduction

This chapter describes alarm functions and alarm settings.

Alarm Safety Information

WARNING

A potential hazard can exist if different alarm presets and default configuration settings are used for the same or similar equipment in the same care area, for example an intensive care unit or cardiac operating room.

If the monitor is connected to the central monitoring system (CMS) or other monitors, alarms can be presented and controlled remotely. Remote suspension, inhibition, or reset of monitor alarms via the CMS or other monitors may cause a potential hazard. For more information, see the operator’s manuals for the CMS and the other monitors.

The monitors in the care area may each have different alarm settings to suit different patients.

Always check that the alarm settings are appropriate for the patient before starting the monitoring.

Always make sure that necessary alarm limits are active and set according to the patient's clinical condition.

Setting alarm limits to extreme values may cause the alarm system to become ineffective. For example, setting the SpO

2

high alarm limit to 100%, is equivalent to switching the alarm off.

When the alarm sound is switched off, the monitor gives no alarm tones even if a new alarm occurs.

Be careful when considering to switch off the alarm sound. When the alarms are off or while alarm audio is paused either temporarily or indefinitely, observe the patient frequently.

When monitoring patients that are not continuously attended by a clinical operator, properly configure the alarm system and adjust alarm settings as per the patient's condition.

Do not rely exclusively on the audible alarm system for monitoring. Adjustment of alarm volume to a low level may result in a hazard to the patient.

8.3

8.3.1

8.3.2

Understanding the Alarms

Alarm Categories

The monitor has two different types of alarms: physiological alarms and technical alarms.

Physiological alarms are triggered by patient measurement exceeding the parameter limits, or by an abnormal patient conditions.

Technical alarms are triggered by an electrical, mechanical, connectivity, or other monitor abnormal operation, or by failure of a sensor or component. Technical alarm conditions may also be caused when an algorithm cannot classify or interpret the available data.

Apart from the physiological and technical alarms, the monitor can also prompt with some messages describing the system status or patient status.

Alarm Priorities

By severity, the alarms are classified into the following priority levels:

High priority alarms: indicate a life threatening situation or a severe device malfunction. High priority alarms require an immediate response.

Medium priority alarms: indicate abnormal vital signs or a device malfunction. Medium priority alarms require a timely response.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 8 - 1

8.3.3

Low priority alarms: indicate a discomfort condition, a device malfunction, or an improper operation. Low priority alarms require you to be aware of this condition.

Messages: provides additional information on the patient or the equipment.

Alarm Indicators

When an alarm occurs, the monitor indicates it through visual or audible alarm indications. For more information, see the following table.

Alarm Indicator High Priority

Alarm

Alarm lamp

Audible tone pattern

Red

Flashing frequency: 1.4 -

2.8 Hz

Duty cycle: 20 -

60% on

ISO Repeat pattern of triple + double + triple

+ double beeps

Mode 1 Repeat pattern of high-pitched single beep

Mode 2 Repeat pattern of high-pitched triple beeps

Alarm message White text inside a red box

Medium

Priority Alarm

Yellow

Flashing frequency: 0.4 -

0.8 Hz

Duty cycle: 20 -

60% on

Repeat pattern of triple beeps

Repeat pattern of double beeps

Repeat pattern of double beeps

Black text inside a yellow box

Low Priority

Alarm

Cy a n

No flashing

Duty cycle:

100% on

Single beep

Low-pitched single beep

Low-pitched single beep

Black text inside a cyan box

Message Comments

None

None

None

None

White text

None

None

Alarm priority indicator

Parameter value

*** ** * None

Alarm messages are displayed in the alarm information area at the top of the screen. You can select the alarm messages to show the alarm list.

The indicator appears in front of the corresponding alarm message.

None White text inside a flashing red box

Black text inside a flashing yellow box

Black text inside a flashing cyan box

None

NOTE

When multiple alarms of different priority levels occur simultaneously, the monitor selects the highest priority alarm to light the alarm lamp and issue the alarm tone.

When multiple alarms of different priority levels occur simultaneously and should be displayed in the same area, the monitor only displays the messages from the highest priority alarm.

When multiple alarms of the same priority levels occur simultaneously alarm messages are displayed cyclically.

Some physiological alarms, such as Asystole, are exclusive. They have identical alarm tones and alarm lights with normal high level physiological alarms, but their alarm messages are displayed exclusively. That is, when an exclusive physiological alarm and a normal high level physiological alarm are triggered simultaneously, only exclusive physiological alarm message will be displayed.

8 - 2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

8.3.4

Alarm Status Symbols

Apart from the alarm indicators as described in 8.3.3 Alarm Indicators , the monitor uses the following symbols

to indicate the alarm status:

Alarm pause: indicates that all the alarms are paused.

Alarm off: indicates that individual measurement alarms are turned off or the system is in the alarm off status.

Audio pause: indicates that audible alarm tones are paused.

Audio off: indicates that audible alarm tones are turned off.

Alarm reset: indicates that the alarm system is reset.

8.4

Accessing On-screen Help for Technical Alarms (AlarmSight)

In the technical alarm list, alarm messages followed by Detail include help messages or pictures to help you identify the problem. This function is called AlarmSight. To access AlarmSight, follow this procedure:

1.

Select an alarm displaying in the technical alarm information area to enter the Alarms dialog.

2.

From the alarm list select the desired alarm.

NOTE

If there are no displayed alarms, then selecting the technical alarm area will not open any dialogs.

8.5

8.6

8.6.1

Checking the Physiological Alarms List

If the patient monitor has more than one physiological alarm, you can see the physiological alarm list by selecting the physiological alarm information area to enter the Alarms dialog. If there is only one physiological alarm, selecting the physiological alarm information area enters the Review dialog for the event created by this alarm.

Changing Alarm Settings

Alarms settings can be changed by selecting Main Menu and choosing from the list in the Alarm column.

Setting Parameter Alarm Properties

To set parameter alarm properties, follow this procedure:

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 8 - 3

1.

Access the Limits page in either of the following WAYS:

◆ Select the Alarm Setup quick key.

Select the Main Menu quick key

from the Alarm column select Limits.

2.

Enter the password if required.

3.

Select a parameter tab and set alarm properties as desired.

You can also change the alarm properties of individual parameter from the corresponding parameter dialog.

8.6.2

NOTE

The monitor can be configured to require an password or login to change alarm settings. For more

information, refer to 32.14.3 Selecting Password for User Authentication

.

Changing the Alarm Volume

To change the alarm volume, follow this procedure:

1.

Access the Setup page in either of the following ways:

◆ Select the Alarm Setup quick key → select the Setup tab.

◆ Select the Main Menu quick key → from the Alarm column select Setup.

2.

Set Alarm Volume . The alarm volume can be set from X to 10, in which X is the minimum volume allowed by your department, and 10 is the maximum volume.

3.

Select High Alarm Volume to set the volume of the high priority alarm.

4.

Select Reminder Volume to set the volume of the reminder tone.

NOTE

Alarm volume can only be set to 0 if the monitor is connected to a Central Monitoring System.

When the alarm volume is set to 0, the alarm sound is turned off and the audio off symbol appears on the screen.

You cannot set the volume of high priority alarms if Alarm Volume is set to 0.

You can set the Alarm Volume to 0 only when your monitor is connected to the CMS. If your monitor is not connected to the CMS, the lowest setting of Alarm Volume is 1.

When the monitor is connected to the CMS and Alarm Volume is set to 0, Alarm Volume automatically changes to 2 if the CMS is disconnected.

8 - 4 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

8.6.3

Setting the Auto Limits for New Patient Switch

If the Auto Limits for New Patient function is enabled, a dialog box can pop up to ask you whether to set alarm limits based on the latest parameter measurements for a newly admitted patient. To set the Auto Limits for New

Patient switch, follow this procedure:

1.

Enter the alarm setup page in any of the following ways:

Select the Alarm Setup quick key

select the Setup tab.

◆ Select the Main Menu quick key

from the Alarm column select Setup .

2.

Set the Auto Limits for New Patient switch.

When Auto Limits for New Patient is switched on, the confirmation dialog box pops up if all of the following requirements are met:

Within 10 minutes after the patient is admitted.

Continuous measurements are stable.

An NIBP measurement has been taken.

HR alarm switch is on.

No lethal alarms are triggered.

The patient is not in poor perfusion condition.

Alarm limits of no parameters were manually changed.

The monitor is not in intubation mode, private mode, or CPB mode.

NOTE

The Auto Limits for New Patient function is intended for newly admitted patients only.

The automatically set alarm limits take effect only after being confirmed.

8.6.4

Initiating Auto Alarm Limits

The monitor provides the auto alarm limits function to automatically adjust alarm limits according to the patient’s vital signs. When auto limits are selected, the monitor calculates safe auto limits based on the latest measured values. To get accurate auto alarm limits, collect a set of measured vital signs as a baseline.

To initiate auto alarm limits, follow this procedure:

1.

Access the Limits page in either of the following ways:

◆ Select the Alarm Setup quick key.

◆ Select the Main Menu quick key → from the Alarm column select Limits.

2.

From the Limits page, select Auto Limits at the left bottom.

3.

Select Ok from the popup dialog box.

Then the monitor will automatically calculate alarm limits based on the latest measured values. Before applying these automatically created alarm limits, confirm if they are appropriate for the patient from the Limits dialog. If not, you can adjust them manually. These alarm limits will remain unchanged until you select auto limits again or adjust them manually.

The monitor calculates auto limits based on the following rules:

Module

ECG

Parameter Lower Limit

HR/PR

(bpm)

Upper Limit Auto Limit Range

Adult/ pediatric

Neonate Adult/ pediatric

HR × 0.8 or 40

(whichever is greater)

(HR - 30) or 90

(whichever is greater)

HR × 1.25 or

240

(whichever is smaller)

Neonate

(HR + 40) or

200

(whichever is smaller)

Adult/pediatric: 35 to 240

Neonate: 55 to 225

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 8 - 5

8 - 6

Module

Resp

SpO

NIBP

2

Temp

(xx refers to temperat ure site)

IBP: ART/

Ao/UAP/

BAP/FAP/

LV/P1-P4

(Arterial pressure)

Txx (°C)

ΔT (°C)

IBP-S

(mmHg)

IBP-D

(mmHg

IBP: PA

Parameter Lower Limit

RR (rpm)

SpO

2

(%)

NIBP-S

(mmHg)

NIBP-D

(mmHg)

NIBP-M

(mmHg)

Upper Limit Auto Limit Range

Adult/ pediatric

RR × 0.5 or 6

(whichever is greater)

Same as the default alarm limit

(SYS × 0.68 +

10)

(Dia × 0.68 +

6)

(Mean × 0.68

+ 8)

(Txx - 0.5)

Same as the default alarm limit

Neonate Adult/ pediatric

Neonate

(RR - 10) or 30

(whichever is greater)

Same as the default alarm limit

(SYS - 15) or

45 (whichever is greater)

(Dia - 15) or

20 (whichever is greater)

(Mean - 15) or

35 (whichever is greater)

(Txx - 0.5)

Same as the default alarm limit

(RR × 1.5) or

30 (whichever is smaller)

Same as the default alarm limit

(SYS × 0.86 +

38)

(Dia × 0.86 +

32)

(RR + 25) or

85 (whichever is smaller)

Same as the default alarm limit

(SYS + 15) or

105

(whichever is smaller)

(Dia + 15) or

80 (whichever is smaller)

(Mean × 0.86

+ 35)

(Txx + 0.5)

Same as the default alarm limit

Adult/pediatric: 6 to 55

Neonate: 10 to 90

Same as the measurement range

Adult: 45 to 270

Pediatric: 45 to 185

Neonate: 35 to 115

(Mean + 15 or

95)

(whichever is smaller)

(Txx + 0.5)

Same as the default alarm limit

Adult: 25 to 225

Pediatric: 25 to 150

Neonate: 20 to 90

Adult: 30 to 245

Pediatric: 30 to 180

Neonate: 25 to105

1 to 49

Same as the measurement range

IBP-M

(mmHg)

SYS × 0.68 +

10

(Dia × 0.68 +

6)

Mean × 0.68 +

8

SYS × 0.75

(SYS - 15) or

45 (whichever is greater)

(Dia - 15) or

20 (whichever is greater)

(Mean - 15) or

35

(whichever is greater)

SYS × 0.75

SYS × 0.86 +

38

(Dia × 0.86 +

32)

Mean × 0.86 +

35

SYS × 1.25

(SYS + 15) or

105

(whichever is smaller)

(Dia + 15) or

80 (whichever is smaller)

(Mean + 15) or 95

(whichever is smaller)

SYS × 1.25

Adult: 45 to 270

Pediatric: 45 to 185

Neonate: 35 to 115

Adult: 25 to 225

Pediatric: 25 to 150

Neonate: 20 to 90

Adult: 30 to 245

Pediatric: 30 to180

Neonate: 25 to 105

3 to 120

IBP: CPP

IBP-S

(mmHg)

IBP-D

(mmHg

IBP-M

(mmHg)

CPP

(mmHg)

Dia × 0.75

Mean × 0.75

Dia × 0.75

Mean × 0.75

CPP × 0.68 + 8 (CPP-15) or

Mean × 0.75

35,

(whichever is greater)

Mean × 0.75

Dia × 1.25

Mean × 1.25

CPP × 0.86 +

35

Mean × 1.25

Dia × 1.25

Mean × 1.25

3 to 120

3 to 120

IBP: CVP/

LAP/

RAP/

UVP/P1-

P4

(Venous pressure)

IBP-M

(mmHg)

(CPP+15) or

95,

(whichever is smaller)

Mean × 1.25

Adult: 20 to 235

Pediatric: 25 to175

Neonate: 25 to 100

3 to 40

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Module

CO

AG

C.O.

SvO rSO

2

2

2

Parameter Lower Limit

EtCO

(mmHg)

FiCO

FiO

2

/EtO

ScvO rSO

2

2 awRR

(rpm)

EtCO

FiCO

2 awRR

(rpm)

FiAA/EtAA

FiN

EtN

TB (°C)

C.I.

TFC

SvO

2

2

O/

2

2

2

O

(%)

2

2

(%)

Upper Limit Auto Limit Range

Adult/ pediatric

0 to 32: remains the same

33 to 35: 29

Neonate

0 to 32: remains the same

33 to 35: 29 awRR × 0.5 or

6 (whichever is greater)

(awRR - 10) or

30 (whichever is greater)

Adult/ pediatric

0 to 32: remains the same

33 to 35: 41

Neonate

0 to 32: remains the same

33 to 35: 41

Same as the measurement range

36 to 45:

(etCO

2

- 6)

46 to 48: 39

>48: remains the same

None

36 to 45:

(etCO

2

- 6)

46 to 48: 39

>48: remains the same

None

36 to 45:

(etCO

2 limit

+ 6)

46 to 48: 51

>48: remains the same

Same as the default alarm awRR × 1.5 or

30 (whichever is smaller)

36 to 45:

(etCO

2

+ 6)

46 to 48: 51

>48: remains the same

Same as the default alarm limit

(awRR+25) or

85 rpm

(whichever is smaller)

Same as the measurement range

Same as the measurement range

Same as the measurement range

Same as the measurement range

Same as the measurement range

Adult/pediatric: 6 to 55

Neonate: 10 to 90

Same as the CO

2

module

Same as the CO

2

module awRR × 0.5 or

6 (whichever is greater)

(awRR - 10) or

30 (whichever is greater)

Same as the default alarm limit

Same as the default alarm limit

Same as the default alarm limit

Same as the default alarm limit

Same as the default alarm limit

Adult: (TB - 1)

Pediatric: N/A

N/A

N/A

SvO

2

- 5

Same as the default alarm limit

N/A

N/A

N/A

N/A awRR × 1.5 or

30 (whichever is smaller)

Same as the default alarm limit

Same as the default alarm limit

Same as the default alarm limit

Adult: (TB + 1)

Pediatric: N/A

N/A

N/A

SvO

2

+ 5

N/A

N/A

N/A

ScvO

2

- 5 N/A ScvO

2

+ 5 awRR+25 or

85 (whichever is smaller)

Same as the default alarm limit

Same as the default alarm limit

Same as the default alarm limit

N/A

Adult/pediatric: 6 to 55

Neonate: 10 to 90

Same as the measurement range

Same as the measurement range

Same as the measurement range

Same as the measurement range

N/A

N/A

N/A

Same as the measurement range

Same as the measurement range

15 to 95 Same as the default alarm limit

Manual mode: same as the default alarm limit

Auto mode: remains unchanged

Same as the default alarm limit

Manual mode: same as the default alarm limit

Auto mode: remains unchanged

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 8 - 7

8.6.5

8.6.6

8.6.7

Setting the Apnea Delay Time

To set the apnea delay time, follow this procedure:

1.

Access the Setup page in either of the following ways:

◆ Select the Alarm Setup quick key → select the Setup tab.

◆ Select the Main Menu quick key → from the Alarm column select Setup.

2.

Select Apnea Delay to set the apnea delay time.

Switching On or Off V-Tach Latching

When V-Tach Latching Lock is switched on, you can switch on or off V-Tach Latching . To do so, follow this procedure:

1.

Access the Setup page in either of the following ways in either of the following ways:

◆ Select the Alarm Setup quick key → select the Setup tab.

◆ Select the Main Menu quick key → from the Alarm column select Setup.

2.

Switching on or off V-Tach Latching .

For information on the V-Tach Latching function, see 32.3.4 Latching Physiological Alarms

.

Restoring the Default Alarm Settings

To reset all alarm settings to the defaults, follow this procedure:

1.

Access the Alarm page in either of the following ways:

◆ Select the Alarm Setup quick key.

◆ Select the Main Menu quick key → from the Alarm column select Limits.

2.

On the Limits page, select Defaults at the bottom.

8.6.8

Setting the Length of Printed Waveforms

You can define the length of printed waveforms when an alarm is triggered. To do so, follow this procedure:

1.

Access the Setup page in either of the following ways:

◆ Select the Alarm Setup quick key

select the Setup tab.

◆ Select the Main Menu quick key

from the Alarm column select Setup.

2.

Set Printing Duration On Alarm .

8.7

Pausing Alarms/Pausing Alarm Tones

You can either pause alarms or pause alarm tones. This depends on the pause setting. To set the pause function, refer to

32.3.3 Defining the Pause Function

.

8.7.1

Pausing Alarms

If the Pause function is configured to alarm pause, pressing the Alarm Pause quick key can temporarily disable alarm indicators. For the configuration of the pause function, see

32.3.3 Defining the Pause Function

. When alarms are paused, the following rules are followed:

No physiological alarm will be presented.

Except battery-related technical alarms, sounds of other technical alarms are paused, but alarm lamps and alarm messages continue to be presented.

The remaining alarm pause time is displayed in the physiological alarm information area.

The alarm pause symbol is displayed in the system information area.

When the alarm pause time expires, the alarm paused state is automatically deactivated. You can also cancel the alarm paused state by pressing the Alarm Pause quick key again.

8 - 8 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

8.7.1.1

Prolonging the Alarm Pause Time

If enabled, you can temporarily prolong the alarm pause time after the monitor enters the alarm paused state.

This capability is disabled by default. To prolong the alarm pause time, follow this procedure:

1.

In the physiological alarm information area, select the alarm pause countdown.

2.

Select Pause 5 min, Pause 10 min , or Pause 15 min .

NOTE

Prolonging alarm pause time does not affect the setting of alarm pause time.

8.7.1.2

Switching Off All Alarms

If Pause Time is set to Permanent

(see section 32.3.3.1 Setting the Alarm Pause Time/Alarm Tone Pause Time

), pressing the Alarm Pause quick key permanently switches off all alarms. The alarm off state has the following features:

Physiological alarms are switched off. The alarm lamp does not flash and alarm sound is not issued.

Alarm sound of technical alarms is switched off, but alarm lamp flashes and alarm messages are presented.

The message Alarm Off with red background is displayed in the physiological alarm information area.

The alarm off symbol is displayed in the system status information area.

To exit the alarm off state, press the Alarm Pause quick key again.

WARNING

Pausing or switching off alarms may result in a hazard to the patient.

8.7.2

Pausing Alarm Sound

If the Pause function is configured to Audio Pause, pressing the Audio Pause quick key pauses alarm tone and

sets the quick key to be highlighted. For the configuration of the pause function, see section 32.3.3 Defining the

Pause Function

. When alarm tones are paused, the following rules are followed:

The sound of all physiological alarms and technical alarms are switched off.

The remaining audio pause time is displayed in the physiological alarm information area.

The audio pause symbol is displayed in the system information area.

When the audio pause time expires, the audio paused state is automatically deactivated. You can also cancel the audio paused state by pressing the highlighted Audio Pause quick key again.

8.7.2.1

Prolonging the Alarm Tone Pause Time

If enabled, you can temporarily prolong the alarm tone pause time after the monitor enters the alarm tone paused state. This capability is disable by default. To prolong the audio pause time, follow this procedure:

1.

In the physiological alarm information area, select the alarm pause countdown.

2.

Select the alarm tone pause time. Options may include Pause 5 min, Pause 10 min , or Pause 15 min .

NOTE

Prolonging alarm pause time does not affect the setting of alarm tone pause time.

8.7.2.2

Switching Off Alarm Sound

If Pause Time is set to Permanent

(see section 32.3.3.1 Setting the Alarm Pause Time/Alarm Tone Pause Time

), pressing the Audio Pause quick key permanently switches off all alarm sound. The audio off state has the following features:

Alarm sound of both physiological alarms and technical alarms is switched off.

The audio off symbol is displayed in the system information area.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 8 - 9

To exit the audio off state, press the Audio Pause quick key again.

WARNING

Pausing or switching off alarm sound may result in a hazard to the patient.

8.8

Resetting Alarms

Press the Alarm Reset quick key to reset the alarm system. When the alarm system is reset, the alarm reset symbol displays in the system status information area for alarm symbols.

NOTE

If a new alarm is triggered after the alarm system is reset, the alarm reset icon will disappear and the alarm light and alarm tone will be reactivated.

8.8.1

8.8.2

8.9

Resetting Physiological Alarms

For physiological alarms, when the alarm system is reset, the following occur:

The alarm sound is silenced.

A √ appears before the alarm message.

The color of the parameter numeric background corresponds with the alarm priority, but the parameter numeric does not flash.

Resetting Technical Alarms

For technical alarms, when the alarm system is reset, the following occur:

Some technical alarms are cleared. The monitor gives no alarm indications.

Some technical alarms are changed to prompt messages.

For some technical alarms, the alarm is silenced and a √ appears before the alarm message.

For details about the indications of technical alarms when the alarm system is reset, see

D.2 Technical Alarm

Messages

.

Latching Alarms

The latching setting for physiological alarms defines how alarm indicators behave if you do not reset the alarms.

If you do not “latch” physiological alarms, their alarm indications disappear when the alarm condition ends.

If you “latch” physiological alarms, all visual and audible alarm indications remain (if configured) until the alarms are reset. For latched alarms the time when the alarm is last triggered is displayed behind the alarm message.

The monitor can be set to separately latch visual indications or simultaneously latch the visual and the audible indications.

When visual indications are latched, visual indications, including alarm lamp, alarm message and its background remain when the alarm condition ends and the time when the alarm last triggered is displayed behind the alarm message.

When audible indications are latched, the monitor continues to issues alarm sounds when the alarm condition ends.

To set how you want to latch the physiological alarms, see section 32.3.4 Latching Physiological Alarms .

NOTE

Changing alarm priority may affect the latching status of the corresponding alarm. Determine if you need to adjust the latching status for the specific alarm when you have changed its alarm priority.

When the alarm system is reset, latched physiological alarms are cleared.

8 - 10 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

8.10

Nurse Call

The monitor provides a nurse call connector to output a nurse call signal when a user-defined alarm occurs. To obtain a nurse call signal, use the nurse call cable to connect the hospital nurse call system with the monitor’s nurse call connector.

Alarms are indicated on the nurse call device only when the following conditions are met:

The nurse call system is enabled.

A user-defined alarm occurs.

Alarms are not paused or reset.

WARNING

Do not rely exclusively on the nurse call system for alarm notification. Remember that the most reliable alarm notification combines audible and visual alarm indications with the patient’s clinical condition.

8.11

Calling for Help

This monitor can be configured to call monitors in the same department so that nearby doctors and nurses can come to help.

8.11.1

Issuing the Call Help Signal

To call help, select the Call Help quick key and select Ok from the popup dialog box. If Ok is not selected within 5 seconds then the monitor will automatically send out the call help signal.

After the call help signal is sent out, the Call Help quick key flashes in red. Select the Call Help quick key again to stop the call help signal.

Monitors receiving the call help signal issue a sound and a dialog box pops up indicating which monitor is calling.

Select Ok to acknowledge the call and stop the sound at this monitor.

NOTE

The call help function works only when the monitor is connected to the network.

The call help sound may disturb patients in the same department.

8.12

CPB Mode

The CPB (Cardiopulmonary Bypass) mode is activated only if you set the department to OR .

In the CPB mode, except for EEG, NMT, tcGas, and rSO

2

related alarms, all the physiological alarms and technical alarms are switched off. So when performing CPB, you can put the monitor in the CPB mode to inactivate unnecessary alarms.

8.12.1

Entering the CPB Mode

To enter the CPB mode, choose either of the following ways:

Select the CPB Mode quick key.

Select the Main Menu quick key

from the Alarm column select CPB Mode .

In the CPB mode, CPB Mode is displayed in the physiological alarm area with a red background color.

NOTE

When the CPB mode is entered, the monitor stops all NIBP measurements. You can restart NIBP measurements after entering the CPB mode.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 8 - 11

8.12.2

Exiting the CPB Mode

To exit the CPB mode, choose either of the following ways:

Select the CPB Mode quick key.

Select the Main Menu quick key

from the Alarm column select Exit CPB Mode .

8.13

Intubation Mode

Intubation mode is available for Resp, CO

2

, and AG monitoring. When performing intubation during general anesthesia, you can put the monitor in the intubation mode in order to inactivate unnecessary alarms.

In the intubation mode, Resp, CO

2

, and AG related physiological alarms are switched off.

8.13.1

Entering the Intubation Mode

To enter the intubation mode, choose either of the following ways:

Select the Intubation Mode quick key.

From the bottom of the Resp , CO2 , or AG dialog, select Intubation Mode .

Select the Main Menu quick key

→ from the Alarm column select Intubation Mode .

8.13.2

Exiting the Intubation Mode

To exit the intubation mode, choose either of the following ways:

Select the Intubation Mode quick key.

From the bottom of the Resp , CO2 , or AG dialog, select Exit Intubation Mode .

Select the Main Menu quick key from the Alarm column

→ select Exit Intubation Mode .

8.14

Testing Alarms

The monitor automatically performs a selftest at startup. Check that an alarm tone is heard, the alarm lamp illuminates, one after the other, in red, yellow, and cyan. This indicates that the visible and audible alarm indicators function correctly.

To further test individual measurement alarms, perform measurements on yourself or using a simulator. Adjust alarm limits and check that appropriate alarm behavior is observed.

8 - 12 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

9

Monitoring ECG, Arrhythmia, ST and QT

9.1

ECG Introduction

The electrocardiogram (ECG) measures the electrical activity of the heart and displays it on the monitor as waveforms and numerics. ECG monitoring provides 3-, 5-, 6-, and 12-lead ECG monitoring, ST-segment analysis, arrhythmia analysis, QT/QTc measurements, and optional 12-lead resting analysis via the Glasgow algorithm.

The associated module incorporating ST analysis has the ST label.

The associated module incorporating 12-lead has the 12-lead label.

(3)

(1)

(4)

(1)

(2)

(1)

9.2

(1) ECG cable connector (2) ST label (3) 12-lead label (4) Glasgow label

ECG Safety Information

WARNING

This equipment is not intended for direct cardiac application.

Ensure that the conductive parts of ECG electrodes and associated connectors, including the neutral electrode, do not come into contact with any other conductive parts including earth.

Always use defibrillation-proof ECG cables during defibrillation.

Do not touch the patient, or metal devices connected to the patient, during defibrillation.

To minimize the hazard of burns during high-frequency surgical procedures, ensure that the monitor’s cables and transducers never come into contact with the electrosurgery unit (ESU).

To minimize the hazard of burns during use of high-frequency surgical unit (ESU), the ECG electrodes should not be located between the surgical site and the ESU return electrode.

CAUTION

Only use parts and accessories specified in this manual. Follow the instructions for use and adhere to all warnings and cautions.

Periodically inspect the electrode application sites to ensure skin integrity. If the skin quality changes, replace the electrodes or change the application site.

Interference from an ungrounded instrument near the patient or electrosurgery usage can induce noise and artifact into the waveforms.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 9 - 1

9.3

ECG Display

The following figures show the ECG waveform and numeric areas. The display may be configured to look slightly different.

(1) (2) (3) (4) (5) (6) (7)

(1) ECG lead label of the displayed waveform

(3) ECG filter mode

(2) ECG waveform gain

(4) Notch filter status

(5) Paced status: If Paced is set to Yes , is displayed. If Paced is set to No , is displayed.

(6) Pace pulse marker: If Paced is set to Yes , pace pulse markers “|” are displayed corresponding to detected pacer for each beat.

(7) HR/PR alarm message

(1)

(2)

(4)

(3)

9.4

9.4.1

9.4.2

(1) Parameter label

(3) HR alarm limits

(2) HR measurement unit

(4) HR value

Preparing for ECG Monitoring

Preparing the Patient Skin

Proper skin preparation is necessary to ensure good signal quality at the electrode sites, as the skin is a poor conductor of electricity. To properly prepare the skin, choose flat areas and then follow this procedure:

1.

Shave hair from skin at chosen electrode sites.

2.

Gently rub skin surface at sites to remove dead skin cells.

3.

Thoroughly cleanse the site with a mild soap and water solution. We do not recommend using ether or pure alcohol, because this dries the skin and increases the resistance.

4.

Dry the skin completely before applying electrodes.

Applying Electrodes and Connecting the Patient

To connect ECG cables, follow this procedure:

1.

Check that electrode packages are intact and the electrodes are not past the expiry date. Ensure the electrode gel is moist. If you are using snap electrodes, attach the snaps to the electrodes before placing electrodes on the patient.

2.

Place the electrodes on the prepared sites. Ensure that all electrodes have good skin contact.

9 - 2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

3.

Connect the leadwires to the patient cable if not already connected.

4.

Plug the patient cable into the ECG connector.

CAUTION

Never mix patient electrode types or brands. This may lead to impedance mismatches issues.

NOTE

When applying the electrodes, avoid bony area, obvious layers of fat, and major muscles. Muscle movement can result in electrical interference. Applying electrodes on major muscles, for example on muscles of the thorax, may lead to erroneous arrhythmia alarms due to excessive muscle movement.

9.4.3

Lead Wire Color Code

The following table lists the color coding of leadwires for AHA standards:

Lead

Right arm

Left arm

Right leg (neutral)

Left leg

Chest 1

Chest 2

Chest 3

Chest 4

Chest 5

Chest 6

Label

V1

V2

V3

V4

RA

LA

RL

LL

V5

V6

Color

White

Black

Green

Red

Brown/Red

Brown/Yellow

Brown/Green

Brown/Blue

Brown/Orange

Brown/Violet

9.4.4

ECG Electrode Placements

In this section, electrode placement is illustrated using the AHA naming convention.

9.4.4.1

3-leadwire Electrode Placement

The following is the electrode configuration when a 3-leadwire cable is used:

RA placement: directly below the clavicle and near the right shoulder.

LA placement: directly below the clavicle and near the left shoulder.

LL placement: on the left lower abdomen.

RA LA

LL

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 9 - 3

9.4.4.2

5-leadwire and 6-leadwire Electrode Placement

The following is the electrode configuration for a 5-leadwires cables:

RA placement: directly below the clavicle and near the right shoulder.

LA placement: directly below the clavicle and near the left shoulder.

RL placement: on the right lower abdomen.

LL placement: on the left lower abdomen.

V placement: on the chest in any of the V1 to V6 positions.

For 6-leadwire placement, you can use the position for the 5 leadwire placement but with two chest leads. The two chest leads (Va and Vb) can be positioned at any two of the V1 to V6 positions. For more information, see

9.4.4.3 Chest Electrode Placement

. The Va and

Vb lead positions are configurable. For more information, see

9.5.3.3 Changing Va and Vb Labels

.

V

RL

RA LA

9.4.4.3

Chest Electrode Placement

The chest electrode can be placed at the following positions:

V1 placement: on the fourth intercostal space to the right of the sternum.

V2 placement: on the fourth intercostal space to the left of the sternum.

V3 placement: midway between the V2 and V4 electrode positions.

V4 placement: on the fifth intercostal space at the left midclavicular line.

V5 placement: on the left anterior axillary line, horizontal with the V4 electrode position.

V1

VE V2 V3

V6 placement: on the left midaxillary line, horizontal with the V4 electrode position.

V3R-V6R placement: on the right side of the chest in positions corresponding to those on the left.

V6

V5

V4

NOTE

For the 5-leadwire and 6-leadwire placement, place the precordial electrode according to the physician's preference.

LL

9 - 4 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

9.4.4.4

10-leadwire Electrode Placement

12-lead ECG uses 10 electrodes, which are placed on the patient’s four limbs and chest. The limb electrodes should be placed on the limb extremities and the chest electrodes placed according to the physician’s preference. The figure at right illustrates the conventional 10-leadwire electrode placement.

9.4.4.5

Lead Placement for Pacemaker Patients

The recommended lead placement for monitoring a pacemaker patient is as follows.

RA

Pacemaker

LA

LL

RA

V1

RL

Pacemaker

LA

LL

3-leadwire Electrode Placement 5-leadwire Electrode Placement

A pacemaker patient usually requires a different electrode patch placement configuration than a nonpacemaker patient.

Do not place an ECG electrode directly over the pacemaker generator. Place the electrode patches 2 to 3 inches away from the pacemaker generator area. For example, if the pacemaker generator is located in the left subclavian area, relocate the Left Arm electrode closer in towards the center of the chest.

9.4.4.6

Lead Placement for Surgical Patients

The surgical site should be taken into consideration when placing electrodes on a surgical patient. For example, for open-chest surgery, the chest electrodes can be placed on the lateral chest or back. To reduce artifacts and interference from electrosurgical units, you can place the limb electrodes close to the shoulders and lower abdomen and the chest electrodes on the left side of the mid-chest. Do not place the electrodes on the upper arm. Otherwise, the ECG waveform will be very small.

WARNING

To reduce the hazard of burns during use of electrosurgical units (ESU), the ECG electrodes should not be located between the surgical site and the ESU return electrode.

Never entangle the ESU cable and the ECG cable together.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 9 - 5

When using ESU, never place ECG electrodes near the grounding plate of the ESU, as this can cause a lot of interference on the ECG signal.

When using ESU, ensure proper contact of the ESU return electrode to the patient to avoid burns at monitor measurement sites. Also ensure that the ESU return electrode is near the operating area.

9.4.5

9.4.6

Choosing the ECG Lead Type

To choose ECG lead type, follow this procedure:

1.

Select the ECG numeric area or waveform area to enter the ECG dialog.

2.

Select the Setup tab.

3.

Set Lead Set according to the lead type you are going to use. The default lead type is Auto . In this case, the monitor automatically detects the lead type.

Checking Paced Status

It is important to correctly set the patient’s paced status before you start monitoring ECG. The paced symbol is displayed when Paced is set to Yes . Pace pulse markers “|” are displayed on each ECG waveform whenever a pacer signal is detected. If Paced is set to No or unspecified, the symbol will be shown in the ECG waveform area.

To change the paced status, follow this procedure:

1.

Select the ECG numeric area or waveform area to enter the ECG dialog.

2.

Select the Pacer tab.

3.

Set Paced to Yes or No .

You can also change the patient’s paced status from the Patient Management dialog. For more information, see

5.2.1 Entering the Patient Management Dialog

.

If you did not set the paced status, the monitor issues a prompt tone when pace pulse is detected. At the same time, the paced symbol flashes and the message Please check if the patient has a pacemaker?

appears in the ECG waveform area. Check and set the patient’s paced status.

WARNING

For paced patients, you must set Paced to Yes. If it is incorrectly set to No, the monitor could mistake a pace pulse for a QRS complex and fail to alarm when the ECG signal is too weak. On ventricular paced patients, episodes of ventricular tachycardia may not always be detected. Do not rely entirely upon the system’s automated arrhythmia detection algorithm.

False low heart rate or false asystole alarms may result with certain pacemakers because of pacemaker artifacts, such as electrical overshoot of the pacemaker overlapping the true QRS complexes.

Do not rely entirely on rate meter alarms when monitoring patients with pacemakers. Always keep these patients under close surveillance.

The auto pacer recognition function is not applicable to pediatric patients, neonatal patients, and patients with NMT monitoring.

9.4.7

Enabling Pacer Rejection

To eliminate the pacing pulse from the ECG waveform of paced patients, it is recommended to enable the pace pulse rejection function. The pace pulse rejection function is disabled by default. To enable this function, follow this procedure:

1.

Select the ECG numeric area or waveform area to enter the ECG dialog.

2.

Select the Pacer tab.

3.

Switch on Pacer Reject .

9 - 6 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

NOTE

When pace pulses are detected, the pace pulse markers “|” are shown on the ECG waveforms. Pacer

Rejection setting has no impact on the display of pace pulse markers “|”.

You can switch on pacer rejection only when Paced is set to Yes. If Paced is set to no, the Pacer Reject function is disabled.

9.5

Changing ECG Settings

9.5.1

9.5.2

Choosing an ECG Screen

When monitoring ECG, several display options are available.

For 3-lead ECG monitoring, only normal screen is available.

For 5-lead ECG monitoring, besides the normal screen, you can also choose Full-screen (7-lead) or Half

Screen (7-lead).

For 6-lead ECG monitoring, besides the normal screen, you can also choose Full-screen (8-lead) or Half

Screen (8-lead).

For 12-lead ECG monitoring, besides the normal screen, you can also choose Full-screen (7-lead), Half

Screen (7-lead), or 12-Lead (12-lead analysis).

To choose the desired screen configuration, follow this procedure:

1.

Select the ECG numeric area or waveform area to enter the ECG dialog.

2.

Select Full-Screen , Half-Screen , or 12-Lead .

To return to the normal screen, press the highlighted button to unselect that screen.

Alternately you can choose the desired screen configuration as follows:

1.

Select the Screen Setup quick key

Choose Screen or select the Main Menu quick key

→ from the

Display column select Choose Screen .

2.

Select ECG Full-Screen , ECG Half-Screen , or ECG 12-Lead .

Setting ECG Alarm Properties

To select the desired ECG alarm options, follow this procedure:

1.

Select the ECG numeric area or waveform area to enter the ECG dialog.

2.

Select the Alarm tab.

3.

Select the desired alarm options.

9.5.3

Changing ECG Wave Settings

9.5.3.1

Selecting the Leads of Displayed ECG Waveforms

To select the desired ECG waveforms to be displayed, follow this procedure:

1.

Select the ECG numeric area or waveform area to enter the ECG dialog.

2.

Select the Setup tab.

3.

Select ECG to set the lead of each displayed ECG waveform.

4.

If more than three ECG waveforms are to be displayed, select the More Leads tab, and then select ECG to set leads of other ECG waveforms.

For optimal ECG analysis, select leads with the following characteristics:

The QRS complex should be either completely above or below the baseline and it should not be biphasic.

The QRS complex should be tall and narrow.

The P waves and T waves should be less than 0.2mV.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 9 - 7

CAUTION

Ensure that you have selected the optimal leads with the best waveform amplitude and the highest signal-to-noise ratio. Selecting the optimal leads is important for detecting beats, classifying beats, and detecting ventricular fibrillation.

9.5.3.2

Changing ECG Waveform Size

If the ECG waveform is too small or clipped, you can change its size by selecting the appropriate Gain setting. To do so, follow this procedure:

1.

Select the ECG numeric area or waveform area to enter the ECG dialog.

2.

Select the Setup tab.

3.

Select ECG Gain to set the size of each ECG waveform.

4.

If more than three ECG waveforms are displayed, select the More Leads tab, and then select ECG Gain to change the sizes of other ECG waveforms. If you select Auto , the monitor automatically adjusts the size of the ECG waveforms for all the leads.

9.5.3.3

Changing Va and Vb Labels

When monitoring ECG with 6-leadwire. You can change the labels of Va and Vb leads. To do so, follow this procedure:

1.

Select the ECG numeric area or waveform area to enter the ECG dialog.

2.

Select the Setup tab.

3.

Set Va and Vb according to the locations of the Va and Vb electrode sites. Default settings are Va and Vb .

9.5.3.4

Changing ECG Waveform Speed

To change ECG waveform speed, follow this procedure:

1.

Select the ECG numeric area or waveform area to enter the ECG dialog.

2.

Select the Setup tab.

3.

Set the desired Speed .

9.5.3.5

Setting the ECG Filter

To set the appropriate ECG waveform filter mode, follow this procedure:

1.

Select the ECG numeric area or waveform area to enter the ECG dialog.

2.

Select the Setup tab.

3.

Set Filter .

◆ Diagnostic : use when diagnostic quality ECG is required. The unfiltered ECG waveform is displayed so that changes such as R-wave notching or discrete elevation or depression of the ST segment are visible.

◆ Monitor : use under normal monitoring conditions.

◆ Surgery : use when the signal is distorted by high frequency or low frequency interference. High frequency interference usually results in large amplitude spikes making the ECG signal look irregular.

Low frequency interference usually leads to wandering or rough baseline. The surgery filter reduces artifacts and interference from electrosurgical units. Under normal measurement conditions, selecting Surgery may suppress certain features or details of the QRS complexes.

◆ ST : recommended for ST monitoring.

9.5.3.6

Switching On or Off the Notch Filter

The notch filter removes the line frequency interference. To switch on or off the notch filter, follow this procedure:

9 - 8 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

1.

Select the ECG numeric area or waveform area to enter the ECG dialog.

2.

Select the Setup tab.

3.

Switch on or off Notch Filter .

NOTE

The notch filter can only be switched on or off when ECG Filter is set to Diagnostic. In other filter modes, the notch filter is always on.

9.5.4

9.5.5

9.5.6

Disabling the Smart Lead Off Function

The monitor provides the smart lead off function. When the lead corresponding to the first ECG wave gets detached but another lead is available, the monitor automatically switches to the available lead to recalculate heart rate, and to analyze and detect arrhythmias. When you reconnect the detached leads, the monitor automatically switches back to the original lead.

The smart lead off function is enabled by default. To disable this function, follow this procedure:

1.

Select the ECG numeric area or waveform area to enter the ECG dialog.

2.

Select the Setup tab.

3.

Switch off Smart Lead .

Adjusting the QRS Volume

To adjust the QRS volume, follow this procedure:

1.

Select the ECG numeric area or waveform area to enter the ECG dialog.

2.

Select the Setup tab.

3.

Set QRS Volume .

When valid SpO

2

measurements are available, the monitor adjusts the pitch of QRS tone based on the SpO

2 value.

Adjusting the Minimum QRS Detection Threshold

To avoid false asystole alarm due to low R wave amplitude, and to avoid tall T waves and P waves being mistaken for QRS complexes, the monitor provides a means to manually adjust the minimum QRS detection threshold.

To adjust the minimum QRS detection threshold, follow this procedure:

1.

Select the ECG numeric area or waveform area to enter the ECG dialog.

2.

Select the Setup tab and set Filter to Monitor .

3.

Select the QRS Threshold tab.

4.

Select up or down arrow buttons to adjust the minimum threshold for QRS detection. Selecting Default resets the QRS threshold to the default value (0.16 mV).

CAUTION

The setting of the QRS detection threshold can affect the sensitivity for arrhythmia, ST, QT/QTc detection, and heart rate calculation.

If the QRS amplitude is low, the monitor might not be able to detect the beats and false asystole calls may occur.

NOTE

The minimum QRS detection threshold can only be adjusted when the ECG filter is set to Monitor.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 9 - 9

9.6

9.6.1

Monitoring Arrhythmia

Arrhythmia monitoring is intended for adult, pediatric, and neonatal patients.

Arrhythmia Safety Information

WARNING

Heart rate reading may be affected by cardiac arrhythmias. Do not rely entirely on heart rate alarms when monitoring patients with arrhythmias. Always keep these patients under close surveillance.

The arrhythmia analysis program may incorrectly identify the presence or absence of an arrhythmia.

Therefore, a physician must analyze the arrhythmia information with other clinical findings.

Pacer not Capture and Pacer not Pacing detection funcitons are not intended for neonatal patients.

CAUTION

Atrial fibrillation (A-Fib) detection function is not intended for pediatric and neonatal patients.

Since the arrhythmia detection algorithm sensitivity and specificity are less than 100%, it is possible for false arrhythmias to be detected and true arrhythmia events to not be detected. The possibility increases when the signal is noisy.

The ECG size and minimum QRS detection threshold settings affect arrhythmia detection and heart rate calculation sensitivity.

If QRS amplitude is low, the monitor might be unable to calculate heart rate and thus create false asystole calls. During the learning phase of the algorithm, arrhythmia detection may not be available. It is recommended to closely monitor patient condition during, and for several minutes after, the learning phase to allow the algorithm to reach optimal detection performance.

9.6.2

Arrhythmia Events

This section lists all arrhythmia events and their criteria.

9.6.2.1

Lethal Arrhythmia Events

Arrhythmia message

Asystole

V-Fib/V-Tach

V-Tach

Description

No QRS complex detected within the set time interval in the absence of ventricular fibrillation or chaotic signal.

A fibrillatory wave for 6 consecutive seconds.

A dominant rhythm of adjacent PVCs and the ventricular rate is greater than the V-tach rate limit.

The number of consecutive PVCs is greater than or equal to the V-Tach PVCs limit, and the ventricular rate is greater than or equal to the V-Tach rate limit.

9.6.2.2

Nonlethal Arrhythmia Events

Arrhythmia message

Vent Brady

Extreme Tachy

Extreme Brady

R on T

Run PVCs

Couplet

Description

The number of consecutive PVCs is greater than or equal to V brady PVC limit and the ventricular rate is less than the V-Brady Rate limit.

The heart rate is greater than the extreme tachycardia limit.

The heart rate is less than the extreme bradycardia limit.

R on T PVC is detected.

More than two consecutive PVCs, but lower than the V-Brady PVCs limit, and the ventricular rate is lower than the V-Tach rate limit.

A Pair of PVCs detected in between normal beats.

9 - 10 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

9.6.3

Arrhythmia message

Multiform PVC

PVC

Bigeminy

Trigeminy

Tachy

Brady

Pacer Not Capture

Pacer Not Pacing

Missed Beat

Nonsus V-Tach

Vent Rhythm

Pause

Irr Rhythm

A-Fib (for adult only)

PVCs/min

Pauses/min

Irr Rhythm End

A-Fib End (for adult only)

*N: normal beat; V: ventricular beat

Description

Multiform PVCs detected in Multif. PVC's Window (which is adjustable).

One PVC detected in between normal beats.

A dominant rhythm of N, V, N, V, N, V * .

A dominant rhythm of N, N, V,N, N, V, N, N, V * .

The heart rate is greater than the tachycardia limit.

The heart rate is lower than the bradycardia limit.

No QRS complex detected for 300 ms following a pace pulse (for paced patients only).

No pace pulse detected for 1.75 x average R-to-R intervals following a QRS complex (for paced patients only).

At least 3 consecutive Ns * , and

The current RR interval is greater than 1.5 x previous RR interval, and

The next RR interval is lower than 1.5 x average RR interval, and

HR lower than 100 and the current RR interval is greater than 1.75 x average RR interval , or

HR is greater than or equal to 100 and the current RR interval is greater than 1000 ms.

The number of consecutive PVCs is lower than the V-Tach PVCs limit but greater than 2, and the ventricular rate is greater than or equal to the V-Tach Rate limit.

The number of consecutive PVCs is greater than or equal to the V-Brady PVCs limit, and ventricular rate is greater than or equal to the V-Brady Rate limit but lower than V-Tach Rate limit.

No QRS complex is detected within the set pause time threshold.

Consistently irregular rhythm (N * , irregular RR interval change is greater than 12.5%)

P wave is absent and normal beat RR intervals are irregular.

PVCs/min exceeds high limit.

Pauses/min exceeds high limit.

Irregular rhythm no longer detected for the irregular rhythm end delay time.

Atrial fibrillation no longer detected for the A-Fib end delay time.

Displaying Arrhythmia Information

You can display the arrhythmia information in the numeric area. To do so, follow this procedure:

1.

Access Tile Layout using either of the following ways:

◆ Select the Screen Setup quick key → select the Tile Layout tab.

◆ Select Main Menu quick key → from the Display column select Tile Layout .

2.

Click the numeric area where you want to display the arrhythmia information, and then select ECG →

Arrhythmia .

9.6.4

Changing Arrhythmia Settings

9.6.4.1

Changing Arrhythmia Alarm Settings

To set the arrhythmia alarm properties, follow this procedure:

1.

Select the ECG numeric area or waveform area to enter the ECG dialog.

2.

Select the Arrhythmia tab

Alarm tab.

3.

Set alarm properties as desired.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 9 - 11

NOTE

You can switch off lethal arrhythmia alarms only when the monitor is set to enableLethal Arrh

Alarms Off. For more information, see 32.3.10.1 Setting the Lethal Arrhythmia Alarms Switch

.

The priority of lethal arrhythmia alarms is always high. It cannot be altered.

9.6.4.2

Changing Arrhythmia Alarm Threshold Settings

You can change threshold settings for certain arrhythmia alarms. When an arrhythmia exceeds its threshold, an alarm will be triggered. To do so, follow this procedure:

1.

Select the ECG numeric area or waveform area to enter the ECG dialog.

2.

Select the Arrhythmia tab

select the Threshold tab.

3.

Enter the password if required. For more information, refer to 32.14.3 Selecting Password for User

Authentication

.

4.

Set the threshold for the desired arrhythmia alarms.

NOTE

The asystole delay time relates to ECG relearning. When heart rate is less than 30 bpm, it is recommended to set Asystole Delay to 10 sec.

9.6.4.3

Arrhythmia Threshold Range

Arrhythmia

Asystole Delay

Tachy (HR High)

Brady (HR Low)

Extreme Tachy

Extreme Brady

Multif PVCs Window

V-Tach Rate

V-Brady Rate

V-Tach PVCs

V-Brady PVCs

PVCs/min

Pauses/min

Pause Threshold

AF/Irr Rhy End Time

Threshold Range

3 sec to 10 sec

60 bpm to 295 bpm

16 bpm to 120 bpm

65 bpm to 300 bpm

15 bpm to 115 bpm

3 beats to 31 beats

100 bpm to 200 bpm

15 bpm to 60 bpm

3 beats to 99 beats

3 beats to 99 beats

1 to 100

1 to 15

1.5 sec, 2.0 sec, 2.5 sec, 3.0 sec

0 min, 1 min, 2 min, 3 min, 4 min, 5 min, 10 min, 15 min, 30 min

9.6.4.4

Setting Thresholds for PVC-Related Alarms

PVC-related alarms are detected on the basis of the current PVC rate and the number of consecutive PVCs.

To set the required thresholds for PVC-related alarms, follow this procedure:

1.

Select the ECG numeric area or waveform area to enter the ECG dialog.

2.

Select the Arrhythmia tab → select the More Threshold tab.

9 - 12 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

3.

Enter the password if required. For more information, refer to 32.14.3 Selecting Password for User

Authentication

.

Adjust V-Tach PVCs , V-Tach Rate , V-Brady PVCs , and V-Brady Rate as necessary to set the thresholds for the desired PVC-related alarms.

The following figure illustrates the conditions under which PVC alarms will be generated if V-Tach PVCs is set to

6, V-Tach Rate is set to 130, V-Brady PVCs is set to 5, and V-Brady Rate is set to 40.

9.6.5

If the number of consecutive PVCs is greater than or equal to the V-Tach PVCs limit (6), and the ventricular rate (Vent Rate) is greater than or equal to the V-Tach Rate limit (130), a V-Tach alarm is generated.

If the number of consecutive PVCs is lower than the V-Tach PVCs limit (6) but greater than 2, and the ventricular rate is greater than or equal to the V-Tach Rate limit (130), a Nonsus V-Tach alarm is generated.

If the number of consecutive PVCs is greater than or equal to the V-Brady PVCs limit (5), and the ventricular rate is lower than the V-Tach Rate limit (130) but greater than or equal to the V Brady Rate limit (40), a Vent

Rhythm alarm is generated.

If the number of consecutive PVCs is lower than the V-Brady PVCs limit (5) but greater than 2, and the ventricular rate is lower than the V-Tach Rate limit (130), a Run PVCs alarm is generated.

If the number of consecutive PVCs is greater than or equal to the V-Brady PVCs limit (5), and the ventricular rate is lower than the V Brady Rate limit (40), a Vent Brady alarm is generated.

Intelligent Arrhythmia Alarm

Normally, an arrhythmia alarm is presented when an alarm condition is detected. However, there are certain situations that can inhibit audible and visible alarm indications even though an alarm condition was detected.

For more information, see 9.6.5.1 Arrhythmia Alarm Chains and 9.6.5.2 Arrhythmia Alarm Shielding Rules .

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 9 - 13

9.6.5.1

Arrhythmia Alarm Chains

If multiple alarms overlap, announcing all of the detected alarm conditions would be confusing, and a more serious condition might be overlooked. So arrhythmia alarms are prioritized by alarm "chains".

Asystole

V Ͳ Fib/V Ͳ Tach

V Ͳ Tach / Vent Brady

PVC Alarm Chain

Nonsus V Ͳ Tach /

Vent Rhythm

Run PVCs

Extreme Tachy/

Extreme Brady

Beat Detection

Alarm Chain

Pauses/min

Rate Alarm Chain

Tachy/Brady

Pause

HR High /

HR Low

Couplet

Pacer not Capture /

Pacer not Pacing /

Missed Beat

R on T

A Ͳ Fib Alarm Chain

A Ͳ Fib/

A Ͳ Fib End

Irr. Rhythm/

Irr.Rhythm End

Bigeminy

Trigeminy

PVCs/min

Multiform PVC

PVC

9.6.5.2

Arrhythmia Alarm Shielding Rules

The following table explains how audible and visual alarms are indicated during arrhythmia alarm shielding period .

Previous alarm

Alarm in high priority chain

Alarm in medium priority chain

Current alarm

Alarm in high priority chain

Alarm in medium priority chain

Alarm indication

Alarm light and alarm tone

During the shielding period, alarm light and alarm tone are disabled. When the shielding period is reached, alarm light and alarm tone are reactivated.

Alarm light and alarm tone

Alarm light and alarm tone

Alarm in high priority chain

Alarm in the same medium priority chain, but with higher priority

The same alarm reoccurs

Alarm in the same medium priority chain, but with lower priority

Alarm in other medium priority chain

During the shielding period, alarm light and alarm tone are disabled. When the shielding period is reached, alarm light and alarm tone are reactivated.

During the shielding period, alarm light and alarm tone are disabled. When the shielding period is reached, alarm light and alarm tone are reactivated.

Alarm light and alarm tone

9 - 14 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

9.6.5.3

Arrhythmia Alarm Refractory Period

For the following eight alarms in medium priority chain, there is a 30-second refractory period. That is to say, if the same alarm is not presented within 30 seconds after the previous alarm disappears.

Couplet

R on T

PVCs/min High

Multiform PVCs

PVC

Missed Beat

Pacer not Capture

Pacer not Pacing

9.7

ST Segment Monitoring

ST segment analysis is intended for adult, pediatric and neonatal patients. To use this feature, the MPM must be labeled to support ST analysis, see

9.1 ECG Introduction .

9.7.1

ST Safety Information

WARNING

ST segment deviation values may be affected by such factors as certain drugs or metabolic and conduction disturbances.

ST deviation is often calculated at a fixed offset from the J point. Changes in heart rate may affect ST.

The significance of ST segment changes needs to be determined by a physician.

This monitor provides information regarding changes in ST deviation levels. The clinical significance of the ST level change information should be determined by a physician.

9.7.2

9.7.3

Enabling ST Monitoring

The ST monitoring function is disabled by default. Before you start ST monitoring, enable the ST function. To do so, follow this procedure:

1.

Select the ECG numeric area or waveform area to enter the ECG dialog.

2.

Select the ST tab → select the Setup tab.

3.

Switch on ST Analysis .

Reliable ST monitoring cannot be ensured under the following situations:

You are unable to get a lead that is not noisy.

Arrhythmias, such as atrial fibrillation or atrial flutter, which may cause an irregular baseline.

The patient is continuously ventricularly paced.

The patient has left bundle branch block.

In these cases, you may consider switching off ST monitoring.

Displaying ST Deviation Numerics

To display ST numerics, follow this procedure:

1.

Access Tile Layout by either of the following ways:

◆ Select the Screen Setup quick key → select the Tile Layout tab.

◆ Select Main Menu quick key → from the Display column select Tile Layout .

2.

Click the numeric area where you want to display the ST numerics, and then select ECG → ST .

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 9 - 15

The ST parameter display area is configured differently according to the ECG cable used:

When you are using the 3-lead ECG leadwires, a separate ST numeric area does not appear on the display.

The ST deviation value displays in the ECG numeric area.

When you are using the 5-lead ECG leadwires, the ST numeric area displays 7 ST deviation values: ST-I, ST-II,

ST-III, ST-aVR, ST-aVL, ST-aVF, ST-V.

When you are using the 6-lead ECG leadwires, the ST numeric area displays 8 ST deviation values: ST-I, ST-II,

ST-III, ST-aVR, ST-aVL, ST-aVF, ST-Va, ST-Vb.

When you are using the 12-lead ECG leadwires, the ST numeric area displays12 ST deviation values: ST-I, ST-

II, ST-III, ST-aVR, ST-aVL, ST-aVF, ST-V1, ST-V2, ST-V3, ST-V4, ST-V5, ST-V6.

This example shows the ST numeric area when 5-lead ECG cable is used. The monitor screen may look slightly different:

(4) (5)

(1)

(2)

(3 )

9.7.4

(1) Parameter label

(3) ST alarm off symbol

(2) ST measurement unit

(4) Lead labels

(5) ST numerics: a positive value indicates ST segment elevation, and a negative value indicates ST segment depression.

Displaying ST Segments in the Waveform Area

You can display ST segments in the waveform area. To do so, follow this procedure:

1.

Access Tile Layout by either of the following ways:

◆ Select the Screen Setup quick key → select the Tile Layout tab.

◆ Select Main Menu quick key → from the Display column select Tile Layout .

2.

Select the waveform area where you want to display the ST segments, and then select ECG → ST .

9 - 16 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

ST segment shows a QRS complex segment for each measured ST lead. The current ST segment is drawn in the same color as the ECG wave, usually green, superimposed over the stored reference segment, drawn in a different color. The information is updated once every ten seconds.

(1) (2)

(3 )

9.7.5

9.7.6

(1) ST lead

(3) Baseline ST deviation value

(4) (5)

(2) Current ST deviation value

(4) 1 mV scale

(5) Current ST segment (green) and baseline ST segment (white)

Entering the ST View

In ST View mode a complete QRS segment for each ST lead is displayed. The color of the current ST segments and ST deviation values is consistent with the color of the ECG waveforms, normally green. The color of the baseline ST segments and ST deviation values is white.

To enter the ST View, follow this procedure:

1.

Select the ECG numeric area or waveform area to enter the ECG dialog.

2.

Select the ST tab.

3.

From the bottom of the dialog, select ST View .

Saving the Current ST Deviation and Segment as Baseline

ST deviation is typically monitored as a relative change from a baseline value. Set an ST baseline when ST deviation values become stable. If you did not set the ST baseline, the monitor automatically saves the baseline when valid ST deviation values appear for 5 minutes. To set the ST baseline, follow this procedure:

1.

From the ST View , select Set Baseline .

2.

From the pop-up dialog box, select Ok to set the current ST segments and values as the baseline.

From the ST View , you can also perform the following operations:

Display or hide ST baseline by selecting Display Baseline or Hide Baseline

Display or hide the position of ISO point, J point and ST point by selecting

.

Display Marker or Hide Marker .

CAUTION

Updating ST baseline affects ST alarms.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 9 - 17

9.7.7

Entering the ST Graphic

To display ST Graphic , follow this procedure:

1.

Select the ECG numeric area or waveform area to enter the ECG dialog.

2.

Select the ST tab.

3.

Select ST Graphic .

The following figure shows the ST Graphic when ST Alarm Mode is set to Absolute . The height of the bar indicates the ST deviation value of corresponding ST lead. The color of the bar indicates ST alarm status: green indicates that corresponding ST deviation value is within alarm limits; cyan, yellow and red indicate that the ST deviation value exceeds the alarm limits. The color matches ST alarm priority.

The following figure shows the ST Graphic when ST Alarm Mode is set to Relative . The height of grey bar indicates the baseline ST deviation value and the green bar (cyan, yellow or red if an alarm occurs) indicates relative ST from that baseline.

9.7.8

Changing ST Settings

9.7.8.1

Setting ST Alarm Properties

To set ST alarm properties, follow this procedure:

1.

Select the ECG numeric area or waveform area to enter the ECG dialog.

2.

Select the ST tab → Alarm tab.

3.

Set ST Alarm Mode to Absolute or Relative .

◆ Absolute : you can separately set the alarm properties for each ST alarm for each lead.

◆ Relative : you can set the alarm properties for ST Single and ST Dual alarms.

4.

Set ST alarm properties.

9 - 18 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

9.7.8.2

Changing Leads for ST Display

The monitor automatically selects the three most deviated leads for ST display. You can also manually select the leads. To do so, follow this procedure:

1.

Select the ECG numeric area or waveform area to enter the ECG dialog.

2.

Select the ST tab → select the Setup tab.

3.

Set ST Segment . You can select up to 3 leads.

9.7.8.3

Showing ISO Point, J Point, and ST Point Markers

In the waveform area, the ISO point, J point, and ST point marker do not display on the ST segments by default.

To display these markers, follow this procedure:

1.

Select the ECG numeric area or waveform area to enter the ECG dialog.

2.

Select the ST tab

select the Setup tab.

3.

Switch on Show Markers .

9.7.9

Adjusting ST Measurement Points

9.7.9.1

About ST Point, ISO Point, and J Point

The ST deviation value for each beat is the potential difference between the isoelectric (ISO) point and the ST point. The ISO point provides the baseline. The ST point is located between the J point and the start of the Twave. The J point is the end of the QRS complex. As the ST point is at a fixed distance away from the J point (40,

60, 80 etc msec), manually adjusting the J point helps you correctly position the ST point.

R

P

(1) (2) (3) T (4)

Q

(1) ISO point (2) J point

S

(3) ST point (4) ST deviation

9.7.9.2

Setting ST Point, ISO Point, and J Point

CAUTION

If Auto Adjust is Off, or if the patient's heart rate or ECG morphology changes significantly, the ISO and ST points need to be occasionally adjusted because this may affect the QT interval and thus the location of the ST point. False ST segment depression or elevation may occur if the isoelectric point or the ST point is incorrectly set.

Always ensure that the locations of ST points are appropriate for the patient.

To set the ST point, ISO point, and J point, follow this procedure:

1.

Select the ECG numeric area or waveform area to enter the ECG dialog.

2.

Select the ST tab → select the Adjust tab.

3.

Set ST Point .

The setting for Auto Adjust defines the method of adjusting the ISO and J point locations. Auto Adjust is enabled by default. In this case, the locations of the ISO point and J point are automatically determined by the

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 9 - 19

9.8

9.8.1

9.8.2

algorithm. When Auto Adjust is disabled, you need to manually adjust the position of the ISO point and J point using the arrows at the right sides of ISO and J .

The ISO point (isoelectric) location shown is relative to the R-wave peak. Position the ISO point in the middle of the flattest part of the waveform (between the P and Q waves).

The J point location is indicated relative to the R-wave peak. Position the J point at the end of the QRS complex and the beginning of the ST segment.

The ST point is positioned a fixed distance from the J point. Position the ST point relative to the J point at

J+60/80ms , J+40ms , J+60ms or J+80ms . When J+60/80ms is selected, the ST point will be positioned 80 ms (heart rate 120 bpm or less) or 60 ms (heart rate more than 120 bpm) from the J point.

QT/QTc Interval Monitoring

The QT interval is the time between the beginning of the Q-wave and the end of the T-wave. It represents the total duration of the depolarization (QRS duration) and repolarization (ST-T) phases of the ventricles. QT interval monitoring can assist in the detection of long QT syndrome.

The QT interval has an inverse relationship to heart rate. Faster heart rates shorten QT interval and slower heart rates prolong QT interval. Therefore, several formulas are commonly used to correct the QT interval for heart rate. The heart rate corrected QT interval is abbreviated as QTc.

QT/QTc interval monitoring is intended for adult, pediatric, and neonatal patients.

QT/QTc Monitoring Limitations

Some conditions may make it difficult to achieve reliable QT/QTc monitoring, for example:

R-wave amplitudes are too low

The presence of frequent ventricular ectopic beats

RR intervals are unstable

P-waves tending to encroach on the end of the previous T-wave at high heart rates

The T-wave is very flat or T-wave are not well defined

The end of the T-wave is difficult to delineate because of the presence of U-waves

QTc measurements are not stable

In the presence of noise, asystole, ventricular fibrillation, atrial fibrillation, or ECG lead off

For reliable QT/QTc monitoring, choose lead with good T-wave amplitude and no visible flutter activity, and without a predominant U-wave or P-wave.

Some conditions such as left or right bundle branch block or ventricular hypertrophy can lead to a widened QRS complex. If a long QTc is observed you should verify it to ensure that it is not caused by QRS widening.

Because normal beats followed by ventricular beats are not included in the analysis, no QT measurement will be generated in the presence of a bigeminy rhythm.

If the heart rate is extremely high (over 150 bpm for adults and over 180 bpm for pediatrics and neonates), QT will not be measured. When the heart rate changes, it can take several minutes for the QT interval to stabilize. For reliable QTc calculation it is important to avoid measurements when the heart rate is changing.

Enabling QT/QTc Monitoring

The QT monitoring function is disabled by default. Before you start QT monitoring, enable the QT function. To do so, follow this procedure:

1.

Select the ECG numeric area or waveform area to enter the ECG dialog.

2.

Select the QT tab → select the Setup tab.

3.

Switch on QT Analysis .

9 - 20 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

9.8.3

9.8.4

Displaying QT Numerics and Segments

To display QT numerics and Segments, follow this procedure:

1.

Access Tile Layout by either of the following ways:

◆ Select the Screen Setup quick key

select the Tile Layout tab.

Select Main Menu quick key

→ from the Display column select Tile Layout .

2.

Click the parameter numeric area where you want to display the QT numerics, and then select ECG

QT/

QTc .

NOTE

QTc values are calculated based on the QT-HR, not the ECG HR. To view the QT-HR, open the QT View.

For more information, see

9.8.4 Entering the QT View .

The following picture shows the QT numeric area. The monitor screen may look slightly different:

(3)

(2)

(4)

(5)

(1 )

(6)

(1) QTc alarm limit (if QTc alarm is off, the alarm off symbol is displayed)

(2) Parameter label (3) QTc value

(4) ΔQTc value (the difference between the current and baseline QTc values)

(5) QT value (6) QT-HR value

Entering

the

QT View

QT View shows the current and baseline QT parameter values and waveforms. To enter QT View, follow this procedure:

1.

Select the ECG numeric area or waveform area to enter the ECG dialog.

2.

Select the QT tab.

3.

Select QT View .

The following picture shows a typical QT view.

The current waveform is shown in the upper half in green.

The baseline waveform is shown below in white.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 9 - 21

9.8.5

■ The waveform of selected lead is highlighted, while the waveforms of other leads are in light green or light gray.

The start of the QRS complex and the end of the T wave are marked with a vertical line.

In some conditions, no QT measurement can be made. Then the cause of failed QT measurement is shown at the bottom of the QT numerics area and the message “Cannot Analyze QT” is shown in the technical alarm area.

Select the left or right arrow to switch leads, and the corresponding waveform will be highlighted.

Saving the Current QTc as Baseline

In order to quantify changes in the QTc value, you can set a QTc baseline. If no baseline has been set for this patient within the first five minutes after getting valid QT values, the monitor will automatically set a baseline. To set the current values as baseline, follow this procedure:

1.

From the QT View , select Set Baseline .

2.

From the pop-up dialog box, select Ok . This baseline will then be used to calculate ΔQTc.

If you set a new baseline the previous baseline is discarded.

From the QT View , you can also perform the following operations:

■ Select the left or right arrow to select a lead label to highlight corresponding waveform.

■ Select Display Baseline or Hide Baseline to display or hide the baseline waveform.

CAUTION

Updating QTc baseline affects the ΔQTc value and alarm.

9.8.6

Changing QT Settings

9.8.6.1

Setting QT Alarm Properties

To set QT alarm properties, follow this procedure:

1.

Select the ECG numeric area or waveform area to enter the ECG dialog.

2.

Select the QT tab

select the Alarm tab.

3.

Set QTc and ΔQTc alarm properties.

9.8.6.2

Selecting Leads for QT Calculation

You can select one lead or all leads for QT calculation. To do so, follow this procedure:

1.

Select the ECG numeric area or waveform area to enter the ECG dialog.

2.

Select the QT tab → select the Setup tab.

3.

Set QT Leads . All is selected by default. This means all leads are used for QT calculation.

9.9

9.9.1

ECG Relearning

Changes in ECG beat morphology could result in incorrect arrhythmia alarms and/or inaccurate heart rate. ECG relearning allows the monitor to learn the new dominant beat morphology and thus reduce false arrhythmia alarms and inaccurate HR values. Once learning is complete, the dominant QRS complex is stored as a reference template. The reference template is used as the normal morphology of that patient and it is compared with incoming beats to identify possible arrhythmias.

Auto ECG Relearning

Auto arrhythmia relearning happens in the following situation:

■ The ECG lead type or lead label is changed.

9 - 22 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

9.9.2

ECG leads are off and are not reconnected within 60 seconds.

The patient‘s paced status is changed.

Initiating an ECG Relearning Manually

If you suspect that false arrhythmia alarms are presented, you may need to manually initiate an ECG relearn. To do so, follow this procedure:

1.

Select the ECG numeric area or waveform area to enter the ECG dialog.

2.

Select Relearn .

CAUTION

Initiate ECG relearning only during periods of predominantly normal rhythm and when the ECG signal is relatively noise-free. If ECG learning takes place during arrhythmias, the ectopic beats may be incorrectly learned as normal QRS complexes. This may result in missed detection of subsequent arrhythmia events.

9.10

Defibrillation Synchronization Pulse Output

The 12-lead MPM module provides an analog out connector to output synchronization pulses for defibrillators. If a defibrillator is connected, it receives a synchronization pulse (100 ms, +5 V) through the analog out connector each time an R-wave is detected. The 3/5/6 lead MPM module does not offer this output.

WARNING

According to IEC 60601-2-4: 2010, the peak of the synchronized defibrillator discharge should be delivered within 60 ms of the peak of the R wave. The signal at the ECG output (sync pulse) on the monitor is delayed by maximum of 30 ms. the biomedical engineer should verify that the ECG/

Defibrillator combination does not exceed recommended maximum delay of 60 ms.

Before defibrillation, ensure that both defibrillator and monitor have passed the system test and can be safely used together.

9.11

ECG Troubleshooting

This section lists some of the problems that might be encountered during ECG monitoring. If you encounter problems when using the monitor or accessories, check the table below before requesting for services. If the problem persists after you have taken corrective actions, contact the service personnel.

NOTE

For the physiological and technical alarm messages, see

D Alarm Messages

.

Problem

Noisy ECG traces

Excessive electrosurgical Interference

Muscle Noise

Corrective Actions

1. Check that the electrodes are not detached or dry. Replace with fresh and moist electrodes if necessary.

2. Check that leadwires are not defective. Replace leadwires if necessary.

3. Check that the patient cable or leadwires are not routed too close to other electrical devices. Move the patient cable or leadwires away from electrical devices and their cables if necessary.

Use ESU-proof ECG cables.

For more information, see

36.1 ECG Accessories .

Inadequate skin preparation, tremors, tense subject, and/or poor electrode placement.

1. Perform skin preparation again and re-place the electrodes. For more

information, see 9.4.1 Preparing the Patient Skin and 9.4.2 Applying Electrodes and Connecting the Patient

.

2. Apply fresh, moist electrodes. Avoid muscular areas.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 9 - 23

Problem

Intermittent Signal

Excessive alarms: heart rate, lead fault

Low Amplitude ECG Signal

No ECG Waveform

Base Line Wander

Corrective Actions

1. Check that cables are properly connected.

2. Check that electrodes are not detached or dry. Perform skin preparation again as described in

9.4.1 Preparing the Patient Skin

and apply fresh and moist electrodes.

3. Check that the patient cable or leadwires are not damaged. Change them if necessary.

1. Check that electrodes are not dry. Perform skin preparation again and re-place the electrodes. For more information, see

9.4.1 Preparing the

Patient Skin and 9.4.2 Applying Electrodes and Connecting the Patient

.

2. Check for excessive patient movement or muscle tremor. Reposition the electrodes. Replace with fresh and moist electrodes if necessary.

1. Check that the ECG gain is not set too low. Adjust the gain as required.

For more information, see 9.5.3.2 Changing ECG Waveform Size .

2. Perform skin preparation again and re-place the electrodes. For more information, see

9.4.1 Preparing the Patient Skin

and

9.4.2 Applying

Electrodes and Connecting the Patient

.

3. Check electrode application sites. Avoid bone or muscular area.

4. Check that electrodes are not dry or used for a prolonged time. Replace with fresh and moist electrodes if necessary.

1. Check that the ECG gain is not set too low. Adjust the gain as required.

For more information, see 9.5.3 Changing ECG Wave Settings .

2. Check that the leadwires and patient cables are properly connected.

3. Change cable and lead wires if necessary.

4. Check that the patient cable or leadwires are not damaged. Change them if necessary.

1. Check for excessive patient movement or muscle tremor. Secure leadwires and cable.

2. Check that electrodes are not detached or dry and replace with fresh and

moist electrodes if necessary. For more information, see 9.4.1 Preparing the Patient Skin and 9.4.2 Applying Electrodes and Connecting the Patient

.

3. Check for ECG filter setting. Set ECG Filter mode to Monitor to reduce baseline wander on the display.

9 - 24 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

10

Resting 12-Lead ECG Analysis

10.1

Resting 12-Lead ECG Analysis Introduction

The monitor can be configured with the Glasgow 12-lead resting ECG analysis algorithm. The MPM module or N1 incorporating the Glasgow algorithm is labelled with the logo of Glasgow as shown in

9.1 ECG Introduction .

The Glasgow algorithm is intended for adult, pediatric, and neonatal patients.

10.2

Entering the 12-Lead Screen

To enter the 12-Lead screen, follow this procedure:

1.

Select the ECG numeric area or waveform area to enter the ECG dialog.

2.

From the bottom of the ECG dialog, select 12-Lead .

You can also enter the 12-Lead screen by following this procedure:

Select the Screen Setup quick key

select the Choose Screen tab

select ECG 12-Lead .

Select Main Menu quick key

→ from the Display column select Choose Screen

select ECG 12-Lead .

NOTE

The ECG 12-Lead screen is available only when the Glasgow algorithm is configured (associated module has the Glasgow label) and Lead Set is set to Auto or 12-Lead.

10.3

Initiating Resting 12-Lead ECG Analysis

Before initiating 12-lead ECG analysis, check that all electrodes are correctly connected to the lead wires and the

ECG trunk cable is properly connected. Check that patient information is correct. Keep the patient still.

To initiate 12-Lead ECG analysis, select Analyze from the bottom left of the 12-Lead screen.

10.4

Changing 12-Lead ECG Analysis Settings

On the ECG 12-Lead screen, you can set the high frequency filter, baseline drift removal (BDR) filter, and the waveform layout.

10.4.1

Setting the High Frequency Filter

The high frequency filter attenuates muscle artifact by restricting the included frequencies. The setting of the high frequency filter is 35 Hz by default. To change the setting, follow this procedure:

1.

On the ECG 12-Lead screen, select the ECG numeric area or waveform area to enter the ECG dialog.

2.

Select the Setup tab.

3.

Set High-Freq Cut-off .

The high frequency filter is a low-pass filter. That is to say, signal frequencies higher than the set frequency are filtered out. For example, if you set High-freq Cutoff to 35 Hz , then signal below 35 Hz is kept in the signal and signal exceeding 35 Hz is attenuated.

10.4.2

Setting the Baseline Drift Removal Filter

The baseline drift removal (BDR) filter suppresses most baseline drift interference and also is able to preserve the fidelity of the ST-segment level. BDR is switched on by default. To set the BDR, follow this procedure:

1.

On the ECG 12-Lead screen, select the ECG numeric area or waveform area to enter the ECG dialog.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 10 - 1

2.

Select the Setup tab.

3.

Switch on or off Baseline Drift Removal as desired. If BDR is switched off, the 0.05 Hz high pass filter is used.

NOTE

BDR introduces around 1-second delay. Using BDR is recommended except when the delay is unacceptable.

10.4.3

Setting the 12-Lead Waveform Layout

To set the 12-lead waveform layout, follow this procedure:

1.

On the ECG 12-Lead screen, select the ECG numeric area or waveform area to enter the ECG dialog.

2.

Select the Setup tab.

3.

Set Waveform Layout .

◆ Standard : the waveform sequence is I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6.

◆ Cabrera : the waveform sequence is aVL, I, -aVR, II, aVF, III, V1, V2, V3, V4, V5, V6.

For the Glasgow algorithm, the sequence of the chest leads depends on the setting of V3 Placement . If V3 placement is set to V4R , the sequence of chest leads is V4R, V1, V2, V4, V5, V6.

10.4.4

Editing Patient Information

Some patient information may directly affect ECG analysis. Complete and correct patient information is helpful for accurate diagnosis and treatment of the patient. Enter patient information before starting a resting ECG analysis.

To enter patient information, follow this procedure:

1.

On the ECG 12-Lead screen, select Setup to enter the 12-Lead Setup dialog.

2.

On the Patient Demographics page, input or edit patient information.

NOTE

Saving information will also save the information as though it was entered in the Patient

Management dialog.

Changes are not saved until you press Save. So changing to another tab without saving and then pressing the exit button “X” will lose any changes.

Check that patient information is correct before starting a resting 12-lead analysis.

Using pediatric lead placement V4R, V1, V2, V4 - V6 is recommended if the patient is under 16 years of age. Please record V4R using the V3 electrode. Also set V3 Electrode to V4R. This is a normal practice for a patient of this age.

10.4.5

Setting Tachycardia and Bradycardia Thresholds

To set tachycardia and bradycardia thresholds, follow this procedure:

1.

On the ECG 12-Lead screen, select Setup to enter the 12-Lead Setup dialog.

2.

Select the Setup tab.

3.

Set Tachy and Brady .

NOTE

The tachycardia threshold only applies to patients whose age exceeds 180 days.

The bradycardia threshold only applies to patients whose age exceeds 2191 days (6 years).

10 - 2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

10.4.6

Setting the 12-Lead Interpretation Report

To set the 12-lead interpretation report, follow this procedure:

1.

On the ECG 12-Lead screen, select Setup to enter the 12-Lead Setup dialog.

2.

Select the Report tab.

3.

Set the format and items included in the 12-lead interpretation report.

10.5

Saving the 12-Lead Interpretation Report

At the completion of 12-lead ECG interpretation, select Save to save the report. You can review the saved 12-

lead interpretation reports. For more information, see 25.2.11 12-Lead ECG Review Page .

10.6

Printing the 12-Lead Interpretation Report

At the completion of 12-lead ECG interpretation, select Print or Record to output the report via the printer or recorder.

10.7

Exiting the ECG 12-Lead Screen

To exit the ECG 12-Lead screen, select Exit on the ECG 12-Lead screen.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 10 - 3

This page intentionally left blank.

10 - 4 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

11

Monitoring Impedance Respiration (Resp)

11.1

Resp Introduction

Impedance respiration is measured across the thorax. When the patient is breathing or ventilated, the volume of air changes in the lungs, resulting in impedance changes between the electrodes. Impedance respiration rate

(RR) is calculated from the signal representing these impedance changes, and a respiration waveform is displayed on the patient monitor screen.

Impedance respiration monitoring is intended for adult, pediatric and neonatal patients.

Respiration may also be measured using a CO2 module or anesthetic gas (AG) module. Refer to 19 Monitoring

Carbon Dioxide (CO

2)

and 20 Monitoring Anesthetic Gas (AG) for respiration information specific to those modules.

11.2

Impedance Resp Safety Information

WARNING

When monitoring the patient’s impedance respiration, do not use ESU-proof ECG cables. The monitor cannot measure impedance respiration with ESU-proof ECG cables.

If you do not set the detection level for the respiration correctly in manual detection mode, it may not be possible for the monitor to detect zero respiration rate. If you set the detection level too low, the monitor is more likely to detect cardiac activity, and to falsely interpret cardiac activity as respiratory activity in case of zero respiration rate.

The respiration measurement does not recognize the cause of zero respiration rates. It only indicates an alarm if a pre-adjusted time has elapsed without a breath being detected since the last detected breath. Therefore, it cannot be used for diagnostic purpose.

If operating under conditions according to the EMC Standard IEC 60601-1-2 (Radiated Immunity 3V/ m), field strengths above 3V/m may cause erroneous measurements at various frequencies.

Therefore, it is recommended to avoid the use of electrically radiating equipment in close proximity to the impedance respiration measurement unit.

The current used to measure impedance respiration may cause rate changes in Minute Ventilation

Rate Responsive Pacemakers. If you suspect the patient monitor is affecting the pacemaker rate then replace the ECG cable with an ECG cable that contains ESU filters.

CAUTION

Only use parts and accessories specified in this manual.

Impedance respiration monitoring is not for use on patients while they are active, as this can cause false alarms.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 11 - 1

11.3

Resp Display

(1) (2) (3) (4) (5)

(1) Resp waveform gain

(4) Respiration rate (RR)

(2) Resp lead label

(5) RR source

(3) Alarm limits

NOTE

If ESU-proof ECG cables are used, the Resp waveform area will display the message “Check Leads”.

Replace the ECG cable if necessary.

11.4

Preparing for Impedance Resp Monitoring

11.4.1

Preparing the Patient

Follow this procedure to prepare the patient:

1.

Shave hair from skin at chosen sites.

2.

Gently rub skin surface at sites to remove dead skin cells.

3.

Thoroughly cleanse the site with a mild soap and water solution.

4.

Dry the skin completely before applying the electrodes.

CAUTION

Proper skin preparation is necessary for good signal quality at the electrode site, as the skin is a poor conductor of electricity.

11 - 2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

11.4.2

Placing the Electrodes

Impedance respiration is acquired using standard ECG electrodes and cables. Either lead I (RA-LA) or lead II (RA-

LL) can be used.

For more information, see 9.4.4 ECG Electrode Placements .

(1) (2)

(1) Lead I (2) Lead II

CAUTION

Correct electrode placement can help to reduce interference from cardiac related impedance changes: avoid including the liver area and the ventricles of the heart between the respiratory electrodes. This is particularly important for neonates.

Some patients with restricted movements breathe mainly abdominally. In these cases, you may need to place the left leg electrode on the left abdomen at the point of maximum abdominal expansion to optimize the respiratory waveform.

Some patients (especially neonates) expand their chests laterally, causing a negative intrathoracic pressure. In these cases, it is better to place the two respiration electrodes in the right midaxillary and the left lateral chest areas at the patient’s maximum point of the breathing movement to optimize the impedance respiration waveform.

To optimize the impedance respiration waveform, place the RA and LA electrodes horizontally when monitoring respiration with ECG Lead I; place the RA and LL electrodes diagonally when monitoring respiration with ECG Lead II.

Periodically inspect electrode application sites to ensure skin integrity. If the skin quality changes, replace the electrodes or change the application site.

NOTE

Store the electrodes at room temperature. Open the electrode package immediately prior to use.

Check that the electrode packages are intact and that the electrodes are not past the expiration date. Ensure the electrode gel is moist.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 11 - 3

11.5

Changing Resp Settings

11.5.1

Setting the Resp Alarm Properties

To set the Resp alarm properties, follow this procedure:

1.

Select the Resp numeric area or waveform area to enter the Resp dialog.

2.

Select the Alarm tab.

3.

Enter the password if required. For more information, refer to

32.14.3 Selecting Password for User Authentication .

4.

Set alarm properties as desired.

11.5.2

Setting the RR Source

To set RR source, follow this procedure:

1.

Select the Resp numeric area or waveform area to enter the Resp dialog.

2.

Select the Setup tab.

3.

Choose RR Source from the dropdown list.

When you select Auto , the monitor automatically selects the RR source following order of priority: first CO2 , and then ECG . When the manually selected RR source is not available, the monitor automatically switches the RR

Source to Auto .

11.5.3

Choosing the Respiration Lead

To set the respiration lead, follow this procedure:

1.

Select the Resp numeric area or waveform area to enter the Resp dialog.

2.

Select the Setup tab.

3.

Set Resp Lead .

If you cannot get an acceptable impedance Resp waveform or you suspect the veracity of the Resp value after choosing the Resp lead, you may need to adjust the electrode placement.

11.5.4

Setting the Resp Waveform Size

To set the Resp waveform size, follow this procedure:

1.

Select the Resp numeric area or waveform area to enter the Resp dialog.

2.

Select the Setup tab.

3.

Set Gain .

11.5.5

Setting the Resp Waveform Speed

To set the Resp waveform speed, follow this procedure:

1.

Select the Resp numeric area or waveform area to enter the Resp dialog.

2.

Select the Setup tab.

3.

Set Speed .

11 - 4 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

11.5.6

Setting the Auto Detection Switch

To set the auto detection switch, follow this procedure:

1.

Select the Resp numeric area or waveform area to enter the Resp dialog.

2.

Select the Setup tab.

3.

Switch on or off Auto Threshold Detection .

◆ If Auto Threshold Detection is switched on, the monitor automatically adjusts the Resp waveform detection level, or threshold.

◆ If Auto Threshold Detection is switched off, you have to manually adjust the Resp waveform threshold. For more information, see

11.5.7 Adjusting the Resp Waveform Detection Threshold .

In the auto threshold detection mode, if you are monitoring Resp and the ECG parameter is switched off, the monitor cannot compare the heart rate and Resp rates to detect cardiovascular artifact. The respiration detection level is automatically set higher to prevent the detection of cardiovascular artifact as respiration.

In the manual detection mode (where auto threshold detection is off ), cardiovascular artifact can, in certain situations, trigger the respiration rate counter. This may lead to a false indication of a high respiration or an undetected zero respiration rate condition. If you suspect that cardiovascular artifact is being registered as respiratory activity, raise the detection level above the zone of cardiovascular artifact. If the Resp wave is so small that raising the detection level is not possible, you may need to optimize the electrode placement.

11.5.7

Adjusting the Resp Waveform Detection Threshold

Use the manual detection mode (where auto threshold detection is off ) in the following situations:

The respiration rate and the heart rate are close.

Patients have intermittent mandatory ventilation.

Respiration is weak. Try repositioning the electrodes to improve the signal.

To set the Resp waveform threshold to the desired level, follow this procedure:

1.

Select the Resp numeric area or waveform area to enter the Resp dialog.

2.

Select the Threshold tab.

3.

Select the up and down arrows below Upper Line and Lower Line to define the Resp waveform threshold.

Once set, the detection level will not adapt automatically to different respiration depths. It is important to remember that if the depth of breathing changes, you may need to change the detection level.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 11 - 5

This page intentionally left blank.

11 - 6 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

12

Monitoring Pulse Oxygen Saturation (SpO

2

)

12.1

SpO

2

Introduction

Pulse Oxygen Saturation (SpO

2

) monitoring is a non-invasive technique used to measure the amount of oxygenated hemoglobin and pulse rate by measuring the absorption of selected wavelengths of light. The light generated in the emitter side of the probe is partly absorbed when it passes through the monitored tissue. The amount of transmitted light is detected in the detector side of the probe. When the pulsative part of the light signal is examined, the amount of light absorbed by the hemoglobin is measured and the pulse oxygen saturation can be calculated. This device is calibrated to display functional oxygen saturation.

SpO

2

monitoring is intended for adult, pediatric and neonatal patients.

You can simultaneously measure SpO

2

from two different measurement sources: the MPM module and the SpO

2 module. The measurement from the MPM module is labelled as SpO

2

and the measurement from the SpO

2 module is labelled as SpO

2 b.

The monitor can be configured with Masimo SpO

2 or Nellcor SpO

2

:

Masimo SpO

2

: the connector is purple and the logo of Masimo SET is on the module’s front panel.

Nellcor SpO

2

: the connector is grey and the logo of Nellcor is on the module’s front panel.

Platinum MP1 Platinum MP1

ECG

T1

T2

SpO

2

IBP2

NIBP

ECG

T1

T2

SpO

2

IBP2

NIBP

(1) (2)

(1) Connector of Masimo SpO

2

(2) Connector of Nellcor SpO

2

NOTE

If you need to measure SpO

2

by both the MPM module and the SpO

2

module, select the same type of

SpO

2

. Otherwise, the SpO

2

module will be disabled. For example, if an MPM module configured with the Nellcor SpO

2

and an SpO

2

module configured with the Masimo SpO

2

are simultaneously applied, the SpO

2

module will be automatically disabled.

The SpO

2

extension cable used must be compatible with the SpO

2 sensor connectors used. For example, only the Masimo SpO

2

extension cable can be connected to the Masimo SpO

2

sensor connectors.

Measurement accuracy verification: The SpO

2

accuracy has been verified in human experiments by comparing with arterial blood sample reference measured with a CO-oximeter. Pulse oximeter measurement are statistically distributed and about two-thirds of the measurements are expected to come within the specified accuracy range compared to CO-oximeter measurements.

A functional tester or SpO

A functional tester or SpO

2

simulator can be used to determine the pulse rate accuracy.

2

simulator cannot be used to assess the SpO

2

accuracy.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 12 - 1

12.2

SpO

2

Safety Information

WARNING

When a trend toward patient deoxygenation is indicated, analyze the blood samples with a laboratory co-oximeter to completely understand the patient’s condition.

Do not use SpO

2

sensors during magnetic resonance imaging (MRI). Induced current could potentially causes burns. The sensor may affect the MRI image, and the MRI unit may affect the accuracy of the oximetry measurements.

Prolonged continuous monitoring may increase the risk of undesirable changes in skin characteristics, such as irritation, reddening, blistering or burns. Inspect the sensor site every two hours and move the sensor if the skin quality changes. Change the application site every four hours.

For neonates, or patients with poor peripheral blood circulation or sensitive skin, inspect the sensor site more frequently.

If the sensor is too tight because the application site is too large or becomes too large due to edema, excessive pressure for prolonged periods may result in venous congestion distal from the application site, leading to interstitial edema and tissue ischemia.

When patients are undergoing photodynamic therapy they may be sensitive to light sources. Pulse oximetry may be used only under careful clinical supervision for short time periods to minimize interference with photodynamic therapy.

Setting alarm limits to extreme values may cause the alarm system to become ineffective. For example, high oxygen levels may predispose a premature infant to retrolental fibroplasia. If this is a consideration, do not set the high alarm limit to 100%, which is equivalent to switching off the alarm.

SpO

2

is empirically calibrated in healthy adult volunteers with normal levels of carboxyhemoglobin

(COHb) and methemoglobin (MetHb).

To protect from electric shock, always remove the sensor before bathing the patient.

The pulse oximetry function of the bedside monitor should not be used for apnea monitoring.

The pulse oximetry function of the bedside monitor should not be used for arrhythmia analysis.

CAUTION

Change the application site or replace the sensor and/or patient cable when a persistent SpO2 Low

Signal Quality message is displayed on the equipment. These messages may indicate that patient monitoring time is exhausted on the patient cable or sensor.

Replace the cable or sensor when a “SpO2 Sensor Off”, “SpO2 No Sensor”, or “SpO2 Low Signal

Quality” message is consistently displayed while monitoring consecutive patients after completing troubleshooting steps listed in this manual.

Variation in measurements may be profound and may be affected by sampling technique as well as the patient's physiological conditions. Any results exhibiting inconsistency with the patient’s clinical status should be repeated and/or supplemented with additional test data. Blood samples should be analyzed by laboratory instruments prior to clinical decision making to completely understand the patient’s condition.

Use only SpO

2

sensors specified in this manual. Follow the SpO

2

sensor’s instructions for use and adhere to all warnings and cautions.

Do not place the patient monitor where the controls can be changed by the patient.

If using pulse oximetry during full body irradiation, keep the sensor out of the radiation field. If the sensor is exposed to the radiation, the reading might be inaccurate or the device might read zero for the duration of the active irradiation period.

NOTE

Additional information specific to the Masimo sensors compatible with the equipment, including information about parameter/measurement performance during motion and low perfusion, may be found in the sensor's directions for use (DFU).

12 - 2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Masimo cables and sensors are provided with X-Cal™ technology to minimize the risk of inaccurate readings and unanticipated loss of patient monitoring. Refer to the Cable or Sensor DFU for the specified duration of the patient monitoring time.

12.3

SpO

2

Measurement Limitations

The following factors may influence the accuracy of SpO

2

measurement:

■ Patient physiological characteristics:

◆ Cardiac arrest

◆ Hypotension

◆ Darkly pigmented skin

◆ Shock

◆ Severe vasoconstriction

◆ Hypothermia

◆ Severe anemia

◆ Ventricular septal defects (VSDs)

◆ Venous pulsations

◆ Poor perfusion

◆ Dysfunctional hemoglobin, such as carboxyhemoglobin (COHb) and methemoglobin (MetHb)

◆ Elevated levels of bilirubin

◆ Vasospastic disease, such as Raynaud’s, and peripheral vascular disease

◆ Hemoglobinopathies and synthesis disorders such as thalassemias, Hb s, Hb c, sickle cell, etc.

◆ Hypocapnic or hypercapnic conditions

◆ Birthmark(s), tattoos, skin discolorations, moisture on skin, deformed or abnormal fingers. etc.

Interfering substances:

◆ Intravascular dyes (such as indocyanine green, methylene blue, indigo carmine, etc.)

Dyes in the measure site, such as nail polish

Environmental conditions:

◆ Excessive ambient light

Electrosurgery equipment

◆ Defibrillation (may cause inaccurate reading for a short amount of time)

◆ Excessive patient/sensor motion

Electromagnetic field

◆ Arterial catheters and intra-aortic balloon

Others

Inappropriate positioning of the SpO

2

sensor, or use of incorrect SpO

2

sensor

◆ Cuff or arterial blood pressure measurement device on the same limb as the SpO

2

sensor

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 12 - 3

12.4

SpO

2

Display

(1) (2) (3)

(1) (2) (3)

(1) Pleth waveform (Pleth/Plethb): visual indication of patient’s pulse. The waveform is not normalized.

(2) Oxygen saturation of arterial blood (SpO

2

/SpO

2 b): percentage of oxygenated hemoglobin in relation to the sum of oxyhemoglobin and deoxyhemoglobin.

(3) Perfusion indicator: the pulsatile portion of the measured signal caused by arterial pulsation.

(4)

(5)

(6)

Perfusion index (PI): gives the numerical value for the pulsatile portion of the measured signal caused by arterial pulsation. PI is an indicator of the pulsatile strength. You can also use it to assess the SpO

2

signal strength.

Pulse rate (derived from the pleth wave): detected pulsations per minute.

SpO

2 difference ( ∆ SpO

2

):

 SpO

2

= SpO

2

– SpO

2 b

NOTE

PI is displayed for Masimo SpO

2

.

12.5

Preparing for SpO

2

Monitoring

To prepare to monitor SpO

2

, follow this procedure:

1.

Select an appropriate sensor according to the module type, patient category and weight.

2.

Clean the contact surface of the reusable sensor.

3.

Remove colored nail polish from the application site.

4.

Apply the sensor to the patient according to the instruction for use of the sensor.

5.

Select an appropriate extension cable according to the connector type and plug the cable into the SpO

2 connector.

6.

Connect the sensor to the extension cable.

CAUTION

Do not apply sensor too tightly as this results in venous pulsation which may severely obstruct circulation and lead to inaccurate measurements.

At elevated ambient temperatures, be careful with measurement sites that are not well perfused, because this can cause burns after prolonged application.

Avoid placing the sensor on extremities with an arterial catheter, an NIBP cuff or an intravascular venous infusion line.

For neonatal patients, make sure that all sensor connectors and adapter cable connectors are outside the incubator. The humid atmosphere inside can cause inaccurate measurements.

(4)

(5)

(4)

(6)

12 - 4 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

NOTE

Up to two measurement sites are available simultaneously.

12.6

Changing the SpO

2

Settings

NOTE

The settings of in the SpO

2

module and SpO

2 b module are linked.

12.6.1

Changing the SpO

2

Alarm Settings

To change the SpO

2

alarm settings, follow this procedure:

1.

Select the SpO

2

numeric area or waveform area to enter the SpO2 dialog.

2.

Select the Alarm tab.

3.

Enter the password if required. For more information, refer to 32.14.3 Selecting Password for User

Authentication

.

4.

Set the alarm properties as desired.

For SpO

2 b, you can also set alarm properties for

SpO

2

.

NOTE

The desat alarm is a high level alarm notifying you of potentially life threatening drops in oxygen saturation. When the SpO

2

value is below the desat alarm limit and desat alarm switch is set on, the message “SpO2 Desat” is displayed.

You can switch off the SpO

2

Desat alarm only when the SpO

2

Desat Alarm Off is enabled. For more information, see section

32.3.7.7 Setting the Switch of the SpO

2

Desat Alarm Off

.

12.6.2

Nellcor SatSeconds

TM

Alarm Management

With traditional alarm management, high and low alarm limits are set for monitoring oxygen saturation. During monitoring, once an alarm limit is violated, an audible alarm immediately sounds. When the patient SpO

2 fluctuates near an alarm limit, the alarm sounds each time the limit is violated. Such frequent alarms can be distracting. Nellcor’s SatSeconds alarm management technique is used to reduce these nuisance alarms.

The SatSeconds feature is available with the Nellcor SpO

2

to decrease the likelihood of false alarms caused by motion artifacts. With SatSeconds alarm management, high and low alarm limits are set in the same way as those with traditional alarm management. A SatSeconds limit is also set. The SatSeconds limit controls the amount of time that SpO

2

saturation may be outside the set limits before an alarm sounds.

The method of calculation is as follows: the percentage points of the SpO

2

saturation falling outside the alarm limit is multiplied by the number of seconds remaining outside the limit. This can be stated as the equation:

SatSeconds = Points × Seconds

Only when the SatSeconds limit is reached, the monitor gives a SatSeconds alarm. For example, the figure below demonstrates the alarm response time with a SatSeconds limit set at 50 and a low SpO

2

limit set at 90%. In this example, the patient SpO

2

drops to 88% (2 points) and remains there for 2 seconds. Then it drops to 86% (4 points) for 3 seconds, and then to 84% (6 points) for 6 seconds. The resulting SatSeconds are:

% SpO

2

Total SatSeconds=

Seconds

2=

3=

6=

SatSeconds

4

12

36

52

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 12 - 5

After approximately 11 seconds, a Sat-Second alarm would sound, because the limit of 50 SatSeconds would have been exceeded.

%SpO

2

Seconds

Saturation levels may fluctuate rather than remaining steady for a period of several seconds. Often, the patient

SpO

2

may fluctuate above and below an alarm limit, re-entering the non-alarm range several times. During such fluctuation, the monitor integrates the number of SpO

2

points, both positive and negative, until either the

SatSeconds limit is reached, or the patient SpO

2

re-enters the non-alarm range and remains there.

NOTE

The SpO

2

Too Low or SpO

2

Too High alarm is presented in the case that SpO

2

value violates the alarm limits 3 times within one minute even if the setting of SatSeconds is not reached.

To set the SatSeconds, follow this procedure:

1.

Select the SpO

2

numeric area or waveform area to enter the SpO2 dialog.

2.

Select the Alarm tab.

3.

Set Sat-Seconds .

12.6.3

Setting SpO

2

Sensitivity (for Masimo SpO

2

)

For Masimo SpO

2

, selects the Sensitivity as per signal quality and patient motion.

Normal sensitivity is the recommended for patients who are experiencing some compromise in blood flow or perfusion. It is advisable for care areas where patients are observed frequently, such as the intensive care unit

(ICU).

Adaptive Probe Off Detection (APOD) sensitivity is the recommended sensitivity mode where there is a high probability of the sensor becoming detached. It is also the suggested mode for care areas where patients are not visually monitored continuously. This mode delivers enhanced protection against erroneous pulse rate and arterial oxygen saturation readings when a sensor becomes inadvertently detached from a patient due to excessive movement.

Maximum sensitivity is recommended for use on patients with weak signals (e.g. high ambient noise and/or patients with very low perfusion) and for use during procedures or when clinician and patient contact is continuous such as in higher acuity settings.

To set SpO

2

sensitivity, follow this procedure:

1.

Select the SpO

2

numeric area or waveform area to enter the SpO2 dialog.

2.

Select the Setup tab.

3.

Set Sensitivity to Maximum, Normal, or APOD .

12 - 6 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

CAUTION

When using the Maximum Sensitivity setting, performance of "Sensor Off" detection may be compromised. If the equipment and the sensor becomes detached from the patient, the potential for false readings may occur due to environmental noise such as light, and vibration.

Configuring the monitor to “Load Latest Configuration” as the default configuration may result in

Masimo SpO

2

being set to Maximum sensitivity mode on power up or after admitting a new patient.

Maximum sensitivity is recommended for use during procedures or when clinician and patient contact is continuous, such as in higher acuity settings. Maximum sensitivity is not recommended for care areas where patients are not monitored visually as “Sensor Off” detection may be compromised. Refer to

Section 31.4 Setting Default Configuration for managing configuration.

12.6.4

Enabling FastSAT (for Masimo SpO

2

)

FastSAT enables rapid tracking of arterial oxygen saturation changes as may be required in urgent situations.

When FastSAT is switched on, the averaging algorithm evaluates all the SpO

2

values and provides an averaged

SpO

2

value that is a better representation of the patient’s current oxygen saturation status.

The reliability of FastSAT is dependent on the setting for the averaging time and the input signal. FastSAT is disabled by default. To enable FastSAT, follow this procedure:

1.

Select the SpO

2

numeric area or waveform area to enter the SpO2 menu.

2.

Select the Setup tab.

3.

Switch on FastSAT .

12.6.5

Displaying SIQ (for Masimo SpO

2

)

The signal quality indicator (SIQ) displays below the Pleth waveform. The SIQ is conveyed by vertical bars. The height of the bar provides an assessment of the confidence in the displayed SpO

2

value. The SpO

2

SIQ can also be used to identify the occurrence of a patient ‘s pulse.

The following picture shows the SpO

2

SIQ:

(1)

(1) Signal quality indicator (SIQ)

To show SpO

2

SIQ, follow this procedure:

1.

Select the SpO

2

numeric area or waveform area to enter the SpO2 menu.

2.

Select the Setup tab.

3.

Switch on Display SIQ .

NOTE

If the monitor connects to the TM70 telemetry monitor for SpO2 monitoring, the SIQ does not display.

12.6.6

Changing Averaging Time (for Masimo SpO

2

)

The SpO

2

value displayed on the monitor screen is the average of data collected within a specific time. The shorter the averaging time is, the quicker the monitor responds to changes in the patient’s oxygen saturation level. Contrarily, the longer the averaging time is, the slower the monitor responds to changes in the patient’s oxygen saturation level, but the SpO

2

measurement is more stable. For critically ill patients, selecting a shorter averaging time will help with understanding the patient’s state.

To set the averaging time, follow this procedure:

1.

Select the SpO

2

numeric area or waveform area to enter the SpO2 dialog.

2.

Select the Setup tab.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 12 - 7

3.

Set Averaging .

12.6.7

Monitoring SpO

2

and NIBP Simultaneously

When monitoring SpO

2

SpO

2

and NIBP on the same limb simultaneously, you can switch on NIBP Simul to lock the

alarm status until the NIBP measurement ends. If you switch off NIBP Simul , low perfusion caused by NIBP measurement may lead to inaccurate SpO

2

readings and therefore cause false physiological alarms.

To set the NIBP Simul , follow this procedure:

1.

Select the SpO

2

numeric area or waveform area to enter the SpO2 dialog.

2.

Select the Alarm tab.

3.

Set NIBP Simul .

12.6.8

Changing the Sweep Speed of the Pleth Waveform

To set the sweep speed of Pleth waveform, follow this procedure:

1.

Select the SpO

2

numeric area or waveform area to enter the SpO2 dialog.

2.

Select the Setup tab.

3.

Set Speed .

12.7

Changing the PR Settings

12.7.1

Changing the PR Alarm Settings

To change the PR alarm settings, follow this procedure:

1.

Select the SpO

2

numeric area or waveform area to enter the SpO2 dialog.

2.

Select the PR tab.

3.

Enter the password if required. For more information refer to 32.14.3 Selecting Password for User

Authentication

.

4.

Set the alarm properties as desired.

12.7.2

Changing the QRS Volume

If the Alarm Source is set to PR , the QRS tone is derived from PR measurements. To set the QRS volume, follow this procedure:

1.

Select the SpO

2

numeric area or waveform area to enter the SpO2 dialog.

2.

Select the PR tab.

3.

Select the Setup tab.

4.

Set QRS Volume .

If the SpO

2

value is valid, the monitor also adjusts the QRS tone (pitch tone) according to the SpO

2

value. For

information, see 32.16.4 Setting the SpO

2

Tone Mode .

NOTE

Changing the QRS volume settings in the PR Setup page also changes the QRS volume settings in the

Setup page of the ECG dialog.

12.7.3

Setting the PR Source

Current pulse source is displayed in the PR numeric area if current PR source is not SpO

2

. The PR from current pulse source has the following characteristics:

■ PR is stored in the monitor’s database and can be reviewed in the graphic and tabular trends.In graphic trends, as the PR curve has the same color as that of current PR source.

12 - 8 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

■ PR is sent via the network to the CMS, if available.

To set which parameter is used as a PR source, follow this procedure:

1.

Select the SpO

2

numeric area or waveform area to enter the SpO2 dialog.

2.

Select the PR tab.

3.

Select the Setup tab.

4.

Set PR Source .

The dropdown list of the PR Source displays the currently available PR sources from top to bottom by priority.

When you select Auto , the system will automatically select the first option as the PR source. If the current PR source is unavailable, the system will automatically switch PR Source to Auto . When you select IBP , the system will automatically select the first pressure label as the PR source.

12.7.4

Showing/Hiding PR

You can set whether to display the PR value in the SpO

2

parameter area. To do so, follow this procedure:

1.

Select the SpO

2

numeric area or waveform area to enter the SpO2 menu.

2.

Select the PR tab.

3.

Select the Setup tab.

4.

Switch on or off Display PR.

12.8

SpO

2

Troubleshooting

This section lists the problems that might occur. If you encounter the problems when using the equipment or accessories, check the table below before requesting service. If the problem persists, contact your service personnel.

NOTE

For the physiological and technical alarm messages, see Appendix D Alarm Messages .

Problem

Do not see SpO

2

numeric area or waveform area on the main screen

Dashes “- -” display in place of numerics

Low amplitude SpO

2

signal

SpO

2

value is inaccurate

Solution

1. Check that the SpO

2 is set to display in the Screen Setup

Tile

Layout page. For more information, see

3.9.2 Displaying Parameter

Numerics and Waveforms

.

2. Check that the SpO

SpO

2

Off a Parameter

.

2 parameter switch is enabled. If not, enable the

measurement .

For more information, see

3.9.1 Switching On or

3. Check that the cable connections of the SpO

2

sensor and the extension cable are tight. Replace the SpO

2

sensor or the extension cable if needed.

1. Check that the cable connections of SpO

2 extension cable are tight. Replace the SpO

2 sensor and the

sensor or the extension cable if needed.

2. Reconnect the SpO

2

sensor if the alarm SpO2 Sensor Off appears.

3. Check the PI value. If the PI value is too low, adjust the SpO

2 or apply the sensor to a site with better perfusion.

sensor,

4. Move the sensor to a place with weaker ambient light, or cover the sensor to minimize the ambient light if the alarm SpO2 Too Much

Light appears.

1. The SpO

2

sensor and NIBP cuff are placed on the same limb.

Change the monitoring site if necessary.

2. Check the PI value. If the PI value is too low. Adjust the SpO

2 or apply the sensor to a site with better perfusion.

sensor,

1. Check the patient’s vital signs.

2. Check for conditions that may cause inaccurate SpO

2

readings. For more information, see

12.3 SpO

2 Measurement Limitations

.

3. Check the monitor, the SpO

2 functioning.

module or the MPM for proper

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 12 - 9

12.9

Nellcor Information

Nellcor Patents

This posting serves as notice under 35 U.S.C.§287(a) for Covidien patents: http://www.covidien.com/patents.

No Implied License

Possession or purchase of this device does not convey any express or implied license to use the device with unauthorized replacement parts which would, alone, or in combination with this device, fall within the scope of one or more of the patents relating to this device.

12.10

Masimo Information

Masimo Patents

This posting serves as notice under 35 U.S.C.§287(a) for Masimo patents: http://www.masimo.com/patents.htm.

No Implied License

Possession or purchase of this device does not convey any express or implied license to use the device with unauthorized sensors or cables which would, alone, or in combination with this device, fall within the scope of one or more of the patents relating to this device.

12.11

Masimo End-User License Agreement

THIS DOCUMENT IS A LEGAL AGREEMENT BETWEEN YOU (“PURCHASER”) AND SHENZHEN MINDRAY. IF YOU DO

NOT AGREE TO THE TERMS OF THIS AGREEMENT, PROMPTLY RETURN THE ENTIRE PACKAGE, INCLUDING ALL

ACCESSORIES, IN THEIR ORIGINAL PACKAGE, WITH YOUR SALES RECEIPT TO SHENZHEN MINDRAY FOR A FULL

REFUND.

1.

Grant of License. In consideration of payment of the license fee, which is part of the price paid for this product, Shenzhen Mindray grants to Purchaser a nonexclusive, nontransferable license, without right to sublicense, to use the copy of the incorporated software/firmware, and documentation in connection with

Purchaser's use of the Masimo Products for their labeled purpose. Shenzhen Mindray reserves all rights not expressly granted to Purchaser.

2.

Ownership of Software/Firmware. Title to, ownership of, and all rights and interests in, any Masimo software and/or firmware and the documentation, and all copies thereof, remain at all times vested in

Masimo Corporation, licensor to Shenzhen Mindray, and they do not pass to Purchaser.

3.

Assignment. Purchaser shall not assign or transfer this License, in whole or in part, by operation of law or otherwise, without Shenzhen Mindray's prior written consent; any attempt without such consent, to assign any rights, duties or obligations arising hereunder shall be void.

4.

Copy Restrictions. The software/firmware, mask works, circuit board layouts, and accompanying written materials are copyrighted. Unauthorized copying of the software, including software that has been modified, merged, or included with other software, or other written materials is expressly forbidden. You may be held legally responsible for any copyright infringement that is cause or incurred by your failure to abide by the terms of this license. Nothing in this license provides any rights beyond those provided by 17

U.S.C. §117.

5.

Use Restriction. As the Purchaser, you may physically transfer the products from one location to another provided that the software/firmware is not copied. You may not electronically transfer the software/ firmware from the products to any other device. You may not disclose, publish, translate, release distribute copies of, modify, adapt, translate, reverse engineer, decompile, disassemble, or create derivative works based on the Masimo Product, the software/firmware, or the written materials without the prior written consent of Masimo. Masimo Sensors that are designated for single use are licensed under Masimo patents for use on a single patient only, and are not sold. There is no license, implied or otherwise, that would allow use of single use Masimo Sensors beyond their intended single use. After use of single use Masimo Sensors, there is no further license granted by Masimo to use the sensors and they must be discarded.

12 - 10 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

6.

Transfer Restrictions. The software/firmware is licensed to the Purchaser, and may not be transferred to anyone, except other end-users, without the prior written consent of Shenzhen Mindray. In no event may you transfer, assign, rent, lease, sell, or otherwise dispose of the software/firmware or the products on a temporary basis.

7.

Beneficiary. Masimo Corporation is a Beneficiary of this Agreement and has the right to enforce its provisions.

U.S. Government Rights: If you are acquiring software (including the related documentation) on behalf of any part of the United States Government, the following provisions apply: the software is deemed to be “commercial software” and “commercial computer software documentation,” respectively pursuant to DFAR Section 227.7202

FAR 12.212, as applicable. Any use, modification, reproduction, release, performance, display or disclosure of the software (including the related documentation) by the U.S. Government or any of its agencies shall be governed solely by the terms of this Agreement and shall be prohibited except to the extent expressly permitted by the terms of this agreement.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 12 - 11

This page intentionally left blank.

12 - 12 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

13

Monitoring Temperature (Temp)

13.1

Temperature Introduction

The monitor can continuously monitor the patient’s skin temperature and core temperature by the MPM module and the Temp modules. Thermally sensitive resistors (thermistors) are used, and are based on the principle that electrical resistance of the thermistor changes as temperature changes. The resistance change of the thermistors is used to calculate the temperature.

You can connect an MPM module and up to three Temp modules. Up to eight temperature sites can be monitored simultaneously and the difference between two measured sites is calculated.

.

Temperature monitoring is intended for adult, pediatric and neonatal patients.

Platinum MP1

(2) (2)

IBP2 T1

T2

SpO

2

(1)

(1)

ECG NIBP

(1)

(1)

(1) Temperature probe connector (2) Temp dialog hard key

13.2

Temperature Safety Information

CAUTION

Verify that the probe detection program works correctly before monitoring. Remove the temperature probe cable from the temperature probe connector, and check that the monitor can display the alarm messages and give alarm tones correctly.

13.3

Temperature Display

The following figure shows the Temp numeric area for temperature monitoring with the MPM module or Temp module. Your display may be configured to look different.

(3)

(1)

(2)

(1)

(2) (4)

(1) Temperature site

(3)

( 2 ) Temperature alarm limits (3) Temperature value

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 13 - 1

(4) Temperature difference ( ∆ T): Difference between two temperature sites. It displays only when ∆ T is switched on.

13.4

Preparing for Temperature Monitoring

To prepare temperature monitoring, follow this procedure:

1.

Select an appropriate probe for your patient according to patient category and measured site.

2.

Plug the probe or temperature cable to the temperature connector. If using a disposable probe, connect the probe to the temperature cable.

3.

Follow the probe manufacturer’s instructions to connect the probe to the patient.

13.5

Changing Temperature Settings

13.5.1

Setting the Temperature Alarm Properties

To set the temperature alarm properties, follow this procedure:

1.

Select the temperature numeric area to enter the Temp dialog.

2.

Select the Alarm tab.

3.

Enter the password if required. For more information, refer to 32.14.3 Selecting Password for User

Authentication

.

4.

Set the alarm properties as desired.

13.5.2

Selecting the Temperature Label

Select the temperature label according to the measurement site. To do so, follow this procedure:

1.

Select the temperature numeric area to enter the Temp dialog.

2.

Select the Setup tab.

3.

Set the temperature label.

Label

T1

T2

T skin

T core

T axil

T naso

T eso

T rect

Description

Temperature of application site 1

Temperature of application site 2

Skin temperature

Core temperature

Axillary temperature

Nasopharyngeal temperature

Esophageal temperature

Rectal temperature

Label

T amb

T airw

T vesic

T blood

T myo

T tymp

T cereb

Description

Ambient temperature

Airway temperature

Vesical temperature

Blood temperature

Myocardial temperature

Tympanic temperature

Cerebral temperature

13.5.3

Displaying the Temperature Difference

To display the temperature difference between two measurement sites monitored by the same temperature module, switch on

T. To do so, follow this procedure:

1.

Select the temperature numeric area to enter the Temp dialog.

2.

Select the Setup tab.

3.

Switch on ∆ T .

13 - 2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

13.6

Temperature Troubleshooting

This section lists the problems that might occur. If you encounter the problems when using the equipment or accessories, check the table below before requesting service. If the problem persists, contact your service personnel.

NOTE

For the physiological and technical alarm messages, see Appendix D Alarm Messages .

Problem

Do not see Temp numeric area on the main screen

Measurement fails/’--’ is displayed in the Temp numeric area

Solution

1. Check that if the Temp parameter switch is enabled. If not, enable

the Temp measurement. For more information, see 3.9.1 Switching

On or Off a Parameter

.

2. Check that the connections of the temperature probe and the temperature cable are tight.

1. If using a disposable probe, check the connection between the probe and the temperature cable.

2. Try using a known good probe in case the sensor is damaged.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 13 - 3

This page intentionally left blank.

13 - 4 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

14

Monitoring Noninvasive Blood Pressure (NIBP)

14.1

NIBP Introduction

The monitor uses the oscillometric method for measuring Non-Invasive Blood Pressure (NIBP). NIBP measurement is based on the principle that pulsatile blood flow through an artery creates oscillations of the arterial wall. The oscillometric device uses a blood pressure cuff to sense these oscillations that appear as tiny pulsations in cuff pressure. The oscillometric devices measure the amplitude of pressure changes in the occluding cuff as the cuff deflates from above systolic pressure. The amplitude suddenly increases as the pulse breaks through the occlusion in the artery. As the cuff pressure decreases further, the pulsations increase in amplitude, reach a maximum (which approximates to the mean pressure), and then diminish. The oscillometric method measures the mean pressure and determines the systolic and diastolic pressures.

The NIBP module is integrated into the MPM module. NIBP monitoring is intended for adult, pediatric, and neonatal patients.

MP1 Platinum

ECG

T1

T2

SpO

2

IBP2

NIBP

(1)

(1) NIBP cuff connector

NOTE

Blood pressure measurements determined with this device are equivalent to those obtained by a trained observer using the cuff/stethoscope auscultatory method or an intra-arterial blood pressure measurement device, within the limits prescribed by the American National Standard: manual, electronic, or automated sphygmomanometers.

NIBP measurement can be performed during electro-surgery and discharge of a defibrillator.

14.2

NIBP Safety Information

WARNING

Be sure to select the correct patient category setting for your patient before NIBP measurement. Do not apply the higher adult settings for pediatric or neonatal patients. Otherwise, it may present a safety hazard.

Do not measure NIBP on patients with sickle-cell disease or on the limb where skin damage has occurred or is expected.

Use clinical judgment to determine whether to perform frequent unattended blood pressure measurements on patients with severe blood clotting disorders because of the risk of hematoma in the limb fitted with the cuff.

Do not use the NIBP cuff on a limb with an intravenous infusion or arterial catheter in place. This could cause tissue damage around the catheter when the infusion is slowed or blocked during cuff inflation.

Do not apply cuff on the arm on the side of a mastectomy or lymph node clearance.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 14 - 1

Continuous cuff pressure due to connection tubing kinking may cause blood flow interference and result in harmful injury to the patient.

NIBP reading can be affected by the measurement site, the position of the patient, exercise, or the patient's physiologic condition. If you doubt the NIBP measurements, determine the patient’s vital signs by alternative means, and then verify that the monitor is working correctly.

Devices that exert pressure on tissue have been associated with purpura, ischemia, and neuropathy.

Inspect the application site regularly to ensure skin quality and inspect the extremity of the cuffed limb for normal color, warmth and sensitivity. If the skin quality changes, or if the extremity circulation is being affected, move the cuff to another site or stop the blood pressure measurements immediately. Check more frequently when making automatic or STAT measurements. Auto NIBP measurements with one and two minute intervals are not recommended for extended periods of time.

Do not modify or replace connectors of the NIBP air hose except with mindray-approved connectors.

Use neonatal and infant cuffs with CM1901 hoses only. Use pediatric/adult cuffs with CM1903 hoses only.

Never connect intra-arterial or intra-venous lines, or any other incompatible connectors to the NIBP hose. This can cause serious injury or death.

NIBP diagnostic significance must be decided by the hospital’s clinician staff.

CAUTION

Measuring NIBP on patients connected to an Intra Aortic Balloon Pump (IABP) machine may result in the NIBP measurement, including PR, to be inaccurate or fail.

Only use parts and accessories specified in this manual. Follow the instructions for use and adhere to all warnings and cautions.

Using IABP may cause NIBP, including PR, measurements inaccurate or failed.

Accuracy of NIBP measurement depends on using a cuff of proper size. It is essential to measure limb circumference and choose a cuff with proper size.

14.3

NIBP Measurement Limitations

Measurements are impossible with heart rate extremes of less than 30 bpm or greater than 300 bpm, or if the patient is on a heart-lung machine. The measurement may be inaccurate or impossible in the following situations:

Regular arterial pressure pulses are hard to detect

Excessive and continuous patient movement such as shivering or convulsions

Cardiac arrhythmias

Rapid blood pressure changes

Severe shock or hypothermia that reduces blood flow to the peripheries

On an edematous extremity

Obesity, where a thick layer of fat surrounding a limb dampens the oscillations coming from the artery

NOTE

The effectiveness of this sphygmomanometer has not been established in pregnant women, including pre-eclamptic patients.

14.4

Measurement Modes

The monitor provides the following NIBP measurement modes:

Manual: measurement on demand.

Auto: repeated measurements at set interval.

STAT: continual rapid series of measurements over a five-minute period.

Sequence: continually automatic measurement at set durations and intervals.

14 - 2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

14.5

NIBP Display

The NIBP display shows only numerics.

(3) (4) (5)

(2)

(1)

(6)

(7)

(8)

(11) (10) (9)

(1) Systolic pressure alarm limits

(2) NIBP unit: mmHg or kPa

(3) The last NIBP measurement time

(4) Time to the next measurement (for Auto mode and Sequence mode)

(5) Measurement mode: for Auto NIBP, interval is displayed; for Sequence mode, the current phase and interval are displayed

(6) Diastolic pressure

(7) Diastolic pressure alarm limits

(8) Mean pressure alarm limits

(9) Mean pressure (displayed after measurement completed) or cuff pressure (displayed during the measurement)

(10) Systolic pressure

(11) Pulse Rate

NOTE

If NIBP measurement fails, “XX” is displayed; if NIBP measurement is not taken or NIBP measurement exceeds its measurement ranges, “--” is displayed.

NIBP numerics in black color with a light outline indicate that the measurement is old and exceeds the configured NIBP Measurement Timeout time (see section

32.5.10 Setting Parameter

Measurement Timeout ). These NIBP values are not recommended for reference.

14.6

Preparing for NIBP Measurements

14.6.1

Preparing the Patient for NIBP Measurements

In normal use, perform NIBP measurement on a patient who is in the following position:

Comfortably seated

Legs uncrossed

Feet flat on the floor

Back, arm and feet supported

NOTE

It is recommended that the patient remains calm and relaxes as much as possible before performing the measurement and that the patient does not talk during the measurement.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 14 - 3

Other factors that have been shown to result in an overestimation of blood pressure are labored breathing, full bladder, pain etc.

14.6.2

Placing the NIBP Cuff

To place the NIBP cuff, follow this procedure:

1.

Verify that the patient category setting is correct.

2.

Connect the air tubing to the NIBP connector.

3.

Select an appropriately sized cuff for the patient, and then wrap it around the limb directly over the patient’s skin as follows: a Determine the patient

’ s limb circumference.

b Select an appropriate cuff by referring to the limb circumference marked on the cuff. The width of the cuff should be 40% (50% for neonates) of the limb circumference, or 2/3 of the length of the upper arm or the thigh. The inflatable part of the cuff should be long enough to encircle to overlap at least

50% to 80% of the limb. c Apply the cuff to the patient ’ s upper arm or leg and make sure the Φ marking on the cuff matches the artery location. The cuff should fit snugly, but with enough room for two fingers to be placed between the cuff and the patient ’ s arm (on adults), and loosely on neonates with little or no air present within the cuff. Otherwise it may cause discoloration and ischemia of the extremities. Make sure that the cuff index line falls within the range markings on the cuff. If it does not, use a cuff that fits better.

d Middle of the cuff should be at the level of the right atrium of the heart. If it is not, use the measurement correction formula to correct the measurement. For more information, see

14.8.7 Setting the NIBP Display Format

.

4.

Connect the cuff to the air tubing. Avoid compression or restriction of pressure tubes. Air must pass unrestricted through the tubing.

CAUTION

Do not touch or apply external pressure against the cuff and air tubing during NIBP measurement.

This may cause inaccurate blood pressure values.

Use care when placing the cuff on an extremity used for monitoring other patient parameters.

14.7

Starting and Stopping NIBP Measurements

Start and stop NIBP measurement by selecting the NIBP quick keys or from the NIBP dialog.

Task

Start a manual measurement

Start auto NIBP series

Start NIBP sequence measurement

Start STAT measurement

Stop the current NIBP measurements

By Quick Key

NIBP Start/Stop quick key

NIBP Start/Stop quick key

NIBP Measure quick key

Interval

NIBP Measure quick key

Sequence

NIBP STAT quick key

→ select

NIBP Measure quick key

NIBP Start/Stop quick key

STAT

From NIBP dialog

Start NIBP button

Setup tab

set Interval

Start NIBP button

Sequence tab → set NIBP sequence

→ Start NIBP button

STAT button

Stop NIBP button

14 - 4 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Task

End auto NIBP series or NIBP

Sequence

Stop STAT measurement and end series

By Quick Key

NIBP Stop All quick key

NIBP Start/Stop quick key

NIBP STAT quick key

From NIBP dialog

NIBP Stop All button

Stop NIBP or NIBP Stop All button

14.8

Changing NIBP Settings

14.8.1

Setting the NIBP Alarm Properties

To set the NIBP alarm properties, follow this procedure:

1.

Select the NIBP numeric area to enter the NIBP dialog.

2.

Select the Alarm tab.

3.

Enter the password if required. For more information, refer to 32.14.3 Selecting Password for User

Authentication

.

4.

Set the alarm properties as desired.

14.8.2

Setting the Initial Cuff Inflation Pressure

To set initial cuff inflation pressure, follow this procedure:

1.

Select the NIBP numeric area to enter the NIBP dialog.

2.

Select Initial Pressure , and then select the appropriate setting.

NOTE

For known hypertensive patients, you need to set initial cuff pressure to a higher value to reduce the measurement time.

14.8.3

Setting the NIBP Interval

For auto NIBP measurement, you need to set the interval between two NIBP measurements. To set the NIBP interval, follow this procedure:

1.

Select the NIBP numeric area to enter the NIBP dialog.

2.

Set Interval . Selecting Manual switches to manual mode.

NOTE

NIBP Measure dialog gives quick access to starting and setting the interval. Selecting an option from the NIBP Measure dialog will change the NIBP interval to that selection and start the NIBP measurement.

14.8.4

Selecting NIBP Start Mode

Start mode defines how future NIBP measurements are initiated with automatic measurement mode. To set the start mode, follow this procedure:

1.

Select the NIBP numeric area to enter the NIBP dialog.

2.

Select Start Mode .

◆ Clock : after the first measurement, the monitor automatically synchronizes NIBP automatic measurements with the real time clock. For example, if Interval is set to 20 min , and NIBP auto measurement is started at 14:03, the next measurement will be taken at 14:20, and then at 14:40,

15:00, and so on.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 14 - 5

◆ Interval : after the first measurement, the monitor automatically repeats measurements at set interval. For example, if Interval is set to 20 min , and NIBP auto measurement is started at 14:03, the next measurement will be taken at 14:23, and then at 14:43, 15:03, and so on.

14.8.5

Enabling the NIBP End Tone

The monitor can issue a reminder tone at the completion of NIBP measurement. The NIBP End Tone is off by default. To switch on the NIBP end tone, follow this procedure:

1.

Select the NIBP numeric area to enter the NIBP dialog.

2.

Switch on NIBP End Tone .

14.8.6

Setting NIBP Sequence

NIBP sequence measurement can have up to five phases: A, B, C, D, and E. The duration and interval of each phase can be set individually.

To set NIBP sequence, follow this procedure:

1.

Select the NIBP numeric area to enter the NIBP dialog.

2.

Select the Sequence tab.

3.

Set Duration and Interval of each phase.

14.8.7

Setting the NIBP Display Format

To set the NIBP display format, follow this procedure:

1.

Select the NIBP numeric area to enter the NIBP menu.

2.

Select the Setup tab.

3.

Set Display Format .

14.8.8

Setting the NIBP Alarm Limits Display Switch

To set whether to display the alarm limits of diastolic NIBP and mean NIBP, follow this procedure:

1.

Select the NIBP numeric area to enter the NIBP menu.

2.

Select the Setup tab.

3.

Switch on or off Display Alarm Limits.

14.8.9

Showing/Hiding PR

You can set whether to display the PR value in the NIBP parameter area. To do so, follow this procedure:

1.

Select the NIBP numeric area to enter the NIBP menu..

2.

Select the Setup tab.

3.

Switch on or off Display PR.

14.8.10

Correcting the NIBP Measurements

The middle of the cuff should be at the level of right atrium. If the limb is not at the heart level, the measurement should be corrected as follows:

Add 0.75 mmHg (0.10 kPa) to the displayed value for each centimetre higher.

Deduct 0.75 mmHg (0.10 kPa) to the displayed value for each centimeter lower.

14.9

Assisting Venous Puncture

You can use the NIBP cuff to cause sub-diastolic pressure to block the venous blood vessel and therefore help venous puncture. To assist venous puncture, follow this procedure:

1.

Select the Venipuncture quick key or select the NIBP numeric area → Setup tab.

14 - 6 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

2.

Set Venipuncture Pressure .

3.

Select Venipuncture at the bottom of the dialog.

4.

Puncture vein and draw blood sample.

5.

Select the NIBP Start/Stop quick key to deflate the cuff. If the cuff is not manually deflated, the cuff automatically deflates after a fixed period of time (170 seconds for adult and pediatric patient, 85 seconds for neonatal patient).

During venous puncture, pay attention to the cuff pressure and the remaining time displayed in the NIBP numerics area.

14.10

NIBP Troubleshooting

This section lists the problems that might occur. If you encounter the problems when using the equipment or accessories, check the table below before requesting for services. If the problem persists, contact your service personnel.

NOTE

For the physiological and technical alarm messages, see Appendix D Alarm Messages .

Problem

Cannot see NIBP numeric area on the main screen

Solution

1. Check that the NIBP is set to display in the Screen Setup

Tile Layout page. For more information, see.

3.9.2 Displaying Parameter Numerics and

Waveforms

2. Check that if the NIBP parameter switch is enabled. If not, enable the NIBP measurement. For more information, see

3.9.1 Switching On or Off a

Parameter

.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 14 - 7

This page intentionally left blank.

14 - 8 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

15

Monitoring Invasive Blood Pressure (IBP)

15.1

IBP Introduction

You can measure invasive blood pressure (IBP) using the MPM, or the IBP module. This patient monitor can monitor up to 8 invasive blood pressures and displays the systolic, diastolic and mean pressures and a waveform for each pressure.

IBP monitoring is intended for adult, pediatric, and neonatal patients. PAWP monitoring is only intended for adult and pediatric patients.

(1) (3) (1)

Platinum MP1

(2)

ECG

T1

T2

SpO

2

IBP2

NIBP

(2)

(2)

(1) Zero IBP hard key (2) IBP cable connector (3) IBP dialog hard key

15.2

IBP Safety Information

WARNING

Use only pressure transducers specified in this manual. Never reuse disposable pressure transducers. See chapter

36 Accessories for specified pressure transducers.

Make sure that the applied parts never contact other conductive parts.

To reduce the hazard of burns during high-frequency surgical procedures, ensure that the monitor’s cables and transducers never come into contact with the high-frequency surgical units.

When using accessories, their operating temperature should be taken into consideration. For more information, see instructions for use of accessories.

All invasive procedures involve risks to the patient. Use aseptic technique. Follow catheter manufacturer's instructions.

Mechanical shock to the invasive blood pressure transducer may cause severe shifts in zero balance and calibration, and cause erroneous readings.

The neutral electrode of the electro-surgery unit (ESU) shall properly contact the patient. Otherwise, burns may result.

CAUTION

Measuring IBP on patients connected to an Intra Aortic Balloon Pump (IABP) machine may result in the IBP measurement, including PR, to be inaccurate or fail.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 15 - 1

15.3

Preparing for IBP Monitoring

15.3.1

IBP Equipment to Patient Connection

(1) mmHg

(4)

(5)

(3)

(2)

(5)

(1) Pressure bag

(3) IBP cable

(5) Three-way valve

(2) IBP connector

(4) IBP transducer

15.3.2

Measuring an Invasive Blood Pressure

To monitor IBP, follow this procedure:

1.

Connect one end of the IBP cable to the IBP cable connector, and the other end to the IBP transducer.

2.

Flush the IBP transducer system to exhaust all air from the tubing according to the manufacturer’s instructions. Ensure that the system is free of air bubbles.

3.

Connect the IBP transducer to the patient, making sure that the transducer is at the same horizontal level as the heart.

4.

Select the proper pressure label for currently measured pressure. For more information, see

15.6.2 Changing the Pressure Label .

5.

Zero the IBP transducer. For more information, see.

15.3.3 Zeroing the IBP transducer . After a successful

zeroing, turn off the three-way valve to the air and turn on the three-way valve to the patient.

CAUTION

Make sure that all the transducers are zeroed correctly before making IBP measurements.

Make sure that all air bubbles have been flushed from the IBP transducer system before making IBP measurements.

If measuring intracranial pressure (ICP) with a sitting patient, level the transducer with the top of the patient’s ear. Incorrect leveling may give incorrect values (not applicable if measuring ICP with the

Codman ICP transducer).

15 - 2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

15.3.3

Zeroing the IBP transducer

To avoid inaccurate pressure readings, the monitor requires a valid zero. Zero the transducer in accordance with your hospital policy. The IBP transducer should be zeroed in the following conditions:

The IBP transducer, adapter cable or module is reconnected.

The monitor restarts.

The readings are in doubt.

The monitor displays the prompt message Zero Required .

To zero the transducer, follow this procedure:

1.

Connect the IBP transducer, the IBP adapter cable and the module.

2.

Turn off the three-way valve (the one near the transducer) to the patient, in order to vent the transducer to the atmospheric pressure.

3.

Zero the transducer by one of the following methods:

◆ Press the Zero hard key on the module. In the Zero IBP dialog, select to zero an IBP, or select to zero all IBP.

◆ Select the numeric area (such as the Art numeric area), and then select Zero button.

◆ Select the Zero IBP quick key. In the Zero IBP dialog, select to zero an IBP, or select to zero all IBP.

4.

After the zero calibration is completed, close the three-way valve to the air and open the three-way valve to the patient.

15.4

Zero calibration may fail in case of pressure fluctuation or pressure exceeding the calibration range. If zero calibration fails, follow this procedure:

1.

Check that the three-way valve (the one near the transducer) is open to the air.

2.

Perform zero calibration again. Do not sway the IBP transducer and tubing during zero calibration.

Measuring ICP Using the Codman ICP Transducer

15.4.1

Zeroing the Codman ICP transducer

The Codman ICP transducer (Model: 82-6653) must be zeroed before use. To zero the ICP transducer, follow this procedure:

1.

Before unpacking the ICP transducer, check that the monitor supports the Codman ICP transducer. a Select the Main Menu quick key → from the Parameters column select Setup → select ICP (If the

ICP button is not in the Parameter Setup dialog, select any IBP button to enter corresponding IBP dialog, and then change the IBP label to ICP ) → select the Zero tab.

b Check that the following icon is displayed in the Zero page. The monitor supports the Codman ICP transducer if the following icon is displayed in the Zero page.

2.

Connect the ICP transducer, the ICP adapter cable and the module.

3.

Follow the manufacturer’s instructions to prepare the ICP transducer.

4.

Zero the ICP transducer: when the message Zero Reference is displayed in the ICP numeric area, select the

ICP waveform area or numeric area to enter the ICP dialog

select the Zero tab

→ select the Zero button.

5.

Record the zero reference value on the blank area of the ICP transducer for reference.

If the ICP transducer zero calibration failed or the zero reference value is in doubt, perform the zero calibration again.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 15 - 3

15.4.2

Measuring ICP

To perform the ICP measurement, follow this procedure:

1.

Zero the Codman ICP transducer. For more information, see section 15.4.1 Zeroing the Codman ICP transducer

.

2.

Disconnect the ICP transducer and ICP adapter cable. Follow the manufacturer’s instructions to apply the

ICP transducer to the patient.

3.

Reconnect the ICP transducer and ICP adapter cable.

4.

Check that the zero reference value displayed on the monitor is consistent with that recorded on the ICP transducer.

◆ Consistent: select Accept .

◆ Inconsistent: input the zero reference value recorded on the ICP transducer, and select Accept .

If the patient being monitored for ICP must be transported, check that the target monitor supports the Codman

ICP transducer. For more information, see 15.4.1 Zeroing the Codman ICP transducer

. If the target monitor does not support the Codman ICP transducer, do not use it for ICP monitoring.

If the target monitor supports the Codman ICP transducer, follow this procedure to transport the patient:

1.

Disconnect the ICP adapter cable from the measurement module, or remove the module from the monitor.

2.

Connect the ICP adapter cable, measurement module, and the transport monitor, or insert the measurement module into the transport monitor.

3.

Check that the zero reference value displayed on the monitor is consistent with that recorded on the ICP transducer.

◆ Consistent: select Accept .

◆ Inconsistent: input the zero reference value recorded on the ICP transducer, and select Accept .

CAUTION

If monitors of different brands are used to zero the Codman ICP transducer, the zero reference values can be different. Use a Mindray monitor to Zero the Codman ICP transducer if you will take

ICP measurement using a Mindray monitor. Otherwise the ICP measurement could be inaccurate.

15.5

IBP Display

The IBP measurement is displayed on the monitor as a waveform and numeric pressures. For arterial pressure, the IBP numeric area displays systolic pressure, diastolic pressure and mean pressure. For venous pressure, the

IBP numeric area displays only the mean pressure. The figure below shows the waveform and numerics for the

Art pressure.

(1) (2) (3) (4) (5)

15 - 4

(7) (6)

(1) Pressure label

(3) Pressure Unit

(5) Diastolic pressure

(7) PPV measurement

(2) Waveform

(4) Systolic pressure

(6) Mean pressure

For some pressures, the parameter window may show the mean pressure only. For different pressures, their defaults unit may be different. If the Art and ICP pressures are measured simultaneously, the ICP parameter area will display numeric CPP, which is obtained by subtracting ICP from the Art mean.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

15.5.1

Overlapping IBP Waveforms

The IBP waveforms can be displayed together. To combine IBP waveforms, follow this procedure:

1.

Access Tile Layout by either of the following ways:

◆ Select the Screen Setup quick key

select the Tile Layout tab.

Select Main Menu quick key

→ from the Display column select Tile Layout .

2.

Select the waveform area where you want to display the overlapped IBP waveforms, and then select the IBP waves to be overlapped on the left side of the same line.

3.

Repeat step 2 in another waveform area if needed.

4.

Select to save the setting and exit the screen. The main screen will display the overlapped IBP waves.

Selecting the overlapped IBP waveforms on the main screen opens the Overlapping Waveform Setup dialog, where you can make the following settings:

■ Scale

Set Left Scale for the arterial pressure.

◆ Set Right Scale for the venous pressure.

◆ Set CVP Scale individually if the CVP waveform is combined and CVP unit is different from IBP unit.

Set ICP Scale individually if the ICP waveform is combined and ICP unit is different from IBP unit.

◆ Set PA Scale individually if the PA waveform is combined.

Switch on or off Gridlines to show or hide gridlines in the overlapped waveform area.

Set Speed for the overlapped waveforms.

NOTE

The unit of CVP scale is consistent with the CVP parameter unit.

15.6

Changing IBP Settings

15.6.1

Changing the IBP Alarm Settings

To change the IBP alarm settings, follow this procedure:

1.

Select the IBP numeric area or waveform area to enter the corresponding pressure dialog.

2.

Select the Alarm tab.

3.

Enter the password if required. For more information, refer to 32.14.3 Selecting Password for User

Authentication

.

4.

Set the alarm properties of pressure.

15.6.2

Changing the Pressure Label

The pressure label is a unique identifier for each type of pressure. Therefore, it is recommended to select the proper pressure label for the source of the pressure being monitored.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 15 - 5

To select the pressure label, follow this procedure:

1.

Select the IBP numeric area or waveform area to enter the corresponding pressure dialog.

2.

Select the Setup tab.

3.

Set IBP1 Label or IBP2 Label .

Label

Art

Ao

BAP

CVP

RAP

UVP

P1 to P4

Description

Arterial blood pressure

Aortic pressure

Brachial arterial pressure

Central venous pressure

Right atrial pressure

Umbilical venous pressure

Non-specific pressure label

Label

PA

UAP

FAP

LAP

ICP

LV

Description

Pulmonary artery pressure

Umbilical arterial pressure

Femoral arterial pressure

Left atrial pressure

Intracranial pressure

Left ventricular pressure

NOTE

The same label cannot be selected for different pressures. When two pressures are detected having the same label, the monitor automatically changes one pressure label to a currently unused one.

15.6.3

Setting the Pressure Type for Display

For the non-specific pressure (P1, P2, P3 or P4), the displayed pressure type is configurable. To set the displayed pressure type, follow this procedure:

1.

Select the numeric area or waveform area of the non-specific pressure to enter the corresponding pressure dialog.

2.

Select the Setup tab.

3.

Set Measure :

If this non-specific pressure is arterial pressure, set the Measure to All . In this case, its corresponding numeric area displays systolic pressure, diastolic pressure and mean pressure.

◆ If this non-specific pressure is venous pressure, set the Measure to Mean Only . In this case, its corresponding numeric area displays only the mean pressure.

15.6.4

Changing the Sensitivity

The IBP value displayed on the monitor screen is the average of data collected within a specific time. The shorter the averaging time is, the quicker the monitor responds to changes in the patient’s blood pressure, and the higher the sensitivity. Contrarily, the longer the averaging time is, the slower the monitor responds to changes in the patient’s blood pressure, the lower the sensitivity, but the measurement accuracy will be improved. For critically ill patients, selecting higher sensitivity will help with understanding the patient’s state.

To set the sensitivity, follow this procedure:

1.

Select the IBP numeric area or waveform area to enter the corresponding pressure dialog.

2.

Select the Setup tab.

3.

Set Sensitivity .

15.6.5

Setting the IBP Waveform

To set the IBP waveform, follow this procedure:

1.

Select the IBP numeric area or waveform area to enter the corresponding pressure dialog.

2.

Select the Setup tab.

3.

Set the following properties of the IBP waveform:

15 - 6 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

◆ Speed

◆ Scale : if Auto is selected, the size of the pressure’s waveform will be adjusted automatically.

15.6.6

Setting the Display Format of Artery Pressure

To set the display format of the artery pressure, follow this procedure:

1.

Select the numeric area or waveform area of any arterial pressure to enter the corresponding menu.

2.

Select the Setup tab.

3.

Set Display Format .

15.6.7

Showing/Hiding the Alarm Limits of Artery Pressure

To set whether to display the alarm limits of the arterial pressure, follow this procedure:

1.

Select the numeric area or waveform area of any arterial pressure to enter the corresponding menu.

2.

Select the Setup tab.

3.

Switch on or off Display Alarm Limits.

15.6.8

Setting the Use PA-D as PAWP Switch

The PA-D value can be configured to replace the PAWP value for hemodynamic calculation. To do so, follow this procedure:

1.

Select the PA numeric area or waveform area to enter the PA dialog.

2.

Select the Setup tab.

3.

Switch on or off Use PA-D as PAWP .

For more information on hemodynamic calculation, see 27.4 Hemodynamic Calculations

.

15.6.9

Enabling PPV Measurement

PPV indicates pulse pressure variation. When measuring the arterial pressure (except PA), the PPV measurement is available. To enable the PPV measurement, follow this procedure:

1.

Select the IBP numeric area or waveform area to enter the corresponding pressure dialog.

2.

Select the PPV Setup tab.

3.

Switch on PPV Measure .

You can select PPV source after enabling the PPV measurement.

WARNING

This monitor can calculate PPV from beat-to-beat values of any arterial pulsatile pressure. The circumstances under which the calculation of a PPV value is clinically meaningful, appropriate and reliable must be determined by a physician.

The clinical value of the derived PPV information must be determined by a physician. According to recent scientific literature, the clinical relevance of PPV information is restricted to sedated patients receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia.

PPV calculation may lead to inaccurate values in the following situations:

◆ at respiration rates below 8 rpm

◆ during ventilation with tidal volumes lower than 8 ml/kg

◆ for patients with acute right ventricular dysfunction (cor pulmonale).

The PPV measurement has been validated only for adult patients.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 15 - 7

15.7

PAWP

Pulmonary Artery Wedge Pressure (PAWP) values, used to assess cardiac function, are affected by fluid status, myocardial contractility, and valve and pulmonary circulation integrity.

Obtain the measurement by introducing a balloon-tipped pulmonary artery flotation catheter into the pulmonary artery. When the catheter is in one of the smaller pulmonary arteries, the inflated balloon occludes the artery allowing the monitor to record changes in the intrathoracic pressures that occur throughout the respiration cycle.

The pulmonary wedge pressure is the left ventricular end diastolic pressure when the airway pressure and valve function are normal. The most accurate PAWP values are obtained at the end of the respiration cycle when the intrathoracic pressure is fairly constant and the artifact caused by respiration is minimal.

WARNING

PAWP monitoring is not intended for neonatal patients.

15.7.1

PAWP Equipment to Patient Connection

(2)

(1)

(3)

(4)

(6)

(7)

(8)

(5)

(1) IBP connector

(3) IBP transducer

(5) PA distal port

(7) Thermodilution catheter

15.7.2

Preparing to Measure PAWP

To prepare to monitor PAWP, follow this procedure:

(2) Flush bag

(4) Three-way valve

(6) Balloon inflation syringe

(8) Balloon

15 - 8 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

1.

Connect one end of the IBP cable to the IBP cable connector, and the other end to the IBP transducer. For more information, see

15.3.2 Measuring an Invasive Blood Pressure

.

2.

Follow the manufacturer’s instructions to connect the PA port of the thermodilution catheter and the patient end of the IBP transducer.

3.

Zero the IBP transducer. For more information, see

15.3.3 Zeroing the IBP transducer

.

4.

Set the IBP label to PA

since the PAWP is measured on PA. For more information, see 15.6.2 Changing the

Pressure Label .

15.7.3

Measuring PAWP

To measure the PAWP, follow this procedure:

1.

Select the PA numeric area or waveform area to enter the PA dialog, and then select PAWP .

2.

Wedge the flotation catheter into the pulmonary artery by observing the PA waveform changes on the screen, referring to the following figure.

3.

Select Start .

4.

Inflate the balloon and pay attention to PA waveform changes on the screen when the prompt message

Ready For Balloon Deflation appears.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 15 - 9

5.

Deflate the balloon when the prompt message Ready For Balloon Deflation appears. If the PA waveform is stable, but the monitor is still not showing the prompt message Ready For Balloon Deflation , select

Freeze to freeze the waveform, and deflate the balloon.

6.

Select Accept to save the PAWP value.

7.

If you need to start a new measurement, repeat step 3 to step 6.

If the measurement fails or you need to adjust the PAWP value, you can use the following buttons to adjust the

PAWP waveform and measurement.

Select the up or down arrow button to adjust the PAWP value.

Select the left or right arrow button to view the frozen waveforms of 40 seconds.

Select Accept to save the PAWP value.

WARNING

Prolonged inflation can cause pulmonary hemorrhage, infarction or both. Inflate the balloon for the minimum time necessary to get an accurate measurement.

If the PAWP is greater than the PA (systolic), deflate the balloon and report the incident in accordance with hospital policy. Because the pulmonary artery could be accidentally ruptured, and the PAWP value derived will not reflect the patient’s hemodynamic state, but will merely reflect the pressure in the catheter or balloon.

If the flotation/thermodilution catheter drifts into the wedge position without inflation of the balloon, the PA waveform assumes a wedged appearance. Take appropriate action, in accordance with standard procedures, to correct the situation.

NOTE

The PA alarm is turned off automatically when the monitor enters the PAWP screen.

15.7.4

Setting the Waveforms of the PAWP Screen

On the PAWP screen, select Setup to enter the PAWP Setup dialog. In the PAWP Setup dialog, you can make the following settings:

Select

Select

Select

Select

Reference Waveform 1

Reference Waveform 2

Speed

Scale to set an ECG lead wave as the first reference wave.

to set a respiration wave as the second reference wave.

to set a sweep speed for the displayed waveforms on the

to set the size of the PA waveform on the PAWP screen.

PAWP screen.

15.7.5

Performing Hemodynamic Calculation

On the PAWP screen, select Hemo Calcs to enter the Calculations dialog. For more information, see

27.4 Hemodynamic Calculations .

15.8

Connecting a Camino Device

The IBP module can interface with the Camino multi-parameter monitor (Model: MPM-1) to measure intracranial pressure (ICP).

To connect the Camino, follow this procedure:

1.

Plug the IBP module into the module rack.

2.

Connect the Camino ICP cable to the IBP module.’

3.

Connect the Camino ICP cable to the ICP adapter.

4.

Connect the ICP connector to the ICP adapter.

5.

Connect the Camino cable to the Camino monitor.

15 - 10 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

(1)

(2) (3) (4) (5)

(6)

(1) IBP module

(3) ICP adapter

(5) Camino cable

(2) Camino ICP cable

(4) ICP connector

(6) Camino monitor

WARNING

Refer to the Camino Operator’s Manual to adjust settings and connect the monitor with the patient.

ICP alarm settings on the Mindray patient monitor are independent of the Camino monitor alarm settings, and thus may be different. Please pay special attention to the alarms on the Camino monitor.

15.9

IBP Troubleshooting

This section lists the problems that might occur. If you encounter the problems when using the equipment or accessories, check the table below before requesting service. If the problem persists, contact your service personnel.

NOTE

For the physiological and technical alarm messages, see Appendix D Alarm Messages .

Problem

Cannot see IBP numeric area or waveform area on the main screen

Cannot see systolic pressure and diastolic pressure for P1/P2/P3/P4

IBP readings seem unstable

Solution

1. Check that the IBP is set to display in the Screen Setup dialog. For more information, see.

3.9.2 Displaying Parameter Numerics and Waveforms .

2. Check that if the IBP parameter switch is enabled. If not, enable the IBP measurement. For more information, see

3.9.1 Switching On or Off a

Parameter

.

3. Check the connection of IBP cable, IBP transducer and monitor.

4. Check that the three-way valve is turned to the correct position.

5. Check that the IBP transducer has been zeroed. For more information, see

15.3.3 Zeroing the IBP transducer .

Set Measure to All in the P1/P2/P3/P4 setup page. For more information, see

15.6.3 Setting the Pressure Type for Display .

1. Make sure there are no air bubbles in the transducer systems.

2. Check that the transducer is properly fixed.

3. Zero the transducer again.

4. Replace the transducer.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 15 - 11

Problem

Zeroing of IBP channel(s) fails.

Solution

1. Ensure that the channels are open to air.

2. Perform zero calibration again.

Do not sway the IBP transducer and tubing during zero calibration.

For more information, see

15.3.3 Zeroing the IBP transducer .

3. If zero calibration still fails, replace the transducer.

15 - 12 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

16

Monitoring Cardiac Output (C.O.)

16.1

C.O. Introduction

The cardiac output (C.O.) measurement invasively measures cardiac output and other hemodynamic parameters using the right heart (atria) thermodilution method. A cold solution of known volume and temperature is injected into the right atrium through the proximal port of a pulmonary artery (PA) catheter. The cold solution mixes with the blood in the right ventricle and the change in blood temperature is measured with a thermistor at the distal end of the catheter in the pulmonary artery. The temperature change is displayed as a curve on the

C.O. split screen, and the monitor calculates the C.O. value from this curve. The C.O. value is inversely proportional to the area under the curve. As cardiac output varies continuously, a series of measurements must be carried out to achieve a reliable C.O. average value. Always use the average of multiple thermodilution measurements for therapy decisions. The monitor is capable of storing 6 measurements.

C.O. monitoring is intended for adult patients only.

(1) (2)

(3)

(1) C.O. dialog hard key (2) C.O. measure dialog hard key (3) C.O. cable connector

16.2

C.O. Safety Information

WARNING

The C.O. measurement results may be erroneous during electrosurgery.

All invasive procedures involve risks to the patient. Use aseptic technique and follow catheter manufacturer's instructions.

Use only accessories specified in this manual. Make sure that the accessories never come into contact with conductive parts.

C.O. monitoring is not intended for pediatric and neonatal patients.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 16 - 1

16.3

C.O. Measurement Limitations

The following factors may influence the accuracy of C.O. measurement: temperature of injectate solution volume of injectate solution baseline of patient’s blood temperature patient’s inspiratory/expiratory cycle placement of catheter with relation to proximity of lung field the catheter itself patient’s heart rate and hemodynamic status any solution infused with intravenous injection during the C.O. measurement

To obtain accurate C.O. measurements, follow these recommendations:

Temperature of injectate solution must be at least 10 ºC cooler than that of the patient’s blood.

Inject solution at end of expiration.

Inject solution rapidly and smoothly.

Finish injection within four to five seconds.

16.4

C.O. Display

The C.O. display shows only C.O., C.I. (cardiac index), and TB (blood temperature) in the C.O. numeric area.

(1)

(2)

(3)

(1) C.O. label

(3) Labels and values for primary parameters

(4)

(2) Primary parameter unit

(4) Labels and values for secondary parameters

NOTE

The C.O. value is invalid if it is displayed in outline font.

16 - 2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

16.5

C.O. Equipment to Patient Connection

(1) C.O. module

(4) In-line probe

(7) Three-way valve

(10) Thermistor connector

(13) Thermistor

(2) 12-pin C.O. cable (Model: CO7702)

(5) Injectate solution

(8) Proximal injectate port

(11) PA distal port

(3) TI cable connector

(6) Injectate syringe

(9) Balloon inflation valve

(12) Thermodilution catheter

16.6

Performing C.O. Measurement

16.6.1

Preparing for C.O. Measurement

1.

Connect the C.O. cable to the C.O. module and thermistor connector, making sure the C.O. numeric area is displayed on the monitor’s main screen.

2.

Follow the hospital’s policy and procedures to prepare the patient for the C.O. measurement.

3.

Follow the manufacturer’s instructions to set up the catheter and other accessories.

4.

Check that all the accessories are properly connected.

For an in-line probe setup, make sure the in-line sensor is securely connected to the tubing. For the bath probe setup, make sure the bath probe is correctly sensing the injectate temperature.

16.6.2

Setting C.O. Measurement

Before performing the C.O. measurement, follow this procedure:

1.

Select the C.O. numeric area to enter the C.O.

Measure dialog.

2.

Select the Setup .

3.

Perform the following check or setup:

Check if the height and weight are appropriate for your patient. Change if necessary. The patient's height and weight values are required for determining cardiac index (C.I.).

◆ Check that the correct computation constant is entered. The computation constant has a close relationship with the entered injectate volume, injectate probe type (in-line probe or bath probe) and temperature. See the Instructions for Use of the pulmonary artery catheter to determine correctness.

To change the computation constant, select Comp Const and then input the correct value. When a

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 16 - 3

new catheter is used, the computation constant should be adjusted in accordance with the manufacturer’s instructions for use.

Switch on or off Auto TI . If Auto TI is switched on, the system automatically detects the injectate temperature, and TI setting is disabled. If Auto TI is switched off, you need to input the injectate temperature at TI .

◆ Switch on or off Auto Start . If Auto Start is switched on, the monitor automatically takes the C.O. measurement after establishing a baseline of blood temperature. If Auto Start is switched off, you need to click the Start button in the C.O. Measure dialog for a new measurement.

16.6.3

Performing C.O. Measurement

To perform the C.O. measurement, follow this procedure:

1.

Select the C.O. numeric area to enter the C.O. Measure dialog.

(1)

(2)

16 - 4

(3)

(4)

(5)

(6)

(1) Average values

(3) Current measurement values

(5) Prompt message area

(2) Historical measurement dialogs

(4) Current C.O. curve

(6) Buttons

2.

Proceed as follows to perform the C.O. measure:

◆ If Auto Start is switched off, select the Start button, and then inject the solution quickly when you see the message Please Wait . As shown in the figure above, during the measurement, the currently measured thermodilution curve is displayed. At the end of the measurement, the thermodilution curve is transferred to one of the 6 measurement areas and the monitor prompts you to wait for a certain period of time before starting a new measurement.

◆ If Auto Start is switched on, inject the solution quickly when you see the message Ready For New

Set Of Measurement . The monitor consecutively takes C.O. measurements automatically without the need for pressing the Start button between two measurements. A new thermodilution measurement is possible as soon as the message Inject Now!

is displayed on the screen. The monitor automatically detects further thermodilution measurements.

3.

Acquire the average value of C.O. and C.I. A maximum of 6 measurements can be stored. If you perform more than six measurements without rejecting any, the oldest will automatically be deleted when a seventh curve is stored. Select from the 6 measurement curves and the system will automatically calculate and display the averaged C.O. and C.I. values. Then select the Accept Average button to accept and store the averaged values.

When injecting, the stopcock to the thermodilution catheter is open and the stopcock to the injectate solution is closed. After completing the measurement, turn off the stopcock to the thermodilution catheter and turn on the stopcock to the injectate solution, and then draw the injectate solution into the injectate syringe.

The button area also provides you with the following functions:

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Select Stop to stop the current measurement. Select Setup to enter the C.O.

dialog.

Select Hemo Calcs to enter the Calculations dialog.

NOTE

Starting a measurement without blood temperature being stable may cause measurement failure.

The TB alarms are inactivated during a C.O. measurement, and will be reactivated automatically after the completion of C.O. measurement.

Please see the Instructions for Use of thermodilution catheter to determine the Comp Const and the volume of injectate solution.

16.7

Changing C.O. Settings

16.7.1

Setting C.O. Alarm Properties

To set the C.O. alarm properties, follow this procedure:

1.

Select the C.O. numeric area to enter the C.O.

Measure dialog.

2.

Select Setup to enter the C.O.

dialog.

3.

Select the Alarm tab.

4.

Enter the password if required. For more information, refer to 32.14.3Selecting Password for User

Authentication

.

5.

Set the alarm properties as desired.

16.8

C.O. Troubleshooting

This section lists the problems that might occur. If you encounter the problems when using the equipment or accessories, the table below may help with troubleshooting before requesting a service personnel visit. If the problem persists, contact your service personnel.

NOTE

For the physiological and technical alarm messages, see Appendix DAlarm Messages .

Problem

Do not see C.O. numeric area on the main screen

C.O. value is inaccurate

Solution

1. Check that the C.O. is set to display in the Screen Setup dialog. For

more information, see 3.9.2Displaying Parameter Numerics and

Waveforms .

2. Check that if the C.O. parameter switch is enabled. If not, enable the

C.O. measurement. For more information, see

3.9.1Switching On or Off a

Parameter .

3.Check that the patient type is adult.

4. Check the connection of C.O. cable, thermodilution catheter and TI sensor.

1. Check that the thermodilution catheter is positioned properly.

2. Check that the computational constant is proper for current injectate temperature, injectate volume and injectate probe type.

3. Inject solution rapidly and smoothly.

4. Finish injection within four to five seconds.

5. Inject more volume, or inject colder solution.

6. Check that the height and weight of patient is properly configured.

7. If Auto TI is switched off, check that the entered temperature is correct.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 16 - 5

Problem

C.O. measurement fails

Solution

1. Inject more volume, or inject colder solution. Make sure that the injectate temperature is at least 10°C colder than the patient blood temperature.

2. Finish injection within four to five seconds.

3. Check the connection of C.O. cable, thermodilution catheter and TI sensor.

16 - 6 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

17

Monitoring Central Venous Oxygen Saturation

(ScvO

2

)

17.1

ScvO

2

Introduction

Central venous oxygen saturation (ScvO

2

) is measured across spectrophotometry. Spectrophotometry involves the use of light emitting diodes (LED) that produce light of various wavelengths in red and infrared spectra. The light is transmitted to the blood through a fiberoptic in the probe, reflected off the red blood cells and transmitted back through a separate fiberoptic to an optical module. The central venous oxygen saturation is calculated through the analysis of the reflected spectra.

ScvO

2

monitoring is intended for adult and pediatric patients.

(1)

(1) Calibration hard key (2) CeVOX cable connector

17.2

ScvO

2

Safety Information

WARNING

ScvO

2

monitoring is not intended for neonatal patients.

17.3

ScvO

2

Display

(2)

(2)

(1)

(1) ScvO

2

value

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

(2) SQI bar graph

17 - 1

17.4

Accessing the On-screen ScvO

2

Guide

The monitor provides the on-screen ScvO

2

guide to help you understand ScvO

2

monitoring principles, functions, and operating procedure. To access the on-screen ScvO

2

guide, follow this procedure:

1.

Select the ScvO

2

numeric area to enter the ScvO2 dialog.

2.

Select the Introduction tab.

3.

Select the desired tab of Summary , Target Patients , Points To Note , or Operation Guides .

17.5

ScvO

2

Equipment to Patient Connection

(1) ScvO

2

module

(3) CeVOX fiberoptic probe

(2) CeVOX optical module

(4) Central venous catheter

17.6

Measuring ScvO

2

To perform the ScvO

2

measurements, follow this procedure:

1.

Apply the central venous catheter.

2.

Place one end of the fiberoptic probe into the central venous catheter through the distal lumina, and connect the other end to the CeVOX optical module.

3.

Plug the CeVOX cable into the ScvO

2

module.

4.

If you see the message Calibration Required , calibrate the ScvO

2

before performing the measurements.

For more information, see 17.7 ScvO

2

Calibration

.

5.

Check the reading in the ScvO

2

numeric area.

WARNING

To avoid application failure, ensure that the proper fiberoptic probe is selected.

The monitor may only be regarded as a device providing early warning. If there is an indication of a trend towards de-oxygenation of the patient, blood samples must be taken and tested on a laboratory oximeter in order to arrive at a decision concerning the condition of the patient.

Incorrect placement of the fiberoptic probe can lead to vessel perforation. Therefore, check the correct position of the probe as indicated in the probe’s instructions for use.

17 - 2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

17.7

ScvO

2

Calibration

Regular in vivo calibration is required using blood gas analysis of a central venous blood sample to ensure accurate measurement of continuous ScvO

2

. For optimal accuracy, it is recommended that an in vivo calibration be performed at least every 24 hours or if hemoglobin is changing (for more details, check the notes below).

To perform calibration, follow this procedure:

1.

Check the central venous catheter and CeVOX probe for proper placement.

2.

Check the quality of the signal. The Signal Quality Indicator (SQI) is used for assessing the quality of fiberoptical signals during probe placement, calibration and measurement. The signal quality is indicated by bars of different height levels. Generally, the higher the level, the better the signal.

3.

Withdraw a sufficient amount of central venous blood from the side port of the CeVOX probe to avoid intermixture of infusion/injection with the withdrawn blood.

4.

Slowly withdraw 2ml blood from the side port of the CeVOX probe. Do not pull too strongly in order to avoid a hemolysis.

5.

Immediately confirm by selecting Sample Drawn in the Calibration page of the ScvO2 dialog.

6.

If necessary put blood sample on ice and perform an analysis by a blood gas analysis device or a laboratory oximeter.

7.

Input lab values for Hb, Hct, and ScvO

2

and select Calibrate to confirm.

NOTE

The SQI signal can be affected by the presence of electrosurgical units. Keep electrocautery equipment and cables away from the monitor and use separate power socket if possible.

To achieve optimal accuracy, it is recommended that the entered hemoglobin and hematocrit values are updated when there is a change of 6% or more in hematocrit, or of 1.8 g/dl (1.1 mmol/l) or more in hemoglobin. A change in hemoglobin may also affect SQI.

Dye (e.g. Indocyanine Green) or other substances, containing dyes which usually modify the light absorption capacities, can lead to faulty measurement values of the oxygen saturation.

17.8

Accessing the HemoSight Dialog

To accessing the HemoSight dialog, follow this procedure:

1.

Select the ScvO

2

numeric area to enter the ScvO2 dialog.

2.

Select the HemoSight

button. For information on HemoSight, see 26.6 HemoSight

TM

.

17.9

Changing ScvO

2

Settings

17.9.1

Changing ScvO

2

Alarm Settings

To change the ScvO

2

alarm settings, follow this procedure:

1.

Select the ScvO

2

numeric area to enter the ScvO2 dialog.

2.

Select the Alarm tab.

3.

Enter the password if required. For more information, refer to 32.14.3 Selecting Password for User

Authentication

.

4.

Set the alarm properties as desired.

17.9.2

Setting Hb or Hct

To set the Hb or Hct, follow this procedure:

1.

Select the ScvO

2

numeric area to enter the ScvO2 dialog.

2.

Select the Calibration tab.

3.

Set Hb or Hct .

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 17 - 3

17.9.3

Inputting the SaO

2

Value

The monitor computes the DO

2

and VO

2

using the SaO

2

value. To input the SaO

2 value, follow this procedure:

1.

Select the ScvO

2

numeric area to enter the ScvO2 dialog.

2.

Select the Setup tab.

3.

Input the SaO2 value.

17 - 4 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

18

Monitoring CCO/SvO

2

18.1

CCO/SvO

2

Introduction

The Edwards Lifesciences Vigilance II monitor, Vigileo monitor, EV1000 monitor, and HemoSphere monitor, measure continuous cardiac output (CCO), mixed venous oxygen saturation (SvO

2

), central venous oxygen saturation (ScvO

2

) etc. They also calculate hemodynamic and oxygenation parameters. The N Series monitors can be connected to the Vigilance II/Vigileo/EV1000/HemoSphere monitors and can display, store, and review the measured and calculated parameter values from these monitors. The N Series monitors can also give alarms of these measured parameters; however, you must set alarm on/off, alarm limits, alarm priority, and alarm record separately on the N Series monitors. The alarm is On by default.

The CCO/SvO

2

module is only intended for adult and pediatric patients.

(1) (2)

(3)

(4)

(1) CCO/SvO2 dialog hard key

(2) Calibration key (for Vigilance II and Vigileo monitor )

(3) Module status indicator (4) CCO/SvO

2

cable connector

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 18 - 1

18.2

CCO/SvO

2

Safety Information

WARNING

Because the alarm limits of the relevant measured parameters can be set on the N Series monitors, the alarms of these parameters may be different from those on the Vigilance II/Vigileo/EV1000/

HemoSphere monitors. Please pay special attention to the alarms on the Vigilance II/Vigileo/

EV1000/HemoSphere monitors.

The CCO/SvO

2 monitoring is not intended for neonatal patients.

CAUTION

Observe the operator’s manuals of Vigilance II/Vigileo/EV1000/HemoSphere monitor to configure settings and to connect the monitor to the patient.

The N Series monitors give disconnection alarms when being disconnected from the Vigilance II/

Vigileo/EV1000/HemoSphere monitors. These alarms may be delayed.

NOTE

The Vigilance II monitor, Vigileo monitor, EV1000 monitor, and HemoSphere monitor are manufactured and trademarked by Edwards Lifesciences, and they provide the technology for measuring and calculating the relevant parameters. Mindray only provides a connection and interface to display the parameters from the Vigilance II/Vigileo/EV1000/HemoSphere monitor on the N Series monitors. If you have any doubts about the operation and maintenance of the Vigilance

II/Vigileo/EV1000/HemoSphere monitors, read the operator’s manuals of the corresponding monitor, or contact Edwards Lifesciences (www.edwards.com) directly.

18.3

Connecting the Device

The CCO/SvO

2

cable is used to connect this monitor to the Vigilance II/Vigileo/EV1000/HemoSphere monitor.

18.3.1

CCO/SvO2 Cable

The following figure shows the CCO/SvO

2

cable.

18.3.2

Connecting to the Vigilance II Monitor

The following figure shows the rear housing connectors of the Vigilance II monitor.

18 - 2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

100-240V-

50/60Hz , 1.0A

T1AL 250V

1

2

ECG

1 2 3

(1) (2) (3)

(1) Serial ports (2) ECG signal input port (3) Analog signal input ports

To connect the Vigilance II monitor, follow this procedure:

1.

Insert the end of the CCO/SvO

2

cable marked CN1 to the Mindray CCO/SvO

2

module.

2.

Insert the ECG signal end of the CCO/SvO

2

cable into the ECG signal input port marked on the rear housing of the Vigilance II monitor.

3.

Insert the MAP signal end of the CCO/SvO

2

cable into the analog signal input port 1 marked , the CVP signal end into port 2 marked , and SpO

2

signal end into port 3 marked respectively on the rear housing of the Vigilance II monitor.

4.

Insert UART end of the CCO/SvO

2

cable into either of the serial ports (marked housing of the Vigilance II monitor.

or ) on the rear

5.

Enter the Serial Port Setup menu of the Vigilance II monitor, and make the following settings:

◆ Device : IFMout

◆ Baud Rate : 19200

◆ Parity : None

◆ Stop Bits : 1

◆ Data Bits : 8

◆ Flow Control : 2 seconds

6.

Enter the Analog Input Setup menu of the Vigilance II monitor, and set port 1, port 2 and port 3 as follows:

Setting Port 1 Port 2 Port 3

Parameter

Voltage Range

MAP

0-5 V

CVP

0-5 V

SaO

2

0-10 V

Full Scale Range 500 mmHg (66.7 kPa) 100 mmHg (13.3 kPa) 100%

Simulated High Value 500 mmHg (66.7 kPa) 100 mmHg (13.3 kPa) 100%

Simulated Low Value 0 mmHg (0.0 kPa) 0 mmHg (0.0 kPa) 0%

For more information, see the Vigilance II operator’s manual for the operation of the monitor.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 18 - 3

WARNING

Calibrate the Vigilance II monitor before monitoring. For more information, see the Vigilance II operator’s manual for the calibration instructions.

18.3.3

Connecting to the Vigileo Monitor

The following figure shows the rear housing connectors of the Vigileo monitor.

(1)

(2)

OM

(3)

(4)

18 - 4

(1) Patient Optical Module (OM)

(3) Analog Input

(2) Analog Input

(4) Serial Port

To connect the Vigileo monitor, follow this procedure:

1.

Insert the end of the CCO/SvO

2

cable marked CN1 to the Mindray CCO/SvO

2

module.

2.

Insert the CVP signal end of the CCO/SvO

2

cable into the analog signal input port(2) on the rear housing of the Vigileo monitor.

3.

Insert UART end of the CCO/SvO

2

cable into the serial port on the rear housing(4) of the Vigileo monitor.

4.

Enter the Serial Port Setup menu of the Vigileo monitor, and make the following settings:

◆ Device : IFMout

◆ Baud Rate : 19200

◆ Parity : None

◆ Stop Bits : 1

◆ Data Bits : 8

◆ Flow Control : 2 seconds

5.

Enter the Analog Input Port Setup menu of the Vigileo monitor, and set the CVP as follows:

◆ Parameter : CVP

◆ Voltage Range : 0-5 v

◆ Full Scale Range : 100 mmHg (13.3 kPa)

◆ Simulated High Value : 100 mmHg (13.3 kPa)

◆ Simulated Low Value : 0 mmHg (0.0 kPa)

For more information, see the Vigileo operator’s manual for the operation of the monitor.

WARNING

Calibrate the Vigileo monitor before monitoring. See the Vigileo operator’s manual for the calibration instructions.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

18.3.4

Connecting the EV1000 Monitor

The following figure shows the rear housing of the EV1000 monitor.

(1)

LAN1 LAN2

- +

DC24V

(1) serial port

To connect the EV1000 monitor, follow this procedure:

1.

Insert the end of the CCO/SvO2 cable marked CN1 to the Mindray CCO/SvO2 module.

2.

Insert UART into the serial port(1) on the rear housing of the EV1000 monitor.

3.

Enter the Serial Port Setup menu of the EV1000 monitor, and make the following settings:

Device : IFMout

Baud Rate : 19200

Parity : None

Stop Bits : 1

Data Bits : 8

Flow Control : 2 seconds

For more information, see the EV1000 operator’s manual for the operation of the monitor.

WARNING

Calibrate the EV1000 monitor before monitoring. Refer to the EV1000 Operator’s Manual for the calibration instructions.

18.3.5

Connecting the HemoSphere Monitor

The following figure shows the rear housing of the HemoSphere monitor.

.*3/

(2)

(3)

+)-

(4)

(1)

(1) COM1 Serial Port

(3) Analog Input

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

(2) Analog Input

(4) ECG Input

18 - 5

To connect the HemoSphere monitor, follow this procedure:

1.

Connect CN1 with the CCO/SvO2 cable connector of the CCO/SvO2 module.

2.

Insert the UART signal end of the CCO/SvO

2

cable into the COM1 serial port on the rear housing of the

HemoSphere monitor.

3.

Insert the CVP and MAP signal ends of the CCO/SvO

2

cable into the Analog Inputs on the rear housing of the HemoSphere monitor.

4.

Insert ECG signal end of the CCO/SvO

2

cable into the ECG Input on the rear housing of the HemoSphere monitor.

5.

Enter the Serial Port Setup menu of the HemoSphere monitor, and make the following settings:

◆ Device : IFMout

◆ Baud Rate : 19200

◆ Parity : None

◆ Stop Bits : 1

◆ Data Bits : 8

◆ Flow Control : 2 seconds

6.

Enter the Analog Input Setup menu of the HemoSphere monitor, and set as follows:

Parameter

Voltage Range

Full Scale Range

Simulated High Value

Simulated Low Value

MAP

0-5 v

500 mmHg (66.7 kPa)

500 mmHg (66.7 kPa)

0 mmHg (0.0 kPa)

CVP

0-5 v

100 mmHg (13.3 kPa)

100 mmHg (13.3 kPa)

0 mmHg (0.0 kPa)

For more information, see the HemoSphere operator’s manual for the operation of the monitor.

WARNING

Calibrate the HemoSphere monitor before monitoring. Refer to the HemoSphere Operator’s Manual for the calibration instructions.

18.4

Accessing the HemoSight dialog

To access the HemoSight dialog, follow this procedure:

1.

Select the CCO numeric area to enter the CCO dialog.

2.

Select the HemoSight button. For more information, see

26.6 HemoSight

TM

.

18.5

Changing CCO Settings

18.5.1

Changing the CCO Alarm Settings

To change the CCO alarm settings, follow this procedure:

1.

Select the CCO numeric area to enter the CCO dialog.

2.

Select the Alarm tab.

3.

Set the alarm properties of CCO and CCI.

18.5.2

Setting Parameters for Display

To set the parameters for display, follow this procedure:

18 - 6 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

1.

Select the CCO numeric area to enter the CCO dialog.

2.

Select the Select Parameter tab.

3.

Select the primary and secondary parameters for display.

18.5.3

Setting the CCO Analog Output Signal

To set the CCO output signal, follow this procedure:

1.

Select the CCO numeric area to enter the CCO dialog.

2.

Select the Signal Output Setup tab.

3.

Set the output signal as follows:

This monitor can output the analog signals of ECG waveform, MAP value, SpO

2

value and CVP value to the Vigilance II monitor. If a signal has several sources, you can select a source.

◆ This monitor can output the CVP analog signal to the Vigileo monitor. If the CVP signal has several sources, you can select a source.

◆ This monitor can output ECG waveform, MAP value, and CVP value to the HemoSphere monitor. If a signal has several sources, you can select a source.

◆ Select Simulated High Value to output the simulated high value calibration signals to Vigilance II,

Vigileo, or HemoSphere monitor. To stop outputting the simulated high value signal, select Stop Cal

Signal Output where the Simulated High Value button was previously.

◆ Select Simulated Low Value to output the simulated low value calibration signals to Vigilance II,

Vigileo, or EV1000 monitor. To stop outputting the simulated low value signal, select Stop Cal Signal

Output where the Simulated Low Value button was previously.

The following table shows values and voltages of the high and low value calibration signals.

Parameter Values Output Voltage Parameters

High Value Calibration Signals

MAP

SpO

2

CVP

Low Value Calibration Signals

MAP

SpO

2

CVP

500 mmHg

100%

100 mmHg

0 mmHg

0%

0 mmHg

0V

0V

0V

5V

10V

5V

CAUTION

The calibration voltage of Vigilance II, Vigileo, EV1000, and HemoSphere monitor should be the same as the output voltage of this monitor. Otherwise, some parameter values may be incorrectly calculated.

NOTE

See the operator’s manuals of Vigilance II, Vigileo, EV1000, and HemoSphere monitors for the calibration instructions.

18.6

Changing SvO

2

/ScvO

2

Settings

18.6.1

Changing the SvO

2

/ScvO

2

Alarm Settings

To change the SvO

2

/ScvO

2

alarm settings, follow this procedure:

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 18 - 7

1.

Select the SvO

2

/ScvO

2

numeric area to enter the SvO2 or ScvO2 dialog.

2.

Select the Alarm tab.

3.

Set the alarm properties of SvO

2

/ScvO

2

.

18.6.2

Setting the SvO

2

/ScvO

2

Analog Output Signal

This monitor can output the analog signals to the Vigilance II/Vigileo/EV1000/HemoSphere monitor. To set the

SvO

2

/ScvO

2

analog output signal, follow this procedure:

1.

Select the SvO

2

numeric area to enter the SvO2 dialog; or select the ScvO

2

numeric area to enter the

ScvO2 dialog.

2.

Select the Signal Output Setup tab.

3.

Set the output signal as follows:

◆ This monitor can output the analog signals of ECG waveform, MAP value, SpO

2

value and CVP value to the Vigilance II monitor. If a signal has several sources, you can select a source.

◆ This monitor can output the CVP analog signal to the Vigileo monitor. If the CVP signal has several sources, you can select a source.

◆ This monitor can output ECG waveform, MAP value, and CVP value to the HemoSphere monitor. If a signal has several sources, you can select a source.

◆ Select Simulated High Value to output the simulated high value calibration signals to Vigilance II,

Vigileo, or HemoSphere monitor. To stop output the simulated high value signal, select Simulated

High Value again.

◆ Select Simulated Low Value to output the simulated low value calibration signals to Vigilance II,

Vigileo, or HemoSphere monitor. To stop output the simulated low value signal, select Simulated

Low Value again.

The following table shows values and voltages of the high and low value calibration signals.

Parameter Values Output Voltage Parameters

High Value Calibration Signals

MAP

SpO

2

CVP

Low Value Calibration Signals

MAP

SpO

2

CVP

500 mmHg

100%

100 mmHg

0 mmHg

0%

0 mmHg

0V

0V

0V

5V

10V

5V

CAUTION

The calibration voltage of Vigilance II, Vigileo, EV1000, and HemoSphere monitor should be the same as the output voltage of this monitor. Otherwise, some parameter values may be incorrectly calculated.

NOTE

See the operator’s manuals of Vigilance II, Vigileo, EV1000, and HemoSphere monitors for the calibration instructions.

18 - 8 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

18.7

CCO/SvO

2

Troubleshooting

This section lists the problems that might occur. If you encounter the problems when using the equipment or accessories, check the table below before requesting for services. If the problem persists, contact your service personnel.

NOTE

For the physiological and technical alarm messages, see Appendix D Alarm Messages .

Problem

The numeric area does not display CCO values when the Vigilance II, Vigileo, EV1000, or

HemoSphere monitor is connected.

Do not see ScvO

2

or SvO

2

numeric area on the main screen

Solution

1. Check that the CCO is set to display in the Screen Setup

Tile

Layout

page. For more information, see 3.9.2 Displaying Parameter

Numerics and Waveforms

.

2. Check that if the CCO parameter switch is enabled. If not, enable the

CCO measurement. For more information, see

3.9.1 Switching On or Off a

Parameter .

3.Check that the Vigilance II, Vigileo, EV1000, or HemoSphere monitor is set properly as described in

18.3 Connecting the Device .

4. Check the connection of CCO/SvO

2

cable. For more information, see

18.3 Connecting the Device

.

1. Check that the ScvO

2

or SvO

2

is set to display in the Screen Setup

Tile Layout page. For more information, see

3.9.2 Displaying Parameter

Numerics and Waveforms

.

2. Check that if the ScvO

2

/SvO

2

parameter switch is enabled. If not, enable the ScvO

2

/SvO

2

measurement. For more information, see

3.9.1 Switching On or Off a Parameter .

3.Check that the Vigilance II , Vigileo, EV1000, or HemoSphere monitor is set properly as described in

18.3 Connecting the Device .

4. Check the connection of CCO/SvO

2

cable. For more information, see

18.3 Connecting the Device

.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 18 - 9

This page intentionally left blank.

18 - 10 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

19

Monitoring Carbon Dioxide (CO

2

)

19.1

CO

2

Introduction

CO

2

monitoring is a continuous, non-invasive technique for determining the concentration of CO

2

in the patient’s airway by measuring the absorption of infrared (IR) light of specific wavelengths. CO

2

has its own absorption characteristic and the amount of light passing the gas probe depends on the concentration of the measured CO

2

. When a specific band of IR light passes through respiratory gas samples, some of IR light will be absorbed by the CO

2

molecules. The amount of IR light transmitted after it has been passed through the respiratory gas sample is measured with a photodetector. From the amount of IR light measured, the concentration of CO

2

is calculated.

CO

2

measurement is used to monitor the patient’s respiratory status. The sidestream/Microstream CO

2 measurement takes a sample of the respiratory gas with a constant sample flow from the patient's airway and analyzes it with a remote CO

2

sensor built into the Sidestream or Microstream CO

2

module.

The sidestream and microstream CO

2

measurement can be used, with specified accessories, with intubated and non-intubated adult, pediatric, and neonatal patients. With intubated patients, a sample of the respiratory gas is drawn from the patient's breathing circuit through an airway adapter and a gas sampling line. With nonintubated patients, the gas sample is drawn through a nasal cannula.

Both sidestream and microstream CO

2

modules have the function of automatic barometric pressure compensation (the system automatically measures the barometric pressure to which the patient monitor is exposed).

CO

2

monitoring is intended for adult, pediatric and neonatal patients.

19.2

Identifying CO

2

Modules

The following figures show the controls and connectors of different CO

2

module.

(1) (2) (1) (2)

(4)

(3)

(5)

(4)

Sidestream CO

2

module Microstream CO

2

module

(1) CO

2

dialog hard key

(3) CO

2

watertrap seat

(5) Sample line connector

(2) CO

2

Measure/standby hard key

(4) Gas outlet

To measure CO

2

using the AG module, see

20 Monitoring Anesthetic Gas (AG)

.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 19 - 1

19.3

CO

2

Safety Information

WARNING

Route all tubing away from the patient’s throat to avoid strangulation.

CAUTION

Remove the airway sample line from the patient’s airway while nebulized medications are being delivered. Measuring CO

2

during nebulization may lead to inaccurate CO

2

readings.

EtCO

2

values measured from the CO

2

module may differ from those from the blood gas analysis.

Avoid mechanical shock to the sidestream CO

2

module.

Shaking the CO2 module during O2 measurement may lead to a distorted O2 waveform or inaccurate O2 measurement

NOTE

The CO

2

module automatically suppresses physiological alarms until breathing waves have been detected. Ensure that the patient is properly connected when using the CO

2

module.

19.4

CO

2

Measurement Limitations

The following factors may influence the measurement accuracy:

Leaks or internal venting of sampled gas

Mechanical shock

Cyclic pressure up to 10 kPa (100 cmH

2

Other sources of interference, if any

O)

For more information, see

A.13.12 CO

2

Specifications

.

CAUTION

Measurement accuracy of the sidestream CO

2

module may be affected by the breath rate and inspiration/expiration (I/E) ratio.

Measurement accuracy of the microstream CO

2

module may be affected by the breath rate.

19.5

CO

2

Display

The CO

2

numeric and waveform areas provide FiCO

2

measurement, EtCO

2

measurement, awRR measurement, and a CO

2

waveform.

19 - 2

(1)

(1) CO

2

waveform

(3) End tidal CO

2

value (EtCO

2

)

(2) (3) (4)

(2) Fraction of inspired CO

2

(FiCO

2

)

(4) Airway respiration rate (awRR)

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

19.6

Measuring CO

2

Using the Sidestream/Microstream CO

2

Module

19.6.1

Preparing to Measure CO

2

Using the Sidestream CO

2

Module

To prepare the CO

2

measurement, follow this procedure:

1.

Select the appropriate gas sample line and watertrap according to the patient category.

2.

Connect the watertrap to the CO

2

module, and connect the gas sample line to the watertrap.

(1)

(2)

(3)

(1) Watertrap receptacle

(3) Gas sample line

(2) DRYLINE II watertrap

3.

Connect the other end of the gas sample line to the patient.

◆ For intubated patients requiring an airway adapter, install the airway adapter between the patient circuit and the ventilator Y-piece.

(1)

(2)

(3)

(4)

(1) Sample line

(3) Airway adapter

(2) Connection to the ventilator

(4) Connection to the patient

For non-intubated patients, place the nasal cannula onto the patient.

4.

Connect the gas outlet to the scavenging system using an exhaust tube.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 19 - 3

After the CO

2

module is connected, it enters measure mode by default and the monitor displays CO2 Starting .

CO

2

can be measured after the start-up sequence is complete.

WARNING

Do not use the adult or pediatric watertrap with a neonate patient. Otherwise, patient injury could result.

Connect the gas outlet to the scavenging system when measuring CO

2

using the sidestream CO

2 module.

CAUTION

Leakage in the breathing or sampling system may cause the displayed EtCO

2

values to be significantly low. Always make sure that all components are securely connected.

Inspect the airway adapter for a tight connection and proper operation before attaching it to the patient.

Squeezing or bending the sample line during sidestream or microstream CO

2

measurement may cause inaccurate CO

2

readings or no reading.

To avoid blocking the airway, empty the DRYLINE II watertrap container whenever half full.

Replacing the DRYLINE II watertrap once a month is recommended. Dispose of accumulated fluids in accordance with hospital policy or your local regulations.

The DRYLINE II watertrap has a filter preventing bacterium, water and secretions from entering the module. Extended use could destroy the filter in watertrap and fail to stop the bacterium, water and secretions entering the module, result in damaging the gas module and having infection risk.

NOTE

To extend the lifetime of the watertrap and module, disconnect the watertrap from the module and set the operating mode to Standby mode when CO

2

monitoring is not required.

The sample rates are different when different types of watertraps are used.

The emptying interval of the DRYLINE II adult/pediatric watertrap is 26 hours @ 120 ml/min, sample gas of 37 °C, room temperature of 23 °C, and 100% RH.

The emptying interval of the DRYLINE II neonatal watertrap is 35 hours @ 90 ml/min, sample gas of

37 °C, room temperature of 23 °C, and 100% RH.

19.6.2

Preparing to Measure CO

2

Using the Microstream CO

2

Module

To prepare the CO

2

module for measurement, follow this procedure:

1.

Connect one end of the sample line to the microstream CO

2

module.

(1)

19 - 4

(2)

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

(1) Sample line connector (2) Sample line

2.

Connect the other end of the sample line to the patient.

◆ For intubated patient requiring an airway adapter, install the airway adapter between the patient circuit and the ventilator Y-piece.

For non-intubated patient, place the nasal cannula onto the patient.

◆ For patient prone to mouth breathing, place the oral-nasal cannula onto the patient.

3.

Connect the gas outlet to the a scavenging system using an exhaust tube.

After the CO

2

module is connected, it enters measure mode by default and the monitor displays CO2 Sensor

Warmup . CO

2

can be measured after the start-up sequence is complete.

WARNING

Connect the gas outlet to the scavenging system when measuring CO

2

using the microstream CO

2 module.

NOTE

Disconnect the sample line from the module when CO

2

monitoring is not required.

19.6.3

Zeroing the Sidestream/Microstream CO

2

Module

The sidestream and microstream CO

2

modules perform a zero calibration automatically when needed. Once the zero calibration is started, the CO

2

module stops measuring and “ Zeroing ” is displayed in the CO

2

numeric area.

After the zero calibration is completed, the CO

2

module reacquires the CO

2

readings. During the reacquisition period, “ Zero Recovering ” is displayed in the CO

2

numeric area. Valid data will reappear 30 seconds after the zero calibration is started. You can hide the display of the “ Zero Recovering ” message, but values displayed

during the reacquisition period may not be accurate. See 32.5.6 Hiding the Invalid Display after Zeroing the CO

2

Module for more information.

The automatic zero calibration will not start under the following conditions:

Physiological alarms related to CO

An apnea alarm is active.

2

or AG are active.

No breath has been detected for over 30 seconds.

You can also perform the zero calibration manually. For more information, see 32.5.5 Manually Zeroing the CO

2

Module

.

19.7

Changing Settings for All CO

2

Modules

19.7.1

Changing CO

2

Alarm Settings

To change the CO

2

alarm settings, follow this procedure:

1.

Select the CO

2

numeric area or waveform area to enter the CO2 dialog.

2.

Select the Alarm tab.

3.

Enter the password if required. For more information, refer to 32.14.3 Selecting Password for User

Authentication

.

4.

Set alarm properties as desired.

19.7.2

Setting the CO

2

Waveform

To set the CO

2

waveform, follow this procedure:

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 19 - 5

1.

Select the CO

2

numeric area or waveform area to enter the CO2 dialog.

2.

Select the Setup tab.

3.

Set Waveform Type , Speed, Scale , or CO2 Scale of the CO

2

waveform.

19.7.3

Setting the RR Source

To set the respiration rate (RR) source, follow this procedure:

1.

Select the CO

2

numeric area or waveform area to enter the CO2 dialog.

2.

Select the Setup tab.

3.

Set RR Source .

When you select Auto , the system automatically selects the RR source according to the priority. The priority of

RR source is first CO

2

, and then ECG. When the current RR source does not have valid measurement, the system automatically switches the RR Source to Auto .

19.7.4

Entering the Standby Mode

Set the CO

2

module to one of the following modes according to the module status:

Select Measure mode when using the CO

2

module for monitoring.

Select Standby mode to prolong the CO

2

module service life when not using the CO

2

module.

The default operating mode is Measure . When not using the CO

2

module, proceed as follows to enter the

Standby mode:

1.

Select the CO

2

numeric area or waveform area to enter the CO2 dialog.

2.

Select the Setup tab.

3.

Set Operating Mode to Standby .

19.7.5

Entering the Intubation Mode

When performing intubation during general anesthesia, enter the intubation mode to reduce unnecessary alarms. To enter the intubation mode, follow this procedure:

1.

Select the CO

2

numeric area or waveform area to enter the CO2 dialog.

2.

Select Intubation Mode .

To exit the intubation mode, choose either of the following ways:

From the bottom of the CO2 dialog, select Exit Intubation Mode .

Select the Main Menu quick key → from the Alarm column select Exit Intubation Mode .

For the details of the intubation mode, see

8.13 Intubation Mode

.

NOTE

When the configured intubation period is over, the monitor automatically exits the intubation mode and gives two alert sounds.

19.7.6

Setting the Auto Standby

The monitor enters standby mode automatically if no breath is detected after the configured period of time. To set the auto standby, follow this procedure:

1.

Select the CO

2

numeric area or waveform area to enter the CO2 dialog.

2.

Select the Setup tab.

3.

Set Auto Standby .

19 - 6 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

19.7.7

Setting Humidity Compensation

Sidestream and microstream CO

2

modules are configured to compensate CO

2

readings for either Body

Temperature and Pressure, Saturated Gas (BTPS), to account for humidity in the patient’s breath, or Ambient

Temperature and Pressure, Dry Gas (ATPD).

ATPD: P

CO2

(mmHg)=CO

2

(vol%) x P amb

/100

BTPS (sidestream): P

CO2

(mmHg)=CO

2

(vol%) x (P amb

- 47)/100

BTPS (microstream): P

CO2

(mmHg)=CO

2

(vol%) x (1- 0.03)x P amb

/100

Where, P

CO2

(mmHg)= partial pressure, vol%=CO

2

concentration, P amb

=ambient pressure, and unit is mmHg.

For the sidestream and microstream CO

2

module, set the humidity compensation on or off according to the actual condition.

To set the humidity compensation, follow this procedure:

1.

Select the CO

2

numeric area or waveform area to enter the CO2 dialog.

2.

Select the Setup tab.

3.

Set BTPS Compensation .

◆ Switch on for BTPS.

◆ Switch off for ATPD.

19.8

Setting the Gas Compensation

The presence of interfering gas affects the CO

2

measurement. To get the best possible measurement result, it is recommended to set the gas compensation. The configured concentration of the interfering gas should be in accordance with its actual proportion.

For the microstream CO

2

module, gas compensations are not required.

CAUTION

Make sure to use the appropriate compensations. Inappropriate compensations may cause inaccurate measurement values and result in misdiagnosis.

For the sidestream CO

2

module, follow this procedure to set the gas compensation:

1.

Select the CO

2

numeric area or waveform area to enter the CO2 dialog.

2.

Select the Setup tab.

3.

Set the compensation according to the actual condition.

19.9

Choosing a Time Interval for Peak-Picking

For microstream CO

2

modules, select a time interval for picking the highest CO

2

as the EtCO

2

and the lowest as the FiCO

2

.

To set the time interval, follow this procedure:

1.

Select the CO

2

numeric area or waveform area to enter the CO2 dialog.

2.

Select the Setup tab.

3.

Set Maximum Hold .

4.

Toggle between Single Breath , 10 s and 20 s .

Single Breath : EtCO

2

and FiCO

2

are calculated for every breath.

10 s and 20 s : EtCO

2

and FiCO

2

are calculated using 10 or 20 seconds of data.

19.10

Performing the Leakage Test

When measuring CO

2

using the sidestream CO

2

module, the leakage test is required every time before the CO

2 measurement. To perform the CO

2

leakage test, follow this procedure:

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 19 - 7

1.

Connect the measuring accessories as per section

19.6.1 Preparing to Measure CO

2 Using the Sidestream CO2

Module

.

2.

Wait until the startup finishes. Completely block the gas inlet on the sidestream CO

2

module. The alarm message “ CO2 Airway Occluded ” will appear on the screen.

3.

Continue to block the gas inlet for another one minute.

4.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

5.

Select the Module tab

CO2 tab.

6.

Check that the current flow rate is less than 10ml/min, and the alarm message “ CO2 Airway Occluded ” does not disappear.

This indicates that the module does not leak. If the alarm message disappears, or the flow rate is equal to

10ml/min or greater, it indicates that the module leaks. Perform the leakage test again. If the problem remains, contact your service personnel for help.

19.11

CO

2

Calibration

For sidestream and microstream CO

2

modules, a calibration is needed every year or when the measured values have a great deviation.

To calibrate the CO

2

module, contact your service personnel.

CAUTION

Connect the gas outlet to the scavenging system when calibrating the CO

2

module.

19.12

Test Method Used to Determine the Respiration Rate Measuring

Range

To determine the RR measurement range, follow this procedure:

1.

Connect the 5% CO

2

+ N monitor.

2

(BAL) gas cylinder, reduction valve, rhythm generator and gas sampling path of a

2.

Respectively set the rhythm generator to generate a respiration rate of 0 rpm, 6 rpm, 20 rpm, 60 rpm,

120rpm, and 150 rpm.

3.

Verify that it meets the specifications.

NOTE

The rhythm generator mentioned above is a sensor/sampling site in Figure 201.101 – Test apparatus for the TOTAL SYSTEM RESPONE TIME of an RGM of ISO 80601-2-55.

19.13

CO

2

Troubleshooting

This section lists the problems that might occur. If you encounter the problems when using the equipment or accessories, check the table below before requesting service. If the problem persists, contact your service personnel.

NOTE

For the physiological and technical alarm messages, see Appendix D Alarm Messages

.

19 - 8 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

19.13.1

Troubleshooting the Sidestream/Microstream CO

2

Module

Problem

EtCO

2

measurements too low

Solution

1. Check the patient status.

2. Check the sample line and connectors for leakage.

3. Ventilate the room if the environmental CO

2

concentration is too high.

19.14

Oridion Information

Oridion Patents

This posting serves as notice under 35 U.S.C.§287(a) for Covidien patents: http://www.covidien.com/patents.

No Implied License

Possession or purchase of this device does not convey any express or implied license to use the device with unauthorized CO

2 sampling consumables which would, alone, or in combination with this device, fall within the scope of one or more of the patents relating to this device and/or CO2 sampling consumable.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 19 - 9

This page intentionally left blank.

19 - 10 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

20

Monitoring Anesthetic Gas (AG)

20.1

AG Introduction

The anesthetic gas (AG) module measures the patient’s anesthetic and respiratory gases by connecting to the airway of intubated patients or collecting the gases with specified accessories. It also incorporates the features of the O

2

module and BIS module.

The AG module determines the concentration of certain gases using the infrared (IR) light absorption measurement. The gases that can be measured by the AG module absorbing IR light. Each gas has its own absorption characteristic. The gas is transported into a sample cell, and an optical IR filter selects a specific band of IR light to pass through the gas. For multiple gas measurements, there are multiple IR filters. The higher the concentration of gas in a given volume, the more IR light is absorbed. This means that higher concentration of IR absorbing gas causes a lower transmission of IR light. The amount of IR light transmitted after it has been passed through an IR absorbing gas is measured. From the amount of IR light measured, the concentration of gas present can be calculated.

Oxygen does not absorb IR light as other breathing gases and is therefore measured relying on its paramagnetic properties. Inside the O

2

sensor are two nitrogen-filled glass spheres mounted on a strong rare metal taut-band suspension. This assembly is suspended in a symmetrical non-uniform magnetic field. In the presence of paramagnetic oxygen, the glass spheres are pushed further away from the strongest part of the magnetic field.

The strength of the torque acting on the suspension is proportional to the oxygen concentration. From the strength of the torque, the concentration of oxygen is calculated.

AG monitoring is intended for adult, pediatric and neonatal patients.

(1)

(2)

(3)

(4)

(5)

(1) AG dialog hard key

(3) AG Measure/standby hard key

(5) BIS cable connector

(2) Gas outlet

(4) Watertrap seat

NOTE

The AG module is configured with automatic barometric pressure compensation function.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 20 - 1

20.2

AG Safety Information

WARNING

To avoid explosion hazard, do not use flammable anesthetic agent such as ether and cyclopropane for this equipment.

The presence of other substances in the patient’s breathing circuit, such as ethanol, acetone, methanol, isopropanol, freon, asthma medication carrier gases, and other infrared absorbing gases, can influence the anesthesia agent identification and lead to incorrect measurements and identification.

Using high-frequency electrosurgical units may increase the risk of skin burn. In this case, do not use antistatic or conductive respiratory tubing.

Route all tubing away from the patient’s throat to avoid strangulation.

CAUTION

Perform the measurement in a well-ventilated environment.

EtCO

2

values measured from the AG module may differ from that of from the blood gas analysis.

NOTE

The AG module automatically suppresses physiological alarms until breathing waves have been detected. Make sure that a patient is properly connected when monitoring with the AG module

20.3

AG Measurement Limitations

The following factors may influence the measurement accuracy:

Leaks or internal venting of sampled gas

Mechanical shock

Cyclic pressure up to 10 kPa (100 cmH

2

Other sources of interference, if any

O)

20 - 2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

20.4

AG Display

The AG module can send waves and numerics for all measured anesthetic gases for display on the monitor, including:

CO

2

, O

2

, N

2

O and AA waves awRR: airway respiratory rate

MAC: minimum alveolar concentration

End tidal (Et) and fraction of inspired (Fi) numerics for CO

2

, O

2

, N

2

O and AA

AA represents one of the following agents: Des (desflurane), Iso (isoflurane), Enf (enflurane), Sev (sevoflurane), or

Hal (halothane).

If only one anesthetic agent is used, the AA waveform area displays the waveform of this anesthetic agent. If several anesthetic agents are used, the AA waveform area displays the waveform of the primary anesthetic agent.

20.5

AG Equipment to Patient Connection

(5)

(4)

( )

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 20 - 3

(1) AG module

(3) Watertrap

(5) Airway adapter (connected to patient)

(2) Gas outlet

(4) Gas sample line

(6) Y-piece (connected to the anesthesia machine)

20.6

Preparing for AG Monitoring

To prepare to monitor AG, follow this procedure:

1.

Select the appropriate gas sample line and watertrap according to the patient category.

2.

Connect the watertrap to the AG module, and connect the gas sample line to the watertrap.

3.

Connect the other end of the gas sample line to the patient via the airway adapter.

4.

Connect the gas outlet to a scavenging system or the patient respiration circuit using an exhaust tube.

5.

Check that the connections are tight.

After the AG module is connected to the SMR, the AG module enters the measurement mode by default and the monitor prompts AG Starting . AG measurement is available after the start-up is completed.

WARNING

When using the AG module, connect the gas outlet to the scavenging system or the patient respiration circuit when in use with the Mindray A7 anesthesia machine.

If the sampled gas is returned to the breathing system, there may be a risk of patient crossinfection.

If the sampled gas is returned to the breathing system, install a breathing system filter compliant with the ISO 23328-1 and ISO 23328-2 standards on the patient end.

Make sure that the connections are tight. Any leak in the system can result in erroneous readings due to ambient air mixing with patient gases.

Always inspect the airway adapter for a tight connection and proper operation before attaching it to the patient.

Squeezing or bending the gas sample line during AG measurement may cause erroneous readings or no readings.

CAUTION

Position the airway adapter so that the part connecting to the gas sample line is pointing upwards.

This prevents condensed water from passing into the gas sample line and causing an occlusion.

The watertrap collects water drops condensed in the sample line and therefore prevents them from entering the module. To avoid blocking the airway, empty the watertrap container whenever half full.

The watertrap has a filter preventing bacterium, water and secretions from entering the module.

After long-term use, dust or other substances may compromise the performance of the filter or even block the airway. In this case, replace the watertrap. Replacing the watertrap once a month is recommended.

NOTE

Do not apply adult watertrap to the neonatal patient. Otherwise, patient injury could result.

To extend the lifetime of the watertrap and module, disconnect the watertrap from the module and set the operating mode to Standby when AG monitoring is not required.

20.7

Zeroing the AG Module

The AG module performs a zero calibration automatically when needed. Once the zero calibration is started, the

AG module stops measuring and “ Zeroing ” is displayed in the AG numeric area.

20 - 4 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

After the zero calibration is completed, the AG module reacquires the AG readings. During the reacquisition period, “ Zero Recovering ” is displayed in the AG numeric area. Valid data will reappear 30 seconds after the zero calibration is started. You can hide the display of the “ Zero Recovering ” message, but values displayed during

the reacquisition period may not be accurate. See 32.5.8 Hiding the Invalid Display after Zeroing the AG Module for

more information.

The automatic zero calibration will not start under the following conditions:

Physiological alarms related to CO

2

An apnea alarm is active.

or AG are active.

No breath has been detected for over 30 seconds.

You can also perform the zero calibration manually. For more information, see 32.5.7 Manually Zeroing the AG

Module .

20.8

MAC Values

Minimum alveolar concentration (MAC) is the minimum concentration of the agent in the alveoli. It is a basic index to indicate the depth of anesthesia. The standard ISO 80601-2-55 defines MAC as this: alveolar concentration of an inhaled anesthetic agent that, in the absence of other anesthetic agents and at equilibrium, prevents 50% of patients from moving in response to a standard surgical stimulus.

MAC values are listed below:

Agent Des Iso Enf Sev

1 MAC 6% 1.15% 1.7% 2.1%

* indicates 1 MAC nitrous oxide can only be reached in hyperbaric chamber.

Hal

0.77%

N

2

O

105%*

NOTE

The types of gases used for calculating the MAC value can be selected. For more information, see

32.5.5 Manually Zeroing the CO

2

Module .

The MAC values shown in the table above are those published by the U.S. Food and Drug

Administration for a healthy 40-year-old adult male patient.

In actual applications, the MAC value may be affected by age, weight and other factors.

The formula to calculate the MAC value is as follows:

MAC =  i = 0

AgentVol age i

Where N is the number of all agents (including N

2

O) that the AG module can measure, EtAgent i

is the concentration of each agent, and AgentVol age i is the concentration of each agent at 1 MAC with age correction.

The formula for calculating age correction of 1 MAC is:

MAC age

= MAC

40

 10

 – 0.00269

  age 40  

For example, the Des concentration at 1 MAC of a 60-year old patient is .

6% 10

 – 0.00269

  60 40  

=

The AG module measures 4% of Des, 0.5% of Hal and 50% of N

2

O in the patient’s end-tidal gas:

MAC = + + = 2.04

NOTE

The formula above is only suitable for patients who are older than one year. If the patient is less than one year, the system uses one year old to do age correction.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 20 - 5

20.9

Changing AG Settings

20.9.1

Changing AG Alarm Settings

To change the AG alarm settings, follow this procedure:

1.

Select the AG numeric area or waveform area to enter the Gas dialog.

2.

Select the desired gas tab.

3.

Select the Alarm tab.

4.

Enter the password if required. For more information, refer to 32.14.3 Selecting Password for User

Authentication .

5.

Set alarm properties as desired.

20.9.2

Setting the O

2

Compensation

If the AG module does not incorporate the O

2

module, you need to set the amount of O

2

in the ventilation gas mixture. To set the O

2

compensation, follow this procedure:

1.

Select the AG numeric area or waveform area to enter the Gas dialog.

2.

Select the Setup tab.

3.

Set O2 Compensation :

◆ Select Off when the amount of O

2

is less than 30%.

◆ Select the other options in accordance with the O

2

concentration in the gas mixture.

The O2 Compensation setting is available only when the AG module does not incorporate the O

2

module. If the

AG module incorporates the O

2

module, the system directly uses the O module to make compensation.

2

concentration detected by the O

2

20.9.3

Entering the Standby Mode

Set the AG module to one of the following modes according to the module status:

Select

Select

Measure

Standby

mode when using the AG module for monitoring.

mode when not using the AG module.

The default operating mode is Measure . If not using the AG module, follow this procedure to enter the Standby mode:

1.

Select the AG numeric area or waveform area to enter the Gas dialog.

2.

Select the desired gas tab.

3.

Select the Setup tab.

4.

Set Operating Mode to Standby .

20.9.4

Setting Auto Standby

The monitor enters the standby mode automatically after the configured period of time if no breath is detected since the last detected breath. To set the auto standby, follow this procedure:

1.

Select the AG numeric area or waveform area to enter the Gas dialog.

2.

Select the desired gas tab.

3.

Select the Setup tab.

4.

Set Auto Standby .

20 - 6 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

20.9.5

Setting the Gas Waveform

To set the gas waveform, follow this procedure:

1.

Select the AG numeric area or waveform area to enter the Gas dialog.

2.

Select the desired gas tab.

3.

Select the Setup tab.

4.

Set the speed and scale of gas waveforms. For CO

2

, set Waveform Type .

20.9.6

Setting the RR Source

To set the RR (respiration rate) source, follow this procedure:

1.

Select the AG numeric area or waveform area to enter the Gas dialog.

2.

Select the desired gas tab.

3.

Select the Setup tab.

4.

Set RR Source .

When the current RR source does not have valid measurement, the system will automatically switch RR Source to

Auto .

20.9.7

Entering the Intubation Mode

When performing intubation during general anesthesia, enter the intubation mode to reduce unnecessary alarms. To enter the intubation mode, follow this procedure:

1.

Select the AG numeric area or waveform area to enter the Gas dialog.

2.

Select Intubation Mode from the bottom of the dialog.

For the details of the intubation mode, see

8.13 Intubation Mode .

20.9.8

Enabling or Disabling MAC Display

Set whether MAC value is displayed in the AG numeric area by following this procedure:

1.

Select the AG numeric area or waveform area to enter the Gas dialog.

2.

Select the desired anesthetic agent tab.

3.

Switch on or off MAC .

20.10

Changing the Anesthetic Agent

When the anesthetic agent used on the patient is changed, the AG module detects the mixed anesthetic gas during the transition of two anesthetic agents. The time required for completing the replacement of anesthetic agent depends on anesthesia type (low flow or high flow) and the characteristics of anesthetic agents

(pharmacokinetics). During the transition of two anesthetic agents, the monitor gives no prompt messages and the MAC value displayed may be inaccurate.

The AG module can identify two anesthetic agents automatically. When the proportion of primary and secondary anesthetic agents in the mixture changes, the AG module can distinguish between them according to their contributions to the MAC value. Then primary and secondary anesthetic agents will be exchanged for display.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 20 - 7

20.11

Performing AG Leakage Test

The AG leakage test is required every time before the AG measurement. To perform the AG leakage test, follow this procedure:

1.

Plug the AG module into the module rack.

2.

Wait for about one minute until the AG module warms up. Completely block the gas inlet of the AG module.

Then the alarm message “ AG Airway Occluded ” will appear on the screen.

3.

Block the gas inlet for another one minute.

4.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

5.

Select the Module tab

AG tab.

6.

Check that the current flow rate is less than 10 ml/min, and the alarm message “ AG Airway Occluded ” does not disappear.

This indicates that the module does not leak. If the alarm message disappears, or the flow rate is equal to

10ml/min or greater, it indicates that the module leaks. Perform the leakage test again. If the problem remains, contact your service personnel for help.

20.12

Calibrating the AG Module

Calibrate the AG module every year or when the measured value is outside the specification. To calibrate the AG module, contact the service personnel.

CAUTION

Connect the gas outlet to the scavenging system when calibrating the AG module.

20.13

AG Troubleshooting

If the AG airway is occluded, the message AG Airway Occluded appears. In this case, check for the follows until the message disappears:

1.

Check the airway adapter for occlusion and replace if necessary.

2.

Check the sample line for occlusion or kinking and replace if necessary.

3.

Check the watertrap for water or occlusion. Empty the watertrap, or replace the watertrap if necessary.

4.

Check the gas outlet and the exhaust tube for any occlusion.

If the message does not disappear, it is probably the module fault. Contact the service personnel in this case.

NOTE

For the physiological and technical alarm messages, see Appendix

D Alarm Messages

.

20 - 8 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

21

Monitoring Electroencephalogram (EEG)

21.1

EEG Introduction

The Electroencephalograph (EEG) module monitors the patient's cerebral function by measuring the electrical activity of the brain.

The EEG module can continuously monitor EEG signal from up to four channels. It can also display Density

Spectral Arrays (DSA) and Compressed Spectral Arrays (CSA).

EEG monitoring is intended for adult, pediatric, and neonatal patients.

(1) (2)

(3)

(4)

(1) EEG dialog hard key

(3) Module status indicator

(2) Sensor check hard key

(4) EEG cable connector

21.2

EEG Safety Information

WARNING

: •

The conductive parts of electrodes and connectors should not contact other conductive parts, including electrically grounded parts.

To reduce the hazard of burns in the high-frequency surgical neutral electrode connection, the EEG sensor should not be located between the surgical site and the electrosurgical unit return electrode.

To avoid burns at the application site, do not use needle electrodes to perform EEG monitoring if electrosurgical unit (ESU) is in use. In case that needle electrodes are used, disconnect the EEG cable from the monitor and move the EEG cable away from the ESU cable before starting the ESU.

The EEG electrode must not be located between defibrillator pads when a defibrillator is used on a patient under monitoring.

To ensure proper defibrillator protection, use only recommended cables and leadwires.

EEG is a complex monitoring technology intended for use only as an adjunct to clinical judgment and training.

CAUTION

Only use parts and accessories specified in this manual. Follow the instructions for use and adhere to all warnings and cautions.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 21 - 1

Implanted devices (for example cardiac pacemakers), other patient connected equipment, and other equipment near the patient (for example high-frequency surgical units) can cause interference on the waveform, numerics, and the CSA presentation.

External radiating devices may disturb the measurement. It is recommended to avoid the use of electrical radiating equipment in close proximity to the monitor.

Interference from ECG can be eliminated by adjusting the filter settings.

NOTE

The EEG accessories are purchased from EB Neuro S.p.A. Please contact EB Neuro or visit its website

(www.ebneuro.com) for more information.

In the event of electrode connection issues, the monitor provides an error indication only when it performs an auto sensor check according to the time interval that is set by the user. It is recommended to start a manual sensor check if an abnormal waveform or high noise is observed.

21.3

EEG Parameters

EEG monitoring provides nine parameters: SEF, MF, PPF, TP, SR, Delta, Theta, Alpha, and Beta. The numerics area of each EEG channel can display up to six parameters.

Parameter Label

SEF

(Spectrum Edge Frequency)

MF

(Median Frequency)

PPF

(Peak Power Frequency)

TP

(Total Power)

SR

(Suppression Ratio)

Delta%, Theta%, Alpha%, Beta%

(Frequency band ratio)

Description

The frequency at which 95% of the total power lies below it and 5% lies above it. Its range is from 0.5 to 30.0 Hz.

The frequency at which 50% of the total power lies below it and 50% lies above it. Its range is from 0.5 to 30.0Hz.

The PPF is the frequency with the highest measured amplitude range from 0.5 to 30 Hz.

TP numeric is a measure of the absolute total power in the frequency band from 0.5 to 30.0 Hz. The useful range is from 40 to 100 dB, where 0 dB = 0.0001uV2..

SR numeric is the percentage of time in the past 60 seconds in which the EEG signal is considered suppressed.

EEG is traditionally divided into four frequency bands: delta, theta, alpha and beta. Frequency band ratio is the percentage of total power falling in corresponding band. For example, Delta % = Power in delta band/Total power.

Alpha waves: 8 to 13 Hz

Beta waves: 13 to 30 Hz

Theta waves: 4 to 8 Hz

Delta waves: 0.5 to 4 Hz

21.4

EEG Display

(1) (2) (3) (4)

21 - 2

(1) Lead label

(2) EEG waveform scale. For more information, see

21.8.1 Changing the EEG Scale .

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

(3)

(4)

EEG waveform

You can configure the displayed EEG waveforms. A maximum of four EEG waveforms can be displayed.

EEG parameters

You can configure the displayed EEG parameters. A maximum of six EEG parameters can be displayed. For more

information, see 21.8.5 Changing Displayed EEG Parameters .

21.5

Accessing the On-screen EEG Guide

The monitor provides the on-screen EEG guide to help you understand EEG monitoring functions and operating procedure.

To access the on-screen EEG guide, follow this procedure:

1.

Select the EEG numerics area or waveform area to enter the EEG dialog.

2.

Select the Introduction tab.

3.

Select a tab as required.

◆ Select the Summary tab to view the EEG monitoring principles, and work modes.

◆ Select the Target Patients tab to view the applicable patients for EEG monitoring.

◆ Select the Operation Guides tab to view the EEG monitoring procedure.

◆ Select the CSA and DSA tab to understand the CSA and DSA expand views.

21.6

Prepare for EEG Monitoring

To monitor EEG, follow this procedure:

1.

Connect EEG module, patient cable and lead wires. For more information, see 21.6.1 EEG Equipment to

Patient Connection .

2.

Select Montage. You can select a predefined montage and you can also customize a montage. See

21.6.2.3 Customizing an Electrode Montage

.

3.

Mark the electrode sites on the patient’s head according to the montage you have chosen.

4.

Prepare the skin of the electrode application sites.

5.

Apply the electrodes. For more information, see 21.6.3 Attaching EEG Electrodes

.

6.

Perform sensor check and observe the results.

21.6.1

EEG Equipment to Patient Connection

The following picture illustrates the connection between the EEG module and accessories.

(1) (2)

(1) EEG module (2) EEG patient cable (3) Lead wires

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

(3)

21 - 3

21.6.2

EEG Electrode Montages

Both bipolar mode and referential mode have four predefined montages. The following table shows electrode locations of each predefined montage in the bipolar mode:

Electrode

Positive (+)

EEG1 (A) Negtive (-)

Label

Positive (+)

EEG2 (B) Negtive (-)

Label

Positive (+)

EEG3 (C) Negtive (-)

Label

Positive (+)

EEG4 (D) Negtive (-)

Label

Montage 1

Fp1

T3

Fp1-T3

Fp2

T4

Fp2-T4

C3

O1

C3-O1

C4

O2

C4-O2

Montage 2

F3

C3

F3-C3

C3

P3

C3-P3

F4

C4

F4-C4

C4

P4

C4-P4

Montage 3

P3

P4

P3-P4

X

X

X-X

X

X

X-X

X

X

X-X

X

X-X

X

X

X-X

Note: X indicates to no electrode is applied in this site.

The following table shows electrode locations of each predefined montage in the referential mode:

Montage 4

P3

C3

P3-C3

P4

C4

P4-C4

X

Electrode

Positive (+)

EEG1 (A) Negtive (-)

Label

Positive (+)

EEG2 (B) Negtive (-)

Label

Positive (+)

EEG3 (C) Negtive (-)

Label

Positive (+)

EEG4 (D) Negtive (-)

Label

Montage 1

Fp1

Ne

Fp1-Ne

Fp2

Ne

Fp2 -Ne

C3

Ne

C3 -Ne

C4

Ne

C4 -Ne

Montage 2

F3

Ne

F3 -Ne

C3

Ne

C3 -Ne

F4

Ne

F4

C4

Ne

C4

-Ne

-Ne

Montage 3

F3

Ne

F3 -Ne

F4

Ne

F4 -Ne

P3

Ne

P3

P4

Ne

P4

-Ne

-Ne

Ne

O1 -Ne

O2

Ne

O2 -Ne

Montage 4

Fp1

Ne

Fp1 -Ne

Fp2

Ne

Fp2 -Ne

O1

You can modify the predefined montage and rename it as a customized montage. For more information, see

21.6.2.3 Customizing an Electrode Montage

.

21 - 4 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

21.6.2.1

EEG Electrode Locations

The following pictures shows the electrode locations. The electrode locations are labeled according to the international 10-20 electrode placement system. The numbers and letters refer to electrode locations:

21.6.2.2

Odd numbered electrodes: placed on the left

Even numbered electrodes: placed on the right

Letters: F = frontal, T = temporal, C = central, P = parietal, O = occipital, Z = midline electrodes

Bipolar Mode and Referential Mode

Measurement can be referential or bipolar.

In bipolar mode, each channel (EEG1, EEG2, EEG3 and EE4) uses two electrodes, a positive and a negative, to measure the potential difference between each pair.

In referential mode, all channels use the same referential electrode (Ne), and only use one electrode

(positive) to measure the potential difference.

21.6.2.3

Customizing an Electrode Montage

To modify a predefined montage and save it as customized montage, follow this procedure:

1.

Select the EEG numerics area or waveform area to enter the EEG dialog.

2.

Select the Montage Setup tab.

3.

Select the Montage Type to set the work mode.

4.

Select a montage (for example Montage 1 ).

5.

Modify the electrode locations: select an electrode position from the channel area on the left, and then select the electrode position from the electrode map on the right. Repeat this operation until you have defined all the electrode positions as desired.

6.

Select Save As and input a new name for this montage, and then select OK to confirm the changes.

You can define up to three customized montages. Selecting Delete can delete a customized montage.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 21 - 5

21.6.3

Attaching EEG Electrodes

CAUTION

Keep the mental part of the EEG electrodes from being contacted. Otherwise impedance results of the sensor check may be inaccurate.

NOTE

Use only one type of electrode in the whole montage.

Make sure that you have attached the ground electrode.

The neutral electrode is usually more prone to artifact.

For best results, use Ag/AgCl electrodes for the EEG measurement.

21.6.3.1

Attaching Cup Electrodes

To attach the cup electrodes, follow this procedure:

1.

Mark the electrode sites on the patient’s head according to the montage you have chosen.

2.

Comb or cut the hair away from the spots and rub the skin with the abrasive paste to remove oil and grease.

3.

Apply the conductive paste on the inside of each electrode and then press the electrodes on the spots.

NOTE

Preferably use cup electrodes if montage includes placements within hair area.

21.6.3.2

Attaching Needle Electrodes

To attach the needle electrodes, follow this procedure:

1.

Clean the skin with alcohol.

2.

Insert the needle into the subcutaneous area.

3.

Secure the needles to prevent excess movement or detachment. You may also use a small amount of paste to attach the cable to the patient's hair. This prevents the cables from pulling the needles out of the skin.

WARNING

Check that the package of needle electrodes is intact before use. Do not use the electrodes if the package is damaged.

Do not open the electrode package until immediately before use.

Needle electrodes are disposable. Do not reuse needle electrodes.

Replace any needle electrodes that are found to be bent. Using needles that have been previously bent may cause damage to the attachment site or inaccurate measurement.

21.7

Performing EEG Sensor Check

The monitor has an EEG sensor check function. The Sensor Check dialog displays the status of each electrode and the result of sensor check.

21.7.1

Setting the Interval of Auto Sensor Check

The sensor check is automatically initiated in the following situations:

The EEG module is connected.

The patient cable is connected.

21 - 6 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

The Electrode montage is changed.

The Sensor Check dialog is entered.

You can set the interval of performing the auto sensor check. To do so, follow this procedure:

1.

Select the EEG numerics area or waveform area to enter the EEG dialog.

2.

Select the Sensor Check tab.

3.

Select an appropriate setting from Interval list.

21.7.2

Displaying/Hiding Impedance Value

To display the impedance value of each electrode on the electrode map, select Display Imped Values(KΩ) .

Selecting Hide Imped Values(KΩ) displays the sensor check result instead of the impedance value.

21.7.3

Manually Starting a Sensor Check

To manually start sensor check, choose either of the following way:

1.

Select the EEG numerics area or waveform area to enter the EEG menu.

2.

Select the Sensor Check tab.

3.

Verify that the Montage Type setting and Montage setting are correct.

4.

Select Start Sensor Check .

You can also press the sensor check hardkey on the EEG module to start the sensor check.

At the completion of sensor check, the electrode status is shown. For more information, see 21.7.4 EEG Electrode

Status .

21.7.4

EEG Electrode Status

EEG electrode status is color coded. The following table lists all electrode status options and actions to be taken.

Color

Red

Grey

Yellow

Status

Off

Noise

High

Description Action

Electrode falls off and has no skin contact. In the Bipolar mode, reconnect the electrode indicated as red, and PGND electrode.

In the Referential mode, reconnect the electrode indicated as red, NE and PGND electrodes.

The EEG signal is too noisy. Impedance cannot be measured.

The impedance is above the need.

The impedance is within the need.

Check the sensor-to-skin contact. If necessary, reconnect the electrodes.

Check the sensor-to-skin contact. If necessary, reconnect the electrodes.

No action is necessary.

Green Pass

For each EEG channel, to get reliable results all electrodes for this channel should be in Pass status (green).

21.7.5

Stopping Sensor Check

Sensor check automatically stops if all the electrodes pass the impedance check. You can also manually stop the sensor check.

To stop sensor check, choose either of the following ways:

Press the sensor check hardkey on the EEG module.

Select Stop Sensor Check from the Sensor Check dialog.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 21 - 7

21.8

Changing EEG Settings

21.8.1

Changing the EEG Scale

To set EEG waveform scale, follow this procedure:

1.

Select the EEG numerics area or waveform area to enter the EEG dialog.

2.

Select an appropriate setting from the Scale list.

21.8.2

Changing the EEG Sweep Speed

To set EEG sweep speed, follow this procedure:

1.

Select the EEG numerics area or waveform area to enter the EEG dialog.

2.

Select an appropriate setting from the Speed list.

21.8.3

Changing the High/Low Filter

The low and high filters can screen out undesirable interference which may come from respiration, movement, etc. The current EEG high and low filter settings are shown at the top of DSA and CSA view.

To change the filter settings, follow this procedure:

1.

Select the EEG numerics area or waveform area to enter the EEG dialog.

2.

Select an appropriate setting from the Low Freq Cut-off and High Freq Cut-off list.

21.8.4

Switching Off the Notch Filter

The notch filter can screen out 50Hz/60Hz power line noise. The notch filter is enabled by default. To disable the notch filter, follow this procedure:

1.

Select the EEG numerics area or waveform area to enter the EEG dialog.

2.

Set Notch Filter to Off .

21.8.5

Changing Displayed EEG Parameters

You can choose the displayed parameters. To do so, follow this procedure:

1.

Select the EEG numerics area or waveform area to enter the EEG dialog.

2.

Select the Select Parameter tab.

3.

Select the parameters you want to display.

You can select at most six parameters. The selected parameters are applied to all EEG channels.

21.9

Displaying the EEG Expand View

To display the EEG expand view, follow this procedure:

1.

Select the EEG numerics area or waveform area to enter the EEG dialog.

2.

From the bottom of the EEG dialog, select EEG Expand .

3.

Select the desired tab to enter corresponding view:

◆ Select the EEG tab, and then select EEG Channels , Scale , and Speed to view corresponding EEG waveforms.

◆ Select the Parameters tab to view parameter values of each EEG channel.

◆ Select the Trends tab, and then select EEG Channels , Parameters , and Trend Length to view the trends of corresponding EEG channels and parameters.

◆ Select the CSA

tab to enter the CSA view. For more information, see 21.9.1 CSA View

.

◆ Select the DSA tab to enter the DSA view. For more information, see

21.9.2 DSA View

.

21 - 8 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

21.9.1

CSA View

The continuous EEG signal is sampled periodically and this value is stored in a frame. Each frame is processed using Fast Fourier Transformation (FFT) to provide a frequency spectrum displayed as a compressed spectral array (CSA).

The CSA View provides an overview of the patient's EEG values over time.

To display the CSA View, follow this procedure:

1.

Select the EEG numerics area or waveform area to enter the EEG dialog.

2.

From the bottom of the EEG dialog, select EEG Expand .

3.

Select the CSA tab.

The CSA View provides up to CSA of four EEG channels. It provides the following information:

Displayed item

Status bar

Frequency scale

Spectral lines

Trendlines

Description

The first line displays the current montage, filter setting, notch frequency, SEF percentage (95%), and trendline labels and color codes.

The second line displays the EEC channel labels and lead labels.

It is the horizontal axis. The scale range depends on the filter settings ( Low Freq Cut-off and

High Freq Cut-off settings. The maximum displayed frequency is 30 Hz, so if you set High Freq

Cut-off to 50 or 70, the upper scale remains 30.

The energy at each frequency is computed and displayed as a spectral line.

EEG values are sampled at configured time intervals and displayed as color-coded trendlines.

Trendlines are available for up to three frequency numerics (SEF, MF, and PPF). SEF trendline is white, MF trendline is purple, and PPF trendline is green.

Appears when artifact is detected, electrodes are off or disconnected, or montage is changed.

“?” mark

From the CSA view you can select the following items:

EEG Channels

Parameters

Trend Length

To change Power Scale and CSA Clipping setting, select CSA Setup .

Changing Power Scale can adjust the amplitude of spectral lines. The wider the power scale range, the greater amplitude of the spectral lines is.

If CSA Clipping is switched on, the latest spectral line displays in a normal shape, in which area other gothrough spectral lines will be cut out. If the CSA Clipping is switched off, all the spectral lines display normally.

21.9.2

DSA View

The Density Spectral Array (DSA) is to show changes in the power spectrum distribution over time.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 21 - 9

To display the DSA View, follow this procedure:

1.

Select the EEG numerics area or waveform area to enter the EEG dialog.

2.

From the bottom of the EEG dialog, select EEG Expand .

3.

Select the DSA tab.

The DSA View displays DSA for up to four EEG channels. It provides the following information:

Displayed item

Color bar

Status bar

Frequency scale

Trendlines

“?” mark

Description

It is located at the right of the DSA view. The color bar color codes the power.

You can change the setting of Power Scale to adjust the color of corresponding power.

Displays the current montage, filter setting, notch frequency, SEF percentage

(95%), and trendline labels and color codes.

It is the vertical axis. The scale range depends on the filter settings ( Low Freq

Cut-off and High Freq Cut-off setting. The maximum displayed frequency is

30 Hz, so if you set High Freq Cut-off to 50 or 70, the upper scale remains 30.

EEG values are sampled at configured time intervals and displayed as colorcoded trendlines. Trendlines are available for up to three frequency numerics

(SEF, MF, and PPF). SEF trendline is white, MF trendline is purple, and PPF trendline is green.

Appears when high electrode impedance or artifact is detected, electrodes are off or disconnected is changed.

From the DSA view you can select the following items:

EEG Channels

Parameters

Trend Length

Power Scale

21 - 10 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

21.10

EEG Troubleshooting

This section lists the problems that might occur. If you encounter the problems when using the equipment or accessories, check the table below before requesting for services. If the problem persists after corrective actions have been taken, contact the service personnel.

Problem

EEG signal is noisy

EEG cable and electrodes properly connected, but no EEG waveforms.

The EEG numerics area displays ”- -”.

EEG waveform baseline fluctuates.

The electrode impedances show “- -” and there is a message prompting to check the ground electrode?

Solution

1. Check that the electrodes are properly connected and not dried out.

2. Check that the electrodes properly contact with skin.

3. Perform electrode impedance check.

4. Calm the patient since frontal muscle activity can cause artifact.

5. Remove sources of external electrical noise (for example, the lamps) from the vicinity of the patient’s head.

6. ECG monitoring may cause artifact; change electrode positioning.

7. Possible mains power noise. Check that the monitor and other medical devices connected to the patient are properly connected to earth ground. If the noise persists, consider stopping EEG monitoring.

1. The number of channels in the montage is smaller than the number of channels connected to the patient. Check the number of channels.

2. Select the Screen Setup quick key → Parameters On/Off. Make sure that EEG parameter is turned on.

The patient has high muscle activity in the head area, or noise from some interfering equipment is coupling to electrode cables. Relax the patient and remove the source of noise.

1. Sweating may cause variations in the electrode impedance. Check the patient.

2. If the fluctuation is disturbing, prepare the skin and replace the electrodes.

The ground electrode is poorly connected to the patient. Check the electrode and cable. If the impedance of the electrode is too high, the measurement fails even if the electrode is properly attached. Use better electrodes or prepare the skin better.

NOTE

For the list of physiological and technical alarm messages, see

D Alarm Messages

.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 21 - 11

This page intentionally left blank.

21 - 12 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

22

Monitoring Bispectral Index (BIS)

22.1

BIS Introduction

Bispectral Index (BIS) monitoring helps to monitor the level of consciousness of a patient under general anesthesia or sedation in OR and ICU. It is designed to monitor the hypnotic state of the brain based on acquisition and processing of EEG signals. Use of BIS monitoring to help guide anesthetic administration may be associated with the reduction of the incidence of awareness with recall during general anesthesia or sedation.

There are two BIS solutions available for use with the BIS module: the BISx or the BISx4. The BISx is for single side

BIS monitoring, and the BISx4 is for both single side and bilateral BIS monitoring. The BISx4 provides bilateral BIS monitoring only when the BIS Bilateral Sensor is connected.

BIS monitoring is intended for adult and pediatric patients.

(1) (2)

(3) (5)

(6) (7)

(4)

(1) BIS dialog hard key

(3) Module status indicator

(5) BIS module

(7) Patient cable

(2) Sensor check hard key

(4) BIS cable connector

(6) BISx or BISx4

(8) BIS sensor

(8)

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 22 - 1

22.2

BIS Safety Information

WARNING

BIS monitoring is not intended for neonatal patients.

The BIS technology used with this monitor is purchased from Covidien. It is important to recognize this index is derived using solely that company's proprietary technology. Therefore, it is recommended that clinicians have reviewed applicable information on its utility and/or risks in published articles and literature/web site information from Covidien, or contact Covidien at www.covidien.com for clinical-based BIS questions. Failure to do so could potentially result in the incorrect administration of anesthetic agents and/or other potential complications of anesthesia or sedation. It is recommend that clinicians also review the following practice advisory (that includes a section on BIS monitoring): The American Society of Anesthesiologists, Practice Advisory for

Intraoperative Awareness and Brain Function Monitoring (Anesthesiology 2006;104:847-64).

Clinicians are also recommended to maintain current knowledge of FDA or other federal-based regulatory, practice or research information on BIS and related topics.

The conductive parts of sensors and connectors should not come into contact with other conductive parts, including earth conductors.

To reduce the hazard of burns in the high-frequency surgical neutral electrode connection, the BIS sensor should not be located between the surgical site and the electrosurgical unit return electrode.

To reduce the hazard of burns during use of brain-stimulating devices (e.g. transcranial electrical motor evoked potential), place stimulating electrodes as far as possible from the BIS sensor and make certain that sensor is placed according to package instructions.

The BIS sensor must not be located between defibrillator pads when a defibrillator is used on a patient connected to the monitor.

The clinical utility, risk/benefit and application of the BIS component have not undergone full evaluation in the pediatric population.

Due to limited clinical experience for patients with neurological disorders, patients taking psychoactive medication, and children under one year old, BIS values from these patients should be interpreted cautiously.

The BIS monitoring is a complex technology, intended for use only as an adjunct to clinical judgment and training. Clinical judgment should always be used when interpreting BIS in conjunction with other available clinical signs. Reliance on BIS alone for intraoperative anesthetic management is not recommended.

Misinterpretation of BIS can result in incorrect administration of anesthetic agents and/or other potential complications of anesthesia or sedation.

BIS values should be interpreted cautiously with certain anesthetic combinations, such as those relying primarily on either ketamine or nitrous oxide/narcotics to produce unconsciousness.

CAUTION

Ensure that the BISx or BISx4 does not come into prolonged contact with the patient's skin, as it may generate heat and cause discomfort.

Do not use the BIS sensor if the sensor gel is dry. To avoid dryout, do not open the pack until you are ready to use the sensor.

When using electro-convulsive therapy (ECT) equipment during BIS monitoring, place ECT electrodes as far as possible from the BIS sensor to minimize the effect of interference. Certain ECT equipment may interfere with the proper function of the BIS monitoring system. Check for compatibility of equipment during patient setup.

The BIS measurement based on measuring the EEG signal is inherently very sensitive. Do not use electrical radiating equipment close to the BISx or BISx4.

Artifact may lead to inappropriate BIS values. Potential artifact may be caused by unusual or excessive electrical interference or high EMG activity like shivering, muscle activity or rigidity, sustained eye movements, head and body motion. Also, improper sensor placement and poor skin contact (high impedance) may cause artifact and interfere with the measurement.

External radiating devices may disturb the measurement.

Poor signal quality may lead to inappropriate BIS values.

22 - 2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

22.3

BIS Parameters

Single side BIS monitoring provides the following parameters:

Parameter Description

BIS

(Bispectral Index)

SQI

(Signal Quality Index)

EMG

(Electromyography)

The BIS numeric reflects the patient’s level of consciousness. It ranges from 100 for wide awake to 0 in the absence of brain activity.

100: the patient is wide awake.

70: the patient is underdosed but still unlikely to become aware.

60: the patient is under general anesthesia and loses consciousness.

40: the patient is overdosed and in deep hypnosis.

0: the EEG is a flat line. The patient has no electrical brain activity.

The SQI numeric reflects signal quality and provides information about the reliability of the BIS, SEF, TP, and SR numerics during the last minute. The greater the SQI value, the better signal quality.

EMG bar graph reflects the electrical power of muscle activity and high frequency artifacts. Low EMG indicates that EMG activity is low. BIS monitoring conditions are optimal when the bar is empty.

SR numeric is the percentage of time over the last 63-second period during which the

EEG is considered to be in a suppressed state.

SR

(Suppression Ratio)

SEF

(Spectral Edge Frequency)

TP

(Total Power)

BC

(Burst Count)

The SEF is a frequency below which 95% of the total power is measured.

TP numeric is a measure of the absolute total power in the frequency ranging from 0.5 to

30.0 Hz. The useful range is from 40 to 100 db.

A burst means a period (at least 0.5 second) of EEG activity followed and preceded by inactivity. The BC numeric helps you quantify suppression by measuring the number of

EEG bursts per minute. For single side BIS monitoring, this parameter is intended for the

BIS module with the Extend sensor only. BC numeric is valid only when SQI≥15% and

SR≥5%.

Bilateral BIS monitoring simultaneously monitors both cerebral hemispheres. When bilateral BIS is monitored, an "L" or "R" will be added to each of the numeric identifiers except ASYM. In addition to the above parameters, it also provides the following parameters:

Parameter sBIS

(BIS Variability Index) sEMG

(EMG Variability Index)

ASYM

(Asymmetry)

Description sBIS numeric represents the standard deviation of the BIS variable over the last three minutes. sEMG numeric represents the standard deviation of the EMG value over the last three minutes.

ASYM is a processed variable indicating the percentage of EEG power presented in the left or right hemisphere with respect to total (left and right) EEG power.

Designation ‘L’ of the asymmetry data indicates asymmetry to the left side.

Designation ‘R’ of the asymmetry data indicates asymmetry to the right side.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 22 - 3

22.4

BIS Display

(1) (2)

The display of BIS waveform area depends on the setting of Display from the BIS dialog. For more information, see

22.7.3 Setting the Display of BIS Waveform Area

.

(3) SQI indicator

Empty: SQI < 15%, unable to calculate BIS and secondary parameter values. BIS and secondary parameter values are displayed as “---”.

1 to 2 bars: SQI 15% - 49%, parameter values are unreliable.

3 to 5 bars: SQI 50% - 100%, parameter values are reliable.

(4) EMG indicator

Empty: EMG < 30 dB. BIS monitoring conditions are optimal.

1 to 4 bars: EMG 30 - 55 dB. BIS monitoring conditions are acceptable.

5 bars: EMG > 55 dB. BIS monitoring conditions are unacceptable.

(5) Secondary parameters

The displayed secondary parameters are configurable. For more information, see 22.7.5 Setting the Displayed BIS

Parameters

.

22.5

Accessing the On-screen BIS Guide

The monitor provides the on-screen BIS guide to help you understand BIS monitoring functions and operating procedure.

To access the on-screen BIS guide, follow this procedure:

1.

Select the BIS numerics area or waveform area to enter the BIS dialog.

2.

Select the Introduction tab.

3.

Select a tab as required.

◆ Select the Summary tab to view the BIS monitoring principles.

◆ Select the Target Patients tab to view the applicable patients for BIS monitoring.

◆ Select the Operation Guides tab to view the BIS monitoring procedure.

(3)

(4)

(5)

22 - 4 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

22.6

Preparing for BIS Monitoring

(1)

(2)

(3)

(4)

(1) BIS module

(3) Patient cable

(2) BISx or BISx4

(4) BIS sensor

To perform BIS monitoring, follow this procedure:

1.

Connect the BISx or BISx4 to the BIS module.

2.

Use the attachment clip to secure the BISx or BISx4 to a convenient location near the patient’s head.

3.

Connect the patient cable to the BISx or BISx4.

4.

Attach the BIS sensor to the patient following the instructions supplied with the sensor.

5.

Insert the BIS sensor into the patient cable connector until it is fully engaged.

6.

Observe the results of the automatic sensor check in the numeric area.

Sensor Check is initiated automatically when the BIS sensor and the patient cable are connected to the BISx or

BISx4. The measurement starts automatically after the sensor has passed the check.

WARNING

To minimize the risk of patient strangulation, the patient cable must be carefully placed and secured.

CAUTION

Make sure the patient’s skin is dry. A wet sensor or a salt bridge could result in erroneous BIS and impedance values.

Do not use if sensor gel is dry. To avoid dry out, do not open pack until ready for use.

Due to intimate skin contact, reuse may pose risk of infection. If skin rash or other unusual symptoms develop, stop using and remove.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 22 - 5

22.7

Changing BIS Settings

22.7.1

Setting BIS Alarm Properties

To set the BIS alarm properties, follow this procedure:

1.

Select the BIS numeric area or waveform area to enter the BIS dialog.

2.

In the BIS dialog, select the Alarm tab.

3.

Enter the password if required. For more information, refer to

32.14.3 Selecting Password for User Authentication .

4.

Set alarm properties as desired.

22.7.2

Choosing the BIS Smoothing Rate

The smoothing rate defines how the monitor averages the BIS value. With the smoothing rate becoming smaller, the monitor provides increased response to changes in the patient’s state. Contrarily, the monitor provides a smoother BIS trend with decreased variability and sensitivity to artifacts.

To change the smoothing rate, follow this procedure:

1.

Select the BIS numeric area or waveform area to enter the BIS dialog.

2.

In the BIS dialog, select the Setup tab.

3.

Set Smoothing Rate to 10 sec , 15 sec , or 30 sec .

22.7.3

Setting the Display of BIS Waveform Area

To set the display of BIS waveform, follow this procedure:

1.

Select the BIS numeric area or waveform area to enter the BIS dialog.

2.

In the BIS dialog, select the Setup tab.

3.

Set Display .

◆ If you set Display to EEG waveforms, set Scale and Speed for EEG waveforms.

◆ If you set Display to BIS parameter trends, set Trend Length .

22.7.4

Switching the Filter On or Off

The filter can remove undesired interference from the raw EEG wave display. It is switched on by default.

To enable/disable the filter, follow this procedure:

1.

Select the BIS numeric area or waveform area to enter the BIS dialog.

2.

In the BIS dialog, select the Setup tab.

3.

Switch on/off Filter .

22.7.5

Setting the Displayed BIS Parameters

Besides the BIS value, you can also display up to four secondary parameters in the BIS numeric area. To select the displayed parameters, follow this procedure:

1.

Select the BIS numeric area or waveform area to enter the BIS dialog.

2.

In the BIS dialog, select the Select Parameter tab.

3.

From the BIS Tile area, select a secondary parameter block, and then select a secondary parameter from the Parameters area.

22 - 6 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

22.8

Sensor Check

22.8.1

Automatic Sensor Check

Once the BIS sensor is connected, an automatic sensor check starts to check the sensor type, status, and impedance of all the electrodes, including the signal electrodes, the reference electrode and the ground electrode. During the sensor check, the message “Sensor Check In Progress” displays in the information area. If this message continuously displays, enter the Sensor Check dialog and check if the impedance of each electrode is acceptable.

After the initial sensor check, the monitor performs automatic checks during BIS monitoring. Automatic sensor checks includes the following items:

The combined impedance of the signal electrodes plus the reference electrode. This is done automatically and continuously and does not affect the EEG waveform. As long as the impedances are within the valid range, there is no prompt message of this check or its results.

The impedance of the ground electrode. This is done every ten minutes and takes approximately four seconds. It causes artifact in the EEG waveform, and the message BIS: Checking Ground is displayed on the monitor during the check. If the ground electrode does not pass this check, re-apply the electrode.

Then another check will automatically initiate.

The monitor continually checks impedance levels during a procedure by generating a 128 Hz test signal.

Occasionally this signal may interfere with other equipment. If this becomes a problem, you may need to disable the auto checking. Impedance levels will still be tested at startup, but once they pass, they will not be tested again until a new case is begun.

NOTE

Automatic sensor check may need to be disabled if the 1nA 128 Hz impedance check signal interferes with other equipment.

22.8.2

Disabling automatic sensor check

To disable automatic sensor check, follow this procedure:

1.

Select the BIS numeric area or waveform area to enter the BIS dialog.

2.

Select the Setup tab.

3.

Switch off Auto Check .

CAUTION

Switching the auto impedance check off will disable automatic prompt to the user of impedance value changes, which may lead to incorrect BIS values. Therefore, this should only be done if the check interferes with or disturbs other measurements.

22.8.3

Manual Sensor Check

To manually start a sensor check, use either of the following methods:

Press the sensor check hard key

Select the Sensor Check tab from the

on the BIS module.

BIS dialog.

The monitor enters the Sensor Check dialog after you start the sensor check. The Sensor Check dialog displays the following items:

Sensor Type

The status of each electrode

Expiration time or usable cycles

NOTE

For different types of sensors the Sensor Check dialog may be different.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 22 - 7

22.8.4

BIS Sensor Status

The color of each electrode indicates its status:

Color

Red

Grey

Yellow

Status

Lead off

Noise

High

Description Action

Electrode has fallen off and has no skin contact.

The EEG signal is too noisy. Impedance cannot be measured.

The impedance is above the limit.

The impedance is within valid range.

Press the sensor more firmly to skin to ensure good sensor-to-skin contact. If necessary, remove the sensor, and then clean and dry the skin. Reapply the senor or replace the sensor.

Press the sensor more firmly to skin to ensure good sensor-to-skin contact.

Press the sensor more firmly to skin to ensure good sensor-to-skin contact.

No action necessary.

Green Pass

Although BIS may still be measured when the electrode status is Noise or High , for best performance, all electrodes should be in the Pass status.

The sensor check may fail for the following reasons:

Impedance too high

Incorrect sensor application

Poor sensor connection

Defective patient interface cable or sensor

To correct the situation:

Recheck the sensor

Reapply the sensor according to instructions

Check sensor connection

Replace patient interface cable or sensor

22.9

Monitoring Bilateral BIS

By using BISx4 and the Bilateral sensor you can perform bilateral BIS monitoring. You can display the BIS expand view during bilateral BIS monitoring.

From the BIS expand view you can monitor 4-channel EEG and the density spectral array (DSA) which shows the degree of asymmetry in EEG power between the left and right hemispheres. The BIS expand view also provides

BIS trend.

22.9.1

Entering the BIS Expand View

To enter the BIS expanded view, follow this procedure:

1.

Select the BIS numeric area or waveform area to enter the BIS dialog.

2.

Select BIS Expand at the bottom left corner to enter the BIS Expand View.

22.9.2

Selecting BIS Expand View Display

To configure what is displayed in the BIS Expand View screen, follow this procedure:

1.

Select the BIS numeric area or waveform area to enter the BIS dialog.

2.

Select BIS Expand at the bottom left corner to enter the BIS Expand View.

3.

Select the EEG , BIS Trend or DSA tab.

22.9.2.1

Displaying the EEG waveforms

The EEG tab of the BIS Expand view shows the selected EEG waveforms. You can configure the EEG waveforms:

■ Select EEG Waveforms to set which EEG waveforms you want to display.

22 - 8 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Select Scale to set EEG waveform scale.

Select Speed to set EEG waveform speed.

22.9.2.2

Displaying the BIS Trend

The BIS Trend tab of the BIS Expand view shows the trend of selected parameters. You can configure the BIS trend:

Select Parameter 1 and Parameter 2 to set the parameters whose trend you want to display.

Set Trend Length .

22.9.2.3

Displaying BIS DSA

The Density Spectral Array (DSA) shows changes in the power spectrum distribution over a certain period of time.

DSA view shows the following information:

■ y-axis: time scale x-axis: signal frequency scale from 0 to 30 Hz

Color bar: shows range of power. Red indicates maximum power and blue indicates minimum power.

Spectral edge frequency (SEF) trend: It is the white Spectral Edge line superimposed on the graph where

95% of the total power lies on one side of the line (toward the inside of the graph) and 5% lies on the other.

The current SEF value: displays above the graph.

ASYM graph: displays in the center of the DSA view. It shows the degree of asymmetry in EEG power between the left and right hemispheres. The ASYM scale begins at 20% at the center line and runs left or right to 100%. Asymmetry data less than 20% are not displayed on the graph, but are available in the tabular trends.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 22 - 9

22.10

BIS Troubleshooting

This section lists the problems that might occur. If you encounter the problems when using the equipment or accessories, check the table below before requesting for services. If the problem persists after corrective actions have been taken, contact the service personnel.

Problem

Measurement does not start

Solution

1. Check the sensor attachment to the patient and the sensor placement. Check the sensor contact with skin.

2. Check the sensor type.

3. Check all connections and the patient cable.

NOTE

For a comprehensive list of physiological and technical alarm messages, see

D Alarm Messages .

22.11

Covidien Information

Covidien Patents

This posting serves as notice under 35 U.S.C.§287(a) for Covidien patents: http://www.covidien.com/patents.

No Implied License

Possession or purchase of this device does not convey any express or implied license to use the device with unauthorized replacement parts or consumables which would, alone, or in combination with this device, fall within the scope of one or more of the patents relating to this device.

22 - 10 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

23

Monitoring Neuromuscular Transmission (NMT)

23.1

NMT Introduction

The neuromuscular transmission (NMT) module evaluates muscle relaxation of patients under neuromuscular block by measuring the strength of muscle reaction after electrically stimulating the dedicated motor nerve.

Two electrodes are placed on the patient’s skin over dedicated nerve. A controllable current source delivers stimulation pulses to the two electrodes, and the acceleration of muscle contractions is measured with a 3-axis accelerometer.

NMT monitoring is intended for adult and pediatric patients.

(1)

(2)

(3)

(4)

(1) Calibration hard key

(3) Module status indicator

(2) Start/stop key

(4) NMT patient cable connector

23.2

NMT Safety Information

WARNING

The NMT measurement is not intended for neonatal patients.

The NMT stimulation should not be applied across or through the head, directly on the eyes, covering the mouth, on the front of the neck, especially the carotid sinus, or from electrodes placed on the chest and the upper back or cross over the heart.

Application of electrodes near the thorax may increase the risk of cardiac fibrillation.

Never apply electrodes to patients in areas where inflammation or injury is evident.

When you are connecting the electrodes or the patient cable, make sure that the connectors do not touch any electrically conductive material including earth conductors.

Patients with nerve damage or other neuromuscular problems may not respond properly to stimulation. The NMT measurement may show unusual patterns when monitoring muscle paralysis in these patients.

NMT stimulation current pulses may interfere with other sensitive equipment, for example, implanted cardiac pacemakers. Do not use the NMT measurement on patients with implanted electronic devices unless so directed by a medical specialist.

Simultaneous use of the NMT with high frequency electrosurgical equipment (ESU) may result in burns at the stimulation site and can also adversely affect measurement accuracy. Make sure the

ESU return electrode is properly applied to the patient.

Do not use the NMT in close proximity to shortwave or microtherapy devices, there is a risk of adversely affecting the NMT measurement.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 23 - 1

Never touch the electrodes unless the stimulation has been stopped.

Check each time before use that the material insulating the NMT sensor and the stimulation cable is intact and does not show signs of wear and tear.

CAUTION

NMT monitoring is intended as an adjunct in patient assessment and must be used in conjunction with observation of clinical signs and symptoms.

NMT stimulation can be painful to a non-sedated patient. It is recommended not to stimulate before the patient is adequately sedated.

Use only electrodes suitable for nerve stimulation deemed appropriate by the attending physician.

Pay special attention to current densities exceeding 2 mA r.m.s/cm

2

for any electrodes.

23.3

Stimulation Modes

The NMT module provides the following stimulation modes. Some stimulation modes require a minimum neurophysiological recovery time and during this recovery phase no new stimulation can be started. So you cannot start a measurement or calibration.

23.3.1

Train-Of-Four (TOF)

TOF mode is recommended for most cases.

In TOF mode, four stimulation pulses are generated at 0.5 second intervals. Each stimulation of the train causes the muscle to contract. The fade in the individual response to each single stimulation provides a basis for evaluation. The response is measured after each stimulus and the ratio of the fourth to the first response of the

TOF sequence is calculated resulting in TOF-Ratio (TOF%).

When relaxation deepens, the TOF% declines until the fourth response disappears and no TOF% is calculated.

Low response to the first stimulus (T1) also results in TOF% unavailable. When no TOF% is available, the degree of neuromuscular block is estimated from the number of responses or TOF counts. The fewer the response count is detected, the deeper is the relaxation.

If NMT calibration establishes the reference response amplitude, response to T1 as percentage of the reference value is calculated resulting in T1%.

In TOF mode, the minimum neurophysiological recovery time is 10 seconds. If NMT measurement or calibration is initiated during this period, it will be automatically delayed.

23.3.2

Single Twitch (ST)

In single twitch (ST) stimulation, the module delivers a single electrical pulse and measures the strength of the resulting twitch, the module then calculates the ratio of measured response to the reference twitch resulting in

ST-Ratio.

ST mode is practical when using depolarizing relaxants since TOF% does not give any additional information about the patient status. Additionally, when the change of patient’s relaxation level is considered, ST stimulation at a frequency of 1 Hz can indicate the relaxation change in a more real-time way.

23.3.3

Post-Tetanic Count (PTC)

When neuromuscular block deepens, different parameters are needed to measure the response. At first, when the response to the fourth TOF stimulation pulse disappears or the first twitch is very weak, the TOF% is not available and only the number of detected counts can be observed. When stimulation pulses no longer give any stimulation response, you do not get the TOF count either. To monitor the relaxation level, you can start tetanic stimulation and estimate the relaxation level from the Post Tetanic Count (PTC).

PTC stimulation mode starts with a sequence of four current pulses delivered at 2 Hz. If a muscle response is detected, the PTC sequence is stopped and the TOF result is reported. If there is no muscle response, the sequence continues with a five seconds long tetanic stimulation delivered at 50 Hz, followed by a pause of 3 seconds, followed by 20 single current pulses delivered at 1 Hz. The number of detected responses is counted and expressed as PTC. The fewer responses are detected, the deeper the relaxation is.

23 - 2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

After tetanic stimulation, NMT measurements and calibration are disabled for 20 seconds and PTC is disabled for

2 minutes.

23.3.4

Double Burst Stimulation (DBS)

Double Burst Stimulation (DBS) enables better visual observing of the fading in the responses. DBS consists of two separate bursts at an interval of 750 ms, where each burst consists of certain pulses directly after each other at a frequency of 50 Hz. The response ratio of the second to the first burst is calculated resulting in DBS-Ratio, while the number of responses is detected and expressed as DBS Count.

The module supports DBS3.2 and DBS3.3.

■ For DBS3.2 mode, the first burst consists of 3 consecutive pulses, and the second burst consists of 2 consecutive pulses.

For DBS3.3 mode, both bursts consist of 3 consecutive pulses.

In DBS mode, the minimum neurophysiological recovery time is 15 seconds. If NMT measurement or calibration is initiated during this period, it will be automatically delayed.

23.4

NMT Parameters

The following table lists the NMT parameters in different stimulation modes.

Stimulation Mode

TOF

ST

PTC

DBS

Parameter Label

TOF-ratio

TOF-count

T1%

ST-ratio

ST-count

PTC

DBS ratio

DBS count

Unit

%

/

/

/

/

%

%

/

Number of Bars

2

2

1

/

/

1

4

4

23.5

NMT Display

(10) (9)

(1)

(2)

(8)

(7)

(6)

(5) (3) (4)

(1) Parameter unit

(2) Parameter value

(3) Parameter label

(4) Stimulation current

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 23 - 3

(5) T1%: responses to the first stimulus as percentage of the reference amplitude in TOF mode. This value is not shown if calibration fails.

(6) Bar graph: indicates the amplitude of response to the stimulation. The maximum height of the bar graph displayed is

120%. Bar graph is not shown if calibration is not completed successfully.

(7) Scale: indicates the amplitude of response to stimulation.

(8) Measurement interval: Manual is displayed if Interval is set to Manual .

(9) Measurement countdown: time to the next measurement. The measurement countdown is not shown if Interval is set to Manual .

(10) The last measurement time .

NOTE

The NMT parameter values become outline font 15 minutes after the NMT measurement is taken.

The PTC value is shown on the display for 20 seconds after which the NMT module returns to the preset stimulation mode.

23.6

Accessing the On-screen NMT Guide

The monitor provides the on-screen NMT guide to help you understand NMT monitoring functions and operating procedure.

To access the on-screen NMT guide, follow this procedure:

1.

Select the NMT numerics area or waveform area to enter the NMT menu.

2.

Select the Introduction tab.

3.

Select a tab as required.

Select the Summary tab to view the NMT monitoring principles.

Select the Target Patients tab to view the applicable patients for NMT monitoring.

Select the Points To Note tab to view the precautions before NMT monitoring.

Select the Operation Guides tab to view the NMT monitoring procedure.

23.7

Preparing for NMT Monitoring

23.7.1

NMT Equipment to Patient Connection

The following picture shows NMT cable and patient connection .

(5)

(6)

(3)

(7)

(1)

(1) NMT module

(3) NMT stimulation cable

(5) Proximal electrode (red)

(7) NMT sensor

(2) (4)

(2) NMT patient cable

(4) NMT sensor cable

(6) Distal electrode (black)

23 - 4 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

23.7.2

Preparing the Skin

Good electrode-to-skin contact is important for good signal quality. Before applying the electrodes, clean the application site of oil and dirt and avoid placing the electrodes over excessive body hair or lesions. Insufficient cleaning of the skin can cause high skin impedance which could cause the stimulation to stop.

To properly prepare the skin, follow this procedure:

1.

Select sites with intact skin, without lesion of any kind.

2.

Shave hair from skin at chosen sites.

3.

Gently rub skin surface at sites to remove dead skin cells.

4.

Thoroughly cleanse the site with a mild soap and water solution.

5.

Dry the skin completely before applying the electrodes.

23.7.3

Placing the Electrodes and Sensor

Stimulation of the ulnar nerve in the wrist and acceleration measurements at the adductor pollicis are preferred for routine monitoring. When monitoring neuromuscular transmission, round surface electrodes with snap connection are a must. Small (pediatric or neonatal) electrodes are advisable to obtain a sufficient current density. In order to ensure a steady signal quality, be sure only to use CE marked electrodes.

Ensure that the thumb can move freely before applying NMT electrodes and sensor. Follow this procedure to place the electrodes and sensor:

1.

Place the distal electrode near the wrist.

2.

Place the proximal electrode 2 to 3 cm proximal of the distal electrode.

3.

Attach the red clamp to the proximal electrode.

4.

Attach the black clamp to the distal electrode.

5.

Affix the sensor with its large flat side against the palmar side of the thumb with a piece of tape. The cable should be attached in such a way that it does not 'pull' at the sensor and that movement of the thumb is not obstructed.

CAUTION

When placing the electrodes, make sure that they do not touch each other.

If the electrodes are placed incorrectly, wrong nerves are stimulated and this causes wrong muscle response.

When multiple nerves are stimulated, the measured response may be affected by activity of other muscles.

If the stimulation electrodes are placed very close to the palm of the hand, the muscles are stimulated directly by the stimulation pulses.

If the current is too strong, it may stimulate the muscles too much.

Moving or touching the patient during measurement may cause incorrect results.

Make sure that the NMT cables does not contact external pacemaker or catheter wires.

To avoid unintentional electrical shocks always make sure that the NMT stimulation has been stopped before touching the electrodes.

Take care to handle the NMT sensor, avoiding forcefully striking the sensor.

After repositioning the patient, check the NMT sensor application site and ensure that the sensor is still properly applied and the thumb can move freely.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 23 - 5

CAUTION

Correct positioning of the electrodes is important. Small displacements may result in considerable changes in stimulation current requirements. Furthermore, the electrodes must be positioned in such a way to avoid direct stimulation of the muscle.

The electrodes should be applied properly to the patient skin. It has been found that slight pressure on the electrodes may improve the stimulation considerably. Therefore, taping the electrodes to the skin may be advisable.

The more distal the sensor is placed on the thumb, the stronger the acceleration signal. This effect can be used to adjust the signal strength.

During NMT measurement, the arm applied with NMT electrodes and sensor should be kept immobile during the whole procedure.

23.8

Calibrating the NMT Measurement

The size of the sensor signal varies from patient to patient. NMT calibration determines supramaximal stimulation current and the reference response amplitude. The reference response amplitude is the twitch at the supramaximal stimulation current when the patient is not paralyzed. The calibration must be done prior to administration of a muscle relaxant drug.

CAUTION

Start calibration before the administration of a muscle relaxant drug (but after the induction of sleep in general anesthesia) to prevent voluntary muscle contraction and tension from interfering with the reference search.

23.8.1

Setting the Calibration Current

If you set Stimulation Current to Supra (60 mA) , the module automatically searches for supramaximal current to determine the reference response amplitude. If you select a value between 0 and 60 mA, the reference response amplitude is determined using the selected stimulation current. For adults, the supramaximal current is usually between 35 and 55 mA. For more information, see

23.11.2 Changing the Stimulation Current

.

23.8.2

Starting NMT Calibration

To Start NMT calibration, follow this procedure:

1.

Select the NMT numeric area to enter the NMT menu.

2.

Select the Setup tab.

3.

Verify that the settings of Stimulation Current and Pulse Width .

4.

Press the calibration hard key on the NMT module or select Calibrate at the bottom of the NMT menu to start calibration.

If calibration succeeds, the message Calibration Completed displays on the NMT menu. If calibration failed, the

NMT module automatically uses the default value as the reference amplitude.

NOTE

It is recommended that the patient be anesthetized before setting up the calibration twitch as nerve stimulation can be painful.

Changing the stimulation current or pulse width after calibration invalidates the stored reference data, and therefore recalibration is required.

23 - 6 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

23.9

Starting NMT Measurements

At the completion of NMT calibration, the monitor automatically starts a TOF mode measurement.

You can also choose either of the following ways to start NMT measurements:

Press the Start/stop key on the NMT module.

From the Measure tab of the NMT menu, select a measurement mode, set Interval , and select Start (XX) to start NMT measurement at corresponding mode.

NOTE

Stop NMT measurements if you need to change the NMT settings.

Take care when removing the sensor from the patient. Do not pull on the cable.

23.10

Stopping NMT Measurements

Choose either of the following ways to stop an on-going NMT measurement.

Press the

Select

Start/stop

Stop All

key on the NMT module.

at the bottom of the NMT menu.

23.11

Changing NMT Measurement Settings

23.11.1

Selecting the NMT Measurement Mode

To select the NMT measurement mode, follow this procedure:

1.

Select the NMT numeric area to enter the NMT menu.

2.

Select the Measure tab.

3.

Select the desired stimulation mode.

4.

If you select TOF Mode , ST Mode , or DBS mode, select Interval to set the time interval between two measurements.

23.11.2

Changing the Stimulation Current

Before calibration and NMT measurement, confirm that the desired stimulus current is selected. To set the simulation current, follow this procedure:

1.

Select the NMT numeric area to enter the NMT menu.

2.

Select the Setup tab.

3.

Set Stimulation Current .

When Stimulation Current is set to Supra , the module automatically searches for supramaximal current to determine the reference response amplitude. For adults, the supramaximal current is usually between 35 and 55 mA. Smaller currents may be desirable for children.

If Stimulation Current is set to a value between 0 and 60 mA, the reference response amplitude is determined using the selected stimulation current.

NOTE

Stimulation current is adjusted in increments of 5 mA.

23.11.3

Changing the Pulse Width

You can increase the pulse width to increase the effect of the stimulation to help finding the supramaximal current. To set the pulse width, follow this procedure:

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 23 - 7

1.

Select the NMT numeric area to enter the NMT menu.

2.

Select the Setup tab.

3.

Set Pulse Width .

23.11.4

Enabling Block Recovery Notification

A message “NMT Block Recovery” is displayed in the technical alarm information area to notify you when the set block recover limit is reached. This indicates that the patient is responding more clearly to the stimuli and the neuromuscular block is decreasing. This notification can be used, for example, to help maintain a certain relaxation level.

To enable the block recovery notation and set the limit for activating the note, follow this procedure:

1.

Select the NMT numeric area to enter the NMT menu.

2.

Select the Setup tab.

3.

Set Block Recovery . If Block Recovery is set to Off , the monitor will not give a note.

23.11.5

Adjusting NMT Stimulation Beep Volume

The monitor gives a beep at the selected volume at each stimulation pulse if the setting is not zero. To adjust the volume of NMT stimulation beep, follow this procedure:

1.

Select the NMT numeric area to enter the NMT menu.

2.

Select the Setup tab.

3.

Set Stimulation Beep Volume .

23.12

Recalling Calibration Information

In the situation that the NMT module is power down, or you want move the NMT module to another monitor along with the patient and you want to continue with the already determined calibration information, including stimulation current, pulse width, and reference response amplitude, you can use the recall function.

To recall the calibration information, follow this procedure:

1.

Select the NMT numeric area to enter the NMT menu.

2.

Select the Recall Reference tab.

3.

Select Recall Reference .

23.13

NMT Troubleshooting

This section lists the problems that might occur. If you encounter the problems when using the equipment or accessories, check the table below before requesting for services. If the problem persists, contact the service personnel.

Problem

NMT calibration and measurement fails.

NMT measurement is disturbed.

Supramaximal current cannot be found.

Solution

1. Check that the electrodes are properly applied.

2. Replacing the electrodes.

1. Do not touch the arm where electrodes are applied.

2. Check that the electrodes and NMT sensor are properly applied.

Check that the electrodes are properly applied. Also, supramaximal current may not be found if the patient is already relaxed.

NOTE

For a comprehensive list of physiological and technical alarm messages, see

D Alarm Messages .

23 - 8 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

24

Monitoring Regional Oxygen Saturation (rSO

2

)

24.1

rSO

2

Introduction

The rSO

2

(Regional Oxygen Saturation) monitoring provides noninvasive and continuous information of changes in regional oxygen saturation of blood. The measurement takes place in real time, providing an immediate indication of a change in the critical balance of regional oxygen delivery and oxygen consumption.

The harmless, near-infrared wavelengths generated by the INVOS System’s light-emitting diodes pass through scalp and bone tissue beneath the sensor. Once in vivo they are either absorbed or scattered back up to the sensor’s shallow and deep detectors. Red-colored hemoglobin molecules within red blood cells have the highest light absorption of the wavelengths used, and the exact shade of red of each hemoglobin molecule indicates the amount of oxygen it is carrying. The type and quantity of absorption data returned to the detectors reflects deoxyhemoglobin and total hemoglobin, from which a regional oxygen saturation (% rSO

2

) value unique to the specific area under the sensor is calculated.

(1)

(2)

(3)

(1)

(2)

(

(3)

(1) Light source

(4)

(2) Shallow detector (3) Deep detector

(5)

(6)

(4) rSO

2

dialog hard key

(6) Module status indicator

(7)

(5) Set the baseline for all rSO

2

channels

(7) rSO

2

cable connector

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 24 - 1

24.2

rSO

2

Safety Information

CAUTION

• rSO

2

monitoring is intended for use in individuals greater than 2.5 kg at risk for reduced-flow or noflow ischemic states.

Do not use the rSO

2 value as the sole basis for making decisions regarding diagnosis or therapy because the rSO

2

values represent a small volume of tissue beneath the sensor and may not reflect oxygenation disturbances that occur elsewhere.

Use only recommended or provided accessories. Using any other sensor will compromise accuracy.

Use of an electrosurgical instrument in the vicinity of the monitor may interfere with the signal and result in inaccurate rSO

2

measurement.

NOTE

Environments with excessive ambient light such as bright sunlight or strong operating room lighting may require loosely covering the area of the sensor with an opaque drape.

24.3

rSO

2

Measurement Limitations

If present, the following may cause inaccurate readings:

Cardiogreen, indigo carmine or other intravascular dyes

Carboxyhemoglobin or other dyshemoglobins

Hemoglobinopathies

Conjugated hyperbilirubinemia (direct)

Myoglobin (Mb) in muscle tissues

24.4

rSO

2

Display

rSO

2

-1 and rSO

2

-2 are from the first inserted rSO

2

module. rSO

2

-1' and rSO

2

-2' are from the second inserted rSO module. Each rSO

2

numeric area displays one rSO

2

channel, which is shown as below:

2

(2) (3)

(1)

(4)

(7)

(6) (5)

(8)

(9)

(1) rSO

2

label

(3) Baseline (BL)

(2) Channel indicator

(4) Signal Strength Indicator (SSI)

24 - 2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

(5) Change percentage of the rSO

2

realtime value and baseline (If the rSO

2

value is greater than or equal to the baseline, the arrowhead is upward; If the rSO

2

value is smaller than the baseline, the arrowhead is downward)

(6) Realtime rSO

(8)rSO

2

2

value

rolling average of latest 60 minutes

(7) Upper and lower rSO

2

alarm limits

(9) Area Under the Curve (AUC)

24.5

Accessing the On-screen rSO

2

Guide

The monitor provides the on-screen rSO

2

guide to help you understand rSO

2

monitoring principles, functions, and operating procedure. To access the on-screen rSO

2

guide, follow this procedure:

1.

Select the rSO

2

numeric area to enter the rSO

2

dialog.

2.

Select the Introduction tab.

3.

Select the desired tab of Summary Target Patients , Points To Note , or Operation Guides .

24.6

Preparing for rSO

2

Monitoring

1.

Select the site for sensor placement. For more information, see 24.6.1 rSO

2

Sensor Site Selection .

2.

Prepare the patient skin. For more information, see

24.6.2 Preparing the Skin

.

3.

Apply the rSO

2

sensor. For more information, see

24.6.3 Applying the rSO

2

Sensor

.

4.

Connect the rSO

2

parts. For more information, see 24.6.4 Connecting the rSO

2

Parts .

24.6.1

rSO

2

Sensor Site Selection

The rSO

2

sensor can be placed on the proper cerebral or somatic site.

■ Cerebral site selection: select sensor site on the right and/or left side of forehead. Placement of the sensor in other cerebral locations, or over hair, may cause inaccurate readings.

■ Somatic site selection: select sensor site over tissue area of interest (site selection will determine which body region is monitored)

CAUTION

For cerebral site selection, do not place the sensor over nevi, sinus cavities, the superior sagittal sinus, subdural or epidural hematomas, injured skin or other anomalies such as arteriovenous malformations, as this may cause readings that are not reflective of brain tissue or no readings at all.

For somatic site selection, avoid placing the sensor over thick fatty deposits, hair or bony protuberances. Do not place the sensor over nevi, hematomas or broken skin, as this may cause readings that are not reflective of somatic tissue or no readings at all.

NOTE

For the somatic site selection, see the instruction for use of corresponding sensor for detailed information.

24.6.2

Preparing the Skin

To achieve the optimal measurement result, the sensor application site should be clean and dry. To properly prepare the skin, follow this procedure:

1.

Shave hair from skin at chosen sites.

2.

Gently rub skin surface at sites to remove dead skin cells.

3.

Thoroughly cleanse the site with a mild soap and water solution.

4.

Dry the skin completely before applying the sensors.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 24 - 3

24.6.3

Applying the rSO

2

Sensor

Follow this procedure to apply the rSO

2

sensor to patient:

1.

Remove the protective backing label from the adhesive side of the sensor and apply the sensor to the skin.

2.

Continue applying the sensor by smoothing it to the skin from the center outward. Ensure edges of the sensor are sealed to prevent light from entering.

CAUTION

The sensor is designed for single use only and can not be reused. Reuse may cause inaccurate readings, erratic readings, or no readings at all. Also, reuse may cause an increased risk of crosscontamination among patients.

The sensor is designed for external use only as described in the instructions. Do not use the sensor internally for any reason.

Do not place the sensor on regions with severe tissue edema to reduce the possibility of skin lesions.

Do not autoclave or gas sterilize the sensor.

NOTE

Use care when placing orremoving sensor. Do not place on broken or undeveloped skin.

For extended monitoring, it is recommended to use a new sensor every 24 hours or if adhesive is inadequate to seal the sensor to the skin.

To avoid pressure sores do not apply external pressure (e.g. headbands, wraps, or tape) to sensor.

24.6.4

Connecting the rSO

2

Parts

1.

Connect Preamplifier(s) to the rSO

2

module. Align red dot on the silvery connector of the Preamplifier with red dot on rSO

2

cable connector.

2.

Connect the following devices as per patient category:

For neonatal patients, connect the neonatal sensor directly to the Preamplifier.

For pediatric or adult patients, a reusable sensor cable is needed for the connection of the

Preamplifier and the sensor. Use color-coding when connecting the Preamplifier and the reusable sensor cable.

(2)

(3)

24 - 4

(1)

(1) rSO

2 module (2) Preamplifier (3) Reusable sensor cable

NOTE

Different sensors (adult, pediatric and infant/neonatal) cannot be used simultaneously on the same monitor. Cerebral sensors can be used with somatic sensors on the same monitor.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

24.7

Changing rSO

2

Settings

24.7.1

Changing rSO

2

Alarm Settings

To change the rSO

2

alarm settings, follow this procedure:

1.

Select the rSO

2

numeric area to enter the rSO2 dialog.

2.

Select the Alarm tab.

3.

Enter the password if required. For more information, refer to 32.14.3 Selecting Password for User

Authentication

.

4.

Set alarm properties as desired.

24.7.2

Setting the rSO

2

Auto Low Limit Switch

To set the rSO

2

auto low alarm limit, follow this procedure:

1.

Select the rSO

2

numeric area to enter the rSO2 dialog.

2.

Select the Alarm tab.

3.

Set Auto Low Limit switch.

◆ If Auto Low Limit is switched on, rSO2-1 Variance and rSO2-2 Variance are activated to allow you to set the percentage of rSO

2

low limits below the baseline. Then the monitor calculates the rSO

2

low limits automatically based on the setting.

◆ If Auto Low Limit is switched off, the rSO

2

low limits should be set manually.

24.7.3

Setting the rSO

2

Label

To set the rSO

2

label, follow this procedure:

1.

Select the rSO

2 numeric area to enter the rSO2 dialog.

2.

Select the Setup tab.

3.

Set rSO2-1 Label and rSO2-2 Label .

24.7.4

Setting the AUC Mode

To set the AUC mode, follow this procedure:

1.

Select the rSO

2

numeric area to enter the rSO2 dialog.

2.

Select the Setup tab.

3.

Set AUC Mode .

Select Fixed to activate the item Fixed Threshold . In this case, AUC is calculated according to the configured fixed threshold.

◆ Select Below Base Percentage to activate the item Percentage Below Baseline . In this case, AUC is calculated according to the configured percentage below baseline.

24.7.5

Setting the Baseline

To set the rSO

2

baseline for the respective channel, follow this procedure:

1.

Select the rSO

2

numeric area to enter the rSO2 dialog.

2.

Select the Baseline tab.

3.

Select Set Baseline . The monitor automatically sets the current rSO

2

value as the baseline.

You can also set the baselines for all rSO

2

channels by the following two methods:

Select Set Baseline s in the lower left corner of the rSO

2

dialog.

Press the hard key at the upper right of the rSO

2

module.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 24 - 5

Then the monitor set the baselines for all rSO

2

channels according to their respective current rSO

2

values.

NOTE

Set the rSO

2

value measured when patient is sober and eupraxic as baseline. The baseline will be set automatically if it is not set within 5 to 6 minutes and current rSO

2

value is effective.

24.7.6

Selecting rSO

2

Parameters for Display

In the rSO

2

numeric area, rSO

2

and SSI are permanently displayed parameters, and other parameters are selectable. To select the parameters for display, follow this procedure:

1.

Select the rSO

2

numeric area to enter the rSO2 dialog.

2.

Select the Select Parameter tab.

3.

Select the parameters for display.

24.8

Covidien Information

Covidien Patents

This posting serves as notice under 35 U.S.C.§287(a) for Covidien patents: http://www.covidien.com/patents.

No Implied License

Possession or purchase of this device does not convey any express or implied license to use the device with unauthorized replacement parts or consumables which would, alone, or in combination with this device, fall within the scope of one or more of the patents relating to this device.

24 - 6 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

25

Review

25.1

Review Overview

Trends are patient data collected over time and displayed in graphic, tabular, or other forms to give you a picture of how the patient's condition is developing. The Review dialog also enables access to review the events, 12lead ECG analysis results and waveforms, full disclosure waveforms, and so on.

25.2

Review Dialog

The Review dialog contains tabs to display trend data in tabular, graphic, or other forms.

25.2.1

Accessing the Review Dialog

Choose one of the following methods to enter the review dialog:

Select the Review quick key. If reviewing patient data is password protected, input the monitor's clinical password (local password).

Select the Main Menu quick key → from the Review column select the desired option. If reviewing patient data is password protected, input the monitor's clinical password (local password).

25.2.2

Example Review Dialog

The review pages have a common structure. The graphic trends review page is taken as an example:

(1)

(2)

(5)

(3)

(4)

(1)

(2)

(3)

(4)

Event type indicator: different color blocks match different types of events:

■ Red: high priority alarm event

■ Yellow: medium priority alarm event

Cyan: low priority alarm event

Green: manual event

White: operation-related event

Current screen time line: indicates the time length of the current screen.

Waveform area: displays trend curves. The color of trend curves is consistent with the color of parameter labels.

Time line: indicates the entire time length.

■ : indicates the time length of trend data.

■ : indicates the time length of no trend data.

can be moved within this time length.

cannot be moved within this time length.

■ Different color blocks along the time line indicate events of different types. See the color definition for the event type indicator.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 25 - 1

(6)

(7)

(8)

(9)

(5)

(6)

(7)

(8)

(9)

Event area: displays the event of the cursor time. Selecting the event accesses the event list. If there is no event at the cursor time, the cursor time is displayed.

Numeric area: displays numeric values at the cursor indicated time. The background color of numeric values matches the alarm priority.

Cursor: this is adjusted by touching different locations. Adjusting this changes the data displayed in the Event area and Numeric area.

Slider: indicates the position of current screen time in the entire time length. Dragging this button left or right enables you to locate the trend data at a specific time and also refreshes trend data in current screen accordingly.

Button area.

25.2.3

Symbols on Review Pages

The following table lists the symbols on review pages.

Symbol

or

or

Description

Slider: indicates the position of current screen time in the entire time length. Dragging the slider left or right enables you to locate the trend data at a specific time and also refreshes data in current screen accordingly.

Goes to the previous or next event.

Event list: displays events in a chronological order. The most recent event is displayed at the top. The number of asterisk symbols before an event indicates alarm priority.

Selecting this symbol displays two review pages simultaneously.

+

Record button: select it to output patient information and data through the recorder.

Print button: select it to output patient information and data through the printer.

Indicates that the followed parameter is from an external device connected to the monitor via the BeneLink module.

25.2.4

Common Operations

This section describes common operations for all review pages.

25.2.4.1

Browsing Trend Data

Browse trend data in one of the following ways:

Move the cursor.

Move the slider .

Slide your finger on the screen.

25.2.4.2

Viewing Events

You can view the following types of events:

Manually triggered events

Parameter-related operation events and alarm-related events, such as starting NIBP measurement

Operation events not related to parameters, such as system time change

View events in either of the following ways:

Select and select the desired event.

Select or to jump to the previous event or next event.

Events are displayed in a chronological order. The most recent event is displayed at the top. The number of asterisk symbols before an event indicates alarm priority as follows:

25 - 2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

***: high priority alarm

**: medium priority alarm

*: low priority alarm

25.2.4.3

Displaying Two Review Pages Simultaneously

For N22/N19, you can display two review pages simultaneously. To do so, follow this procedure:

1.

Enter the desired review page by one of the following methods:

◆ Select the Review quick key → select the desired tab.

◆ Select the Main Menu quick key → from the Review column select the desired item.

2.

Select (for landscape display) or navigated independently.

(for portrait display) to open a duplicate view that can be

25.2.5

Tabular Trends Review Page

The tabular trends review page displays trend data in a tabular form.

25.2.5.1

Entering the Tabular Trends Review Page

Choose one of the following methods to enter the tabular trends review page:

Select the Review quick key

→ select the Tabular Trends tab.

Select the Main Menu quick key

from the Review column select Tabular Trends .

25.2.5.2

Changing the Tabular Trend Group

To change the tabular trend group, follow this procedure:

1.

Enter the tabular trends review page.

2.

Set Trend Group .

25.2.5.3

Editing the Tabular Trend Group

The setting of the Trend Group defines the contents of displayed trends. It is possible to edit any trend group except for the All and Standard trend groups. To do so, follow this procedure:

1.

Enter the tabular trends review page:

2.

Select Group Setup

→ select the desired tab.

◆ Rename the trend group (optional): select the desired name.

symbol at the upper left corner and input the

◆ Add parameters: select desired parameters from the Choices column, and then select Add . Selecting

Add All adds all parameters (up to the maximum of 83).

◆ Delete parameters: select desired parameters from the Selected column, and then select Delete .

Selecting Delete All deletes all parameters except for HR.

◆ Move the position of parameters displayed on the trend page: select desired parameters from the

Selected column, and then select Move up , Move Down , Move to Top , or Move to Bottom .

◆ Display two parameters side by side: select two parameters from the Selected column, and then select Overlap . Then, the symbol appears at the right of the overlapped parameters. The overlapped parameters will be displayed side by side in one numerics area and waveform area.

◆ Unoverlap two parameters: select the overlapped parameters from the Selected column, and then select Unoverlap .

Selecting Default will restore the trend group settings to factory defaults.

NOTE

Trend groups labeled All or Standard cannot be edited.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 25 - 3

The HR parameter and waveform are always displayed in the first row on the trend page. It cannot be deleted or moved.

25.2.5.4

Changing the Resolution of Tabular Trend Data

The interval of tabular trends defines the interval of displaying trend data. A short interval is especially suited for neonatal applications, where the clinical situation may change very quickly. In adult monitoring, where the patient's status typically changes more gradually, a longer interval may be more informative.

To change the interval of trend data, follow this procedure:

1.

Enter the tabular trends review page.

2.

Select Interval .

◆ 5 sec or 30 sec : select to view up to 4 hours of tabular trends at an interval of 5 seconds or 30 seconds.

◆ 1 min, 5 min, 10 min, 15 min, 30 min, 1 hr, 2 hrs, or 3 hrs : select to view up to 120 hours of tabular trends at selected interval.

Select parameters, such as NIBP, or GCS, to view the tabular trends when these parameter measurements are acquired.

NOTE

Changing between intervals does not typically change the position of the slider or cursor except to round it to the nearest interval, however changing intervals to be 30 seconds or less will move any cursor position more than 4 hours ago to the oldest data less than 4 hours ago.

25.2.5.5

Printing a Tabular Trends Report

To print a tabular trends report, follow this procedure:

1.

Enter the tabular trends review page.

2.

Select to enter the Print Setup dialog.

3.

Set the tabular trends report as described in 29.6.3 Configuring Tabular Trends Reports .

4.

Select Print or Print (All) .

◆ Print : the trend data before the above set Time , for the above set Period , and at the above set

Interval will be printed.

◆ Print (All) : all the stored trend data will be printed at the review interval.

25.2.5.6

Recording Tabular Trends

To print tabular trends with the recorder, follow this procedure:

1.

Enter the tabular trends review page.

2.

Select to enter the Record Setup dialog.

3.

Respectively set Start Time and End Time of the tabular trends data.

4.

Select Record or Record (All) .

◆ Record : the trend data between the Start Time and End Time will be recorded.

◆ Record (All) : all the stored trend data at the review interval will be recorded.

25.2.6

Graphics Trends Review Page

The graphic trends review page displays trend data in a visual format.

25.2.6.1

Entering the Graphic Trends Review Page

Choose one of the following methods to enter the graphic trends review page:

Select the Review quick key → select the Graphic Trends tab.

Select the Main Menu quick key → from the Review column select Graphic Trends .

25 - 4 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

25.2.6.2

Changing the Graphic Trend Group

To change the graphic trend group, follow this procedure:

1.

Enter the graphic trends review page.

2.

Set Trend Group .

25.2.6.3

Editing the Graphic Trend Group

1.

Enter the graphic trends review page.

2.

Select Group Setup .

For more information, see 25.2.5.3 Editing the Tabular Trend Group for detail on how to edit the group.

25.2.6.4

Changing the Resolution of Trend Data

To change the length of trend data displayed on the current screen, follow this procedure:

1.

Enter the graphic trends review page.

2.

Select Zoom .

8 min: the screen displays eight minutes of trend data. You can view the most recent one hour of data.

◆ 30 min, 1 hr, 2 hrs, 4 hrs : the screen displays 30 minutes, one hour, two hours, or four hours of trend data. You can view the most recent four hours of data.

8 hrs, 12 hrs, 24 hrs, 48 hrs : the screen displays 8 hours, 12 hours, 24 hours, or 48 hours of trend data.

You can view the most recent 120 hours of data.

25.2.6.5

Changing the Number of Waveforms

To change the number of waveforms displayed on the trend review page, follow this procedure:

1.

Enter the graphic trends review page.

2.

Select Trends .

25.2.6.6

Printing a Graphic Trends Report

To print a graphic trends report, follow this procedure:

1.

Enter the graphic trends review page.

2.

Select to enter the Print Setup dialog.

3.

Set the graphic trends report as described in

29.6.4 Configuring Graphic Trends Reports

.

4.

Select Print to print the graphic trend data before the above set Time and for the above set Period .

25.2.7

Events Review Page

The monitor stores events in real time, including technical alarm events, physiological alarm events, manual events, and operational events. When an event occurs, all the measurement numerics and three event-related waveforms 16 seconds before and after the event are stored.

NOTE

Alarms are saved as events and will be maintained if the equipment is powered down. The time of equipment power down is not recorded as an event and cannot be reviewed.

Earlier events will be overwritten by later ones if the storage capacity is reached.

A total loss of power does not affect the events already stored.

25.2.7.1

Entering the Events Review Page

Choose one of the following methods to enter the events review page:

■ Select the Review quick key → select the Events tab.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 25 - 5

■ Select the Main Menu quick key → from the Review column select Events .

The Event page displays the event list. Events are displayed in descending chronological order, with the most recent displayed at the top. The number of asterisk symbols before an event indicate alarm priority as described

in 8.3.3 Alarm Indicators

.

Different color blocks are displayed on the left of each event to indicate different event types.

Red: high priority alarm event

Yellow: medium priority alarm event

Cyan: low priority alarm event

Green: manual event

White: operation-related event

The number of currently selected events and the total number of events are displayed at the top right corner of the event list. For example, 2/4 indicates that the selected event is the second event in the filtered events and the total number of filtered events is 4. Total indicates the total number of events. For example: Total : 28 means that there are a total of 28 events.

25.2.7.2

Configuring the Filter

You can filter events to facilitate event review. To configure the filter, follow this procedure:

1.

Enter the Events page.

2.

Select Filter . From the drop-down list, select the desired filter item.

You can create up to two sets of custom filter. To do so, follow this procedure:

1.

From the Filter drop-down list, select Custom 1 or Custom 2 to enter the Filter Setup menu.

2.

Select the Name field to edit the name of the custom filter criterion.

3.

Select the desired items.

If you want to review events which happened around certain time, select the button → set the time → select Ok

.

The cursor then jumps to the event happened closest to the time entered.

25.2.7.3

Editing Events

To edit events, follow this procedure:

1.

Enter the Events page and tick off the desired events.

2.

Select to edit the selected events.

◆ Lock : manually lock the event. Locked events cannot be deleted.

◆ Note : enter comments for the event.

◆ Rename : allow renaming an event name. Only manual events and arrhythmia events can be renamed if enabled by the hospital’s settings. For more information, see

32.6.2 Renaming Events

.

25.2.7.4

Viewing Event Details

To view waveforms and parameter values at the event time, follow this procedure:

1.

Enter the Events page.

2.

Select Detail .

To display beat labels on the first ECG waveform, switch on Beat Anno . The white beat labels indicate heart beats classification and may explain suspected, missed, or false arrhythmia calls. Heart beats are classified as follows:

N = Normal

V = Ventricular ectopic

S = Supraventricular premature

P = Paced

L = Learning

? = Insufficient information to classify beat

25 - 6 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

I = Inoperative (for example, Lead Off )

M = Missed beat

If you switch on Beat Anno on the Events page, beat labels will also be displayed on the Full Disclosure page, and vise versa. Beat labels can be printed out.

25.2.7.5

Printing Event Reports

To print event reports via a printer, follow this procedure:

1.

Enter the events review page.

2.

Select to enter the Print Setup dialog.

3.

Select the desired options.

◆ Print All Event List : print the entire event list.

◆ Print List of Selected Events : print the list of selected events.

◆ Print Detail of Selected Events : print the details of selected events.

◆ Print Displayed Event Detail : print the waveforms and parameters of the currently displayed event.

4.

Select Print .

25.2.7.6

Recording Event Details

To print events details via a recorder, follow this procedure:

1.

Enter the events review page.

2.

Select the Detail button.

3.

Select to print the details of the event.

NOTE

When a parameter label is prefixed with the plus sign (+) on a printout, it indicates that the data is from an external device.

25.2.8

Full Disclosure Review Page

You can review up to 48 hours of waveform data on the full disclosure review page. You can view both the compressed waveforms and numeric values.

25.2.8.1

Entering the Full Disclosure Review Page

Choose one of the following methods to enter the full disclosure review page:

Select the Review quick key → select the Full Disclosure tab.

Select the Main Menu quick key → from the Review column select Full Disclosure .

25.2.8.2

Selecting Waveforms

You can select the waveforms to be displayed and stored. Up to 48 hours of waveforms for all parameters can be stored. When closing the Select Waveform dialog, a prompt will inform you of the duration of waveforms that can be saved if the amount is less than 48 hours. To save and display the desired waveforms, follow this procedure:

1.

Enter the full disclosure review page.

2.

Select Setup to enter the Select Waveform page.

3.

Select the Storage tab and set the desired waveforms to be stored in the monitor. Select the Display

(Maximum: 3) tab and set the desired waveforms to be displayed on the Full Disclosure page.

NOTE

Storing more waveforms may shorten the storage time to less than 48 hours. Please adjust the waveforms stored with this consideration.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 25 - 7

Parameter labels from external devices are prefixed with the plus sign (+).

In case of alarms, the background of the compressed waveform is highlighted with a colored block during the alarm time:

Red: high alarm priority

Yellow: medium alarm priority

Cyan: low alarm priority

25.2.8.3

Setting the Scale and Duration

To set the length and size of displayed compressed waveforms, follow this procedure:

1.

Enter the full disclosure review page.

2.

Set the waveforms scale in either of the following ways:

Select the Scale button. If you do not see Scale , select the

Select the waveform label.

3.

Select Duration to set the length of displayed waveforms.

button to show it.

4.

Select the parameter label beside the waveforms to set scale for each parameter.

25.2.8.4

Viewing Details of Compressed Waveforms

To view the full waveforms and numeric values, follow this procedure:

1.

Enter the full disclosure review page.

2.

Select Detail .

You can perform the following operations on the this page:

Switch on Beat Anno . For more information, see

25.2.7.4 Viewing Event Details

.

Select Speed , ECG Gain , or Save As Event . If you do not see an option, then select the show it.

Select Overview to switch to the compressed waveform view.

button to

25.2.8.5

Printing the Full Disclosure Waveform Report

To print a compressed waveform report, follow this procedure:

1.

Enter the full disclosure review page.

2.

Select to enter the Print Setup dialog.

3.

Select waveform type you want to print:

Print Waveform Overview : compressed waveform of selected period will be printed. This option is only available when printing from the Overview or compressed waveform view.

◆ Print Displayed Waveform Detail : parameter values and all stored waveforms around the cursor time will be printed. This option is only available when printing from the Detail view.

4.

Select Print .

NOTE

When a parameter label is prefixed with the plus sign (+) on a printout, it indicates that the data is from an external device.

25.2.8.6

Recording the Details of the Compressed Waveform

To record a Full Disclosure Detail Record Report, follow this procedure:

1.

Enter the Full Disclosure page.

2.

Select Detail .

25 - 8 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

3.

Select .

NOTE

When a parameter label is prefixed with the plus sign (+) on a printout, it indicates that the data is from an external device.

25.2.9

Arrhy Statistics Review Page

The Arrhy Statistics review page shows arrhythmia events happened over the lastest 24 hours. This page is only available with the ECG 24h Summary option.

To enter the Arrhy Statistics review page, choose one of the following methods:

Select the Review quick key

→ select the Arrhy Statistics tab.

Select the Main Menu quick key

from the Review column select Arrhy Statistics .

Selecting each type of arrhythmia statistics bar shows compressed waveforms, the time and duration of each arrhythmia event. Selecting the compression waveform shows the event details.

25.2.10

OxyCRG Review Page

You can review up to 48 hours worth of 4 minute trend curves on the OxyCRG review page. The OxyCRG review functionality is available for neonatal monitoring only.

NOTE

OxyCRG Review Page is available only when Patient Category is set to Neo.

25.2.10.1 Entering the OxyCRG Review Page

Choose one of the following methods to enter the OxyCRG review page:

From the OxyCRG screen, select the ABD events list area.

Select the Review quick key

→ select the OxyCRG tab.

Select the Main Menu quick key

from the Review column select OxyCRG .

25.2.10.2 The Display of the OxyCRG Review Page

The following figure shows the OxyCRG screen:

(1)

(2) (2)

(4)

(3)

(1) Event title area: displays information of the selected event, such as the event type and time.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 25 - 9

(2) Event detail area: displays 4 minutes of parameter trends and compressed waveform, and parameter values of the selected event.

(3) Event summary area: displays ABD events within the Zoom period. The selected event is enclosed in a white frame.

(4) Event statistics area: displays the total number of ABD events and the numbers of each event within the Zoom period.

25.2.10.3 Changing the Resolution of Trend Curves

To set the resolution of trend curves, follow this procedure:

1.

Enter the OxyCRG review page.

2.

Set Zoom .

NOTE

If the Zoom is set for a large duration (for example, 8 hrs), the screen may show "No Patient Data" for the first several minutes of data collection (for example, 6 minutes), but data will be shown after enough data has been collected.

25.2.10.4 Printing an OxyCRG Review Report

To print an OxyCRG review report, follow this procedure:

1.

Enter the OxyCRG review page.

2.

Select to enter the Print Setup dialog.

3.

Select Print . The report duration corresponds to the duration of data shown on the screen.

25.2.11

12-Lead ECG Review Page

When 12-lead ECG analysis is performed, you can review the most recent 20 events of 12-lead analysis. For more

information, see 10 Resting 12-Lead ECG Analysis

.

25.2.11.1 Entering the 12-Lead Review Page

Choose one of the following methods to enter the 12-lead ECG review page:

Upon completion of 12-lead ECG analysis, select Review from the 12-Lead Interpretation page. For more

information, see 10 Resting 12-Lead ECG Analysis

.

Select the Review quick key → select 12-Lead ECG .

Select the Main Menu quick key → from the Review column select 12-Lead ECG .

25.2.11.2 Switching to Median Complex (for Glasgow Algorithm Only)

The median complex template displays 12-lead ECG waveforms on one page in 4 columns, with 3 lines in each column, and one rhythm lead waveform at the bottom. Additionally, a short vertical bar appears above each waveform, marking the start and end positions of the P-wave and the QRS-wave and the end position of T-wave.

To view Median Complex, follow this procedure:

1.

Enter the 12-lead review page.

2.

Select Median Complex .

Selecting Waveform can return to the 12-lead ECG waveform page.

25.2.11.3 Setting 12-Lead ECG Waveforms

To set the 12-lead ECG waveforms on the review page, follow this procedure:

1.

Enter the 12-lead review page.

25 - 10 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

2.

Set Speed , Gain , and Layout .

25.2.11.4 Printing the 12-Lead ECG Report

To print the 12-Lead ECG report, follow this procedure:

1.

Enter the 12-lead review page.

2.

Select .

3.

Configure the 12-lead ECG report.

4.

Select Print .

25.2.11.5 Recording the 12-Lead ECG Report

To print the 12-Lead ECG report via the recorder, follow this procedure:

1.

Enter the 12-lead review page.

2.

Select .

25.2.12

ST Review Page

When ST analysis is enabled, the monitor saves ST segments and values at an interval of one minute. You can review the latest 120 hours of ST data.

25.2.12.1 Entering the ST Review Page

Choose either of the following methods to enter the ST review page:

Select the Review quick key

→ select the ST tab.

Select the Main Menu quick key → from the Review column select ST .

25.2.12.2 Setting the ST Reference

You can set the currently displayed ST as reference. To do so, follow this procedure:

1.

Enter the ST review page.

2.

Select Set Reference .

NOTE

The ST baseline is used as ST reference by default.

25.2.12.3 Displaying/Hiding the ST Reference

To display or hide ST reference, follow this procedure:

1.

Enter the ST review page.

2.

Select Display Reference or Hide Reference .

25.2.12.4 Displaying/Hiding Markers

To display or hide markers, follow this procedure:

1.

Enter the ST review page.

2.

Select Display Marker or Hide Marker .

25.2.12.5 Printing ST Data

To print ST data, follow this procedure:

1.

Enter the ST review page.

2.

Select .

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 25 - 11

25.3

Reviewing Discharged Patients

For discharged patients, you can review the trend data in the review page. You can also review the events and

12-lead ECG analysis results.

Access the Discharged Patients dialog box by either of the following ways:

◆ Select the Discharged Patients quick key. If viewing discharged patients is password protected, input the user name and password (the user name and password saved in the MLDAP server).

◆ Select the Main Menu quick key

→ from the Patient Management column select Discharged

Patients .

If viewing discharged patients is password protected, input the user name and password (the user name and password saved in the MLDAP server).

25.3.1

Checking the Data of a Discharged Patient

1.

Access the data of the discharged patient as described in

25.3 Reviewing Discharged Patients .

2.

From the patient list select the desired patient.

3.

Select Detail .

25.3.2

Checking the Information of a Discharged Patient

1.

Access the data of the discharged patient as described in

25.3 Reviewing Discharged Patients .

2.

Select the icon to enter the Patient Management dialog box.

3.

Select Ok to exit the Patient Management dialog box.

25 - 12 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

26

Clinical Assistive Applications (CAA)

The Clinical Assistive Applications (CAA) function integrates some commonly used clinical guidelines and tools into the monitor. It puts the currently monitoring parameter measurements together and provides comprehensive analysis results.

CAA can improve the clinician’s working efficiency. However, it is not directly used for diagnosis and cannot not replace the clinician’s judgement.

26.1

Checking Software Licenses

Licenses are required to run the following CAA functions in the monitor:

HemoSight

Early Warning Score (EWS)

SepsisSight

ECG 24h Summary

To check the licenses, select the Main Menu quick key → select License → Local .

To install the licenses, follow this procedure:

1.

For N17/N15/N12, connect the USB drive with the licenses in to the monitor’s USB connector. For N22/N19. connect the USB drive to the monitor’s MSB connector.

2.

Select the Main Menu quick key

→ select License

→ select External .

3.

Select Install .

26.2

Early Warning Score (EWS)

The Early Warning Scores (EWS) can help you recognize the early sign of deterioration in patients based on vital signs and clinical observations. It is based on the “Royal College of Physicians (2015) National Early Warning

Score (NEWS2): Standardizing the assessment of acute illness severity in the NHS, Report of a Working Party. RCP,

London. Depending on the score calculated, appropriate recommendations are displayed.

The monitor supports the following scores:

MEWS (Modified Early Warning Score)

NEWS (National Early Warning Score)

NEWS2 (National Early Warning Score 2)

Custom Score

A subscore is given for each parameter based on the measured or entered value. When all the required parameters are entered or measured, the subscores are added together to calculate the total early warning score. Each subscore has a color coding to indicate associated level of risk. When the total score is outside of the thresholds, actions are recommended. MEWS, NEWS and NEWS2 can give total scores.

Custom Score is based on user-defined parameters.

MEWS, NEWS and NEWS2 are intended for adult patients only. The patient category applied to the Custom Score is defined by the Mindray Clinical Score Configuration Tool. For more information, see Mindray Clinical Scoring

Config Tool Instruction for Use (P/N: 046-012986-00).

WARNING

EWS is not intended for pediatric and neonatal patients.

The EWS scores and recommended actions are for reference only and cannot be directly used for diagnostic interpretation.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 26 - 1

EWS cannot be used as an prognosis index. It is not a clinical judgement tool. Clinicians must use their clinical judgement in conjunction with the EWS tool at all times.

MEWS and NEWS are intended for adult patients only. NEWS is not applicable to pregnant woman and COPD (Chronic Obstructive Pulmonary Disease) patients.

NOTE

The EWS scores does not trigger any alarms and cannot used for continuous monitoring.

26.2.1

Displaying the EWS Numerics Area

To display the EWS numerics area, follow this procedure:

1.

Access Tile Layout in either of the following ways:

◆ Select the Screen Setup quick key → select the Tile Layout tab.

◆ Select the Main Menu quick key → from the Display column select Tile Layout .

2.

Select the parameter area where you want to display the EWS score, and then from the popup list select

EWS .

(2)

(1)

(3)

(8)

(4)

(5)

(6)

(5)

(6)

(7)

(8)

(1)

(2)

(3)

(4)

(7)

EWS protocol label

Scoring countdown: time to the next scoring.

Scoring interval

The scoring time

Single parameter whose score reaches 3

Risk level indicator. The level of risk increases from top down. The current level is enclosed by a white square frame.

History total score. The rightmost one is the latest history score.

Total score. The color of the circle indicates the level of risk.

26.2.2

Accessing the EWS Screen

Access the EWS window in any of the following ways:

Select the EWS parameter area

Select the EWS quick key.

Select the Screen Setup quick key → select the Choose Screen tab → select EWS .

Select the Main Menu quick key → from the CAA column select EWS .

When the EWS screen is hidden as

EWS screen.

, you can also choose one of the following methods to quickly enter the

Swipe left or right across the touchscreen with two fingers until you switch to the EWS screen.

Swipe right across the touchscreen with a single finger ,

26 - 2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

■ Select the button ,

Take NEWS2 as an example, the EWS screen is shown as follows. Your screen may be slightly different due to the configuration.

(12) (11) (10) (9)

(1)

(2)

(8)

(7)

(6)

(3)

(4)

(1)

(2)

(3)

(4)

(5)

(6)

(7)

(8)

(9)

(10)

(11)

(12)

(5)

Total score. The color of the circle indicates the level of risk.

Risk level indicator. The level of risk increases from top down. The current level is enclosed by a white frame.

Parameter area: display the subscore and parameter value of each parameter. The keyboard symbol indicates that the parameter value is manually entered.

History total scores area: selecting this area or swiping up with one finger can review the trends of total score and each subscore

Selecting this button expands the History Total Scores area where your can review the trends of the total score and each subscore.

Selecting this button or swiping right across the screen with one finger displays the trends of total score and parameter values for scoring.

Clinician ID (displays only when the Clinician ID is enabled): allows inputing the Clinician ID to associate with the

EWS score. For more information, see

32.4.1 Enabling Inputing the Clinician ID .

Select this button to see the clinical response to the current score

Scoring interval

EWS protocol label

Scoring countdown: time to the next scoring.

The scoring time

26.2.3

Performing EWS Scoring

CAUTION

The decision to use Scale 2 of the SpO2 Scale should be made by a competent clinical decision maker and should be recorded in the patient’s clinical notes.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 26 - 3

NOTE

Before calculating the score, select Reset to clear the previous score.

The keyboard symbol at the right of the parameter value indicates that the value is manually entered.

You can get the score only when all required parameters have been measured or entered.

When a patient is discharged, or the monitor is turned off, the clinician ID is cleared

To perform scoring, follow this procedure:

1.

Select Reset to clear the previous score and update values of currently monitored parameters and relevant subscores.

2.

For NEWS2, set the SpO2 Scale .

Scale 1 : for patient without hypercapnic respiratory failure.

Scale 2 : f or patients with a prescribed oxygen saturation requirement of 88–92% (for example, in patients with hypercapnic respiratory failure).

3.

Measure or manually enter other required parameters and observations.

4.

If the clinician ID is enabled, input the clinician information by selecting and manually entering the information, or by scanning the clinician’s barcode.

5.

Select Calculate to get the total score.

6.

If Score Confirmation is enabled, select Confirm to save current scoring, or select Cancel to give up

current scoring. Refer to section 26.2.4.2 Setting the Scoring Confirmation Switch for more information.

26.2.4

Changing EWS Settings

26.2.4.1

Changing the Scoring Protocol

The monitor is configured with a default scoring protocol. To change the scoring protocol, follow this procedure:

1.

From the EWS page select Setup .

2.

Set Score .

26.2.4.2

Setting the Scoring Confirmation Switch

To select if confirmation is required before saving score, follow this procedure:

1.

From the EWS page select Setup .

2.

Set Score Confirmation switch.

Off: the monitor automatically saves the scoring result after the scoring is completed.

On: you need to confirm that whether the scoring result is saved or not after the scoring is completed.

26.2.4.3

Setting the Manual Data Timeout

The manually input parameter data can be configured to become invalid after a preset time. To set the timeout period for the input data, follow this procedure:

1.

From the EWS screen select Setup .

2.

From the Manual Data Timeout area, select a desired parameter and set its timeout period.

NOTE

If the data is expired and not updated, the monitor displays the corresponding parameter score in outline font, and gives a timeout alarm.

26.2.4.4

Setting Auto Scoring

The monitor automatically starts scoring at the preset interval. To set auto scoring, follow this procedure:

1.

From the EWS page select Setup .

26 - 4 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

2.

Set Auto Scoring :

◆ Interval : the monitor automatically starts scoring at the preset interval.

◆ NIBP : the monitor automatically starts scoring at the completion of each NIBP measurement.

◆ Alarm : the monitor automatically starts scoring when an alarm occurs to the parameter for scoring.

◆ If no option is selected, the monitor does not initiate auto scoring.

26.2.4.5

Setting Auto Scoring Interval

1.

From the EWS page select Setup .

2.

Set Interval :

◆ By Score : the monitor automatically starts scoring as per the interval selected for corresponding total score.

5 min - 24 h : If Auto Scoring is set to Interval , the monitor automatically starts scoring as per the selected interval. If Auto Scoring is not set to Interval , the countdown timer of manual scoring is selected.

26.2.5

Viewing Historical Scores

From the EWS screen, you can view the total score or subscores of the past 24 hours. To do so, choose either of the following ways:

Select the history total score area.

From the history total score area, swipe up with a finger.

To return to the EWS screen, select

from the top of the historical score screen. Refer to 26.2.2 Accessing the

EWS Screen for the position of the history total score area.

26.2.6

Viewing Parameter Trends

From the EWS screen, you can view the 24-hour graphic trends of each parameter used for scoring. To do so, choose either of the following ways:

Select the button.

Swipe right across the EWS screen with a finger.

From the screen of parameter trends, selecting the button or swiping left across the screen with a finger can return to the EWS screen.

Refer to 26.2.2 Accessing the EWS Screen for the position of the

button.

26.3

Glasgow Coma Scale (GCS)

The Glasgow Coma Scale (GCS) function is based on Teasdale’s Assessment of Coma and Impaired Consciousness-A

Practical Scale (Lancet, 1974) . Three aspects of behavior are independently measured: eye opening, verbal response, and motor response. The scores are added together to indicate that patient’s level of consciousness.

GCS is intended for adults and pediatric patients.

CAUTION

GCS is intended as an adjunct in patient assessment and must be used in conjunction with observation of clinical signs and symptoms.

GCS is not applied to patients that are sedated, muscularly relaxed, with artificial airway, drunk, hypothermia or in status epilepsies.

Please refer to hospital policy regarding acceptable use of GCS.

When applied to children younger than five years old or to the elderly, the GCS score may not reflect actual function.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 26 - 5

26.3.1

Displaying the GCS Parameter Area

To display the GCS parameter area, follow this procedure:

1.

Access Tile Layout in either of the following ways:

◆ Select the Screen Setup quick key

select the Tile Layout tab.

Select the Main Menu quick key

→ from the Display column select Tile Layout .

2.

Select the parameter area where you want to display the GCS score, and then from the popup list select

GCS .

The following figure shows the GCS parameter area. The display may be configured to look slightly different.

(4)

(1)

(5)

(2)

(3)

(1)

(2)

(3)

(4)

(5)

(6)

GCS label

Total score and level of consciousness. The color of the circle indicates the level of risk.

Scoring time

Scoring countdown: time to the next scoring

Scoring interval

Subscores

E: eye opening

◆ V: verbal response

M: motor response

(6)

26.3.2

Accessing the GCS Dialog

Enter the GCS dialog in any of the following ways:

Select the GCS parameter area.

Select the GCS quick key.

Select the Main Menu quick key

→ from the CAA column select GCS .

26 - 6 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

(1)

(2)

(1) Subscore (2) Total score

26.3.3

Performing GCS Scoring

To perform scoring, follow this procedure:

1.

From the Eye Opening area, Verbal Response area, and Motor Response area, respectively select an item that represents the patient’s status.

2.

Select Ok to accept the total score.

The following table lists the default score range and color of relevant consciousness level.

Level

Mild damage

Moderate damage

Severe damage

Range

13-15

9 - 12

3 - 8

Color

White

Yellow

Red

NOTE

The GCS score range is adjustable. Refer to

32.4.4 Setting GCS Threshold for Each Consciousness Level

for the method of changing the high limit and low limit of each consciousness level.

26.3.4

Setting GCS Scoring Interval

From the GCS dialog, select Interval to set GCS scoring interval. When the scoring interval is reached and you do not perform another scoring, the score will be invalidated and displayed as outline fonts.

26.3.5

Reviewing GCS Trend Data

From the GCS dialog, select Review to enter the Review dialog and view the GCS trend data from the Tabular

Trends .

26.4

SepsisSight

TM

The SepsisSight

TM

function is based on Third International Consensus Definitions for Sepsis and Septic Shock

(Sepsis-3) and Surviving Sepsis Campaign: International Guidelines for Management of Sepsis and Septic Shock:

2012 and 2016 (SSC Guideline 2016). It assists you in recognizing the early signs and symptoms of sepsis by comparing the state of the patient to the defined criteria and then guides you through the recommended treatment protocol.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 26 - 7

CAUTION

SepsisSight is not a diagnostic or therapeutic tool and is not intended to replace the competent judgment of a clinician.

SepsisSight should not be used with intravenous immunoglobulins in adult patients with severe sepsis or septic shock.

SepsisSight should not be used with intravenous selenium for the treatment of severe sepsis.

SepsisSight should not be used with sodium bicarbonate therapy for the purpose of improving hemodynamics or reducing vasopressor requirements in patients with hypoperfusion-induced lactic acidemia with pH ≥7.15.

NOTE

SepsisSight is intended to perform routine screening of potentially infected seriously ill patients for severe sepsis to allow earlier implementation of therapy.

SepsisSight is intended to be followed by application of hospital–based performance improvement efforts in severe sepsis.

The Surviving Sepsis Campaign recommends to discuss goals of care and prognosis with patients and families.

The Surviving Sepsis Campaign recommends to incorporate goals of care into treatment and end-oflife care planning, utilizing palliative care principles where appropriate.

The SepsisSight goals of care are intended to be applied in a time defined by the SSC guidelines.

Due to space limitations on the screen, the recommendations cannot always be reproduced in the full detail available in the printed and on-line guidelines.

SepsisSight is intended for adult patients suffering from sepsis or suspicion of sepsis.

26.4.1

Accessing the SepsisSight Dialog

Enter the SepsisSight dialog in any of the following ways:

Select the SepsisSight quick key.

Select the Main Menu quick key → from the CAA column select SepsisSight .

26.4.2

Screening

As per Sepsis-3, SepsisSight supports quick Sepsis-Related Organ Failure Assessment (qSOFA, or quick SOFA) and

Sepsis-Related Organ Failure Assessment (SOFA). SOFA is the default assessment tool for ICU, while qSOFA is the default for other departments.

26.4.2.1

Performing qSOFA qSOFA evaluates the patient’s respiration rate, systolic blood pressure and altered mental status.

RR and BP-S being monitored are automatically obtained. You can also manually enter these values by selecting the symbol. Select whether the patient’s mental status is altered. Then qSOFA score is calculated.

Select Confirm to record the calculation time.

If the qSOFA score is greater than or equal to 2, or sepsis is suspected, select SOFA>> to perform SOFA.

To clear the score, select Reset .

NOTE

The keyboard symbol indicates that the parameter value is manually entered.

The question mark (?) in the score circle indicates that more parameter values are required before a score can be calculated.

26 - 8 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

26.4.2.2

Performing SOFA

SOFA score is used to identify organ function or the rate of sepsis-related organ failure.

To perform SOFA, enter the value or select a range for each item, SOFA score will be automatically calculated.

To clear the score, select Reset .

If Sepsis criteria is met, make a comprehensive judgement on the clinical features.

26.4.2.3

Changing Screening Settings

From the Screening page select Setup . You can change the following settings:

In the Screening area, set RR (rpm) high limit and BP-S (mmHg) low limit for qSOFA scoring.

In the Unit area, set the unit of Bilirubin and Creatinine .

26.4.3

Recommendations

The SSC Bundles page lists goals and treatments to be completed in the defined time. Pages Treatment I and

Treatment II list graded recommendations as per the SSC Guidelines 2016,

You can define the time and goals for initial resuscitation, as well as treatments to be completed in one hour, 3

hours, and 6 hours. For more information, see 32.4.5 Changing SepsisSight Settings

.

26.4.3.1

Viewing Detailed Recommendations

On the Pages Treatment I and Treatment II select the arrow symbol at the right side of each item to view detailed recommendations of SSC Guideline 2016. The star symbol

indicates the grade of recommendation:

★★

: strong recommendation

: weak recommendation

■ No star symbol: best practice statement

To hide the detailed recommendations, select the arrow symbol .

26.4.3.2

Marking Implemented Items

Check off implemented items. Then the time and date are automatically recorded and displayed.

You can select the symbol to change the date and time.

Select Reset to clear the current results.

26.4.4

Reviewing SepsisSight Trend Data

Select the Graphic Trends tab to view the trend of parameters of resuscitation.

When a recommended treatment is checked off on pages Treatment I and Treatment II , relevant event is marked in the tabular trend. Vertical lines of different colors indicate the event type:

White: inspection performed

Blue: medication

Green: goal achieved

Purple: other treatment

26.5

ECG 24h Summary

The ECG 24h Summary provides ECG statistics of the current patient over the latest 24 hours. You can view the following information through the ECG 24h Summary:

Heart rate statistics

Arrhythmia event statistics

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 26 - 9

QT/QTc measurement statistics

Maximum and minimum ST statistics of each lead

Pacer statistics

Typical ECG strips

NOTE

The ECG 24h Summary is intended for the current patient. It is not intended for discharged patients.

Pacer statistics is intended for paced patients.

Patient data is saved, collected and displayed together in the ECG 24h Summary. Data displayed in the ECG 24h Summary is not recalculated.

If BeneView T1 is used to monitor a patient, when BeneView T1 is connected with the host monitor, data sent to the host monitor is not included in the ECG 24h Summary.

A license is required for the ECG 24h Summary function.

26.5.1

Viewing the ECG 24h Summary

To view the ECG 24h Summary, choose either of the following ways:

Select the ECG 24h Sum quick key.

Select the Main Menu quick key

from the CAA column select ECG 24h Summary

.

26.5.2

Selecting Typical ECG Strips

The Typical Strips area displays ECG strips of the following situations:

Maximum heart rate

Minimum heart rate

Four arrhythmia events

You can select a typical ECG strip of each situation. For example, to select the typical ECG strip of asystole, follow this procedure:

1.

Select the currently displayed asystole strip.

2.

From the popup strips, select the desired strip as the typical strip of asystole.

26.5.3

Reviewing ECG Summary

From the ECG 24h Summary dialog, you can review corresponding trends and events.

Select the Heart Rate area to review HR graphic trends.

Select the Max ST/Min ST area to review the current ST reference and ST graphic trends.

Select the Pace area to review historic pace events.

Select the QT area to review QT/QTc graphic trends.

Select the Arrhythmia area to review arrhythmia statistics. For more information, refer to

25.2.9 Arrhy

Statistics Review Page .

Select Full Disclosure to review ECG full disclosure waveforms. For more information, refer to

25.2.8 Full

Disclosure Review Page

.

26.6

HemoSight

TM

This monitor provides the HemoSight TM function. You can easily view and review hemodynamic parameters through the HemoSight dialog.

The HemoSight dialog is available when you are using the following modules:

C.O. module

ScvO

2

module

26 - 10 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

■ CCO/SvO

2

module (measured from the Vigilance II, Vigileo, HemoSphere, and EV1000 monitor)

For detailed information on monitoring and setting hemodynamic-related parameters, see

16 Monitoring

Cardiac Output (C.O.),

17 Monitoring Central Venous Oxygen Saturation (ScvO

2)

, and 18 Monitoring CCO/

SvO

2

. This chapter only introduces functions operated through the

HemoSight dialog.

26.6.1

Accessing the HemoSight Dialog

Access the HemoSight dialog by either of the following ways:

Select the HemoSight quick key.

Select the CCO numeric area, ScvO2 numeric area, or SvO2 numeric area → from the bottom of the dialog select HemoSight .

Select the Main Menu quick key

→ from the CAA column select HemoSight .

26.6.2

Viewing Hemodynamic Parameters

To view hemodynamic parameters, select the desired tab on the Diagnosis page of the HemoSight dialog:

◆ Select All to view all hemodynamic parameters. For more information, see 32.2.1 Hemodynamic

Parameters .

◆ Select Physiology Graphics to view dynamic graphics for parameters’ changes. For more

information, see 26.6.3 Physiology Graphics

.

◆ Select Physiology Relationship to view the realtime relationship of parameters. For more

information, see 26.6.4 Physiology Relationship .

The symbols beside the hemodynamic parameters have the following meanings:

*: indicates an intermittent parameter.

**: indicates an oxygenation parameter.

*Measure Time: refers to the measurement time of the intermittent parameter.

↑ or ↓ : indicates that a parameter value exceeds its upper or lower limit. For setting the upper and lower limits of the hemodynamic parameters, see section

26.6.8 Setting Hemodynamic Parameter Ranges

.

26.6.2.1

Hemodynamic Parameters from ScvO

2

module

The following table lists the hemodynamic parameters from the ScvO

2

module.

Oxygenation

Parameters

Abbreviation

ScvO

2

Hb

Hct

DO

2

DO

2

I

VO

2

VO

2

I

SaO

2

Full Spelling

Central Venous Oxygen Saturation

Hemoglobin

Haematocrit

Oxygen Delivery

Oxygen Delivery Index

Oxygen Consumption

Oxygen Consumption Index

Arterial Oxygen Saturation

Unit

% g/L, g/dl, mmol/L

% ml/min ml/min/m

2 ml/min ml/min/m

2

%

26.6.2.2

Hemodynamic Parameters from CCO/SvO

2

module

The following table lists the hemodynamic parameters from the Vigilance II, Vigileo, HemoSphere, and EV1000 monitors.

Output

Abbreviation

CCO

Full Spelling

Continuous Cardiac Output

Unit

L/min

Applicable Device

All

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 26 - 11

26 - 12

Contractility

Preload

Volume

Afterload

SVR

SVRI eArt-S eArt-M eArt-D

Oxygenation

Parameters

SaO

2

ScvO

2

EDV

EDVI

SVV

GEDV

GEDI

ITBV

ITBI

PPV

MAP

PR

ESV

ESVI

RVEF

Abbreviation

CCI

C.O.

CI

SV

SVI

HR

Full Spelling

Continuous Cardiac Index

Cardiac Output

Cardiac Index

Stroke Volume

Stroke Volume Index

Heart Rate

GEF

CFI

Pulse Rate

End Systolic Volume

End Systolic Volume Index

Right Ventricular Ejection

Fraction

Global Ejection Fraction

Cardiac Function Index

Unit

L/min/m

2

L/min

L/min/m

2 ml ml/m

2 bpm bpm ml ml/m

2

%

% l/min cmH

2

O, kPa , or mmHg

End Diastolic Volume

End Diastolic Volume Index

Stroke Volume Variation ml ml/m

%

2

Vigilance II,

HemoSphere

Vigilance II,

HemoSphere

Vigileo, EV1000,

HemoSphere

EV1000

EV1000

Global End Diastolic Volume

Global End Diastolic Volume

Index

Intrathoracic Blood Volume

Intrathoracic Blood Volume

Index

Pulse Pressure Variation

Mean Arterial Pressure ml ml/m

2 ml ml/m

2

% mmHg or kPa

EV1000

EV1000

HemoSphere

Vigilance II

Systemic Vascular Resistance

Systemic Vascular Resistance

Index

Systolic Arterial Pressure from the EV1000 monitor

Mean Arterial Pressure from the EV1000 monitor

Diastolic Arterial Pressure from the EV1000 monitor

Arterial Oxygen Saturation

DS/cm

5

or kPa-s/l

DS•m

2

/cm

5

or kPa-s-m

2

/l mmHg or kPa mmHg or kPa mmHg or kPa

%

All

All

EV1000,

HemoSphere

EV1000,

HemoSphere

EV1000,

HemoSphere

Vigilance II

Central Venous Oxygen

Saturation

% All

Applicable Device

All

Vigilance II

Vigilance II

All

All

Vigilance II,

HemoSphere

EV1000,

HemoSphere

Vigilance II

Vigilance II

Vigilance II,

HemoSphere

EV1000

EV1000

All

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Organ

Function

VO2I

VO2E

VO2IE

O

2

EI

Hb

Abbreviation

SvO

2

DO

2

Full Spelling

Mixed Venous Oxygen

Saturation

Oxygen Delivery

DO

2

I Oxygen Delivery Index

VO

2

Unit

% ml/min ml/min/m

2

Oxygen Consumption

Oxygen Consumption Index ml/min ml/min/m

2

Estimated Oxygen

Consumption

Estimated Oxygen

Consumption Index

Oxygen Extraction Index

Hemoglobin ml/min ml/min/m

%

2 g/L, g/dl, or mmol/L

SpO2

EVLW

Arterial Oxygen Saturation from Pulse Oximetry

Extravascular Lung Water

% ml

ELWI

PVPI

TB

Extravascular Lung Water

Index

Pulmonary Vascular

Permeability Index

Blood Temperature ml/kg no unit

ºC or °F

Applicable Device

All

Vigilance II, EV1000,

HemoSphere

EV1000,

HemoSphere

Vigilance II, EV1000,

HemoSphere

EV1000,

HemoSphere

EV1000, EV1000,

HemoSphere

EV1000,

HemoSphere

Vigilance II

EV1000,

HemoSphere

EV1000,

HemoSphere

EV1000

EV1000

EV1000

EV1000,

HemoSphere

26.6.3

Physiology Graphics

The lungs, heart and circulatory system vary according to the patient’s condition at the time of the thermodilution measurement. The Physiology Graphics displays monitored parameters using animation, which gives a visual representation of the interaction between the heart, lungs, blood, and vascular system. The continuous parameter values are displayed in realtime. When the intermittent data is available, the animation changes to reflect this change. The heart rate showed by the animation is also a visual reflection of the patient’s heart rate.

(1) When ScvO

2

ScvO

2

or SvO

or SvO

2 is available, the color varies according to the value, so a change in color indicates a change of

2

value.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 26 - 13

(2)

(3)

When CVP and SVRI are available, the change in blood vessel diameter indicates the change of SVRI values. For more information, see

26.6.3.1 Systemic Vascular Resistance Status .

When SpO

2

SpO

2

value.

is available, the color varies according to the SpO

2

value, so a change in color indicates a change of

(4)

(5)

(6)

When EDVI or GEDI is available, the size change in the diastole indicates the change of end diastolic volume index.

For more information, see

26.6.3.3 End Diastolic Volume Status .

When ELWI is available, the change in liquid level of lung indicates the change of ELWI values. For more information,

see 26.6.3.2 Lung Water Status

.

When RVEF or GEF is available, the size change in the systole indicates the change of ejection fraction. For more information, see

26.6.3.4 Ejection Fraction Status .

26.6.3.1

Systemic Vascular Resistance Status

Systemic vascular resistance status is indicated by SVRI. The following pictures show the systemic vascular resistance status.

Normal resistance High resistance Low resistance

26.6.3.2

Lung Water Status

Lung water status is indicated by ELWI. The following picture shows the lung water status, which is indicated by the liquid level of lung. The dotted line marks the high limit of ELWI. The ELWI is considered too high if the liquid level exceeds the dotted line. When ELWI is not available, the lung is displayed in gray color.

26.6.3.3

End Diastolic Volume Status

End diastolic volume status is indicated by EDVI (from the CCO/SvO

2

module). The following pictures show different end diastolic volume statuses. The dotted line marks the normal end diastolic volume status. When EDVI or GEDI is not available, the heart in the diastole is displayed in gray color.

26 - 14 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

High end diastolic volume Normal end diastolic volume Low end diastolic volume

26.6.3.4

Ejection Fraction Status

Ejection fraction status is indicated by RVEF (from the CCO/SvO

2

module). The following pictures show different ejection fraction statuses. The dotted line marks the normal ejection fraction status. When RVEF or GEF is not available, the heart in the systole is displayed in gray color.

Low ejection fraction Normal ejection fraction High ejection fraction

26.6.4

Physiology Relationship

The Physiology Relationship screen depicts the balance between oxygen delivery (DO

2

) and estimated oxygen consumption (VO

2

). It automatically updates as parameter values change so the displayed values are always current. The connecting lines highlight the relationship of the parameters to each other.

In the Physiology Relationship screen, the connecting lines have the following meanings:

Solid line: indicates that connected parameters have a direct relationship.

Dotted line: indicates that connected parameters have an indirect relationship.

Red frame: the parameter value is out of its normal range.

Green frame: the parameter value is within its normal range.

Gray frame: the parameter value is not available.

In the Physiology Relationship screen, each parameter frame displays the following contents:

Parameter name

Parameter value (”--” if parameter value is not available)

* (if the parameter is intermittent)

** (if the parameter value is input manually)

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 26 - 15

↑ or ↓ (if the parameter value exceeds its upper or lower limit, as defined in the Setup dialog. See section

26.6.8 Setting Hemodynamic Parameter Ranges )

NOTE

When measuring the hemodynamic parameters with different modules or devices, parameters displayed in the Physiology Relationship screen are different.

26.6.5

Hemodynamic Test

The Test page in the HemoSight window provides the trends before and after a hemodynamic test. To access the Test page, follow this procedure:

1.

Select the numeric area of CCO , ScvO2, or SvO2 .

2.

Select HemoSight .

3.

Select the Test tab.

26.6.5.1

Renaming a Test

The passive leg raising (PLR) test and rapid fluid loading (RFL) test are default tests, and their names cannot be changed. To rename a custom defined test, follow this procedure:

1.

Access the HemoSight dialog. For more information, refer to

26.6.1 Accessing the HemoSight Dialog

.

2.

Select the Test tab.

3.

Select Custom1 , Custom 2 , Custom 3 or Custom 4 .

4.

Select the symbol at the upper left corner of the current dialog.

5.

Input the test name in the input field.

26.6.5.2

Setting Test Time Duration

To set test time duration, follow this procedure:

1.

Access the HemoSight dialog. For more information, refer to

26.6.1 Accessing the HemoSight Dialog

.

2.

Select the Test tab.

3.

Set Test Duration .

26.6.5.3

Selecting Parameters for Testing

To set test parameters, follow this procedure:

1.

Access the HemoSight dialog. For more information, refer to

26.6.1 Accessing the HemoSight Dialog

.

2.

Select the Test tab.

3.

Select Setup .

4.

From desired parameter page, select parameters for testing.

26.6.5.4

Performing a Hemodynamic Test

To measure and record a hemodynamic test, follow this procedure, along with hospital guidelines for the test to be administered:

1.

Access the HemoSight dialog. For more information, refer to

26.6.1 Accessing the HemoSight Dialog

.

2.

Select the Test tab.

3.

Select Setup to select the involved parameters.

4.

Select Test Duration to set test duration.

5.

Prepare the patient and select the Start button to begin recording the measurements.

6.

Wait until the test finishes automatically, or select the Stop button to end the test.

26 - 16 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

7.

View the parameter trends. After the test, the Test page displays the current parameter value, reference parameter value, and Delta value (the variation of current value and reference value). For more information,

see 25 Review .

NOTE

Hemodynamic tests can also run in background. If the Test page is closed during a test or the

HemoSight dialog is closed, then the monitor gives a prompt tone and pops up the Test page after the test ends.

26.6.6

Following Up the Patient Hemodynamic Status

You can view the cardiac function curve, lung water curve and SVV/PPV slope indicator in the Follow-up page.

To view these curves or indicator, follow this procedure:

1.

Access the HemoSight

dialog. For more information, refer to 26.6.1 Accessing the HemoSight Dialog

.

2.

Select the Follow-up tab.

3.

View the following curves or indicator:

◆ Cardiac function curve

◆ SVV/PPV slope indicator

◆ Lung water curve

26.6.6.1

Cardiac Function Curve and Lung Water Curve

Cardiac function curve and the lung water curve are used to assess the fluid responsiveness.

(1)

(2)

(1) Cardiac function curve (2) Lung water curve

For more information on the setup of X-axis or Y-axis parameter, see 26.6.6.2 Setting the Cardiac Function Curve and Lung Water Curve

.

26.6.6.2

Setting the Cardiac Function Curve and Lung Water Curve

To set the cardiac function curve and lung water curve, follow this procedure:

1.

Access the HemoSight

dialog. For more information, refer to 26.6.1 Accessing the HemoSight Dialog

.

2.

Select the Follow-up tab.

3.

Make the following settings:

◆ Interval : set the time interval between coordinate points on the X-axis and Y-axis of the cardiac function curve and lung water curve. This setting is effective only when continuous parameters are selected at both the X-axis and Y-axis. When ELWI is unavailable and all the other parameters selected are continuous parameters, coordinate points will not be drawn on the lung water curve, but will be drawn on the cardiac function curve at the configured interval.

◆ Output : set the Y-axis parameter for cardiac function curve.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 26 - 17

◆ Preload Volume : set the X-axis parameter for both cardiac function curve and lung water curve.

NOTE

If intermittent parameters are selected at either the X-axis or Y-axis, coordinate points will be drawn for all the intermittent parameters. In this case, coordinates drawn for continuous parameters and intermittent parameters are consistent.

26.6.6.3

SVV/PPV Slope Indicator

The SVV/PPV slope indicator is a visual representation of the cardiac function curve used when assessing the stroke volume variation value or pulse pressure variation. The curved line indicates the SVV/PPV slope.

The dot moves up and down the curved line according to the SVV or PPV value. The color of the dot changes based upon set target ranges.

SVV/PPV≤10%: the dot is green, and may predict that the patient is fluid unresponsive.

SVV/PPV>10%: the dot is red, and may predict that the patient is fluid responsive.

CAUTION

PPV and SVV are only applicable in patients on controlled mechanical ventilation and sinus rhythm.

26.6.6.4

Setting the SVV/PPV Slope Indicator

To set the SVV or PPV slope, follow this procedure:

1.

Access the HemoSight dialog. For more information, refer to

26.6.1 Accessing the HemoSight Dialog

.

2.

Select the Follow-up tab.

3.

Select SVV or PPV to set which Curve is the indicator. You can also set it to Off to hide the indicator.

26.6.7

Evaluating the Hemodynamic Parameters

You can review the hemodynamic parameter trends on the Evaluation page. The changes of hemodynamic parameters are shown by the trends or Spidervision diagram. The trends and Spidervision diagram display both current values and reference values of parameters, which let you conveniently see the hemodynamic status at different timepoints.

To enter the Evaluation page, follow this procedure:

1.

Access the HemoSight dialog. For more information, refer to

26.6.1 Accessing the HemoSight Dialog

.

2.

Select the Evaluation tab.

26.6.7.1

Viewing the Spidervision Diagram

The Spidervision diagram shows hemodynamic parameters in dynamic conjunction. To enter the Spidervision diagram, follow this procedure:

1.

Access the HemoSight dialog. For more information, refer to

26.6.1 Accessing the HemoSight Dialog

.

2.

Select the Evaluation tab.

26 - 18 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

(1)

(2)

(3) (4) (5) (6)

(1) Current value

(4) Middle segment

(2) Reference value

(5) Inner segment

(3) Outer segment

(6) Link between legs of the graoh

Each leg of the graph is divided into 3 segments indicating different value ranges for the respective parameters.

The middle segment indicates the normal range for the respective parameter. The outer segment will be highlighted when corresponding parameter value exceeds the upper limit. The inner segment will be highlighted when its corresponding parameter value exceeds the lower limit.

The links between legs of the graph are displayed in green when all displayed parameters are within the normal range.

The links between legs of the graph and the corresponding spider segment are displayed in yellow when one of the displayed parameters is outside the normal range.

The links between legs of the graph and the corresponding spider segments are displayed in red when two or more displayed parameters are outside the normal range.

26.6.7.2

Viewing Trends

To view the trends of hemodynamic parameters, follow this procedure:

1.

Access the HemoSight

dialog. For more information, refer to 26.6.1 Accessing the HemoSight Dialog

.

2.

Select the Evaluation tab.

3.

Select the Trends tab.

For more information on trends review, see 25.2.2 Example Review Dialog

.

26.6.7.3

Viewing Hemodynamic Parameters

You can view the hemodynamic parameters on the Evaluation page. The timeline below the Spidervision diagram indicates the entire time length. The timeline below the Trends page indicates the time length of

current window (for more information, see 26.6.7.4 Changing the Time Length of Trends ). Different color markers

on the timeline indicate different types of events:

Red: high priority physiological alarm event

Yellow: medium priority physiological alarm event

Cyan: low priority physiological alarm event

Green: manual event

White: operation-related event, such as accepting the C.O. average

To locate a hemodynamic event, choose any of the following methods:

Move the slider to the target position.

Use the following buttons:

◆ Select or beside to go to the previous or next hemodynamic event.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 26 - 19

◆ Select

◆ Select

to view the event list and select the desired hemodynamic event.

or beside to go to the previous or next C.O. average acceptance event.

◆ Select to enter the list of C.O. average acceptance events, and select the desired C.O. average acceptance event.

◆ Select or to move the white vertical line to the left or right. Alternately, the cursor will move to any location you touch on the graph.

NOTE

If a physiological alarm event or manual event occurs simultaneously with a C.O. average acceptance event, this event will be displayed in preference to a C.O. average acceptance event.

26.6.7.4

Changing the Time Length of Trends

To change the time length of trends, follow this procedure:

1.

Access the HemoSight dialog. For more information, refer to

26.6.1 Accessing the HemoSight Dialog

.

2.

Select the Evaluation tab.

3.

Select Trends Tab.

4.

Select Zoom and set the time length of trends.

26.6.7.5

Selecting the Spidervision Pattern

The pattern defines the hemodynamic parameters and their numbers displayed in the Spidervision diagram and Trends page. To select the pattern, follow this procedure:

1.

Access the HemoSight dialog. For more information, refer to

26.6.1 Accessing the HemoSight Dialog

.

2.

Select the Evaluation tab.

3.

Select the Spidervision tab or Trends tab.

4.

Click the Pattern button and select a Spidervision pattern.

For more information, see

26.6.10 Setting the Pattern

for detailed information of creating or updating a pattern.

26.6.7.6

Saving Reference Values

To save the hemodynamic parameter values of any moment as reference values, follow this procedure:

1.

Access the HemoSight dialog. For more information, refer to

26.6.1 Accessing the HemoSight Dialog

.

2.

Select the Evaluation tab.

3.

Select the Spidervision tab or Trends tab.

4.

Move the timeline below the trends or Spidervision graph to select a desired time, or refer to

26.6.7.3 Viewing Hemodynamic Parameters

to select an event.

5.

Select Set Reference to save the parameter values at the selection time as reference values.

26.6.8

Setting Hemodynamic Parameter Ranges

To set the ranges of hemodynamic parameters, follow this procedure:

1.

Access the HemoSight dialog. For more information, refer to

26.6.1 Accessing the HemoSight Dialog

.

2.

Select the Setup button.

3.

Select the Range tab.

4.

Set the normal ranges of the hemodynamic parameters.

26.6.9

Restoring Default HemoSight Range Values

To restore the default values, follow this procedure:

1.

Access the HemoSight dialog. For more information, refer to

26.6.1 Accessing the HemoSight Dialog

.

2.

Select the Setup button.

26 - 20 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

3.

Select the Defaults button.

NOTE

Selecting Defaults restores all the parameter ranges to default values.

26.6.10

Setting the Pattern

To set the pattern, follow this procedure:

1.

Access the HemoSight

dialog. For more information, refer to 26.6.1 Accessing the HemoSight Dialog

.

2.

Select the Setup button.

3.

Select the Spidervision tab.

4.

Make the following settings:

◆ Select the number of parameters (three to seven).

Select the parameters to be displayed in Spidervision. The same parameters will show in the Trends as they are listed in Spidervision clockwise, starting at the top.

◆ Select Save , Save As or Delete to save, create or delete a pattern.

NOTE

You cannot delete the default pattern or the pattern that is in use.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 26 - 21

This page intentionally left blank.

26 - 22 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

27

Calculation

27.1

Calculation Overview

The monitor provides calculation functions. The calculated values, which are not directly measured, are computed based on the values you provide. The calculation function is independent of other monitoring functions and can therefore be used for patients being monitored by other monitors. Any operation in a calculation dialog does not affect the patient monitored by the current monitor.

You can perform the following calculations:

Drug calculations

Hemodynamic calculations

Oxygenation calculations

Ventilation calculations

Renal calculations

27.2

Calculation Safety Information

WARNING

Decisions on the choice and dosage of drugs administered to patients must always be made by the physician in charge. The drug calculations are based on the values input; it does not check the plausibility of the calculation performed.

Check that the entered values are correct and the calculated values are appropriate. Mindray assumes no responsibility for any consequences caused by wrong entries and improper operations.

27.3

Drug Calculations

27.3.1

Performing Drug Calculations

To perform drug calculations, follow this procedure:

1.

Access the drug calculator by either of the following ways:

◆ Select the Calculations quick key → select the Drug tab.

◆ Select the Main Menu quick key → from the Calculations column select Drug .

2.

Set Drug Name and Patient Category.

If the dose of drug is weight dependent, you must input the patient’s weight. The dose calculation program has a library of commonly used drugs, while Drug A through Drug E are user defined.

3.

Enter the known values, for example Drug Amount and Solution Volume.

4.

Select Calculate . The calculated values are indicated by red arrows.

NOTE

If available, the patient category and weight from the Patient Demographics dialog are automatically entered when you first access drug calculation. You can change the patient category and weight. This will not change the patient category and weight stored in the patient demographic information.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 27 - 1

27.3.2

Checking the Titration Table

The titration table shows information on the currently used drugs. Use the titration table to see what dose of a drug your patient will receive at different infusion rates. To access the titration table, follow this procedure:

1.

Access drug calculator by either of the following ways:

◆ Select the Calculations quick key.

◆ Select the Main Menu quick key → from the Calculations column select Drug .

2.

Select the Titration Table tab.

3.

Select Dose Type to set the type of dose unit in the titration table.

4.

Select Interval to set the interval between two adjacent titration table items.

You can select how to display the titration table:

Dose : the titration table is listed in the sequence of increased drug dose.

Infusion Rate : the titration table is listed in the sequence of increased infusion rate. By default the resolution of the infusion rate is one (1). By selecting Exact Rate the resolution of the infusion rate can reach 0.01 so that you can display the infusion rate more accurately.

27.3.3

Drug Calculation Formula

Description

Dose

Dose (weight based)

Drug Amount

Drug Amount (weight based)

Duration

Duration (weight based)

Concentration

Solution volume

Infusion rate

Infusion rate (weight based)

Unit Formula

Dose/hr

Dose/min

Dose/kg/hr

Dose/kg/min g series: mcg, mg, g unit series: Unit, KU, MU mEq series: mEq g series: mcg, mg, g unit series: Unit, KU, MU mEq series: mEq hr hr mcg/ml, mg/ml, g/ml, Unit/ml,

KU/ml, MU/ml, mEq/ml ml ml/hr g•ml/hr

Dose = Infusion Rate × Concentration

Dose (weight based) = Infusion Rate ×

Concentration/Weight

Drug Amount =Dose × Duration

Drug Amount (weight based) = Dose × Duration ×

Weight

Duration = Amount/Dose

Duration (weight based) = Amount/(Dose × Weight)

Concentration = Drug Amount/Solution Volume

Volume = Infusion Rate × Duration

Infusion Rate = Dose/Concentration

Infusion Rate = Dose × Weight/Concentration

27.3.4

Titration Table Calculation Formula

Description

Infusion Rate

Infusion Rate (weight based)

Dose

Dose (weight based)

Unit ml/hr ml/hr

Dose/hr

Dose/min

Dose/kg/hr

Dose/kg/min

Formula

Infusion Rate = Dose/Concentration

Infusion Rate = Weight × Dose/Concentration

Dose = Infusion Rate × Concentration

Dose (weight based) = Infusion Rate ×

Concentration/Weight

27 - 2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

27.4

Hemodynamic Calculations

The monitor provides the hemodynamic calculation function and can save the results of up to 10 calculations, which are displayed in groups.

27.4.1

Performing Hemodynamic Calculations

To perform hemodynamic calculation, follow this procedure:

1.

Access hemodynamic calculation by either of the following ways:

◆ Select the Calculations quick key → Hemodynamics tab.

◆ Select the Main Menu quick key → from the Calculations column select Hemodynamics .

2.

Enter the known values. For a patient who is being monitored, the currently measured values are automatically entered.

3.

Select Calculate .

A calculated value greater than the normal upper limit is indicated by an up arrow “ ↑ ”. A calculated value lower than the normal lower limit is indicated by a down arrow “

”.

You can select Range to show the normal range of each parameter.

27.4.2

Input Parameters for Hemodynamic Calculations

Input Parameter cardiac output heart rate pulmonary artery wedge pressure mean arterial pressure pulmonary artery mean pressure central venous pressure end-diastolic volume height weight

Label

C.O.

HR

PAWP

MAP

PA-M

CVP

EDV

Height

Weight

Unit

L/min bpm mmHg mmHg mmHg mmHg ml cm kg

NOTE

If you enable Use PA-D as PAWP, PA-D value will be used to replace PAWP value for hemodynamic

calculation. For more information, refer to 15.6.8 Setting the Use PA-D as PAWP Switch

.

27.4.3

Calculated Parameters and Formulas for Hemodynamic Calculations

Calculated Parameters cardiac index body surface area

Label

C.I.

BSA stroke volume stroke index systemic vascular resistance

SV

SVI

SVR

Unit

L/min/m

2 m

2 ml ml/m

2

DS/cm

5

Formula

C.I. (L/min/m

2

) = C.O. (L/min)/BSA (m

2

)

BSA (m

2

) = Wt

0.425

(kg) × Ht

0.725

(cm) ×

0.007184

SV (ml) = 1000× C.O. (L/min)/HR (bpm)

SVI (ml/m

2

) = SV (ml)/BSA (m

2

)

SVR (DS/cm

5

) = 79.96 × [APMAP (mmHg) -

CVP (mmHg)]/C.O. (L/min)

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 27 - 3

Calculated Parameters systemic vascular resistance index pulmonary vascular resistance pulmonary vascular resistance index left cardiac work

Label

SVRI

PVR

PVRI

LCW left cardiac work index left ventricular stroke work left ventricular stroke work index right cardiac work

LCWI

LVSW

LVSWI

RCW right cardiac work index right ventricular stroke work right ventricular stroke work index

RCWI

RVSW

RVSWI ejection fraction end-diastolic volume index

EF

EDVI end-systolic volume ESV end-systolic volume index ESVI

Unit

DS•m

2

/cm

5

DS/cm

5

DS•m

2

/cm

5 kg•m kg•m/m

2 g•m g•m/m

2 kg•m kg•m/m

2 g•m g•m/m

2

% ml/m2 ml ml/m

2

Formula

SVRI (DS•m

2

/cm

5

) = SVR (DS/cm

5

) × BSA (m

2

)

PVR (DS/cm

5

) = 79.96 × [PAMAP (mmHg) -

PAWP (mmHg)]/C.O. (L/min)

PVRI (DS•m

2

/cm

5

) = PVR (DS/cm

5

) × BSA (m

2

)

LCW (kg•m) = 0.0136 × APMAP (mmHg) × C.O.

(L/min)

LCWI (kg•m/m

2

) = LCW (kg•m)/BSA (m

2

)

LVSW (g•m) = 0.0136 × APMAP (mmHg) × SV

(ml)

LVSWI (g•m/m

2

) = LVSW (g.m)/BSA (m

2

)

R CW (kg•m) = 0.0136 × PAMAP (mmHg) × C.O.

(L/min)

R CWI (kg•m/m

2

) = RCW (kg.m)/BSA (m

2

)

R VSW (g•m) = 0.0136 × PAMAP (mmHg) × SV

(ml)

R VSWI (g•m/m

2

) = RVSW (g•m)/BSA (m

2

)

EF (%) = 100 × SV (ml)/EDV (ml)

EDVI (ml/m

2

) = EDV (ml)/BSA (m

2

)

ESV (ml) = EDV (ml) –SV (ml)

ESVI (ml/m

2

) = ESV (ml)/BSA (m

2

)

27.5

Oxygenation Calculations

The monitor provides the oxygenation calculation function and can save the results of up to 10 calculations, which are displayed in groups.

27.5.1

Performing Oxygenation Calculations

To perform oxygenation calculations, follow this procedure:

1.

Access oxygenation calculation by either of the following ways:

◆ Select the Calculations quick key

Oxygenation tab.

◆ Select the Main Menu quick key

from the Calculations column select Oxygenation .

2.

Enter the known values. For a patient who is being monitored, the currently measured values are automatically entered.

3.

Select Calculate .

A calculated value greater than the normal upper limit is indicated by an up arrow “

”. A calculated value lower than the normal lower limit is indicated by a down arrow “ ↓ ”.

In the Oxygenation page, you can also perform the following operations:

Select OxyCont Unit , Hb Unit , and Pressure Unit . Then corresponding parameter values will be automatically converted and updated accordingly.

Select Range to show the normal range of each parameter.

27 - 4 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

27.5.2

Input Parameters for Oxygenation Calculations

Input Parameter cardiac output percentage fraction of inspired oxygen partial pressure of oxygen in the arteries partial pressure of carbon dioxide in the arteries arterial oxygen saturation partial pressure of oxygen in venous blood venous oxygen saturation hemoglobin respiratory quotient atmospheric pressure height weight

Label

C.O.

FiO

2

PaO

2

PaCO

2

SaO

2

PvO

2

SvO

2

Hb

RQ

ATMP

Height

Weight

27.5.3

Calculated Parameters and Formulas for Oxygenation Calculations

Unit

L/min

% mmHg, kPa mmHg, kPa

% mmHg, kPa

% g/L, g/dl, mmol/L

None mmHg, kPa cm, inch kg, lb

Calculated Parameters body surface area

Label

BSA alveolar-arterial oxygen difference capillary oxygen content

AaDO

2

CcO

2 venous admixture QS/QT

Unit m

2 oxygen consumption arterial oxygen content venous oxygen content

VO

2

ml/min

CaO

2 ml/L, ml/dL

CvO

2 ml/L, ml/dL arteriovenous oxygen content difference oxygen extraction ratio oxygen transport partial pressure of oxygen in the alveoli

C(a-v)O

2

O

2

ER

DO

2

PAO

2 ml/L, ml/dl

% ml/min mmHg, kPa mmHg, kPa ml/L, ml/dl

% oxygen transport index oxygen consumption

DO

2

I

VO

2

I ml/min/m

2 ml/min/m

2

Formula

BSA (m

2

) = Wt

0.425

(kg) × Ht

0.725

(cm) ×

0.007184

VO

2

(ml/min) = C(a-v)O

2

(ml/L)× C.O. (L/min))

CaO

2

(ml/L) = 10× (0.0134 × Hb (g/dl) × SaO

2

(%)) +0.031× PaO

2

(mmHg)

CvO

2

(ml/L) = 10× (0.0134 × Hb (g/dl) × SvO

2

(%)) +0.031 × PvO

2

(mmHg)

C(a-v)O

2

(ml/L) = CaO

2

(ml/L) - CvO

2

(ml/L)

O

2

ER (%) = 100×C(a-v)O

2

(ml/L)/CaO

2

(ml/L)

DO

2

(ml/min) = C.O. (L/min) × CaO

2

(ml/L)

PAO

2

(mmHg) = [ATMP (mmHg) - 47 mmHg] ×

FiO

2

(%)/100 - PaCO

2

(mmHg) × [FiO

2

(%)/100

+ (1 - FiO

2

(%)/100)/RQ]

AaDO

2

(mmHg) = PAO

2

(mmHg) - PaO

2

(mmHg)

CcO

2

(ml/L) = Hb (g/L) × 1.34 + 0.031 × PAO

2

(mmHg)

QS/QT (%) = 100× [1.34 × Hb (g/L) × (1 - SaO2

(%)/100) + 0.031 × (PAO2 (mmHg) - PaO2

(mmHg))]/[1.34 × Hb (g/L) × (1 - SvO2 (%)/

100) + 0.031× (PAO2 (mmHg) - PvO2

(mmHg))]

DO2I (ml/min/m

2

) = CaO2 (ml/L) × (C.O. (L/ min)/BSA (m

2

))

VO2I (ml/min/m

2

) = C (a-v) O2 (ml/L) ×(C.O. (L/ min)/BSA (m

2

))

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 27 - 5

27.6

Ventilation Calculations

The monitor provides the ventilation calculation function and can save the results of up to 10 calculations, which are displayed in groups.

27.6.1

Performing Ventilation Calculations

To perform ventilation calculations, follow this procedure:

1.

Access ventilation calculation by either of the following ways:

◆ Select the Calculations quick key → Ventilation tab.

◆ Select the Main Menu quick key → from the Calculations column select Ventilation .

2.

Enter the known values. For a patient who is being monitored, the currently measured values are automatically taken. If the anesthesia machine or ventilator is connected, measured values for ventilation calculation are also automatically entered.

3.

Select Calculate .

A calculated value greater than the normal upper limit is indicated by an up arrow “ ↑ ”. A calculated value lower than the normal lower limit is indicated by a down arrow “

”.

On the Ventilation page, you can also perform the following operations:

Select Pressure Unit . Then corresponding parameter values will be automatically converted and updated accordingly.

Select Range to show the normal range of each parameter.

27.6.2

Input Parameters for Ventilation Calculations

Input Parameter percentage fraction of inspired oxygen respiration rate partial pressure of mixed expiratory CO2 partial pressure of carbon dioxide in the arteries partial pressure of oxygen in the arteries tidal volume respiratory quotient atmospheric pressure

Label

FiO

2

RR

PeCO

2

PaCO

2

PaO

2

TV

RQ

ATMP

27.6.3

Calculated Parameters and Formulas for Ventilation Calculations

Unit

% rpm mmHg, kPa mmHg, kPa mmHg, kPa ml

None mmHg, kPa

Calculated Parameters Label partial pressure of oxygen in the alveoli

PAO

2 alveolar-arterial oxygen difference oxygenation ratio

AaDO

2

Pa/FiO

2 arterial to alveolar oxygen ratio minute volume a/AO

2

MV

Unit mmHg, kPa mmHg, kPa mmHg, kPa

%

L/min

Formula

PAO

2

(mmHg) = [ATMP (mmHg) - 47 mmHg] ×

FiO

2

(%)/100 - PaCO

2

(mmHg) × [FiO

2

(%)/100

+ (1 - FiO

2

(%)/100)/RQ]

AaDO

2

(mmHg) = PAO

2

(mmHg) - PaO

2

(mmHg)

Pa/FiO

2

(mmHg) = 100 × PaO

2

(mmHg)/FiO

2

(%) a/AO

2

(%) = 100 × PaO

2

(mmHg)/PAO

2

(mmHg)

MV (L/min) = [TV (ml) × RR (rpm)]/1000

27 - 6 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Calculated Parameters Label volume of physiological dead space physiologic dead space in percent of tidal volume alveolar volume

Vd

Vd/Vt

VA

Unit ml

%

L/min

Formula

Vd (ml) = TV (ml) × [1 - PeCO

2

(mmHg)/PaCO

2

(mmHg)]

Vd/Vt (%) = 100 × Vd (ml)/TV (ml)

VA (L/min) =[TV (ml) - Vd (ml)] × RR (rpm)/

1000

27.7

Renal Calculations

The monitor provides the renal calculation function and can save the results of up to 10 calculations, which are displayed in groups.

27.7.1

Performing Renal Calculations

To perform renal calculations, follow this procedure:

1.

Access renal calculation by either of the following ways:

◆ Select the Calculations quick key

select the Renal tab.

◆ Select the Main Menu quick key

from the Calculations column select Renal .

2.

Enter the known values.

3.

Select Calculate .

A calculated value greater than the normal upper limit is indicated by an up arrow “

”. A calculated value lower than the normal lower limit is indicated by a down arrow “ ↓ ”.

You can select Range to show the normal range of each parameter.

27.7.2

Calculated Parameters and Formulas for Renal Calculations

Input Parameter urine potassium urinary sodium urine plasm osmolality urine osmolality serum sodium creatinine urine creatinine blood urea nitrogen height weight

Label

URK

URNa

Urine

Posm

Uosm

SerNa

Cr

UCr

BUN

Height

Weight

Unit mmol/L mmol/L ml/24 hrs mOsm/kgH

2

O mOsm/kgH

2

O mmol/L

μmol/L

μmol/L mmol/L cm kg

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 27 - 7

27.7.3

Calculated Parameters and Formulas for Renal Calculations

Calculated Parameters urine sodium excretion

Label

URNaEx urine potassium excretion URKEx sodium potassium ratio Na/K clearance of sodium fractional excretion of sodium osmolar clearance

CNa creatinine clearance rate Clcr

FENa

Cosm free water clearance CH2O urine to plasma osmolality ratio blood urea nitrogen creatinine ratio urine-serum creatinine ratio

U/P osm

BUN/Cr*

U/Cr

*: BUN/Cr is a ratio under the unit of mol.

Unit mmol/24 hrs mmol/24 hrs

% ml/24 hrs ml/min

% ml/min ml/hr

None

Mmol/L

Formula

URNaEx (mmol/24 hrs) = Urine (ml/24 hrs) ×

URNa (mmol/L)/1000

URKEx (mmol/24 hrs) = Urine (ml/24 hrs) ×

URK (mmol/L)/1000

Na/K (%) = 100 × URNa (mmol/L)/URK (mmol/

L)

CNa (ml/24 hrs) = URNa (mmol/L) × Urine (ml/

24 hrs)/SerNa (mmol/L)

Clcr (ml/min) = Ucr (μmol/L) × Urine (ml/24 hrs)/[Cr (μmol/L) × (BSA (m

2

)/1.73) × 1440]

FENa (%) = 100 × URNa (mmol/L) × Cr (μmol/

L)/[SerNa (mmol/L) × Ucr (μmol/L)]

Cosm (ml/min) = Uosm (mOsm/kgH

2

O) ×

Urine (ml/24 hrs)/(Posm (mOsm/kgH

2

O) ×

1440)

CH2O (ml/hr) = Urine (ml/24 hrs) × [1 - Uosm

(mOsm/kgH

2

O)/Posm (mOsm/kgH

2

O)]/24

U/P osm = Uosm (mOsm/kgH

2

O)/Posm

(mOsm/kgH

2

O)

BUN/Cr = 1000 × BUN (mmol/L)/Cr (μmol/L)

27 - 8 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

28

Recording

28.1

Recorder

The thermal recorder records patient information, measurement data, and up to three waveforms.

N17/N15/N12 can be configured with a built-in recorder. If the built-in recorder is not configured, the monitor can be configured to use an external recorder module.

N22 and N19 do not have a built-in recorder. To perform recording, insert the recorder module into the SMR.

(1)

(3)

(2)

(5)

(4)

(1)

(2)

(3)

(4)

(5)

Start/Stop key: press to start a recording or stop the current recording.

Module status indicator

On: when the recorder is powered and ready for use.

Off: when the monitor is switched off.

Flashes: if an error has occurred with the recorder.

Paper outlet

Recorder door

Latch: pull it away from the recorder to open the recorder door.

28.2

Starting Recordings

Recordings can be started manually or automatically.

28.2.1

Manually Starting Recordings

To manually start a recording, you can either:

Press the

Select

hardkey on the front of the recorder to start a Realtime Segment Waveform Recording.

on the current dialog to start a recording corresponding to that dialog or page.

28.2.2

Automatic Recordings

In the following conditions, you can set the recorder to automatically start recording:

At a preset interval. For more information, see

28.5 Setting up the Recorder .

When a parameter alarm is triggered. For more information, see 32.3.10.3 Enabling Automatic Recordings on an Alarm

.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 28 - 1

28.3

Stopping Recordings

Recordings can be stopped manually or automatically.

28.3.1

Stopping Recordings Manually

To manually stop a recording, press the

Record Tasks in the Record Setup menu.

hardkey while the recorder is creating a strip or select Clear All

28.3.2

Stopping Recordings Automatically

Recordings stop automatically in the following conditions:

The recording is completed.

The recorder runs out of paper.

The recorder has an alarm condition.

The recorder module is removed from the monitor.

The monitor is turned off.

28.4

Recording Related Flags

You can find the following flags on the recording reports:

For automatically stopped recordings, there are two columns of asterisks “*” at the end of the report.

For manually or abnormally stopped recordings, there is one column of asterisks “*” at the end of the report.

If the parameter data is from external devices connected to the monitor via the BeneLink module, the parameter label is prefixed with the plus sign “+”.

28.5

Setting up the Recorder

To configure the recorder for waveforms and printing intervals, follow this procedure:

1.

Select the Main Menu quick key

→ from the Report column select Record Setup .

2.

In the Record Setup dialog, select the desired waveform for Waveform 1 , Waveform 2 and Waveform 3 in turn. The recorder can record up to 3 waveforms at a time.

3.

Switch on or off IBP Overlap to enable or disable IBP recordings in the overlapping format.

◆ When the IBP Overlap is enabled: If two or more waveforms in the selected waveforms for recording are IBP waveforms, the IBP waveforms will be recorded in the overlapping format.

◆ When the IBP Overlap is disabled: IBP waveforms will be recorded without overlap.

4.

Select Recording Duration to set the duration of real-time recording.

5.

Select Interval to set the time interval for automatic recording.

6.

Select Recorder Paper Speed to set the speed for recording waveforms.

28.6

Clearing Recording Tasks

To clear recording tasks, follow this procedure:

1.

Select the Main Menu quick key → from the Report column select Record Setup .

2.

In the Record Setup dialog, select Clear All Record Tasks . This clears all queued recording tasks and stops the current recording.

28 - 2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

28.7

Loading Paper

To load paper, follow this procedure:

1.

Use the latch at the upper right of the recorder door to pull the door open.

2.

Insert a new roll into the compartment as shown below. Feed the paper through and pull some paper out from the top of the roller.

3.

Close the recorder door.

CAUTION

Use only specified thermal paper. Otherwise, it may cause damage to the recorder’s printhead, the recorder may be unable to print, or poor print quality may result.

Never pull the recorder paper with force when a recording is in process. Otherwise, it may cause damage to the recorder.

Do not leave the recorder door open unless you are reloading paper or cleaning a paper jam.

28.8

Removing Paper Jam

If the recorder works incorrectly or produces unusual sounds, check if there is a paper jam. If a paper jam is detected, follow this procedure to remove it:

1.

Open the recorder door.

2.

Take out the paper and tear off the draped part.

3.

Reload the paper and close the recorder door.

28.9

List of Recording Reports

The recorder can output the following reports:

Realtime reports

Realtime Segment Waveform Record

Realtime Continuous Waveform Record

Auto Realtime Record

ST Record

QT Record

12-Lead Interpretation Record

Record on Alarm Report

12-Lead Waveform Record

History reports

Event Detail

Full Disclosure Detail Record

12 Lead Review Record

ST Review Record

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 28 - 3

◆ Tabular Trend Report

◆ Graphic Trend Report

28 - 4 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

29

Printing

29.1

Supported Printer

The monitor can output patient reports via a connected network printer. The monitor supports the following printers:

HP LaserJet Pro M202dw

HP LaserJet Pro 400 M401n

HP LaserJet P4015n

HP LaserJet 600 M602

HP LaserJet Enterprise M605

HP LaserJet Enterprise M608n

NOTE

For more details about the printer, refer to the document accompanying the printer. With product upgrades, the monitor may support additional printers without prior notice. If you have any doubts about the compatibility of the printer used with this monitor, contact Mindray.

29.2

End Case Reports

29.2.1

Printing the End Case Report

To print the end case report, choose one of the following ways:

Select Print from the End Case Report menu.

Select Print End Case Report when you discharge a patient.

Select the End Case Report quick key.

29.2.2

Setting a Report as An End Case Report

The following reports can be set as end case reports:

Tabular Trends Report

Graphic Trends Report

Event Report

12-lead Interpretation

Alarm Limits Report

Realtime Report

ECG Report

To set a report as an end case report, follow this procedure:

1.

Select the Main Menu quick key

from the Report column select End Case Report .

2.

From the Select Reports page, select the desired report tab, for example ECG Report .

29.2.3

Configuring the End Case Reports

To configure the end case reports, follow this procedure:

1.

Select the Main Menu quick key → from the Report column select End Case Report .

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 29 - 1

2.

From the Report Setup page, set the following end case reports:

◆ Select the Tabular Trends Report , Graphic Trends Report , Realtime Report , and ECG Report tab, and set these end case report by referring to section

29.6 Configuring Reports .

◆ Select the Event Report tab, and select the event that needs to be printed.

◆ Select the 12-Lead Interpretation tab, and set the switch of Median Complex , Measurements ,

Interpretation , or Interpretation Summary . For other settings, refer to section

29.6.1 Configuring

ECG Reports .

29.2.4

Setting the End Case Report Period

To set the end case report print period, follow this procedure:

1.

Select the Main Menu quick key → from the Report column select End Case Report .

2.

From the Select Reports page, set the Period .

NOTE

End case report print period is calculated from the patient discharged time to the configured period.

Period setting is applicable to all of the end case reports.

29.3

Manually Starting a Printing Task

This section describes different methods for manually printing a report.

29.3.1

Printing from the Screen

The button will be shown at the top of dialogs that have associated reports. If the button is gray, then there is not enough information to begin the report, or the printer has not been set up.

When the icon is white, pressing it opens a Print Setup dialog or starts printing a report. If the Print Setup dialog is entered, configure the report and press Print to start printing.

29.3.2

Printing Realtime Reports

Select the Print quick key to print a realtime report. You can also print a realtime report from the Report Setup

dialog. For more information, see 29.3.3 Printing Most Common Reports .

29.3.3

Printing Most Common Reports

The following most common reports can be printed:

ECG Report

Realtime Report

Tabular Trends Report

Graphic Trend Report.

To print these reports, follow this procedure:

1.

Select the Main Menu quick key

→ from the Report column select Report Setup .

2.

Select the desired report tab.

3.

Check the settings.

4.

Select Print .

29.4

Automatically Printing Reports

When a parameter alarm switch is set to on and an alarm is triggered for this parameter, you can set the monitor to automatically print a Print on Alarm Report.

To do so, follow this procedure:

29 - 2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

1.

Access alarm related tabs such as the Alarm tab for a parameter in one of the following ways:

◆ Select the Alarm Setup quick key at the bottom of the screen.

◆ Select the parameter or waveform area of the desired parameter → select the Alarm tab.

◆ Select the Parameters Setup quick key at the bottom of the screen → select the desired parameter

select the Alarm tab.

Select the Main Menu quick key

from the Parameters column select Setup

select the desired parameter → select the Alarm tab.

2.

Switch on Alarm Outputs for desired parameters.

29.5

Stopping a Printing Task

To stop a printing task, follow this procedure:

1.

Select the Main Menu quick key

→ from the Report column select Print Queue .

2.

Select desired printing tasks and then select Delete . Alternately, select Delete All to stop all the printing tasks.

29.6

Configuring Reports

This section focuses on how to configure ECG reports, realtime reports, tabular trends reports, and graphic trends reports.

29.6.1 Configuring ECG Reports

To configure ECG reports, follow this procedure:

1.

Select the Main Menu quick key

→ from the Report column select Report Setup .

2.

Select the ECG Report tab.

3.

Set the desired options.

The following table explains some of the less familiar options.

Menu item

Speed

Auto Interval

12-Lead Format

Rhythm Lead 1

Rhythm Lead 2

Rhythm Lead 3

Function Description

Set the print speed of

ECG waveforms

25 mm/sec : prints 25 mm of ECG waveform per second.

50 mm/sec : prints 50 mm of ECG waveform per second.

Defines the spacing between the ECG waveforms on a printout

On: automatically adjusts the space between waveforms to avoid overlapping.

Off : each waveform area has the same size on a printout.

Note: Auto Interval is only enabled when 12×1 is selected for 12-Lead Format .

Select the format of 12lead ECG waveforms on a printout.

12×1: displays 12-lead ECG waveforms on one page in one column.

6×2 : displays 12-lead ECG waveforms on one page in two columns, with 6 lines in each column.

6×2+1 : displays 12-lead ECG waveforms on one page in two columns, with 6 lines in each column, and one rhythm lead waveform at the bottom.

3×4+1 : displays 12-lead ECG waveforms on one page in 4 columns, with 3 lines in each column, and one rhythm lead waveform at the bottom.

3×4+3 : displays 12-lead ECG waveforms on one page in 4 columns, with 3 lines in each column, and three rhythm lead waveforms at the bottom.

Select the lead that will be used as Rhythm

Lead 1, 2, or 3.

I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6

Note: Rhythm Lead setting is only enabled when 6×2+1 , 3×4+1 , or 3×4+3 is selected for

12-Lead Format .

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 29 - 3

Menu item

Format Sequence

Function

Select the recording method of ECG report generated by auto measurement

Description

Sequential : 12-lead ECG data are recorded sequentially and displayed in 3 lines and 4 columns with 2.5 seconds of ECG data for each column.

Simultaneous : Record simultaneous 12-lead ECG data.

NOTE

When ECG Lead Set is set to 3-Lead, the ECG report cannot be printed.

29.6.2

Configuring Realtime Reports

To configure tabular realtime reports, follow this procedure:

1.

Select the Main Menu quick key

from the Report column select Report Setup .

1.

Select the Realtime Report tab.

2.

Set the desired options.

The following table explains some of the options.

Menu item

Select Waveform

Function

Select the desired waveform to print

Description

Current Waveforms : prints the realtime report for the currently displayed waveforms and all the currently measured parameter values.

Selected Waveforms : prints the realtime report for the waveforms configured in the Select Waveform dialog, accessed by clicking the

Select Waveform button.

29.6.3

Configuring Tabular Trends Reports

To configure tabular trends reports, follow this procedure:

1.

Select the Main Menu quick key

→ from the Report column select Report Setup .

2.

Select the Tabular Trends Report tab.

3.

Set the desired options.

The following table explains some of the options.

Menu Item

Time

Period

Interval

Function Description

Select the time before which a tabular trends report will be printed

Select the period during which a tabular trends report will be printed.

/

Select the resolution of the tabular trends printed on a report.

Auto : one page of a tabular trends before the current time will be printed at the selected Interval .

All : all stored tabular trends will be printed at the selected

Interval .

30 min to 96 hrs : 30 min to 96 hrs of tabular trends before the selected Time will be printed at the selected Interval .

NIBP, EWS , GCS , C.O

.: at an interval of acquiring the values of selected parameter.

Auto : using the Interval setting of the Tabular Trends review page.

5 sec to 3 hrs : the tabular trends will be printed at the interval of 5 sec to 3 hrs .

29 - 4 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Menu Item

Report Format

Function

Select the printing principle.

Description

Parameter Oriented : print one page span of report with parameters listed by row and time listed by column when

Interval is set to Auto .

Time Oriented : print one page span of report with time listed by row and parameter listed by column when Interval is set to Auto .

Note that Interval on Auto only limits the number of rows or columns for Time. The report may have multiple pages to accommodate the number of parameters. To limit the parameters, change the Trend Group .

29.6.4

Configuring Graphic Trends Reports

To set graphic trends reports, follow this procedure:

1.

Select the Main Menu quick key → from the Report column select Report Setup .

2.

Select the Graphic Trends Report tab.

3.

Set the desired options.

Menu Item

Period

Function

Select the period during which a graphic trends report will be printed.

Description

Auto : one page of a graphic trends before the current time will be printed.

All : all stored graphic rends will be printed..

30 min to 96 hrs : 30 min to 96 hrs of graphic trends before the selected Time will be printed.

29.7

Viewing Printer Status

You can view the status of the most recent ten printing tasks in the Print Queue dialog. To view the status of printing tasks, follow this procedure:

1.

Select the Main Menu quick key → from the Report column select Print Queue .

2.

View the status of printing tasks in the Print Queue dialog.

Each printing task includes the following information:

Print time

Report title

Printer name (when using the printer server) or IP address (when using the network printer)

Printing status, for example, printing, failed, retrying, and waiting.

29.8

Printer Out of Paper

When the printer runs out of paper, the print request will remain incomplete. If there are too many print jobs that remain incomplete, then a printer error may occur. To resolve the error, install paper. Some print requests may be lost if the incomplete tasks overflow the printer buffer. Missing or lost reports must be recreated and printed.

To reduce risks of printer errors, maintain enough paper in the printer.

29.9

List of Reports

The monitor can output the following reports to the printer:

■ Realtime reports

◆ Realtime Report

Freeze Report

◆ ECG Report (5-lead, 6-lead , 12-lead)

12-Lead Interpretation Report (realtime)

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 29 - 5

◆ Print on Alarm Report

Alarm Limits Report

◆ Titration Table Report

Hemodynamic Calculation Report

◆ Oxygenation Calculation Report

Ventilation Calculation Report

◆ Renal Calculation Report

EEG Report

◆ CSA Report (EEG expand View - CSA)

EEG Report (EEG expand View - EEG)

◆ Hemodynamic Report (hemodynamic parameters)

◆ Hemodynamic Report (hemodynamic graphics)

◆ ST Report (Realtime)

◆ QT Report (Realtime)

◆ OxyCRG Report

◆ C.O. Measure Report

◆ PAWP Report

◆ ECG 24h Summary & Typical Strips

◆ Device Integration - Loops Report (BeneLink)

History reports

◆ Tabular Trends Report

◆ Graphic Trends Report

◆ Event List Report

◆ Event Detail Report

◆ Full Disclosure Overview Report

◆ Full Disclosure Detail Report

◆ 12-Lead Review Interpretation Report (review)

◆ ST Report (review)

◆ OxyCRG review Report (review)

End Case Report

◆ End Case Report Realtime Report

◆ End Case Report ECG Report

◆ End Case Report Tabular Trends Report

◆ End Case Report Graphic Trends Report

29 - 6 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

30

Using the On-Screen Timers

The monitor has a Timer function to notify you when a preset time period is expired or how much time has elapsed. You can simultaneously display up to four timers.

30.1

Displaying Timers

To display a timer, follow this procedure:

1.

Access Tile Layout in either of the following ways:

◆ Select the Screen Setup quick key → select the Tile Layout tab.

◆ Select the Main Menu quick key → from the Display column select Tile Layout .

2.

Click the numeric tile area where the timer is to be displayed, and then select a timer from the popup list.

30.2

Controlling the Timer

The timer provides the following controls, depending on the state and settings of the timer:

Start

Reset

: starts the timer.

Pause : pauses the timer.

Resume : resumes the timer.

: clears the timer.

30.3

Setting the Timer

You can set each timer independently. To set the timer, follow this procedure:

1.

Select the timer area to enter the Timer Setup dialog.

2.

Set Timer Type :

Normal : The timer has a single and defined run time, and stops with a notification beep when the run time is reached.

◆ Advanced : The timer has a single and defined run time. When the run time is reached, the timer makes a notification beep and continuously displays the time beyond the end of run time.

Cycled : The timer has a single and defined run time. When the run time is reached, the timer makes a notification beep and restarts automatically. The number of elapsed cycles is also displayed on the left of the Timer tile.

Unlimited : The timer displays the time elapsed since the timer was started.

Clock : The timer displays the system time.

3.

Set Direction .

Up : the timer counts up, from zero to Run Time.

Down : the timer counts down, from Run Time to zero.

4.

Set Run Time .

5.

Set Reminder Volume .

A progress bar is shown with the run time. When the remaining time is 10 seconds, the monitor issues a reminder tone and the timer flashes in red, prompting you that the run time is about to expire.

NOTE

You cannot change timer settings, other than Reminder Volume when a timer is running.

You can set Direct, Run Time, and Reminder Volume only for normal, advanced, and cycled timers.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 30 - 1

This page intentionally left blank.

30 - 2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

31

Managing Configurations

31.1

Configuration Introduction

When performing continuous monitoring on a patient, the clinical professional often needs to adjust the monitor’s settings according to the patient’s condition. The collection of all these settings is called a configuration. The system configuration items can be classified as: parameter configuration items, conventional configuration items, and user maintenance items. Allowing you to configure the monitor more efficiently, the monitor provides different sets of configurations to accommodate the varying patient categories and departments. You can change some settings from a certain set of configuration and then save the changed configuration as a user configuration.

The default configurations provided for your monitor are department-oriented. You can choose either from:

General

OR

ICU

Neonatology

CCU

Each department has three different sets of configurations tailored for adult, pediatric and neonatal patients respectively.

WARNING

The configuration management function is password protected. The configuration management tasks must be performed by clinical professionals.

31.2

Changing the Department

If the current department configuration is not the one you want to view, you can change the department by following this procedure:

1.

Select the Main Menu quick key → from the Configuration column select Manage → input the required password

→ select .

2.

Select Change Department .

3.

Select a department.

4.

Select Ok .

CAUTION

Changing the department will delete all current user configurations.

31.3

Setting Default Patient Category

To set the default patient category when admitting a new patient, follow this procedure:

1.

Select the Main Menu quick key

from the Configuration column select Manage

→ input the required password → select .

2.

Set Default Patient Category .

31.4

Setting Default Configuration

The monitor will load the pre-set default configuration in the following cases:

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 31 - 1

A patient is admitted.

A patient is discharged.

Patient category is changed.

To set the default configuration, follow this procedure:

1.

Select the Main Menu quick key

from the Configuration column select Manage

→ input the required password → Select .

2.

Select Select Default Config .

3.

Select Load the Latest Config or Load Specified Config .

◆ When you select Load the Latest Config , the latest configuration is loaded when the monitor is started or a patient is admitted.

When you select Load Specified Config , the selected configuration of Default Adult Config , Default

Ped Config , or Default Neo Config is loaded when the monitor is started or a patient is admitted. The specified configuration can be the factory default configuration, the age segments configuration, or a saved user defined configuration. As an example, select Default Neo Config and then select Factory

Default , Neo GA Segments , or a user configuration. For more information on defining age segments, see

31.5 Defining Age Segments

.

31.5

Defining Age Segments

You must define age segments for any patient category you want to load configurations based on the patient’s age. To do so, follow this procedure:

1.

Select the Main Menu quick key → from the Configuration column select Manage → input the required password

Select .

2.

Select Select Default Config .

3.

Respectively select the edit icons followed Customize Configurations for Adult Age Segments ,

Customize Configurations for Ped Age Segments , Customize Configurations for Neo Gestational Age

Segments to define the age segments for each patient category. The age segment of the neonatal patient is based on the baby’s gestational age.

31.6

Saving Current Settings

Current settings can be saved as a user configuration. Up to 25 user configurations can be saved.

To save current settings, follow this procedure:

1.

Select the Main Menu quick key → from the Configuration column select Manage → input the required password

Select .

2.

Select Save Current Settings .

3.

Input the configuration name.

4.

Select Ok to save current settings as a user configuration.

31.7

Deleting a Configuration

To delete a configuration, follow this procedure:

1.

Select the Main Menu quick key → from the Configuration column select Manage → input the required password

Select .

2.

Select Delete Configuration .

3.

Select the configuration you want to delete:

◆ In the Delete Configuration dialog, selecting Local tab shows the existing user configurations on the monitor.

In the Delete Configuration dialog, selecting USB Drive tab shows the existing user configurations on the USB drive.

4.

Select Delete .

5.

Select Ok .

31 - 2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

31.8

Transferring a Configuration

When installing several monitors with identical user configurations, it is not necessary to set each unit separately. Use a USB drive to transfer the configuration from monitor to monitor.

31.8.1

Exporting a Configuration

To export the current monitor’s configuration, follow this procedure:

1.

Connect the USB drive to the monitor’s USB port.

2.

Select the Main Menu quick key → from the Configuration column select Manage → input the required password

Select .

3.

Select Export Configuration .

4.

Select the configurations and User Maintenance Settings to export.

5.

Select Export .

31.8.2

Importing a Configuration

To import the configuration from the USB drive to the monitor, follow this procedure:

1.

Connect the USB drive to the monitor’s USB port.

2.

Select the Main Menu quick key → from the Configuration column select Manage → input the required password

Select .

3.

Select Import Configuration .

4.

Select the configurations and User Maintenance Settings to import.

5.

Select Import .

31.8.3

Loading a Configuration

You may make changes to some settings during operation. However, these changes or the pre-selected configuration may not be appropriate for the newly admitted patient. Therefore, the monitor allows you to load a desired configuration to ensure that all the settings are appropriate for your patient.

To load a configuration, follow this procedure:

1.

Select the Main Menu quick key

from the Configuration column select Load .

2.

Select the desired configuration.

◆ Select the configuration on this monitor in the Local page.

◆ Select the configuration on the USB drive in the USB Drive page.

3.

Select Load .

NOTE

The monitor may configure some settings by default when you load a configuration of different software version with the current configuration.

31.9

Printing Configurations

To print both factory configurations and user configurations, follow this procedure:

1.

Select the Main Menu quick key → from the Configuration column select Manage → input the required password

select .

2.

Select Print Configuration .

3.

Select desired configurations.

4.

Select Print .

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 31 - 3

31.10

Modifying Configuration Password

To modify the configuration password, follow this procedure:

1.

Select the Main Menu quick key → from the Configuration column select Manage → input the required password

Select .

2.

Select Modify Password .

3.

Respectively input the old password and new password.

4.

Select OK .

31 - 4 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

32

Password-protected User Settings

This chapter describes the password-protected functions accessed in the Maintenance dialog.

32.1

Setting the Device Location

32.1.1

Setting Monitor Information

To set monitor information, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the Device Location tab.

3.

Input monitor name, facility name and department name.

32.1.2

Setting Monitor Location

If your monitor is installed at a fixed position, or you do not need clinicians to change the room number and bed number, you can set Monitor Location to Fixed . To set the monitor location, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the Device Location tab.

3.

Set Location .

◆ Fixed : the Patient Management dialog displays Bed No.

and Room No.

, but you cannot change them.

Unfixed : you can change Bed No.

and Room No.

from the Patient Management dialog.

4.

Input the room number and bed number.

NOTE

If Location is set to Unfixed, Bed No.

and Room No.

are cleared from the Device Location tab each time you discharge a patient.

32.1.3

Enabling the Auto Obtain Bed Number Function

If the monitor is connected to the wired network, the monitor can automatically set the patient’s bed number according to the bed number information bonded to the bedside network connector.

To enable this function, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the Device Location tab.

3.

Select Unfixed .

4.

Switch on Auto Obtain Bed No..

NOTE

The Auto Obtain Bed Number function is available only when the switch connected to the monitor supports the LLDP or CDP protocol, and the corresponding protocol is enabled.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 32 - 1

32.2

Changing Patient Management Settings

32.2.1

Selecting Displayed Patient Information

You can define which items can be displayed and edited from the Patient Management tab. To do so, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the Patient Management tab.

3.

Select the fields you want to display in the Patient Management dialog.

4.

If necessary, select the customized fields and input names for these fields.

NOTE

If the monitor is connected with the CMS, the patient information items and customized fields are loaded from the CMS.

32.2.2

Setting the Range of Finding Patient

You can configure which patient can be found in the CMS or the ADT server. To do so, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password

select .

2.

Select the Patient Management tab.

3.

Select Find Patient :

All Patients : search from all patients in the CMS or ADT server.

Current Department Patients : only searches from the current department in the CMS or ADT server.

32.2.3

Setting the Length of Data Transferred by N1 or T1

When using N1 or T1 to transfer the patient data, you can set the data transfer length. To do so, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password

→ select .

2.

Select the Patient Management tab

→ select the Transfer tab.

3.

Set Transfer Data Length .

32.2.4

Setting the Strategy of Transferring Patient Information

When N1, T1or MPM is connected to the monitor, N1, T1 or MPM automatically uploads the data to the monitor if the patient demographic information in the monitor are consistent with those of in N1, T1 or MPM. However, the monitor needs a data transfer strategy if the patient demographic information in the monitor are not consistent with those of in the N1/T1/MPM. To set the data transfer strategy, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password

→ select .

2.

Select the Patient Management tab → select the Transfer tab.

3.

Select Data Transfer Strategy .

◆ Always Ask : always prompts a dialog box to ask for strategy.

◆ Continue Patient in Module : continue to use the patient data in N1/T1/MPM. The monitor discharges the patient, and automatically admits a new patient and copies all data from N1/T1/MPM.

Continue Patient in Monitor : continue to use the patient information in the monitor. The monitor deletes all patient information in the N1/T1/MPM and copies the parameter settings in the monitor to the N1/T1/MPM.

32 - 2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

32.2.5

Transferring the N1 or T1 Settings

When transferring the N1 or T1 data, the settings can also be transferred. To transfer settings, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the Patient Management tab

→ select the Transfer tab.

3.

Set Apply Module Settings to On .

32.2.6

Auto Discharging a Patient after Monitor Power Off

You can let the monitor automatically discharge after the monitor has been switched off for a period of time.

To set the time period of discharging a patient, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the Patient Management tab → select the Discharge tab.

3.

Select a time for Auto Discharge When Power Off . The monitor will automatically discharge the patient when it is turned off for the designated period of time. The default is Never . That is to say the monitor will not automatically discharge a patient, regardless of how long the monitor has been switched off.

32.2.7

Configuring Alarms for Auto Deleting Discharged Patients

The monitor automatically delete earlier discharged patients when the memory is limited. You can select whether an alarm is issued when discharged patients are automatically deleted and the monitor memory is very low.

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the Patient Management tab

Discharged tab.

3.

Configure the alarms as needed:

◆ Prompt on patient auto deleted: an alarm is issued when the monitor automatically deletes earlier discharged patients.

◆ Alarm on storage is nearly full : selects whether an alarm is issued when the monitor memory is very low and the priority of this alarm.

32.2.8

Configuring Discharged Patients Data

To configure discharged patient data, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the Patient Management tab → Discharged tab.

3.

Switch on Include Patient Demographics When Exporting Patient Data if you want to include patient demographics when exporting the patient data.

4.

Selects whether patient data is deleted when the patient is discharged.

◆ Auto : patient data is not deleted when the patient is discharged. The oldest discharged patient data will be deleted when the storage space of the monitor is full.

◆ Right Now : deletes patient data as soon as the patient is discharged.

◆ 7 days : deletes discharged patient data seven days after the patient was discharged.

◆ 1 Month : deletes discharged patient data one month after the patient was discharged

32.2.9

Clearing All Patient Data

You can delete all patient information and data in the monitor. To do so, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

select .

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 32 - 3

2.

Select the Patient Management tab → select the Discharge tab.

3.

Select Clear All Patient Data .

4.

Select OK . Select OK again for confirmation until a message indicating you that all data is cleared.

CAUTION

Clearing patient data will discharge the current patient. Make sure there is no patient is being monitored before you clear all data.

32.2.10

Selecting Query Criteria for Searching Patients in ADT Server

You can define which criteria can be used to search patients in the ADT server. To do so, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the Patient Management tab → ADT Query tab.

3.

Select the query criteria as desired.

32.2.11

Setting Patient Location Options for the Discharge Screen

To set options for patient location, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the Patient Management tab → select the Location tab.

3.

Input or edit patient location options.

32.2.12

Hiding Patient Name

To hide patient name on specific screens, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the Patient Management tab → select the Display tab.

3.

Set the following switches as desired:

◆ Switch off Primary Screen Display Full Name : does not display patient name in the patient information area on the primary display.

Switch off Secondary Screen Display Full Name : does not display patient name in the patient information area on the secondary display, if configured.

◆ Switch off Remote View Display Full Name : does not display patient name in the patient information area on the remote monitors when this monitor is viewed by other monitors.

Switch off Remote View Bedlist Display Full Name : does not display patient name in bed list on the remote monitors when this monitor is viewed by other monitors.

32.3

Changing the Alarm Settings

32.3.1

Setting Alarm Tone Properties

32.3.1.1

Setting the Minimum Alarm Volume

To set the minimum alarm volume, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the Alarm tab

Audio tab.

32 - 4 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

3.

Set Minimum Alarm Volume .

32.3.1.2

Setting the Audible Alarm Tone Pattern

You can distinguishes the heart beat tone, pulse tone, and keystroke tone by frequency. You can choose any of the three tone patterns: ISO, Mode 1, and Mode 2. For more information, see

8.3.3 Alarm Indicators .

To set the alarm tone pattern, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the Alarm tab → Audio tab.

3.

Set Alarm Sound to ISO , Mode 1 or Mode 2 .

32.3.1.3

Setting the Interval between Alarm Sounds

If you choose the ISO pattern, you can change the interval between alarm tones. To change the interval between alarm tones, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the Alarm tab → Audio tab.

3.

Set High Alarm Interval , Med Alarm Interval , and Low Alarm Interval .

32.3.1.4

Setting Alarm Volume Escalation

The monitor provides the function of automatically escalating the alarm tone volume. When this function is enabled, if an alarm is not reset within the designated delay time after the alarm occurs, the volume of the alarm tone increases automatically.

To set the alarm volume escalation, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the Alarm tab → Audio tab.

3.

Set Auto Increase Volume .

◆ 2 Steps : if an alarm is not reset within the designated delay time after the alarm occurs, the alarm volume automatically increases by two levels.

1 Step : if an alarm is not reset within the designated delay time after the alarm occurs, the alarm volume automatically increases by one level.

◆ Off : if an alarm is not reset within the designated delay time after the alarm occurs, the volume of the alarm tone does not change.

4.

Select Increase Volume Delay to set the delay time of alarm volume escalation.

NOTE

The alarm volume escalation function is not applied to the latched alarms.

32.3.2

Setting Alarm Pausing and Resetting Properties

32.3.3

Defining the Pause Function

You can either pause alarms or pause alarm tones. This depends on the pause setting. To set the pause function, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password

select .

2.

Select the Alarm tab → Pause/Reset tab.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 32 - 5

3.

Set Pause to Alarm Pause or Audio Pause .

32.3.3.1

Setting the Alarm Pause Time/Alarm Tone Pause Time

The alarm pause time or alarm tone pause time can be set to 1 min , 2 min , 3 min , or Permanent . The default audio pause time is two minutes.

To set the alarm tone pause time, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the Alarm tab → Pause/Reset tab.

3.

Set Pause Time .

32.3.3.2

Setting the Priority of Audio Paused Alarms

To select alarm of what priority can be paused or alarm sound of what priority can be paused, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the Alarm tab → Pause/Reset tab.

3.

Set Pause Priority .

◆ All : pressing the Alarm Pause quick key pauses all alarms.

◆ Med & Low : pressing the Alarm Pause quick key pauses alarms of medium and low priority. The high priority alarms will not be paused.

Disable : the Alarm Pause quick key is disabled.

32.3.3.3

Disabling the Options of Prolonging Alarm Pause Time/Alarm Tone Pause Time

To disable specific options for the clinician when prolonging alarm pause time or the alarm tone pause time, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the Alarm tab → Pause/Reset tab.

3.

Switch off Pause 5 min , Pause 10 min , or Pause 15 min .

32.3.3.4

Setting Alarm Light Status on Alarm Reset

When the alarm system is reset, the monitor presents the alarm light by default , but you can switch off the alarm light. To change the alarm light setting, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password

→ select .

2.

Select the Alarm tab → Pause/Reset tab.

3.

Set Alarm Light .

◆ On When Reset : when the alarm system is reset, the alarm tones of the current alarms are switched off, but the alarm lamp remains flashing.

Off When Reset : when the alarm system is reset, both the alarm tone and alarm lamp of the current alarms are switched off.

32.3.3.5

Setting the Reminder Tones

When the alarm volume is set to zero, or the alarm is reset or switched off, the monitor issues a periodical reminder tone. The reminder tone is enabled by default.

To set the reminder tone, follow this procedure:

32 - 6 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password → select .

2.

Select the Alarm tab

Pause/Reset tab.

3.

Set Alarm Reset Reminder and Alarm Off Reminder .

On : the monitor issues reminder tones at a designated interval.

Re-alarm : if the alarm condition persists the alarms marked with “√” will be re-generated after the designated reminder tone interval.

◆ Off : the monitor does not issue reminder tones at a designated interval. The alarms marked with “√” will be silenced.

4.

Set Reminder Interval to 10 min , 5 min , 3 min , 2 min , or 1 min .

32.3.4

Latching Physiological Alarms

To latch physiological alarms, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the Alarm tab

Latching tab.

3.

Select how you want to latch the alarms.

Alarm latching rules are as follows:

You can separately latch visible alarm signals and audible alarm signals.

Latching audible alarm signals simultaneously latches the corresponding visible signals.

Selecting alarms of lower priority simultaneously latches higher priority alarms.

Switching V-Tach Latching Lock off allows you to switch on or off V-Tach Latching .

When V-Tach Latching Lock is switched on, V-Tach latching follows the setting of Lethal alarm latching and you cannot control V-Tach Latching .

32.3.5

Changing Remote Alarm Settings

32.3.5.1

Resetting Alarms for Remote Devices

You can reset the alarms occurring on the remote devices that are viewed on the Remote View screen of the monitor. To enable this function, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the Alarm tab

Remote View tab.

3.

Switch on Reset Remote Bed Alarms . Then the Alarm Reset button appears on the bottom left of the

Remote View screen.

To reset any remote device alarms, the clinician may now select the Alarm Reset button at the bottom of the

Remote View screen.

NOTE

You can reset remote device alarms only if the Alarm Reset By Other Bed switch is on at the remote devices and alarms are occurring on the remote device.

32.3.5.2

Authorizing the Alarm Reset to Other Devices

Alarms on your monitor can be reset by remote devices if you enable this function. To do so, follow this procedure:

1.

Select the Main Menu quick key

→ from the System column select Maintenance

→ input the required password

→ select .

2.

Select the Alarm tab

Remote View tab.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 32 - 7

3.

Switch on Alarm Reset By Other Bed .

32.3.5.3

Selecting the Alarm Reminder for Remote Devices

You can configure what alarm indicators are necessary for the remote devices. To do so, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the Alarm tab

Remote View tab.

3.

Select Alarm Reminder :

◆ Visible + Audible : the monitor provides visual alarm indication, and continuous audible alarm indication if the alarm persists at the remote device.

Visible + Single Tone : the monitor provides visual alarm indication, and a single tone when the alarm occurs at the remote device.

◆ Visible Only : the monitor only provides visual alarm indication.

32.3.5.4

Presenting Alarm Sound for Remote Devices as per Alarm Priority

You can configure what priority of remote device alarms are presented for audible notification. To do so, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the Alarm tab

Remote View tab.

3.

Select Alarm Priority :

◆ All : the monitor sounds if an alarm occurs.

◆ High & Med : the monitor sounds if a high or medium priority alarm occurs.

◆ High Only : the monitor sounds only if a high priority alarm occurs.

32.3.5.5

Setting the Tone Pattern for Alarms from Remote Devices

The monitor provides the same alarm tone pattern for the remote device alarms as those for your monitor

alarms. For more information on alarm tone pattern, see 8.3.3 Alarm Indicators

.

To set the alarm tone pattern for the remote device alarms, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the Alarm tab

→ select the Remote View tab.

3.

Set Alarm Sound to ISO , Mode 1 or Mode 2 .

32.3.5.6

Switching Off the Remote Device Disconnection Alarm

The monitor can provide an alarm if remote devices, for example, a bedside monitor or a telemetry, are disconnected. By default, the function is enabled. To disable the alarm, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the Alarm tab

Remote View tab.

Switch off Remote Disconnected Alarm .

32.3.6

Setting Nurse Call Properties

32.3.6.1

Selecting Alarms to be Sent to the Nurse Call System

To set the type and priority of alarms that are sent to the nurse call system, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

32 - 8 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

2.

Select the Alarm tab → Nurse Call tab.

3.

Select Signal Type

◆ Pulse : the nurse call signal is a pulse signal and each pulse lasts one second. When multiple alarms simultaneously occur, only one pulse signal is outputted. If an alarm occurs but the previous one is not cleared, a new pulse signal will also be outputted.

◆ Continuous : the nurse call signal lasts until the alarm ends. That is to say the duration of a nurse call signal is equal to that of the alarm condition.

4.

Select Contact Type to set the work mode of the nurse call relay.

5.

Select Alarm Priority to set the priority of alarms sent to the nurse call system.

6.

Select Alarm Type to set the type of alarms sent to the nurse call system.

32.3.6.2

Disabling the Receiving Call Help Function

If a monitor in the same department calls for help, you can receive the calling signal. This function is enabled by default. To disable this function, follow this procedure:

1.

Select the Main Menu quick key

→ from the System column select Maintenance

input the required password → select .

2.

Select the Alarm tab

Nurse Call tab.

3.

Switch off Receive Call Help.

32.3.6.3

Selecting the Source of Call Help Signals

You can choose which calling for help signals the monitor can receive. To do so, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the Alarm tab → Nurse Call tab.

3.

Set the switch of Only Beds from Remote View.

◆ On : The monitor can only receive the calling for help signals from the remote monitors being viewed.

◆ Off : The monitor can receive the calling for help signals from all the monitors in the same department.

32.3.6.4

Enabling Presenting the Nurse Call from the Telemetry

If the telemetry is connected, the monitor can pops up the nurse call dialog if nurse call is initiated from the telemetry. To enabling this function, follow this procedure:

1.

Select the Main Menu quick key

→ from the System column select Maintenance

input the required password → select .

2.

Select the Alarm tab

Nurse Call tab.

3.

Switch on Telemetry Nurse Call.

32.3.7

Changing Alarm Priority Settings

32.3.7.1

Setting the Priority of the ECG Lead Off Alarm

To set the priority of the ECG lead off alarm, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the Alarm tab

select the Other tab.

3.

Set the priority of ECG Lead Off .

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 32 - 9

32.3.7.2

Setting the Priority of the SpO

2

Sensor Off Alarm

To set the priority of the SpO

2

sensor off alarm, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the Alarm tab

select the Other tab.

3.

Set the priority of SpO2 Sensor Off .

32.3.7.3

Setting the Priority of the IBP No Sensor Alarm

To set the priority of the IBP no sensor alarm, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the Alarm tab

select the Other tab.

3.

Set the priority of IBP No Sensor .

32.3.7.4

Setting the Priority of the Integrated Device Disconnected Alarm

To set the priority of the alarm when the external device is disconnected, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the Alarm tab

select the Other tab.

3.

Set the priority of Integrated Device Disconnected .

32.3.7.5

Setting the Switch of the CMS and eGateway Disconnection Alarm

You can choose whether to issue an alarm when the monitor is not connected or disconnected from the CMS/ eGateway. To do so, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password → select .

2.

Select the Alarm tab

Other tab.

3.

In the Other block, switch on or off CMS/eGW Disconnected Alarm . If CMS/eGW Disconnected Alarm is switched off, the “Offline” alarm is not presented when the monitor is not connected or disconnected from the CMS/eGateway.

32.3.7.6

Setting the Priority of the CMS and eGateway Disconnection Alarm

To set the priority of the CMS and eGateway disconnection alarm, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password

select .

2.

Select the Alarm tab

Other tab.

3.

Set CMS/eGW Disconnected .

32.3.7.7

Setting the Switch of the SpO

2

Desat Alarm Off

You can choose whether switching off the SpO

2

Desat alarm is permissible or not. To do so, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the Alarm tab

→ select the Other tab.

3.

Set SpO2 Desat Alarm Off .

Disable : the SpO

2

Desat alarm is always on. You cannot switch it off.

32 - 10 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

◆ Enable : you can switch off the SpO

2

Desat alarm. For more information, see 12.6.1 Changing the SpO

2

Alarm Settings

.

WARNING

If a clinician switches off the SpO2 Desat alarm, the monitor will not alarm when the patient’s SpO

2

is extremely low. This may result in a hazard to the patient.

32.3.7.8

Setting the Switch of the Apnea Alarm Off

You can choose whether switching off the apnea alarm is permissible or not. To do so, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the Alarm tab → select the Other tab.

3.

Set Apnea Alarm Off .

◆ Disable : the apnea alarm is always on. The clinician cannot switch it off.

◆ Enable : the clinician can switch off the apnea alarm. For more information, see

11.5.1 Setting the Resp

Alarm Properties .

WARNING

If a clinician switches off the apnea alarm, the monitor will not issue the apnea alarm when that situation occurs. This may result in a hazard to the patient. Keep the patient under close surveillance.

32.3.8

Changing Alarm Delay Settings

32.3.8.1

Setting the Alarm Delay Time

For continuously measured parameters, you can set the alarm delay time. If the alarm condition is resolved within the delay time, the monitor does not present the alarm.

To set the alarm delay time, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the Alarm tab → Other tab.

3.

Set Alarm Delay .

◆ Off : an alarm is presented as soon as the alarm condition occurs.

◆ 1 sec ~ 8sec : for continuously measured parameters, an alarm is not presented if the alarm condition is resolved within the designated delay time.

The setting of Alarm Delay is not applied to the apnea alarms and the ST alarms. You can set Apnea Delay and

ST Alarm Delay separately.

WARNING

The alarm delay time can be set to a maximum of 8 seconds. Changing this setting to an inappropriate level could result in a hazard to the patient.

32.3.8.2

Setting the ST Alarm Delay Time

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

select .

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 32 - 11

2.

Select the Alarm tab → Other tab.

3.

Set ST Alarm Delay .

32.3.9

Adjusting the Alarm Light Brightness

To adjust the alarm light brightness, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password

→ select .

2.

Select the Alarm tab → Other tab.

3.

From the Alarm Light Brightness area, set Primary Screen (note that you may have to scroll up the dialog to see this option). If you set alarm light brightness to Auto , the monitor automatically adjusts the alarm light brightness according to the ambient light. The stronger the ambient light is, the brighter the alarm light is.

For N22/N19, if you are using a secondary display, set the Secondary Screen too.

32.3.10

Changing Other Alarm Settings

32.3.10.1 Setting the Lethal Arrhythmia Alarms Switch

To allow a clinician to disable the lethal arrhythmia alarms, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the Alarm tab

select the Other tab.

3.

From the Other block, set Lethal Arrh Alarms Off as follows.

Lethal Arrh Alarms Off is set to Disable by default. In this case, a clinician cannot switch off lethal arrhythmia alarms.

◆ If you set Lethal Arrh Alarms Off to enable , a clinician can switch off lethal arrhythmia alarms from the ECG

dialog. For more information, see 9.6.4.1 Changing Arrhythmia Alarm Settings .

WARNING

If any of lethal arrhythmia alarms are switched off, the ECG waveform area displays a message to indicate that corresponding alarms are off.

If a clinician switches off all arrhythmia alarms, the monitor will not present arrhythmia alarms.

Always keep the patient under close surveillance.

32.3.10.2 Setting the Intubation Time

The default intubation time is 2 minutes. To change the intubation time, following this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the Alarm tab

Other tab.

3.

Set the Intubation Mode Period .

32.3.10.3 Enabling Automatic Recordings on an Alarm

Print on Alarm is set to Printer by default. To enable automatic recording via recorder when a parameter alarm is triggered, set Print on Alarm to Recorder . To do so, follow this procedure:

1.

Select the Main Menu quick key

→ from the System column select Maintenance

→ input the required password

→ select .

2.

Select the Alarm tab

→ select Other .

3.

Set Print on Alarm to Recorder .

32 - 12 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

32.3.10.4 Setting Arrhythmia Alarm Timeout Period

The arrhythmia algorithm can disable alarm light and alarm tone for designated period of time when certain arrhythmia alarms are detected.

To set the arrhythmia alarm timeout period, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the Alarm tab → select the Other tab.

3.

From the Other block, set Arrhy Shield Time .

NOTE

For the following alarms, alarm light and alarm tone cannot be disabled: HR high, HR low,

Tachycardia, Bradycardia, Afib End, Irr Rhythm End.

The arrhythmia alarm timeout period is only applied to the medium priority chains and atrial fibrillation chain. For the alarms in the high priority chain, alarm tone and alarm light are presented

as soon as the alarm condition is detected. Refer to 9.6.5.1 Arrhythmia Alarm Chains for more

information on the arrhythmia alarm chain.

Alarm indication rules for alarms in the atrial fibrillation chain are the same with those for the medium priority chains.

32.4

Changing CAA Settings

32.4.1

Enabling Inputing the Clinician ID

To display the clinician ID on the EWS screen, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the CAA tab

→ select the EWS tab.

3.

Switch on Clinician ID .

4.

Select Clinician ID Timeout to set how long the clinician ID will remain valid.

NOTE

The EWS tab is only available if a license for EWS is installed in the monitor.

32.4.2

Selecting the Default Scoring Tool

To set the default scoring tool for different patient category, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the CAA tab → select EWS tab.

3.

From the Select Default Score area, set Default Adult Score , Default Ped Score , and Default Neo Score .

32.4.3

Managing the Scoring Tools

To manage the scoring tool, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the CAA tab

→ select EWS tab.

3.

Select the Manage Score button.

Delete local scoring tools: from the Local page, delete any unnecessary scoring tools.

Import the desired scoring tools to the monitor: from the USB drive page, select the scoring tools on the USB drive, and then select Import .

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 32 - 13

NOTE

The monitor provides MEWS and NEWS by default. You cannot delete them.

32.4.4

Setting GCS Threshold for Each Consciousness Level

You can configure the threshold and color of each consciousness level. To do so, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password

→ select .

2.

Select the CAA tab → select the GCS tab.

3.

Set high limit, low limit for each level.

32.4.5

Changing SepsisSight Settings

To change SepsisSight settings, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password

→ select .

2.

Select the CAA tab

→ select the SepsisSight tab.

3.

Select The first of resuscitation to define the desired period of initial resuscitation. Select and edit the goals for initial resuscitation.

4.

In the Bundles area, select and edit treatments to be completed in 1 hour, 3 hour, and 6 hours.

32.5

Changing Module Settings

32.5.1

Selecting the ECG Standard

Select the ECG standard according to the leadwires you are using. To select the ECG standard, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the Module tab

ECG tab.

3.

Set ECG Standard to AHA or IEC .

32.5.2

Changing the QTc Formula

The monitor uses the Hodges correction formula by default to correct the QT interval for heart rate. To select the

QTc formula, follow this procedure:

1.

Select the Main Menu quick key

→ from the System column select Maintenance

input the required password → select .

2.

Select the Module tab → ECG tab.

3.

Set QTc Formula .

◆ Hodges:

QTc = QT + 1.75

  HeartRate – 60 

◆ Bazett:

QTc = QT 

HeartRate

60

2

◆ Fridericia:

QTc = QT 

HeartRate 

3

◆ Framingham:

QTc = QT + 154 

1 –

60

HeartRate

32 - 14 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

32.5.3

Setting the 12-Lead Order Switch

To set whether to send the order number of 12-lead interpretation report to the hospital information system while saving the report, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → i nput the required password

→ select .

2.

Select the Module tab.

3.

Select the ECG tab.

4.

Set 12-Lead Order .

32.5.4

Calibrating ECG

The ECG signal may be inaccurate due to hardware or software problems. As a result, the ECG waveform amplitude becomes greater or smaller. In that case, you need to calibrate the ECG module. To do so, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

select .

2.

Select the Module tab

ECG tab.

3.

From the bottom left corner of the dialog select Calibrate .

32.5.5

Manually Zeroing the CO

2

Module

To zero the CO

2

module, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the Module tab → CO2 tab.

3.

From the bottom left corner of the dialog select Zero .

32.5.6

Hiding the Invalid Display after Zeroing the CO

2

Module

After the zero calibration is completed, the CO

2

module reacquires the CO

2

readings. During the reacquisition period, “ Zero Recovering ” is displayed in the CO

2

numeric area. Valid data will reappear 30 seconds after the zero calibration is started.

To hide the display of the “ Zero Recovering ” message, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the Module tab

CO2 tab.

3.

Switch off Zero Recovery for 30s .

Selecting Defaults from the bottom of the dialog can restore this setting to the factory default.

32.5.7

Manually Zeroing the AG Module

To zero the AG module, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the Module tab

AG tab.

3.

From the bottom left corner of the dialog select Zero .

32.5.8

Hiding the Invalid Display after Zeroing the AG Module

After the zero calibration is completed, the AG module reacquires the AG readings. During the reacquisition period, “ Zero Recovering ” is displayed in the AG numeric area. Valid data will reappear 30 seconds after the zero calibration is started.

To hide the display of the “ Zero Recovering ” message, follow this procedure:

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 32 - 15

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the Module tab

AG tab.

3.

Switch off Zero Recovery for 30s .

Selecting Defaults from the bottom of the dialog can restore this setting to the factory default.

32.5.9

Selecting the Gas Type for Measurement

You can select what gases detected by the monitor are used for measurements. To do so, follow this procedure:

1.

Ensure that the AG module is inserted in the monitor.

2.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

3.

Select the Module tab

AG tab.

4.

Set Types of Gas Measurement . Any gas that is not selected, will not be displayed or used for calculating the MAC value. When the concentration of any unselected gas is equal to or greater than 1%, you will be prompted to select the gas and enable the measurement ( Set XX measurement to On) .

Selecting Defaults from the bottom of the dialog can restore this setting to the factory default.

32.5.10

Setting Parameter Measurement Timeout

NIBP measurements become outline fonts after a preset time. This feature prevents older values being misinterpreted as current measurements. To adjust the timeout period for PAWP, C.O., and NIBP, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password → select .

2.

Select the Module tab → select the Other tab.

3.

Set the timeout period for PAWP, C.O., and NIBP.

32.5.11

Setting Flow Rate (For Sidestream CO

2

Module Without O

2

)

When you are using the sidestream CO

2

module without the O

2

monitoring function to monitor a neonatal patient, you can select flow rate. To do so, follow this procedure:

1.

Ensure that the sidestream module without O

2

is inserted in the monitor.

2.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

3.

Select the Module tab → Other tab.

4.

Set CO2 Flow Rate For Neo .

32.6

Changing the Review Settings

32.6.1

Hiding Undesired Review Tabs

If you do not need to review some items, you can hide them. To do so, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the Review tab.

3.

From the Tabs page, Event page or Arrh Mark page deselect those that do not need to be available to clinicians during review.

32.6.2

Renaming Events

You can rename manual events and arrhythmia events. To do so, follow this procedure:

32 - 16 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the Review tab.

3.

Switch on Rename Event .

32.6.3

Exporting Patient Data

To export data of the current and discharged patients, follow this procedure:

1.

For N17/N15/N12, connect the USB drive to the monitor’s USB connector. For N22/N19, connect the USB drive to the monitor’s MSB connector.

2.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

3.

Select the Review tab

Export tab.

4.

Select Export Patient Data .

5.

From the patient list select desired patients.

6.

Select Export Patient Data .

32.7

Setting the Printer

32.7.1

Setting the Printer Properties

To set a network printer, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the Print tab.

3.

In the Printer tab, select Connection Type to choose whether you want to output patient reports via the print server or a network printer.

4.

If you set Connection Type to Print Server , perform the following settings:

◆ Print Server Address : select

◆ Port : selects to input the print server name or IP address.

to input the print server port. If the CMS is used as the print server, set Port to 6603 .

◆ Set the report type in the Report Type area. For more information, see section

32.7.2 Setting the

Report Type .

5.

If you set Connection Type to Printer , perform the following settings:

◆ Select Printer IP Address . Select

◆ Select Paper Size .

◆ Select Resolution .

to input the IP address of the desired network printer.

After finishing configurations, you can select the Print Test Page button to verify that the printer works properly.

32.7.2

Setting the Report Type

If the Connection Type is set to Print Server , you need to set the report type:

1.

Select the Main Menu quick key

→ from the System column select Maintenance

→ i nput the required password → select .

2.

Select the Print tab.

3.

Set Connection Type to Print Server .

4.

Select the report you intend to set, for example, General Report .

5.

Set the Print Action :

◆ Paper : outputs paper report.

◆ PDF : outputs the electronic report in PDF format.

◆ Paper And PDF : outputs both the paper report and the PDF report.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 32 - 17

6.

For paper report, respectively select Printer and Printer Resolution to set the default printer and the resolution for the default printer.

7.

For PDF report, select the PDF Resolution to set the resolution.

NOTE

General reports refer to the reports other than the end case report and realtime alarm report.

32.7.3

Setting the Report Layout

To set the patient information you want to display on general reports, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the Print tab.

3.

Select the Report Layout tab.

4.

Select the desired items under Report Name . N/A indicates that this item is not displayed on a report.

32.7.4

Setting Patient Information on 12-lead ECG Reports

To set the patient information you want to display on 12-lead ECG reports, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the Print tab.

3.

s elect the ECG Report tab.

4.

On the right of the ECG Report page, select the desired patient information items. Patient ID, Patient

Name, Age, and Gender are displayed on an ECG report by default.

NOTE

You can only set the patient information to display on the ECG report from the ECG Report page.

Patient information configured in the Report Layout page is not applied to ECG reports.

32.7.5

Configuring the Name of PDF Files

If you use the print server to output patient reports, you can use the PDF printer to output PDF format reports. To configure the name of PDF files, follow this procedure:

1.

Select the Main Menu quick key

→ from the System column select Maintenance

→ input the required password → select .

2.

Select the Print tab → PDF File Name tab.

3.

Select the desired items and their sequence under PDF File Name . N/A refers to no information.

32.7.6

Showing the Second Mark

To set whether you want to show second marks on the report output by the printer, follow this procedure.

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password

→ select .

2.

Select the Print tab

Other tab.

3.

Enable or disable Second Mark (Printer) .

32.7.7

Setting the Arrhythmia Information to Output by the Recorder

To set whether you want to include arrhythmia thresholds and QRS thresholds in the report output by the recorder, follow this procedure.

32 - 18 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the Print tab

Other tab.

3.

Enable or disable Arrhy Settings (Recorder) .

32.7.8

Setting Parameter Unit

To set parameter unit, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Set the Unit tab.

3.

Set unit for each parameter.

32.8

Setting the Time

32.8.1

Setting the Time Synchronization

If the NTP Time Synchronization function is enabled, the monitor can automatically synchronize its time with a time server on which NTP (Network Time Protocol) is implemented. If there are many monitors connected within one network, you can enable the time synchronization to synchronize the time, and make it unnecessary to manually set the time for individual monitors.

32.8.1.1

Enabling NTP Time Synchronization

To enable the NTP Time Synchronization function, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the Time tab

Time Synchronization tab.

3.

Switch on Start NTP Time Sync .

32.8.1.2

Setting the Time Server Address

To set the DNS name of the time server address, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the Time tab

Time Synchronization tab.

3.

Input the name or IP address of the time server.

After configuring the NTP time server, select Network Test to verify that the NTP server is properly connected.

32.8.1.3

Setting the Time Synchronization Interval

To set the time synchronization interval, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the Time tab → Time Synchronization tab.

3.

Set Interval .

32.8.1.4

Defining the Hours of Nighttime

To define the nighttime, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password

→ select .

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 32 - 19

2.

Select the Time tab → Time Synchronization tab.

3.

Select From and To to define the beginning and end of Nighttime.

32.8.2

Enabling Auto Daylight Savings Time

By default, daylight savings time is disabled. If this function is needed, you need to manually enable the daylight savings time. To auto start the daylight savings time, follow this procedure:

1.

Select the Main Menu quick key

→ from the System column select Maintenance

input the required password → select .

2.

Select the Time tab → Daylight Savings Time tab..

3.

Switch on Auto Daylight Savings Time .

4.

Adjust daylight savings time settings as necessary.

NOTE

If you enable Auto Daylight Savings Time you cannot manually switch it on or off from the System

Time dialog. For more information, see 3.8.2 Setting the Date and Time.

32.9

Viewing Version Information

To view the system software version, module hardware and software version, and firmware version , follow this procedure:

1.

Select the Main Menu quick key

→ from the System column select Maintenance

input the required password

→ select .

2.

Select the Version tab.

You can also view the system software version by selecting the Main Menu quick key → from the System column select Version .

32.10

Checking Battery Information

To check the battery information, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password

→ select .

2.

Select the Battery Information tab.

32.11

Changing Secondary Display Settings (for N22/N19)

You can separately set the orientation, brightness, alarm indications, and the display contents of the secondary display.

NOTE

If you change the Maintenance dialog settings for the secondary display, you need to restart the monitor for the changes to take effect.

If you plug in a secondary display when the monitor is on, you need to restart the monitor before you use the secondary display.

32.11.1

Configuring Secondary Display Screen Contents

(for N22/N19/N17)

The contents of the secondary display is configurable. To do so, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the Display tab.

3.

Set Screen Contents .

32 - 20 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

◆ Mirrored: The contents of the secondary display is exactly the same with the primary display. The orientation of the secondary display is also the same with the primary display. It is not available for the N17 monitor.

◆ Independent : You can separately configure the contents and layout of the primary display and secondary display. The independent secondary display cannot share the mouse and keyboard with the primary display. For N22/N19, separate mouse and keyboard connected to the MSB connectors of the secondary display is required.

Extended : You can separately configure the contents and layout of the primary display and secondary display. The extended secondary display shares the mouse and keyboard with the primary display. You cannot use separate mouse and keyboard to operate the extend secondary display.

4.

If the Screen Contents is set to Extended , set Secondary Screen Location .

32.11.2

Enabling Secondary Display Alarm Indications (for N22/N19)

The secondary display can provide visual and audible indications. This function is disabled by default. To enable the secondary display to present alarm light and alarm sound, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the Display tab.

3.

Switch on Alarm Sound/Light .

32.12

Changing Scanner Settings

For the 2D barcode reader, you can change the scanner settings.

32.12.1

Establish the Relationship between Patient Demographics and 2D Barcode

Fields (for the Mindray Custom 2D Barcode Reader)

To configure 2D barcode reader, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

select .

2.

Select the Scanner tab

2D Barcode tab.

3.

Establish the relationship between the monitor data and hospital’s barcode data for selectable patient demographics. For example, the monitor has an option of Ped for patient category. In your hospital barcode, the text may read as Pediatric . So input Pediatric for the field Ped to establish their relationship.

32.12.2

Setting the Barcode Reader Information

To set the barcode reader information, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the Scanner tab → Scanner Info.

tab.

3.

Set Scanner Type :

◆ 1D Scanner : select this option when using a 1D scanner or a 2D scanner other than the Mindray custom 2D scanner.

◆ 2D Scanner : select this option when using the Mindray custom scanner.

NOTE

When you set Scanner Type to 2D Scanner, default settings are applied to Data Encoding Type and

Data Parse Mode. You do not need to change these setting.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 32 - 21

32.12.3

Identifying the Barcode Reader (for the non-Mindray Custom 2D Barcode

Reader)

When you are using barcode readers other than the JADAK Flexpoint HS-1R or HS-1M, you should select the barcode reader from the USB device list, so that the monitor can identify the barcode reader. To do so, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the Scanner tab → Identify Scanner tab.

3.

From the USB device list, select the barcode reader you are using.

32.12.4

Selecting Patient Information Read by the Barcode Reader (for the Mindray

Custom 2D Barcode Reader)

To select desired patient information that will be read by the barcode reader, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

select .

2.

Select the Scanner tab

Field tab.

3.

Select the patient information. The selected information can be read by the barcode reader.

NOTE

The patient information read by the barcode reader is dependent on the patient information encoded in the hospital's barcodes.

32.13

Configuring the Network

You can connect the monitor to the central monitoring system (CMS), eGateway, and other monitors through wired LAN or wireless LAN.

32.13.1

Network Safety Information

CAUTION

Wireless network designing, deploying, debugging, and maintenance should be executed by

Mindray service personnel or authorized technicians.

Always set the wireless network according to local wireless regulations.

Using 5G frequency band is recommended whenever possible. There are more interference sources in 2.4G frequency band.

Private APs and wireless routers are not allowed. These devices may cause radio interference and result in monitor and CMS data loss.

32 - 22 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Data communication must be performed within a closed network or within a virtually isolated network provided by a hospital for all network functions. The hospital is responsible for ensuring the security of the virtually isolated network.

WPA2-PSK and WPA2-Enterprise verification and encryption should be used if possible. Otherwise, the equipment may not be able to work or patient information may be divulged. WPA2-Enterprise and a long password are recommended.

Keep network authentication information, for example passwords, safe, to protect the network from being accessed by unauthorized users.

Do not connect non-medical devices to the monitor network.

If wireless network signal is poor, there may be a risk of CMS-destined data loss.

Maximum number of monitors connected to a single AP is 16 for N17/N15/N12 or 12 for N22/N19.

Too many monitors connected to the same AP may result in network disconnection.

RF interference may result in wireless network disconnection.

Disconnecting from the network may result in CMS-destined data loss and function failure. Check the patient in case of network disconnection and solve the network problem as soon as possible.

Ensure that the monitor IP address setting is correct. Changing the network settings may result in network disconnection. Contact your service personnel if there are any problems with the IP address.

If the device remains in service following the end of support, the cybersecurity risks for end-users can be expected to increase over time.

Network disconnection does not have an impact on local monitoring functions.

Contact Mindray service personnel if cybersecurity problems occur.

32.13.2

Selecting a Network Type

To select network type, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the Network Type tab.

3.

Set Monitor to Auto , LAN1 IP , or WLAN according to your network type. The default is Auto , which means the monitor will automatically use LAN1 IP when detected and the available WLAN when LAN1 IP is not detected..

32.13.3

Setting the Wired Network

To set the wired network, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the Network Setup tab → LAN1 IP tab.

3.

Select how to get the IP address.

◆ Obtain IP Address Automatically : the monitor automatically gets the IP address.

◆ Use the following Address : you need to input the IP address , Subnet mask , and Gateway .

4.

Select how to get the DNS address:

◆ Obtain DNS Address Automatically : the monitor automatically gets the DNS address.

◆ Use the following DNS Address : you need to input the IP address of Preferred DNS server and

Alternate DNS server .

32.13.4

Setting the Wireless Network

To add and set the wireless network, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the Network Setup tab → WLAN tab.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 32 - 23

3.

Select Add WLAN . In the popup dialog, input network name, SSID, and password. Set Security .

4.

Select the WLAN IP tab and select how to get the IP address and DNS address.

◆ Obtain IP Address Automatically : the monitor automatically gets the IP address.

◆ Use the Following Address : you need to input the IP address , Subnet Mask , and Gateway .

◆ Obtain DNS Address Automatically : the monitor automatically gets the DNS address.

◆ Use the following DNS Address : you need to input the IP address of Preferred DNS server and

Alternate DNS server .

5.

Select the WLAN Setup tab and set WLAN band and channels. The default is Auto , which means the monitor can automatically identify the WLAN band.

After configuring the wireless network, select Network Test to verify that the wireless network can be properly connected.

If you need to change network settings, select beside the desired wireless network. Selecting deletes the wireless network.

32.13.5

Managing Certifications

You can delete certifications from the monitor, or import certifications from the USB memory device. To do so, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the Network Setup tab → WLAN tab.

3.

Select the Certificate Management button from the bottom left corner of the dialog.

◆ From the Local tab, select certifications you want to delete from the monitor, and then select Delete .

◆ From the USB Drive tab, select certifications you want to import from the USB memory device, and then select Import .

32.13.6

Enabling Selecting a CMS

To enable selecting a CMS, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

select .

2.

Select the Network Setup tab

Central Station Setup tab.

3.

Switch on Select CMS .

32.13.7

Adding CMSs

You can add up to 30 central monitoring systems (CMS) for your monitor. To add CMSs, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the Network Setup tab → Central Station Setup tab.

3.

Select Add Central Station , and then input the CMS name and IP address of this department.

32.13.8

Setting Multicast Parameters

Multicast helps device discovery between monitors and other devices on the network. Devices in the same multicast group can be mutually discovered.

To set multicast parameters, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the Network Setup tab

Device Discover tab.

3.

Set Multicast TTL .

4.

Set Multicast Address .

32 - 24 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

32.13.9

Setting the Master Server Address

To set the DNS name of the master server, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the Network Setup tab → Device Discover tab.

3.

Select Master Server Address to input the DNS name or IP address of the master server.

32.13.10 Setting the Network Service Quality Level

To set the quality of service (QoS), follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the Network Setup tab

QoS tab.

3.

Select QoS Level For Realtime Monitoring to set the service quality of network connection for realtime monitoring, for example parameter measurements and waveforms, alarms, etc.

4.

Select QoS Level For Others to set the service quality of network connection for non-realtime monitoring, for example history data, printing, etc.

32.13.11 Using the ADT Gateway

The ADT (admit-discharge-transfer) gateway is normally deployed in the eGateway. You can obtain patient information from the hospital ADT server through the ADT gateway.

To configure the ADT gateway, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the Network Setup tab

ADT tab.

3.

Select Server Address to input the host name or IP address of the ADT gateway.

4.

Select Port to input the port of the ADT gateway.

ADT Query is switched off by default. You can load patient information to the monitor from the ADT server only when this function is enabled.

After configuring the ADT gateway, select Network Test to verify that the ADT server is properly connected.

32.13.12 Sending Realtime Data, Waveforms and Alarms via HL7 Protocol

You can send the realtime data, waveforms, and alarms from the monitor to the hospital servers via HL7 protocol. To do so, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the Network Setup tab

HL7 Configuration tab.

3.

From the Data + Waveforms area, select Server Address to input the name or IP address for the server receiving the realtime data and waveform. Set Port .

4.

Switch on Send Data and Send Waveform , as needed. Set Data Interval .

5.

From the Alarms area, select Server Address to input the name or IP address for the server receiving the alarm data. Set Port .

6.

Switch on Send Alarms as needed.

7.

From the Compatibility area, set HL7 Protocol Version .

This page also displays the server connection status.

NOTE

Licenses are required for sending realtime data, waveforms, and alarms via HL7. For more

information on checking the licenses, see 26.1 Checking Software Licenses .

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 32 - 25

32.13.13 Selecting Data Encryption Type

To set data encryption type, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the Network Setup tab → Information Security tab.

3.

Select Encryption Connection Type to set data encryption type when connecting devices:

◆ Only Private Encryption : Mindray private encryption is used to encrypt the transmitted data. You cannot connect devices supporting SSL (secure sockets layer) encryption.

◆ SSL Encryption Priority : for devices supporting SSL encryption, SSL encryption is used when connecting the devices. For devices not supporting SSL encryption, private encryption is used when connecting the devices.

32.13.14 Disabling the Broadcasting Patient Demographics Function

When viewing other patients, device location and patient information of remote devices are displayed in the remote device list by default. To protect patient privacy, you can disable the broadcasting of patient demographics function to prevent the monitor from sending patient information via broadcast. To do so, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select Network Setup tab → Information Security tab.

3.

Switch off Broadcast Patient Demographics .

Thus, patient information will not display in the remote device list.

32.14

MLDAP

MLDAP refers to Mindray LDAP (Lightweight Directory Access Protocol). It is an independent process which can be installed on the eGateway or other application server (Windows). MLDAP provides user identity and authentication.

The MLDAP server is connected with the hospital LDAP server. All monitoring devices are connected to the

MLDAP server to implement identity and authentication for the following operations:

Changing alarm settings

Changing arrhythmia settings

Accessing the Maintenance dialog

32.14.1

Setting MLDAP

To access the MLDAP server, you should set your monitor as follows:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the Network Setup tab

MLDAP tab.

3.

Set Server address and port of the MLDAP server.

32.14.2

Testing MLDAP Server Connection

To test whether the monitor is properly connected with the MLDAP server, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the Network Setup tab

MLDAP tab.

3.

Select Network Test .

32 - 26 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

32.14.3

Selecting Password for User Authentication

You can select what password is used when changing alarm settings, arrhythmia settings, viewing current or discharged patient data, and accessing the Maintenance dialog. To do so, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the Authorization Setup tab.

3.

From the Maintenance area, set User Maintenance to select the password for accessing the monitor’s

Maintenance dialog.

◆ Local Password : the monitor’s password for accessing the Maintenance dialog is required.

◆ User Password : the user name and password saved in the MLDAP server are required.

4.

From the Clinical Setting area, set passwords for Alarm Setup , Arrhythmia , View Discharged Patients , and Viewing Patient Review Data .

◆ No Password : changing alarm settings and arrhythmia settings, reviewing data of the current patient and discharged patients are not password protected.

◆ Local Password : changing alarm settings, arrhythmia settings, and reviewing data of the current patient are password protected. The clinical password is required.

◆ User Password : changing alarm settings, arrhythmia settings, and reviewing data of discharged patients are password protected. The user name and password saved in the MLDAP server are required.

From the Maintenance area, selecting Modify Local Password can change the monitor’s password for accessing the Maintenance dialog.

From the Clinical Setting area, selecting Modify Local Password can change the monitor’s password for accessing alarm settings, arrhythmia settings, and reviewing data of the current patient and discharged patients.

32.14.4

Setting the Password Timeout Period

If you use the password saved in the MLDAP server to access the Maintenance dialog, alarm settings and arrhythmia settings, you can set the password timeout period. Once the timeout period is reached, you will need to re-enter the password. To do so, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the Authorization Setup tab.

3.

Set Automatic Logout Time .

32.15

32.16

Defining Other Functions

32.16.1

Setting Notch Filter Frequency

Set notch filter frequency according to the power line frequency of your country. To set notch filter frequency, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the Other tab.

3.

Set Notch Frequency to 50 Hz or 60 Hz according to the power line frequency.

32.16.2

Setting Mouse Sensitivity

The mouse sensitivity is adjustable. To do so, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 32 - 27

2.

Select the Other tab.

3.

Set Mouse Sensitivity .

32.16.3

Setting the Manual Event Edit Switch

You can set whether selecting and editing the name of a manual event is allowed. To do so, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the Other tab.

3.

Switch on or off Manual Event Edit .

32.16.4

Setting the SpO

2

Tone Mode

The monitor adjusts the QRS tone (pitch tone) according to the SpO

2

values. To set the SpO

2

tone mode, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password

→ select .

2.

Select the Other tab.

3.

Set SpO2 Tone to Mode 1 or Mode 2 . The pitch changes more rapidly in Mode 2 than in Mode 1, as saturation levels change.

The SpO

2

tone mode chart for SpO

2

between 40% to 100% is as follows:

CAUTION

It is highly recommended to use the same SpO

2

tone mode for the same monitors in a single area to prevent issues arising from inconsistent operation.

32.16.5

Selecting the Language

To set the user interface (UI) language, follow this procedure:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the Other tab

3.

Set Language .

32 - 28 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

32.16.6

Switching On or Off a Parameter

If there is no Parameters On/Off tab in the Screen Setup dialog then setting parameter switches is password protected. Follow this procedure to set parameter switches:

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

2.

Select the Other tab.

3.

Select Parameters On/Off .

4.

Enable or disable desired parameters.

When a parameter is switched off, the monitor stops data acquisition and alarming for this measurement.

NOTE

When a parameter is manually switched off and the corresponding parameter module is plugged in, you cannot monitor this parameter.

32.16.7

Setting If Setting Parameter Switches is Protected

You can set whether setting parameter switches is password protected. To do so, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password

→ select .

2.

Select the Other tab.

3.

Switch on or off Parameters On/Off Protected .

32.16.8

Setting If Parameter Switch is Influenced by Configuration

You can set if the settings of parameter switches are influenced by loading a configuration. To do so, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password

→ select .

2.

Select the Other tab.

3.

Switch on or off Parameters On/Off Config Influenced .

32.16.9

Setting Whether the Monitor Stands by if MPM module is Removed

You can set whether the monitor enters the Standby mode if the MPM module, including the N1 and T1, is removed from the monitor. To do so, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the Other tab.

3.

Set Auto Standby after MPM Module Removed . If Off is selected, the monitor does not enter the standby mode if the MPM module is removed.

32.16.10 Browsing System Log

To browse the system log, follow this procedure:

1.

Select the Main Menu quick key

from the System column select Maintenance

input the required password → select .

2.

Select the Other tab.

3.

Select Browse System Log to enter the System Log page.

4.

Select the log classifications you want to view, and then select Search . To view logs of certain date and time, select Jump To and define the date and time.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 32 - 29

32.16.11 Exporting System Log

To export the system log, follow this procedure:

1.

For N17/N15/N12, connect the USB drive to the monitor’s USB connector. For N22/N19, connect the USB drive to the monitor’s MSB connector.

2.

Select the Main Menu quick key → from the System column select Maintenance → input the required password

→ select .

3.

Select the Other tab.

4.

Select Export System Log .

32 - 30 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

33

Battery

33.1

Battery Introduction

This monitor is designed to operate on battery power when the mains power is not available. The monitor uses mains power as primary power source. In case of mains power failure, the monitor automatically runs on the battery power.

NOTE

If the mains power fails and the monitor runs on battery power, the display brightness automatically lowers to the value of Brightness on Battery (set in Screen Setup

Display). You can manually adjust the display brightness as required.

33.2

Battery Safety Information

WARNING

Keep batteries out of the reach of children.

Use only Mindray specified batteries. Use of a different battery may present a risk of fire or explosion.

Keep the batteries in their original package until you are ready to use them.

Do not expose batteries to liquid.

Do not crush, drop or puncture the battery. Mechanical abuse can lead to internal damage and internal short circuits. If a battery has been dropped or banged against a hard surface, whether damage is externally visible or not, remove the battery from use and dispose of it properly.

If the battery shows signs of damage or signs of leakage, replace it immediately. Use caution in removing the battery. Avoid contacting the leakage.

Batteries should be charged only in this monitor.

Extremely high ambient temperature may cause battery overheat protection, resulting in monitor shutdown.

The lithium-ion battery has a service life of two years (for N22/N19) or three years (for N17/N15/

N12). Replace the battery when it reaches the end of its service life. Failure to replace the battery may cause serious damage to the equipment from battery overheating.

Do not open batteries, heat batteries above 60 °C, incinerate batteries, or short battery terminals.

They may ignite, explode, leak or heat up, causing personal injury.

CAUTION

Remove the battery before shipping the monitor or when it will not be used for an extended period of time.

When the monitor runs on battery power, the monitor may automatically shutdown due to high power consumption if too many external modules are connected.

For N22/N19, the battery run time is about one hour. Connect the mains power as long as the mains power is available.

Use only the AC mains to power the monitor when the iView system is in use.

33.3

Installing or Replacing the Battery

No battery is installed when the monitor is shipped from the factory.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 33 - 1

For N22/N19, the battery must only be installed or replaced by service personnel trained and authorized by

Mindray.

For N12, to install the battery, follow this procedure:

1.

Turn off the monitor. Disconnect the power cable and other cables.

2.

Place the monitor on the worktable with monitor face down. Ensure that the worktable is free of dirt and debris where the monitor face will rest to reduce risk of damage to the monitor screen.

3.

If the monitor is installed on a mounting arm, remove the monitor from the mounting arm and remove any mounting plates.

4.

Open the battery door as indicated by Figure 1 below.

5.

Turn the latch aside.

6.

Insert the battery into the battery compartment with the battery terminal inwards. Turn the latch back to the middle as indicated by Figure 2 below. To replace the battery, remove the old battery and insert a new one.

7.

Close the battery door.

Figure 1

Figure 2

For N17/N15, to install the battery, follow this procedure:

1.

Turn off the monitor. Disconnect the power cable and other cables.

2.

Place the monitor on the worktable with monitor face down.

3.

If the monitor is installed on a mounting arm, remove the monitor from the mounting arm and remove any mounting plates.

4.

Pull up the battery door to open the battery compartment as indicated by Figure 1 below.

5.

Insert the battery into the battery compartment. Push the battery downwards till the battery terminal is plugged into the battery connector as indicated by Figure 2 below. To replace the battery, remove the old battery and insert a new one.

6.

Close the battery door.

Figure 1

Figure 2

33.4

Battery Indications

The battery LED, on-screen battery power indicator, and related alarm messages indicate the battery status.

33.4.1

Battery LED

The battery LED indications are as follows:

33 - 2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Green: the battery is fully charged.

Yellow: the battery is being charged.

Flashing green: the monitor is running on battery power.

Off: no battery is installed, or the battery malfunctioned, or the AC mains is not connected when the monitor is powered off.

33.4.2

Battery symbols

The on-screen battery symbols indicate the battery status as follows:

■ indicates that the battery is working correctly. The green portion represents the remaining charge.

indicates that the battery power is low and needs to be charged.

indicates that the battery is almost depleted and needs to be charged immediately. Otherwise, the monitor will automatically shut down soon.

indicates that the battery is being charged.

indicates that no battery is installed or the battery has failed.

33.4.3

Battery-related Alarms

The capacity of the battery is limited. When the battery level is low, the monitor presents the Low Battery alarm, the alarm lamp flashes, and the monitor produces an alarm sound.

If the battery is almost depleted, the monitor presents the Critically Low Battery alarm. In this case, immediately connect the AC mains to power the monitor and charge the battery. Otherwise, the monitor will automatically shut down soon.

For N22/N19, if the battery is aged, the Battery Service Required alarm is presented each time the monitor is turned on, indicating that the battery has reached its lifetime.

For more information on battery-related alarms, see Appendix D Alarm Messages .

33.5

Charging the Battery

The battery is recharged automatically when the monitor is connected to AC mains power.

33.6

Maintaining the Battery

33.6.1

Conditioning the Battery

The performance of batteries deteriorates over time. For N22/N19, you should condition the batteries periodically or when you notice a degradation in performance. If the battery is not conditioned for a prolonged time, its charge indication may not be accurate and you may wrongly evaluate the remaining battery runtime.

To condition a battery, follow this procedure:

1.

Disconnect the monitor from the patient and stop all monitoring and measuring procedures.

2.

Turn off the monitor, and connect the monitor to the external power source.

3.

Allow the battery to be charged uninterruptedly until it is fully charged.

4.

Disconnect the monitor from the external power source, and turn on the monitor.

5.

Allow the monitor to run on the battery until the battery is completely depleted and the monitor automatically shuts down.

6.

Fully charge the battery again for use or charge it to 40 - 60% for storage.

NOTE

Do not use the monitor to monitor a patient during battery conditioning.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 33 - 3

Do not interrupt battery conditioning. Interrupting battery conditioning to use the monitor will prevent completion of the conditioning cycle and require repeating the conditioning process completely to properly condition the battery.

33.6.2

Checking Battery Performance

The performance of a rechargeable battery deteriorates over time. You should check the battery performance periodically or if you are concerned that the battery may fail.

See steps 1 to 5 of 33.6.1 Conditioning the Battery

to check battery performance. The operating time of the batteries reflects their performance directly. If the operating time of a battery is noticeably shorter than that stated in the specifications, the battery may have reached its service life or be malfunctioning. If the battery performance meets the requirement, fully charge the battery again for use or charge it to 40 - 60% for storage.

NOTE

Battery operating time depends on equipment configuration and operation. For example, high display brightness or measuring NIBP repeatedly will shorten the battery operating time.

33.7

Storing Batteries

When storing batteries, make sure that the battery terminals do not come into contact with metallic objects. If batteries are stored for an extended period of time, place the batteries in a cool place with a partial charge of

40% to 60% capacity.

Condition the stored batteries every three months. For more information, see 33.6.1 Conditioning the Battery .

NOTE

Remove the battery from the equipment if the equipment is not used for a prolonged time (for example, several weeks). Otherwise the battery may overdischarge.

Storing batteries at high temperature for an extended period of time will significantly shorten their life expectancy.

Storing batteries in a cool place can slow the aging process. Ideally the batteries should be stored at

15 °C.

33.8

Recycling Batteries

Discard a battery in the following situations:

The battery has visual signs of damage.

The battery fails.

The battery is aged and its runtime significantly less than the specification.

The battery service life is reached.

Properly dispose of batteries according to local regulations.

WARNING

Do not open batteries, heat batteries above 60 °C, incinerate batteries, or short the battery terminals. They may ignite, explode, leak or heat up, causing personal injury.

33 - 4 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

34

Care and Cleaning

34.1

Care and Cleaning Introduction

In this chapter we only describe cleaning and disinfection of the main unit, modules, satellite module rack (SMR) and certain accessories. For the cleaning and disinfection of other reusable accessories, refer to their instructions for use of the corresponding accessories.

Keep your equipment and accessories clean. To avoid damage to the equipment, follow these guidelines:

Always follow the manufacturer’s instructions for each cleaning / disinfecting agent.

Do not immerse any part of the equipment or accessories into liquid.

Do not pour liquid onto the equipment or accessories.

Do not allow liquid to enter the case.

Never use abrasive materials (such as steel wool or silver polish), or erosive cleaners (such as acetone or acetone-based cleaners).

34.2

Care and Cleaning Safety Information

WARNING

Be sure to turn off the system and disconnect all power cables from the outlets before cleaning the equipment.

CAUTION

If you spill liquid on the equipment or accessories, contact Mindray or your service personnel.

Avoid wetting the pins and metal parts of the equipment, mounting kits or accessories during cleaning and disinfection.

Use only Mindray approved cleaners and disinfectants and methods listed in this chapter to clean or disinfect your equipment. Warranty does not cover damage caused by unapproved substances or methods.

We make no claims regarding the efficacy of the listed chemicals or methods as a means for controlling infection. For the method to control infection, consult your hospital’s Infection Control

Officer or Epidemiologist.

The responsible hospital or institution shall carry out all cleaning and disinfection procedures specified in this chapter.

Refer to the respective instructions for use of the cleaning agents and disinfectants.

Do not mix disinfecting solutions (such as bleach and ammonia) as hazardous gases may result.

NOTE

To clean or disinfect reusable accessories, refer to the instructions delivered with the accessories.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 34 - 1

34.3

Cleaning and Disinfecting the Equipment and Mounting Kits

Use approved cleaning and disinfecting agents for cleaning or disinfecting the main unit, parameter modules, and SMR.

34.3.1

Approved Cleaning and Disinfecting Agents

The following table lists approved cleaning and disinfecting agents:

:

Product Name Product Type Active Ingredients

Liquid Sodium hypochlorite bleach

Hydrogen peroxide

Isopropanol

1-Propanol

Metrex CaviCide1™

Virex® II 256 (1:256)

Virex® TB

Rely+OnTM Virkon® High Level

Surface Disinfectant

Alpet® D2 Surface Sanitizing

Wipes

Powder

Wipes

Sodium hypochlorite bleach 0.5%

Hydrogen peroxide 3%

Isopropanol 70%

1-Propanol 50%

Diisobutylphenoxyethoxyethyl dimethyl benzyl ammonium chloride 0.28%,

Isopropanol 17.2%

Didecyl dimethyl ammonium chloride 8.704%, n-Alkydimethyl benzyl ammonium chloride 8.190% n-Alkyl dimethyl benzyl ammonium chlorides 0.105%, n-Alkyl dimethyl ethylbenzyl ammonium chlorides 0.105%

Used as 1% solution

Biocidal active:

Pentapotassium bis (peroxymonosulphate) bis (sulphate)(500g/ kg),

Contains dipotassium peroxodisulphate.

Isopropyl Alcohol 58.6000%,

Octyl Decyl Dimethyl Ammonium chloride 0.0075%,

Dioctyl Dimethyl Ammonium Chloride 0.0030%

Sodium Hypochlorite 0.65% Clorox Dispatch® Hospital

Cleaner Disinfectant Towels with

Bleach

Clorox Healthcare® Bleach

Germicidal Wipes

Clorox Healthcare® Hydrogen

Peroxide Cleaner Disinfectant

Wipes

Diversey Oxivir® TB Wipes

Metrex CaviWipes™

Sodium Hypochlorite 0.55%

Hydrogen Peroxide 1.4%

Hydrogen Peroxide 0.5%

Diisobutylphenoxyethoxyethyl dimethyl benzyl ammonium chloride 0.28%,

Isopropanol 17.2% n-Alkyl dimethyl ethylbenzyl ammonium chlorides 0.14%, n-Alkyl dimethyl benzyl ammonium chlorides 0.14%

Sodium Hypochlorite 0.63%,other ingredients 99.37%

PDI Sani-Cloth® AF3 Germicidal

Disposable Wipe

PDI Sani-Cloth® Bleach

Germicidal Disposable Wipe

PDI Sani-Cloth® HB Germicidal

Disposable Wipe

PDI Sani-Cloth® Plus Germicidal

Disposable Cloth

PDI Super Sani-Cloth® Germicidal

Disposable Wipe n-Alkyl dimethyl ethylbenzyl ammonium chlorides 0.07%, n-Alkyl dimethyl benzyl ammonium chlorides 0.07% n-Alkyl dimethyl ethylbenzyl ammonium chlorides 0.125%, n-Alky dimethyl benzyl ammonium chlorides 0.125% n-Alkyl dimethyl ethylbenzyl ammonium chlorides 0.25%, n-Alkyl dimethyl benzyl ammonium chlorides 0.25%,

Isopropyl Alcohol 55.0%

34 - 2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Product Name

VIRAGUARD Hospital Surface

Disinfectant Towelettle

Product Type

Wipes

Active Ingredients

Isopropanol 70%,

Other ingredients 30%

NOTE

For equipment with the symbol , all the listed cleaning and disinfecting agents are available for use.

34.3.2

Cleaning the Equipment and Mounting Kits

The main unit, parameter modules, and SMR should be cleaned on a regular basis. Before cleaning the equipment and mounting kits, consult your hospital’s regulations for cleaning the equipment.

To clean your equipment, follow this procedure:

1.

Shut down the monitor and disconnect it from the AC power.

2.

Clean the display screen with wipes or a soft cloth moistened with one of the cleaning agents listed in

34.3.1 Approved Cleaning and Disinfecting Agents only.

3.

Clean the exterior surface of the equipment, module, and SMR with wipes or a soft cloth moistened with

one of the cleaning agents listed in 34.3.1 Approved Cleaning and Disinfecting Agents

only.

4.

Wipe off all the cleaning agent residue with a dry cloth.

5.

Dry your equipment in a ventilated, cool place.

34.3.3

Disinfecting the Equipment and Mounting Kits

Disinfect the main unit, parameter modules, and SMR as required in your hospital’s servicing schedule using the disinfecting agents listed in the table above. Cleaning equipment before disinfecting is recommended.

CAUTION

Never use EtO or formaldehyde for disinfection.

34.4

Cleaning and Disinfecting the Accessories

For the NIBP air hose, SpO

2

cable, and NMT accessories, you should clean and disinfect them using the cleaners and disinfectants and methods listed in this section. For other accessories, you should consult the instructions delivered with the accessories.

CAUTION

Fluids entering the NIBP air hose can damage the equipment. When cleaning and/or disinfecting the

NIBP air hose, prevent liquid from entering the hose.

Periodically inspect the NIBP air hose and connector for signs of wear or deterioration. Replace the

NIBP air hose if you detect a leak. Dispose of damaged NIBP air hose according to local laws for disposal of hospital waste.

Never immerse or soak the accessories in any liquid.

Never clean or disinfect the metalic connectors at either end of the accessories.

Use only Mindray approved cleaners and disinfectants and methods listed in this section to clean or disinfect the accessories. Warranty does not cover damage caused by unapproved substances or methods.

To avoid long term damage, the accessories should be disinfected only when necessary as determined by your hospital’s policy.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 34 - 3

34.4.1

Approved Accessories Cleaning and Disinfecting Agents

The following table lists approved NIBP air hose cleaning and disinfecting agents:

:

Product Name Product Type Active Ingredients

Isopropanol

1-Propanol

Metrex CaviCide1™

Virex® TB

Rely+OnTM Virkon® High Level

Surface Disinfectant

Alpet® D2 Surface Sanitizing

Wipes

Liquid

Powder

Wipes

Isopropanol 70%

1-Propanol 50%

Diisobutylphenoxyethoxyethyl dimethyl benzyl ammonium chloride 0.28%,

Isopropanol 17.2% n-Alkyl dimethyl benzyl ammonium chlorides 0.105%, n-Alkyl dimethyl ethylbenzyl ammonium chlorides 0.105%

Used as 1% solution

Biocidal active:

Pentapotassium bis (peroxymonosulphate) bis (sulphate)(500g/ kg),

Contains dipotassium peroxodisulphate.

Isopropyl Alcohol 58.6000%,

Octyl Decyl Dimethyl Ammonium chloride 0.0075%,

Dioctyl Dimethyl Ammonium Chloride 0.0030%

Sodium Hypochlorite 0.65% Clorox Dispatch® Hospital

Cleaner Disinfectant Towels with

Bleach

Metrex CaviWipes™

PDI Sani-Cloth® AF3 Germicidal

Disposable Wipe

PDI Sani-Cloth® Plus Germicidal

Disposable Cloth

PDI Super Sani-Cloth® Germicidal

Disposable Wipe

VIRAGUARD Hospital Surface

Disinfectant Towelettle

Diisobutylphenoxyethoxyethyl dimethyl benzyl ammonium chloride 0.28%,

Isopropanol 17.2% n-Alkyl dimethyl ethylbenzyl ammonium chlorides 0.14%, n-Alkyl dimethyl benzyl ammonium chlorides 0.14% n-Alkyl dimethyl ethylbenzyl ammonium chlorides 0.125%, n-Alky dimethyl benzyl ammonium chlorides 0.125% n-Alkyl dimethyl ethylbenzyl ammonium chlorides 0.25%, n-Alkyl dimethyl benzyl ammonium chlorides 0.25%,

Isopropyl Alcohol 55.0%

Isopropanol 70%,

Other ingredients 30%

:

The following table lists approved Masimo SpO

2

cable cleaning and disinfecting agents:

Product Name Product Type Active Ingredients

Isopropanol Liquid Isopropanol 70%

34 - 4 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

:

The following table lists approved Nellcor SpO

2

cable cleaning and disinfecting agents:

Product Name Product Type Active Ingredients

Liquid Sodium hypochlorite bleach

Isopropanol

1-Propanol

Virex® TB

Rely+OnTM Virkon® High Level

Surface Disinfectant

Powder

Sodium hypochlorite bleach 0.5%

Isopropanol 70%

1-Propanol 50% n-Alkyl dimethyl benzyl ammonium chlorides 0.105%, n-Alkyl dimethyl ethylbenzyl ammonium chlorides 0.105%

Used as 1% solution

Biocidal active:

Pentapotassium bis (peroxymonosulphate) bis (sulphate)(500g/ kg),

Contains dipotassium peroxodisulphate.

Sodium Hypochlorite 0.65% Clorox Dispatch® Hospital

Cleaner Disinfectant Towels with

Bleach

Clorox Healthcare® Bleach

Germicidal Wipes

Clorox Healthcare® Hydrogen

Peroxide Cleaner Disinfectant

Wipes

Diversey Oxivir® TB Wipes

PDI Super Sani-Cloth® Germicidal

Disposable Wipe

Wipes

VIRAGUARD Hospital Surface

Disinfectant Towelettle

Sodium Hypochlorite 0.55%

Hydrogen Peroxide 1.4%

Hydrogen Peroxide 0.5% n-Alkyl dimethyl ethylbenzyl ammonium chlorides 0.25%, n-Alkyl dimethyl benzyl ammonium chlorides 0.25%,

Isopropyl Alcohol 55.0%

Isopropanol 70%,

Other ingredients 30%

:

The following table lists approved NMT accessory cleaning and disinfecting agents:

Product Name Product Type Active Ingredients

Liquid Sodium hypochlorite bleach

Isopropanol

1-Propanol

Metrex CaviCide1™

Virex® TB

Rely+OnTM Virkon® Powder

(Used as 1% solution)

Powder

Sodium hypochlorite bleach 0.5%

Isopropanol 70%

1-Propanol 50%

Diisobutylphenoxyethoxyethyl dimethyl benzyl ammonium chloride 0.28%,

Isopropanol 17.2% n-Alkyl dimethyl benzyl ammonium chlorides 0.105%, n-Alkyl dimethyl ethylbenzyl ammonium chlorides 0.105%

Used as 1% solution

Biocidal active:

Pentapotassium bis (peroxymonosulphate) bis (sulphate)(500g/ kg),

Contains dipotassium peroxodisulphate.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 34 - 5

Product Name

Alpet® D2 Surface Sanitizing

Wipes

Clorox Dispatch® Hospital

Cleaner Disinfectant Towels with

Bleach

Clorox Healthcare® Bleach

Germicidal Wipes

Clorox Healthcare® Hydrogen

Peroxide Cleaner Disinfectant

Wipes

Diversey Oxivir® TB Wipes

Metrex CaviWipes™

PDI Sani-Cloth® AF3 Germicidal

Disposable Wipe

PDI Sani-Cloth® Bleach

Germicidal Disposable Wipe

PDI Sani-Cloth® Plus Germicidal

Disposable Cloth

PDI Super Sani-Cloth® Germicidal

Disposable Wipe

VIRAGUARD Hospital Surface

Disinfectants

Product Type

Wipes

Active Ingredients

Isopropyl Alcohol 58.6000%,

Octyl Decyl Dimethyl Ammonium chloride 0.0075%,

Dioctyl Dimethyl Ammonium Chloride 0.0030%

Sodium Hypochlorite 0.65%

Sodium Hypochlorite 0.55%

Hydrogen Peroxide 1.4%

Hydrogen Peroxide 0.5%

Diisobutylphenoxyethoxyethyl dimethyl benzyl ammonium chloride 0.28%,

Isopropanol 17.2% n-Alkyl dimethyl ethylbenzyl ammonium chlorides 0.14%, n-Alkyl dimethyl benzyl ammonium chlorides 0.14%

Sodium Hypochlorite 0.63%,other ingredients 99.37% n-Alkyl dimethyl ethylbenzyl ammonium chlorides 0.125%, n-Alky dimethyl benzyl ammonium chlorides 0.125% n-Alkyl dimethyl ethylbenzyl ammonium chlorides 0.25%, n-Alkyl dimethyl benzyl ammonium chlorides 0.25%,

Isopropyl Alcohol 55.0%

Isopropanol 70%,

Other ingredients 30%

34.4.2

Cleaning the Accessories

You should clean the accessories (NIBP air hose, Masimo SpO

2

cable, Nellcor SpO

2

cable and NMT accessories) on a regular basis. Before cleaning the accessories, consult your hospital’s regulations for cleaning the accessories.

To clean the accessories, follow this procedure:

1.

Clean the accessories with wipes or a soft cloth moistened with one of the cleaning agents listed in

34.4.1 Approved Accessories Cleaning and Disinfecting Agents

only.

2.

Wipe off all the cleaning agent residue with a dry cloth.

3.

Allow the accessories to air dry.

34.4.3

Disinfecting the Accessories

We recommend that the accessories (NIBP air hose, SpO

2

cables, and NMT accessories) should be disinfected only when necessary as determined by your hospital’s policy, to avoid long term damage to the accessories.

Cleaning the accessories before disinfecting is recommended.

34.5

Sterilization

Sterilization is not recommended for this equipment, related products, accessories, or supplies unless otherwise indicated in the Instructions for Use that accompany the products, accessories or supplies.

34 - 6 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

35

Maintenance

35.1

Maintenance Introduction

Regular maintenance is essential to ensure that the equipment continues to function properly. This chapter contains information on periodic testing and maintenance.

35.2

Maintenance Safety Information

WARNING

To avoid electric shock, stop using the equipment if you find the housing is damaged, and contact the service personnel.

Failure on the part of the responsible individual hospital or institution using this equipment to implement a recommended maintenance schedule may cause undue equipment failure and possible health hazards.

No modification of this equipment is allowed.

This equipment contains no user serviceable parts.

Software upgrades to the monitor can only be done by Mindray service engineers with Mindray upgrading tools.

The safety checks or maintenance involving any disassembly of the equipment should be performed by professional service personnel. Otherwise, undue equipment failure and possible health hazards could result.

Do not open batteries, heat batteries to above 60 °C, incinerate batteries, or short the battery terminals. Batteries may ignite, explode, leak or heat up, causing personal injury.

The service personnel must be properly qualified and thoroughly familiar with the operation of the equipment.

CAUTION

The equipment and accessories shall not be serviced or maintained while in use on a patient.

If you discover a problem with any of the equipment, contact your service personnel or Mindray.

Use and store the equipment within the specified temperature, humidity, and altitude ranges.

When disposing of the packaging material, be sure to observe the applicable waste control regulations and keep it out of the reach of children.

At the end of its service life, the equipment, as well as its accessories, must be disposed of in compliance with the guidelines regulating the disposal of such products. If you have any questions concerning disposal of the equipment, please contact Mindray.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 35 - 1

35.3

Maintenance and Testing Schedule

Follow the maintenance and testing schedule or local regulations to perform testing and maintenance such as defined in section

35.5 Testing Methods and Procedures

. Follow your hospital policy on cleaning and disinfecting the equipment before testing and maintenance.

The following table lists the maintenance and testing schedule:

Test/Maintenance Item

Performance Tests

Visual inspection

Measurement module performance test and calibration

Analog output test

Defibrillation synchronization test

Nurse call test

Electrical Safety Tests

Electrical safety tests

Other Tests

Power-on test

NMT sensor check

Recorder check

Network printer tests

Device integration check

Battery check Functionality test

Performance test

Recommended Frequency

Every day, before first use.

1. If you suspect that the measurement values are incorrect.

2. Follow any repairs or replacement of relevant module.

3. Once a year for CO

2

and AG tests.

4. Once every two years for other parameter module performance tests.

If you suspect that the analog output function does not work properly.

If you suspect that the defibrillation synchronization function does not work properly.

If you suspect that the nurse call function does not work properly.

Once every two years.

Before use

Once a year.

1. When the recorder is used for the first time.

2.Follow any repair or replacement of the recorder.

1. When first installed.

2.Follow any repair or replacement of the printer.

1. When first installed.

2. Following any repair or replacement of the external device.

1. When first installed.

2. When battery is replaced.

Every three months or if the battery runtime is reduced significantly.

35.4

Checking System Software Version

To view system software version information, select the Main Menu quick key

from the System column select

Version .

You can check the detailed system software version, module hardware and software version, and firmware version only from the Maintenance dialog.

35.5

Testing Methods and Procedures

Except for the following maintenance tasks, all other test and maintenance tasks should be performed by

Mindray-qualified service personnel only.

Regular checks, including visual inspection and power-on test

NMT sensor check

35 - 2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Printer and recorder tests

Battery check

If your monitor needs a safety test and performance test, contact the service personnel.

35.5.1

Performing Visual Inspection

Visually inspect the equipment before its first use every day. If you find any signs of damage, remove the monitor from use and contact the service personnel.

Verify that the equipment meets the following requirements:

Environment and power supply specifications are met.

The monitor housing and display screen are free from cracks or other damage.

The power cord is not damaged and the insulation is in good condition.

Connectors, plugs, and cables are not damaged or kinked.

Power cord and patient cables are securely connected to the equipment and modules.

35.5.2

Performing Power-on Test

The monitor automatically performs a power-on test at startup. Verify the following items for the power-on test:

The equipment powers on properly.

The alarm system works properly, denoted by alarm sounds and display of all 3 alarm light colors.

The monitor displays properly.

35.5.3

Checking the NMT Sensor

NMT sensor check is required once a year or when you doubt the measured values.

To calibrate the NMT sensor,

1.

Select the Main Menu quick key → from the System column select Maintenance → input the required password → select .

2.

Select the Module tab → NMT tab.

3.

Follow the on-screen instructions to check the NMT sensor in four ways.

If sensor check completes successfully, the message “Test passed. The function of NMT sensor is OK” is presented. If any of the four steps fails, check if the sensor is placed correctly as instructed and perform the sensor check again. Replace the sensor or contact your service personnel if it cannot pass the sensor check.

NOTE

Stop NMT measurement or calibration before starting an NMT sensor check.

Take care when handling the NMT sensor. Avoid forceful strikes to the sensor.

35.5.4

Testing the Recorder

To test the recorder, follow this procedure:

1.

Start a recording task to print waveforms and reports.

2.

Check that the recorder functions correctly.

3.

Check that the printout is clear without missing dots.

35.5.5

Testing the Network Printer

To check the printer, follow this procedure:

1.

Start a printing task to print waveforms and reports.

2.

Check that the printer is properly connected and functions correctly.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 35 - 3

3.

Check that the printout is clear with no missing information.

35.5.6

Checking the Battery

For information on battery check, see 33.6.2 Checking Battery Performance

.

35.6

NIBP Maintenance

35.6.1

NIBP Leakage Test

The NIBP leakage test checks the integrity of the system and of the valve. The NIBP leakage test should be performed once every two years or when you doubt the NIBP measurements. The NIBP leakage test should be performed by Mindray-qualified service personnel only.

35.6.2

NIBP Accuracy Test

The NIBP accuracy test should be performed once every two years or when you doubt the NIBP measurements.

The NIBP accuracy test should be performed by Mindray-qualified service personnel only.

35.7

Disposing of the Monitor

Dispose of the monitor and its accessories when its service life is reached. Follow local regulations regarding the disposal of such product.

WARNING

For disposal of parts and accessories, where not otherwise specified, follow local regulations regarding disposal of hospital waste.

35 - 4 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

36

Accessories

The accessories listed in this chapter comply with the requirements of IEC 60601-1-2 when in use with the patient monitor. The accessory material that contacts the patients has undertaken the bio-compatibility test and is verified to be in compliance with ISO 10993-1. For details about the accessories, refer to the instructions for use provided with the accessory.

WARNING

Use accessories specified in this chapter. Using other accessories may cause damage to the patient monitor or not meet the claimed specifications.

Single-use accessories are not designed to be reused. Reuse may cause a risk of contamination and affect the measurement accuracy.

CAUTION

The accessories may not meet the performance specifications if stored or used outside the specified temperature and humidity ranges. If accessory performance is degraded due to aging or environmental conditions, contact your service personnel.

Check the accessories and their packages for any sign of damage. Do not use them if any damage is detected.

Use the accessories before the expiration date if an expiration date is indicated.

Discard disposable accessories according to hospital regulations.

NOTE

This manual describes all the accessories that are validated for use. Not all accessories are available in every market.

36.1

ECG Accessories

36.1.1

ECG Electrodes Available for Purchase

Model Part No.

31499224 0010-10-12304

1050NPSMKittycat 0681-00-0098-01

1051NPSMKittycat 0681-00-0098-02

Description

Electrode, Kendall, pkg of 10

NEO pre-wired electrode radio

Opaque, pkg of 100

NEO pre-wired electrode radio

Translucent, pkg of 100

36.1.2

ECG Electrodes Also Compatible

Model

2245-50

Part No.

9000-10-07469

Description

ECG electrode, 3M, pkg of 50

Usage

Disposable

Disposable

Disposable

Applicable patient

Adult

Neonate

Neonate

Usage

Disposable

Applicable patient

Pediatric

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 36 - 1

36.1.3

12-Pin Trunk Cables Available for Purchase

Model

EV6206

EV6207

EV6216

EV6217

EV6208

EV6202

EV6212

EV6222

Part No.

009-005266-00

009-005267-00

009-005268-00

009-005269-00

040-003528-00

0010-30-42720

0010-30-42724

040-000754-00

Description

ECG cable, 10' (3.1 m), defibrillation-proof, for N/T

ECG cable, 20' (6.2 m), defibrillation-proof, for N/T

ECG cable, 10' (3.1 m), ESU-proof, for N/T

ECG cable, 20' (6.2 m), ESU-Proof , for N/T

ECG cable, 10' (3.1 m), 12-lead, defibrillation-proof, for N/T 5-lead leadwires

12Pin 3-Lead ECG trunk cable, defibrillation-proof

12Pin 3-Lead ECG trunk cable, ESUproof

12Pin 3-Lead trunk cable, defibrillation-proof, DIN Conn

Usage

Reusable

Reusable

Reusable

Reusable

Reusable

Reusable

Reusable

Reusable

Note: 040-003528-00 uses 5-Lead N/T leadwires for the limb leads.

36.1.4

12-Pin Trunk Cables Also Compatible

Applicable patient

Adult/Pediatric

Adult/Pediatric

Adult/Pediatric

Adult/Pediatric

Adult/Pediatric

Pediatric/Neonate

Pediatric/Neonate

Pediatric/Neonate

/

/

Model

EV6201

EV6203

EV6211

Part No.

0010-30-42719

0010-30-42721

0010-30-42723

040-001416-00

009-003652-00

Description

12Pin 3/5-Lead ECG trunk cable, defibrillation-proof

12Pin 12-Lead trunk cable, defibrillation-proof

12Pin 3/5-Lead ECG trunk cable,

ESU-proof

12Pin 3/5-Lead ECG trunk cable, defibrillation-proof

12Pin 3/5-Lead ECG trunk cable,

ESU-proof

Usage

Reusable

Reusable

Reusable

Reusable

Reusable

Applicable patient

Adult/Pediatric

Adult/Pediatric

Adult/Pediatric

Adult/Pediatric

Adult/Pediatric

36.1.5

3-lead ECG Leadwires Available for Purchase

Model

EY6316B

EY6305B

EY6316A

EY6305A

EY6310B

EY6310B

EL6305A

Part No.

009-004765-00

009-004766-00

009-004771-00

009-004772-00

009-004777-00

115-032954-00

0010-30-42896

Description

3-Lead, N/T, AHA, snap, 24’’

3-Lead, N/T, AHA, snap, 36’’

3-Lead, N/T, AHA, clip, 24’’

3-Lead, N/T, AHA, clip, 36’’

3-Lead, N/T, AHA, snap, 24’’

3-Lead, N/T, AHA, snap, 24’’, pkg of

20

3-Lead leadset, AHA, clip Reusable

Usage

Reusable

Reusable

Reusable

Reusable

Disposable

Disposable

Applicable patient

Adult/Pediatric

Adult/Pediatric

Adult/Pediatric

Adult/Pediatric

Adult/Pediatric

Adult/Pediatric

Neonate

36 - 2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Model

EL6311B

EL6311A

Part No.

040-000146-00

040-000148-00

Description

3-Lead leadset, AHA, snap

3-Lead leadset, AHA, clip

Usage

Disposable

Disposable

36.1.6

3-lead ECG Leadwires Also Compatible

Model

EL6303A

EL6301B

/

/

/

/

Part No.

0010-30-42731

0010-30-42734

0012-00-1503-05

0012-00-1503-06

0012-00-1514-05

0012-00-1514-06

Description Usage

3-Lead leadset, AHA, clip, Long

3-Lead leadset, AHA, snap

3-Lead leadset, snap, 24" (61.0 cm),

AAMI

3-Lead leadset, snap, 36" (101.6 cm), AAMI

3-Lead leadset, clip, 24" (61.0 cm),

AAMI

3-Lead leadset, clip, 36" (101.6 cm),

AAMI

Reusable

Reusable

Reusable

Reusable

Reusable

Reusable

36.1.7

5-lead ECG Leadwires Available for Purchase

Applicable patient

Pediatric/Neonate

Pediatric/Neonate

Applicable patient

Adult/Pediatric

Adult/Pediatric

Adult/Pediatric

Adult/Pediatric

Adult/Pediatric

Adult/Pediatric

Model

EY6511B

EY6512B

EY6511A

EY6512A

EY6507B

EY6507B

Part No.

009-004782-00

009-004783-00

009-004786-00

009-004787-00

009-004790-00

115-032955-00

Description

5-Lead, N/T, AHA, snap, 24’’

5-Lead, N/T, AHA, snap, 36’’

5-Lead, N/T, AHA, pinch, 24’’

5-Lead, N/T, AHA, pinch, 36’’

5-Lead, N/T, AHA, snap, 24’’

5-Lead, N/T, AHA, snap, 24’’, pkg of

20

Usage

Reusable

Reusable

Reusable

Reusable

Disposable

Disposable

Applicable patient

Adult/Pediatric

Adult/Pediatric

Adult/Pediatric

Adult/Pediatric

Adult/Pediatric

Adult/Pediatric

36.1.8

5-lead ECG Leadwires Also Compatible

/

/

Model

EL6503A

EL6501B

EL6501A

/

/

Part No.

0010-30-42729

0010-30-42735

0010-30-42727

0012-00-1503-02

0012-00-1503-03

0012-00-1514-02

0012-00-1514-03

Description Usage

5-Lead leadset, AHA, clip, Long

5-Lead leadset, AHA, snap

5-Lead leadset, AHA, clip

5-Lead leadset, snap, 24" (61.0 cm),

AAMI

5-Lead leadset, snap, 36" (101.6 cm), AAMI

5-Lead leadset, clip, 24" (61.0 cm),

AAMI

5-Lead leadset, clip, 36" (101.6 cm),

AAMI

Reusable

Reusable

Reusable

Reusable

Reusable

Reusable

Reusable

Applicable patient

Adult/Pediatric

Adult/Pediatric

Adult/Pediatric

Adult/Pediatric

Adult/Pediatric

Adult/Pediatric

Adult/Pediatric

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 36 - 3

36.1.9

6-lead ECG Leadwires Available for Purchase

Model

EY6601B

EY6602B

EY6601A

EY6602A

Part No.

009-004794-00

009-004795-00

009-004798-00

009-004799-00

Description

6-Lead, N/T, AHA, snap, 24’’

6-Lead, N/T, AHA, snap, 36’’

6-Lead, N/T, AHA, pinch, 24’’

6-Lead, N/T, AHA, pinch, 36’’

36.1.10

12-lead ECG Chest Leadwires Available for Purchase

Usage

Reusable

Reusable

Reusable

Reusable

Model

EL6803A

EL6803B

Part No.

0010-30-42904

0010-30-42908

Description

12-Lead leadset, chest, AHA, clip

12-Lead leadset, chest, AHA, snap

Usage

Reusable

Reusable

36.1.11

12-lead ECG Leadwires Also Compatible

Applicable patient

Adult/Pediatric

Adult/Pediatric

Adult/Pediatric

Adult/Pediatric

Applicable patient

Adult/Pediatric

Adult/Pediatric

Model

EL6801A

EL6801B

Part No.

0010-30-42902

0010-30-42906

Description

12-Lead leadset, limb, AHA, clip

12-Lead leadset, limb, AHA, snap

Usage

Reusable

Reusable

36.2.2

Extension Cables and Adapter Cables Also Compatible

Model

582A

Part No.

115-020768-00

Description

8-pin SpO2 extension cable, Masimo

Usage

Reusable

Applicable patient

Adult/Pediatric

Adult/Pediatric

36.2

SpO

2

Accessories

Wavelength emitted by the sensors is between 600 nm and 1000 nm. The maximum photic output consumption of the sensor is less than 18 mW.

The information about the wavelength range and maximum photic output consumption can be especially useful to clinicians, for example, when photodynamic therapy is performed.

36.2.1

Extension Cables and Adapter Cables Available for Purchase

Model

572A

583A

4089

4092

Part No.

0010-20-42712

040-003310-00

040-003381-00

040-003426-00

Description

8-pin SpO2 extension cable, Nellcor

8 pin, RD SET, Masimo

RD to LNCS adapter cable, Masimo

LNCS to RD adapter, Masimo

Usage

Reusable

Reusable

Reusable

Reusable

Applicable patient

/

/

/

/

/

Applicable patient

36 - 4 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

36.2.3

Masimo SpO

2

RD Set Sensors Available for Purchase

4002

4003

4004

4005

Model

4050

4051

4053

4000

4001

Part No.

040-003376-00

040-003377-00

040-003380-00

040-003382-00

040-003383-00

040-003384-00

040-003385-00

040-003386-00

040-003387-00

Description Usage

SpO2 sensor, RD SET DCI

SpO2 sensor, RD SET DCI

SpO2 sensor, RD Set TC-I, tip-clip ear sensor, 3ft

SpO2 sensor, RD SET Adt, adhesive, pkg of 20

SpO2 sensor, RD SET PDT, adhesive, pkg of 20

SpO2 sensor, RD Set Inf, adhesive, pkg of 20

SpO2 sensor, RD Set Neo, adhesive, pkg of 20

SpO2 sensor, RD Set NeoPt, adhesive, pkg of 20

SpO2 sensor, RD Set NeoPt-500, nonadhesive, pkg of 20

Reusable

Reusable

Reusable

Disposable

Disposable

Disposable

Disposable

Disposable

Disposable

Applicable patient

Adult (> 30 kg)

Pediatric (10-50 kg)

Adult (> 30 kg)

Adult (> 30 kg)

Pediatric (10 to 50 kg)

Infant (3 to 20 kg)

Neonatal /Adult

(<3Kg or >40Kg)

Neonate (<1 kg)

Neonate (<1 kg)

36.2.4

Masimo SpO

2

LNCS Sensors Also Compatible

Model Part No.

Description

LNCS Adtx

LNCS Pdtx

LNCS DCI

LNCS DCIP

LNCS NeoPt

LNCS Neo-L

0600-00-0121

0600-00-0122

0600-00-0126

0600-00-0127

0600-00-0156

0600-00-0157

SpO2 sensor, finger-clip, 20 pcs/box

SpO2 sensor, finger-clip, 20 pcs/box

SpO2 sensor, finger-clip

SpO2 sensor, finger-clip

SpO2 sensor, finger-clip

SpO2 sensor, 20 pcs/box

Usage

Disposable

Disposable

Reusable

Reusable

Disposable

Disposable

Disposable

Applicable patient

Adult (>30 kg)

Pediatric (10-50 kg)

Adult (>30 kg)

Pediatric (10-50 kg)

Neonate (<1Kg)

Neonate /Adult

(<3 kg or > 40 kg)

Infant (3-20 kg) LNCS Inf 0600-00-0158 SpO2 sensor, 20 pcs/box

36.2.5

Nellcor SpO

2

Sensors Also Compatible

Model

DS100A

D-YS

OXI-P/I

OXI-A/N

MAXAI

MAXPI

MAXII

MAXNI

Part No.

9000-10-05161

0010-10-12476

9000-10-07308

9000-10-07336

0010-10-12202

0010-10-12203

0010-10-12204

0010-10-12205

Description

SpO2 sensor, finger-clip

SpO2 sensor, with wraps

SpO2 sensor, finger, band

SpO2 sensor, finger/foot, band

SpO2 sensor, 24 pcs/box

SpO2 sensor, 24 pcs/box

SpO2 sensor, 24 pcs/box

SpO2 sensor, 24 pcs/box

Usage

Reusable

Reusable

Reusable

Reusable

Disposable

Disposable

Disposable

Disposable

Applicable patient

Adult

/

Pediatric/Infant

Adult/Neonate

Adult (>30 kg)

Pediatric (10-50 Kg)

Infant (3-20 kg)

Adult/Neonate

(<3 kg or >40 kg)

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 36 - 5

36.3

Temp Accessories

36.3.1

Temp Cable Available for Purchase

Model

MR420B

Part No.

040-001235-00

36.3.2

Temp Cable Also Compatible

Description

2-pin Temperature adapter cable

Usage

Reusable

Model

MR420B

Part No.

0011-30-37391

Description

2-pin, extension cable

36.3.3

Temp Probes Available for Purchase

Usage

Reusable

/

Applicable patient

/

Applicable patient

Model

MR401B

MR402B

MR403B

MR404B

MR411

MR411

MR412

MR412

ES400-12

ES400-18

ER400-9

ER400-12

STS-400

Part No.

0011-30-37392

0011-30-37394

0011-30-37393

0011-30-37395

040-003292-00

040-003294-00

040-003293-00

040-003295-00

0206-03-0112-02

0206-03-0118-02

0206-03-0209-02

0206-03-0212-02

0206-03-0300-02

Description Usage

Temperature probe, endocavity

Temperature probe, endocavity

Temperature probe, skin surface

Temperature probe, skin surface

Temperature probe, esophageal/ rectal, 9FR

Temperature probe, esophageal/ rectal, 9FR, pkg of 20

Reusable

Reusable

Reusable

Reusable

Disposable

Disposable

Temp probe, skin surface Disposable

Temp probe, skin surface, pkg of 20 Disposable

PROBE,D TEMP,ES400-12 (box of 20) Disposable

PROBE,D TEMP,ES400-18 (box of 20) Disposable

PROBE,D TEMP,ER 400-9 (box of 20) Disposable

PROBE,D TEMP,ER400-12 (box of 20) Disposable

PROBE,D TEMP,STS-400 (box of 20) Disposable

All

All

All

Adult

Adult

Adult

Adult

Adult

Applicable patient

Adult

Pediatric/Infant

Adult

Pediatric/Infant

All

36.3.4

Temp Probes Also Compatible

Model

MR411

MR412

Part No.

0011-30-90446

0011-30-90447

Description

Disposable temperature probe, esophageal

Disposable temperature probe, skin

Usage

Disposable

Disposable

Applicable patient

All

All

36 - 6 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

36.4

NIBP Accessories

36.4.1

NIBP Hoses Available for Purchase

Model

CM1903

CM1901

Part No.

6200-30-09688

6200-30-11560

Description

NIBP hose, inbuilt connector, 3m

NIBP hose, inbuilt connector, 3m

Usage

Reusable

Reusable

36.4.2

Cuffs Available for Purchase

Applicable patient

Adult/Pediatric

Neonate

CM1504

CM1505

CM1506

CM1507

CM1500A

CM1500B

CM1500C

CM1500D

/

Model

CM1301

CM1302

CM1303

CM1304

CM1305

CM1306

CM1307

CM1501

CM1502

CM1503

115-027566-00

115-027567-00

115-027568-00

115-027569-00

125-000051-00

125-000052-00

125-000053-00

125-000054-00

115-031807-00

Part No.

115-027713-00

115-027714-00

115-027715-00

115-027716-00

115-027717-00

115-027718-00

115-027719-00

115-027563-00

115-027564-00

115-027565-00

Description

Cuff, 10-19 cm, bladderless

Cuff, 18-26 cm, bladderless

Cuff, 25-35 cm, bladderless

Cuff, 33-47 cm, bladderless

Cuff, 46-66 cm, bladderless

Cuff, 24-35 cm, bladderless

Cuff, 33-47 cm, bladderless

Cuff, 10-19 cm, bladderless, pkg of

10

Cuff, 18-26 cm, bladderless, pkg of

10

Cuff, 25-35 cm, bladderless, pkg of

10

Cuff, 33-47 cm, bladderless, pkg of

10

Cuff, 46-66 cm, bladderless, pkg of

5

Cuff, 25-35 cm, bladderless, pkg of

10, long

Cuff, 33-47 cm, bladderless, pkg of

10, long

Cuff, 3.1-5.7 cm, pkg of 20

Cuff, 4.3-8.0 cm, pkg of 20

Cuff, 5.8-10.9 cm, pkg of 20

Cuff, 7.1-13.1 cm, pkg of 20

Cuff starter kit, bladderless

Usage

Reusable

Reusable

Reusable

Reusable

Reusable

Reusable

Reusable

Disposable

Disposable

Disposable

Disposable

Disposable

Disposable

Disposable

Disposable

Disposable

Disposable

Disposable

Reusable

Large adult

Adult thigh

Adult

Large adult

Neonate

Neonate

Neonate

Neonate

Child/Small adult/

Adult/Adult thigh

Applicable patient

Infant

Small Adult

Adult

Adult

Adult thigh

Adult

Adult

Child

Small adult

Adult

36.4.3

Cuffs Also Compatible

Model

CM1500A

CM1500B

Part No.

001B-30-70692

001B-30-70693

Description

Cuff, 3.1-5.7 cm, pkg of 20

Cuff, 4.3-8.0 cm, pkg of 20

Usage

Disposable

Disposable

Applicable patient

Neonate

Neonate

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 36 - 7

Model

CM1500C

CM1500D

Part No.

001B-30-70694

001B-30-70695

36.5

IBP Accessories

Description

Cuff, 5.8-10.9 cm, pkg of 20

Cuff, 7.1-13.1 cm, pkg of 20

Usage

Disposable

Disposable

Applicable patient

Neonate

Neonate

36.5.1

IBP Accessories Available for Purchase

Model

IM2202

IM2201

/

Part No.

001C-30-70757

001C-30-70759

125-000123-00

Description

12 Pin IBP cable, Argon

12 Pin IBP cable, ICU Medical

12 Pin IBP cable, Edwards (not for transportation)

Usage

Reusable

Reusable

Reusable

36.5.2

ICP Accessories Available for Purchase

Applicable patient

/

/

/

Model

82-6653

/

CP12601

Part No.

040-002336-00

115-025257-00

009-005460-00

36.5.3

ICP Accessories Also Compatible

Description

Codman ICP sensor kit, Johnson &

Johnson, pkg of 1

Camino ICP cable kit

12-pin ICP cable

Usage

Disposable

Reusable

Reusable

/

/

Applicable patient

/

626631US / Codman microsensor basic kit for subdural/intraparenchyma1 procedure

Disposable /

36.6

C.O. Accessories Available for Purchase

Model

CO7702

9850A

SP4042

SP5045

TR0012

93522

Part No.

0010-30-42743

0012-00-1519

6000-10-02079

6000-10-02080

040-006154-00

0012-00-1520

Description

12Pin C.O. cable, Edwards

Cable, bath probe adapter

In-line injection temperature sensor, BD

In-line injection temperature sensor housing, BD, pkg of 25

12cc control syringe w/Rotator

(MEDEX MX387)

In-line inject temperature probe

Usage

Reusable

Reusable

Reusable

Disposable

Disposable

Reusable

/

/

Applicable patient

/

/

/

/

36 - 8 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

36.7

ScvO2 Accessories Available for Purchase

Model

PC3030

PV2022-37

PV2022-35

Part No.

115-008191-00

040-000919-00

040-000920-00

Description Usage

ScvO2 optical module

CeVOX fiberoptic probe, 37 cm, pkg of 1

CeVOX fiberoptic probe, 35 cm, pkg of 1

Reusable

Disposable

Disposable

Applicable patient

/

/

/

36.8

CO

2

Accessories

36.8.1

Sidestream CO

2

Accessories Available for Purchase

Model

4000

4100

4200

60-15200-00

60-15300-00

60-14100-00

040-001187-00

60-14200-00

100-000080-00

100-000081-00

Part No.

115-043001-00

115-043002-00

115-043003-00

115-043017-00

115-043018-00

115-043020-00

115-043019-00

115-043021-00

115-058733-00

115-058734-00

Description

CO2 Nasal sample cannula, 7'’ line, pkg of 25

CO2 Nasal sample cannula, 7'’ line, pkg of 25

CO2 Nasal sample cannula, 7'’ line, pkg of 25

Sampling line, 2.5 m, pkg of 25

Sampling line, 2.5 m, pkg of 25

Dryline airway adapter, straight, pkg of 10

Airway Adapter pkg of 10

Dryline airway adapter, elbow pkg of 10

DRYLINE II water trap, pkg of 10

DRYLINE II water trap, pkg of 10

Usage

Disposable

Disposable

Disposable

Disposable

Disposable

Disposable

Disposable

Disposable

Reusable

Reusable

Applicable patient

Adult

Pediatric

Neonatal

Adult/Pediatric

Neonatal

Adu/Ped

Neonatal

/

Adult/Pediatric

Neonatal

36.8.2

Microstream CO

2

Accessories Also Compatible

Model

XS04620

XS04624

006324

007768

007737

007738

009818

007266

009822

007269

Part No.

0010-10-42560

0010-10-42561

0010-10-42562

0010-10-42563

0010-10-42564

0010-10-42565

0010-10-42566

0010-10-42567

0010-10-42568

0010-10-42569

Description Usage

FilterLine Set Disposable

FilterLine H Set Disposable

FilterLine H Set

FilterLine Set, long

Disposable

Disposable

FilterLine H Set, long

FilterLine H Set, long

S CapnoLine Plus

Smart CapnoLine

S CapnoLine Plus O2

Smart CapnoLine O2

Disposable

Disposable

Disposable

Disposable

Disposable

Disposable

Applicable patient

Adult, pediatric

Neonate

Adult, pediatric

Adult, pediatric

Neonate

Adult

Pediatric

Adult

Pediatric

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 36 - 9

Model

009826

007743

008177

008178

008179

008180

008181

008174

008175

36.9

AG Accessories

Part No.

0010-10-42570

0010-10-42571

0010-10-42572

0010-10-42573

0010-10-42574

0010-10-42575

0010-10-42576

0010-10-42577

0010-10-42578

Description Usage

S CapnoLine Plus O2, long

S CapnoLine O2, long

CapnoLine H

CapnoLine H

Disposable

Disposable

Disposable

Disposable

CapnoLine H

CapnoLine H O2

Disposable

Disposable

CapnoLine H O2 Disposable

NIV-Line Disposable

NIV-Line Disposable

36.9.1

AG Accessories Available for Purchase

Applicable patient

Adult

Pediatric

Adult

Pediatric

Neonate

Adult

Pediatric

Adult

Pediatric

Model

60-15200-00

60-15300-00

60-14100-00

60-14200-00

100-000080-00

100-000081-00

115-052370-00

115-052162-00

Part No.

115-043017-00

115-043018-00

115-043020-00

115-043021-00

115-058733-00

115-058734-00

115-052370-00

115-052162-00

Description

Sampling line, 2.5 m, pkg of 25

Sampling line, DRYLINE, 2.5m, pkg of 25

DRYLINE airway adapter, straight, pkg of 10

DRYLINE airway adapter, elbow, pkg of 10

DRYLINE II water trap, pkg of 10

DRYLINE II water trap, pkg of 10

A7 Recirculation line

Generic scavenging line

Usage

Disposable

Disposable

Disposable

Disposable

Reusable

Reusable

Reusable

Reusable

Applicable patient

Adult/Pediatric

Neonate

Adult/Pediatric

Adult/Pediatric

/

/

Adult/Pediatric

Neonate

36.9.2

AG Accessories Also Compatible

Model

115-052160-00

Part No.

115-052160-00

Description

A3/A4/A5 Scavenging line

36.10

EEG Accessories Available for Purchase

Usage

Reusable /

Applicable patient

Model Part No.

B8830085010 040-001594-00

B9721104003 040-001598-00

B9721105004

B9600085001

040-001602-00

040-001596-00

Description

Cup electrodes for EEG

Cup electrodes for EEG (D=6 mm)

Subdermal for EEG

Usage

Reusable

Reusable

Reusable

Disposable

Applicable patient

Adult/Pediatric

Adult/Pediatric

Pediatric

Adult/Pediatric

36 - 10 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

36.11

BIS Accessories

36.11.1

BIS Accessories Available for Purchase

Model

186-0195-MR

186-0224-MR

Part No.

6800-30-50761

115-005707-00

Description

Bispectral Index kit, Covidien

BISx4 cable kit, Covidien

36.11.2

BIS Accessories Also Compatible

Usage

Reusable

Reusable

Applicable patient

/

/

Model

186-0106

186-0200

186-0212

186-0160

Part No.

0010-10-42672

0010-10-42673

040-000392-00

6800-10-50652

Description Usage

BIS Sensor Quatro(186-0106), Covidien Disposable

BIS Pediatric sensor(186-0200),

Covidien

Disposable

BISx4 Sensor(Bilateral)(186-0212),

Covidien

Disposable

BIS Extend Sensor(186-0160), Covidien Disposable

Applicable patient

Adult

Pediatric

Adult

Adult

36.12

NMT Accessories Available for Purchase

Model

NM13101

NM13401

NM13701

NM13901

Part No.

040-001462-00

040-001463-00

040-001464-00

040-002258-00

Description

NMT cable

NMT sensor cable

NMT stimulation cable

Bandage for NMT sensor, pkg of 20

36.13

rSO

2

Accessories

36.13.1

rSO

2

Accessories Available for Purchase

Usage

Reusable

Reusable

Reusable

Disposable

Applicable patient

Adult, pediatric

Adult, pediatric

Adult, pediatric

Adult, pediatric

Model

RSC-1

RSC-2

5100C-PA

Part No.

100-000164-00

100-000165-00

100-000173-00

Description Usage

Sensor cable, channel 1, INVOS 5100C

Sensor cable, channel 2, INVOS 5100C

Reusable

Reusable

Preamplifier, channel 1 & 2, INVOS 5100C Reusable

36.13.2

rSO

2

Accessories Also Compatible

Model

SAFB-SM

SPFB

SNN

Applicable patient

/

/

/

Part No.

100-000168-00

100-000169-00

100-000181-00

Description

SomaSensor (>40 kg)

SomaSensor (<40 kg)

OxyAlert NIRSensor (<5kg), somatic, with sensor cable

Usage

Disposable

Disposable

Disposable

Applicable patient

Adult

Pediatric

Neonate

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 36 - 11

Model

CNN

CNN/SNN

Part No.

100-000180-00

100-000182-00

Description

OxyAlert NIRSensor (<5kg), cerebral, with sensor cable

OxyAlert NIRSensor (<5kg), somatic/ cerebral, with sensor cable

Usage

Disposable

Disposable

36.14

CCO/SvO2 Accessories Available for Purchase

Model

/

Part No.

009-000259-00

Description

CCO/SvO

2

cable

Usage

Reusable

36.15

BeneLink Accessories

Applicable patient

Neonate

Neonate

Applicable patient

/

Part No.

009-001767-00

009-001768-00

009-001769-00

009-002943-00

009-004613-00

009-008485-00

009-008624-00

115-011465-00

115-007277-00

047-004857-00

047-004859-00

009-001770-00

36.16

External Modules

Module

MPM module

Part No.

M51CE-PA00001

MPM module

MPM module

MPM module

MPM module

MPM module

Masimo SpO

2 module

Nellcor SpO

2 module

M51CE-PA00002

M51CE-PA00003

M51CE-PA00005

M51CE-PA00006

M51CE-PA00007

115-047244-00

115-053719-00

Description

Serial port adapting cable, type A

Serial port adapting cable, type B

Serial port adapting cable, type C

Serial port adapting cable, type D

Serial port adapting cable, type E

Serial port adapting cable, type F

Serial port adapting cable, type G

ID Adapter kit (contains ID Adapter with label and RJ45 cable with label)

ID Adapter

ID adapter label

Network line label

RJ45 connecting cable

Comments

Integrates 3/5/6-lead ECG, arrhythmia, ST, Resp, Masimo SpO

2

, Temp,

NIBP, IBP

Integrates 3/5/6-lead ECG, arrhythmia, Resp, Masimo SpO

2

, Temp,

NIBP, IBP

Integrates 12-lead ECG, arrhythmia, ST, Glasgow algorithm, Resp,

Masimo SpO

2

, Temp, NIBP, IBP, analog output

Integrates 3/5/6-lead ECG, arrhythmia, ST, Resp, Nellcor SpO

2

, Temp,

NIBP, IBP

Integrates 3/5/6-lead ECG, arrhythmia, Resp, Nellcor SpO

2

, Temp,

NIBP, IBP

Integrates 12-lead ECG, arrhythmia, ST, Glasgow algorithm, Resp,

Nellcor SpO

2

, Temp, NIBP, IBP, analog output

Supports SpO2 monitoring, Masimo SpO

2

Supports SpO2 monitoring, Nellcor SpO

2

36 - 12 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Module

Temp module

C.O. module

IBP module

CCO/SvO

2 module

Part No.

115-053714-00

115-047283-00

115-047286-00

115-053718-00

ScvO

2 module

EEG module

BIS module

NMT module

INVOS rSO

2 module

Microstream CO

2 module

Sidestream CO

2 module

Sidestream CO

2 module

AG module

AG module

115-013199-00

115-053713-00

115-013194-00

115-053717-00

115-037264-00

6800-30-50820

115-049134-00

115-053711-00

115-051561-00

115-049189-00

BeneLink module

Recorder module

115-053710-00

115-053716-00

Comments

Supports temperature monitoring

Supports C.O. monitoring

Supports IBP monitoring

Connects Edwards Vigilance II®, Vigileo

TM

, EV1000, or HemoSphere monitor, supports CCO and SvO2 monitoring

Supports ScvO2 monitoring

Supports EEG monitoring‘

Supports BIS monitoring

Supports NMT monitoring

Supports rSO

2

monitoring

Supports CO2 monitoring

Supports CO2 monitoring, with adult and pediatric accessories

Supports CO2 monitoring, with neonatal accessories

Supports AG monitoring, integrates O

2

(paramagnetic) monitoring

Supports AG monitoring, integrates O

2

(paramagnetic) and BIS monitoring

Connects external devices

Supports recording (for N22/N19)

36.17

Mount and Mounting Accessories

Part No.

115-033887-00

0010-30-11972

034-000457-00

034-000458-00

042-014101-00

043-006060-00

115-035995-00

115-033911-00

115-045621-00

Description

8-slot satellite module rack (SMR), with handle, hook, and 2 m cable

SMR clamp assembly

GCX M series pivot arm, 12" VESA mount

GCX M series articulating arm, 12" x 12" for SMR

NIBP cuff holder

Cable hooks

SMR clamp mounting kit, w/ 6" post

SMR cable management mounting kit

SMR clamp mounting kit, w/ 9" post

The following installation accessories are for N17/N15/N12 only.

Part No.

045-000915-00

045-003424-00

045-000955-00

045-001190-00

045-003425-00

Description

N17/N15 rolling stand with mounting plate

N12 quick release mount for rolling stand

N17/N15 quick release mount for rolling stand

N12 transport stand (MR) with GCX shoe

N17/N15/N12 Mount Transition Plate

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 36 - 13

045-003426-00

045-003253-00

045-003255-00

115-045865-00

045-003254-00

N17/N15/N12 M Series 12" arm with Transition Plate

N17/N15/N12 GCX VHM Series arm with Transition Plate

N12 rolling stand with quick release mount

N12 Bedrail Hook

TDS cross clamp with 9" riser post (for N17/N15/N12)

The following installation accessories are for N22/N19 only.

Part No.

034-000782-00

034-000454-00

045-001976-00

045-002198-00

115-033880-00

115-033871-00

115-051379-00

041-017853-00

Description

GCX VHM arm

GCX M series arm, 16", with 6” riser

Rotatory assembly kit for display

Dock install to bracket package

Host to GCX mounting kit

Keyboard kit with 9” downpost

Separate host & display mounting kit, for rotation

Host mounting block

36.18

Miscellaneous Accessories

Part No.

009-005121-00

009-011808-00

009-005122-00

009-006593-00

009-005123-00

009-006594-00

009-009766-00

009-005391-00

8000-21-10361

023-001566-00

0683-00-0505-02

115-056983-00

115-055522-00

023-001788-00

115-049288-00

115-049289-00

115-034030-00

115-037270-00

009-005115-00

009-007740-00

Description

SMR cable, 2m

SMR cable, 4m

SMR cable, 10m

N Series to N1/T1 docking station cable, 2 m

N Series to N1/T1 docking station cable, 4 m

N Series to N1/T1 docking station cable, 10 m

N Series to N1/T1 docking station cable, 20 m

MPM analog output and defib sync cable

Nurse call cable

HP LaserJet Enterprise M608n, black and white

Recording paper, 50 mm*20 m, pkg of 12 A30-000001---

Clinical Scoring Custom CD

Field-installed iView module (N22/N19)

External display, 21.5’(for N17/N15/N12)

Display assembly kit, 22", with AC adapter and 5 m video cable (for N22)

Display assembly kit, 19", with AC adapter and 5 m video cable (for N19)

Display handle, without encoder (for N22/N19)

Display handle, with encoder (for N22/N19)

Video cable, 2.3 m (for N22/N19)

Video cable, 5 m (for N22/N19)

36 - 14 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Part No.

009-005117-00

009-005118-00

009-008237-00

009-005120-00

115-065140-00

115-034132-00

DA8K-10-14452

1000-21-00122

009-003648-00

009-003903-00

023-000247-00

023-000248-00

023-000524-00

023-000525-00

115-039575-00

115-039635-00

115-045643-00

042-007382-00

034-000666-00

Description

Video cable, 10 m (for N22/N19)

USB cable, 2.3 m (for N22/N19)

USB cable, 5 m (for N22/N19)

USB cable, 10 m (for N22/N19)

Lithium-ion battery, LI23S002A (for N17/N15/N12)

Lithium-ion battery, LI23S003A (for N22/N19)

Power cord, USA

Grounding cable

Cable protecting tube

Accessory management tape

Keyboard, wired USB connector, black

Mouse, wired USB connector, black

Wireless keyboard and mouse set

Wired keyboard and mouse set

2D Barcode reader, HS-1M, JADAK

2D Barcode reader, HS-1R, JADAK

Remote control

Cable management hook

Barcode reader holder (for JADAK Barcode reader)

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual 36 - 15

This page intentionally left blank.

36 - 16 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

A

Product Specifications

A.1

A.2

Monitor Safety Specifications

The monitor is classified, according to IEC 60601-1: 2005:

Degree of protection against electrical shock

Type CF defibrillation proof for ECG, Resp, TEMP, IBP, SpO

2

, C.O., NIBP, EEG, and

NMT

Type BF defibrillation proof for ScvO

2

, CO

2

, AG, BIS, and rSO

2

Class I Type of protection against electrical shock

Degree of protection against harmful ingress of water

Degree of safety of application in the presence of flammable anesthetic mixture with air or with oxygen or nitrous oxide

Mode of operation

IPX1

The equipment is not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide

Continuous

Physical Specifications

A.2.1

BeneVision N22/N19

Item Maximum

Weight (kg)

10.30

N19 monitor

(main unit and primary display installed together)

N22 monitor

(main unit and primary display installed together)

Main unit

Display, 19 inch

Display, 22 inch

11.50

3.40

6.20

7.40

W × H × D (mm) Comments

509 × 423 × 115

(display horizontally installed)

584 × 348 × 115

(display vertically installed)

566 × 458 × 115

(display horizontally installed)

641 × 383 × 115

(display vertically installed)

268 × 268 × 68

509 × 348 × 48

566 × 383 × 48

Including the battery, iView module, Wi-Fi module, display with the handle and navigation knob.

Excluding SMR, modules, and accessories.

Including the battery, iView module, Wi-Fi module, display with the handle and navigation knob.

Excluding SMR, modules, and accessories.

Including the battery.

Excluding the handle.

Excluding the handle.

A.2.2

BeneVision N17/N15/N12

Item

N17

N15

N12

Maximum

Weight (kg)

7.3

5.4

4.1

W × H × D (mm)

466×355×210

396×313×193

313×290×161

Comments

Excluding iView module, Wi-Fi module, SMR, measurement modules, recorder, battery and accessories.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual A - 1

A.2.3

SMR and Modules

Item Maximum

Weight (kg)

Satellite module rack (SMR) 2.30

MPM module 0.63

SpO

2 module

Temp module

0.29

0.25

C.O. module

IBP module

CCO/SvO

2 module

ScvO

2 module

EEG module

BIS module

NMT module rSO

2 module

Microstream CO

2 module

Sidestream CO

2 module

AG module

0.25

0.26

0.26

0.26

0.28

0.26

0.29

0.30

BeneLink module

Recorder

0.38

0.63

1.03

0.35

0.40

W × H × D (mm)

403 × 221 × 145

136.5 × 80.5 × 102

136.5 × 40 × 102

136.5 × 40 × 102

136.5 × 40 × 102

136.5 × 40 × 102

136.5 × 40 × 102

136.5 × 40 × 102

136.5 × 40 × 102

136.5 × 40 × 102

136.5 × 40 × 102

136.5 × 40 × 102

136.5 × 40 × 102

136.5 × 40 × 102

136.5 × 80.5 × 102

136.5 × 40 × 102

136.5 × 80.5 × 102

A.3

Comments

/

/

/

/

Including the handle and cable hooks.

/

Masimo SpO

2

or Nellcor SpO

2

/

/

/

/

/

/

/

With built-in O

2

module

/

/

Environmental Specifications

WARNING

The monitor may not meet the performance specifications if stored or used outside the specified temperature and humidity ranges. If the performance of the equipment is degraded due to aging or environmental conditions, contact your service personnel.

When the monitor and related products have differing environmental specifications, the effective range for the combined products is that range which is common to the specifications for all products.

NOTE

The environmental specification of unspecified parameter modules are the same as those of the main unit.

Components

Main Unit

Property (unit)

Temperature (°C)

Relative humidity

(noncondensing) (%)

Barometric (mmHg)

Operating Condition

0 to 35 (for N17 configured with iView module)

0 to 40 for other configuration

15 to 95

Storage Condition

-20 to 60

10 to 95

427.5 to 805.5

120 to 805.5

A - 2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Components

Microstream CO

2 module

Sidestream CO

2 module

AG module

ScvO

2 module rSO

2 module

Property (unit)

Temperature (°C)

Relative humidity

(noncondensing) (%)

Barometric (mmHg)

Temperature (°C)

Relative humidity

(noncondensing) (%)

Barometric (mmHg)

Temperature (°C)

Relative humidity

(noncondensing) (%)

Barometric (mmHg)

Temperature (°C)

Relative humidity

(noncondensing) (%)

Barometric (mmHg)

Temperature (°C)

Relative humidity

(noncondensing) (%)

Barometric (mmHg)

Operating Condition

0 to 40

15 to 95

430 to 790

5 to 40

15 to 95

430 to 790

10 to 40

15 to 95

525 to 805.5

10 to 40

15 to 75

427.5 to 805.5

16 to 32

20 to 80

525 to 805.5

A.4

A.4.1

Power Supply Specifications

External Power Supply Specifications

Line voltage

Input Current

Frequency

100 to 240 VAC (±10%)

N22/N19: 2.8 A to 1.6 A

N17/N15/N12: 2.0 A to 0.9 A

50/60 Hz (± 3 Hz)

A.4.2

Battery Specifications

A.4.2.1

N22/N19 Battery Specifications

Battery type

Voltage

Rechargeable lithium-ion battery

10.8 VDC

Run time

Charge time

At least 1 hour continuously on a typical configuration, powered by a new, fully-charged battery.

Note: A typical configuration is 12-lead ECG, Resp, SpO

2

, 4-channel IBP, 2channel Temp, CO

2

, C.O., and NIBP module set at an interval of 15 minutes.

WI-Fi is enabled.

The screen brightness is set to the factory default.

For a new battery:

5 hours to 90% when the monitor is off

9 hours to 90% when the monitor is on.

Storage Condition

-20 to 60

10 to 95

430 to 790

-20 to 60

10 to 95

430 to 790

-20 to 60

10 to 95

525 to 805.5

-20 to 60

10 to 90

120 to 805.5

-20 to 70

10 to 95

435.7 to 822

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual A - 3

Shutdown delay At least 15 minutes (after a low battery alarm first occurs)

A.4.3

N17/N15/N12 Battery Specifications

Battery type

Voltage

Rechargeable lithium-ion battery

10.95 VDC

A.5

A.6

A.7

Run time

Charge time

Shutdown delay

At least 2 hours continuously for N17/N15 or at least 4 hours continuously for

N12 on a typical configuration, powered by a new, fully charged battery.

The monitor is powered by a new, fully-charged battery at 25 °C with a 5-lead

ECG and SpO2 cable connected, auto NIBP measurements at an interval of 15 minutes, and the screen brightness is set to 1.

No more than 4.5 hours to 90% when the monitor is off

No more than 10 hours to 90% when the monitor is on

At least 15 minutes (after a low battery alarm first occurs)

Display Specifications

Screen type

Screen Size (diagonal)

Resolution

Medical-grade color TFT LCD

N22: 22 inches

N19: 19 inches

N17: 18.5 inches

N15: 15.6 inches

N12: 12.1 inches

N22/N19: 1680 x 1050 pixels

N17/N15: 1920 x1080 pixels

N12: 1280 x 800 pixels

Recorder Specifications

Method

Horizontal resolution

Vertical resolution

Paper width

Paper length

Paper speed

Number of waveform channels

Thermal dot array

16 dots/mm (25 mm/s paper speed)

8 dots/mm

50 mm

20 m

25 mm/s, 50 mm/s

Accuracy: ±5%

A maximum of 3

LEDs

Alarm lamp

Power-on LED

AC power LED

Battery LED

1 (three color-coded: red, yellow, and cyan)

1 (green)

1 (green)

1 (two color-coded: yellow and green)

A - 4 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

A.8

Audio Indicator

Speaker Give alarm tones (45 to

85

dB), reminder tones, key tones, QRS tones; support PITCH TONE and multi-level tone modulation; alarm tones comply with IEC 60601-1-8:2012.

A.9

A.9.1

Monitor Interface Specifications

Interface Specifications of the N22/N19 Main Unit

AC power input

Network connector (LAN1, LAN2, LAN3)

Serial bus connector (MSB)

USB connector

Satellite module rack (SMR) connector

Video output connector (VP1, VP2)

Nurse call connector (NC)

Equipotential grounding terminal

Primary display connectors

1

3 standard RJ45 connectors (one on the iView module), 100 Base-TX, IEEE

802.3

6: 3 connectors and each connector contains 2 sub-connectors

4, USB 2.0 (for iView only)

3

2, VP1 connects the secondary display (DisplayPort). VP2 connects the display for iView system (DisplayPort)

1, standard BNC

1

1 set (VP3, SIG1, MSB4, and DC-in), for connecting to separate primary display, hidden when the display is integrated with the main unit.

A.9.2

Interface Specifications of the N22/N19 Separate Primary Display

Serial bus connector (MSB)

Serial bus hub connector (SBH)

Signal input connector (SIG1)

DC-in connector

Video input connector (VP)

1

1

3

1

1

A.9.3

Interface Specifications of the N22/N19 Integrated Primary Display

Serial bus connector (MSB) 1

A.9.4

Interface Specifications of the N22/N19 Secondary Display

Serial bus connector (MSB)

Serial bus hub connector (SBH)

Signal input connector (SIG1)

DC-in connector

Video input connector (VP)

1

1

3

1

1

A.9.5

Interface Specifications of the N17/N15/N12

AC power input 1

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual A - 5

Network connector

USB connector

Satellite module rack (SMR)/Docking station connector

Video output connector

Nurse call connector

Equipotential grounding terminal

N17: 2 standard RJ45 connectors (one on the iView module)

N15/N12: 1 standard RJ45 connector

4, USB 2.0

1 (For N17/N15, it connects the SMR or the docking station. For N12, it connects the N1/T1 docking station)

N17: 2, (configuration A: 1 mirrored (DVI), 1 independent (DVI); configuration

B: 1 mirrored (DVI), 1 iView (DVI).)

N15/N12: 1 mirrored (DVI)

1, standard BNC

1

A.10

Signal Outputs Specifications

Auxiliary Output

Standard Meets the requirements of IEC 60601-1:2005 for short-circuit protection and leakage current

ECG Analog Output

Bandwidth

(-3dB; reference frequency: 10Hz)

Maximum QRS delay

Gain (reference frequency 10Hz)

Pace enhancement

Diagnostic mode: 0.05 to 150 Hz

Monitor mode: 0.5 to 40 Hz

Surgical mode: 1 to 20 Hz

ST mode: 0.05 to 40 Hz

25 ms (in diagnostic mode, and non-paced)

1V/mV (±5%)

Signal amplitude: V oh

≥2.5V

Pulse width: 10 ms±5%

Signal rise and fall time: ≤100μs

IBP Analog Output

Bandwidth (-3dB; reference frequency:1Hz)

Maximum transmission delay

Gain (reference frequency 1 Hz)

Nurse Call Signal

Amplitude

0 to 40 Hz

30 ms

1 V/100 mmHg, ±5%

High level: 3.5 to 5 V, ±5%, providing a minimum of 10 mA output current;

Low level: < 0.5 V, receiving a minimum of 5 mA input current.

≤1 ms Rise and fall time

Defib Sync Pulse

Output impedance

Maximum time delay

Amplitude

Pulse width

Maximum rise and fall time

Alarm output

≤100 Ω

35 ms (R-wave peak to leading edge of pulse)

High level: 3.5 to 5 V, ±5%, providing a maximum of 10 mA output current;

Low level: < 0.5 V, receiving a maximum of 5 mA input current.

100 ms ±10%

1 ms

A - 6 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Alarm delay time from the monitor to remote equipment

Alarm signal sound pressure level range

The alarm delay time measured at the monitor signal output connector:

From the monitor to the CMS and remote monitors: ≤2 seconds

From the monitor to the TM80 telemetry monitor: ≤3 seconds

From the monitor to the TM70 telemetry monitor: ≤4 seconds

45 dB(A) to 85 dB(A) within a range of one meter

A.11

Data Storage

Trends

Events

NIBP measurements

Interpretation of resting 12-lead ECG results

Full-disclosure waveforms

ST view

OxyCRG view

A minimum of 120 hours of trend data for trend interval as low as 1 minute.

1000 events

1000 sets

20 sets

48 hours at maximum. The specific storage time depends on the waveforms stored and the number of stored waveforms.

A maximum of 120 hours of ST segment waveforms. One group of ST segment waveforms is stored every minute.

A maximum of 48 hours of oxyCRG events

A.12

Wi-Fi Specifications

A.12.1

Wi-Fi Technical Specifications (MSD45N)

Protocol

Modulation mode

Operating frequency

Channel spacing

Wireless data rate

Output power

Operating mode

Data security

IEEE 802.11a/b/g/n

BPSK, QPSK, 16QAM, 64QAM

2.4 GHz to 2.495 GHz.

5.15 GHz to 5.25 GHz, 5.725 GHz to 5.85 GHz

IEEE 802.11b/g: 5 MHz

IEEE 802.11n (at 2.4 GHz): 5 MHz

IEEE802.11a: 20 MHz

IEEE802.11n (at 5 GHz): 20 MHz

IEEE 802.11b: 1 Mbps to 11 Mbps

IEEE 802.11g: 6 Mbps to 54 Mbps

IEEE 802.11n: 6.5 Mbps to 72.2 Mbps (MCS0-MCS7)

IEEE 802.11a: 6 Mbps to 54 Mbps

<20 dBm (CE requirements, detection mode: RMS)

<30 dBm (FCC requirements, detection mode: peak power)

As station, access AP for data transmission

Standards: WPA-PSK, WPA2-PSK, WPA-Enterprise, WPA2-Enterprise, WPA CCKM, WPA2

CCKM

EAP method: EAP-FAST, EAP-TLS, EAP-TTLS, PEAP-GTC, PEAP-MSCHAPv2, PEAP-TLS, LEAP

Encryption: AES

A.12.2

Wi-Fi Technical Specifications (SX-SDMAC-2832S+)

Protocol

Modulation mode

IEEE 802.11a/b/g/n

BPSK, QPSK, 16QAM, 64QAM

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual A - 7

Operating frequency

Channel spacing

Wireless data rate

Output power

Operating mode

Data security

2412 MHz to 2462 MHz

5180 MHz to 5320 MHz, 5500 MHz to 5700 MHz, 5745 MHz to 5825 MHz

WARNIGN: SX-SDMAC-2832S+ supports the DFS channels. When using the DFS channels, Wi-Fi performance stability and roaming time can be undermined due to avoiding interfering with Radar systems. DFS channels are disabled by default and not recommended. The operator should comprehensively assess the risk before using the DFS channels.

IEEE 802.11b/g: 5 MHz

IEEE 802.11n (at 2.4 GHz): 5 MHz

IEEE802.11a: 20 MHz

IEEE802.11n (at 5 GHz): 20 MHz

IEEE 802.11b: 1 Mbps to 11 Mbps

IEEE 802.11g: 6 Mbps to 54 Mbps

IEEE 802.11n: MCS0 - MCS7

IEEE 802.11a: 6 Mbps to 54 Mbps

<20 dBm (CE requirements, detection mode: RMS)

<30 dBm (FCC requirements, detection mode: peak power)

As station, access AP for data transmission

Standards: WPA-PSK, WPA2-PSK, WPA-Enterprise, WPA2-Enterprise, WPA CCKM, WPA2

CCKM

EAP method: EAP-FAST, EAP-TLS, EAP-TTLS, PEAP-GTC, PEAP-MSCHAPv2, PEAP-TLS, LEAP

Encryption: AES

A.12.3

Wi-Fi Performance Specifications

WARNING

All network functions of data communication are designed to operate on a private network.

A.12.3.1

System Capacity

Number of the monitors supported by a single AP: ≤ 12 (for N22/N19) or ≤ 16 (for N17/N15/N12).

Testing conditions are as follows:

Each monitor can communicate with the CMS.

Two monitors are used to view other monitors.

Only one monitor can transmit history data.

The weakest strength of the AP signal where the monitor is located cannot be less than -65 dBm.

A.12.3.2

Resistance to Wireless Interference

The distance between the interfering devices and the monitor is greater than 20 cm. A Wi-Fi interference (no greater than -85 dBm) in the same channel and a Wi-Fi interference (no greater than -50 dBm) in an adjacentchannel are presented synchronously. The interfering devices include, but are not limited to, 2.4 GHz wireless devices, cellular mobile networks, microwave ovens, intercoms, cordless phones, and ESU equipment. The interfering devices do not include Wi-Fi devices.

The monitoring network must support the following requirements:

All the monitors do not encounter communication loss.

The total delay of data transmission from the monitor to the CMS: ≤ 2 seconds.

The delay for monitor-related settings configured at the CMS to be effective: ≤ 2 seconds.

The total delay of data transmission from one monitor to the other: ≤ 2 seconds.

The delay for the monitor to reset alarms of another to be effective: ≤ 2 seconds.

A - 8 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

NOTE

When the monitor works at 2.4G, the most serious interference source is other 2.4G Wi-Fi devices not sharing the same network with the monitor and LTE devices. When the monitor works at 5G, the most serious interference source is other 5G Wi-Fi devices not sharing the same network with the monitor.

To make sure that the monitor works well, the separation distance between the monitor and unintended interference device or the distance between the AP and interference devices should be greater than the distance listed below:

Working frequency band

Interference devices

2.4G (2412-

2462MHz)

Adjacent channel Wi-Fi

Co-channel Wi-Fi lower adjacent band LTE upper adjacent band LTE

5G (5190 -5 210MHz) Adjacent channel Wi-Fi

Co-channel Wi-Fi

5G (5270 -5 290MHz) Adjacent channel Wi-Fi

Co-channel Wi-Fi

5G (5590 -5 610MHz) Adjacent channel Wi-Fi

Co-channel Wi-Fi

5G (5755 -5 775MHz) Adjacent channel Wi-Fi

Co-channel Wi-Fi

0.

5 m

3.0 m

0.

6 m

0.

6 m

2 .

5 m

4.0 m

2.0 m

3.5 m

2.0 m

3.5 m

2 .

5 m

4.0 m

Separation distance between the monitor and interference devices

Separation distance between the AP that the monitor is connected to and interference devices

2.0 m

3 .

5 m

2.0 m

3 .

5 m

2.0 m

3 .

5 m

2.0 m

3 .

5 m

0.

5 m

4.0 m

0.

8 m

0.

8 m

A.12.3.3

Wi-Fi Network Stability

The ratio of the communication data loss on the CMS from any monitor does not exceed 0.1% over a 24-hour period.

Testing conditions are as follows:

Number of the monitors supported by a single AP: ≤ 12 (for N22/N19) or ≤ 16 (for N17/N15/N12).

For N17/N15/N12, 12 of the 16 monitors connected to the network roam for 30 times

Each monitor can communicate with the CMS.

Two monitors are used to view other monitors.

Only one monitor can transmit history data.

The weakest strength of the AP signal where the monitor is located cannot be less than -65 dBm.

A.12.3.4

Line of Sight Distance

The line of sight distance between the monitor and the AP is no less than 50 meters.

A.13

Measurement Specifications

The adjustable range of alarm limits is the same with the measurement range of signals unless otherwise specified.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual A - 9

A.13.1

ECG Specifications

ECG

Standards

Lead set

ECG standard

Display sensitivity

Sweep speed

Bandwidth (-3dB)

Common mode rejection ratio

(with Notch filter off )

Notch filter

Differential input impedance

Input signal range

Accuracy of signal reproduction

Electrode offset potential tolerance

Lead-off detection current

Input offset current

Defibrillation protection

Patient leakage current

Calibration signal

ESU protection

Meets the requirements of IEC 60601-2-27: 2011 and IEC 60601-2-25: 2011

3-lead: I, II, III

5-lead: I, II, III, aVR, aVL, aVF, V

6-lead: I, II, III, aVR, aVL, aVF, Va, Vb

12-lead: I, II, III, aVR, aVL, aVF, V1 to V6

AHA, IEC

1.25 mm/mV (×0.125), 2.5 mm/mV (×0.25), 5 mm/mV (×0.5), 10 mm/mV (×1),

20 mm/mV (×2), 40 mm/mV (×4), Auto, less than 5% error

6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s, less than 5% error

Diagnostic mode:

Monitor mode:

Surgery mode:

ST mode:

High frequency cut-off (for 12lead ECG analysis):

Diagnostic mode:

Monitor mode:

Surgery mode:

ST mode:

0.05 to 150 Hz

0.5 to 40 Hz

1 to 20 Hz

0.05 to 40 Hz

350 Hz (0.05 to 350 Hz), 150 Hz (0.05 to 150

Hz), 35 Hz (0.05 to 35 Hz), 20 Hz (0.05 to 20

Hz) selectable

>90 dB

>105 dB (with notch filter on)

>105 dB (with notch filter on)

>105 dB (with notch filter on)

50/60 Hz

Monitor, surgical, and ST mode: notch filter turns on automatically

Diagnostic mode and High Freq Cut-off: notch filter is turned on/off manually

≥5 MΩ

±8 mV (peak-to-peak value)

Use A and D methods based on IEC 60601-2-25: 2011 to determine frequency response.

±500 mV

Measuring electrode: <0.1 μA

Drive electrode: <1 μA

≤0.1 μA, (drive lead≤1μA)

Withstands 5000V (360 J) charge without data loss or corruption

Baseline recovery time: <5 s

Polarization recovery time: <10 s

Defibrillation energy absorption: ≤10% (100 Ω load)

<10 uA

1mV (peak-to-peak value) ±5%

Cut mode: 300 W

Coagulate mode: 100 W

Recovery time: ≤10 s

In compliance with the requirements in clause 202.6.2.101 of IEC 60601-2-27:

2011

Pace Pulse

A - 10 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Pace pulse markers

Pace pulse rejection

Pace pulses meeting the following conditions are labelled with a PACE marker:

Amplitude:

Width:

Rise time:

No overshoot

±2 mV to ±700 mV

0.1 ms to 2 ms

10 μs to 100 μs (< 10% of pulse width)

When tested in accordance with the IEC 60601-2-27: 2011: 201.12.1.101.13, the heart rate meter rejects all pulses meeting the following conditions.

Amplitude:

Width:

Rise time:

No overshoot

±2 mV to ±700 mV

0.1 ms to 2 ms

10 μs to 100 μs (< 10% of pulse width)

HR

Measurement range

Resolution

Accuracy

Sensitivity

HR averaging method

Response to irregular rhythm

Response time to heart rate change

Time to alarm for tachycardia

Tall T-wave rejection capability

Arrhythmia Analysis Classifications

Neonate: 15 to 350 bpm

Pediatric: 15 to 350 bpm

Adult: 15 to 300 bpm

1 bpm

±1 bpm or ±1%, whichever is greater.

200 μV (lead II)

In compliance with the requirements in Clause 201.7.9.2.9.101 b) 3) of IEC

60601-2-27: 2011, the following method is used:

If the last 3 consecutive RR intervals are greater than 1200 ms, the 4 most recent RR intervals are averaged to compute the HR. Otherwise, heart rate is computed by subtracting the maximum and minimum ones from the most recent 12 RR intervals and then averaging the remaining 10 RR intervals.

The HR value displayed on the monitor screen is updated every second.

In compliance with the requirements in Clause 201.7.9.2.9.101 b) 4) of IEC

60601-2-27: 2011, the heart rates for the specified test signals after 20 seconds of stabilization are:

Ventricular bigeminy (waveform A1): 80±1 bpm

Slow alternating ventricular bigeminy (waveform A2): 60±1 bpm

Rapid alternating ventricular bigeminy (waveform A3): 120±1 bpm

Bidirectional systoles (waveform A4): 90±2 bpm

Meets the requirements of IEC 60601-2-27: 2011: Clause 201.7.9.2.9.101 b) 5).

From 80 to 120 bpm: less than 11 s

From 80 to 40 bpm: less than 11 s

Meets the requirements in Clause 201.7.9.2.9.101 b) 6) of IEC 60601-2-27: 2011.

Waveform

B1h-range:

B1-range:

B1d-range:

B2h-range:

B2-range:

B2d-range:

<11 s

<11 s

<11 s

<11 s

<11 s

<11 s

When the test is performed based on Clause 201.12.1.101.17 of IEC 60601-2-27:

2011, the heart rate calculation is not affected for QRS of 1 mV amplitude and

100 ms duration, T-wave duration of 180 ms and amplitude lower than 1.2 mV, and QT interval of 350 ms.

Asystole, V-Fib/V-Tach, V-Tach, Vent Brady, Extreme Tachy, Extreme Brady, Vent

Rhythm, PVCs/min, Pauses/min, Couplet, Bigeminy, Trigeminy, R on T, Run

PVCs, PVC, Tachy, Brady, Missed Beats, Pacer Not Pacing, Pacer Not Capture,

Multiform PVC, Nonsus V-Tach, Pause, Irr Rhythm, A-Fib

ST Segment Analysis

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual A - 11

Measurement range

Accuracy

Resolution

QT/QTc Analysis

Measurement range

-2.0 mV to +2.0 mV RTI

-0.8 mV to +0.8 mV: ±0.02 mV or ±10%, whichever is greater

Beyond this range: not specified

0.01mV

Accuracy

Resolution

12-lead ECG Interpretation

Sampling rate

Amplitude quantisation

PVCs/min

Measurement range

QT: 200 to 800 ms

QTc: 200 to800 ms

QT-HR: 15 bpm to 150 bpm for adult, 15 bpm to 180 bpm for pediatric and neonate

QT: ±30 ms

QT: 4 ms

QTc: 1 ms

1000 samples/s (A/D)

500 samples/s (ECG algorithm)

24 bits

Adult: 0 to 300

Pediatric, neonate: 0 to 350

Alarm limit

HR High

HR Low

ST High

ST Low

QTc High

ΔQTc High

A.13.2

Resp Specifications

Technique

Lead

Respiration excitation waveform

Minimum respiration impedance threshold

Baseline impedance range

Bandwidth

Sweep speed

Respiration Rate

Range

HR≤40bpm: (low limit + 2 bpm) to 40 bpm

HR > 40 bpm: (low limit + 5 bpm) to 295 bpm (Adult/Pediatric)

(low limit + 5 bpm) to 300 bpm (Neonate)

HR≤40bpm: 16 bpm to (high limit - 2 bpm) (Adult/Pediatric)

15 bpm to (high limit - 2 bpm) (Neonate)

HR > 40 bpm: 40 bpm to (high limit - 5 bpm)

(low limit + 0.2 mV) to 2.0 mV (ST alarm mode: Absolute)

0 mV to 2.0 mV (ST alarm mode: Relative)

-2.0 mV to (high limit - 0.2 mV) (ST alarm mode: Absolute)

-2.0 mV to 0 mV (ST alarm mode: Relative)

200 ms to 800 ms

30 ms to 200 ms

Step

HR≤40bpm: 1 bpm

HR > 40 bpm: 5 bpm

0.05 mV

10 ms

Trans-thoracic impedance

Options are lead I, II, and Auto.

<300 μA RMS, 62.8 kHz (±10%)

0.3Ω

200 to 2500 Ω (using an ECG cable with 1kΩ resistance)

0.2 to 2.5 Hz (-3 dB)

3mm/s, 6.25 mm/s, 12.5 mm/s, 25 mm/s or 50 mm/s, less than 10% error

A - 12 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Measurement range

Resolution

Accuracy

Apnea alarm delay

Alarm limit

RR High

RR Low

Adult: 0 to 120 rpm

Pediatric, neonate: 0 to 150 rpm

1 rpm

7 to 150 rpm: ±2 rpm or ±2%, whichever is greater

0 to 6 rpm: not specified

10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s

Range (rpm)

Adult, pediatric:

RR≤20

RR>20

Neonate:

RR≤20

RR>20

RR≤20:

RR>20:

(low limit + 2) to 20

(low limit + 5) to 100

(low limit + 2) to 20

(low limit + 5) to 150

0 to (high limit - 2)

20 to (high limit - 5)

Step (rpm)

RR≤20: 1

RR>20: 5

A.13.3

SpO

2

Specifications

Refer to Appendix F SpO

2

Sensor Accuracy for the clinical study results of SpO

2

sensor accuracy.

Alarm limit

SpO

2

High

SpO

2

Low

ΔSpO

2

High

SpO

2

Desat Low

Range (%) Step (%)

(low limit + 2) to 100

Masimo: (Desat+1) to (high limit - 2)

Nellcor: (Desat+1) or 20 (whichever is greater) to (high limit - 2)

0 to 50

0 to (SpO2 low limit - 1)

1

Masimo SpO

2

Module

Standards

Measurement range

Resolution

Response time

Accuracy

1

Refresh rate

SpO

2

averaging time

Low perfusion conditions

Low perfusion SpO

2

accuracy

2

PI measurement range

Meets the requirements of ISO 80601-2-61: 2017

1 to 100%

1%

≤20 s (normal perfusion, no disturbance, SpO2 value sudden change from 70% to 100%)

70 to 100%: ±2%ABS (adult/pediatric, without motion)

70 to 100%: ±3%ABS (adult/pediatric, with motion)

70 to 100%: ±3%ABS (neonate)

1% to 69%: Not specified.

≤ 1 s

2-4 s, 4-6 s, 8 s, 10 s, 12 s, 14 s, 16 s

Pulse amplitude: >0.02%

Light penetration: >5%

±2%

0.02 to 20%

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual A - 13

1

The Masimo pulse oximeter with sensors have been validated for no motion accuracy in human blood studies on healthy adult volunteers in induced hypoxia studies in the range of 70% to 100% SpO2 against a laboratory co-oximeter and ECG monitor. This variation equals plus or minus one standard deviation. Plus or minus one standard deviation encompasses

68% of the population. One percent was added to the accuracies for neonatal sensors to account for accuracy variation due to properties of fetal hemoglobin.

The Masimo pulse oximeter with sensors has been validated for motion accuracy in human blood studies on healthy adult volunteers in induced hypoxia studies while performing rubbing and tapping motions at 2 to 4 Hz. At an amplitude of 1 to

2 cm and non-repetitive motion between 1 to 5 Hz. At an amplitude of 2 to 3 cm in induced hypoxia studies in the range of

70% to 100% SpO2 against a laboratory co-oximeter and ECG monitor. This variation equals plus or minus one standard deviation. Plus or minus one standard deviation encompasses 68% of the population.

2

The Masimo pulse oximeter has been validated for low perfusion accuracy in bench top testing against a Biotek Index 2 simulator and Masimo's simulator with signal strengths of greater than 0.02% and a % transmission of greater than 5% for saturations ranging from 70 to 100%. This variation equals plus or minus one standard deviation. Plus or minus one standard deviation encompasses 68% of the population.

Nellcor SpO

2

Module

Standard

Measurement range

Resolution

Refresh rate

Meet the requirements of ISO 80601-2-61 2017

0 to 100%

1%

≤1 s

Response time

Accuracy

≤30 s (normal perfusion, no disturbance, SpO

2

value sudden changes from

70% to 100%)

70 to 100%: ±2%ABS (adult/pediatric)

70 to 100%: ±3%ABS (neonate)

0 to 69%: Not specified.

When the SpO

2

sensor is applied for neonatal patients as indicated, the specified accuracy range is increased by ±1%, to compensate for the theoretical effect on oximeter measurements of fetal hemoglobin in neonatal blood.

A.13.4

PR Specifications

Alarm limit

PR High

PR Low

Range (bpm)

PR≤40: (low limit + 2) to 40

PR > 40: (low limit + 5) to 295 (Adult/Pediatric),

(low limit + 5) to 300 (Neonate)

PR≤40: 16 to (high limit - 2) (Adult/Pediatric),

15 to (high limit - 2) (Neonate)

PR > 40: 40 to (high limit - 5)

PR from Masimo SpO

2

Measurement range

Resolution

Response time

Accuracy

Step (bpm)

PR≤40: 1

PR>40: 5

25 to 240 bpm

1 bpm

≤20 s (with normal perfusion, no disturbance, and a PR value transition from

25 to 220 bpm)

±3 bpm (without motion)

±5 bpm (with motion)

≤1 s Refresh rate

PR from Nellcor SpO

2

Measurement range

Resolution

20 to 300 bpm

1 bpm

A - 14 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Response time

Accuracy

Refresh rate

PR from NIBP

Measurement range

Resolution

Accuracy

PR from IBP

Measurement range

Resolution

Accuracy

≤30 s (with normal perfusion, no disturbance, and a PR value transition from

25 to 250 bpm)

20 to 250 bpm: ±3 bpm

251 to 300 bpm: not specified

≤1 s

30 to 300 bpm

1 bpm

±3 bpm or ±3%, whichever is greater

25 to 350 bpm

1 bpm

±1 bpm or ±1%, whichever is greater

A.13.5

Temp Specifications

Standard

Technique

Operating mode

Measurement range

Resolution

Accuracy

Minimum time for accurate measurement

Alarm limit

Txx High (xx refers to temperature site)

Txx Low (xx refers to temperature site)

ΔT High

Meets the requirements of ISO 80601-2-56: 2017

Thermal resistance

Direct mode

0 to 50 °C (32 to 122 °F)

0.1°C

±0.1 °C or ±0.2 °F (excluding probe error)

Body surface: <100 s

Body cavity: <80 s

Range

(low limit +1.0) to 50.0 °C

(low limit +2.0) to 122.0 °F

0.1 to (high limit - 1.0) °C

32.2 to (high limit - 2.0) °F

0.1 to 50.0 °C

0.2 to 90.0 °F

Step

0.1 °C

0.1 °F

A.13.6

NIBP Specifications

Standard

Technique

Mode of operation

Auto mode repetition intervals

STAT mode cycle time

Max measurement time

Heart rate range

Meets the requirements of IEC 80601-2-30: 2018

Oscillometry

Manual, Auto, STAT, Sequence

1, 2, 2.5, 3, 5, 10, 15, 20, 30, 60, 90, 120, 180, 240 or 480 min

5 min

Adult, pediatric: 180 s

Neonate: 90 s

30 to 300 bpm

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual A - 15

A - 16

Measurement ranges

(mmHg)

Accuracy

Resolution

Initial cuff inflation pressure range

(mmHg)

Default initial cuff inflation pressure

(mmHg)

Re-inflation increment (mmHg)

Software overpressure protection

Static pressure measurement range

Static pressure measurement accuracy

Alarm limit

NIBP-S High

NIBP-S Low

NIBP-M High

NIBP-M Low

NIBP-D High

NIBP-D Low

NIBP-S Extreme High

NIBP-S Extreme Low

Systolic:

Diastolic:

Mean:

Max mean error: ±5 mmHg

Max standard deviation: 8 mmHg

1mmHg

Adult: 80 to 280

Pediatric: 80 to 210

Neonate: 60 to 140

Adult: 160

Pediatric: 140

Neonate: 90

Adult/Pediatric: 40

Neonate: 15

Adult: 297±3 mmHg

Pediatric: 297±3 mmHg

Neonate: 147±3 mmHg

0 mmHg to 300 mmHg

±3 mmHg

Adult

25 to 290

10 to 250

15 to 260

Range (mmHg)

Adult: (low limit + 5) to 285

Pediatric: (low limit + 5) to 235

Neonate: (low limit + 5) to 135

26 to (high limit - 5)

Adult: (low limit + 5) to 255

Pediatric: (low limit + 5) to 210

Neonate: (low limit + 5) to 120

16 to (high limit - 5)

Adult: (low limit + 5) to 245

Pediatric: (low limit + 5) to 195

Neonate: (low limit + 5) to 110

11 to (high limit - 5)

NIBP-S high limit < 50

Adult: (NIBP-S high limit + 1) to 290

Pediatric: (NIBP-S high limit + 1) to 240

Neonate: (NIBP-S high limit + 1) to 140

NIBP-S high limit ≥ 50

Adult: (NIBP-S high limit + 5) to 290

Pediatric: (NIBP-S high limit +5) to 240

Neonate: (NIBP-S high limit +5) to 140

NIBP-S low limit ≤ 50

25 to (NIBP-S low limit - 1)

NIBP-S low limit > 50

25 to (NIBP-S low limit - 5)

Pediatric

25 to 240

10 to 200

15 to 215

Step (mmHg)

NIBP ≤ 50: 1

NIBP > 50: 5

NIBP ≤ 50: 1

NIBP > 50: 5

Neonate

25 to 140

10 to 115

15 to 125

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

NIBP-M Extreme High

NIBP-M Extreme Low

NIBP-D Extreme High

NIBP-D Extreme Low

NIBP-M high limit < 50

Adult: (NIBP-M high limit + 1) to 260

Pediatric: (NIBP-M high limit + 1) to 215

Neonate: (NIBP-M high limit + 1) to 125

NIBP-M high limit ≥ 50

Adult: (NIBP-M high limit + 5) to 260

Pediatric: (NIBP-M high limit + 5) to 215

Neonate: (NIBP-M high limit + 5) to 125

NIBP-M low limit ≤ 50

15 to (NIBP-M low limit - 1)

NIBP-M low limit > 50

15 to (NIBP-M low limit - 5)

NIBP-D high limit < 50

Adult: (NIBP-D high limit + 1) to 250

Pediatric: (NIBP-D high limit + 1) to 200

Neonate: (NIBP-D high limit + 1) to 115

NIBP-D high limit ≥ 50

Adult: (NIBP-D high limit + 5) to 250

Pediatric: (NIBP-D high limit + 5) to 200

Neonate: (NIBP-D high limit + 5) to 115

NIBP-D low limit ≤ 50

10 to (NIBP-D low limit - 1)

NIBP ≤ 50: 1

NIBP > 50: 5

NIBP ≤ 50: 1

NIBP > 50: 5

NIBP-D low limit > 50

10 to (NIBP-D low limit - 5)

*Measurement accuracy verification: In adult and pediatric modes, the blood pressure measurements measured with this device are in compliance with the Standard for Non-invasive sphygmomanometers (ISO 81060-2: 2013) in terms of mean error and standard deviation by comparing with intra-arterial or auscultatory measurements

(depending on the configuration) in a typical patient population. For auscultatory reference, the 5th Korotkoff sound was used to determine the diastolic pressure.

In neonatal mode, the blood pressure measurements measured with this device are in compliance with the

American National Standard for Non-invasive sphygmomanometers (ISO 81060-2) in terms of mean error and standard deviation by comparing with intra-arterial measurements (depending on the configuration) in a typical patient population.

A.13.7

IBP Specifications

Standard

Technique

IBP

Measurement range

Resolution

Accuracy

PPV

Measurement range

Pressure transducer

Excitement voltage

Sensitivity

Meets the requirements of IEC 60601-2-34: 2011

Direct invasive measurement

-50 to 300 mmHg

1 mmHg

±2% or ±1 mmHg, whichever is greater (excluding sensor error)

0% ~ 50%

5 VDC, ±2%

5 μV/V/mmHg

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual A - 17

Zero adjustment range

Impedance range

Volume displacement

Alarm limit

Sys High

Mean High

Dia High

Sys Low

Mean Low

Dia Low

Art-S Extreme High

Art-M Extreme High

Art-D Extreme High

Art-S Extreme Low

Art-M Extreme Low

Art-D Extreme Low

A.13.8

C.O. Specifications

Measurement method

Measurement range

±200 mmHg

300 to 3000Ω

<0.04 mm

3

/100 mmHg

Range (mmHg)

IBP ≤ 50: (low limit + 2) to 50

IBP > 50: (low limit + 5) to 295

IBP ≤ 50: -49 to (high limit - 2)

IBP > 50: 50 to (high limit - 5)

High limit < 50: (High limit+ 1) to 300

High limit ≥ 50: (High limit+ 5) to 300 low limit ≤ 50: -50 to (low limit- 1) low limit > 50: -50 to (low limit- 5)

Step (mmHg)

IBP ≤ 50: 1

IBP > 50: 5

IBP ≤ 50: 1

IBP > 50: 5

TB Operating mode

Minimum time for accurate TB measurement

Resolution

Accuracy

Repeatability

Alarm range

Alarm limit

TB High

TB Low

Thermodilution method

C.O.:

TB:

TI:

0.1 to 20 L/min

23 to 43 °C

0 to 27 °C

Direct mode

10 s

C.O.:

TB, TI:

C.O.:

TB, TI:

C.O.:

TB:

Range

(low limit + 1) to 43 °C

(low limit + 2) to 109.4 °F

23 to (high limit - 1) °C

73.4 to (high limit - 2) °F

0.1 L/min

0.1 °C (excluding sensor)

±5% or ±0.1 L /min, whichever is greater

±0.1 °C (excluding sensor)

±2% or ±0.1 L/min, whichever is greater

23 to 43 °C

Step

0.1 °C

0.1 °F

A.13.9

ScvO2/SvO2 Specifications from the Edwards Monitor

Operating mode

Measured parameters

Parameter alarms

Interfaces with Edwards Vigilance II®, Vigileo TM monitor, EV1000, or

HemoSphere monitor

Consistent with SvO

2

/ScvO

2

-related parameters output by Vigilance II®,

Vigileo TM monitor, EV1000, or HemoSphere monitor

SvO

2

, ScvO

2

A - 18 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Signal Output

Standard

Output impedance

Isolation voltage

SpO

2

Analog Signal Output

Output voltage

Output voltage error

Meets the requirements of IEC 60601-1 for short-circuit protection and leakage current

≤1000Ω

1500 VAC

0 to 10V (0 to 100%)

±5%

Alarm Limit

SvO

2

/ScvO

2

High

SvO

2

/ScvO

2

Low

Range

ScvO

2

<60%: (Low limit + 5%) to 60%

ScvO

2

≥60%: (Low limit + 1%) to 99%

ScvO

2

<60%: 0 to (High limit - 5%)

ScvO

2

≥60%: 60% to (Low limit - 1%)

Step

ScvO2<60%: 5%

ScvO2≥60%: 1%

A.13.10 CCO Specifications from the Edwards Monitor

Operating mode

Measured parameter

Parameter alarm

Interfaces with Edwards Vigilance II®, Vigileo

TM

monitor, EV1000, or

HemoSphere monitor

Consistent with CCO-related parameters output by Vigilance II®, Vigileo

TM monitor, EV1000, or HemoSphere monitor

CCO, CCI

Signal Outputs

Standard

Output impedance

Isolation voltage

ECG Analog Output

Bandwidth (-3dB; reference frequency:

10Hz)

Meets the requirements of IEC 60601-1 for short-circuit protection and leakage current

≤1000Ω

1500 VAC

Diagnostic mode: 0.05~150Hz

Monitor mode: 0.5~40Hz

Surgery mode: 1~20Hz

ST mode: 0.05~40Hz

2 V/mV ±5% Sensitivity

MAP Analog Signal Output

Output voltage

Output voltage error

CVP Analog Signal Output

Output voltage

Output voltage error

Alarm Limit

0 to 5 V (0 to 500 mmHg)

±5%

0 to 5 V (0 to 100 mmHg)

±5%

Range Step

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual A - 19

CCO High

CCO Low

CCI High

CCI Low

(Low limit + 0.1) to 25 L/min

0.3 to (high limit - 0.1)L/min

(Low limit + 0.1) to 15 L/min/m

2

0.1 to (high limit - 0.1)L/min/m

2

0.1 L/min

0.1 L/min/m

2

A.13.11 ScvO

2

Specifications

Measured parameters

ScvO

2

Alarm Limit

ScvO

2

High

ScvO

2

Low

Measurement range

0 to 99%

Range

ScvO

2

<60%: (Low limit + 5%) to 60%

ScvO

2

≥60%: (Low limit + 1%) to 99%

ScvO

2

<60%: 0 to (High limit - 5%)

ScvO

2

≥60%: 60% to (High limit - 1%)

Measurement accuracy

50% to 80%: ±3%

Other ranges: Not specified.

Step

ScvO

2

<60%: 5%

ScvO

2

≥60%: 1%

A.13.12 CO

2

Specifications

Measurement mode

Technique

Apnea delay

Alarm limit

Sidestream, microstream

Infrared absorption

10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s

Range

EtCO

2

High

EtCO

2

Low

FiCO

2

High

Sidestream CO

2

Module

(low limit + 2) to 99 mmHg

1 to (high limit - 2) mmHg

1 to 99 mmHg

Step

1 mmHg

Standard Meets the requirements of ISO 80601-2-55: 2018

0 to 150 mmHg CO

2

Measurement range

CO

2

absolute accuracy* Full accuracy mode:

0 - 40 mmHg: ±2mmHg

41 - 76 mmHg: ±5% of reading

77 - 99 mmHg: ±10% of reading

100 - 150 mmHg: ±(3mmHg + 8% of reading

ISO accuracy mode: Add ±2mmHg to the full accuracy mode

Accuracy specifications are affected by the breath rate and I:E. The EtCO

2

accuracy is within specification for breath rate ≤

60 rpm and I/E ratio ≤ 1:1, or breath rate ≤ 30 rpm and I/E ratio ≤ 2:1.

CO

2 resolution

Accuracy drift

Sample flowrate

Sample flowrate tolerance

Start-up time

1 mmHg

Meets the requirements for measurement accuracy within 6 hours

Connected a DRYLINE II watertrap for adult and pediatric patient: 120 ml/min

Connected a DRYLINE II watertrap for neonatal patient: 70 ml/min, 90 ml/min

±15% or ±15 ml/min, whichever is greater.

Maximum: 90 s

Typically: 20 s

A - 20 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Rise time

Response time

Measured with a DRYLINE II neonatal watertrap and a 2.5-meter neonatal sampling line:

≤250 ms@70 ml/min

≤250 ms@90 ml/min

Measured with a DRYLINE II adult watertrap and a 2.5-meter adult sampling line:

≤300 ms@120 ml/min

Measured with a DRYLINE II neonatal watertrap and a 2.5-meter neonatal sampling line:

≤5.0 s @ 70 ml/min

≤4.5 s @ 90 ml/min

Measured with a DRYLINE II adult watertrap and a 2.5-meter adult sampling line:

≤5.0 s @ 120 ml/min awRR measurement range awRR measurement precision awRR resolution

Data sample rate

0 to 150 rpm

≤60 rpm: ±1 rpm

6 1 to 150 rpm: ±2 rpm

1 rpm

50 Hz

Effect of interference gases on CO

2 measurements

Gas Concentration (%) Quantitative effect*

N

2

O

Hal

Sev

Iso

≤60

≤4

≤5

≤5

±1 mmHg

Enf

Des

≤5

≤15 ±2 mmHg

*: means an extra error should be added in case of gas interference when CO

2

measurements are performed between 0 to

40 mmHg.

Microstream CO

2

Module

Standard

CO

2

Measurement range

Accuracy*

Meets the requirements of ISO 80601-2-55: 2018

0 to 99 mmHg

0 to 38 mmHg:

39 to 99 mmHg:

±2 mmHg

±5% of the reading (0.08% of the reading increased in error for every 1 mmHg if the reading is more than 38 mmHg)

Accuracy drift Meets the requirements for measurement accuracy within 6 hours

* Accuracy applies for respiration rate up to 80 rpm. For respiration rate above 80 rpm and EtCO

2

exceeding 18 mmHg, the accuracy is 4 mmHg or ±12% of the reading, whichever is greater. For respiration rate above 60 rpm, the above accuracy can be achieved by using the CapnoLine H Set for Infant/Neonatal. In the presence of interfering gases, the above accuracy is maintained to within 4%.

Resolution

Sample flow rate

1 mmHg

50 (- 7.5/+15) ml/min

Initialization time Maximum 180 s

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual A - 21

Response time awRR measurement range awRR measurement accuracy awRR resolution

Data sample rate

A.13.13 AG Specifications

Standard

Technique

Warm-up time

Sample flow rate

Measurement range

Resolution

Iso accuracy

2.9 s (typical)

(The response time is the sum of the rise time and the delay time when using a FilterLine of standard length)

Rise time: <190 ms (10% to 90%)

Delay time: 2.7 s (typical)

0 to 150 rpm

0 to 70 rpm:

71 to 120 rpm:

121 to 150 rpm:

1 rpm

40 Hz

±1 rpm

±2 rpm

±3 rpm meets the requirements of ISO 80601-2-55: 2018

Infrared absorption, paramagnetic properties for O

2

monitoring

Iso accuracy mode:

Full accuracy mode:

≤45 s

≤10 min

Adult, pediatric:

Neonate:

Accuracy:

CO

2

:

O

2

:

N

2

O:

Des:

Sev:

Enf:

Iso:

Hal: awRR:

200 ml/min

120 ml/min

±10 ml/min or ±10%, whichever is greater

0 to 30%

0 to 100%

0 to 100%

0 to 30%

0 to 30%

0 to 30%

0 to 30%

0 to 30%

2 to 100 rpm

CO

2

:

O

2

:

N

2

O:

Des:

Sev:

Enf:

Iso:

Hal: awRR:

0.1%

1%

1%

0.1%

0.1%

0.1%

0.1%

0.1%

1 rpm

As full accuracy specifications, but derated as follows:

Add ±0.3%

ABS

to accuracy for CO

2

Add ±8%

REL

to accuracy for all anesthetic gases

N

2

O accuracy is ±(8%

REL

+2%

ABS

)

A - 22 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Full accuracy

Accuracy drift

Apnea alarm delay

Refresh rate

Rise time

(10% ~ 90%)

Delay time

Gases

CO

O

2

Sev

2

N2O

Des

Enf, Iso, Hal

Range (%REL)

0≤CO

2

≤1

1<CO

2

≤5

5<CO

2

≤7

7<CO

2

≤10

CO

2

>10

0≤NO

2

≤20

20<NO

2

≤100

0 to 25

25 to 80

80 to 100

0 to 1

1 to 5

5 to 10

10 to 15

15 to 18

>18

0 to 1

1 to 5

5 to 8

>8

0 to 1

1 to 5

>5

1

Accuracy (%ABS)

±0.1

±0.2

±0.3

±0.5

Not specified

±2

±3

±1

±2

±3

±0.15

±0.2

±0.4

±0.6

±1

Not specified

±0.15

±0.2

±0.4

Not specified

±0.15

±0.2

Not specified awRR 2 to 60 rpm

>60 rpm

±1 rpm

Not specified

Note

1

: The highest GAS LEVEL for a single halogenated anaesthetic gas in a gas mixture that is concealed when the anaesthetic concentration falls is 0.15/0.3%

(Full/ISO accuracy).

Meet the requirement for measurement accuracy within 6 hours

10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s

1 s

Gas sample flow rate 120 ml/min, using a DRYLINE II neonatal watertrap and sampling line (2.5 m):

CO

2

: ≤250 ms

N

2

O: ≤250 ms

Hal, Iso, Sev, Des: ≤300 ms

Enf: ≤350 ms

O

2

: ≤600 ms

Gas sample flow rate 200 ml/min, using the adult DRYLINE II water trap and sampling line (2.5 m):

CO

2

: ≤250 ms

N

2

O: ≤250 ms

Hal, Iso, Sev, Des: ≤300 ms

Enf: ≤350 ms

O

2

: ≤500 ms

<4 s

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual A - 23

A - 24

Response time

Anesthetic agent limit

Measured with a DRYLINE II neonatal watertrap and a 2.5-meter neonatal sampling line:

120 ml/min:

CO

2

:

N

2

O:

O

2

:

Hal, Iso, Sev, Des, Enf:

≤4s

≤4.2s

≤4s

≤4.4s

Measured with a DRYLINE II adult watertrap and a 2.5-meter adult sampling line:

200 ml/min:

CO

2

:

N

2

O:

Hal, Iso, Sev, Des, Enf:

O

2

:

Primary anesthetic agent

In full accuracy mode: 0.15%

≤4.2s

≤4.3s

≤4.5s

≤4s

Secondary anesthetic agent:

In full accuracy mode: 5% of primary agent if primary agent is greater than 10%,

0.3% if primary agent is less than or equal to 10%.

Data sample rate 25 Hz

Accuracy specifications are affected by the breath rate and I:E change. The end-tidal gas reading is within specification for breath rate below 15 rpm and I:E ratio smaller than 1:1 relative to the gas readings without breath; Add ±6%REL to inaccuracy for Hal and O

2

for breath rate larger than 15 rpm; Add ±6%REL to inaccuracy for all gases for breath rate larger than 30 rpm (inaccuracy for HAL and O

2

are unspecified in this case); inaccuracy is unspecified for breath rate larger than

60 rpm.

Effect of interference gases on AG measurements

(low limit + 2) to 99 mmHg

1 to (high limit - 2) mmHg

0 to 99 mmHg

Quantitative effect(%ABS)3)

Gas

Alarm limit

EtCO

2

High

EtCO

2

Low

FiCO

2

High

Concentration(

%)

CO

2

N

2

O Agent 1)

CO

2

N

2

O

Agent 1) 2)

/

/

/

0.1

0.1

/

0

0.1

0.2

0.2

/

<100%

0.1

0.1

0.1

0

0.1

0

1

0.5

Xenon

Helium <50%

<0.1%

0.1

0

0

0

0

0

0.5

0.5

Ethanol

Acetone <1%

<1%

0.1

0.1

0.1

0.1

0

0

0.5

0.5

Methane

Saturated Isopropanol vapour /

/

0.1

Unspecified

0

Unspecified

0

Unspecified

0.5

Unspecified Metered dose inhaler propellants

O

2

/ 0.2

0.2

1.0

/

1) Agent represents one of Des, Iso, Enf, Sev, and Hal.

2) Multiple agent interference on CO

2

, N

2

O and O

2

is typically the same as single agent interference.

3) For CO

2

, N

2

O and Agents, maximum interference from each gas at concentrations within specified accuracy ranges for each gas. The total interference of all gases is never larger than 5%REL.

Range

O

2

Step

1 mmHg

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

EtO

2

High

EtO

2

Low

FiO

2

High

FiO

2

Low

EtN

2

O High

EtN

2

O Low

FiN

2

O High

FiN

2

O Low

EtHal/Enf/Iso High

EtHal/Enf/Iso Low

FiHal/Enf/Iso High

FiHal/Enf/Iso Low

EtSev High

EtSev Low

FiSev High

FiSev Low

EtDes High

EtDes Low

FiDes High

FiDes Low

A.13.14 EEG Specifications

Standard

Bandwidth

Input Signal Range

Max. DC Input Offset

Common Mode Rejection Ratio

Noise Level

Differential Input Impedance

Electrode Impedance

(low limit + 2%) to 100%

0 to (high limit - 2)%

(low limit + 2%) to 100%

18% to (high limit - 2)%

(low limit + 2) to 100%

0 to (high limit - 2)%

(low limit + 2) to 100%

0 to (high limit - 2)%

(low limit + 0.2) to 5.0%

0 to (high limit - 0.2)%

(low limit + 0.2) to 5.0%

0 to (high limit - 0.2)%

(low limit + 0.2) to 8.0%

0 to (high limit - 0.2)%

(low limit + 0.2) to 8.0%

0 to (high limit - 0.2)%

(low limit + 0.2) to 18.0%

0 to (high limit - 0.2)%

(low limit + 0.2) to 18.0%

0 to (high limit - 0.2)%

1%

1%

0.1%

0.1%

0.1% meets the requirements of IEC 60601-2-26: 2012

0.5 to 50 Hz (-3 dB)

-2 mVp-p to +2 mVp-p

± 500 mV

≥100 dB @51 kΩ imbalance and 60 Hz

≤0.5 uV rms (1 Hz to 30 Hz)

>15 MΩ @10 Hz

Range:

Accuracy:

0 to 90 kΩ,

± 1 kΩ or ± 10%, whichever is greater

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual A - 25

A.13.15 BIS Specifications

Standard

Technique

Measured parameters

Calculated parameters

Impedance range

Sweep speed

Input impedance

Noise (RTI)

Input signal range

EEG bandwidth

Patient leakage current

Alarm limit

BIS High

BIS Low

Meet the standard of IEC 60601-2-26: 2012

Bispectral index

EEG

BIS, BIS L, BIS R: 0 to 100

SQI, SQI L, SQI R:0 to 100%

EMG, EMG L, EMG R: 0 to 100 dB

SR, SR L, SR R: 0 to 100%

SEF, SEF L, SEF R: 0.5 to 30.0 Hz

TP, TP L, TP R: 40 to 100 dB

BC, BC L, BC R: 0 to 30 sBIS L, sBIS R: 0 to 10.0

sEMG L, sEMG R: 0 to 10.0

ASYM: 0 to 100%

0 to 999 kΩ

6.25 mm/s, 12.5 mm/s, 25 mm/s or 50 mm/s, ±10% error

>5 MΩ

<0.3 μV (0.25 to 50 Hz)

±1 mV

0.25 to 100 Hz

<10 μA

Range

(low limit + 5) to 100

0 to (high limit - 5)

Step

5

A.13.16 NMT Specifications (NMT from Mindray NMT module)

Standard

Stimulation output

ST mode

TOF mode

PTC mode meets the requirements of IEC 60601-2-10: 2016

Pulse width 100, 200, or 300 μs; monophasic rectangle pulse

Accuracy: ±10%

Current range 0 to 60 mA in increments of 5 mA

Accuracy: ± 5% or ±2 mA, whichever is greater

Max. skin impedance

Max. output voltage

ST-Ratio

Measurement interval

3 kΩ @ 60 mA, 5 kΩ @ 40 mA

300 V

0 to 200%

Manual, 1 s, 10 s, 20 s

TOF-Count

TOF-Ratio

TOF-T1%

Measurement interval

PTC-Count

Measurement interval

0 to 4

5% to 160%

0 to 200%

Manual, 12s, 15s, 20s, 30s, 1min, 5 min, 15 min, 30 min, 60 min

0 to 20

Manual

A - 26 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

DBS mode Measurement interval

DBS-Count

DBS-Ratio

Manual, 15s, 20s, 30s, 1min, 5 min, 15 min,

30 min, 60 min

0 to 2

5% to 160%

Threshold

Off, 1, 2, 3, 4, 5%, 10%, 20%, 30%, 40%, 50%, 60%, 70%, 80%, 90%, 100%

NMT message

Block Recovery

A.13.17 rSO

2

Specifications

Measurement range 15% to 95%

A.14

Software Specifications

Mindray’s software

OTS software

The BeneVision N12/N15/N17/N19/N22 patient monitor system software

BeneVision N22/N19 OS: Linux 3.8.0

BeneVision N17/N15/N12 OS: Linux 3.2.0

iView OS: Windows 10

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual A - 27

This page intentionally left blank.

A - 28 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

B

EMC and Radio Regulatory Compliance

B.1

EMC

The device meets the requirements of IEC 60601-1-2: 2014.

WARNING

Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.

The non-ME EQUIPMENT (e.g. information technology equipment (ITE)) that is a part of an ME

SYSTEM may be disrupted by the electromagnetic interference of nearby equipment. It may be necessary to take mitigation measures, such as re-orienting or relocating the non-ME EQUIPMENT or shielding the location.

Use of this equipment adjacent to, or stacked with, other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.

This equipment is intended for use in professional healthcare facility environment only. If it is used in a special environment, such as a magnetic resonance imaging environment, the equipment/ system may be disrupted by the operation of nearby equipment.

Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the this device, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.

Guidance and Declaration - Electromagnetic Emissions

The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment.

Emission tests Compliance Electromagnetic environment - guidance

Conducted and radiated RF

EMISSIONS CISPR 11

Conducted and radiated RF

EMISSIONS CISPR 11

Harmonic distortion

EMISSIONS IEC61000-3-2:

2014

Voltage Fluctuations/Flicker

EMISSIONS IEC61000-3-3:

2013

Group 1

Class A

Class A

Complies

The device uses RF energy only for its internal function.

Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.

The device is suitable for use in all establishments other than domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes

If the system is operated within the electromagnetic environment listed in Table Guidance and Declaration -

Electromagnetic Immunity, the system will remain safe and provide the following essential performance:

Operating mode

Accuracy

Function

Accessories identification

Data stored

Alarm

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual B - 1

B - 2

■ Detect for connection

NOTE

If the essential performance is lost or degraded, it may be necessary to take mitigation measures, such as re-orienting or relocating the ME EQUIPMENT or ME SYSTEM or shielding the location.

During this time, the user should stop using the monitor and contact the service personnel.

The device needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided below.

Other devices may interfere with this device even though they meet the requirements of CISPR.

When an input signal is below the minimum amplitude provided in technical specifications, erroneous measurements could result.

The EMISSIONS characteristics of this device make it suitable for use in industrial areas and hospitals

(CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B is normally required) this device might not offer adequate protection to radio-frequency communication services. The user might need to take mitigation measures, such as relocating or re-orienting the device.

Guidance and Declaration - Electromagnetic Immunity

The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment.

Immunity test IEC60601test level Compliance level Electromagnetic environment - guidance

Electrostatic discharge (ESD) IEC

61000-4-2: 2008

Electrical fast transient/burst IEC

61000-4-4: 2012

Surge IEC 61000-4-5:

2014

Voltage dips and voltage interruptions

IEC 61000-4-11: 2004

±8 kV contact

±15 kV air

±2 kV for power supply lines

±1 kV for input/ output lines

(length greater than 3 m)

±1 kV line(s) to line(s)

±2 kV line(s) to earth

0 % U

T

for 0.5 cycle: at

0°, 45°, 90°, 135°, 180°,

225°, 270° and 315°

0 % U

T

for 1 cycle and

70 % U

T

for 25/30 cycles: at 0°

±8 kV contact

±15 kV air

±2 kV for power supply lines

±1 kV for input/ output lines

(length greater than 3 m)

±1 kV line(s) to line(s)

±2 kV line(s) to earth

0 % U

T

for 0.5 cycle: at

0°, 45°, 90°, 135°, 180°,

225°, 270° and 315°

0 % U

T

for 1 cycle and

70 % U

T

for 25/30 cycles: at 0°

Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.

Mains power quality should be that of a typical commercial or hospital environment.

Mains power quality should be that of a typical commercial or hospital environment. If the user of our product requires continued operation during power mains interruptions, it is recommended that our product be powered from an uninterruptible power supply or a battery.

RATED power frequency magnetic fields

IEC 61000-4-8: 2009

0 % U

T

for 250/300 cycle

30 A/m

50 Hz/60 Hz

0 % U cycle

T

30 A/m

for 250/300

50 Hz/60 Hz

Note: U

T

is the AC mains voltage prior to application of the test level.

Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.

Guidance and Declaration - Electromagnetic Immunity

The device is intended for use in the specified electromagnetic environment. The customer or the user of the device should assure that it is used in such an environment as described below.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Guidance and Declaration - Electromagnetic Immunity

Immunity test IEC60601 test level Compliance level Electromagnetic environment - guidance

Conducted disturbances induced by RF fields

IEC61000-4-6: 2013

Radiated RF EM fields

IEC61000-4-3:

2006+A1:2007+A2:20

10

3 Vrms

150 kHz to 80 MHz

6 Vrms in ISM bands a between 0.15 MHz and 80 MHz

3 V/m

80 MHz to 2.7 GHz

3Vrms

6 Vrms

3V/m

Portable and mobile RF communications equipment should be used no closer to any part of the system, including cables, than the recommended separation distance calculated from the equation appropriate for the frequency of the transmitter. Recommended separation distances: d = 1.2

P

Recommended separation distances:

80 MHz to 800 MHz: d = 1.2

P

800MHz - 2.7GHz: d = 2.3

P

Where, P is the maximum output power rating of the transmitter in Watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters

(m) b

.

Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey a

should be less than the compliance level in each frequency range b

.

Interference may occur in the vicinity of equipment marked with the following symbol:

Proximity fields from

RF wireless communications equipment

IEC61000-4-3:

2006+A1:2007+A2:20

10

27 V/m

380 - 390 MHz

28 V/m

430 - 470 MHz,

800 - 960 MHz

1700 - 1900 MHz,

2400 - 2570 MHz

27 V/m

28 V/m

/

9 V/m

704 - 787 MHz,

5100 - 5800 MHz

9 V/m

Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.

Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

a

: The ISM (industrial, scientific, and medical) bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.

b

: Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered.

If the measured field strength in the location in which the ME EQUIPMENT or ME SYSTEM is used exceeds the applicable

RF compliance level above, the ME EQUIPMENT or ME SYSTEM should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the ME EQUIPMENT or

ME SYSTEM .

c

: Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.

WARNING

The device is configured with a wireless network connector to receive wireless signal. Other devices may interfere with this device even though they meet the requirements of CISPR.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual B - 3

B.2

Recommended separation distances between portable and mobile RF communications equipment and the device

The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the device as recommended below, according to the maximum output power of the communications equipment.

Separation distance in meters (m) according to frequency of the transmitter

Rated maximum output power of transmitter (W)

150 kHz to 80 MHz d = 1.2

P

80 MHz to 800 MHz d = 1.2

P

800 MHz to 2.7 GHz d = 2.3

P

0.01

0.1

1

10

0.12

0.38

1.20

3.80

0.12

0.23

0.38 0.73

1.20 2.30

3.80

7.30

100 12.00

12.00 23.00

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres

(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in Watts (W) according to the transmitter manufacturer.

Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.

Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

Radio Regulatory Compliance

This device complies with Part 15 of the FCC Rules and RSS-210 of Industry Canada. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.

Any changes or modifications to this equipment not expressly approved by Mindray may cause harmful radio frequency interference and void your authority to operate this equipment.

The maximum antenna gain permitted complies with the e.i.r.p. limits as stated in RSS-210.

The maximum antenna gain permitted complies with the e.i.r.p. limits specified for point-to-point operation, as stated in RSS-210.

WARNING

Changes or modifications not expressly approved by the party responsible compliance could void the user’s authority to operate the equipment.

B - 4 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

C

Default Settings

C.1

C.1.1

ECG, Arrhythmia, ST and QT Default Settings

ECG Default Settings

Item

HR Alarm switch

High limit

Priority

Alarm Outputs

Alarm Source

ECG1

ECG2 (5-lead, 6-lead, 12-lead)

Va (for 6-lead only)

Vb(for 6-lead only)

ECG Gain

Speed

Filter

Notch Filter

Lead Set

Smart Lead

QRS Volume

Baseline Drift Removal (for 12-lead only)

Waveform Layout (for 12-lead only)

QRS Threshold

Paced

Pacer Reject

Low limit

Default Setting

On

Adult: 120 bpm

Pediatric: 160 bpm

Neonate: 200 bpm

Adult: 50 bpm

Pediatric: 75 bpm

Neonate: 100 bpm

Med

Off

Auto

II

V (for 5-lead).Va (for 6-lead), V1 (for 12-lead)

Va

Vb

×1

25 mm/sec

OR: Surgery

CCU: Diagnostic

Other departments: Monitor

On

Auto

On

General and OR: 2

Other departments: 0

On

Standard

0.16 mV

Adult: Unspecified

Pediatric/neonate: No

Off

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual C - 1

C.1.2

Arrhythmia Default Settings

C.1.2.1

Arrhythmia Alarm Default Settings

Item Alarm Switch

Asystole

V-Fib/V-Tach

V-Tach

Vent Brady

Extreme Tachy

Extreme Brady

R on T

Run PVCs

Couplet

Multiform PVC

PVC

Bigeminy

Trigeminy

Tachy

Brady

Pacer Not Capture

Pacer Not Pacing

Missed Beat

Nonsus V-Tach

Vent Rhythm

Pause

Irr Rhythm

A-Fib

PVCs/min

Pauses/min

On

On

On

On

On

On

CCU: On

Other departments: Off

Off

Off

Off

Off

CCU: On

Other departments: Off

CCU: On

Other departments: Off

Off

Off

Off

Off

Off

CCU: On

Other departments: Off

CCU: On

Other departments: Off

Off

Off

Off

CCU: On

Other departments: Off

CCU: On

Other departments: Off

Med

Low

Prompt

Prompt

Med

Med

Low

Prompt

Med

Prompt

Med

Med

Med

Med

Prompt

Prompt

Prompt

Med

Priority

High, unadjustable

High, unadjustable

High, unadjustable

High, unadjustable

High, unadjustable

High, unadjustable

Med

Off

Off

Off

Off

Off

Off

Off

Off

Off

Off

Off

Off

Off

Off

Off

Off

Off

Off

Alarm Outputs

Off

Off

Off

Off

Off

Off

Off

C - 2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

C.1.2.2

Arrhythmia Threshold Default Settings

Item Default Setting

Asystole Delay

Tachy

Brady

Extreme Tachy

Extreme Brady

Multif PVCs Window

PVCs/min

Pauses/min

Pause Threshold

AF/Irr Rhy End Time

V-Tach Rate

V-Brady Rate

V-Tach PVCs

V-Brady PVCs

Adult

5 sec

120 bpm

50 bpm

160 bpm

35 bpm

15 beats

10

8

6

5

2.0 sec

2 min

130 bpm

40 bpm

C.1.3

ST Default Settings

Item

ST Alarm Mode

ST Alarm Mode set to Absolute :

ST-I, ST-II, ST-III, ST-aVR, ST-aVL, ST-aVF, ST-V1, ST-

V2, ST-V3, ST-V4, ST-V5,ST-V6, ST-Va, ST-Vb

ST Alarm Mode set to

ST Single,

ST Dual

Relative :

Alarm switch

High limit

Low limit

Priority

Alarm Outputs

Alarm switch

High limit

Low limit

Priority

Alarm Outputs

ST Analysis

ST Segment

Show Marker

ST Point

Auto Adjust

J

ISO

Pediatric

5 sec

160 bpm

75 bpm

180 bpm

50 bpm

15 beats

10

8

6

5

2.0 sec

2 min

130 bpm

40 bpm

Default Setting

Auto

Off

J+60

On

-0.1 mV

Med

Off

Off

48

-80

Absolute

Off

0.2 mV

-0.2 mV

Med

Off

Off

0.1 mV

Neonate

5 sec

200 bpm

100 bpm

220 bpm

60 bpm

15 beats

10

8

6

5

2.0 sec

2 min

160 bpm

40 bpm

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual C - 3

C.1.4

C.1.5

QT Default Settings

Item

QTc

ΔQTc

QT Analysis

QT Lead

Alarm switch

High limit

Priority

Alarm Outputs

Alarm switch

High limit

Priority

Alarm Outputs

Default Setting

Med

Off

Off

All

Off

Adult: 500

Pediatric: 480

Neonate: 460

Med

Off

Off

60

Glasgow 12-lead ECG Algorithm Default Settings

Item

High Freq Cut-off

Baseline Drift Removal

Tachy

Brady

Waveform Layout

Median Complex

Measurements

Interpretation

Interpretation Summary

Default Setting

35 Hz

On

On

On

On

100

50

Standard

Off

C.2

Respiration Default Settings

Item

RR

Apnea

Alarm switch

High limit

Low limit

Priority

Alarm Outputs

Alarm switch

Priority

Alarm Outputs

Default Setting

On

Adult: 30

Pediatric: 30

Neonate: 100

Adult: 8

Pediatric: 8

Neonate: 30

Med

Off

On

High, unadjustable

Off

C - 4 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

C.3

Item

Apnea Delay

RR Source

Resp Lead

Gain

Speed

Auto Threshold Detection

Default Setting

Adult: 20 sec

Pediatric: 20 sec

Neonate: 15 sec

Auto

Adult: Auto

Pediatric: Auto

Neonate: II

×2

6.25 mm/s

On

SpO

2

/SpO

2

b Default Settings

Item

SpO

2

/SpO

2

High limit

SpO

2

/SpO

Desat

2 b

Low limit

Priority

Alarm Outputs

Alarm switch

Low limit

Priority

Alarm Outputs

ΔSpO

2

High limit

Priority

Alarm Outputs

Satsecond (for Nellcor SpO

2

)

NIBP Simul

FastSAT (for Masimo SpO

2

)

Display SIQ (for Masimo SpO

2

)

Display PI (for Masimo SpO

2

)

Speed

Default Setting

10%

Med

Off

Off

80%

High

Off

On

On

Adult: 100%

Pediatric: 100%

Neonate: 95%

90%

Med

Off

On

Off

Off

Off

On

25 mm/s

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual C - 5

C.4

C.5

Item

PR Alarm switch

High limit

Low limit

Priority

Alarm Outputs

Alarm Source

PR Source

Display PR

QRS Volume

Default Setting

On

Adult: 120

Pediatric: 160

Neonate: 200

Adult: 50

Pediatric: 75

Neonate: 100

Med

Off

Auto

Auto

On

General, OR: 2

Other departments: 0

Temperature Default Settings

Item

Txx

(xx refers to temperature site)

Alarm switch

High limit

Low limit

Priority

Alarm Outputs

High limit

Priority

Alarm Outputs

NIBP Default Settings

Item

Default Setting

On

38.0 °C

35.0 °C

Med

Off

On

2.0 °C

Med

Off

High limit

Low limit

Priority

Alarm Outputs

Default Setting

On

Adult: 160 mmHg

Pediatric: 120 mmHg

Neonate: 90 mmHg

Adult: 90 mmHg

Pediatric: 70 mmHg

Neonate: 40 mmHg

Med

Off

C - 6 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Item

NIBP-M

High limit

Low limit

Priority

Alarm Outputs

Alarm switch

High limit

Low limit

NIBP-S Extreme

Priority

Alarm Outputs

Alarm switch

High limit

Low limit

NIBP-D Extreme

Priority

Alarm Outputs

Alarm switch

High limit

Low limit

NIBP-M Extreme

Priority

Alarm Outputs

Alarm switch

High limit

Low limit

Priority

Alarm Outputs

Default Setting

Adult: 75 mmHg

Pediatric: 60 mmHg

Neonate: 35 mmHg

High

Off

Off

Adult: 105 mmHg

Pediatric: 80 mmHg

Neonate: 65 mmHg

Adult: 35 mmHg

Pediatric: 30 mmHg

Neonate: 15 mmHg

High

Off

Off

Adult: 125 mmHg

Pediatric: 100 mmHg

Neonate: 75 mmHg

Adult: 45 mmHg

Pediatric: 40 mmHg

Neonate: 20 mmHg

High

Off

On

Adult: 90 mmHg

Pediatric: 70 mmHg

Neonate: 60 mmHg

Adult: 50 mmHg

Pediatric: 40 mmHg

Neonate: 20 mmHg

Med

Off

On

Adult: 110 mmHg

Pediatric: 90 mmHg

Neonate: 70 mmHg

Adult: 60 mmHg

Pediatric: 50 mmHg

Neonate: 25 mmHg

Med

Off

Off

Adult: 175 mmHg

Pediatric: 130 mmHg

Neonate: 95 mmHg

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual C - 7

Item

Initial Pressure

Interval

C.6

Start Mode

NIBP End Tone

Venipuncture Pressure

Display Format

Display Alarm Limits

Display PR

IBP Default Settings

Item

IBP-S Alarm switch

High limit

Default Setting

Adult: 160 mmHg

Pediatric: 140 mmHg

Neonate: 90 mmHg

OR: 5 min

Neonatology: 30 min

Other departments: 15 min

Clock

Off

Auto

Sys/Dia (Mean)

Off

Off

Low limit

Priority

Alarm Outputs

Default Setting

On

■ Art/Ao/UAP/BAP/FAP/LV/P1-P2 arterial pressure

Adult: 160 mmHg

Pediatric: 120 mmHg

Neonate: 90 mmHg

■ PA

Adult: 35 mmHg

Pediatric and neonate: 60 mmHg

■ Art/Ao/UAP/BAP/FAP/LV/P1-P2 arterial pressure

Adult: 90 mmHg

Pediatric: 70 mmHg

Neonate: 55 mmHg

■ PA

Adult: 10 mmHg

Pediatric and neonate: 24 mmHg

Med

Off

C - 8 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Item

IBP-D

IBP-M

Alarm switch

High limit

Low limit

Priority

Alarm Outputs

Alarm switch

High limit

Low limit

Priority

Alarm Outputs

Default Setting

On

■ Art/Ao/UAP/BAP/FAP/LV/P1-P2 arterial pressure

Adult: 90 mmHg

Pediatric: 70 mmHg

Neonate: 60 mmHg

■ PA

Adult: 16 mmHg

Pediatric and neonate: 4 mmHg

■ Art/Ao/UAP/BAP/FAP/LV/P1-P2 arterial pressure

Adult: 50 mmHg

Pediatric: 40 mmHg

Neonate: 20 mmHg

■ PA

Adult: 0 mmHg

Pediatric and neonate: -4 mmHg

Med

Off

On

■ Art/pArt/Ao/UAP/BAP/FAP/LV/P1-P2 arterial pressure

Adult: 110 mmHg

Pediatric: 90 mmHg

Neonate: 70 mmHg

■ PA

Adult: 20 mmHg

Pediatric and neonate: 26 mmHg

■ CVP

Adult: 14 cmH

2

O

Pediatric and neonate: 5 cmH

2

O

■ ICP/RAP/LAP/UVP/P3-P4 venous pressure

Adult: 10 mmHg

Pediatric and neonate: 4 mmHg

■ Art/Ao/UAP/BAP/FAP/LV/P1-P2 arterial pressure

Adult: 70 mmHg

Pediatric: 50 mmHg

Neonate: 35 mmHg

■ PA

Adult: 0 mmHg

Pediatric and neonate: 12 mmHg

■ CVP

Adult: 0 cmH

2

O

Pediatric and neonate: 0 cmH

2

O

■ ICP/RAP/LAP/UVP/P3-P4 venous pressure

Adult: 0 mmHg

Pediatric and neonate: 0 mmHg

Med

Off

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual C - 9

C - 10

Item

Art-S Extreme Alarm switch

High limit

Low limit

Art-D Extreme

Priority

Alarm Outputs

Alarm switch

High limit

Low limit

Art-M Extreme

Priority

Alarm Outputs

Alarm switch

High limit

CPP

Low limit

Priority

Alarm Outputs

Alarm switch

High limit

Low limit

Measure (for P1, P2)

Measure (for P3, P4)

Sensitivity

Speed

Priority

Alarm Outputs

Default Setting

Adult: 55 mmHg

Pediatric: 40 mmHg

Neonate: 30 mmHg

High

Off

On

Adult: 130 mmHg

Pediatric: 100 mmHg

Neonate: 90 mmHg

Adult: 50 mmHg

Pediatric: 40 mmHg

Neonate: 30 mmHg

Med

Off

All

Mean Only

Med

25 mm/sec

Off

Adult: 175 mmHg

Pediatric: 130 mmHg

Neonate: 95 mmHg

Adult: 75 mmHg

Pediatric: 60 mmHg

Neonate: 50 mmHg

High

Off

Off

Adult: 105 mmHg

Pediatric: 80 mmHg

Neonate: 65 mmHg

Adult: 35mmHg

Pediatric: 30 mmHg

Neonate: 15 mmHg

High

Off

Off

Adult: 125 mmHg

Pediatric: 100 mmHg

Neonate: 75 mmHg

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

C.7

Item

Scale

Default Setting

0- 20 mmHg ICP/RAP/LAP/UVP venous pressure

Art/Ao/BAP/FAP/LV/P1/

P2 arterial pressure

UAP/P3/P4 venous pressure

CVP/PA

0-160 mmHg

0-80 mmHg

CVP: 0-30 cmH2O

PA: 0 - 30 mmHg

Sys/Dia (Mean)

Off

Off

Display Format

Display Alarm Limits

Use PA-D as PAWP (only available for independent external display)

PPV Measure

PPV Source

PAWP Reference Waveform 1

Reference Waveform 2

Speed

PA Scale

Overlapping

Waveform Setup

Left Scale

Right Scale

CVP Scale

ICP Scale

PA Scale

Speed

Gridlines

Off

Auto

II

Resp

12.5 mm/sec

0-30 mmHg

0-160 mmHg

0-20 mmHg

0-30 cmH2O

0-20 mmHg

0-30 mmHg

25 mm/sec

Off

C.O. Default Settings

Item

TB Alarm switch

High limit

Low limit

Priority

Alarm Outputs

Comp const

Auto Start

Auto TI

Default Setting

Off

0.542

On

On

On

39.0 °C

36.0 °C

Med

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual C - 11

C.8

ScvO

2

Default Settings

Item

ScvO

2

Alarm switch

High limit

Low limit

Priority

Alarm Outputs

Hb/Hct

Default Setting

On

90

40

Med

Off

Hb

C.9

CCO Default Settings(Vigilance II/Vigileo/EV1000/HemoSphere)

Item

CCO

CCI

SVR Unit

Select Parameter

Alarm switch

High limit

Low limit

Priority

Alarm Outputs

Alarm switch

High limit

Low limit

Priority

Alarm Outputs

Default Setting

On

14.0

2.0

Med

Off

On

10.0

1.0

Med

Off

DS/cm

5

CCI, GEF, SVRI, ELWI, GEDI

C.10

ScvO

2

/SvO

2

Default Settings(Vigilance II/Vigileo/EV1000/

HemoSphere)

Item

SvO

2

/ScvO

2 switch

High limit

Low limit

Priority

Alarm Outputs

Default Setting

On

90

40

Med

Off

C - 12 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

C.11

CO

2

Default Settings

C.11.1

General Settings

Item

EtCO

2

High limit

Low limit

Priority

Alarm Outputs

FiCO

2

Apnea Delay

RR Source

Speed

Scale

Waveform Type

High limit

Priority

Alarm Outputs

C.11.2

Sidestream CO

2

Default Settings

Item

BTPS Compensation

O

2

Compensation (for CO

2

module without

O

2

measurement)

N

2

O Compenation

AG Compenation

Auto Standby

Operating Mode

Default Setting

Off

OR: 100%

Other departments: 21%

0%

0%

60 min

Measure

C.11.3

Microstream CO

2

Default Settings

Item

BTPS Compen

Maximum Hold

Auto Standby

Operating Mode

Default Setting

Off

20 sec

Off

Measure

Default Setting

On

Adult and pediatric: 50 mmHg

Neonate: 45 mmHg

Adult and pediatric: 25mmHg

Neonate: 30mmHg

Med

Off

On

4 mmHg

Med

Off

Adult and pediatric: 20 s

Neonate: 15 s

Auto

6.25 mm/s

50 mmHg

Draw

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual C - 13

C.12

Gas Default Settings

Item

EtCO

2

FiCO

EtO

FiO

2

2

2

High limit

Low limit

Priority

Alarm Outputs

High limit

Priority

Alarm Outputs

High limit

Low limit

Priority

Alarm Outputs

Alarm switch

High limit

Low limit

Priority

Alarm Outputs

EtN

2

High limit

Low limit

Priority

Alarm Outputs

FiN

2

O Alarm

High limit

EtAA/FiAA

Low limit

Priority

Alarm Outputs

Alarm switch

High limit

Low limit

Priority

Alarm Outputs

Default Setting

Off

On

30%

0.0%

Med

Off

On

88%

18%

Med

On

4 mmHg

Med

Off

On

Adult and pediatric: 50 mmHg

Neonate: 45 mmHg

Adult and pediatric: 25 mmHg

Neonate: 30 mmHg

Med

Off

On

53%

0%

Med

55%

0%

Med

Off

18%

Med

Off

On

Off

On

Adult and pediatric: 100%

Neonate: 90%

C - 14 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Item

EtHal/EtEnf/EtIso

FiHal/FiEnf/FiIso

Alarm switch

High limit

Low limit

Priority

Alarm Outputs

Alarm switch

High limit

Low limit

Priority

Alarm Outputs

FiSev

Apnea Delay

RR Source

Operating Mode

Auto Standby

RR Source

Speed

High limit

Low limit

Priority

Alarm Outputs

High limit

Low limit

Priority

Alarm Outputs

Alarm switch

High limit

Low limit

Priority

Alarm Outputs

High limit

Low limit

Priority

Alarm Outputs

Default Setting

0.0%

Med

Off

On

Med

Off

On

6.0%

Off

On

2.0%

0.0%

On

3.0%

0.0%

Med

On

8.0%

0.0%

Med

5.0%

0.0%

Med

Off

Off

On

6.0%

0.0%

Med

Off

Adult and pediatric: 20 sec

Neonate: 15 sec

Auto

Measure

Off

Auto

6.25 mm/sec

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual C - 15

Item

Scale

Waveform Type

O2 compensation

Default Setting

O2: 400 mmHg

CO2: 50 mmHg

N2O: 50 %

Hal, Enf, and Iso: 2.5%

Sev: 4.0%

AA and Des: 9.0 %

Draw (for CO

2 only )

OR: 100%

Other departments: Off

C.13

EEG Default Settings

Item

Scale

Speed

Low Filter

High Filter

Notch Filter

EEG Expand View

EEG

Trends

DSA

EEG Channels

Scale

Speed

EEG Channels

Parameters

Trend Length

EEG Channels

Parameters

CSA

Trend Length

Power Scale

EEG Channels

Parameters

Trend Length

Power Scale

CSA Clipping

Default Setting

100 μV

30 mm/s

0.5 Hz

30 Hz

On

Default Setting

All

100 μV

30 mm/s

EEG1, EEG2, EEG3, EEG4

SEF

60 min

All

SEF

20 min

1 - 64 dB

All

SEF

20 min

1 - 64 dB

On

C - 16 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

C.14

BIS Default Settings

Item

BIS

Smoothing Rate

Display

Scale

Speed

Filter

Auto Check

BIS Expand View

Parameter 1

Parameter 2

Trend Length

Alarm switch

High limit

Low limit

Priority

Alarm Outputs

C.15

NMT Default Settings

Item

Stimulation Current

Pulse Width

Block Recovery

Stimulation Beep Volume

Interval

DBS Mode

C.16

rSO

2

Default Settings

Item rSO2-1/rSO2-2/ rSO2-1’/rSO2-2’

Alarm switch

High limit

Low limit

Priority

Alarm Outputs

Default Setting

On

70

20

Med

Off

30 sec

OR: BIS Trend

Other departments: EEG LT/RT

100 μV

25mm/s

On

On

Default Setting

BIS L

EMG

60 min

Default Setting

Supra (60 mA)

200 μs

Off

2

TOF Mode: 1 min

ST Mode: 10 sec

DBS Mode: 1 min

3.3

Default Setting

On

90

40

Med

Off

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual C - 17

Item rSO2-1/rSO2-2/ rSO2-1’/rSO2-2’

Variance

Auto Low Limit rSO2-1 Label rSO2-2 Label rSO2-1’ Label rSO2-2’ Label

AUC Mode

Low Limit

Fixed Threshold

Percentage Below Baseline

Select Parameter

Default Setting

-20

Off

L

R

S1

S2

Below Baseline Percentage

50

25

Baseline, Baseline Variance

C.17

Alarm Default Settings

Item

Alarm Volume

High Alarm Volume

Reminder Volume

Apnea Delay

Printing Duration on Alarm

Auto Limits for New Patient

Default Setting

2

Alarm Volume + 2

2

Adult/Pediatric: 20 sec

Neonate: 15 sec

20 sec

On

C.18

Display Default Settings

Item

Choose Screen

Display

Night Mode

Screen Lock Duration

Brightness

Brightness On Battery

Brightness

Alarm Volume

QRS Volume

Key Volume

NIBP End Tone

Stop NIBP

Default Setting

1

0

Off

Off

1

2

5

1

Normal Screen

General: Permanent

CCU: Permanent

Other departments: 10 sec

C - 18 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

D

Alarm Messages

D.1

D.1.1

This chapter lists only the physiological and technical alarm messages. Some messages appearing on your monitor may not be included. In the “Cause and solution” column, corresponding solutions are given instructing you to troubleshoot problems. If the problem persists, contact your service personnel.

Physiological Alarm Messages

General Physiological Alarm Messages

D.1.2

Alarm messages

XX High

XX Low

Default priority

Med

Med

Cause and solution

XX value has risen above the high alarm limit or fallen below the low alarm limit. Check the patient’s condition and check if the patient category and alarm limit settings are correct.

Note: XX represents a measurement or parameter label, such as HR, NIBP, PVCs, RR, SpO

2

, PR, etc.

Arrhythmia Alarm Messages

Alarm message

Asystole

V-Fib/V-Tach

V-Tach

Vent Brady

Extreme Tachy

Extreme Brady

R on T

Run PVCs

Couplet

Multiform PVC

PVC

Bigeminy

Trigeminy

Tachy

Brady

Pacer Not Pacing

Pacer Not Capture

Missed Beats

Nonsus. Vtac

Vent Rhythm

Pause

Default priority

Prompt

Med

Prompt

Med

Med

Med

Med

Prompt

High

High

Med

Low

High

High

High

High

Prompt

Prompt

Med

Med

Low

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual D - 1

D.1.3

Alarm message

Irr Rhythm

PVCs/min

Pauses/min

A-Fib

Default priority

Prompt

Med

Med

Prompt

Note: When arrhythmia alarms occur, check the patient’s condition and the ECG connections.

Resp Physiological Alarm Messages

Alarm message

Resp Artifact

Apnea

Default priority

High

High

Cause and solution

The patient’s heartbeat has interfered with the impedance respiration measurement. Check the patient’s condition and the ECG connections.

The respiration signal is so weak that the monitor cannot perform respiration analysis. Check the patient’s condition, module and patient connections.

D.1.4

SpO

2

Physiological Alarm Messages

Alarm message

SpO2/SpO2b Desat

ΔSpO2 High

Default priority

High

Med

Cause and solution

The SpO

2

or SpO

2 b value falls below the desaturation alarm limit.

Check the patient’s condition and check if the alarm limit settings are correct.

The ΔSpO2 value exceeds the alarm limit. Check the patient’s condition.

D.1.5

PR Physiological Alarm Messages

Alarm message

No Pulse

Default priority

High

Cause and solution

The pulse signal is so weak that the monitor cannot perform pulse analysis. Check the patient’s condition and the source of the PR, such as

IBP or SpO2 sensor, and measurement site.

D.1.6 NIBP Physiological Alarm Messages

Alarm message

NIBP-S Extremely

High/NIBP-D

Extremely High/NIBP-

M Extremely High

NIBP-S Extremely

High/NIBP-D

Extremely High/NIBP-

M Extremely Low

Default priority

High

High

Cause and solution

The NIBP value is higher than the NIBP Extreme alarm high limit. Check the patient’s condition and check if the alarm limit settings are correct.

The NIBP value is lower than the NIBP Extreme alarm low limit. Check the patient’s condition and check if the alarm limit settings are correct.

D - 2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

D.1.7 IBP Physiological Alarm Messages

Alarm message

Art-S/Art-D/Art-M

Extremely High

Art-S/Art-D/Art-M

Extremely Low

Default priority Cause and solution

High

High

The Art value is higher than the Art Extreme alarm high limit. Check the patient’s condition and check if the alarm limit settings are correct.

The Art value is lower than the Art Extreme alarm low limit. Check the patient’s condition and check if the alarm limit settings are correct.

D.1.8

CO

2

Physiological Alarm Messages

D.1.9

Alarm message

FiO2 Shortage

Default priority

High

Cause and solution

FiO

2

concentration is less than 18%. Check the patient’s condition, the ventilated O

2

content and the airway connection.

AG Physiological Alarm Messages

Alarm message

FiO2 Shortage

Mixed Agent and MAC≥3

Apnea

Default priority

High

Med

High

Cause and solution

FiO

2

concentration is less than 18%. Check the patient’s condition, the ventilated O

2

content and the AG connections.

The mixed anaesthetic gases concentration is too high. Verify the MAC parameters are correct and adjust the anaesthetic gases concentration as necessary.

The respiration signal was so weak that the monitor cannot perform respiration analysis. Check the patient’s condition, module and patient connections.

D.2

Technical Alarm Messages

This section lists technical alarms, their default priority, responses on alarm reset, and the actions that can be taken when an alarm occurs. Some of alarms may not be included.

Various technical alarms respond differently when the alarm system is reset. For easy clarification, in this section the technical alarms are classified into three categories of response when the alarm system is reset:

A: technical alarms are cleared. The monitor gives no alarm indications.

B: technical alarms are changed to the prompt messages.

C: the alarm is silenced and a √ appears before the alarm message.

In the following tables A, B, and C are used to refer to the responses on alarm reset.

D.2.1

General Technical Alarm Messages

Alarm message Default priority

Indication on alarm reset

C

Cause and solution

XX Module Error High XX module does not work properly. Reinsert the module, if the alarm persists, contact your service personnel.

Note: XX represents a measurement or parameter label, such as HR, RR, SpO2, or EtCO2, etc.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual D - 3

D.2.2

ECG Technical Alarm Messages

D.2.3

Alarm message Default priority

Indication on alarm reset

A

Cause and solution

ECG Noisy

ECG Amplitude Too Small

ECG (XX) Lead Off

ECG Signal Invalid

ECG Learning

Low/Prompt

Low

Low

Low

Prompt

C

B

A

/

The ECG signal is noisy. Check for any possible sources of signal noise around the cable and electrode, and check the patient for excessive motion.

The ECG amplitude does not reach the detected threshold. Check for any possible source of interference around the cable and electrode.

The electrode has become detached from the patient or the lead wire has become disconnected from the adapter cable. Check the connections of the electrodes and leadwires.

Patient skin impedance is too high. Check ECG electrode application.

ECG learning is manually or automatically triggered.

/ Cannot Analyze QT Prompt /

Note: XX represents ECG lead name, for example RL, LL, V, Va, Vb, etc.

Resp Technical Alarm Messages

Alarm message

Resp Interference

Electrode Poor Contact

Default priority

Prompt

Indication on alarm reset

/

Prompt /

Cause and solution

The respiration circuit is disturbed. Check for any possible sources of signal noise.

Check the electrode application. Reposition or replace the electrodes if necessary.

D.2.4

SpO

2

Technical Alarm Messages

Alarm message

SpO2 Sensor Off/SpO2b

Sensor Off

Default priority

Low

Indication on alarm reset

B

SpO2 No Sensor/SpO2b

No Sensor

SpO2 Excess Light/SpO2b

Excess Light

SpO2 No Pulse/SpO2b No

Pulse

Low

Low

Low

A

C

C

Cause and solution

The SpO

2

sensor has become detached from the patient or the module. Check the sensor connection. If the alarm persists, replace the sensor.

The SpO

2

extension cable is detached from the

SpO

2

module, or the SpO

2

sensor is detached from the SpO

2

extension cable. Check the SpO cable and the sensor connection. If the alarm

2 persists, replace the sensor.

Ambient light is too strong. Move the sensor to a place with lower level of ambient light or cover the sensor to minimize the ambient light.

The SpO

2

sensor failed to obtain pulse signal.

Check the patient’s condition and replace the sensor application site. If the alarm persists, replace the sensor.

D - 4 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Alarm message Default priority

Indication on alarm reset

C SpO2 Sensor

Incompatible/SpO2b

Sensor Incompatible

SpO2 Low Signal Quality/

SpO2b Low Signal Quality

Low

Low C

SpO2 Interference

/SpO2b Interference

Low

SpO2 Sensor Error/SpO2b

Sensor Error

SpO2 Searching Pulse/

SpO2b Searching Pulse

SpO2 Low Perfusion/

SpO2b Low Perfusion

Low

Prompt

Prompt

C

C

/

/

Cause and solution

Incompatible or an unspecified SpO

2

sensor is used. Use specified sensors.

1. Check the sensor and sensor position.

2. Make sure the patient is not shivering or moving.

3. The patient’s pulse may be too low to be measured.

The SpO

2

signal has interference. Check for any possible sources of signal noise and check the patient for excessive motion.

Replace the sensor and measure again.

SpO

2

is searching for pulse.

The SpO

2

sensor is not properly placed or the patient’s perfusion index is too low.

1. Check the sensor and sensor position.

2. Reposition the sensor if necessary.

D.2.5

Temp Technical Alarm Messages

Alarm message

T1/T2 Sensor Off

Default priority

Low

Indication on alarm reset

A

Cause and solution

Check the sensor connection and reconnect the sensor.

D.2.6

NIBP Technical Alarm Messages

Alarm message

NIBP Loose Cuff

NIBP Air Leak

NIBP Airway Error

NIBP Weak Signal

NIBP Overrange

NIBP Excessive Motion

NIBP Cuff Overpress

Default priority

Low

Indication on alarm reset

A

Low

Low

Low

Low

Low

Low

A

A

A

A

A

A

Cause and solution

There is a leak in the cuff or air tubing. Use a cuff of correct type based on the patient size. Apply the cuff and connect the air tubing as instructed in the manual.

Check the NIBP cuff and pump for leakages.

The air tubing may be occluded. Check the air tubing for an occlusion or kinking. If the alarm persists, contact your service personnel.

The patient’s pulse is weak or the cuff is loose.

Check the patient’s condition and replace the cuff application site.

The measured NIBP value exceeds the module measurement range. Check the patient’s condition.

Check the patient’s condition and reduce patient motion.

The NIBP airway may be occluded. Check the airway and measure again. If the alarm persists, contact your service personnel.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual D - 5

Alarm message

NIBP Timeout

NIBP Cuff and Patient

Mismatch

NIBP Airway Leak

Default priority

Low

Indication on alarm reset

A

Low

Low

A

A

Cause and solution

The measurement time exceeds 120 seconds in the adult or pediatric mode, or exceeds 90 seconds in the neonatal mode, and the BP value cannot be obtained. Check the patient’s condition and NIBP connections, or replace the cuff and measure again.

The cuff type mismatches the patient category.

Verify the patient category or replace the cuff if necessary. If patient category is correct, check that the tubing is not bent and the airway is not occluded.

Airway leakage is found during the NIBP leakage test. Check the NIBP cuff and pump for leakages.

D.2.7

IBP Technical Alarm Messages

D.2.8

Alarm message

XX Sensor Error

XX No Sensor

XX No Pulse

XX Disconnected

Monitor Version Low

Default priority

Med

Med

Low

High

Low

C

A

Indication on alarm reset

A

C

A

Cause and solution

The IBP sensor fails. Replace the sensor.

The IBP patient cable and/or corresponding IBP sensor is not connected or detached. Check the cable and sensor connection.

The catheter may be occluded. Please flush the catheter.

The tubing is disconnected from the patient, or the three-way valve is open to the air. Check the connection of the tubing, or check the valve is open to the patient. If the alarm persists, contact your service personnel.

The N Series system software version is too low.

In order for the N Series monitor to connect to the N1 which connects to the external IBP module and perform IBP monitoring, the system software of N 1 and N series monitor should be

V01.11.00 and above.

Note: XX represents an IBP label, for example PA, CVP, FAP, P1, etc.

C.O. Technical Alarm Messages

Alarm message

TB Sensor Off

TI Sensor Off

Default priority

Low

Indication on alarm reset

A

Low A

Cause and solution

Check the sensor connection and reconnect the sensor.

Check the sensor connection and reconnect the sensor.

D - 6 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

D.2.9

ScvO

2

Technical Alarm Messages

Alarm message

ScvO2 High Signal

ScvO2 Low Signal

ScvO2 Excess Light

Optical Module

Disconnected

Unsupported CeVOX

Version

Disconnected from

Vigilance

Disconnected from

Vigileo

Disconnected from

EV1000

Disconnected from

HemoSphere

Default priority

Low

Indication on alarm reset

C

Low

Low

Low

Low

Low

C

C

A

A

A

Low

Low

Low

A

A

A

Cause and solution

Check the sensor and reposition the catheter.

Recalibrate the sensor.

Check the sensor and reposition the catheter.

Recalibrate the sensor.

Avoid excessively strong backlight.

Connect the optical module.

The module version is not compatible with the system. Contact your service personnel.

The Vigilance monitor is disconnected. Check the connection between the CCO/SvO

2

module and the Vigilance monitor.

The Vigileo monitor is disconnected. Check the connection between the CCO/SvO

2

module and the Vigileo monitor.

The EV1000 monitor is disconnected. Check the connection between the CCO/SvO

2

module and the EV1000 monitor.

The HemoSphere monitor is disconnected. Check the connection between the CCO/SvO

2

module and the HemoSphere monitor.

D.2.10

CO

2

Technical Alarm Messages

Alarm message

CO2 Module High Temp

Default priority

Low

Indication on alarm reset

C

CO2 Module Low Temp

CO2 Zero Failed

CO2 No Watertrap

CO2 High Airway Press.

Low

Low

Low

Low

C

C

B

C

Cause and solution

Ambient temperature is too high or there is a module failure.

1. Lower the operating temperature.

2. Reinsert the module.

3. If the alarm persists, the CO

2

module may have failed, contact your service personnel.

Ambient temperature is too low or there is a module failure.

1. Raise the operating temperature.

2. Reinsert the module.

3. If the alarm persists, the CO

2

module may have failed, contact your service personnel.

Reinsert the module. If the alarm persists, contact your service personnel.

Check the watertrap connections.

1. Check the airway pressure settings of the ventilator/anesthesia machine.

2. Disconnect the module from the ventilator/ anesthesia machine.

3. Reinsert the module.

4. If the alarm persists, contact your service personnel.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual D - 7

Alarm message

CO2 Low Airway Press.

High Barometric

Low Barometric

CO2 Airway Occluded

CO2 No Filterline

CO2 Calibration Required

CO2 Airway Error

CO2 Adapter Error

CO2 No Sensor

CO2: Change Watertrap

CO2 Watertrap Mismatch

Default priority

Low

Indication on alarm reset

C

Low

Low

Low

Low

Low

Low

Low

Low

Low

Low

C

C

C

A

C

C

A

A

C

C

Cause and solution

1. Check the airway pressure settings of the ventilator/anesthesia machine.

2. Disconnect the module from the ventilator/ anesthesia machine.

3. Reinsert the module.

4. If the alarm persists, contact your service personnel.

The ambient pressure exceeds the operating pressure range or CO

2

module fails.

1. Make sure that the ambient pressure meets the specifications, and check for sources that affect the ambient pressure.

2. Reinsert the module. If the alarm persists, contact your service personnel.

The ambient pressure exceeds the operating pressure range or CO

2

module fails.

1. Make sure that the ambient pressure meets the specifications, and check for sources that affect the ambient pressure.

2. Reinsert the module. If the alarm persists, contact your service personnel.

1. Check if the sample line is kinked or occluded.

2. Replace the sample line.

3. Reinsert the module.

4. If the alarm persists, contact your service personnel.

Make sure that the filterline is connected.

Perform a calibration.

1. Check if the sample line is kinked or occluded.

2. Replace the sample line.

3. Reinsert the module.

4. If the alarm persists, contact your service personnel.

Check, clean or replace the airway adapter.

Perform a zero calibration.

Make sure that the CO

2 transducer is connected.

Replace the watertrap.

Check the patient category and use a correct watertrap.

D.2.11

AG Technical Alarm Messages

Alarm message

AG No Watertrap

Default priority

Low

Indication on alarm reset

B

AG: Change Watertrap

AG Watertrap and Patient

Mismatch

Low

Low

C

C

Cause and solution

Check the connections of the watertrap and reconnect it.

Replace the watertrap.

Check the patient category and use a correct watertrap.

D - 8 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Alarm message

AG Zero Failed

Anesthetic Mixture

AG Airway Occluded

Default priority

Low

Indication on alarm reset

C

Low

Low

C

C

Cause and solution

There is external electromagnetic interference, airway occlusion or module failure.

1. Check for external inference sources.

2. Check for “AG Airway Occluded” alarm message. Remove the occlusion.

3. If the alarm persists, contact your service personnel.

Anesthetic mixture is detected.

1. Check if the sample line is occluded.

2. Replace the sample line.

3. Reinsert the module.

4. If the alarm persists, contact your service personnel.

D.2.12

EEG Technical Alarm Messages

Alarm message Default priority

Indication on alarm reset

A

Cause and solution

EEG No Sensor

EEG Sensor Off

EEG Electrode X + Y Off

EEG Electrode X + Y Poor

Contact

EEG: Module Protected

Low

Low

Low

Low

High

A

A

A

C

The EEG patient cable is not connected. Connect the cable.

EEG electrodes are not connected or detached.

Check electrode connection and reconnect the electrodes.

X + Y electrode is detached. Check the electrode connection and reconnect the electrode.

X + Y electrode does not properly contact the patient. Check the electrode connection and reconnect the electrode.

EEG module may have failed. Disconnect the EEG module and reconnect it. If the alarm persists, replace the module, or contact your service personnel.

Note: X represents EEG channel and polarity, for example A+, A-, A+, A-, etc. Y represents electrode location, for example Fp1, T3, etc.

D.2.13

BIS Technical Alarm Messages

Alarm message

BIS Sensor Off

BIS Electrode XX Off

BIS Electrode XX Poor

Contact

BISx Error

Default priority

Low

Indication on alarm reset

A

Low

Low

High

A

A

C

Cause and solution

Check and reconnect the BIS sensor. If the alarm persists, replace the sensor.

Check the electrode connection, and re-attach the electrodes if necessary.

Enter the Sensor Check menu, and check the connections of the sensor and electrodes.

Reinsert the module. If the alarm persists, contact your service personnel.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual D - 9

Alarm message Default priority

Indication on alarm reset

A

Cause and solution

BIS No Sensor Low

BIS Sensor Too Many Uses Low

BIS Sensor Old Low

BIS XX Signal Quality Too

Low

Low

BIS XX Low Signal Quality

BIS Wrong Sensor Type

BIS Sensor Error

BISx Disconnected

Low

Low

Low

Low

A

C

A

A

A

A

A

BIS sensor is not properly connected, BIS cable failure, BISx or BISx4 failure.

1. Check the BIS sensor connection.

2. Reinsert the BIS module.

3. Replace the BIS cable.

4. Replace BISx or BISx4.

Replace the sensor.

Replace the sensor.

SQI < 15%

1. Check the patient’s condition.

2. Check the sensor position, and its contact with the patient’s skin.

3. Check that BISx or BISx4 is away from electrically radiating equipment.

SQI between 50% and 15%.

1. Check the patient’s condition.

2. Check the sensor position, and its contact with the patient’s skin.

3. Check that BISx or BISx4 is away from electrically radiating equipment.

Check or replace the sensor.

Replace the sensor.

BISx or BISx4 is not connected properly or fails, or

BIS patient cable fails.

1. Check that BISx or BISx4 is connected properly.

2. Replug the BIS Module.

3. Replace the BIS patient cable.

4. Replace BISx or BISx4.

Reinsert the BIS Module.

Reconnect BISx Low A

Note: XX represents BIS label, for example G, C, LE, LT, RL-RA, L-R, F-R, 1, 2, etc.

D.2.14

NMT Technical Alarm Messages

Alarm message

NMT No Cable

NMT No Sensor

NMT Stimulation

Electrode Off

NMT Sensor Error

Default priority

Low

Indication on alarm reset

A

Low A

Low

Low

A

C

Cause and solution

Check that NMT patient cable is properly connected to the NMT module.

Check that NMT sensor is properly connected to the NMT patient cable. If the alarm persists, replace the sensor.

Check that NMT sensor is properly connected to the NMT patient cable. If the alarm persists, check the application of electrodes.

Contact your service personnel.

D - 10 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

D.2.15

rSO

2

Technical Alarm Messages

Alarm message Default priority

Indication on alarm reset

A rSO2-1 No Sensor rSO2-2 No Sensor rSO2-1’ No Sensor rSO2-2’ No Sensor rSO2-1 Excess Light rSO2-2 Excess Light rSO2-1’ Excess Light rSO2-2’ Excess Light rSO2-1 Low Signal Quality rSO2-2 Low Signal Quality rSO2-1’ Low Signal Quality rSO2-2’ Low Signal Quality rSO2 -1/rSO2-2 No

Preamplifier rSO2-1’/rSO2-2’ No

Preamplifier rSO2 -1: Change Sensor rSO2-2: Change Sensor rSO2-1’: Change Sensor rSO2: Change Sensor

Low

Low

Low

Low

Low

C

C

A

C rSO2 -1/rSO2-2

Interference rSO2-1’/rSO2-2’

Interference rSO2-1 Auto Baseline rSO2-2 Auto Baseline rSO2-1’ Auto Baseline rSO2-2’ Auto Baseline

Disconnect/ Reconnect rSO2-1/rSO2-2

Disconnect/ Reconnect rSO2-1’/rSO2-2’

Low

Prompt

Low

C

/

A

Cause and solution rSO

2

sensor is detached, or the sensor cable is disconnected from the pre-amplifier. Re-connect the sensor to the pre-amplifier.

The sensor is detached or ambient light is too strong. Re-attach the sensor or reduce the level of ambient light.

The signal acquired by the sensor was unstable or weak due to power supply noise. Check the sensor connection. If the alarm persists, contact your service personnel.

Properly connect the pre-amplifier.

The sensor type mismatches patient category, the reusable sensor cable fails, the sensor reaches the lifetime, or the sensor type mismatches the monitor and patient category setting. Verify patient category and use sensors of correct patient category.

There is noise interference (such as electrosurgical unit). Check for any possible sources of signal noise.

The monitor automatically sets the rSO

2

baseline.

Remove the rSO

2

module and reconnect it.

D.2.16

Power Supply Technical Alarm Messages

Alarm message

Low Battery

Critically Low Battery

Default priority

Med

Indication on alarm reset

C

High C

B Battery Service Required

(N22/N19)

Battery Overload ( N22/

N19 )

Low

High C

Power Board Comm Error High C

Cause and solution

Connect the monitor to an AC power source and allow the batteries to charge.

Connect the monitor to an AC power source and allow the batteries to charge.

The battery has reached its lifetime. Contact the service personnel.

Too many parameter modules are connected, causing system overload and high power consumption. Use AC power supply.

Shutdown and restart the monitor. If the alarm persists, contact your service personnel.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual D - 11

Alarm message

Battery Err

Default priority

High

Indication on alarm reset

C

C Battery Charging Error

( N22/N19 )

RT Clock Need Reset

RT Clock Not Exist

XX V Too High

XX V Too Low

High

High

High

High

High

Note: XX represents 2.5, 3.3, 5, or 12.

C

C

C

C

Cause and solution

The battery may have failed. Contact your service personnel.

The charging circuit fails or the battery fails.

Contact your service personnel.

Contact your service personnel.

Contact your service personnel.

There is a problem with the system power supply.

Shutdown and restart the monitor.

D.2.17

Recorder Technical Alarm Messages

Alarm message

Recorder Init Error

Recorder Comm Error

Recorder Unavailable

Recorderhead

Overheated

Recorder Initializing

Recorder No Paper

Recorder Busy

Recorder Not Found

Default priority

Low

Indication on alarm reset

A

Low

Low

Low

Prompt

Prompt

Prompt

Prompt

A

A

C

/

/

/

/

Cause and solution

An error occurred during the recorder initialization. Reinsert the module. If the alarm persists, contact your service personnel.

Reinsert the recorder. Shutdown and restart the monitor. If the alarm persists, contact your service personnel.

Recorder module failure. Replace the module.

The recorder has been printing for too long. Stop the recording and resume the recording after the recorder’s print head cools down.

Wait until the recorder initialization is completed.

The recorder paper is not loaded or the recorder door is not closed. Check the recorder, load the recorder paper or close the recorder door.

The buffer queue for recording is full.

The recorder module is not inserted in. Insert the module.

D.2.18

Printer Technical Alarm Messages

Alarm message

Printer Buffer Full

Printing Failed

Printing Stopped

Printer Unavailable

PDF storage space is nearly full

Default priority

Prompt

Indication on alarm reset

/

Prompt

Prompt

Prompt

Prompt

/

/

/

/

Cause and solution

The printer buffer is full. Wait until the printer finishes the printing task.

The printer ran out of paper or cannot be found on the network. Check the printer.

Printing was manually stopped.

The printer may have failed. Check the printer.

Delete the files saved under the PDF file path to release storage space. Otherwise you cannot save new PDF files.

D - 12 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Alarm message

Error storing PDF file

Change the print server language to be consistent with this monitor

Default priority

Prompt

Indication on alarm reset

/

Prompt

Print Server Disconnected Prompt

/

/

Cause and solution

The PDF file path settings on the print server and the PDF Creator are not consistent or the PDF storage space is full. Check the PDF file path settings for consistency, or delete the files saved under the PDF file path to release storage space.

Verify that the language settings of the print server and the monitor are consistent, Otherwise you cannot perform printing.

Check that the monitor is properly connected with the print server.

D.2.19

Technical Alarm Messages Related to Networked Monitoring

Alarm message Default priority

Indication on alarm reset

B

Cause and solution

CMS/eGW Disconnected

View Bed XX YY-ZZ,

Network Disconnected.

Viewed by Bed XX YY-ZZ,

Network Disconnected.

WLAN IP Address Conflict

LAN1 IP Address Conflict

Fail To Get WLAN IP

Address

Fail To Get LAN1 IP

Address

Low

Low

Low

Low

Low

Low

Low

A

A

C

C

C

C

The monitor is disconnected from the CMS.

Check the network connection.

The network is interrupted when the monitor is viewing the remote device. Check the network connection.

The network is interrupted when the monitor is viewed by another remote device. Check the network connection.

There is a wireless network IP address conflicts.

Check the network settings.

There is a wired network IP address conflicts.

Check the network settings.

Unable to automatically obtain the wireless network IP address. Check the network settings.

Unable to automatically obtain the wired network IP address. Check the network settings.

Note: XX refers to the department name, YY refers to the room number, and ZZ refers to the bed number.

D.2.20

Technical Alarm Messages Related to Telemetry Monitors

Alarm message

Telemetry Disconnected

Default priority

High

Indication on alarm reset

B

Telemetry Error High C

Cause and solution

The telemetry is powered off, the monitor or the telemetry is not connected to the wireless network, or out of the network coverage.

1. Power on the telemetry.

2. Connect the monitor and the telemetry to the network.

3. Move the telemetry in the network coverage.

4. The exchanger network connected by the telemetry and the monitor does not support multicast data transfer. Contact your service personnel.

An error occurred to the telemetry.

Restart the telemetry. If the problem still persists, replace with a known good telemetry.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual D - 13

Alarm message

Telemetry Low Battery

Telemetry Battery

Depleted

Telemetry Battery

Maintenance Required

Telemetry Battery Error

Telemetry Battery Type

Error

Default priority

Med

Indication on alarm reset

C

High

Med

Med

Med

C

C

C

C

Cause and solution

The telemetry battery charge is low.

Replace with a known good battery.

The telemetry battery charge is critically low.

Replace with known good batteries.

The telemetry battery may reach its life. Replace with known good batteries.

The telemetry battery communication encounters an error. Replace with known good batteries.

The telemetry battery contacts may fail. Replace with known good batteries.

D.2.21

Other System Technical Alarm Messages

Alarm message

Primary Screen Alarm

Error

Secondary Screen Alarm

Error

High

XX: Disconnected

(XX refers to the name of the external device)

Storage Error

High

High

Loading Default Config

Failed

Low

XX Conflicts

(XX refers to the module label)

XX Measurement has been closed

(XX refers to the module label)

The display setup for XX is disabled.

(XX refers to the parameter label)

Prompt

Prompt

Prompt

Default priority

High

Indication on alarm reset

C

C

A

C

A

/

/

/

Cause and solution

The alarm signal communication between the primary display and the main unit is interrupted.

The primary display cannot provide alarm indications. Check the connection between the main unit and the primary display.

The alarm signal communication between the secondary display and the main unit is interrupted. The secondary display cannot provide alarm indications. Check the connection between the main unit and the secondary display.

Corresponding external device is disconnected.

Check the connection between the monitor and the external device.

The storage card failed or files are damaged.

Shutdown and restart the monitor. Format the storage card (CAUTION: this will clear all patient data stored on this monitor). If the alarm persists, contact your service personnel.

The default configuration was not correctly loaded. The monitor will restore to the factory default configuration for the current patient category.

The same type of corresponding module being used exceeds the supported number. Remove the conflict module.

The parameter module is disabled. Switch on the

module if you want to use it. For more information, see

3.9.1 Switching On or Off a

Parameter .

The parameter of the newly inserted module is not displayed on the screen. Select a desired area to display the parameter numerics and waveforms. For more information, see

3.9.2 Displaying Parameter Numerics and

Waveforms .

D - 14 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Alarm message Default priority

Indication on alarm reset

B The patient data storage space is nearly full. Please delete some discharged patients.

Med

Cause and solution

The storage space of the monitor is full. Delete unnecessary discharged patients.

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual D - 15

This page intentionally left blank.

D - 16 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

E

SpO

2

Sensor Accuracy

E.1

The Accuracy of Masimo SpO

2

Sensors

Table information for the plots below show A

RMS

values measured with Masimo SET Oximetry Technology in a clinical study.

Adtx/Pdtx/4000/4001

Inf/Neo/NeoPt/4002/4003/4004/4005

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual E - 1

DCI/DCIP/4050/4051

LNCS YI

4053

E - 2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

E.2

The Accuracy of Nellcor SpO

2

Sensors

SpO

2

Accuracy for Nellcor Sensors vs. Co-Oximeters (Arms)

SpO

2

Range

DS-100A

D-YS, OXI-P/I, OXI-A/N

MAXAI, MAXPI, MAXII

MAXNI

100% to 70%

1.64%

2.41%

1.62%

1.85%

100% to 90%

1.16%

1.38%

1.49%

1.71%

90% to 80%

1.67%

2.50%

1.57%

1.51%

80% to 70%

2.25%

3.60%

2.50%

1.59%

Modified Bland-Altman for SpO

2

- MAXAI, MAXPI, MAXII, MAXNI Sensors: (SpO

2 vs.SaO

2

- SaO

2

)

Modified Bland-Altman for SpO

2

- DS-100A Sensors: (SpO

2

- SaO

2

) vs.SaO

2

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual E - 3

Modified Bland-Altman for SpO2 - D-YS, OXI-A/N, OXI-P/I Sensors: (SpO

2

- SaO

2

) vs.SaO

2

E - 4 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

F

Units, Symbols and Abbreviations

F.1

Units

Abbreviation mEq mg min ml mm m mAh

Mb mcg

L lb kg kPa in k h

Hz

°F g

GHz

GTT cc cm dB

DS

Ah bpm bps

°C

μs

A

μA

μV

In Full hour hertz inch kilo kilogram kilopascal litre pound meter milliampere hour mega byte microgram milli-equivalents milligram minute milliliter millimeter microampere microvolt microsecond

Ampere

Ampere hour beat per minute bits per second

Celsius cubic centimeter centimeter decibel dyne second

Fahrenheit gram gigahertz gutta

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual F - 1

F.2

Abbreviation s

V

VA

Ω

W mW

MΩ nm rpm mmHg cmH2O ms mV

Symbols

Symbol

%

/

×

©

±

In Full millimeters of mercury centimeters of water millisecond millivolt milliwatt megaohm nanometer breaths per minute second volt volt ampere ohm watt

Explanation negative, minus percent per; divide; or to plus equal to less than greater than less than or equal to greater than or equal to plus or minus multiply copyright

F - 2 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

F.3

Abbreviations

Abbreviation

CCO

CCU

CE

CFI

C.I.

CIS

CISPR

CMOS

CMS

C.O.

CO

2

COHb

Compl

BTPS

CAA

CaO

2

CCI

BC

BL

BSA

BT aVL aVR awRR

BAP

ATMP

AUC

Avg aVF

AG

AHA

Ao

Art

AaDO

2

AC

ACI

Adu

In Full alveolar-arterial oxygen gradient alternating current acceleration index adult anaesthesia gas

American Heart Association aortic pressure arterial barometric pressure area under the curve rSO

2

average left foot augmented lead left arm augmented lead right arm augmented lead airway respiratory rate brachial arterial pressure burst count baseline body surface area blood temperature body temperature and pressure, saturated

Clinical Assistive Application arterial oxygen content continuous cardiac index continuous cardiac output cardiac (coronary) care unit

Conformité Européenne cardiac function index cardiac index clinical information system

International Special Committee on Radio Interference complementary metal oxide semiconductor central monitoring system cardiac output carbon dioxide carboxyhemoglobin compliance

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual F - 3

F - 4

Abbreviation

EtDes

EtEnf

EtHal

EtIso

EtSev

EtCO

2

EtN

2

O

EtO

ESVI

ESU

Et

EtAA

EMG

EMI

Enf

ESV

EtO

2

EVLW

ELWI

EWS

FAP

FCC

EDV

EEC

EEG

EMC dPmx

DVI

DO

2

DO

2

I

ECG

DC

Des

Dia dpi

COPD

CPI

CPO

CVP

In Full chronic obstructive pulmonary disease cardiac power index cardiac power output central venous pressure direct current desflurane diastolic dot per inch left ventricular contractility digital video interface oxygen delivery oxygen delivery index electrocardiograph end-diastolic volume

European Economic Community electroencephalogram electromagnetic compatibility electromyograph electromagnetic interference enflurane end systolic volume end systolic volume index electrosurgical unit end-tidal end-tidal anesthetic agent end-tidal anesthetic agent end-tidal carbon dioxide end-tidal nitrous oxide ethylene oxide end-tidal oxygen extravascular lung water extravascular lung water index

Early Warning Score femoral arterial pressure

Federal Communication Commission

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Abbreviation

Iso

ITBI

ITBV

LA

I:E

IEC

IEEE

IP

ICP

ICT/B

ICU

ID

HIS

HR

IBP

IBW

LAP

LCD

LCW

GEF

Hal

Hb

Hct

FV

GCS

GEDV

GEDI

FiEnf

FiHal

FiIso

FiSev

FDA

Fi

FiAA

FiDes

FiCO2

FiN

2

O

FiO

2

FPGA

In Full

Food and Drug Administration fraction of inspired inspired anesthetic agent inspired anesthetic agent fraction of inspired carbon oxygen fraction of inspired nitrous oxide fraction of inspired oxygen field programmable gate array flow-volume

Glasgow Coma Scale global end diastolic volume global end diastolic volume index global ejection fraction halothane hemoglobin haematocrit hospital information system heart rate invasive blood pressure ideal body weight intracranial pressure intracranial catheter tip pressure transducer intensive care unit identification inspiratory time: expiratory time ratio

International Electrotechnical Commission

Institute of Electrical and Electronic Engineers internet protocol isoflurane intrathoracic blood volume index intrathoracic blood volume left arm left atrial pressure liquid crystal display left cardiac work

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual F - 5

F - 6

Abbreviation

MAP

Mb

MetHb

MEWS

MLDAP

MRI

MV

MVe

LCWI

LDAP

LED

LL

LVET

LVSW

LVSWI

MAC

NEWS

NIBP

NIF

O

2

O

2

%

OR

MVi

N/A

N

2

N

2

O

Neo

PI

PIF

PIP

Pleth

Ped

PEEP

PEF

PEP

OxyCRG

PA

Paw

PAWP

In Full left cardiac work index

Lightweight Directory Access Protocol light emitting diode left leg left ventricular ejection time left ventricular stroke work left ventricular stroke work index minimum alveolar concentration mean arterial pressure myoglobin methemoglobin

Modified Early Warning Score

Mindray LDAP, Mindray lightweight directory access protocol magnetic resonance imaging minute volume expiratory minute volume inspiratory minute volume not applied nitrogen nitrous oxide neonate

National Early Warning Score noninvasive blood pressure negative inspiratory force oxygen oxygen concentration operating room oxygen cardio-respirogram pulmonary artery airway pressure pulmonary artery wedge pressure pediatric positive end expiratory pressure peak expiratory flow pre-ejection period perfusion index peak inspiratory flow peak inspiratory pressure plethysmogram

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

Abbreviation

SV

SVI

SVR

SVRI

SVV

SR

SSC

SSI

STR

Sev

SI

SIQ

SMR

RL

RR

RSBI

RVEF rSO

2

SaO

2

ScvO

2

SEF

SOFA

SpO

2

SQI

RAP

Raw

Rec

Resp

PVR

PVRI qSOFA

RA

Pmean

PO

2

Pplat

PPV

PR

PVC

PVPI

In Full mean pressure oxygen supply pressure plateau pressure pulse pressure variation pulse rate premature ventricular contraction pulmonary vascular permeability index pulmonary vascular resistance pulmonary vascular resistance index quick Sepsis-Related Organ Failure Assessment right arm right atrial pressure airway resistance record, recording respiration right leg respiration rate rapid shallow breathing index right ventricular ejection fraction regional oxygen saturation arterial oxygen saturation central venous oxygen saturation spectral edge frequency sevoflurane stroke index signal quality index satellite module rack

Sepsis-Related Organ Failure Assessment arterial oxygen saturation from pulse oximetry signal quality index suppression ratio

Surviving Sepsis Campaign signal strength index systolic time ratio stroke volume stroke volume index systemic vascular resistance systemic vascular resistance index stroke volume variation

BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual F - 7

Abbreviation

TVi

TV

UAP

UPS

TI

TP

TRC

TVe

SvO

2

Sync

Sys

TB tcpCO

2 tcpO

2

TD

Temp

TFC

TFI

TFT

USB

UVP

VAC

VEPT

VI

VPB

WOB

In Full venous oxygen saturation synchronization systolic pressure blood temperature transcutaneous carbon dioxide partial pressures transcutaneous oxygen partial pressures temperature difference temperature thoracic fluid content thoracic fluid index thin-film technology injectate temperature total power tube resistance compensation expiratory tidal volume inspiratory tidal volume tidal volume umbilical arterial pressure uninterruptible power supply universal serial bus umbilical venous pressure volts alternating current volume of electrically participating tissue velocity index ventricular premature beat per minute work of breathing

F - 8 BeneVision N22/N19/N17/N15/N12 Patient Monitor Operator’s Manual

P/N: 046-011259-00(1

7

.0)