User Manual: Averagers


Add to my manuals
44 Pages

advertisement

User Manual: Averagers | Manualzz

U

SER

M

ANUAL

A

GILENT

A

CQIRIS

A

VERAGERS

Models covered:

AP235/AP240 with Avgr

U1082A-AVG

AP100/AP200

U1081A

Manual Part Number

U1092-90011

Edition

D-RevL, June 2008

The information in this document is subject to change without notice and may not be construed in any way as a commitment by Agilent Technologies, Inc. While Agilent makes every effort to ensure the accuracy and contents of the document it assumes no responsibility for any errors that may appear.

All software described in the document is furnished under license. The software may only be used and copied in accordance with the terms of license. Instrumentation firmware is thoroughly tested and thought to be functional but it is supplied “as is” with no warranty for specified performance. No responsibility is assumed for the use or the reliability of software, firmware or any equipment that is not supplied by Agilent or its affiliated companies.

You can download the latest version of this manual from http://www.agilent.com/ by clicking on Manuals in the

Technical Support section and then entering a model number. You can also visit our web site at http://www.agilent.com/find/acqiris . At Agilent we appreciate and encourage customer input. If you have a suggestion related to the content of this manual or the presentation of information, please contact your local

Agilent Acqiris product line representative or the dedicated Agilent Acqiris Technical Support

([email protected]) .

Acqiris Product Line Information

USA (845) 782-6544

Asia - Pacific 61 3 9210 2890

Europe 41 (22) 884 32 90

© Copyright Agilent 2008

Adobe® and Acrobat® are trademarks of Adobe Systems Incorporated.

Windows® is a U.S. registered trademark of Microsoft Corporation.

MATLAB® is a registered trademark of The MathWorks, Inc.

User Manual: Agilent Acqiris Averagers Page 2 of 44

CONTENTS

1.

OUT OF THE BOX ............................................................................................................................................. 5

1.1.

Message to the User ....................................................................................................................................... 5

1.2.

Using this Manual .......................................................................................................................................... 5

1.3.

Conventions Used in This Manual ................................................................................................................. 5

1.4.

Model Names ................................................................................................................................................. 6

1.5.

Disclaimer and Safety .................................................................................................................................... 6

1.6.

Warning Regarding Medical Use................................................................................................................... 6

1.7.

Packaging and Handling ................................................................................................................................ 6

1.8.

Warranty......................................................................................................................................................... 7

1.9.

Warranty and Repair Return Procedure, Assistance and Support .................................................................. 7

1.10.

System Requirements................................................................................................................................. 7

1.11.

Transport & Shipping................................................................................................................................. 8

1.12.

Maintenance ............................................................................................................................................... 8

1.13.

Cleaning ..................................................................................................................................................... 8

1.14.

Disposal and Recycling.............................................................................................................................. 8

2.

INSTALLATION................................................................................................................................................. 9

2.1.

Installing the Software under Windows ......................................................................................................... 9

2.1.1.

Warnings ................................................................................................................................................ 9

2.1.2.

Multiple Versions................................................................................................................................... 9

2.1.3.

Installation.............................................................................................................................................. 9

2.2.

Installing the Software for Linux ................................................................................................................. 10

2.2.1.

Installing the Linux device driver ........................................................................................................ 10

2.2.2.

Loading the Linux device driver .......................................................................................................... 11

2.2.3.

Environment variables for the Firmware.............................................................................................. 12

2.3.

Installing the Hardware ................................................................................................................................ 13

2.4.

After Restarting............................................................................................................................................ 13

2.4.1.

Windows 2000 ..................................................................................................................................... 13

2.4.2.

Windows XP ........................................................................................................................................ 13

2.5.

LabVIEW RT............................................................................................................................................... 14

2.6.

Installing the IVI-COM/C Driver................................................................................................................. 14

2.7.

Distribution for Windows 2000/XP and Linux ............................................................................................ 14

3.

PRODUCT DESCRIPTION ............................................................................................................................. 15

3.1.

Overview...................................................................................................................................................... 15

3.2.

Channel Input............................................................................................................................................... 17

3.2.1.

Coupling & Impedance ........................................................................................................................ 17

3.2.2.

Input Protection.................................................................................................................................... 17

3.2.3.

Mezzanine Front-end............................................................................................................................ 17

3.2.4.

Bandwidth and Rise Time .................................................................................................................... 18

3.2.5.

Input Voltage and Offset ...................................................................................................................... 18

3.2.6.

Vertical Resolution............................................................................................................................... 18

3.2.7.

DC Accuracy........................................................................................................................................ 18

3.2.8.

Linearity and Dithering ........................................................................................................................ 18

3.3.

Trigger.......................................................................................................................................................... 18

3.3.1.

Trigger Source...................................................................................................................................... 18

3.3.2.

Trigger Coupling.................................................................................................................................. 19

3.3.3.

Trigger Level........................................................................................................................................ 19

3.3.4.

Trigger Slope........................................................................................................................................ 19

3.3.5.

External Trigger Output ....................................................................................................................... 19

3.3.6.

Trigger Status....................................................................................................................................... 20

3.4.

Data Acquisition – Common........................................................................................................................ 20

3.4.1.

Timing.................................................................................................................................................. 20

3.4.2.

Sampling Rate ...................................................................................................................................... 20

3.5.

Data Acquisition - Digitizer Mode............................................................................................................... 20

3.5.1.

Acquisition Memory and Time Base Range - Digitizer Mode............................................................. 20

3.5.2.

Pre- and Post-Trigger Delay - Digitizer Mode ..................................................................................... 21

3.5.3.

Single and Sequence Acquisitions - Digitizer Mode............................................................................ 21

User Manual: Agilent Acqiris Averagers Page 3 of 44

3.6.

Data Acquisition - Averaging Mode ............................................................................................................ 22

3.6.1.

Averaging Mode................................................................................................................................... 22

3.6.2.

Time Base Range ................................................................................................................................. 22

3.6.3.

Averaging Memory .............................................................................................................................. 22

3.6.4.

Single Average ..................................................................................................................................... 23

3.6.5.

Sequence Average Mode...................................................................................................................... 23

3.6.6.

Trigger Overview ................................................................................................................................. 24

3.6.7.

Random and Synchronized Triggers .................................................................................................... 25

3.6.8.

Noise Suppressed Accumulation.......................................................................................................... 25

3.6.9.

TDC Histogram mode .......................................................................................................................... 27

3.6.10.

Real-time Add/Subtract control............................................................................................................ 27

3.6.11.

Pre Acquisition Triggers ...................................................................................................................... 28

3.7.

External Clock and Reference...................................................................................................................... 28

3.8.

Front Panel Inputs and Controls................................................................................................................... 29

3.9.

Internal Calibration ...................................................................................................................................... 29

3.10.

Electrical, Environmental and Physical Specifications ............................................................................ 30

3.10.1.

Electrical .............................................................................................................................................. 30

3.10.2.

Environmental and Physical................................................................................................................. 30

4.

RUNNING THE ACQIRISLIVE APPLICATION......................................................................................... 31

4.1.

Getting Started with AcqirisLive ................................................................................................................. 31

4.2.

Control Panel and Functions ........................................................................................................................ 32

4.2.1.

Choice of Instrument Class .................................................................................................................. 32

4.2.2.

Control Panel Mode ............................................................................................................................. 32

4.2.3.

Numeric Editor..................................................................................................................................... 33

4.2.4.

Keyboard shortcuts............................................................................................................................... 33

4.2.5.

Input Voltage Scale, Offset and Coupling............................................................................................ 34

4.2.6.

Trigger.................................................................................................................................................. 34

4.2.7.

Acquisition Mode................................................................................................................................. 35

4.2.8.

Timebase and Memory......................................................................................................................... 35

4.2.9.

Segmented Memory ............................................................................................................................. 36

4.2.10.

Display Features, Zoom and Persistence.............................................................................................. 36

4.2.11.

Using averagers/digitizers of different models simultaneously............................................................ 37

4.2.12.

Displaying Multiple Traces.................................................................................................................. 37

4.3.

Top Line Menu of AcqirisLive .................................................................................................................... 38

4.4.

Setup ............................................................................................................................................................ 38

4.5.

Store, Autostore and Conversion of Waveforms.......................................................................................... 38

4.6.

Calibrate....................................................................................................................................................... 39

4.7.

Averager settings.......................................................................................................................................... 40

4.8.

External 10 MHz Reference and External Clock ......................................................................................... 40

4.9.

Additional Waveform Information............................................................................................................... 40

4.10.

Display Features....................................................................................................................................... 40

4.11.

Command Line Switches ......................................................................................................................... 41

5.

APPENDIX A: U1092A-HVB XA100 BNC INPUT OVERVOLTAGE PROTECTION............................ 43

6.

APPENDIX B: U1091AX02 XP102 FAN UNIT FOR THE AP FAMILY OF MODULES ........................ 44

User Manual: Agilent Acqiris Averagers Page 4 of 44

1. Out of the Box

1.1. Message to the User

Congratulations on having purchased an Agilent Technologies Acqiris data conversion product. Acqiris

Digitizers/Averagers are high-speed data acquisition modules designed for capturing high frequency electronic signals. To get the most out of the products we recommend that you read this manual carefully. We trust the product you have purchased will meet with your expectations and provide you with a high quality solution to your data conversion applications.

1.2. Using this Manual

This guide assumes you are familiar with the operation of a personal computer (PC) running a Windows 2000/XP or other supported operating system. It also assumes you have a basic understanding of the principles of data acquisition using either a waveform digitizer or a digital oscilloscope.

The manual is divided into 4 separate sections. To understand the elements of operation for the module it is essential that you read them all.

Chapter 1

OUT OF THE BOX, describes what to do when you first receive your new Acqiris product.

Special attention should be paid to sections on safety, packaging and product handling. Before installing your product please ensure that your system configuration matches or exceeds the requirements specified.

Chapter 2

INSTALLATION, covers all elements of installation and performance verification. Before

attempting to use your Acqiris product for actual measurements we strongly recommend that you read all sections of this chapter.

Chapter 3

PRODUCT DESCRIPTION, provides a full description of all the functional elements of the

digitizer.

Chapter 4

RUNNING THE ACQIRISLIVE APPLICATION, describes the operation of AcqirisLive 3.1, an

application that enables basic operation of Acqiris Digitizers and Averagers in a Windows 2000/XP environment. Note: AcqirisMAQS is an alternate software application offering many interesting possibilities for the control of acquisition systems in a single or multi-machine environment. Ask your sales representative or Agilent for more information.

For information necessary for writing your own software to control Acqiris products you should refer to the

Programmer’s Guide and the Programmer’s Reference Manual.

1.3. Conventions Used in This Manual

The following conventions are used in this manual:

WARNING

CAUTION

NOTE

Italic

This icon to the left of text warns that an important point must be observed.

Denotes a warning, which advises you of precautions to take to avoid being electrically shocked.

Denotes a caution, which advises you of precautions to take to avoid electrical, mechanical, or operational damages.

Denotes a note, which alerts you to important information. text denotes a warning, caution, or note. text is used to emphasize an important point in the text or a note text is used for sections of code, programming examples and operating system commands.

Bold Italic mono

B,KB,MB,GB is for Byte, KiloByte = 1024 bytes, MegaByte = 1024*1024 bytes, GigaByte = 1024*1024*1024 bytes

User Manual: Agilent Acqiris Averagers Page 5 of 44

1.4. Model Names

Agilent Technologies Inc. acquired Acqiris SA and its product lines in December 2006. Use the tables below to cross reference the legacy model name and new Agilent numbers

Agilent

Model

Number

U1081A-001

Acqiris

Model

Name

U1081A-003

AP100

AP200

U1082A-002-AVG AP235

U1082A-001-AVG AP240

1.5. Disclaimer and Safety

The AP Series PCI Averager cards have been designed to operate in a standard PCI slot found inside most personal computers. Power for the modules is provided by plugging them into a free slot (refer to the installation procedure).

CAUTION

:

Do not exceed the maximum input voltage rating! The maximum input voltage for 50

Ω input

impedance is

±5 V.

1.6. Warning Regarding Medical Use

The AP Series PCI Averager cards are not designed with components and testing intended to ensure a level of reliability suitable for use in treatment and diagnosis of humans. Applications of these cards involving medical or clinical treatment can create a potential for accidental injury caused by product failure, or by errors on the part of the user. These cards are not intended to be a substitute for any form of established process or equipment used to monitor or safeguard human health and safety in medical treatment.

WARNING

:

The modules discussed in this manual have not been designed for making direct measurements on the human body. Users who connect an Acqiris module to a human body do so at their own risk.

1.7. Packaging and Handling

Your Averager is shipped with the following components:

• A compact disc in an Agilent Technologies paper CD envelope that includes o

10 product User Manuals in electronic form (8-bit Digitizers, 10-bit Digitizers, 12-bit Digitizers,

Averagers, Analyzers, Signal Analyzers, Streamer Analyzers, Time-to-Digital Converters, 3-, 5-, and 8-slot CompactPCI Crates, and the 21-slot CompactPCI Crate), o

1 Programmer’s Guide and 1 Programmer’s Reference Manual, o device drivers with sample software for different operating systems, environments and languages, o the AcqirisAnalyzers application, a demonstration program for the AC/SC Analyzer products, o the AcqirisLive application, a demonstration program for our digitizer and averager products, o the DemoSSR application, a demonstration program for the Acqiris AP235/AP240 Analyzers, o the DemoAPX01 application, a demonstration program for the Acqiris AP101/AP201 Analyzers, o the DemoTC application, a demonstration program for the Acqiris TC840/TC890 Time-to-Digital

Converters, o full installation procedures for use with Microsoft Windows, National Instruments LabVIEW RT,

Wind River VxWorks, IVI-COM/C, and Linux software.

• A declaration of conformity

• Optional documentation such as a model-dependent document giving Specifications & Characteristics, a

Calibration Certificate, or a Performance Verification

After carefully unpacking all items, inspect each to ensure there are no signs of visible damage. Also check that all the components received match those listed on the enclosed packing list. Agilent cannot accept responsibility for missing items unless we are notified promptly of any discrepancies. If any items are found to be missing or are

User Manual: Agilent Acqiris Averagers Page 6 of 44

received in a damaged condition please contact the Agilent service center or your local supplier immediately. Retain the box and packing materials for possible inspection and/or reshipment.

1.8. Warranty

All Agilent Acqiris Digitizer products are warranted to operate within specification, assuming normal use, for a period of at least one year from the date of shipment. Units sold before April 2008 had three year warranties, as do some more recent ones; in case of doubt examine your invoice. It is recommended that yearly calibration be made in order to verify product performance. All repairs, replacement and spare parts are warranted for a period of 3 months.

Warranty extensions are available as an option.

Agilent endeavors to provide leading edge technology that includes the latest concepts in hardware and software design. As such software and firmware used with the products is under continual refinement and improvement. All software and instrument firmware is supplied “as is” with no warranty of any kind. Software and firmware is thoroughly tested and thought to be functional at the time of shipment. At Agilent’s discretion software and firmware may be revised if a significant operational malfunction is detected.

In exercising this warranty, Agilent will repair or replace any product returned to the Agilent service center, within the warranty period. The warranty covers all defects that are a result of workmanship or materials. This excludes defects that are caused by accident, misuse, neglect, or abnormal operation.

The purchaser is responsible for returning the goods to the nearest Agilent service center. This includes transportation costs and insurance. Agilent will return all warranty repairs with transportation prepaid.

1.9. Warranty and Repair Return Procedure, Assistance and Support

Agilent acquired Acqiris SA and its product lines in December 2006. Please contact your nearest Agilent Service

Center before returning any product for repair.

You can find information about technical and professional services, product support, and equipment repair and service on the Web, see http://www.agilent.com/find/service (or http://www.agilent.com/ and after selecting your country click on Contact Us). The service center will ask for your name, company, phone number and address, the model and serial numbers of the unit to be repaired, and a brief description of the problem.

Before issuing a Service Order the service center may ask you to communicate with us by phone or eMail so that we can learn as much as needed about the problems observed. If a unit returned under guarantee is found to be working normally and this procedure was not followed we reserve the right to charge you for the work done.

For your nearest customer support center please contact Acqiris Technical Support

([email protected]) or come visit our web site at http://www.agilent.com/find/acqiris .

Alternatively, contact Acqiris at 1-845-782-6544 in the USA, +41 22 884 32 90 in Europe or +61 3 9210 2890 in the

Asia-Pacific region. The Agilent Support Centers can also help redirect you for any questions concerning the installation and operation of your equipment.

1.10. System Requirements

Acqiris products need the following minimum PC System Requirements in order to obtain reasonable performance from your averager:

Processor: 150 MHz Pentium (higher recommended). Some PowerPC systems running Wind River VxWorks are supported; please contact us for details.

Memory: 64 MB RAM. The previous number is a very rough estimate. Assuming that you are using AcqirisLive or an application of your own that operates on the acquired data it seems reasonable to ask for 10 times the total acquisition memory that you will be using at the same time in the application. Performance is likely to be degraded if less memory is available.

Display resolution: At least 800 x 600 pixels and 256 colors for use of AcqirisLive or AcqirisDemo

Operating System: Microsoft Windows 2000/XP including 2003 Server, Wind River VxWorks 5.5.1 and 6.4, and

Linux with kernels 2.4 and 2.6. Support for Windows 95/98/NT4 is included “as is” since these operating systems are no longer supported by Microsoft.

Hard Drive Space: 20 MB Minimum

CD Drive (or any method to copy the software installation files from CD to the hard drive such as LAN, floppy drive, etc.)

User Manual: Agilent Acqiris Averagers Page 7 of 44

LabVIEW: A driver is available for National Instruments LabVIEW versions 7.1or 8.0

LabVIEW RT: A driver is available for National Instruments LabVIEW RT version 7.1 or higher. The VISA driver must be version 3.0 or higher.

MATLAB: The MEX interface can be used with MathWorks MATLAB

7.3 or a newer version.

Visual BASIC: The interface files and examples are available for Microsoft Visual Basic versions 5 or 6 and the interface files only for .NET.

Tornado: The example files are useable with Wind River Tornado 2.2.1

1.11. Transport & Shipping

CAUTION:

Cards can be safely transported in their original shipping packages. The transport of AP cards

mounted in a PC is a more delicate issue. Because of their mass the cards can vibrate loose unless they are properly secured and braced. However, properly mounted AP cards with XP102 fans can be sufficiently well held; the Adjustable retainer must be used so as to hold the card in place.

To package the instrument for shipping:

Step Notes

1. Place the instrument in its original packaging materials.

If the original packaging materials are not available, use a professional packaging service. Contact your

Agilent Service Center for more information.

2. Surround the instrument with at least 3 to 4 inches (8 to 10 cm) of its original packing material or bubble-pack to prevent the instrument from moving in its shipping container.

3. After wrapping it with packing material, place the instrument in its original shipping container or a strong shipping container that is made of double-walled corrugated cardboard with 159 kg (350 lb) bursting strength.

The shipping container must be large and strong enough to accommodate your instrument and allow at least 3 to 4 inches (8 to 10 cm) on all sides for packing material.

4. Seal the shipping container securely with strong nylon adhesive tape.

5. Mark the shipping container “FRAGILE, HANDLE

WITH CARE” to help ensure careful handling.

6. Use the address obtained from your Agilent

Technologies Service Center.

7. Retain copies of all shipping papers.

CAUTION: Damage can result if the original packaging materials are not used. Packaging materials should be anti-static and cushion the instrument on all sides. NEVER USE STYRENE PELLETS IN ANY SHAPE AS

PACKAGING MATERIALS. They do not adequately cushion the instrument or prevent it from moving in the shipping container. Styrene pellets can also cause equipment damage by generating static electricity or by lodging in fan motors.

1.12. Maintenance

The cards do not require any maintenance. There are no user serviceable parts inside. A periodic calibration can be obtained on request.

1.13. Cleaning

Cleaning procedures consist only of exterior cleaning.

Clean the exterior surfaces of the module with a dry lint-free cloth or a soft-bristle brush. If any dirt remains, wipe with a cloth moistened in a mild soap solution. Remove any soap residue by wiping with a cloth moistened with clear water. Do not use abrasive compounds on any parts.

1.14. Disposal and Recycling

Electronic equipment should be properly disposed of. Digitizers and their accessories must not be thrown out as normal waste. Separate collection is appropriate and may be required by law.

User Manual: Agilent Acqiris Averagers Page 8 of 44

2. Installation

This chapter describes how to install the Acqiris hardware and software for Windows 2000/XP, National Instruments

LabVIEW RT, Linux, or Wind River VxWorks.

NOTE:

For a first time installation we strongly recommend installing the software before inserting the hardware into the PC.

2.1. Installing the Software under Windows

2.1.1. Warnings

If Setup detects a previous installation of Acqiris software on your system, a warning screen will be displayed. It is recommended to exit Setup and uninstall older versions.

The installer from software releases prior to Acqiris Software 2.0 installed the Digitizer Driver DLL files into the

System directory. These will be removed by Setup. If you wish to keep the old installation on your system, you should exit Setup, and move all Acqiris driver files (acqiris*, acqrs* and acqir*) to some archive directory.

The DLL files will be installed into the bin subdirectory of the Acqiris software root, and the corresponding path will be added to the PATH environment variable.

2.1.2. Multiple Versions

With the software installation from Acqiris Software 2.0 (or above), it is possible to keep multiple versions on the same system, but you must specify a different root directory (i.e. Install Folder). If you keep the same directory,

Setup will overwrite your previous installation.

To go back to a previous version, you must change the PATH environment variable and reinstall the Kernel driver.

Under Windows 2000/XP:

1. Copy the SYS file from <old_AcqirisSoftware_root>\bin\kernel to the

Windows\System32\drivers

directory.

2. Change the AcqirisDxRoot, AcqirisDxDir and PATH environment variables to the old root.

3. Reboot the computer.

2.1.3. Installation

Before installing the Acqiris hardware, you should complete the following steps to install the software for Windows

2000/XP.

NOTE: You will need administrator privileges to complete the software installation under Windows

2000/XP.

1. Insert the Acqiris Software CD into the CD-ROM drive of your computer. If the Autorun program does not start automatically (Autoplay disabled), you can start it manually, or navigate to the AcqirisSoftware\Windows folder in order to display the files included.

2. Choose Install AcqirisSoftware for Windows 2000/XP/2003 Server (or run Setup.exe from the

AcqirisSoftware\Windows folder). After several seconds for initialization the first of many screens will appear.

Please note the following points:

ƒ It is good practice to remove any previously installed version o0f Acqiris software. If the program finds that there is still Acqiris software installed on your machine a warning panel will appear.

ƒ In the Select Install Type window selecting Custom installation will let you select individual packages for loading. The space indicated for LabVIEW, Firmware and UserManual packages is incorrect. The correct values are 7 MB, 40 MB, and 30 MB respectively. A full installation requires just under 150 MB of disk.

ƒ If MATLAB is installed on your machine, you will be asked to point the installer to the MATLAB root directory. You should do this if you want the installer to modify the standard startup.m file to incorporate the Acqiris adaptor.

User Manual: Agilent Acqiris Averagers Page 9 of 44

ƒ In the Installation Folder window you will give the name of the root directory of the Acqiris software installation. If User Manuals (30 MB) and Firmware (40 MB) are loaded more space than indicated here will be required on the drive. For the case of a Tornado 2.2 installation the folder name should not contain any spaces.

ƒ AcqirisLive needs the LabWindows/CVI 7.0 Run-Time Engine to run. If Setup has detected that a

LabWindows/CVI Run-Time Engine is already installed on your system, it will ask you if you would like to install it locally for AcqirisLive anyway. If you are not sure about the version of the CVI Run-Time Engine on your system, it is recommended to install it locally.

ƒ The Installation Summary window will be shown to allow you to check what you have asked for. At this point it is not too late to go back and make changes. The actual installation will only be started after clicking on “Install” in the next window.

ƒ Please read the IMPORTANT Information window text. It could help you avoid serious problems.

ƒ Registration of your installation will help Agilent provide you with better support. You will also be notified of updates and upgrades. All information submitted will be treated in accordance with Agilent’s privacy policy. Setup will prepare a registration e-mail in your e-mail client application upon termination of the setup procedure. You can then decide whether or not you wish to send it. You may also add comments.

Uncheck the box if you do not want to register your installation.

ƒ After the software installation is complete you can either accept the suggestion to restart the computer or you should shutdown your computer and proceed with the hardware installation.

2.2. Installing the Software for Linux

The Acqiris software is ready to install on Linux systems running kernels 2.4 and 2.6. The following archives all contain the driver and library compiled with the gcc version shown:

AcqirisLinux-3.2a-gcc-3.3.tar.gz

- compiled under Debian ‘sarge’ 3.1 with gcc 3.3.

AcqirisLinux-3.2a-gcc-3.4.tar.gz

- compiled under SL 4.4 with gcc 3.4.

AcqirisLinux-3.2a-gcc-4.1.tar.gz

- compiled under Debian ‘etch’ 4.0 with gcc 4.1.

The appropriate tar file should be copied to a local directory (e.g. your home directory) and then unpacked by using the following command tar xzf AcqirisLinux-3.2a-gcc-X.x.tar.gz

The resulting directory AcqirisLinux contains an install script drv-install and a graphical Demo program

demo/AcqirisDemo. Before running the software you have to install the device driver into the running kernel

2.2.1. Installing the Linux device driver

The acqrsPCI device driver is a loadable kernel module. It is provided both as source files and already compiled binary files for some kernel versions.

The compiled loadable kernel module is available for:

• Debian ‘sarge’ 3.1: kernel 2.6.8-3-686

• Suse Linux 9.2: kernel 2.6.8-24-smp

• Scientific Linux 4.4: kernel 2.6.9-42.0.10.EL-i686

• Debian ‘etch’ 4.0: kernel 2.6.18-4-686

For other Linux distributions, the loadable kernel module will have to be compiled from the sources.

Using a compiled kernel module

You can find the running version of your Linux kernel by running the command:.

uname -r

Then, if there is a kernel module that matches yours, you can rename the file:

cd AcqirisLinux/lib/modules

cp acqrsPCI.ko[.n] acqrsPCI.ko where [.n] is the running version of the Linux kernel.

To install the device driver you have to rename the kernel module whose name matches the version of your kernel to a name without this version number.

Compiling the kernel module

User Manual: Agilent Acqiris Averagers Page 10 of 44

All files needed to compile a new kernel mode driver are in AcqirisLinux/linuxdriverpci for Linux kernels 2.4, or AcqirisLinux/linux2.6driverpci for Linux kernels 2.6.

To compile the driver, the header files of the Linux kernel need to be installed. Furthermore, depending on the Linux distribution, the kernel source files may need to be installed too.

Compiling for Linux kernels 2.6

The makefile will try to guess where the kernel headers are located. If it fails, see the instructions inside the makefile itself to specify manually where these kernel header files are located.

To compile a new kernel module, use the following commands:

cd linux2.6driverpci

make clean all

make install to generate a new kernel module from scratch to copy the kernel module where it should reside

(this last command require super user privileges).

In the end, the loadable kernel module should be present in the system's loadable kernel modules directory. You can verify that it is present by running:

ls -l /lib/modules/$(uname -r)/extra/acqrsPCI.ko

For Debian users, in order to compile the kernel module on a system, the kernel packages that match the current kernel-image must be installed. For instance, on a Debian ‘etch’ 4.0 system 686 the required packages are:

linux-image-2.6-686

linux-image-2.6.18-4-686

linux-headers-2.6-686

linux-headers-2.6.18-4

linux-headers-2.6.18-4-686

linux-kbuild-2.6.18

Note that this version of the loadable kernel module had been tested on Linux kernel versions up to 2.6.18. However, starting with kernel 2.6.17, a few kernel functions relating to the device classes are no longer available to proprietary modules. Thus automatic creation of the /dev/acqrsPCI node has to be done manually in the /etc/rc.local file (see below).

Compiling for Linux kernels 2.4

The makefile will try to guess where they are located. If it fails, see the instructions inside the makefile itself to specify manually where these kernel header files are located.

The path variable INCLUDEDIR in the makefile has to point to the correct kernel header files. The default is

INCLUDEDIR=/usr/src/linux-2.4

. This path usually is a link to the actual header files (driver source). To compile for a new kernel, issue the following commands:

cd linuxdriverpci

make clean

Make

drv-install rem

drv-install add to remove all *.o files in AcqirisLinux/linuxdriverpci. to generate a new kernel mode driver acqrsPCI.o and also copy it to the directory AcqirisLinux/lib/module where the install script (drv_install) can access it. to remove the previous installed kernel mode driver. to install the new driver.

In the end, the loadable kernel module should be present in the system's loadable kernel modules directory. You can verify that it is present by running:

ls -l /lib/modules/$(uname -r)/ACQIRIS/acqrsPCI.o.

2.2.2. Loading the Linux device driver

To install the device driver and the load script to the system, you have to get super user privileges and execute the driver install script by typing, cd AcqirisLinux

./drv-install add

You can check that the driver is loaded properly with lsmod.

/sbin/lsmod | grep acqrsPCI

User Manual: Agilent Acqiris Averagers Page 11 of 44

If lsmod does not report a module named acqrsPCI, then the kernel module is not loaded into the kernel. Read the next section

You can check that a device node has been created.

ls -l /dev/acqrsPCI

If the device node /dev/acqrsPCI is not present on your system, the Acqiris Software will not be able to access the device. Read more in the section below.

The drv-install script has the following additional functionality:

drv-install rem to remove the driver and load script

drv-install res to restart the driver

Loading the module into the kernel

If you are running a Linux distribution that does not use the standard paths for the load scripts, you have to load the device driver manually.

The kernel module should be present in the kernel modules directory. Check with the following command:

ls -l /lib/modules/$(uname -r)/extra/acqrsPCI.ko

First update the kernel modules dependencies:

/sbin/depmod -a

Then load the kernel module:

/sbin/modprobe -v /lib/modules/$(uname -r)/extra/acqrsPCI.ko

The acqrsPCI kernel module should now be loaded correctly. Check with lsmod.

You can add the command above to your /etc/rc.local file. It is run at system startup and will load the acqrsPCI kernel module automatically when your computer is restarted.

Creating the device driver node

For systems using udev (the dynamic /dev/ hierarchy) or running kernels above 2.6.17, the device driver node may not be created automatically. You can do it with mknod running the following command:

/bin/mknod -m 666 /dev/acqrsPCI c 124 0

You can check that the device node has been created:

ls -l /dev/acqrsPCI

You can add the mknod command to your /etc/rc.local file. It is run at system startup and will create the

/dev/acqrsPCI

device node automatically when your computer is restarted.

Note that the permissions that are set in this example (-m 666) gives access to the Acqiris instruments to all the users.

This may not be what you expect and these permissions can be adjusted to fit your requirements. The basic rule is that any user that wants to use the Acqiris instruments has to get read and write access to /dev/acqrsPCI.

2.2.3. Environment variables for the Firmware

Automatic loading of the firmware needed by 12-bit, analyzer, and averager modules relies on the environment variable AcqirisDxDir pointing to the directory containing the file AqDrv4.ini which in turn points to the directory containing the Firmware .bit files. Therefore, assuming that your Firmware is in /usr/local/AcqirisLinux/Firmware and that AqDrv4.ini is in /usr/local/AcqirisLinux/demo then you must edit AqDrv4.ini so that it contains the line

fpgaPath=/usr/local/AcqirisLinux/Firmware

Then, if your shell is csh or tcsh modify the /etc/csh.login file to contain the line

setenv AcqirisDxDir /usr/local/AcqirisLinux/demo

or, if your shell is bash, ksh, zsh or sh, modify the /etc/profile file to contain the lines

AcqirisDxDir=/usr/local/AcqirisLinux/demo export AcqirisDxDir

User Manual: Agilent Acqiris Averagers Page 12 of 44

2.3. Installing the Hardware

1. Turn off the power of the PC and the crate in the case of a CompactPCI module.

CAUTION:

For PCI modules the PC may have to be unplugged to ensure that the PCI bus has no power

available. However, CompactPCI crates can be left plugged in since this ensures proper

grounding.

CAUTION:

Touch the antistatic package to a grounded object before removing the card from the package.

Electrostatic discharge can damage the card.

2. Open the PC, identify a free PCI slot and carefully insert the DP Series card into it. Be sure to ground yourself by touching the grounded PC frame and avoid touching any components on the DP Series card.

Make sure that the grounding of the card’s mounting bracket to the back panel rail of the computer is done correctly. If present make sure that the fan's adjustable retainer is correctly positioned and tightened for mechanical support. Close the PC.

3. Turn on the power of the PC and start the operating system.

NOTE: Acqiris modules are equipped with a LED. If this LED is not glowing orange or red when the power is applied there is a severe problem. Either the module is broken or the necessary voltages for its use are not

available.

4. If devices were installed using a previous version of Acqiris software the instruments in these logical positions may still appear as Unknown Devices. This can be changed to the new Acqiris type category with the Grey Diamond icon by Uninstalling the device and then Installing again. Instructions on this procedure can be found in the ReadMe.txt file in the manuals folder of your Acqiris software installation.

2.4. After Restarting

2.4.1. Windows 2000

Under Windows 2000, you must login with administrator privileges after the first boot following the hardware installation; the Plug&Play system must have the appropriate privileges to be able to complete your hardware installation successfully. After a successful hardware installation, you will be able to use your Acqiris module(s) with normal privileges.

At the first boot following the hardware installation, Windows will detect the new hardware and will install the devices automatically. The following image will appear.

NOTE: In some systems an application program (such as AcqirisLive) will not yet work correctly at this point. One

additional boot cycle may be needed if this is the first time that a hardware board is being installed.

2.4.2. Windows XP

Under Windows XP, you must login with administrator privileges after the first boot following the hardware installation; the Plug&Play system must have the appropriate privileges to be able to complete your hardware installation successfully. After a successful hardware installation, you will be able to use your Acqiris digitizer(s) with normal privileges.

If you login with administrator privileges after the first boot following the hardware installation, Windows will detect the new hardware and start the “Found New Hardware Wizard” after a few seconds. There is no need to use

Windows Update to search for the software. You can “Install the software automatically”. The final screen should appears as shown here:

User Manual: Agilent Acqiris Averagers Page 13 of 44

NOTE: In some systems an application program (such as AcqirisLive) will not yet work correctly at this point. One additional boot cycle may be needed if this is the first time that a hardware board is being installed.

2.5. LabVIEW RT

During program development you can choose whether you use LabVIEW or LabVIEW RT compatible libraries by switching the version present in National Instruments\LabVIEW m.n\instr.lib\. This swap can be facilitated by using the Install VI library for LabVIEW or LabVIEW Real-Time shortcut available in the Shortcut folder under

Start

→ Programs.

There is only one Acqiris Driver. It supports all Acqiris Instruments. The instructions below concern LabVIEW RT as used in NI PXI processors.

The Aq_RT.inf and AqRT_4.ini files must be uploaded to the target. To do this,

• start the MAX application,

• right click on the target

• select file transfer

• select the Aq_RT.inf file on your host machine and upload ('To Remote') to the LabVIEW RT working directory (/NI-RT/system) on the target

• select the Aq_RT.ini file on your host machine and upload ('To Remote') to the LabVIEW RT working directory (/NI-RT/system) giving it the name AqDrv4.ini

For Acqiris modules which need FPGA files you should,

• create the folder \firmware in the /NI-RT/system directory using the file transfer application

• select the FPGA files (from <AcqirisDxRoot>\Firmware) you want to copy to the target and upload them into the firmware directory

Restart the target after finishing the file transfers.

Restart the MAX Explorer and you should have Acqiris digitizers detected in your PXI system.

2.6. Installing the IVI-COM/C Driver

Please install Acqiris software for Windows first. Then the Acqiris Software CD Autorun program gives access to two installers:

ƒ IVI Shared Components 1.4

ƒ IVI-COM/C IviAqD1 driver

These must be installed in the order shown above. For more information you can then consult the Readme.txt file in the IVI\Drivers\IviAqD1 folder or the documentation through the program shortcut present under Ivi/IviAqD1.

2.7. Distribution for Windows 2000/XP and Linux

The manuals/ReadMe.txt file contains a list of files to be found after a complete installation of Acqiris software on

Windows systems. Similarly the ReadMeLinux file gives the list of files corresponding to that installation.

User Manual: Agilent Acqiris Averagers Page 14 of 44

3. Product Description

3.1. Overview

Acqiris Averagers are designed to provide superior measurement precision and accuracy. Key acquisition specifications (such as DC accuracy, integral and differential non-linearity) have been optimized to deliver maximum measurement fidelity. Careful circuit layout, custom IC’s and special packaging techniques have all been employed to reduce the overall random and correlated system noise. The use of custom IC’s also dramatically reduces the total number of discrete components required. This has tremendous benefits on reliability and also allows the modules to use a minimal amount of power.

The AP series Averagers are PCI modules that plug directly into any vacant PCI slot in a PC. They are fully programmable over the PCI bus and can deliver 2 GS/s (AP240/AP200) and 1 GS/s (AP235/AP100) real time averaging performance. Signals on 2 channels with an Analog bandwidth limit at 1 GHz (AP240) or 0.5 GHz

(AP235) can be analyzed simultaneously at sampling rates up to 1 GS/s and 0.5 GS/s, respectively. Alternatively, the

AP100 and AP200 are single channel instruments with 0.5 GHz Analog bandwidth. Averaged Data can be transferred to a host processor over the PCI bus at speeds up to 100 Mbytes/s.

For complete technical specifications concerning your particular digitizer please refer to the product’s Specifications

& Characteristics. In addition, we maintain up-to-date versions of all product data sheets on our web site

( http://www.agilent.com/find/acqiris ). The data sheets are available in pdf format and are best viewed using Adobe

Acrobat software. If you have trouble accessing our web site, or viewing the data sheets, please contact your nearest sales office.

Front

Panel

Signals

CH1

CH 2

Dual Channel

M ezzanine

C H1

AP235/AP240 BLOCK

DIAGRAM

Cross

Point

Switch

ADC1

1 G S/s 8-bit

O ffset and G ain C ontrol

C H2

Ditherer

8-bit DA C

EXT

CK in

Trig

O ut

I/O A

I/O B

MULTI_IO

M ezzanine

Trigger

Circuit

ADC2

1 G S/s 8-bit

COS

Synchronizer

Clock Synch

& Start

P1

P2

Internal

Reference

Clock

M AC 1

D em ux

&

M em ory

Processing

Block

M AC 2

D em ux

&

M em ory

2xFPG A

XC2V 3000

24x m em ory

CPLD

PLX

PCI Interface

100 M B /s

PCI

Bus

User Manual: Agilent Acqiris Averagers Page 15 of 44

Key features

ƒ High Speed and Precision – All averager models provide both an ultra-fast real-time summing capability and exceptionally high data transfer rates to the host processor.

The two-channel AP240 delivers a real-time sampling rate of up to 1 GS/s with 1 GHz of analog bandwidth simultaneously on both channels, while the AP235 offers sampling to 0.5 GS/s and a bandwidth of 0.5 GHz.

Alternatively they can be used as one channel instruments with double the sampling rate. The one-channel

AP200 delivers a real-time sampling rate of up to 2 GS/s with 500 MHz of analog bandwidth, while the

AP100 offers sampling to 1 GS/s.

The high sampling rate is achieved in realtime for each trigger and does not require the use of interleaved sampling techniques. Compared to real-time sampling, interleaved techniques significantly slow down averaging speeds and reduce the total data throughput.

ƒ Signal conditioning The AP235/AP240 models include all the required input signal conditioning such as gain, from 50 mV to 5 V Full Scale, offset adjustment and 50 Ω DC coupling. The models AP100/AP200

Averager offer similar functionality with gains ranging from 50 mV to 500 mV Full Scale. The inputs are fully protected, and calibration is a built-in feature. A crystal controlled time base and advanced trigger circuits are used to allow fast and stable acquisitions.

ƒ Improved Timing – In order to preserve the timing accuracy of successively summed acquisitions, the aveargers include two operating modes, with respect to the trigger. These are the Non-Resync Mode with jitter (clock positioning with respect to the trigger from acquisition to acquisition) of

± one-half clock cycle

(minimum of 500 ps), and the ReSync Mode with jitter of

± 200 ps. The ReSync Mode also offers a trigger output with its edge synchronized to the sampling clock.

ƒ Better Accuracy - Vertical resolution can also be improved with a special dithering circuit that can increases the dynamic range from 8 to 12 bits (when averaging more than 256 spectra). Dithering a signal involves the addition of a small, systematic, pseudo-random amount of DC offset to the acquired signal. The dithering range is user programmable within

± 15 ADC LSB’s. Dithering improves measurement resolution and can dramatically reduce the errors associated with ADC effects such as differential non-linearity.

ƒ Fast Averaging - The AP240 and AP235 systems have 2 MSamples of 24-bit accumulation memory/channel

(optionally 8 MSamples/channel). The AP200 and AP100 systems have 2 MSamples of accumulation memory (optionally 8 MSamples). The accumulation memory receives the converted data from the digitizer, which in real-time manages the high-speed accumulation function. A high-speed on-board computing circuit performs the retrieve, add and store functions necessary to accumulate the signal in real time at the maximum sampling rate. At the highest sampling rate, summed events can be acquired with less than 1 μs dead time between them. For example, multiple acquisitions made at 2 GS/s sampling rate and containing 250

KSamples can be summed at sustained rates of up to 8000 shots per second. If 100 KSample records are used, the sustained summing rate can reach 20,000 shots per second. Once a sum is accumulated to the desired depth, a number of events selectable from one to 65,536, the signal is read out over the PCI bus and the division function necessary to obtain the average is performed inside the PC (if desired).

ƒ Add/Subtract Feature – For the AP240 and AP235 systems the DPU Ctrl (I/O P1 and P2) allow hardware to choose, at the instant the trigger arrives, whether the data of the new trace will be added to or subtracted from the accumulated sum. This gives the user control over how background noise is taken into account in the accumulated sum.

ƒ Sequence Average Modes – In time resolved applications, where multiple averaged waveforms must be acquired with very low dead time, the user can partition the averaging memory to store from 1 to 8191 separate accumulations of summed data. In standard segment accumulation mode the user selects the desired number of triggers per segment, N. The next segment will be started after the previous N triggers have been processed. For the AP240 and AP235 a round robin segment accumulation is also available. In this mode consecutive triggers can be accumulated into different segments. In particular, for stimulus-response experiments, one segment could be the averaged stimulus and another could be the averaged response. For the AP235 and AP240 in dual-channel mode, the length of each segment is adjustable from 16 samples up to the maximum 2 MS in steps of 16 points, while in the single-channel mode it is adjustable from 32 samples in steps of 32 samples. In the AP200, the length of each segment is adjustable from 32 samples up to the maximum 2 MS in steps of 32 samples, while in the AP100, it is adjustable from 16 samples to the maximum

2 MS in steps of 16 samples.

ƒ Numerous front panel control signals – In order to ease the integration and synchronization of the

Averager in a large variety of systems, a number of signals are made available on MMCX connectors on the instruments front panel. The following high-speed timing signals are available for all models: Inputs for

Clock, Reference Clock, and Trigger Enable, Outputs for Trigger, Trigger Ready, Acquisition state, and

Reference Clock. They also can accept a hardware input that controls the start of an average accumulation.

Furthermore, the AP240 and AP235 have Add/Subtract control inputs as described above.

User Manual: Agilent Acqiris Averagers Page 16 of 44

ƒ High Data Throughput –All of the Acqiris Averagers can easily be integrated into any standard computer with 1 free long PCI slot that provides the 3.3 V supply and adequate power as required by the cards.

Averaged signals can be transferred, in DMA mode, directly to the processor over the PCI bus at rates of up to 100 Mbytes per second (4 ms for a 100 KS record, 10 ms for a 250 KS record, including DMA set up time overhead). In the case of the AP235 and AP240, if the full precision of the accumulated sum data is not desired, the 16-bit fast readout capability can further shorten the data transfer time.

ƒ Noise Suppressed Accumulation – The Acqiris Averagers provide a noise suppression feature which makes if possible for the AP cards to enhance rarely occurring events, which, under most circumstances, disappear in signals with baseline noise. Only data points that exceed a user-defined threshold are added to the result.

This considerably reduces the noise contribution and the signal-to-noise ratio is enhanced. Furthermore, the

NSA mode implements a fixed baseline subtraction feature that is applied to the data values that are over the threshold. This can be used to greatly reduce the occurrence of data overflows. Therefore many more events can be accumulated before running the risk of data overflow.

ƒ TDC mode – The Acqiris Averagers can also be used to accumulate a histogram of Time to Digitial

Conversion (TDC) data. This mode exploits the threshold and baseline features defined in the NSA section above. The data of each segment is scanned to locate events above a user-defined threshold. Assuming that the event has at least the desired duration, a user-selectable value, the histogram will be incremented at the time of the middle of the event. The increment can be either a single count or the baseline corrected value at the middle of the event.

3.2. Channel Input

The principle characteristics of the input channels are given in the table below:

Model

Agilent #

Bandwidth into 50

Ω

Minimum

Rise Time

BW Limiter selections

Maximum

Full Scale

1 GHz 0.35 ns 20, 200, 700

MHz

5 V

AP240

U1082A

AP235

U1082A

AP100/

AP200

U1081A

0.5 GHz

0.5 GHz for

> 50 mV FS

200 MHz @

50 mV FS

0.7 ns

0.7 ns

1.8 ns

20, 200 MHz

25 MHz

5 V

0.5 V

Maximum

Offset

5 V

Other particularities

2 channels

5 V

2 V

2 channels

1 channel with 2 selectable inputs

3.2.1. Coupling & Impedance

Two 50

Ω signal input BNC connectors are provided. The AP240 and AP235 are two channel modules. In the case of the AP100 and AP200 these two connectors are multiplexed to the module's input. The active input may be selected by software to allow connection of one of the two possible signal sources. The 50

Ω coupling offers high quality termination with better than

± 1% precision. It is ideally suited for use with 50 Ω transmission lines.

All modules offer DC coupling. The AP240 and AP235 also provide AC coupling with a 32 Hz cutoff frequency.

3.2.2. Input Protection

The input amplifiers are fully protected against over-voltage signals. The device can tolerate an input current of up to

40A for 1

μs without damage to the front-end electronics. For extreme cases, such as detector sparking, an external

coaxial spark gap and attenuator can be supplied on request (See APPENDIX A: U1092A-HVB XA100 BNC

INPUT OVERVOLTAGE PROTECTION).

3.2.3. Mezzanine Front-end

The front-end electronics are all mounted on a removable mezzanine card. In the event of accidental damage or as components fatigue over time (e.g. relays in high duty cycle automated testing applications), the mezzanine card allows for fast and efficient replacement.

User Manual: Agilent Acqiris Averagers Page 17 of 44

3.2.4. Bandwidth and Rise Time

The bandwidth specification indicates the frequency at which an input signal will be attenuated by 3 dB

(approximately 30% loss of amplitude). The bandwidth also affects the minimum rise and fall times that can be passed through the front-end electronics. A pulse with a very sharp edge will be observed to have a minimum rise time (τ min

) determined by the front-end electronics. In general a pulse with a given 10-90% rise time (τ observed with a slower value given by:

10-90real

) will be

τ

10-90

2

=

τ

10-90real

2

+

τ

min

2 where

τ

min

(ns) ≈ 0.35 (GHz-ns) / BW (GHz)

If desired, hardware bandwidth limiters can be selected.

3.2.5. Input Voltage and Offset

The input channel provides a fully programmable amplifier with variable input voltage and offset. Full Scale (FS) input voltages are selectable from 50 mV to the maximum indicated in the table above in a 1, 2, 5 sequence. Care should be taken to select an input voltage range that will allow the signal to be recorded using as much dynamic range of the digitizer as possible. Signals going outside of the FS Range will be clipped and data values for the clipped portion of a signal should be regarded as erroneous. The Variable Offset is programmable in the range of

±2 V when using an FS Input Voltage setting of 500 mV or below, increasing to ± 5 V for FS settings above

500 mV.

3.2.6. Vertical Resolution

The AP Series Averagers use an ADC system with 8 bits of vertical resolution (256 levels). The dynamic range of the ADC covers the Full Scale (FS) of the Input Voltage setting. For example, if the Input Voltage is set to 500 mV then the ADC resolution is equivalent to 1.95 mV. To obtain the best dynamic range from the ADC care should be taken to ensure that the input signal varies over more than 50% of the Input Voltage Full Scale (FS) setting. The highest and lowest levels of the ADC correspond to underflow and overflow conditions. For the Averaging mode, any sums that contain such data are subject to unknown errors.

For the Averaging mode, vertical resolution increases as the number of acquisitions per average increases. It can be improved to up to 12 bits when averaging 256 or more traces.

3.2.7. DC Accuracy

The AP Series Averagers use low noise front-end electronics in order to ensure voltage measurement is made with accuracy and precision. DC voltage accuracy is better than

± 2% (± 1% typical) of the input voltage full-scale.

3.2.8. Linearity and Dithering

In the Digitizer mode the differential linearity is better than ±0.9 LSB for the AP240 and ±0.7 LSB for

AP100/AP200. In the Averaging mode, an offset dither of up to

± 15 ADC codes can be added to the input signal.

This dithering improves the Averager’s differential and integral non-linearity and can reduce quantization nonlinearity errors for low noise signals

The dithering range is selectable and has a resolution of a waveform. The offset sequence is

1

/

8 of the ADC Least Significant Bit (LSB). The dithering is added as a controlled offset at the input of the ADC. The offset amount is different for each successive acquisition of

0, -

1

/

8

,

1

/

8

, -

2

/

8

,

2

/

8

, -

3

/

8

,

3

/

8

, . . . - (N-

1

/

8

), (N-

1

/

8

), -N, N, 0, -

1

/

8

,

1

/

8

etc

N is programmable between 0 (no dithering) and 15. There are 16*N+1 states in the complete cycle.

Dithering over

± 15 ADC codes is expected to improve the Differential Non-Linearity (DNL) down to 1

/

16 of one

ADC code and the Integral Non-Linearity (INL) by a factor of two (~ ½ ADC code). However, underflow/overflow conditions must be avoided.

3.3. Trigger

3.3.1. Trigger Source

The trigger source can be a signal applied to either of the Input Channels (for internal triggering) or the External

Trigger Input.

User Manual: Agilent Acqiris Averagers Page 18 of 44

The modules provide a front panel BNC External Trigger Input. The External Input provides a fully functional trigger circuit with selectable coupling, level and slope. The AP240 and AP235 have 50

Ω termination impedance with diode protection against overload. A

±5 V limit on trigger signals should be respected, although somewhat higher voltages for short time periods will not damage the unit.

The AP100 and AP200 have both 1 M

Ω and 50 Ω termination impedance. Overload protection will automatically switch the coupling from 50

Ω to 1 MΩ if the signal is greater than ±5 V DC. If 50 Ω termination is selected a ±5

V limit on trigger signals should be respected, although somewhat higher voltages for short time periods will not damage the unit. For 1 M

Ω input, signals up to ±100 V (DC + peak AC < 10 KHz) are allowed.

3.3.2. Trigger Coupling

Trigger coupling is used to select the coupling mode applied to the input of the trigger circuitry. The AC LF Reject mode couples signals capacitively and removes the input signal's DC component and frequencies below 50 Hz for the AP240 and AP235 models (and 50 KHz for the AP100 and AP200). DC mode allows all signal components to be passed through to the trigger circuit. The AP240 and AP235 models have an HF Reject mode that removes signal components above 50 KHz. They also implement an HF trigger that allows triggers to be reliably accepted at rates above

∼ 1 GHz. In this mode, triggers occur on every fourth positive edge. In the HF mode negative slope and window triggers are not available.

3.3.3. Trigger Level

The trigger level specifies the voltage at which the selected trigger source will produce a valid trigger. The trigger level is defined as a set voltage. Using the internal trigger with DC coupling, the level (expressed in %) is set with respect to the midpoint voltage (V m

= – Offset voltage) of the digitizer’s vertical scale. All trigger circuits have sensitivity levels that must be exceeded in order for reliable triggering to occur.

The AP240 and AP235 offer level control for all trigger coupling modes. Internal trigger level settings must be within V m

± 0.5 FS, where FS is the channel Full Scale. In addition, they implement a Window trigger. Two trigger level thresholds are used to define the desired range. The trigger can then be chosen to occur either when the signal exits or enters the window range. This mode can be thought of as the appropriate OR of two edge triggers of opposite slope.

For the AP100 and AP200 the AC coupled mode is implemented with an auto-level trigger. Internal trigger level settings for DC coupling must be within V m

± 0.6 FS, where FS is the channel Full Scale.

The AP240 and AP235 models allow the user to choose the external trigger Full Scale from the set of values 0.5, 1.0,

2.0 or 5.0 V. The external trigger level can then be set to values in the range ± 0.5 FS. The AP100 and AP200 models have an external trigger range of ± 3 V.

The AP series Averagers will trigger on signals with a peak-peak amplitude > 15% FS from DC to their bandwidth limit.

3.3.4. Trigger Slope

The trigger slope defines the direction of the signal that will be used to initiate the acquisition when it passes through the specified trigger level. Positive slope indicates that the signal is transitioning from a lower voltage to a higher voltage. Negative slope indicates the signal is transitioning from a higher voltage to a lower voltage.

3.3.5. External Trigger Output

When the module is ready to be triggered and a valid trigger signal occurs, a trigger output is generated for external use. It is always available on the Front Panel Trigger Out MMCX connector.

NOTE: The External Trigger Output functionality is implemented in the hardware. No Trigger Out signal occurs for

software-generated triggers such as those of the AUTO mode of AcqirisLive or through the use of the function

AcqrsD1_forceTrigger.

Normally, the trigger output signal occurs when the module is ready to be triggered and a valid trigger signal occurs.

As an aid in applications where a complete system has to be warmed-up before use, the averager can be put into a mode (TrigAlways) in which triggers received when the Averager is not in use are passed through and used to generate a trigger output. This mode is controlled using the function AcqrsD1_configAvgConfig.

User Manual: Agilent Acqiris Averagers Page 19 of 44

Trigger Output Block diagram:

Signal: 1.6V swing

0v centered

G=1

50 Ohm

Trigger Out

Offset: +/-2.5 V

The output swing is 1.6 V (

± 0.8 V) when unloaded and 0.8 V when terminated on 50 Ω. The rise and fall times are

2.5 ns typical. The offset can be adjusted, by software control in the range [–2.5 V, +2.5 V] unloaded, or [-1.25 V,

+1.25 V] into 50

Ω. The maximum output current capability is ± 15 mA. As the output is retro-terminated, it is possible to drive a 50

Ω line unterminated (HiZ) without loss of performance.

For a TTL compatible signal, set the offset to 1.0 V and the swing at destination will be +0.2 to +1.8 V.

For an ECL compatible signal, terminated on 50

Ω to –1.2

V, set the offset to –1.2 V and the output will be in the range

[–0.8 V, –1.6 V]).

GND

R

68

ECL

Signal

+5V

R

220

TTL signal

Alternatively, to reduce the current drawn from the digitizer, the terminations here can be used:

Trigger

Out

With

Offset

-1.18V

R

220

-5V

Trigger Out

With

Offset

+1.18V

R

68

GND

3.3.6. Trigger Status

The front panel includes a tri-color LED indicator to show the status of the trigger. When the LED is green it indicates the trigger is armed and waiting for a valid trigger to occur. Red indicates that the trigger has occurred, the acquisition is complete and the data is waiting to be readout. The user can override the default functions and program the LED color in an application-specific manner.

3.4. Data Acquisition – Common

The AP series Averagers can be used either for averaging or for normal digitizer data acquisition. These two modes are quite different in their functioning. After describing some common characteristics they will be treated individually in the sections below. The digitizer mode is particularly useful when trying to configure an AP module for later use as an Averager.

3.4.1. Timing

A crystal controlled time base is used to generate the clock of the digitizer. Clock accuracy is better than

± 2 ppm, and individual sample points are recorded with better than

± 10 ps (< 2 ps rms) of sampling jitter.

3.4.2. Sampling Rate

The AP series Averagers include an analog-to-digital converter (ADC) that can sample waveforms in real time, at rates from 1 or 2 GS/s (0.5 ns per point) down to 100 S/s (10 ms per point). The sampling rate can be programmed and is selectable in a 1, 2, 2.5, 4, 5 sequence (i.e. 100 MS/s, 200 MS/s, 250 MS/s, 400 MS/s, 500 MS/s, 1GS/s,

2GS/s). The data of all of the active channels is acquired synchronously; all of the ADC’s are acquiring data at the same time, to within a small fraction of the maximum sampling rate.

3.5. Data Acquisition - Digitizer Mode

3.5.1. Acquisition Memory and Time Base Range - Digitizer Mode

Data from the ADC is stored in on-board acquisition memory. The amount of memory in use for acquisition can be programmed and is selectable from 2 samples to 256KS, the full amount of acquisition memory available (with

AP235/AP240 in single-channel mode). Note that the Averaging memory is much greater. The raw ADC data is treated as a value in the range [-128, +127], which corresponds to the voltage range [–FS/2 – Offset, +FS/2 – Offset].

For technical reasons, a certain memory “overhead” is required for each waveform, reducing the available memory by a small amount. In order to simplify programming, an interface function recommends the best sampling rate and

User Manual: Agilent Acqiris Averagers Page 20 of 44

the maximum possible number of data points, taking into account the available memory, the requested time window, the number of segments (in Sequence mode), as well as the required memory overhead.

The Time Base Range defines the time period over which data is being acquired. For example, the AP200 in the digitizer mode has an acquisition memory of just under 256 KS and maximum sampling rate of 2 GS/s. Thus, at the maximum sampling rate, the digitizer can record a signal over a time period of up to 130

μs (256 KS * 0.5 ns/point).

The time base range can be adjusted by varying the amount of acquisition memory or the sampling rate of the digitizer.

3.5.2. Pre- and Post-Trigger Delay - Digitizer Mode

To increase trigger flexibility a pre- or post-trigger delay can be applied to the trigger position. However, pre-trigger is not available in the Averaging mode.

The amount of pre-trigger delay can be adjusted between 0 and 100% of the acquisition time window (i.e. sampling interval x number of samples), whereas the post-trigger delay can be adjusted between 0 and 200 million samples.

Pre- or post-trigger delays are just different aspects of the same trigger positioning parameter:

• The condition of 100% pre-trigger indicates that all data points are acquired prior to the trigger, i.e. the trigger point is at the end of the acquired waveform.

• The condition of 0% pre-trigger (which is identical to a post-trigger of 0) indicates that all data points are acquired immediately after the trigger, i.e. the trigger point is at the beginning of the acquired waveform.

• The condition of a non-zero post-trigger delay indicates that the data points are acquired after the trigger occurs, at a time that corresponds to the post-trigger delay, i.e. the trigger point is before the acquired waveform.

The digitizer hardware accepts pre- and post-trigger adjustments in increments of 16 samples. By definition posttrigger settings are a positive number and pre-trigger settings are a negative number.

Thus it is only natural that the software drivers provided treat pre- and post-trigger delays as a single parameter in seconds that can vary between –nbrSamples * samplingInterval (100% pre-trigger) and +maxPostTrigSamples *

samplingInterval (max post-trigger). Since the Acqiris software drivers provide very accurate trigger position information upon waveform readout, the accepted resolution of the user-requested pre-/post-trigger delay is much better than 16 samples. For more details, refer to the Programmer’s Reference Manual.

3.5.3. Single and Sequence Acquisitions - Digitizer Mode

Digitizers acquire waveforms in association with triggers. Each waveform is made of a series of measured voltage values (sample points) that are made by the ADC at a uniform clock rate. To maximize sampling rates and utilize memory as efficiently as possible, the digitizers include both Single and Sequential storage modes.

The Single Acquisition mode is the normal operation of most digitizer products. In this mode an acquisition consists of a waveform recorded with a single trigger. The user selects the sampling rate and acquisition memory size and sets the number of segments to 1 (default value).

The modules also include a built-in Trigger Time Interpolator (TTI) that measures the time from the trigger point to the first sample point. This information is essential for determining the precise relation between the trigger or other event of interest and the digitized samples of the signal. The TTI resolution is 80 ps.

The Sequence Acquisition mode allows the capture and storage of consecutive “single” waveforms. Sequence

Acquisition mode is useful as it can optimize the digitizer’s sampling rate and memory requirements for applications where only portions of the signal being analyzed are important. The mode is extremely useful in almost all impulseresponse type applications (RADAR, SONAR, LIDAR, Time-of-Flight, Ultrasonics, Medical and Biomedical

Research, etc.).

In Sequence Acquisition mode the acquisition memory is divided into a pre-selected number of segments.

Waveforms are stored in successive memory segments as they arrive. Each waveform requires its own individual trigger. The memory can be divided into any number of segments between 2 and 200. In Sequence Acquisition mode the user needs to specify the sampling rate, the total acquisition memory and number of segments. Note that the

Single Acquisition mode is just a special case of the Sequence Acquisition mode with the number of segments set to

1.

Sequence acquisition enables successive events, which can occur within a very short time, to be captured and stored without loss. A crucial feature of Sequence Acquisition mode is that it has a very fast trigger rearm time. A fast trigger rearm helps produce very low “dead time” (less than 800 ns for the highest available sampling rates) between the segments of a sequence acquisition. The “dead time” is the period after the end of an event when the card cannot digitize data for a new trigger event. To complement this mode of operation the digitizer can also measure and store the arrival time of each trigger using the information from the on board TTI (Trigger Time Interpolator). Readout of

User Manual: Agilent Acqiris Averagers Page 21 of 44

the individual trigger time stamps makes it possible to determine the time from one trigger to any other trigger in the sequential acquisition. The TTI resolution sets the resolution of the trigger time stamps.

3.6. Data Acquisition - Averaging Mode

Model -

Mode

AP240 –

Single-channel

Channels Maximum Maximum

Sampling length for

Average

Rate

Default

Memory

Maximum Memory length for

Average

Minimum/

Increment

Optional

Memory

samples

8M samples 32

AP240 –

Dual-channel

AP235 –

Single-channel

AP235 –

Dual-channel

AP200

AP100

Start

Delay

Maximum

16M

Samples

GS/s 8M 32 16M

2 0.5 16 16M

3.6.1. Averaging Mode

In Averaging mode, newly acquired waveforms are summed in real-time with the corresponding samples of previous waveforms. The corresponding samples are those with identical delay calculated from the beginning (trigger time) of the acquisition. While the acquisition is running, the previous sum is read from the averaging memory, added to the incoming sample and written back to the averaging memory. A high-speed on-board computing circuit performs the retrieve, add and store functions necessary to average the signal in real time. Real time averaging is possible up to the maximum 2 GS/s sampling rate. In Averaging mode, the trigger starts the sampling clock, and only post-trigger data are acquired. The sums are stored in the 24-bit deep averaging memory, allowing the summation of up to 65,536 waveforms, without overflow. Each raw ADC datum is treated as a value in the range [0, 255], which corresponds to the voltage range [–FS/2 – Offset, +FS/2 – Offset].

3.6.2. Time Base Range

The Time Base Range defines the time period over which data is being acquired. For example, the AP240 in singlechannel mode has an averaging acquisition memory of 2 MS and maximum sampling rate of 2 GS/s. Thus, at the maximum sampling rate, the digitizer can record a signal over a time period of up to 1.0 ms (2 MS * 0.5 ns/point).

With the optional 8 MS averaging memory, this range can be extended to 4 ms. The time base range can be adjusted by varying the amount of acquisition memory or the sampling rate of the digitizer.

3.6.3. Averaging Memory

Data from the ADC is stored in the on-board averaging memory. The actual averaging memory size used for each acquisition can be programmed. The maximum values are given in the table above.

Segment time stamps and descriptors are also stored in the averaging memory.

To ensure the use of maximum sampling rate and high timing resolution, we strongly recommend the use of long averaging memories whenever possible. For example, with the optional 8 MS memory, the model AP240 can record a signal over a 4 ms period at full sampling rate. The fast sampling rate ensures that all high frequency signal components, up to the full bandwidth of the Averager, are accurately recorded. If the averaging memory were reduced to just 20 KS then the sampling rate would need to be reduced to just 5 MS/s (20,000/4 ms) to record the same 4 ms period. All frequencies above 2.5 MHz would then be incorrectly digitized and important signal characteristics may be distorted or even completely missed.

User Manual: Agilent Acqiris Averagers Page 22 of 44

3.6.4. Single Average

When a single average is acquired, the series of events shown in the diagram below takes place:

• The trigger starts the acquisition (Clock pre-trigger is not possible in Average mode).

• After each waveform acquisition, the clock will stop and is restarted again only by the next trigger.

• When the whole average is complete, the module will stop acquiring.

Single average:

The example shows an Average of depth 3

Software Start Acquisition (out=21)

Trigger is armed (ACQRDY, out=22)

Average (AP235/AP240 P1/P2 Out)

Trigger Input

Trigger Output

Acquisition Status

PCI interrupt

Software

Acquisition

Enable

Acquire Sum Sum

Readout

&

Processing

End of Sequence acquisition disabled

3.6.5. Sequence Average Mode

For time resolved applications where multiple averaged waveforms must be acquired with minimal dead time between them, the Sequence Average mode is available. This mode partitions the averaging memory into a pre-selected number of segments. Each memory segment is then used to store accumulations of summed data. The number of segments can be chosen from 1 to 8191. The length of each segment is adjustable from the Memory minimum value in the table above, in steps of that value, up to the maximum consistent with the memory size and the number of segments.

As with a Single Average, Sequence Average sums are stored in the 24-bit deep averaging memory; this allows the summation of up to 65,536 8-bit waveforms. Post acquisition processing of sequence mode acquisitions in the host computer can further enhance the sums. E.g. in order to obtain 32-bit sums, up to 256 sequence-averaged waveforms may be added together. The maximum number of acquisitions for an average can thus be increased to over 16 million waveforms.

It should be noted that the normal Sequence Average mode is not at all like the Sequence Acquisition Mode of digitizers as described earlier. The data associated with successive triggers is summed in a segment until the desired depth, N succ

, is attained. The next trigger initiates the sum of the following segment.

The AP240 and AP235 Averagers also implement a round-robin Sequence Average mode. In this case, successive triggers go into the same segment until the desired count, N succ

is reached. The next trigger goes into the next segment and the entire loop over segments is repeated until each segment has been worked on N robin

times. This implementation will allow >65536 triggers to be accumulated into a 24-bit word. Therefore, there is a possibility for the accumulated sum to wrap around 0; For the particular case, Segments = 2, N succ

=1 and N robin

= N the acquisition will contain the accumulated data for the N odd triggers in segment 1 and for the N even triggers in segment 2.

To complement the Sequence Average mode, the digitizer can also measure and store the times when the first trigger of each segment occurred. The timestamp records the count of the number of cycles of the 33 MHz system clock.

Readout of the time stamps allows the determination of the time from the beginning of one Average to that of any

User Manual: Agilent Acqiris Averagers Page 23 of 44

other Average in the sequential acquisition. Timing measurements between trigger events is possible with a resolution of 30 ns.

The averager can also store the state of 4 bits at the time of the first trigger of each segment of each acquisition.

These bits, which will be called hardware markers, are those of logic signals at the P1, P2, MMCX I/O A, and

MMCX I/O B inputs. See section 3.8 Front Panel Inputs and Controls.

In the Sequence Average mode, the module needs as many triggers as defined by the sequence setting, (N succ

* number of segments.

* N robin

)

3.6.6. Trigger Overview

Before the Averager starts acquiring signals it must be armed and in a ready state. Arming the module is done by software (ACQUIRE command). The Averager will then wait for a trigger to start an acquisition (see the timing diagram below).

When the Signal Averager is armed (via the AcqrsD1_acquire command) and ready to acquire data, it generates, after a time Tst (which may vary between 50

μs and 10 ms, depending on the processor and the operating system used) a Ready signal (ACQRDY). This signal is available on the front panel I/O MMCX connector. The Ready signal is normally used to activate an external event, such as firing a laser.

This external event in turn can produce a signal that can be used to generate the trigger for the Signal Averager. The trigger-input pulse can have a delay Ttid that can vary from zero to as long as required. The Ready signal

(ACQRDY) is reset with the arrival of the trigger pulse. When the acquisition is running, or when the module has not

Software ACQUIRE

Tst

ACQRDY

Trigger-Input

Ttid

Ttoj

Trigger-Output

Ttod Ttow

StartDelay NbrSamples StopDelay

Data Acquisition status

Acquire

1st data

Acquire last data

1st TRIGGER been enabled to start the acquisition, ACQRDY is set low.

When the trigger is received, the Averager generates a trigger output which is then used internally to start the acquisition after a predetermined delay shown as StartDelay in the diagram above. The trigger output is provided for applications where an external signal source must be synchronized with the internal sampling clock. The Averager trigger output edge is synchronized to the sample clock to within

< 100 ps. The trigger output accurately preserves the sample timing with respect to the event of interest from shot-to-shot. Ideally, the trigger output is used to synchronously activate an event serving as the zero-time reference for the time-dependent phenomena that are measured in each accumulated waveform. As an example, the trigger output could activate the application of an electric or magnetic field to a sample under test. This method ensures that the application of the field is identically synchronized to the sample clock for each shot contained within the average. It also means that each sample in the accumulated sum is identically positioned shot-to-shot with respect to a fixed zero-time reference, provided that the time between the trigger output and the event waveform is kept constant.

The trigger output has, with respect to the trigger input, a delay of around Ttod

± Ttoj/2, where Ttod is about 10 ns and Ttoj is either 400 ps or 1 clock pulse, depending on the mode of operation as explained in the next section. The digitizing clock is started at a precise, user configurable time after the trigger output pulse. This is shown as

StartDelay in the diagram above. StartDelay is adjustable from its minimum as given in the table below, in steps of the time of the memory increment value, up to the maximum value given in the table.

User Manual: Agilent Acqiris Averagers Page 24 of 44

StartDelay

/Sampling

Rate

2GS/s

1

0.5

0.4

AP240

Singlechannel

18.8 ns

22.6

38.2

AP235

Singlechannel

NA

22.6

38.2

AP240

Dualchannel

NA

18.8 ns

22.6

25.8

AP235

Dualchannel

NA

NA

22.6

25.8

AP200 AP100

18.8 ns

22.6

38.2

NA

18.8 ns

22.6

25.8

Zero points

10 10 5 5 10 5

These results apply to current versions of the FPGA firmware. It should also be noted that when the user configurable StartDelay is 0, the first few data points (as shown in the table above) of 32-bit summed waveform data are automatically set to 0 (bottom of the range) and should be ignored.

The Averager can be used to limit the system’s overall event rate. A rate adjustment delay StopDelay can be programmed in steps from zero to the model maximum with the same range and resolution as StartDelay. For maximum acquisition rates StopDelay should be set to zero.

3.6.7. Random and Synchronized Triggers

The Averagers have two trigger synchronization modes, Non-Resync or Resync. Normally the user’s trigger input is not synchronized with respect to the internal sampling clock.

In Non-Resync mode, the internal trigger signal, which generates the trigger output and starts the digitizing clock after StartDelay, is not resynchronized with respect to the internal sampling clock. The trigger output is synchronized with respect to the sampling clock to better than

± 100 ps. The jitter, i.e. the variation of the position of the clock with respect to the input trigger time, as a function of sampling rate is shown in the table below. In order to preserve the timing accuracy of successively summed acquisitions, the Averagers also offer a Resync mode, where the jitter is limited to

± 200 ps (at the highest sampling rate).

The Resync mode is the preferred method for applications where the acquisition jitter (and the trigger out jitter) must be minimized to avoid loss of effective system bandwidth typically associated with the higher jitter found when using the Non-Resync mode.

Sampling Clock Synchronization with respect to Trigger Input

Sample Rate Resynchronized Mode Non-Resynchronized Mode

AP240/AP235

Single-channel

AP100 or

AP240/AP235

Dual-channel

AP200 or

AP240/AP235

Single-channel

AP100 or

AP240/AP235

Dual-channel

2 GS/s

± 200 ps --

± 500 ps --

1 GS/s

± 325 ps ± 200 ps ± 1 ns ± 500 ps

500 MS/s

± 325 ps ± 325 ps ± 1 ns ± 1 ns

400 MS/s

± 400 ps ± 400 ps ± 1.25 ns ± 1.25 ns

≤250 MS/s

± 400 ps ± 400 ps ± 1.25 ns ± 1.25 ns

3.6.8. Noise Suppressed Accumulation

In some applications, such as time of flight mass spectroscopy, the signal is a rare event sitting on top of a noisy baseline. The averaging process will reduce the random noise. However, to enhance the Averager’s ability to detect such signals in the presence of synchronous noise, the Averagers allow the user to set a threshold that has to be exceeded for each data value to be entered into the sum. The threshold should be set above the level of the noise

(synchronous and asynchronous). Furthermore, as an aid to the user or in order to avoid overflow in the summed data, a fixed number, the NoiseBase, can be subtracted from each data value before the summation is done. Both of these values are specified as absolute voltages, V thresh

and V base

, which are converted into their ADC code equivalent values, base and threshold. This is shown in the following picture:

User Manual: Agilent Acqiris Averagers Page 25 of 44

Sample above the Threshold

Thus, for each trigger and in each time bin, i, if the data value exceeds the threshold then the threshold minus the base value will be added to the sum, s

i

:

s i

=

data

(

data

>

threshold i

base

)

and

<

V i

>=

1 /

N

V i

(

>

V thresh

V i

V base

)

where N is the total number of acquisitions and where base = (V base

V lower edge of range

)/FS*256 and

threshold = (V thresh

- V lower edge of range

)/FS*256

Note that V base

can be below V lower edge of range

if so desired.

Similarly, if the signal of interest is negative going and the data of interest lie below a threshold, data inversion can be enabled and the application of the same recipe gives the same result for the following input:

Sample below the Threshold

s i

=

data

<

( 255

255

threshold

data i

base

)

and

<

V i

>=

1 /

N

V i

(

<

V thresh

V base

V i

)

where base = (V upper edge of range

V base

)/FS*256 and

threshold = (V upper edge of range

V thresh

)/FS*256

Note that V base

can be above V upper edge of range

if so desired.

The result of this is that the summed or averaged data always shows positive peaks. In order to ensure that the threshold can be applied and that the values contributing to the sum are always positive and because the base value is stored as a signed 8-bit number it is required that:

For positive going signals (data above a threshold)

0 < threshold < 255 or, equivalently, –FS/2 – Offset < V thresh

< +FS/2 – Offset

threshold ≥ base or, equivalently, V thresh

V base

and

-128 ≤ base ≤ 127 or, equivalently, –FS – Offset < V base

< – Offset

For negative going signals (data below a threshold)

0 < threshold < 255 or, equivalently, –FS/2 – Offset < V thresh

< +FS/2 – Offset

threshold ≥ base or, equivalently, V thresh

V base

and

User Manual: Agilent Acqiris Averagers Page 26 of 44

-128 ≤ base ≤ 127 or, equivalently, – Offset < V base

< +FS – Offset

The average value is obtained by dividing the sum by the total number of triggers which is often much greater than the number of triggers that actually contributed to the sum. The interpretation of the absolute voltage value is not obvious.

3.6.9. TDC Histogram mode

This mode implements the accumulation of a histogram of peak position values based on the NSA analysis introduced above. Defining a peak as a number of consecutive data values above the threshold, a peak will be accepted if its width (the number of values) is greater than the user-defined minimum in the range [0,3]. The peak's time t i

is defined as that of the datum in the middle or just to the left of the middle of the peak. The peak's value is chosen to be either 1 or the value V i

. The peak will be rejected if its width is greater than 16 values.

The TDC histogram will have the same length as that of the associated waveform with segments. The accumulated value for a single bin is stored in a 24-bit sum. The acquisition will run for the number of triggers defined by the number of segments, the round-robin count, and the number of waveforms to be entered into a segment accumulation before incrementing the segment. Each waveform can contain an arbitrary number of peaks satisfying the definition above.

The process of accumulating a TDC histogram is very similar to the process of accumulating the sum used in

ADC Window

averaging. The difference is that instead of adding the current waveform to the sum we convert that waveform into one in which only the acceptable peaks have a single non-zero point and we sum that derived waveform into the accumulated sum.

3.6.10. Real-time Add/Subtract control

Some applications may have the possibility to have triggers with and without the usual background. The background can include both contributions from the Acqiris module as well as those from the experimental apparatus. If the user provides a digital signal to indicate whether the data associated to the current trigger should be added to or subtracted from the accumulated sum it becomes possible to cancel the Background contributions to the sum. We know of two such examples;

1. Some of the triggers provide an analog input V

1

= + Signal + Background while the other triggers provide V

– Signal + Background. In this case if we provide equal numbers of triggers of the two types then

2

=

<V> = 1/n * (Σ V

1

– Σ V

2

) = Signal

2. Some of the triggers provide an analog input V

1

= + Signal + Background while the other triggers provide V

+ Background. In this case if we provide equal numbers of triggers of the two types then

2

=

<V> = 1/n * (Σ V

1

– Σ V

2

) = Signal/2

Note that the technique poses no requirements on the relative magnitudes of the Signal and the Background. To successfully exploit the method the vertical range of the channel has to be chosen so as to avoid both underflow and overflow for both V

1

and V

2

. However, the final result is independent of the offset of the channel. When reading

User Manual: Agilent Acqiris Averagers Page 27 of 44

back the data the driver will not take the offset into account in the usual way. If the user reads back 32-bit sums their values can be negative. Real-time Add/Subtract should not be combined with NSA or TDC modes.

3.6.11. Pre Acquisition Triggers

For applications where thermal effects must be taken into account, the Averager module has the capability of generating trigger output pulses that can be used to activate the system even when the associated event will not be acquired. The trigger output pulses can be used to stabilize systems prior to starting the acquisition. A call to

AcqrsD1_configAvgConfig using the parameter string "TrigAlways" will cause these pulses to be generated. In this case the Control I/O Trigger armed signal is also active and the trigger rate should not exceed that which can normally be taken under the chosen acquisition conditions.

The Acquisition Ready signal can be observed on either of the I/O A or B MMCX connectors when appropriately configured. The Average Enable signal can be sent in to the module using the I/O A or B connector for the

AP100/AP200 or the P1/P2 control for the AP235/AP240. The Average Enable can be configured to act as either a low-high transition event enabling the start of the averaging process (Start Veto Enable) or as a level which must be high to allow an individual trigger to be accepted (Average Trigger Enable).

3.7. External Clock and Reference

For applications where the user wants to replace the internal clock of the digitizer, and drive the ADC with an external source, an External Clock or Reference signal input is available. The Clock or Reference signals can be entered into the digitizer via the MMCX CK connector on the front panel.

When using an External Clock, the user must ensure that the input signal has a frequency between 20 MHz and

2000 MHz for the AP235/AP240 models (or between 10 MHz and 500 MHz for the AP100/AP200 models) and a minimum amplitude of at least 1 V peak-to-peak (2 V for the AP100/AP200 models). The External Clock allows the

Pre Acquisition Triggers: Readout

&

Processing

The example shows a 2 segment Sequence Average where the segments have a depth of 2.

Software Start Acquisition

Software Trigger Output Enable

Acquisition Ready

Averaging activity

Trigger Input

Trigger Output

Average Enable

Acquisition Status

PCI interrupt

Average1

Acquire Sum

Average2

Acquire Sum

Software

Trigger Enable

TrigAlways

Software

Acquisition

Enable

Hardware

Average

Enable

End of

1st Segment

Average

End of Sequence acquisition disable digitizer to make a voltage measurement when the clock signal passes through a predefined threshold.

Normally the external clock sampling rate is the same as the external clock frequency; however for the

AP240/AP235 single-channel mode the sampling rate is ½ of the clock frequency. The threshold range is variable and user selectable between

± 2 V. The signals should not exceed ±5 V amplitude.

For applications that require greater timing precision and stability than is obtainable from the internal clock, a

10 MHz Reference signal can be used. The amplitude and threshold conditions, for an External Reference, are the same as for the External Clock. If phase synchronization between several digitizers is required, the reference signal should be applied to all of them. The allowed frequency range is [9.0 MHz, 10.2 MHz].

When using an External Clock, the Resync trigger mode is not functional.

User Manual: Agilent Acqiris Averagers Page 28 of 44

3.8. Front Panel Inputs and Controls

Name Function

AP100/AP200 AP240/AP235

Signal input A Channel 1 input INPUT

A

INPUT

B

Channel

1

Channel

2

TRIGGER IN

Signal input B

Trigger input

Channel 2 input

Connector

BNC 1

BNC 2

TRIGOUT

User configurable (see below).

Signal occurs after an accepted TRIGGER. It is synchronous to the acquisition Clock and can be used to trigger events synchronously to the acquisition clock.

User configurable

BNC 3

MMCX I/O A

MMCX TR

MMCX I/O B

MMCX CK

NA

NA

User configurable (see below)

User configurable (see below)

P1

P2

The I/O A, I/O B, P1 and P2 signals are 3.3 V compatible CMOS. This means that, on input, low is < 0.7 V and high must be in the range [1.7 V, 5.0 V]. An unconnected signal will be high. This definition ensures TTL compatibility.

On output, the low level will be in the range [0 V, 0.7 V] and the high level in the range [1.7 V, 3.3 V] for HiZ. The high level output will typically give 0.8 V into 50

Ω. The I/O A or B connectors can be used for the following output signals: 10 MHz reference clock, Acquisition skipping to next segment, Acquisition is active, Trigger is armed. They can also be used for Enable trigger input or Averager Start Veto for the AP100/AP200. The AP240/AP235 P1 and

P2 connectors can be used for Enable trigger input, Averager Start Veto, or Real-time Add/Subtract Control inputs and the Average output.

To ensure proper operation of the segment recording of hardware markers, the logical state of P1, P2, MMCX I/O A, and MMCX I/O B inputs, the input must be stable for a time interval starting 10 ns before the trigger and continuing until 20 ns after it. This implies that the HF Reject trigger coupling should not be used. If the individual inputs are not in the disable state the corresponding bits will be useless.

3.9. Internal Calibration

The software drivers supplied include calibration functions for the timing, gain and offset settings, which can be executed upon user request. The digitizers are never calibrated in an “automatic” way, i.e. as a side effect of another operation. This ensures programmers have full control of all calibrations performed through software in order to maintain proper event synchronization within automated test applications.

The model AP series Averagers include a high precision voltage source and a 16-bit DAC, used to determine the input voltage and offset calibration.

For accurate time and voltage measurements it is recommended to perform a calibration once the module has attained a stable operating temperature (usually reached with a few minutes after power on). Further calibration should not be necessary unless temperature variations occur. Calibration can usually be performed with signals present at the channel, external, and clock inputs. However, if the calibration is found to be unreliable, as shown by a calibration failure status, it may be necessary to remove such signals.

User Manual: Agilent Acqiris Averagers Page 29 of 44

3.10. Electrical, Environmental and Physical Specifications

3.10.1. Electrical

Power Consumption (W)

Model Maximum / Typical

AP240

AP235

56 / 46

56 / 46

CURRENT REQUIREMENTS (A)

+12 V

1.1

1.1

+5 V

2.7

2.7

+3.3 V

8.3

8.3

−12 V

0.04

0.04

AP100

AP200

The Maximum Power Consumption has been increased by 5% over the value calculated with the currents shown to take into account higher allowed values for the PCI voltages.

Except for the high total power consumption all modules conform to Revision 2.2 of the PCI Local Bus

Specification. They use the PCI Bus at 33 MHz. All of these modules are capable of DMA transfers at rates ~100

MB/s.

3.10.2. Environmental and Physical

The modules have a Declaration of Conformity according to ISO/IEC Guide 22 and EN45014 and CE Marks of

Compliance.

Operating Temperature

0° to 40°C

The above values are for the ambient temperature of the room (or equivalent) where the PC is located. The temperature as measured on the board may well be significantly higher. On-board temperatures above 60°C should be avoided

Relative Humidity

5 to 95% (non-condensing)

Dimensions

All AP modules conform to the PCI standard and require a long card slot. When used with the XP102 Fan the overall length can be adjusted from ~325 mm up to the long card slot maximum of 342.3 mm.

Safety

Complies with EN61010-1

EMC Immunity

Complies with EN61326-1: Industrial Environment

EMC Emissions

Complies with EN61326-1: Class A for radiated emissions

Required Airflow

XP102 Fan unit required. AP Averagers are delivered with this Fan unit.

User Manual: Agilent Acqiris Averagers Page 30 of 44

4. Running the AcqirisLive Application

AcqirisLive is an application to control and demonstrate the capabilities of Acqiris digitizers and averagers on a single machine. AcqirisMAQS is a more advanced application offering many interesting possibilities for the control of acquisition systems in a single or multi-machine environment. Ask your sales representative or Acqiris for more information.

4.1. Getting Started with AcqirisLive

Once the software and hardware installation described in Section 2 of this manual is complete, you will find

AcqirisLive in the start menu of your computer. Start AcqirisLive from the menu. An initial AcqirisLive load screen will appear followed by a panel of controls and a waveform display window.

During startup, AcqirisLive searches for all Acqiris Averagers and Digitizers on the PCI bus. If no modules are found on the PCI bus, AcqirisLive will display an error message indicating that none were found, and the program will continue in simulation mode. If this is not wanted, turn off the computer, install and turn ON the hardware on the

PCI bus and, then, restart the computer.

AcqirisLive is intended to verify the functionality of the card and to serve as an easy to use and convenient program for capturing, displaying, and archiving captured data into the format of your choice, binary or ASCII. AcqirisLive is provided free with the purchase of any Acqiris Digitizer, Averager, or Analyzer.

We supply drivers on the CDROM for National Instruments LabWindows/CVI, LabVIEW, Microsoft Visual C++, and Visual Basic. We also provide some examples (GetStarted applications) as a guide or starting point for programmers developing their own applications to interface with the DC or DP series digitizer cards.

When you start AcqirisLive a display like that shown below appears.

AcqirisLive displays two independent windows, a control window and a display window. The control window contains functions that allow you to manipulate the acquisition parameters of the card. In order to display an acquired waveform, connect a signal to the input, and then select the Auto acquisition mode at the bottom of the control panel.

A waveform should appear in the display.

Other controls in the control panel allow you to select the optimal acquisition settings for your particular signal. Each of the other control panel functions will be discussed in detail in the following Control Panel Functions section. Here is a sample result.

User Manual: Agilent Acqiris Averagers Page 31 of 44

4.2. Control Panel and Functions

4.2.1. Choice of Instrument Class

Averager cards can be used either as such or as Digitizers. To use an AP240/AP235/AP200/AP100 as an averager, you must select the Averagers class from the Instrument menu. This will slightly modify the appearance of the

Instrument Control window and enable the Averager Settings and Logic Device Loading commands from the

Options menu. Please note that the Select Class command is only enabled when the Acquisition is stopped. The transition will take a moment and when it is over only Averagers will be shown in the Instrument Control and an

AVG

icon will appear in the Waveform Display window.

DSO TR

4.2.2. Control Panel Mode

In AcqirisLive there are two control panel modes available, Oscilloscope Mode and Transient

Recorder Mode. The choice of mode is entirely determined by the preference of the user.

Both modes provide the same set of functional capabilities but display the settings slightly differently.

For example in Transient Recorder Mode, the user selects the Full Scale Range of the amplifier whereas in

Oscilloscope Mode the user selects the input voltage scale set in Volts/Div. The desired mode is selected using one of the two buttons shown at the top of the control panel. The distinction between the control panels for the two operating modes can be seen in the images below.

In both modes it is possible to set the memory depth and the sample speed independently within the limits of the maximum available memory. However the two modes differ in the following way.

In Oscilloscope Mode, the user selects the time window (time per div) and the sampling rate will be adjusted to the fastest possible within the limits of the available memory. The maximum available memory is programmable.

In Transient Recorder Mode, the user selects the capture time window, the sampling rate, or the number of samples.

When selecting the capture time window, the number of samples is adapted, keeping the current sampling rate and number of segments, up to the maximum available memory on the card. When selecting the sampling rate, the number of samples is adapted, keeping the current capture time window and number of segments, up to the maximum available memory on the card. When selecting the number of samples, the capture time window is adapted, keeping the current sampling rate and number of segments.

User Manual: Agilent Acqiris Averagers Page 32 of 44

Digitizer-Oscilloscope Mode Digitizer-TR Mode Averager-TR Mode

4.2.3. Numeric Editor

The numeric editor is shown in the control window at locations where numeric values are set.

The Acqiris Numeric Editor enables easy input of numeric values, particularly those ranging over many decades. The numeric value can be changed in multiple ways.

A number can be typed directly into the white number display area or alternatively, the Acqiris Numeric Editor may be used. In order to use the Acqiris Numeric Editor, click on the green numeric editor icon at the left of the number display. A numeric editor control will appear as shown.

The methods to change the values are:

1. Increment or decrement any digit with the arrow buttons by pressing the increment or decrement button. When the button is held down, the digit will continue to increment or decrement automatically.

2. Click on a digit and drag the mouse up or down to increase or decrease the value.

3. Type a new value with the keyboard for any highlighted digit.

The value is constrained to fall within its allowed range whenever the user attempts to enter a larger or smaller value.

Digits that are below the resolution or granularity cannot be modified. A message is displayed in the numeric setting control window when the pointer is positioned over a digit that is below the allowed granularity and cannot be changed.

4.2.4. Keyboard shortcuts

Two shortcuts have been defined to simplify applying changes of a parameter to a group of channels. When the mouse is over the control of a channel parameter -

ƒ 's' can be used to set the current value to all channels of all available instruments

ƒ 'c' can be used to set the current value to all channels of the current instrument.

Other shortcuts are:

ƒ 'Ctrl-S' as a synonym for 'Save Setup'

User Manual: Agilent Acqiris Averagers Page 33 of 44

ƒ 'Ctrl-R' as a synonym for 'Recall Setup'

ƒ 'Ctrl-C' as a synonym for Calibrate all (active) instruments'

ƒ 'Ctrl-E' functions as a toggle between 'Internal Clock' and 'External Clock' operation.

4.2.5. Input Voltage Scale, Offset and Coupling

Settings for input voltage scale, offset, coupling and impedance are available in the portion of the control panel shown to the right. The selected channel is shown on the upper section of the display. Only this channel will be affected by changes in any settings on the control panel. The input voltage scale is set by one of three possible methods:

1. Clicking on the white numeric display of the voltage scale will show a pop-up list. Select the desired gain setting with the mouse pointer.

2. Clicking on the Decrement / Increment icons will step the voltage scale up or down to the next level for each click of the mouse.

3. Clicking Max / Min will toggle to the least sensitive gain setting or to the most sensitive gain setting.

The input Coupling can be set for AC, DC, or Ground. The selected setting is shown as an actuated indicator with a green backlight. AC coupling blocks DC signal components from the input. The AC coupling mode acts as a high pass filter passing frequencies above 10 Hz. In DC coupling all frequency components are passed through the input.

The input voltage offset control applies an offset to the input amplifier in order to better digitize and display signals offset away from 0 volts. Signals riding on a DC level can be most accurately characterized when the offset is adjusted so that the acquisitions are made about the midpoint of the digitizer’s Full Scale Range, and when as much of the 8-bit ADC range is used as possible. The ability to select an offset voltage enables accurate measurement when offsets are present in the signal.

The Adv button in this section gives access to the choice of hardware bandwidth limits for the channel.

4.2.6. Trigger

Trigger characteristics are set using the settings in the control panel

Trigger section. The trigger source setting can be either Internal

(trigger on any one of the input channels) or External (trigger on an external input). The trigger source is indicated in the display by the name appearing under Source.

In order to select the trigger source, click the drop-down source menu in the trigger Source section of the control panel and choose the desired one.

Notice that, for most digitizer models, when external trigger is selected, the controls for the trigger termination, 50

Ω or 1 MΩ, are enabled in the Coupling section. Select the desired termination for the external trigger.

The Coupling section also contains a selection for AC or DC coupling. AC coupling on the trigger sets the trigger circuit to an AC Low Frequency Reject mode.

Use the Level section to set the desired trigger level. The relationship between coupling, trigger source, and trigger

level was discussed in section 3.3.3 Trigger Level.

(Digitizers) The trigger Delay setting positions the acquisition with respect to the trigger. Pre-trigger values are negative and are selectable in the range of [0 to – (10 * timebase)]. The value (10 * timebase) is equivalent to the time window or capture-time. 100% pre-trigger is achieved by setting the Delay to -capture-time. A 100% pre-trigger delay means the trigger point is at the right edge of the display and all of the data in the acquired waveform were captured just prior to the trigger event. Maximum Post trigger delay settings depend on the sample interval used. Post trigger delays are adjustable up to 200 MS.

(Averagers) When the Averagers Instrument Class is selected, the digitzer oriented Delay control of the Trigger section is replaced by a more appropriate control. This allows the independent control of the Start and Stop delays expressed in samples. The Trigger time arrow is no longer displayed. The choice of memory samples will be

constrained by the requirement of it being a multiple of 16 or 32 samples as discussed in Section 3.6 Data

User Manual: Agilent Acqiris Averagers Page 34 of 44

Acquisition - Averaging Mode. The Averager Settings window allows you to select the number of acquisitions

needed to complete an average. It also lets you choose the type of trigger synchronization and the dithering to be applied to the offset.

When a Slope box is shown next to the delay section, selecting the trigger slope is achieved by depressing either the positive or negative slope buttons. Alternatively, the AP240/AP235 models allow Window or HF modes to be selected.

The Adv button in this section gives access to the choice of offset of the trigger out signal and the external trigger range.

4.2.7. Acquisition Mode

The acquisition mode of the digitizers is selected using one of four acquisition mode buttons in the Acquisition section of the control panel.

Available acquisition modes are Auto, Normal, Single and Stop.

This Acquisition section of the control panel also indicates the number of waveforms acquired per second by the digitizer. The use of each of the acquisition modes is described below. In the discussion, a valid trigger indicates a trigger signal meeting the trigger conditions at a time when the digitizer is armed and ready to acquire data.

Stop will stop the acquisition and hold the latest complete acquisition on the display.

Single mode is used in order to capture either, for instruments in the digitizer class, one event at the first valid trigger, or, for instruments in the Averager class, an average over the desired number of triggers. It freezes this result in the module’s memory, and on the display, until the user requests another acquisition. After an acquisition is taken in Single mode, the module will ignore subsequent trigger events until the Single button is pressed again or another acquisition mode is selected. Pressing the single button re-arms the trigger and captures another acquisition of the same type.

Normal mode is used to continuously acquire averages or waveforms into memory for valid trigger events. The display is updated with the new acquisitions in accordance with the Display Refresh Rate setting under the Options menu on the top line of the AcqirisLive Control window. Acquisitions will happen for each valid trigger until either single or Stop is pressed.

If the trigger indicator at the lower left of the waveform display does not blink green, the digitizer is not receiving valid triggers. Check the trigger settings or, for instruments in the digitizer class, select Auto mode in order to obtain a display of the waveform.

Auto mode, for instruments in the digitizer class only, will acquire and display waveforms according to the trigger settings if a valid trigger is present within a timeout interval. If a valid trigger is not available within this interval, the digitizer generates its own trigger in order to digitize and display whatever signal is at the input at that time.

If valid triggers are received at a high enough rate, Auto mode behavior differs little from normal mode behavior on the display. However if the trigger rate is too low or trigger settings are not appropriate for the characteristics of the signal, the input signal will not be stable on the display. This occurs because the internal auto-trigger generated by the digitizer is asynchronous to the input signal.

Auto mode is often used to aid in setup when the input signal must be quickly characterized in order to determine proper trigger settings for Normal or Single mode acquisitions.

4.2.8. Timebase and Memory

The timebase and memory settings should be chosen together in order to optimize the desired sample rate for the acquisition.

In Oscilloscope Mode the timebase settings range from 1 ns/div to 5 S/div. In

Transient Recorder Mode time windows of 8 ns to 20 Ks are available when using a card with the 2 MS memory option.

In order to change the timebase click on the white area in the timebase display window and select the desired timebase from the list.

Alternatively, the Decrement / Increment timebase selection buttons may be used in order to change the timebase in steps with each click of the mouse. The sample rate and sample interval resulting from the combination of memory and timebase are indicated under the Sampling section of the contro l panel.

The Memory setting shows the maximum number of points to be acquired. The actual number of points acquired could be less than the maximum number set if the acquisition time window selected fills less than the selected

User Manual: Agilent Acqiris Averagers Page 35 of 44

amount of memory at the full sample rate. Alternatively, in Oscilloscope Mode, if a slower timebase is selected, in order to ensure that the selected amount of memory will be filled before the acquisition is complete, the digitizer will reduce its sample rate as necessary. This action differs from the same case in Transient Recorder Mode where preference is given to the selected sample interval.

In Transient Recorder Mode, the digitizer will continue to allocate more memory with increasing capture times. This is done in order to maintain the selected sample interval for as long a time window as is possible given the available memory. When the card memory is fully utilized, the sample rate will decrease with increasing capture times.

4.2.9. Segmented Memory

Segmented memory enables the sequential acquisition mode having less than 800 ns dead time between segments.

This feature is particularly useful when capturing impulse-response type waveforms in a variety of applications.

Segmented memory is available in all acquisition modes. However, it is generally used in either Normal or Single mode. Waveform display occurs only after all segments are filled. Segments can be displayed sequentially on the display or they can be overlaid on top of each other. To select an overlaid display, select the item Overlay Segments under the Options menu. Segmented data can be archived to any storage device on the computer in a similar way to non-segmented acquisitions. If the acquisition needs to be stopped before all segments are filled, press the Stop button.

In order to set up the sequential acquisition mode, simply select more than 1 segment on the segment selection control under the Memory section of the control window and start the acquisition.

4.2.10. Display Features, Zoom and Persistence

A waveform zoom feature is available under the View menu of the display window. The screen image below shows the normal display and the zoom window that appears when zoom is selected.

The zoom scale and center position are modified using the horizontal and vertical controls shown on the zoom display window. To the right side of the zoom window click on the “Horizontal” button, then modify the zoom expansion factor by clicking the “In” or “Out” buttons.

The vertical blue cursor, or box, (depending on the zoom factor), shown on the main display window indicates the center point and range of the zoom. The vertical box or line can be dragged right or left with the mouse in order to re-position the center point of the waveform displayed in the zoom display window. Alternatively, the right and left arrows at the bottom of the zoom display window will reposition the horizontal center of the zoom.

In order to expand the zoom region vertically, select the

“Vertical” button on the zoom display and click the “In”

/ “Out” buttons. Alternatively, the up / down arrows on the scale indicator just below the vertical button can be used. When the vertical scale is expanded, the top and bottom lines forming the blue box in the main display window show the extent of the vertical zoom region. The vertical position of the box, and thus the data region shown on the zoom display, may be changed by moving the box vertically with the mouse or by clicking on the up and down arrows at the left of the zoom display window.

When expanding a waveform to the point where individual data points are seen, it may be desirable to use the “Connect data points” option under the Options menu in AcqirisLive. When this option is selected, larger data points are drawn on the display and a line connects each point.

User Manual: Agilent Acqiris Averagers Page 36 of 44

The main display window contains a drop down selector, under VIEW, that controls the display type for both the main display and the zoom window. There are two modes available, the Live Display, or the Persistence Display.

When persistence display is selected, all points drawn to the display windows remain on while new points are added from subsequent acquisitions. This mode is particularly suited for examining the jitter on an edge of a periodic signal as shown in the image below.

To clear the display in persistence display mode, press the “Clear” button that appears on the main display window.

4.2.11. Using averagers/digitizers of different models simultaneously

AcqirisLive features a single timebase (i.e. the set of acquisition parameters pertaining to the time axis of the waveform, such as the sampling rate and the number of samples). It is therefore not possible to specify different timebase settings for different modules. However, it is possible to run each module at its maximum available sampling rate performance in the following way:

• Select the Oscilloscope Mode.

• In the channels list, select a channel from the fastest module (that will make that module the current

module).

• Make sure the maximum number of samples (MAX samples in Memory panel) is large enough.

• Select the desired time per division setting (timebase panel): the fastest sampling rate for the current module, compatible with the maximum number of samples, will be selected and applied to all modules in the system. Any module not capable to satisfy that sampling rate will run at its closest setting.

Finally, note that the sampling rate menu (visible in Digitizer mode) contains the sampling rate capability of all modules currently in use. You must turn off the channels of 'slower' modules in order to view the highest sampling rates of the 'faster' modules.

4.2.12. Displaying Multiple Traces

If multi-channel modules, or several single channel modules, are installed on the PCI bus, multiple waveform displays, one from each installed channel, are overlaid on the display.

The current channel is selected by clicking on the channel name within the upper portion of the control panel. The vertical settings of the current channel and the timebase and trigger of the correspondent digitizer are then indicated.

A unique name is given to each channel installed on the PCI bus. The current channel is indicated by a dark highlight over the name. Settings changed on the main control panel will only affect the active channel. Always verify proper acquisition setting for each channel prior to collecting data.

There is a small rectangular indicator button to the left of each channel name in the channel list. If this indicator is black, the channel waveform is active; the channel is turned on and displayed in the display window. Clicking on this indicator once turns it transparent. This deactivates the reading of the data from the channel and clears the waveform display. Clicking once again on the indicator turns the channel back on and the indicator black. After making such changes the channel data will not be displayed until after the next acquisition.

The image at the left shows a channel list with four active channels in the system. All are turned on and displayed in the display window as indicated by the black indicators at the left. The DC271

Ch1 is the currently selected channel.

User Manual: Agilent Acqiris Averagers Page 37 of 44

The channel list image at the right shows the same system with the

DC271 Ch2 selected as the current channel. The other indicators are transparent and therefore their channels are turned off and not displayed.

As an alternative to controlling the channels to be displayed from the control panel you can also control them by selecting the desired channels from the list of channel gains shown at the right hand side of the display window.

Pressing the Finder button at the top of the control panel will identify the module associated with the current channel. When the finder button is depressed, the LED on the front panel lights up yellow on the digitizer module with the current channel.

4.3. Top Line Menu of AcqirisLive

The top line menu contains functions offering additional utility to the AcqirisLive application. Provisions are included for storing instrument setups, waveforms, converting the format of stored data files, identifying models and versions and in the Options Menu, calibrating the digitizer, extended control of event acquisition and controlling the display properties. These possibilities are discussed in the next few sections.

4.4. Setup

The Setup menu allows you to store and recover instrument configurations for digitizers. It is not active for

Averagers. When you store a configuration the parameters used to configure each of the channels of each module together with the timebase and trigger configurations will be saved in a file. You will be asked to choose the destination directory for the file and be allowed to give it an appropriate name. When you choose to recover a file you will be asked to give the name or navigate to it using the browser functionality. If you are using many modules the settings in the configuration file will be then with the digitizer in the same location. associated preferentially with the digitizer that has that serial number or

Only the parameters of the modules are stored in the configuration file. This means that parameters of the

AcqirisLive program, such as channel activation and display mode choices are not included.

4.5. Store, Autostore and Conversion of W aveforms

The store menu gives you two mechanisms to store waveforms .

1. The Store Waveform option allows you to store the last acquired waveform for the current channel. To store the last acquired waveform for each of the active channels, you must sequentially select each one to make it the current channel and use the Store

Waveform menu command. The title of the Store

Waveform dialog indicates the name of the channel that will be stored. The data is stored in ASCII format,

Volts). as Raw ADC values or Amplitudes (in

2. The Autostore option allows you to store consecutively acquired waveforms from all of the active channels into files. Configuration of

Autostore is achieved by selecting the Autostore item under the Store menu. Autostore brings up the dialog box shown here. Three formats are available for stored waveforms, Binary, Raw ADC values, and Amplitude values (in Volts). The destination folder can be selected or created on any system drive using the Browse button. Waveforms will be stored in a new folder under the destination folder.

User Manual: Agilent Acqiris Averagers Page 38 of 44

The new folder will be named AcqirisRun_{date}_{time}. Each waveform will be stored in a file with a name {channel identifier}_{i}.{suffix}, where i goes from 1 to N and the suffix indicates the data type. If the data is stored in Binary form, the suffix will be .data and an additional ASCII format .awh file is written which contains necessary details about each of the other files. Otherwise, the suffix will be .awd.

The number of generations of waveforms to be stored with this function is selectable from 1 to N. N is a maximum number dependent on the number of samples taken in each waveform. There i s also a user selectable limit on the total amount of data allowed in any single Autostore acquisition series.

When using segmented memory with Autostore, a waveform is considered to be a complete set of acquired segments. When storing Amplitudes, it is possible to write each segment as a separate column of data in the waveform file. This is achieved by clicking the check box option Segments as columns in the Autosto re settings window. Formatting with segments in columns, in Amplitude format, is limited to 200 segments.

The Binary file format selection provides the fastest storage performance and generates the smallest files on the hard drive. Binary files can be subsequently converted to either Raw values using the Data Format Conversion utility provided with AcqirisLive.

ADC values or to Amplitude

Selecting the “Write files at end option” acquires all waveforms into memory first, before converting to the selected format and writing the data to disk. This option allows for faster data capture ti

“Show waveforms” allows the user to see the waveforms before writing them to disk. mes. Selecting the

When all parameters are selected, click OK, and then select Normal, Auto or Single acquisition mode.

When the waveforms are recorded a progress bar will appear to show the status of t he acquisition and of writing the data to disk. The Autostore function can be aborted from the progress bar.

Raw ADC values are the raw 8-bit ADC levels in ASCII format. The channel’s gain and offset are needed in order to obtain the signal level in Volts. The required information to convert the Raw ADC values to voltage is included in the header information of each Raw ADC ASCII file or in the .awh file accom convert the Raw ADC values into Volts t fo f e used: panying the binary files. In order to

V = vGain * data – vOffset for Digitizer mode data

V = vGain * (data/NumberOfWaveforms – 128) – vOffset for Averager mode data

V = |vGain| * (127 – data/NumberOfWavef orms) – v Offs et for Inverted Averager mode data

V = vGain * data / NumberOfWaveforms for real-time Add/Subtract mode

The Amplitudes [V] selection stores the waveforms as voltage samples in ASCII format. This data has already been adjusted for gain and offset and can be directly read out as a voltage. However, it requires roughly twice as much space as Raw ADC value data.

For the sequential acquisition mode, the s

Autostore and Save Waveform functions. egment time stamps are also stored when waveforms are saved using the

Data stored in binary format with Autostore must be converted before it can be imported into a data processing package or a spreadsheet (e.g.

Excel). The Data Format Conversion option is found under the store menu.

Select Data Format Conversion in order to bring up the fo llowing window.

The acquisition must be stopped in order to access this menu.

To convert the binary data, simply select the sou rce folder where the data is located, select the output format and click OK.

All binary data files found in the source directory will be converted to the desired format. If you are writing amplitudes with multiple segments, formatting with segments in columns is limited to 200 segments.

4.6. Calibrate

The Calibrate item in the Options menu leads to a sub-menu where a particular type of calibration can be invoked. In order to use this ite poss ibilities are m the acquisition must be stopped; otherwise the Calibrate option is not selectable.. The

ƒ Calibrate instruments - to perform a self-calibration on the digitizers, calibrating the gain, offset, trigger level and the timing. Performing a ca libration compensates for any drift in these properties, typically due to component temperature variations.

ƒ Calibrate cur instrument. rent combine configuration – to perform a calibration only at the current configuration in each

User Manual: Agilent Acqiris Averagers Page 39 of 44

4.7. Averager settings

The Averager setting function gives the user control of many

details of the averaging operation as were explained in 3.6

Data Acquisition - Averaging Mode. This includes: the

number of triggers required for each sum, the kind of trigger synchronization, the dithering to be applied, and flags for data inversion and activating the Trigger OUT always option.

Noise Suppressed Accumulation (NSA) can also be controlled here. The Number of Rounds shown here corresponds to N robin

and Waveforms Per Round to N succ

of the round-robin mode. Note that the displayed waveform is that of the average data value in each bin so the NSA signal may be very small. A zoom can be used to examine it in detail. If Add/Subtract control is enabled an appropriate signal must be made available at the P1 Control MMCX connector.

The Averager settings will appear for the currently selected channel. If a different channel is desired, select it in the

Instrument Control window and then simply click anywhere in the Averager Settings window to force it to update to the new choice.

4.8. External 10 MHz Reference and External Clock

Either one of the External 10 MHz Reference or External Clock options can be selected in the Options Menu. The external input threshold option, available in the options menu, must also be set appropriately.

The External Clock function is available in AcqirisLive only when using the control panel in Transient Recorder

Mode. The input frequency should be keyed into the clock frequency selector (outlined in red) on the control panel.

Then the appropriate sampling frequency must be selected in order for the proper timebase to be displayed on the waveform display window. The time window displayed can be manipulated by allocating more or less memory to the acquisition.

The External 10 MHz Reference in AcqirisLive is designed to work either in Oscilloscope Mode or in Transient

Recorder Mode. The timebase and sample rate for waveform capture is fully selectable when using the external reference, just as it is when using the digitizer’s own internal reference clock.

4.9. Additional Waveform Information

The Options menu also includes possibilities to obtain additional information about the waveform. In both cases described below the display will be updated after each acquisition.

The “Show Segment Timestamps” option for the digitizer instrument class shows a display like the one here:

It gives you the information needed to better understand the trigger times of the individual segments of an acquisition. By definition the first trigger of an acquisition occurs at t

1

=0.0. There are two columns for each digitizer. For the n’th row, the first column contains the difference t

n

column contains t

n

column and the standard deviation of the t ps.

n

– t

n-1 and the second

. The top line contains the average value of the t

n

in the left

in the right one. The times are in

The “Show Parameters” option activates a display of pulse parameters for each active channel. Inactive channels will have parameter values of 0.0. The parameters will be calculated in a model in which the distribution of measured voltages has two distinct peaks. If this condition is not satisfied the results will be unreliable.

4.10. Display Features

Disp lay features in the Options menu include items for

• turning the waveform display off (by de

-selecting “Display Waveforms”),

• overlaying segments of a sequence,

• connecting displayed data points,

• modifying the display refresh rate,

User Manual: Agilent Acqiris Averagers Page 40 of 44

• currently selected module

4.11. Command Line Switches

When st more sw arting the itches to ch

AcqirisLive application, the executable's name in the command line can be followed by one or ange the running conditions.

Syntax

The synt ax of the command line is the fo llowing:

path\AcqirisLive.exe options where

options

are chosen from the following list

[-a][-b][-B][-c][d][-h][-i][-l][-m][-n][-p][-s][t][-v][-x]

Available Swit ches

List of a ab

-a start in Averager mode

-b Buffer size for large waveforms

-B Number of bank for S AR mode

-c no initial calibration

-d use DMA for data transfers (de fault) priority lt)

-i use acquisition interr upts

-n no DMA for data transf ers

-r start in Transient Recorder (digitizer) Mode off

Operation Mode

Acqiris averagers can be operated in Averager Mode or in standard Digitizer Mode.

AcqirisLive operates in either mode, but since th ese are mutually exclusive, only instruments that support the chosen mode are accessible in any given mode. E.g. only averagers are accessible in Averager Mode, while both di gitizers and averagers are accessible in Digitizer Mode.

The ope ration mode can be switched within the application. By default, AcqirisLive starts in Digitizer Mode.

Buffer size

Upon startup of the application, the available physical memory on the computer is checked. The total data buffer size is then limited to the available memory minus 10% of the tota l memory. It is also possible to limit the total data buffer size to a fixed value with the -bn option (where n is the limit in MB, e.g. -b200 will limit the buffer to 200

MB). The data buffer is then shared among the active channels.

When the total acquisition memory (sum over all active channels) exceeds the data buffer size limit, readout of the data will be partial. The special operation modes such as Segments Average, Average, Digital Filter and FFT are not supported with partial data readout.

Partial readout is controlled with the Very Long Memory panel (menu Optio segment and the first sample to read out can be specified. Note that if the buffer is large en segment, only full segments will be read (i.e. first sample control is ignored). ns ->Very Long Memory...). The first ough for at least 1 full

When sa ving data to file, you will be prompted for a maximum number of segments per file.

User Manual: Agilent Acqiris Averagers Page 41 of 44

Banks for SAR

Some of the digitizers are capable of acquiring data and being read out simultaneously. The -Bn option allows the internal memory of the digitizer to be divided into n banks. To observe the improved acquisition trigger rate achievable in this mode it is suggested to set up the desired conditions and then turn the event display off. The trigger out signa l can be used to monitor the trigger rate.

Calibration

Before operation, the instruments must undergo an internal calibration. As this calibration can only proceed sequentially for each channel, the initialization time of AcqirisLive can become significant for systems with many channels. Therefore, it can sometimes be desirable to be able to start the ap calibrati on (which must then be done "m anually" before collecting correct data). plication quickly by skipping the

DMA Data Transfers

By default, data transfers on the PCI bus between the digitizer memory and the PC are done in DMA (Direct

Memory Access) mode, which allows the fastest transfer speeds (up to about 130 MB/s peak, 100 MB/s sustained acquisition data rate). These very high data rates are not well supported by all hardware configurations (PC motherboard), and system hangs due to endless retries on the PCI bus have been observed on some machines. Such problems disappear when switching DMA mode off (data transfer rate is then about 2.8 MB/s).

Please no therefor te that switching DMA da e be ignored when –n i ta tran s used. sfer mode off precludes the use of acquisition interrupts. The –i switch will

Process Priority

When the program checks for an event from the operating system (e.g. to react to user input), it can put itself to sleep for a specified period of time. This gives other applications more processor time, but limits the performance of

AcqirisLive in terms of maximum acquisition rate. The default is not to go to sleep, but this can be overridden to allow be running.

tter performance for other app

When the acquisition stops, th lications. Note that this sleep policy is only in effect wh e sleep policy is always reset to the longest sleep period. ile the acquisition is

Acquisition Interrupts

When performing acqu

hard

wa e interrupt. isitions, there are 2 possible ways to wait for the acquisition to complete: polling or using a

-p Polling: AcqirisLive runs as a single thread process. After an acquisition is started, the instruments are regularly polled to check when the acquisition is com for reading the data after the acquisition may vary. plete. The user interface is still alive, but the latency

-i Hardware interrupt: In this mode, AcqirisLive is run as a multi-thread process. After starting an acquisition, a new thread is created to wait for the hardware interrupt indicating that the acquisition completed. This thread is put to sleep and wakes up upon reception of the interrupt from the instrument, or after a 2 seconds timeout.

Please note that switching the DMA data tr will ther efore be ignored when –n is used. ansfer mode off precludes the use of acquisition interrupts. The –i switch

Transient Recorder Mode

The app lication will start in the T ransient Recorder Mode as opposed to the Oscilloscope Mode

Simulation Mode

Simulati on mode allows you to run A cqirisLive without having any digitizer installed in the PC.

Temperature Update

The internally measured temperat this featu re off. ure of the currently selected module is normally displayed. Using this switch turns

Acquisition State

This switch turns on the display of the current acquisition state (Stopped, Ready, Started, Reading, Data ready). This is helpful when running with time-consuming setting etc.), bu t will slow down the overall p erformance. s (e.g. many segments overlaid, many channels, large records,

Multi-instruments

By default, AcqirisLive tries to detect instruments connected through AS bus, and digitizers as a multi-instrument. This behavior can be inhibited by the -x switch.

will initialize any connected

User Manual: Agilent Acqiris Averagers Page 42 of 44

5. Appendix A: U1092A-HVB XA100 BNC Input Overvoltage Protection

Specifications and User Instructions

User Manual: Agilent Acqiris Averagers Page 43 of 44

6. Appendix B: U1091AX02 XP102 Fan Unit for the AP Family of Modules

Specification and Assembly Instructions

User Manual: Agilent Acqiris Averagers Page 44 of 44

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!