Baumer DFON is a versatile graphical display designed for use in hazardous environments, offering a range of features to enhance process monitoring and control. With its ability to operate as a stand-alone display or integrated into instruments like CombiPress, CombiTemp, and CombiLyz, it provides flexibility for various applications. The DFON's unique background color setting, featuring white, red, and green options, allows for easy customization and clear visibility.
Advertisement
Advertisement
Operating instructions
CombiView™, type DFON
DFON graphical displays are used as separate units, but also used in complete instruments, e.g. CombiPress™, CombiTemp™ and CombiLyz™
Safety instructions
This instrument is built and tested according to the current EUdirectives and packed in a technically safe condition. In order to maintain this condition and to ensure safe operation, the user must follow the hints and warnings given in this instruction.
During the installation the valid national rules have to be observed. Ignoring the warnings may lead to severe personal injury or substantial damage to property.
The product must be operated by trained staff. Correct and safe operation of this equipment is dependent on proper transport, storage, installation and operation.
All electrical wiring must conform to local standards. In order to prevent stray electrical radiation, we recommend twisted and shielded input cables, as also to keep power supply cables separated from the input cables. The connection must be made according to the connection diagrams.
Before switching on or off the power supply take care that other equipment is not effected. Ensure that the supply voltage and the conditions in the environment comply with the specification of the device.
Description
The DFON is powered by the 4…20 mA loop and need no additional power supply. When mounted on new instrument series like the CombiPress™ PFMx it is via the internal UnitCom ribbon cable supplied with power and a digital signal, instead of integrated in the 4…20 mA loop. In instruments with UnitCom both the transmitter and display can be programmed at the same time alternatively separate programming of each device can be selected.
The DFON has an unique background colour setting. Three colours are available – White / Red / Green and further the red and green colour can be set as flashing in warning mode.
The DFON has two integrated galvanic separated relays. The instrument can be delivered with or without activated relays. If supplied without activated relays, those can be activated by purchasing a license code from Baumer. By entering the code to the instrument via the FlexProgrammer the relays are enabled.
The DFON can be programmed by the touch screen on the display or by the FlexProgrammer 9701 Baumer programming unit with the Baumer FlexProgram installed on a PC.
( Free download from www.baumer.com)
The DFON can be delivered as a stand-alone display or as an integrated part of an instrument e.g. PFMx, TFRx and AFIx.
Alternatively it can be delivered as a loop-powered local display for any instrument in a SS case for wall mounting, panel mounting or tube mounting.
Display mode
There are 10 different display modes available www.baumer.com
Digital small large
Analogue analogue same w. bar graph same w. value
Bar graph horizontal vertical
Operating instructions: 11112759 15 EN
Tank tank illustration bottle illustration
Page 1 / 7
Operating instructions
CombiView™, type DFON
Dimensions for mounting of stand-alone
Wall mounting Panel mounting Tube mounting
PCD for holes: Ø95 mm Panel cut out: 80 mm
PCD for holes: 82 mm
Tube diameter: Ø38 mm
Ø51 mm
Ø64 mm
Electrical connection
Pin 1
Pin 2
Pin 3
Pin 4
Pin 5
Pin 6
Com 1
Com 2
-
+ 4…20 mA
4…20 mA
Relay 21
Relay 22
Relay 11
Relay 12
FlexProgrammer ① red
FlexProgrammer
②
black
Cable gland
Cable diameter M16 plastic
M16 stainless steel
M20 plastic
M20 stainless steel
5 … 10 mm
5 … 9 -
8 … 13 -
11 … 13 -
Check the maximum temperature for the cable used.
Be sure to fixate the instrument before tightening the cable gland.
When using M16 stainless steel and M20 stainless steel the maximum tightening torque is 4 Nm. UnitCom
Pin 3 and 5 can be jumpered together if common supply is used for the two relays, e.g. via a M12 5-pin connector.
When instruments like PFMx are delivered with DFON display terminal 1 and 2 are not connected. In this case the DFON is powered and will have data through a special ribbon cable (UnitCom). When connected via UnitCom the DFON and transmitter can be programmed together.
When upgrading a product from the CombiSeries without display with a DFON touch screen, remember to remove the Oring from the sealing. Otherwise the sealing won’t be tight.
Special connection types
For connection ‘type B’ all the necessary wiring is done inside the DFON housing. It is a ”Plug-and-play” solution. Connect:
PF20x (3-wire) DFON
M12 male, 4-pin ↔ M12 female, 4-pin
For connection ‘type C’ all the necessary wiring is done inside the DFON housing. It is a ”Plug-and-play” solution. Connect:
4...20 mA transmitter (2-wire) DFON
M12 male, 4-pin ↔ M12 female, 4-pin www.baumer.com Operating instructions: 11112759 14 EN Page 2 / 7
Electrical connection
For connection ‘type D’ the following connection diagram has to be followed:
PF20x (3-wire) DFON
Pin 2 (Iout flow) ↔ Pin 1 (+ 4...20 mA)
M12 male, 4-pin
Pin 1 (+ Vs) ↔ Pin 1 (+ Vs)
Pin 3 (GND)
↔
Pin 3 (GND)
Operating instructions
CombiView™, type DFON
For connection ‘type E’ the following connection diagram has to be followed:
4...20 mA transmitter (2-wire) DFON
Pin 3 (- 4...20 mA)
Pin 1 (+ 4...20 mA)
↔ Pin 1 (+ 4...20 mA)
↔
M12 male, 4-pin
Pin 1 (+ 4...20 mA)
M12 male, 4-pin
M12 male, 4-pin
Terminal
Terminal
Touch screen programming the CombiView™ , type DFON
MENU and the display may start with one out of two possible selections.
With traditional 4…20 mA loop configuration (connected to pin 1 and 2 on the display)
Configuration
Menu Display menu Display setup
Diagnostics
With a transmitter communicating with the DFON via UnitCom ribbon cable
Product menu
Menu Display menu
Data display mode
Product menu
Enabling the user to set up the connected transmitter
Display menu
Enabling the user to program the DFON display
Data display mode
Enabling the user to program the connected transmitter. The programmed values will also be valid for the DFON display
(Ignoring whatever setup already set on the display).
Only programming of background colours and relays must be done under “ Display menu ” → “ Configuration ”
Data display mode will communicate with the transmitter digitally, which is more accurate than using the 4…20 mA communication.
For menu tree please see page 7.
www.baumer.com Operating instructions: 11112759 14 EN Page 3 / 7
Operating instructions
CombiView™, type DFON
CombiView™, type DFON - ATEX specifications and instruction
Zone 0/1 Gas:
Zone 20/21 Dust:
Zone 2 Gas:
II 1 G Ex ia IIC T5 Ga
II 1 D Ex ia IIIC T100°C Da
II 3 G Ex ec IIC T5
Please ensure the special requirements for installation in the specific environment is followed, as described below.
ATEX approval requires the CombiView ™ , type DFON to be installed in IP67 certified housing.
Safety instructions
This instrument is constructed and tested according to the current EU directives and packed in a technically safe condition. In order to ensure safe conditions and operation, the user must follow the instructions and warnings given in this instruction and the standard operation instruction.
During the installation the valid national rules have to be observed. Ignoring the instructions may lead to severe personal injury or substantial damage to property.
The product must be installed and operated by trained staff. Correct and safe operation of this equipment is dependent on proper installation and operation.
All electrical wiring must conform to local standards and the connection must be made according to the connection diagrams on the following pages.
Before switching on the power supply take care that other equipment is not affected. Ensure that the supply voltage and conditions in the environment comply with the specification of the device.
Before switching off the supply voltage check the possible effects on other equipment and processing system.
To obtain the specified ingress protection degree, the CombiView ™ , type DFON must be mounted with a compliant cable.
Warning
This product is allowed to be operated in an explosion hazardous atmosphere of zone 0 only if atmospheric conditions exists
(temperature –20°C…+60°C and pressure from 0.8 … 1.1 bar). Under other atmospheric conditions the certificate may be used as a guide.
Use of FlexProgrammer 9701 is only allowed in the safe area, not in the hazardous area.
Impact test of the display cover is performed according to EN 60079-0, with low impact energy of 2J and does not create a crack or other intrusion into the housing. However the display, which is mounted close to the front cover may be damaged, but this will not create any external sparking. The housing is impact texted also according to EN 60079-0, with low impact energy of 4J on housing, connectors and cable gland.
This product contains no replaceable parts. In case of malfunction the products must be shipped to Baumer for repair.
Operating instructions: 11112759 14 EN Page 4 / 7 www.baumer.com
Operating instructions
CombiView™, type DFON
ATEX Gas ia and for ATEX Dust ia
For ATEX ia Gas zone 0/1/2 and ATEX ia Dust zone 20/21/22 a zener barrier must separate the hazardous and safe area, and must be installed in accordance with prevailing guidelines for the zone.
ATEX ia Gas / Dust data:
Approval: ATEX II 1G Ex ia IIC T5 Ga
ATEX II 1D Ex ia IIIC T100°C Da
Voltage drop U
Disp
4.5 or 6.5 VDC
Temperature class T1…T5 Zone 0 and 20
Internal inductivity
Internal capacity
L
C i i
<10 µH
<15 nF
-20°C…60°C
Zone 1/2 and 21/22 -40°C…65°C
Barrier data U i
I i
<30 VDC
<0.1 A
P i
<0.75 W
ZEX-ALL.B28RD100 Suitable barrier
The display is supplied by the 4…20 mA loop from the transmitter and can thus be placed near the transmitter or near the power supply, as preferred. However to make the display intrinsic safe a zener barrier must be inserted before the DFON. The transmitter may be placed inside or outside the hazardous area.
If the CombiView ™ , type DFON is attached to a Baumer transmitter using the UnitCom ribbon cable, it is considered to be an integrated part of the instrument. Please see the ATEX instruction for the relevant instrument.
If the relays are enabled, each relay must be protected by a zener barrier. Use a barrier for each relay or a barrier with multiple channels. However the two relays must have each a barrier.
Barrier data U
I
<30 VDC
<75 mA
P
Suitable barrier
<0.75 W
ZEX-ALL.B30RS075
Display in the zone, transmitter outside the zone
1 +
Zener barrier
4…20 mA
15…30 VDC
2 -
Display and transmitter in the zone
1 +
2 -
4…20 mA
Zener barrier 15…30 VDC
With relay output
Display in the zone/transmitter outside the zone
4…20 mA
1 +
Zener barrier
15…30 VDC
2 -
3
Relay 1
4
Zener barrier
0…30 VDC
75 mA
5
Relay 2
6
Zener barrier
0…30 VDC
75 mA www.baumer.com
With relay output
Display and transmitter in the zone
Operating instructions: 11112759 14 EN
1 +
2 -
3
4
5
6
4…20 mA
Relay 1
Zener barrier 15…30 VDC
Zener barrier
0…30 VDC
75 mA
Relay 2
Zener barrier
0…30 VDC
75 mA
Page 5 / 7
Operating instructions
CombiView™, type DFON
ATEX Gas nA
For ATEX ec zone 2 is approved without using zener barrier and must be installed in accordance with prevailing guidelines for zone 2.
ATEX data:
Approval: Gas Zone 2 II 3 G Ex ec II T5
Voltage drop U
Disp
4.5 or 6.5 VDC
Temperature class T1…T5 -30 < T amb
<65°C
Internal inductivity L i
<10 µH
Internal capacity Ci
Maximum voltage U max
Maximum current I max
<15 nF
<35 VDC
<0.1 A
Display without relay output Display with relay output
1 +
4…20 mA
1 +
4…20 mA
2 -
3
Relay 1
2 - 4
5
6
Relay 2
Programming the CombiView™, type DFON in ATEX area
Programming the CombiView ™ , type DFON in hazardous area with the FlexProgrammer is not allowed, as the FlexProgrammer
(and/or the PC) is not ATEX approved.
Follow below procedure to programme the instrument: a Disconnect mains from the 4…20 mA loop circuit b c d
Disconnect the instrument from the circuit within the hazardous area
Uninstall the instrument and bring it to safe area
Connect the FlexProgrammer and perform the configuration session.
(Please see: www.baumer.com/downloads/product documents/manuals - DFON - Programming ) e f g
Reinstall the instrument in the hazardous area
Connect the instrument to the circuit
Connect mains from the 4…20 mA loop circuit www.baumer.com Operating instructions: 11112759 14 EN Page 6 / 7
Operating instructions
CombiView™, type DFON
Touch screen programming the CombiView™, type DFON
Display menu Configuration Identification
Input
Tag, S/N, Date, Prod. date
Input at 0%
Input at 100 %
Damping
Lin. Correction
Data display mode Transmitter value
Display conversion
Display output
Error / warning
Display at 0%
Display at 100%
Decimals
Unit
Abs. / Rel.
High error
Display setup
Diagnostics
Relay setup
Screen layout
Backlight
Language
Password
High warning
Low error
Low warning
Relay 1 mode
Relay 1 set point
Relay 1 reset point
Relay 2 mode
Relay 2 set point
Relay 2 reset point
Standard screens
Product specific screens
Colour
Intensity
EN / DK / DE / FR
Password enable
New password
Timeout value Menu timeout
Statistics
Demo setup
Factory setting
Service menu
Min/Max / Errors / Uptime
Select demo mode
Static demo value
Accept / cancel
Service info
Service code
Relay activation
Operating instructions: 11112759 14 EN
Current value
Current value
Damping value
Enable / Disable
Display value
Display value
Select from list
Select from list
Select, when pressure
High error limit
High error indication
High error backlight
See High error
See High error
See High error
AO / AC / NO / NC
Set point value
Set point value
AO / AC / NO / NC
Set point value
Set point value
Select display design
Select display design
White / Green / Red
10%-140% www.baumer.com
Enable / Disable
Create new password
Disabled / Static / Cyclic
Current value
Internal temp, Start up, Ver.
Enter code
Enter code