Welch Allyn SpiroPerfect Spirometer User Manual
Reklamë
Reklamë
CardioPerfect Workstation
SpiroPerfect Module - User Manual
Welch Allyn
Staalweg 50
2612 KK Delft
The Netherlands
0297
DIR 80012334 Ver. C
Welch Allyn, Inc
4341 State Street Road
Skaneateles Falls, NY
13153-0220 USA
CardioPerfect Workstation
SpiroPerfect Module – User Manual
Copyright
© Copyright 2008, Welch Allyn. All rights are reserved. No one is permitted to translate, reproduce or duplicate this manual or any part thereof, in any form, without prior written permission from Welch Allyn. Welch Allyn assumes no responsibility for any injury to anyone, or for any illegal or improper use of the product, that may result from failure to use this product in accordance with the instructions, cautions, warnings, or statement of intended use published in this manual.
Unauthorized copying of this publication may not only infringe copyright but also reduce the ability of Welch Allyn to provide accurate and up-to-date information to users and operators alike.
Welch Allyn
®
, CardioPerfect
®
Workstation and SpiroPerfect
®
are registered trademarks of
Welch Allyn.
Software in this product is Copyright 2008, Welch Allyn. All rights are reserved. The software is protected by United States of America copyright laws and international treaty provisions applicable worldwide. Under such laws, the licensee is entitled to use the copy of the software provided on the original distribution medium. The software may not be copied, decompiled, reverse-engineered, disassembled or otherwise reduced to human-perceivable form. This is not a sale of the software or any copy of the software; all right, title and ownership of the software remains with Welch Allyn.
The information contained in this manual is subject to change without notice.
All changes will be in compliance with regulations governing manufacture of medical equipment.
User responsibility
This product is designed to perform in conformity with the description thereof contained in this manual and accompanying labels and inserts, when assembled, operated, maintained and repaired in accordance with the instructions provided. A defective product should not be used. Parts that are broken, plainly worn, missing or incomplete, distorted or contaminated should be replaced immediately. Should any repair or replacement become necessary, we recommend that service be performed at the nearest approved service center. The user of the product shall have the sole responsibility for any malfunction, which results from improper use, faulty maintenance, improper repair, damage or alteration by anyone other than Welch
Allyn or their authorized service personnel.
Accessories
The Welch Allyn warranty can only be honored if you use Welch Allyn approved accessories and replacement parts.
Caution
Use of accessories other than those recommended by Welch Allyn may compromise product performance.
DIR 80012334 Ver. C 2 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
Warranty, Service, and Spare Parts
Warranty
All repairs on products under warranty must be performed or approved by Welch Allyn.
Unauthorized repairs will void the warranty. In addition, whether or not covered under warranty, any product repair shall exclusively be performed by Welch Allyn certified service personnel.
Assistance and Parts
If the product fails to function properly or if assistance, service, or spare parts are required, contact the nearest Welch Allyn Technical Support Center.
Latin America (+1) 305-669-9591 South Africa (+27) 11-777-7509
European Call Center (+353) 469-067-790 Australia 2-9638-3000
United Kingdom
France
(+44) 207-365-6780
(+33) 1-60-09-33-66
Singapore (+65) 6291-0882
Japan (+81) 3-5212-7391
Germany (+49) 7477-927-173 China (+86) 21-6327-9631
Before contacting Welch Allyn it is helpful to attempt to duplicate the problem and to check all accessories to ensure that they are not the cause of the problem.
When calling, please be prepared to provide:
Product name and model number and complete description of the problem
The serial number of your product (if applicable)
The complete name, address and phone number of your facility
For out-of-warranty repairs or spare parts orders, a purchase order (or credit card) number
For parts order, the required spare or replacement part number(s)
Repairs
If your product requires warranty, extended warranty, or non-warranty repair service, please call first the nearest Welch Allyn Technical Support Center. A representative will assist you troubleshooting the problem and will make every effort to solve it over the phone, avoiding potential unnecessary return.
In case the return cannot be avoided, the representative will record all necessary information and will provide a Return Material Authorization (RMA) number, as well as the appropriate return address. A Return Material Authorization (RMA) number must be obtained prior to any return.
Note
Welch Allyn does not accept returned products without an RMA.
Packing Instructions
If you have to return goods for service, follow these recommended packing instructions:
Remove all hoses, cables, sensors, power cords, and ancillary products (as appropriate) before packing, unless you suspect they are associated with the problem.
Wherever possible use the original shipping carton and packing materials.
Include a packing list and the Welch Allyn Return Material Authorization (RMA) number.
It is recommended that all returned goods be insured. Claims for loss or damage to the product must be initiated by the sender.
DIR 80012334 Ver. C 3 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
Limited Warranty Statement
Welch Allyn, Inc. warrants that the SpiroPerfect computer based Spirometer you have purchased (the Product) meets the labeled specifications of the Product and will be free from defects in materials and workmanship that occur within 1 year after the date of purchase.
Accessories used with the Product are warranted for 90 days after the date of purchase. Such accessories include: disposable flow transducers, pressure tubing, and nose clip.
The date of purchase is: 1) the date specified in our records, if you purchased the Product directly from us, 2) the date specified in the warranty registration card that we ask you to send to us, or 3) if you don’t return the warranty registration card, 120 days after the date on which the Product was sold to the dealer from whom you bought the Product, as documented in our records.
This warranty does not cover damage caused by: 1) handling during shipping, 2) use or maintenance contrary to labeled instructions, 3) alteration or repair by anyone not authorized by Welch Allyn, and 4) accidents.
If a Product or accessory covered by this warranty is determined to be defective because of defective materials, components, or workmanship, and the warranty claim is made within the warranty period described above, Welch Allyn will, at its discretion, repair or replace the defective Product or accessory free of charge.
You must obtain a return authorization from Welch Allyn to return your Product before you send it to Welch Allyn’s designated service center for repair.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED,
INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY
AND FITNESS FOR A PARTICULAR PURPOSE. WELCH ALLYN'S OBLIGATION UNDER
THIS WARRANTY IS LIMITED TO REPAIR OR REPLACEMENT OF PRODUCTS
CONTAINING A DEFECT. WELCH ALLYN IS NOT RESPONSIBLE FOR ANY INDIRECT
OR CONSEQUENTIAL DAMAGES RESULTING FROM A PRODUCT DEFECT COVERED
BY THE WARRANTY.
DIR 80012334 Ver. C 4 / 95
CardioPerfect Workstation
Table of Contents
SpiroPerfect Module – User Manual
Using the Spirometer Safely .................................................................... 10
2 General information .......................................................................... 14
Intended Use / Indications for Use........................................................... 14
Important Considerations......................................................................... 15
3 Installing the SpiroPerfect Spirometer ............................................ 16
Connecting the Flow Sensor with USB connection to the Computer ....... 16
Configuring the Welch Allyn CardioPerfect Workstation .......................... 19
4 The Spirometer Window ................................................................... 20
5 Customizing the Spirometry Module ............................................... 22
6 Ambient Settings /Temperature, Humidity and Pressure .............. 32
Why the Workstation needs Ambient Setting Information........................ 32
When to Adjust the Ambient Settings ...................................................... 32
Adjusting the Ambient Settings ................................................................ 32
7 Calibration of Flow Sensor ............................................................... 34
Error Messages Associated with Failed Calibration................................. 40
DIR 80012334 Ver. C 5 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
8 Recording Spirometry Tests ............................................................ 43
Record a Spirometry Test ........................................................................ 43
Add or Change Information in the Comment Editor ................................. 49
9 Viewing Spirometry tests.................................................................. 50
View and Add Information to a Test......................................................... 50
Common Features for each Tab .............................................................. 53
10 Interpreting Spirometry Tests .......................................................... 61
Editing and Confirming an Interpretation ................................................. 61
Reanalyze a Spirometry Test................................................................... 63
11 Printing Spirometry tests.................................................................. 65
Norm-Related Clinical Studies ................................................................. 69
Composite Norm Values .......................................................................... 70
DIR 80012334 Ver. C 6 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
Understanding Interpretation Results ...................................................... 72
13 Maintaining the Spirometer – Welch Allyn ...................................... 74
Ordering Information for Replacement Parts ........................................... 75
16 Statutory and Regulatory Requirements......................................... 81
17 Guidance and Manufacturer’s Declarations.................................... 82
18 Spiro Perfect VCT-400....................................................................... 86
Connect the flow sensor to the computer ................................................ 86
Configure Welch Allyn CardioPerfect Workstation ........................................... 86
Maintaining the flow sensor ..................................................................... 87
DIR 80012334 Ver. C 7 / 95
1 Introduction
CardioPerfect Workstation
SpiroPerfect Module – User Manual
1.1 About This Manual
This manual is written for clinical professionals performing pulmonary function testing.
Users must be familiar with measurements and the clinical significance of basic spirometry products.
Caregivers need to know how to properly coach patients, recognize acceptable waveforms and know whether results are reproducible or not and whether they meet ATS criteria or not.
The hospital's Biomedical/IT support staff shall require primary skills including disciplines related to maintenance and servicing computer controls/platforms. It is recommended that users attend a certified spirometry training course. The instructions given here are only a guide and should not be used to train a technician.
For definitions of specialized terms and abbreviations related to spirometry, see the Glossary.
Before using the spirometer, all users and technicians must read and understand this manual and all other information accompanying the SpiroPerfect spirometry option and the
CardioPerfect workstation.
Note
This manual supplements the CardioPerfect workstation manual, entitled
CardioPerfect Workstation User Manual. For information that the workstation and spirometry functions share — for example, instructions for moving through the menus, searching for patient data — see the CardioPerfect workstation manual.
Please take note of all safety precautions and warnings provided with this device before using this device and the accompanying software.
We at Welch Allyn are dedicated to provide safe products to our customers. It is the user’s responsibility to follow the rules of safety as established for their protection and for the protection of their patients as described in this manual. Please take special note of the safety
and precautions as described in Using the Spirometer Safely on page 10.
1.2 Symbols
The symbols shown below may appear on the spirometer components, on the packaging, on the shipping container, or in this manual.
Documentation Symbols
WARNING
Indicates conditions or practices that, if continued or not corrected immediately, could lead to illness, injury, or death.
CAUTION
Indicates conditions or practices that, if continued or not corrected immediately, could damage the equipment.
Shipping, Storing, and Environment Symbols
Single use
DIR 80012334 Ver. C 8 / 95
200x-xx
CardioPerfect Workstation
SpiroPerfect Module – User Manual
Expiration date
Keep away from sunlight
0297
Stacking limits
Do not dispose of this product as unsorted municipal waste. Prepare this product for reuse or separate collection as specified by Directive
2002/96/EC of the European Parliament and the Council of the
European Union on Waste Electronic and Electrical Equipment
(WEEE). If this product is contaminated, this directive does not apply.
Fore more specific disposal information contact Welch Allyn
Customer Service, see page 3 for telephone numbers.
Certification Symbols
Meets essential requirements of European Medical Device Directive
93/42/EEC
The authorized representative in the European Community
DIR 80012334 Ver. C 9 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
1.3 Using the Spirometer Safely
Before using or servicing the spirometer, you must read and understand the following safety-related information.
WARNING:
Do not perform spirometry test if any of the following conditions apply to the patient:
hemoptysis of unknown origin (forced expiratory maneuver may aggravate the underlying condition);
pneumothorax;
unstable cardiovascular status (forced expiratory maneuver may worsen angina or cause changes in blood pressure) or recent myocardial infarction or pulmonary embolus;
thoracic, abdominal, or cerebral aneurysms (danger of rupture due to increased thoracic pressure);
recent eye surgery (e.g., cataract);
presence of an acute disease process that might interfere with test performance (e.g., nausea, vomiting);
recent surgery of thorax or abdomen.
WARNING The spirometer captures and presents data reflecting a patient’s physiological condition. When reviewed by a trained physician or clinician, this data can be useful in determining a diagnosis. However, the data should not be used as a sole means for determining a patient’s diagnosis.
WARNING To minimize chances of a misdiagnosis, it is the physician’s responsibility to assure that spirometry tests are properly administered, evaluated, and interpreted.
WARNING People may become light-headed, dizzy, or even faint during a spirometry effort. Watch patients closely. If they choose to stand during testing, keep a chair immediately behind them. If there is any reason for concern, stop the test and take appropriate action.
WARNING To prevent cross-contamination, do not try to clean the flow transducers and nose clips. Discard these items after a single patient use.
WARNING The American Thoracic Society (ATS) recommends using rubber gloves when replacing disposable flow transducers, and washing hands after touching them.
DIR 80012334 Ver. C 10 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
WARNING:
The CardioPerfect family of devices is an integral part of a personal computer based diagnostic system. The user shall adhere to warnings in order to ensure safe and reliable performance of the system.
The personal computer (non-medical electrical equipment) shall be situated outside the patient environment (reference IEC 60601-1-1).
The personal computer used should adhere to the appropriate safety standard for non-medical electrical equipment (IEC 60950, or its national variants), and use of an isolation transformer is recommended.
The personal computer used should adhere to the appropriate electromagnetic compatibility (EMC) standard for non-medical electrical equipment (CISPR 22/24 -
FCC Part 15 - CE, or related national variants).
If it is required for the personal computer to be situated within the patient environment, it is the responsibility of the user to ensure that the system provides a level of safety in compliance with IEC 60601-1.
WARNING A color printer and a color printout are recommended for printing
Spirometry reports. Printing these reports with a monochrome printer or in black and white can lead to confusion as it is not easy to identify which curve is a Pre and which is a Post effort.
CAUTION
Do not clean the pressure tubing or sensor. Trapped moisture could affect their accuracy. Replace the pressure tubing when it becomes dirty. Replace the sensor when it becomes faulty.
CAUTION Do not use aromatic hydrocarbons, rubbing alcohol, or solvents for cleaning the spirometer.
CAUTION Do not immerse any part of the spirometer into a cleaning liquid or sterilize it with hot water, steam, or air.
CAUTION You cannot clean the spirometer or any of its components. If you choose to clean the calibrations syringe, wipe its external surfaces as needed with a cloth dampened with water only.
CAUTION Use only parts and accessories supplied with the device and available through Welch Allyn. The use of accessories other than those specified my result in degraded performance of the device.
CAUTION When you put the spirometer away, store its pressure tubing in a basket or drawer or other place that prevents compression or kinking.
CAUTION Avoid installing the spirometer in direct sunlight or in a location where it may be affected by significant changes in humidity, ventilation, or airborne particles containing dust, salt or sulfur.
CAUTION Keep the spirometer away form splashing fluids.
CAUTION Do not over tighten the rotor when assembling the Viasys VCT400
Spirometer. This can lead to inaccurate registration of low volume rates.
DIR 80012334 Ver. C 11 / 95
CardioPerfect Workstation
1.4 Product Overview
SpiroPerfect Module – User Manual
SpiroPerfect performs FVC, SVC and MVV testing, including pre-post testing.
It instantly displays flow-volume curves and depicts inspiratory and expiratory measurements.
For details, see the following sections:
Features
Ordering Information for Replacement Parts (page 75)
Specifications
Figure 1.1 Components of the SpiroPerfect Spirometer
Disposable Flow Transducer
For single patient use only to minimize the risk of cross contamination.
Pressure Tubing
Connects the flow transducers to the Spirometer sensor.
Serial Sensor
Connects to the serial port of your PC.
Converts pressure to airflow.
USB Sensor
Connects to the USB port of your PC.
Converts pressure to airflow.
Assembled SpiroPerfect Spirometer
Consists of: Disposable Flow transducer, Pressure tubing and Sensor, either Serial or USB.
Note: Only one sensor is shipped with the Spirometer.
It is either the Serial or USB sensor.
Nose Clip
Highly recommended during testing to avoid air leaks.
Unless a medical condition makes it uncomfortable or unpractical to use it, in which case, the clinician should record that the nose clip was not used.
3 liter Calibration syringe
For daily use, to calibrate the Welch Allyn SpiroPerfect spirometer for accuracy.
DIR 80012334 Ver. C 12 / 95
CardioPerfect Workstation
1.5 Features
SpiroPerfect Module – User Manual
Automatic interpretation and comparison to best pre-bronchodilator.
Real-time flow/volume and volume/time graphs.
Incentive graphic for pediatric patient coaching.
Multiple predicted norms.
Customizable report formats.
Validated to meet the American Thoracic Society spirometry accuracy standards for both ambient and BTPS humidified air.
Instant quality and variability check for proper test performance.
Single-stroke and multiple-stroke calibration protocols.
Reduced risk of cross contamination with Welch Allyn single-use, disposable flow transducers.
Meets all industry standards, including ATS, NIOSH, OSHA and Social Security.
Trending of several different tests from the same patient.
DIR 80012334 Ver. C 13 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
2 General information
2.1 Welcome
Welcome to the SpiroPerfect module of the Welch Allyn CardioPerfect Workstation. With this module, you can record, view and interpret spirometric tests. You can also use it to print spirometry tests in various formats.
The SpiroPerfect module exceeds the recommendations for spirometry of the American
Thoracic Society (ATS).
This manual contains specific information about the SpiroPerfect module of the Welch Allyn
CardioPerfect Workstation. For all general information about the workstation software, please refer to the Workstation manual, which describes:
Creating and editing Patient cards
General information about printing
For further information on installation and configuration please refer to the Workstation
Installation manual.
2.2 Intended Use / Indications for Use
Using the optional spirometry module and associated accessories to acquire, view, store, and print measures and waveforms of pulmonary function. The spirometer should only be used with patients able to understand the instructions for performing the test.
Indications for spirometry include, but are not limited to, the following:
Shortness of breath cough
Occupational exposure to dust and chemicals
Assist in the diagnosis of Bronchitis
Assist in the diagnosis of Asthma
Wheezing
Assist in the monitoring of bronchodilators
2.3 Contraindications
Relative contraindications to performing spirometry are [AARC Clinical Practice Guideline
Spirometry, 1996 Update]:
hemoptysis of unknown origin (forced expiratory maneuver may aggravate the underlying condition);
pneumothorax;
unstable cardiovascular status (forced expiratory maneuver may worsen angina or cause changes in blood pressure) or recent myocardial infarction or pulmonary embolus;
thoracic, abdominal, or cerebral aneurysms (danger of rupture due to increased thoracic pressure);
recent eye surgery (e.g., cataract);
presence of an acute disease process that might interfere with test performance (e.g., nausea, vomiting);
recent surgery of thorax or abdomen.
DIR 80012334 Ver. C 14 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
2.4 Important Considerations
The Spirometer should not be used if any of the following conditions exist or are thought to exist:
The spirometer is not regularly calibrated.
The maintenance instructions listed in section 13 are not satisfactorily completed.
Any part of the equipment or system is known, or suspected, to be defective.
DIR 80012334 Ver. C 15 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
3 Installing the SpiroPerfect Spirometer
The SpiroPerfect Spirometer consists of two elements: the spirometry sensor, and the software that runs on the computer to which the sensor is connected. Before you can start recording spirometry tests, you need to:
Connect the sensor to the computer.
Configure the software.
Warm up the Spirometer
After connecting the device it is recommended to let the Spirometer warm up.
1. Connect the Spirometer to the computer.
2. Open the Spiro module.
The sensor starts to warm up as soon as the SpiroPerfect module is opened.
3. Wait for at least 5 minutes before starting a new test.
Flow sensor with USB connection:
SpiroPerfect from Welch Allyn
OEM SpiroPerfect manufactured by Medikro Oy, Finland for Welch Allyn Inc, USA.
0537
For information on connecting the Flow sensor with USB connection see section 3.1
Flow sensor with serial connection:
SpiroPerfect from Welch Allyn
OEM SpiroPerfect manufactured by Medikro Oy, Finland for Welch Allyn Inc, USA.
0537
The Flow sensor with serial connection is ready for use after plugging it into the computer. No further driver needs to be installed.
3.1 Connecting the Flow Sensor with USB connection to the Computer
If the Spirometer has never been connected to the computer before, follow the instructions under First Time Installation. This procedure will copy and install the driver files from the
CardioPerfect Workstation Installation CD to the computer.
If the Spirometer was already installed, but is now connected to a USB port it hasn’t been connected to before, Windows shows the Found New Hardware Wizard. In this case the
Spirometer can be installed without needing the CardioPerfect Workstation Installation
CD. Therefore, follow the abridged instructions under Connecting to a different USB port.
DIR 80012334 Ver. C 16 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
First time installation
1. For a USB Spirometer, you will need to install the USB Spirometer drivers if you have not installed these already.
2. If you have a Serial Spirometer no drivers will have to be installed.
To connect the USB flow sensor to the computer:
3. Connect the device to a free USB port on the computer.
4. If you don’t have the appropriate rights for installing device drivers, Windows will display the following window asking you to type in a user name and password for a user with sufficient rights. Otherwise continue at the following step.
Type in a user name and password for a user with administrator rights and press the
OK button.
Select No, not this time and press the Next button.
6. The following window will appear:
DIR 80012334 Ver. C 17 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
Select Install the software automatically (Recommended).
Press the
7. If you are installing on a Windows XP system, the following driver signature warning window will appear:
Press the Continue Anyway button.
DIR 80012334 Ver. C 18 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
8. The system now installs the device. Please wait until the following window appears.
Press the Finish button.
3.2 Configuring the Welch Allyn CardioPerfect
Workstation
After connecting the spirometry sensor, you need to configure Welch Allyn CardioPerfect
Workstation.
To configure Welch Allyn CardioPerfect Workstation for use with the sensor:
1. Start Welch Allyn CardioPerfect Workstation.
4. Select Welch Allyn SpiroPerfect.
DIR 80012334 Ver. C 19 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
4 The Spirometer Window
This section guides you through the various parts of SpiroPerfect. The structure of the workspace is similar to the other Welch Allyn CardioPerfect Workstation modules and conforms to the Microsoft UI guidelines.
Figure 4.1 Main Window
Search area Toolbar Workspace Menu bar Title bar
Effort selector
Shortcut menu
Graph area
Title bar
Menu bar
Toolbar
Status bar
Search area
Interpretation area
Parameters area
The title bar displays the name of the program. Three buttons located on the right of the title bar can be used to maximize, minimize and close
CardioPerfect Workstation.
The menu bar contains the File, Edit, Mail, View, Action, Tools and Help menus. When a menu is grayed out you cannot access its functionality.
The toolbar contains the Patient, ECG, Exercise ECG, Recollect,
Spirometry, ABP, Print, and Print Preview buttons. It provides easy access to other CardioPerfect Workstation applications and most common tasks in the SpiroPerfect module.
The search area on the left hand side contains search and display functionality. In the search area, you can find a patient, see the date and type of tests recorded for a patient. You can create search patterns, so you can easily locate frequently needed information.
DIR 80012334 Ver. C 20 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
Workspace
Shortcut Menu
Status bar
The workspace displays tests and test-related data, such as graphs and measurements. This is where you record, view and interpret the data.
The workspace is divided into three elements:
Graph area:
This area displays spirograms and flow curves.
Interpretation
area:
The interpretation area displays the automatic or confirmed interpretation for the test, lung age and
ATS reproducibility data.
Parameters
area:
The parameters area displays each effort and up to 6 user-defined measured parameters.
In the workspace, you can use shortcut menus to access the most common tasks. You can access these tasks by clicking on the workspace with your right mouse button. Shortcut menus are context sensitive, meaning that they show only relevant tasks for the area clicked.
The status bar at the bottom of the window shows the name of the user currently logged on, the patient’s race, height, weight and the Prediction norm used in the Spirometry test currently viewed.
DIR 80012334 Ver. C 21 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
5 Customizing the Spirometry Module
This chapter shows how to adjust various settings like selecting prediction schemes, determining which parameters to view and print, and set various display options.
Customize features in the Spirometry settings.
To open the Spirometry settings:
The following screen appears:
Figure 5.1 Settings Screen
5.1 General Tab
To display the General tab:
The following screen appears:
DIR 80012334 Ver. C 22 / 95
CardioPerfect Workstation
Figure 5.2 Spirometry General Tab
SpiroPerfect Module – User Manual
Setting Description
Prediction
Select the prediction to use. The list contains all supported predictions.
VC Parameter
VC parameters, FEV1% formula:
The FEV1% formula determines the calculation method for the FEV1% value, which affects the automatic interpretation. The variable part of this formula is the denominator; the numerator is always the best effort’s FEV1 value.
To determine the way in which FEV1% is calculated, choose from these options:
FVC (FEV1% = FEV1/FVC)
FIVC (FEV1% = FEV1/FIVC)
Max (FVC, FIVC, SVC*) (FEV1% = FEV1/FVC or FIVC or SVC, the largest)
*Note: The SVC parameter is only included if Final Result is set to
Best composite.
Reversibility Reversibility is the percentage difference between pre-test and post-test data. This measurement indicates the effect of medication on lung function. Reversibility applies to each parameter separately.
DIR 80012334 Ver. C 23 / 95
Final Result
CardioPerfect Workstation
SpiroPerfect Module – User Manual
A patient’s best effort is a measurement calculated from a set of efforts. To determine the method in which best effort is calculated, choose from these options:
Calibration syringe
Calibration reminder
Pressure
Flow
Best effort Defines best effort as the single best effort in a set of efforts per effort type (best FVC-pre, best FVC post, best SVC). This ATS-recommended method uses the effort with the highest sum of FVC + FEV1, or the effort with the highest SVC value. (For details, see the
document noted in reference 5.)
Best Composite Defines best effort as a composite of the highest parameter values across all selected efforts.
Default value for the volume of the calibration syringe. Select the Syringe
Volume from the list.
Check this box to receive a calibration reminder pop-up daily, weekly or monthly.
Determines the unit of Pressure. Check the preferred unit.
Determines the unit of Flow on the axis of the graph, possible options are
L/s or L/m.
Determines the unit of Temperature, possible options are °C or °F.
Temperature
5.2 Viewing Tab
To display the Viewing tab:
The following screen appears:
DIR 80012334 Ver. C 24 / 95
CardioPerfect Workstation
Figure 5.3 Spirometry Viewing tab
SpiroPerfect Module – User Manual
Setting Description
Parameters in
Columns or Rows. Changes the layout of the six parameter table.
Show
Predictive
Points and/or Curve.
If Points is checked, predictive points display and print in the FVC graph,
Predictive points definition see page 93.
If Curve is selected a prediction curve will be displayed in the FVC graph.
Trending
Manual selection of best effort
Display ATS
Acceptability per effort
Superimpose
FV Curves
% Reference Value or % Predicted. When % Reference Value is selected, parameters values are graphed as a percentage of the selected reference value. When % predicted is selected, parameters will trend as a percentage of predictive values.
If checked, you are allowed to manually select the best effort, when the
Final Result is set to Best Effort.
If checked, a row or column appears in the Parameter and Measurement tables displaying whether or not each individual effort meets the ATS 2005 acceptability criteria.
If checked, curves are offset on the graph. If unchecked, all curves are superimposed.
DIR 80012334 Ver. C 25 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
Display Lung
Age
Effort Label
If checked, the estimated Lung Age will be shown while viewing a test and
in the printed reports for patients of 20 years or older. For details, see Lung
Time or Number. If Time is selected, each effort is labeled with the time it was recorded. If Number is selected, each effort is labeled with a number and stage. For example, FVC Pre3 means it is the 3rd effort of a FVC test.
X Axis Position
Table Color
Scheme
Bottom or Top. If Bottom is selected, spirograms are displayed with the horizontal axis at the bottom of the graph. If Top is selected spirograms are displayed with the horizontal axis at the top of the graph.
Defines the background color and font type and color of the Spirometry module. The default setting is Welch Allyn. To customize the settings select User Defined from the drop-down menu.
Scheme Editor
Select the User Defined option from the Table Color Scheme drop-down menu. Once selected, the Scheme Editor button becomes highlighted.
Click on the Scheme Editor button. The Styles properties editor dialog box appears. You can customize the properties for the Spirometry module in the Styles properties editor dialog box.
5.3 Parameters Tab
To display the Parameters tab:
The following screen appears:
DIR 80012334 Ver. C 26 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
Figure 5.4 Spirometry Parameters tab
Select parameters for three categories:
Setting Description
Measurements
Parameters selected in the Measurements column are displayed in the
Measurements Tab of the SpiroPerfect module.
Six parameters
Parameters selected in the Six parameters column are displayed in the six parameters table of the module’s Parameter area. A maximum of six parameters can be selected per test type. For FVC, a minimum of three parameters is required.
Printer
P arameters selected in the Printer column are printed on the reports.
WARNING
Selecting more than 15 parameters for printing may result in the list of printed parameters being truncated on all but the
Best FVC report.
Parameters Measured
FVC testing
FVC
FEV2
FEV1%
PEF
FEF75-85
FIVC
FEV3
FEV1/FVC
FEF25
PIF
FIV1
FEV5
FEV2%
FEF50
FIF50
FIV1%
FEV6
FEV3%
FEF75
FEF50/FIF50
FEV0.5
FIV0.5
FEV5%
FEF0.2-1.2
FEV1/FEV6
FEV1
FEV0.5%
FEV6%
FEF25-75
FET
DIR 80012334 Ver. C 27 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
MEF25
SVC testing
SVC
MV
MEF50
ERV
Tin
MEF75
IRV
Tex
VT
Tin/Tex
IC BF
MVV testing
MVV MV VT BF DFRC
Selection
None
All
Inverted
Default
Blanks all previously selected parameters boxes in the Measurements and
Printer columns. It does not affect the Six parameters column.
Selects all parameters in the Measurements and Printer columns. It does not affect the Six parameters column.
Deselects the selected parameters, and selects the parameters that are not for the Measurements and Printer columns. It does not affect the Six parameters column.
Selects factory Default parameter selections for the Measurement, Six parameters and Printer columns.
5.4 Printing Tab
WARNING
To display the Printing tab:
A color printer is recommended for printing Spirometry reports.
Printing these reports with a black and white printer can lead to confusion as it is not easy to identify which curve is a Pre and which is a Post effort.
The following screen appears:
DIR 80012334 Ver. C 28 / 95
CardioPerfect Workstation
Figure 5.5 Spirometry Printing tab
SpiroPerfect Module – User Manual
Setting
Default report templates
Show on
Report
Description
A list of available templates used for printing reports.
To print multiple reports, select the preferred formats from the list.
Unconfirmed If checked, Unconfirmed is printed on the reports if the test is not yet confirmed.
Confirmed By If checked, Confirmed By is printed on the reports. It provides a space for the clinician’s signature.
Print in color
If checked, the spirometry reports are printed in color when using a color printer.
Graph scaling
Select the type of scaling (graph resizing) to use when printing volume/time curves.
Fixed scale (volume 10 mm/L, time 20 mm/sec, flow 5mm/(L/s))
Auto scale—both x and y-axes (volume and time) scale automatically.
DIR 80012334 Ver. C 29 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
To display the Recording tab:
The following screen appears:
Figure 5.6 Spirometry Recording tab
DIR 80012334 Ver. C 30 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
Setting
Sensor Type
Description
Select Spiro Perfect VCT-400 or Welch Allyn SpiroPerfect.
Connected To
Select the COM port to which the Spiro Perfect VCT-400 is connected.
Select the port to which the Welch Allyn SpiroPerfect sensor is connected.
The port for the Welch Allyn SpiroPerfect sensor with USB connector is automatically detected.
Calibration protocol
Select the Single calibration stroke protocol (this is recommended for the
Welch Allyn Spiro Perfect)
or
Select Three calibration stroke protocol (1, 2 and 3 seconds flow)
Calibration
Error Tolerance
Incentive file
Select Social Security (1%) for increased accuracy required by US Social
Security Administration guidelines
or
Standard (3%)
Select the file that is used for the incentive screen for testing the pediatric population.
Sensor calibration information
Enter the Lot code and the Calibration code and Confirm. For more
information see page 34. If the Spiro Perfect VCT 400 is selected this area
is not applicable.
5.6 Customize the spiro.txt file
Statements used in the Comment editor can be customized. Please refer to the Workstation manual for general instructions on editing this file.
Medication list
In addition to the pre-defined comment and interpretation statements, this file also contains the medications shown in the medication list. These items are immediately followed by an asterisk (*) in the spiro.txt file.
Note: If no spiro_cmt.txt file is available the spiro.txt file is used.
DIR 80012334 Ver. C 31 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
6 Ambient Settings /Temperature, Humidity and Pressure
Adjust the Ambient Settings (the temperature, humidity and air pressure) before calibrating the flow sensor.
CAUTION
Adjust ambient settings before calibrating the flow sensor. If the ambient settings are not adjusted before calibration, the device will not be properly calibrated and could give false readings.
You must recalibrate if there is a significant change in the ambient settings.
Ambient settings are stored locally by the program and passed on to the flow sensor before each measurement. This means that when using different PC's with the same flow sensor you have to set the ambient settings on each PC before starting the measurements. Also, when another person logs in to the PC, he/she needs to enter the ambient settings.
6.1 Why the Workstation needs Ambient Setting
Information
Ambient settings information is necessary for calculating the Ambient Temperature Pressure
Saturation (ATPS) to Body Temperature Pressure Saturation (BTPS) correction in the Flow sensor.
6.2 When to Adjust the Ambient Settings
Adjust the ambient settings:
Daily, the first time logging into the Spirometry module.
When ambient settings have changed significantly during the day.
When the same flow sensor is used on different computers. In this case, adjust the ambient settings on each computer.
Before a calibration takes place, in the pre-calibration window.
6.3 Adjusting the Ambient Settings
1. Make sure the SpiroPerfect module is loaded.
2. Press F9 or choose Ambient settings from the Tools menu.
The following screen appears:
DIR 80012334 Ver. C 32 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
Figure 6.1 The Ambient settings dialog box
1. Enter the Temperature value. (The value for the ambient temperature.)
Tip:
2. Enter the Humidity value. (The value for the ambient air humidity.)
3. Enter the Pressure value. (The value for the ambient barometric pressure.)
In the spirometry settings, the ambient units for temperature and pressure can be changed.
Tip:
There is an additional option available to update the Ambient settings:
1. Select Calibrate located on the Toolbar (or press F10)
2. Enter the ambient setting information in the Pre-Calibration dialog box. Updating ambient settings is recommended when a calibration is going to be performed.
DIR 80012334 Ver. C 33 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
7 Calibration of Flow Sensor
The American Thoracic Society and Welch Allyn recommend
CAUTION
calibrating spirometers every day before use
Welch Allyn guarantees accurate calibration only with the use of a Welch Allyn 3L calibration syringe. Although SpiroPerfect provides other calibration syringe volumes for use, Welch
Allyn is not responsible for the system’s accuracy if these syringes are used.
Flow Transducers
Flow Transducers are manufactured to high precision and it is not necessary to calibrate the spirometer system with each Flow Transducer separately.
CAUTION
Consult Accompanying Documents
Perform a new calibration when using a new lot of Flow
Transducers.
7.1 Preparing calibration
Calibration Protocol
SpiroPerfect supports two calibration protocols:
Single Stroke Calibration
Three Stroke Calibration
The calibration protocol can be set on the Recording tab in the Spirometry Settings. See
To calibrate the Welch Allyn SpiroPerfect flow sensor it is strongly recommended to use the
Single Stroke Calibration Protocol while calibrating. This method will increase the accuracy of the flow sensor. To calibrate the Spiro Perfect VCT-400 use the Three Stroke Calibration
Protocol for the best results. The protocol can be changed in the spirometry settings.
Warm up the Spirometer
Before calibrating it is recommended to let the Spirometer warm up. If the Spirometer has already been used shortly before calibration, this warm up period is not needed.
1. Connect the Spirometer to the computer.
2. Open the Spiro module.
The sensor starts to warm up as soon as the spirometer module is opened.
3. Wait for at least 5 minutes before starting the calibration process.
7.2 The Calibration Process
Make sure the Spirometer is plugged in before continuing.
1. To start the calibration, select the Calibrate button located on the Toolbar (or press
F10). The following dialog box will be presented to configure the calibration process.
DIR 80012334 Ver. C 34 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
Figure 7.1 The Pre-calibration dialog box
Fill in the appropriate settings. For a description of the options see the following table.
Setting
Lot code
Description
Enter the Lot code for the Flow transducers located on the box the
Transducers came in.
Calibration code
Enter the Calibration code for the Flow transducers located on the box the
Transducers came in. See example of a schematic cutout of the label from the Flow transducers box below.
Note: the sensor calibration information can also be set in the
Spirometry settings (recording tab). Please make sure that Lot code and
Calibration code are still accurate before calibrating.
Vica test:
If the VCT400 was selected the Lot Code and Calibration Code options are not applicable and grayed out.
Syringe Volume Select the appropriate Syringe Volume.
Tip: See page 22 on how to change the default setting.
Current
Calibration
Factor
This value cannot be changed and indicates the correction factor applied to the calibration data from the previous session. Once the calibration is performed, this value will be updated. The factor displayed is the average of the inspiration and expiration calibration factor.
DIR 80012334 Ver. C 35 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
Setting
Calibration
Error Tolerance
Temperature
Humidity
Description
Select the measured calibration accuracy to be within 1% or 3% of the syringe volume.
Tip: See page 30 on how to change the default setting.
The ambient temperature
See Operating Environment Specifications on page 80
The ambient humidity
See Operating Environment Specifications on page 80
Pressure
The ambient pressure
See Operating Environment Specifications on page 80
WARNING
For the ambient settings pressure field please enter the pressure as given by a barometer in the immediate vicinity.
Do not enter the normalized sea-level pressure as commonly listed on internet sites on meteorological data resources.
2. Press the Next button to continue.
The following screen appears depending on your setup:
Figure 7.2 Calibration window
Beneath the Calibration window the calibration instruction and message window is displayed, giving instruction for the calibration procedure.
Note: Please check the calibration and lot code if you cannot continue to the
Calibration window.
3. Connect the syringe to the new flow transducer.
4. Fill the syringe by pulling the plunger completely out.
6. Wait until the messages ‘Initializing sensor. Opening sensor, please wait...’ disappears.
DIR 80012334 Ver. C 36 / 95
Figure 7.3 Start Calibration
CardioPerfect Workstation
SpiroPerfect Module – User Manual
7. Verify the syringe is completely filled and press the OK button.
Note: If the syringe was emptied before calibration, the “No valid stroke recorded”
message will appear.
8. Follow the instruction on screen. The blue calibration bar can be used as a guide line by giving you an indication of the speed.
Figure 7.4 Calibration bar
For a single flow calibration protocol
Push the plunger entirely in and pull the plunger out as far as possible, while following the blue bar as closely as possible. The calibration procedure will stop automatically, and inform you of the results.
For multiple-stroke calibrations
Push the plunger entirely in and pull the plunger out as far as possible, three times, while following the blue bar as closely as possible.
At the end of each stroke a message appears; you can either choose to accept the calibration stroke or redo the last stroke.
DIR 80012334 Ver. C 37 / 95
Figure 7.5 Accept stroke
CardioPerfect Workstation
SpiroPerfect Module – User Manual
The following options apply:
Yes:
Continue to the next stroke, or show calibration results.
No:
Redo the current stroke with the same speed.
Cancel: Stop calibrating. The sensor will not be calibrated.
9. If the calibration was successful this will be displayed. You can either accept the results or recalibrate. If the ATS standard has not been met you have to recalibrate.
See the following section on how to deal with the results.
If you have pressed the Recalibrate button: Press the Repeat Calibration button in the Calibration window and follow the instructions above from step 6.
Note: If you have trouble getting the results within the Calibration Error Tolerance try:
waiting 1 second between emptying and filling the syringe.
a single flow calibration before the three flow calibration, if the three flow
calibration is unsuccessful
DIR 80012334 Ver. C 38 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
7.3 View Calibration Results
After calibration the Verify Calibration Results window will appear.
Figure 7.6 Verify Calibration Results
(|Exp.%| + |Insp.%|)/2=Abs.Avg.%
Note:
Abs.Avg% should be:
-
<1% to meet the social security standard
-
<3% to meet ATS standard
Each row gives the result of a stroke, the last row gives the Averages. The following columns are given:
L/s Results per stroke:
The speed in the L/s plunger should be moved, as indicated by the blue
Exp. Vol
Exp.%
calibration bar during the strokes.
The expiration volume reached by pushing in the plunger.
The expiration deviation from the actual volume in percent.
Insp. Vol
The inspiration volume reached by pulling out the plunger.
Insp.%
Avg.Vol
The inspiration deviation percentage
The averages of inspiration volume and expiration volume.
Abs.Avg.% The absolute average deviation percentage between inspiration volume and expiration volume.
The Averages row gives the averages reached for all strokes. Please refer to this row for improving the calibration results.
After you have accepted the results you can view these values in the table below the graph by selecting the Calibrate Results button.
DIR 80012334 Ver. C 39 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
Figure 7.7 Single stroke calibration window with calibrated results
Figure 7.8a Single stroke calibration
Figure 7.8b Three stroke calibration
Calibrated results table:
Selection Description
Calibrate
Press the "Calibrate" icon when you are ready to start the calibration process. After one calibration, the icon label changes to ‘Repeat
Calibration’. It is recommended to repeat the calibration more than once.
Calibration
Results
Calibration
Logs
Print Log
Press the Calibration Results icon to view the results of the calibration.
This can only be viewed between calibration attempts.
Press the Calibration Logs icon to view previous calibration attempts.
Press Print log icon to print the currently selected or displayed log.
Print Preview
Done
Press Print Preview.
The Print dialog box appears.
Press OK on the Print dialog box to the view the calibration results before printing.
Press Done to exit the Calibration window.
7.4 Error Messages Associated with Failed Calibration
Figure 7.9 No valid stroke recorded
When the calibration attempt was not valid or if the volume read back by the sensor is not within 35% of the selected syringe volume, the calibration fails.
This message also appears if the calibration attempt was performed in the wrong order, first emptying the syringe instead of filling it before starting the calibration.
DIR 80012334 Ver. C 40 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
Caution
It is the user’s responsibility to determine whether to accept or reject failed calibration data. If the device does not pass calibration it could give false readings.
Note for Spiro Perfect VCT-400 users:
If the following message appears after a failed calibration, turn the Vicatest sensor off and on again and proceed with a new calibration.
Figure 7.10 Old calibration factors warning message
7.5 Calibration log
Figure 7.11 Calibration log window
Use the Calibration log to view calibration information of current and previous calibration efforts. Each time the sensor is calibrated, results are stored in the calibration log. Select a calibration effort from the list to see the curve that belongs to it.
Calibration Log
Selection Description
Date & time
Date and time of the calibration.
User
Location
Name of the user that performed the calibration.
Location specified in the general settings.
DIR 80012334 Ver. C 41 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
Device ID
Insp. CF
Insp. SD
Exp. CF
The spirometry sensor hardware used.
The Calibration Factor of the inspiratory strokes.
The stroke difference between the inspiratory strokes.
The Calibration Factor of the expiratory strokes.
Exp. SD
Calibrated
The stroke difference between the expiratory strokes.
A check mark shows if the sensor was actually calibrated (yes) or only a log entry was saved (no).
To view the calibration log:
1. Choose Tools
2. Select Calibration log
DIR 80012334 Ver. C 42 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
8 Recording Spirometry Tests
Various types of efforts can be recorded with the Spirometer module:
FVC: Forced Vital Capacity.
MVV: Maximum Voluntary Ventilation.
SVC: Slow Vital Capacity.
The following tags can be assigned to each effort:
Pre
Post
When recording a post stage effort the medication administered to the patient can be entered.
8.1 Record a Spirometry Test
Follow these steps to record a test.
1. In the Workstation, find or create a patient (see the Workstation manual for instructions).
located in the toolbar at the top of the screen.
Tip:
The following screen appears if a new effort or test is being added to the patient’s profile within 24 hours since the last test or effort.
Figure 8.1 Warning window
Caution
Use numerals only to set the date format. Alpha characters cannot be used in the date field.
DIR 80012334 Ver. C 43 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
Figure 8.2 New spirometry test window
3. Complete Patient information fields in the New spirometry test window. Check the
Smoker and/or Asthmatic boxes if applicable.
4. Select the Specialty and Referring physician, by whom the test was ordered.
5. Select the Prediction Norm for the test.
Caution
To obtain predictive values for certain parameters, the patient’s age, gender, race and height must be entered into the Patientcard dialog box (Choose Edit>patientcard or Alt+P), otherwise no predictive data is reported. The patient’s weight is only obligatory for certain prediction norms.
Note: If patient data is missing these will be displayed in red in the
New spirometry test window. You must fill in the blanks before you
are able to continue.
The Norm Profiles (see section 12.1) indicate valid demographic
ranges for each norm.
6. Select Ambient Settings. If the humidity, temperature or pressure have changed since the last calibration, adjust as necessary.
DIR 80012334 Ver. C 44 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
7. Select Next
The following screen appears:
Tip:
Double click on the left graphic to expand it to full window size.
Double click again to reduce it to half window size.
Figure 8.3 Recording window
8. Select the effort type to perform by selecting the FVC, SVC or MVV.
9. Select the effort stage. If you have selected Post, enter the medication dosage and unit.
Note: The medication and dosage fields are only active if a post-effort is selected.
Post-effort is only available after a pre-effort has been recorded.
10. Select the type of curve from the drop-down menu located at the bottom of the right graph.
Figure 8.4 Type of Curve menu
11. Instruct the patient to hold the SpiroPerfect sensor still.
Note: Make sure the rear of the flow-tube is not blocked. The extra resistance will result in faulty measurements.
12. Select Record to start recording.
13. Ask the patient to perform the effort according to the appropriate procedures. See
14. When the patient has completed the test, select Done. The recording window closes and the main view displays all efforts of the recorded effort stage (Pre/Post).
Note: The effort along with six corresponding parameter values are displayed in the parameters area.
DIR 80012334 Ver. C 45 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
15. The status bar of the Recording window displays, ATS acceptability criteria met, duration of the effort, number of FVC, SVC and MVV efforts completed in a test and if the reproducibility criteria are met.
16. When the patient finishes testing, select Done. The Spirometry view appears displaying all the efforts.
8.2 Incentive Screen
The incentive screen is used to encourage pediatric patients to blow into the flow transducer the best they can.
To display the Incentive screen:
Select the Incentive button from the Recording/Test toolbar
or
Select Incentive from the Type of Curve drop-down menu
The following screen appears
Figure 8.5 Recording window with incentive screen selection
Note: Incentive screen
The fireman extinguishes the fire if the patient’s effort reaches 80% of the predicted for PEF & FVC values. If the patient’s effort is below 80%, the fire is not extinguished.
To remove the Incentive screen:
Select Volume/Time or Flow/Time from the Type of Curve drop-down menu or select the
FVC, SVC or MVV button.
Note: If the patient’s demographics are outside of the Prediction norm demographics no prediction values will be calculated. The incentive screen will not operate without predicted values, but will be visible.
DIR 80012334 Ver. C 46 / 95
CardioPerfect Workstation
8.3 Patient Procedures
SpiroPerfect Module – User Manual
WARNING
Patients may become faint, light-headed, dizzy, or short of breath during spirometry testing. Watch patients closely. If they choose to stand during testing, keep a chair immediately behind them. If there is any reason for concern, stop the test and take proper action.
Recommendations
Practice the procedure with the patient before recording the effort.
American Thoracic Society recommends ending recording after eight successful FVC efforts to avoid fainting.
To prepare patients for any spirometry test, explain the entire procedure for the type of effort you want them to perform. Remind patients that the test is painless. Demonstrate at least one effort for the patient.
The accuracy of a spirometry test is highly dependent on the patient's understanding and cooperation. So, be prepared to coach and encourage the patient with your “body language” and your words — for example, ”Blow, blow, blow, keep blowing until you can't blow out any more” — to ensure a good effort with reproducible results.
Instruct patients to do the following:
Loosen any tight articles of clothing that might constrict lung function, for example, a tight belt, tie, vest, bra, girdle, or corset.
Remove any foreign objects from the mouth, including loose dentures.
The use of a nose clip is highly recommended. If used, check for proper fit.
Place lips and teeth around a new flow transducer, sealing their lips tightly around the transducer. Grip slightly with teeth in the groove.
Keep tongue away from the flow transducer to avoid blocking it.
Keep the rear of the Flow sensor free.
Keep chin up so as not to restrict the airway.
Once the patients have the flow transducer in place, ask them to perform the effort using the guide below for the patients’ predicted performance effort breathing instructions.
Note: Place the mouthpiece in the patient’s mouth after stabilization.
For an FVC effort, instruct the patient to:
1. Breathe in (until the Total Lung Capacity is reached).
2. Blow out forcefully (until the Residual Volume is reached). Allow sufficient time.
DIR 80012334 Ver. C 47 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
For an FVC loop, instruct the patient to:
1. Breathe in (until the Total Lung Capacity is reached).
2. Exhale forcefully (until the Residual Volume is reached)
3. Breathe in forcefully (until the Total Lung Capacity is reached). Allow sufficient time.
-or-
1. Start normal breathing (tidal breathing).
2. Breathe out (until the Residual Volume is reached).
3. Breathe in forcefully (until the Residual Volume is reached). Allow sufficient time.
4. Exhale forcefully (until the Total Lung Capacity is reached).
For an SVC effort, instruct the patient to:
1. Start normal breathing (tidal breathing).
2. Breathe in calmly, (until the Total Lung Capacity is reached).
3. Exhale calmly, (until the Residual Volume is reached). Allow sufficient time.
4. If necessary, repeat steps 3 and 4.
Steps 3 & 4 can be reversed, meaning: a maximum expiration followed by a maximum inspiration.
For an MVV effort, instruct the patient to:
Breathe in and out forcefully at a pace of approximate 30 breaths per minute (2 seconds per complete breath) for 15 seconds (the program automatically stops gathering data after 15 seconds).
About Quality Feedback
The spirometer provides two kinds of quality feedback: effort-quality messages and test quality grades, as described in the following sections.
About Effort-Quality Messages
One of the following effort-quality messages appears on the screen after each effort is completed. These messages indicate whether an effort was acceptable and reproducible, and if not, what the patient needs to do differently.
The term “match” here means “variation” or “difference with respect to best test.
Effort-Quality Message Criteria
Don’t hesitate Back-extrapolated volume > 150 mL or 5%, whichever is greater.
Blast out faster
Blow out longer
Good effort
PEF time > 120 ms.
FET < 6.0 seconds, and end-of-test volume > 100 mL (invalid
FEV6).
Effort meets above criteria.
> 25 mL within the last second of exhalation No plateau
8.4 Deleting an Effort
You can easily delete an effort after recording it.
To delete an effort:
Option 1: In the Recording Test window
DIR 80012334 Ver. C 48 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
The following screen appears
Figure 8.6 Confirmation dialog box
Option 2: While viewing the test
Highlight the effort to delete in the Parameters area, located in the lower right side of
the of workspace window. See Figure 4.1 Main Window
1. Select Action> Delete Effort or Ctrl+D, located on the menu bar, the confirmation
dialog box appears. See Figure 8.6.
8.5 Add or Change Information in the Comment Editor
When creating a new spirometry test, the SpiroPerfect offers space for adding or changing comments, while recording.
To add or change comments:
1. Choose Patient, and start a new Spirometry test
3. Select the Comment button from the toolbar.
Note: The Comment editor is displayed containing previously added comments.
4. Select interpretations and or medication from the statement tree on the left side, or type in comments in the comment pane.
5. Select Save
The Recording/Test window will appear again.
The comment editor is also available form the menu bar, select Edit comment from the Action menu or type CTRL+T.
DIR 80012334 Ver. C 49 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
9 Viewing Spirometry tests
9.1 View a Spirometry test
To view a spirometry test:
1. Select a patient. The patient’s previously recorded tests appear in the test list.
2. From the test list, select a spirometry test to view.
Note: Spirometry tests are indicated with a .
3. SpiroPerfect launches and the test is displayed in the workspace.
4. Use the tabs and the Effort selector (in the toolbar) for selecting information to view.
Figure 9.1
Patient list database
9.2 Setting the Best Effort
Follow these steps for setting the Best Effort:
1. Choose File
2. Select Settings > Spirometry
3. Select the General tab
4. Under the Final result, check Best effort
5. Select the Viewing tab
6. Check Manual selection of the best effort
Note: This action is not available if “Best composite” is set as Final result in the
Spirometry General Settings.
7. Select Ok. The Spirometry settings window closes
8. Set the effort selector to pre or post, see Figure 9.3.
9. Select the pre effort you consider best.
10. Next, from the menu bar, select Action> Set Current Effort As Best
11. Repeat steps 8-10 for the post effort selection.
9.3 View and Add Information to a Test
To view and/or add information to a test:
Select Tools > Information
The following screen appears
DIR 80012334 Ver. C 50 / 95
Figure 9.2 Information dialog box
CardioPerfect Workstation
SpiroPerfect Module – User Manual
To enter comments:
1. Type comments in the Comment section
9.4 Test Modes and Tabs
There are four views available from the Effort selector in the toolbar:
Figure 9.3 Effort selector
All Efforts:
Pre:
Post:
Final/Best result:
view and compare all efforts of current test. view and compare only the pre efforts of current test. view and compare only post efforts of current test. view and compare only the best effort/ final result of current test.
DIR 80012334 Ver. C 51 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
Figure 9.4 Five tabs
Tab Description
FVC
Select to view only the currently selected FVC efforts
A flow/volume curve of the current FVC effort and the flow/volume curves of all selected FVC efforts. The dotted line marks the predicted values.
SVC
MVV
Select to view only the currently selected SVC efforts.
A spirogram for the current SVC effort.
Select to view only the currently selected MVV efforts.
Trend
A spirogram for the current MVV effort.
Select to view only the trends to a maximum of six parameters.
Trends of the FVC effort of the test.
Measurements Select to view all the parameters calculated of all effort types.
All parameters values based on user settings for each stage and effort.
DIR 80012334 Ver. C 52 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
9.5 Common Features for each Tab
Figure 9.5 Tab overview
View multiple flow/volume curves of one test
It is possible to view and compare multiple efforts previously recorded in one test. The right hand side of the window displays a flow curve of all selected efforts. The left hand side of the window displays the flow curve of the currently selected effort.
To view multiple efforts in one flow/volume graph:
1. In the Spirometer window, move the mouse arrow to the Parameters table.
2. Check the box of each effort to view in the Selected row/column.
3. Uncheck the boxes of each effort to hide it from view.
DIR 80012334 Ver. C 53 / 95
9.5.1 Parameters Area
Figure 9.6 Parameters Table
CardioPerfect Workstation
SpiroPerfect Module – User Manual
Effort
Selection Description
Parameters area
The Parameters area holds the parameters table. It is displayed under the
FVC, SVC MVV and Trend tabs.
The Parameters table lists up to six user-defined parameters. See page 26
to select the parameters.
The following information is displayed in the table:
predictive norm
ATS acceptability criteria predictive values per parameter effort stage & parameter value
final result parameter values
% predictive
% change (in “All efforts” and “Post”view)
The color in front of the effort name corresponds to the color of the curve in the graph.
Select the check box in the selected row or column and the curve is displayed in the graph. You can select to show the parameters in rows or
columns in the settings menu>viewing tab, see page 24.
Deselect a check box and the curve is hidden.
9.5.2 Interpretation Area
The interpretation area displays automatic or confirmed interpretation, medications,
comments, lung age (if enabled in the settings), and reproducibility information. See page 61
for more information.
DIR 80012334 Ver. C 54 / 95
Figure 9.7 Interpretation Area
CardioPerfect Workstation
SpiroPerfect Module – User Manual
9.6 FVC Tab
Figure 9.8 FVC Tab
Selection Description
Left Graph
The left graph always represents the selected effort in the parameters table as a flow volume loop or as tidal volume.
Right Graph
The right graph displays curves all efforts selected for a particular stage.
Different curve views are selected from the drop down menu:
Flow/Volume
Volume/Time
Flow/Time
Axes
Units
In a flow volume graph, the flow is plotted against the volume.
In a volume/time graph, the volume is plotted against time.
Volume is expressed in liters.
Time is expressed in seconds.
Flow is expressed in liters per second, or liters per minute based on settings.
DIR 80012334 Ver. C 55 / 95
9.7 SVC Tab
Figure 9.9 SVC Tab
CardioPerfect Workstation
SpiroPerfect Module – User Manual
SVC Test
Review results under the SVC tab.
Only volume/time (spirogram) graphs are displayed along with six SVC parameters.
Calipers mark the beginning and the end of the tidal area. Each effort line displayed on the graph has a different color.
Note: Calipers can be manually adjusted. If so, affected parameters will automatically be recalculated.
If no SVC test is performed, the SVC tab is disabled.
DIR 80012334 Ver. C 56 / 95
9.8 MVV Tab
Figure 9.10 MVV Tab
CardioPerfect Workstation
SpiroPerfect Module – User Manual
MVV Test
Review the results under the MVV tab.
Only volume/time (spirogram) graphs are displayed along with six MVV parameters.
Calipers (vertical lines) mark the beginning and the end of the ventilation volume (not the tidal area).
Note: Calipers can be manually adjusted. If so, affected parameters will automatically be recalculated.
Tip: You can select or deselect a curve in the parameters table.
If no MVV test is performed, the MVV tab is disabled.
DIR 80012334 Ver. C 57 / 95
9.9 Trend Tab
Figure 9.11 Trend Tab
CardioPerfect Workstation
SpiroPerfect Module – User Manual
The Trends tab displays trends of:
the FVC efforts of the test or
the best pre and best post efforts of the several different tests of the same patient.
There is no limit to the number of tests that you can trend. You can simultaneously view three parameters and see how the parameters evolve during the test.
The interpretation area shows the interpretation of the most recent test.
Figure 9.12 Parameters menus
Selection Description
Parameters
Three parameters are always trended. The choice of parameters trended depends on the selected parameters in the settings (File> Settings>
Spirometry> Parameters tab> Six parameter column). When you exit the trend view, SpiroPerfect remembers the last three parameters selected and recalls them when you enter the trend view again.
Axes
The horizontal axis displays the date & time of the efforts.
The vertical axis displays the parameter values:
As a % of Predictive (effort x/pred value) x 100.
As a percentage of a reference value. The value of the parameters is a relative value. It depends on the currently selected effort. For example, if a test has three efforts and effort 1 is selected, all parameter values for effort 1 are set to 100%. The values for other efforts are expressed as lower or higher percentages in relation to effort 1. (Effort x / Effort 1) x 100.
Example with Effort 1 currently selected
Real value of FEV1%
Relative value of FEV1% as displayed in trends
Effort 1
3.49
100%
Effort 2
3.70
106%
Effort 3
3.77
108%
Effort 4
3.46
99%
DIR 80012334 Ver. C 58 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
To view trends:
1. Select three parameters from the drop down menus located beneath the Trends graph. The curve for these parameters is displayed in the Trends graph.
2. Select or deselect efforts by checking the boxes beneath the efforts listed in the parameters area. The efforts are added or removed from the Trends graph.
3. Select which effort is used as the reference point, by clicking on one of the efforts located in the parameters area. The parameter value of this effort is set to 100% in the Trend graph, and the parameter values of all other efforts are expressed as a percentage proportional to the reference value. The percentage of deviation is given in the parameters area.
Figure 9.13 Parameters Area
4. You can select a line from the Trends graph by clicking on one of the points in the line. The percentage deviation for each point is shown when you move the mouse over a point in the line. By clicking on a different colored point you select the line of that color and the percentage deviation for that line is displayed when moving over the points of that line.
9.10 Measurements Tab
Figure 9.14 Measurement Tab
DIR 80012334 Ver. C 59 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
The Measurements Tab contains a number of parameter values for each FVC, SVC and MVV efforts. Each effort is represented by a separate column.
The measurement table only displays the efforts belonging to the selected stage. It only
lists the parameters selected in the settings. See page 26 to select parameters for display
in the Measurement table.
Information on test reproducibility is displayed in the reproducibility table that is placed below the Measurements Table. In particular, the FVC and FEV1 absolute value variance
(difference) between the best effort and the second best effort is analyzed for both pre and post tests.
When a value is in
bold red text
, that value is below the lower prediction limit.
9.11 Compare Tests
With SpiroPerfect, you can compare final results from different tests recorded for the same patient.
Selecting various tests
The following screen appears
Figure 9.15 Comparison dialog box
Click on the check box in front of each test to select tests. The best pre and best post efforts of the patient’s selected tests are compared.
Available views are:
FVC
Measurements
Trend
The trend view displays a graphical overview of the patient’s performance over time.
DIR 80012334 Ver. C 60 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
10 Interpreting Spirometry Tests
The Spirometer module can automatically interpret FVC efforts.
WARNING
A computer generated interpretation cannot replace sound medical reasoning by a trained professional. Therefore, a physician should always review the interpretation.
10.1 Editing and Confirming an Interpretation
Figure 10.1 Interpretation editor
Interpretation edit area
Comment pane
Statement Tree
In the Interpretation editor, text or interpretation statements are added to the interpretation area. Once an interpretation is edited, confirm it. Otherwise, the edits are not saved.
Opening the interpretation editor
Choose Tools and select Interpretation. or
Select Interpretation on the Toolbar, see Figure 4.1 Main Window
Automatically generated interpretation:
The generated interpretation is shown in the interpretation editor automatically if the interpretation is unconfirmed. You can keep this interpretation and add text to it or replace it.
The automatic interpretation statements can be inserted by clicking the Best effort FVC interpretation button.
Confirming an interpretation and closing the interpretation editor:
Select Confirm to save your comments and to return to the spirometry window.
Adding comments to the interpretation
Click in the comment pane and start typing the comment.
Adding text to the interpretation edit area
Click in the interpretation edit area and start typing the text.
DIR 80012334 Ver. C 61 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
Adding an interpretation statement to the interpretation edit area using the statement tree
1. Select a category to display the statements.
2. In the statement tree look up the statement to include in the interpretation.
3. Click on the statement to add it to the interpretation edit area.
Deleting an interpretation statement from the interpretation edit area
Select the statement text and press BACKSPACE or DEL to delete it.
Deleting a comment from the comment pane
Select the comment and press BACKSPACE or DEL to delete it.
Tips for editing and confirming an interpretation
Automatically insert the current date and time by selecting the Date/time button.
Clear the interpretation editor by selecting the Clear button.
The statement tree can be changed. Please consult your system administrator or local dealer for new or changed statements.
10.2 Automatic Interpretation
The spirometer module automatically calculates interpretive results as described in the
document noted in reference 2 on page 73.
The automatic interpretation is shown in the interpretation area if the interpretation is not confirmed. If the interpretation is confirmed the confirmed interpretation is shown in the interpretation area.
10.3 View Interpretation History
When making changes in an interpretation, the original interpretation is not changed, but a new one is created. A copy of all interpretations is kept in the interpretation history.
DIR 80012334 Ver. C 62 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
Figure 10.2 Interpretation History screen
To view the interpretation history:
1. Choose Tools
2. Select Interpretation history. The Interpretation history window is displayed.
The left hand pane displays the interpretations sorted by date. The right hand pane displays the content of each interpretation, including the date, time and editor.
3. Select a date to view an interpretation.
10.4 Reanalyze a Spirometry Test
Retrieve overwritten automatic interpretations by reanalyzing the spirometry test.
To reanalyze a spirometry test:
Choose Actions
Select Reanalyze test.
Reanalyzing the test will result in the following;
A new interpretation is appended to the test containing the automatic interpretation statements.
The state of the interpretation is set to unconfirmed.
All parameter values are re-calculated.
10.5 Recalculate prediction
With this option you can re-calculate the predicted values for the test with a different
Prediction norm.
To recalculate a prediction:
1. Go to the Action menu
2. Select Recalculate Prediction.
3. Select the preferred Prediction norm from the list.
DIR 80012334 Ver. C 63 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
Note: for a more elaborate description of the Prediction norms see section 12.
4. Press the OK button.
DIR 80012334 Ver. C 64 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
11 Printing Spirometry tests
11.1 Printing reports
To print a particular report for the test currently on your screen:
Select File > Print or
The Print dialog box appears:
Figure 11.1 Print dialog box
Select the desired print report type. The report type corresponding with the current view is already selected. If you should so desire, you can select another report type.
Press the OK button to start printing.
To print multiple reports for the test currently on your screen:
Select File > Print selected formats or
To control which reports will be printed, see page 28.
DIR 80012334 Ver. C 65 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
11.2 Print Report Formats
The SpiroPerfect module prints the following report formats:
All plus best effort FVC
Best FVC
Best MVV
Best SVC
Best Three FVC
Measurements MVV
Measurements SVC
Measurements FVC
Social Security
Trend
Each format contains the patient’s personal information, test information, interpretation, parameter table and all but the measurement report contain a graphs section.
Please refer to the Workstation manual for further information on printing a test.
11.3 Print Preview
To Preview a Test
Select File > Print Preview. The print dialog box appears. See Figure 11.1.
Select type of report to preview.
The name of the report type appears at the top of the dialog box.
The print preview window appears.
Figure 11.2 Print preview dialog box
Name of the report type
DIR 80012334 Ver. C 66 / 95
12 Predictions
CardioPerfect Workstation
SpiroPerfect Module – User Manual
12.1 Norm Profiles
Each predictive norm supports a particular subset of parameters and covers a specific population, as detailed in the profile charts below
Norm Name
FVC
FEV1
FEV1%
FEV0.5
FEV3
FEV3%
FEV6
FEV1/
FEV6
PEF
X X X X X X X X X X
X X X X X X X X X X
X X X X X
X
X
X
FEF0.2-1.2
FEV0.5%
MVV
SVC X
Male
Female
Pediatric
Adult
Height (cm)
Weight (kg)
Caucasian X X X X X X
Black
Hispanic
Native
American
*adult population only
X
X X *
X *
X X
**Caution: Pediatric use in the US > 6 years old
X X X X X
X
X
DIR 80012334 Ver. C 67 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
Norm Name
FVC
FEV1
FEV1%
FEV0.5
X X X
FEV3
FEV3%
FEV6
FEV1/ FEV6
PEF
X
X
X
X
X
X
X X X X X X X X X X
FEF25-75 X X X X X X X X
FEF75 X X X X X X X X
FEF50 X X X X X X X X X
FEF25 X X
FEF0.2-1.2
FEV0.5%
MVV
X
X
Male
Female
Pediatric
Adult
Height (cm)
Weight (kg)
Caucasian X X X X X X X X
Black
Hispanic
X
X
X
Asian
Native
American
**Caution: Pediatric use in the US > 6 years old
X
DIR 80012334 Ver. C 68 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
12.2 Norm-Related Clinical Studies
Each of the following studies provides expected values for various spirometric parameters by measuring significant samples of a specific population.
Berglund
1963
Spirometric Studies in Normal Subjects. I. Forced Expirograms in Subjects 7-70 Years of
Age, Berglund E., et. al., Acta Medica Scandinavica, vol. 173(2): 185-192, 1963.
Crapo
1981
Reference Spirometric Values using Techniques and Equipment that Meet ATS
Recommendations, Crapo RO, et. al., American Review of Respiratory Disease 1981,
123:659-664.
Dockery
1983
Distribution of Forced Vital Capacity and Forced Expiratory Volume in One Second in
Children 6-11 Years of Age, Dockery DW, et. al., American Review of Respiratory Disease
1983, 128:405-412.
ECCS/Quanjer
1993
Falaschetti
2004
Forche II
1988
Langhammer
2001
Hedenström
1986
Hibbert
1989
Hsu
1979
Knudson
1976
Knudson
1983
Koillinen
1998
Kory
1961
Morris
1971
Lung Volumes and Forced Ventilatory Flows: Official Statement of the European Respiratory
Society, Quanjer Ph. H., et. al., European Respiratory Journal, 1993, vol. 6, Suppl. 16: 5-40.
Prediction equations for normal and low lung function from the Health Survey for England,
Falaschetti E., et.al., European Respiratory Journal, 2004, 23: 456-463
Neue spirometrische Bezugswerte für Kinder, Jugendliche und Erwachsene; Forche G.,
Harnoncourt K., Stadlober E.; Österreichische Ärztezeitung 43, 15-16, 1988.
Forced Spirometry Reference Values for Norwegian Adults: The Bronchial Obstruction in
Nord-Trondelag Study, Langammer A., Gulsvik A., et. al., European Respiratory Journal
2001, 18: 770-779.
Reference Values for Lung Function Tests in Men: Regression Equations With Smoking
Variables, Hedenström, H. et. al., Upsala Journal of Medicine Science 91:299-310, 1986.
Reference Values for Lung Function Tests in Females: Regression Equations With Smoking
Variables, Hedenström, H. et. al., Bull. Eur. Physiopathol. Respir. 1985, 21, 551-557.
Lung function values from a longitudinal study of healthy children and adolescents. Hibbert
ME, Lanigan A., Landau LI, Phelan PD, Pediatric pulmonology, 7:101-109, 1989.
Ventilatory Functions of Normal Children and Young Adults—Mexican-American, White and
Black. I. Spirometry, Katharine HK Hsu, et. al., The Journal of Pediatrics; volume
95(1):14-23, July 1979.
The Maximal Expiratory Flow-Volume Curve. Normal Standards, Variability, and Effects of
Age, Ronald J. Knudson, Ronald C. Slatin, Michael D. Lebowitz, and Benjamin Burrows.
American Review of Respiratory Disease, volume 113:587-600, 1976.
Changes in the Normal Maximal Expiratory Flow-Volume Curve With Growth and Aging,
Ronald J. Knudson, et. al., American Review of Respiratory Disease 1983 127: 725-734.
Terveiden suomalaislasten spirometrian ja uloshengityksen huippuvirtauksen viitearvot,
Hannele Koillinen, et. al., Suomen Laakarilehti, 1998, 5 vsk 53, p. 395-402.
The Veterans Administration-Army Cooperative Study of Pulmonary Function. I. Clinical
Spirometry in Normal Men, Kory RC, et. al., American Journal of Medicine, February 1961,
243-258.
Spirometric Standards for Healthy Nonsmoking Adults, James F. Morris, et. al., American
Review of Respiratory Disease, 103: 57-67, 1971.
1999
Polgar
1971
Hankinson, John R. Odencrantz, and Kathleen B. Fedan, Division of Respiratory Disease
Studies, National Institute for Occupational Safety and Health, Centers for Disease Control and Prevention, Morgantown, West Virginia, 1999. The Third National Health And Nutrition
Examination Survey (NHANES III). Am J Respir Crit Care Med Jan 1999; 159:179-187.
Pulmonary Function Testing in Children: Techniques and Standards, Polgar G. and
Promadhat V. Philadelphia, WB Saunders, 1971.
DIR 80012334 Ver. C 69 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
Roca
1986
Schoenberg
1978
Solymar
1980
Viljanen
1981
Wang
1993
Zapletal
1969
Spirometric Reference values from a Mediterranean population, J. Roca, J. Sanchis, A.
Agusti-Vidal, F. Segarra, D. Navajas, R. Rodriguez-Roisin, P. Casan, S. Sans. Bull. Eur.
Physiopathol. Respir. 1986, 22, 217-224
Growth and Decay of Pulmonary Function in Healthy Blacks and Whites, Janet B.
Schoenberg, Gerald J. Beck, and Arend Bouhuys, Respiration Physiology, 1978, 33,
367-393.
Nitrogen Single Breath Test, Flow-Volume Curves and Spirometry in Healthy Children, 7 -18
Years of Age, L. Solymar, P. H. Aronsson, B. Bake, and J. Bjure. European Journal of Respir.
Dis. 1980, 61:275-286.
Spirometric Studies in Non-smoking, Healthy Adults, AA Viljanen, et. al., The Scandinavian
Journal of Clinical Lab Investigation, 41 supplement 159, 5-20, 1981.
Wang X, Dockery DW, Wypij D, Fay ME, Ferris BG Jr., Pulmonary function between 6 and 18 years of age. Pediatric Pulmonology 1993; 15: 75–88.
Maximum Expiratory Flow-Volume Curves and Airway Conductance in Children and
Adolescents, A Zapletal, EK Motoyama, KP Van De Woestijne, VR Hunt and A. Bouhuys,
Journal of Applied Physiology, vol. 26, no. 3:308-316, March 1969.
12.3 Norm Extrapolation
Extrapolation is the practice of applying a norm’s formula to a patient whose profile doesn’t fit that norm’s profile. For example, if you were testing an 88-year-old man, and the primary
(selected) norm was based on males 85 or younger, the predicted values are extrapolated values.
When it takes place, extrapolation is indicated in the test record.
Pediatric norms do not provide any age, weight, or height extrapolation.
Adult norms allow extrapolation of age up, but not down.
Adult norms allow extrapolation of height, weight, up and down.
12.4 Composite Norm Values
When the Composite norm, see tables in section 12.1, is selected, predictive parameter
values are filled in from one of the alternative (composite) norm sources listed here.
NHANESIII FVC, FEV1, FEV1%, FEV6, FEV1/FEV6, FEV6/FVC, PEF,
FEF25-75
Crapo 1981
Morris 1971
FEV0.5, FEV3, FEV3/FVC
FEF0.2-1.2
ECCS/Quanjer 1993 FEF25, FEF50, FEF75
Note: If an adult norm is selected but pediatric patient data is used – no prediction value will be calculated and displayed.
The following combinations of norms are supported:
Predicition Norm
Solymar
ECCS
Age range Composite norm
7-18
19-70
ECCS/Solymar
Zapletal
ECCS
Solymar
Hedenström
6-18
19-70
7-18
20-70
ECCS/Zapletal
Hedenström/Solymar*
*The Composite of Hedenstrom/Solymar cannot be used for age 19
DIR 80012334 Ver. C 70 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
For a listing of the parameters included in each norm, see section 12.1 Norm Profiles.
12.5 Lung Age
Lung age is a calculated value based on a patient’s demographics and spirometric performance. This provides a relative indication of the health of the patient's lungs. This value is used primarily to encourage smoking cessation.
The SpiroPerfect spirometer, calculates lung age values according to the document cited in
Reference 4 (Morris, 1985). For single-effort tests, lung age is based on the current effort.
Otherwise, it is based on the patient’s “best” pre effort as defined in the settings.
Lung age calculations are provided only for patients 20 and older. For patients older than 84 years the lung age is extrapolated. This limitation is derived from the subject population on which Morris based his research. The lung age is one floating point number in years: the average of the 4 formulas in the Morris article (FVC, FEV1, FEF25-75%, and FEF0.2-1.2).
Specifically, lung age is calculated as follows:
Gender Lung Age Formula
Men [5.920 (height) – 40.000 (FVC) – 169.640 +
2.870 (height) – 31.250 (FEV1) – 39.375 +
2.319 (height) – 21.277 (FEF200-1200) + 42.766 +
1.044 (height) – 22.222 (FEF25%-75%) + 55.844 ] / 4
Women [4.792 (height) – 41.667 (FVC) – 118.833 +
3.560 (height) – 40.000 (FEV1) – 77.280 +
4.028 (height) – 27.778 (FEF200-1200) – 70.333 +
2.000 (height) – 33.333 (FEF25%-75%)+18.367 ] / 4 height in inches
12.6 Ethnic group correction
Studies have demonstrated that expected values for certain spirometric parameters can vary significantly from one ethnic group to another. Some norm studies include separate regression equations for different races but most others do not.
In the latter case Welch Allyn CardioPerfect applies ethnic group correction to all non-Caucasian adult patients in the prediction formulas. The interpretation area will state if the norm values are extrapolated. The ATS (for blacks) or NIOSH (for Asians) recommendations will be used for extrapolation.
Race Choices
Caucasian
Black
Asian
Hispanic
Native American
Note
FVC&FEV1
No adjustment
88%
94%
No adjustment
No adjustment
Recommendation Source
-
ATS
NIOSH
None found
None found
Race adjustment applies for adults only and applies to all supported parameters within the norm study.
If a race adjustment percentage is used, the same adjustment is applied to the LLN value.
DIR 80012334 Ver. C 71 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
12.7 Understanding Interpretation Results
The following diagram illustrates the process of collecting and interpreting spirometry data.
For details, see the document noted in reference 8.
Figure 12.1 Data Interpretation Process
DIR 80012334 Ver. C 72 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
12.8 References
1. Disability Evaluation Under Social Security (the “blue book”), Social Security
Administration SSA publication number 64-039, Office of Disability Programs ICN
468600, January 2003.
See in particular the calibration and reporting sections of this document.
2. Lung Function Testing: Selection of Reference Values and Interpretive Results,
American Thoracic Society, March 1991.
This document describes the methods of selecting the reference values and the algorithm for interpretative results.
3. National Occupational Respiratory Mortality System, National Institute for
Occupational Safety and Health (NIOSH).
4. Short Report Spirometric “Lung Age” Estimation for Motivating Smoking Cessation,
James F. Morris, M.D., and William Temple, Preventive Medicine 14, 655-662, 1985.
5. ATS/ERS Task Force: Standardisation of Lung Function Testing, European
Respiratory Journal, Volume 26 Number 2, 319-338, 2005.
This document describes the methods of acquiring the output parameters and the required accuracy. For details on ATS/ERS acceptability criteria, see these sections:
“Start of Test Criteria,” page 324
“Manoeuvre repeatability,“ page 325
6. Standardized Lung Function Testing, European Respiratory Journal, volume 26, supplement number 16, April 2005.
7. U.S. Pulmonary Function Standards for Cotton Dust Standard, 29 CFR 1910.1043,
Appendix D.
8. Lung Function Testing: Selection of reference values and interpretive strategies.
American Thoracic Society, American Review of Respiratory Disease,
144:1202-1218 (1991).
DIR 80012334 Ver. C 73 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
13 Maintaining the Spirometer – Welch Allyn
13.1 Maintaining the Sensor
The Spirometer sensor needs little maintenance to stay in good working condition. Change the flow transducer for each patient. Check periodically for damages. Check that all connections are properly aligned and tight. Visually check the pressure tubing for leaks and kinks. Check for irreversible bending or compression of the pressure tubing between flow transducers and device.
Ensure spirometer is calibrated and that the proper lot code and calibration code is used. The lot code and calibration code can be found on the flow transducer package. For more detailed information please refer to chapter 7 the Calibration chapter of this manual.
Avoid placing spirometer and any of its components in direct sunlight or in a dusty environment.
Caution
To make reliable recordings, calibrate the sensor on a daily basis.
Keep track of the calibrations in the calibration log.
13.2 Cleaning the Spirometer
You cannot clean the spirometer or any of its components. If you choose to clean the calibration syringe, wipe its external surfaces as needed with a cloth dampened with water only.
Warning
Satisfactory maintenance procedures must be implemented, or equipment failure and health hazards may result. Only qualified service personnel should repair the equipment. See “Limited
Warranty” and “Service Policy”, page 4.
To prevent cross-contamination, do not try to clean the flow transducers and nose clips. Discard these items after a single patient use. Wear rubber gloves when replacing flow transducers, and wash hands after touching them.
DIR 80012334 Ver. C 74 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
Caution
Do not clean the pressure tubing or sensor. Trapped moisture could affect accuracy.
Replace the pressure tubing when it becomes dirty. Recalibrate after replacement.
Replace the sensor when it becomes faulty. See section 13.3
Ordering Information for Replacement Parts on page 75.
Do not immerse any part of the spirometer into a cleaning liquid or sterilize it with hot water, steam, or air.
Do not use aromatic hydrocarbons, rubbing alcohol, or solvents for cleaning the Spirometer.
13.3 Ordering Information for Replacement Parts
The following parts must be replaced as noted:
flow transducers & nose clips – Replace for each new patient.
pressure tubing – Replace when dirty.
sensor – Replace when faulty.
To order parts, call the Welch Allyn Technical Support Center.
Warning
Discard all spirometry components according to local regulations.
Use of components other than those recommended by Welch Allyn may compromise product performance. The Welch Allyn warranty can only be honored if you use Welch Allyn approved components and replacement parts.
DIR 80012334 Ver. C 75 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
Figure 13.1 Ordering Information for Replacement Parts
Item
Numbers
703418
703419
Quantities
25 pk
100 pk
Disposable Flow
Transducer
(CPWS,CP200)
Package includes Lot code and Calibration code
Pressure Tubing
(CPWS, CP200, 2m)
703415 1
Sensor
Spirometer USB Kit
703554 1
Sensor
Spirometer Serial Kit
Nose Clip
703554
703552
1
1
58550-0000 1
Calibration Syringe
3L,CPWS,CP200,SPIRO
703480 1
DIR 80012334 Ver. C 76 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
14 Troubleshooting
Condition Solution
The Device (sensor) is not responding
Disconnect and reconnect the sensor.
Check if the port settings in the settings menu correspond with the used COM-port.
Measured values are incorrect
Verify LOT number and perform a verification test.
Check the flow transducer for potential obstruction.
Do a volume calibration to check the gain-factor and to recalibrate the device if necessary.
Values are too high (intermittent)
Flow data is out of range (measured flow has exceeded the allowable limits)
The program does not predict values or the values appear incorrect
Retest with fingers positioned properly around the flow transducer. Do not block the end of the flow transducer with your fingers or hand.
Recalibrate with a 3-liter syringe.
Unable to calibrate
Check in the settings menu to see if the correct author is selected.
Verify that the date of birth, gender, race and the height of the patient are correctly entered in the patient card; these are needed for the calculation of the predicted values. For some prediction norms the patient’s weight is also obligatory.
Verify sensor calibration information.
Check the connection between flow transducer and sensor.
Replace the flow transducer.
Check that the connection between the syringe and the flow transducer is tight and without leaks.
Use even strokes in calibration.
DIR 80012334 Ver. C 77 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
Condition Solution
Error message:
No valid stroke recorded.
Wait with pushing the plunger in until the blue calibration bar starts moving.
Report does not print parameters or graphs
Indistinguishable Pre and Post curves on printed reports
Pull the plunger completely out before pressing the OK button on the start calibration window.
Check print and parameters settings.
A color printer and a color printout are recommended for printing Spirometry reports. Printing these reports with a monochrome printer or in black and white can lead to confusion as it is not easy to identify which curve is a Pre and which is a
Post effort.
Patient test values differ from values expected by physician
Verify sensor calibration information.
Verify the barometric pressure.
Recalibrate.
Replace transducer.
Verify the patient data. The norm selection is dependent upon accurate input of patient data in the SpiroPerfect database.
Eliminate any leaks in the pressure tubing.
Replace the sensor if damaged.
Make sure the patient remains still during recording.
Recalibrate.
The flow sensor has been dropped
Loss of network connection during spirometry test
DIR 80012334 Ver. C 78 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
15 Specifications
Specification Description
SpiroPerfect Computer based full diagnostic spirometer
Tests FVC, SVC, MVV, Pre-Post BD
Sensor Type
Power Equipment
Pneumotach
None, obtained from USB port
Accuracy
Reproducibility
Volume Range
Flow Range
Predictive Norms
Meets or exceeds ATS/ERS 2005 standard
Meets or exceeds ATS/ERS 2005 standard
0-14 L
+- 14 L/sec
For Predictive Norms included, see 12.2
Additional predictive norms can be added upon customer request
Interpretation
Reports
Incentive graphic
Parameters
1991 ATS Interpretation Standards.
Automatic interpretation can be disabled.
Manual Interpretation available.
Lung Age calculation
FVC - Volume / Time
FVC - Flow / Volume
FVC - Both – Volume / Time and Flow / Volume
SVC – Volume / Time
Fireman
FVC, FIVC, FIV1, FIV1%, FEV0.5, FEV1, FEV2, FEV3, FEV5, FEV6, FEV0.5,
FEV0.5%, FEV1%, FEV1/FVC, FEV2%, FEV3%, FEV5%, FEV6%, PEF, FEF25,
FEF50, FEF75, FEF0.2-1.2, FEF25-75, FEF75-85, PIF, FIF50, FEF50/FIF50,
FEV1/FEV6, FET, MEF25, MEF50, MEF75
Quality Checks
Connectivity
Storage and
Environment
SVC, ERV, IRV, VT, IC, BF, MV, Tin, Tex, Tin/Tex
MVV, MV, VT, BF, DFRC
ATS Acceptability and ATS Reproducibility checks
Audio and visual incentive for assistance in coaching patients
Compatible with CardioPerfect Workstation
Export compatible with most electronic medical records programs
Available in multi user network
Telemedicine option for e-mail transfer
Temperatures between –20 °C (-4 °F) and 50 °C (122 °F).
Relative humidity between 15 and 95% (non-condensing).
Atmospheric Pressure of 500 hPa (mbar) to 1,060 hPa (mbar).
DIR 80012334 Ver. C 79 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
Specification Description
Operation Environment
Temperatures between 10 °C (50 °F) and 40 °C (104 °F),
Relative humidity between 15% and 90% (non-condensing),
Atmospheric Pressure of 700 hPa (mbar) to 1,060 hPa (mbar),
Warm up period of 5 minutes.
DIR 80012334 Ver. C 80 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
16 Statutory and Regulatory Requirements
MDD - Medical Device Directive (MDD) 93/42/EEC
IEC/EN 60601-1-1, Medical Electrical Equipment, General Requirements for safety, Safety requirements for medical electrical systems.
IEC/EN 60601-1 Medical Electrical Equipment – General Requirements for Safety
IEC 60601-1-2 Medical Electrical Equipment - Safety Requirements - EMC
IEC/EN 60601-1-4 Collateral Standard for Programmable Medical Systems
CAN/CSA C22.2 No. 601.1-M90/UL 60601-1, Medical Electrical Equipment – General
Requirements for Safety
The authorized representative in the European Community
Welch Allyn LTD.
Navan Business Park
Dublin Road
Navan, County Meath, Republic of Ireland
Tel.: 353-46-90-67700
Fax: 353-46-90-67755
4341 State Street Road
Skaneateles Falls, NY 13153 USA
Tel: 800 535-6663 (North America Only) or 315 685-4560
Fax: 315 685-3361 www.welchallyn.com
DIR 80012334 Ver. C 81 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
17 Guidance and Manufacturer’s Declarations
Caution
The Welch Allyn SpiroPerfect spirometer needs special precautions regarding EMC and needs to be installed and put into service according to the following EMC information provided.
Portable and mobile RF communications equipment can affect the
Welch Allyn SpiroPerfect spirometer.
Electromagnetic Emissions
The Welch Allyn SpiroPerfect is intended for use in the electromagnetic environment specified below. The customer or the user of the Spirometer should assure that it is used in such an environment.
Emissions test Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1 The Welch Allyn SpiroPerfect uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
Class A
Not applicable
Not applicable
The Welch Allyn SpiroPerfect is suitable for use in all establishments other than domestic, and may be used in domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes, provided the following warning is heeded:
Warning: This equipment/system is intended for use by healthcare professionals only. This equipment/system may cause radio interference or may disrupt the operation of nearby equipment. It may be necessary to take mitigation measures, such as re-orienting or relocating the Welch Allyn SpiroPerfect or shielding the location.
DIR 80012334 Ver. C 82 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
Electromagnetic Immunity
The Welch Allyn SpiroPerfect is intended for use in the electromagnetic environment specified below. The customer or the user of the Spirometer should assure that it is used in such an environment.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
IEC 60601 test level
6 kV contact
8 kV air
Compliance level Electromagnetic environment
–
guidance
6 kV contact
8 kV air
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %.
Mains power quality should be that of a typical commercial or hospital environment.
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
±2 kV for power supply lines
±1 kV for input/output lines
Not Applicable
1 kV differential mode
2 kV common mode
Not Applicable
Not Applicable Voltage dips, short interruptions and voltage variations on power supply input lines
IEC 61000-4-11
<5 % U
T
(>95 % dip in U
T
) for 0,5 cycle
40 % U
T
(60 % dip in U
T
) for 5 cycles
70 % U
T
(30 % dip in U
T
) for 25 cycles
<5 % U
T
(>95 % dip in U
T
) for 5 sec
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
3 A/m 3 A/m
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment. If the user of the Welch Allyn SpiroPerfect requires continued operation during power mains interruptions, it is recommended that the Welch Allyn SpiroPerfect be powered from an uninterruptible power supply or a battery.
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
NOTE U
T is the AC mains voltage prior to application of the test level.
DIR 80012334 Ver. C 83 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
Electromagnetic Immunity
The Welch Allyn SpiroPerfect is intended for use in the electromagnetic environment specified below. The customer or the user of the Welch Allyn SpiroPerfect should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment guidance
–
Portable and mobile RF communications equipment should be used no closer to any part of the Welch Allyn SpiroPerfect, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2,5 GHz
3 Vrms
3 V/m
d
P d
P
80 to 800 MHz
d
P
800 MHz to 2,5 GHz where P is the maximum output power rating of the transmitter in watts (W) and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey a
, should be less than the compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Welch Allyn SpiroPerfect is used exceeds the applicable RF compliance level above, the Welch Allyn SpiroPerfect should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Welch Allyn SpiroPerfect. b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
DIR 80012334 Ver. C 84 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
Recommended separation distances between portable and mobile RF
communications equipment and the Welch Allyn SpiroPerfect
The Welch Allyn SpiroPerfect is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Welch Allyn SpiroPerfect can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Welch Allyn SpiroPerfect as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of transmitter
W
d
Separation distance according to frequency of transmitter
150 KHz to 80 MHz
P
m
80 MHz to 800 MHz
d
P
800 MHz to 2,5 GHz
d
P
0,01 0.12 0.12 0.23
0,1 0.37 0.37 0.74
1 1.2 1.2 2.3
10 3.7 3.7 7.4
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
DIR 80012334 Ver. C 85 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
18 Spiro Perfect VCT-400
The Spiro Perfect VCT-400 product was sold only in Europe and is discontinued. It will be supported until December 31, 2009.
The Product consists of two elements: the spirometry sensor, and the software that runs on the computer to which the sensor is connected. Before you can start recording spirometry tests, you need to:
Connect the sensor to the computer.
Configure the software.
OEM Spirometer: VIASYS Healthcare GmbH, Hoechberg, Germany, Vicatest 400
0123
18.1 Connect the flow sensor to the computer
Connecting the flow sensor to the computer:
1. Turn off your computer.
2. Connect the serial connector of the flow sensor to a free serial port on the computer.
3. Disconnect the mouse from your computer and connect it to the female PS2 connector of the Spiro Perfect VCT-400 flow sensor.
4. Connect the male PS2 connector of the Spiro Perfect flow sensor to the mouse port of your computer.
5. Start the computer.
18.1.1 Configure Welch Allyn CardioPerfect Workstation
After connecting the flow sensor, configure Welch Allyn CardioPerfect Workstation.
Configuring Welch Allyn CardioPerfect Workstation for use with the flow sensor:
1. Start Welch Allyn CardioPerfect Workstation.
2. Choose File> Settings> Spirometry.
3. Select the Recording tab.
4. Select the Spiro Perfect VCT 400.
5. Set the COM-port number to the number of the serial port connecting the flow sensor to the Welch Allyn CardioPerfect Workstation.
6. Select OK to save the settings.
18.1.2 Preparing the flow sensor
Preparing the flow sensor for recording:
1. Switch on the flow sensor
Note: The power switch is located on the bottom of the flow sensor.
2. Wait for the green LED indicator glow.
3. Insert a disposable mouthpiece into the flow sensor.
Note: The flow sensor is ready for recording.
DIR 80012334 Ver. C 86 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
18.2 Maintaining the flow sensor
The Spirometer flow sensor needs little maintenance to stay in good working condition. Only clean and sterilize it.
Caution
To make reliable recordings, calibrate the sensor on a daily basis.
Keep track of the calibrations in the calibration log.
Cleaning and Disinfection Procedure Spirometer:
Mouthpiece
The cardboard mouthpiece is for single patient use only. Discard it after the test.
Housing
Clean the outside of the housing with a non-aggressive cleaning agent daily.
Holder and sensor
Clean and disinfect the holder and sensor after each test.
Cleaning:
1. Use an ultrasound bath with a household cleaning agent.
2. Rinse thoroughly with distilled water afterwards
3. Let it dry.
Note: Clean holder and sensor immediately after a test. It will be more difficult to
clean when saliva dries on the sensor.
Disinfection
Leave the sensor and holder emerged in alcohol (max. 70%) for 5 to 10 minutes (if it is left emerged for a longer period, the glue inside the sensor may dissolve).
Caution
Never use Isopropyl alcohol, this dissolves the glue in the sensor.
The use of “Spiritus Ketonatus Dilutus" (70%) is recommended. This alcohol solution consists of 70% ethanol and 0.5% methylethylketon.
Other recommended disinfection substances are: (best used in an ultrasound bath):
Secusept forte in a concentration of 1.5%
3000
Descogen
After disinfection, rinse the parts thoroughly with distilled water.
DIR 80012334 Ver. C 87 / 95
18.3 Troubleshooting
CardioPerfect Workstation
SpiroPerfect Module – User Manual
Condition Solution
Flow sensor is not responding
Check if the flow sensor is switched on
(switch located on the bottom of the device)
Check if the device is connected both to a COM port (for communication) and to a
PS2 port (for power)
Check if the COM-port settings in the
Measured values are incorrect
CPWS settings menu correspond with the used COM port
Check the turbine. Take it out of the device and move it freely in the room so that air streams through it. The fan should rotate when you move it and stop quite abrupt when the movement is stopped (provided there is no draft in the room. If the vane does not rotate freely, the turbine probably needs to be replaced.
Do a volume calibration to check the gain-factor and to recalibrate the device if necessary.
Registration of low volume rates
Indistinguishable Pre and Post curves on printed reports
Rotor has been overtightend after disassembly or there is some mechanical friction. Disassemble device and re-assemble correctly.
A color printer and a color printout are recommended for printing Spirometry reports. Printing these reports with a monochrome printer or in black and white can lead to confusion as it is not easy to identify which curve is a Pre and which is a Post effort.
DIR 80012334 Ver. C 88 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
19 Function keys
The Welch Allyn CardioPerfect module is, just like all Windows applications, designed for working with the mouse. However, there might be situations in which working with the keyboard can be quicker. Therefore a number of functions within the Welch Allyn
CardioPerfect module can also be selected directly using the keyboard. Here is a list of all available keyboard shortcuts in this module, for a more general function key description please refer to the Workstation manual:
Spirometry functions
Key Function
F6 Reanalyze
F7 Comparison
F9
F10
Opens the Ambient settings dialog
Starts the Calibration process
[CTRL]+[L]
[CTRL]+[E]
Opens the Calibration log
Add new effort
[CTRL]+[H]
[CTRL]+[D]
Opens the Interpretation History
Delete current effort
Recording new test
Key Function
Exit the recording, cannot exit when recording is active. [Esc]
DIR 80012334 Ver. C 89 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
20 Glossary
adult. Generally, 18 or older. Age limits vary with each norm.
ASS. American Security Society.
ATS. American Thoracic Society. An organization that provides standards for spirometry common practice and equipment.
ATS acceptability criteria. Applicable to FVC testing only. (1) Criteria ensuring that an individual effort started and ended satisfactorily (no leaks or coughs). (2) Criteria ensuring that the patient has made at least two efforts of the same kind (two FVC-pre or two FVC-post),
and that these efforts are reproducible. For details, see document noted in reference 5.
ATS interpretive results. The software calculates interpretive results as described in the
document noted in reference 2.
best effort. A measurement calculated from a set of efforts. The formula for calculating best effort is user-selectable: (1) the single best effort or (2) a composite of best parameter values.
BF. Breathing frequency. See also MV and tidal breathing.
bronchospasm evaluation. See post-test.
BTPS. Body conditions, normal body temperature (37° C), ambient pressure, saturated with water vapor. The BTPS correction factor converts ambient conditions—temperature, humidity, and pressure— to BTPS.
CardioPerfect workstation. A PC using Welch Allyn CardioPerfect software. Stores ECG and spirometry test data. Can communicate with other electronic patient-information systems, such as billing and medical records.
composite norm value. A value that is filled in from another norm—a “composite norm source”—when the primary (selected) norm does not support a given parameter. Applicable only when “composite norm values” is enabled.
COPD. Chronic obstructive pulmonary disease. Characterized by airflow obstruction that is primarily caused by smoking. Examples include emphysema, chronic bronchitis, and asthmatic bronchitis.
curve. A graphical display of spirometry data. During SVC testing, only one curve type is available: volume/ time. During FVC testing, three curve types are available: volume/time, flow/volume, and flow/time.
effort. A single spirometry maneuver, for example, one blow. A test typically comprises
multiple efforts. See also best effort and test.
ERS. European Respiratory Society.
ERV. Expiratory reserve volume (in liters). The maximum volume that can be expired from the
level of the functional residual capacity (FRC). See also tidal breathing.
extrapolation. The practice of applying a norm’s formula to a patient who doesn’t fit that norm’s demographics. For example, if you were testing an 88-year-old man, and the primary
DIR 80012334 Ver. C 90 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
(selected) norm were based on males 85 or younger, the predicted values would be extrapolated values.
FEF50/FIF50. The ratio of these two parameters. See FEF50 and FIF50.
FEF25. Forced expiratory flow (in L/s) at 25% of FVC.
FEF50. Forced expiratory flow (in L/s) at 50% of FVC.
FEF75. Forced expiratory flow (in L/s) at 75% of FVC.
FEF85. Forced expiratory flow (in L/s) at 85% of FVC.
FEF0.2-1.2. Forced expiratory flow average (in L/s) between 0.2 and 1.2 liters of FVC.
FEF25-75. Forced expiratory flow average (in L/s) during the middle half of FVC.
FEF75-85 (“late” FEF). Forced expiratory flow average (in L/s) between 75% and 85% of
FVC.
FET. Forced expiratory time (in seconds). The elapsed time from the beginning of expiration until a specified percentage of FVC.
FEV0.5. Forced expiratory volume (in liters) at 0.5 seconds.
FEV1. Forced expiratory volume (in liters) at 1 second. An important parameter because it reflects the severity of COPD.
FEV1/FEV6. The ratio of these two parameters. See FEV1 and FEV6.
FEV2. Forced expiratory volume (in liters) at 2 seconds.
FEV3. Forced expiratory volume (in liters) at 3 seconds.
FEV5. Forced expiratory volume (in liters) at 5 seconds.
FEV6. Forced expiratory volume (in liters) at 6 seconds.
FEV0.5%. FEV0.5 as % of FVC.
FEV1%. FEV1 as % of FVC. Same as FEV1/FVC. A parameter for a single FVC effort.
FEV1% formula. A user-selectable formula that determines the calculation method for a test’s (not an effort’s) overall FEV1% value, which affects the automatic interpretation.
FEV2%. FEV2 as % of FVC.
FEV3%. FEV3 as % of FVC.
FEV5%. FEV5 as % of FVC.
FEV6%. FEV6 as % of FVC.
FEVt. Timed forced expiratory volume (in liters). Volume of air exhaled in the specified time during an FVC effort.
DIR 80012334 Ver. C 91 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
FIF50. Forced inspiratory flow (in L/s) at 50% of FIVC.
FIV1. Forced inspiratory volume (in liters) at one second.
FIV1%. FIV1 as % of FIVC.
FIVC. Forced inspiratory vital capacity (in liters). The maximum volume of air that can be inspired during forced inspiration starting from full expiration.
FIVt. Timed forced inspiratory volume (in liters). Volume of air inhaled in the specified time (t).
flow. The speed at which air is inhaled or exhaled (in L/s).
flow/volume. Same as flow over volume or flow = f(V). A type of data curve available during
FVC testing. The y axis represents flow (L/s); the x axis represents volume (liters).
flow loop. A flow/volume curve that includes inspiratory data (negative values on the y axis).
FRC. Functional residual capacity (in liters). Volume of air remaining in the lungs and airway at the average end-expiratory level.
FVC. Forced vital capacity. (1) A type of test in which patients inhale fully and exhale forcefully for as long as they can. The goal: to measure the volume and flow of air. May or may not include forced inhaling. When forced inhaling is included, it may be done either
before or after exhaling. See flow loop. (2) An important parameter (in liters): the maximum
volume of air that can be delivered during forced expiration starting from full inspiration.
IC. Inspiratory capacity (in liters). The maximum volume of air that can be inhaled after a
normal— unforced—exhalation. See also tidal breathing.
incentive screen. An animated screen that gives patients—usually children—a goal to achieve while exhaling. This screen is listed as a type of “curve” (data display) available during FVC testing.
IRV. Inspiratory reserve volume (in liters). The maximum volume that can be inspired from the
average end-inspiratory level. See also tidal breathing.
LLN. Lower limits of normal. The lowest expected value for a spirometric parameter. The method of determining this value varies from norm to norm.
lung age. A calculated value based on a patient’s demographics and spirometric performance that gives a relative indication of the health of the subject's lungs. This value is used primarily to encourage smoking cessation. Lung age is not available for patients younger than 20 years.
MV. Minute volume (in liters). The volume of air expired per minute measured over at least
one minute. MV = BF • VT. See also tidal breathing.
norm. A research-based spirometry data set with a specific profile for race, gender, age, and height. The software compares each patient’s results with data in the primary (selected) norm, reporting the results as percentages of the predicted (normal) values.
normal. Consistent with norm data.
DIR 80012334 Ver. C 92 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
parameter. A commonly defined attribute of a spirometric waveform (FVC, FEV1, and so on).
pediatric. Generally, under 18 years old. Age limits vary with each norm. Also young children’s lung sizes vary greatly. Norm values and interpretive results are not available for patients under 3 years of age. Pediatric use in the US for ages 6 and above.
PEF. Peak expiratory flow (in L/s). The largest expiratory flow achieved with a forced effort.
PIF. Peak inspiratory flow (in L/s). The largest inspiratory flow achieved with a forced effort.
post-test. A test that provides data to compare with pre-test data. Sometimes called post-Rx or post-BD (bronchodilator). A post-test must follow a pre-test within 24 hours. See also
predictive curve. A curve that follows a set of predictive points.
predictive points. Key values from the selected norm and from composite norms (if enabled). Applicable for FVC tests only. For flow/volume curves, predictive values are PEF,
FEF25, FEF50, FEF75, and FVC (all are represented as points). For volume/time curves, predictive values are FEV1 (represented as point) and FVC (represented as horizontal line). If predictive points are enabled, all available values appear on the screen and the printout.
pre-test. A test that provides a baseline for comparison with a post-test taken by the same patient. Sometimes called pre-Rx or pre-BD (bronchodilator). Pre-tests and post-tests are
commonly used to evaluate the effectiveness of medication. See also reversibility.
reversibility. The percentage difference between pre-test and post-test data. This measurement indicates the effect of medication on lung function. Reversibility applies to each parameter separately. The reversibility formula, which determines the way in which reversibility is calculated, is user-selectable.
SVC. Slow (relaxed) vital capacity. (1) A type of test in which patients breathe normally several times, then inhale maximally and exhale maximally, or vice versa. (2) An important parameter (in liters): the maximum volume of air exhaled from the point of maximum inhalation, or maximum volume of air inhaled from a point of maximum exhalation.
test. A set of efforts—at least 1 and no more than 12—in various possible combinations of
FVC efforts, SVC efforts, or both. Tests may include pre- and post-efforts (FVC or SVC to measure the effectiveness of medication.
Tex. Tidal breathing expiration time (in seconds). See also tidal breathing.
tidal breathing. Spontaneous or normal breathing. See also Tin and Tex.
tidal volume. See VT.
Tin. Tidal breathing inspiration time (in seconds). See also tidal breathing.
Tin/Tex. The ratio of these two parameters. See also Tin and Tex.
variance. The difference between the best and second best effort’s parameter for FEV1 and
FVC. Pretest and post-test variance are reported separately. See also best effort.
VC. Vital capacity. See also FVC and SVC.
volume = f(t). See volume/time.
DIR 80012334 Ver. C 93 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
volume/time. Same as volume over time or volume = f(t). A type of data curve available during both FVC and SVC testing. The y axis represents liters; the x axis represents seconds.
VT. Tidal volume (in liters). Also called TV, although VT is the preferred abbreviation. The volume of air that enters the lungs during inspiration and leaves the lungs during expiration in
a normal breathing cycle. See also MV and tidal breathing.
workstation. See CardioPerfect workstation.
DIR 80012334 Ver. C 94 / 95
CardioPerfect Workstation
SpiroPerfect Module – User Manual
Index
add test information .............................. 50 ambient Settings................................... 32
ATS acceptability criteria ...................... 25 best composite ..................................... 24 best effort.............................................. 24 manual selection ............................... 25 calibration of flow sensor ...................... 34 clean the spirometer ............................. 75 comment editor..................................... 31 add or change comments ................. 49 compare.................................... 51, 53, 60 composite norm values......................... 71 connect flow sensor.............................. 87 delete effort........................................... 48 flow transducer disposable................................... 12, 77 incentive screen.................................... 46 interpretation......................................... 62 area ....................................... 21, 54, 58 automatic .............................. 23, 62, 63 confirm .............................................. 62 history ......................................... 63, 73
Interpretation editor................................................. 62 history ............................................... 64 lung age................................................ 72 patient procedures
FVC effort ......................................... 47
FVC loop........................................... 48
MVV effort......................................... 48
SVC effort ......................................... 48 serial sensor ......................................... 12
Spiro Perfect VCT 400 ......................... 87
SVC .......................................... 28, 52, 56 patient procedures ............................ 48 symbols .................................................. 8 troubleshooting..................................... 78 warranty.............................................. 2, 3
DIR 80012334 Ver. C 95 / 95