Elis FLONET FN50 Series Design, Assembly And Service Manual

Add to My manuals
32 Pages

advertisement

Elis FLONET FN50 Series Design, Assembly And Service Manual | Manualzz

Design, Assembly and Service Manual

ELIS PLZEŇ a. s.

Electromagnetic flowmeter FLONET FN50xx.1

Electromagnetic Flowmeter

FLONET FN50xx.1

Page 1 of 32

ELIS PLZEN a. s., Lucni 425/15, 301 00 Plzen, Czech Republic, Phone: +420/377 517 711, fax: +420/377 517 722 Es90716K

ELIS PLZEŇ a. s.

Design, Assembly and Service Manual

Electromagnetic flowmeter FLONET FN50xx.1

Page 2 of 32

ELIS PLZEN a. s., Lucni 425/15, 301 00 Plzen, Czech Republic, Phone: +420/377 517 711, fax: +420/377 517 722 Es90716K

Design, Assembly and Service Manual

Electromagnetic flowmeter FLONET FN50xx.1

Page 3 of 32

ELIS PLZEŇ a. s.

Content

1. APPLICATION ...............................................................................................................................................4

2. MEASUREMENT PRINCIPLE ......................................................................................................................4

3. TECHNICAL DESCRIPTION .........................................................................................................................5

3.1.

G ENERAL ................................................................................................................................................................... 5

3.2.

M ETER DESIGN ........................................................................................................................................................... 5

3.3.

P ROTECTION OF METERS AGAINST UNPROFESSIONAL HANDLING ..................................................................................... 7

4. TECHNICAL PARAMETERS .........................................................................................................................8

4.1.

F

LOW SENSOR

............................................................................................................................................................ 8

4.1.1. Selection of correct sensor size........................................................................................................................ 8

4.1.2. Operational pressure of measured liquid ........................................................................................................ 12

4.1.3. Material of electrodes ..................................................................................................................................... 12

4.1.4. Selection of sensor lining ............................................................................................................................... 12

4.1.5. Dimensions of flangeless sensor ISX.Mxx (for heat meters) .......................................................................... 13

4.1.6. Dimensions of flanged sensor IS0.Mxx (for heat meters) ............................................................................... 14

4.1.7. Dimensions of flangeless sensor ISX.1xx ..................................................................................................... 15

4.1.8. Dimensions of flanged sensor ISX.1xx ........................................................................................................... 16

4.1.9. Technical data of sensor ................................................................................................................................ 17

4.2.

E LECTRONIC UNIT BOX .............................................................................................................................................. 18

4.2.1. Electronic unit specifications .......................................................................................................................... 18

5. METER APPLICATION RULES ............................................................................................................... 19

5.1.

S ENSOR PLACEMENT IN PIPING ................................................................................................................................... 19

5.2.

S

ENSOR EARTHING

................................................................................................................................................... 21

6. FLOWMETER INSTALLATION AND OPERATIONAL START ................................................................. 22

6.1.

S ENSOR INSTALLATION .............................................................................................................................................. 22

6.2.

E LECTRIC CONNECTIONS OF ELECTROMAGNETIC FLOWMETER ....................................................................................... 22

6.2.1. Connection to power source ........................................................................................................................... 22

6.2.2. Output signal connections .............................................................................................................................. 23

6.3.

O

PERATIONAL START

................................................................................................................................................ 24

7. OPTIONAL SETTINGS ............................................................................................................................... 24

7.1. Samples ............................................................................................................................................................ 24

7.2. Pulse output OUT1 ............................................................................................................................................ 25

8. ERROR REMOVAL AND METER REPAIR PROCEDURES FN50X0 ....................................................... 25

8.1.

FLOW-METER REPAIR PROCEDURE ................................................................................................................. 26

9. PACKAGING ............................................................................................................................................... 27

10. PRODUCT ACCEPTANCE ....................................................................................................................... 27

11. WARRANTY CONDITIONS ...................................................................................................................... 27

12. PRODUCT ORDERING ............................................................................................................................ 28

ELIS PLZEN a. s., Lucni 425/15, 301 00 Plzen, Czech Republic, Phone: +420/377 517 711, fax: +420/377 517 722 Es90716K

Design, Assembly and Service Manual

Page 4 of 32

Electromagnetic flowmeter FLONET FN50xx.1

ELIS PLZEŇ a. s.

1. APPLICATION

The FLONET FN50xx.1 electromagnetic flowmeter has been designed to measure volume flow rates of electrically conductive liquids in closed piping systems. It enables flow measurement with high accuracy in a wide range of flow rates (0.05 to 10 m/s) unlike the previous type FLONET FN20xx.1 (range 1:100) in extended range 1:200.

It is supplied in a compact design (dismountable piece of flow sensor and electronic unit) as a “blind”

(without a display) low-cost flowmeter for heat and cold measurement systems.

There is also a remote design of this flowmeter with a junction box on the top of chimney connected to a cable with an electronic unit fitted with L-shaped steel angle bracket at the other end of the cable.

Power supply range is 230 VAC (or110 VAC). The flowmeter has only one galvanic isolated passive pulse output OUT1.

The basic parameters can be set during activation process using special USB channel and sw FLOSET (it isn’t standard in a delivery) .

2. MEASUREMENT PRINCIPLE

The function of an electromagnetic flowmeter is based on Faraday’s induction law. The meter sensor consists of a non-magnetic and non-conductive tube with two embedded measuring electrodes to pick up the induced voltage. To create an alternating magnetic field, two coils are fitted onto the tube in parallel with the plane defined by the active parts of the measuring electrodes. Now if a conductive liquid flows across magnetic field B , voltage U will appear on the measuring electrodes proportional to the flow velocity v and the conductor length l .

U = B x l x v

U induced voltage

B

l

v magnetic flux density distance between the measuring electrodes liquid flow velocity

As the magnetic flux density and distance between the electrodes are constant, the induced voltage is proportional to the liquid flow velocity in the tube. The value of the volume flow rate can then be readily determined as a product of the flow velocity and square section of the tube, Q = v x S.

ELIS PLZEN a. s., Lucni 425/15, 301 00 Plzen, Czech Republic, Phone: +420/377 517 711, fax: +420/377 517 722 Es90716K

Design, Assembly and Service Manual

Page 5 of 32

Electromagnetic flowmeter FLONET FN50xx.1

ELIS PLZEŇ a. s.

3. TECHNICAL DESCRIPTION

3.1. General

The electromagnetic flowmeter consists of a sensor through which the measured liquid flows and an electronic unit where the low-level signal from the sensor is modified to a standardized form suitable for further processing in various industrial electronic devices. The output signal is proportional to the volume flow rate of the measured liquid. The only factor limiting the application of electromagnetic flowmeters is the requirement that the measured liquid shall be conductive and non-magnetic. The electromagnetic flowmeter can be designed as a compact or remote device. The meter sensor is with different type of lining (hard rubber, soft rubber and special hard rubber for drinking water, PTFE or E-CTFE), with electrodes from different material and in flanged or wafer (flangeless) version.

Cold meters always must be in a remote design !

The reason is a risk of moisture condensation inside of the cooled electronic unit in a compact design that can be caused by drawing in a surrounding air.

Flow meters for heat meters in a compact design may have a sensor with a prolonged chimney (made of stainless material). Such a sensor can measure media with temperatures up to 150ºC provided that the sensor and pipeline are thermally insulated and the ambient temperature doesn’t exceed 60ºC.

When choosing a sensor for cold or heat measurement, physical parameters of lining must be taken into account – see 4.1.4 Selection of flow sensor lining

3.2. Meter design

Compact design solution for a flangeless sensor and associated electronic unit

Compact design solution for a flanged sensor with associated electronic unit

ELIS PLZEN a. s., Lucni 425/15, 301 00 Plzen, Czech Republic, Phone: +420/377 517 711, fax: +420/377 517 722 Es90716K

Design, Assembly and Service Manual

Electromagnetic flowmeter FLONET FN50xx.1

ELIS PLZEŇ a. s.

Example of flowmeter assembly in the remote design version

Dimensions of the box to accommodate the flowmeter in the compact design version

Page 6 of 32

ELIS PLZEN a. s., Lucni 425/15, 301 00 Plzen, Czech Republic, Phone: +420/377 517 711, fax: +420/377 517 722 Es90716K

Design, Assembly and Service Manual

Electromagnetic flowmeter FLONET FN50xx.1

Page 7 of 32

ELIS PLZEŇ a. s.

3.3. Protection of meters against unprofessional handling

The meter shall be provided with an assembly seals after installation, which can do duly authorized organization, which is trained from the producer.

Placement of official and assembly seals on meters.

ASSEMBLY

SEAL

ELIS PLZEN a. s., Lucni 425/15, 301 00 Plzen, Czech Republic, Phone: +420/377 517 711, fax: +420/377 517 722 Es90716K

Design, Assembly and Service Manual

Page 8 of 32

Electromagnetic flowmeter FLONET FN50xx.1

ELIS PLZEŇ a. s.

4. TECHNICAL PARAMETERS

4.1. Flow sensor

The sensor environment must be free of any strong magnetic fields.

4.1.1. Selection of correct sensor size

The following table shows minimum and maximum flow rates for various sensor sizes and flow velocities ranging from 0,05 to10 m/s. The best operational properties will be achieved at the flow-velocity range of 0.5 to 5 m/s. For lower flow velocities, the measurement accuracy is worse while at higher flow velocities the turbulences at contact edges may cause undesirable interference.

Minimum and maximum flow rates for various sensor sizes

Q min

corresponds to flow velocity 0.05 m/s

Q max

corresponds to flow velocity 10.0 m/s m3 / h

DN

15

20

25

32

40 l / s

Q min

0.0072

0,013

0,025

0,04165

Q max

1,8

3,3

5

8,33

Q min

0,026

0,048

0,09

0,15

Q max

6,5

12

18

30

50

65

80

100

125

150

200

250

0,0625

0,1

0,16665

0,25

12,5

20

33,33

50

0,38885

0,5972

77,77

119,44

0,90275 180,55

1,597 319,4

0,225

0,36

0,6

0,9

1,4

2,15

3,25

5,75

45

72

120

180

280

430

650

1150

300

350

400

500

600

700

800

2,5

3,5

4,86

6,25

10

11,1

15,5

20,0

500

700

972

1250

2000

2778

3889

5000

36

40

56

72

9

12,6

17,5

22,5

1800

2520

3500

4500

7200

10000

14000

18000

ELIS PLZEN a. s., Lucni 425/15, 301 00 Plzen, Czech Republic, Phone: +420/377 517 711, fax: +420/377 517 722 Es90716K

Design, Assembly and Service Manual

Page 9 of 32

Electromagnetic flowmeter FLONET FN50xx.1

ELIS PLZEŇ a. s.

The basic parameters of flow meters for heat/cold meters are designed in compliance of standard EN1434

(OIML R 75).

Below is issued ratio of following flow rates: q s q p

= 2

Size DN

Overload flow q s

[m3/h]

Permanent flow q p

[m3/h]

Min. flow q i

[m3/h]

Ratio qp/pi

15 6,5 3 0,0163

20 12 6 0,0300

25 18 9 0,0450

32 30 15 0,0750

40 45 23 0,1125

50 72 36 0,1800

65 150 75 0,3750

80 180 90 0,4500

100 280 140 0,70

125 430 215 1,08

150 650 325

200 1 150 575

350 3 500 1 750

1,6

2,9

250 1 800 900 4,5

300 2 520 1 260 6,30

9

400 4 500 2 250 11

500 7 200 3 600 18,0

600 10 000

700 14 000

5 000

7 000

25

35

200

800 18 000 9 000 45

Data in the table are valid for accuracy higher than ±2% in range q min x% ≤ q i

≤ q s

Legend: q s

= overload (maximum) flow rate of a measured liquid q p

= permanent (nominal) flow rate of a measured liquid q i

= min flow rate for an accuracy of measurement

ELIS PLZEN a. s., Lucni 425/15, 301 00 Plzen, Czech Republic, Phone: +420/377 517 711, fax: +420/377 517 722 Es90716K

Design, Assembly and Service Manual

Page 10 of 32

Electromagnetic flowmeter FLONET FN50xx.1

ELIS PLZEŇ a. s.

The basic parameters of flow meters are designed in compliance of standard ISO 4064 (OIML R 49).

Below is issued ratio of following flow rates:

Q

4

Q

3

= 1.25

Q

2

Q

1

= 1.6

An accuracy complies of standard EN ISO 4064-1 (OIML R 49)

6

4

2

0

-2

-4

-6

Q1 Q2 Q3 Q4

Values in the table are acc to standard ISO 4064-1 (OIML R 49).

ELIS PLZEN a. s., Lucni 425/15, 301 00 Plzen, Czech Republic, Phone: +420/377 517 711, fax: +420/377 517 722 Es90716K

Design, Assembly and Service Manual

Page 11 of 32

Electromagnetic flowmeter FLONET FN50xx.1

ELIS PLZEŇ a. s.

Values of flow Q

1

, Q

2

, Q

3

a Q

4

, which apply to individual design variants and to dimensions, are written in the table 2 below:

Table 2

Size DN

Over load flow

Q4 [m3/h]

Permanent flow

Q3 [m3/h]

Transition flow

Q2 [m3/h]

Min. flow

Q1 [m3/h]

Insensitivity

[m3/h]

Q1 35%

Insensitivity

[m3/h]

Q1 50%

Ratio

Q3/Q1

72

120

180

280

430

650

1 150,0

1 800,0

6,5

12

18

30

45

2 520,0

3 500,0

4 500,0

7 200,0

10 000,0

14 000,0

125

150

200

250

50

65

80

100

15

20

25

32

40

300

350

400

500

600

700

0,0091

0,0168

0,0252

0,042

0,063

0,1008

0,168

0,252

0,392

0,602

0,91

1,61

2,52

3,528

4,9

6,3

10,08

14

19,6

0,288

0,48

0,72

1,12

1,72

2,6

4,6

7,2

0,026

0,048

0,072

0,12

0,18

10,08

14

18

28,8

40

56

0,0416

0,0768

0,1152

0,192

0,288

0,4608

0,768

1,152

1,79

2,75

4,16

7,36

11,52

16,13

22,4

28,8

46,08

64

89,6

57,6

96

144

224

344

520

920

1 440,0

5,2

9,6

14,4

24

36

2 016,0

2 800,0

3 600,0

5 760,0

8 000,0

11 200,0

200

800 18 000,0 14 400,0 115,2 72 25,2 36

Q

1

Q

3

Q

2 transition flow

Legend:

Q

4 overload (maximum) low (maximum) permanent (nominal) flow min. flow for specific accuracy

ELIS PLZEN a. s., Lucni 425/15, 301 00 Plzen, Czech Republic, Phone: +420/377 517 711, fax: +420/377 517 722

0,144

0,24

0,36

0,56

0,86

1,3

2,3

3,6

0,013

0,024

0,036

0,06

0,09

5,04

7

9

14,4

20

28

Es90716K

Design, Assembly and Service Manual

Page 12 of 32

Electromagnetic flowmeter FLONET FN50xx.1

ELIS PLZEŇ a. s.

4.1.2. Operational pressure of measured liquid

The standard flow-sensor versions have the following pressure ratings:

Sensor size Pressure rating

DN 15 to DN 50

DN 65 to DN 200

DN 250 to DN 800

PN 40 (4,0 MPa)

PN 16 (1,6 MPa)

PN 10 (1,0 MPa)

Relationship between operational pressure and temperature of the measured liquid.

4.1.3.

Material of electrodes

Measurement electrodes are made of stainless steel 1.4571 (316Ti). Some special applications require use of higher-quality and more resilient materials. We provide electrodes made of Hastelloy C276, Tantalum,

Titanium and Platinum-Rhodium (PtRh10).

4.1.4. Selection of sensor lining

The sensors are produced with linings of different materials. Selection of lining material depends on properties of measured liquid.

Soft rubber, low aggressive liquids, -35 o

C to 80ºC, higher content of abrasive particles

Hard rubber, low aggressive liquids, +5 to 80ºC

Special rubber for potable water, +5 to 8 0ºC

Teflon PTFE, aggressive liquids with operating temperature –20 to +110ºC, on request –35 to +150ºC

E-CTFE, operating temperature -20 to +11 0ºC, for dimensions over DN300, on request –35 to +130ºC

ELIS PLZEN a. s., Lucni 425/15, 301 00 Plzen, Czech Republic, Phone: +420/377 517 711, fax: +420/377 517 722 Es90716K

Design, Assembly and Service Manual

Electromagnetic flowmeter FLONET FN50xx.1

ELIS PLZEŇ a. s.

4.1.5. Dimensions of flangeless sensor ISX.Mxx (for heat meters)

Page 13 of 32

DN

L

Flangeless sensor dimensions for various rated diameters (DN)

DN D A* L Hmotnost

[ kg ]

PN40

25

32

40

50

72

82

92

107

165

175

186

199

104

104

104

104

2

2

2

3

PN16

65

80

100

127

142

162

219

234

254

104

104

104

3

4

4

125

150

192

218

284

310

134

134

6

8

200 274 366 219 10

* Height is considered without the electronic unit, without the junction box respectively. Weight of sensor has an indicative character.

ELIS PLZEN a. s., Lucni 425/15, 301 00 Plzen, Czech Republic, Phone: +420/377 517 711, fax: +420/377 517 722 Es90716K

Design, Assembly and Service Manual

Electromagnetic flowmeter FLONET FN50xx.1

ELIS PLZEŇ a. s.

4.1.6. Dimensions of flanged sensor IS0.Mxx (for heat meters)

Page 14 of 32

DN l

L D

Dimensions of flanged sensor according to DN

Flanges according to the norm EN 1092-1

DN D d A* L l

PN40

PN16

PN10

PN6

15

20

25

32

40

50

65

80

100

125

150

200

250

300

350

400

500

600

700

800

95

105

115

140

150

165

185

200

220

250

285

340

395

445

505

565

670

780

895

975

62

62

72

82

92

107

127

142

162

192

218

274

370

420

480

530

640

760

880

960 x x

187

206

216

230

251

267

287

317

347

405

487

542

591

649

759

877

997

1107

200

250

250

300

350

450

200

200

200

200

200

200

200

500

550

600

600

600

700

800

66

66

96

96

96

96

96

96

96

126

126

211

211

320

320

320

320

320

420

420

* Height is considered without the electronic unit, without the junction box respectively. Weight of flow sensor has an indicative character.

Weight

[ kg ]

41

54

77

92

116

167

288

356

427

6

9

14

16

19

25

2

2,5

3

4

4

ELIS PLZEN a. s., Lucni 425/15, 301 00 Plzen, Czech Republic, Phone: +420/377 517 711, fax: +420/377 517 722 Es90716K

Design, Assembly and Service Manual

ELIS PLZEŇ a. s.

Electromagnetic flowmeter FLONET FN50xx.1

4.1.7. Dimensions of flangeless sensor ISX.1xx

Page 15 of 32

L

Dimensions of flangeless sensor according to DN

DN D A* L Hmotnost

[ kg ]

PN40

25

32

40

50

65

80

72

82

92

107

127

142

158

168

179

192

212

227

104

104

104

104

104

104

3

3

4

2

2

2

PN16 100

125

150

162

192

218

247

277

303

104

134

134

4

6

8

200 274 359 219 10

* Height is considered without the electronic unit, without the junction box respectively. Weight of flow sensor has an indicative character.

ELIS PLZEN a. s., Lucni 425/15, 301 00 Plzen, Czech Republic, Phone: +420/377 517 711, fax: +420/377 517 722 Es90716K

Design, Assembly and Service Manual

ELIS PLZEŇ a. s.

Electromagnetic flowmeter FLONET FN50xx.1

4.1.8. Dimensions of flanged sensor ISX.1xx

Page 16 of 32 l

L

D

Dimensions of flanged flow sensor according to DN

Flanges according to the norm ČSN EN 1092-1

DN D d A* L l Weight

[ kg ]

15

20

25

95

105

115

62

62

72 x

170

180

200

200

200

66

66

96

2,5

3

PN40

PN16

32

40

50

65

80

100

125

140

150

165

185

200

220

250

82

92

107

127

142

162

192

199

209

223

244

260

280

310

200

200

200

200

200

250

250

96

96

96

96

96

96

126

3

4

4

6

9

14

PN10

150

200

250

300

350

400

500

285

340

395

445

505

565

670

218

274

370

420

480

530

640

340

398

480

535

584

642

752

300

350

450

500

550

600

600

126

211

211

320

320

320

320

16

19

25

41

54

77

92

600

700

800

780

895

975

760

880

960 x x x

600

700

800

320

420

420

116

167

288 x

427 PN6

* Height is considered without the electronic unit, without the junction box respectively. Weight of flow sensor has an indicative character.

ELIS PLZEN a. s., Lucni 425/15, 301 00 Plzen, Czech Republic, Phone: +420/377 517 711, fax: +420/377 517 722 Es90716K

Design, Assembly and Service Manual

Connection of sensor

Earthing

Flow rate of measured liquid

Temperature of measured liquid

Minimum conductivity of measured liquid

Lining

Electrodes

Ingress protection

Storage temperature

Page 17 of 32

ELIS PLZEŇ a. s.

Dimensions of sensor

Electromagnetic flowmeter FLONET FN50xx.1

4.1.9. Technical data of sensor

Flangeless DN 25 to DN 200

Flanged DN 15 to DN 800

Operating pressure PN 40 (4,0 MPa) for DN 15 to DN 50

PN 16 (1,6 MPa) for DN 65to DN 200

PN 10 (1,0 MPa) for DN 250to DN 800

Flangeless

Flanged

Earthing electrode

0,05 m/s to 10 m/s (range 1:200)

110 o

C

Up to 150ºC on a request (based on lining type and when using sensor with prolonged chimney made of stainless material) – see article 4.1.4

20 μS/cm, based on agreement with manufacturer up to 5 μS/cm

Soft rubber

Hard rubber

Rubber for potable water

Teflon PTFE

E - CTFE

Stainless steel 1.4571 (316Ti)

Hastelloy C276

Tantalum

Titanium

Platinum-Rhodium (PtRh10)

IP 65

10ºC to +70ºC, at the max. relative humidity 70 %

ELIS PLZEN a. s., Lucni 425/15, 301 00 Plzen, Czech Republic, Phone: +420/377 517 711, fax: +420/377 517 722 Es90716K

Design, Assembly and Service Manual

Electromagnetic flowmeter FLONET FN50xx.1

Page 18 of 32

ELIS PLZEŇ a. s.

4.2. Electronic unit box

The signal-processing electronic unit is accommodated in a cast aluminum box coated on the surface with paint of hue RAL 1017. The box is held by four (2x hexagonal head and 2x countersunk head) M5 bolts with hexagonal socket heads. Inside the box is the terminal board accessible by removing the cover fastened by

4 bolts. The side of the box is equipped by cable glands. Prior to putting the meter in service, check the correct sealing of all active glands, and tightening of the bolts holding the cover of the box.

The remote version of electronic unit is equipped with L- shaped angle bracket for wall mounting. Through the cable bushing fitted at the angle bracket, the connecting cable connects electronic unit and junction box of sensor.

4.2.1. Electronic unit specifications

Power source

Power consumption

Line fuse

Electric shock protection according to standard

ČSN 332000-4-41

Cable bushing Pg9

Box material

Weight

Ambient temperature

Storage temperature

85 to 305 VAC /47 ÷ 440 Hz

3 VA max.

T2 A

Automated disconnection from power source in TN-S network

For power supply cable with a diameter of 5 to 10 mm

Aluminum casting

0,4 kg

5 °C to 55 °C (protected from direct sun light)

10 °C to 70 °C at relative air humidity not exceeding 70%

Flow velocity range

Maximum flow error

Output 1 - passive output, insulated

Serial communication ports

0,05 to 10 m/s

Acc to standard EN1434 or ISO4064-1 class 2 (standard)

0,2 % for 10 to 100 % Q max

(on request)

0,5 % for 5 to 100 % Q max

(on request)

Binary multi-function optocoupler 5-30 V / 50 mA

Frequency or pulse or binary *)

USB not insulated, only for service

Protection class IP 65

*) A variant must be specified by an order. The calibration is done for a chosen output!

ELIS PLZEN a. s., Lucni 425/15, 301 00 Plzen, Czech Republic, Phone: +420/377 517 711, fax: +420/377 517 722 Es90716K

Design, Assembly and Service Manual

Page 19 of 32

Electromagnetic flowmeter FLONET FN50xx.1

ELIS PLZEŇ a. s.

5. METER APPLICATION RULES

5.1. Sensor placement in piping

No chemical injection or batching unit (such as chlorine compound injector) should be located at the input side of the sensor. The insufficient homogeneity of the flowing liquid may affect the flow-rate values indicated by the meter.

The meter performance will be the best if the liquid flow in the piping is well stabilized; therefore it is necessary to observe specific rules for the sensor placement in piping. In the contact planes between the sensor and the adjoining piping sections should be no edges as these would cause flow turbulence. Make sure that straight piping sections are provided before and after the sensor; their required length is proportional to the inner diameter of the piping concerned.

As required by clause 4.2.1 of standard EN 29104, the inner diameter of the connected pipe should not differ by more than 3% from that of the sensor.

If more than one flow-disturbing element such as pipe bend or fitting are located near the sensor, the required length of straight piping section on the sensor side concerned should be multiplied by the quantity of such elements.

In the cases of bi-directional flow-rate measurement, the same conditions concerning flow stability shall be met at the input and output sides of the sensor.

Required straight piping sections Pipe narrowing

In the cases where the pipe size larger than that of the meter sensor, it is necessary to use conical reduction pieces with the angle of taper not exceeding 15° (see the picture). In the cases of bi-directional flow measurement, the minimum length of straight piping sections on both sides is 5 DN. In horizontal sensor installations, to prevent bubbling, use eccentrically-fitted reduction pieces (see standard EN ISO 6817).

Pipe narrowing sections with angles not exceeding 8° can be taken for straight sections.

In the cases where the liquid is pumped, the flow sensor shall always be placed at the output side of the pump to prevent under pressure in the piping which might damage the sensor. The required length of the straight piping section between the pump and sensor is then at least 25 DN.

Pump in the piping Closing valve in the piping

ELIS PLZEN a. s., Lucni 425/15, 301 00 Plzen, Czech Republic, Phone: +420/377 517 711, fax: +420/377 517 722 Es90716K

Design, Assembly and Service Manual

Page 20 of 32

Electromagnetic flowmeter FLONET FN50xx.1

ELIS PLZEŇ a. s.

For the same reason, the sensor shall be always placed before the closing valve in the piping.

The sensor can be fitted in the piping in either horizontal or vertical position. However, make sure that the electrode axis is always horizontal and, if the sensor is mounted in a horizontal position, the flange section for attachment of the electronic unit box faces upwards.

ELECTRODE

AXIS

Electrode axis Sensor mounted in a vertical position

In the cases where the sensor is mounted in a vertical position, the flow direction shall always be upwards.

Risk of liquid aeration Permanent flooding of sensor

To ensure correct meter function at all times, the measured liquid shall completely fill up the sensor and no air bubbles shall be permitted to accumulate or develop in the sensor tube. Therefore the sensor shall never be placed in the upper pocket of the piping or in a vertical piping section where the flow direction is downwards.

In piping systems where complete flooding of the piping cannot always be guaranteed, consider placing the sensor in a bottom pocket where full flooding is ensured.

If the sensor is located near a free discharge point, such point shall be by at least 2 DN higher than the top part of the sensor.

ELIS PLZEN a. s., Lucni 425/15, 301 00 Plzen, Czech Republic, Phone: +420/377 517 711, fax: +420/377 517 722 Es90716K

Design, Assembly and Service Manual

Page 21 of 32

Electromagnetic flowmeter FLONET FN50xx.1

ELIS PLZEŇ a. s.

Sensor placement near free discharge point

Make sure that the adjoining piping is clamped/supported as close to the sensor as possible, to prevent vibrations and damage to the sensor.

Undesirable sensor vibrations Sensor bypass

In applications where continuous liquid flow is essential, a bypass shall be provided to allow for sensor servicing. A sensor bypass may also be a reasonable solution in the cases where, to dismantle the flow sensor from the piping, liquid from a very long piping section would have to be discharged.

5.2. Sensor earthing

The correct meter function requires that the sensor and adjoining piping sections be connected by lowimpedance earthing conductors to the earth potential and the protection conductor of the power source.

Measured liquid is earthed using a earthing electrode embedded in the sensor. If the sensor is not equipped with a earthing electrode is in terms of measurement the required electrical wiring between the screw on the chimney and flanges or earthing rings as shown in the pictures bellow.

This connection is recommended even if the sensor with earthing electrode (indeed fulfills the contact hazard protection and stray currents).

ELIS PLZEN a. s., Lucni 425/15, 301 00 Plzen, Czech Republic, Phone: +420/377 517 711, fax: +420/377 517 722 Es90716K

Design, Assembly and Service Manual

Page 22 of 32

Electromagnetic flowmeter FLONET FN50xx.1

ELIS PLZEŇ a. s.

6. FLOWMETER INSTALLATION AND OPERATIONAL START

The meter installation work shall be performed in strict observance of the procedures and rules described in this manual.

To prevent undesirable interference, the power cables shall be laid at least 25cm away from all signal cables. All cables shall be laid outside the thermal insulation layer on the piping (if any). Only shielded conductors shall be used and shielding shall be connected to the earth potential on the side of the plant control system.

6.1. Sensor installation

The measurement point chosen for the sensor installation should ensure that the internal part of the sensor is fully flooded with the measured liquid at all times. If the sensor is mounted in vertical position, the only permitted liquid flow direction is upwards.

No thermal insulation shall be used on the sensor body. If the flowmeter is to be installed in a pipeline with thermal insulation, the insulation shall be removed at the sensor insulation point.

The internal diameters of the piping and the sensor tube shall be identical. The input and output piping sections including seals shall be perfectly aligned, with no protruding edges. The arrow on the sensor body indicates the required fluid flow direction (positive flow direction).

Do not expose the electronic unit box to direct sunlight; in the cases of outdoor installation, use a suitable protection shield.

6.2. Electric connections of electromagnetic flowmeter

The terminals for connecting cables can be accessed upon removal of the upper lid of the electronics box.

The lid is screwed by four bolts. A schematic diagram of the connecting is shown on the bottom side of the lid.

Examples of labels showing power supply and impulse output

6.2.1. Connection to power source

Terminal

BN

BU

N neutral conductor

L phase conductor

To connect the power source, use a standard cable of three conductors of square section not exceeding 2 x

1.5mm

2 . For ambient temperatures, over 50°C, use a cable with rated operating temperature of at least

90°C. The box grommets will only accommodate cables with outer diameter between 4 and 8mm. Use of any other cable would disturb the protection of the IP 65 box.

The earthing conductor shall be longer than both the phase and neutral conductors. This is a safety requirement as in the case of loosening the cable clamping in the gland, the earthing conductor shall be the last to be disconnected from the terminal.

The power supply line shall be protected by an overcurrent circuit breaker. A seal should be applied on the breaker to prevent unauthorized handling. The electronic unit has no independent power switch. The recommended rating of the overcurrent circuit breaker is 4 to 6A.

ELIS PLZEN a. s., Lucni 425/15, 301 00 Plzen, Czech Republic, Phone: +420/377 517 711, fax: +420/377 517 722 Es90716K

Design, Assembly and Service Manual

Electromagnetic flowmeter FLONET FN50xx.1

Page 23 of 32

ELIS PLZEŇ a. s.

6.2.2. Output signal connections

Terminal Polarity Function Comments

1

2

Optocoupler collector (+) Binary output OUT1

Optocoupler emitter (-) Pulse output

Passive output, required external power source and loading resistor

The output terminals can be connected to co-operating electronic equipment using standard shielded signal cables of external diameter 3 to 6.5mm and conductor cross-section 0.5 to 1.5mm

2

. Shielded conductors shall also be used to connect all output signals where the shielding shall be connected to the earth potential on the side of the plant control system.

Upon connecting the conductors to the terminals, tighten the bolts holding the electronic box cover and check the grommet sealing.

Example of output OUT1 interconnection

6.2.3. Connection of sensor in remote design

The sensor in remote design has a junction box mounted on the top of a chimney. The connecting cable goes from the electronic unit through the cable bushing into the junction box. The cable shall be connected with the junction box as shown in the table below. The bushing has to be well tightened.

Junction box Meaning Color of conductor

Electrode E1 Brown BN

Electrode E2 White WH

Earthing electrode

Excitation winding

Excitation winding

Earthing electrode

Blue BU (shielding of couple BN, WH)

Yellow YE

Green GN

Blue BU (shielding of couple GN, YE)

Protective conductor Yellow-green GNYE (cable shielding)

ELIS PLZEN a. s., Lucni 425/15, 301 00 Plzen, Czech Republic, Phone: +420/377 517 711, fax: +420/377 517 722 Es90716K

Design, Assembly and Service Manual

Page 24 of 32

ELIS PLZEŇ a. s.

Electromagnetic flowmeter FLONET FN50xx.1

6.3. Operational start

The electromagnetic flowmeter must first be fitted mechanically and then the power supply and output terminals be interconnected. Then switch on the supply voltage. Within a short time, the meter will be initialized and its operational conditions stabilized. This version of the flowmeter contains neither display nor buttons.

The meter configuration is always customized. Changes in the configuration and/or setting can be performed via special serial communication line USB ( ELIS PLZEŇ a. s. delivers as a producer).

7. OPTIONAL SETTINGS

The electromagnetic flowmeter is standardly set at the factory. If the customer needs the parameters of optional settings, it must be specified in the order.

7.1. Samples

The number of samples “N”, on the basis of which the average flow rate value is determined, can be set within the range of 1 to 255. The averaging feature is useful in the cases where the flow through the meter sensor is unstable, the fluid is turbulent or where there air bubbles are trapped in the fluid flow.

Average flow rage as measured and displayed is the parameter used to calculated impulse output OUT1.

The averaging function helps suppress fast changes in the fluid flow rate.

Current flow rate

Average flow rate

Sampling

Suppression of step changes in flow rate

ELIS PLZEN a. s., Lucni 425/15, 301 00 Plzen, Czech Republic, Phone: +420/377 517 711, fax: +420/377 517 722 Es90716K

Design, Assembly and Service Manual

Page 25 of 32

ELIS PLZEŇ a. s.

7.2. Pulse output OUT1

Electromagnetic flowmeter FLONET FN50xx.1

In any of the impulse modes, an impulse will be generated as soon as a defined (preset) fluid volume passes through the meter sensor. Pulse width is fixed at 100 ms. For the duration of the impulse output is closed.

The following table shows the optimal choice of impulse numbers depending on the dimension of the flowmeter.

Q max

Pulse number

DN l/s l/pulse

15

20

25

1,6

3

5

32

40

8,33

12,5

50 20

65 33,33

80 50

100 77,77

125 119,44

150 180,55

200 319,4

250 500

300

350

700

972

400 1250

500 2000

600 2778

700 3889

0,1

0,1

0,1

0,1

1

1

1

1

1

10

10

10

10

10

10

100

100

100

100

800 5000 100

7.3. Production data

The production label consist all the parameters that are set at the factory.

Constant K1

Pulse number

Numbers of samples

Suppressed flow

Constant of a sensor K1 – defines a basic calibration factor of a sensor

Pulse number

Numbers of samples for averaging – it stabilizes a flow value by flow drifting. It is possible to set it by production on values 25, 50, 100, 150, 200, 250 of samples.

Suppressed flow – it defines a flow area (symmetric around zero), where all outputs values are set on zero.

8. ERROR REMOVAL AND METER REPAIR PROCEDURES FN50X0

ELIS PLZEN a. s., Lucni 425/15, 301 00 Plzen, Czech Republic, Phone: +420/377 517 711, fax: +420/377 517 722 Es90716K

Design, Assembly and Service Manual

Page 26 of 32

Electromagnetic flowmeter FLONET FN50xx.1

ELIS PLZEŇ a. s.

Error removal and meter repair activities should be reserved to duly qualified staff skilled is maintenance of electronic equipment and acquainted with the labor safety rules applicable to the plant concerned. On request, the meter manufacturer will provide training for such staff. The manufacturer’s responsibility for any meter damage due to incorrect handling is precluded.

Prior to any work on the meter such as disconnection or removal of printed circuit boards, sensor disconnection, etc., make sure that the supply voltage is disconnected. At all times, be aware of the danger of electric shock.

8.1. FLOW-METER REPAIR PROCEDURE

Prior to any maintenance or repair action on the internal parts of the meter control unit including the power source, sensor disconnection, PC boards, etc. make sure that the supply voltage is disconnected. Warning: disregarding this instruction implies risk of electric shock.

The self-diagnostic feature of the electronic unit concerns basic errors indicated via LED diode – lighting

(lightless), flashing (not flashing), flashing regularly (irregularly).

On the basis of these conditions possible to check status of the flowmeter. Correct status is that when the

LED is flashing regularly about 3 times per second for small sizes, for bigger sizes is flashing slowly. A velocity of flushing matches with velocity of excitation of coils. The second LED is red and it must flush 1 per second. If a fault occurs it will appear by changing LEDs flashing.

Of course, it can ’t be evaluated all possible faults, that even when the LED lighting correctly number of output pulses and their frequency may not correspond to the measured flow.

Because the flowmeter has no display, it is the only visual check of status of the meter via LED on the circuit board (after removing the cover and cover plates electronics). It is necessary to connect PC with diagnostic software and in the section “ERRORS” to check, which error is signalized:

Code error

E00

E01

Description of error

No error

Overload of a range AD

Possible reason/removing of an error

E02

E03

E04

E05

E06

E07

E08

E09

E10

Step change of flow

Error of reading /recording to memory

Other error of the transmitter block

Alert -it is not possible to calibrate zero flow.

Low of current of excitation

Short circuit of excitation coils

Information

– zeroing is doing, any measurement

Hight resistance of measured medium

Error of output OUT1

Over limit flow rate

(Short time of step change of flow in the range from 0 to Q

4

)

If this error lasts, the PC board is in defect.

See code error E01

In case of big error of time RTC it is necessary to change of backup battery on the processor board. If this error lasts still, it is necessary to change the block of a transmitter.

If this error lasts, it is necessary to change the block of a transmitter

(processor board, board of outputs, board of measurement).

To repeat a calibration of zero flow after several seconds. If this error lasts, it is necessary to change the block of a transmitter (processor board, board of outputs, board of measurement).

Open circuit of excitation = defect of a sensor or interrupted connection between a transmitter and a sensor

Defected board of measurement

If this error lasts, it is necessary to change the block of a transmitter

(processor board, board of outputs, board of measurement).

Short circuit of excitation = defected sensor or shorted connection between a transmitter and a sensor

Checking of signal cables

Check a sensor by Simulator SF1.0.

A flowmeter does not measure for time of 1 min. This information is missing after finishing of zeroing and a measure again normally.

Indication of empty pipe.

E14 Exceeded over flow Q4

Frequency output over 10 kHz.

Frequency of pulse output over 100 Hz.

Gap between pulses is shorten than duration of pulse.

Checking of Q

4

(100%)

If this error lasts, it is necessary to change the block of a transmitter

(processor board, board of outputs, board of measurement).

ELIS PLZEN a. s., Lucni 425/15, 301 00 Plzen, Czech Republic, Phone: +420/377 517 711, fax: +420/377 517 722 Es90716K

Design, Assembly and Service Manual

Page 27 of 32

Electromagnetic flowmeter FLONET FN50xx.1

ELIS PLZEŇ a. s.

9. PACKAGING

The product packaging shall meet the requirements regarding safe domestic and international transport or other conditions agreed to with the customer. In that, the manufacturer uses its own in-company packaging directives and standards.

10. PRODUCT ACCEPTANCE

The product acceptance procedure consists of visual inspection and check on the completeness of the delivered items with reference to the delivery note. On delivery to the customer, enclosed to the flowmeter

FLONET FN50XX shall be a delivery note, operation and maintenance manual and a statement on the product compliance with the respective standards.

11. WARRANTY CONDITIONS

Unless agreed otherwise between the manufacturer and the customer, the warranty period for electromagnetic flowmeters is 12 months counted from the delivery date. Within the warranty period, the manufacturer shall repair, free of charge, any product defects due to faulty materials or parts. In the case of a warranty repair, the warranty period shall be extended by the time the flowmeter was inoperative because of such repair. Manufacturer’s warranty shall not cover product defects or malfunctions due to incorrect product installation, operation, intentional damage, pilferage or damage due to force majeure circumstances.

ELIS PLZEN a. s., Lucni 425/15, 301 00 Plzen, Czech Republic, Phone: +420/377 517 711, fax: +420/377 517 722 Es90716K

Design, Assembly and Service Manual

Electromagnetic flowmeter FLONET FN50xx.1

Page 28 of 32

ELIS PLZEŇ a. s.

12. PRODUCT ORDERING

Position in order number 1 2 3 4 5 6 7 - 8 9 10 11

ORDERING No.:

TYPE IDENTIFICATION

Sensor design

F N 5 0 1 flange (cold and other application) 1 wafer (cold and other application) 2 flange (heat meters) wafer (heat meters)

3

4

Flow meter design and equipment

ECONOMIC, compact

ECONOMIC, remote

0

1

TECHNICAL PARAMETERS

Dimension of sensor DN/max flow rate Qmax

[m3/h],

15/6,5 (1/2"/1717)

20/12 (3/4"/3170)

25/18 (1"/4755)

0 1

0 2

0 3

(inch/max flow

32/30 (1 1/4"/7920) 0 4 rate Qmax

40/45 (1 1/2"/11887) 0 5

[US G/h]) 50/72 (2"/19019) 0 6

65/120 (2 1/2"/31697) 0 7

Sensor flanges

80/180 (3"/47551)

100/280 (4"/73961)

125/430 (5"/113590)

150/650 (6"/171706)

200/1150 (8"/303756)

250/1800 (10"/475510)

300/2520 (12"/665712)

350/3500 (14"/924391)

400/4500 (16"/1188774)

500/7200 (20"/1902039)

ČSN EN 1092-1

ANSI B 16.5

600/10000 24" /2 641 721)

700/14000 (28"/3 698 380)

800/18000 32"/4 755 060)

0 8

0 9

1 1

1 2

1 3

1 4

1 5

1 6

1 7

1 8

1 9

2 0

2 1

1

2

JIS B2210

AS 4087

3

4

Material of sensor, surface treatment non-standard X

Sensor body and flanges made of carbon steel, painted of polyurethane 1 complete stainless steel 1.4301 2 non-standard X

ELIS PLZEN a. s., Lucni 425/15, 301 00 Plzen, Czech Republic, Phone: +420/377 517 711, fax: +420/377 517 722 Es90716K

Design, Assembly and Service Manual

Sensor protection

Grounding

Electromagnetic flowmeter FLONET FN50xx.1

Page 29 of 32

ELIS PLZEŇ a. s.

Position in order number

Material of electrodes and grounding rings

Sensor lining

Nominal pressure PN

[bar]/psi [lb/in²]

Maximum temperature of measured fluid [ °C ]

Length of cables

Power supply

FLOW METER SETTINGS

Measurement in sensitivity

Number of samples for averaging stainless steel 1.4571 hastelloy C-276 non-standard

13 14 15 16 17 18 19 20 - 21 22

1

2

X hard, soft rubber (0 to 80 °C, DN20 to

DN800) teflon PTFE (-20 to +150 °C, DN15 to DN250)

E-CTFE (-20 to +130 °C, DN300 to DN800) non-standard

1

2

3

X standard IP 65 1 sensor to piping grounding electrode grounding rings

10/150

16/150

40/600

50

80

130

150

3

6

10

15 non-standard

85 ÷ 305 VAC, 47 ÷ 440 HZ standard ±0,5 % Qmax non-standard

25

50

100

150

200

250 non-standard

1

2

3

1

2

3

1

2

3

4

1

2

3

4

X

1

1

X

1

2

3

4

5

6

X

ELIS PLZEN a. s., Lucni 425/15, 301 00 Plzen, Czech Republic, Phone: +420/377 517 711, fax: +420/377 517 722 Es90716K

Design, Assembly and Service Manual

Electromagnetic flowmeter FLONET FN50xx.1

Page 30 of 32

ELIS PLZEŇ a. s.

Position in order number 22 23 - 24 - 25 26 27 - 28

Pulse output constant l/pulse

10

100

1000 non-standard

3

X

1

2 pulse output pulses for Q+ standard setting pulse length 100ms, 1

CALIBRATION

Calibration without calibration standard calibration acc to EN1434 without cal certificate standard calibration acc to EN1434 with cal certificate

1

2

3 above standard calibration acc to EN1434, accuracy ±0.5% without cal certificate 4 above standard calibration acc to EN1434, accuracy ±0.5% with cal certificate 5 standard calibration acc to ISO 4064 without cal certificate 6

7 standard calibration acc to ISO 4064 with cal certificate above standard calibration acc to ISO 4064, accuracy ±0.5% without cal certificate above standard calibration acc to ISO 4064, accuracy ±0.5% with cal certificate non-standard

8

9

X

PURCHASE CONDITIONS

Packaging no packaging standard export non-standard

Delivery personally by shipping agent on supplier’s costs by shipping agent on buyer’s costs non-standard

Warranty 6 months

12 months

18 months

24 months

36 months non-standard

IDENTIFICATION CODE

Number of manual flow meter

Es90716K

3

X

1

2

1

2

3

X

3

4

1

2

5

X

1

ELIS PLZEN a. s., Lucni 425/15, 301 00 Plzen, Czech Republic, Phone: +420/377 517 711, fax: +420/377 517 722 Es90716K

Design, Assembly and Service Manual

ELIS PLZEŇ a. s.

Electromagnetic flowmeter FLONET FN50xx.1

ELIS PLZEN a. s., Lucni 425/15, 301 00 Plzen, Czech Republic, Phone: +420/377 517 711, fax: +420/377 517 722

Page 31 of 32

Es90716K

ELIS PLZEŇ a. s.

Design, Assembly and Service Manual

Electromagnetic flowmeter FLONET FN50xx.1

Manufacturer’s address:

ELIS PLZEŇ a. s.

Luční 15,

301 00

Plzeň

Czech Republic

Tel.: +420 377 517 711

Fax: +420 377 517 722

E-mail: [email protected]

http://www.elis.cz

Issue No. 1

Page 32 of 32

ELIS PLZEN a. s., Lucni 425/15, 301 00 Plzen, Czech Republic, Phone: +420/377 517 711, fax: +420/377 517 722 Es90716K

advertisement

Related manuals