INSTRUCTION MANUAL
1
CONTENTS:
It is recommended that you read this manual thoroughly before using your
Sterling Impulse HVG.
SUBJECT
SAFETY NFORMATION
INDICATIONS
CONTRAINDICATIONS
PRECAUTION
CONTROLS & FEATURES
TROUBLESHOOTING
WARRANTY
FUNCTIONAL OPERATION
SPECIFICATIONS
PAGE
2
3
3
6
8
11
14
15
16
SAFETY INFORMATION Please read the following prescription information carefully before using the Sterling Impulse HVG. If you have any questions about this information, consult with your clinician before proceeding.
2
Indications
: High Voltage Pulsed Galvanic
(HVPG) devices should only be used under medical supervision for adjunctive therapy in the treatment of medical diseases and conditions.
HVG Applications
Maintain or increase range of motion
Prevent or retard disuse atrophy
Re-educate muscle spasms
Increase local blood circulation
Prevent venous thrombosis by
immediate post-surgical stimulation in selective muscles
Contraindications:
Cardiac Pacemakers: HVG should not be used on patients with demand-type cardiac pacemakers.
Cancer Patients: Do not use HVG on cancer patients.
3
Heart Patients: Adequate precautionary measures should be taken prior to using
HVG on patients suspected of having heart disease.
Trans thoracic Stimulation: Use caution in Trans Thoracic application of HVG as introduction of electrical current may cause arrhythmias.
Trans cerebral Stimulation: do not apply
HVG Tran cerebrally.
Epilepsy: Adequate precautions should be taken in the case of persons with suspected or diagnosed epilepsy.
Children: HVG devices should be kept out of reach of children.
Carotid Sinus: Do not use HVG over the carotid sinus nerves, especially in patients with a known sensitivity to the carotid sinus reflex.
Neck Stimulation: Severe spasms of the laryngeal and pharyngeal muscles may occur when the electrodes are placed across the neck or the mouth. This may be strong enough to close off the airway or cause difficulty in breathing.
4
Machinery Operation: Patients should never operate potentially dangerous machinery such as power saws, automobile, etc. during HVG stimulation.
Pregnancy: The safety of HVG stimulation during pregnancy, labor or delivery has not been established for either mother or fetus.
Skin and Vascular Problems: Do not use
HVG over swollen, infected or inflamed areas or skin eruptions, e.g., phlebitis, thrombophlebitis or varicose veins.
Skin Irritation: Some patients may experience skin irritation hypersensitivity due to electrical stimulation or the electrical conductive medium used.
Should this occur, remove the electrodes, discontinue stimulation and consult your clinician to determine the cause of irritation. Irritation may be reduced by alternate conductive medium or alternate electrode placement.
The long-term effects of chronic use of HVG are unknown
5
Precaution should be observed:
Where there is a tendency to hemorrhage following acute trauma or fracture.
Following recent surgical procedures when muscle contraction may disrupt the healing process.
When stimulating over the menstruating uterus. Where sensory nerve damage is present resulting in a loss of normal skin sensation.
Adverse Effects: Skin irritation or electrodes burns have been reported with the use of electrical muscle stimulator.
Additional Precautions:
Do not immerse unit in water or other liquids.
Turn the unit off when applying or removing electrodes.
6
Do not use the stimulator on any muscles or for any purpose other than that for which it was prescribed.
Do not use the device on any individuals other than those for whom the device has been prescribed.
Stimulation over areas where bone is near the skin can be painful in some patients.
Be sure to adjust amplitude slowly.
If HVG treatment is ineffective or unpleasant, discontinue therapy and consult your physician or therapist.
For external use only.
Do not use over the eyelids.
PLEASE NOTE: Use only Sterling Medical
Technologies approved accessories. Nonapproved accessories may damage your device and or void the warranty.
CAUTION: Improper use of electrodes may cause burns. Care should be exercised to use only electrodes specified by Sterling Medical
Technologies.
7
STERLING IMPULSE HVG CONTROLS
The Sterling Impulse HVG was designed with the patient usage in mind. This unit’s functions are built with Programs 1, 2, & 3. Program 4 is self-adjustable. Please consult with your clinician or health care equipment provider to determine which program is best suited to meet your needs.
ON/OFF button: allows a patient to turn on or turn off the Sterling Impulse HVG.
UP ARROW and DOWN ARROW buttons:
These buttons allow patient to increase or decrease intensity level. Note by pressing and
holding the UP / DOWN ARROW buttons the intensity will increase or decrease rapidly.
P button: Program button. This button allows changing between pre-set Programs. Please consult with your clinician or health care equipment provider.
S button: Select button confirms your selection regarding program changes.
8
Note if you make a change and do not press the select button within ten seconds, the
Sterling Impulse HVG will revert to the previous settings.
Note the arrow buttons are also used when changing options in Program 4.
LCD SCREEN: allows a patient to view on screen pre-program selections and ensure unit is operating properly. If a patient is using the nighttime program the LCD will read idle between treatments. If your unit indicated an error reading, please consult with the trouble shoot section (pg. 15).
BATTERY COMPARTMENT: If the unit’s LCD reads LO b, please turn unit off, remove battery cover and replace with a new 9 Volt Battery.
LEAD WIRE CONNECTIONS: With the unit turned off, take hold of the three-hole plug, and place into the keyed jack receptacle on top of the unit. Push plug firmly into unit. Note: Forcing plug into receptacle may cause plug to bend improperly and damage the wire. Also, when removing plug, first shut unit off, grab wire plug end and remove.
9
Electrodes should be stored according to the
instructions on the electrode package.
TWO LEAD WIRES: Black lead wire is plugged into the Dispersive electrode outlet. Red lead wire is plugged into active electrode outlet.
THREE LEAD WIRES: Black lead wire is plugged into the Dispersive Electrode outlet. The
Two Red lead wires are plugged into active electrode outlets.
Program 1 is a Nighttime protocol at 100 PPS continuous negative polarity for 15 minutes and then 20 PPS continuous negative polarity for 15 minutes for a total of 30 minutes on. Then 30 minutes off (idle mode), then repeats five more times for a total of 360 total minutes or six hours of treatment time.
Program 2 is a PRN Chronic protocol at 100
PPS continuous negative polarity for 15 minutes and then 20 PPS continuous negative polarity for 15 minutes for a total of 30 minutes on. Then the Sterling Impulse HVG turns itself OFF.
10
Program 3 is a PRN Acute protocol at 100 PPS continuous positive polarity for 15 minutes and then 20 PPS continuous positive polarity for 15 minutes for a total of 30 minutes on. Then the
Sterling Impulse HVG turns itself OFF.
Program 4 is Adjustable.
To view the total compliance time on the LCD, turn the Sterling Impulse HVG off by pushing the ON/OFF button. Depress and hold down the
UP Arrow button and the ON/OFF button at the same time. Release the ON/OFF button then release the UP Arrow button as soon as you see the LCD display 8888.
To reset compliance time to zero {when displayed} push S button once.
Functional Operation
All Programs P1, P2, P3, and P4 are selected by the practitioner or health care equipment provider. Once a program is selected, and the treatment is finished, the Sterling Impulse HVG will automatically shut it self off. The
Sterling
11
Impulse HVG
will remember the selected program including Program P4 even if manually turned off or when the battery is replaced.
1. Push the ON/OFF button to turn the
Sterling Impulse HVG on, then push the
Up Arrow button to increase the intensity to begin in all of the selected programs.
Note: by holding down the up or down button the intensity will increase or decrease rapidly.
2. To select other programs press the P button until the desired program P1, P2, P3, or P4
{the Sterling Impulse HVG will reset the output intensity to zero}, then press the S button to use the selected program. Push the Up Arrow button to increase the intensity to desired level and the selected program will begin.
3. Program P4 The adjustable program when selected allows for specific settings of the frequency from 0 - 125 PPS, the on time from 0 - 60 minutes of treatment time and the polarity: continuous negative Cn,
12
continuous positive CP, alternating positive
AP, or alternating negative An Stimulation.
A. Use the Up/Down Arrow buttons to change the frequency from 0-125 Hz
then push the select button to hold the selection.
B. Use the Up/Down buttons to change, on
time from 0- 60 minutes then push the select button to hold the selection.
C. Use the Up/Down buttons to change alternating positive AP, alternating negative AN, continuous positive CP, continuous negative CN then push the select button to hold the selection and write program to memory.
D. Push the Up button to increase intensity to desired level for treatment.
13
STERLING IMPULSE HVG SPECIFICATIONS
Channels
Option 1: One active (red) and one dispersive (black)
Option 2: Two active (red) and one dispersive (black)
On-Time
5 seconds; fixed in alternating mode and Continuous in continuous mode
Waveform
Pulsed galvanic Monophasic exponential spikes, delivered in pairs with a fixed interpulse interval of 100 Microseconds.
Pulse Width
5 microseconds at 50% of peak Voltage
Pulse Rate
1-125 pulse pairs/ seconds, Adjustable
Output Volt
0-330 Volts, adjustable into 500 ohm load
Polarity
Negative and Positive Pre-set in P1, P2, P3
P4 Selectable: Consult your Clinician
Power Source: 9-Volt Battery
Dimensions:
4.5” x 2.75” x 0.80”
Weight: 6.3 oz. including battery
14
TROUBLE SHOOTING...SERVICE
If your Sterling Impulse HVG seems to be malfunctioning, use the information below to determine what may be wrong. If none of these measures corrects the problem, the unit should be serviced. Contact your health care equipment provider.
“On” button is pushed and unit turns on, but unit does not function properly. Check LEAD WIRES
Make sure all connectors are firmly seated. If faulty
LEAD WIRES, replace with approved Sterling Medical
Technologies Lead wires.
Low Battery LO b Indicator is shown on LCD Screen
Replace with a new 9 Volt battery. Make sure to put
Battery cover back on and then presses ON / OFF button. Unit should return to program settings. Press
UP arrow to increase intensity level to achieve desired level.
If new 9-volt battery does not work, contact your healthcare equipment provider to find out where the device was obtained.
You may need to send in your complete Sterling
Impulse HVG (stimulator and lead wires) with a letter
describing the problem.
15
Send Postage-paid and insured to:
Sterling Medical Technologies
1871 Harroun St, Ste 200
McKinney, TX 75069
If outside the USA, please contact your distributor.
STERLING IMPULSE HVG LIMITED
THREE-YEAR WARRANTY
Sterling Medical Technologies Promises to the original consumer-purchaser to repair or, at the option of Sterling Medical Technologies replace any electrical stimulator which malfunctions or proves defective in materials or workmanship under normal use for a period of three years from the date of purchase. During the three years, Sterling Medical
Technologies will provide all labor and parts necessary to correct such defects or malfunctions free of charge. It is the duty of the consumer-purchaser to deliver the unit to a service facility of the factory as described under service.
EXLCUSIONS This warranty shall not apply to damage resulting from failure to follow the operating instructions in the user manual, accident, abuse, alterations, or disassembled by unauthorized personnel. This warranty does not extend to
16
accessory items such as batteries, electrodes, electrodes lead, and conductive gel, which are not an integral part of the stimulator. Your service representative can provide these items, but costs for repair or replacement will be the responsibility of the consumer-purchaser. Sterling Medical Technologies
Shall not be liable for incidental or consequential damages resulting from the sales or use of the item.
Some states do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation or exclusion may not apply to you.
NO OTHER WARRANTY
This limited warranty is the only express warranty given by Sterling Medical Technologies Implied warranties, including, but not limited to, warranties of merchantability and for a particular purpose are limited to the warranty period set forth above. Some states do not allow limitations on how long an implied warranty lasts, so the above limitation may not apply to you. This warranty gives you specific legal rights, and may also have rights, which vary state to state.
17