Fuji Electric ZKJ-2 Instruction Manual

Add to My manuals
84 Pages

advertisement

Fuji Electric ZKJ-2 Instruction Manual | Manualzz

Instruction Manual

NDIR TYPE

INFRARED GAS

ANALYZER

TYPE: ZKJ-2

INZ-TN2ZKJ-E

PREFACE

We are grateful for your purchase of Fuji Electric’s Infrared Gas Analyzer, TYPE:ZKJ.

• First read this instruction manual carefully until an adequate understanding is acquired, and then proceed to installation, operation and maintenance of the analyzer. Wrong handling may cause an accident or injury.

• The specifications of this analyzer are subject to change without prior notice for further product improvement.

• Modification of this analyzer is strictly prohibited unless a written approval is obtained from the manufacturer. Fuji Electric will not bear any responsibility for a trouble caused by such a modification.

• This instruction manual shall be stored by the person who actually uses the analyzer.

• After reading the manual, be sure to store it at a place easier to access.

• This instruction manual should be delivered to the end user without fail.

Manufacturer:

Type:

Fuji Electric Instruments Co., Ltd.

Described in Fuji Electric’s company nameplate on main frame

Date of manufacture: Described in Fuji Electric’s company nameplate on main frame

Product nationality: Japan

Delivered Items

Name

Analyzer main frame

Input/Output terminal module

Connection cable

Power cable

Fuse

Cell window mounting tool

Slide rail

Instruction manual

Quantity

1

1

1

1

2

1

2

1

Remarks

250V AC/3A delay type

With mounting block cell

Option

Request

• It is prohibited to transfer part or all of this manual without Fuji

Electric’s permission in written format.

• Description in this manual is subject to change without prior notice for further improvement.

© Fuji Electric Instruments Co., Ltd. 2001

Issued in September, 2001

INZ-TN2ZKJ-E i

ii

CAUTION ON SAFETY

First of all, read this “Caution on safety” carefully, and then use the analyzer in the correct way.

• The cautionary descriptions listed here contain important information about safety, so they should always be observed. Those safety precautions are ranked in 3 levels, “DANGER”, “CAUTION” and

“PROHIBITION”.

DANGER

CAUTION

PROHIBITION

Wrong handling may cause a dangerous situation, in which there is a risk of death or heavy injury.

Wrong handling may invite a dangerous situation, in which there is a possibility of medium-level trouble or slight injury or only physical damage is predictable.

Items which must not be done are noted.

Caution on installation and transport of gas analyzer

DANGER

• This unit is not explosion-proof type. Do not use it in a place with explosive gases to prevent explosion, fire or other serious accidents.

CAUTION

• For installation, observe the rule on it given in the instruction manual and select a place where the weight of gas analyzer can be endured.

Installation at an unsuited place may cause turnover or fall and there is a risk of injury.

• For lifting the gas analyzer, be sure to wear protective gloves.

Bare hands may invite an injury.

• Before transport, fix the casing so that it will not open. Otherwise, the casing may be separated and fall to cause an injury.

• The gas analyzer is heavy. It should be transported carefully by two or more persons if manually required.

Otherwise, body may be damaged or injured.

• During installation work, care should be taken to keep the unit free from cable chips or other foreign objects. Otherwise, it may cause fire, trouble or malfunction of the unit.

INZ-TN2ZKJ-E

INZ-TN2ZKJ-E

DANGER

CAUTION

DANGER

CAUTION

Caution on piping

In piping, the following precautions should be observed.

Wrong piping may cause gas leakage.

If the leaking gas contains a toxic component, there is a risk of serious accident being induced.

Also, if combustible gas is contained, there is a danger of explosion, fire or the like occurring.

• Connect pipes correctly referring to the instruction manual.

• Exhaust should be led outdoors so that it will not remain in the locker and installation room.

• Exhaust from the analyzer should be relieved in the atmospheric air in order that an unnecessary pressure will not be applied to the analyzer. Otherwise, any pipe in the analyzer may be disconnected to cause gas leakage.

• For piping, use a pipe and a pressure reducing valve to which oil and grease are not adhering. If such a material is adhering, a fire or the like accident may be caused.

Caution on wiring

• Wiring work must be performed with the main power set to

OFF to prevent electric shocks.

• Enforce construction of class-D grounding wire by all means.

If the specified grounding construction is neglected, a shock hazard or fault may be caused.

• Wires should be the proper one meeting the ratings of this instrument. If using a wire which cannot endure the ratings, a fire may occur.

• Be sure to use a power supply of correct rating. Connection of power supply of incorrect rating may cause fire.

Caution on use

• For correct handling of calibration gas or other reference gases, carefully read their instruction manuals beforehand.

• Before leaving unused for a long time or restarting after left at such a status for an extended length of time, follow the directions of each instruction manual because they are different from normal starting or shutdown. Otherwise, the performance may be poor and accidents or injuries may be caused.

• Do not operate the analyzer for a long time with its door left open. Otherwise, dust, foreign matter, etc. may stick on internal walls, thereby causing faults.

iii

iv

PROHIBITION

Caution on use

• Do not allow metal, finger or others to touch the input/output terminals in the instrument. Otherwise, shock hazard or injury may occur.

• Do not smoke nor use a flame near the gas analyzer. Otherwise, a fire may be caused.

• Do not allow water to go into the gas analyzer. Otherwise, hazard shock or fire in the instrument may be caused.

DANGER

CAUTION

Caution on maintenance and check

• When doors are open during maintenance or inspection, be sure to purge sufficiently the inside of the gas analyzer as well as the measuring gas line with nitrogen or air, in order to prevent poisoning, fire or explosion due to gas leak.

Be sure to observe the following for safe operation avoiding the shock hazard and injury.

• Remove the watch and other metallic objects before work.

• Do not touch the instrument wet-handed.

• If the fuse is blown, eliminate the cause, and then replace it with the one of the same capacity and type as before. Otherwise, shock hazard or fault may be caused.

• Do not use a replacement part other than specified by the instrument maker. Otherwise, adequate performance will not be provided. Besides, an accident or fault may be caused.

• Replacement parts such as a maintenance part should be disposed of as incombustibles.

CAUTION

Others

• If the cause of any fault cannot be determined despite reference to the instruction manual, be sure to contact your dealer or Fuji

Electric’s technician in charge of adjustment. If the instrument is disassembled carelessly, you may have a shock hazard or injury.

INZ-TN2ZKJ-E

CONTENTS

PREFACE .....................................................................................................................i

CAUTION ON SAFETY ............................................................................................ ii

CONTENTS ................................................................................................................. v

1.

OVERVIEW ..................................................................................................... 1-1

2.

NAME AND DESCRIPTION OF EACH UNIT ............................................. 2-1

2.1

Name and description of main unit ..................................................................... 2-1

2.2

Input/Output terminal module ............................................................................ 2-2

3.

INSTALLATION ............................................................................................. 3-1

3.1

Installation .......................................................................................................... 3-2

3.1.1

Installation of analyzer main frame ............................................................ 3-2

3.1.2

Mounting input/output terminal module .................................................... 3-3

3.2

Piping .................................................................................................................. 3-4

3.3

Sampling ............................................................................................................. 3-7

3.3.1

Conditions of sampling gas ........................................................................ 3-7

3.3.2

Sampling gas flow ...................................................................................... 3-7

3.3.3

Preparation of standard gas ........................................................................ 3-7

3.3.4

Purging of instrument inside ...................................................................... 3-7

3.3.5

Pressure at sampling gas outlet................................................................... 3-8

3.3.6

Example configuration of gas sampling system ......................................... 3-8

3.4

Wiring ................................................................................................................. 3-9

3.4.1

Power inlet .................................................................................................. 3-9

3.4.2

Input/output terminal module ..................................................................... 3-9

4. OPERATION ...................................................................................................... 4-1

4.1

Preparation for operation .................................................................................... 4-1

4.2

Warm-up operation and regular operation .......................................................... 4-1

5.

DESCRIPTION OF DISPLAY AND OPERATION PANELS ....................... 5-1

5.1

Name and description of operation panel ........................................................... 5-1

5.2

Overview of display and operation panels.......................................................... 5-2

5.3

Outline of display screen .................................................................................... 5-3

5.4

General operation ................................................................................................ 5-6

6.

SETTING AND CALIBRATION .................................................................... 6-1

6.1

Changeover of range ........................................................................................... 6-1

6.2

Calibration Setting .............................................................................................. 6-2

6.2.1

Setting of calibration concentration ............................................................ 6-2

6.2.2

Setting of manual zero calibration .............................................................. 6-4

6.2.3

Setting of calibration range ........................................................................ 6-6

INZ-TN2ZKJ-E v

vi

6.2.4

Setting of auto calibration component ........................................................ 6-8

6.3

Alarm setting .................................................................................................... 6-10

6.3.1

Setting of alarm values ............................................................................. 6-10

6.3.2

Hysteresis setting ...................................................................................... 6-12

6.4

Setting of auto calibration ................................................................................. 6-13

6.4.1

Auto calibration ........................................................................................ 6-13

6.4.2

Forced stop of auto calibration ................................................................. 6-15

6.5

Setting of auto zero calibration ......................................................................... 6-17

6.5.1

Auto zero calibration ................................................................................ 6-17

6.5.2

Forced stop of auto zero calibration ......................................................... 6-19

6.6

Peak alarm setting ............................................................................................. 6-21

6.7

Parameter setting .............................................................................................. 6-23

6.8

Maintenance mode ............................................................................................ 6-27

6.9

Calibration ........................................................................................................ 6-30

6.9.1

Zero calibration ........................................................................................ 6-30

6.9.2

Span calibration ........................................................................................ 6-31

7.

MAINTENANCE ............................................................................................. 7-1

7.1

Daily check ......................................................................................................... 7-1

7.2

Daily check and maintenance procedures........................................................... 7-1

7.3

Maintenance of analyzer unit .............................................................................. 7-2

7.3.1

Cleaning method for sample cell (pipe cell)............................................... 7-2

7.3.2

Cleaning method for sample cell (block cell)............................................. 7-4

7.3.3

Optical zero adjustment method (optical balance adjustment)................... 7-6

7.3.4

Moisture interference compensation adjustment method ........................... 7-7

7.3.5

Replacement of fuse on analyzer unit ........................................................ 7-8

8 TROUBLE SHOOTING FOR ANALYZER....................................................... 8-1

8.1

Error message ..................................................................................................... 8-1

9.

SPECIFICATIONS .......................................................................................... 9-1

9.1

General specifications ......................................................................................... 9-1

9.2

Code symbols...................................................................................................... 9-4

9.3

Outline diagram .................................................................................................. 9-6

INZ-TN2ZKJ-E

1. OVERVIEW

This infrared gas analyzer (type: ZKJ) measures the concentration of NO, SO

2

, CO

2

, CO and CH

4 contained in sampling gas on the principle that different atomic molecules have an absorption spectrum in the wave band of infrared rays, and the intensity of absorption is determined by the Lambert-Beer law.

Since this instrument incorporates a compact paramagnetic O

2

sensor, it allows measuring up to 5 components simultaneously by using the built-in O

2

sensor (up to 4 components if O

2

sensor is excluded).

Furthermore, use of a microprocessor or large sized liquid crystal display realizes improvement of operability, accuracy and multi-functions.

This instrument is optimum for measuring combustible gas exhausted from boilers or incinerators, and it is effective for steel gas analysis (blast furnace, steel converter, thermal treatment furnace, sintering

(Pellet equipment), coke furnace), storage and maturity of vegetable and fruit, biochemistry (microbe),

[fermentation], air pollution [incinerator, exhaust gas desulfurization, denitration], automotive emission

(excluding tester), protection against disasters [detection of explosive gas and toxic gas, combustion gas analysis of new building material], growth of plants, chemical analysis [petroleum refinery plant, petroleum chemistry plant, gas generation plant], environment [landing concentration, tunnel concentration, parking lot, building management] and various physical and chemical experiments.

INZ-TN2ZKJ-E

1 - 1

2. NAME AND DESCRIPTION OF EACH UNIT

2.1 Name and description of main unit

POWER

DISPLAY

〈  Rear panel 〉

Fig. 2-1 q

Name

Handle

Description

Used for withdrawing the main unit from the panel.

w Power switch e Switch for back light

Used for ON/OFF the analyzer.

Used for ON/OFF the back light of display.

r Display/operation panel t Sampling gas inlet

Liquid crysral diaplay and keys for setting various functions

For connecting to the measuring gas tube y Sampling gas outlet

Connect to the exhaust line. (A pair of sampling gas inlet/outlet is provided for each measuring unit. When ordered with purge, the piping to measuring unit 2 is built inside. In this case, the sample gas outlet for measuring unit 2 is used for purge gas inlet.) u

Name

Sector motor

Description

For driving the rotation of sector i Light source cover Infrared light source is arranged in the cover.

o Input/output terminal connector For connecting to the external input/output terminal module

!0

Power inlet

For connecting the power cable

INZ-TN2ZKJ-E

2 - 1

2.2 Input/Output terminal module

This analyzer provides input/output of various signals from the supplied input/outpt terminal module by connecting the instrument to this module.

Fig. 2-2 q

Name

Mounting hole w Input/output terminal block

(TN 1 to TN 5)

Description

Used for mounting input/output terminal module.

φ

4.5, 6 places

Name e Calibration solenoid valve drive signal output connector

Input/output terminal for signals of analog output, range identification contact, alarm contact output, etc.

r Input/output terminal module connection cable

Description

Provides outputs of solenoid valve drive signal during auto calibration.

Transistor output

(5V, 50mA, 1 contact)

Used for connecting the analyzer main unit to the input/output terminal module.

t Communication connector

Connect communication cable.

*Please refer to another manual

(INZ-TN513327-E) about communication function.

2 - 2

INZ-TN2ZKJ-E

3. INSTALLATION

DANGER

This unit is not explosion-proof type. Do not use it in a place with explosive gases to prevent explosion, fire or other serious accidents.

CAUTION

• Entrust the installation, movement or re-installation to a specialist or the supplier. A poor installation may cause accidental tipover, shock hazard, fire, injury, etc.

• The gas analyzer is heavy. It should be installed with utmost care. Otherwise, it may tip over or drop, for example, causing accident or injury.

• For lifting the gas analyzer, be sure to wear protective gloves. Bare hands may invite an injury.

• This unit should be installed in a place which conforms to the conditions noted in the instruction manual. Otherwise, it may cause electric shocks, fire or malfunction of the unit.

• During installation work, care should be taken to keep the unit free from entry of cable chips or other foreign objects. Otherwise, it may cause fire, trouble or malfunction of the unit.

INZ-TN2ZKJ-E

3 - 1

3.1 Installation

3.1.1 Installation of analyzer main frame

Installation methods for the analyzer main unit are divided into 2 types;

Type

19 inch rack mounting guide rail method

External dimensions Mounting dimensions Mounting method

19 inch rack mounting slide rail method

429

483

M6

450 or more

465

Slide rail

(Supports mass)

Slide rail

(Supports mass)

Note 1 Check and maintenance of the analyzer main unit may be carried out with the top cover detached.

The guide rail method may be used if a space accessible for maintenance is provided at the top of the main unit. If maintenance space is not provided specially, it is recommended to use the slide rail method.

Recommended slide rail: Product No.: 305A-24 manufactured by Accuride International Co.

Note 2 For 19 inch rack mounting, the weight of the analyzer is supported with the bottom of the case

(with the side of the case in case of slide rail method). For mounting dimensions of the slide rail, see “Item 9.3 External diagram”.

Don’t install the analyzer at a place which is exposed to direct sunlight.

The analyzer should be installed at a place where ambient temperature is within 0 to 40

°

C, and temperature fluctuation during use is minimum.

3 - 2

INZ-TN2ZKJ-E

3.1.2 Mounting input/output terminal module

Mount the input/output terminal module on the panel; observing the following method.

(Note) To avoid the effect of noise generated from external units, mount the I/O terminal module mounting plate on the panel for continuity at the mounting surface and connect the panel to the same ground as the analyzer main unit.

External dimensions Mounting dimensions Mounting method

Panel plate

4.5 mounting hole (x 6 places) M4 screw

150 150

Terminal

Connection cable

316 302

Cut a hole at 6 places.

Drill a 302 x 142 rectangular hole in the center.

To analyzer rear panel

Screwed to panel

Note) How to ground analyzer main unit and I/O terminal module

To avoid the effect of noises, etc. from external units, it is recommended to ground them by the procedure described below.

Bring I/O terminal module sheet metal and panel into continuity at ///// portion.

Bring analyzer main unit and panel into continuity at ///// portion or rail mounting portion.

(No grounding is required at the power terminal).

Mount the analyzer and I/O terminal module on the same panel.

Ground the panel casing.

INZ-TN2ZKJ-E

* Don’t separate the analyzer and I/O terminal module, and be sure to

ground them together.

3 - 3

3.2 Piping

Piping should be connected to the gas inlets and outlets of the front panel of the analyzer.

Use a corrosion resistant tube of Teflon, stainless or polyethylene to connect the instrument to a sampling system. Even if there is a danger of corrosion, refrain from using a tube of rubber or soft vinyl. The instrument provides inaccurate indication due to gas absorption by piping materials.

Pipe connection port is Rc1/4 female thread (or NPT1/4). Piping should be cut as short as possible for a quick response. About 4 mm inner diameter is recommended.

Entry of dust into the instrument may result in defective operation. Use a clean piping or coupling.

Connect the gas tube by the following method.

Sampling gas outlet

For measuring unit 1

Sampling gas inlet

Sampling gas inlet

Sampling gas outlet

(Purge gas inlet)

For measuring unit 2

Sampling gas inlet: Attach the gas tube to introduce gas to be measured such as one that has completed dehumidification process and standard gases for zero and span calibration to this inlet.

Gas flow to be introduced should be constant within the range of 0.5 L/min

±

0.2 L/min.

Sampling gas outlet: Exhaust measured gas through the outlet. Attach the tube to exhaust measured gas outdoors or to the atmosphere.

Purge gas inlet: It is used for purging the inside of the total gas analyzer . When the analyzer must be purged, refer to Item 3.3.4 Purging of instrument inside.

Use dry gas N

2

or instrumentation air for purge gas. (flow rate of 1L/min or more should be used and no dust or mist is contained).

3 - 4

INZ-TN2ZKJ-E

Internal piping diagram

Measuring unit 2

Measuring unit 1

Built-in

O

2

sensor

Sampling cell

Note)

2 cells may be used by combination of range.

OUTLET INLET OUTLET INLET

Sampling gas inlet

Sampling gas outlet

Sampling gas inlet

Sampling gas outlet

For measuring unit 2

For measuring unit 1

Note) When the purge gas inlet is provided, the piping to measuring unit 2 is built inside.

Correspondence of measured components and measuring units

Measuring components

1-component meter for

NO, SO

2

, CO

2

, CO and

CH

4

2-component meter for

NO/SO

2

and CO

2

/CO

2-component meter for

NO/CO

3-component meter for

NO/SO

2

/CO

4-component meter for

NO/SO

2

/CO

2

/CO

Measuring unit 1

Each component

NO/SO

2

CO

2

/CO

NO

NO/SO

2

NO/SO

2

Measuring unit 2

None

None

CO

CO

CO

2

/CO

Note) When the built-in O

2

sensor is provided, it must be connected to the measuring unit 1.

INZ-TN2ZKJ-E

3 - 5

Example of connecting each measuring unit

• One pair of gas inlet/outlet

Sampling gas

Filter

Flow meter

0.5L/min

ZKJ

Sampling gas inlet

Sampling gas outlet

Exhaust

Release to atmosphere.

• Two pair of gas inlet/outlet - (1)

Sampling gas

NO

2

/NO converter

Filter

Flow meter

0.5L/min

Filter

Flow meter

0.5L/min

ZKJ

Sampling gas inlet

Sampling gas outlet

For measuring unit 1

Sampling gas inlet

Sampling gas outlet

For measuring unit 2

Note) The NO

2

/NO converter is used when NO measurement is used for NOx measurement.

Exhaust

Release to atmosphere.

• Two pair of gas inlet/outlet - (2)

Sampling gas

Filter

Flow meter

0.5L/min

NO

2

/NO converter

Filter

ZKJ

Sampling gas inlet

Sampling gas outlet

For measuring unit 1

Sampling gas inlet

Sampling gas outlet

For measuring unit 2

3 - 6

Note) The NO

2

/NO converter is used when NO measurement is used for NOx measurement.

Exhaust

Release to atmosphere.

INZ-TN2ZKJ-E

3.3 Sampling

3.3.1 Conditions of sampling gas q Dust contained in the sampling gas should be completely removed with a filter. For the final stage filter, use a filter that allows removing dust particles of 0.3

µ m.

w Dew point of sampling gas must be lower than the ambient temperature to avoid occurrence of drain in the gas analyzer. If vapor is contained in the sampling gas, dew point should be lowered to 0

°

C by using a dehumidifier.

e If SO

3

mist is contained in the sampling gas, use a mist filter or cooler to remove SO

3 mist.

Other mists should be removed by using a mist filter or cooler.

r Corrosive gases such as Cl

2

, F

2

and HCl, if they are contained in the sampling gas in considerable amounts, will shorten the life of instruments.

t Temperature of sampling gas should be within 0 to 50

°

C. Provide a means that prevents entry of hot gas directly into the instrument.

3.3.2 Sampling gas flow

Flow of sampling gas should be 0.5L/min

±

0.2L/min.

Avoid flow fluctuation during measurement.

Observe the flow reading by a flowmeter provided as shown in the example of the sampling system configuration (Item 3.3.6).

3.3.3 Preparation of standard gas

Routine calibration is required by standard gas for keeping this instrument under normal operation condition (once a week). Prepare a standard gas cylinder for zero calibration and span calibration.

Zero gas

Analyzer without O

2 measurement

N

2 gas

Analyzer with builtin O

2 sensor

N

2

gas

Analyzer with external zirconia O

2 sensor

Dry air or atmospheric air (Eliminated with CO

2

provided)

Span gas other than for O

2 measurement

Span gas for O

2 measurement

Gas with concentration of 90% or more of full scale

Gas with concentration of 90% or more of full scale

Gas with concentration of 90% or more of full scale

Gas with concentration of 90% or more of full scale

1 to 2% O

2

3.3.4 Purging of instrument inside

The inside of instrument need not be purged generally except for the following cases.

q A combustible gas component is contained in sample gas.

w Corrosive gas is contained in the atmospheric air at the installation site.

e The same gas as the sample gas component is contained in the atmospheric air at the installation site.

In such cases as above, the inside of analyzer should be purged with the air for instrumentation or N

2

.

Purging flow rate should be about 1L/min.

If dust or mist is contained in purging gas, it should be eliminated completely in advance.

INZ-TN2ZKJ-E

3 - 7

3.3.5 Pressure at sampling gas outlet

Pressure at the sampling gas outlet should be adjusted to atmospheric pressure.

3.3.6 Example configuration of gas sampling system

The following illustrates a typical system configuration for five component gas measurement for monitoring combustion exhaust gas from boiler, refuse incinerator, etc.

Contact Fuji Electric for system configuration matching the particular use or further information.

q w

Name

Gas extractor

Gas conditioner e Gas aspirator r Electronic gas cooler t Solenoid valve y Membrane filter

Description

Gas extractor with a heating type stainless steel filter of standard mesh 40

µ m

For separation of drain, prevention of drain from being sucked through secondary filter and composite operation of constant-pressure bubbler

For aspiration of sample gas

Dries the moisture in sample gas to a dew point of approx.

2

°

C.

Used for introducing calibration gas.

PTFE filter used to eliminate fine dust particles and permit monitoring of dust adhering condition on the front panel of the gas analyzer.

u i

Name

Flowmeter

Standard gas o Zirconia O

2

sensor

!0

NO

2

/NO converter

Description

Adjusts and monitors the flow rate of sample gas.

Reference gas used for calibrating zero and span of the analyzer. Total 6 cylinders required for zero gas air, span gas NO,

SO

2

, CO, CO

2

and O

2

.

External zirconia oxygen sensor used for measuring the oxygen concentration (0 to

25%) in sample gas.

(This is not necessary in case when O

2

sensor is built-in.)

Added to NOx analyzer.

A special catalyst material for efficient conversion of NO

2 gas to NO is used.

3 - 8

INZ-TN2ZKJ-E

3.4 Wiring

3.4.1 Power inlet

The power inlet is provided at the rear panel.

Connect supplied power cable to this power inlet (Commercial power of 85V AC to 264V AC is available).

Rear panel

Power inlet

Input/output terminal connector

Grounding 2-pole plug

When noise source is in the vicinity

Avoid installing this instrument near an electrical unit (high frequency furnace or electric welder) that generates much electrical noise. If using the instrument near such a noise generating unit is unavoidable, use a different power line to avoid noise.

Main unit power supply

Varistor or spark killer

Mount a noise suppressor such as varister (ENA211-2 by

Fuji Elctric) or spark killer (S1202 by OKAYA) to the noise generating unit when noise is generated from relays or solenoid valves.

Mount the suppressor near the noise generating source, or it will have no effect.

Install (connect) near the source.

Noise generating source

3.4.2 Input/output terminal module

This analyzer should be connected to the input/output terminal module by supplied exclusive cable.

Plug this cable connector into the receptacle at the rear panel of the analyzer and the receptacle on the PC board of the input/output module.

INZ-TN2ZKJ-E

Analyzer Exclusive cable

(1 meter long)

Input/output terminal module

3 - 9

(1) Analog output signal (AO): terminal block 1 q to @0 , terminal block 2 e to y

Output signal: 4 to 20 mADC or 0 to 1 VDC (selected when ordering)

Non-insulated output

Allowable load: 4 to 20 mADC, 550

or less

0 to 1 VDC, 100k

or less

Analog output is provided from each terminal corresponding to the channel displayed in the measurement screen.

All of analog output signals for the instrument are not isolated. It is recommended to isolate signals individually to prevent interference from unnecessary signals or to prevent external interference.

(2) O

2

sensor input: terminal block 2 q – w

Input signal:

External zirconia O

2

analyzer: Zirconia O

2

External O

2

analyzer:

sensor signal (Fuji ZFK7 output)

0 to 1 VDC (DC input resistor of 1M

or more)

It is used when the external zirconia O

2

analyzer or external O

2

analyzer is specified as order.

To connect to the output of the external Zirconia analyzer or external O

2

analyzer prepared separately.

In case of an external O

2

analyzer, input a signal of 0 to 1 VDC with respect to O

2

full scale of the analyzer.

In case of built-in O

2

analyzer, do not use the terminals.

O

2

sensor input is not isolated. It is recommended to isolate when an external O

2

analyzer is installed apart from this analyzer.

(3) Contact input (DI): terminal block 2 !1 to @0 , terminal block 3 t to !0

It is for a contact input at no voltage. An input is provided when switching to short circuit

(on) or open (off).

No voltage is applied to the terminals.

(4) Contact output (DO): terminal block 3 !1

to @0 , terminal block 4 and terminal block 5

Contact rating:    250VAC/2A, load resistance

An output is for a relay contact output. An output is provided when switching to conductive

(on) or open (off).

Wiring of analog output signal, O

2

sensor input and contact input should be fixed separately from the wiring of power supply and contact output.

Note) To avoid the effect of noise generated from external units, be sure to ground the analyzer main unit. Continue between the I/O module mounting plate and the panel and connect the panel casing to the same ground as the analyzer.

3 - 10

INZ-TN2ZKJ-E

(5) List of terminal blocks

Terminal block 1

<TN1>

CH5 output –

(CH5_OUT) +

1

2

11

12

+

CH10 output

(CH10_OUT)

CH4 output

(CH4_OUT)

+

3

4

13

14

+

CH9 output

(CH9_OUT)

CH3 output

(CH3_OUT)

+

5 15

+

CH8 output

(CH8_OUT)

6

7

16

17 CH2 output

(CH2_OUT)

+

+

CH7 output

(CH7_OUT)

8 18

CH1 output

(CH1_OUT)

+

9 19

10 20

(M3.5 screw)

+

CH6 output

(CH6_OUT)

Unassigned

Terminal block 3

<TN3>

1 11 CH5 range identification contact output (RNG_IDCH5)

2 12

Unassigned

3

4

13

14

CH4 range identification contact output (RNG_IDCH4)

Remote hold input

(R_HOLD)

Average value reset

input (RESET)

Auto calibration

remote start

input (R_CAL)

7

8

5

6

9

15

16

17

18

19

10 20

(M3.5 screw)

CH3 range identification contact output (RNG_IDCH3)

CH2 range identification contact output (RNG_IDCH2)

CH1 range identification contact output (RNG_IDCH1)

Note 1

O

2

sensor input

(O

2

_IN)

Terminal block 2

<TN2>

+

1

2

11

12

CH5 remote range changeover input

(R_RNG_CH5)

CH12 output

(CH12_OUT)

CH11 output

(CH11_OUT)

3

4

5

13

14

15

CH4 remote range changeover input

(R_RNG_CH4)

CH3 remote range changeover input

(R_RNG_CH3) 6 16

Unassigned

Unassigned

7

8

9

17

18

19

CH2 remote range changeover input

(R_RNG_CH2)

CH1 remote range changeover input

(R_RNG_CH1) 10 20

(M3.5 screw)

Note 1 : For external O

2

sensor input.

Peak count alarm output (PEAK_ALM)

Terminal block 4

<TN4>

1 11

2 12

Unassigned

Auto calibration status/ maintenance status contact output (ACAL/MNT)

3

4

13

14

Unassigned

Pump ON/OFF

contact output (PUMP)

Calibration error contact output

(CAL_ALM)

Instrument error contact output

(FAULT)

7

8

5

6

9

15

16

17

18

19

10 20

(M3.5 screw)

Unassigned

Unassigned

Unassigned

Unassigned

Terminal block 5

<TN5>

1 11 Unassigned

CH3 alarm output

(ALM_CH3)

CH2 alarm output

(ALM_CH2)

CH1 alarm output

(ALM_CH1)

4

5

2

3

6

7

8

17

18

9 19

10 20

(M3.5 screw)

12

13

14

15

16

Power disconnection alarm output

(POWER_OFF)

CH5 alarm output

(ALM_CH5)

CH4 alarm output

(ALM_CH4)

Connector

<CN3>

Solenoid valve drive signal output for calibration

(Transister output)

Contact output for sample gas selection

1

2

Contact output for zero calibration

3

Contact output for CH1 span calibration

4

Contact output for CH2 span calibration

5

6

7

8

Contact output for CH3 span calibration

Contact output for CH4 span calibration

Contact output for CH5 span calibration

9 5V DC relay drive power

<D-sub 9P>

INZ-TN2ZKJ-E

3 - 11

(6) Description on terminal block

Terminal block 1

<TN1>

CH5 output –

(CH5_OUT) +

1

2

11

12

+

CH10 output

(CH10_OUT)

CH4 output

(CH4_OUT)

+

CH3 output

(CH3_OUT)

+

CH2 output

(CH2_OUT)

+

CH1 output

(CH1_OUT)

+

3

4

13

14

+

CH9 output

(CH9_OUT)

5 15

+

CH8 output

(CH8_OUT)

6

7

16

17

+

CH7 output

(CH7_OUT)

8

9

18

19

10 20

(M3.5 screw)

+

CH6 output

(CH6_OUT)

Note 1

O

2

sensor input

(O

2

_IN)

Terminal block 2

<TN2>

+

1

2

11

12

CH5 remote range changeover input

(R_RNG_CH5)

CH12 output

(CH12_OUT)

CH11 output

(CH11_OUT)

Unassigned

Unassigned

8

9

6

7

3

4

5

13

14

15

16

17

18

19

CH4 remote range changeover input

(R_RNG_CH4)

CH3 remote range changeover input

(R_RNG_CH3)

CH2 remote range changeover input

(R_RNG_CH2)

CH1 remote range changeover input

(R_RNG_CH1) 10 20

(M3.5 screw)

Note 1: For external O

2

sensor input.

3 - 12

Terminal block 2 <TN2>

Terminal block for analog output (non-isolated output)

Between 1–2:

Between 3–4:

Between 5–6:

Between 7–8:

CH5 output

CH4 output

CH3output

CH2 output

Between 9–10: CH1 output

Between 11–12: CH10 output

Between 13–14: CH9 output

Between 15–16: CH8 output

Between 17–18: CH7 output

Between 19–20: CH6 output

Terminal block 2 <TN2>

Between 1–2:

Between 3–4:

O

2

sensor input

(For input of Fuji’s zirconia oxygen sensor and externally oxygen sensor. Must not be used unless external O

2

sensor is provided.)

CH12 output

Between 5–6:

Between 7–10

CH11 output

For internal connection. Must not be wired. (Must not be used as junction terminal).

Between 11–12: CH5 remote range changeover input

Between 13–14: CH4 remote range changeover input

Between 15–16: CH3 remote range changeover input

Between 17–18: CH2 remote range changeover input

Between 19–20: CH1 remote range changeover input

Action of remote range changeover

High range is selected when open. Low range is selected when short-circuited. For details of action, see “Item 6.7 Parameter

Setting, Remote Range”.

INZ-TN2ZKJ-E

Unassigned

Terminal block 3

<TN3>

1 11 CH5 range identification contact output (RNG_IDCH5)

2 12

Unassigned

Remote hold input

(R_HOLD)

Average value reset

input (RESET)

Auto calibration

remote start

input (R_CAL)

5

6

3

4

7

8

13

14

15

16

17

18

9 19

10 20

(M3.5 screw)

CH4 range identification contact output (RNG_IDCH4)

CH3 range identification contact output (RNG_IDCH3)

CH2 range identification contact output (RNG_IDCH2)

CH1 range identification contact output (RNG_IDCH1)

Terminal block 3 <TN3>

Between 1–4: For internal connection. Must not be wired. (Must not be used as junction terminal.)

Between 5–6:

Between 7–8:

Remote hold input. No hold when open. Output hold when short-circuited.

For details, refer to “Item 6.7

Parameter setting, Output Hold”.

Average value reset input. shortcircuitting the contact input (for at 1.5 sec min.) resets O

2

average and O

2 corrected average simultaneously. Opening it restarts the average value.

For details, refer to “Item 6.7

Parameter setting, Average Value

Resetting”

Between 9–10: Automatic calibration remote start input

After shorting for 1.5 sec. or more, automatic calibration is started by the opening input whether the automatic calibration setting is ON/OFF.

For details, refer to “Item 6.4

Setting of auto calibration”

Between 11–12: CH5 range identification contact output

Between 13–14: CH4 range identification contact output

Between 15–16 CH3 range identification contact output

Between 17–18: CH2 range identification contact output

Between 19–20: CH1 range identification contact output

Action of range identification signal

Range identification contact is conductive at low range and open at high range.

INZ-TN2ZKJ-E

3 - 13

Peak count alarm output (PEAK_ALM)

Terminal block 4

<TN4>

1 11

2 12

Unassigned

Auto calibration status/ maintenance status contact output (ACAL/MNT)

3

4

13

14

Unassigned

Pump ON/OFF

contact output (PUMP)

5 15

Unassigned

Calibration error contact output

(CAL_ALM)

Instrument error contact output

(FAULT)

8

9

6

7

16

17

18

19

10 20

(M3.5 screw)

Unassigned

Unassigned

Terminal 4 <TN4>

Between 1–2:

Between 3–4:

Peak count alarm contact output

It is conductive when peak count exceeds the setting time. It remains open below the setting time. For setting and operation, refer to “Item 6.6 Peak alarm setting”.

Contact output of auto calibration status and maintenance contact

When the auto calibration is carried out and remote hold is

ON, it is conductive. Remains open otherwise.

Between 5–6:

Between 7–8:

Pump ON/OFF contact output

Used when turning ON/OFF the pump. It is open during auto and manual calibration status and conductive during measurement.

Calibration error contact output

It is conductive when an error occurs during zero calibration or span calibration. It is normally open.

Between 9–10: It is conductive when an error occurs to the analyzer unit. It is normally open.

Between 11–20: For internal connection, wiring is not allowed. (Do not use it as junction terminal).

3 - 14

INZ-TN2ZKJ-E

Unassigned

Terminal block 5

<TN5>

1 11 Unassigned

CH3 alarm output

(ALM_CH3)

2

3

12

13

Power disconnection alarm output

(POWER_OFF)

CH2 alarm output

(ALM_CH2)

CH1 alarm output

(ALM_CH1)

6

7

4

5

8

9

14

15

16

17

18

19

CH5 alarm output

(ALM_CH5)

CH4 alarm output

(ALM_CH4)

10 20

(M3.5 screw)

Terminal 5 <TN5>

Between 2, 3 and 4:

CH3 alarm output

When the output exceeds the set value, it is conductive between 2 and 3, and open between 3 and 4. Otherwise, it is open between 2 and 3 and conductive between 3 and 4.

Between 5, 6 and 7:

CH2 alarm output

When the output exceeds the set value, it is conductive between 5 and 6, and open between 6 and 7. Otherwise, it is open between 5 and 6, and conductive between 6 and 7.

Between 8, 9 and 10:

CH1 alarm output

When the output exceeds the set value, it is conductive between 8 and 9, and open between 9 and 10. Otherwise, it is open between 8 and 9.

Between 12, 13 and14:

Analyzer unit power OFF output

When the analyzer unit is turned ON, it is conductive between 12 and 13, and open between 13 and 14. When the analyzer unit is turned OFF, it is open between 12 and 13, and conductive between 13 and 14.

Between 15, 16 and 17:

CH5 alarm output

When the output exceeds the set value, it is conductive between 15 and 16, and open between 16 and 17. Otherwise, it is open between 15 and 16, and conductive between 16 and 17.

Between 18, 19 and 20:

CH4 alarm output

When the output exceeds the set value, it is conductive between 18 and 19, and open between 19 and 20. Otherwise, it is open between 18 and 19, and conductive between 19 and 20.

For detailed action of the alarm contact, refer to “Item 6.3 Alarm setting”.

INZ-TN2ZKJ-E

3 - 15

Connector <CN3>

Solenoid valve drive signal output for calibration

Connector

<CN3>

(Transister output)

Contact output for sample gas selection

1

2

Contact output for zero calibration

3

Contact output for CH1 span calibration

4

Contact output for CH2 span calibration

5

6

7

8

Contact output for CH3 span calibration

Contact output for CH4 span calibration

Contact output for CH5 span calibration

9

5V DC relay drive power

<D-sub 9P>

Connector <CN3> provides outputs in combination with calibration action during auto calibration and manual calibration.

An output is from a transistor (ratings: 5V/50mA).

A transistor is turned ON before starting each calibration.

Sample selection output is ON during measurement and OFF during calibration.

If calibration is not performed, the other transistors are OFF.

In case of auto calibration, sequential output is ON/OFF according to the setting.

Refer to “Item 6.4 Setting of auto calibration”.

Note) No. 9 pin is for solenoid valve ON/OFF relay drive power (5V DC/0.5A, max).

Use No. 9 with reference to the diagram.

Example of using solenoid valve drive signal output for calibration

<Electrical system>

5V Contact relay

9

1

3

4

5

CN3

Cable

Relay board I/O terminal module

<Gas sampling system>

Membrane filter

SV1

SV2

SV3

SV4

Solenoid valve drive power

SV1 to SV4: solenoid valves

SV1 SV2 SV3 SV4

Refer to “Item 3.3.6 Example configuration of gas sampling system”.

Electronic gas cooler

Zero CH1 span

CH2 span

Standard gas for calibration

Relay board (TK7H2949C4) and exclusive cable (D-sub 9p straight cable: 1.5 meters) are available on request.

3 - 16

INZ-TN2ZKJ-E

(7) Timing of solenoid valve drive signal for calibration q Manual calibration (see “Item 6.9 Calibration”.)

• Zero calibration

Pump ON/OFFcontact

Sample selection output

Zero calibration output

CH1 to 5 span calibration output

Output hold function

(with hold ON setting)

ZERO ENT ENT Calibration end

Calibration gas flow off on off on off on off

Output signal hold

Hold extension time.

• Span calibration

Pump ON/OFFcontact

Sample selection output

Zero calibration output

CH1 to 5 span calibration output

For CH1 span calibration

Output hold function

(with hold ON setting)

SPAN When selecting

CH2 using

DOWN and UP keys.

ENT ENT

Calibration end

Calibration gas flow

Output signal hold

Hold extemsion time.

Note) The hold extension time depends on the gas flow time of the automatic calibration settings.

INZ-TN2ZKJ-E

3 - 17

w In case of automatic calibration (example shown in Item 6.4.1, Automatic calibration settings)

Pump ON/OFF contact

Automatic calibration start

CH1 span calibration

Zero calibration

CH2 span calibration

CH3 span calibration

CH4 span calibration

CH5 span calibration

Sample selection output

Zero calibration output

CH1 span calibration output

Zero gas

350

CH1 span gas

350 CH2 span calibration output

CH3 span calibration output

CH4 span calibration output

CH5 span calibration output

Automatic calibration contact

Output hold function

Hold extension time.

3 - 18

INZ-TN2ZKJ-E

4. OPERATION

4.1 Preparation for operation

(1) Tube and wiring check

Double-check if tubes of the gas sampling and exhaust ports are correctly connected.

Double-check for proper wiring.

4.2 Warm-up operation and regular operation

(1) Operation procedure q Turn ON the power switch on the front panel of the analyzer unit.

The measurement screen appears on the front display panel in 1 or 2 seconds.

w Wait for about 4 hours until the instrument is warmed up.

Since about 4 hours are required until the instrument allows accurate measurement, warm up the instrument with the power ON.

Note) When in warm-up, the concentration reading may be beyond.

upper limit of range or

lower limit of range.

But, it is not an error.

e Setting of various set values

Perform the various settings according to “Chapter 6. Setting and Calibration”.

r Zero calibration and span calibration

Perform zero calibration and span calibration after warm-up operation.

Refer to “Chapter 6.9. Calibration”.

t Introduction and measurement of measuring gas

Introduce the measuring gas into the analyzer unit before starting measurement.

INZ-TN2ZKJ-E

4 - 1

5.

DESCRIPTION OF DISPLAY AND OPERATION PANELS

This section describes the display unit and operation panel of the analyzer unit. It also explains the name and description of function on the operation panel.

5.1 Name and description of operation panel

POWER

DISPLAY

MODE ESC

ENT

ZERO

SPAN

Fig. 5-1 q

Name

MODE key

Description

Used to switch the mode.

w e

SIDE key

UP key r DOWN key

Used to change the selected item

(by moving the cursor) and numeral digit.

Used to change the selected item

(by moving the cursor) and to increase numeral value.

Used to change the selected item

(by moving the cursor) and to decrease numeral value.

t y

Name

ESC key

ENT key u ZERO key

Description

Used to return to a previous screen or cancel the setting midway.

Used for confirmation of selected items or values, and for execution of calibration.

Used for zero calibration.

i SPAN key

Used for span calibration.

Note) The switch for back light is used for turning ON/OFF the back light of display.

The life time of back light is 50000 hours.

Please turn OFF the switch when indication is not necessary, so the life time will be longer.

INZ-TN2ZKJ-E

5 - 1

5.2 Overview of display and operation panels

Measurement mode

* 1

MODE ESC

Measurement mode

MODE ESC

Measurement mode

MODE ESC

MODE

Menu mode

ESC

Menu mode

1. Changeover of Range

2. Setting about Calibration

3. Alarm Setting

4. Setting of Auto Calibration

5. Setting of Auto Zero calibration

6. Setting of Peak Alarm

7. Parameter Setting

* 2) The peak alarm setting is

added according to the

code symbol when CO

and O

2

components exist.

ESC

ESC

ESC

ESC

ESC

ESC

Changeover of

Range

Setting about

Calibration

Alarm Setting

ESC

Setting of Auto

Calibration

Setting of Auto

Zero calibration

Setting of

Peak Alarm

Parameter

Setting

* 1) The panel configuration is changed depending on the

display channel. (The measurement mode screen can

be viewed by scrolling the arrow key up and down).

ZERO

ZERO Calibration

SPAN Calibration

SPAN

Selection of items

Calibration value

Zero calibration

Calibration range

Auto calibration component

Selection of items

Start Time

Cycle

Flow Time

Auto calibration ON/OFF

Stop Auto Calibration

Selection of items

Start Time

Cycle

Flow Time

ON/OFF

Stop Auto zero Calibration

Selection of items

Peak Alarm ON/OFF

Peak Value

Peak Count

Hysteresis

Selection of items

Current time

Key lock

Remote range

: Current time setting

: Key lock ON/OFF

: Remote range ON/OFF

Hold

Response time

: ON/OFF

: Response time (filter)

Average time : Average time setting

Average value reset : Average value resetting

Maintenance mode : Maintenance mode

(entry of password)

Fig. 5-2

5 - 2

INZ-TN2ZKJ-E

5.3 Outline of display screen

(1) Measurement mode screen (appears when the power is turned ON)

The measurement screen depends on the number of components. The following screen configuration as shown as an example is for NO, SO

2

, CO

2

, CO and O

2

(output: 12 channel).

* corrected instantaneous value ..... CV.

** correctted average value .............

CV.

AV.

Fig. 5-3 Name and function of measurement mode screen

No.

q w e r

* For outputs of more than 5 channels, scroll the arrow key or to view.

Name

Component display

Concentration display

Range display

Unit display

Function

Displays component of instantaneous value, corrected instantaneous value, corrected average value, etc.

Displays measured value of concentration.

Displays range values.

Displays unit with ppm and vol%.

No.

t

Name

Peak alarm component display y u i

Peak alarm concentration display

Peak alarm times

Peak alarm unit display

Description

Displays peak alarm component.

(Only CO)

Displays peak alarm concentration display.

(Upper limit value)

Displays the alarm times exceeding the peak value.

Displays units of peak alarm with times/H.

INZ-TN2ZKJ-E

5 - 3

Instantaneous value and concentration value:

The concentration display of CH (component) where sampling components such as “CO

2

”,

“CO” or “O

2

are displayed in the component display, indicates current concentration values of the measured components contained in gas that is now under measurement.

O

2

correction concentration values:

CH components where “cv**” is displayed as “cv CO” in the component display are calculated from the following equation, by setting sampling components, O

2

instantaneous/concentration values and O

2

correction reference value (see item 6.8).

Correction output=

21 - On

21 - Os

K

×

Cs

On: The value of the O

2

correction referance value

(Value set by application)

Os: Oxygen concentration (%)

Cs: Concentration of relevant measured component

K: The value of the fractional part.

where, K is When K 4, K = 4 . When K < 0, K = 4.

When Cs < 0, K = 0.

The converted sampling components are NO

X

, SO

2

and CO only.

O

2

correction concentration average value:

In the CH (component) and O

2

CV

average value where “ **” is displayed as “ CO” in the component display, a value obtained by averaging O

2

correction concentration value or O

2 correction value in a fixed time is output every 30 seconds.

Averaging time can be changed between 1 minute and 59 minutes or 1 hour and 4 hours according to the average time settings (See 6.7, Parameter setting).

(The set time is displayed as “1h” , for instance, in the range display.)

* The measurement ranges of O

2

correction concentration value and O

2

correction concentration average value are the same as that of the measuring components. Also, the measurement range of O

2 average value is the same as that of O

2

.

(2) Setting/selection screen

The setting/selection screen is configured as shown below:

• In the status display area, the current status is displayed.

• In the message display area, messages associated with operation are displayed.

• In the setting item and selection item display area, items or values to be set are displayed,

as required. To work on the area, move the cursor to any item by using UP, DOWN and

SIDE keys.

Message display area

• LCD screen

Status display area

Setting item/selection item display area

Cursor

Fig. 5-4

5 - 4

INZ-TN2ZKJ-E

(3) Contents of measured channel (CH)

The following table gives measurement channels and their contents according to the symbols.

B

E

P

A

M

N

G

L

F

E

F

H

Code symbol

5th digit 6th digit 21th digit

D

B

P

A

Y

Y

Y

Y

Y

Y

Y

Y

Y

Y

Y

Y

Y

Y

Y

Y

Y

Y

A, B, C

A, B, C

A, B, C

A, B, C

A, B, C

A

A

A

A

Y

Y

Y

Y

A

Contents

CH1: NO

CH1: SO

2

CH1: CO

2

CH1: CO

CH1: CH

4

CH1: NO, CH2: SO

2

CH1: NO, CH2: CO

CH1: CO

2

, CH2: CO

CH1: NO, CH2: SO

2

, CH3: CO

CH1: NO, CH2: SO

2

, CH3: CO

2

, CH4: CO

CH1: NO, CH2: CH

4

, CH3: CO

2

, CH4: CO

CH1: NOx, CH2: O

2

, CH3: Corrected NOx, CH4: Corrected NOx average, CH5: O

2 average

CH1: SO

2

, CH2: O

2

, CH3: Corrected SO

2

, CH4: Corrected SO

2 average, CH5: O

2 average

H

G

L

M

L

M

B

B

H

G

H

G

L

M

A, B, C

A, B, C

A, B, C

A, B, C

A, B, C

A, B, C

A, B, C

A, B, C

A, B, C

A, B, C

A, B, C

A, B, C

A, B, C

A, B, C

A

A

A

A

B

B

B

B

B

C

C

C

C

C

CH1: CO, CH2: O

2

, CH3: Corrected CO, CH4: Corrected CO average, CH5: O

2 average

CH1: CH

4

, CH2: O

2

, CH3: O

2 average

CH1: NOx, CH2: SO

2

, CH3: O

2

, CH4: Corrected NOx, CH5: Corrected SO

2

,

CH6: Corrected NOx average, CH7: Corrected SO

2 average, CH8: O

2 average

CH1: NOx, CH2: CO, CH3: O

2

, CH4: Corrected NOx, CH5: Corrected CO,

CH6: Corrected NOx average, CH7: Corrected CO average, CH8: O

2 average

CH1: CO

2

, CH2: CO, CH3: O

2

, CH4: Corrected CO, CH5: Corrected CO average,

CH6: O

2 average

CH1: NOx, CH2: SO

2

, CH3: CO, CH4: O

2

, CH5: Corrected NOx, CH6: Corrected SO

2

,

CH7: Corrected CO, CH8: Corrected NOx average, CH9: Corrected SO

2 average,

CH10: Corrected CO average, CH11: O

2 average

CH1: NOx, CH2: SO

2

, CH3: CO

2

, CH4: CO, CH5: O

2

, CH6: Corrected NOx,

CH7: Corrected SO

2

, CH8: Corrected CO, CH9: Corrected NOx average,

CH10: Corrected SO

2

average, CH11: Corrected CO average, CH12 : O

2 average

CH1: CO, CH2: O

2

, Peak alarm

CH1: NO, CH2: CO, CH3: O

2

, Peak alarm

CH1: CO

2

, CH2: CO, CH3: O

2

, Peak alarm

CH1: NO, CH2: SO

2

, CH3: CO, CH4: O

2

, Peak alarm

CH1: NO, CH2: SO

2

, CH3: CO

2

, CH4: CO, CH5: O

2

, Peak alarm

CH1: CO, CH2: O

2

, CH3: Corrected CO, CH4: Corrected CO average, CH5: O

2

average,

Peak alarm

CH1: NOx, CH2: CO, CH3: O

2

, CH4: Corrected NOx, CH5: Corrected CO,

CH6: Corrected NOx average, CH7: Corrected CO average, CH8: O

2 average, Peak alarm

CH1: CO

2

, CH2: CO, CH3: O

2

, CH4: Corrected CO, CH5: Corrected CO average,

CH6: O

2

average, Peak alarm

CH1: NOx, CH2: SO

2

, CH3: CO, CH4: O

2

, CH5: Corrected NOx, CH6: Corrected SO

2

,

CH7: Corrected CO, CH8: Corrected NOx average, CH9: Corrected SO

2 average,

CH10: Corrected CO average, CH11: O

2 average, Peak alarm

CH1: NOx, CH2: SO

2

, CH3: CO

2

, CH4: CO, CH5: O

2

, CH6: Corrected NOx,

CH7: Corrected SO

2

, CH8: Corrected CO, CH9: Corrected NOx average,

CH10: Corrected SO

2

average, CH11: Corrected CO average, CH12 : O

2 average, Peak alarm

INZ-TN2ZKJ-E

5 - 5

5.4 General operation

• Measurement mode

The measurement mode can be displayed up to 5 channels in a single screen. If 5 channels or more are to be displayed in a single screen, press the or key to scroll the channel one by one.

C H

C H

C H

C H

C H

NO

X

0-100

SO

2

0-100

CO

2

0-10

CO

0-100

O

2

0-25

C H

C H

C H

C H

C H

NO

X

0-100

SO

2

0-100

CO

2

0-10

CO

0-100

O

2

0-25 ppm ppm vol%

0 0 .

ppm vol% ppm ppm vol%

0 0 .

ppm vol%

ZERO

SPAN

Zero calibration

See 6.9.1.

Span calibration

See 6.9.2.

C H CV.

0-100

NO

X ppm

C H

0-100

SO

2 ppm

C H

0-100

CO ppm

C H

AV.

O

2

0-25

PEARK COUNTER

CO 500ppm 0 times/H vol%

5 - 6

• Menu mode displays;

Changeover of Range

Setting about Calibration

Alarm Setting

Setting of Auto Calibration

Setting of Auto Zero Calibration

Setting of Peak Alarm

Parameter Setting.

For the setting contents, refer to

“Chapter 6. Setting and calibration”.

ESC MODE

Measurement Mode

ESC MODE

INZ-TN2ZKJ-E

6. SETTING AND CALIBRATION

6.1 Changeover of range

This mode is used to select the ranges of measured components.

q During measurement, press the

MODE

key to display the User mode.

w Point the cursor to “Changeover of

Range”. Press the

ENT

key.

MODE

( )

ENT e The “Channel Selection” screen appears.

Press the or key until the “ ” cursor moved selects a desired CH

(component).

r After selection, press the

ENT

key.

Note) The range of O

2 correction instantaneous values and O

2

correction average values is automatically switched by changing the range of instantaneous value of each CH

(component).

t In the Range Setting screen that appears, move the cursor by pressing the or

key to select the range. (The range with a mark of is currently selected.) y After selection, press the

ENT

key.

u Measurement is conducted within the selected range. The range identification contact is closed with the low range

(Range 1), and open with the high range

(Range 2).

Note) If the Remote Range is set to ON, the changeover of range cannot be performed on the screen.

To close “Changeover of Range”

To close Changeover of Range, or cancel

A previous screen will return.

INZ-TN2ZKJ-E

( )

ENT

( )

ENT

End of Range Selection

6 - 1

6.2 Calibration setting

This mode is used to set calibration concentration and actions. The calibration setting involves calibration concentration, zero calibration, calibration range and auto calibration component.

6.2.1 Setting of calibration concentration

It allows you to set concentrations of the standard gas (zero and span) of each channel used for calibration.

MODE q During measurement, press the

MODE

key to display the User mode.

w Point the cursor to “Setting about Calibration” by pressing the or key.

Press the

ENT

key.

ENT e In the “Setting about Calibration” screen that appears, point the cursor to “Calibration Value” by pressing the or key. Press the

ENT

key.

r In the “Calibration Concentration CH

Selection” screen that appears, point the cursor to CH you want to set by using the

or key. Press the

ENT

key.

( )

ENT

6 - 2

( )

ENT

INZ-TN2ZKJ-E

t In the “Calibration Concentration Selection” screen that appears, select any concentration item you want to set by pressing the ,

key.

y In the “Calibration Concentration Value

Setting” screen that appears, enter calibration gas concentration values (zero and span).

For value entry, press the or key, and a 1-digit value increases or decreases. By pressing the , the digit moves.

After setting, save the entry by pressing the

ENT

key. The saved value becomes valid from the next calibration process.

Note) Enter the set values corresponding to each range. When the O

2

measurement uses atmospheric for the zero gas, set the concentration value to 20.60. When the cylinder air is used, set to the concentration value as indicated on the cylinder.

To close the setting

To close the calibration concentration value setting process or cancel this mode midway,

( )

ENT

Cursor for setting value

ENT

End of Calibration

Concentration Setting

A previous screen will return.

Setting range of values

NOx, SO

2

, CO

2,

CO, CH

4

, external O

2 measurement and buit-in paramagnetic O

2 sensor

Span gas: 1 to 100% of full scale

(Full scale (FS) is the same as each range value.)

External Zirconia O

2

measurement

The setting cannot be performed beyond the range.

Zero gas: 5 to 25 vol%

Span gas: 0.01 to 5 vol%

INZ-TN2ZKJ-E

6 - 3

6.2.2 Setting of manual zero calibration

When zero calibration is made manually, set either all measurement components should be calibrated simultaneously or each component should be calibrated while selecting one by one.

MODE q During measurement, press the

MODE

key to display the User mode.

w Point the cursor to “Setting about Calibration” by pressing the or key.

Press the

ENT

key.

e In the “Setting about Calibration” screen that appears, point the cursor to “About

ZERO Calibration” by pressing the or

key. Press the

ENT

key.

( )

ENT r In the “Manual Calibration CH Selection” screen that appears, point the cursor to CH you want to set by using the or

key. Press the

ENT

key.

( )

ENT

6 - 4

( )

ENT

INZ-TN2ZKJ-E

t In the “Manual Calibration Selection” screen that appears, select “at once” or

“each” by pressing the or key.

When selecting “at once”, the CH (components) to be set can be zero-calibrated at the same time. When selecting “each”, either of the CH (components) to be selected is zero-calibrated. After setting, press the

ENT

key.

To close “About ZERO Calibration”

To close “About ZERO Calibration” setting or to

A previous screen will return.

( )

ENT

End of Manual Calibration Setting

Example

Whether “each” or “at once” can be determined for each CH (component).

•Setting “each”

Select the CH (component) on the manual zero calibration screen and then perform

zero calibration.

•Setting “at once”

At a manual zero calibration, zero of CH (components) for which “at once” was selected can simultaneously be calibrated.

* When the cylinder air or atmospheric air is used for the zero gas, select “At once”.

Manual Calibration screen

When setting all components to “each”:

When setting all components to “at once”:

A single cursor will appear.

INZ-TN2ZKJ-E

Cursors will appear at all components where “at once” is set.

6 - 5

6.2.3 Setting of calibration range

This mode is used to set if the range of each CH (component) at the calibration (manual calibration or auto calibration) should be calibrated with a single range or 2 ranges.

MODE q During measurement, press the

MODE

key to display the User mode.

w Point the cursor to “Setting about Calibration” by pressing the or key.

Press the

ENT

key.

e In the “Setting about Calibration” screen that appears, point the cursor to “About

Calibration Range” by pressing the or

key. Press the

ENT

key.

( )

ENT

6 - 6 r In the “Calibration Range CH Selection” screen that appears, point the cursor to the CH you want to set by pressing the

or key. Press the

ENT

key.

( )

ENT

( )

ENT

INZ-TN2ZKJ-E

t In the “Calibration Selection” screen that appears, select “both” or “current” by pressing the or key.

• When selecting “both”, Range 1 and

Range 2 of the set CH are calibrated together. When selecting “current”, the range alone displayed at the set CH is calibrated.

To close “Setting of Calibration Range”

To close “Setting of Calibration Range” or to cancel this mode midway, press the

( )

ENT

End of Manual Calibtation Setting

Example

CH1

NOx

Range 1: 0 to 100 ppm

Range 2: 0 to 2000 ppm both

CH2

SO

2

Range 1: 0 to 100 ppm

Range 2: 0 to 2000 ppm current

CH1: Range 1 and Range 2 are calibrated together with zero and span calibration.

CH2: Only currently displayed range is calibrated with zero and span calibration.

Caution

When calibrating the span gas in the interlocked ranges, the same values should be set in both ranges.

Manual Calibration screen

When setting NOx and CO to “both”

Two cursors will appear in both ranges (CH1 and CH4).

INZ-TN2ZKJ-E

6 - 7

6.2.4 Setting of auto calibration component

It sets the CH (component) to be calibrated in the auto calibration.

q During measurement, press the

MODE

key to display the User mode.

w Point the cursor to “Setting about Calibration” by pressing the or key.

Press the

ENT

key.

MODE

( )

ENT e In the “Setting about Calibration” screen that appears, point the cursor to “Auto

Calibration Components” by pressing the

or key. Press the

ENT

key.

6 - 8 r In the “Auto Calibration Components” selection screen that appears, point the cursor to the CH you want to set by pressing the or key. Press the

ENT

key.

( )

ENT

( )

ENT

INZ-TN2ZKJ-E

t In the “Auto Calibration Selection” screen that appears, select “enable” or

“disable” by pressing the or key.

After setting, press the

ENT

key.

To close "Auto Calibration

Component" setting

To close “Setting of Auto Calibration

Component” or to cancel this mode midway, press the ESC key.

A previous screen will return.

Example

( )

ENT

End of Manual Calibtation Setting

Caution

The components which were set to “enable” is calibrated to Zero at once at the time of auto calibration regardless of setting in “6.2.2 Setting of manual zero calibration”.

INZ-TN2ZKJ-E

6 - 9

6.3 Alarm setting

6.3.1 Setting of alarm values

This mode is used to set the upper or lower limit value to provide an alarm output during measurement. Before changing the alarm setting, set the ON/OFF to OFF.

MODE q During measurement, press the

MODE

key to display the User mode.

w Point the cursor to “Alarm Setting” by pressing the or key. Press the

ENT key.

( )

ENT e After the alarm setting CH selection screen has appeared, point the cursor to the CH you want to set by pressing or

key .

Press the

ENT

key.

ENT

6 - 10 r After the alarm item selection screen has appeared, operate the or key until the cursor is aligned with a desired item and press the

ENT

key.

Caution

Set the values so that upper limit value > lower limit value and that (upper limit value lower limit value) > hysteresis.

( )

ENT

INZ-TN2ZKJ-E

After setting, the alarm setting is now completed by pressing the

ENT

key.

Cursor for setting value

To close the "Alarm Setting"

To close the “Alarm Setting” or to cancel this mode midway, press the ESC key.

A previous screen will return.

( )

ENT

End of Alarm Setting

Description of setting items

Upper limit value: Sets the upper limit value (concentration) of alarm.

Lower limit value: Sets the lower limit value (concentration) of alarm.

Contact action: Selects one of upper limit alarm, lower limit alarm, and upper limit or lower limit alarm.

Upper ... Alarm contact closes when above upper limit alarm.

Lower ... Alarm contact closes when below lower limit alarm.

Upper or lower ... Alarm contact closes when above upper limit value or below lower limit value.

ON/OFF: Enables the alarm function if set at ON, or disables the alarm function if set at OFF.

* The upper limit value cannot be set below the lower limit value, and the lower limit value cannot be set above the upper limit value.

If it is desired to set the upper limit value below the lower limit value already stored in the memory, reduce the lower limit value beforehand, and vice versa.

Typical on-screen display when an alarm occurs

When an upper limit alarm has occurred, the “High alarm” message lights at CH (component)

(“Low alarm” at lower limit alarm).

C H NO

X

0-100

C H

SO

2

0-100

C H CO

2

0-10

C H

CO

0-100

C H O

2

0-25

- - - ppm ppm vol% ppm vol%

Caution

For 10 minutes after turning on power, the alarm judgment is inactive.

INZ-TN2ZKJ-E

6 - 11

6.3.2 Hysteresis setting

To prevent chattering of an alarm output near the alarm setting values, set the value of hysteresis.

q In the “Alarm Setting CH” Selection” screen that appears, point the cursor to

“Hysteresis” by pressing the or key. Press the

ENT

key.

w In the “Hysteresis Value Setting” screen that appears, enter hysteresis values.

For the value entry, 1-digit value is increased or decreased by pressing the

or key, and pressing the key moves the digit. After setting, press the

ENT

key.

To close "Hysteresis Setting"

To close the “Hysteresis Setting” or cancel the mode midway, press the

ESC

key.

A previous screen will return .

( )

ENT

Setting range

0 to 20% of full scale

A full scale means each range provides a full scale of width.

( )

ENT

End of Hysteresis Setting

Caution

The hysteresis is common to all CHs

(components).

Hysteresis (In case of upper limit alarm)

An alarm output is turned ON if measurement value exceeds the upper limit value as shown below. Once the alarm output has been turned ON, it is not turned OFF as long as the indication does not fall below the hysteresis width from the upper limit value.

Upper limit value

Hysteresis width

6 - 12

Alarm

ON (conductive)

OFF (open)

INZ-TN2ZKJ-E

6.4 Setting of auto calibration

6.4.1 Auto calibration

Auto calibration is automatically carried out at the time when zero calibration and span calibration are set.

Before changing the setting of auto calibration, set the ON/OFF to OFF.

q During measurement, press the

MODE

key to display the User mode.

w Point the cursor to “Setting of Auto

Calibration” by pressing the or key. Press the

ENT

key.

MODE e In the “Setting of Auto Calibration” screen that appears, point the cursor to any item you want to set by pressing the

or key. Press the

ENT

key.

r In the “Auto Calibration Parameter

Setting” screen that appears, perform the value entry or the setting. For the value entry or setting change, use the or

key. To change the setting, use the

key to move the cursor to the right.

After setting, press the

ENT

key, and auto calibration is carried out by the entered setting value.

Description of setting items

• Start Time : Setting at the first calibration

(day of the week, hour, minute)

• Cycle : A period between the start time of one calibration and another

(unit : hour/day)

• Flow Time : The time required for the calibration gas

• ON/OFF to be replaced in the cell

: ON/OFF of auto calibration

To close "Setting of Auto calibration"

To close the "Setting of Auto calibration" or cancel this mode midway, press the ESC key.

A previous screen will return.

INZ-TN2ZKJ-E

( )

ENT

( )

ENT

ENT

End of Aoto Calibration Setting

6 - 13

“Auto calibration status” and “maintenance status” contact outputs are closed during auto calibration, and are open in other cases. When hold is set to ON, the contacts are closed during holding.

Example

Start Time

Cycle

Flow Time

ON/OFF

SUN

1

350

ON

12:00 day sec

In case where auto calibration is carried out at the above setting.

Sunday

12:00

Cycle

Monday

12:00

Tuesday

12:00

: Auto calibration

Zero calibration

CHI

Span calibration

CH2

Span calibration

CH3

Span calibration

CH4

Span calibration

CH5

Span calibration

350sec 350sec 350sec

Flow time

(An example of “CH1: through CH5: enable”, as given in Item 6.2.4 “Auto Calibration

Components”)

Setting range

Cycle

Flow time

: 1 to 99 hours or 1 to 40 days (initial value 7days)

: 60 to 599 sec        (initial value 300sec)

Caution

When an auto calibration starts, the measurement screen automatically appears.

Any operation other than “Stop Auto Calibration” (see Item 6.4.2) is not permitted during auto calibration. “Stop Auto Calibration” cannot be performed with the key lock to ON. To cancel auto calibration forcedly, set the key lock to OFF and then execute “Stop Auto Calibration”.

Remote start

Whether the auto calibration is set at ON or OFF, an auto calibration is available by keeping the remote start input closed for at least 1.5 seconds.

Remote start input

Closed (keep at least 1.5 sec.)

Open

6 - 14

INZ-TN2ZKJ-E

6.4.2 Forced stop of auto calibration

This mode is used to cancel the auto calibration forcedly.

q In the User mode that is displayed, point the cursor to “Setting of Auto Calibration” by pressing the or key.

Press the

ENT

key.

( )

ENT w In the “Setting of Auto Calibration” item selection screen that appears, point the cursor to “Stop Auto Calibration” by pressing the or key. Press the

ENT key.

e “Stop Auto Calibration” is inverted. A message appears, prompting you to verify that you want to cancel or continue auto calibration. To cancel the auto calibration, press the

ENT

key. If you press the

ESC

key, auto calibration is not stopped.

( )

ENT

INZ-TN2ZKJ-E

6 - 15

“Auto Calibration” screen

Example

In case where setting the auto calibration components (see Item 6.2.4) to

“CH1: enable” and “CH2: enable”

• Zero calibration

A message, “Zero cal.” blinks at CH1 and CH2.

C H

C H

C H

C H

C H

NO

0-100

X

SO

0-100

2

CO

2

0-10

CO

0-100

O

2

0-25 ppm ppm vol% ppm vol%

• CH1 span calibration

A message, “Span cal.” blinks at CH1.

C H

C H

C H

C H

C H

NO

0-100

X

SO

2

0-100

CO

2

0-10

CO

0-100

O

2

0-25 ppm ppm vol% ppm vol%

• CH2 span calibration

A message, “Span cal.” blinks at CH2.

C H

C H

C H

C H

C H

NO

X

0-100

SO

0-100

2

CO

2

0-10

CO

0-100

O

2

0-25 ppm ppm vol% ppm vol%

Caution

During auto calibration, any key operation is not permitted other than operations such as key lock ON/OFF and “Stop Auto Calibration”.

When the key lock is set at ON, even the “Stop Auto Calibration” cannot be used.

To stop “Auto Calibration” forcedly, set the key lock to OFF and then execute

“Stop Auto Calibration”.

6 - 16

INZ-TN2ZKJ-E

6.5 Setting of auto zero calibration

6.5.1 Auto zero calibration

Auto zero calibration is automatically carried out at the time when zero calibration is set.

Components for which a calibration is to be made are determined by setting of auto calibration component in Item 6.2.4.

Before changing the setting of auto zero calibration, set the ON/OFF to OFF.

q During measurement, press the

MODE

key to display the User mode.

w Point the cursor to “Setting of Auto Zero

Calibration” by pressing the or key. Press the

ENT

key.

e In the “Setting of Auto Zero Calibration” screen that appears, point the cursor to any item you want to set by pressing the

or key. Press the

ENT

key.

r In the “Auto Zero Calibration Parameter

Setting” screen that appears, perform the value entry or the setting. For the value entry or setting change, use the or

key. To change the setting, use the

key to move the cursor to the right.

After setting, press the

ENT

key, and auto zero calibration is carried out by the entered setting value.

Description of setting items

• Start Time : Setting at the first calibration

(day of the week, hour, minute)

• Cycle : A period between the start time of one calibration and another

(unit : hour/day)

• Flow Time : The time required for the calibration gas

• ON/OFF to be replaced in the cell

: ON/OFF of auto zero calibration

MODE

( )

ENT

( )

ENT

To close "setting of Auto Zero Calibration"

To close the Setting of Auto Zero Calibration or cancel this mode midway, press the ESC key.

A previous screen will return.

ENT

End of Auto Zero Calibration Setting

INZ-TN2ZKJ-E

6 - 17

“Auto calibration status” and “maintenance status” contact outputs are closed during auto calibration, and are open in other cases. When hold is set to ON, the contacts are closed during holding.

Example

Start time

Cycle

Flow time

ON/OFF

SUN

1

300

ON

12:00 hour sec

In case where auto zero calibration is carried out at the above setting.

Sunday

12:00

Cycle

Monday

12:00

Tuesday

12:00

: Auto zero calibration

Zero calibration

300sec

Flow time

(An example of “CH1: through CH5: enable”, as given in Item 6.2.4 “Auto Calibration

Components”)

Setting range

Cycle

Flow time

: 1 to 99 hours or 1 to 40 days (initial value 7days)

: 60 to 599 sec        (initial value 300sec)

Caution

When an auto zero calibration starts, the measurement screen automatically appears.

Any operation other than "Stop Auto Zero Calibration" (see Item 6.5.2) is not permitted during auto zero calibration. “Stop Auto Zero Calibration” cannot be performed with the key lock to ON. To cancel auto zero calibration forcedly, set the

• key lock to OFF and then execute “Stop Auto Zero Calibration”.

If the auto calibration period and auto zero calibration period have overlapped, the auto calibration is retained, ignoring the auto zero calibration of that period.

6 - 18

INZ-TN2ZKJ-E

6.5.2 Forced stop of auto zero calibration

This mode is used to cancel the auto zero calibration forcedly.

q In the User mode that is displayed, point the cursor to “Setting of Auto Zero

Calibration” by pressing the or key. Press the

ENT

key.

( )

ENT w In the “Setting of Auto Zero Calibration” item selection screen that appears, point the cursor to “Stop Auto Zero Calibration” by pressing the or key.

Press the

ENT

key.

e “Stop Auto Zero Calibration” is inverted.

A message appears, prompting you to verify that you want to cancel or continue auto zero calibration. To cancel the auto zero calibration, press the

ENT

key. If you press the

ESC

key, auto zero calibration is not stopped.

( )

ENT

INZ-TN2ZKJ-E

6 - 19

“Auto Zero Calibration” screen

Example

In case where setting the auto calibration components (see Item 6.2.4) to

“CH1: enable”, “CH2: enable” and “CH3 to CH5:disable”

• Zero calibration

A message, “Zero cal.” blinks at CH1 and CH2.

C H

C H

C H

C H

C H

NO

0-100

X

SO

0-100

2

CO

2

0-10

CO

0-100

O

2

0-25 ppm ppm vol% ppm vol%

Caution

During auto zero calibration, any key operation is not permitted other than operations such as key lock ON/OFF and “Stop Auto Zero Calibration”.

When the key lock is set at ON, even the “Stop Auto Zero Calibration” cannot be used.

To stop “auto zero calibration” forcedly, set the key lock to OFF and then execute

“Stop Auto Zero Calibration”.

6 - 20

INZ-TN2ZKJ-E

6.6 Peak alarm setting

When the peak number of times CO concentration exceeds the upper limit value during measurement exceeds the set number, an alarm is provided.

The peak alarm and this setting screen appear only when an option is added.

MODE q Press the

MODE

key in the Measurement mode, and the User mode appears.

w Point the cursor to “Setting of Peak

Alarm” by pressing the or key.

Press the

ENT

key.

e In the “Peak Alarm Setting” item selection screen that appears, point the cursor to any item you want to set by pressing the or key. Press the

ENT

key.

( )

ENT r Then, enter numeric values and perform the setting.

Entering the numeric values or setting the items should be carried out by using the or key.

After setting, press the

ENT

key, and the set values are saved.

Description of setting items

• Peak Alarm : ON/OFF of peak alarm

• Alarm Value : If measuring value exceeds the set alarm value, a peak counter counts

1 time.

• Alarm Count : When a peak in excess of the setting

time occurs, a peak count alarm output is provided.

• Hysteresis : To prevent possible chattering when the measuring value may exceed the set peak concentration by only

1 time, the peak count has an allowance in the hysteresis width.

( )

ENT

ENT

End of Peak Alarm Setting

INZ-TN2ZKJ-E

6 - 21

Action of peak alarm

Example

If CO concentration exceeds the alarm value, counting will begin. If the number of peaks is over the set times per hour, a peak alarm contact output becomes closed (ON). If it is less than the set times per hour, it is open (OFF). Since 5 times of peaks /hour is marked at q section from the above graph, the peak count alarm is turned ON. Since peaks of more than 5 times per 1 hour occur at the interval between q and w , the peak count alarm remains ON. Since at w , peaks are reduced to 4 times per hour, it is turned OFF.

Like the hysteresis of the alarm setting , the hysteresis prevents possible chattering when measured gas is fluctuated near the alarm value.

* For 10 minutes after the power is turned ON, a peak alarm counting is not carried out.

Releasing peak alarm

To release the peak alarm, set the peak alarm to OFF.

Turning on the peak alarm initiates counting from 0.

6 - 22

INZ-TN2ZKJ-E

6.7 Parameter setting

It allows you to carry out the parameter setting such as time, key lock, etc., as required. Items to be set are as follows:

Description of setting items

• Current Time

• Key Lock

• Remote Range

• Output Hold

• Average Value Reset

• Response time

• Average Period

• Maintenance mode

: Sets the current day of the week, hour and time.

: Sets with ON/OFF so that any key operation except the key lock OFF cannot be performed.

: Sets with ON/OFF whether the Range Selection is made valid or invalid by external input.

: Sets whether Calibration Output is held or not.

: Resets the average value.

: Sets the response time of electrical system.

: Sets the moving average time.

: Enters passwords to switch to the Maintenance mode.

* For the maintenace mode, see Item 6.8.

MODE q To display the User mode, press the

MODE key in the measurement mode.

w Point the cursor to “Parameter Setting” by pressing the or key. Press the

ENT

key.

( )

ENT e In the “Parameter Setting” screen that appears, point the cursor to any item you want to set by pressing the or key. Press the

ENT

key.

( )

ENT

INZ-TN2ZKJ-E

6 - 23

r In the Parameter Setting screen that appears, enter the numeric values and set the items. Entering the numeric values or setting the items should be carried out by using the or key. To move the cursor to the right, press the key.

After setting, press the

ENT

key, that the parameter setting is carried out with the value you set.

To close Parameter Setting screen

To close the “Parameter Setting” screen or cancel this mode midway, press the

A previous screen will return.

Setting Range

• Response time

• Average period

• Maintenance mode

: 1 to 60sec.

: 1 to 59 min or 1 to 4 hours

ENT

End of Parameter Setting

When setting the unit of 1 to 59 minutes is terms of minute

or 1 to 4 hours with hour

: 0000 to 9999

(Initial value: 15 sec)

(Initial value: 1 hour)

(Initial value: 0000)

Remote Range

A range can be switched via an external input by setting the Remote Range to ON (The switching action affects all of instantaneous value, O

2

correction value, O

2 correction average value and O

2 average value.) If the Remote Range is set to OFF, the external input becomes invalid.

Opening the input gives the High range, or shorting the input gives the Low range.

For the terminal input, refer to Item 3.4, Wiring.

Switching the range cannot be performed by on-screen operation when the Remote Range is set to

ON.

Note) In case of 1 range meter, this function is overridden.

Output Hold

By setting an output hold to ON, an output signal of each channel are held during the calibration

(manual calibration and auto calibration) and for the gas flow time (refer to Item 6.4, Setting of

Auto Calibration). Regardless of Hold ON/OFF setting, an output signal can be held via an external input.

(1) Manual calibration

ZERO or

SPAN

Calibration

6 - 24

Output hold

Hold extending time.

Time set to gas flow time (See Item 6.4 Auto Calibration.)

INZ-TN2ZKJ-E

(2) Auto calibration

Auto calibration start

Calibration

Auto calibration end

Output hold

(3) External hold

Remote hold

Hold extending time.

Time hold set to the gas flow time

Close

Open

Output hold

Note) With the external hold input, auto calibration status and maintenance status contact are conductive.

(4) Screen display during Holding

The “Holding” message blinks on the measuring screen.

Since the screen displays the process of calibration is displayed during the manual calibration,

“Holding” is not displayed even if the output signal is held, but the screen is displayed with the hold extending time.

(5) If calibration is cancelled after the calibration gas is supplied regardless of during manual calibration or auto calibration, the holding extending time will be performed.

Average value reset

This mode is used to clear all average values O

2

correction average and O

2

average, and restarts averaging. All average values are reset at a time. The indication value and output value is 0 ppm, vol% or so at the time of the reset input (Refer to the average peripd).

Close (hold at least 1.5 sec.)

Open Reset input

So long as close, resetting lasts.

At the edge of changing from closing to opening, the average action restarts.

Response time

The response time of the electrical system can be changed.

Setting is available by components.

Note) It does not provide exact seconds for the setting time, but it gives a guide of the setting time.

The setting value can be modified as

requested by the customer.

INZ-TN2ZKJ-E

6 - 25

Average period

It allows you to set an average period of the average value of O

2 correction and O

2 average.

It enables you to set an average time of 1 to 59 minutes (1-minute step) or 1 to 4 hours (1-hour step).

Changing the setting resets the average value of

O

2 correction and O

2

average value. (Pressing the

ENT

validates the resetting only for components whose setting was changed.)

Example of average action

In case the average period was set to 1 hour.

Reset

Average value

Time

1 Time

· Sampling occurs every 30 seconds.

·

Every 30 seconds, the average for last 1 hour (time setting) is output.

· At the instant of resetting, zero is assumed for all past values. It means that the average value will not be correct for 1 hour after resetting.

Maintenance mode

To open the maintenance mode, enter a password. After entering the password, press the

ENT

key.

The password can be set for the Password Setting in the Maintenance mode. A password is set to

“0000” before factory-shipment. This value is available for the Maintenance mode.

6 - 26

INZ-TN2ZKJ-E

6.8 Maintenance mode

This mode is used for check of sensor input values, display of error log files or setting of passwords, etc. First, enter a password and then use it from the next operation. This mode is displayed by selecting the Maintenance Mode from “Item 6.7 Parameter Setting”.

q Select the Maintenance mode from the

Parameter Setting screen to display the

Password Setting screen.

w Enter the password, and the Maintenance

Mode item selection screen will be displayed. Point the cursor to the item you want to set by pressing the or key and press the

ENT

key.

e Next, each Maintenance screen is displayed.

Note) “To Factory Mode” is used for our service engineers only. Refrain from using this mode.

( )

ENT

Each “Maintenance” screen

• Sensor Input Value screen

Description of Sensor Input Value screen

NOx M

NOx C

: NOx sensor input value

: NOx interference compensation sensor input value

SO

2

SO

2

M

C

: SO

: SO

2

2

sensor input value

interference compensation sensor input value

CO

2

M

CO

2

C

: CO

: CO

2

2

sensor input value

interference compensation sensor input value

CO M

CO C

: CO sensor input value

: CO interference compensation sensor input value

Temperature: temperature sensor input value

O

2

: O

2

sensor input value

• Error Log screen

Description of Error Log screen

Error history

For error number, date and time (days, period) of occurrence, channel and other details of error, refer to Item 8.1 Error message.

INZ-TN2ZKJ-E

6 - 27

• Password Setting screen

Description of Password Setting screen

It enables you to set a password to be used when switching the parameter setting mode to the maintenance mode.

Note) The password set herein must be managed for safety. Failure to enter the correct password will not open the Maintenance mode.

Caution

If the following operation is maladjusted, the measurement may be adversely and excessively affected. Carry out the operation with utmost attention.

. Optical adjustment screen

For details of this item, refer to “Item 7.3.3

Optical zero adjustment method”.

The selenoid valve driving signal for zero gas is turned ON by pressing the

ENT

key.

. Moisture interference adjustment screen

For details of this item, refer to “Item 7.3.4

Moisture interference adjustment method”.

Description of moisture interference adjustment screen

In values on the right side of screen, the moisture interference for each component is already offset.

By “ ” or “ ” key, move “ ” cursor to a desired CH (component).

Make sure the gas for moisture interference flows. By “ “ or “ ” key, get the value after moisture interference offset will get near 0 and press the “

ENT

” key to register the moisture interference corrected value into the memory.

6 - 28

ENT

INZ-TN2ZKJ-E

Caution

Since an interference compensation detector is not provided if the 1st range is beyond 0 to

10 vol%, no interference adjustment can be performed (no need).

• Setting of O

2

correction reference value

Set the O

2

correction reference value by selecting a numerical value by or key or shifting a digit by key.

Calculated value shoud be referred to O

2

correction values given in “Item 5.3 Outline of display screen”.

Setting range

01 to 19%

• Setting of station No. for communication function

Set the station NO. by selecting a numerical value by or key or shifting a digit by key.

00 to 31

Setting range

(Initial = 00)

* Please refer to another manual (INZ-TN513327-E) about the communication function.

INZ-TN2ZKJ-E

6 - 29

6.9 Calibration

6.9.1 Zero calibration

It is used for zero point adjustment. For zero calibration gas, suited for an application should be used according to “(3) Standard gas in Item 3.3 Sampling”.

q Press the

ZERO

key on the Measurement screen to display the Manual Zero Calibration screen.

w Select the CH (component) to be calibrated by pressing the or key.

After selection, press the

ENT

key, and zero gas will be supplied.

Caution

For the CH (components) that is set to

“both” in the “Zero Calibration” of the

Calibration Setting mode, zero calibration is also carried out at the same time.

e Wait until the indication is stabilized with the zero gas supplied. After the indication has been stabilized, press the

ENT

key. Zero calibration in range selected by the cursor is carried out.

To close "Zero Calibration"

To close the Zero Calibration or cancel this mode midway, press the ESC key. A previous screen will return.

ZERO

( )

ENT

ENT

6 - 30

ENT

To Measurement screen after executing Manual Zero Calibration

INZ-TN2ZKJ-E

6.9.2 Span calibration

It is used to perform a span point adjustment. Supply calibration gas with concentration set to the span value to perform the span calibration. For the span calibration gas for the NO

X

, SO

2

, CO

2

,

CO measurement, use the standard gas with a concentration of 90% or more of the range value.

For the span calibration gas for the O

2

measurement, use the standard gas with a concentration of

90% or more of the range value when measuring with the built-in O

2

sensor, and use the standard gas of about 2 vol% when measuring with an external zirconia O

2

sensor.

SPAN q Press the

SPAN

key on the Measurement screen to display the Manual Span Calibration screen.

( )

ENT w Select CH (component) to be calibrated by pressing the or key and press the

ENT key. The calibration gas is supplied.

Caution

When “both” from “Calibration Range” of the

Calibration Setting mode is set, span calibration is performed together with 2 Ranges.

e Wait until the indication is stabilized in the state where the calibration gas is supplied.

After the indication has been stabilized, press the

ENT

key. Span calibration of

Range selected by the cursor is performed.

To close "Span Calibration"

To close the “Span Calibration” or cancel

A previous screen will return.

INZ-TN2ZKJ-E

ENT

ENT

To Measurement screen after executing Manual Span Calibration

6 - 31

7. MAINTENANCE

7.1 Daily check

(1) Zero calibration and span calibration q Perform zero calibration. For the calibration procedures, refer to “Item 6.9.1 Zero calibration”.

w Then, perform span calibration. For the calibration procedures, refer to “Item 6.9.2 Span calibration”.

e Zero calibration and span calibration should be carried out once a week, as required.

(2) Flow rate check q Sampling gas flow and purge gas flow are as follows:

Sampling gas flow:

Purge gas flow:

0.5L/min

±

0.2L/min

About 1L/min w Check and maintenance should be carried out every day, as required.

7.2 Daily check and maintenance procedures

Table 7.1 Maintenance and check table

INZ-TN2ZKJ-E

7 - 1

7.3 Maintenance of analyzer unit

7.3.1 Cleaning method for sample cell (pipe cell)

This section is strictly factory adjusted. Handle it with utmost attention.

If it is absolutely required, contact us.

q Turn off the power switch, stop the sample gas, and allow the zero gas to flow for several minutes to purge the cell interior.

Loosen the setscrew (2 pieces) from the top cover and remove it.

w Remove the internal gas inlet tube.

e Loosen both right and left screws for cell holding plate.

Remove the sample cell only.

r Turn to the left the sample cell window and remove it from the sample cell (see

Fig. 7-1).

t For cleaning the window and cell inside surface, first eliminate coarse dust by soft brush or the like and then wipe them by soft rag.

The window is easy to get scratched.

Pay utmost attention so as not to damage it.

y After the end of sample cell cleaning, mount the cell in place and proceed to running.

After cleaning sample cell, be sure to perform optical zero adjustment (see

Item 7.3.3) and moisture interference compensation adjustment (see Item

7.3.4).

Top cover

Screw

Cell holding plate

Gas outlet

Sample cell

Screw

Cell holding plate

Gas inlet

Reference cell

Caution

A slightly corroded infrared transmission window or sample cell can be remedied by gently rubbing with chromium oxide powder on cleaning cloth but an excessively corroded one must be replaced.

When cleaning, do not exert an excessive stress.

7- 2

INZ-TN2ZKJ-E

INZ-TN2ZKJ-E

Infrared transmission window

Window holder

O ring

Bonded together

Sample gas inlet

Cell

Sample gas outlet

Fig. 7-1 Structure of sample cell (pipe cell)

7 - 3

7.3.2 Cleaning method for sample cell (block cell) q Turn off the power switch, stop the sample gas, and allow the zero gas to flow for several minutes to purge the cell interior.

Loosen the setscrew (2 pieces) from the top cover and remove it.

w Remove the internal gas inlet tube.

e Loosen the 2 detector set bolts.

Note) The distribution cell, block cell and detector are fastened by the same bolts.

r Using the furnished cell mounting tool, turn the window fixture to the left and remove it from the cell. (See the structure of sample cell (block cell) in

Fig. 7-2.) t For cleaning the infrared transmission window and cell inside surface, first eliminate coarse dust by soft brush or the like and then wipe them by soft rag.

The window is easy to get scratched.

Pay utmost attention so as not to damage it.

y After the end of sample cell cleaning, mount the cell in place and proceed to running.

After cleaning sample cell, be sure to perform optical zero adjustment (see

Item 7.3.3) and moisture interference compensation adjustment (see Item

7.3.4).

Top cover

Gas inlet tube set cap nut

Detector set bolt

Infrared transmission window

Inner O ring

Outer O ring

O ring holder

Window fixture

Caution

A slightly corroded infrared transmission window or sample cell can be remedied by gently rubbing with chromium oxide powder on cleaning cloth but an excessively corroded one must be replaced.

When cleaning, do not exert an excessive stress.

7- 4

INZ-TN2ZKJ-E

Cell mounting tool

Window fixture

O ring holder

Outer O ring

Inner O ring

Infrared transmission window

Sample gas outlet

Sample cell unit

Sample gas inlet

Reference cell unit

Hole for bolt for fixing together with distribution cell and detector

Structure of sample cell (of 32, 16, 8, 4, 2 mm long)

(sample cell and reference cell are integrated)

Note) Use the dedicated cell mounting tool (furnished).

Fig. 7-2 Structure of sample cell (block cell)

INZ-TN2ZKJ-E

7 - 5

7.3.3 Optical zero adjustment method (optical balance adjustment)

Caution

If the following operation is maladjusted, the measurement may adversely be affected. If you are not trained for adjustment, do not carry out this operation but contact the distributor or our serviceman.

The adjustment is performed at reassembly after removing the sample cell, etc. for cleaning, etc.

q Remove the top cover. Allow dry N

2 or air to flow through the analyzer unit sample gas inlet until the reading stabilizes. The sample gas is introduced directly to the INLET of analyzer unit through the gas cylinder.

Piping at optical balance adjustment

Optical system 2

(measuring unit 2)

Short

(not required for measuring unit 1 only)

Open

Optical system 1

(measuring unit 1)

Introduce directly from cylinder w Proceed to an optical adjustment in the maintenance mode. The display on the operation panel of the main unit is as illustrated on the right. (If values are within

±

50, the balance adjustment is unnecessary.) a b c d e f g h

7- 6

<Correspondence between measurement detector and indicated position>

No. of components to be measured

1-component meter a b c d e

– f

– g

– h

NO/SO 2

CO 2 /CO

Main

NO

Main

CO 2

Main

Comp

NO

Comp

CO 2

Comp

SO 2

Main

CO

Main

SO 2

Comp

CO

Comp

NO/CO

NO

Main

NO

Comp

– –

CO

Main

CO

Comp

– –

3-component meter

NO/SO 2 /CO

NO

Main

NO

Comp

SO 2

Main

SO 2

Comp

CO

Main

CO

Comp

– –

4-component meter

NO/SO 2 /CO 2 /CO

NO

Main

NO

Comp

SO 2

Main

SO 2

Comp

CO 2

Main

CO

Main

CO

Comp

* O 2 is excluded from the number of components.

* “Main” is signal input value from the main detector of each component.

“Comp” is signal input value from interference compensation detector of each component.

If low range exceeds the range of 0 to 10vol%, detector signal of “comp” is not usable.

Sensor values of which are not included in measuring components should be ignored.

INZ-TN2ZKJ-E

e Carry out the adjustment in the procedure in r and subsequent.

When optical system 1 is adjusted. Get the values at a to d in 1-1 and 1-2 within ±

50 and as near 0 as possible.

When optical system 2 is adjusted. Get the values at e to h in 2-1 and 2-2 within ±

50 and as near 0 as possible.

Optical zero adjustment knob r Operate the optical zero adjustment knob to change the value displayed at a (or e ).

Optical system 2 t Move the beamadjustment plate sideview to change the value displayed at b (or f ).

y Move the beam adjustment plate sidewise to change the value displayed at c (or g ).

u Move the beam adjustment plate sidewise to change the value displayed at d (or h ).

i Sequentially repeat steps r to u so that all values will be within

±

50 and as near 0 as possible.

Detector set bolt

Optical system 1

* Adjust the beam adjustment plate which is the nearest to the zero adjustment knob first, and sequentially.

Beam adjustment plate o After the optical balance adjustment, mount the top cover of the analyzer unit, then carry out a moisture interference compensation adjustment, and perform zero and span calibrations.

* Before moving the beam adjustment plate, loosen the detector set bolts (just enough to make the plate movable for snug adjustment).

7.3.4 Moisture interference compensation adjustment method

Caution

If the following operation is maladjusted, the measurement may adversely be affected. If you are not trained for adjustment, do not carry out this operation but contact the distributor or our serviceman.

Proceed to an adjustment if excessively

(beyond

±

2% FS) affected by moisture inteference.

After the end of optical balance adjustment, be sure to carry out moisture inteference compensation adjustment.

q After warm-up, select the low range, allow dry gas (N

2

, air) to flow at 0.5 L/min and carry out zero calibration.

N

2

or air

Water Bubbler

INZ-TN2ZKJ-E

7 - 7

w Display the moisture interference compensation screen of the analyzer unit (see “6.8 Maintenance mode”). Then, introduce bubbled N

2

or air gas to the analyzer (shown on right).

e On the screen, select a CH (component) by

ENT

” key, press the “ ” or “ ” key until the analytical value is within

±

10 (as near 0 as possible), and press the “

ENT

” key to register it in the memory. (Exiting by “

ESC

” cancels the adjustment.)

Or, selecting the ALL CH and pressing the

ENT

” key, zeroes all components integrally.

(First, adjust all components by selecting

ALL and then perform fine adjustment for components one by one using UP and

DOWN keys.)

* If any components exceed the range of 0 to 10vol%, no adjustment can be performed (No interference compensation is required).

r After the end of adjustment for all components, return the piping to the original status and carry out zero and span calibrations.

7.3.5 Replacement of fuse on analyzer unit

Note) Before the following work, be sure to remove the cause of blown fuse (short, etc) and repair.

q Turn OFF the main power supply SW of the analyzer.

w Loosen setscrews (4 pcs.) from the top of the analyzer and remove the top cover.

e Turn the fuse holder (shown in Fig. 7-3) counterclockwise and put it out, and the cap will be removed. Remove the fuse

(250V AC/3A) from the holder. Replace it with a new one .

r Mount the fuse holder cap and cover for analyzer indication unit in the reverse procedure. Turn ON the power supply SW.

The work will be completed if the analyzer is normally worked.

Fuse holder

Front

Fig. 7-3

7- 8

INZ-TN2ZKJ-E

8 TROUBLE SHOOTING FOR ANALYZER

8.1 Error message

If errors occur, the following contents are displayed.

Error display

Error No.1

Error No.4

Error No.5

Error No.6

Error No.7

Error No.8

Error No.9

Error No.10

Error contents Probable causes

Motor rotation detection signal faulty

Zero calibration is not within.

Amount of zero calibration

(indication value) is over 50% of full scale.

• Motor rotation is faulty or stopped.

• Motor rotation detector circuit is faulty.

Note) Sector motor is a consumption part.

It is recommendable to exchange the motor once two years.

• Zero gas is not supplied.

• Zero is deflected much due to dirty cell.

• Detector is faulty.

• Optical balance is maladjusted.

Span calibration is not within the allowable range.

• Span gas is not supplied.

• Calibrated concentration setting does not

Amount of span calibration

(difference between indication value

match cylinder concentration.

• Zero calibration is not performed normally.

• Span is deflected much due to dirty cell.

and calibrated concentration) is over

50% of full scale.

• Detector sensitivity has deteriorated.

Measured values fluctuate too much during zero and span calibration.

• Calibration gas is not supplied.

• Time for flowing calibration gas is short.

Calibration is abnormal during auto calibration.

• Error corresponding to No. 4 to No. 8

occurred during auto calibration.

Output cable connection is improper.

• Wiring is detached between analyzer and

interface module.

• Wiring is disconnected between analyzer and

interface module.

When errors No. 1 and No. 10 occur, analyzing block error contact output is closed.

When errors No. 4 to No. 9 occurs, calibration error contact output is closed.

Screen display and operation at the occurrence of error

In case of Error No. 1 to No. 4, No. 6, No. 8 to No. 10

Display of error contents

Measurement screen

C H

NO

0-25

2

C H

SO

2

0-100

C H CO

2

0-10

C H CO

0-100

C H

O

2

0-25 ppm ppm vol% ppm vol%

ENT

ESC

the cause of an error, the error will be

displayed again.

• When more than one error occurs, pressing

the key moves to another error display.

INZ-TN2ZKJ-E

8 - 1

In case of Error No. 5 and No. 7

ENT

ENT

Calibrated forcedly

Calibration is continued. Unless another calibration error occurs, calibration is carried out to the end, the

Measurement screen returns.

C H

C H

C H

C H

NO

2

0-25

SO

2

0-100

CO

2

0-10

CO

0-100

C H

O

2

0-25

ESC ppm ppm vol% ppm vol%

Error log file

If error occurs, the history is saved in an error log file. The error log file exists in the maintenance mode.

Error log screen

Day of the week and time when an error occurred.

New

Component with which the error occurred.

Errors that occurred

Old

* Up to 14 errors can be saved in the error history; the oldest error will be deleted one by one every time a new occurs.

* If the power supply is turned OFF, the contents in the error log file will not be lost or damaged.

Deletion of error history

Press the

ENT

key on the above screen, and the “Error Log Clear” will be inverted. Further pressing the

ENT

key will clear the error history.

8 - 2

INZ-TN2ZKJ-E

9.

SPECIFICATIONS

9.1

General specifications

1. Standard specifications

Measurable component and measurement range:

NO

SO

2

CO

2

CO

CH

4

O

2

(built in)

O

2

(external zirconia)

Minimum range Maximum range

0 to 100 ppm 0 to 5000 ppm

0 to 100 ppm

0 to 50 ppm

0 to 10 vol%

0 to 100 vol%

0 to 100 ppm 0 to 100 vol%

0 to 500 ppm 0 to 100 vol%

0 to 5 vol% 0 to 25 vol%

Measured values can also be displayed from externally mounted O

2

analyzer. In case of Zirconia O

2

sensor, measurable range is shown above.

Max. 5 components including O

2

sensor can be measured and 2 measuring ranges are available. For measurable components and possible combinations of measuring range, refer to Tables 1 to 5.

Principle of measurement: No, SO

2

, CO

2

, CO, CH

4

Non-dispersion infrared-ray absorption method, Single light source and double beams (double-beam system)

O

2

; Paramagnetic method (Built-in O

2

meter)

Measured value indication:

Digital indication in 4 digits

(LCD with back light)

Instantaneous value of each component

Display of instantaneous value of each component measurement concentration, O

2 corrected instantaneous value, O

2

correvted average value and O

2

average value

(O

2 average value when only O

2

sensor is connected. Instantaneous value converted into O

2

and average value converted into O

2 are added with measurement of NO, SO

2 and CO only. Display is selectable.

Analog output signals:

4 to 20mA DC or 0 to 1V DC, non-isolated output ; 12 points max.

Analog output corresponds to measured value indication in 1:1.

Permissible load resistance;

550

max. for 4 to 20 mA

DC100k

min. for 0 to 1V DC

Analog input signal:

For signal input from externally mounted zirconia O

2

sensor (when ZFK7 is used)

For signal input from externally mounted O

2 analyzer.

0 to 1 VDC from an O

2

analyzer.

Input section is not non-isolated. This feature is effective when an O

2

sensor is not built in.

Input signal triggers measured concentration indication and O

2

correction.

Contact output:

1a contact (250V AC/2A, resistive load)

Instrument error, calibration error, range discrimination, auto calibration status and maintenance status, pump ON/OFF, peak alarm.

1c contact (250V AC/2A, resistive load)

Upper/lower alarm contact output.

(for each channel)

Power disconnection alarm.

* All relay contacts are isolated mutually and from the internal circuit.

Contact input: No-voltage contact (ON/0V, OFF/5V DC,

5mA flowing at ON).

Remote range changeover, auto calibration remote start, remote holding, average value resetting, pump ON/OFF.

Isolated from the internal circuit with photocoupler. Contact inputs are not isolated from one another.

Transmission output:

Solenoid valve drive signal for automatic calibration; open collector output (50 mA or less), used in combination with sampling module.

Rated operating conditions:

Power supply; 85 to 264 V AC, 50/60Hz (3pin inlet terminal used)

Power consumption; 60VA

Ambient temperature; -5˚C to 45˚C

Ambient humidity; 90% RH max.

Storage conditions:

Ambient temperature; -5˚C to 45˚C

Ambient humidity; 100% RH max., non-condensing

Dimensions (H

×

W

×

D):

19

rack nounted type: 177

×

483

×

690mm

(Analyzer main unit)

164

×

318

×

55mm

(Input/output terminal module;)

Mass: Approx. 22 kg

Finish color: Front panel; Off-white (Munsell 10Y7.5/0.5 or equivalent)

Casing; Steel-blue (gray)

Enclosure: Steel casing, for indoor use

Material of gas-contacting parts:

Gas inlet/outlet; SUS304

Infrared-ray sample cell;

SUS304/neoprene rubber

Infrared-ray transmitting window;

CaF

2

O

2 sensor sample cell : SUS316

Internal piping; Toaron tube,

Teflon tube

Gas inlet/outlet: Rc1/4 or NPT1/4 internal thread

Purge gas flow rate: 1L/min ( when required)

INZ-TN2ZKJ-E

9 - 1

2. Standard Functions

Output signal holding:

Output signals are held during manual and auto calibrations by activation of holding (turning "ON" its setting).

The values to be held are the ones just before start calibration mode.

Indication values will not be held.

Remote output holding:

Output signal is held at the latest value by short-circuiting the remote output holding input terminals.

Holding is maintained while the terminals are short-circuited. Indication values will not be held.

Remote range changeover:

Measuring range can be changed according to an external signal when remote range changeover input is received.

Changeover is effective only when remote range setting is turned on. In this case, measuring range cannot be changed manually.

When the contact input terminals for each component are short-circuited, the first range is selected, and it is changed over to the second range when the terminals are open.

Range identification signal:

The present measuring range is identified by a contact signal.

The contact output terminals for each component are shortcircuited when the first range is selected, and when the second range is selected, the terminals are open.

Auto calibration:

Auto calibration is carried out periodically at the preset cycle.

When a standard gas cylinder for calibration and a solenoid valve for opening/closing the gas flow line are prepared externally by the customer, calibration will be carried out with the solenoid valve drive contacts for zero calibration and each span calibration turned on/off sequentially at the set auto calibration timing.

Auto calibration cycle setting:

Auto calibration cycle is set.

Setting is variable within 1 to 99 hours (in increments of

1 hour) or 1 to 40 days (in increments of 1 day).

Gas flow time setting:

The time for flowing each calibration gas in auto calibration is set.

Settable within 60 to 599 seconds (in increments of 1 second)

Auto calibration remote start:

Auto calibration is carried out only once according to an external input signal. Calibration sequence is settable in the same way as the general auto calibration.

Auto calibration is started when a no-voltage rectangular wave is applied to the auto calibration remote start input terminals (opened after short-circuiting for 1.5 seconds or longer).

Auto zero calibration:

Auto zero calibration is carried out periodically at the preset cycle.

This cycle is independent on “Auto calibration” cycle.

When zero calibration gas and solenoid valve for opening/ closing the calibration gas flow line are prepared externally by the customer, zero calibration will be carried out with the solenoid valve drive contact for zero calibration turned on/ off at the set auto zero calibration timing.

Auto zero calibration cycle setting:

Auto zero calibration cycle is set.

Setting is variable within 1 to 99 hours (in increments of 1 hour) or setting is variable within 1 to 40 days (in increments of 1 day).

Gas flow time setting:

The timing for flowing zero gas in auto zero calibration is set. Settable 60 to 599 seconds (in increments of 1 second)

Upper/lower limit alarm:

Alarm contact output turns on when measurement value is reached to the preset upper or lower limit alarm value.

Contacts close when the instantaneous value of each component becomes larger than the upper alarm limit value or smaller than the lower alarm limit value.

Instrument error contact output:

Contacts close at occurrence of analyzer error No. 1, 3 or

10.

Calibration error contact output:

Contacts close at occurrence of manual or auto calibration error (any of errors No. 4 to 9).

Auto calibration status and maintenance status contact

outputs: Contacts close during auto calibration and during input of the remote hold signal.

CO concentration peak count alarm:

(added only for CO/O

2

measurement)

Alarm output turns on according to the preset concentration and count.

Whenever the instantaneous concentration value of CO exceeds the preset concentration value, count increments. If the count exceeds the preset value in one hour, the alarm contacts close.

3. Optical functions

O

2

correction:

Correction of measured NO, SO

2

and CO gas concentrations into values at reference O

2

concentration

Correction formula:

21-O n

C = –––––––– x Cs

21-O s

C: Sample gas concentration after O

2

correction

Cs: Measured concentration of sample gas

Os: Measured O

2

concentration

On: Reference O

2

concentration

(value changeable by setting)

* The upper limit value of the fractional part in this calculation is 4.

The result of calculation is indicated and output in an analog output signal.

Average value after O

2

correction and O

2

average value calculation:

The result of O

2

correction or instantaneous O

2

value can be output as an average value in the determined period of time.

Used for averaging is the moving average method in which sampling is carried out at intervals of 30 seconds.

(Output is updated every 30 seconds. It is the average value in the determined period of time just before the latest updating.)

Averaging time is settable within 1 to 59 minutes (in increments of 1 minute) or 1 to 4 hours (in increments of 1 hour).

9 - 2

INZ-TN2ZKJ-E

Average value resetting:

The above-mentioned output of average value is started from the initial state by applying a no-voltage rectangular wave to the average value resetting input terminals

(opened after short-circuiting for 1.5 seconds or longer).

Output is reset by short-circuiting and restarted by opening.

Pump ON/OFF contact output:

During measurement, this contact closes. While calibration gas is flowing, this contact opens. This contact is connected in power supply of pump, and stop the sample gas while calibration gas flowing.

Communication function:

RS-232C (9pins D-sub)

Half-duplex bit serial

Start-stop synchronization

Modbus TM protocol

Contents:

Read/Write parameters

Read measurement concentration and instrument status

Remark:

When connecting via RS-485 interface, an RS-232 RS-485 converter should be used.

6. Installation Requirements

• Select a place where the equipment does not receive direct sunshine, draft/rain or radiation from hot substances.

If such a place cannot be found, a roof or cover should be prepared for protection.

• Avoid a place where receives heavy vibration.

• Select a place where atmospheric air is clean.

4.

Performance

Repeatability:

±

0.5% of full scale

Linearity:

±

1% of full scale

Zero drift:

Span drift:

±

2% of full scale/week

±

2% of full scale/week

Response time (for 90% FS response) :

Within 60 seconds including replacement time of sample gas (when gas flow rate is

0.5L / min).

5.

Standard Requirements for Sample Gas

Flow rate: 0.5L / min

±

0.2L / min

Temperature: 0 to 50˚C

Pressure: 10 kPa or less (Gas outlet side should be open to the atmospheric air.)

100

µ g/Nm 3 in particle size of 1

µ m or Dust: less

Mist:

Moisture:

Unallowable

Below a level where saturation occurs at

2˚C (condensation unallowable).

Corrosive component:

HCl 1 ppm or less

Standard gas for calibration:

Zero gas; Dry N

2

Span gas; Each sample gas having concentration 90 to 100% of its measuring range (recom mended).

In case a zirconia O

2

analyzer is installed externally and calibration is carried out on the same calibration gas line:

Zero gas; Dry air or atmospheric air

(provided without CO

2

sensor)

Span gas; For other than O

2

measure ment, each sample gas having concentration 90 to 100% of its measuring range.

For O

2

measurement, O

2

gas of 1 to 2 vol%.

INZ-TN2ZKJ-E

9 - 3

9.2 Code symbols

Digit

4

Description

<Custom specifications>

Standard

5

6

7

8

9

<Measurable component (SO

2

,CO

2

,CO,CH

4

)>

NO

SO

2

CO

2

CO

CH

2

NO+SO

2

NO+CO

CO

2

+CO

NO+SO

2

+CO

NO+SO

2

+CO

2

+CO

Others

<Measurable component (O

2

)>

None

External zirconia type sensor

External O

2

analyzer

Built-in paramagnetic type O

2

sensor

<Power supply and gas inlet/outlet>

85 to 264V AC Rc1/4

85 to 264V AC Rc1/4, with purging

85 to 264V AC NPT1/4

85 to 264V AC NPT1/4, with purging

<Revision code>

<Structure>

19-inch rack mounting type

19-inch rack mounting type with side rail

10

11

12

<Indication>

In Japanese

In English

<Measuring range> 1st component, 1st range

0 to 50ppm

0 to 100ppm

0 to 200ppm

0 to 250ppm

0 to 500ppm

0 to 1000ppm

0 to 2000ppm

0 to 5000ppm

0 to 1%

0 to 2%

0 to 3%

0 to 5%

0 to 10%

0 to 20%

0 to 40%

0 to 50%

0 to 70%

0 to 100%

<Measuring range> 1st component, 2nd range

None

0 to 100ppm

0 to 200ppm

0 to 250ppm

0 to 500ppm

0 to 1000ppm

0 to 2000ppm

0 to 5000ppm

0 to 1%

0 to 2%

0 to 3%

0 to 5%

0 to 10%

0 to 20%

0 to 40%

0 to 50%

0 to 70%

0 to 100%

Basic type: ZKJ M M M M 2 M M M M M M M M M M M M M M M

ZKJ

4 5 6 7 8 9 10 11 12

2

F

E

F

H

G

L

M

Z

P

A

D

B

Y

A

B

C

2

3

0

1

2

B

C

J

E

W

P

X

R

Q

L

M

N

J

K

G

H

C

D

E

F

Y

B

W

P

X

R

Q

L

M

N

G

H

J

K

C

D

E

F

A

B

Digit

13

14

15

16

13 14 15 16

Description

<Measurable component> 2nd component, 1st range

None

0 to 100ppm

0 to 200ppm

0 to 250ppm

0 to 500ppm

0 to 1000ppm

0 to 2000ppm

0 to 5000ppm

0 to 1%

0 to 2%

0 to 3%

0 to 5%

0 to 10%

0 to 20%

0 to 40%

0 to 50%

0 to 70%

0 to 100%

<Measurable component> 2nd component, 2nd range

None

0 to 200ppm

0 to 250ppm

0 to 500ppm

0 to 1000ppm

0 to 2000ppm

0 to 5000ppm

0 to 1%

0 to 2%

0 to 3%

0 to 5%

0 to 10%

0 to 20%

0 to 40%

0 to 50%

0 to 70%

0 to 100%

<Measurable component> 3rd component, 1st range

None

0 to 100ppm

0 to 200ppm

0 to 250ppm

0 to 500ppm

0 to 1000ppm

0 to 2000ppm

0 to 5000ppm

0 to 1%

0 to 2%

0 to 3%

0 to 5%

0 to 10%

0 to 20%

0 to 40%

0 to 50%

0 to 70%

0 to 100%

<Measurable component> 3rd component, 2nd range

None

0 to 200ppm

0 to 250ppm

0 to 500ppm

0 to 1000ppm

0 to 2000ppm

0 to 5000ppm

0 to 1%

0 to 2%

0 to 3%

0 to 5%

0 to 10%

0 to 20%

0 to 40%

0 to 50%

0 to 70%

0 to 100%

E

F

G

H

J

C

D

Y

B

K

Q

L

M

N

W

P

X

R

L

M

N

W

P

X

R

E

F

G

H

Y

C

D

J

K

Q

N

W

P

X

R

K

Q

L

M

H

J

F

G

Y

C

D

E

G

H

J

E

F

C

D

Y

B

K

Q

L

M

N

W

P

X

R

9 - 4

INZ-TN2ZKJ-E

Digit

17

18

19

20

21

22

23

Description

<Measurable range> 4th component, 1st range

None

0 to 100ppm

0 to 200ppm

0 to 250ppm

0 to 500ppm

0 to 1000ppm

0 to 2000ppm

0 to 5000ppm

0 to 1%

0 to 2%

0 to 3%

0 to 5%

0 to 10%

0 to 20%

0 to 40%

0 to 50%

0 to 70%

0 to 100%

<Measurable range> 4th component, 2nd range

None

0 to 200ppm

0 to 250ppm

0 to 500ppm

0 to 1000ppm

0 to 2000ppm

0 to 5000ppm

0 to 1%

0 to 2%

0 to 3%

0 to 5%

0 to 10%

0 to 20%

0 to 40%

0 to 50%

0 to 70%

0 to 100%

<O

2

analyzer, 1st range>

None

0 to 5%

0 to 10%

0 to 25%

<O

2

analyzer, 2nd range>

None

0 to 10%

0 to 25%

<Output>

4 to 20mA DC

0 to 1V DC

4 to 20mA DC + Communication function

0 to 1V DC + Communication function

<O

2

correction and O

2

average value output>

None

With O

2

correction output

With peak count alarm

With O

2

correction and O

2

average value

<Adjustment>

For combustion exhaust gas

Others

17 18 19 20 21 22 23

Y

B

C

D

E

F

G

H

J

K

Q

L

M

N

W

P

X

R

A

B

C

D

L

M

N

W

P

X

R

G

H

E

F

Y

C

D

J

K

Q

M

V

Y

L

Y

M

V

Y

A

B

C

A

Z

INZ-TN2ZKJ-E

9 - 5

9.3 Outline diagram

<Analyzer main unit>

<Upper> <Side>

M4(for mounting slide rail)

9 - 6

Power switch

DISPLAY

465

483

Switch for back light

429

Connector for input/output

terminal module Power inlet

<Rear>

38

Sample gas outlet Rc1/4 or NPT1/4

Sample gas inlet Rc1/4 or NPT1/4

<Front>

INZ-TN2ZKJ-E

<Input/output terminal module>

150

316

6-

φ

4.5

hole

150 8

<Cable for connecting input /output terminal>

TN1 TN2 TN3 TN4 TN5

(280)

CN3

8

1000

±

30

INZ-TN2ZKJ-E

9 - 7

Outline diagram of accessory slide rail (unit: mm)

* The slide rails are attached to this equipment when designated.

Closed

Model: 305A-24/Accuride International Inc.

Cabinet member

1.5t

Intermediate member

Drawer member

1.5t

632

(22.7)

15.9

±

0.3

12.7

±

0.3

The same or less

Open

577.8

±

0.3

577.8

±

0.3

111.1

±

0.3

609.6

±

0.3

577.8

±

0.3

12.7

±

0.3

Reinforcement plate

19.1

±

0.3

12.7

±

0.3

15.9

±

0.5

123.8

±

0.3

288.9

±

0.3

4.5

466.7

±

0.3

606.6

±

0.8

×

5.3

4.5

×

5.3

19 inch rack mounting method:

The instrument weight should be supported at the base (at the sides in case of slide rail method).

For easy maintenance, it is recommended to select the method to allow withdrawing along the slide rail.

Slide rail mounting method Guild rail mounting method

Rack dimensions Rack dimensions

450 or more

465

Mounting dimensions

450 or more

465

Mounting dimensions

Slide rail Guide rail

9 - 8

Guide rail

In selecting the guide rail mounting method, provide a maintenance space (200 mm or more) at top of the rack.

INZ-TN2ZKJ-E

advertisement

Related manuals