BK Precision 4055 User manual

Add to My manuals
78 Pages

advertisement

BK Precision 4055 User manual | Manualzz

Model: 4052, 4053, 4054, 4055

Function/Arbitrary Waveform

Generator

USER MANUAL

2

Safety Summary

The following safety precautions apply to both operating and maintenance personnel and must be observed during all phases of operation, service, and repair of this instrument. Before applying power, follow the installation instructions and become familiar with the operating instructions for this instrument.

If this device is damaged or something is missing, contact the place of purchase immediately.

This manual contains information and warnings that must be followed to ensure safe operation as well as maintain the meter in a safe condition.

GROUND THE INSTRUMENT

To minimize shock hazard, the instrument chassis and cabinet must be connected to an electrical ground.

This instrument is grounded through the ground conductor of the supplied, three-conductor ac power cable. The power cable must be plugged into an approved three-conductor electrical outlet. Do not alter the ground connection. Without the protective ground connection, all accessible conductive parts

(including control knobs) can render an electric shock. The power jack and mating plug of the power cable must meet IEC safety standards.

DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE

Do not operate the instrument in the presence of flammable gases or fumes. Operation of any electrical instrument in such an environment constitutes a definite safety hazard.

KEEP AWAY FROM LIVE CIRCUITS

Instrument covers must not be removed by operating personnel. Component replacement and internal adjustments must be made by qualified maintenance personnel. Disconnect the power cord before removing the instrument covers and replacing components. Under certain conditions, even with the power cable removed, dangerous voltages may exist. To avoid injuries, always disconnect power and discharge circuits before touching them.

DO NOT SERVICE OR ADJUST ALONE

Do not attempt any internal service or adjustment unless another person, capable of rendering first aid and resuscitation, is present.

DO NOT SUBSTITUTE PARTS OR MODIFY THE INSTRUMENT

Do not install substitute parts or perform any unauthorized modifications to this instrument. Return the instrument to B&K Precision for service and repair to ensure that safety features are maintained.

WARNINGS AND CAUTIONS

WARNING and CAUTION statements, such as the following examples, denote a hazard and appear throughout this manual. Follow all instructions contained in these statements.

A WARNING statement calls attention to an operating procedure, practice, or condition, which, if not followed correctly, could result in injury or death to personnel.

A CAUTION statement calls attention to an operating procedure, practice, or condition, which, if not followed correctly, could result in damage to or destruction of part or all of the product.

3

WARNING:

WARNING:

CAUTION:

CAUTION:

Do not alter the ground connection. Without the protective ground connection, all accessible conductive parts (including control knobs) can render an electric shock. The power jack and mating plug of the power cable meet IEC safety standards.

To avoid electrical shock hazard, disconnect power cord before removing covers. Refer servicing to qualified personnel.

Before connecting the line cord to the AC mains, check the rear panel AC line voltage indicator.

Applying a line voltage other than the indicated voltage can destroy the AC line fuses. For continued fire protection, replace fuses only with those of the specified voltage and current ratings.

This product uses components which can be damaged by electro-static discharge (ESD). To avoid damage, be sure to follow proper procedures for handling, storing and transporting parts and subassemblies which contain ESD-sensitive components.

4

Compliance Statements

Disposal of Old Electrical & Electronic Equipment (Applicable in the European

Union and other European countries with separate collection systems)

This product is subject to Directive 2002/96/EC of the

European

Parliament and the Council of the European Union on waste electrical and electronic equipment (WEEE) , and in jurisdictions adopting that Directive, is marked as being put on the market after August 13, 2005, and should not be disposed of as unsorted municipal waste. Please utilize your local WEEE collection facilities in the disposition of this product and otherwise observe all applicable requirements.

5

CE Declaration of Conformity

The power supply meets the requirements of 2006/95/EC Low Voltage Directive and

2004/108/EC Electromagnetic Compatibility Directive with the following standards.

Low Voltage Directive

-

EN61010-1: 2001

-

EN61010-031: 2002+A1: 2008

EMC Directive

-

EN 61326-1:2006

-

EN 61000-3-2: 2006+A2: 2009

-

EN 61000-3-3: 2008

6

Safety Symbols

Refer to the user manual for warning information to avoid hazard or personal injury and prevent damage to instrument.

Chassis (earth ground) symbol.

On (Power). This is the In position of the power switch when instrument is ON.

Off (Power). This is the Out position of the power switch when instrument is OFF.

On (Supply). This is the AC mains connect/disconnect switch at the back of the instrument.

Off (Supply). This is the AC mains connect/disconnect switch at the back of the instrument.

Notations

TEXT - Denotes a front panel button.

TEXT – Denotes a menu option.

7

Table of Contents

1 General Information .............................................................................. 10

1.1

Product Overview ...................................................................................................... 10

1.2

Package Contents ...................................................................................................... 10

1.3

Front Panel Overview ................................................................................................ 11

Front Panel Description .................................................................................................... 11

1.4

Rear Panel Overview ................................................................................................. 12

Rear Panel Description ..................................................................................................... 12

1.5

Display Overview ....................................................................................................... 13

Display Description ........................................................................................................... 13

2 Getting Started ...................................................................................... 13

2.1

Input Power Requirements ....................................................................................... 13

Input Power ...................................................................................................................... 13

2.2

Output Connections .................................................................................................. 14

Impedance Matching ....................................................................................................... 14

2.3

Preliminary Check ...................................................................................................... 15

Check Model and Firmware Version ................................................................................ 17

Output Check .................................................................................................................... 17

3 Front Panel Operation ........................................................................... 18

3.1

Menu Options ............................................................................................................ 18

3.2

Selecting a Channel ................................................................................................... 21

3.3

Configure Waveform Output ..................................................................................... 21

Configure Waveform Shape ............................................................................................. 21

Configure Frequency ........................................................................................................ 23

Configure Amplitude ........................................................................................................ 23

Configure DC Offset .......................................................................................................... 25

Configure Phase ............................................................................................................... 25

Configure Duty Cycle and Symmetry ................................................................................ 26

Configure Pulse Waveform ............................................................................................... 28

Configure Noise Waveform .............................................................................................. 30

Configure Arbitrary Waveform ........................................................................................ 31

Configure DC Output ........................................................................................................ 34

3.4

Configure Modulation ............................................................................................... 35

AM Modulation ................................................................................................................ 35

8

DSB-AM ............................................................................................................................ 36

ASK (Amplitude Shift Keying) ........................................................................................... 38

FM Modulation ................................................................................................................ 39

FSK (Frequency Shift Keying) ............................................................................................ 40

PM Modulation ................................................................................................................ 41

PWM Modulation ............................................................................................................. 42

3.5

Configure Sweep Output ........................................................................................... 44

3.6

Configure Burst .......................................................................................................... 47

3.7

Store and Recall ......................................................................................................... 50

Accessing Store/Recall Menu ........................................................................................... 51

Store and Recall Instrument State .................................................................................... 52

Store and Recall Arbitrary Waveforms ............................................................................. 54

3.8

Utility Functions ........................................................................................................ 56

DC Waveform Output ....................................................................................................... 56

Remote Interface .............................................................................................................. 57

Output Configurations ..................................................................................................... 58

Frequency Counter ........................................................................................................... 61

3.9

System Configurations ............................................................................................... 63

System Settings ................................................................................................................ 64

Self Test and Adjustment .................................................................................................. 67

Check Instrument Information ......................................................................................... 68

Updating Firmware .......................................................................................................... 68

3.10

Help System ........................................................................................................... 68

4 Troubleshooting Guide .......................................................................... 70

5 Specifications ........................................................................................ 71

SERVICE INFORMATION ................................................................................ 76

LIMITED ONE-YEAR WARRANTY ................................................................... 77

9

1 General Information

1.1 Product Overview

The B&K Precision 4050 series are dual channel function / arbitrary waveform generators, capable of generating up to 50 MHz sine wave (model 4055). They have an informative easyto-read color display, user-friendly controls and a numeric keypad which allows users to easily configure waveform properties. In addition, they feature non-volatile built-in memory to create, store, and recall arbitrary waveforms up to 16,000 points with 14-bit vertical resolution. 48 predefined arbitrary waveforms are also available for output. A standard USB interface on the rear panel allows users to easily interface with application software to create and load arbitrary waveforms into the instrument.

Features:

• Generate sine wave from 1 µHz up to 50 MHz (4055)

• 3.5-inch TFT-LCD color display

• USB interface

• 16k pts arbitrary waveform memory

• 48 built-in predefined arbitrary waveforms

• Store/recall up to 10 instrument settings

1.2 Package Contents

Please inspect the instrument mechanically and electrically upon receiving it. Unpack all items from the shipping carton, and check for any obvious signs of physical damage that may have occurred during transportation. Report any damage to the shipping agent immediately.

Save the original packing carton for possible future reshipment. Every instrument is shipped with the following contents:

4050 series function/arbitrary waveform generator

Getting started manual (printed)

Full instruction manual on CD

AC power cord

USB type A to Type B cable

Certificate of calibration

Verify that all items above are included in the shipping container. If anything is missing, please contact B&K Precision.

An optional USB to GPIB adapter model AK40G is also available

10

1.3 Front Panel Overview

12 11

10

13

2

3

4

7

8

5

6

9

10

1

2

Front Panel Description

1

Power On/Off switch

11

12

13

3 4 5

6 7

TFT LCD color display

Menu function keys

Waveform buttons

Modulation/Sweep/Burst/Store-Recall/Utility/Help menu buttons

Channel 2 output BNC / Counter input

Channel 1 output BNC

Channel 1 and 2 output On/Off buttons

Arrow keys

Rotary dial knob

Numeric keypad

Channel selection button

USB host port / *USB-to-GPIB adapter interface

Accepts USB flash drive to save/recall instrument settings and waveforms.

*This port can be used for connecting the USB-to-GPIB adapter (AK40G) accessory. It can also be used for connecting an external USB flash drive.

8

9

11

1.4 Rear Panel Overview

7 6

5

1 2 3

Rear Panel Description

1

Modulation Input BNC

2

External Trigger/Gate/FSK/Burst BNC Connector

3

USB interface

4

AC input receptacle

5

Chassis ground

6

Sync Output BNC

7

Trigger Input BNC

12

4

1.5 Display Overview

5

1

2

3

4

Display Description

1

Waveform information display

2

Waveform parameters display

3

Menu options

5

4

Output impedance indicator

Selected channel indicator tab

2 Getting Started

Before connecting and powering up the instrument, please review and go through the instructions in this chapter.

2.1 Input Power Requirements

Input Power

The supply has a universal AC input that accepts line voltage and frequency input within:

100 – 240 V (+/- 10%), 50 – 60 Hz (+/- 5%)

100 – 127 V , 45 – 440 Hz

13

Before connecting to an AC outlet or external power source, be sure that the power switch is in the OFF position and verify that the AC power cord, including the extension line, is compatible with the rated voltage/current and that there is sufficient circuit capacity for the power supply. Once verified, connect the cable firmly.

WARNING:

The included AC power cord is safety certified for this instrument operating in rated range. To change a cable or add an extension cable, be sure that it can meet the required power ratings for this instrument.

Any misuse with wrong or unsafe cables will void the

warranty.

2.2 Output Connections

The waveform generator output circuits operate as a 50 Ω voltage source working into a 50

Ω load. At higher frequencies, a non-terminated or improperly terminated output may cause aberrations on the output waveform. In addition, loads with an impedance less than 50 Ω will reduce the waveform amplitude, while loads with an impedance greater than 50 Ω will increase waveform amplitude.

Excessive distortion or aberrations caused by improper termination are less noticeable at lower frequencies, especially with sine and triangle waveforms. To ensure waveform integrity, follow these precautions:

1. Use good quality 50 Ω coaxial cable and connectors.

2. Make all connections tight and as short as possible.

3. Use good quality attenuators, if it is necessary to reduce waveform amplitudes applied to sensitive circuits.

4. Use termination or impedance-matching devices to avoid reflections.

5. Ensure that attenuators and terminations have adequate power handling capabilities.

If there is a DC voltage across the output load, use a coupling capacitor in series with the load. The time constant of the coupling capacitor and load must be long enough to maintain pulse flatness.

Impedance Matching

If the waveform generator is driving a high impedance, such as a 1 MΩ input impedance

(paralleled by a stated capacitance) of an oscilloscope vertical input, connect the transmission line to a 50 Ω attenuator, a 50 Ω termination and to the oscilloscope input. The attenuator isolates the input capacitance of the device and terminates the waveform generator properly.

14

2.3 Preliminary Check

Complete the following steps to verify that the generator is ready for use.

1.

Verify AC Input Voltage

Verify and check to make sure proper AC voltages are available to power the instrument. The AC voltage range must meet the acceptable specification as

explained in section 2.1.

2.

Connect Power

Connect AC power cord to the AC receptacle in the rear panel and press the power switch to the ON position to turn ON the instrument. The instrument will have a boot screen while loading, after which the main screen will be displayed.

3.

Self Test

Press Utility, select 1/2 from the menu to enter the second menu page, and select

Test/Cal option. Then, select SelfTest option. The instrument has 3 self test options to test the screen, keys, and the LED back lights of the function, menu, and channel output keys.

ScrTest Select this option to do a screen test. The screen will turn into a solid color background.

15

Press 7 to toggle through red, green, and blue colors. This test checks for dead pixels and color consistencies of the TFT LCD display. Press 8 to exit the test.

KeyTest Select this option to do a key test. The screen will display a silhouette of buttons that represent the front panel buttons.

Press any of the front panel buttons and its corresponding button in the silhouette will turn green. This tests the key press response. If any of the buttons do not change colors on the silhouette when pressed, there may be something faulty with the key. Press 8 three times to exit the test.

LEDTest Select this option to test the function, menu, and channel output buttons’ backlight.

Like the key test, the screen will display a silhouette of buttons that represent the front panel buttons. Press 7 to begin the test. This test will start with the Sine button, which should light up. Its corresponding button in the silhouette should also turn green. Continue to press 7 to toggle through all the buttons with backlights available. At the end of the test, all function, menu, and channel output buttons will light up, and all of their corresponding buttons in the silhouette on the display will turn green.

Press 8 to exit the test.

16

4. Self Adjust

This option runs an internal self adjustment procedure that will check and adjust the instrument to specification. Use this option only if the instrument is suspected to be out of calibration.

Press Utility, select 1/2 from the menu to enter the second menu page, and select

Test/Cal option. Then, select SelfAdjust option. The instrument will run the procedure internally, and a progress bar will be displayed in the waveform information area. When finished, it will show 100% on the progress bar and prompt you to press any function key to continue. Follow the prompt to return to the main display.

Check Model and Firmware Version

The model and firmware version can be verified from within the menu system.

Press Utility, select 1/2 from the menu to enter the second menu page, and select EditInfo option. The software/firmware version, hardware version, model, and serial number will be displayed. Press any function key to exit.

Output Check

Follow the steps below to do a quick check of the settings and waveform output.

1. Turn on the instrument and set the instrument to default settings. To set to default, press Utility, press 1/2 to go to page 2, select System from the menu, and select Set

to Default. The instrument will set both channels with the following parameters:

Waveform Shape: Sine

Frequency: 1.000000 kHz

Amplitude: 4.000 Vpp

Offset: 0.0 mVdc

Phase: 0.0 °

Output Impedance: Hi-Z

17

2. Connect the BNC output of CH1 (yellow) into an oscilloscope.

3. Press the Output button on top of CH1 output BNC to turn on the output and observe a sine wave with the parameters above.

4. Press the Freq or Period option in the menu and use the rotary knob or the numeric keypad to change frequency. Observe the changes on the oscilloscope display.

5. Press the Ampl option in the menu and use the rotary knob or the numeric keypad to change the amplitude. Observe the changes on the oscilloscope display.

6. Press the Offset option in the menu and use the rotary knob or the numeric keypad to change the DC offset. With the oscilloscope set for DC coupling, observe the changes on the display.

7. Now, connect the BNC output of CH2 (blue) into an oscilloscope and follow steps 3 to 6 to check its output.

3 Front Panel Operation

3.1 Menu Options

All settings and parameters can be configured from the menu system of the instrument. The menu options that are channel specific are the same for both channel 1 and channel 2. Use

CH1/2

Some many options are grouped in pairs and can be selected by toggling their corresponding menu function keys. For example:

Function key will toggle selection between frequency or period.

Function key will toggle selection between amplitude or HI level.

Function key will toggle selection between DC offset or LO level.

Function key will toggle selection between phase or equal phase.

Function key will go to page 2/2 menu.

The menu system is organized as follows:

Sine

Freq/Period

Configures the frequency or period of the waveform.

Ampl/HLevel

Configures the amplitude or the high level of the waveform.

Offset/LLevel

Configures the DC offset or the low level of the waveform.

Phase/EqPhase Configures the phase or set equal phase relative to the other

18

Sweep

Arb

Pulse

Noise

Square

Ramp

Mod

channel.

Freq/Period

Configures the frequency or period of the waveform.

Ampl/HLevel

Configures the amplitude or the high level of the waveform.

Offset/LLevel

Configures the DC offset or the low level of the waveform.

Phase/EqPhase Configures the phase or set equal phase relative to the other channel.

Duty

Configures the duty cycle of the waveform.

Freq/Period

Configures the frequency or period of the waveform.

Ampl/HLevel

Configures the amplitude or the high level of the waveform.

Offset/LLevel

Configures the DC offset or the low level of the waveform.

Phase/EqPhase Configures the phase or set equal phase relative to the other

Symmetry

channel.

Configures the symmetry of the waveform.

Freq/Period

Configures the frequency or period of the waveform.

Ampl/HLevel

Configures the amplitude or the high level of the waveform.

Offset/LLevel

Configures the DC offset or the low level of the waveform.

PulWidth/Duty Configures the pulse width or the duty cycle of the pulse.

Delay

Configures the delay of the pulse waveform.

Stdev

Mean

Configures the standard deviation of the noise waveform.

Configures the mean value of the noise waveform.

Freq/Period

Configures the frequency or period of the waveform.

Ampl/HLevel

Configures the amplitude or the high level of the waveform.

Offset/LLevel

Configures the DC offset or the low level of the waveform.

Phase/EqPhase Configures the phase or set equal phase relative to the other channel.

Load Wforms

Built-In – Access selectable built-in waveforms for output.

Stored Wforms – Access arbitrary waveforms stored in internal memory for output.

Configures parameters for AM, FM, PM, ASK, FSK, DSB-

AM,PWM modulation.

AM Freq/AM

Depth/DSB

Freq/Key

Freq/FM

Freq/FM

Dev/PM

Freq/Phase

Dev/PWM

Freq/Width

Dev

Type

Shape

Source

Configures the type of modulation.

Configures the modulating waveform shape.

Selects modulating source.

SwpTime

Configures the sweep time for sweep output.

19

StopFreq/FrqSp an

StartFreq/MidF req

Configures the sweep stop frequency or the frequency span of the sweep output.

Configures the sweep start frequency or the center frequency.

Source

Selects the sweep source.

Trig Out/Edge

Internal source: Selects to output a trigger signal at the start

Linear/Log

Direction

Trig

of a sweep operation.

External source: Selects to use rising or falling edge from an external trigger for sweep operation.

Selects between linear or logarithmic sweep operation.

Selects the sweep direction

Manually triggers the sweep operation if the source is set to manual.

Burst

Period

Configures the burst period.

StartPhase

Configures the start phase of the burst output.

Ncycle/Gated

Selects to burst by number of cycles or selects for external gated burst.

Source/Polarity Selects the source for burst output or selects the polarity for external gated burst output.

Selects signal triggering on rising or falling edge.

Trig Out

Cycles/Infinite

Configures number of cycles for burst or infinite burst

(external source only).

Configures the delay of each burst.

Store/Recall

Delay

FileType

Browser

Save

Recall

Delete

Selects the type of files for save/recall.

Selects to browser through folders or files on display.

Saves the selected file into internal or external USB flash memory.

Recalls the selected file from internal or external USB flash memory.

Select to delete the selection.

Utility

DC

Interface

Selects to toggle ON/OFF a DC output. Selectable DC voltages are: 1 V, -1 V, 2 V, -2 V, 5 V, -5 V.

Selects the remote interface. .

Output Setup

Load – Select output impedance.

Normal/Invert – Select to toggle between normal and inverted output.

Counter

System

Test/Cal

EditInfo

Update

CHCopy – Copy channel settings between channels.

Sync – Select to enable/disable synchronization output with respect to either channel 1 or 2.

Selects built-in counter function.

Access to system settings.

Built-in self test and adjustment options.

Shows information about the generator.

For firmware updates.

Help

20

3.2 Selecting a Channel

The 4050 series function/arbitrary waveform generators have dual channel outputs. They can be operated independently or in sync with each other. To select between channel 1 and

CH1/2

When Channel 1 is selected, the display will look like below:

When Channel 2 is selected, the display will look like below:

3.3 Configure Waveform Output

Configure Waveform Shape

The instrument can generate many standard as well as arbitrary waveforms. There are waveform keys on the front panel that will switch between different waveform shape to output. Below is a table of waveforms and the corresponding keys to press to generate those waveforms.

21

Waveforms

Sine

Square

Ramp

Pulse

Noise

Arbitrary

Keys

Sine

Square

Ramp

Pulse

Noise

Arb

After the waveform selection, menu options relevant for that waveform shape will be shown on the right side of the display. Below illustrates the screenshots shown when each of the waveform types is selected.

Sine Waveform Square Waveform

Ramp Waveform Pulse Waveform

Noise Waveform Arbitrary Waveform

22

Configure Frequency

Note: This section does not apply for noise waveform.

Follow the steps below to configure the frequency or period of the output. The adjustable frequency range is different for each model and for each type of waveforms. See the specification section for the adjustable ranges.

1. After selecting the waveform shape, there will be a Freq/Period menu option. Press the corresponding menu function key to toggle between selecting frequency or the period to adjust.

2. The cursor position will now highlight the first digit of the frequency parameter display.

3. Use the rotary knob or the numeric keypad to change the frequency. If numeric keypad is used, the below screen will display:

After entering the numeric value, use the menu function keys to select the unit to apply for the frequency setting. Available units are: MHz, kHz, Hz, mHz, uHz.

Configure Amplitude

Note: This section does not apply for noise waveform.

Follow the steps below to configure the amplitude of the output. The adjustable amplitude range for Channel 1 is 2 mVpp – 10 Vpp into 50 Ω, or 4 mVpp – 20 Vpp into HighZ. For

Channel 2, it is 2 mVpp – 3 Vpp into 50 Ω, or 4 mVpp – 6 Vpp into HighZ.

1. From the menu, select Ampl.

2. The cursor position will now highlight the first digit of the amplitude parameter display.

3. Use the rotary knob or the numeric keypad to change the amplitude. If numeric keypad is used, the below screen will display:

23

After entering the numeric value, use the menu function keys to select the unit to apply for the amplitude setting. Available units are: Vpp, mVpp, Vrm, mVrm, dbm.

Vrm is for Vrms and mVrm is for mVrms.

Note: Only using the numeric keypad will allow for changing the units from V to

Vrms, mVrms, or dbm.

User Defined High and Low Level

The user has the option to adjust the high and low level of the waveform. To do this, toggle the menu function key to select HLevel for high level or LLevel for low level.

Note: DC offset settings will be automatically adjusted if the high and low levels reflect a

DC offset.

24

Configure DC Offset

Note: This section does not apply for noise waveform.

Follow the steps below to configure the DC offset of the output.

1. From the menu, select Offset.

2. The cursor position will now highlight the first digit of the offset parameter display.

3. Use the rotary knob or the numeric keypad to change the offset. If numeric keypad is used, the below screen will display:

After entering the numeric value, use the menu function keys to select the unit to apply for the offset setting. Available units are: Vdc, mVdc.

Configure Phase

Note: This section does not apply for pulse and noise waveforms.

Follow the steps below to configure the phase of the output.

1. From the menu, select Phase.

2. The cursor position will now highlight the first digit of the phase parameter display.

3. Use the rotary knob or the numeric keypad to change the phase. If numeric keypad is used, the below screen will display:

25

After entering the numeric value, use the menu function keys to select the degree (°) unit to apply for the phase setting.

Equal Phase

The unit can set both channel 1 and channel 2 to be in phase with a push of one button.

From the menu, simply select EqPhase and both channels will be in-phase with respect to each other.

Configure Duty Cycle and Symmetry

Note: This section does not apply for pulse and noise waveforms.

Duty Cycle

Note: For square waveform only.

Follow the steps below to configure the duty cycle of a square waveform.

1. Set the instrument for square wave output by pressing the Square button.

2. From the menu, select Duty.

3. The cursor position will now highlight the first digit of the duty parameter display.

26

4. Use the rotary knob or the numeric keypad to change the duty cycle. If numeric keypad is used, the below screen will display:

After entering the numeric value, use the menu function keys to select the percent

(%) unit to apply for the duty cycle setting.

Note: The adjustable duty cycle range varies depending on the frequency of the waveform. Some frequencies do not allow change of the duty cycle and is fixed at

50%.

Symmetry

Note: For ramp/triangle waveform only.

Follow the steps below to configure the symmetry of a ramp/triangle waveform.

1. Set the instrument for ramp wave output by pressing the Ramp button.

2. From the menu, select Symmetry.

3. The cursor position will now highlight the first digit of the symmetry parameter display.

27

4. Use the rotary knob or the numeric keypad to change the symmetry. If numeric keypad is used, the below screen will display:

After entering the numeric value, use the menu function keys to select the percent

(%) unit to apply for the symmetry setting.

Configure Pulse Waveform

Note: This section applies to pulse waveform only.

Pulse Width and Pulse Duty Cycle

The pulse width and duty cycle parameters are related and both control the length of the high level of the pulse. Users have the option to specify the pulse width in units of seconds or specify duty cycle as a percentage.

Note: The instrument allows for adjusting pulse width at a minimum of 1 ns, however its actual minimum resolution is 8 ns. Therefore, the pulse width change can be observed in increments of 8 ns.

28

Follow the steps below to configure the pulse width or duty cycle.

1. Set the instrument for pulse output by pressing the Pulse button.

2. From the menu, select PulWidth for pulse width adjustment or Duty for duty cycle adjustment.

3. The cursor position will now highlight the first digit of the width or duty parameter display.

4. Use the rotary knob or the numeric keypad to change the width or the duty.

After entering the numeric value, use the menu function keys to select the s, ms, us, or ns for pulse width or select percent (%) unit for duty cycle.

Pulse Delay

Follow the steps below to configure the pulse delay.

1. Set the instrument for pulse output by pressing the Pulse button.

2. From the menu, select Delay.

3. The cursor position will now highlight the first digit of the delay parameter display.

4. Use the rotary knob or the numeric keypad to change the pulse delay.

After entering the numeric value, use the menu function keys to select the s, ms, us, or ns.

29

Configure Noise Waveform

Note: This section applies to noise waveforms only.

Standard Deviation and Mean

There are two parameters that can be adjusted of the noise waveform: Standard deviation, mean.

Follow the steps below to configure these parameters.

1. Set the instrument for noise output by pressing the Noise button.

2. From the menu, select Stdev for the standard deviation adjustment or Mean for the mean adjustment.

3. The cursor position will now highlight the first digit of the standard deviation or mean parameter display.

30

4. Use the rotary knob or the numeric keypad to change the two parameters. Both parameters can be specified in V or mV units.

Configure Arbitrary Waveform

There are two ways to generate arbitrary waveforms. Users can output an arbitrary waveform selected from the built-in selection of pre-defined arbitrary waveforms, or create and output a user defined waveform by specifying point by point arbitrary data. This section will explain how to do both.

Generate Predefined Built-in Waveforms

There are a total of 48 (including sine, square, and ramp/triangle) predefined built-in arbitrary waveforms that can be output from the generator. They are divided into four categories: Common waveforms, math waveforms, special (project) waveforms,

Windowing/trigonometric waveforms.

Note: PPulse and NPulse waveforms are limited to 4.999999 MHz.

Common Waveforms

Command waveforms consist of:

StairUp

StairUD

NPulse

Upward staircase

Down and up staircase

Negative pulse

31

StairDn

PPulse

Trapezia

Downward staircase

Positive pulse

Trapezoidal

UpRamp

Upward ramp

Math Waveforms

DnRamp

Downward ramp

Math waveforms consist of:

ExpFall

LogFall

Sqrt

Exponential fall

Logarithmic fall

X^2

Sinc

Dlorentz

Lorentz

Square root function

X

2

function

Sinc function

D-lorentz function

Lorentz function

Gmonpuls

Gaussian mono pulse

Special (project) Waveforms

ExpRise

LogRise

Root3

X^3

Gaussian

Tripuls

Exponential rise

Logarithmic rise

Root three (√𝑥 ) function

X

3

function

Gaussian function

Haversine

Haversine function

Gauspuls

Gaussian modulated sinusoidal pulse

Triangle pulse

Special (project) waveforms consist of:

Cardiac

Chirp

Cardiac signal

Frequency sweep cosine

Sine with white noise

SNR

Window and Trigonometric Waveforms

Quake

TwoTone

Earthquake

Two tone signal

Window and trigonometric waveforms consist of:

Hamming

Hamming window

Kaiser

Gaussian

Haris

Tan

Sec

Asin

Atan

Kaiser window

Gaussian window

Haris window

Tangent function

Secant function

Inverse sine function

Inverse tangent function

Hanning

Blackman

Triangle

Bartlett

Cot

Csc

Acos

ACot

Hanning window

Blackman window

Triangle window

Bartlett window

Cotangent function

Cosecant function

Inverse cosine function

Inverse cotangent function

32

Follow the steps below to browse and select a predefined arbitrary waveform.

1. Press the Arb button to enter the arbitrary waveform function menu.

2. Select 1/2 from the menu to go to the second page, then select Load Wform.

3. Select Built-In and there will be a table on the waveform display area and four categories to select from the menu. Select Common to browse the common waveforms. Select Math to browse the math waveforms. Select Project to browse the special (project) waveforms. Select Winfun\Triangle to browse the window and trigonometric waveforms.

4. After making the selection, use the up, down, left, or right arrow keys to select the desired predefined waveform.

5. Once selected, press the Select option from the menu.

6. The generator will return to the main Arb menu and the waveform display area will show the waveform shape of the selected pre-defined arbitrary waveform. For example, below shows the display of the pre-defined Cardiac signal.

Generate User defined Waveforms

The generator has built-in non-volatile memory to store up to 10 user defined arbitrary waveforms, each with a maximum of 16k points.

33

To create an arbitrary waveform, users must use the USB interface in the rear panel for PC connectivity to the EasyWave software, which can be downloaded at www.bkprecision.com

.

Note: EasyWave software supports Windows XP/Vista/7 (32 bit and 64 bit) only.

EasyWave software will allow users to easily create a custom arbitrary waveform and load it into the internal memory.

To select and output a saved arbitrary waveform from internal memory, follow the steps below.

1. Press the Arb button to enter the arbitrary waveform function menu.

2. Select 1/2 from the menu to go to the second page, then select Load Wform.

3. Select Stored Wforms and there will be a table on the waveform display area with the file names of all arbitrary waveforms that were created and loaded from

EasyWave software.

4. Use the up, down, left, or right arrow keys to select the arbitrary waveform to output from memory. Then, select the Select option from the menu to load the waveform. The instrument will return to the main arb menu.

Note: Arbitrary waveforms created in EasyWave software can also be saved into a .csv file to an external USB flash drive, which can also be transferred and stored into the

internal non-volatile arbitrary waveform memory. See “3.7 Store and Recall” for

details.

Configure DC Output

The instrument can output a DC waveform output at a few predefined voltage levels. See

section “3.8 Utility Functions” for details.

34

3.4 Configure Modulation

The modulation functions can be accessed by pressing the Mod button from the front panel.

The supported types of modulation are: AM, DSB-AM, ASK, FM, FSK, PM.

Note: Pulse, Noise, and DC waveforms cannot be modulated.

AM Modulation

Follow the steps below to setup AM modulation output.

1. Configure your waveform with the desired frequency and amplitude, then press the

Mod button. Under Type, select AM.

2. There are four parameters that can be adjusted: AM Freq, AM Depth, Shape, and

Source.

3. To adjust AM frequency, select AM Freq option from the menu. AM frequency parameter will be highlighted by the cursor on the first digit. Use the rotary knob or the numeric keypad to enter the modulating frequency.

AM Freq adjustable range: 2 mHz – 20 kHz

Note: Adjustable range is dependent on the carrier frequency.

4. Select AM Depth to adjust the depth of the amplitude modulation. Use the rotary knob or the numeric keypad to enter a depth percentage.

35

5. Select Shape to choose the shape of the modulating waveform. Options are: Sine,

Square, Triangle, UpRamp, DnRamp, Noise, Arb.

UpRamp = Upward ramp waveform

DnRamp = Downward ramp waveform

Arb = This will be the waveform that is currently loaded from within the arb function menu. To change, follow the instructions in the previous section.

6. Select Source and choose between Internal or External.

Internal – The modulating source will be generated internally using the AM

Freq, AM Depth, and Shape as defined in the previous steps.

External – Use external modulating source. Users can connect an external signal into the rear BNC terminal labeled Modulation In as the modulating source. AM Freq, AM Depth, and Shape options will not be available.

WARNING: Do not connect more than ± 6 V into the Modulation In terminal. This will damage the instrument and void its warranty.

DSB-AM

Follow the steps below to setup DSB-AM modulation output.

1. Configure your waveform with the desired frequency and amplitude, then press the

Mod button. Under Type, select DSB-AM.

2. There are three parameters that can be adjusted: DSB Freq, Shape, and Source.

36

3. To adjust DSB frequency, select DSB Freq option from the menu. DSB-AM frequency parameter will be highlighted by the cursor on the first digit. Use the rotary knob or the numeric keypad to enter the DSB modulating frequency.

DSB Freq adjustable range: 2 mHz – 20 kHz

Note: Adjustable range is dependent on the carrier frequency.

4. Select Shape to choose the shape of the modulating waveform. Options are: Sine,

Square, Triangle, UpRamp, DnRamp, Noise, Arb.

UpRamp = Upward ramp waveform

DnRamp = Downward ramp waveform

Arb = This will be the waveform that is currently loaded from within the arb function menu. To change, follow the instructions in the previous section.

5. Select Source and choose between Internal or External.

Internal – The modulating source will be generated internally using the DSB

Freq and Shape as defined in the previous steps.

External – Use external modulating source. Users can connect an external signal into the rear BNC terminal labeled Modulation In as the modulating source. DSB Freq, and Shape options will not be available.

37

WARNING: Do not connect more than ± 6 V into the Modulation In terminal. This will damage the instrument and void its warranty.

ASK (Amplitude Shift Keying)

Follow the steps below to setup ASK modulation output.

1. Configure your waveform with the desired frequency and amplitude, then press the

Mod button. Under Type, select ASK.

2. There are two parameters that can be adjusted: Key Freq and Source.

3. To adjust ASK frequency, select Key Freq option from the menu. Key frequency parameter will be highlighted by the cursor on the first digit. Use the rotary knob or the numeric keypad to enter the modulating frequency. This is the frequency at which the output amplitude shifts between the carrier amplitude and 0.

4. Select Source and choose between Internal or External.

Internal – The modulating source will be generated internally.

External – Use external modulating source. Users can connect an external signal into the rear BNC terminal labeled ExtTrig/Gate/FSK/Burst as the modulating source.

WARNING: Do not connect more than ± 6 V into the rear terminal. This will damage the instrument and void its warranty.

38

FM Modulation

Follow the steps below to setup FM modulation output.

1. Configure your waveform with the desired frequency and amplitude, then press the

Mod button. Under Type, select FM.

2. There are four parameters that can be adjusted: FM Freq, FM Dev, Shape, and

Source.

3. To adjust FM frequency, select FM Freq option from the menu. FM frequency parameter will be highlighted by the cursor on the first digit. Use the rotary knob or the numeric keypad to enter the modulating frequency.

FM Freq adjustable range: 2 mHz – 20 kHz

Note: Adjustable range is dependent on the carrier frequency.

4. Select FM Dev to adjust the maximum FM deviation. Use the rotary knob or the numeric keypad to enter the frequency.

5. Select Shape to choose the shape of the modulating waveform. Options are: Sine,

Square, Triangle, UpRamp, DnRamp, Noise, Arb.

UpRamp = Upward ramp waveform

DnRamp = Downward ramp waveform

39

Arb = This will be the waveform that is currently loaded from within the arb function menu. To change, follow the instructions in the previous section.

6. Select Source and choose between Internal or External.

Internal – The modulating source will be generated internally using the FM

Freq, FM Dev, and Shape as defined in the previous steps.

External – Use external modulating source. Users can connect an external signal into the rear BNC terminal labeled Modulation In as the modulating source. FM Freq and Shape options will not be available.

WARNING: Do not connect more than ± 6 V into the Modulation In terminal. This will damage the instrument and void its warranty.

FSK (Frequency Shift Keying)

Follow the steps below to setup FSK modulation output.

1. Configure your waveform with the desired frequency and amplitude, then press the

Mod button. Under Type, select FSK.

2. There are three parameters that can be adjusted: Key Freq, Hop Freq, and Source.

3. To adjust FSK frequency, select Key Freq option from the menu. Key frequency parameter will be highlighted by the cursor on the first digit. Use the rotary knob or the numeric keypad to enter the modulating frequency. This is the frequency at which the output frequency shifts between the carrier frequency and the hop frequency.

4. Select Hop Freq to adjust the hop frequency.

5. Select Source and choose between Internal or External.

Internal – The modulating source will be generated internally.

External – Use external modulating source. Users can connect an external signal into the rear BNC terminal labeled ExtTrig/Gate/FSK/Burst as the modulating source.

40

WARNING: Do not connect more than ± 6 V into the rear terminal. This will damage the instrument and void its warranty.

PM Modulation

Follow the steps below to setup phase modulation output.

1. Configure your waveform with the desired frequency and amplitude, then press the

Mod button. Under Type, select PM.

2. There are four parameters that can be adjusted: PM Freq, Phase Dev, Shape, and

Source.

3. To adjust PM frequency, select PM Freq option from the menu. PM frequency parameter will be highlighted by the cursor on the first digit. Use the rotary knob or the numeric keypad to enter the modulating frequency.

PM Freq adjustable range: 2 mHz – 20 kHz

Note: Adjustable range is dependent on the carrier frequency.

4. Select Phase Dev to adjust the phase deviation. Use the rotary knob or the numeric keypad to enter the frequency.

41

PM Deviation adjustable range: 0 ° – 360 °

5. Select Shape to choose the shape of the modulating waveform. Options are: Sine,

Square, Triangle, UpRamp, DnRamp, Noise, Arb.

UpRamp = Upward ramp waveform

DnRamp = Downward ramp waveform

Arb = This will be the waveform that is currently loaded from within the arb function menu. To change, follow the instructions in the previous section.

6. Select Source and choose between Internal or External.

Internal – The modulating source will be generated internally using the PM

Freq, Phase Dev, and Shape as defined in the previous steps.

External – Use external modulating source. Users can connect an external signal into the rear BNC terminal labeled ExtTrig/Gate/FSK/Burst as the modulating source. PM Freq and Shape options will not be available.

WARNING: Do not connect more than ± 6 V into the rear terminal. This will damage the instrument and void its warranty.

PWM Modulation

Follow the steps below to setup pulse modulation output.

1. Press the Pulse button to select pulse waveform. Then configure your waveform with the desired frequency and amplitude, then press the Mod button.

2. There are four parameters that can be adjusted: PWM Freq, Width Dev, Shape, and

Source.

42

3. To adjust PWM frequency, select PWM Freq option from the menu. PWM frequency parameter will be highlighted by the cursor on the first digit. Use the rotary knob or the numeric keypad to enter the modulating frequency.

PWM Freq adjustable range: 2 mHz – 20 kHz

Note: Adjustable range is dependent on the carrier frequency.

4. Select Width Dev to adjust the pulse width deviation. Use the rotary knob or the numeric keypad to enter the frequency.

Note: Width deviation adjustable range is dependent on the carrier frequency.

5. Select Shape to choose the shape of the modulating waveform. Options are: Sine,

Square, Triangle, UpRamp, DnRamp, Noise, Arb.

UpRamp = Upward ramp waveform

DnRamp = Downward ramp waveform

Arb = This will be the waveform that is currently loaded from within the arb function menu. To change, follow the instructions in the previous section.

6. Select Source and choose between Internal or External.

43

Internal – The modulating source will be generated internally using the

PWM Freq, Width Dev, and Shape as defined in the previous steps.

External – Use external modulating source. Users can connect an external signal into the rear BNC terminal labeled Modulation In as the modulating source. PWM Freq and Shape options will not be available.

WARNING: Do not connect more than ± 6 V into the rear terminal. This will damage the instrument and void its warranty.

3.5 Configure Sweep Output

Note: Sweep function is available for sine, square, ramp, and arbitrary waveforms. It is not available for pulse, noise, or DC waveforms.

Burst function cannot operate when in sweep mode.

Follow the steps below to setup a sweep output from the generator.

1. Configure your waveform with the desired frequency and amplitude, then press the

Sweep button.

2. There are many adjustable parameters to configure the sweep function: SwpTime,

StopFreq, FrqSpan, StartFreq, MidFreq, Source, Trig Out (internal source), Edge

(external source), Linear/Log, Direction, and Trig (manual source).

3. To set the sweep time/rate, select SwpTime from the menu and use the rotary knob or the numeric keypad to enter the sweep time in seconds.

Sweep time is the number of seconds required to sweep from the start to stop frequency. The generator will calculate the number of points required in the sweep based on this time.

4. To set the stop frequency, select StopFreq from the menu and use the rotary knob or the numeric keypad to enter the stop frequency in Hz.

44

Stop Frequency is the frequency to which the sweep ends.

5. To set the frequency span, select FrqSpan from the menu and use the rotary knob or the numeric keypad to enter the frequency span.

Frequency Span provides added control to the direction of the sweep.

6. To set the start frequency, select StartFreq from the menu and use the rotary knob or the numeric keypad to enter the start frequency in Hz.

Start Frequency is the frequency to which the sweep begins.

7. To set the center frequency, select MidFreq from the menu and use the rotary knob or the numeric keypad to enter the center frequency in Hz.

Center Frequency is the frequency at the center of the sweep.

8. Select Source and choose between Internal, External, or Manual.

Internal – Internal trigger source.

External – External trigger source. Users can connect an external signal into the rear BNC terminal labeled ExtTrig/Gate/FSK/Burst to trigger the sweep.

Manual – Manual trigger source.

9. Press the 1/2 option to go to the second page of the menu.

If Internal source is previously selected, the Trig Out option will be available.

On – A square waveform with 50% duty cycle will output from the ExtTrig

BNC connector at the beginning of the sweep. The frequency corresponds to the sweep time.

Off – Trigger output be disabled

If External source is previously selected, the Edge option will be available, and the

Trig On option will not be available.

The up arrow means it will trigger off the rising edge of the external signal.

The down arrow means it will trigger off the falling edge of the external signal.

If Manual source is previously selected, the Trig option in RED will be available. Each press of this option using the function key on the front panel will trigger a sweep.

Trig Out option is not available for manual source.

45

10. Use the function key to toggle selection between Linear or Logarithmic sweep. You can see which type is selected from the waveform display area.

Linear Sweep Logarithmic Sweep

11. The direction of the sweep can be changed by toggling the function key corresponding to the Direction option.

Linear Down Direction Logarithmic Down Direction

The up arrow means the sweep will be in the normal direction (sweep up).

The down arrow means the sweep will be in the reverse direction

(sweep down).

46

3.6 Configure Burst

Note: Burst function is available for all waveforms, with the limitation on noise waveform in which only gated burst function is supported.

Sweep function cannot operate when in burst mode.

Gated burst operation is for use with an internal or external source only.

Follow the steps below to setup a burst output from the generator.

1. Configure your waveform with the desired frequency and amplitude, then press the

Burst button.

2. There are many adjustable parameters to configure the sweep function: Period,

StartPhase, Ncycle, Gated, Source, Trig Out (internal and manual source), Edge

(external source), Cycles/Infinite, Delay, and Trig (manual source).

3. To set the burst period (rate), select Period from the menu and use the rotary knob or the numeric keypad to enter in seconds.

Burst Period/Rate is the frequency of the burst.

4. To set the start phase, select StartPhase from the menu and use the rotary knob or the numeric keypad to enter in degrees.

Start Phase is the phase of the waveform where the burst will begin.

5. There are two configurations that can be set for the burst function. One is to burst with control over the number of cycles. The other is gated burst. Toggle the menu function key to select Ncycle or Gated from the menu.

Ncycle – Use specified number of cycles of the waveform from the Cycles menu option to burst.

Gated – Use internal or external source to control the burst output. To use with internal source, select Internal for source and Period to adjust the burst period prior to selecting this option.

47

6. Select Source and choose between Internal, External, or Manual.

Internal – Internal trigger source. Burst will be controlled by the specified burst period.

External – External trigger source. Users can connect an external signal into the rear BNC terminal labeled ExtTrig/Gate/FSK/Burst to trigger the burst.

This is for both Ncycle and Gated burst operations.

Manual – Manual trigger source.

7. If Gated option is selected, a Polarity option will be available. Toggle the corresponding menu function key to select between Positive or Negative.

Positive – The positive polarity (High level) will cause the gated burst output.

Negative – The negative polarity (Low level) will cause the gated burst output.

8. Press the 1/2 option to go to the second page of the menu.

If Internal source is selected previously, the Trig On option will be available.

On – A square waveform with 50% duty cycle will output from the ExtTrig

BNC connector at the beginning of the burst. The frequency corresponds to

48

the burst period.

Off – Trigger output be disabled

If External source is selected previously, the Edge option will be available, and the

Trig On option will not be available.

The up arrow means it will trigger off the rising edge of the external signal.

The down arrow means it will trigger off the falling edge of the external signal.

Note: These options are not available for Gated burst operations.

If Manual source is previously selected, the Trig option in RED will be available on the first page of the menu. Each press of this option using the function key on the front panel will trigger a burst. The Trig On option will also be available.

On – The ExtTrig BNC connector in the rear panel will output a pulse (>1

µHz) at the start of each sweep.

Off – The pulse output from the ExtTrig BNC connector is disabled.

49

9. If Ncycle is selected in the previous steps, the Cycles/Infinite option will be available.

Use the corresponding menu function key to toggle between the two options. If

Cycles is selected, user the rotary knob or the numeric keypad to enter the number of cycles to burst.

Cycles – Select to specify the number of cycles of the waveform to burst.

Adjustable range is 1 – 50,000 cycles

Infinite – Select to burst an infinite number of cycles of the waveform. This will create a continuous waveform which will not be stopped until a trigger even happens.

Note: For Infinite cycle, external or manual trigger is required to trigger the burst.

10. If Ncycle is selected in the previous steps, the Delay option will be available. Select it and use the rotary knob or the numeric keypad to adjust the burst delay.

Maximum delay is 240 ns.

Note: If required, the burst period will increase to accommodate the specified number of cycles. If Infinite cycle is selected, external or manual source is required to trigger the burst output. The default is external source.

3.7 Store and Recall

The instrument can easily store or recall instrument settings and arbitrary waveforms from internal memory or an external USB flash drive.

Note: If a USB flash drive is connected to the front port, wait two seconds (time required

to read the flash drive information) after pressing the Store/Recall button for the file

browser screen to load up.

50

Accessing Store/Recall Menu

Folder/Drive

Directory

Select Instrument

State or Arbitrary

Waveforms

Select Folder or

File Directory

File Directory

To access the store and recall menu, press the Store/Recall button from the front panel. The above screen will display.

The left column is the folder or drive directory. Internal storage is labeled as Local (C:) . If an external USB flash drive is connected, USB Device (0:) will show in the directory.

The right column is the file directory, which lists all the internal storage locations for instrument state as well as arbitrary waveforms. If external USB flash drive is selected from the folder/drive directory, the file directory will show the root folder of the USB flash drive.

Press the corresponding menu function key to select between Folder or File under Browser.

This will select either the folder or drive directory (left column) or the file directory (right column). The selected column will have the selection cursor in blue.

51

Browse Folder Directory Browse Files Directory

Press the corresponding menu function key to select between State or Data under FileType.

Select State to show the list of internal instrument states. Select Data to show the list of internal arbitrary waveforms.

Instrument States List Arbitrary Waveforms List

Use the up or down arrow keys to move the selection cursor between the lists.

Store and Recall Instrument State

Follow the instructions in this section to store and instrument state to internal or external memory.

Store to and Recall from Internal Memory

1. Configure the instrument with all the settings that you want to save, then press the

Store/Recall button.

2. From the Store/Recall menu, select Folder under Browser and select Local (C:).

52

3. Press the menu function key corresponding to Browser again to change to File.

4. From the menu, change FileType to State.

5. Use the up and down arrow keys to move the selection cursor and highlight any item

(state memory locations) on the list from STATE1 – STATE 10.

6. Once selected, press the Save option from the menu to store current instrument settings into the selected state memory location.

7. The screen will ask you to enter a file name. Use the rotary knob to change the highlighted selection from the on-screen keyboard to select each character for the file name. Once selected, press Select option from the menu to enter the highlighted character. To delete characters, press Delete from the menu. To cancel, press Cancel from the menu.

8. When finished, select Save from the menu. If saved successfully, a “Store state

success!” message will be shown. Wait a few seconds before the screen returns to the Store/Recall main display.

9. Now, the selected state memory location will have a file name next to the STATE#:.

To recall stored instrument states:

10. Select File under Browser from the menu and use the up and down arrow keys to select the state memory location.

11. From the menu, select Recall and the instrument settings saved at the selected location will load. If loaded successfully, a message will be shown “Read finished,

read state success!”

Store to and Recall from External USB Drive

1. Configure the instrument with all the settings that you want to save, then press the

Store/Recall button.

2. From the Store/Recall menu, select Folder under Browser and select USB Device

(0:).

3. From the menu, change FileType to State.

4. Press the Save option from the menu.

5. The screen will ask you to enter a file name. Use the rotary knob to change the highlighted selection from the on-screen keyboard to select each character for the

53

file name. Once selected, press Select option from the menu to enter the highlighted character. To delete characters, press Delete from the menu. To cancel, press Cancel from the menu.

6. When finished, select Save from the menu. If saved successfully, a “Store state

success!” message will be shown. Wait a few seconds before the screen returns to the Store/Recall main display.

7. The file directory displaying the contents of the USB flash drive will now have a new file with the entered filename with a .SET extension. This is the file type used to store instrument states.

To recall stored instrument states, select the USB Device (0:) from the folder directory, then use the up and down arrow keys to highlight the saved .SET file and select Recall from the menu.

Store and Recall Arbitrary Waveforms

Follow the instructions in this section to store and recall arbitrary waveforms to internal or external memory.

Store Arbitrary Waveforms

Note: To store arbitrary waveforms into internal memory or external USB flash drive requires the use of EasyWave software. The software will have an option to load the waveforms into the generator, which will then be stored into the internal memory. Waveforms created in EasyWave can also be saved into a .csv file onto a

USB flash drive, which can be loaded into the generator’s arbitrary waveform

memory.

Recall Arbitrary Waveforms

Follow the steps below to recall internally saved arbitrary waveforms.

1. From the Store/Recall menu, select Folder under Browser and select Local (C:).

2. From the menu, change FileType to Data. Then, change Browser to File.

3. The list of arbitrary waveform memory locations are listed here as ARB1 – ARB10.

4. The file name used to load from EasyWave will be listed here if it was transferred successfully.

5. Use the up and down arrow keys to move the selection cursor and select the location with the arbitrary waveform you want to output.

6. Press the Recall option, and the instrument will automatically load the instrument into the arb menu with the selected waveform loaded and shown on display.

Follow the steps below to recall arbitrary waveforms stored in an external USB flash drive.

54

1. Connect the USB flash drive to the front panel USB port, and a message will appear on the display after a second “A USB drive has been detected.

2. Press the Store/Recall button and select Folder under Browser and select USB

Device (0:).

3. From the menu, change FileType to Data. Then, change Browser to File.

4. Use the up and down arrow keys to move the selection cursor and select the .csv file from the USB flash drive that contains the arbitrary waveform data saved from

EasyWave software.

5. Select the Recall option from the menu and wait for a few seconds for the instrument to load the data file. Once finished, a message will appear “Your data

has been read. Please choose a memory location.

6. Use the up and down arrow keys again to move the selection cursor and select any of the arbitrary waveform memory locations listed (ARB1 – ARB10).

Note: The instrument can only load arbitrary waveforms that are stored into the internal memory, so waveforms that are saved onto an external USB flash drive must be loaded back the internal memory.

7. Select Save from the menu to transfer the loaded arbitrary waveform data from the external USB flash drive into the selected internal arbitrary waveform memory location.

8. The instrument will then automatically go into the arb menu with the selected waveform loaded and shown on display.

Note: If you browse through the internal arbitrary waveform memory locations, you will see the filename that was loaded from the USB flash drive now indicated next to the selected memory location.

55

3.8 Utility Functions

The utility menu can be accessed by pressing the Utility button from the front panel. This section describes all the settings configurable under this menu.

DC Waveform Output

The instrument can output a DC waveform by turning the DC menu option to On. There are six available pre-defined DC offset output levels that can be selected from the menu: 1 V, -1

V, 2 V, -2 V, 5 V, -5 V. Use the menu function keys to select any of these DC offset voltages.

56

Remote Interface

The instrument has a USB interface in the rear panel for remote communication with

EasyWave software. From the utility menu, press Interface to select the interface. Press

USB Setup and select USBTMC, then select Done.

Note: USBRAW is reserved for factory use only.

A type A to type B USB cable is required for PC connectivity.

To connect with EasyWave software, you must install the USB driver. For Windows® 7 users, this may install automatically. For other users, visit www.bkprecision.com

to download the driver.

Note: Users who have LabVIEW™ or NI-VISA installed will automatically have this driver

in their system. In this case, driver download is not required.

GPIB Interface

The generator can be remotely controlled via GPIB using the optional AK40G USB-to-GPIB

adapter.

Note: Be sure all devices are powered off before connecting the adapter to the USB port of the generator and/or the GPIB port on your computer.

Connecting AK40G

1. Connect the USB end of the AK40G to the front USB host port of the instrument.

Since this port is a shared port, an external USB flash drive cannot be used during

GPIB operation.

2. Connect the GPIB end of the AK40G to the computer via a GPIB cable.

57

3. On the AK40G adapter, a RED LED light indicates power to the adapter. A YELLOW

LED light indicates communication is in process.

Power indicator LED

GPIB

USB

PC Generator

AK40G

Communication indicator LED

Figure 1 - Connecting AK40G GPIB adapter

Note: Once connected, do not unplug the adapter on either end, before powering down the instrument and the computer first.

Configure GPIB address

1. Press Utility to enter the utility menu, and press Interface to select the interface.

2. Press GPIB and use the rotary knob to set the GPIB address of the generator. The address range is 1 to 30.

3. Be sure that GPIB is highlighted in blue to indicate GPIB as the selected interface for remote communication. Then, select Done.

Note: EasyWave software does not support GPIB interface. It must be used with USB interface only.

Note: All supported remote commands are described in the programming manual. To

obtain a copy, please contact B&K Precision.

Output Configurations

Select Output Setups from the utility menu to configure output parameters.

58

Output Impedance

The output impedance of the instrument can be switched between 50 Ω and HighZ (high impedance). Select the menu function key corresponding to Load to toggle between the two settings.

Note: If you are connecting directly into an oscilloscope, note that most scopes have 1

MΩ impedance. In this case, it is best to set the output impedance to HighZ to match the load impedance. However if your load impedance is 50 Ω, keep output

impedance set to 50 Ω to match.

Invert Waveform

The main waveform can be inverted by simply selecting Invert from the utility menu. The corresponding menu function key can be pressed to toggle between Normal and Invert output. Inverting a waveform will not change the offset.

Duplicate Channel Parameters

The utility menu has an option that can allow quick copying between channel parameters.

First, press CHCopy option from the menu, and the below menu will display.

Select CH1 -> CH2 to copy channel 1’s parameter settings into channel 2. Likewise, select

CH2 -> CH1 to copy channel 2’s parameter settings into channel 1. After selection, press

Done, and now both channels are identical with the same parameter settings.

59

To cancel, select Cancel.

Sync Output

Note: Sync output is not available for noise and DC waveforms.

The rear panel has a Sync Out BNC terminal available to provide a synchronization output signal. This output can be enabled and can be setup to provide a sync signal that correlates with channel 1 or channel 2.

To set up, press the Sync option from the utility menu.

Under Channel, use the corresponding menu function key to select CH1 or CH2. This will select the channel the output signal will be synchronized with.

To enable the sync output, select On for the State option.

To disable the sync output, select Off for the State option.

When finished, select Done. To cancel, select Cancel.

Note:

When waveform is Inverted, the corresponding sync signal does not invert.

For non-modulated waveform, the sync output reference is the carrier.

For internal modulating AM, FM, and PM, the sync output reference is the

modulated signal (not the carrier).

For ASK and FSK, the sync output reference is the keying frequency.

When sweep is enabled, the sync output becomes TTL level high at the start of

the sweep, and the sync frequency will be the same as the specified sweet time.

When burst is enabled, the sync output will be TTL level high at the start of the

burst.

For external gated burst, the sync output follows the external gated signal.

60

Frequency Counter

The instrument has a built-in frequency counter. The counter input shares the same BNC terminal as the channel 2 main output terminal, labeled CNT. To access and enable the counter function, select the Counter option from the utility menu.

Note: When the counter function is enabled, channel 2 will automatically be disabled

and the same BNC terminal will become the counter’s input terminal.

There are several measurement parameters that can be displayed when counter is enabled.

Measure Frequency/Period

The instrument can measure in frequency or period by selecting Freq or Period respectively from the counter menu.

Measure Width and Duty

The width of the signal can be measured. This parameter is especially useful for measuring square and pulse signals. Users can select between measuring positive or negative width by selecting PWidth or NWidth respectively from the counter menu. Users can also measure the duty cycle in percent of the signal by selecting Duty from the menu.

Adjust Reference Frequency

61

The default reference frequency used by the counter is 10 MHz. Users can change this reference frequency by selecting RefeFreq and use the rotary knob or the numeric keypad to change the reference frequency value.

Adjust Trigger Level

The trigger level of the counter can be adjusted by selecting TrigLev from the menu. Use the rotary knob or the numeric keypad to change the trigger level. The maximum trigger level value is 1.8 V, and the minimum is – 3.0 V.

Counter Settings

There are some additional counter settings that can be adjusted, which may help improve the counter measurements of some signals.

To access these settings, select Setup from the counter menu, and an expanded set of menu options will be available.

62

Coupling

The counter can be set to AC or DC coupling. Select Mode from the menu and toggle the corresponding menu function key to change between DC or AC for DC and AC coupling respectively.

HFR

A high frequency reject filter can be enabled or disabled. Enabling this filter may help improve the frequency measurements of certain signals. From the menu, select HFR and toggle the corresponding menu function key to turn it On or Off.

Default Settings

All of the counter settings can be configured to default values by selecting Default from the menu. Below is a table of the default settings.

Counter Parameters

Reference Frequency (RefeFreq)

Trigger Level (TrigLev)

Coupling (Mode)

High Frequency Reject Filter (HFR)

Select Done to return to the counter main menu.

Default Settings

10.000000 MHz

0.0 V

AC

Off

3.9 System Configurations

The second page of the utility menu contains system parameters that can be adjusted.

63

System Settings

In the utility menu, select 1/2 from the menu to go to the second page of the menu, then select the System option.

Numerical Format

The number format can be changed on the instrument. Select Number Format from the menu.

Toggle the corresponding menu function key to select Point to choose between using period

(.) or comma (,) as the decimal representation of all numerical values on display.

Numeric separators can also be configured by selecting the Separator option. Choose between On, Off, or Space. Below illustrates the difference between the three on all applicable numerical values on the display.

64

Separator: Space Separator: Off

Separator: On

Select Done when finished.

Language

Default language is English. It can be changed to simplified Chinese by selecting the

Language option from the second page of the utility menu. Press the corresponding menu function key to toggle between the two languages.

Power-On Settings

The power-on instrument settings can be configured to load the default values or values configured during the last power-on state. This function can be used to retain instrument configurations in the case of power interruptions.

Press the PowerOn option from the menu to select between Default and Last.

Default will load default instrument settings during power up.

Last will load instrument settings from the last power up state.

Note: The output states of both channels cannot be recalled at power-on. For safety

reasons, the states are disabled by default.

Set to Factory Default

You can set the instrument settings to factory default by selecting Set to Default. The instrument will change all settings to default, and the screen will return to the main display with Sine wave active.

65

Below is a table of the factory default settings:

Parameters

Function

Frequency

Amplitude

Offset

Phase

Output Impedance

Modulation

Carrier Frequency

Modulating Frequency

Modulation Shape

AM Depth

FM Deviation

Key Frequency

FSK Hop Frequency

Phase Deviation

Sweep

Start/Stop Frequency

Sweep Time

Trig Out

Mode

Direction

Burst

Period

Phase

Count

Trig

Source

Counter

Measurement Mode

Reference Frequency

Trigger Level

Mode

HFR

Select 1/2 to go to the second page of the system menu.

Default Settings

Sine Wave

1.000000 kHz

4 Vpp

0 V

0 °

HighZ

1.000000 kHz

100 Hz

Sine Wave

100 %

500 Hz

100 Hz

1 MHz

180 °

100 Hz/1.9 kHz

1 second

Off

Linear

10.000 ms

0 °

1 Cycle

Off

Internal

Frequency

10.000000 MHz

0.0 V

AC

Off

66

Key Sound

Select Beep and toggle the corresponding menu function key to turn the instrument key press sound On or Off. This will also disable the sound for any errors that may occur during front panel or remote operation.

Screen Saver

The instrument has a built-in screen saver function that automatically dims the display when left idle over a period of time specified by a timer. The timer can be changed by selecting the ScrnSvr option from the menu. Press the corresponding menu function key to change between: 1 min, 5 min, 15 min, 30 min, 1 hour, 2 hours, 5 hours, Off.

Clock Source

By default, the instrument uses an internal clock source to generate output waveforms. An external 10 MHz clock source can also be used, especially for synchronization purposes. To change to use an external clock source, select CLKSource and set it to External by pressing the menu function key. To change it to Internal, press the key again.

When set to External, connect a 10 MHz reference signal to the rear BNC terminal labeled

10MHz In.

Self Test and Adjustment

See section “2.3 Preliminary Check” under Output Check for details.

67

Check Instrument Information

Press Utility and press 1/2 to go to the second page of the utility menu. Select EditInfo to see information about the instrument. You will be able to find the following: Boot up Times,

Software version (firmware), Hardware version, Model, Serial Number.

To exit, press any of the menu function keys or any of the waveform keys.

Updating Firmware

When updates are available, users can update the firmware of the instrument by following the below steps.

1. Go to www.bkprecision.com

and download the firmware update file. The file format will have a .ADS extension.

2. Save the file onto a USB flash drive, then connect the drive into the front USB connector of the instrument.

3. Once the USB drive is detected, press Utility and press 1/2 to go to the second page of the utility menu.

4. Select Update option from the menu and the instrument will go into the

Store/Recall main menu.

5. Select Browser and select Folder. Then use the rotary knob or the up and down arrow keys to select USB Device (0:). Then select Browser again and change it to

File.

6. On the right column, use the rotary knob or the up and down arrow keys to select the .ADS firmware file that was saved in the USB flash drive.

7. Then select Recall from the menu. Wait a few seconds and the firmware will begin updating the device.

8. When finished, a message will prompt you to restart the generator. At this point, power off the instrument and wait a few seconds, then power it back on.

9. The instrument will now have the updated firmware. You can verify this by checking the instrument information (follow instructions from the previous section to check).

3.10 Help System

A built-in help system is available to provide quick references on setting up common output configurations with the generator. To access the system, press the Help button from the front panel. The screen below will be displayed:

68

You can use the menu keys, the rotary knob, or the up and down buttons to select any of the ten items listed in the help system as shown above. Once selected (the cursor highlights the selection in blue), push the corresponding menu function key for the Select option and the screen will display a description of the selected help item.

When finished, press the Cancel option from the menu to go back to the main help menu.

69

4 Troubleshooting Guide

Below are some frequently asked questions and answers. Please check if any apply to your power supply before contacting B&K Precision.

Q: I cannot power up the generator

-

Check that the power cord is securely connected to the AC input and there is live power from your electrical AC outlet.

-

Verify that the AC power coming from the mains is the correct voltage. The generator

can accept a specific range of AC input voltages. Refer to section “2.1”.

Q: I cannot see any output out of the output terminals

-

Make sure that the Output buttons above channel 1 and/or channel 2 is lit. If not, press it once. The backlight of these Output buttons indicates that the respective channel’s output is enabled.

Q: I have connected my signal to an oscilloscope, but the amplitude is double of what I set

-

This is often because the impedance of the generator is not matched with the oscilloscope. When the generator is set to 50 Ω impedance, connecting it directly to an oscilloscope with a 1 MΩ input impedance will cause this to happen. To eliminate this issue, connect a 50 Ω terminator to the input of the oscilloscope, then connect a BNC cable between the generator and the terminator to have matching impedance.

70

5 Specifications

Note: All specifications apply to the unit after a temperature stabilization time of 15 minutes over an ambient temperature range of 23 °C ± 5 °C. Specifications are subject to change without notice.

Environmental Conditions:

This instrument is designed for indoor use and operated with maximum relative humidity of

95%.

Model

Channels

Frequency Characteristics

Sine

4052

2

Square

4053

1 µHz – 5 MHz 1 µHz – 10

MHz

1 µHz – 5 MHz 1 µHz – 10

MHz

4054

1 µHz – 25

MHz

4055

1 µHz – 50

MHz

1 µHz – 25 MHz

Triangle, Ramp

Pulse

Accuracy

1 µHz – 300 kHz

500 µHz – 5 MHz

Gaussian Noise (-3 dB) > 5 MHz

Arbitrary

> 10 MHz

1 µHz – 5 MHz

± 50 ppm (90 days)

± 100 ppm (1 year)

1 µHz Resolution

Arbitrary Characteristics

Built-in Waveforms

Waveform Length

Vertical Resolution

Sampling Rate

Minimum Rise/Fall

Time

Jitter (pk-pk)

Non-volatile memory storage

Output Characteristics

Amplitude Range

> 25 MHz

48 Built-in waveforms (includes DC)

16,000 points / Ch

14 bits

125 MSa/s

7 ns (typical)

8 ns (typical)

10 waveforms

> 50 MHz

Channel 1: 2 mVpp – 10 Vpp into 50 Ω, ≤ 10 MHz

2 mVpp – 5 Vpp into 50 Ω, > 10 MHz

4 mVpp – 20 Vpp into open circuit, ≤ 10 MHz

4 mVpp – 10 Vpp into open circuit, > 10 MHz

Channel 2: 2 mVpp – 3 Vpp into 50 Ω

4 mVpp – 6 Vpp into open circuit

Amplitude Resolution Up to 4 digits

Amplitude Accuracy ± (0.3 dB + 1 mVpp of setting value)

(100 kHz)

Amplitude Flatness ± 0.3 dB

71

(relative to 100 kHz, 5

Vpp)

Cross Talk

Offset Range (DC)

Offset Resolution

Offset Accuracy

Output Impedance

Output Protection

Waveform Characteristics

Harmonic Distortion

< -70 dBc

Channel 1: ± 5 V into 50 Ω

± 10 V into open circuit

Channel 2: ± 1.5 V into 50 Ω

± 3 V into open circuit

Up to 4 digits

± ( |Offset setting value| * 1% + 3 mV)

50 Ω (typical)

Short-circuit protection

DC – 1 MHz, < - 60 dBc

1 MHz – 5 MHz, < -53 dBc

5 MHz – 25 MHz, < - 35 dBc

25 MHz – 50 MHz, < -32 dBc

DC – 20 kHz at 1 Vpp, < 0.2 % Total Harmonic

Distortion

Spurious (nonharmonic)

Phase Noise 10 kHz offset, - 108 dBc/Hz (typical)

Rise/Fall Time (Square) < 12 ns (10 % - 90 %) at full amplitude into 50 Ω

Variable Duty Cycle

(Square)

DC – 1 MHz, < -70 dBc

1 MHz – 10 MHz , < -70 dBc + 6 dB/spectrum phase

20% - 80% to 10 MHz

40% - 60% to 20 MHz

50% > 20 MHz

1% of period + 20 ns (typical,1 kHz, 1 Vp-p)) Asymmetry (50 % duty cycle)

Jitter square)

Ramp Symmetry

Linearity (Triangle, ramp at 1 kHz, 1 Vp-p,

100% symmetry)

Pulse

Pulse Width

0.1 % of period (typical, 1 kHz, 1 Vp-p)

0 % - 100 %

< 0.1 % of peak output (typical)

Rise/Fall Time

Duty Cycle

Overshoot

Jitter (pk-pk)

Burst

Waveform

Type

Start/Stop Phase

Internal Period

Gated Source

16 ns minimum,

8 ns resolution

7 ns (typical) at 1 kHz, 1 Vpp from 10 % – 90%

0.1 % resolution

< 5 %

8 ns

Sine, Square, Ramp, Pulse, Arbitrary (except DC)

Cycle (1 – 50,000 cycles), Infinite, Gated

0 ° – 360 °

1 µs – 500 s

External Trigger

72

Trigger Source

Phase Offset

Range

Resolution

Trigger Characteristics

Trigger Input

Input Level

Slope

Pulse Width

Internal, External, Manual

0 ° – 360 °

0.1 °

TTL compatible

Rising or Falling, selectable

> 100 ns

Input Impedance > 5 kΩ, DC coupling

1 MHz

Source

Modulation

Waveform

Frequency

FSK

Deviation

Carrier

Source

Modulation

Waveform

ASK

Carrier

Source

Modulation

Waveform

DSB-AM

Carrier

Source

Maximum

Frequency

Input Latency

Trigger Output

Voltage Level

Pulse Width

< 300 ns

TTL compatible

> 400 ns

Output Impedance 50 Ω

Maximum 1 MHz

Frequency

Modulation Characteristics

Amplitude Modulation

Carrier

Source

Modulation

Sine, Square, Ramp, Arbitrary (except DC)

Internal, External

Sine, Square, Ramp, Noise, Arbitrary (2 mHz – 20 kHz)

Waveform

Modulation Depth 0 % - 120 %

Frequency Modulation

Carrier Sine, Square, Ramp, Arbitrary (except DC)

Internal, External

Sine, Square, Ramp, Noise, Arbitrary (2 mHz – 20 kHz)

0 – 0.5*bandwidth, 10 µHz resolution

Sine, Square, Ramp, Arbitrary (except DC)

Internal, External

50 % duty cycle square waveform (2 mHz – 50 kHz)

Sine, Square, Ramp, Arbitrary (except DC)

Internal, External

50 % duty cycle square waveform (2 mHz – 50 kHz)

Sine, Square, Ramp, Arbitrary (except DC)

Internal, External

73

Modulation

Waveform

Phase Modulation

Carrier

Source

Modulation

Waveform

Phase Deviation

PWM Modulation

Frequency

Source

Modulation

Waveform

External

Modulation

2 mHz – 1 kHz

Sine, Square, Ramp, Arbitrary (except DC)

Internal, External

Sine, Square, Ramp, Noise, Arbitrary (2 mHz – 20 kHz)

0 – 360 °, 0.1 ° resolution

500 µHz – 20 kHz

Internal, External

Sine, Square, Ramp, Noise, Arbitrary (except DC)

- 6 V to + 6 V (max. width deviation)

Duty Cycle 2 mHz – 20 kHz

Modulating

Frequency

Sweep Characteristics

Waveforms

Sweep Shape

Sweep Time

Sweep Trigger

Inputs and Outputs

Output Impedance

Sync OUT

Sine, Square, Ramp, Pulse, Arbitrary (except DC)

Linear or Logarithmic, up or down

1 ms – 500 s

Internal, External, Manual

Modulation In

50 Ω , High impedance

TTL compatible

> 50 ns width, not adjustable

50 Ω (typical) output impedance

2 MHz max. frequency

± 6 Vpp for 100% modulation

> 5 kΩ input impedance

Max. voltage input: ± 6 V

External Clock 10 MHz ± 100 Hz, TTL compatible for external unit synchronization

Ext Trig/Gate/FSK/Burst TTL compatible

Max. voltage input: ± 6 V

Frequency Counter

Measurement Frequency, Period, Positive/Negative pulse width, Duty cycle

Measurement Range Single channel: 100 mHz – 200 MHz

Pulse width/Duty cycle: 1 Hz – 10 MHz

Frequency Resolution 6 bits

Voltage range (nonmodulated signal)

DC Coupling DC Offset Range: ± 1.5 VDC

100 mHz – 100 MHz, 50 mVrms - ± 2.5 V

100 MHz – 200 MHz, 100 mVrms - ± 2.5 V

74

AC Coupling

Pulse width/Duty cycle

Voltage Range

Input Impedance

Coupling

Trigger Level Range

General

Display Resolution

Remote Control

1 Hz – 100 MHz, 50 mVrms - 5 Vpp

100 MHz – 200 MHz, 100 mVrms – 5 Vpp

50 mVrms – 5 Vpp

1 MΩ

AC, DC

-3 V to +1.8 V

3.5” TFT-LCD Display, 320 x 240

USBTMC (standard), GPIB (optional)

Interface

Storage Memory 10 Instrument settings, 10 Arbitrary waveforms

Dimensions (WxHxD) 213mm x 89 mm x 281 mm (8.4” x 3.5” x 11.1”)

Weight

Power

Power Consumption

Temperature

Humidity

Altitude

2.6 kg (5.7 lbs)

100 – 240 VAC ± 10%, 50 / 60 Hz ± 5%

100 – 120 VAC ± 10%, 45 – 440 Hz

50 W max.

Operation: 0 °C – 40 °C

Storage: -20 °C – 60 °C

< 35 °C, ≤ 90 % RH

35 °C – 40 °C, ≤ 60 % RH

Operation: Below 3,000 m

Storage: Below 15,000 m

75

SERVICE INFORMATION

Warranty Service: Please go to the support and service section on our website at www.bkprecision.com

to obtain a RMA #.

Return the product in the original packaging with proof of purchase to the address below. Clearly state on the RMA the performance problem and return any leads, probes, connectors and accessories that you are using with the device.

Non-Warranty Service: Please go to the support and service section on our website at www.bkprecision.com

to obtain a

RMA #. Return the product in the original packaging to the address below. Clearly state on the RMA the performance problem and return any leads, probes, connectors and accessories that you are using with the device. Customers not on an open account must include payment in the form of a money order or credit card. For the most current repair charges please refer to the service and support section on our website.

Return all merchandise to B&K Precision Corp. with prepaid shipping. The flat-rate repair charge for Non-Warranty Service does not include return shipping. Return shipping to locations in North America is included for Warranty Service. For overnight shipments and non-North American shipping fees please contact B&K Precision Corp.

B&K Precision Corp.

22820 Savi Ranch Parkway

Yorba Linda, CA 92887 www.bkprecision.com

714-921-9095

Include with the returned instrument your complete return shipping address, contact name, phone number and description of problem.

76

LIMITED ONE-YEAR WARRANTY

B&K Precision Corp. warrants to the original purchaser that its products and the component parts thereof, will be free from defects in workmanship and materials for a period of one year from date of purchase.

B&K Precision Corp. will, without charge, repair or replace, at its option, defective product or component parts. Returned product must be accompanied by proof of the purchase date in the form of a sales receipt.

To help us better serve you, please complete the warranty registration for your new instrument via our website www.bkprecision.com

Exclusions: This warranty does not apply in the event of misuse or abuse of the product or as a result of unauthorized alterations or repairs. The warranty is void if the serial number is altered, defaced or removed.

B&K Precision Corp. shall not be liable for any consequential damages, including without limitation damages resulting from loss of use. Some states do not allow limitations of incidental or consequential damages. So the above limitation or exclusion may not apply to you.

This warranty gives you specific rights and you may have other rights, which vary from state-to-state.

B&K Precision Corp.

22820 Savi Ranch Parkway

Yorba Linda, CA 92887 www.bkprecision.com

714-921-9095

77

22820 Savi Ranch Parkway

Yorba Linda, CA 92887 www.bkprecision.com

© 2013 B&K Precision Corp.

v042213