Atal EPD-TRHP-INT Manual

Add to My manuals
41 Pages

advertisement

Atal EPD-TRHP-INT Manual | Manualzz

EPD-TRHP-INT

Gebruikershandleiding/Manual

ATAL B.V.

Ampèrestraat 35-37

NL-1446 TR PURMEREND

Postbus 783

NL-1440 AT PURMEREND

T (+31) 0299 630 610

F (+31) 0299 630 611

E [email protected]

I www.atal.nl

1

2

Table of contents

Introduction ...................................................................................................................................................... 3

General safety rules ..................................................................................................................................... 3

Device description and important notices ................................................................................................... 4

Getting started ................................................................................................................................................. 6

What is needed for operation ...................................................................................................................... 6

Mounting the device .................................................................................................................................... 6

Device settings ............................................................................................................................................. 7

Checking functions ....................................................................................................................................... 8

Device setup ..................................................................................................................................................... 9

Setup using web interface............................................................................................................................ 9

Setup using TSensor software .................................................................................................................... 19

Factory defaults ......................................................................................................................................... 20

Communication protocols .............................................................................................................................. 22

Website ...................................................................................................................................................... 22

SMTP – sending e-mails ............................................................................................................................. 22

SNMP ....................................................................................................................................................... 22

Modbus TCP ............................................................................................................................................... 25

SOAP ....................................................................................................................................................... 26

Syslog ....................................................................................................................................................... 28

SNTP ....................................................................................................................................................... 28

Troubleshooting ............................................................................................................................................. 29

I forgot the device IP address .................................................................................................................... 29

I cannot connect to the device................................................................................................................... 29

I forgot the password for setup ................................................................................................................. 29

Factory defaults ......................................................................................................................................... 29

Display is off ............................................................................................................................................... 30

Device error states ..................................................................................................................................... 31

Technical specifications .................................................................................................................................. 32

Dimensions ................................................................................................................................................. 32

General parameters ................................................................................................................................... 34

Measured values ........................................................................................................................................ 35

Temperature measurement ....................................................................................................................... 36

Relative humidity measurement ................................................................................................................ 36

Atmospheric pressure measurement ........................................................................................................ 37

Computed quantities ................................................................................................................................. 37

Operating terms ......................................................................................................................................... 39

End of operation ........................................................................................................................................ 40

Technical support and service.................................................................................................................... 40

Preventive maintenance ............................................................................................................................ 40

Optional accessories ....................................................................................................................................... 41

Introduction

This chapter provides basic information about device. Before starting please read this manual carefully.

Chapter 1

The Web Sensor devices are designed for measurement of temperature, relative humidity and atmospheric pressure in air without aggressive substances. Measured quantities are according device type. Devices with relative humidity measurement can show one of computed quantities: dew point temperature, absolute humidity, specific humidity, mixing ratio and specific enthalpy. Temperature units are °C or °F. Pressure units are hPa, PSI, inHg, mBar, oz/in 2 , mmHg, inH

Ethernet. Device types:

2

O and kPa. Communication interface of Web Sensors is

Type

EPD-T-INT

EPD-TRH-

INT

EPD-TRHP-

INT

EPD-P-INT

EPD-TRH(P)

EPD-T

EPD-TRHP

Temperature Relative humidity Pressure

 

Computed quantity

Web Sensor EPD-xxx-xxx-P devices are designed for measurement in compressed air up to 25 bars. The removable probe is integral part of the device. It is not possible to interchange probe between devices. Probe has connector with protection IP67.

Models marked EPD-xxx-Z are custom-specified devices. Description of these devices is not included in this manual.

General safety rules

The following summary is used to reduce the risk of injury or damage the device.

To prevent injury, please follow instructions.

The device can be services only by a qualified person. The device contains no serviceable parts inside.

Don’t use the device, if it doesn’t work correctly. If you think, that the device is not working correctly, let check it by qualified service person.

3

It’s forbidden to use the device without the cover. Inside the device can be a dangerous voltage and may be risk of electric shock.

Use only the appropriate power supply adapter according to manufacturer specifications and approved according to relevant standards. Make sure, that the adapter does not have damaged cables or covers.

Connect the device only to network parts approved according to relevant standards.

Connect and disconnect the device properly. Don’t connect or disconnect Ethernet cable or probes, if the device is powered.

The device may be installed only in prescribed areas. Never expose the device to higher or lower temperatures than is allowed. The device has not improved resistance to moisture.

Protect it from dripping or splashing water and do not use at areas with condensation.

Devices are not designed for locations with chemically aggressive environment.

Temperature and humidity sensors must not be exposed to direct contact with water or other liquids. It is not allowed to remove the sensor cover to avoid any mechanical damage of the sensors.

Before you removing probe of the Web Sensor EPD-xxx-xxx-P, make sure that pressure in a pressure chamber and ambient pressure are equal.

Don’t use device in explosive environments.

Don’t stress the device mechanically.

Device description and important notices

This chapter contains information about basic features. Also there are important notices concerning to functional safety.

Measured values can be displayed on LCD display or can be read using an Ethernet connection. The following formats are supported:

 Web pages with user changeable look and XML files

 Modbus TCP protocol

 SNMPv1 protocol

 SOAP protocol

The device can also be used to check measured values and if the limit is exceeded, device sends warning messages. Possible ways to sending warning messages:

 Sending e-mails up to 3 e-mail addresses

 Sending SNMP traps up to 3 configurable IP addresses

 Displaying the alarm status on web page

 Sending messages to Syslog server

The device setup can be made by the TSensor software or web interface. TSensor software can be free downloaded from the manufacturer’s website. Also you will find there latest firmware for your device. Do not upload to your device firmware which is not designed for it. Unsupported firmware can damage your device.

4

Device does not support powering over Ethernet cable (PoE). PoE splitter must be used.

Compatible PoE splitter can be purchased as optional accessories. Splitter must have 12V output with approximately 1W.

Reliability of warning messages delivering (e-mail, trap, syslog), depends on actual availability of necessary network services. The device should not be used for critical applications, where malfunction could cause to injury or loss of human life. For highly reliable systems, redundancy is essential. For more information please see standard IEC 61508.

Never connect the device directly to the Internet. If it is necessary connect the device to the Internet, properly configured firewall must be used. Firewall can be partially replaced by the NAT.

5

6

Getting started

Here you can find information necessary to put newly purchased equipment to operation. This procedure is only informative.

Chapter 2

What is needed for operation

To install the unit you need to the following equipment. Before installation check if it’s available.

 Web Sensor Tx5xx device

 power supply adapter 9 - 30V/200mA

 RJ45 LAN connection with appropriate cable

 free IP address in your network

 for Web Sensor EPD-T is required temperature probe Pt1000/3850ppm, connected by shielded cable with maximum length 10m

Mounting the device

 check if the equipment from previous chapter are available

 install the latest version of TSensor software. This software is used to all device settings.

 TSensor software can be free downloaded from the manufacturer’s website.

Software can be also supplied on CD. Device configuration can be made using web interface. For web configuration is not TSensor software necessary.

 contact your network administrator to obtain following information for the connection to the network:

IP address: _____._____._____._____

Default gateway: _____._____._____._____

DNS serer IP: _____._____._____._____

Subnet mask: _____._____._____._____

 check if there is no IP address conflict when you connect the device into network for the first time. The device has from factory set the IP address to

192.168.1.213

. This address must be changed according to information from the previous point. When you installing several new devices, connect them to the network one after another.

 connect probe Pt1000 to Web Sensor EPD-T

 connect the Ethernet connector

 connect the power adapter 9 - 30V (e.g. 12V/200mA)

 LEDs on LAN connector should blink after connecting the power and LCD start showing measured values.

Web Sensor EPD-T-INT, EPD-TRH-INT, EPD-TRHP-INT, EPD-P-INT, EPD-

TRH(P), EPD-TRHP:

Web Sensor EPD-T:

Device settings

 run configuration software TSensor on your PC

 switch to an Ethernet communication interface

 press button Find device...

 the window shows all available devices on your network

7

 click to Change IP to set new address according to network administrator instructions. If your device is not listed, then click Help! My device wasn’t found! Then follow the instructions. MAC address is on product label. The device is factory set to IP 192.168.1.213

.

8

 gateway may not be entered if you want to use the device only in local network. If you set the same IP address which is already used, the device will not work correctly and there will be collisions on the network. If the device detects a collision of IP address then reboot is performed automatically.

 after changing IP address device is restarted and new IP address is assigned.

Restart of the device takes about 10 seconds.

 connect to device using TSensor software and check the measured values. Set the other parameters (alarm limits, SMTP server, etc.). Settings are saved after click on button Save changes .

Checking functions

The last step is to check measured values on the device website. In the address bar of the web browser, enter the device IP address. If the default IP address was not changed, then insert http://192.168.1.213

.

Displayed web page lists actual measured values. If the web pages are disabled, you can see text Access denied . In the case of measurement error a text Error is shown instead value.

Device setup

This chapter describes basic device configuration. There is a description of settings using web interface.

Chapter 3

Setup using web interface

Device can be setup using web interface or TSensor software. Web interface can be managed by the web browser. Main page will be shown when you insert device address into address bar of your web browser. There you find actual measured values. Page with history graphs is shown when you click to tile with actual values. Access to device setup is possible via tile Settings .

9

General

Device name can be changed using item Device name . Measured values are stored into memory according History storage interval field. After changing of this interval all history values will be cleared. Changes must be confirmed by Apply settings button.

10

Network

Network parameters can be obtain automatically from DHCP server using option Obtain an IP address automatically . Static IP address is configurable via field IP address . It is not necessary setup Default gateway while you use device inside one subnet only. DNS server IP is required to set for proper function of DNS. Option Standard subnet mask sets network mask automatically according A, B or C network class. Subnet mask field must be set manually, when network with nonstandard range is used. Periodic restart interval enables to restart device after selected time since device start.

11

Alarm limits

For each measurement channel is possible to set upper and lower limits, time-delay for alarm activation and hysteresis for alarm clearing.

Example of setting the limit to the upper alarm limit:

12

In Point 1 the temperature exceeded the limit. From this time, the time-delay is counting.

Because at point 2 the temperature dropped below the limit value before the time delay expired, alarm was not set.

In Point 3 the temperature has risen over limit again. During the time-delay the value does not drop below the set limit, and therefore was in Point 4 caused alarm. At this moment were sent e-mails, traps and set alarm flag on website, SNMP and Modbus.

The alarm lasted up to Point 5, when the temperature dropped below the set hysteresis

(temperature limit – hysteresis). At this moment was active alarm cleared and e-mail sent.

When alarm occurs, alarm messages will be sent. In case of power failure or device reset

(e.g. changing the configuration) will new alarm state evaluated and new alarm messages will be sent.

Measuring

Page contains settings of the measuring part of the device. Settings can be saved only if jumper inside device is closed. Fields in this page are shown according device type.

It is possible set visibility of measured values on device LCD. Display can be switched off.

Units of measured and computed quantities are selectable by fields Temperature unit ,

Computed value and Atmospheric pressure unit . Devices without atmospheric pressure measurement have field Pressure value for computed values . Here is required to set pressure for calculation of computed quantities. Pressure compensation for altitude above sea-level is possible via Atmospheric pressure offset . This offset depends on current altitude. This value is added to measured atmospheric pressure, final value must be in range

300hPa to 1350hPa.

13

SOAP protocol

SOAP protocol can be enabled by option SOAP protocol enabled . Destination SOAP server can be set via SOAP server address . For setup of server port can be used option

SOAP server port . Device sends SOAP message according selected Sending interval .

Option Send SOAP message when alarm occurs sends message when an alarm on channel occurs or alarm is cleared. These SOAP messages are sent asynchronously to selected interval.

14

Email

Email sending enabled option allows email features. It is necessary set address of the

SMTP server into SMTP server address field. Domain name for SMTP server can be used. Default port of the SMTP server can be changed using item SMTP server port .

SMTP authentication can be enabled using SMTP authentication option. When authentication is enabled Username and Password must be set.

For successfully email sending it is necessary insert Email sender address . This address is usually same as username of the SMTP authentication. Into fields Recipient 1 to

Recipient 3 it is possible set address of email recipients. Option Short email enable sending emails in short format. This format is usable when you need to forward emails into

SMS messages.

When option Alarm email repeat sending interval is enabled and there is active alarm on channel, then emails with actual values are sent repeatedly. Info email sending interval option enables sending emails at selected time interval. CSV history file can be sent together with the repeat/info emails. This feature can be enabled by Alarm and Info emails attachment option.

It is possible to test email function using button Apply and test . This button save a new settings and send a testing email immediately.

15

Modbus a Syslog protocols

ModbusTCP and Syslog protocol settings are configurable via menu Protocols . Modbus server is enabled by default. Deactivation is possible via Modbus server enabled option.

Modbus port can be changed via Modbus port field. Syslog protocol can be enabled using item Syslog enabled . Syslog messages are sent to IP address of the Syslog server - field

Syslog server IP address .

16

SNMP

For reading values via SNMP it is necessary to know password - SNMP read community .

SNMP Trap can be delivered up to three IP address - IP address of the Trap recipient .

SNMP Traps are sent at alarm or error state on the channel. Trap feature can be enabled by option Trap enabled .

17

Time

Time synchronization with SNTP server can be enabled by Time synchronization enabled option. IP address of the SNTP is necessary to set into SNTP server IP address item. SNTP time is synchronized at UTC format, and due to be necessary set corresponding time offset - GSM offset [min] . Time is synchronized every 24 hours by default. Option NTP synchronization every hour decrease this synchronization interval to one hour.

18

WWW and security

Security features can be enabled by the Security enabled option. When security is enabled it is necessary to set administrator password. This password will be required for device settings. When secured access is required even to actual values reading it is possible to enable User account only for viewing enabled . Port of the www server can be changed from the default value 80 using filed WWW port . Web pages with actual values are refreshed according to Web refresh interval field.

Setup using TSensor software

TSensor software is an alternative to web configuration. Some less important parameters are configurable only by the TSensor software. TSensor software allows user adjustment of measured values.

Parameter MTU size can reduce size of the Ethernet frame. Lowering of this size can solve some communication problems mainly with Cisco network infrastructure.

19

Factory defaults

Factory defaults field set the device into factory configuration. Network parameters (IP address, Subnet mask, Gateway, DNS) are left without changes. Settings relating to measurement are not restored by factory defaults.

20

Network parameters are changed while you close jumper and push button inside device during connecting power supply. Factory parameters settings:

Parameter

SMTP server address

SMTP server port

Alarm email repeat sending interval

Info email repeat sending interval

Alarm and Into emails attachment

E-mail recipients addresses

E-mail sender

SMTP authentication

SMTP user/SMTP password

E-mail sending enabled

IP addresses SNMP traps recipients

Password for SNMP reading

Sending SNMP Trap

Website refresh interval [sec]

Website enabled

Website port

Security

Default value example.com

25 off off off cleared sensor@IP off cleared off

0.0.0.0 public off

10 yes

80 off

Administrator password

User password

Modbus port

Modbus server enabled

History storage interval [sec]

SOAP server port

SOAP server address

SOAP sending interval [sec]

SOAP message when alarm occurs

SOAP protocol enabled

Syslog server IP address

Syslog protocol enabled

SNTP server IP address

GMT offset [min]

NTP synchronization every hour

SNTP synchronization enabled

MTU

Periodic restart interval

Demo mode

Temperature

Relative humidity

High limit

Low limit

Hysteresis

Time delay [sec]

High limit

Low limit

Hysteresis

Time delay [sec]

Computed quantity

Atmospheric pressure

High limit

Low limit

Hysteresis

Time delay [sec]

High limit

Low limit

Hysteresis

Time delay [sec]

Device name

20

1

30

50

0

1

30

1000

0

1

30

80

1400 off off

50

700

12

30

Web Sensor cleared cleared

502 yes

60

80 cleared

60 yes off

0.0.0.0 off

217.31.205.226

0 off off

21

22

Communication protocols

Short introduction to communication protocols of the device. To use some communication protocols is necessary software, which can use the protocol. This software is not included. For detailed description of protocols and application notes please contact your distributor.

Chapter 4

Website

The device supports displaying of measured values, history graphs and configuration using web browser. History graphs are based on HTML5 canvas. Web browser must support this feature for proper function of graphs. Firefox, Opera, Chrome or Internet Explorer 9 can be used. If the device has IP address 192.168.1.213

type into your browser http://192.168.1.213

. Using TSensor software or web interface can be set automatic webpages refresh in interval. The default value is 10sec. Actual measured values can be obtained using XML file values.xml

.

Values from history can be exported in CSV format. History storage interval can be set using TSensor software or web interface. History is erased after every reboot of the device.

Reboot of the device is performed when the power supply is disconnected and also after configuration change.

The device allows you to customize the design of web pages. For detailed information please contact your distributor.

SMTP – sending e-mails

When measured values are over the set limits, the device allows send e-mail to a maximum of 3 addresses. E-mail is send when alarm condition on the channel is cleared or a measuring error occurs. It is possible to set repeat interval for email sending. For correct sending of e-mails it is necessary to set address of SMTP server. Domain address can be used as SMTP server address too. For proper function of DNS is required to set DNS server IP address. SMTP authentication is supported but SSL/STARTTLS not. Standard

SMTP port 25 is used by default. SMTP port can be changed. Contact your network administrator to obtain configuration parameters of your SMTP server.

E-mail sent by the device cannot be answered.

SNMP

Using SNMP protocol you can read actual measured values, alarm status and alarm parameters. Via SNMP protocol is also possible to get last 1000 measured values from history table. Writing via SNMP protocol is not supported. It is supported SNMPv1 protocol version only. SNMP used UDP port 161 . OID keys description can be found in the MIB table, which can be obtained from device website or from your distributor. The

password for reading is factory set to public . The changes can be made using TSensor software or web interface. OID keys:

OID Description Type

.1.3.6.1.4.1.22626.1.2.1

.1.3.6.1.4.1.22626.1.2.1.1

.1.3.6.1.4.1.22626.1.2.1.2

.1.3.6.1.4.1.22626.1.2.1.3

.1.3.6.1.4.1.22626.1.2.1.4

.1.3.6.1.4.1.22626.1.2.1.5

.1.3.6.1.4.1.22626.1.2.1.6

.1.3.6.1.4.1.22626.1.2.1.7

.1.3.6.1.4.1.22626.1.2.1.8

.1.3.6.1.4.1.22626.1.2.1.9

.1.3.6.1.4.1.22626.1.2.1.10

.1.3.6.1.4.1.22626.1.2.1.11

.1.3.6.1.4.1.22626.1.2.1.12

.1.3.6.1.4.1.22626.1.2.1.13

.1.3.6.1.4.1.22626.1.2.1.14

.1.3.6.1.4.1.22626.1.2.1.15

.1.3.6.1.4.1.22626.1.2.1.16

.1.3.6.1.4.1.22626.1.2.1.17

.1.3.6.1.4.1.22626.1.2.1.18

.1.3.6.1.4.1.22626.1.2.1.19

.1.3.6.1.4.1.22626.1.2.1.20

.1.3.6.1.4.1.22626.1.2.2

.1.3.6.1.4.1.22626.1.2.2.1

.1.3.6.1.4.1.22626.1.2.2.2

.1.3.6.1.4.1.22626.1.2.2.3

.1.3.6.1.4.1.22626.1.2.3

.1.3.6.1.4.1.22626.1.2.3.1

.1.3.6.1.4.1.22626.1.2.3.2

.1.3.6.1.4.1.22626.1.2.3.3

.1.3.6.1.4.1.22626.1.2.3.4

.1.3.6.1.4.1.22626.1.2.3.5

Actual measured temperature

Actual measured relative humidity

Actual measured computed quantity

Actual measured atmospheric pressure String

Temperature alarm state ("none",

"high", "low")

Relative humidity alarm state

Computed quantity alarm state

Atmospheric pressure alarm state

Temperature unit

String

String

String

String

String

Relative humidity unit

Computed quantity unit

Atmospheric pressure unit

Min. temperature memory

String

String

String

String

Min. relative humidity memory

Min. computed value memory

String

String

Min. atmospheric pressure memory String

Max. temperature memory String

Max. relative humidity memory

Max. computed value memory

String

String

Max. atmospheric pressure memory String

Device name

Device serial number

Device type

Actual measured temperature

Actual measured relative humidity

String

String

String

String

String

String

Actual measured computed quantity Int*10

Actual measured atmospheric pressure Int*X

Temperature alarm (0–none, 1–high, 2– low)

Relative humidity alarm state

Computed quantity alarm state

Atmospheric pressure alarm state

Int*10

Int*10

Integer

Integer

Integer

Integer

.1.3.6.1.4.1.22626.1.2.3.6

.1.3.6.1.4.1.22626.1.2.3.7

.1.3.6.1.4.1.22626.1.2.3.8

.1.3.6.1.4.1.22626.1.2.4

.1.3.6.1.4.1.22626.1.2.4.1

.1.3.6.1.4.1.22626.1.2.4.2

.1.3.6.1.4.1.22626.1.2.4.3

.1.3.6.1.4.1.22626.1.2.4.4

.1.3.6.1.4.1.22626.1.2.4.5

.1.3.6.1.4.1.22626.1.2.4.6

.1.3.6.1.4.1.22626.1.2.4.7

.1.3.6.1.4.1.22626.1.2.4.8

Temperature low limit

Temperature high limit

Relative humidity low limit

Relative humidity high limit

Computed quantity low limit

Computed quantity high limit

Temperature time-delay

Relative humidity time-delay

Int*10

Int*10

Int*10

Int*10

Int*10

Int*10

Integer

Integer

23

24

Trap

0/0

1/0

6/0

6/1

6/2

6/3

6/4

6/5

6/6

6/7

6/8

6/9

6/10

6/11 – 6/14

6/21 – 6/24

6/31 – 6/34

6/41 – 6/44

.1.3.6.1.4.1.22626.1.2.4.9

.1.3.6.1.4.1.22626.1.2.4.10

.1.3.6.1.4.1.22626.1.2.4.11

.1.3.6.1.4.1.22626.1.2.4.12

.1.3.6.1.4.1.22626.1.2.4.13

.1.3.6.1.4.1.22626.1.2.4.14

.1.3.6.1.4.1.22626.1.2.4.15

.1.3.6.1.4.1.22626.1.2.4.16

Computed quantity time-delay

Temperature hysteresis

Relative humidity hysteresis

Computed quantity hysteresis

Atmospheric pressure low limit

Atmospheric pressure high limit

Atmospheric pressure time-delay

Atmospheric pressure hysteresis

.1.3.6.1.4.1.22626.1.5.5.1.0 SNMP Trap

.1.3.6.1.4.1.22626.1.5.6.1.1.1.nr Temperature history values

.1.3.6.1.4.1.22626.1.5.6.1.1.2.nr Relative humidity history values

.1.3.6.1.4.1.22626.1.5.6.1.1.3.nr Computed quantity history values

String

Int*10

Int*10

Int*10

.1.3.6.1.4.1.22626.1.5.6.1.1.4.nr Atmospheric pressure history values Int*X

Integer

Int*10

Int*10

Int*10

Int*X

Int*X

Integer

Int*X

When alarm occurred a warning messages (trap) can be sent to selected IP addresses.

Addresses can be set using TSensor software or web interface. Traps are sent via UDP protocol on port 162 . The device can send following traps:

Description

Reset of the device

Settings was changed

Testing Trap

NTP synchronization error

E-mail sending error

SMTP server login error

SMTP authentication error

Some error occurred during SMTP communication

TCP connection to server cannot be opened

SOAP message sending error

DNS error

SOAP file not found inside web memory

DNS error or Host error

TCP connection to server cannot be opened

Wrong response code from the SOAP server

Upper alarm on channel

Lower alarm on channel

Clearing alarm on channel

Measuring error

Modbus TCP

Device supports Modbus protocol for communication with SCADA systems. Device use

Modbus TCP protocol. TCP port is set to 502 by default. Port can be changed using

TSensor software or web interface. Only two Modbus clients can be connected to device at one moment. Modbus device address (Unit Identifier) can be arbitrary. Modbus write command is not supported. Specification and description of the Modbus protocol is free to download on: www.modbus.org.

Supported Modbus commands (functions):

Command Code Description

Read Holding Register (s) 0x03 Read 16b register(s)

Modbus device registers. Address could be by 1 higher, depending on type used communication library:

Address [DEC] Address [HEX]

57

4149

4150

4151

20481

20482

20483

20484

53

54

55

56

49

50

51

52

20485

20486

20487

20488

20489

20490

20491

20492

20493

20494

20495

20496

20497

20498

20499

20500

20501

0x0031

0x0032

0x0033

0x0034

0x0035

0x0036

0x0037

0x0038

0x0039

0x1035

0x1036

0x1007

0x5001

0x5002

0x5003

0x5004

0x5005

0x5006

0x5007

0x5008

0x5009

0x500A

0x500B

0x500C

0x500D

0x500E

0x500F

0x5010

0x5011

0x5012

0x5013

0x5014

0x5015

Value Type

Measured temperature

Measured relative humidity

Measured computed quantity

Measured atmospheric pressure

Dew point temperature

Absolute humidity

Specific humidity

Mixing ration

Specific enthalpy

Serial number high

Serial number low

Device type

Temperature low limit

Temperature high limit

Relative humidity low limit

Relative humidity high limit

Computed quantity low limit

Computed quantity high limit

Temperature hysteresis

Temperature time-delay

Relative humidity hysteresis

Relative humidity time-delay

Computed quantity hysteresis

Computed quantity time-delay

Temperature alarm state

Relative humidity alarm state

Int*10

BCD

BCD

HEX

Int*10

Int*10

Int*10

Int*10

Int*10

Int*10

Int*10

Int*X

Int*10

Int*10

Int*10

Int*10

Int*10

Int*10

Int*10

DEC

Int*10

DEC

Int*10

DEC

String

String

Computed quantity alarm state

Atmospheric pressure low limit

Atmospheric pressure high limit

String

Int*X

Int*X

Atmospheric pressure hysteresis Int*X

Atmospheric pressure alarm state String

Atmospheric pressure time-delay DEC

Min. temperature memory Int*10

25

20502

20503

20504

20505

20506

20507

20508

Type:

0x5016

0x5017

0x5018

0x5019

0x501A

0x501B

0x501C

Max. temperature memory

Min. relative humidity memory

Max. relative humidity memory

Min. computed value memory

Int*10

Int*10

Int*10

Int*10

Max. computed value memory Int*10

Min. atmospheric pressure memory Int*X

Max. atmospheric pressure memory Int*X

DEC

BCD

HEX

String register is in range 0 – 4500 (16bit) register is in BCD format (16bit) number in HEX format (16bit) two characters in one 16bit register: no – no alarm hi – value is lower than set limit lo – value is higher than set limit

Int*10 register is in format integer*10 – 16 bits:

(125=12.5°C; error = 9999 or -9999)

Int*X register format depends on value type (error = -9999): hPa – integer*10 (9760 = 976.0hPa)

PSI – integer*1000 (14156 = 14.156PSI) inHg – integer*100 (2882 = 28.82inHg) mBar – integer*10 (9761 = 976.1mBar) oz/in 2 – integer*10 (2265 = 226.5oz/in 2 ) mmHg – integer*10 (7321 = 732.1mmHg) inH

2

O – integer*10 (3919 = 391.9inH

2

O) kPa – integer*100 (9761 = 97.61kPa)

SOAP

The device allows you to send currently measured values via SOAP v1.1

protocol. The device sends values in XML format to the web server. The advantage of this protocol is that communication is initialized by the device side. Due to it is not necessary use port forwarding. If the SOAP message cannot be delivered, warning message via SNMP Trap or

Syslog protocol is sent. SOAP message example:

<?xml version="1.0" encoding="utf-8"?>

<soap:Envelope xmlns:soap="http://schemas.xmlsoap.org/soap/envelope/"

xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance"

xmlns:xsd="http://www.w3.org/2001/XMLSchema">

<soap:Body>

<InsertTx5xxSample xmlns="http://cometsystem.cz/schemas/soapTx5xx_v2.xsd">

<passKey>13960932</passKey>

<device>4145</device>

26

<temp>1.4</temp>

<relHum>91.9</relHum>

<compQuant>0.3</compQuant>

<pressure>-9999</pressure>

<alarms>hi,no,no,no</alarms>

<compType>Dew point</compType>

<tempU>C</tempU>

<pressureU>n/a</pressureU>

<timer>60</timer>

</InsertTx5xxSample>

</soap:Body>

</soap:Envelope>

Element Description

<passKey>

<device>

Contains the device serial number (an eight digit number).

Device type identification number (code):

Device

EPD-TRH

Code[DEC]

4107

EPD-T 4106

EPD-TRHP 4129

EPD-P-INT 4124

EPD-T-INT 4144

<temp>

EPD-TRH-INT 4145

EPD-TRHP-

INT

4146

Contains the value of temperature (a decimal part of number is separated by a dot). Error on channel is signaled by 9999 or -9999 .

<relHum> Contains the value of relative humidity. Error value: 9999 or -9999 .

<compQuant> Contains the value of computed quantity. Error value: 9999 or -9999 .

<pressure>

<alarms>

Contains the value of atmospheric pressure. Error value: -9999 .

State of alarm on temperature, relative humidity, computed quantity and pressure channel. Format: tm,rh,cq,pr where: tm – temperature alarm, rh – relative humidity alarm, cq – computed quantity alarm, pr – atmospheric pressure alarm and alarm values: no – no alarm or value is not supported, hi – high alarm, lo – low alarm.

<compType> Computed quantity type: Absolute humidity , Specific humidity ,

Mixing proportion , Specific enthalpy , Dew point or n/a .

<tempU> Temperature and dew point unit. Values: temperature in °F.

C – temperature in °C, F –

<pressureU> Atmospheric pressure unit. Values: hPa , PSI , inHg , mBar , oz/in^2 , mmHg , inH2O and kPa .

<timer> SOAP sending interval [sec].

Example of the web service for incoming SOAP messages from Web Sensor Tx5xx.

Example uses Apache (2.2.10) web server and PHP (5.2.6). It is required to install and enable SOAP extension for PHP. This example store incoming messages to the hard drive:

<?

function InsertTx5xxSample($passKey,$device,$temp,$relHum,$compQuant,$pressure,

27

28

$alarms,$compType,$tempU,$pressureU,$timer) {

$data = "Time: ".StrFTime("%y/%m/%d %H:%M:%S", Time()).", Temp: ".$temp.

", RH: ".$relHum.", CQ: ".$compQuant. ", Pressure: ".$pressure."\n";

$file_write = FOpen("soap.log", "a");

FWrite($file_write, $data);

FClose($file_write);

}

$server = new SoapServer(null, array('uri' => "http://test-uri/"));

$server->addFunction('InsertTx5xxSample');

$server->handle();

?>

Syslog

The device allows sending text message to selected Syslog server. Events are send using

UDP protocol on port 514 . Syslog protocol implantation is according to RFC5424 and

RFC5426. Events when Syslog messages are send:

Text Event

SNTP

Sensor - fw 1-5-5.x

Settings changed

NTP synchronization error

Testing message

Email login error

Email auth error

Email some error

Email socket error

Reset of the device

Settings was changed

NTP synchronization error

Test Syslog message

E-mail sending error

Email dns error

SOAP file not found

SOAP host error

SOAP sock error

SOAP message sending error

SOAP delivery error

SOAP dns error

Text XXXX is set according channel type: Temperature, Relative humidity, Specific humidity, Mixing proportion, Specific enthalpy, Atmospheric pressure.

High alarm XXXX

Low alarm XXXX

Clearing XXXX alarm

Error XXXX

Upper alarm on channel

Lower alarm on channel

Clearing alarm on channel

Measuring error

The device allows time synchronization with NTP (SNTP) server. SNMP protocol version

3.0 is supported (RFC1305). Time synchronization is made every 24 hours. Time synchronization every hour can be enabled. For time synchronization it is necessary set IP address to the SNTP server. It is also possible set GMT offset for correct time zone. Time is used in graphs and history CSV files. Maximum jitter between two time synchronization is 90sec at 24 hours interval.

Troubleshooting

The chapter describes the common problems with devices Web Sensor Tx5xx and methods how to fix these problems. Please read this chapter before you will call technical support.

Chapter 5

I forgot the device IP address

IP address is factory set to 192.168.1.213

. If you had changed it and forgot new IP address, run the TSensor software and press Find device...

In the window are displayed all available devices.

I cannot connect to the device

In search window is only IP and MAC address displayed

Other details are marked N/A . This problem occurs if IP address of the device is set to another network.

Select the window Find device in TSensor software and press Change IP address . Follow the software instructions. To assign IP address automatically using DHCP server, set the device IP address to 0.0.0.0

.

Device IP address is not displayed in window Find device

In TSensor software menu press Help! My device was not found!

in window Find device .

Follow the software instructions. MAC address of the device can be found on product label.

The device is not found even after manually setting MAC address

This problem occurs especially in cases when the IP address of the device belongs to another network and also Subnet mask or Gateway are incorrect.

In this case is DHCP server in the network necessary. In TSensor software menu press Help!

My device was not found!

in window Find device . As new IP address set 0.0.0.0

. Follow the software instructions. An alternative is to reset device to factory defaults using factory-defaults button.

I forgot the password for setup

Please reset device to factory defaults. Procedure is described at following point.

Factory defaults

This procedure restore device to factory settings including network parameters (IP address,

Subnet mask, etc.). Settings relating to measurement are not restored by factory defaults. For factory-defaults follow these steps:

 disconnect the power supply

 unscrew upper cover of the device case

 close the jumper inside device

 press button inside device and power on device at same time

 keep the button pressed for 10 sec

 close the device

Display is off

Check if power supply is connected. Disconnect power connector and connect connector again

– watch the display at the moment of connecting the power. If all LCD segments light for 1 second and go out again, the display is turned off by software.

30

Device error states

Device still make self-tests. If error occurred, LCD shows error codes:

Error LCD display Description

Error 0 Err0

Error 1 Err1

Error 2 Err2

Error 3 Err3

Error 4 Err4

Internal memory CRC error. In this state device doesn't work.

This is a critical error, contact the distributor.

Measured or computed quantity is over the upper limit. Error code 9999. This state appears in case of:

Measured temperature is higher than approximately

+600°C (i.e. high non-measurable resistance of temperature sensor, probably open circuit)

Relative humidity is higher than 100%RH (i.e. damaged humidity sensor or humidity calculation is not possible due to temperature error)

Unable to calculate computed quantity (temperature or humidity measurement error)

Measured or computed quantity is under lower limit or is error in pressure measuring. Error code -9999. This state appears in case of:

Measured temperature is lower than approximately -

210°C (i.e. resistance is too small, probably short circuit)

Relative humidity is lover than 0%RH (i.e. damaged humidity sensor or humidity calculation is not possible due to temperature error)

Measured atmospheric pressure with offset is over range

300hPa to 1350hPa or the pressure sensor is damaged

Unable to calculate computed quantity (temperature or humidity measurement error)

Internal A/D converter error. Error code -9999. In this state device doesn't work. Contact the distributor.

Internal pressure sensor error. Error code -9999. In this state device not measure pressure. Contact the distributor.

31

Technical specifications

Information about technical specifications of the device.

Dimensions

Web Sensor EPD-T-INT:

Chapter 6

Web Sensor EPD-P-INT:

32

Web Sensor EPD-TRH-INT, EPD-TRHP-INT:

Web Sensor EPD-TRH(P), EPD-TRHP:

33

Web Sensor EPD-T:

General parameters

Supply voltage:

DC voltage from 9V to 30V, coaxial connector, 5x2.1mm diameter, positive pole in the middle

Consumption:

~ 1W depending on the operating mode

Measuring interval:

2sec

Display switching interval:

4sec (when more than two values are displayed)

Not allowed manipulations:

It is not allowed to operate the device under conditions other than specified in technical parameters. Devices are not designed for locations with chemically aggressive environment. Temperature and humidity sensors must not be exposed to direct contact with water or other liquids. It is not allowed to remove the sensor cover to avoid any mechanical damage of the sensors.

Case material:

ABS

Material of EPD-TRHP probe: duralumin with the black eloxal surface finish

Mechanical connection of probe EPD-TRHP:

G1/2 with O-ring

Weight:

EPD-P-INT ~130g

EPD-T-INT ~145g

EPD-T ~145g

34

EPD-TRH-INT ~155g

EPD-TRHP-INT ~155g

EPD-TRH/1m probe ~210g, EPD-TRH/2m probe ~250g, EPD-

TRH/4m probe ~330g

EPD-TRHP/1m probe ~260g, EPD-TRHP/2m probe ~300g, EPD-

TRHP/4m probe ~380g

EPD-TRHP/1m probe ~210g, EPD-TRHP/2m probe ~250g, EPD-

TRHP/4m probe ~330g

Mounting the device:

With two holes at the bottom of the unit

Communication port:

RJ45 connector, 10Base-T/100Base-TX Ethernet (Auto-Sensing)

Recommended Connector Cable: for industrial use is recommended Cat5e STP cable, in less demanding applications can be replaced by Cat5 cable, maximum cable length

100m

Supported protocols:

TCP/IP, UDP/IP, ARP, ICMP, DHCP, TFTP, DNS

HTTP, SMTP, SNMPv1, ModbusTCP, SNTP, SOAPv1.1, Syslog

Supported SNMP authentication types:

AUTH LOGIN

Supported web browsers:

Internet Explorer 9 and later, Mozilla Firefox 12 and later, Google

Chrome 18 and later, Opera 11 and later

Recommended minimum screen resolution:

1024 x 768

Memory:

1000 values for each channel inside non-backup RAM memory

EMC emission:

EN 55022, Class B

EMC resistance:

EN 61000-4-2, levels 4/8kV, Class A

EN 61000-4-3, intensity of electromagnetic filed 3V/m, Class A

EN 61000-4-4, levels 1/0.5kV, Class A

EN 61000-4-6, intensity of electromagnetic filed 3V/m, Class A

Measured values

Type Temperature Relative humidity Pressure Computed quantity

EPD-T-INT 

35

EPD-TRH-

INT

EPD-TRHP-

INT

EPD-P-INT

EPD-TRH(P)

EPD-T

EPD-TRHP

Temperature measurement

Temperature and humidity measuring range is limited - see graph below.

Type Operating temperature of

Accuracy electronic

EPD-T-

INT

EPD-

TRH-INT

EPD-

TRHP-

INT

EPD-T

-30 to +80°C

(-22 to +176°F)

±0.6°C

(±1.1°F)

±0.2°C

(±0.4°F)

EPD-

TRH(P)

EPD-

TRHP

±0.4°C

(±0.7°F)

Range

-30 to +80°C

(-22 to +176°F)

-200 to +600°C

(-225 to +999°F)

-30 to +105°C

(-22 to +221°F)

Resolution

0.1°C

(0.2°F)

0.1°C

(0.2°F)

0.1°C

(0.2°F)

Temperature response time with stainless steel sensor cover (F5200) in air flow 1m/s:

EPD-TRH-INT, EPD-TRH, EPD-TRHP-INT, EPD-TRHP

t90 < 6min, temperature step 20°C (36°F)

EPD-TRHP t90 < 16min, temperature step 20°C (36°F)

Temperature probe for Web Sensor EPD-T:

Pt1000/3850ppm, connected by shielded cable max. length 10m

Relative humidity measurement

Temperature and humidity measuring range is limited - see graph below.

Type Operating humidity of electronic

Accuracy

EPD-TRH-

INT

EPD-

TRHP-INT

EPD-

TRH(P)

EPD-TRHP

0 to 100%RH without condensation

±2.5%RH in range 5 to

95%RH in 23°C (73.4°C)

Range

0 to 100%RH temperature compensated

Resolution

0.1%RH

Humidity response time with stainless steel sensor cover (F5200) in air flow 1m/s: t90 < 30s (humidity step 65%RH, constant temperature)

36

Relative humidity and temperature restrictions:

Atmospheric pressure measurement

Unit Range Accuracy hPa mBar

PSI mmHg inHg inH

2

O

±oz/in 2 kPa

T = 23°C

600 to 1100

600 to 1100

±1.3

±1.3

8.70 to 15.95 ±0.02

450.0 to 825.1 ±1.0

17.72 to 32.48 ±0.04

240.9 to 441.6 ±0.5

139.2 to 255.3 ±0.3

60.00 to 110.00 ±0.13

0 ≤ T ≤ 40 °C else

±1.5

±1.5

±0.02

±1.1

±0.04

±0.6

±0.3

±0.15

±2.0

±2.0

±0.03

±1.5

±0.06

±0.8

±0.5

±0.20

Atmospheric pressure response time: t90 < 44s (Response time is selectable. More detail you find in calibration manual, which is included with TSensor software installation.)

Computed quantities

Dew point temperature

Range:

-60 to +80°C (-20 to 176°F)

Accuracy:

±1.5°C (±2.7°F) at ambient temperature T < 25°C (77°F) and RH >

30%, for more details see graphs:

37

Specific humidity

This value depends on atmospheric pressure. Pressure for quantities calculation is stored in device memory. Default value is 1013hPa and can be changed by software.

Range:

0 to 550g/kg

Accuracy:

±2.1g/kg at ambient temperature T < 35°C (95°F)

Mixing ratio

This value depends on atmospheric pressure. Pressure for quantities calculation is stored in device memory. Default value is 1013hPa and can be changed by software.

Range:

0 to 995g/kg

Accuracy:

±2.2g/kg at ambient temperature T < 35°C (95°F)

Specific enthalpy

This value depends on atmospheric pressure. Pressure for quantities calculation is stored in device memory. Default value is 1013hPa and can be changed by software.

Range:

0 to 995kJ/kg

Accuracy:

±4kJ/kg at ambient temperature T < 25°C (77°F)

Absolute humidity

Range:

0 to 400g/m 3

38

Accuracy:

±3g/m 3 graph:

at ambient temperature T < 40°C (104°F), for more details see

Operating terms

Temperature and humidity range in case with electronic:

-30°C to +80°C, 0 to 100%RH (no condensation)

It is recommended to switch off in temperature over +70°C.

Temperature and humidity range of external probe, cable including (EPD-TRH(P), EPD-

TRHP):

-30 to +105°C, 0 to 100%RH (no condensation, limited by graph)

Temperature of Pt1000 probe (EPD-T): depends on probe model

Protection:

IP30 case with electronic

IP40 temperature-humidity probe (EPD-TRH, EPD-TRHP, EPD-

TRHP)

Operating pressure range of probe EPD-TRHP: up to 25bar

Air flow velocity around EPD-TRHP probe: up to 25m/s at pressure of 1bar (1m/s at a pressure of 25bar)

Other environmental conditions (external influences) according to HD 60364-5-51: normal

Working position EPD-P-INT, EPD-TRH(P), EPD-T, EPD-TRHP: arbitrary

Working position EPD-T-INT, EPD-TRH-INT, EPD-TRHP-INT:

39

with sensor cover downwards. When mounting in RACK 19" with universal holder ATS-AC20 (optional accessories) then sensor cover can be placed horizontally.

End of operation

Disconnect the device and dispose it according to current legislation for dealing with electronic equipment. Electronic devices must be professionally destroyed in accordance with EU Directive 2002/96/ES of 27th January 2003.

Technical support and service

Technical support and service is provided by distributor. Contact is included in warranty certificate.

Preventive maintenance

Make sure the cables are not damaged periodically. We recommend you periodic calibration for measurement accuracy validation. Recommended calibration interval:

Type

EPD-T-INT

EPD-TRH-INT

EPD-TRHP-INT

EPD-P-INT

EPD-TRH(P)

EPD-T

EPD-TRHP

Calibration interval

2 years

1 year

1 year

1 year

1 year

2 years

1 year

40

Chapter 7

Optional accessories

This chapter contains list of optional accessories, which can be ordered by extra cost.

Manufacturer recommends using only original accessories.

Temperature probe Pt1000 for Web Sensor EPD-T

Wide ranges of Pt1000/3850ppm probes are available as optional accessories. Please contact your local distributor to help you with choice of correct probe according your application.

Power supply Adapter EPD-AC01

Power supply adapter with CEE 7 plug, 100-240V 50-60Hz/12V DC, 0.5A for Web Sensor devices.

Device case holder for RACK 19" ATS-AC20

ATS-AC20 is a universal holder for mounting of Web Sensor devices into RACK 19".

Probes holder for RACK 19" ATS-AC21

Universal holder for easy mounting probes in RACK 19".

Flow chamber SH-PP for Web Sensor EPD-TRHP

Flow chamber for compressed air measurement up to 25 bars. Made from stainless steel DIN

1.4301, inlet and outlet connection - G1/8, thread for humidity probe connection - G1/2.

41

advertisement

Related manuals

advertisement