Chemglass CG-1978-C-35 Chiller CS-35 Specifications

Add to My manuals
2 Pages

advertisement

Chemglass CG-1978-C-35 Chiller CS-35 Specifications | Manualzz

Huber CS 35

Chiller with air-cooled refrigerating unit and circulation pump (stainless steel). Housing , atmospheric open tank and copper soldered evaporator made of stainless steel. With digital level indicator. Condenser in air-cooled design, performanceoptimized by a built-in high-efficiency fan motor. Powerful feed pump with integrated overtemperature protection. The flow rate can be adjusted via the manual bypass valve on the backside of the chiller.

Control unit B400 / RB400:

Capacitive operating interface with OLED display and multi-coloured status notification for instant identification of the current operating status. Choice of eight different system languages (DE, EN, ES, FR, IT, PT, RU, TR). Separate operating option for the feed pump and the cooling unit with convenient adjustment of the desired setpoint. Operating of the system can be evaluated on a PC or notebook via an integrated RS232 interface.

Technical data according to DIN 12876

Operating temperature range

Temperature stability at -10°C temperature set point / display

Internal temperature sensor

Interface digital

Safety classification

Cooling power at +15°C at +10°C at ±0°C at -10°C at -20°C

Refrigeration machine

-20...+15 °C

±1.5 K colour LED Touchscreen

Pt100

RS232

Class I / NFL

4.0 kW

3.5 kW

2.1 kW

2.0 kW

0.8 kW air-cooled, CFC- and HCFCfree

R449A

1,6 kg without

Refrigerant

Refrigerant quantity

Gas warning sensor

Pressure pump max. delivery max. delivery pressure

Pump connection

Bath volume min.

Bath volume max. sound pressure level +/- 4 dB(A)

Overall dimensions WxDxH **

Net weight

Power supply requirement max. current

Fuse on power network side

Degree of Protection max. ambient temperature min. ambient temperature

30 l/min

4.0 bar

19 A

IP20

+40 °C

+5 °C

2x 20 A

G ½” male

10 l

11 l

62 dB(A)

580x660x820 mm

135 kg

208V 2~ 60Hz

Order-No.: VDH1101336 from Serial-No.: 1861 20 1.0/20

Huber USA Inc. http://www.huber-usa.com Tel 919-674-4266 Toll Free +1 800-726-4877 [email protected]

Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.

Accessories and periphery: 2pcs Hose nozzles ∅ 8mm*, bath cover*, overpressure. Bypass valve*, Bypass valve*, drain valve, thermofluid, RS232 cable

* standard accessories

Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.

In accordance with EN60034-1 the following voltage and frequency tolerances are valid:

Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%

Example: 5% voltage and +2% frequency -> not allowed! -

5% voltage and - 2% frequency -> allowed

Information to Electromagnetic compatibility:

Classification (disturbance) to EN55011: Class A, Group 1

Recommended thermofluid: Water - Monoethylene Glycol 50:50

Standard delivery conditions - Power cable configuration:

1.

Single-phase devices (115V) -> with cable and plug

2.

Two-phase devices (208V) -> with cable, without plug

3.

Three-phase devices (208V/460V) with current consumption less than 63A -> with cable, without plug

4.

Three-phase devices (208V/460V) with current consumption greater than 63A -> with cable, without plug

** Please respect space requirements. See operating conditions at www.huber-online.com

Huber USA Inc. http://www.huber-usa.com Tel 919-674-4266 Toll Free +1 800-726-4877 [email protected]