GASSONIC SURVEYOR User Manual
Add to My manuals28 Pages
GASSONIC Surveyor, an ultrasonic gas leak detector, is a high-tech solution for detecting pressurised gas leaks by sensing the ultrasound generated by escaping gas. The detector, designed for harsh environments and extreme weather conditions, is ideal for complex pipeline systems, high-pressure gas facilities, petrochemical processing plants, gas pipeline compressor stations, and gas storage facilities. With a detection radius of 5 to 20 meters, depending on ultrasonic background noise level, the device offers reliable performance.
advertisement
uSeR Guide FoR
GaSSonic SuRVeYoR ulTRaSonic GaS leaK deTecToR
conTenTS
1.0 The Gassonic Surveyor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
1.1 Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
1.2 Working principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
1.3 Main features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
2.0 Background noise mapping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
2.1 Noise levels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
3.0 Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
4.0 Mechanical construction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
4.1 Mechanical operation and safety . . . . . . . . . . . . . . . . . . . . . . . . . .7
4.2 Spare parts and accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
4.3 Mechanical operation and safety . . . . . . . . . . . . . . . . . . . . . . . . . .7
5.0 Alarm delay and trigger level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
5.1 Output methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
5.2 Using the 4 - 20 mA output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
5.3 Alarm relay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
5.4 Error relay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
5.5 NAMUR or dry contact . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
6.0 Test and calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
7.0 Gain test procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
8.0 Delay test procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
9.0 Calibration procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
10.0 Step test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
11.0 Gassonic Surveyor LED interface . . . . . . . . . . . . . . . . . . . . . . . . . . .16
12.0 Trouble shooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
13.0 Power consumption . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
14.0 Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
15.0 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
16.0 Suppliers declaration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
17.0 Warranty Statement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
18.0 Appendix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
Installation and User Guide for Gassonic Surveyor (BB6019-01) Page 2
1.0 -
The GaSSonic SuRVeYoR
The Gassonic Surveyor is an ultrasonic gas leak detector. It is designed to detect pressurized gas leaks by sensing the ultrasound generated by escaping pressurized gas. The detector is designed for use in harsh environments and extreme weather conditions, making it ideal for detecting leaks in complex pipeline systems both onshore and offshore.
Gassonic Surveyor is developed, marketed and serviced by Gassonic A/S. Gassonic A/S is part of the General Monitors family of companies. The Gassonic Surveyor and other selected Gassonic products are manufactured by General
Monitors, Ireland.
1.1 -
applicaTion
The Gassonic Surveyor is ideal for:
• High pressure gas facilities (methane, hydrogen etc.)
• Petrochemical processing plants
• Gas pipeline compressor stations
• Gas storage facilities
1.2 -
WoRKinG pRinciple
When gas leaks from a pressurized system, it emits a characteristic “hissing” sound. A part of this sound is high frequency ultrasound, which can not be heard by humans.
As the detector senses ultrasonic noise, the Gassonic Surveyor does not have to be in contact with the leaking gas itself.
It will detect the gas leak instantly even if the gas is carried away by wind or diluted in the air. The Gassonic Surveyor can be used in configurations with other detection methods.
1.3 -
main FeaTuReS
Detection radius
Sensor technology
Outputs
Visual output
Housing
Certifications
5 to 20 meters, depending on ultrasonic background noise level.
Stainless steel microphone technology
Analogue 4-20 mA interface and Alarm relay or Error relay. (NAMUR and dry contact).
Red and green LED
Acid-proof stainless steel (316L)
ATEX and IEC Ex, Intrinsically Safe (Ex-i) (Available on Gassonic website)
Installation and User Guide for Gassonic Surveyor (BB6019-01) Page 3
2.0 -
BacKGRound noiSe mappinG
The Gassonic Surveyor filters out frequencies below 25 kHz. However, a few sources other than leaking gas may produce ultrasonic noise above 25 kHz, and could therefore trigger the detector. To avoid these sources influencing the gas leak detector, it is recommended to perform a background noise mapping survey of the plant using the Gassonic 1700
Ultrasonic Mapping Meter.
The mapping survey makes it easy to adjust the trigger level and the delay time of each detector so that it is unaffected by background noise in the area covered. Gassonic can be contacted for further details on performing this mapping survey.
2.1 -
noiSe leVelS
It is helpful to know the actual noise level before installation, as this has influence on the Gassonic Surveyor’s detection range. Most environments can be divided in three overall noise levels.
1 2 3
3 m
1m 2m 3m 4m 5m 6m 7m 8m 9m 10m 11m 12m 13m 14m 15m 16m
If the Gassonic Surveyor is mounted above floor grating, the detection range will continue downwards.
Very low-noise areas 1
In ”very low-noise areas”, where background noise is below 58 dB, the trigger level must be set to at least 64 dB. This corresponds to a detection radius of 13-20 m.
Typical areas:
• Onshore wellhead areas in calm environment
• Salt dome gas storage facilities in calm environment low-noise areas
In ”low-noise areas”, where background noise is below than 68 dB, the trigger level must be set to at least 74 dB.
This corresponds to a detection radius of 9-12 m.
2
Typical areas:
• Areas with no machinery
• Areas with low frequency machine made noise
HigH-noise areas
3
In ”high-noise areas”, where background noise is below 78 dB, the trigger level must be set to at least 84 dB. This corresponds to a detection radius of 5-8 m.
Typical areas:
• Turbo compressor areas
• Complete open offshore weather deck
• Next to very noisy machinery
Installation and User Guide for Gassonic Surveyor (BB6019-01) Page 4
3.0 -
mounTinG
To ensure optimal performance, correct mounting of the gassonic surveyor is essential.
For optimal detection coverage the Gassonic Surveyor should be mounted at a height of 1 to 2 meters above potential leak spots, and if possible in a position which makes it accessible for maintenance purposes. In special cases, such as installation over compressors, the height can be increased.
The Gassonic Surveyor can be mounted on a pole or a wall using the optional mounting bracket UA1352A for standard horizontal mounting or on the 45° mounting bracket UA1353. It can also be mounted on a ceiling, on structural beams or similar non-vibrating structures.
As it primarily detects in a downward direction the unit must be mounted with the microphone compartment facing down. This will also prevent dirt from collecting in the wind screen, which may block the sound.
17m 16m 15m 14m 13m 12m 11m 10m 9m 8m 7m 6m 5m 4m 3m 2m 1m
1 2 3
1 Full coVerage 2 obsTrucTion e.g. a concreTe wall 3 reduced coVerage
Mount the detector so it is clear of solid physical structures such as walls and heavy machinery, as these will dampen any ultrasonic noise, leading to reduced detection coverage. Normal piping will not affect the coverage.
1m 2m 3m 4m 5m 6m 7m 8m 9m 10m 11m 12m 13m 14m 15m 16m
45˚
Installation and User Guide for Gassonic Surveyor (BB6019-01)
If necessary, the Gassonic Surveyor can be mounted with an inclination not exceeding
45°. If mounted next to a wall, the inclination will provide improved coverage. The optional mounting bracket UA1353 can be used for mounting at a 45° angle.
Page 5
4.0 -
mechanical conTRucTion
The Gassonic Surveyor is an intrinsically safe (Ex-i) ultrasonic gas leak detector. The cables are connected through M20 x 1.5 cable entries in the top part using glands according to Ingress Protection class IP66. The inner cores of the cable penetrating the detector should be 200 mm long. This will ensure no tension on the wires and the connectors when the unit is opened. The two mounting bolts are on the top part of the detector. This means that the cables will enter on the fixed part of the detector. The bottom part is attached to the top part by means of four allen screws with water sealing washers. Loosening these screws will expose the connectors. The screws will be retained in the bottom part and will not drop out. The bottom part of the detector is supported by the Load-strap, which is connected to the top part.
The Gassonic Surveyor is mounted from the top with two 8 mm stainless steel bolts. Maximum thread depth is 15 mm. To open the unit, loosen the four screws. The screws are secured, and will not drop out. When closing the unit use a torque of 6Nm on the 4 screws.
The lower part of the unit is supported by a solid cable strap on which it can hang while connecting the cables.
When connecting the cables and moving the jumpers, the power of the unit must be switched off.
See wiring diagram on page 18.
13
3
14
17
2
4
7
6
17
15
15
1
11
12
8
7
6
5
10
9
1. Top
2. Bottom
3. Mounting Bracket UA1352A
4. Sealing flange and sealing gasket
5. Windscreen DS0592
6. Microphone MM4190
7. M5 x 16 Allen screw
8. M5 Sealing washer
}
9. User LED
EU0001
Installation and User Guide for Gassonic Surveyor (BB6019-01)
10. Communication LED
11. Earth connection inside
12. Earth connection outside
13. Cable entries M20 x 1.5
14. Earth wire and strap between Top and Bottom part
15. Protective plate permanently attached
16. Rotary switches
17. Jumpers
18. Eex-i connectors
Page 6
4.1-
mechanical inSTallaTion
Two M8 stainless steel bolts (not supplied), 88mm apart, attached to the top of the detector are used to fix the Gassonic
Surveyor in its operating position. These bolts may penetrate the detector top by a maximum of 14 mm. The detector can be mounted using a Gassonic mounting bracket. The bracket is an optional accessory and is supplied with two M8 mounting U-bolts which can fit around a pole with a maximum dimension of 63 mm.
4.2 -
SpaRe paRTS and acceSSoRieS
DESCRIPTION
Screw & Sealing Washer Kit
Microphone*
Magnet stick
Mounting bracket (standard)
Mounting bracket (45° angle)
Torque wrench
Windscreen*
GASSONIC PART NUMBER
EU0001
MM4190
UC5352
UA1352A
UA1353
EZ1000
DS0592
*The following parts are field replaceable but not consumable parts: Allen screw, Sealing washer, Microphone and
Windscreen.
4.3 -
mechanical opeRaTion and SaFeTY
When closing the top chamber ensure that the Load-strap and wires are not caught between the bottom part and the top part of the detector. Check the status of the rubber sealing and the flange. Send the unit to Gassonic for repair if the flange and/or the rubber sealing is damaged. Check the sealing washers and replace them if they are damaged. The four allen screws with sealing washers must be tightened to 6 Nm using a torque-wrench.
Installation and Maintenance must be carried out by suitably skilled and competent personnel only. Before opening and working on the Gassonic Surveyor in hazardous locations, be sure that power is off.
High readings of ultrasonic sound level may indicate an explosive concentration of gas at the sensor. A subsequent fall in the ultrasonic sound level does not imply that safe working conditions have been restored.
Install and maintain all hazardous area equipment in accordance with the relevant regulations and practices of the country concerned.
warning:
The inner three screws should not be unscrewed and the bottom part should not be opened.
The warranty will be void if the bottom part is opened.
Installation and User Guide for Gassonic Surveyor (BB6019-01) Page 7
5.0 -
alaRm delaY and TRiGGeR leVel
80
60
40 dB
120
100
Trigger level
Delay time Alarm
Ambient background noise
20
0
10 20 30 40
In this example the trigger level is set to 74 dB (corresponding to 12 mA) and the alarm delay time to 10 sec. sec mA
20
12
4
An alarm delay time prevents false alarms when the
Gassonic Surveyor detects spikes of high level ultrasonic noise. After the delay time has run out an alarm will be triggered, but only if the noise level is still above the trigger level. It is imperative to incorporate an alarm delay time of at least 10 seconds. This can be set locally in the unit if using the relay or in the Fire & Gas system if using the 4-20mA output. The trigger level is set at least 6 dB higher than the ambient ultrasonic background noise.
5.1 -
ouTpuT meThodS
The Gassonic Surveyor has two output methods: standard 4-20 ma output.
When using this output method a trigger level and delay time should be set in the Fire & Gas system. alarm relay or error relay.
The unit has one relay output. This can be configured to be either an Alarm or Error relay. When using the Alarm relay, the trigger level and delay time should be set internally.
The relay (Alarm or Error) output is also changeable from NAMUR to Dry contact. It must be decided which output is the best suited for each application. Factory default settings are: Alarm relay and
NAMUR output.
89
94
99
104
69
74
79
84 dB
44
54
59
64 mA
4,0
6,7
8,0
9,3
10,7
12,0
13,3
14,7
16,0
17,3
18,7
20,0
Installation and User Guide for Gassonic Surveyor (BB6019-01)
5.2 -
uSinG The 4-20 ma ouTpuT
When using the 4-20 mA output, the trigger level and the delay time are set in the Fire
& Gas system. The 4-20 mA output is a straight output. This means that if the noise level increases the output current increases as well. The trigger level must be set at least 6 dB higher than the ambient ultrasonic background noise level.
It is good practise to set the internal trigger level to the same level as the 4-20 mA trigger level in the F&G system.
◄ Noise levels (trigger levels) and the relating output values. Other values can be calculated using this formula: mA=(((dB-44)/60)*16)+4 or dB=(((mA-4)*60)/16)+44
Page 8
5.3 -
alaRm RelaY
When using the Alarm relay output, the trigger level must be set internally, together with a delay time. The rotary switches are marked with numbers from 0 to 9 representing a specific value (see table below). The Alarm relay is NC (normally closed) as default factory setting.
It is good practice always to set the internal trigger level at the same level as that on the external system when using the 4 -20 mA output. Doing so will ease the test procedure and the maintenance of the system.
Trigger level switch
The trigger level switch position and the corresponding ultrasonic noise level values.
7
8
9
5
6
3
4
1
2
Position
0
89
94
99
69
74
79
84
Trigger level, dB
54
59
64
▲ Rotary switches for setting trigger level and delay time.
alarm delay switch
The alarm delay switch position and the corresponding delay times. The delay must be at least 10 seconds.
8
9
6
7
4
5
2
3
Position
0
1
15
20
30
45
Delay time, seconds
0
10
60
120 (2 minutes)
240 (4 minutes)
480 (8 minutes)
5.4 -
eRRoR RelaY
In Error relay mode the Gassonic Surveyor will indicate low supply voltage by deactivating the relay, swiching off the user LED and flashing the communication LED. The
Error relay is always NC (normally closed).
The figure shows the jumper in the lower position (factory default setting) to provide Alarm relay functionality. To choose Error relay function instead of Alarm relay function, move the relay jumper to the upper position.
5.5 -
namuR oR dRY conTacT
The figure shows the two jumpers in the lower position
(factory default settings) to provide NAMUR output functionality to either the Alarm or Error relay. To switch the relay mode from NAMUR to dry contact move the two jumpers as shown.
Installation and User Guide for Gassonic Surveyor (BB6019-01) Page 9
6.0 -
TeST and caliBRaTion
The Gassonic Surveyor is tested and calibrated as part of the manufacturing process. However, prior to use, it is good practice to calibrate and test the unit. This can be done using the Gassonic 1701 test and calibration unit.
Start any test and calibration procedure by clipping the
Gassonic 1701 onto the Gassonic Surveyor sensor head.
Lock the tool on the mounting bayonet by rotating it clockwise.
Installation and User Guide for Gassonic Surveyor (BB6019-01)
To activate the calibration mode of the
Gassonic Surveyor place the magnet stick in the notch as illustrated.
The magnet stick is supplied with the Gassonic 1701.
Page 10
1
7.0 -
Gain TeST pRoceduRe
acTion
Switch the Gassonic 1701 on. The version number will display briefly. You must now select the detector type.
2
Select “Gassonic Surveyor” by pressing the Down button and then press Enter.
BuTTon
3
Use default filter 25 kHz by pressing the
Enter button.
diSplaY
Select type
Gassonic
Surveyor
Filter 25 kHz
4
Select test with the Down button.
Select test
5
Select gain test by pressing the Enter button.
Test gain
6
Activate the test by pressing the Enter button.
Activate
“TEST”
7
When the test is finished, the Gassonic 1701 beeps twice, and shows a message, corresponding to the Gassonic Surveyor’s internal settings. In this example the trigger level is 79 dB and the rotary switch is in position 5. If the test is more than ± 3 dB out of tolerance, perform calibration.
If the test was faulty the Gassonic 1701 beeps three times and a No response message displays on the screen. Press Enter to try again, or check the “Trouble shooting”-chapter
Meas: 79
Pos: 5
No response
BuTTon
Installation and User Guide for Gassonic Surveyor (BB6019-01) Page 11
8.0 -
delaY TeST pRoceduRe
After a successful gain test, you must move on to the delay test. Please note that this test will trigger an alarm regardless of output mode (Alarm relay or 4-20 mA). You should be in contact with the staff in the fire and gas control room to measure the delay time and to avoid causing a false alarm. When the control room observes an alarm, press ”ENTER” or
“TEST” to stop the test. The time read-out on the display corresponds to the alarm delay setting of the system. The user interface LED will turn red when the test starts and then flash when the internal delay time has passed.
1
acTion
Switch the 1701 on. The version number will display briefly. You must now select the sensor type.
2
Select “Gassonic Surveyor” by pressing the
Down button and then press Enter.
BuTTon diSplaY
Select type
Gassonic
Surveyor
BuTTon
3
Use default filter 25 kHz by pressing the Enter button.
Filter 25 kHz
4
Select test with the Down button.
Select test
5
Select the delay test by pressing the Enter button.
Test delay
6
To test the internal delay, you must keep an eye on the unit LED as it changes from constant red to flashing red. To test the external alarm delay time you must be in contact with staff in the control room.
7
Press Enter when the LED flashes red. The delay shows on the screen. If you use the analog 4-20 mA output, press Enter when the control room observes an alarm.
Activate
“TEST”
Delay on:
0:10
Installation and User Guide for Gassonic Surveyor (BB6019-01) Page 12
9.0 -
caliBRaTion pRoceduRe
A calibration can be performed at any time. Along with the Gassonic 1701, you will need the magnet-stick to set the
Gassonic Surveyor in calibration mode.
1
acTion
Switch the Gassonic 1701 tool on. The version number will display briefly. You must now select the sensor type.
BuTTon diSplaY
Select type
BuTTon
2
Select “Gassonic Surveyor” by pressing the
Down button and then press Enter.
Gassonic
Surveyor
3
Use default filter 25 kHz by pressing the Enter button.
Filter 25 kHz
4
Select test with the Down button.
Select test
5
Select the Gassonic Surveyor calibration by pressing the Down button, and then press
Enter.
Surveyor
Calibr.
Installation and User Guide for Gassonic Surveyor (BB6019-01) Page 13
6
acTion
Activate the test and calibration mode on the Gassonic Surveyor with the enclosed magnet-stick. Place the magnet-stick in the magnet stick notch next to the interface LED and hold it for more than 3 seconds.
diSplaY
> 3 Sec.
7
The interface LED will flash green and red to indicate that the unit is now in test mode.
The test mode will time out after 60 sec of inactivity and return to normal mode. You may also return the Gassonic Surveyor to normal service by placing the magnet stick in the magnet stick notch again.
8
acTion
Activate the calibration by pressing Enter on the Gassonic 1701.
diSplaY
Activate
Calibr.
9
The calibration is carried out automatically, and will stop when finished. The Gassonic
Surveyor LED turns green.
10
If the test was faulty the red LED on the Gassonic 1701 flashes and the message “No response” is displayed on the screen. The
Gassonic Surveyor user LED will flash green for 5 seconds and then the unit will return to normal operation. Check the chapter
“Trouble shooting” and try calibrating again.
Calibr. complete
No response
BuTTon
Installation and User Guide for Gassonic Surveyor (BB6019-01) Page 14
10.0 -
STep TeST
1
The step test can be carried out to validate the 4-20 mA output to the control room. The test makes it possible to step through different sound pressure levels. Note that this test might trigger an alarm if the sound level exceeds the trigger level.
acTion BuTTon diSplaY BuTTon
Switch on the Gassoinc 1701. The version number will display briefly. You must now select the sensor type.
Select type
2
Select “Gassonic Surveyor” by pressing the
Down button and then press Enter.
Gassonic
Surveyor
3
Use default filter 25 kHz by pressing the
Enter button.
Filter 25 kHz
4
Select test with the Down button.
Select test
5
Select the step test by pressing the Enter button.
Step test
6
Push UP-button to increase the test sound level from 44 to 101 dB.
049 dB
7
Push DOWN-button to decrease the test sound level.
045 dB
8
When all levels are tested, push ENTER to stop the step test.
Step test
Installation and User Guide for Gassonic Surveyor (BB6019-01) Page 15
11.0 -
GaSSonic SuRVeYoR led inTeRFace
The Gassonic Surveyor has two interface LEDs, which will flash red or green depending on the state of the unit. The LED on the front of the unit indicates the current mode of the Gassonic Surveyor, and the LED next to the sensor head communicates with the test- and calibration tool Gassonoc 1701. During service operations with the Gassonic 1701 the communication LED is not visible.
Normal oPeratioN
Normal Operation
Voltage Error
Internal Trigger level reached
Internal Delay time elapsed
Service oPeratioN
Calibration Mode
Calibration Failure
Internal Delay time testing
Internal Delay time elapsed
Front led
Green
Off
Red
Flashing Red
Flashing Red and Green
Flashing Green
Red
Flashing Red communication led
Off
On
On
On
12.0 -
TRouBle ShooTinG
If the calibration or gain test was faulty the Gassonic 1701 beeps three times and a No response message is displayed on the screen.
• Check if there is dirt on the optical window between the 1701 and the Gassonic Surveyor.
• Check if there is dirt on the windscreen or on the microphone.
• Replace microphone if nessesary.
13.0 -
poWeR conSumpTion
The unit can operate with a supply voltage of 9 to 28 Vdc. Power consumption patterns below are calculated with a 33% safety factor, and with the minimum voltage of 9 Vdc.
When using the 4-20 mA and the relay in Error mode
When using the 4-20 mA and the relay in Alarm mode
When using only relay output in Alarm mode
I min
=600 mW
I min
=500 mW
I min
=350 mW
Unit in alarm. Relay energized
Unit in alarm. Relay de-energized
Unit in alarm. Relay de-energized
Installation and User Guide for Gassonic Surveyor (BB6019-01) Page 16
Normal oPeratioN
Normal Operation
Voltage Error
Internal Trigger level reached
Internal Delay time elapsed
Service oPeratioN
Calibration Mode
Calibration Failure
Internal Delay time testing
Internal Delay time elapsed
Front led
Green
Off
Red
Flashing Red
Flashing Red and Green
Flashing Green
Red
Flashing Red communication led
Off
On
On
On
(1.) 4-20mA with 500ohm load, Alarm Relay (De-energized), 101dB background noise, Unit in alarm (2.) 4-20mA with 500ohm load, Error Relay (Energized), 101dB background noise, Unit in alarm (3.) 4-20mA not connected Alarm Relay (De-energized), 101dB background noise, Unit in alarm (4.) 4-20mA not connected Alarm Relay (Energized), 101dB background noise, Unit in alarm
Installation and User Guide for Gassonic Surveyor (BB6019-01) Page 17
(5.) 4-20mA with 500ohm load, Alarm Relay (Energized), 44dB background noise, Unit not in alarm (6.) 4-20mA with 500ohm load, Alarm Relay (Energized), 64dB background noise, Unit not in alarm (7.) 4-20mA with 500ohm load, Alarm Relay (Energized), 84dB background noise, Unit not in alarm (8.) 4-20mA with 500ohm load, Alarm Relay (Energized), 94dB background noise, Unit not in alarm
Installation and User Guide for Gassonic Surveyor (BB6019-01) Page 18
14.0 -
WiRinG
Hazardous area
Gassonic Surveyor
Minimum specification:
Supply voltage: 9 V.
Supply power: 600, mW.
(350 mW. with Io= 0mA)
Safety description:
Specification for the sum of
Pin 1, 3, 4 and 5 in respect to Pin 2:
Ui: 28V dc
Ii: 220mA dc
Pi: 950mW
Ci: 14nF
Li: 10µH
V+
GND
I_SOURCE
3
1
2
NB: Values given to I.S. parameters are the total sum of inputs.
When input voltage is 28 Vdc the current can not exceed 120 mA, and when the input current is equal to
220 mA Vdc the voltage must not exceed 17,5 Vdc.
REL1
REL2
4
5
Safe area
Barriers / Isolators
This could be a number of different barriers or isolator.
(see example drawings)
NB: When choosing barriers or isolators ensure that safety description is fulfilled.
Control room
24VDC
U_SUP
24VDC
0V
U
0V
Min.
5K6
0V
U
Installation in accordance with 60079-14 and -25 or corresponding local regulations and laws.
Specific safety descriptions
Power supply:
U i = 28/17.3 Vdc
I i = 120/220 mAdc
P i = 950 mW
4-20mA source: output:
U o = Ui
I o = Ii
Relay output:
I
U
P i i i = 28/17.3 Vcd
= 120/220 mAdc
= 950 mW
Cabel param from En60079-14:
220 pF/meter
1 uH/meter or 30uH/ohm
IS Earth
Generic wiring diagram for Gassonic Surveyor. Connection data and example wiring diagrams of MTL and Peppel &
Fuchs barriers/isolators can be found in the Appendix. Any barriers/isolators can be used as long as their output safety description does not exceed that of the Gassonic Surveyor. This product is designed to be installed according to IEC
Standard EN 60079-14.
Installation and User Guide for Gassonic Surveyor (BB6019-01)
Wire connectors on the unit are individually marked for correct wiring and reference.
Page 19
15.0 -
Technical daTa
General description
Detector type Ultrasonic Gas Leak
Detector for fixed installations
Sensor technology Stainless steel Microphone technology
Detector frequency range 25 kHz – 70 kHz
Dynamic range
Response time
44 dB – 104 dB SPL
Instant
Detection coverage (leak rate 0.1 kg/s)
5 - 20 meter radius
Power requirements (minimum)
When using 4-20mA and Relay output
Input Voltage
Input Current
9 Vdc
57 mA
Pi
Ci
Ui
Ii
Li
Input Power 600 mW
When using relay output only
Input Voltage 9 Vdc
Input Current
Input Power
Safety description
Power supply
39 mA
350 mW
28 Vdc
220 mA
950 mW
14 nF
10 µH
182 mm / 7.2 inch
Installation and User Guide for Gassonic Surveyor (BB6019-01) output methods
Analogue interface
0 mA
3 mA
Low supply voltage/No power
Unit locally inhibit
4 - 20 mA 44 dB – 104 dB SPL
Relay (option 1 or 2 operator selectable)
Option 1 Alarm relay (switch on preset trigger level)
Option 2 Error relay naMur/dry contact (operator selectable)
NAMUR output
Normal (NC)
Alarm (NO)
Environmental data
760 Ω ±1%
10,422k Ω ±1%
Operating temperature
- 40 °C to + 75 °C
Humidity 0 to 100 % relative humidity
Ingress Protection
Atmospheric conditions
Construction
Housing
IP66
Normal air pressure and oxygen content
Stainless steel AISI 316L
Weight
Vibration sensitivity
2.24 kg
0.13 m/s² perpendicular to the membrane will trigger the unit in the most sensitive range (44dB SPL)
Dimensions
182 x 145 mm
(7.2 x 5.7 inch)
Two gland/conduit entries
M20 x 1.5
RF emission and immunity
Tested according to:
EN61010-1 General electrical safety
RF emission EN61000-6-4
EN61000-6-2 RF immunity
Certifications
ATEX: II 2G EEx ia
IIC T4
IECEX: EEx ia IIC T4
SIL 1/2 rated
Page 20
16.0 -
SupplieRS declaRaTion
supplier’s declaration of conformity in accordance with ec, iec and aTeX directives
No.: BC1019 issuers name and address: General Monitors Ireland Ltd., Ballybrit Business Park, Galway, Republic of Ireland object of the declaration:
Ultrasonic Gas Leak Detector, Gassonic Surveyor
The object of the declaration described above is in conformity with the requirements of the following documents:
DS/IEC 60079-0 2006
DS/IEC 60079-11 2007
Electrical apparatus for explosive gas atmospheres
Part 0: General requirements
Electrical apparatus for explosive gas atmospheres Part 11: Intrinsic Safety ‘i’
DS-EN50014, 1.
DS-EN50020, 2.
2000 Electrical apparatus for potentially explosive atmospheres General requirements
2002-07-01 Electrical apparatus for potentially explosive atmospheres - Intrinsic safety ‘i’
DS-EN61010-1, 2. 2001-12-7 Safety requirements for electrical equipment for measurement, control, and laboratory use Part 1: General requirements.
DS-EN61000-6-3, 1. 2002-05-30 Electromagnetic compatibility (EMC) Part 6-3: Generic standards -
Emission standard for residential, commercial and light-industrial environments.
DS-EN61000-6-2, 2. 2002-05-30 Electromagnetic compatibility (EMC) Part 6-2: Generic standards -
Immunity for industrial environments
ISO 9001 2000 Quality Management System certificate Markings:
DEMKO 08 ATEX 142547X
IECEx ULD 08.0001X
Ex ia IIC T4
0518
II 2G
-40°C to +75°C additional information:
This declaration shall cease to be valid if modifications are made to the equipment without our approval.
Supplier will provide an Instruction Manual in a European Language upon request placed in adequate time.
Responsible Person
Cecil Lenihan
General Manager
Date
The signatory acts on behalf of company management, and with full power of attorney
Installation and User Guide for Gassonic Surveyor (BB6019-01) Page 21
17.0 -
WaRRanTY STaTemenT
Gassonic A/S warrants the Gassonic Surveyor to be free from defects in workmanship or material under normal use and service within two (2) years from the date of shipment. Gassonic A/S will repair or replace without charge any equipment found to be defective during the warranty period. Full determination of the nature of, and responsibility for, defective or damaged equipment will be made by Gassonic A/S personnel. Defective or damaged equipment must be shipped prepaid by appointment with Gassonic
A/S or the representative from which shipment was made. In all cases this warranty is limited to the cost of the equipment supplied by Gassonic A/S. The customer will assume all liability for the misuse of this equipment by its employees or other personnel. All warranties are contingent upon proper use in the application for which the product was intended and do not cover products which have been modified or repaired without Gassonic A/S approval or which have been subjected to neglect, accident, improper installation or application, or on which the original identification marks have been removed or altered. Except for the express warranty stated above, Gassonic A/S disclaims all warranties with regard to the products sold, including all implied warranties of merchantability and fitness and the express warranties stated herein are in lieu of all obligations or liabilities on the part of Gassonic A/S for damages including, but not limited to, consequential damages arising out of/or in connection with the use or performance of the product.
Installation and User Guide for Gassonic Surveyor (BB6019-01) Page 22
18.0 -
appendix
Installation and User Guide for Gassonic Surveyor (BB6019-01) Page 23
Installation and User Guide for Gassonic Surveyor (BB6019-01) Page 24
Installation and User Guide for Gassonic Surveyor (BB6019-01) Page 25
Installation and User Guide for Gassonic Surveyor (BB6019-01) Page 26
GaSSonic SuRVeYoR ulTRaSonic GaS leaK deTecToR
From serial number: 197590
BB6019-01
CAPITOL COPENHAGEN +45 4580 0831
GaSSoNic a/S
Energivej 42A
DK-2750 Ballerup
Denmark
Tel +45 44 700 910
Fax +45 44 700 911 [email protected]
www.gassonic.com
advertisement
Key Features
- Detects pressurized gas leaks by sensing escaping gas's ultrasound
- Detection radius of 5 to 20 meters depending on background noise
- Stainless steel microphone technology for enhanced durability
- Analogue 4-20 mA interface and Alarm or Error relay outputs
- Red and green LED visual output for clear status indication
- Acid-proof stainless steel (316L) housing for corrosion resistance
- ATEX and IEC Ex, Intrinsically Safe (Ex-i) certifications for hazardous area operation
Related manuals
Frequently Answers and Questions
What type of gas leaks can the GASSONIC Surveyor detect?
What is the detection range of the GASSONIC Surveyor?
What types of output does the GASSONIC Surveyor provide?
Is the GASSONIC Surveyor suitable for use in hazardous areas?
advertisement