1, 2, 3, and 4-Gas Detector
Operator’s Manual
Limited Warranty and Limitation Liability
BW Technologies LP (BW) warrants the product to be free from defects in material and workmanship under normal use and service for a period of two years, beginning on the date of shipment to the buyer. This warranty extends only to the sale of new and unused products to the original buyer.
BW’s warranty obligation is limited, at BW’s option, to refund of the purchase price, repair or replacement of a defective product that is returned to a BW authorized service center within the warranty period. In no event shall BW’s liability hereunder exceed the purchase price actually paid by the buyer for the Product.
This warranty does not include: a) fuses, disposable batteries or the routine replacement of parts due to the normal wear and tear of the product arising from use; b) any product which in BW’s opinion, has been misused, altered, neglected or damaged, by accident or abnormal conditions of operation, handling or use; c) any damage or defects attributable to repair of the product by any person other than an authorized dealer, or the installation of unapproved parts on the product; or
The obligations set forth in this warranty are conditional on: a) proper storage, installation, calibration, use, maintenance and compliance with the product manual instructions and any other applicable recommendations of BW; b) the buyer promptly notifying BW of any defect and, if required, promptly making the product available for correction. No goods shall be returned to BW until receipt by the buyer of shipping instructions from BW; and c) the right of BW to require that the buyer provide proof of purchase such as the original invoice, bill of sale or packing slip to establish that the product is within the warranty period.
THE BUYER AGREES THAT THIS WARRANTY IS THE BUYER
’
S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES
,
EXPRESS OR IMPLIED
,
INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE
.
BW SHALL NOT BE LIABLE FOR ANY
SPECIAL
,
INDIRECT
,
INCIDENTAL
,
OR BASED ON CONTRACT
,
TORT OR RELIANCE OR ANY OTHER THEORY
.
Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or limitation of incidental or consequential damages, the limitations and exclusions of this warranty may not apply to every buyer. If any provision of this warranty is held invalid or unenforceable by a court of competent jurisdiction, such holding will not affect the validity or enforceability of any other provision.
Contacting BW Technologies by Honeywell
USA: 1-888-749-8878
Europe: +44(0) 1295 700300
Canada: 1-800-663-4164
Other countries: +1-403-248-9226
Email us at: [email protected]
Visit BW Technologies by Honeywell’s website at: www.gasmonitors.com
Introduction
The operator’s manual provides basic information to operate the GasAlertQuattro gas detector. For complete operating instructions, refer to the GasAlertQuattro Technical Reference
Guide provided on the CD-ROM. The GasAlertQuattro gas detector (“the detector”) is designed to warn of hazardous gas levels above user-defined alarm setpoints.
The detector is a personal safety device. It is your responsibility to respond properly to the alarm.
Note
The detector is shipped with English as the default displayed language. Additional languages provided are French, German, Portuguese, and Spanish. The screens for the additional languages are displayed on the detector and in the corresponding operator’s manual.
Zeroing the Sensors
To zero the sensors, refer to steps #1-3 in Calibration on page 8.
GasAlertQuattro
Safety Information - Read First
Use the detector only as specified in this operator’s manual and the technical reference guide, otherwise protection provided by the detector may be impaired. Read the following Cautions before using the detector.
a
Cautions
•
Warning: Substitution of components may impair Intrinsic
Safety.
• Before using the detector, refer to
.
• Protect the combustible sensor from exposure to lead compounds, silicones, and chlorinated hydrocarbons.
Although certain organic vapors (such as leaded gasoline and halogenated hydrocarbons) may temporarily inhibit sensor performance, in most cases the sensor will recover after calibration.
• Caution: For safety reasons, this equipment must be operated and serviced by qualified personnel only. Read and understand the technical reference guide completely before operating or servicing.
1
GasAlertQuattro
Operator’s Manual
• Charge the detector before first-time use. BW Technologies by Honeywell recommends the detector be charged after every workday.
• Calibrate the detector before first-time use and then on a regular schedule, depending on use and sensor exposure to poisons and contaminants. BW recommends that the sensors must be calibrated regularly and at least once every 180 days (6 months).
• Calibrate only in a safe area that is free of hazardous gas in an atmosphere of 20.9% oxygen.
• The combustible sensor is factory calibrated to 50% LEL methane. If monitoring a different combustible gas in the
% LEL range, calibrate the sensor using the appropriate gas.
• Only the combustible gas detection portion of this instrument has been assessed for performance by CSA
International.
• BW recommends that the combustible sensor be checked with a known concentration of calibration gas after any exposure to contaminants/poisons such as sulfur compounds, silicon vapors, halogenated compounds, etc.
• BW recommends to bump test the sensors before each day’s use to confirm their ability to respond to gas by exposing the detector to a gas concentration that exceeds the alarm setpoints. Manually verify that the audible, visual, and vibrator alarms are activated. Calibrate if the readings are not within the specified limits.
• Caution: High off-scale LEL readings may indicate an explosive concentration.
2
• Any rapid upscaling reading followed by a declining or erratic reading may indicate a gas concentration beyond upper scale limit, which can be hazardous.
• For use only in potentially explosive atmospheres where oxygen concentrations do not exceed 20.9% (v/v).
• Extended exposure of the GasAlertQuattro to certain concentrations of combustible gases and air may stress the detector element that can seriously affect its performance. If an alarm occurs due to a high concentration of combustible gases, a calibration should be performed, or if needed, the sensor replaced.
• Warning
: The lithium battery (QT-BAT-R01) may present a risk of fire or chemical burn hazard if misused. Do not disassemble, heat above 212°F (100°C), or incinerate.
• Warning: Do not use any other lithium batteries with the
GasAlertQuattro detector. Use of any other cell can cause fire and/or explosion. To order and replace the QT-BAT-R01 lithium battery, contact BW Technologies by Honeywell .
•
•
Warning: Lithium polymer cells exposed to heat at 266°F
(130°C) for 10 minutes can cause fire and/or explosion.
ec
Warning : This instrument contains a lithium polymer battery. Dispose of used lithium cells immediately. Do not disassemble and do not dispose of in fire. Do not mix with the solid waste stream. Spent batteries must be disposed of by a qualified recycler or hazardous materials handler.
• Keep lithium cells away from children.
• Deactivating the detector by removing the battery pack may cause improper operation and harm the detector.
GasAlertQuattro
Parts of the GasAlertQuattro
Parts of the GasAlertQuattro
Item
1
2
3
Description
IntelliFlash
(green LED)
Visual alarm indicator
(red LED)
Hydrogen sulfide
(H
2
S) sensor
Item
4
Description
Pushbutton
5
6
Combustible (LEL) sensor
Carbon monoxide
(CO) sensor
Item
7
Description
Liquid crystal display
(LCD)
8
Audible alarm
9
Oxygen (O
2
) sensor
Item
10
Description
Alligator clip
11
12
Battery pack
Charging connector and IR interface
3
GasAlertQuattro
Operator’s Manual
Screen Elements
4
GasAlertQuattro
Pushbutton
Pushbutton
Pushbutton
C
Description
• To activate the detector, press and hold C in a safe area that is free of hazardous gas and in an atmosphere of 20.9% oxygen.
• To deactivate the detector, press and hold C during the powering off countdown. Release
C when OFF displays.
• To view the date/time, current battery power, calibration due date, bump test due date,
TWA, STEL, and peak readings, press C twice rapidly. To clear the TWA, STEL, and peak readings, press and hold C when the LCD displays Hold C to reset peaks, TWA,
STEL.
• To initiate calibration, press and hold C while the detector performs the OFF countdown.
Continue holding C while the LCD briefly deactivates and then reactivates to begin the calibration countdown. Release C when Calibration started displays.
• To activate the backlight, press C and release.
• To acknowledge latching alarms, press C.
• To acknowledge a low alarm and disable the audible alarm, press C. The Low Alarm
Acknowledge option must be enabled in Fleet Manager II.
• To acknowledge any of the “due today” messages (calibration and bump test) press C.
If enabled, the force calibration and force bump features cannot be bypassed.
5
GasAlertQuattro
Operator’s Manual
Sensor Poisons and Contaminants
Several cleaners, solvents, and lubricants can contaminate and cause permanent damage to sensors. Before using cleaners, solvents, and lubricants in close proximity to the detector sensors, read the following caution and table.
a
Caution
Use only the following BW Technologies by Honeywell recommended products and procedures:
• Use water based cleaners.
• Use non-alcohol based cleaners.
• Clean the exterior with a soft, damp cloth.
• Do not use soaps, polishes, or solvents.
The following table lists common products to avoid using around sensors.
Cleaners and
Lubricants
Silicones Aerosols
Brake cleaners
Bug repellents and sprays
Lubricants
Rust inhibitors
Silicone cleaners and protectants
Silicone based adhesives, sealants, and gels
Hand/body and medicinal creams that contain silicone
Tissues containing silicone
Mold releasing agents
Lubricants
Window and glass cleaners
Dishsoaps
Citrus based cleaners
Alcohol based cleaners
Hand sanitizer
Anionic detergents
Methanol
(fuels and antifreezes)
Polishes
Rust inhibitors
Window cleaners
6
Connecting the Gas Cylinder to the Detector
GasAlertQuattro
Connecting the Gas Cylinder to the Detector
7
GasAlertQuattro
Operator’s Manual
Calibration
Calibration is performed to adjust the sensitivity levels of sensors to ensure accurate responses to gas.
This calibration procedure is written as the procedure is intended.
If an error or alarm screen displays, refer to Calibration Troubleshooting in the GasAlertQuattro Technical Reference Guide.
a
Caution
Calibrate only in a safe area that is free of hazardous gas in an atmosphere of 20.9% oxygen.
If performing single gas calibration, calibrate O
2
first.
Note
The maximum hose length for calibration is 3 ft. (1 m).
The following steps are written for use with a standard quad gas cylinder.
Calibration can only be aborted after the sensors have
been zeroed. If C is pressed to abort, CALIBRATION
cancelled displays.
1. Press and hold C as the detector performs the Powering
off countdown.
Continue holding C when OFF displays and the detector briefly deactivates.
2. The detector activates again and performs the calibration countdown. Continue holding until
Starting Calibration displays.
3. The detector enters the zero function. zeroing displays while the detector zeros all of the sensors.
a
Caution
If a sensor fails to zero, it cannot be calibrated. Refer to Startup Self-Test
Troubleshooting in the
GasAlertQuattro Technical Reference
Guide.
8
If the IR Lock option is enabled, the following screen displays to indicate calibration can only be performed using an IR device (MicroDock II or
IR Link).
4. When the following screen displays, attach the calibration cap and apply calibration gas at a flow rate of 250-500 ml/min. Refer to
Connecting the Gas Cylinder to the Detector .
If a sensor is not yet due for calibration, its box will have a greyed out checkmark.
5. The detector initially tests for gas. When a sufficient amount of gas is detected, displays beside each gas that is detected.
6. The detector then begins calibrating the sensors. The following activities occur during the span:
• calibrating displays at the bottom of the screen.
• Gas values adjust during the span.
• Target gas values that are defined in Fleet Manager II display above or below the adjusting gas value.
To abort calibration after the sensors have been zeroed, press C.
7. When the following screen displays, close the valve on the gas cylinder and remove the calibration cap from the detector.
A check mark displays beside each sensor that has calibrated successfully.
GasAlertQuattro
Calibration
9
GasAlertQuattro
Operator’s Manual
8. When calibration is complete, the following screen displays.
Note
The calibration due date cannot be reset for a sensor that fails calibration. If a sensor fails or an error screen displays, refer to
Calibration Troubleshooting in the
GasAlertQuattro Technical
Reference Guide.
9. All successfully calibrated sensors automatically reset to the number of days defined in the Cal
Interval field in Fleet Manager II.
The calibration due dates can be changed in Fleet Manager II.
10. The detector now enters normal operation.
10
Bump Test
A bump test applies test gas to force the detector into alarm. A bump test should be performed regularly to confirm the sensors are responding correctly to gas, and that the audible, visual and vibrator alarms activate during an alarm condition.
The detector can also prompt for a bump test during startup when the Bump Test Interval is defined. Refer to the GasAlert-
Quattro Technical Reference Guide.
a
Caution
BW recommends to bump test the sensors before each day’s use to confirm their ability to respond to gas by exposing the sensors to a gas concentration that exceeds alarm setpoints.
1.
Connect the calibration hose to the 0.5 l/min regulator on the gas cylinder. Refer to
Connecting the Gas Cylinder to the Detector
.
To bump test using the MicroDock II station, refer to the MicroDock II User Manual.
2.
Connect the calibration hose to the intake inlet on the calibration cap. Arrows on the calibration cap indicate the direction of gas flow.
3.
Attach and tighten the calibration cap onto the detector and apply gas. Verify the visual, audible, and vibrator alarms activate.
4.
Close the regulator and remove the calibration cap.
The detector temporarily remains in alarm until the gas clears from the sensors.
GasAlertQuattro
Alarms
Alarms
Refer to the following table for information about alarms and corresponding screens. For more information about alarms refer to the
GasAlertQuattro Technical Reference Guide.
Alarm Screen Alarm Screen
Low Alarm
• Slow siren (upward tone)
• Slow flash
• Black box around gas flashes
• Vibrator alarm activates
TWA Alarm
• Fast siren (downward tone)
• Fast flash
• Black box around gas flashes
• Vibrator alarm activates
High Alarm
• Fast siren (downward tone)
• Fast flash
• Black box around gas flashes
• Vibrator alarm activates
STEL Alarm
• Fast siren (downward tone)
• Fast flash
• Black box around gas flashes
• Vibrator alarm activates
11
GasAlertQuattro
Operator’s Manual
Alarm
Multi Alarm
• Alternating low and high alarm siren and flash
• Black box around gas flashes
• Type of alarm alternates
• Vibrator alarm activates
Sensor Failure Alarm
Screen Alarm
Over Limit (OL) Alarm
• Fast siren (downward tone)
• Fast flash
• Black box around gas flashes
• Vibrator alarm activates
Note: LCD may also display an under limit reading (-OL)
Normal Deactivation
• Sequence of alternating beeps and alternating flashes
• Vibrator alarm activates
• Countdown initiates
• OFF displays
Screen
12
Note
If enabled, during an alarm condition the Latching Alarms option causes the low and high gas alarms (audible, visual, and
vibrator) to persist until the alarm is acknowledged by pressing C and the gas concentration is below the low alarm setpoint.
The peak concentration values display continually until the alarm no longer exists. Enable/disable Latching Alarms in Fleet
Manager II. Local regulations may require the Latching Alarms option be enabled.
Alarm
Low Battery Alarm
• Sequence of 10 rapid sirens and alternating flashes with
7 seconds of silence in between
(continues for 15 minutes)
• Vibrator alarm pulses
• After 15 minutes of the low battery alarm sequence, the detector enters critical battery alarm (see
Critical Battery Alarm below)
Screen
Critical Battery Alarm
• Fifteen minutes after low battery alarm activates, sequence of 10 rapid sirens and alternating flashes with 1 second of silence in between (sequence reactivates seven times)
• Vibrator alarm pulses
• Low Battery Powering Off displays and the detector deactivates
GasAlertQuattro
Alarms
Alarm Screen
Confidence/compliance Beep
• One beep every 1-120 seconds
(beep frequency is defined with
Confidence/compliance Beep
Interval option)
IntelliFlash
(default: one flash every 1 second)
• One flash every 1-120 seconds
(flash frequency is defined with
IntelliFlash Interval option)
Heartbeat
• pulses once every second to verify detector is operating correctly
Note: Confidence/ compliance beep and
IntelliFlash automatically deactivate during a low battery alarm, calibration fail, bump test fail, self-test fail, and in an alarm condition.
Note
If the Low Alarm Acknowledge option is enabled, the
audible alarm can be disabled during a low alarm condition.
The LED and visual alarm indicators remain active until the alarm condition changes or the detector deactivates. Press
C to acknowledge the low alarm and deactivate the audible
alarm. If the alarm escalates to a high, TWA, or STEL alarm, the audible alarm reactivates.
13
GasAlertQuattro
Operator’s Manual
User Options and Sensor Configuration
The following items are required to modify the user options and sensor configuration:
• Detector
• IR Link adapter or MicroDock II
• Fleet Manager II software
Refer to the GasAlertQuattro Technical Reference Guide and
Fleet Manager II Operator’s Manual for complete information.
Maintenance
To maintain the detector in good operating condition, perform the following basic maintenance as required.
• Calibrate, bump test, and inspect the detector on a regular schedule.
• Maintain an operations log of all maintenance, bump tests, calibrations, and alarm events.
• Clean the exterior with a soft damp cloth. Do not use solvents, soaps, or polishes. Refer to
Rechargeable Battery Capacity
A rechargeable battery’s runtime decreases approximately 20% over a two-year period of typical use.
Battery Pack Retaining Screw
The retaining screw (QAQD-20x) provided with the detector must be used to lock the battery pack on all European and
14
IECEx scheme detectors, and on all Canadian and U.S. Zone
Certified detectors.
The screwdriver included with the detector has a double-ended driver. Loosen the brass nut to switch between a Phillips head and a hex head.
A hex tool is required to tighten and loosen the retaining screw.
Tighten the screw 1-2 turns using 3-4 in-lbs of torque. Do not overtighten the screw.
Replacing the Battery Pack
The alkaline and rechargeable battery packs can be changed in hazardous locations.
1.
Press and hold C to deactivate the detector.
2.
If using the retaining screw, loosen it 1-2 turns. Push the battery release latch toward the top of the detector to release the battery pack.
3.
From the top of the battery pack, lift upward to remove.
4.
Insert a new battery pack. Insert the bottom of the battery pack first, then lower the top into place. Press until the release tab engages. Tighten the retaining screw if required.
Charging the Rechargeable Battery Pack
a
Warning
To avoid personal injury and/or damage to the detector, adhere to following:
Charge only in a safe area that is free of hazardous gas within temperatures of 32°F to 104°F (0°C to 40°C).
Charge the battery immediately when the detector emits a low battery alarm.
Charge the lithium battery pack using the BW supplied charger and charger adapter only. The charging adapter is specific to your region. Use of the charging adapter outside your region will damage the charger
GasAlertQuattro
Maintenance
and the detector. Failure to adhere to this caution can lead to fire and/or explosion.
Charge the lithium battery after each workday.
1.
Press and hold C to deactivate the detector, then plug the charger into an AC outlet.
Note
The time required to charge will increase if the detector is activated.
2.
Connect the charging adapter to the detector IR receptacle. Refer to the following illustration.
3.
The lithium battery may require 6 hours to reach full capacity.
15
GasAlertQuattro
Operator’s Manual
Replacing the Alkaline Batteries
a
Warning
To avoid personal injury and/or damage to the detector, use only BW recommended alkaline batteries. Refer to
Change the alkaline batteries only in safe area that is free of hazardous gas.
1.
Press and hold C to deactivate the detector.
2.
If using the retaining screw, loosen the retaining screw 1-2 turns. Remove the alkaline battery pack.
Refer to
.
3.
Unhook the ejector bar from the release clasp. Move the ejector bar towards the top of the battery pack until it is aligned horizontally over the batteries.
4.
Using the tab, pull on the ejector bar.
5.
To the left of the tab, pull up on the ejector bar.
6.
Remove the spent batteries. Reset the ejector bar to its original flat position. Ensure the ejector bar engages the release clasp.
16
7.
Insert the new batteries. Position the positive end of the battery at a 30° angle and insert into the battery pack before pushing the negative end down. Ensure the batteries are not inserted over the tab.
8.
Replace the battery pack by inserting the bottom first, then lower top into place. Ensure the tab is tucked in before replacing the battery pack.
Press until the release tab engages. If required, tighten the retaining screw using 3-4 in. lbs torque.
GasAlertQuattro
Maintenance
17
GasAlertQuattro
Operator’s Manual
Replacing the Sensors
18
a
Warning
To avoid personal injury and/or property damage, only use sensors that are specifically designed for the detector.
Note
Detectors that are configured for 1, 2, or 3 gases may contain a dummy sensor in one of the four sensor locations.
To replace a sensor or sensor filter, refer to the illustration
and the following steps #1-8.
1.
Press and hold C to deactivate the detector. Press the release latch, and remove the battery pack.
2.
Remove the six machine screws from the rear shell.
3.
Remove the front shell.
4.
Remove the spent sensor(s). Ensure no damage occurs to the LCD.
5.
Insert the new sensor(s).
6.
Reassemble the detector. Press the front and rear shells together firmly to ensure a proper seal. Ensure the front and rear shells have a tight, uniform 1/16 in.
(1.5 mm) seal on all sides of the detector.
7.
Replace the six machine screws using 3-4 in. lbs torque. Do not overtighten the screws. Replace the battery pack.
8.
New sensors must be calibrated prior to use.
Calibrate the new sensor(s) immediately. Refer to
.
GasAlertQuattro
Maintenance
Replacing the Sensor Filter
To replace the filter, refer to the illustration Replacing the Sensors
and the following steps #1-6.
1.
Press and hold C to deactivate the detector. Press the release latch, and remove the battery pack.
2.
Remove the six machine screws from the rear shell.
3.
Remove the front shell. Remove the sensor filter.
4.
Refer to the following illustration before inserting the new filter. Ensure the filter is laying flat and that the holes are correctly aligned over the filter posts.
5.
Replace the front shell. Press the front and rear shells together firmly to ensure a proper seal. Ensure the front and rear shells have a tight, uniform 1/16 in.
(1 mm) seal on all sides of the detector.
6.
Replace the six machine screws using 3-4 in. lbs torque. Do not overtighten the screws. Replace the battery pack.
19
GasAlertQuattro
Operator’s Manual
Specifications
Instrument dimensions: 12.9 x 8.0 x 3.8 cm
(5.1 x 3.2 x 1.5 in.)
Weight: 330 g (11.7 oz.)
Operating temperature: -20°C to +50°C (-4°F to +122°F)
Storage temperature: -40°C to +60°C (-40°F to +140°F)
Operating humidity: 10% to 100% relative humidity
(non-condensing)
Dust and moisture ingress: IP66/67 (with screw engaged)
Alarm setpoints: May vary by region and are user-defined
Detection range:
H
2
S: 0 - 200 ppm (0.1 ppm increments)
CO: 0 - 1000 ppm (1 ppm increments)
O
2
: 0 - 30.0% vol. (0.1% vol. increments)
Combustible (LEL): 0 - 100% (1% LEL increments) or
0 - 5.0% v/v methane
Sensor type:
H
2
S, CO, O
2
: Single plug-in electrochemical cell
Combustibles: Plug-in catalytic bead
O
2
measuring principle: Capillary controlled concentration sensor
Bump test specified limits: BW recommends using a gas cylinder that will ensure the combustible sensor has an accuracy of -0 to +20% of actual reading (reference CAN/CSA C22.2 No.
152)
Alarm conditions: TWA alarm, STEL alarm, low alarm, high alarm, multi alarm, over limit (OL) alarm, low battery alarm, critical low battery alarm, sensor failure alarm, IntelliFlash, confidence/compliance beep
20
Audible alarm: 95 dB at 30 cm (12 in.) variable pulsed beeper
Visual alarm: Red light-emitting diodes (LEDs)
IntelliFlash: Green light-emitting diode. Flash frequency is user-defined with IntelliFlash interval option
Confidence/compliance beep: Audible beep from variable pulsed beeper. Beep frequency is user-defined with confidence/ compliance beep interval option
Display: Alphanumeric liquid crystal display (LCD) with flip display (0° or 180°) capability (user-defined in Fleet Manager II)
Backlight: Activates upon startup and deactivates when selftest is complete. Activates when the pushbutton is pressed and deactivates after 10 seconds. Also activates during an alarm condition and remains lit until alarm ceases
Internal vibrator: Vibrates during activation, deactivation, and all alarms
Self-test: Initiated during activation, self-test runs continuously on the battery and electrochemical sensors (H
2 detector is operational
S and CO) while
Calibration: Zero and automatic span
User options: Startup message, lockout on self-test error, safe mode, IntelliFlash, confidence/compliance beep, latching alarms, force calibration, force bump, calibration IR lock, flip display stealth mode, datalog interval, IntelliFlash interval, confidence/compliance beep interval, and language selection
Sensor options: Sensor enable/disable, calibration gas values, calibration interval, bump test interval, alarm setpoints
(low/high/TWA/STEL), STEL interval, TWA period, auto zero at startup enable/disable, LEL correction factor, 10% (of reading) over-span, low alarm acknowledge, O
2
measurement, LEL gas measurement, %vol methane measurement
Year of manufacture: The detector's year of manufacture is determined from the serial number. The second and third number after the second letter determines the year of manufacture.
E.g., QA109-001000 = 2009 year of manufacture
Approved lithium battery for GasAlertQuattro product:
Lithium-ion polymer (QT-BAT-R01) as per standards UL913,
EN60079-11, EN60079-0, C22.2 No. 157
Rechargeable battery (QT-BAT-R01)
Lithium polymer -20°C
≤ Ta ≤ +50°C
Temperature code
T4
Lithium battery operating time: One rechargeable lithium polymer battery provides the following operating runtimes:
20 hours at 20°C (68°F)
18 hours at -20°C (-4°F)
Approved alkaline battery pack for GasAlertQuattro
(QT-BAT-A01): as per standards UL913, EN60079-11,
EN60079-0, C22.2 No. 157
Approved alkaline batteries for GasAlertQuattro product:
Duracell MN1500 -20°C
≤ Ta ≤ +50°C
Energizer E91VP -20°C
≤ Ta ≤ +50°C
T4 (129.9°C)
T3C (135.3°C)
AA alkaline battery operating time:
14 hours at 20°C (68°F)
Battery charger: Charging adapter
First-time charge: 6 hours
Normal charge: 6 hours
Warranty: 2 years including sensors
GasAlertQuattro
Specifications
Approvals:
Approved by CSA to both U.S. and Canadian Standards
CAN/CSA C22.2 No. 157 and C22.2 152
ANS/UL - 913 and ANSI/ISA - S12.13 Part 1
CSA
ATEX
IECEx
Class I, Division 1, Group A, B, C, D
CE 0539 g II 1 G Ga Ex ia IIC T4 for Zone 0
Group IIC
KEMA 09 ATEX 0137
EN 60079-0, EN 60079-11, EN 60079-26
Ga Ex ia IIC T4 IECEx CSA 09.0006
IEC 60079-0, IEC 60079-11, IEC 60079-26
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules and ICES-003 Canadian EMI requirements. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
21
GasAlertQuattro
Operator’s Manual
22
GasAlertQuattro
Specifications
23
GasAlertQuattro
Operator’s Manual
24
iERP: 127890
D6445/0 [English]
© BW Technologies 2009. All rights reserved.
1, 2, 3, and 4-Gas Detector
Operator’s Manual
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 2 Limited Warranty and Limitation Liability
- 2 Contacting BW Technologies by Honeywell
- 3 Introduction
- 3 Zeroing the Sensors
- 3 Safety Information - Read First
- 5 Parts of the GasAlertQuattro
- 6 Screen Elements
- 7 Pushbutton
- 8 Sensor Poisons and Contaminants
- 9 Connecting the Gas Cylinder to the Detector
- 10 Calibration
- 12 Bump Test
- 13 Alarms
- 16 User Options and Sensor Configuration
- 16 Maintenance
- 16 Rechargeable Battery Capacity
- 16 Battery Pack Retaining Screw
- 17 Replacing the Battery Pack
- 17 Charging the Rechargeable Battery Pack
- 18 Replacing the Alkaline Batteries
- 20 Replacing the Sensors
- 21 Replacing the Sensor Filter
- 22 Specifications