Fuse Holder Assembly Replacement
Model 410 MKIII
Tools and Materials Needed
1. Fuse Holder Assembly (Spare) (#109584)
2. 5/16" Wrench
3. 11/16" Wrench
8 Amp Fuse Retaining Nut Cap
Instructions
1. Disconnect the Pump from the power supply.
2. Remove the Fuse from the Fuse Holder by twisting the
Cap on the outside of the Pump Housing, and removing the Fuse from the Cap.
(Not Included) Fuse Holder
Fuse Holder Assembly
5. Disconnect the red and white wires from the Fuse Holder, by pulling off the quick-connect fittings.
6. Use the 11/16" wrench to remove the Retaining Nut from the Fuse Holder.
Retaining Nut
3. Using the 5/16" wrench, remove the four screws from the
End Plate (opposite end from Pump Head).
4. Carefully slide the Control Board out of the enclosure to allow access to the Fuse Holder.
Control Board
Quick-connect
Fittings
7. Pull out to remove the Fuse Holder from the Pump
Housing.
8. Insert the new Fuse Holder through the Pump Housing
(remove the Retaining Nut from the Fuse Holder first).
9. Use the 11/16" wrench to secure the Fuse Holder to the
Pump Housing with the Retaining Nut.
10. Reconnect the red and white wires to the new Fuse
Holder. The white wire goes to the connection on the end of the Fuse Holder, and the red wire to the connection on the side.
11. Slide the Control Board back into the enclosure in the slots, plugs facing down.
12. Carefully push all wires into the housing and screw the End
Plate back on the Pump.
Printed in Canada
May 12, 2014
(#110794)