NI 4060 User Manual


Add to my manuals
64 Pages

advertisement

NI 4060 User Manual | Manualzz

Computer-Based

Instruments

NI 4060 User Manual

Digital Multimeter for PCI and

PXI/CompactPCI

NI 4060 User Manual

March 2000 Edition

Part Number 321760B-01

Worldwide Technical Support and Product Information

www.ni.com

National Instruments Corporate Headquarters

11500 North Mopac Expressway Austin, Texas 78759-3504 USA Tel: 512 794 0100

Worldwide Offices

Australia 03 9879 5166, Austria 0662 45 79 90 0, Belgium 02 757 00 20, Brazil 011 284 5011,

Canada (Calgary) 403 274 9391, Canada (Ontario) 905 785 0085, Canada (Québec) 514 694 8521,

China 0755 3904939, Denmark 45 76 26 00, Finland 09 725 725 11, France 01 48 14 24 24,

Germany 089 741 31 30, Greece 30 1 42 96 427, Hong Kong 2645 3186, India 91805275406,

Israel 03 6120092, Italy 02 413091, Japan 03 5472 2970, Korea 02 596 7456, Mexico (D.F.) 5 280 7625,

Mexico (Monterrey) 8 357 7695, Netherlands 0348 433466, New Zealand 09 914 0488, Norway 32 27 73 00,

Poland 0 22 528 94 06, Portugal 351 1 726 9011, Singapore 2265886, Spain 91 640 0085,

Sweden 08 587 895 00, Switzerland 056 200 51 51, Taiwan 02 2528 7227, United Kingdom 01635 523545

For further support information, see the

Technical Support Resources

appendix. To comment on the documentation, send e-mail to [email protected]

© Copyright 1998, 2000 National Instruments Corporation. All rights reserved.

Important Information

Warranty

The NI 4060 is warranted against defects in materials and workmanship for a period of one year from the date of shipment, as evidenced by receipts or other documentation. National Instruments will, at its option, repair or replace equipment that proves to be defective during the warranty period. This warranty includes parts and labor.

The media on which you receive National Instruments software are warranted not to fail to execute programming instructions, due to defects in materials and workmanship, for a period of 90 days from date of shipment, as evidenced by receipts or other documentation. National Instruments will, at its option, repair or replace software media that do not execute programming instructions if National Instruments receives notice of such defects during the warranty period. National Instruments does not warrant that the operation of the software shall be uninterrupted or error free.

A Return Material Authorization (RMA) number must be obtained from the factory and clearly marked on the outside of the package before any equipment will be accepted for warranty work. National Instruments will pay the shipping costs of returning to the owner parts which are covered by warranty.

National Instruments believes that the information in this document is accurate. The document has been carefully reviewed for technical accuracy. In the event that technical or typographical errors exist, National Instruments reserves the right to make changes to subsequent editions of this document without prior notice to holders of this edition. The reader should consult

National Instruments if errors are suspected. In no event shall National Instruments be liable for any damages arising out of or related to this document or the information contained in it.

E

XCEPT AS SPECIFIED HEREIN

, N

ATIONAL

I

NSTRUMENTS MAKES NO WARRANTIES

,

EXPRESS OR IMPLIED

,

AND SPECIFICALLY DISCLAIMS ANY

WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE

. C

USTOMER

S RIGHT TO RECOVER DAMAGES CAUSED BY FAULT OR

NEGLIGENCE ON THE PART OF

N

ATIONAL

I

NSTRUMENTS SHALL BE LIMITED TO THE AMOUNT THERETOFORE PAID BY THE CUSTOMER

. N

ATIONAL

I

NSTRUMENTS WILL NOT BE LIABLE FOR DAMAGES RESULTING FROM LOSS OF DATA

,

PROFITS

,

USE OF PRODUCTS

,

OR INCIDENTAL OR

CONSEQUENTIAL DAMAGES

,

EVEN IF ADVISED OF THE POSSIBILITY THEREOF

. This limitation of the liability of National Instruments will apply regardless of the form of action, whether in contract or tort, including negligence. Any action against National Instruments must be brought within one year after the cause of action accrues. National Instruments shall not be liable for any delay in performance due to causes beyond its reasonable control. The warranty provided herein does not cover damages, defects, malfunctions, or service failures caused by owner’s failure to follow the National Instruments installation, operation, or maintenance instructions; owner’s modification of the product; owner’s abuse, misuse, or negligent acts; and power failure or surges, fire, flood, accident, actions of third parties, or other events outside reasonable control.

Copyright

Under the copyright laws, this publication may not be reproduced or transmitted in any form, electronic or mechanical, including photocopying, recording, storing in an information retrieval system, or translating, in whole or in part, without the prior written consent of National Instruments Corporation.

Trademarks

CVI

, LabVIEW

, NI-DAQ

, National Instruments

, ni.com

, and PXI

are trademarks of National Instruments Corporation.

Product and company names mentioned herein are trademarks or trade names of their respective companies.

WARNING REGARDING USE OF NATIONAL INSTRUMENTS PRODUCTS

(1) NATIONAL INSTRUMENTS PRODUCTS ARE NOT DESIGNED WITH COMPONENTS AND TESTING FOR A LEVEL

OF RELIABILITY SUITABLE FOR USE IN OR IN CONNECTION WITH SURGICAL IMPLANTS OR AS CRITICAL

COMPONENTS IN ANY LIFE SUPPORT SYSTEMS WHOSE FAILURE TO PERFORM CAN REASONABLY BE

EXPECTED TO CAUSE SIGNIFICANT INJURY TO A HUMAN.

(2) IN ANY APPLICATION, INCLUDING THE ABOVE, RELIABILITY OF OPERATION OF THE SOFTWARE PRODUCTS

CAN BE IMPAIRED BY ADVERSE FACTORS, INCLUDING BUT NOT LIMITED TO FLUCTUATIONS IN ELECTRICAL

POWER SUPPLY, COMPUTER HARDWARE MALFUNCTIONS, COMPUTER OPERATING SYSTEM SOFTWARE

FITNESS, FITNESS OF COMPILERS AND DEVELOPMENT SOFTWARE USED TO DEVELOP AN APPLICATION,

INSTALLATION ERRORS, SOFTWARE AND HARDWARE COMPATIBILITY PROBLEMS, MALFUNCTIONS OR

FAILURES OF ELECTRONIC MONITORING OR CONTROL DEVICES, TRANSIENT FAILURES OF ELECTRONIC

SYSTEMS (HARDWARE AND/OR SOFTWARE), UNANTICIPATED USES OR MISUSES, OR ERRORS ON THE PART OF

THE USER OR APPLICATIONS DESIGNER (ADVERSE FACTORS SUCH AS THESE ARE HEREAFTER

COLLECTIVELY TERMED “SYSTEM FAILURES”). ANY APPLICATION WHERE A SYSTEM FAILURE WOULD

CREATE A RISK OF HARM TO PROPERTY OR PERSONS (INCLUDING THE RISK OF BODILY INJURY AND DEATH)

SHOULD NOT BE RELIANT SOLELY UPON ONE FORM OF ELECTRONIC SYSTEM DUE TO THE RISK OF SYSTEM

FAILURE. TO AVOID DAMAGE, INJURY, OR DEATH, THE USER OR APPLICATION DESIGNER MUST TAKE

REASONABLY PRUDENT STEPS TO PROTECT AGAINST SYSTEM FAILURES, INCLUDING BUT NOT LIMITED TO

BACK-UP OR SHUT DOWN MECHANISMS. BECAUSE EACH END-USER SYSTEM IS CUSTOMIZED AND DIFFERS

FROM NATIONAL INSTRUMENTS' TESTING PLATFORMS AND BECAUSE A USER OR APPLICATION DESIGNER

MAY USE NATIONAL INSTRUMENTS PRODUCTS IN COMBINATION WITH OTHER PRODUCTS IN A MANNER NOT

EVALUATED OR CONTEMPLATED BY NATIONAL INSTRUMENTS, THE USER OR APPLICATION DESIGNER IS

ULTIMATELY RESPONSIBLE FOR VERIFYING AND VALIDATING THE SUITABILITY OF NATIONAL

INSTRUMENTS PRODUCTS WHENEVER NATIONAL INSTRUMENTS PRODUCTS ARE INCORPORATED IN A

SYSTEM OR APPLICATION, INCLUDING, WITHOUT LIMITATION, THE APPROPRIATE DESIGN, PROCESS AND

SAFETY LEVEL OF SUCH SYSTEM OR APPLICATION.

Compliance

FCC/Canada Radio Frequency Interference Compliance*

Determining FCC Class

The Federal Communications Commission (FCC) has rules to protect wireless communications from interference.

The FCC places digital electronics into two classes. These classes are known as Class A (for use in industrialcommercial locations only) or Class B (for use in residential or commercial locations). Depending on where it is operated, this product could be subject to restrictions in the FCC rules. (In Canada, the Department of

Communications (DOC), of Industry Canada, regulates wireless interference in much the same way.)

Digital electronics emit weak signals during normal operation that can affect radio, television, or other wireless products. By examining the product you purchased, you can determine the FCC Class and therefore which of the two

FCC/DOC Warnings apply in the following sections. (Some products may not be labeled at all for FCC; if so, the reader should then assume these are Class A devices.)

FCC Class A products only display a simple warning statement of one paragraph in length regarding interference and undesired operation. Most of our products are FCC Class A. The FCC rules have restrictions regarding the locations where FCC Class A products can be operated.

FCC Class B products display either a FCC ID code, starting with the letters EXN, or the FCC Class B compliance mark that appears as shown here on the right.

Consult the FCC web site http://www.fcc.gov

for more information.

FCC/DOC Warnings

This equipment generates and uses radio frequency energy and, if not installed and used in strict accordance with the instructions in this manual and the CE Mark Declaration of Conformity**, may cause interference to radio and television reception. Classification requirements are the same for the Federal Communications Commission (FCC) and the Canadian Department of Communications (DOC).

Changes or modifications not expressly approved by National Instruments could void the user’s authority to operate the equipment under the FCC Rules.

Class A

Federal Communications Commission

This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.

Canadian Department of Communications

This Class A digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations.

Cet appareil numérique de la classe A respecte toutes les exigences du Règlement sur le matériel brouilleur du

Canada.

Class B

Federal Communications Commission

This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:

• Reorient or relocate the receiving antenna.

• Increase the separation between the equipment and receiver.

• Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.

• Consult the dealer or an experienced radio/TV technician for help.

Canadian Department of Communications

This Class B digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations.

Cet appareil numérique de la classe B respecte toutes les exigences du Règlement sur le matériel brouilleur du

Canada.

European Union - Compliance to EEC Directives

Readers in the EU/EEC/EEA must refer to the Manufacturer's Declaration of Conformity (DoC) for information** pertaining to the CE Mark compliance scheme. The Manufacturer includes a DoC for most every hardware product except for those bought for OEMs, if also available from an original manufacturer that also markets in the EU, or where compliance is not required as for electrically benign apparatus or cables.

* Certain exemptions may apply in the USA, see FCC Rules

§

15.103 Exempted devices, and

§

15.105(c).

Also available in sections of CFR 47.

** The CE Mark Declaration of Conformity will contain important supplementary information and instructions for the user or installer.

»

Conventions

bold

bold italic

italic

NI 4060

The following conventions are used in this manual:

The » symbol leads you through nested menu items and dialog box options to a final action. The sequence File»Page Setup»Options»Substitute

Fonts directs you to pull down the File menu, select the Page Setup item,

select Options, and finally select the Substitute Fonts options from the last dialog box.

This icon to the left of bold italicized text denotes a note, which alerts you to important information.

This icon to the left of bold italicized text denotes a caution, which advises you of precautions to take to avoid injury, data loss, or a system crash.

This icon to the left of bold italicized text denotes a warning, which advises you of precautions to take to avoid being electrically shocked.

Bold text denotes the names of menus, menu items, parameters, dialog boxes, dialog box buttons or options, icons, windows, Windows 95 tabs, or LEDs.

Bold italic text denotes a note, caution, or warning.

Italic text denotes variables, emphasis, a cross reference, or an introduction to a key concept. This font also denotes text from which you supply the appropriate word or value, as in Windows 3.x.

Refers to either of the devices in the NI 4060 family—NI 4060 for PCI or

NI 4060 for PXI.

Contents

Chapter 1

Taking Measurements with the NI 4060

Connecting Signals ........................................................................................................1-1

Cable and Probes ...........................................................................................................1-2

Introduction to the VirtualBench-DMM Soft Front Panel.............................................1-3

Use the Soft Front Panel ................................................................................................1-5

Measure DC and AC Voltage..........................................................................1-6

Measure 2-Wire Resistance .............................................................................1-7

Measure 4-Wire Resistance .............................................................................1-7

Measure DC and AC Current ..........................................................................1-9

Measure the Voltage Drop Across a Diode .....................................................1-10

Measure Temperature ......................................................................................1-11

Chapter 2

NI 4060 Operation

Safety Instructions .........................................................................................................2-1

Measurement Fundamentals ..........................................................................................2-2

Warm Up .........................................................................................................2-2

Selecting the Resolution ..................................................................................2-2

Autozeroing .....................................................................................................2-2

Grounding........................................................................................................2-3

Voltage Measurements ..................................................................................................2-3

DC Voltage ......................................................................................................2-3

Input Ranges .....................................................................................2-3

Measurement Considerations ............................................................2-3

Input Impedance .................................................................2-3

Thermal EMF .....................................................................2-4

Noise Rejection...................................................................2-4

AC Voltage ......................................................................................................2-6

Input Ranges .....................................................................................2-7

Measurement Considerations ............................................................2-7

AC Offset Voltage ..............................................................2-7

Frequency Response ...........................................................2-7

Resistance Measurements ..............................................................................................2-8

2-Wire Resistance Measurements ...................................................................2-8

Input Ranges .....................................................................................2-8

Continuity Measurements................................................................................2-9

4-Wire Resistance Measurements ...................................................................2-9

© National Instruments Corporation

ix

NI 4060 User Manual

Contents

Input Ranges ..................................................................................... 2-10

Current Measurements................................................................................................... 2-10

DC Current ...................................................................................................... 2-10

Input Ranges ..................................................................................... 2-11

AC Current ...................................................................................................... 2-11

Input Ranges ..................................................................................... 2-11

Measurement Considerations ........................................................... 2-12

Burden Voltage .................................................................. 2-12

Resistor Heating ................................................................. 2-12

Errors in AC Current Measurements.................................. 2-12

Diode Measurements....................................................................................... 2-13

Scanning ........................................................................................................................ 2-13

AUX I/O Connector and Optional Trigger Cables ......................................... 2-13

Scanning Communication Methods ................................................................ 2-14

Handshaking Mode........................................................................... 2-14

Synchronous Mode ........................................................................... 2-16

Measurement Considerations ........................................................... 2-18

Settling Time ...................................................................... 2-18

Mode and Range Changes.................................................. 2-18

Relay Life........................................................................... 2-18

Appendix A

Specifications

Appendix B

Fuse Replacement

Appendix C

Technical Support Resources

Glossary

Index

Figures

Figure 1-1.

Connectors on the NI 4060 Family of Products.................................... 1-2

Figure 1-2.

NI-DMM Soft Front Panel.................................................................... 1-3

Figure 1-3.

Digits of Precision................................................................................. 1-5

Figure 1-4.

Connecting Signals for Voltage Measurements.................................... 1-6

Figure 1-5.

Connecting Signals for 2-Wire Resistance ........................................... 1-7

NI 4060 User Manual

x www.ni.com

Contents

Figure 1-6.

Connecting Signals for 4-Wire Resistance............................................1-8

Figure 1-7.

Connecting Signals for Current Measurement ......................................1-9

Figure 1-8.

Connecting Signals for Diode Test .......................................................1-10

Figure 1-9.

Connecting Signals for RTDs and Thermistors. ...................................1-11

Figure 2-1.

Effect of Input Impedance on Signal Measurements ............................2-4

Figure 2-2.

Normal Mode Measurement Effects .....................................................2-5

Figure 2-3.

Common Mode Measurement Effects...................................................2-6

Figure 2-4.

Circuit for 2-Wire Resistance Measurements .......................................2-8

Figure 2-5.

Circuit for 4-Wire Measurements..........................................................2-10

Figure 2-6.

Circuit for Current Measurements.........................................................2-11

Figure 2-7.

Effect of Burden Voltage on Current Measurements ............................2-12

Figure 2-8.

Circuit for Diode Measurements ...........................................................2-13

Figure 2-9.

Auxiliary Trigger Cable (SH9MD-2BNC)............................................2-14

Figure 2-10.

Hardware Connections to an External Multiplexer in

Handshaking Mode................................................................................2-15

Figure 2-11.

Handshaking Mode Trigger Timing......................................................2-16

Figure 2-12.

Hardware Connections to an External Multiplexer in

Synchronous Mode................................................................................2-17

Figure 2-13.

Synchronous Mode Trigger Timing ......................................................2-18

Figure B-1.

Insulator Screw Location on NI 4060 (PCI) .........................................B-1

Figure B-2.

Fuseholder Location on NI 4060 (PCI) .................................................B-2

Figure B-3.

Fuseholder Location on NI 4060 (PXI).................................................B-3

© National Instruments Corporation

xi

NI 4060 User Manual

1

Taking Measurements with the NI 4060

Thank you for buying a National Instruments 4060 digital multimeter. A system based on the NI 4060 offers the flexibility, performance, and size that makes it ideal for service, repair, and manufacturing as well as for use in industrial and laboratory environments. The NI 4060, used in conjunction with your computer, is a versatile, cost-effective platform for high-resolution measurements.

For the most current versions of manuals and example programs, visit www.ni.com/instruments

for free downloads.

Detailed specifications for the NI 4060 are in Appendix A,

Specifications

.

Note

Before using any measurement equipment, it is important that you thoroughly

understand the safety instructions for that product. The beginning of Chapter 2,

NI 4060

Operation

, covers the safety guidelines for your NI 4060.

Connecting Signals

Figure 1-1 shows the front panels for the NI 4060 device. These front

panels contain five connectors—four banana jacks and one 9-pin mini circular DIN connector.

The four banana jack connectors are high-voltage, safety signal connectors.

The 9-pin mini circular DIN connector labeled AUX I/O is a digital signal connector, which carries TTL-level triggering signals for use with external

scanning equipment. Scanning is discussed in Chapter 2,

NI 4060

Operation

.

NI 4060 User Manual

© National Instruments Corporation 1-1

Chapter 1 Taking Measurements with the NI 4060

PXI

I

N

N

A

S

T

T

I

R

O

U

N

M

A

E

L

N TS

PCI

IN

250V

Max

HI

V

Input

250V

Max

LO

!

HI mA /

4W

Sense

200mA

Max

250V

Max

LO

AUX I/O

5V

Max mA/

250V

Max

AUX I/O

Figure 1-1. Connectors on the NI 4060 Family of Products

Cable and Probes

The NI 4060 kit contains a pair of test probes with safety banana plugs.

These probes meet international safety requirements, including

UL 3111 and IEC 1010-1, for the full range of applications supported by the NI 4060. Before using any probes or accessories not supplied by

National Instruments, ensure that they meet applicable safety requirements for the signal levels you may encounter.

To use the NI 4060, connect the test probes to the NI 4060 via safety banana jacks. The shrouds around the banana jacks prevent you from contacting potentially hazardous voltages connected to the test probes. You can also connect the cable to standard, unshrouded banana jack probes or

NI 4060 User Manual

1-2 www.ni.com

Chapter 1 Taking Measurements with the NI 4060

accessories; however, use unshrouded probes or accessories only when the voltages are less than 30 V rms and 42 V pk

, or 60 VDC.

Caution

To prevent possible safety hazards, the maximum voltage between either of the inputs and the ground of the computer should never exceed ±300 VDC or 300 V rms

. The maximum current that can be measured between the current inputs is ±200 mA DC or

200 mA rms

.

Introduction to the VirtualBench-DMM Soft Front Panel

The following sections explain how to make connections to your NI 4060 and take simple measurements using the VirtualBench-DMM, as shown in

Figure 1-2. To launch the soft front panel, select Start»Programs»

National Instruments DMM»Soft Front Panel.

Figure 1-2. NI-DMM Soft Front Panel

The following text describes the options available on the soft front panel.

Refer to Help»Online Reference located in the soft front panel help menu for information on front panel menus.

The range selector determines the range of measurements

VirtualBench-DMM makes. The range differs for each measurement mode.

If the measurement exceeds the range, +OVER or –OVER appears in the measurement display. Auto selects the range that best matches the input signal.

The value indicator displays the value measured by your NI 4060

(The value shown is an example only.).

© National Instruments Corporation

1-3

NI 4060 User Manual

Chapter 1 Taking Measurements with the NI 4060

The unit indicator displays the measurement units of the value you are measuring. The units are expressed as VAC, VDC, mVAC, mVDC,

, k

,

M

, mA, AC, or mA DC. The indicator also displays the digits of resolution, and by clicking on the indicator, you can select the resolution.

The Function selector allows you select a measurement mode. Select

Edit»Settings and click on the tabs for Current and Resistance or

Temperature to control the data type acquired by VirtualBench-DMM.

DC volts measures the DC component of a voltage signal.

AC volts measures the AC component of a voltage signal.

DC current measures the DC component of a current source.

AC current measures the AC component of a current source.

2-wire measures resistance using the 2-wire method.

4-wire measures resistance using the 4-wire method.

Diode measures the voltage drop across a diode. The maximum voltage

VirtualBench-DMM measures is 2 V.

Temperature measures temperature.

The run button starts and stops continuous DMM measurements.

The single button performs a single measurement.

The math buttons allow you to manipulate readings mathematically.

Null starts relative mode. VirtualBench-DMM makes all subsequent

measurements relative to the measurement it makes when you click on

Null.

Max/Min displays the maximum and minimum values that occur after you start Relative mode.

mX+B enables the mX+B calculation on all readings.

NI 4060 User Manual

1-4 www.ni.com

Chapter 1 Taking Measurements with the NI 4060

dB compresses a large range of measurements into a much smaller range

by expressing DC or AC voltage in decibels.

dBm shows decibels above or below a 1 mW reference.

% selects the percentage calculation. VirtualBench-DMM expresses the displayed reading as a percent deviation from the reference value entered in the Math Settings. Refer to Help»Online Reference, Math Settings topic for more information about dB, dBm, mX+B, and percentage calculations.

The log button enables data logging. To configure the datalog file and log interval, select Edit»Settings. Refer to Help»Online Reference, Logging

Measurements to Disk topic for more details.

Digits of Precision—A pop-up ring control in the DMM front panel display allows you to set measurement accuracy to 3 1/2, 4 1/2, or 5 1/2. A larger value gives greater precision but slower measurement performance. Refer

to Figure 1-3.

Figure 1-3. Digits of Precision

Use the Soft Front Panel

The following sections describe procedures for measuring DC and AC voltage; 2-wire and 4-wire resistance; DC and AC current; diode; and

temperature using the soft front panel. Figures 1-4 through 1-9 depict the

NI 4060 for PXI; the same connections apply for all NI 4060 devices.

© National Instruments Corporation

1-5

NI 4060 User Manual

Chapter 1 Taking Measurements with the NI 4060

Measure DC and AC Voltage

This section gives the procedures for measuring DC and AC voltage using the soft front panel.

1.

Connect your signal to the top terminal pair, as shown in Figure 1-4.

For DC voltages, the HI (red) input terminal is positive and the LO

(black) input terminal is negative. For AC voltages, positive and negative terms are irrelevant.

I

N

N

A

S

T

T

I

R

O

U

N A

M E

L

N TS

NI 4060 User Manual

HI

V

Input

250V

Max

LO

!

HI mA /

4W

Sense

200mA

Max

250V

Max

LO

AUX I/O

5V

Max

+

V

SIGNAL

NI 4060 (PXI)

Figure 1-4. Connecting Signals for Voltage Measurements

2.

Select the mode you will measure:

• DC Volts

• AC Volts

3.

Select the range for your measurement or autoranging:

• DC Volts—± 20 mV, ± 200 mV, ± 2 V, ± 25 V, and ± 250 V

• AC Volts—20 mV rms

, 200 mV rms

, 2 V rms

, 25 V rms

, and 250 V rms

The value indicator displays the voltage measured.

1-6 www.ni.com

Chapter 1 Taking Measurements with the NI 4060

Measure 2-Wire Resistance

This section gives the procedure for measuring 2-wire resistance using the soft front panel:

1.

Connect the signal to the top terminal pair, as shown in Figure 1-5.

I

N

N

A

S

T

T

I

R

O

U

N A

M E

L

N TS

HI

V

Input

250V

Max

LO

!

HI mA /

4W

Sense

200mA

Max

250V

Max

LO

AUX I/O

5V

Max

R

NI 4060 (PXI)

Figure 1-5. Connecting Signals for 2-Wire Resistance

2.

Select 2-wire resistance mode.

3.

Select the range for your measurement—200

, 2 k

, 20 k

, 200 k

,

2 M

, 200 M

, or autorange.

The value indicator indicates the resistance measured. See the

2-Wire

Resistance Measurements

section of Chapter 2,

NI 4060 Operation

, for

more information on 2-wire resistance measurements.

Measure 4-Wire Resistance

4-wire resistance measurements use both pairs of terminals. This configuration allows you to measure low resistances accurately by

© National Instruments Corporation

1-7

NI 4060 User Manual

Chapter 1 Taking Measurements with the NI 4060

eliminating the effects of lead resistance. This section gives the procedures for measuring 4-wire resistance using the soft front panel.

1.

Connect the top signal pair to the top pair of banana jack connectors.

2.

Connect the bottom signal pair to the bottom pair of banana jack connectors.

These connections are shown in Figure 1-6.

I

N

N

A

S

T

T

I

R

O

U

N

M

A

E

L

N TS

HI

V

Input

250V

Max

LO

!

HI mA /

4W

Sense

200mA

Max

250V

Max

LO

R

AUX I/O

5V

Max

NI 4060 (PXI)

Figure 1-6. Connecting Signals for 4-Wire Resistance

3.

Select 4-wire resistance mode.

4.

Select the range for your measurement—200

, 2 k

, 20 k

, 200 k

,

2 M, or autorange.

The value indicator indicates the resistance measured. See the

4-Wire

Resistance Measurements

section of Chapter 2,

NI 4060 Operation

, for

more information on 4-wire resistance measurements.

NI 4060 User Manual

1-8 www.ni.com

Chapter 1 Taking Measurements with the NI 4060

Measure DC and AC Current

This section gives the procedures for measuring DC and AC current using the soft front panel.

1.

Connect the signal to the bottom terminal pair, as shown in Figure 1-7.

The HI (red) input terminal is positive, and the LO (black) input terminal is negative.

I

N

N

A

S

T

T

I

R

O

U

N A

M E

L

N TS

HI

V

Input

250V

Max

LO

!

HI mA /

4W

Sense

200mA

Max

250V

Max

LO

I

SIGNAL

AUX I/O

5V

Max

NI 4060 (PXI)

Figure 1-7. Connecting Signals for Current Measurement

2.

Select the mode you will measure:

• Amps DC

• Amps AC

3.

Select the range for your measurement or autoranging. The NI 4060 has two input ranges available for measuring DC current: ± 20 mA and

± 200 mA. The NI 4060 also has two input ranges available for measuring AC current: 20 mA rms

and 200 mA rms

.

For 10 A range (with external shunt, CSM-10A), the DMM must operate in voltage mode to be able to read the voltage drop across the shunt.

© National Instruments Corporation

1-9

NI 4060 User Manual

Chapter 1 Taking Measurements with the NI 4060

The value indicator indicates the current measured.

Measure the Voltage Drop Across a Diode

The NI 4060 can excite a device under test and read the resulting voltage drop. Diode mode is useful for testing diodes. Voltage up to 2 V can be measured in this mode.

1.

Connect the signal to the top terminal pair, as shown in Figure 1-8.

I

N

N

A

S

T

T

I

R

O

U

N A

M E

L

N TS

NI 4060 User Manual

HI

V

Input

250V

Max

LO

!

HI mA

4W

/

Sense

200mA

Max

250V

Max

LO

+

V

DIODE

AUX I/O

5V

Max

NI 4060 (PXI)

Figure 1-8. Connecting Signals for Diode Test

2.

Select diode mode.

3.

Select the range for your measurement. Only the 2 V range is available for diode measurements.

The value indicator will indicate the voltage drop measured. If the display indicates 2.200 VDC, the diode is either reverse biased or defective. See the

Diode Measurements

section of Chapter 2,

NI 4060 Operation

, for more

information on diode measurements.

1-10 www.ni.com

Chapter 1 Taking Measurements with the NI 4060

Measure Temperature

You can measure temperature using common temperature transducers such as resistive temperature devices (RTD) and thermistors. You can measure

resistive transducers in the 2-wire resistance mode, as shown in Figure 1-9.

I

N

N

A

S

T

T

I

R

O

U

N

M

A

E

L

N TS

HI

V

Input

250V

Max

LO

!

HI mA /

4W

Sense

200mA

Max

250V

Max

LO

AUX I/O

5V

Max

NI 4060 (PXI)

Figure 1-9. Connecting Signals for RTDs and Thermistors.

Although the soft front panel does not support automatic conversion to temperature, you can include this mathematical operation in the program you will write in your Application Development Environment (ADE), such as LabVIEW or LabWindows/CVI, convert and scale the transducers’ value to temperature programmatically through software.

You can also measure the devices in 4-wire resistance mode to eliminate the

effects of lead resistance, as shown in Figure 1-6.

© National Instruments Corporation

1-11

NI 4060 User Manual

2

NI 4060 Operation

This chapter contains safety instructions; measurement fundamentals and considerations; and scanning information.

Safety Instructions

!

Caution

To avoid personal injury or damage to electronic equipment, observe the following cautions:

Do not operate this instrument in an explosive atmosphere or where there may be flammable gases or fumes.

Equipment described in this document must be used in an Installation Category II environment per IEC 664. This category requires local level supply mains-connected installation.

To prevent safety hazards, the maximum voltage between either of the inputs and the ground of the computer should never exceed ±300 VDC or 300 V rms

.

The NI 4060 must be used in a NRTL-listed chassis or personal computer.

Do not operate damaged equipment. The safety protection features built into this instrument can become impaired if the instrument becomes damaged in any way. If the instrument is damaged, do not use it until service-trained personnel can check its safety. If necessary, return the instrument to National Instruments for service and repair to ensure that its safety is not compromised.

Do not operate this instrument in a manner that contradicts the information specified in this document. Misuse of this instrument could result in a shock hazard.

Do not substitute parts or modify equipment. Because of the danger of introducing additional hazards, do not install unauthorized parts or modify the instrument. Return the instrument to National Instruments for service and repair to ensure that its safety is not compromised.

Connections that exceed any of the maximum signal ratings on the NI 4060 can create a

© National Instruments Corporation

2-1

NI 4060 User Manual

Chapter 2 NI 4060 Operation

shock or fire hazard or can damage any or all of the devices connected to the NI 4060.

National Instruments is not liable for any damages or injuries resulting from incorrect signal connections.

Clean the instrument and accessories by brushing off light dust with a soft, nonmetallic brush. Remove other contaminants with a stiff nonmetallic brush. The unit must be completely dry and free from contaminants before returning to service.

Measurement Fundamentals

Warm Up

The required warm-up time for the NI 4060 is 30 minutes. This warm-up time is important because measurements made with the NI 4060 multimeter can change with temperature. This change is called a thermal drift and is influenced by temperature coefficients. To minimize the effects of thermal drift and ensure the specified accuracies, take all measurements after the

NI 4060 has had a chance to fully warm up. Depending on your environment, the NI 4060 can operate significantly above ambient temperature. Therefore, measurements made immediately after powering up the system can differ significantly from measurements made after the system has fully warmed up. The NI 4060 temperature coefficient specifications are listed in the Accuracy sections in Appendix A,

Specifications.

Selecting the Resolution

The resolution on the NI 4060 multimeter is programmable. You can select from three different resolutions: 5 1/2 digits, 4 1/2 digits, or 3 1/2 digits.

These settings allow you to trade off speed for resolution. The 5 1/2 digit setting has the highest resolution and slowest reading rate, while the

3 1/2 digit setting gives you the least resolution and fastest reading rate.

Measurement mode and range affect the reading rate by requiring different conversion times to obtain a given resolution for the different modes and ranges.

Autozeroing

Autozeroing is a technique that removes the effect of temperature drift on the measurement accuracy. Autozeroing disconnects the external signals from the circuit, internally shorts the input of the NI 4060, and takes a measurement. Ideally, this value should be zero, but real components can cause an offset from zero to occur. By measuring the value with the input

NI 4060 User Manual

2-2 www.ni.com

Chapter 2 NI 4060 Operation

leads internally shorted, autozeroing automatically eliminates this error.

Autozeroing is not supported in AC voltage, AC current, and diode modes.

Autozeroing reduces the reading rate of your measurements by approximately 50%.

Grounding

When measuring ground-referenced signals, connect the ground referenced side of your signal to the IN HI + terminal for best performance.

Voltage Measurements

DC Voltage

Your NI 4060 multimeter uses a high-resolution delta sigma, A/D converter (ADC) to sample signals and convert them into a digital form.

The ADC is preceded by a series of gain and attenuation circuitry that allow both small and large signals to be measured using the same converter. The

NI 4060 uses a digital filter, which heavily rejects powerline frequencies

(50–60 Hz) and their harmonics, as well as high frequency noise.

Input Ranges

The NI 4060 has five input ranges available for measuring DC voltages.

These ranges are ±20 mV, ±200 mV, ±2.0 V, ±25V, and ±250 V. Each range has a 10% overrange. The 250 V range can overrange to 300 V. The

250 V and 25 V input ranges have a 1 M

Ω input impedance; the 2 V,

200 mV, and 20 mV ranges have an input impedance greater than 1 G

.

Take these values into consideration when measuring high-impedance sources. When the NI 4060 is powered off, the 250 V and 25 V input range have a 1 M

Ω input impedance and the 2 V, 200 mV, and 20 mV ranges have an input impedance of 100 k

Ω.

If you are taking measurements that require a high degree of accuracy, you should consider problems associated with input impedance, noise effects, and thermal electromotive forces (thermal EMFs). These effects are discussed in the

Measurement Considerations

section.

Measurement Considerations

Input Impedance

Figure 2-1 illustrates the input impedance of an NI 4060 and its effect on

the measurement of a circuit under test. If you know the source impedance

© National Instruments Corporation

2-3

NI 4060 User Manual

Chapter 2 NI 4060 Operation

of the circuit being tested, you can correct for the attenuation caused by the

NI 4060 in software. Since R in

is large, at least 1 M

, it will require a large source impedance, R s

, to cause a large change in the measured voltage, V m

.

Measured

Voltage

V m

+

Input

Impedance

R in

HI

External Source

Impedance R s

Input

V

+

Source

Voltage V s

LO

V m

=

V s

R

---------------in

-----------

R s

+ R in

Figure 2-1. Effect of Input Impedance on Signal Measurements

Thermal EMF

Thermal EMFs, or thermoelectric potentials, are voltages generated at the junctions of dissimilar metals and are functions of temperature. Thermal

EMFs in a circuit under test can cause higher than expected offsets that change with temperature.

Noise Rejection

The NI 4060 filters out AC voltages in the DC voltage measurement ranges. However, if the amplitudes of the AC voltages are large compared to the DC voltages, or if the peak value (AC + DC) of the measured voltage is outside the overrange limits, the NI 4060 may exhibit additional errors.

To minimize these errors, keep the NI 4060 away from strong AC magnetic sources and minimize the area of the loop formed by the test leads.

Choosing the 5 1/2 digit resolution will also help minimize noise from

AC sources. If the peak value of the measured voltage is likely to exceed the selected input range, select the next highest input range.

Normal Mode Rejection

Normal mode rejection (NMR) is the NI 4060’s ability to reject a normally

(differentially) applied signal. The ability is quantified in the normal mode rejection ratio (NMRR) specification, which indicates the capability of the

NI 4060 to reject 50 or 60 Hz and is valid only at the specified frequency and useful only when taking DC measurements. The NMRR is specified at

NI 4060 User Manual

2-4 www.ni.com

Chapter 2 NI 4060 Operation

the powerline frequency because this is typically where most measurement noise occurs.

Figure 2-2 shows a 60 Hz signal connected differentially to the NI 4060 in

DC Volts mode. V m is the voltage that will be measured after the signal is rejected. NMR is very useful when trying to measure DC voltages in the presence of large powerline interference.

Measured

Voltage

V m

+

HI

Input

V

LO

Source

Voltage

V s at 60 Hz

V m

=

V s

×

10

– N MRR

--------------------

20

Figure 2-2. Normal Mode Measurement Effects

If you are measuring signals in the presence of large normal mode voltages,

consult Appendix A,

Specifications

, to calculate the additional error to your

system. Use the equation in Figure 2-2 to calculate the voltage error due to

normal mode voltage.

Common Mode Rejection

Common mode rejection (CMR) is the NI 4060’s ability to reject signals that are common to both input terminals. The ability is quantified in the common mode rejection ratio (CMRR) specification. Theoretically, the floating measurement circuitry of the NI 4060 should have an infinite

CMRR. Parasitic resistances and capacitances to earth ground limit the

CMR of the NI 4060. This effect is most noticeable when measuring small signals in the presence of a large common mode voltage, as shown in

Figure 2-3.

© National Instruments Corporation

2-5

NI 4060 User Manual

Chapter 2 NI 4060 Operation

AC Voltage

NI 4060 User Manual

Measured

Voltage

V m

+

HI

Input

V

LO

+

Source

Voltage

V s

+

Common

Voltage

V c

V error

=

V

------

2 s

V

×

10

– CMRR

20

V m

= V s

+ V error

Figure 2-3. Common Mode Measurement Effects

Using the equation in Figure 2-3, you can calculate the voltage error due to

the common mode voltage. If you are measuring signals in the presence of

large common mode voltages, consult Appendix A,

Specifications

, to

calculate the additional error to your system.

Effective Common Mode Rejection

Effective common mode rejection is the sum of the CMRR and the NMRR at a given frequency. It is the effective rejection on a given noise signal that is applied to both input leads as it gets rejected first by the CMR capability of the instrument then again by its NMR capability. This specification is most useful at the powerline frequency where most of the noise resides and is only valid for DC measurements.

In the AC voltage ranges, the NI 4060 measures the AC-coupled RMS value of a signal. The RMS value of a signal is a fundamental measurement of the magnitude of an AC signal. The RMS value of an AC signal can be defined mathematically as the square root of the average of the square of the signal.

In practical terms, the RMS value of an AC signal is the DC value required to produce an equivalent amount of heat in the same resistive load. The

2-6 www.ni.com

Chapter 2 NI 4060 Operation

NI 4060 first AC-couples the measured signal to remove any DC components and then measures the RMS value of the AC component. This method lets you measure a small AC signal in the presence of a large DC offset.

Input Ranges

The NI 4060 has five input ranges available for measuring AC voltages.

These ranges are 20 mV rms

, 200 mV rms

, 2.0 V rms

, 25 V rms

, and 250 V rms

.

The impedance in each of these ranges is a 0.068 µF capacitor followed by

1 M

. When the NI 4060 is powered off, the 250 V, 25 V, and 2 V input ranges have a 0.068 µF capacitor, followed by a 1 M

Ω input impedance.

The 200 mV and 20 mV ranges have a 0.068 µF capacitor, followed by an approximate 100 k

Ω input impedance.

The NI 4060 can measure AC voltages to its specified accuracy as long as the voltage is at least 10% and no more than 100% of the selected input range. The DC component in any of these ranges can be as high as

250 VDC. Each range has a 10% overrange. The 250 VAC range can overrange up to 300 V.

The AC voltage measurement accuracy depends on many factors, including the signal amplitude, frequency, and waveform shape.

Measurement Considerations

AC Offset Voltage

The AC measurements of the NI 4060 are specified over the range of 10% to 100% of the full-scale input range. Below 10% of the input range, errors due to the AC voltage offset become significant. This offset, unlike DC voltage offsets, cannot simply be subtracted from the readings or zeroed out because the offset gets converted in the RMS conversion. To minimize the errors due to the AC offset voltage try to choose an input range that keeps the measured voltage between 10% and 100% of full scale.

Frequency Response

The accuracy of the NI 4060’s AC voltage measurements is a function of the input signal frequency. Your NI 4060 is calibrated at the factory using a 60 Hz sine wave. Your frequency-dependent error will be minimal around this frequency. The error will then increase as you approach the upper and lower bandwidth limits. This additional error is added to the accuracy errors in computing the absolute error.

© National Instruments Corporation

2-7

NI 4060 User Manual

Chapter 2 NI 4060 Operation

These additional errors are shown in Appendix A,

Specifications

. While the

NI 4060 is characterized and specified over the 20 Hz to 25 kHz frequency range, measurements outside of this range can still be made with decreased accuracy.

Resistance Measurements

2-Wire Resistance Measurements

The NI 4060 measures 2-wire resistance by passing a current through the device under test and reading the resulting voltage drop through the same

connections, as illustrated in Figure 2-4. The resistance value is then

computed using Ohm’s Law (R=V/I). To accurately measure the value of a resistor, make sure the resistor is not connected to any other circuits.

Erroneous or misleading readings can result if the resistor you are measuring is connected to external circuits that supply voltages or currents, or to external circuits that change the effective resistance of that resistor.

V sense

+

I ex

HI

Input

V

I ex

LO

I ex

R unknown

=

V sense

I ex

I ex

= 100

µ

A (200

, 2 k

, 20 k

ranges)

1

µ

A (200 k

, 2 M

, 200 M

ranges)

R unknown

Figure 2-4. Circuit for 2-Wire Resistance Measurements

Input Ranges

The NI 4060 has five basic input ranges for 2-wire resistance as well as an extended range. The basic ranges are 200

, 2.0 k

, 20 k

, 200 k

, and

2 M

. With the extended range, measurements up to at least 200 M

Ω are possible.

NI 4060 User Manual

2-8 www.ni.com

Chapter 2 NI 4060 Operation

In the extended ohms range, the NI 4060 adds a 1 M

resistor in parallel with the test resistor, and then calculates the value of the resistor being tested. The test current for the 200

, 2.0 k

, and 20 k

ranges is 100 µA.

The test current for the 200 k

, 2 M

, and 200 M

ranges is 1 µA.

Continuity Measurements

Many traditional multimeters can take continuity measurements, which test for the presence or absence of continuity between the two test probes.

These measurements are simply resistance measurements, where the resistance between the two probes is measured and compared to a set value.

You can perform continuity measurements on a circuit by setting the

NI 4060 to the 200

Ω range and comparing the measured value to some low resistance value, typically 10

. If the measured value is less than 10

, there is continuity between the test probes.

4-Wire Resistance Measurements

4-wire resistance measurements work using a similar principle as 2-wire resistance measurements. However, the 4-wire method is more accurate when measuring small resistances (< 100

).

4-wire resistance measurements separate the current excitation leads from the voltage sense leads. This method allows the test currents to flow through the device under test while allowing a separate path for the voltage

drop to be sensed as illustrated in Figure 2-5. The only drawback is that the

2-wire ohms method doubles the number of connections for a resistance measurement.

© National Instruments Corporation

2-9

NI 4060 User Manual

Chapter 2 NI 4060 Operation

I ex

V sense

+

HI

Input

V

LO

HI mA/

4W

Sense

I ex

LO

I ex

R unknown

R unknown

=

V sense

I ex

I ex

= 100

µ

A (200

, 2 k

, 20 k

ranges)

1

µ

A (200 k

, 2 M

ranges)

Figure 2-5. Circuit for 4-Wire Measurements

Figure 2-5 illustrates the different paths for the excitation current and the

sense voltage. The sense leads should be connected as close as possible to the device to be measured. Any extra lead length between the actual device to be measured and the sense leads will add error to the measurement.

Input Ranges

The NI 4060 has five input ranges for 4-wire resistance: 200

, 2.0 k

,

20 k

, 200 k

, and 2 M

. Extended ohms range is not supported in the

4-wire configuration.

Current Measurements

DC Current

The NI 4060 uses the same input circuitry to measure current as it does DC voltage, with the exception that it switches in a 1

current shunt in parallel

with the input. Figure 2-6 illustrates how DC current measurements are

made.

NI 4060 User Manual

2-10 www.ni.com

AC Current

Chapter 2 NI 4060 Operation

The current shunt is protected by a 500 mA/250 V fast-acting fuse. Refer

to Appendix B,

Fuse Replacement

, for instructions on replacing this fuse.

V sense

+

Fuse

R shunt

= 1

HI mA/ 4W

Sense

LO

I signal

V sense

I signal

=

R shunt

R shunt

= 1

Figure 2-6. Circuit for Current Measurements

Input Ranges

The NI 4060 has two input ranges for DC current: ± 200 mA and ± 20 mA.

Overranging allows you to measure an additional 10% of your current range. You can measure larger currents using an optional external shunt accessory such as the CSM-10A current shunt, available from National

Instruments, or a third-party shunt. With the CSM-10A current shunt, you can measure currents up to 10 A.

To take measurements using an external shunt, set the voltmeter to

DC voltage. Using the value of the resistor, the resulting voltage measured can be scaled and converted to current in software.

The NI 4060 uses the same input circuitry to measure AC current as it does

AC voltage, with the exception that it switches in the same 1

current shunt used in the DC current mode in parallel with the input. The current shunt is protected by a 500 mA/250 V fast-acting fuse. Refer to

Appendix B,

Fuse Replacement

, for instructions on replacing this fuse.

Input Ranges

The NI 4060 has two input ranges for AC current: 200 mA rms

and

20 mA rms

. Overranging allows you to measure an additional 10% of your current range.

© National Instruments Corporation

2-11

NI 4060 User Manual

Chapter 2 NI 4060 Operation

You can measure larger currents using an optional external shunt accessory such as the CSM-10A current shunt available from National Instruments, or a third-party shunt. You can take measurements using an external shunt by setting the voltmeter to AC voltage. Using the value of the resistor, you can scale and convert the resulting voltage measured to current in software.

Measurement Considerations

Burden Voltage

To measure current, the current must pass through a resistance. While the shunt resistance is where the actual current is being sensed, the current will experience other voltage drops such as the lead resistance and the fuse resistance. These voltage drops add up to the burden voltage, as shown

in Figure 2-7.

R lead1

R fuse

+

V burden

R shunt

I signal

R lead2

V burden

= Isignal • (R shunt

+ R lead1

+ R lead2

+ R fuse

)

Figure 2-7. Effect of Burden Voltage on Current Measurements

Resistor Heating

Larger current measurements can heat the current shunt resistor and cause it to drift. This should not present a problem because the maximum current range is relatively small compared with the shunt power rating and because the current shunt has an excellent temperature coefficient of resistance.

However, you should consider this when measuring larger currents over a long period of time.

Errors in AC Current Measurements

The additional errors that apply for AC voltage also apply for AC current.

Resistor heating and burden apply in the same manner as they apply in DC current measurements mentioned above. Burden voltage can actually be increased due to the inductance in the measurement path.

NI 4060 User Manual

2-12 www.ni.com

Chapter 2 NI 4060 Operation

Diode Measurements

To properly measure the forward voltage of a diode, make sure that the diode is not connected to any other circuits. The NI 4060 biases the diode with a current of 100

µ

A and measures the resulting voltage drop, as

illustrated in Figure 2-8. Diode measurements are made with a fixed range

of 2.0 V.

V sense

+

I ex

HI

Input

V

LO

I ex

+

V diode

I ex

V sense

= V diode

Figure 2-8. Circuit for Diode Measurements

Scanning

You can use the NI 4060 with National Instruments multiplexers as well as third-party multiplexers that use traditional voltmeter signaling.

Traditional voltmeter signaling is a handshake protocol. In this protocol, the voltmeter starts a measurement when a trigger pulse occurs on its

EXT TRIG IN line, and emits a digital pulse on the voltmeter complete

(VMC) line when it finishes taking a measurement.

AUX I/O Connector and Optional Trigger Cables

This 9-pin mini DIN connector is on the front panel of your NI 4060.

Warning

This connection is not isolated. It is not referenced to your measurement circuit.

It is referenced to the ground of your computer. The digital signals on this connector should not operate beyond -0.5 to 5.5 V of your computer ground. The trigger signals are

TTL-compatible.

Using the optional triggering cable available for purchase for the NI 4060, you can break out the trigger signals to two female BNC connectors, as

shown in Figure 2-9.

© National Instruments Corporation

2-13

NI 4060 User Manual

Chapter 2 NI 4060 Operation

Figure 2-9. Auxiliary Trigger Cable (SH9MD-2BNC)

Scanning Communication Methods

The NI 4060 controls scanning through two methods: handshaking and synchronous.

Handshaking Mode

In handshaking mode, the voltmeter emits a digital pulse on the VMC line when it finishes taking a measurement on one channel of the multiplexer.

This pulse causes the multiplexer to advance to the next channel. When the analog circuitry of the multiplexer has settled, the multiplexer emits a digital pulse on the EXT TRIG IN line. When the NI 4060 receives this

pulse, it takes another measurement. Figure 2-10 illustrates the hardware

configuration for scanning using this method.

NI 4060 User Manual

2-14 www.ni.com

Chapter 2 NI 4060 Operation

HI

IN

LO

NI 4060

External Scanner

HI

Out 0

LO

HI

Out 1

LO

HI

Sense

LO

Connect

Trigger Out

(Scanner Advanced)

EXT TRIG IN

AUX I/O

Connect

Voltmeter Complete

(VMC)

Trigger In

(Voltmeter Complete)

Figure 2-10. Hardware Connections to an External Multiplexer in Handshaking Mode

Figure 2-11 depicts the VMC waveform that is generated by the NI 4060

and shows the specifications for the EXT TRIG IN waveform. The measurement time is not a physical signal; it illustrates how the measurement can be delayed from the trigger.

© National Instruments Corporation

2-15

NI 4060 User Manual

Chapter 2 NI 4060 Operation

500 ns min.

25 ms max.

Trigger Delay

External Trigger Input (EXT TRIG IN) to NI 4060

Measurement Time

(Internal to NI 4060)

Measurement in Progress

3

µ s typ.

Voltmeter Complete (VMC) from NI 4060

Measurement

Complete

Figure 2-11. Handshaking Mode Trigger Timing

The polarities shown in Figure 2-11 are for illustration only and are

programmable via the instrument driver. The trigger delay time is also user-configurable via the instrument driver. Its defaults are set to allow the

NI 4060 to settle to the new input value after the external multiplexer changes channels.

Synchronous Mode

In synchronous mode, the voltmeter emits a digital pulse on the VMC line when it finishes taking a measurement. This pulse causes the multiplexer to advance to the next channel. After a programmable delay has occurred, the voltmeter takes another measurement without requiring an external trigger.

This delay allows the external multiplexer and NI 4060 to settle to the new channel. This method allows you to use multiplexers that do not have the

capability to trigger the NI 4060. Figure 2-12 illustrates the hardware

configuration for scanning using this method.

NI 4060 User Manual

2-16 www.ni.com

Chapter 2 NI 4060 Operation

HI

IN

LO

NI 4060

External Scanner

HI

Out 0

LO

HI

Out 2

LO

HI

Sense

LO

Not Used

Leave Unconnected

EXT TRIG IN

AUX I/O

Connect

Trigger In

Voltmeter Complete

(VMC)

Figure 2-12. Hardware Connections to an External Multiplexer in Synchronous Mode

Figure 2-13 depicts the VMC waveform that is generated by the NI 4060.

The measurement time is not a physical signal; it illustrates how the measurement can be delayed from the trigger. The polarities on the timing diagram are for illustration only; they are programmable via the instrument driver. The trigger delay time is also user-configurable via the instrument driver. Its defaults are set to allow the NI 4060 to settle to the new input value after the external multiplexer changes channels, because there is no way for the NI 4060 to know when the external multiplexer has settled. You can use the trigger delay to add additional delay to compensate for the multiplexer setting time.

© National Instruments Corporation

2-17

NI 4060 User Manual

Chapter 2 NI 4060 Operation

Trigger Delay

3

µ s typ.

Voltmeter Complete (VMC) from NI 4060

Measurement

Complete

Software Start

(Internal to NI 4060)

Measurement Time

(Internal to NI 4060)

Measurement in Progress

Figure 2-13. Synchronous Mode Trigger Timing

Measurement Considerations

Settling Time

Settling time is the time required for a measurement system to stabilize after an input change before that input can be measured accurately. Take care when trying to measure rapidly changing inputs. The NI 4060 settling time is dictated by the measurement range, cable properties, source impedance, and change in input level. The cable used should be short, have low dielectric absorption, and minimal capacitance—Teflon cable is recommended. Your source should have a low output impedance. Settling time becomes especially important in scanning systems. The scanner or multiplexer requires an additional settling time before the measurement can be taken. The NI 4060 software allows for a programmable delay between channels so that both the multimeter and the multiplexer can settle.

Mode and Range Changes

When you change the DMM mode or range, an additional default delay time is inserted before the first measurement. This delay time varies with the digit of resolution, range, and mode selected.

Relay Life

When you use the NI 4060 with an external multiplexer, take care in designing the signal groupings. The internal mode and range changes occur through electromechanical relays. As with any electromechanical device, the relays are subject to wear. A good practice is to group signals according

NI 4060 User Manual

2-18 www.ni.com

Chapter 2 NI 4060 Operation

to mode to limit the number of mode changes the NI 4060 must cycle through, thus extending the component’s life.

© National Instruments Corporation

2-19

NI 4060 User Manual

A

Specifications

This appendix lists the specifications of the NI 4060. These specifications are guaranteed between 15 and 35

°

C unless otherwise specified.

DC Voltage

Accuracy

(% of reading ± µV)

Range

250.000 V*

25.0000 V

2.00000 V

24 Hour

(25 °C ± 1 °C)

0.0032% ± 1.25 mV

0.0032% ± 1 mV

0.0029% ± 10 µV

90 Day

(25 °C ± 10 °C)

0.021% ± 1 mV

0.014% ± 10 µV

1 Year

(25 °C ± 10 °C)

0.021% ± 1.25 mV 0.024% ± 1.25 mV

0.024% ± 1 mV

0.017% ± 10 µV

Temperature Coefficient

(% of reading/ °C ± µV/ °C)

0.0017% ± 480 µV

0.0017% ± 480 µV

0.0009% ± 5 µV

200.000 mV

20.0000 mV

0.0029% ± 6 µV

0.0029% ± 6 µV

0.014% ± 6 µV

0.014% ± 6 µV

0.017% ± 6 µV

0.017% ± 6 µV

0.0009% ± 1 µV

0.0009% ± 1 µV

Accuracy numbers are for 5 1/2 digits with autozero on and include the effects of full-scale and zero-scale errors, temperature variation, linearity, and noise.

*The NI 4060 can overrange to 300 V.

Noise Rejection

NMRR

(10 Hz filter setting, 50/60 Hz powerline frequency ±1%)..................... 80 dB

DC ECMRR

(with a 1 k

Ω imbalance in HI lead)........ 140 dB

AC ECMR (RDC to 50/60 Hz)

(with a 1 k

Ω imbalance in HI lead)........ 150 dB

© National Instruments Corporation

A-1

NI 4060 User Manual

Appendix A Specifications

Input Characteristics

Input bias current ....................................1 nA max

Input resistance .......................................> 1 G

(2 V, 200 mV,

20 mV ranges);

1 M

(250 V, 25 V)

DC Current

Accuracy

(% of reading ± µA

)

Range

20.0000 mA

200.000 mA

10.000 A*

24 Hour

(25 °C ± 1 °C)

0.015% ± 10 µA

0.015% ± 10 µA

0.11% ± 1 mA

90 Day

(25 °C ± 10 °C)

0.039% ± 10 µA

0.039% ± 10 µA

0.035% ± 2 mA

1 Year

(25 °C ± 10 °C)

0.042% ± 10 µA

0.042% ± 10 µA

0.0035% ± 2 mA

Temperature Coefficient

(% of reading/ °C ± µA/ °C)

0.0035% ± 1 µA

0.0035% ± 1 µA

0.007% ± 0.1 mA

Accuracy numbers are for 5 1/2 digits with autozero on and include the effects of full-scale and zero-scale errors, temperature variation, linerarity, and noise.

*Requires 10 A shunt, CSM-10A.

Input Characteristics

Maximum input ......................................200 mA/250 V

Input protection.......................................Fuse F1 500 mA/250 V fast fusing

Shunt resistor ..........................................1

Burden voltage........................................< 400 mV at 200 mA DC

NI 4060 User Manual

A-2 www.ni.com

Appendix A Specifications

AC Voltage

Accuracy

(% of reading ± mV)

Range

250.000 V*

25.0000 V

2.00000 V

24 Hour

(25 °C ± 1 °C)

0.6% ± 250 mV

0.16% ± 30 mV

0.28% ± 3 mV

90 Day

(25 °C ± 10 °C)

0.62% ± 680 mV

0.18% ± 210 mV

0.30% ± 21 mV

1 Year

(25 °C ± 10 °C)

0.62% ± 680 mV

0.18% ± 210 mV

0.30% ± 21 mV

Temperature Coefficient

(% of reading/ °C ± mV/ °C)

0.007% ± 20 mV

0.007% ± 20 mV

0.019% ± 2 mV

200.000 mV

20.0000 mV

0.16% ± 0.22 mV

0.28% ± 100 µV

0.18% ± 1.20 mV

0.30% ± 170 µV

0.18% ± 1.20 mV

0.30% ± 170 µV

0.007% ± 0.110 mV

0.019% ± 12 µV

Accuracy numbers are for 5 1/2 digits and include the effects of full-scale and zero-scale errors, temperature variation, linerarity, and noise, applies for sine waves

10% of input range. Accuracy may be affected by source impedance, cable capacitances dielectric absorption, or slew rate.

*The NI 4060 can overrange to 300 V.

Noise Rejection

AC CMRR at 50/60 Hz

(with a 1 k

Ω imbalance in HI lead)........ > 80 dB

Input Characteristics

Input resistance ...................................... 1 M

Bandwidth .............................................. 20 Hz–25 kHz

© National Instruments Corporation

A-3

NI 4060 User Manual

Appendix A Specifications

Additional AC Errors

Frequency-dependent errors

Input Frequency

20–50 Hz

50–100 Hz

100 Hz–20 kHz

20–25 kHz

Additional Error (% of Full-Scale)

2.5%

0%

1%

2.5%

AC Current

Accuracy

(% of reading ± mA)

Range

200.000 mA

20.0000 mA

24 Hour

(25 °C ± 1 °C)

0.18% ± 0.22 mA

0.30% ± 100 µA

90 Day

(25 °C ± 10 °C)

0.20% ± 1.2 mA

0.32% ± 170 µA

1 Year

(25 °C ± 10 °C)

0.20% ± 1.2 mA

0.32% ± 170 µA

Temperature Coefficient

(% of reading/ °C ± mA/ °C)

0.009% ± 0.110 mA

0.022% ± 12 µA

10.000 A* 0.3% ± 22 mA 0.32% ± 120 mA 0.32% ±120 mA 0.026% ±11 mA

Accuracy numbers are for 5 1/2 digits and include the effects of full-scale and zero-scale errors, temperature variation, linerarity, and noise.

*Requires 10 A shunt, CSM-10A.

Input Characteristics

Maximum input ......................................200 mA/250 V

Input protection.......................................Fuse F1 500 mA/250 V fast fusing

Shunt resistor ..........................................1

Burden voltage........................................< 400 mV at 200 mA AC

NI 4060 User Manual

A-4 www.ni.com

Appendix A Specifications

Resistance

Accuracy

(% of reading ±

)

Range

24 Hour

(25 °C ± 1 °C)

0.1% ± 6 k

90 Day

(25 °C ± 10 °C)

0.1% ± 60 k

1 Year

(25 °C ± 10 °C)

0.1% ± 60 k

Temperature Coefficient

(% of reading/ °C ±

/ °C)

0.0072% ± 6 k

Extended Ohm

(> 2 M

)

2.00000 M

200.000 k

20.0000 k

2.00000 k

200.000

* 0.012% ± 9

0.012% ± 5

0.006% ± 0.09

0.006% ± 0.05

0.006% ± 0.05

0.077% ± 27

0.077% ± 22

0.024% ± 0.3

0.024% ± 0.2

0.024% ± 0.2

0.080% ± 27

0.080% ± 22

0.027% ± 0.2

0.027% ± 0.3

0.027% ± 0.2

0.0072% ± 2

0.0072% ± 2

0.0020% ± 0.02

0.0020% ± 0.02

0.0020% ± 0.02

Accuracy numbers are for the 4-wire resistance mode 5 1/2 digits with autozero on and include the effects of full-scale and zero-scale errors, temperature variation, linearity, and noise.

*With autozero on or while scanning, and when large resistance with capacitive loads is measured, additional delay time is required.

Measurement modes

Ohms ............................................... 2-wire Ohms or 4-wire Ohms

Extended ohms................................ 2-wire mode only

Maximum lead resistance....................... 10

Ω (200 Ω range),

1 k

(all other ranges)

Test current ............................................ 100 µA for 200

Ω, 2 k

Ω,

20 k

Ω ranges

1 µA for 2 M

Ω, 200 k

Ω ranges

1

µA and

1

M

Ω in parallel for extended Ohms mode

Additional error for 2-wire resistance .... 0.6

© National Instruments Corporation

A-5

NI 4060 User Manual

Appendix A Specifications

Diode Testing

Accuracy

(% of reading ± µV)

Range

24 Hour

(25 °C ± 1 °C)

90 Day

(25 °C ± 10 °C)

1 Year

(25 °C ± 10 °C)

Temperature Coefficient

(% of reading/ °C ± µV/ °C)

2 V 0.006% ± 7 µV 0.024% ± 22 µV 0.027% ± 22 µV 0.0020% ± 2 µV

Accuracy numbers are for 5 1/2 digits and include the effects of full-scale and zero-scale errors, temperature variation, linearity, and noise.

Test current .............................................100 µA

General Specifications

Settling time............................................Affected by source impedance and input signal changes

Warm-up time.........................................30 minutes for measurements accurate within specifications

Bus type

PCI...................................................Slave

PXI...................................................Slave

CompactPCI ...................................Slave

Altitude ...................................................Up to 2,000 m; at higher altitudes the installation category must be derated

Working voltage .....................................300 V maximum between either input terminal and earth ground

Power requirement..................................+5 VDC, 250 mA in operational mode

Safety ......................................................Designed in accordance with

IEC 61010-1 and UL 3111-1 for electrical and testing equipment;

Installation Category II, Pollution

Degree 2, Reinforced Insulation,

Indoor Use Only

NI 4060 User Manual

A-6 www.ni.com

Appendix A Specifications

Physical

Environment

Dimensions

PCI .................................................. 10.8 by 17.5 cm

(4.25 by 6.9 in.)

PXI .................................................. 10 by 16 cm

(3.9 by 6.33 in.)

Operating temperature............................ 0 to 55 °C

Storage temperature ............................... –20 to 70 °C

Relative humidity ................................... 10% to 90% noncondensing

© National Instruments Corporation

A-7

NI 4060 User Manual

B

Fuse Replacement

This appendix explains how to replace the fuse in your NI 4060. Replace the fuse with one of the fuse types listed in the chart below.

Caution

For continued protection against fire, replace only with fuses of the same type and rating. See the following chart for fuse types.

Fuse Rating

500 mA/250 V

500 mA/250 V

Fuse Type

Fast acting

Fast acting

Manufacturer

Schurter

LittelFuse

Part Number

FSF 034.1513

217.500

NI 4060 for PCI

1.

Remove all signal connections from your NI 4060. Power down your computer and remove the board.

2.

Turn to the back of the board and remove the four screws, as shown in

Figure B-1. These screws hold the top and bottom insulators onto the

board.

Remove

Remove

© National Instruments Corporation

Remove

Remove

Figure B-1. Insulator Screw Location on NI 4060 (PCI)

B-1

NI 4060 User Manual

Appendix B Fuse Replacement

3.

Remove the bottom and top insulators.

4.

Turn to the front of the board and locate the 5 by 20 mm glass fuse, as

shown in Figure B-2. Visually verify that the fuse is blown and

remove it.

F1-F 500 mA/250 V

Figure B-2. Fuseholder Location on NI 4060 (PCI)

5.

Press the new fuse into the silver holding fixture until you hear a snap.

6.

Reattach the top and bottom covers in the same order as you removed them.

Caution

Do not operate the NI 4060 without both insulators replaced and fastened.

NI 4060 for PXI

1.

Remove all signal connections from your NI 4060. Power down your computer and remove the board.

2.

Hold the board as shown in Figure B-3 and locate the fuse holder.

NI 4060 User Manual

B-2 www.ni.com

Appendix B Fuse Replacement

Fuse Holder

Figure B-3. Fuseholder Location on NI 4060 (PXI)

3.

Insert a screwdriver into the slot on the fuse holder.

4.

Turn counterclockwise.

5.

Pull out the holder and remove the blown fuse.

6.

Insert the 5 by 20 mm glass fuse into the holder and slide the holder in.

7.

Turn the fuse holder clockwise until it snaps shut.

© National Instruments Corporation

B-3

NI 4060 User Manual

C

Technical Support Resources

This appendix describes the comprehensive resources available to you in the Technical Support section of the National Instruments Web site and provides technical support telephone numbers for you to use if you have trouble connecting to our Web site or if you do not have internet access.

NI Web Support

To provide you with immediate answers and solutions 24 hours a day,

365 days a year, National Instruments maintains extensive online technical support resources. They are available to you at no cost, are updated daily, and can be found in the Technical Support section of our Web site at www.ni.com/support

.

Online Problem-Solving and Diagnostic Resources

KnowledgeBase—A searchable database containing thousands of

frequently asked questions (FAQs) and their corresponding answers or solutions, including special sections devoted to our newest products.

The database is updated daily in response to new customer experiences and feedback.

Troubleshooting Wizards—Step-by-step guides lead you through

common problems and answer questions about our entire product line.

Wizards include screen shots that illustrate the steps being described and provide detailed information ranging from simple getting started instructions to advanced topics.

Product Manuals —A comprehensive, searchable library of the latest

editions of National Instruments hardware and software product manuals.

Hardware Reference Database—A searchable database containing

brief hardware descriptions, mechanical drawings, and helpful images of jumper settings and connector pinouts.

Application Notes—A library with more than 100 short papers

addressing specific topics such as creating and calling DLLs, developing your own instrument driver software, and porting applications between platforms and operating systems.

© National Instruments Corporation

C-5

NI 4060 User Manual

Appendix C Technical Support Resources

Software-Related Resources

Instrument Driver Network —A library with hundreds of instrument

drivers for control of standalone instruments via GPIB, VXI, or serial interfaces. You also can submit a request for a particular instrument driver if it does not already appear in the library.

Example Programs Database—A database with numerous,

non-shipping example programs for National Instruments programming environments. You can use them to complement the example programs that are already included with National Instruments products.

Software Library—A library with updates and patches to application

software, links to the latest versions of driver software for National

Instruments hardware products, and utility routines.

Worldwide Support

National Instruments has offices located around the globe. Many branch offices maintain a Web site to provide information on local services. You can access these Web sites from www.ni.com/worldwide

If you have trouble connecting to our Web site, please contact your local

National Instruments office or the source from which you purchased your

National Instruments product(s) to obtain support.

For telephone support in the United States, dial 512 795 8248. For telephone support outside the United States, contact your local branch office:

Australia 03 9879 5166, Austria 0662 45 79 90 0, Belgium 02 757 00 20,

Brazil 011 284 5011, Canada (Calgary) 403 274 9391,

Canada (Ontario) 905 785 0085, Canada (Québec) 514 694 8521,

China 0755 3904939, Denmark 45 76 26 00, Finland 09 725 725 11,

France 01 48 14 24 24, Germany 089 741 31 30, Greece 30 1 42 96 427,

Hong Kong 2645 3186, India 91805275406, Israel 03 6120092,

Italy 02 413091, Japan 03 5472 2970, Korea 02 596 7456,

Mexico (D.F.) 5 280 7625, Mexico (Monterrey) 8 357 7695,

Netherlands 0348 433466, New Zealand 09 914 0488,

Norway 32 27 73 00, Poland 0 22 528 94 06, Portugal 351 1 726 9011,

Singapore 2265886, Spain 91 640 0085, Sweden 08 587 895 00,

Switzerland 056 200 51 51, Taiwan 02 2528 7227,

United Kingdom 01635 523545

NI 4060 User Manual

C-6 www.ni.com

Glossary

Prefix

pn-

µmk-

M-

G-

Meanings

pico nanomicromillikilomegagiga-

Numbers/Symbols

/

%

+

°

±

Ω percent positive of, or plus negative of, or minus per degree plus or minus ohm

Value

10 – 12

10 – 9

10

– 6

10 – 3

10

3

10 6

10

9

A

A

AC

AC coupled

A/D amperes alternating current the passing of a signal through a filter network that removes the

DC component of the signal analog-to-digital

© National Instruments Corporation

G-1

NI 4060 User Manual

B

b

B bus

Glossary

ADC

ADC resolution

ADE amplification amplitude flatness aperture time attenuate autozero analog-to-digital converter—an electronic device, often an integrated circuit, that converts an analog voltage to a digital number the resolution of the ADC, which is measured in bits. An ADC with16 bits has a higher resolution, and thus a higher degree of accuracy, than a 12-bit

ADC.

Application Development Environment a type of signal conditioning that improves accuracy in the resulting digitized signal and reduces noise a measure of how close to constant the gain of a circuit remains over a range of frequencies the period of time over which a measurement is averaged; also called the number of powerline cycles to reduce in magnitude technique of internally shorting the internal circuit while disconnecting the measurement to compensate for temperature effects bit—one binary digit, either 0 or 1 byte—eight related bits of data, an eight-bit binary number. Also used to denote the amount of memory required to store one byte of data.

the group of conductors that interconnect individual circuitry in a computer.

Typically, a bus is the expansion vehicle to which I/O or other devices are connected. Examples of PC buses are the PCI and ISA bus.

the voltage drop across the input section of the current mode burden voltage

C

C

CMRR

Celsius common-mode rejection ratio—a measure of an instrument’s ability to reject interference from a common-mode signal, usually expressed in decibels (dB)

NI 4060 User Manual

G-2

© National Instruments Corporation

CompactPCI conversion device conversion time coupling

CPU crest factor

CSM

D

DAQ

Glossary

refers to the core specification defined by the PCI Industrial Computer

Manufacturer’s Group (PICMG) device that transforms a signal from one form to another. For example, analog-to-digital converters (ADCs) for analog input, digital-to-analog converters (DACs) for analog output, digital input or output ports, and counter/timers are conversion devices.

the time required, in an analog input or output system, from the moment a channel is interrogated (such as with a read instruction) to the moment that accurate data is available the manner in which a signal is connected from one location to another central processing unit the ratio of the peak value of the signal to the RMS value of the signal current shunt module dB

DC default setting device data acquisition—(1) collecting and measuring electrical signals from sensors, transducers, and test probes or fixtures and inputting them to a computer for processing; (2) collecting and measuring the same kinds of electrical signals with A/D and/or DIO boards plugged into a computer, and possibly generating control signals with D/A and/or DIO boards in the same computer decibel—the unit for expressing a logarithmic measure of the ratio of two signal levels: dB=20log10 V1/V2, for signals in volts direct current a default parameter value recorded in the driver. In many cases, the default input of a control is a certain value (often 0) that means use the current

default setting. a plug-in data acquisition board, card, or pad that can contain multiple channels and conversion devices. Plug-in boards, PCMCIA cards, devices such as the DAQPad-1200, which connects to your computer parallel port, are all examples of DAQ devices.

© National Instruments Corporation

G-3

NI 4060 User Manual

Glossary

dielectric absorption differential input

DMM

DNL double insulated

DLL drivers

E

ECMR a parasitic phenomenon related to capacitors that can cause unexpectedly long settling times in circuits using capacitors with poor dielectric absorption specifications an analog input consisting of two terminals, both of which are isolated from computer ground, whose difference is measured digital multimeter differential nonlinearity—a measure in LSB of the worst-case deviation of code widths from their ideal value of 1 LSB a device that contains the necessary insulating structures to provide electric shock protection without the requirement of a safety ground connection dynamic-link library software that controls a specific hardware instrument

EEPROM

EXT TRIG IN

F

F filtering

Effective Common Mode Rejection—a measure of an instrument’s ability to reject interference from a common-mode signal. This includes both the effects of normal mode rejection and common mode rejection.

electrically erasable programmable read-only memory—ROM that can be erased with an electrical signal and reprogrammed external trigger input signal farads a type of signal conditioning that allows you to filter unwanted signals from the signal you are trying to measure

G

gain the factor by which a signal is amplified, sometimes expressed in decibels

NI 4060 User Manual

G-4

© National Instruments Corporation

Glossary

H

harmonics half-power bandwidth hardware

Hz multiples of the fundamental frequency of a signal the frequency range over which a circuit maintains a level of at least –3 dB with respect to the maximum level the physical components of a computer system, such as the circuit boards, plug-in boards, chassis, enclosures, peripherals, cables, and so on hertz—per second, as in cycles per second or samples per second

I

I ex

IEC

IEEE in.

inductance input bias current input impedance the relationship of induced voltage to current the current that flows into the inputs of a circuit

Installation Category

(Overvoltage Category) instrument driver the measured resistance and capacitance between the input terminals of a circuit classification system for expected transients on electrical supply installations a set of high-level software functions that controls a specific plug-in DAQ board. Instrument drivers are available in several forms, ranging from a function callable language to a virtual instrument (VI) in LabVIEW.

interrupt excitation current

International Electrotechnical Commission

Institute of Electrical and Electronics Engineers inches interrupt level

I/O a computer signal indicating that the CPU should suspend its current task to service a designated activity the relative priority at which a device can interrupt input/output—the transfer of data to/from a computer system involving communications channels, operator interface devices, and/or data acquisition and control interfaces

© National Instruments Corporation

G-5

NI 4060 User Manual

M

m

MB

N

NI-DAQ

NMRR noise

Glossary

ISA isolation isolation voltage industry standard architecture a type of signal conditioning in which you isolate the transducer signals from the computer for safety purposes. This protects you and your computer from large voltage spikes and makes sure the measurements from the DAQ device are not affected by differences in ground potentials.

the voltage that an isolated circuit can normally withstand, usually specified from input to input and/or from any input to the amplifier output, or to the computer bus

O

Ohm’s Law overrange meters megabytes of memory

National Instruments driver software for DAQ hardware normal mode rejection ratio—a measure of an instrument’s ability to reject a signal applied directly to the differential inputs of the instrument an undesirable electrical signal—Noise comes from external sources such as the AC power line, motors, generators, transformers, fluorescent lights, soldering irons, CRT displays, computers, electrical storms, welders, radio transmitters, and internal sources such as semiconductors, resistors, and capacitors. Noise corrupts signals you are trying to send or receive.

(R=V/I)—the relationship of voltage to current in a resistance a segment of the input range of an instrument outside of the normal measuring range. Measurements can still be made, usually with a degradation in specifications.

NI 4060 User Manual

G-6

© National Instruments Corporation

P

PCI peak value

PXI

R

R

RAM range error reading error reading rate resolution rms

ROM

R sense

Glossary

Peripheral Component Interconnect—a high-performance expansion bus architecture originally developed by Intel to replace ISA and EISA; it is achieving widespread acceptance as a standard for PCs and workstations and offers a theoretical maximum transfer rate of 132 Mbytes/s the absolute maximum or minimum amplitude of a signal (AC + DC)

PCI eXtensions for Instrumentation. PXI is an open specification that builds off the CompactPCI specification by adding instrumentation-specific features.

resistor random-access memory an error in accuracy that is determined by the input range that is selected.

The range error is independent of the value of the signal being measured.

an error in accuracy that is determined by the input range, as well as the value being measured the rate at which a new measurement is taken. In addition to the measurement speed, the selection of the reading rate affects the filtering, and thus the noise level, of measurements.

the smallest signal increment that can be detected by a measurement system. Resolution can be expressed in bits or in digits. The number of bits in a system is roughly equal to 3.3 times the number of digits.

root mean square—a measure of signal amplitude; the square root of the average value of the square of the instantaneous signal amplitude read-only memory the sense resistor. The voltage across this resistor is measured and converted to a current.

© National Instruments Corporation

G-7

NI 4060 User Manual

Glossary

S

s

S sense settling time

S/s system noise

T

temperature coefficient thermal drift thermoelectric potentials thermal EMFs transfer rate

U

UL seconds samples in four-wire resistance the sense measures the voltage across the resistor being excited by the excitation current the amount of time required for a voltage to reach its final value within specified limits samples per second—used to express the rate at which an instrument samples an analog signal a measure of the amount of noise seen by an analog circuit or an ADC when the analog inputs are grounded the percentage that a measurement will vary according to temperature. See

also thermal drift.

measurements that change as the temperature varies

See thermal EMFs.

thermal electromotive forces—voltages generated at the junctions of dissimilar metals that are functions of temperature. Also called thermoelectric potentials.

the rate, measured in bytes/s, at which data is moved from source to destination after software initialization and set up operations; the maximum rate at which the hardware can operate.

Underwriters Laboratory

NI 4060 User Manual

G-8

© National Instruments Corporation

Glossary

V

V

VAC

VDC

V error

VI

VMC

V rms

V sense

W

waveform shape working voltage volts volts alternating current volts direct current voltage error virtual instrument—(1) a combination of hardware and/or software elements, typically used with a PC, that has the functionality of a classic stand-alone instrument (2) a LabVIEW software module (VI), which consists of a front panel user interface and a block diagram program voltmeter complete signal volts, root mean square value the voltage that is created across the device under test when excited by a current the shape the magnitude of a signal creates over time the highest voltage that should be applied to a product in normal use, normally well under the breakdown voltage for safety margin

© National Instruments Corporation

G-9

NI 4060 User Manual

Index

A

AC current measurements, 2-12 to 2-13 burden voltage, 2-12 errors, 2-13 frequency dependent errors (table), A-3 input ranges, 2-12 measurement considerations, 2-12 to 2-13 resistor heating, 2-13 using Soft Front Panel, 1-9 to 1-10

AC current specifications, A-4

AC voltage measurement, 2-6 to 2-8

AC offset voltage, 2-7 frequency response, 2-8 input ranges, 2-7 measurement considerations, 2-7 to 2-8 using Soft Front Panel, 1-6

AC voltage specifications, A-3 autozeroing, 2-2 to 2-3

AUX I/O connector and optional trigger cables, 2-14 auxiliary trigger cable (SH9MD-2BNC)

(figure), 2-14

B

burden voltage, AC current measurement, 2-12

C

cables and probes, 1-2 to 1-3 common mode noise rejection, 2-5 to 2-6 communication methods, 2-15 to 2-19 handshaking mode, 2-15 to 2-16 synchronous mode, 2-17 to 2-18 connectors

AUX I/O connector and optional trigger cables, 2-14 front panel connectors, 1-1 to 1-2 continuity measurements, 2-9 conventions used in this manual, vi current measurements, 2-11 to 2-13

AC current, 1-9 to 1-10, 2-12 to 2-13

DC current, 1-9 to 1-10, 2-11 diode measurements, 1-10 to 1-11, 2-13

D

DC current measurements, 2-11 circuit (figure), 2-11 input ranges, 2-11 using Soft Front Panel, 1-9 to 1-10

DC current specifications, A-2

DC voltage measurement, 2-3 to 2-6 input impedance, 2-4 input ranges, 2-3 measurement considerations, 2-4 to 2-6 noise rejection, 2-4 to 2-6 common mode rejection, 2-5 to 2-6 effective common mode rejection, 2-6 normal mode rejection, 2-5 overview, 2-3 thermal EMFs, 2-4 using Soft Front Panel, 1-6

DC voltage specifications, A-1 to A-2 diagnostic resources, online, C-1 diode measurements accuracy specifications, A-6 circuitry (figure), 2-13

© National Instruments Corporation

I-1

NI 4060 User Manual

Index

connecting signals for diode test

(figure), 1-10 procedure, 2-13 voltage drop measurement, 1-10 to 1-11

E

environment specifications, A-7 errors

AC current measurements, 2-13 additional AC errors (table), A-3

F

four-wire resistance measurements,

2-9 to 2-10 circuit (figure), 2-10 input ranges, 2-10 using Soft Front Panel, 1-8 frequency response, AC voltage measurement, 2-8 front panel, soft. See VirtualBench-DMM Soft

Front Panel.

front panel connectors, 1-1 to 1-2 fuse replacement, B-1 to B-3 fuse types (table), B-1

NI 4060 (PCI), B-1 to B-2

NI 4060 (PXI), B-3

G

grounding, 2-3

H

handshaking mode, 2-15 to 2-16 hardware connections (figure), 2-15 trigger timing (figure), 2-16

NI 4060 User Manual

I-2

I

input impedance, DC voltage measurement, 2-4 input ranges

AC current measurements, 2-12 to 2-13

AC voltage measurement, 2-7

DC current measurements, 2-11

DC voltage measurement, 2-3 four-wire resistance measurements, 2-10 two-wire resistance measurements, 2-9

M

measurements current measurements, 2-11 to 2-13

AC current, 2-12 to 2-13

DC current, 2-11 diode measurements, 2-13 fundamentals autozeroing, 2-2 to 2-3 fundamentals, 2-2 to 2-3 grounding, 2-3 selecting resolution, 2-2 warm-up time for NI 4060, 2-2 resistance measurements, 2-8 to 2-10 continuity, 2-9 four-wire, 2-9 to 2-10 two-wire, 2-8 to 2-9 using VirtualBench-DMM Soft Front

Panel, 1-5 to 1-11

DC and AC current, 1-9 to 1-10

DC and AC voltage, 1-6 diode mode, 1-10 to 1-11 four-wire resistance, 1-8 temperature, 1-11 two-wire resistance, 1-7 voltage measurements, 2-3 to 2-8

AC voltage, 2-6 to 2-8

DC voltage, 2-3 to 2-6 mode changes, scanning effects, 2-19

www.ni.com

N

National Instruments Web support, C-1 to C-2

NI 4060. See also VirtualBench-DMM Soft

Front Panel.

cables and probes, 1-2 to 1-3 connecting signals, 1-1 to 1-2 current measurements, 2-11 to 2-13 fuse replacement, B-1 to B-3 measurement fundamentals, 2-2 to 2-3 overview, 1-1 resistance measurements, 2-8 to 2-10 safety instructions, 2-1 to 2-2 scanning, 2-14 to 2-19 specifications, A-1 to A-7 voltage measurements, 2-3 to 2-8 noise rejection

AC voltage specifications, A-3

DC voltage measurement, 2-4 to 2-6 common mode rejection, 2-5 to 2-6 effective common mode noise rejection, 2-6 normal mode rejection, 2-5

DC voltage specifications, A-1 normal mode noise rejection, 2-5

O

online problem-solving and diagnostic resources, C-1

P

physical specifications, A-7 probes and cables, 1-2 to 1-3 problem-solving and diagnostic resources, online, C-1

Index

R

range changes, scanning effects, 2-19 relay life, scanning effects, 2-19 resistance accuracy specifications, A-5 resistance measurements, 2-8 to 2-10 continuity, 2-9 four-wire resistance, 2-9 to 2-10 circuit (figure), 2-10 input ranges, 2-10 using Soft Front Panel, 1-8 two-wire resistance, 2-8 to 2-9 circuit (figure), 2-8 input ranges, 2-9 using Soft Front Panel, 1-7 resistor heating, AC current measurement, 2-13 resolution, selecting, 2-2

S

safety instructions, 2-1 to 2-2 scanning, 2-14 to 2-19

AUX I/O connector and optional trigger cables, 2-14 communication methods, 2-15 to 2-19 handshaking mode, 2-15 to 2-16 synchronous mode, 2-17 to 2-18 measurement considerations, 2-19 mode and range changes, 2-19 relay life, 2-19 settling time, 2-19 settling time, 2-19

Soft Front Panel. See VirtualBench-DMM

Soft Front Panel.

software-related resources, C-2 specifications, A-1 to A-7

AC current, A-4

AC voltage, A-3 additional AC errors (table), A-3

© National Instruments Corporation

I-3

NI 4060 User Manual

Index

DC current, A-2

DC voltage, A-1 to A-2 diode testing, A-6 environment, A-7 general, A-6 physical, A-7 resistance, A-5 synchronous mode, 2-17 to 2-18 hardware connections (figure), 2-17 trigger timing (figure), 2-18

T

technical support resources, C-1 to C-2 temperature, measuring, 1-11 thermal EMFs, DC voltage measurement, 2-4 trigger cable, optional, 2-14 two-wire resistance measurements, 2-8 to 2-9 circuit (figure), 2-8 input ranges, 2-9 using Soft Front Panel, 1-7

V

VirtualBench-DMM Soft Front Panel,

1-3 to 1-11 illustration, 1-3 options available, 1-3 to 1-5 using for measurements, 1-5 to 1-11

DC and AC current, 1-9 to 1-10

DC and AC voltage, 1-6 diode mode, 1-10 to 1-11 four-wire resistance, 1-8 temperature, 1-11 two-wire resistance, 1-7 voltage measurements, 2-3 to 2-8

AC voltage, 2-6 to 2-8

AC offset voltage, 2-7 frequency response, 2-8 input ranges, 2-7 measurement considerations,

2-7 to 2-8

DC voltage, 2-3 to 2-6 input impedance, 2-4 input ranges, 2-3 measurement considerations,

2-4 to 2-6 noise rejection, 2-4 to 2-6 common mode rejection,

2-5 to 2-6 effective common mode rejection, 2-6 normal mode rejection, 2-5 overview, 2-3 thermal EMFs, 2-4 using Soft Front Panel, 1-6

W

warm-up time for NI 4060, 2-2

Web support from National Instruments,

C-1 to C-2 online problem-solving and diagnostic resources, C-1 software-related resources, C-2

Worldwide technical support, C-2

NI 4060 User Manual

I-4 www.ni.com

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

advertisement

Table of contents