CAEN C207 Programmable Discriminator User's Manual

Below you will find brief information for Programmable Discriminator C207. This manual describes the C207, a 16-channel programmable discriminator that accepts 16 negative inputs and produces 16 differential ECL outputs. The module can be controlled either via CAMAC or locally, and each channel has its own discriminating threshold programmable via CAMAC. The device also features a current sum output that generates a current proportional to the input multiplicity, an OR output that provides a global OR of the shaped outputs, and a TEST input that triggers all enabled channels at once. This allows a complete test of the module without removing any input cable, and allows generation of a pattern of pulses suitable to test any following electronics.

PDF
Document
Programmable Discriminator C207 User's Manual | Manualzz

Advertisement

Advertisement

/

Advertisement

Key features

  • 16-channel programmable discriminator

  • Accepts 16 negative inputs

  • Produces 16 differential ECL outputs

  • Controllable via CAMAC or locally

  • Programmable threshold per channel

  • Current sum output

  • OR output

  • TEST input

Frequently asked questions

You can set independent thresholds for each channel by performing a F(16) N A(i) CAMAC Function, where subaddress i selects the i-th channel. The chosen value of the thresholds must be present on the CAMAC Write lines W1-W8.

You can enable or disable each of the 16 channels via CAMAC by performing a F(17) N CAMAC Function with the Write Lines W1-W16 set to 1 or 0 according to the chosen configuration (16 bit Pattern of Inhibit). A channel is enabled if the corresponding bit of the Pattern is high.

The two selectable ranges for the output pulse width are 4 ns - 40 ns and 40 ns - 300 ns. You can adjust the range for each channel by setting the corresponding jumpers (JP1 to JP16).

You can test the module's activity by performing a F(2) N CAMAC Function. If at least one of the unmasked channels is over threshold, this function will return a Q=1 value.

You can reset the module by performing a C, Z command or a F(9) N CAMAC Function. This will initialize the module and set all channels to enabled, and clear the Q response on the F(2) N Function.