Percent Weighing – Quick Reference Guide – AD Series
Percent Weighing (
Percent
)
Percent Weighing is used to compare the relative weight of an unknown item to that of a previously stored sample.
An
“E” indicator will appear on the screen to indicate the scale is in Percent Weighing mode.
1.
To enter Percent Weighing Mode, press the “F” key.
Select “ percent ” by pressing the “T” key.
PERCENT
2.
Place an empty weighing container on the pan.
The container’s weight will be displayed.
Wait for the weight to
stabilize and press the “T” key to tare.
100 .
000
------
3.
The scale will display the weight of the previously stored sample and the command “ SAMMPLE ” sequentially.
To accept the current sample weight, press the “T” (Yes) key.
To change the sample weight, press the “ Æ 0 Å ” key.
100 .
000 SAMMPLE
Note: If this is the first time the function is used, the sample weight will be
0.000g.
To change the sample weight, place a new sample in the weighing container when “ SAPL ON ” is displayed.
Once stabilized, press the “T” (Yes) key to accept the new weight.
Percent Weighing – Quick Reference Guide – AD Series
SAPL ON
100 .
000
Note : The reference sample must be equal to or greater than 100d.
Attempting to set a sample lower than 100d will result in a “ SAPL LO ” error.
4.
After setting the sample weight, the command “ READY ” will be
displayed.
Remove the sample and place the object on the pan.
READY
5.
The scale will express the weight of the object as a percentage of the stored sample.
50
59
Note : To change the initial sample press “CLR” and repeat Steps 1 to 3.
If the same sample weight is used, the scale can be tared without clearing the function and new objects can be weighed.
1.
To exit Percent Weighing, press the “C” key.