Kobold EPS Magnetic Inductive Flow Meter Operating instructions
Advertisement
Advertisement
Magnetic-inductive Flow meter
Installation and operation manual
EPS with
UMF2B
This operating manual contains important information for the operation
Please read the instructions carefully and store them in a safe place
Heinrichs Messtechnik GmbH Operating Manual EPS / UMF2B
November 2020
Heinrichs Messtechnik
Contents
Installation and Operating Manual EPS
1INTRODUCTION .................................................................................................................................... 4
1.1Hazard warnings ................................................................................................................................ 4
1.2Installation, commissioning and maintenance .............................................................................. 4
2IDENTIFICATION ................................................................................................................................... 5
2.1Supplier/manufacturer ...................................................................................................................... 5
2.2Product type ....................................................................................................................................... 5
2.3Application ......................................................................................................................................... 5
2.4Product name ..................................................................................................................................... 5
2.5
Issue date ........................................................................................................................................... 5
2.6Version no. ......................................................................................................................................... 5
2.7Designation/rating plate.................................................................................................................... 5
3AREA OF APPLICATION ...................................................................................................................... 6
4SAFETY WARNINGS ............................................................................................................................ 6
4.1Installation, mounting, commissioning and operating personnel ............................................... 6
4.2Intended use ...................................................................................................................................... 6
4.3Packaging / storage / transport ........................................................................................................ 6
4.4
Warranty ........................................................................................................................................... 7
4.5Return of the device for repairs or servicing ................................................................................. 7
5OPERATION AND SYSTEM STRUCTURE .......................................................................................... 7
5.1Mode of operation ............................................................................................................................. 7
5.2System configuration ........................................................................................................................ 8
5.2.1
Integrally mounted transmitter ........................................................................................... 8
5.2.2
Remote mounted transmitter ............................................................................................. 9
6PERFORMANCE CHARACTERISTICS ................................................................................................ 9
6.1Measuring accuracy .......................................................................................................................... 9
6.1.1
Measured error .................................................................................................................. 9
6.1.2
Repeatability ....................................................................................................................10
6.1.3
Reference conditions .......................................................................................................10
6.2Fluid conductivity ............................................................................................................................10
6.3Influence of ambient temperature ..................................................................................................10
6.4Influence of fluid temperature ........................................................................................................10
6.5
Materials .........................................................................................................................................10
6.5.1
Wetted parts ....................................................................................................................10
6.5.2
6.5.3
Non-wetted parts .............................................................................................................10
External power supply / electrical connections ...............................................................10
Page 2 of 38
Installation and Operating Manual EPS Heinrichs Messtechnik
7INSTALLATION/CONDITIONS FOR USE ..........................................................................................10
7.1Receipt of goods and transport .....................................................................................................10
7.1.1
Receipt of goods..............................................................................................................10
7.1.2
Transport .........................................................................................................................10
7.2Installation requirements ................................................................................................................11
7.2.1
7.2.2
Bypass pipes ...................................................................................................................11
Sensor tube lining ............................................................................................................11
7.3Installation ........................................................................................................................................11
7.3.1
Installation in pipelines with larger nominal sizes ............................................................12
7.3.2
7.3.3
7.3.4
7.3.5
7.3.6
Horizontal or vertical Installation .....................................................................................12
Installation examples .......................................................................................................12
Earthing – potential equalisation .....................................................................................15
Torques for screws and bolts ..........................................................................................17
Remote mount transmitter ...............................................................................................18
7.4
Wiring .........................................................................................................................................19
7.4.1
Mounted transmitter ........................................................................................................20
7.4.2
7.4.3
Transmitter mounted separately with a pigtail cable .......................................................20
Connection plan...............................................................................................................21
7.5Nominal size and ranges ................................................................................................................21
7.5.1
Flow - Diagram ................................................................................................................22
7.5.2
Flow specific values ........................................................................................................23
7.6Ambient conditions .........................................................................................................................24
7.6.1
Ambient temperature limits ..............................................................................................24
7.6.2
Storage temperature range .............................................................................................25
7.6.3
7.6.4
7.6.5
Climate category..............................................................................................................25
Ingress protection ............................................................................................................25
Shock and vibration resistance .......................................................................................25
7.7Process pressure ............................................................................................................................26
7.8Medium temperature and assignment of the temperature class ................................................27
8DIMENSIONS AND WEIGHT...............................................................................................................28
8.1Dimensional drawing EPS – DN 10 to DN 1200, flange design ...................................................28
8.2
Dimension DN 2; 3; 6; 10 incl. G ½“ (ISO) or ½“ NPT connection ..............................................29
8.3Dimension in DIN 11851, Tri-Clover® ............................................................................................29
8.4Transmitter UMF2(B) .......................................................................................................................30
8.4.1
Compact design...............................................................................................................30
8.4.2
Junction box of the sensor at remote design ..................................................................30
8.4.3
8.4.4
8.4.5
Wall mounting ..................................................................................................................30
Pipe mounting vertically ..................................................................................................31
Pipe mounting horizontally ..............................................................................................31
8.5Dimension grounding rings ...........................................................................................................31
9MAINTENANCE ...................................................................................................................................32
10AUXILIARY POWER / ELECTRICAL CONNECTION ......................................................................32
11CE TYPING ........................................................................................................................................32
12STANDARDS AND GUIDELINES, CERTIFICATES AND APPROVALS ........................................32
13DECLARATION OF DECONTAMINATION .......................................................................................33
14DECLARATION OF CONFORMITY ..................................................................................................34
15MODELCODE ....................................................................................................................................36
Page 3 of 38
Heinrichs Messtechnik
1 Introduction
Installation and Operating Manual EPS
It is essential that you read these operating instructions, prior to installation and commissioning of the device. The operating instructions must be read completely and fully understood.
If you do not have a copy of the latter instructions or any part is missing, please request one from Heinrichs Messtechnik GmbH or download the instructions from our website.
Custom designs and special applications are not be addressed in this manual.
All devices are thoroughly tested and checked for order compliance and functionality prior to shipping. If however you have any queries or problems concerning your purchased product, please contact our head office in Cologne.
Heinrichs Messtechnik GmbH accepts no liability for any loss or damage of any kind arising from improper operation of any product, improper handling or use of any replacement part, or from external electrical or mechanical effects, overvoltage or lightning. Any such improper operation, use or handling shall automatically invalidate the warranty for the product concerned.
1.1 Hazard warnings
The purpose of the hazard warnings listed below is to ensure the safety of the device operators and maintenance personnel, and that the flowmeter and any devices connected to it are not damaged.
The safety advisories and hazard warnings in this document are defined in the four categorised terms below, and are aimed to prevent putting operators and maintenance personnel at risk, or to avoid material damage. The used terms have, with respect to this document and the products described within, the following meanings:
Danger means that failure to take the prescribed precautions will result in death, severe bodily injury, or substantial material damage.
Warning means that failure to take the prescribed precautions could result in death, severe bodily injury, or substantial material damage.
Caution means that failure to take the prescribed precautions could result in incorrect operation, malfunction or damage to the device.
Note means that the accompanying text contains important information about the product, handling of the product or is about a section of the documentation that is of particular importance.
1.2 Installation, commissioning and maintenance
The devices described in this manual are to be installed and serviced only by qualified technical personnel such as a qualified Heinrichs Messtechnik electronics engineer or service technician.
Warning
To avoid danger of electric shock, service personnel should abide to the safety regulations pertaining to the use of electrical and automated technical devices and with the applicable laws and regulations in their own country before connecting this device to any periphery device or power supply.
Observe the technical data on the type plate and the safety instructions in the separate operating instructions of the associated converter!
Page 4 of 38
Installation and Operating Manual EPS
2 Identification
Heinrichs Messtechnik
2.1 Supplier/manufacturer
Heinrichs Messtechnik GmbH
Robert-Perthel-Str. 9
D-50739 Köln
Telephone: +49 221 49708 – 0, Fax: +49 221 49708 - 178
Internet: http://www.heinrichs.eu
, Email: mailto:[email protected]
2.2 Product type
Magneticinductive flowmeter based on Faraday’s law of induction
2.3 Application
Bidirectional measurement of liquids with a minimum conductivity of 5μS / cm
2.4 Product name
EPS
2.5 Issue date
11/20/2020
2.6 Version no.
File: EPS_BA_20.02_en.docx
2.7 Designation/rating plate
The rating plate states the following information:
Logo
Address
CE
Manufacturer’s logo
Manufacturer’s address (Internet address)
CE Marking in accordance with the applied EC Directives
Type
Code
Ser. No.
Tag No.
T amb
T m
C
DN
PN
PS
PED
Materials
MF-Date
Degrees of protection
Type designation
Code of the model
Serial number (for tracking reasons)
Operator’s measuring point number (if stated in the order)
Ambient temperature range
Medium temperature range
Sensor constant
Flange designation
Pressure stage of flange
Max. permissible process pressure
Information about the Pressure Equipment Directive 2014/68/EU.
- For devices with a process connection =< DN 25: o There is no CE Marking in accordance with Section. 4 para. 3 of the PED.
Under PED ( P ressure E quipment D irective) the reason for exception in accordance with Section 4 para. 3 of the PDE is stated. The device is rated as SEP ( S ound E ngineering P ractice).
- For devices with a process connection > DN 25: o CE Marking with the number of the indicated notified body that certified the device manufacture. o Category ( I , II , III or IV) in accordance with the PED; fluid group 1 comprises “dangerous fluids”.
Material of wetted parts such as pipe lining, material of electrodes and seal
Year of manufacture
Degrees of protection in accordance with DIN EN 60529:2014
Page 5 of 38
Heinrichs Messtechnik
3 Area of application
Installation and Operating Manual EPS
The electromagnetic flowmeter EPS is used to measure or monitor the volume flow of fluids with and without solids concentration, slurries, pastes and other electrically conductive media while minimizing pressure drops. The conductivity of the medium must be at least 5 µS/cm. Pressure, temperature, density and viscosity do not affect the volume measurements. Small quantities of solid particles and small gas pockets are also measured as part of the volume flow. A larger number of solid particles or gas pockets will result in failures.
4 Safety warnings
4.1 Installation, mounting, commissioning and operating personnel
Installation, mounting, commissioning and maintenance of safety relevant, are to be performed by a technician trained to work with this kind of devices, or by a Heinrichs
Messtechnik service technician.
It is a necessity that the qualified personal have read and understood this operating manual and follow its instructions.
The principle rules and regulations in the country of the operator must be observed.
The technical data on the rating plates as well as the safety warnings in the separate operating manual of the associated transmitter are to be observed!
4.2 Intended use
The electromagnetic flowmeter is to be used solely for measuring the volume flow of liquids, suspensions and pastes with a conductivity of
5 µS/cm (
20 µS/cm demineralized cold water). The responsibility for the use and installation of the flowmeter lies solely by the operator. Heinrichs Messtechnik will accept no liability for any damage resulting from its improper or unintended use. Damage caused by improper use will lead to loss of warranty.
Before using corrosive or abrasive fluids, the operator must test the resistance of all wetted materials. We will be happy to assist you in testing the corrosion resistance of wetted parts
(for special fluids including cleaning fluids). However, sole responsibility for ensuring that the device is used in accordance with the manufacturer’s recommendations rests with the system operator. Minor changes of temperature, concentration or the degree of contamination in the process may cause changes in corrosion resistance. Therefore the manufacturer accepts no responsibility for any damage with respect to corrosion resistance of wetted materials in a certain application.
4.3 Packaging / storage / transport
Be careful not to damage the device whilst removing it from its packaging. The device should be stored in a clean dry room until installation so as to prevent impurities from entering the device. Make certain that the ambient temperature in the room in which the device is stored lies within the prescribed range.
Upon receipt of the product, check the contents of the packaging and the product accessories against the information on the delivery slip and order form so as to ensure that all ordered components have been delivered.
If, after the unpacking of the device, it is to sent elsewhere to be installed, the original packaging and transport protection inserts should be used.
Transit damage identified before or after the unpacking of the goods is to be reported immediately.
Delayed notification of the damage cannot be honoured. Besides a fault description, the type of device in question and its serial number will be required.
Page 6 of 38
Installation and Operating Manual EPS
4.4 Warranty
Heinrichs Messtechnik
Your flowmeter was manufactured in accordance with the highest quality standards and was thoroughly tested prior to shipment. However, in the event any problem arises with your device, we will be happy to resolve the problem for you as quickly as possible under the terms of the warranty which can be found in the terms and conditions of delivery. Your warranty will only be honoured if the device was installed and operated in accordance with the instructions for your device. Any mounting, commissioning and/or maintenance work is to be carried out by qualified and authorized technicians only.
4.5 Return of the device for repairs or servicing
Note: According to German waste disposal legislation, it is the owner’s or customer’s responsibility to dispose of hazardous waste. Thus, before any devices is sent back to Heinrichs Messtechnik for servicing or repairs the following measures must be carried out:
When returning the unit for an inspection, please enclose as detailed a description as possible of the fault and the specific application, as well as the chemical-physical properties of the measurement medium.
Remove all residue of measurement medium which may be present. Paying special attention to the gasket grooves and crevices. This is especially important if the medium is detrimental to health and safety, for example: corrosive, poisonous, carcinogenic or radioactive etc.
Costs, which result from insufficient cleaning, (disposal and/or personal injuries) will be invoiced to the customer.
A decontamination certificate, to be found in section Mode of Operational and system design
5 Operation and system structure
5.1 Mode of operation
According to the faraday principle of electrodynamic induction, a conductive fluid flowing through a tube with an insulating lining and a flow velocity V, produces a voltage Um on the two electrodes to be found at right angles to the direction of flow and the magnetic field B generated by the field coils. The strength of this voltage Um is proportional to the mean flow velocity and therefore the volume flow rate.
Page 7 of 38
Heinrichs Messtechnik
5.2 System configuration
Installation and Operating Manual EPS
The electromagnetic flowmeter EPS consists of a sensor, which picks up an induced measuring signal from the medium flowing through the pipe, and a transmitter which transforms this signal into standardized output signals (4-20 mA or pulses). The sensor is installed in the pipe while the transmitter is mounted directly on the sensor (integral mount) or separately at an external location (remote mount), depending on the device version. compact-Version remote-Version
5.2.1 Integrally mounted transmitter
This type of configuration ensures easy and trouble-free installation.
Page 8 of 38
Installation and Operating Manual EPS
5.2.2 Remote mounted transmitter
Heinrichs Messtechnik
This type of configuration is recommended for confined spaces or if the temperature of the measured fluid is high. The connection between the sensor and the transmitter is established with a cable with separately shielded circuits for field coils and electrodes.
Transmitter version with a pigtail cable (maximal cable length 10 m)
The maximum cable length between the transmitter and sensor may not exceed 10 m.
The sensor is equipped with a terminal box, whereas the transmitter has a direct cable entry.
Before the system can be commissioned, the cable from the transmitter must be connected to the sensors terminal box.
Transmitter version with a terminal box
For cable lengths of 10 m or longer, it is mandatory that the cable is wired on both sides in a separately mounted terminal box.
6 Performance characteristics
6.1 Measuring accuracy
6.1.1 Measured error
+/- [0.3 % of actual value + 0.0001 * (Q at 10 m/s)]
Accuracy
Flow velocity v [m/s]
Page 9 of 38
Heinrichs Messtechnik
6.1.2 Repeatability
+/- [0.15 % of actual value + 0.00005 * (Q at 10 m/s)]
Installation and Operating Manual EPS
6.1.3 Reference conditions
In accordance with DIN EN 29104
Fluid temperature 22 °C ± 4 K
Ambient temperature 22 °C ± 2 K
Inlet section ≥ 10 x DN and outlet section ≥ 5 x DN
Sensor and transmitter are earthed
6.2 Fluid conductivity
5 µS/cm (
20 µS/cm for demineralized water)
6.3 Influence of ambient temperature
See Operating Instructions of the corresponding transmitter
6.4 Influence of fluid temperature
None
6.5 Materials
6.5.1 Wetted parts
Parts
Lining
Measuring and grounding electrodes
Grounding disk
Tri-Clamp® connection
Hygienic Conn. DIN 11851
Standard
EPDM
Hastelloy C276/C22
St.st. 1.4571
St.st. 1.4404 (316L)
St.st. 1.4404 (316L)
Others
Ceramics, PFA, soft rubber,
Hard rubber, PTFE
St.st. 1.4571,
Tantalum, Platinum, Titanium
Hastelloy C4,Tantalum
(on request)
(on request)
6.5.2 Non-wetted parts
Parts
Flow tube
Housing
DN 10 – 300
Standard
Stainless steel 1.4571
Varnished steel
Varnished steel
Stainless steel 1.4404
Aluminum pressure casting, varnished
Others
Stainless steel [1.4404/316L]
( standard with connections
DIN11851, Tri-Clamp® ® and
Wafer DN 2;3;6;10)
Flange threaded adapter
Terminal box for remote mount transmitter
6.5.3 External power supply / electrical connections
See rating plate and/or the operating manual of the transmitter.
7 Installation/conditions for use
7.1 Receipt of goods and transport
7.1.1 Receipt of goods
Check the packaging and contents for damage.
Inspect the supplied goods to ensure complete delivery and compare the consignment with your order specifications.
7.1.2 Transport
Please observe the following tips when unpacking your device, or transporting it to its measuring point:
Page 10 of 38
Installation and Operating Manual EPS Heinrichs Messtechnik
If possible the devices should be forwarded in the packaging in which they were delivered.
Do not remove any protection disks or caps from the process connections. This is particularly important in the case of sensors with a PTFE flow tube lining. The protection caps should only be removed immediately before installation of the device in the pipe.
Never lift the devices by the mounted transmitter housing or terminal box for transport. When transporting heavy devices, use slings. Place these around both process connections. Do not use chains as these can damage the surface coating and the housing.
When transporting devices without lugs, and when looping the slings around the flow tube, the center of gravity of the entire device can be higher than both attachment points of the slings. When transporting the device ensure that it does not rotate or slip accidentally. This could cause injury.
Sensors with a nominal sizes of more than DN 150 should not be lifted by the sheet metal of the shell with a forklift truck. This could dent the sheet metal of the shell and damage the internal solenoid coils.
There is also the risk that the device could roll off the forks.
If the sensor is provided with PTFE (Teflon) lining, then, upon delivery, the sensor is mounted with two wooden discs, to hold the lining in position during transport and storage. These wooden discs should remain on the sensor until installed. Without the discs, the liner creeps back to its original shape and installation is more difficult to do. The sensor should be left for at most a few hours without the discs.
Remove the discs immediately before installation.
7.2 Installation requirements
The installation location in the pipe must be selected so that the sensor is always fully filled with the fluid and cannot run empty. This can best be guaranteed if it is installed in an ascending pipe or drain.
The measuring principle is generally independent of the flow profile of the fluid provided no standing vortices reach into the area of measurement, such as from elbows or half-open sliding valves upstream from the sensor. In these cases measures must be taken to normalize the flow profile. Practical experience has shown that in most cases a straight inlet section of ≥ 5 x DN and an outlet section of ≥
2 x DN of the nominal sizes of the sensor is sufficient. The occurrence of strong electromagnetic fields in the vicinity of the installed sensor must be avoided.
For the Implementation of forward and backward flow measurements, both sides of the sensor must be provided with a straight pipe section with the nominal sizes of the sensor and a length of 5 DN of the nominal sizes of the sensor. It is advisable to install actuators, such as regulating or shut-off devices, downstream from the sensor. The forward flow direction is marked on the sensor with an arrow. When mounting sensors, always observe the specified screw torques.
After the installation of the sensor and the electrical connections between the sensor and the transmitter have been made, the system can be taken into operation. To prevent measuring errors caused by gas pockets in the fluid and damage to the lining of the sensor caused by negative pressure, the following points must be observed:
7.2.1 Bypass pipes
To allow for a problem-free dismount, emptying and cleaning of the sensor, a bypass pipe may be installed.
The bypass with a blind flange permits the fluid pipe to be cleaned without having to dismount the flowmeter.
This is recommended for highly soiled fluids.
7.2.2 Sensor tube lining
If the flow tube is lined with PTFE, the flowmeter must be installed with special care. The tube lining is bordered at the flanges (seal). This must not be damaged or removed as it prevents the fluid from penetrating between the flange and flow tube destroying the electrode insulation.
7.3 Installation
Screws, bolts, nuts and seals are not supplied by Heinrichs Messtechnik GmbH and must therefore be provided by the operator. The sensor is to be installed between the pipes. Please observe the required torques stated Section “Torques for screws and bolts”. The installation of additional grounding rings is
with DIN EN 1514-1. Mounted gaskets must not reach into the pipe cross-section.
Page 11 of 38
Heinrichs Messtechnik Installation and Operating Manual EPS
Caution!
Do not use conductive sealing compounds such as graphite. This could result in a conductive layer building up on the inside of the flow tube, short-circuiting the measuring signal.
7.3.1 Installation in pipelines with larger nominal sizes
The flowmeter can also be installed in pipes with larger nominal sizes by using pipe tapers (e.g. flange transition pieces in accordance with DIN EN 545). However, the resulting pressure loss must be taken into consideration. To avoid flow interruptions in the flow tube, a reducing angle of
8° for the tapers should be exceeded.
7.3.2 Horizontal or vertical Installation
The installed position of the flowmeter is arbitrary, however the intended x-y electrode axis should run approximately horizontal. A vertical Electrode axis should be avoided, since gas pockets or solid particles carried along in the fluid could affect the accuracy of the device.
7.3.3 Installation examples
To avoid measuring errors evoked by gas pockets or lining damage caused by negative pressure, the following points must be observed:
Page 12 of 38
Installation and Operating Manual EPS Heinrichs Messtechnik
Vibrations
To eliminate the effects of vibrations and prevent premature damage to the transmitter, the sensor shall be supported in the near vicinity of the flanges.
Horizontal pipeline routing
Install preferably in slightly ascending pipes.
Open inlet or outlet
Where possible, the device should be installed in a syphon. The empty pipe detection circuit of the transmitter is an additional safety feature for recognizing empty or partially filled pipes.
Caution!
There is the danger of accumulation of solids in the siphon. The installation of a cleaning aperture in the pipe is therefore advisable.
Down pipes
Where down pipes are to be found in the pipe system, a syphon or a ventilation valve should be placed after the sensor. By these means, negative pressure can be avoided in the pipeline, which may otherwise damage the sensor lining. This measure will also prevent a breakdown of the flow reducing the risk of air inclusions in the measurement medium.
Page 13 of 38
Heinrichs Messtechnik Installation and Operating Manual EPS
Long pipelines
In long pipelines there is always a danger of pressure surges. Therefore regulating and shut-off devices should be placed behind the sensor. However, when installed in vertical piping, especially when using sensors with PTFE-lining and high operating temperatures, the regulating and shut-off devices should be placed in front of the sensor (danger of vacuum).
Installation of pumps
To avoid negative pressure and eventual damage to the tube lining, never install flowmeters on the suction side of pumps.
If necessary, arrange for pulsation dampeners when using piston, diaphragm or hose pumps.
Consider the space requirements beforehand with respect to a potential deinstallation of the device.
Page 14 of 38
Installation and Operating Manual EPS
7.3.4 Earthing – potential equalisation
Heinrichs Messtechnik
The sound grounding concept of the flowmeter is a necessity for both safety reasons as well as to ensure a faultless operation. In accordance with VDE 0100 Part 410 and VDE 0100 Part 540 the grounding connections must be at protective conductor potential. For metrological reasons, this potential must be identical to the potential of the fluid . The grounding cable should not transmit any interference voltage.
For this reason do not simultaneously ground other electrical devices with this cable.
The measuring signal tapped at the electrodes amounts to only a few millivolts. Correct grounding of the electromagnetic flowmeter is therefore an important prerequisite for exact measurement. The transmitter requires a reference potential to evaluate the measured voltage on the electrodes. In the simplest case the non-insulated metal pipe and/or the connecting flange may be used as a reference potential.
Where pipes are lined with electrically insulating materials or pipes are made of plastic, the reference potential can be obtained from a grounding ring or grounding electrode. These establish the necessary conductive connection to the fluid and are made of a chemical-resistant material. The material used should be identical to that of the measuring electrodes
7.3.4.1 Earthing with earthing electrodes
The device can be optionally equipped with grounding electrodes. With plastic pipes this version is the easiest grounding method. As the surface of the grounding electrode is relatively small, the use of grounding rings on both sides of the sensor is preferable in systems in which high equalizing currents along the pipeline can be expected to occur.
7.3.4.2 Earthing with earthing rings
The outside diameter of the grounding ring should be at least equal to the diameter of the flange or be dimensioned in such a way that the grounding ring is positioned inside the flange bolts and is centered by these. The terminal lugs routed to the outside must be connected to the FE terminal in the junction box of the sensor. During installation ensure that the inner diameter of the seals do not protrude over the grounding disk!
The grounding cables are not included in the scope of your delivery and must be provided by the plant operator. The grounding rings can be ordered as accessories.
7.3.4.3 Examples of earthing the EPS
7.3.4.3.1 Metal piping electrically conductive
Page 15 of 38
Heinrichs Messtechnik
7.3.4.3.2 Plastic piping, or internally coated metal piping
Installation and Operating Manual EPS
7.3.4.3.3 Cathodic piping protection
Special care must be taken with cathodic protection pipelines.
For compact installation:
The transmitter must be powered by an isolation transformer.
The connection "PE" must never be connected.
With separate installation:
The shield must be connected to the sensor end via a 1.5 μF capacitor.
The shield must never be connected to both ends.
For isolated installation:
If the above connections are unacceptable, isolate the sensor from the piping.
Page 16 of 38
Installation and Operating Manual EPS
7.3.5 Torques for screws and bolts
Heinrichs Messtechnik
Due to the fact that the flow pipe lining is made of plastic or vulcanized materials such as hard rubber. Or
PTFE, electromagnetic flowmeters must be installed in pipe system with special care since these materials are malleable under pressure. If the flange screws are overtightened, the sealing surface will deform. If the seals are to function properly, the correct torque is highly important.
Tighten the screws crosswise so that the process connections are tight. When tightening the screws for the first time approx. 50 percent of the required torque should be reached, for the second time the torque should be 80 percent. The required torque should reach 100 percent when the screws are tightened for the third time. For higher torques it is advisable to use protectors.
The maximum permissible torques can be obtained from the following table
2
3
6
10 mm inch PN10 PN16
NBR
PN40 Class 150 AWWA
Ebonit/Weichgummi
PN10 PN16 PN40 PN6 PN10
PTFE
PN16 PN25 PN40
Nm f/lbs Nm f/lbs Nm f/lbs Nm f/lbs Nm f/lbs Nm f/lbs Nm f/lbs Nm f/lbs Nm f/lbs Nm f/lbs Nm f/lbs Nm f/lbs Nm f/lbs
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
15 ½" -
25 1" -
40 1½" -
50 2" -
65 2½" -
80 3" -
-
-
-
-
-
-
-
10 7
16 12 9
-
7
25 18 -
25 18 -
25 18 -
-
-
-
25
25
34
-
5
7
18
18
25 -
-
-
-
-
-
100 4" -
125 5" -
150 6" -
25 18 -
29 21 -
50 37 -
200 8" 50 37 50 37 -
250 10" 50 37 82 61 -
300 12" 57 42 111 82 -
-
-
-
26
42
57
19
31
42
88 65 -
99 73 -
132 97
-
-
-
-
350 14" 60 44 120 89
400 16" 88 65 170 125 -
-
450 18" 92 68 170 125 -
500 20" 103 76 230 170 -
600 24" 161 119 350 258 -
700 28" 200 148 304 224 -
750 30" -
800 32" 274 202 386 285 -
-
900 36" 288 213 408 301 -
1000 40" 382 282 546 403 -
42" -
1100 44" -
1200 48" 395 292 731 539 -
-
-
-
-
-
-
-
-
225 166 -
210 155 -
220 162 -
-
16 12 -
34
-
-
46 34 -
25 18 10
-
7
-
200 148 -
280 207 -
-
-
103 76 230 170
161 119 350 258
200 148 200 148 304 224 -
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
50 37 52 38 -
50 37 88 65 -
60 44 120 89 -
88 65 170 125 -
92 68 170 125 -
240 177
62
260 192 274 202 386 285
240 177 288 213 408 301
280 207 382 282 546 403 -
280 207 -
290 214 -
-
-
-
-
-
-
-
-
-
-
-
-
-
-
46 117 86
-
-
-
-
-
25 18 -
25 18 -
32 24 -
50
-
-
-
-
37
-
-
310 229 395 292 731 539 -
-
-
-
-
-
-
-
-
25
-
-
-
-
-
-
-
-
-
-
-
-
-
25 18 -
25
25
25
18
18
18 -
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
25
-
-
-
-
25 18 -
18
25 18 -
32 24 -
50 37 -
-
-
-
-
-
-
-
-
-
-
-
10
16
34
42
7
12
25
46 34
34 25
31
72 53
114 84
144 106
25 18 50 37 52 38 105 77 185 137
25 18 50 37 88 65 160 118 300 221
50 37 62 46 117 86 170 125 320 236
50 37 60 44 120 89 240 177 450 332
50 37 88 65 170 125 330 244 650 480
56 41 92 68 170 125 320 236 570 421
-
53 39 103 76 230 170 390 288 740 546
81 60 161 119 350 258 560 413 1220 900
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
Torque - calculation:
1) All bolts are new and the material selection complies with EN 1515-1 Table 2
2) Sealing material of maximum 75 Shore A hardness will be used between the flow meter and the associated flanges
3) All bolts are galvanized and greased accordingly
4) The values are calculated for use with carbon steel flanges
5) Flow meter and associated flanges are properly aligned
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
90
90
90
115
-
-
DN2+DN3 Zirkonium;
DN6-100Keramik
Nm
13 f/lbs
10
16
30
54
13
13
13
10
10
10
12
22
40
-
-
-
-
-
-
84
-
-
66
66
66
-
-
-
-
-
-
-
-
-
-
Page 17 of 38
Heinrichs Messtechnik
7.3.6 Remote mount transmitter
Installation and Operating Manual EPS
A separately installed transmitter and sensor is necessary if:
the installation area is difficult to access, space is restricted,
the fluid and ambient temperatures are high,
there is strong vibration.
Caution!
The cable between transmitter and sensor must be shielded. The outer cable shield must be electrically connected at both ends with special EMC cable glands (e.g. type Hummel HSK-M-
EMV).
For the remote mount version, the minimum permissible conductivity of the fluid is determined by the distance between the sensor and the transmitter. To ensure accuracy, a maximum cable length of 200 m should not be exceeded.
The electrode cable must be fixed during installation. If the conductivity of the fluid is low, cable movements may change the capacity considerably and thus disturb the measuring signals.
Do not lay the cables close to electrical machines or switching elements.
Do not connect or disconnect the field coil cable before the primary power of the flowmeter has been disconnected.
Maximum cable length when installed separately
Page 18 of 38
7.4
Installation and Operating Manual EPS
Wiring
Heinrichs Messtechnik
Please observe the following wiring hints. (Inappropriate wiring will invalidate any warranty claims)
Cable glands are not included in the standard scope of delivery
Customer used cable glands must adhere to the applied
Ex-provisions, and be suitable for the specified screw-threads
The cable glands must be appropriate for the diameter of the used cable
Make sure that the cable is routed to the cable gland in the form of a loop (Water trap") see picture on the right.
The cable glands should not be positioned with the opening pointing vertically upwards.
The used blind plugs may not be removed or replaced with cable glands,
when the corresponding bushing is not in use with electrical cables.
The sealing or tightening of the cable glands must be performed in accordance with the manufactures guidelines. Too tight or too loosely tightened cable glands could lead to water or other liquids penetrating into the enclosure.
Installation work or maintenance and repair work on the sensor, transmitter or the terminal box may only be carried out in a non-explosive atmosphere! Before any dismantling can take place, it must be ensured that the device is no longer energised
For the use of remote mounted transmitters:
Only sensors and transmitters with the same serial number may be interconnected. The connection of units with different serial numbers can lead to incorrect measurements.
Ensure that the stripped and twisted inner cable shield ends in the terminal box are kept as short as possible up to the terminals themselves If necessary these must be covered with an insulating sleeve to prevent short circuits.
The outer cable shield must be connected to the EMC cable gland connectors at both ends of the cable.
Important tightening torques for cable glands
KLE (cable glands) to the enclosure
Cap nut to the KLE (cable glands)
12 Nm
8 Nm
Page 19 of 38
Heinrichs Messtechnik
7.4.1 Mounted transmitter
Installation and Operating Manual EPS
On the integral mounted transmitter the connections to the sensor are internally wired. The terminal assignment is described in the operating manual of the UMF2 transmitter.
7.4.2 Transmitter mounted separately with a pigtail cable
Transmitter with pigtail cable (maximum cable length 10 m)
In this version the UMF2 transmitter, the sensor cable is conducted out of the transmitter as a pigtail. The cable is permanently fixed to the transmitter, and is regarded as part of the internal wiring of the transmitter.
The cable length is defined during the order process.
For connecting to the transmitter the sensor is equipped with a terminal box. It contains certified terminals and other elements of the WAGO series.
The Sensor cable is a double chamber cable with separated circuits for the exciter coils and electrode circuits.
Caution!
The pigtail is an integral part of the approval, and may neither be replaced, repaired nor may it be removed from the transmitter UMF2. A repair or replacement may only be carried out by the manufacturer. The securing Cap nut of the pigtail cable gland may not be removed or loosened.
Page 20 of 38
Installation and Operating Manual EPS
7.4.3 Connection plan
Heinrichs Messtechnik
Applying the outer cable shield to the EMC cable gland
For an optimal disturbance rejection, the outer cable shield of the sensor cable must be electrically connected to the special metal cable gland.
Schematic representation
7.5 Nominal size and ranges
Volume flow depends on the flow velocity and the nominal size of the flowmeter. The following flow rate nomogram shows the flow range which can be measured by a device with a specific nominal size as well as which nominal size is suitable for a specific flow rate. The electromagnetic flowmeter has been designed in such a way that it operates within the range of the flow velocities occurring in practical applications. The flow velocities have an upper range value of between 0.5 m/s and 10 m/s.
The nominal size DN of the sensor must be selected, if possible, in such a way that the flow velocity does not drop below the upper range value of 0.5 m/s. In case of fluids with solid particles, the flow velocity should range between 3 m/s and 5 m/s in order to prevent sedimentation depositing in the sensor.
The flow nomogram shows the volume flow in m³/h and the flow velocity in m/s in relation to the nominal size DN of the sensor. The y-axis shows the flow values in m³/h. The nominal size DN of the sensor has been selected as a parameter for the plotted straight lines. The upper range measuring value m³/h is taken as a basis for determining the sought nominal size DN. This value is given on the y-axis. The value for the flow velocity in m/s is shown on the x-axis. The straight line of the nominal size DN is found at the intersection of the two variables.
Page 21 of 38
Heinrichs Messtechnik
7.5.1 Flow - Diagram
Installation and Operating Manual EPS
Page 22 of 38
Installation and Operating Manual EPS
7.5.2 Flow specific values
Connection
Di
DN ASME
2 10 ½“
3 10 ½“
6 10 ½“
10 10 ½“
15 15 ½“
25 25 1"
40 40 1 ½“
50 50 2"
65 65 2 ½“
80 80 3"
100 100 4"
125 125 5"
150 150 6"
200 200 8"
250 250 10"
300 300 12"
350 350 14"
400 400 16"
450 450 18"
500 500 20"
600 600 24"
700 700 28"
800 800 32"
900 900 36"
1000 1000 40"
1200 1200 44"
1400 1400 48"
1600 1600 -
1800 1800 -
2000 2000 liter per second
35,34
48,11
62,83
79,52
98,17
141,4
192,4
251,3
318,1
392,7
565,5
769,7
1005
1272
1571
0,982
1,659
2,513
3,927
6,136
8,836
15,71
24,54
Qmin
0,002
0,004
0,014
0,039
0,088
0,245
0,628
706,9
962,1
1257
1590
1963
2827
3848
5027
6362
7854
11310
15394
20106
25447
31416
19,63
33,18
50,27
78,54
122,7
176,7
314,2
490,9
Qmax
0,031
0,071
0,283
0,785
1,767
4,909
12,57
Heinrichs Messtechnik m³ per hour
2545
3464
4524
5726
7069
10179
13854
18096
22902
28274
40715
55418
72382
91609
113097
70,69
119,5
181,0
282,7
441,8
636,2
1131
1767
Qmax
0,113
0,254
1,02
2,83
6,36
17,67
45,24
127,2
173,2
226,2
286,3
353,4
508,9
692,7
904,8
1145
1414
2036
2771
3619
4580
5655
3,534
5,973
9,048
14,14
22,09
31,81
56,55
88,36
Qmin
0,006
0,013
0,051
0,141
0,318
0,884
2,262
Page 23 of 38
Heinrichs Messtechnik
7.6 Ambient conditions
Installation and Operating Manual EPS
7.6.1 Ambient temperature limits
Medium temperature > 60 °C
When thermally insulating the sensor, it is important to avoid including the transmitter or the terminal box into the insulation.
Since the sensors are an element of the pipeline, when installed these are normally thermally isolated to save energy and prevent accidental physical contact. Due to the process temperature heat is introduced through the support for securing the transmitter or the terminal box. For this reason the thermal insulation of the sensor should not extend over more than the half of this support.
It is essential to prevent inclusion of the installed transmitter or the terminal box in the thermal insulation.
The maximum permissible fluid temperature range is stated on the rating plate of the respective version.
7.6.1.1 Maximum ambient temperature in dependence of the medium temperature for the mounted transmitter version.
7.6.1.2 Maximum ambient temperature for the sensor in dependence of the medium temperature for the remote mounted transmitters.
Page 24 of 38
Installation and Operating Manual EPS Heinrichs Messtechnik
7.6.1.3 Remote mounted transmitter
By remote mounted transmitters, the permissible ambient temperature for the sensor shall not exceed
-20 °C to + 60 °C.
7.6.2 Storage temperature range
The storage temperature range shall not exceed -40 °C to 80°C.
7.6.3 Climate category
In accordance with DIN EN 60654-1; Non weather-protected Class D1 locations with direct exposure to an open-air climates.
7.6.4 Ingress protection
The sensor meets the requirements of the protection class IP 67 . The following must be observed to ensure compliance with protection class IP67 when the device has been installed or serviced:
The housing seals must be clean and undamaged when placed in the sealing groove. If necessary the seals must be cleaned or replaced.
Tighten the cover of the terminal box also tightening the safety bug screw to secure the lid or tighten the screw cap of the transmitter (integral mount version).
The cables used for connection must comply with the specified outer diameter for the cable glands used.
Tighten the cable glands firmly, taking the specified torque into account
Loop the cable in front of the cable gland. Any moisture running along the cable can then drip off and not penetrate the device. Always install the device so that the opening of the cable gland does not face upwards.
Any unused cable glands must be closed with a plug which is suitable for the respective protection class.
The sensors are also available in an IP 68 version. The maximum permissible immersion depth in water is
5 m . In this case the transmitter is installed separately from the sensor. The cable at the sensor is connected by the manufacturer and the terminal box is resin filled (standard, non EX!). For the connection cable a special cable suitable for IP68 applications is used.
7.6.5 Shock and vibration resistance
The flowmeter should be protected from extreme shocks and vibrations, which could cause damage.
Maximum permissible shock/vibration: 15 m/s 2 (10 to150 Hz
Page 25 of 38
Heinrichs Messtechnik
7.7 Process pressure
Installation and Operating Manual EPS
The maximal permissible process pressure is stated on the rating plate, and is dependent on the medium temperature. The maximal process pressure for the appropriate flange can be obtained from the following table.
Flange Connection according to EN1092-1 (DIN 2501)
Flange Connection according to ASME B16.5
Page 26 of 38
Installation and Operating Manual EPS
7.8 Medium temperature and assignment of the temperature class
Heinrichs Messtechnik
The maximum allowable medium temperature of the sensor depends on the used version as well as from the pipe lining used in the sensor. The device-specific temperature is designated on the rating plate.
The Industrial Safety Act stipulates that equipment with very cold or hot components must be provided with protective guards to prevent accidental physical contact of employees with the respective parts.
For this reason, as well as from an energy aspect, in practical applications where temperatures of 60 °C and higher prevail, all pipes and installed measuring instruments are normally thermally insulated.
Thermal insulation of the sensor
The sensor can also be provided with a thermal insulation. However, the insulation should only reach maximal half way up the connecting tube, on which the terminal box or the transmitter is mounted.
In the following table the temperature ranges with respect to the used lining.
Lining Medium temperature range
EPDM -10 to +70 °C (14 to 158 °F)
PTFE -20 to +150 °C (-4 to +302 °F)
Ceramics (EPDM gasket) -20 to +150 C (-4 to +302 °F)
Ceramics (PTFE gasket) -20 to +130 C (-4 to +266 °F)
Hard rubber (Ebonit) 0° to +95 °C (32 to +203 °F)
0 to +70 ° C (32 to 158 °F) Soft rubber
Page 27 of 38
Heinrichs Messtechnik
8 Dimensions and weight
8.1
Installation and Operating Manual EPS
Dimensional drawing EPS – DN 10 to DN 1200, flange design
D
D
D in mm E in mm
(standard - pressure level)
PN 40
PN 16
PN10 *
DN ASME ASME
150lbs 300lbs
15
25
½”
1”
88,9 95,2
108 124
40 1½" 127 155,4
50 2" 152,4 165,1
65 2½" 177,8 190,5
80 3" 190,5 209,6
100 4" 228,6 254
125 5" 254 279,4
150 6" 279,4 317,5
200 8" 342,9 381
250 10" 406,4 444,5
PN 10
PN 16 *
300 12" 482,6 520,7
350 14“ 533,4 584,2
400
16“
596,9 647,7
450
18“
500
20“
635 711,2
698 774,7
600 24“ 812,8 914,4
700
800
-
-
-
-
900
1000
-
-
-
-
1200 -
EN1092-1
D1 in mm
EN 1092-1
340
395
445
505
565
615
670
780
90
115
150
165
185
200
220
250
285
880
1015
1115
1230
1455
104
104
124
139
154
174
214
239
282
338
393
444
451
502
563
614
715
816
927
1032
1136
1348
159
164,5
192
207
224,5
242
272
299,5
338,5
394
449
499,5
533
588,5
644
697
802,5
903
1026
1128,5
1238
1456,5
150lbs 300lbs
159 159
161 169
180,5 194,7
EN 1092-1
312
317,5
345
ASME
150lbs 300lbs
312,0 312,0
314,0 322,0
333,5 347,7
200,7
220,9
237,3
276,3
301,5
335,7
604,5
654
711
818,9
-
-
-
-
-
ASME
207,1
227,3
246,8
289
314,2
354,8
395,5 414,5
454,7 473,8
518,3 537,4
547,2 572,6
629,9
692,1
749,4
869,7
-
-
-
-
with transmitter
360
377,5
395
425
452,5
491,5
547
602
652,5
686
741,5
797
850
955,5
1056
1179
1281,5
1391
1609,5
353,7 360,1
373,9 380,3
390,3 399,8
429,3 442,0
454,5 467,2
488,7 507,8
548,5 567,5
607,7 626,8
671,3 690,4
700,2 725,6
757,5 782,9
807,0 845,1
864,0 902,4
971,9 1022,7
-
-
-
-
-
-
-
-
-
A in mm L with junction box
273,5
291
308,5
338,5
366
405
460,5
515,5
566
599,5
B in mm
ASME
EN 1092-1
150lbs 300lbs
225,5
231
258,5
225,5 225,5
227,5 235,5
247,0 261,2
267,2 273,6
287,4 293,8
303,8 313,3
342,8 355,5
368,0 380,7
402,2 421,3
59
59
82
72
72
72
85
85
85 mm Inch
200 7,9
200 7,9
200 7,9
200 7,9
200 7,9
200 7,9
250 9,8
250 9,8
300 12
462,0 481,0 137 350 14
521,2 540,3 157 450 18
584,8 603,9 157 500 20
613,7 639,1 270 550 22
655
710,5
763,5
869
969,5
1092,5
1195
1304,5
1523
671,0 696,4 270 600 24
720,5 758,6 310 600 24
777,5 815,9 350 600 24
885,4 936,2 320 600 24
-
-
-
-
-
-
-
-
-
-
450
560
700
800
28
32
630 900 35
670 1000 39
792 1200 47 weight / Mass
[ kg ]
9
11
12
16
19
27
4
5
8
40
60
80
110
125
175
200
287
330
450
530
660
1180
Weights are proximate (for PN16) without transmitter. For transmitter additional weight of 2.4 kg
(* higher pressure on request)
Page 28 of 38
Installation and Operating Manual EPS
8.2 Dimension DN 2; 3; 6; 10 incl. G ½“ (ISO) or ½“ NPT connection
Heinrichs Messtechnik
G ½“ (ISO) or ½“ NPT
6 x M4
20 mm
35 mm
52 mm
Size
1/12"
1/8"
1/4"
3/8"
DN
2
3
6
10
L (mm) (w/o gasket)
150
L (inch) (w/o gasket)
5,9
8.3 Dimension in DIN 11851, Tri-Clover®
L (mm) EPDML (mm) GraphitL (mm) Teflon
150
5,9
152
6
156
6,1
Size
DN
10
15
25
40
50
65
80
100
Ansi
3/8"
1/2"
1"
1 ½"
2"
2 ½
3
4
A (mm)
99
99
113
126
154
165
200
225
A (inch)
3,9
3,9
4,45
4,96
6,06
6,5
7,87
8,86
B (mm) C (mm)
159,2
159,2
180
207
240,7
261,1
297
336,5
B (inch)
6,26
6,26
7,09
8,15
9,47
10,28
11,69
13,25
Page 29 of 38
64
64
77,5
91
119
130
155
183
C (inch)
2,52
2,52
3,05
3,58
4,68
5,12
6,1
7,2
11851
L (mm)
146
146
161
176
186
223
258
288
L (inch)
5,75
5,75
6,34
6,93
7,32
8,78
10,16
11,34
Standard
Tri-Clover®
L (mm)
146
146
161
176
186
223
258
288
L (inch)
5,75
5,75
6,34
6,93
7,32
8,78
10,16
11,34
Heinrichs Messtechnik
8.4 Transmitter UMF2(B)
8.4.1 Compact design
Installation and Operating Manual EPS
8.4.2 Junction box of the sensor at remote design
8.4.3 Wall mounting
Page 30 of 38
Installation and Operating Manual EPS
8.4.4 Pipe mounting vertically
Heinrichs Messtechnik
8.4.5 Pipe mounting horizontally
8.5 Dimension grounding rings
Weight kg
Weight kg
Grounding ring type C is only for sensors with Neoprene, EPDM, LinatexR and Ebonite linings.
Grounding ring type E is only for sensors with PTFE lining.
Page 31 of 38
Heinrichs Messtechnik
9 Maintenance
Installation and Operating Manual EPS
The device requires no maintenance under normal operation. By measuring media, which tend to deposit and soiling of the electrodes and the measuring tube, a cleaning may be necessary.
10 Auxiliary power / electrical connection
See type plate or operating instructions of the associated converter.
11 CE typing
The measuring system complies with the legal requirements of the EMC Directive, the Low Voltage
Directive and the Pressure Equipment Directive.
We as the manufacturer confirm the conformity with the guidelines by affixing the CE mark.
12 Standards and guidelines, certificates and approvals
Low voltage directive 2014/35/EU (LVD)
EN 61010-1:2011 Safety requirements for electrical measuring, control and laboratory devices.
Directive 2014/30/EU (EMC)
EN 61000-6-2:2011
EN 61000-6-3:2011
EN 55011:2018
Immunity for Industrial environments
Emission residential environments
Group 1, class B (emitted interference)
Directive 2014/68/EU (PED)
AD-2000 Guidelines
EN60529:2010 Degrees of protection provided by enclosures (IP Code)
Page 32 of 38
Installation and Operating Manual EPS
13 Declaration of Decontamination
Heinrichs Messtechnik
for the cleaning of the device
Company: ...................................... Town/City: ..........................................
Department: ..................................
Tel.-No.: ........................................
The enclosed flowmeter
Name: ................................................
Model: EPS........................
Was operated using the following fluid: ..................................................................
Since the used fluid is potentially hazardous to water / toxic / corrosive / combustible*, we have:
- checked all cavities in the device to ensure that they are free of fluid residues *
- washed and neutralized all cavities in the device*
* Delete if not applicable.
We hereby confirm that no health or environmental hazard will arise from any fluid residues on or in the device enclosed for return.
Date: ............................. Signature: ...........................
Company stamp
Page 33 of 38
Heinrichs Messtechnik
14 Declaration of Conformity
Installation and Operating Manual EPS
Page 34 of 38
Installation and Operating Manual EPS Heinrichs Messtechnik
Page 35 of 38
Heinrichs Messtechnik
15 Modelcode
Model code Description
Sensor
EPS-
E
H
W
P
A
Material Lining
EPDM
Hard Rubber (Ebonit)
Soft Rubber
_
PTFE
Ceramics
Process connection
DIN connections
002E
003E
006E
010E
H02E
H03E
H06E
H10E
DN 2 G1/2"(316L) / incl. Gasket/EPDM
DN 3 G1/2"(316L) / incl. Gasket/EPDM
DN 6 G1/2"(316L) / incl. Gasket/EPDM
DN10 G1/2"(316L) / incl. Gasket/EPDM
DN 2 G1/2"(Hastelloy) / incl. Gasket/PTFE
DN 3 G1/2"(Hastelloy) / incl. Gasket/PTFE
DN 6 G1/2"(Hastelloy) / incl. Gasket/PTFE
DN10 G1/2"(Hastelloy) / incl. Gasket/PTFE
DIN flanges
Electrodes
Standard 2x Measuring- + 2x Groundelectrodes
Standard 2x Measuring- + 2x Groundelectrodes
Standard 2x Measuring- + 2x Groundelectrodes
Standard 2x Measuring Electrodes
Standard 2x Measuring Electrodes
Material
Lining / Electrodes / Sealing
Ceramics / Platinum / EPDM
Ceramics / Platinum / EPDM
Ceramics / Platinum / EPDM
Ceramics / Platinum / EPDM
Keramik / Platin / PTFE
Keramik / Platin / PTFE
Keramik / Platin / PTFE
Keramik / Platin / PTFE
201R
219R
239R
220R
240R
2A1R
2A2R
2A3R
2A4R
2A5R
2A6R
215R
235R
216R
236R
217R
237R
218R
238R
227R
208R
228R
210R
230R
211R
231R
212R
221R
203R
223R
205R
225R
206R
226R
207R
232R
213R
233R
214R
234R
395B
396B
ANSI connections
002A
003A
006A
010A
H02A
H03A
H06A
H10A
3B2B
380B
381B
384B
385B
38AB
389B
390B
391B
392B
393B
394B
362B
363B
365B
369B
370B
375B
376B
3B1B
331B
335B
336B
340B
341B
345B
346B
349B
305B
309B
317B
320B
321B
325B
326B
330B
350B
351B
355B
356B
358B
DN15 PN40 Form B1 DIN EN 1092-1
DN25 PN40 Form B1 DIN EN 1092-1
DN40 PN40 Form B1 DIN EN 1092-1
DN50 PN16 Form B1 DIN EN 1092-1
DN50 PN40 Form B1 DIN EN 1092-1
DN65 PN16 Form B1 DIN EN 1092-1
DN65 PN40 Form B1 DIN EN 1092-1
DN80 PN16 Form B1 DIN EN 1092-1
DN80 PN40 Form B1 DIN EN 1092-1
DN100 PN16 Form B1 DIN EN 1092-1
DN100 PN40 Form B1 DIN EN 1092-1
DN125 PN16 Form B1 DIN EN 1092-1
DN125 PN40 Form B1 DIN EN 1092-1
DN150 PN16 Form B1 DIN EN 1092-1
DN150 PN40 Form B1 DIN EN 1092-1
DN200 PN10 Form B1 DIN EN 1092-1
DN200 PN16 Form B1 DIN EN 1092-1
DN200 PN40 Form B1 DIN EN 1092-1
DN250 PN10 Form B1 DIN EN 1092-1
DN250 PN16 Form B1 DIN EN 1092-1
DN250 PN40 Form B1 DIN EN 1092-1
DN300 PN10 Form B1 DIN EN 1092-1
DN300 PN16 Form B1 DIN EN 1092-1
DN300 PN40 Form B1 DIN EN 1092-1
DN350 PN10 Form B1 DIN EN 1092-1
DN350 PN16 Form B1 DIN EN 1092-1
DN400 PN10 Form B1 DIN EN 1092-1
DN400 PN16 Form B1 DIN EN 1092-1
DN450 PN10 Form B1 DIN EN 1092-1
DN450 PN16 Form B1 DIN EN 1092-1
DN500 PN10 Form B1 DIN EN 1092-1
DN500 PN16 Form B1 DIN EN 1092-1
DN600 PN10 Form B1 DIN EN 1092-1
DN600 PN16 Form B1 DIN EN 1092-1
DN700 PN10 Form B1 DIN EN 1092-1
DN800 PN6 Form B1 DIN EN 1092-1
DN800 PN10 Form B1 DIN EN 1092-1
DN900 PN6 Form B1 DIN EN 1092-1
DN900 PN10 Form B1 DIN EN 1092-1
DN1000 PN6 Form B1 DIN EN 1092-1
DN1000 PN10 Form B1 DIN EN 1092-1
DN1200 PN6 Form B1 DIN EN 1092-1
DN1200 PN10 Form B1 DIN EN 1092-1 only EPDM
DN 2 - 1/2"NPT(316L) / incl. Gasket/EPDM
DN 3 - 1/2"NPT(316L) / incl. Gasket/EPDM
DN 6 - 1/2"NPT(316L) / incl. Gasket/EPDM
DN10 - 1/2"NPT(316L) / incl. Gasket/EPDM
Keramik / Platin / EPDM
Keramik / Platin / EPDM
Keramik / Platin / EPDM
Keramik / Platin / EPDM
DN 2 - 1/2"NPT(Hastelloy) / incl. Gasket/PTFE Keramik / Platin / PTFE
DN 3 - 1/2"NPT(Hastelloy) / incl. Gasket/PTFE Keramik / Platin / PTFE
DN 6 - 1/2"NPT(Hastelloy) / incl. Gasket/PTFE Keramik / Platin / PTFE
DN10 - 1/2"NPT(Hastelloy) / incl. Gasket/PTFE Keramik / Platin / PTFE
ANSI flanges
½ "Class 150 RF ASME B16.5-2003
½ "Class 300 RF ASME B16.5-2003
1" Class 150 RF ASME B16.5-2003
1" Class 300 RF ASME B16.5-2003
1½" Class 150 RF ASME B16.5-2003
1½" Class 300 RF ASME B16.5-2003
2" Class 150 RF ASME B16.5-2003
2" Class 300 RF ASME B16.5-2003
2½" Class 150 RF ASME B16.5-2003
2½" Class 300 RF ASME B16.5-2003
3" Class 150 RF ASME B16.5-2003
3" Class 300 RF ASME B16.5-2003
4" Class 150 RF ASME B16.5-2003
4" Class 300 RF ASME B16.5-2003
5" Class 150 RF ASME B16.5-2003
5" Class 300 RF ASME B16.5-2003
6" Class 150 RF ASME B16.5-2003
6" Class 300 RF ASME B16.5-2003
8" Class 150 RF ASME B16.5-2003
8" Class 300 RF ASME B16.5-2003
10" Class 150 RF ASME B16.5-2003
10" Class 300 RF ASME B16.5-2003
12" Class 150 RF ASME B16.5-2003
12" Class 300 RF ASME B16.5-2003
14" Class 150 RF ASME B16.5-2003
14" Class 300 RF ASME B16.5-2003
16" Class 150 RF ASME B16.5-2003
16" Class 300 RF ASME B16.5-2003
18" Class 150 RF ASME B16.5-2003
18" Class 300 RF ASME B16.5-2003
20" Class 150 RF ASME B16.5-2003
20" Class 300 RF ASME B16.5-2003
24" Class 150 RF ASME B16.5-2003
24" Class 300 RF ASME B16.5-2003
28" Class 150 RF ASME B16.5-2004
30" Class 150 RF ASME B16.5-2005
32" Class 150 RF ASME B16.5-2006
36" Class 150 RF ASME B16.5-2007
40" Class 150 RF ASME B16.5-2008
48" Class 150 RF ASME B16.5-2010
Process-Temperature
-10...+70°C
0°C...95°C
0°C...70°C
-20°C…+150°C
-20°C...+150°C
_
Max. measuring range
(Water 20°C, 1 mPas)
156 l/h
254 l/h
1017 l/h
2827 l/h
156 l/h
254 l/h
1017 l/h
2827 l/h
Installation and Operating Manual EPS
Page 36 of 38
156 l/h
254 l/h
1017 l/h
2827 l/h
156 l/h
254 l/h
1017 l/h
2827 l/h
6,3 m³/h
2544 m³/h
2544 m³/h
3463 m³/h
3463 m³/h
4523 m³/h
4523 m³/h
5725 m³/h
5725 m³/h
7068 m³/h
7068 m³/h
10178 m³/h
10178 m³/h
13854 m³/h
15904 m³/h
18095 m³/h
22902 m³/h
28274 m³/h
40715 m³/h
6 m³/h
17,6 m³/h
17,6 m³/h
45 m³/h
45 m³/h
70 m³/h
70 m³/h
119 m³/h
119 m³/h
180 m³/h
180 m³/h
282 m³/h
282 m³/h
441 m³/h
441 m³/h
636 m³/h
636 m³/h
1130 m³/h
1130 m³/h
1767 m³/h
1767 m³/h
2544 m³/h
2544 m³/h
2544 m³/h
3463 m³/h
3463 m³/h
4523 m³/h
4523 m³/h
5725 m³/h
5725 m³/h
7068 m³/h
7068 m³/h
10178 m³/h
10178 m³/h
13854 m³/h
18095 m³/h
18095 m³/h
22902 m³/h
22902 m³/h
27274 m³/h
27274 m³/h
6,3 m³/h
17,6 m³/h
45 m³/h
70 m³/h
70 m³/h
119 m³/h
119 m³/h
180 m³/h
180 m³/h
282 m³/h
282 m³/h
441 m³/h
441 m³/h
636 m³/h
636 m³/h
1130 m³/h
1130 m³/h
1130 m³/h
1767 m³/h
1767 m³/h
1767 m³/h
40715 m³/h
40715 m³/h
_
Installation length
150 mm
150 mm
150 mm
150 mm
156 mm
156 mm
156 mm
156 mm
150 mm
150 mm
150 mm
150 mm
156 mm
156 mm
156 mm
156 mm
200 mm
500 mm
500 mm
550 mm
550 mm
600 mm
600 mm
600 mm
640 mm
600 mm
730 mm
600 mm
860 mm
800 mm
900 mm
900 mm
1100 mm
1100 mm
1400 mm
200 mm
200 mm
200 mm
200 mm
200 mm
200 mm
200 mm
200 mm
272 mm
272 mm
272 mm
250 mm
310 mm
250 mm
335 mm
300 mm
300 mm
350 mm
350 mm
450 mm
450 mm
500 mm
500 mm
550 mm
550 mm
550 mm
600 mm
600 mm
600 mm
600 mm
600 mm
600 mm
600 mm
600 mm
700 mm
800 mm
800 mm
900 mm
900 mm
1000 mm
1000 mm
200 mm
200 mm
200 mm
200 mm
200 mm
200 mm
200 mm
200 mm
272 mm
250 mm
250 mm
250 mm
250 mm
300 mm
300 mm
350 mm
350 mm
350 mm
450 mm
450 mm
450 mm
1200 mm
1200 mm
Installation and Operating Manual EPS
JIS flanges
483R
486R
488R
491R
493R
496R
498R
463R
466R
468R
471R
473R
476R
478R
481R
441R
443R
446R
448R
451R
453R
456R
458R
461R
416R
418R
426R
428R
431R
433R
436R
438R
Food design
8" JIS K20
10" JIS K10
10" JIS K20
12" JIS K10
12" JIS K20
14" JIS K10
14" JIS K20
16" JIS K10
16" JIS K20
18" JIS K10
18" JIS K20
20" JIS K10
20" JIS K20
24" JIS K10
24" JIS K20
1" JIS K10
1" JIS K20
1½" JIS K10
1½" JIS K20
2" JIS K10
2" JIS K20
2½" JIS K10
2½" JIS K20
3" JIS K10
3" JIS K20
4" JIS K10
4" JIS K20
5" JIS K10
5" JIS K20
6" JIS K10
6" JIS K20
8" JIS K10
7000
7010
7020
7030
7040
7050
7060
6610
6630
6650
6660
6690
XXXX
-
0
1
2 x
S
H
T
N
M
X
DN15 Tri-Clamp-Connection
DN25 Tri-Clamp-Connection
DN40 Tri-Clamp-Connection
DN50 Tri-Clamp-Connection
DN65 Tri-Clamp-Connection
DN80 Tri-Clamp-Connection
DN100 Tri-Clamp-Connection
_
DN 15 Sanitary connection DIN11851
DN 25 Sanitary connection DIN11851
DN 40 Sanitary connection DIN11851
DN 50 Sanitary connection DIN11851
DN100 Sanitary connection DIN11851
Special on request special on request
Material Process connection w/o
Flange steel painted
Flange st.st. 1.4301 (DN2…10 in 1.4404) special on request
Material electrodes st.st. 1.4571 (316 -TI)
Hastelloy
Tantalum
Platinum / Iridium
Titanium
Special on request
Ceramics / Platinum
Ceramics / Platinum
Ceramics / Platinum
Ceramics / Platinum
Ceramics / Platinum
Ceramics / Platinum
Ceramics / Platinum
Ceramics / Platinum
Ceramics / Platinum
Ceramics / Platinum
Ceramics / Platinum
Ceramics / Platinum
(Standard for hard rubber, soft rubber)
(Standard for PTFE / EPDM - lining)
(Standard for ceramic lining)/ PTFE max. DN300/12"
H
T
N
M
X
0
S
Earthing electrode w/o st.st. 1.4571 (316 TI)
Hastelloy
Tantalum
Platinum
Titanium
Special on request
(Standard for PTFE / ceramic - lining)
(Standard for hard rubber, soft rubber)
EPDM; PTFE max DN600/24" (not available for ceramic lining)
PTFE max. DN600/24" (not available for ceramic lining)
PTFE max. DN300/12" (not available for ceramic lining)
PTFE max. DN600/24" (not available for ceramic lining)
6,3 m³/h
17,6 m³/h
45 m³/h
70 m³/h
119 m³/h
180 m³/h
282 m³/h
_
6,3 m³/h
17,6 m³/h
45 m³/h
70 m³/h
282 m³/h xx
1130 m³/h
1767 m³/h
1767 m³/h
2544 m³/h
2544 m³/h
3463 m³/h
3463 m³/h
4523 m³/h
4523 m³/h
5725 m³/h
5725 m³/h
7068 m³/h
7068 m³/h
10178 m³/h
10178 m³/h
17,6 m³/h
17,6 m³/h
45 m³/h
45 m³/h
70 m³/h
70 m³/h
119 m³/h
119 m³/h
180 m³/h
180 m³/h
282 m³/h
282 m³/h
441 m³/h
441 m³/h
636 m³/h
636 m³/h
1130 m³/h
Heinrichs Messtechnik special for hard rubber, soft rubber special for soft rubber, EPDM special for soft rubber, EPDM special for hard rubber rubber, EPDM special for hard rubber, soft rubber special for soft rubber, EPDM special for soft rubber, EPDM special for hard rubber, EPDM
350 mm
450 mm
450 mm
500 mm
500 mm
550 mm
550 mm
600 mm
600 mm
600 mm
640 mm
600 mm
680 mm
600 mm
800 mm
200 mm
200 mm
200 mm
240 mm
200 mm
240 mm
200 mm
272 mm
200 mm
272 mm
250 mm
310 mm
250 mm
335 mm
300 mm
300 mm
350 mm
144 mm
159 mm
174 mm
184 mm
221 mm
256 mm
286 mm
144 mm
159 mm
174 mm
184 mm
286 mm xx
0
1
2
B
C
1
2
3
Transmitter mounting
Integrated transmitter
Remote Transmitter
Remote Transmitter
Protection class, sensor
IP67
IP 67 terminal conn.box via M 20 x 1,5
IP 68, term.conn.box via M 20 x 1,5, encapsulated
Certificates w/o
Certificate of compliance with order 2.1
Test report 2.2
Inspection / material certificate 3.1 DIN/EN 10204:2008
Inspection / material certificate 3.2 DIN/EN 10204:2008
cable > 10m add. term.conn.box on transmitter add. term.conn.box on transmitter
Page 37 of 38
Heinrichs Messtechnik
UMF2-
L
M
G
H
A
B
C
D
1
1
2
4
F
G
0BH
0BK
Installation and Operating Manual EPS
Mounting
IP 67
Integrated transmitter IP 67 standard
Integrated transmitter IP 67 standard
Remote transmitter
Remote transmitter
IP 68 compact transmitter IP68 compact transmitter IP68 remote transmitter remote transmitter
Display and control unit integrated
Power supply
230V AC (+10%, -15%), 50/60Hz
115V AC (+10%, -15%), 50/60Hz
24V DC (±15%)
Output signal current output: 1x 0/4-20 mA current output: 1x 0/4-20 mA c/w HART protocoll
Design
Heinrichs
Kobold
Connection to sensor incl 2,5 m cable, c/w pipe-/wall mounting bracket incl 2,5 m cable, c/w pipe-/wall mounting bracket incl 2,5 m cable, c/w pipe-/wall mounting bracket incl 2,5 m cable, c/w pipe-/wall mounting bracket
Thread - electrical connection transmitter c/w term.conn.box transmitter c/w term.conn.box
½" NPT (f)
M20 x 1,5
1/2" NPT (f)
M20x1,5
½" NPT (f)
M20 x 1,5 modelcode options
3
4
5
6
7
0
1
2
A
B
C
D
E
F
G
H
X
Longer cable
IP 67
2,5m standard on remote version
5m
10m
15m
20m
30m
40m
50m
IP 68
2,5m standard on remote version
5m
10m
15m
20m
30m
40m
50m
Special on request breakout cable breakout cable breakout cable add junction box to transmitter add junction box to transmitter add junction box to transmitter add junction box to transmitter add junction box to transmitter add junction box to transmitter add junction box to transmitter add junction box to transmitter add junction box to transmitter add junction box to transmitter add junction box to transmitter add junction box to transmitter add junction box to transmitter
* Reference: EN 1092-1:2013
Consequently, whilst the mating dimensions, the flange and facing types and designations are compatible with those given in ISO 7005-1, it is important to take account of the following differences which exist in
EN1092-1:
1. The p/T ratings of this standard have been reduced in many cases by either limiting the lower temperature ratings which can no longer exceed the PN value, or by increasing the rate at which allowable pressures shall reduce with increase in temperature.
2. In addition to the range of PN 2,5 to PN 40 DIN origin flanges contained in the ISO standard,EN1092-1 also includes PN63 and PN100 flanges.
3. This standard specifies grades of European steels similar to those specified in ISO 7005-1, but in addition permits the use of those grades of ASTM steels which are specified in the ISO standard for use with the ANSI/ASME based flanges.
4. This standard gives an informative annex concerning requirements of EU Directives.
Subject to change.
Heinrichs Messtechnik GmbH
P.O. Box 600260
D-50682 Köln
Robert-Perthel-Straße 9
D-50739 Köln
Tel. +49-221-49708-0
Fax +49-221-49708-178 www.heinrichs.eu [email protected]
Page 38 of 38