ResMed Astral Series Invasive and Non Invasive Ventilators Installation Guide
ResMed Astral Series: Your path to better breathing. This innovative device offers tailored ventilation solutions for a range of respiratory conditions, including COPD, neuromuscular disorders, and restrictive thoracic disorders. With Astral, you can enjoy the freedom of mobile oxygen therapy and maintain an active lifestyle. The Astral Series is user-friendly and designed for easy operation, providing reliable and effective respiratory support.
Reklama
Reklama
Mouthpiece ventilation
Quick setup guide
English
Mouthpiece ventilation
The settings below are suggested for ‘open’ or ‘sip’ mouthpiece ventilation where the patient exhales to atmosphere frequently or continuously, eg, for on-demand daytime ventilation via a
15 mm non-vented mouthpiece. Other modes and settings are also available if required.
Mouthpiece circuit settings:
Ventilation Setting Selection
Patient type
Circuit
Adult
Mouthpiece circuit (tube only)
Detail/explanation
Ventilation mode
Tidal volume
Resp. rate
PEEP
Trigger
(A)CV
As appropriate
As appropriate
Off
Touch, High, Medium,
Low, Off
15 mm or 22 mm circuit without intentional leak or expiratory valve.
Note: Not designed to support continuous exhalation into the circuit.
(A)CV mode allows the patient to stack breaths as a set volume is delivered with each breath.
Set based on patient comfort and preference.
Resp. Rate shall be set appropriately for patients who may rely on the backup rate. Otherwise, it can be turned Off.
Not available in this configuration.
The Touch trigger setting will allow a breath to be delivered upon engagement of the mouthpiece or inspiratory effort is detected.
High, Medium and Low settings only deliver a breath when inspiratory effort is detected. If the patient experiences false triggering then the trigger sensitivity should be reduced.
Select Mouthpiece circuit.
Select ACV.
Set therapy settings as appropriate.
Set alarm settings as appropriate.
Single limb/Double limb circuit settings:
Ventilation Setting Selection
Patient type
Circuit
Interface
Ventilation mode
Adult
Single limb circuit with valve or Double limb
Mouthpiece
(A)CV
Detail/explanation
Tidal volume
Resp. rate
As appropriate
As appropriate
(A)CV mode allows the patient to stack breaths as a set volume is delivered with each breath.
Set based on patient comfort and preference.
Resp. Rate shall be set appropriately for patients who may rely on the backup rate. Otherwise, it can be turned Off.
PEEP
Trigger
Off
Very High, High, Medium,
Low, Very Low
If the patient experiences false triggering then the trigger sensitivity should be reduced.
Note: Vented (intentional leak) modes are not recommended for highly discontinuous ventilation such as ‘sip’ mouthpiece ventilation.
Safety considerations for mouthpiece ventilation
The Disconnect Alarm allows detection of circuit disconnection (for example, when the mouthpiece has fallen out of reach of the patient), and whether the patient is able to reliably trigger ventilation or not.
Regardless of whether the Disconnection Alarm is active or not, other mitigations may need to be put in place to ensure that patient safety is not compromised, such as the Apnea alarm, external monitoring, an SpO
2
alarm, or full-time supervision.
Alarm setting Selection
Disconnection Alarm On
Disconnection
Tolerance (%)
As appropriate
Alarm Activation Time As appropriate
Apnoea Response Off
Detail/explanation
Enables Disconnection Alarm.
Sets a higher or lower tolerance to the degree of circuit disconnection required to activate the Disconnection Alarm.
The time it takes for the alarm to activate once the disconnection threshold is satisfied. It can be adjusted from 5 seconds to 15 minutes for mouthpiece interface, as appropriate for the patient’s ventilator dependency.
It may be appropriate to configure Apnoea Response to OFF if the
Disconnection Alarm is appropriately configured.
Low pressure alarms are sometimes used to imply circuit disconnection and are quick to activate.
Should this be an annoyance, for example when the patient is receiving a partial breath or missing a breath, or if a false triggered breath occurs, it is at the discretion of the Clinician to turn OFF. Other mitigations may need to be put in place to ensure that patient safety is not compromised. This may include external monitoring, SpO
2
alarm, or full-time supervision.
Mouthpiece Ventilation with Astral 100/150 mouthpiece circuit is not intended to support continuous exhalation into the circuit. The non-user adjustable NV Mask/Rebreathing alarm will activate if the device detects continuous exhalation into the circuit. For patients that may prefer continuous exhalation into the circuit, a circuit with an expiratory valve should be considered.
Note: All alarms except for High pressure and Disconnection alarms are defaulted to Off when the
Mouthpiece circuit or Mouthpiece interface are selected via Setup Assistant in Adult configuration.
Because quality of life matters
ResMed’s mouthpiece ventilation solution has been thoughtfully designed to reflect the needs of patients and their carers and support good quality of life.
The EasySpeak mouthpiece is the ideal daytime solution for patients using Astral. It helps to make eating, breathing and talking more natural and can also help to reduce the risk of infections and complications 1 .
The circuit support arm holds the mouthpiece in a comfortable, accessible position. Quick to install, set up and clean, it readily adapts to any care environment, making it a practical choice in hospital, at home and on the move.
1
2
Ordering information
Product name
3
EasySpeak mouthpiece angled with 17cm flexi-tube 22F
EasySpeak mouthpiece angled for 15M connection
EasySpeak mouthpiece angled for 22M/15F connection
4 MPV Circuit Support Arm
4
Code
21353
21351
21354
27955
1 Boitano LJ, Benditt JO. An evaluation of home volume ventilators that support open-circuit mouthpiece ventilation. Respir Care. 2005 Nov ; 50(11):1457-61.