OKOndt GROUP Sonocon B Quick Start Manual

Add to My manuals
32 Pages

advertisement

OKOndt GROUP Sonocon B Quick Start Manual | Manualzz
ТМ
Ultrasonic Flaw Detector
SONOCON В
Quick start guide
С
Common features
В
UT version features
Т
Thickness Gauge + version features
www.oko-ndt.com
Battery compartment door
А-SCAN MANIPULATION
С
H,mm
- Full Screen
H,us
Y,mm
RECTIFICATION CHOICE
С
Negative halfwave
Radio Frequency
Positive halfwave
Full wave
-3-
GATES MANIPULATION
С
H,mm
GAIN MANIPULATION
Gain adjustment
Gain step
variation
С
С
0dB
N, dB
38dB
Add / remove some
number of dB
Reference
gain
В
80%
50%
Put signal to the
reference level
В (80% or 50%)
-4-
В
MENU NAVIGATION AND С
PARAMETERS ADJUSTMENT
В
INDICATION PANEL
Gain
Sound
Probe
frequency velocity
Indicators
panel
Probe
angle
Probe
zero
-5-
Measurands
INDICATORS
N dB addition
to the current
gain:
Reference
gain mode:
В
+dB / +dB
- Not added / Added
dBr / dBr
- Off / On
AGC:
AGC / AGC
- Inactive / Active
TCG:
TCG / TCG
- Inactive / Active
DAC system:
DAC / DAC
- Inactive / Active
DGS system:
DGS / DGS
- Inactive / Active
AWS rating:
AWS / AWS
- Inactive / Active
Measurement
threshold:
Measurement
mode:
1
/
- Edge / Peak
/ 1-2 / 2-C1 /
-
...
- Reference level for amplitude
Reference level: 50% / 80%
measurement
in % of vertical scale
Probe type:
/
Rectification
mode:
/
/
/
Signal freeze
modes:
/
/
/
Zoom modes:
/
/
Direct
manipulation
modes:
Battery
indicator:
1, 2 – gates 1 and 2;
C1, C2 – cursors 1 and 2.
- Single element / Dual element
Radio Frequency / Full wave /
Positive half waves /
Negative half waves
- Off / Frozen / Frozen + Live signals /
Estimate mode
- Off / Zooming gate 1 / Zooming gate 2
/ ASC / G1 / G2
-
Off / A-Scan manipulation /
Gate 1 manipulation /
Gate 2 manipulation
- Full / … / Almost empty
...
-6-
MENU STRUCTURE
In UT software version there are 3 sets of main menu - up to your choice:
user chosen menus
user chosen menus
В
In Thickness Gauge+ software version there is only one menu type:
Т
-7-
QUICK ACCESS
В
TO MAIN PARAMETERS
UT-B 1. Measurement and alarm
«Gates choice for Flaw Alarm System»
«Time/distance measurement
threshold choice»
«Measurement point marker»
PEAK
EDGE
«Choice of the gate(s) level
for the sound alarm»
«Choice of gate(s) and cursor(s)
for absolute or relative measurements»
Absolute by gate 1
ALARM
SOUND
ACCEPT.
ALARM
SOUND
ALARM
SOUND
OFF
RECORD
Relative
between
gate 1 and
gate 2
ALARM
SOUND
SEARCH
Relative between gate 1 and cursor 2
-8-
QUICK ACCESS
В
TO MAIN PARAMETERS
UT-B 2. Measurands choice
Maximum signal amplitude,
relative to the reference level
∆
A
off
Par. 1
AWS rating
Par. 4
Main
parameter
D
Par. 2
Equivalent area of
the indication
Gain, necessary to get
maximum signal to
the reference level
Par. 3
Se
Equivalent diameter
of the indication
H
Distance to
the indication
along the beam
X
De
Y
-9-
Distance to the
indication projection
to the scanning plane
Indication depth
QUICK ACCESS
В
TO MAIN PARAMETERS
UT-B 3. Display options
«Switching measurement units»
«Switching color schemes»
Light
«Display brightness adjustment»
-10-
Dark
«Interface language choice»
INDICATION SEARCH,
В
EVALUATION AND DOCUMENTATION AIDS
Evaluate mode
Helps finding the “best” signal. Just turn it on and scan.
P
55dB
G1
P P P
42dB
80%
G1
Review all readings and save the indication if necessary
- switches evaluate mode off
and retrives the original gain
-11-
INDICATION SEARCH,
В
EVALUATION AND DOCUMENTATION AIDS
Peak hold mode
Saves the peak values of all signals along the A-Scan.
P P P
- Peak curve reset
- Peak hold mode off
Signals compare mode
Displays both “frozen” and current signals.
P
1
2
P
1
2
- Signals compare mode off
-12-
INDICATION SEARCH,
В
EVALUATION AND DOCUMENTATION AIDS
Memory shortcuts. Results saving
Smart zoom mode
or
-13-
Probe
STRAIGHT-BEAM
В
PROBES CALIBRATION
-14-
ANGLE-BEAM
В
PROBES CALIBRATION
Calibration of the sound velocity
and probe zero at V2 calibration block
Probe
40 50 60 65
70
65
-15-
75
0
ANGLE-BEAM
В
PROBES CALIBRATION
Probe angle calibration at V2
Set up probe zero and sound velocity in V2 before strating this calibration.
Probe
V2-1
0
40 50
60 65
α
65
0
70
40 50
60 65
α
70
65
Pr
75
V2-2
75
or
ob
e
-16-
ANGLE-BEAM
В
PROBES CALIBRATION
Calibration of sound velocity and probe zero by
block or testing object end
Set up probe angle and X-value before starting this calibration
Probe
a
Probe
Probe
a
2 xa
-17-
ANGLE-BEAM
В
PROBES CALIBRATION
Calibration of sound velocity and probe zero by
any reflectors with known location depth
Set up probe angle before starting this calibration
Probe
Н1
1
2
Probe
Н2
1
-18-
2
RAYLEIGH WAVE
В
PROBES CALIBRATION
Calibration of probe zero at V1 block
Set up probe angle = 90° before starting this calibration
Probe
60
-19-
70
76
RAYLEIGH WAVE
В
PROBES CALIBRATION
Calibration of probe zero at V2 block
Set up probe angle = 90° before starting this calibration
Probe
40 50
70
65
-20-
75
0
60 65
RAYLEIGH WAVE
В
PROBES CALIBRATION
Calibration of sound velocity at testing object
Set up probe angle = 90° before starting this calibration
«Х01»
-21-
Probe
BASIC PREPARATION
FOR THICKNESS MEASUREMENT
Т
1. Connect probe and select it in the probes list – PROBE → Probe type (add it to
the list if necessary);
2. Calibrate probe zero (for details refer to next pages);
3. Set or calibrate sound velocity (for details refer to next pages);
4. Choose measurement type – MEASURE → Meas. type (0-1 by default);
5. Choose AGC mode MEASURE → AGC mode (Peak by default);
6. Specify such Maximum Gain Mode (MEASURE → MaxGain mode) and value of
Maximum Gain (MEASURE → MaxGain) for manual mode, that Maximum Gain is
high enough to raise back wall echoes at all thicknesses to be measured at
necessary level and low enough that no noises come in the gate when probe
is in the air;
7. Adjust scan and gate (for details please refer to the beginning of this guide) so
that back wall echoes at all thicknesses to be measured are in the gate limits;
8. Choose PRF mode (MEASURE → PRF mode)– High by default or other in case
if acoustic ghost signals are detected.
ADDITIONAL PREPARATION
Т
FOR THICKNESS MEASUREMENT
1. Choose measurement threshold (MEASURE →Threshold: Edge / 0-Edge);
2. Choose measurement mode (MEASURE → Mode: Conventional / Min & Max /
Differential / % of wear) and specify rated thickness for “Differential” and
“% of wear” modes;
3. Specify measurement frequency;
4. Switch on measurement point marker (GTS & ALRM → Marker)(visible in
zoom mode);
5. Set alarm bounds (GTS & ALRM →Tmin and Tmax), alarm mode (GTS & ALRM →
Alarms), visualization (GTS & ALRM →Show alarm zones) and sound
(GTS & ALRM →Sound).
-22-
ADDING NEW PROBE TO THE LIST
Continued at the next page
-23-
Т
CHOOSING CONNECTED
Т
PROBE FROM THE LIST
-24-
PROBE ZEROING
Т
Choose zero calibration mode (PROBE → Zero cal. mode: in air / on block)
Zeroing in “in air” mode
Pr
ob
e
Zeroing in “on block” mode
e
ob
Pr
-25-
FREE CALIBRATION
Т
Should be chosen in PROBE → Calibr.mode
Prior to use free calibration all necessary preparations should be made and a
thickness reading (can be incorrect) at the calibration block (or other object with
known thickness) should be acquired.
Probe
Calibrating probe zero (sound velocity is already set)
Н
Probe
Calibrating sound velocity (the probe is already zeroed)
Н
-26-
1-POINT CALIBRATION
Т
Should be chosen in PROBE → Calibr.mode
Prior to use 1-point calibration it is necessary to specify the calibration
block parameters in OPTIONS → Block 1pt menu.
Probe
Calibrating probe zero (sound velocity is already set)
Probe
Calibrating sound velocity (the probe is already zeroed)
-27-
2-POINTS CALIBRATION
Т
Probe
Should be chosen in PROBE → Calibr.mode
Prior to use 2-point calibration it is necessary to specify the calibration
blocks parameters in OPTIONS → Block 2pts-1 and Block 2pts-2 menus.
Probe
Н1
Н2
Thicknesses of calibration blocks for 2-points calibration should differ
at least 2 times.
-28-
INDICATION PANEL
Gain in gate 1
Т
Gain in gate 2
AGC maximum gain
Maximum gain mode
Alarm lower thickness threshold
Alarm upper thickness
threshold
Probe short name
Probe number
Probe zeroing mode
Sound velocity
Probe zero value
Current file name
Current file type
Current place / point
index along the 1st file
dimension
Current place / point
index along the 3d file
dimension
Current place / point
index along the 2nd file
dimension
Thickness value
recorded in the current
point or “free” or N/A
1
Indicators panel
INDICATORS
Т
Radio Frequency / Full wave /
- Positive half waves /
Negative half waves
Rectification mode:
/
Measurement
threshold:
/
- Edge / 0-Edge (by zero crossing)
/ 1-2
-
Direct manipulation:
modes
/ ASC / G1 / G2
- Gate 1 manipulation /
Indication of data
collection condition
at obtaining of slices /
thickness maps:
/
/
- Stopped / Active /
Zoom modes:
/
/
- Off / Zooming gate 1 /
Measurement mode:
Battery indicator:
1
/
/
Absolute (P-E) /
Relative (E-E) measurements
Off / A-Scan manipulation /
Gate 2 manipulation
Paused
Zooming gate 2
- Full / … / Almost empty
...
-29-
QUICK ACCESS
Т
TO MAIN PARAMETERS
TG+ 1. Gates choice for absolute or relative
measurements
«Choice of the measurement
mode: absolute or relative»
Absolute
Relative
«Measurement point marker»
«Time/distance measurement
threshold choice»
EDGE
Edge
«Flaw alarm logic choice»
Edge zero
crossing
Both "Too thin" and " Too thick”
«Choice of the gate(s) level
for the sound alarm»
"Too Thick" only
Оff
"Too Thin" only
-30-
QUICK ACCESS
Т
TO MAIN PARAMETERS
TG+ 2. Display options
«Resolution choice»
«Interface language choice»
«Display brightness adjustment»
«Measurement type choice»
Normal
«Switching color schemes»
Light
Dark
Percents of wear
Differential
Min. and Max. fixation
-31-
CONTENTS
С
С
С
С В
С
В
В
В
Т
В
В
В
В
В
Т
Т
Т
Т
Т
Т
Т
Т
Т
Т
Т
A-Scan manipulation
Rectification Choice
Gates manipulation
Gain manipulation
Menu navigation and parameters adjustment
Indication panel
Indicators
Menu structure
Quick access to main parameters:
UT-B 1. Measurement and alarm
UT-B 2. Measurands choice
UT-B 3. Display options
Indication search, evaluation and documentation AIDS:
Evaluate mode
Peak hold mode
Signals compare
Memory shortcuts. Results saving
Zoom mode
Straight-beam probes calibration
Angle-beam probes calibration:
Calibration of the sound velocity and probe zero at V2 calibration block
Probe angle calibration at V2
Calibration of sound velocity and probe zero by block or testing object end
Calibration of sound velocity and probe zero by any reflectors with
known location depth
Rayleigh wave probes calibration:
Calibration of probe zero at V1 block
Calibration of probe zero at V2 block
Calibration of sound velocity at testing object
Basic preparation for thickness measurement
Additional preparation for thickness measurement
Adding new probe to the list
Choosing connected probe from the list
Probe zeroing
Free calibration
1-point calibration
2-points calibration
Indication panel
Indicators
Quick access to main parameters:
TG+ 1. Gates choice for absolute or relative measurements
TG+ 2. Display options
3
3
4
4
5
5
6
7
8
9
10
11
12
12
13
13
14
15
16
17
18
19
20
21
22
22
23
24
25
26
27
28
29
29
30
31

advertisement

Key Features

  • Ultrasonic flaw detection
  • Thickness gauge
  • Multiple measurement modes
  • Advanced signal processing
  • Data logging and analysis
  • User-friendly interface
  • A-scan manipulation
  • Gate manipulation
  • Zoom mode
  • Probe calibration

Related manuals

Frequently Answers and Questions

What types of probes can I use with the Sonocon B?
The Sonocon B supports both straight-beam and angle-beam probes. You can calibrate the probe angle and sound velocity for accurate measurements.
How do I calibrate the probe zero?
You can calibrate the probe zero in two modes: in air or on a block. Refer to the manual for detailed instructions.
What is the difference between the UT and Thickness Gauge+ software versions?
The UT version offers three sets of main menus for advanced functionality, while the Thickness Gauge+ version provides a single menu optimized for thickness measurement.