Extech Instruments RF41 Portable Battery Coolant/Glycol Refractometer User's Guide

Extech Instruments RF41 Portable Battery Coolant/Glycol Refractometer User's Guide

Below you will find brief information for Battery/Coolant Brix Refractometer RF41. This refractometer can be used to measure freeze point of propylene glycol, ethylene glycol and cleaning fluid. Specific gravity readings can be obtained for battery acid. This instrument also features Automatic Temperature Compensation (ATC) to automatically account for temperature changes.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

Extech Battery/Coolant Brix Refractometer RF41 User's Guide | Manualzz

User's Guide 

 

 

 

 

Battery/Coolant Brix Refractometer (ATC) 

 

 

 

 

Model RF41 

Introduction 

Congratulations on your purchase of this Extech Refractometer. 

Precision optical instruments should be handled gently; avoid  touching the optical surface. Careful use of these instruments will  provide years of reliable service.  

Specifications 

 

 

Range 

 

Resolution 

Dimensions 

Weight 

0 to ‐50ºC Propylene Glycol freeze point 

0 to ‐50ºC Ethylene Glycol freeze point 

1.15 to 1.30 specific gravity battery acid 

0 to ‐40ºC cleaning fluid freeze point 

1ºC 

6.3 x 1.6 x 1.2" (160 x 40 x 30mm); 

6.4 oz. (180g) 

Description 

 

 

 

 

1. Eyepiece 

2. Mirror Tube 

3. Adjustment screw 

4. Cover Plate 

5. Prism 

 

Scale 

 

5 4 3 2 1

 

Operation 

This instrument measures the refractive index of a sample and  displays the results in ºC or Specific gravity.  

1. Zero Adjustment 

Cover the prism with a few drops of distilled water. Close the  cover plate and view the light/dark boundary (shadowline). If  necessary, turn the adjustment screw until the shadowline aligns  with the scale’s zero line. The zero calibration should be  performed at ambient temperature and should be repeated every 

30 minutes if the temperature changes. After the zero adjustment,  clean the prism with a soft cloth. 

2. Sample Preparation and Reading 

Cover the prism with a few drops of distilled water from the  included vial. Close the cover plate so that the liquid spreads  across the entire surface of the prism without air bubbles or dry  spots. Allow the sample to  remain on the prism for  approximately 30 seconds. 

X X

While holding the instrument  under a light source, look  through the eyepiece. The brix  concentration is determined by  the intersection of the  boundary of the light and dark  fields  on the scale. If the scale  appears out of focus, adjust by rotating the eyepiece until the  scale is in focus. The instrument also features an eye‐guard to  prevent stray light from entering the eyepiece and causing  reflections.  

It may be necessary to adjust the position of the light source to  maximize the contrast of the shadowline. Under normal  conditions, optimal contrast is obtained by holding the instrument  underneath and perpendicular to a light source.  

Once a reading has been taken, wipe dry with a clean cloth (do not  wash or rinse) and place the instrument in the supplied plastic  case. Store the instrument in a safe, dry environment. 

 

3. Automatic Temperature Compensation (ATC) 

Temperature compensation relieves the user of the responsibility  to measure temperature and apply a correction factor when  taking readings. This refractometer makes this correction 

automatically. When ambient temperature varies from 20°C,  readings are automatically adjusted to compensate for  temperature variance between 10°C to 30°C.  

Warranty 

FLIR Systems, Inc. warrants this Extech Instruments brand device to be free of  defects in parts and workmanship for one year from date of shipment (a six month  limited warranty applies to sensors and cables). If it should become necessary to  return the instrument for service during or beyond the warranty period, contact  the Customer Service Department for authorization. Visit the website  www.extech.com for contact information. A Return Authorization (RA) number  must be issued before any product is returned. The sender is responsible for  shipping charges, freight, insurance and proper packaging to prevent damage in  transit. This warranty does not apply to defects resulting from action of the user  such as misuse, improper wiring, operation outside of specification, improper  maintenance or repair, or unauthorized modification. FLIR Systems, Inc. specifically  disclaims any implied warranties or merchantability or fitness for a specific  purpose and will not be liable for any direct, indirect, incidental or consequential  damages. FLIR’s total liability is limited to repair or replacement of the product. 

 

The warranty set forth above is inclusive and no other warranty, whether written  or oral, is expressed or implied. 

Calibration, Repair, and Customer Care 

Services 

FLIR Systems, Inc. offers repair and calibration services for the 

Support Lines: U.S. (877) 439‐8324; International: +1 (603) 324‐7800 

Technical Support: Option 3; E‐mail: [email protected] 

Repair & Returns: Option 4; E‐mail: [email protected] 

Product specifications are subject to change without notice 

Please visit our website for the most up‐to‐date information  www.extech.com 

FLIR Commercial Systems, Inc., 9 Townsend West, Nashua, NH 03063 USA 

ISO 9001 Certified 

Extech Instruments products we sell. NIST certification for most  products is also provided. Call the Customer Service 

Department for information on calibration services available for  this product. Annual calibrations should be performed to verify  meter performance and accuracy. Technical support and  general customer service is also provided, refer to the contact 

  information provided below. 

Copyright © 2013‐2015 FLIR Systems, Inc. 

All rights reserved including the right of reproduction in whole or in part in any  form 

www.extech.com 

advertisement

Key Features

  • Measures freeze point of propylene glycol, ethylene glycol and cleaning fluid
  • Specific gravity readings for battery acid
  • Automatic Temperature Compensation (ATC)
  • Zero adjustment for accurate readings
  • Easy to use, simply place a few drops of sample on the prism
  • Durable construction for long-lasting use
  • Comes with a plastic case for storage and protection
  • Includes a vial of distilled water for calibration
  • Warranty coverage for one year from date of shipment

Frequently Answers and Questions

What is the range of this refractometer?
The refractometer has a range of 0 to -50°C for propylene glycol freeze point, 0 to -50°C for ethylene glycol freeze point, 1.15 to 1.30 for specific gravity battery acid, and 0 to -40°C for cleaning fluid freeze point.
What is the resolution of this refractometer?
The resolution of this refractometer is 1°C.
How do I calibrate the refractometer?
To calibrate the refractometer, cover the prism with a few drops of distilled water and close the cover plate. Adjust the adjustment screw until the shadow line aligns with the scale's zero line. Calibration should be performed at ambient temperature and repeated every 30 minutes if the temperature changes.
How do I use the refractometer?
To use the refractometer, cover the prism with a few drops of the sample and close the cover plate. Allow the sample to remain on the prism for approximately 30 seconds. While holding the instrument under a light source, look through the eyepiece. The measurement is determined by the intersection of the boundary of the light and dark fields on the scale.
What is the warranty on this refractometer?
This refractometer is warranted to be free of defects in parts and workmanship for one year from the date of shipment. A six month limited warranty applies to sensors and cables.

Related manuals

Download PDF

advertisement