ELx50 Operator`s Manual Revision K


Add to my manuals
188 Pages

advertisement

ELx50 Operator`s Manual Revision K | Manualzz

Microplate Strip Washer

ELx50

Operator’s Manual

All Models

.

ELx50™

Microplate Strip Washer

Operator’s Manual

August 2008

© 2008

Part Number 4071000

Revision M

BioTek

®

Instruments, Inc.

ii

| Preface

Notices

BioTek

®

Instruments, Inc.

Highland Park, P.O. Box 998

Winooski, Vermont 05404-0998 USA

All Rights Reserved

© 2008, BioTek

®

Instruments, Incorporated. No part of this publication may be reproduced, transcribed, or transmitted in any form, or by any means electronic or mechanical, including photocopying and recording, for any purpose other than the purchaser’s use without written permission of BioTek Instruments, Inc.

Trademarks

BioTek

®

is a registered trademark, and ELx50™ and Liquid Level Alert™ are trademarks of BioTek Instruments, Inc. Microsoft

®

and Excel

®

are either registered trademarks or trademarks of Microsoft Corporation in the United States and/or other countries. All other trademarks are the property of their respective holders.

Restrictions and Liabilities

Information in this document is subject to change and does not represent a commitment by BioTek Instruments, Inc. Changes made to the information in this document will be incorporated in new editions of the publication. No responsibility is assumed by BioTek for the use or reliability of software or equipment that is not supplied by BioTek or its affiliated dealers.

BioTek Instruments, Inc.

Contents |

iii

Contents

Notices ................................................................................. ii

Contact Information .................................................................. viii

Document Conventions............................................................... ix

Revision History.......................................................................... x

Intended Use Statement ............................................................ xiii

Quality Control ......................................................................... xiii

Repackaging and Shipping ......................................................... xiii

Warnings................................................................................. xiii

Hazards and Precautions............................................................xiv

CE Mark Information .................................................................xvi

Electromagnetic Interference and Susceptibility ........................... xvii

User Safety............................................................................. xvii

Safety Symbols ...................................................................... xviii

Chapter 1: Introduction ................................................................. 1

Introducing the ELx50 Microplate Strip Washer ............................... 2

ELx50 Washer Variations ............................................................. 3

Manifold Variations................................................................. 3

Carrier Variations ................................................................... 5

Software Variations ................................................................ 5

Hardware Features...................................................................... 7

Accessories Included ................................................................... 8

Optional Accessories ................................................................... 9

Specifications ........................................................................... 10

Microplates ......................................................................... 10

Electrical............................................................................. 10

Physical .............................................................................. 10

Environmental ..................................................................... 10

Hardware............................................................................ 10

Performance Specifications .................................................... 11

Product Support & Service ......................................................... 12

Contacting the Technical Assistance Center.............................. 12

Returning Instruments for Service/Repair ................................ 12

Chapter 2: Installation................................................................. 13

Unpacking and Inspecting the Instrument .................................... 14

Setting Up the ELx50 ................................................................ 20

Operating Environment ......................................................... 20

Installing the Carrier and Manifold .......................................... 20

Microplate Carrier............................................................ 20

8-, 8s-, or 12-Channel Manifold ......................................... 21

16-Channel Manifold ........................................................ 22

ELx50 Operator's Manual

iv

| Preface

Aspiration Carrier (ELx50/8FMW)....................................... 23

Connecting the Tubes and Bottles........................................... 25

Before Connecting the Tubes and Bottles ............................ 25

Waste System................................................................. 28

Supply System................................................................ 31

Liquid Level Alert Option................................................... 34

Final Check..................................................................... 35

Attaching the Mist Shield....................................................... 35

Connecting the Power Supply and Cord ................................... 36

Serial Port for Communication With Other Devices .................... 36

Verifying Performance ............................................................... 37

Repackaging and Shipping ......................................................... 38

Before Repackaging the Instrument ........................................ 38

Repackaging the ELx50 and Its Accessories ............................. 38

Chapter 3: Operation ................................................................... 51

ELx50 Front Panel..................................................................... 52

Keypad Description ................................................................... 53

Entering Program Names ........................................................... 54

System Startup ........................................................................ 54

Main Menu ............................................................................... 55

General Usage Guidelines .......................................................... 56

Before Running Any Program ................................................. 56

Maintenance........................................................................ 58

Programming Considerations ................................................. 58

Liquid Level Alert Option ................................................... 59

Wash ...................................................................................... 60

Run Wash Program............................................................... 60

Run Wash Program Options .............................................. 60

Define Wash Program ........................................................... 61

Define Wash Program Options ........................................... 61

Define Wash METHOD Program Options .............................. 62

Wash METHOD Defaults and Ranges................................... 63

Define Wash DISPENSE (DISP) Program Options.................. 64

Wash DISPENSE Defaults and Ranges ................................ 66

Define Wash ASPIRATION (ASPIR) Program Options............. 67

Wash ASPIRATE Defaults and Ranges................................. 69

Edit Wash Program............................................................... 71

Copy Wash Program ............................................................. 71

Delete Wash Program ........................................................... 71

Prime...................................................................................... 72

Define Prime Program........................................................... 72

Define Prime Program Options........................................... 72

PRIME Program Defaults and Ranges.................................. 73

BioTek Instruments, Inc.

Contents |

v

Run Prime Program .............................................................. 74

Edit Prime Program .............................................................. 74

Copy Prime Program............................................................. 75

Delete Prime Program........................................................... 75

Dispense ................................................................................. 76

Define Dispense Program ...................................................... 76

Run Dispense Program.......................................................... 76

Edit Dispense Program.......................................................... 77

Copy Dispense Program ........................................................ 77

Delete Dispense Program ...................................................... 78

Aspiration ................................................................................ 79

Define Aspiration Program ..................................................... 79

Define Aspiration Program Options..................................... 79

Define Aspirate-only Program Options ................................ 80

Aspirate-only Program Defaults and Ranges ........................ 80

Run Aspiration Program ........................................................ 80

Edit Aspiration Program ........................................................ 81

Copy Aspiration Program....................................................... 81

Delete Aspiration Program..................................................... 82

Soak ....................................................................................... 83

Define Soak Program............................................................ 83

Edit Soak Program ............................................................... 84

Copy Soak Program.............................................................. 84

Delete Soak Program............................................................ 85

Linking Programs ...................................................................... 86

Define Link Program ............................................................. 86

View Link Program ............................................................... 87

Run Link Program ................................................................ 87

Delete Link Program ............................................................. 88

Utility...................................................................................... 89

Tests.................................................................................. 89

Self-Test (SLFCHK).......................................................... 89

Configuration Information (CHKSUM) ................................. 89

Setup ................................................................................. 90

Manifold (MANIFLD)......................................................... 90

SENSOR......................................................................... 90

AutoPrime........................................................................... 91

ELx50™ Microplate Strip Washer Menu Maps ................................ 92

Chapter 4: Maintenance and Performance Verification

(IQ/OQ/PQ) ................................................................................ 97

Overview ................................................................................. 98

Recommended Maintenance Schedule.......................................... 99

Recommended Performance Verification Schedule ........................100

ELx50 Operator's Manual

vi

| Preface

Warnings & Precautions ............................................................101

Maintenance Programs .............................................................102

Prime Programs.......................................................................103

Daily Maintenance....................................................................104

DAY_RINSE ........................................................................104

AUTOPRIME........................................................................104

Overnight Maintenance .............................................................106

OVERNIGHT_LOOP ..............................................................106

RINSE_AND_SOAK ..............................................................107

Monthly Maintenance................................................................108

Tools and Supplies ..............................................................108

Cleaning the Bottles ............................................................109

Cleaning the Manifold ..........................................................109

Cleaning the Aspirate and Dispense Tubes ..............................110

8-, 8s-, or 12-Channel Manifold ........................................110

16-Channel Manifold .......................................................110

Cleaning the Carrier ............................................................111

Microplate Carrier...........................................................111

Aspiration Carrier (ELx50/8FMW only) ...............................112

Cleaning the Check Valves....................................................114

Long-Term Storage Maintenance................................................115

LONG_SHUTDOWN..............................................................115

Disinfect the Aspiration Carrier (ELx50/8FMW) ........................116

Storing the ELx50™ .................................................................116

Decontamination .....................................................................117

Purpose .............................................................................117

Tools and Supplies ..............................................................118

Procedure for External Components .......................................119

Procedure for Tubing and Manifold.........................................119

Alternate Decontamination Procedure for Tubing and Manifold ...120

Procedure for Aspiration Carrier (ELx50/8FMW) .......................120

Performance Verification ...........................................................122

Installation Qualification (IQ).....................................................123

Operational Qualification (OQ) ...................................................123

Performance Qualification (PQ) ..................................................124

Self-Test and Checksum Test.....................................................125

Self-Test............................................................................125

Configuration Information (CHKSUM) .....................................125

Liquid Tests for the 8-, 8s- and 12-Channel Manifolds ...................126

Evacuation Efficiency and Dispense Precision Tests ..................126

Materials .......................................................................126

Evacuation Efficiency Test Procedure .................................127

Dispense Precision Test Procedure ....................................128

Liquid Tests Worksheet (8-, 8s-, and 12-Channel Manifolds)......129

BioTek Instruments, Inc.

Contents |

vii

Liquid Tests for the 16-Channel Manifold .....................................130

Evacuation Efficiency and Dispense Precision Tests ..................130

Materials .......................................................................130

Evacuation Efficiency Test Procedure .................................131

Dispense Precision Test Procedure ....................................132

Liquid Tests Worksheet (16-Channel Manifold) ........................133

Liquid Test for the Aspiration Carrier (ELx50/8FMW) .....................134

Evacuation Efficiency Test ....................................................134

Materials .......................................................................134

Evacuation Efficiency Test Procedure .................................134

Liquid Test Worksheet (Aspiration Carrier) ..............................136

Maintenance/Performance and Operational Logs...........................137

Chapter 5: Troubleshooting ....................................................... 141

Overview ................................................................................142

Troubleshooting Checklist.....................................................142

Troubleshooting Charts.............................................................143

Washer Start-Up .................................................................143

Fluid Aspiration...................................................................144

Fluid Delivery .....................................................................148

Fluid Leakage .....................................................................150

Carrier Movement ...............................................................151

Microplate Carrier...........................................................151

Aspiration Carrier (ELx50/8FMW)......................................151

Washer Manifold Movement ..................................................151

Microplate Scratches............................................................152

Syringe Drive Movement ......................................................152

Appendix A: Error Codes ............................................................ 153

Appendix B: ELx50 Manifold Variations...................................... 159

Appendix C: Chemical Compatibility .......................................... 163

Appendix D: Default Programs................................................... 165

ELx50 Operator's Manual

viii

| Preface

Contact Information

BioTek

®

Instruments, Inc.

Highland Park, P.O. Box 998

Winooski, Vermont 05404-0998 USA

Customer Service and Sales

Internet: www.biotek.com

Phone: 888-451-5171 (toll free in the U.S.)

Fax:

802-655-4740 (outside the U.S.)

802-655-7941

E-Mail: [email protected]

Service/TAC

Phone: 800-242-4685 (toll free in the U.S.)

802-655-4740 (outside the U.S.)

Fax: 802-654-0638

E-Mail: [email protected]

European Coordination Center/Authorized European

Representative

BioTek

®

Instruments GmbH

Kocherwaldstrasse 34

D-74177 Bad Friedrichshall

Germany

Internet: www.biotek.de

Phone: +49 (0) 7136 9680

Fax: +49 (0) 7136 968 111

E-Mail: [email protected]

BioTek Instruments, Inc.

Document Conventions |

ix

Document Conventions

This manual uses the following typographic conventions:

Example Description

This icon calls attention to important

safety

notes.

Warning!

Caution

DEFINE

A

Warning

indicates the potential for bodily harm and tells you how to avoid the problem.

A

Caution

indicates potential damage to the instrument and tells you how to avoid the problem.

Text in COURIER font represents menu options as they appear on the instrument’s display.

Note:

Bold

text is primarily used for emphasis.

This icon calls attention to important information.

About This Manual

The intent of this Operator’s Manual is to instruct the new user how to set up and operate BioTek’s ELx50™. To help you read and understand this manual, certain document conventions have been used.

Major topic headings start a new page (such as

Document Conventions

, above) to give you a visual and style clue that a new major subject is being introduced. One or more subheadings may appear below each major heading.

ELx50 Operator's Manual

x

| Preface

Revision History

Revision Date Changes

B

C

D

E

9/97

4/99

4/99

6/99

Added valve tubing connection diagram. Corrected rear panel figure.

Updated menu sequences, addresses.

Revised operator’s manual to reflect BOM revision.

Added support for 16-channel manifold/384-well plates.

Performance Verification, and added it to Chapter 4. Added a log to record maintenance.

Updated Notices to include current company contact information.

Updated Safety Symbols and Intended Use. Added Chemical

Compatibility information to Precautions. Updated Warranty to include

Bio-Tek’s current warranty statement. Changed operating environment temperature range from 15°C - 30°C to 15°C - 40°C. Added Registration

Card information. Clarified bleach dilutions for Decontamination. selectable ranges for the following throughout manual to conform to software specs: Prime Volume from 1 to 999 ml to 1 to 200 ml; Shake

Duration from 0 to 600 seconds to 1 to 600 seconds, Bottom Dispense

Volume from 25 to 3000 μl/well to 50 to 3000 μl/well. Updated introductory material, enhanced Hardware Features, Package Contents, and Optional Accessories. Corrected specs for physical dimensions and storage and operating temperatures. Updated contact information, enhanced Unpacking and Packing instructions with new text and 14 new figures. Enhanced Operation chapter with new sections: Keypad

Description, System Startup, Main Menu, and Utility. Added new

Defaults and Ranges tables for Wash Method, Wash Dispense, Wash

Aspirate, and Prime Programs. Restructured Chapter 4, Maintenance and

Performance Verification for continuity, enhanced Decontamination section, updated Liquid Testing section. Enhanced/updated

Troubleshooting and Error Codes. Added two new appendices C,

Chemical Compatibility, and D, Default Programs.

Added instructions and information in all chapters of the manual for the aspiration carrier (ELx50™/8FMW model). Removed hyphen from

“Bio-Tek” (“BioTek”) and changed “ELx50 Automated Strip Washer” to

“ELx50 Microplate Strip Washer.” Replaced the previous cover with new cover design from Marketing.

Updated regulatory/safety information and removed Warranty and

Registration Card sections in the Preface. Revised ranges for prime flow rate, dispense flow rate, bottom flow rate, and dispense volume for the

12-channel manifold in Chapter 3, to reflect recent revisions in the washer’s software. Revised instructions for flushing BSA in Chapter 4.

Incorporated manual update H2: added dispense volume ranges for 96- and 384-well plates for the 16-channel manifold in Chapter 3 and in

Appendix A, and revised the Recommended Maintenance/Performance

Verification Schedule in Chapter 4 by moving the Liquid Tests from the

IQ to the Initial/Annual OQ and monthly PQ.

BioTek Instruments, Inc.

Revision History |

xi

Revision Date Changes

aspiration carrier in the ELx50/8FMW washer: the vacuum grate is now secured in place (under the plate seal gasket) by six screws around the perimeter of the top of the carrier. Also revised vacuum levels for the aspiration carrier: from –240 mmHg to –64 mmH for the 0.047” diameter hole of the carrier’s vent port, and from –64 mmHg to –240 mmHg for the 0.020” diameter hole in the black vent plug.

General. Added information on the Liquid Level Alert™ option, which is available for all ELx50 models. Added information on the new ‘First Strip’ run-time parameter. Additional cleanup and clarifications as described below.

Chapter 1, Introduction. Software Variations: Clarified which features and functions are supported for the different basecode part numbers.

Hardware Features: Simplified the description of the supply and waste bottles provided with the instrument. Optional Accessories: Added information on Liquid Level Alert option.

Chapter 2, Installation. Connecting Tubes and Bottles: Added information on the Liquid Level Alert option components. Added Liquid

Alert port to drawings. Added instructions for connecting the sensor cables to the select box, and the select box to the washer.

Chapter 3, Operation. Throughout: Added information on the ‘First Strip’ run-time parameter. Updated Menu Maps. General Usage Guidelines:

Added section for the Liquid Level Alert option. Linking Programs: Added troubleshooting tip for resolving Link program creation issues. Setup:

Added information on new SENSOR option.

Chapter 4, Maintenance/Performance Verification. Recommended

Maintenance and Performance Verification Schedules: Divided the former

Maintenance AND Performance Verification Schedule table into two separate tables. Decontamination: Added warning not to prepare a stronger-than-recommended bleach solution.

Appendix A, Error Codes. General Errors: Added error codes 2800 and

2900, related to the Liquid Level Alert option.

General. Added information on the optional vacuum pump muffler for

ELx50 washers equipped with the new vacuum pump (washers manufactured after January 2008). Additional changes are described below.

Preface. Updated Directives in the CE Mark section and added

Unspecified Use Warning to Hazards and Precautions.

Chapter 1, Introduction. Added the optional vacuum pump muffler to the

Hardware Features and Accessories Included sections. Updated the

Performance Specifications section for the 96-well: changed residual volume specs from ≤ 4.0 µl to ≤ 2.0 µl per well, and dispense precision specs from ≤ 4.0% CV to ≤ 3.0% CV.

Chapter 2, Installation. Updated unpacking/packing artwork and instructions to reflect new packaging materials, and added a graphic of the vacuum pump muffler to illustrations of the washer’s rear panel.

Chapter 4, Maintenance and Performance Verification (IQ/OQ/PQ). In the Evacuation Efficiency and Dispense Precision Tests and Worksheets for the 8-, 8s-, and 12-channel manifolds, changed the residual volume specs from ≤ 0.004 grams per well to ≤ 0.002 grams per well, and the dispense precision specs from ≤ 4.0% CV to ≤ 3.0% CV.

ELx50 Operator's Manual

xii

| Preface

Revision Date Changes

references to the vacuum pump muffler wherever references to the inline vacuum filter were present.

Appendix B, Manifold Variations. Removed instructions for returning washers to BioTek for reconfiguration to work with another manifold.

(The user can perform all the steps necessary to reconfigure the washer for a different manifold.)

Preface: Updated Service/TAC fax number. Added cautions for

Electromagnetic Environment and Compatibility. Updated Directives.

Ch. 1, Introduction: Under ‘Accessories Included’ updated the part numbers for the styluses. Updated Service/TAC fax number.

Ch. 2, Installation: Under ‘Liquid Level Alert’ added instructions for enabling the option via the main menu option UTIL.

Ch. 3, Operation: Correct the default Sensor detection setting for the

Liquid Level Alert™ option (it is disabled by default).

Ch. 4, Maintenance and Performance Verification (IQ/OQ/PQ): Added a note to the Recommended Maintenance Schedule recommending that the user develop a maintenance schedule based on the characteristics of the fluids used and the washer’s activity level. Added a note to the

Recommended Verification Schedule to state that the risk and performance factors associated with the user’s assays may require that some or all of the procedures be performed more frequently than presented in the schedule. For the Evacuation Efficiency liquid test procedures, added text to state that the user’s own wash program or one of the ELx50’s onboard Wash programs can be used. In the Liquid

Test Worksheets, for the Mean Residual Weight (Evacuation Efficiency) and Mean Absorbance (Dispense Precision) calculations, changed references to ‘96’ and ‘384’ wells to ‘number of wells processed.’

BioTek Instruments, Inc.

Intended Use Statement |

xiii

Intended Use Statement

• The ELx50™ Microplate Strip Washer provides microplate priming and washing for ELISA™, fluorescence and chemiluminescence immunoassays, cellular and agglutination assays.

• The intended use of this instrument is dependent on the instrument’s rear panel label. If there is an IVD label, then the instrument may be used for clinical, research and development, or other non-clinical purposes. If there is no such label, then the instrument may

only

be used for research and development, or for other non-clinical purposes.

Quality Control

It is considered good laboratory practice to run laboratory samples according to instructions and specific recommendations included in the assay package insert for the test to be conducted. Failure to conduct Quality Control checks could result in erroneous test data.

Repackaging and Shipping

If you need to ship the instrument to BioTek for service or repair, contact BioTek for a

Return Materials Authorization (RMA)

number, and be sure to use the original packing. Other forms of commercially available packing are not recommended and can void the warranty. If the original packing materials have been damaged or lost, contact BioTek for replacement packing.

Warnings

Operate the instrument on a flat surface and away from excessive humidity.

When operated in a safe environment according to the instructions in this document, there are no known hazards associated with the ELx50™. However, the operator should be aware of certain situations that could result in serious injury; these may vary depending on the instrument model.

ELx50 Operator's Manual

xiv

| Preface

Hazards and Precautions

Hazards

Warning! Power Rating.

The ELx50 Microplate Strip Washer’s power supply must be connected to a power receptacle that provides voltage and current within the specified rating for the system. Use of an incompatible power receptacle may produce electrical shock and fire hazards.

Warning! Electrical Grounding.

Never use a two-prong plug adapter to connect primary power to the ELx50 power supply. Use of a two-prong adapter disconnects the utility ground, creating a severe shock hazard. Always connect the power cord directly to a three-prong receptacle with a functional ground.

Warning! Internal Voltage.

Always turn off the power switch and unplug the power supply before cleaning the outer surface of the instrument.

Warning! Liquids.

Avoid spilling liquids on the instrument; fluid seepage into internal components creates a potential shock hazard. Wipe up all spills immediately. Do not operate the instrument if internal components have been exposed to fluid.

Warning! Potential Biohazards.

Some assays or specimens may pose a biohazard. Adequate safety precautions should be taken as outlined in the assay’s package insert. Always wear safety glasses and appropriate protective equipment, such as chemically resistant rubber gloves and apron.

Warning! Unspecified Use.

Failure to operate this equipment according to the guidelines and safeguards specified in this manual could result in a hazardous condition.

Warning! Software Quality Control.

The operator must follow the manufacturer’s assay package insert when modifying software parameters and establishing wash methods, using the washer’s onboard software.

Warning! Pinch Hazard.

Do not reach under the instrument during operation, as the syringe pump may pinch your fingers.

Precautions

The following precautions are provided to help avoid damage to the instrument:

Caution: Service.

The washer should be serviced by BioTek authorized service personnel. Only qualified technical personnel should perform troubleshooting and service procedures on internal components.

BioTek Instruments, Inc.

Hazards and Precautions |

xv

Caution: Environmental Conditions.

Do not expose the instrument to temperature extremes. For proper operation, ambient temperatures should remain between 15°-40°C (59°-104°F). Performance may be adversely affected if temperatures fluctuate above or below this range. Storage temperature limits are broader (see

Storing the ELx50

in

Chapter 4

).

Caution: Sodium Hypochlorite.

Do not expose any part of the instrument to the recommended diluted sodium hypochlorite solution (bleach) for more than

20 minutes. Prolonged contact may damage the instrument surfaces. Be certain to rinse and thoroughly wipe all surfaces.

Caution: Chemical Compatibility - Washers.

Some chemicals may cause irreparable damage to washers. The following chemicals have been deemed safe for use in washers: buffer solutions (such as PBS), saline, surfactants, deionized water, 70% ethyl, isopropyl, or methyl alcohol, 40% formaldehyde, and 20% sodium hydroxide. Never use acetic acid, DMSO, or organic solvents. Other chemicals may cause severe damage to the instrument. Contact BioTek prior to using any other chemicals. Refer to

Appendix C

,

Chemical Compatibility

.

Caution: Washers and Wash Buffer Solution.

Although many precautions have been taken to ensure that the instrument is as corrosion-proof as possible, the washer is not sealed and liquids can seep into sensitive components. Make sure that any spilled wash buffer solution is wiped off the washer. Prolonged exposure to salt solution may corrode parts of the carrier, movement rail, springs, and other hardware.

Caution: Bovine Serum Albumin.

Solutions containing proteins, such as bovine serum albumin (BSA), will compromise the washer’s performance over time unless a strict maintenance protocol is adhered to. See

Chapter 4

,

Maintenance and Performance Verification (IQ/OQ/PQ)

for cleaning instructions regarding BSA.

Caution: Warranty.

Failure to follow preventive maintenance protocols may

void the warranty

. See

Chapter 4

.

Caution: Disposal.

This instrument contains printed circuit boards and wiring with lead solder. Dispose of the instrument according to Directive 2002/96/EC,

“on waste electrical and electronic equipment (WEEE).”

Caution: Electromagnetic Environment.

Per IEC 61326-2-6 it is the user’s responsibility to ensure that a compatible electromagnetic environment for this instrument is provided and maintained in order that the device will perform as intended.

Caution: Electromagnetic Compatibility.

Do not use this device in close proximity to sources of strong electromagnetic radiation (e.g. unshielded intentional RF sources), as these may interfere with the proper operation.

ELx50 Operator's Manual

xvi

| Preface

Based on the testing described below and information contained herein, this instrument bears the CE mark.

Directive 2004/108/EC: Electromagnetic Compatibility

Emissions - CLASS A

EN 50081-1 and IEC 61326-1

EN 55022 Class A

This equipment has been designed and tested to CISPR 11 Class A. In a domestic environment it may cause radio interference, in which case, you may need to mitigate the interference.

Immunity

EN 50082-1 and IEC 61326-1

IEC 1000-4-2 Electrostatic Discharge

IEC 1000-4-3 Radiated EM Fields

IEC 1000-4-4 Electrical Fast Transient/Burst

IEC 1000-4-5 Surge Immunity

EN 61000-4-6 Conducted Disturbances

EN 61000-4-11 Voltage Dips, Short Interruptions and Variations

Directive 73/23/EEC Low Voltage

EN 61010-1:

“Safety requirement for electrical equipment for measurement, control and laboratory use. Part 1, General requirements.”

Directive 2002/96/EC: Waste Electrical and Electronic Equipment

Disposal Notice.

This instrument contains printed circuit boards and wiring with lead solder. Dispose of the instrument according to Directive 2002/96/EC, “on waste electrical and electronic equipment (WEEE).”

Directive 98/79/EC: In Vitro Diagnostics (some models)

• Product registration with competent authorities

• Traceability to the U.S. National Institute of Standards and Technology (NIST):

Dispense precision and average residual volume is traceable to NIST.

BioTek Instruments, Inc.

Electromagnetic Interference and Susceptibility |

xvii

Electromagnetic Interference and Susceptibility

USA FCC CLASS A

Warning:

Changes or modifications to this unit not expressly approved by the manufacturer could

void the user's authority

to operate the equipment.

This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules.

These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. Like all similar equipment, this equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause interference, in which case the user will be required to correct the interference at his own expense.

Canadian Department of Communications Class A

This digital apparatus does not exceed Class A limits for radio emissions from digital apparatus set out in the Radio Interference Regulations of the Canadian Department of

Communications.

Le present appareil numerique n'met pas du bruits radioelectriques depassant les limites applicables aux appareils numerique de la Class A prescrites dans le Reglement sur le brouillage radioelectrique edicte par le ministere des Communications du Canada.

User Safety

This device has been type tested by an independent laboratory and found to meet the requirements of the following:

North America

Canadian Standards Association CAN/CSA C22.2 No. 61010-1-04

“Safety Requirements for Electrical Equipment for Measurement, Control and

Laboratory Use, Part 1: General Requirements.”

Underwriters Laboratories UL 61010-1:2004

“Safety Requirements for Electrical Equipment for Measurement, Control and

Laboratory Use, Part 1: General Requirements.”

International

EN 61010-1

“Safety Requirements for Electrical Equipment for Measurement, Control and

Laboratory Use, Part 1: General Requirements.”

ELx50 Operator's Manual

xviii

| Preface

Safety Symbols

Alternating current

Courant alternatif

Wechselstrom

Corriente alterna

Corrente alternata

Direct current

Courant continu

Gleichstrom

Corriente continua

Corrente continua

Both direct and alternating current

Courant continu et courant alternatif

Gleich - und Wechselstrom

Corriente continua y corriente alterna

Corrente continua e corrente alternata

Earth ground terminal

Borne de terre

Erde (Betriebserde)

Borne de tierra

Terra (di funzionamento)

Protective conductor terminal

Borne de terre de protection

Schutzleiteranschluss

Borne de tierra de protección

Terra di protezione

On (Supply)

Marche (alimentation)

Ein (Verbindung mit dem Netz)

Conectado

Chiuso

Off (Supply)

Arrêt (alimentation)

Aus (Trennung vom Netz)

Desconectado

Aperto (sconnessione dalla rete di alimentazione)

BioTek Instruments, Inc.

Safety Symbols |

xix

Caution (refer to accompanying documents)

Attention (voir documents d’accompanement)

Achtung siehe Begleitpapiere

Atención (vease los documentos incluidos)

Attenzione, consultare la doc annessa

Warning, risk of electric shock

Attention, risque de choc électrique

Gefährliche elektrische Schlag

Precaución, riesgo de sacudida eléctrica

Attenzione, rischio di scossa elettrica

Warning, risk of crushing or pinching

Attention, risque d’écrasement et pincement

Warnen, Gefahr des Zerquetschens und Klemmen

Precaución, riesgo del machacamiento y sejeción

Attenzione, rischio di schiacciare ed intrappolarsi

Warning, hot surface

Attention, surface chaude

Warnen, heiße Oberfläche

Precaución, superficie caliente

Attenzione, superficie calda

Consult instructions for use

Consulter la notice d’emploi

Gebrauchsanweisung beachten

Consultar las instrucciones de uso

Consultare le istruzioni per uso

In vitro diagnostic medical device

Dispositif médical de diagnostic in vitro

Medizinisches In-Vitro-Diagnostikum

Dispositivo médico de diagnóstico in vitro

Dispositivo medico diagnostico in vitro

Separate collection for electrical and electronic equipment

Les équipements électriques et électroniques font l’objet d’une collecte sélective

Getrennte Sammlung von Elektro- und Elektronikgeräten

Recogida selectiva de aparatos eléctricos y electrónicos

Raccolta separata delle apparecchiature elettriche ed elettroniche

ELx50 Operator's Manual

xx

| Preface

BioTek Instruments, Inc.

Chapter 1

Introduction

This chapter introduces the ELx50™ Microplate Strip Washer, and describes its hardware and software features and technical specifications. Instructions on how to contact BioTek for Product

Support & Service are included on page 12.

Introducing the ELx50 Microplate Strip Washer ............................. 2

ELx50 Washer Variations ........................................................... 3

Variations ............................................................... 3

Patented Dual-Manifold Design ............................................ 4

Variations ................................................................. 5

Software

Hardware Features.................................................................... 7

Accessories Included ................................................................. 8

Optional Accessories ................................................................. 9

Specifications ..........................................................................10

Microplates.........................................................................10

Electrical ............................................................................10

Physical .............................................................................10

Environmental ....................................................................10

Hardware ...........................................................................10

Specifications ...................................................11

Product Support & Service ........................................................12

Contacting the Technical Assistance Center .............................12

Returning Instruments for Service/Repair ...............................12

2

| Chapter 1: Introduction

Introducing the ELx50 Microplate Strip Washer

Fast, versatile, and easy to use, the ELx50™ is a user-programmable, fully automated microplate strip washer. Its key features include the following:

• The precision syringe pump with its long-lasting seal ensures precise fluid delivery, as well as reproducibility from wash to wash.

• Programmable dispense volumes, flow, and height, and aspiration rates and depths provide for a wide range of washing capabilities, from gentle washing for cellular assays to vigorous washing for ELISA™.

• The optional bottom washing routine can be applied to lower the background absorbance. The optional crosswise aspiration routine can be applied to reduce residual volumes.

• The ELx50 can be used to wash all microplate-based assays, including ELISA, fluorescence, chemiluminescence, RIA, DNA probes, and cellular assays.

• The intuitive menu-driven software allows you to create and store up to 75 washer programs. When you are ready to run a program, simply select the program name and follow the screen prompts. The ELx50 will then process the plate according to the program parameters.

• All ELx50 models support Wash, Prime, Dispense, and Aspirate programs. Some models also support Soak and Link programs.

• Several pre-defined programs are provided to simplify preventative maintenance, which should be performed regularly to ensure optimum washer performance.

• All ELx50 models can be used to wash standard 96-well microplates, and some models are also equipped to wash standard 384-well microplates.

• “V” versions of the ELx50 feature automatic wash buffer switching with three valves.

• The Liquid Level Alert™ option allows users the convenience of continuous monitoring for both supply and waste bottles. At the beginning and end of a wash protocol, the liquid level is verified to ensure sufficient wash buffer remains in the designated dispense bottle to complete a microplate wash. Sufficient storage capacity in the waste bottle is also verified. Each Liquid Level Alert kit includes a quantity of supply and waste bottles with level-sensing capability. Up to three different supply bottles can be connected to the washer through the Liquid Level

Alert Select Box.

• An aspiration carrier is provided with the ELx50/8FMW model for bottom aspiration of filter-bottom microplates. A vent port on the carrier and two selectable vent plugs for altering the diameter of the port provide three vacuum levels: low, medium, and high. Plates with filter pore sizes from 0.45 μm to 1.2 μm are supported.

BioTek Instruments, Inc.

ELx50 Washer Variations

|

3

ELx50 Washer Variations

There are several different ELx50™ washer models available: the ELx50/8, ELx50/12,

ELx50/16, and ELx50/8FMW.

Each washer comes with one of four possible washer manifolds (8-, 8s-, 12-, or 16-channel positioned above the microplates), and “V” version washers include automatic buffer switching with three valves. All models support the Liquid Level Alert™ option.

The washer manifold, carrier, and software variations are described on the next few pages.

For additional information, contact BioTek Instruments (see

Product Support & Service

, page 12).

Manifold Variations

The

manifold

is the washer component positioned above the microplate that delivers fluid to, and removes it from, the microwells. Each ELx50 washer is equipped with a manifold containing 8, 12, or 16 sets of aspirate and dispense tubes, or “channels.” The following table shows the washing capabilities of each manifold:

With this manifold: You can wash:

8-channel

8s-channel

12-channel

16-channel

8-well strips, 96-well plates

8-well strips, 96-well plates

12-well strips, 96-well plates

8-well strips, 96-well plates, 384-well plates

The ELx50 supports the washing of standard flat, round, and ‘V’ bottom microwells; the ELx50/8FMW model also supports bottom aspiration of filter-bottom plates.

The ELx50 can be reconfigured to support a different manifold. See

Appendix B

,

ELx50 Manifold Variations

for more information.

The 8s-channel manifold has shorter dispense tubes for special applications.

The 16-channel manifold uses two aspirate tubes for each well when used with 8-well strips and 96-well plates. For these formats, it may be difficult for the tubes to reach the very bottom of round or ‘V’ bottom wells, possibly leading to higher residuals.

The ELx50/8FMW model does not support the 16-channel manifold.

ELx50 Operator's Manual

4

| Chapter 1: Introduction

Patented Dual-Manifold Design

The patented Dual-Action™ 16-channel manifold has 16 sets of dispense and aspirate tubes arranged in a fixed horizontal position relative to each other but located in two separate manifolds (see

Figure 1

below). The two manifolds move independently of each other on the vertical axis. This dual-manifold design overcomes the problem of fitting both a dispense and an aspirate tube into the much smaller 384-well at the same time, yet it allows for overfill washing of the smaller 384 wells. The dispense tubes are held above the plate when the dispense manifold reaches a mechanical stop, while the aspirate tubes are able to continue descending into the wells.

Dispense manifold

Tubes are positioned for overfill washing

Aspirate manifold

Tubes are positioned

Figure 1: Patented Design of the 16-Channel Manifold

BioTek Instruments, Inc.

ELx50 Washer Variations

|

5

Carrier Variations

The

carrier

is the washer component that transports microplates and positions the wells underneath the manifold tubes. There are three possible carrier types, and their usage depends on the manifold type installed and the microplate type being used:

• Microplate carrier that supports 8-, 8s-, and 12-channel manifolds and solid-bottom plates.

• Microplate carrier that supports 16-, 8-, and 8s-channel manifolds and solid-bottom plates.

• “Aspiration carrier” that supports 8- and 8s-channel manifolds and 96-well filter-bottom plates. This carrier is used on ELx50/8FMW model only, and aspirates fluid from below the microplates, so that fluid is drawn down through the filters at the bottom of the wells.

The basecode software onboard the ELx50/8FMW has been pre-configured to use either the microplate carrier or the aspiration carrier. The aspirate-only program parameters ASPIRATE TYPE and BOT ASPIRATE TIME only appear if the washer model is the ELx50/8FMW, and the manifold is configured as either 8- or 8s-channel.

Software Variations

The ELx50™ washer is operated by onboard software, controllable through the washer’s keypad. This onboard software, referred to as the “basecode,” has an identifying

part number.

Your washer is installed with one of three possible ELx50 basecode part numbers:

4070230, 4070261, or 4070201. Each part number in turn has an identifying

version number.

If you need to contact BioTek Instruments for technical assistance, you will be asked for the part number and version of the basecode currently residing on your washer. See

Product Support & Service

, page 12, for instructions for retrieving this information from the washer.

The

4070230

basecode supports:

• Compatibility with the 8-, 8s-, 12, and 16-channel manifolds. The 16-channel manifold supports both 96- and 384-well plate washing.

• An easy to use, menu-driven interface.

• Ability to create Wash, Prime, Dispense, Aspirate, Soak, and Link programs for automated microplate washing.

• Storage of up to 75 washer programs.

• User-definable washer program names, for rapid recall at program run time.

• Configurable program parameters such as variable soaking, multispeed shaking, bottom washing, and crosswise aspiration.

• Pre-defined washer programs that can be copied and used as the basis for new programs.

ELx50 Operator's Manual

6

| Chapter 1: Introduction

• Pre-defined maintenance programs, including AutoPrime, used to keep the manifold tubes moist and free from clogging.

• Support for the following languages: English, French, German, Italian, and

Spanish.

Note: Contact BioTek for information on changing to a different language.

• Compatibility with the ELx50/8FMW washer’s aspiration carrier, for bottom aspiration of 96-well filter-bottom plates. Note: The 16-channel manifold cannot be used with the aspiration carrier.

• Compatibility with the Liquid Level Alert™ option.

The

4070261

basecode supports the above features, but it is not compatible with the

Liquid Level Alert option and does not support the 16-channel manifold with the

ELx50/8FMW washer’s aspiration carrier.

The

4070201

basecode is not compatible with the 16-channel manifold, the Liquid

Level Alert option, or the ELx50/8FMW washer’s bottom aspiration carrier. This basecode does not provide the ability to create Soak or Link programs.

BioTek Instruments, Inc.

Hardware Features

|

7

Hardware Features

• 2-line x 24-character LCD display.

• Membrane keypad with 25 alphanumeric keys.

• Washes:

96-well microplates with 0.355” / 9 mm well centers

384-well microplates with 0.177” / 4.5 mm well centers (washer model-dependent)

96-well filter-bottom microplates (ELx50™/8FMW model only)

1 x 8 or 1 x 12 microwell strips

• Crosswise aspiration for optimizing washing performance.

• External 24-volt power supply.

• One serial RS-232 COM port (25-pin male connector).

• Supply and waste systems appropriate for the model/purchased options.

• Internal positive displacement pump provides controllable flow rates and volumes to make washing cellular assays and immunoassays equally convenient. Settings are adjustable via the onboard software, for low to high velocity dispensing.

• Internal vacuum pump aspirates liquid from the microwells into the waste collection vacuum vessel.

• Stepper motors provide accurate and repeatable carrier and manifold positioning.

• Accommodates 1 x 8, 1 x 12, or 1 x 16 manifolds. See

Appendix B

,

ELx50 Manifold

Variations

for more information.

• Flat, round and ‘V’ bottomed strips and plates are washed.

• Carrier and manifold are easily removed for cleaning.

• Priming/soaking trough built into the carrier.

• Five shaking speeds for more intense washing.

• Programmable aspiration depth: Ensures that the tubes do not touch the well bottom, allowing the handling of sensitive cell assays.

• ELx50/8FMW model: one vent port (0.047” in diameter) on the aspiration carrier and two selectable vent plugs (with 0.032” and 0.020” diameter holes) for varying the amount of vacuum used with the aspiration carrier.

• Mist shield protects the microplate from contaminants.

ELx50 Operator's Manual

8

| Chapter 1: Introduction

Accessories Included

• 24 VDC power supply (PN 61062)

• Power cord (part number varies according to country of use)

• 8-, 8s-, 12-, or 16-channel manifold

• Microplate carrier (specific to manifold type)

• Aspiration carrier (ELx50™/8FMW model only) with two vent plugs:

One plug with 0.032” diameter hole (PN 4072100)

One plug with 0.020” diameter hole (PN 4072099)

• Mist shield (PN 4070517)

• Two 2-liter supply bottles (three with “V” version washers) (PN 4070515)

• One 2-liter waste bottle (PN 48140)

• Waste tubing set:

All models except ELx50/8FMW (PN 4070511)

ELx50/8FMW model only (PN 4073002)

• Supply tubing set (PN 4070510)

• Optional in-line vacuum filter (PN 48146)

• Optional vacuum pump muffler (PN 4073009) for ELx50 washers that are equipped with the new vacuum pump (washers manufactured after January 2008)

• Aspirate tube stylus (PN 4072042) and Dispense tube stylus (PN 4072043)

• Operator’s Manual (PN 4071000)

• Packing instructions (PN 4071015)

• Set of shipping materials (PN 4073001)

• Shipping document (PN 94075) that includes a Warranty Statement and Certificate of Compliance and Calibration

• Declaration of Conformity (PN 4071006 for clinical models or PN 4071034 for non-clinical models)

Part numbers are subject to change over time. Please contact BioTek

Customer Care if you have any questions.

BioTek Instruments, Inc.

Optional Accessories

|

9

Optional Accessories

• 8-channel manifold (PN 4070512)

• 8s-channel manifold (PN 4070519)

• 12-channel manifold (PN 4070513)

• 16-channel manifold (PN 4070527)

• Liquid Level Alert™ option:

PN 4070035, for non-“V” version washers: Liquid Level Alert select box with

2-liter supply bottle (1) and 2-liter waste bottle (1), with sensors

PN 4070036, for “V” version washers: Liquid Level Alert select box with

2-liter supply bottles (3) and 2-liter waste bottle (1), with sensors

• Spare vent plugs for aspiration carrier (ELx50™/8FMW model):

One plug with 0.032” diameter hole (PN 4072100)

One plug with 0.020” diameter hole (PN 4072099)

• Millipore 96-well filter bottom microplates:

0.45 μm filter pore size (Millipore PN MSHVN4550)

1.2 μm filter pore size (Millipore PN’s MSBVN1B50 or MSBVN1210)

• RS-232 serial cable (PN 75053)

• In-line vacuum filter replacement (PN 48146)

• BioTek blue test dye solution (PN 7773001) and wetting agent solution

(PN 7773002) for liquid testing

• Installation-Operational-Performance Qualification (IQ-OQ-PQ) package

(PN 4070530)

• ELx50 Service Manual (PN 4071008)

Part numbers are subject to change over time. Please contact BioTek

Customer Care if you have any questions.

ELx50 Operator's Manual

10

| Chapter 1: Introduction

Specifications

Microplates

• Microplates:

96-well, 384-well solid-bottom plates and plates,

1.2 μm filter pore size (ELx50™/8FMW)

• Microstrips:

• Microwells:

1 x 8, 1 x 12

Flat, round, or “V” bottom

384-well support is available with the 16-channel manifold only.

Electrical

• Voltage range:

100 to 250 V~, 50 to 60 Hz

• Power Consumption:

40 W

Physical

• Dimensions:

• Weight:

16 in. D x 14 in. W x 6.5 in. H

(40.6 cm x 35.6 cm x 16.5 cm)

≤ 20 lbs. (8.9 kg) all models except ELx50/8FMW;

≥ 22 lbs. (9 kg) ELx50/8FMW model

Environmental

• Operating conditions: 15° - 40°C (59° - 104°F)

• Relative humidity:

10% - 85% (non-condensing)

Hardware

• Manifold type:

8-, 8s-, 12-, or 16-channel manifold, depending on washer model

• Waste bottle volume:

2 liters

• Supply bottle volume: 2 liters

• User interface (LCD):

2 x 24 LCD display; 25 alphanumeric soft keys

BioTek Instruments, Inc.

Specifications

|

11

Performance Specifications

Residual Volume (Evacuation Efficiency)

96-Well Performance: Average residual of < 2.0 µl per well after a 3-cycle wash, when dispensing 300 µl per well, into a minimum of 4 strips, using a solution of deionized water with 0.1% Tween

®

20 or equivalent buffer solution. The aspirate height adjustment should be optimized for the plate prior to testing.

384-Well Performance:

(16-channel only)

Average residual of < 4.0 µl per well after a 1-cycle wash, when dispensing 100 µl per well, into a minimum of 4 strips, using a solution of deionized water with 0.1% Tween 20 or equivalent buffer solution. The aspirate height adjustment should be optimized for the plate prior to testing.

Residual Volume specifications do not apply to the ELx50™/8FMW washer when using the aspiration carrier for filter-bottom plates. Residual Volume specifications are only applicable when using the microplate carrier for solid-bottom plates.

Dispense Precision

96-Well Performance: < 3.0% CV when measured over 6 strips from an 8- or

8s-channel manifold, or 4 strips from a 12- channel manifold, dispensing 300 µl per well of deionized water with 0.1% Tween

20 and colored dye at a rate of 300 µl per well per second.

384-Well Performance:

(16-channel only)

< 4.0% CV when measured over 6 strips from a 16-channel manifold, dispensing 100 µl per well of deionized water with

0.1% Tween 20 and colored dye.

Processing Speed

96-Well Performance < 130 seconds for 12 strips (3 cycles, 300 µl/well, no soak)

384-Well Performance

(16-channel only )

< 260 seconds for 24 strips (3 cycles, 100 µl/well, no soak)

ELx50 Operator's Manual

12

| Chapter 1: Introduction

Product Support & Service

All of BioTek’s products are backed by a superior support staff. If your instrument ever fails to function perfectly, if you have questions about how to use or maintain it, or if you need to send the instrument to BioTek for service or repair, please contact our Technical

Assistance Center (TAC).

Contacting the Technical Assistance Center

Our Technical Assistance Center is open from 8:30 AM to 5:30 PM (EST), Monday through Friday, excluding standard U.S. holidays. You can send a fax or an e-mail any time.

Phone:

800-242-4685 (in the U.S.) or 802-655-4740 (outside the U.S.)

Fax:

802-654-0638

E-Mail:

[email protected]

Please be prepared to provide the following information:

• Your name and company information

• A daytime phone or fax number, and/or an e-mail address

• The product name, model, and serial number

• The software part number and basecode version (available via the washer’s main menu by selecting

UTIL TESTS CHKSUM

)

• For troubleshooting assistance or instruments needing repair: the specific steps that produce your problem, and any error codes displayed (see also

Appendix A, Error Codes

).

Returning Instruments for Service/Repair

If you need to return an instrument to BioTek for service or repair, please contact the

TAC for a Return Materials Authorization (RMA) number before shipping the instrument. Repackage the instrument properly (see

Chapter 2

,

Installation

), write the RMA number on the shipping box, and ship to this address:

BioTek Instruments, Inc.

Technical Assistance Center

100 Tigan Street

Highland Park

Winooski, Vermont 05404 USA

BioTek Instruments, Inc.

Chapter 2

Installation

This chapter includes instructions for unpacking and setting up the

ELx50™ Microplate Strip Washer, installing the washer's components, and repackaging the instrument for shipment.

Unpacking and Inspecting the Instrument ...................................14

Setting Up the ELx50 ...............................................................20

Operating Environment ........................................................20

Installing the Carrier and Manifold .........................................20

Connecting the Tubes and Bottles ..........................................25

Attaching the Mist Shield ......................................................35

Connecting the Power Supply and Cord ..................................36

Serial Port for Communication With Other Devices ...................36

Verifying Performance ..............................................................37

Repackaging and Shipping ........................................................38

Before Repackaging the Instrument .......................................38

Repackaging the ELx50 and Its Accessories.............................39

14

| Chapter 2: Installation

Unpacking and Inspecting the Instrument

Important! Save all packaging materials.

If the washer is shipped to the factory for repair or replacement, it must be carefully repackaged, according to the instructions on pp. 38 to 49, using the original packing materials. Using other forms of commercially available packing materials, or failure to follow the repackaging instructions may

void your warranty

. If the original packing materials have been damaged, contact BioTek and ask for part number

4073001

.

The ELx50™ washer and its accessories are securely packaged inside custom-designed shipping materials. This packaging should protect the instrument from damage during shipping. Inspect the shipping box, packaging, instrument, and accessories for signs of damage.

If the washer is damaged:

Notify the carrier and your manufacturer's representative.

Keep the shipping cartons and packing material for the carrier's inspection. The manufacturer will arrange for repair or replacement of your instrument immediately, before the shipping-related claim is settled.

Refer to the unpacking instructions and

Figures 2

through

6

on the following pages when removing the instrument and its accessories from the shipping container.

See

Repackaging and Shipping

at the end of this chapter for complete shipping instructions.

Unpacking the 16-Channel Manifold, Mist Shield, and Bottles

1.

2.

3.

Carefully open the outer shipping container and lift the top packing container and shipping tray that contain the accessories out of the container.

Remove the 16-channel manifold, mist shield, and waste and supply bottles, and then set aside their shipping materials.

Refer to

Figure 3

on the following page for unpacking the shipping tray for the

8-, 8s-, or 12-channel manifold, microplate carrier, power supply, tubing, in-line filter, and other accessories.

The bottles include two 2-liter supply bottles and one 2-liter waste bottle. The “V” version of the ELx50™ (washers equipped with the optional valve module) will include three 2-liter supply bottles.

BioTek Instruments, Inc.

Unpacking and Inspecting the Instrument

|

15

Figure 2: Unpacking the 16-Channel Manifold. Mist Shield, and Bottles

Unpacking the 8-, 8s-, or 12-Channel Manifold, Microplate Carrier,

Power Supply, Tubing, In-Line Filter, and Optional Pump Muffler

1.

2.

Carefully open the shipping tray for the 8-, 8s-, or 12-channel manifold, microplate carrier, power supply and power cord, tubing, in-line vacuum filter, optional vacuum pump muffler, manual, and Declaration of Conformity or other documents.

Remove the accessories listed above and set aside the shipping materials.

ELx50 Operator's Manual

16

| Chapter 2: Installation

Figure 3: Unpacking the 8-, 8s-, or 12-Channel Manifold, Microplate Carrier, Power Supply,

Tubing, In-Line Vacuum Filter, and Optional Vacuum Pump Muffler

BioTek Instruments, Inc.

Unpacking and Inspecting the Instrument

|

17

Unpacking the Inner Shipping Container and Foam Corner Blocks

1.

2.

3.

Lift the inner shipping container with the eight foam corner blocks out of the outer shipping container as shown in the figure below and place it upon a level surface. This package contains the washer.

ELx50™/8FMW washers

include an aspiration carrier and two vent plugs.

Remove and set aside the foam corner blocks.

Carefully open the inner shipping container, using a blunt tool.

Figure 4: Unpacking the Inner Shipping Container and Foam Corner Blocks

ELx50 Operator's Manual

18

| Chapter 2: Installation

Unpacking the Aspiration Carrier (ELx50/8FMW models only)

1.

2.

Remove the aspiration carrier from its storage compartment in the inner shipping container.

Remove and set aside any bubble wrap, plastic bag, or other shipping materials.

Figure 5: Unpacking the Aspiration Carrier (ELx50™/8FMW)

BioTek Instruments, Inc.

Unpacking and Inspecting the Instrument

|

19

Unpacking the ELx50™ Washer

1.

2.

3.

4.

5.

Lift insert #2, then insert #1 out of the inner shipping container (as shown below).

Carefully lift the washer out of the inner shipping container and place it on a level surface.

Remove the washer from the plastic bag.

Slide the foam shipping block out from underneath the manifold mounting bracket.

Store the inner shipping container and all shipping materials (including those noted on the previous pages) in the outer shipping container.

ELx50 Operator's Manual

Figure 6: Unpacking the ELx50 Washer

20

| Chapter 2: Installation

Setting Up the ELx50

Important!

Avoid

excessive humidity.

Condensation directly on the sensitive electronic circuits can cause the instrument to fail internal self checks.

The manifold dispense tubes have a protective Teflon collar at the tip. This is to prevent dripping.

Do not remove these coverings!

Operating Environment

The ELx50™ washer is sensitive to extreme environmental conditions. For optimal operation, install the washer

• on a level surface,

• in an area where ambient temperatures between 15°C (59°F) and 40°C

(104°F) can be maintained, and

• away from excess humidity.

See

Specifications

in

Chapter 1

,

Introduction

, for a list of environmental specifications.

Installing the Carrier and Manifold

All ELx50 washers have a microplate carrier that is specific to their manifold type.

See

ELx50 Washer Variations

in

Chapter 1

,

Introduction

, for information on the various manifold and carrier types.

Refer to

Figures 7

and

8

when installing the microplate carrier and manifold.

ELx50/8FMW: Refer to

Figures 9

and

10

when installing the aspiration carrier.

Microplate Carrier

1.

2.

3.

Hold the microplate carrier over the rail so that the priming/soaking trough is closest to the rear of the washer.

Line up the pin on the underside of the carrier with the slot on the carrier transport.

Snap the two carrier rail guides onto the carrier transport rail. The pin should sit in the slot.

BioTek Instruments, Inc.

Setting Up the ELx50

|

21

8-, 8s-, or 12-Channel Manifold

1.

2.

3.

Orient the manifold so that the aspirate/dispense tubes are pointed down, and the fittings for the supply and waste tubing are pointed toward the back of the washer.

Slide the manifold gently down over the two support pins on the manifold mounting bracket that are closest to the front of the washer.

Insert the thumbscrews (2). Do not overtighten.

Figure 7: Installing the Microplate Carrier and 8-, 8s, or 12-Channel Manifold

ELx50 Operator's Manual

22

| Chapter 2: Installation

16-Channel Manifold

1.

2.

3.

Orient the manifold so that the aspirate/dispense tubes are pointed down, and the fittings for the supply and waste tubing are pointed toward the back of the washer.

Slide the manifold gently down over the three support pins on the manifold mounting bracket.

Insert the thumbscrews (3). Do not overtighten.

Figure 8: Installing the Microplate Carrier and 16-Channel Manifold

BioTek Instruments, Inc.

Setting Up the ELx50

|

23

Aspiration Carrier (ELx50™/8FMW)

For bottom aspiration using the aspiration carrier, the 8- or 8schannel manifold must be installed. The aspiration carrier does not support the 12-channel manifold.

1.

2.

3.

4.

Hold the aspiration carrier over the rail so that the priming/soaking trough is closest to the rear of the washer.

Line up the pin on the underside of the carrier with the slot on the carrier transport.

Snap the two carrier rail guides onto the carrier transport rail. The pin should sit in the slot.

Follow the instructions on the next page for setting the vacuum levels using the vent port and vent plugs.

Figure 9: Installing the Aspiration Carrier (ELx50/8FMW model)

ELx50 Operator's Manual

24

| Chapter 2: Installation

5.

6.

Slide the latch on the plate clamp to the right to open the clamp, whenever inserting or removing a plate.

Use the vent port (labeled “C”) on the carrier and the two vent plugs

(located in storage slots “A” and “B”) to vary the vacuum levels. The vent port is a 0.047” diameter hole, which is equal to approximately -64 mmHg.

The gray vent plug contains a 0.032”diameter hole, which is equal to approximately -116 mmHg. The black vent plug contains a 0.020” diameter hole, which is equal to approximately -240 mmHg.

For a

low

vacuum level, leave the vent port on the carrier

open

.

For a

medium

vacuum level, insert the

gray

vent plug into the vent port.

For a

high

vacuum level, insert the

black

vent plug into the vent port.

The approximate vacuum levels listed above (-64, -116, and

-240 mmHg) are “typical” only, and may vary, depending upon factors such as relative humidity, barometric pressure, and mechanical tolerances.

Figure 10: Close-Up View of Aspiration Carrier Vent Port and Vent Plugs

BioTek Instruments, Inc.

Setting Up the ELx50

|

25

Connecting the Tubes and Bottles

Caution.

The washer manifold, the vacuum port, and the supply bottles have Luer fittings.

Finger-tighten only!!

Important!

Review

Figures 11

through

17

on the following pages, as well as the diagram on the back of the washer.

Identify the different tubes and bottles that came with your washer.

For optimal operation of the ELx50™ washer, all tubing, cables, and fittings for the waste and supply systems must be properly connected. The figures on the following pages illustrate the rear panel of the instrument and the locations of the ports and connections for the waste and supply systems.

Liquid Level Alert™

option: If you purchased the Liquid Level Alert option, its components are shipped in a separate box. Install the waste and supply bottle(s) as described on the following pages, and then install the select box and connect the cables as described on page 34.

• For PN 4070035 (non-“V” version washers), components include a select box, one 2-liter supply bottle, and one 2-liter waste bottle.

• For PN 4070036 (“V” version washers), components include a select box, three 2-liter supply bottles, and one 2-liter waste bottle.

Vacuum Pump Muffler:

An optional vacuum pump muffler (PN 4073009) is provided with ELx50 washers that are equipped with the new vacuum pump (washers manufactured after January 2008). If you wish to use the muffler, attach the female luer end of the muffler to the vacuum pump output/exhaust port on the rear panel of the washer. See

Figure 11b

and

Figure 12b

on the following pages for illustrations of the vacuum pump muffler.

Before Connecting the Tubes and Bottles

1.

2.

Rinse all bottles with deionized or distilled water before using. This rinsing eliminates particles that may have entered during packing or unpacking.

Place the supply and waste bottles on the same horizontal plane as the washer. This ensures optimum pump performance.

ELx50 Operator's Manual

26

| Chapter 2: Installation

Figure 11a: Washer Rear Panel (all models except ELx50™/8FMW)

Figure 11b: Attaching the Vacuum Pump Muffler (all models except ELx50/8FMW)

BioTek Instruments, Inc.

Setting Up the ELx50

|

27

Figure 12a: Washer Rear Panel (ELx50™/8FMW model)

Figure 12b: Attaching the Vacuum Pump Muffler (ELx50/8FMW model)

ELx50 Operator's Manual

28

| Chapter 2: Installation

Waste System

All models except the ELx50™/8FMW:

Follow steps 1 through 4 below for connecting the waste tubing and bottle. Refer to

Figure 11

on page 26 for an illustration of the rear panel of the washer, and to

Figure 13

on page 29 for an illustration of the waste tubing connections.

1.

2.

Attach the paired (Y-shaped) waste tubing to the two waste ports on the back of the washer manifold.

Finger-tighten only!

Remove the waste bottle cover and replace it with the stopper from the waste tubing set.

3.

4.

To install the optional* in-line vacuum filter, cut the tubing from the waste bottle to the vacuum pump output port in half and attach both cut ends to the vacuum filter ports.

Attach the free end of the tubing from the waste bottle to the vacuum port for waste fluid on the back of the washer.

Finger-tighten only!

ELx50/8FMW model (with aspiration carrier installed):

Follow steps 1 through 7 below for connecting the waste tubing and bottle. Refer to

Figure 12

on the previous page for an illustration of the rear panel of the washer, and to

Figure

14

on page 30 for an illustration of the waste tubing connections.

1.

2.

3.

4.

5.

6.

7.

Attach the paired (Y-shaped) waste tubing to the two waste ports on the back of the washer manifold.

Finger-tighten only!

Attach the free end of the waste tubing from the manifold to the top aspiration port (“TOP ASP”) on the back of the washer.

Finger-tighten only

!

Remove the waste bottle cover and replace it with the stopper from the waste tubing set.

Attach one tube from the waste bottle to the vacuum pump output port on the back of the washer.

To install the optional* in-line vacuum filter, cut the tubing from the waste bottle to the vacuum pump output port in half and attach both cut ends to the vacuum filter ports.

Attach the free end of the tubing from the waste bottle to the vacuum port for waste fluid on the back of the washer.

Finger-tighten only!

Attach the waste tubing from the aspiration carrier to the bottom aspiration port (“BOT ASP”) on the back of the washer.

Finger-tighten only!

The in-line vacuum filter is “optional” in the sense that the washer will function without it. It is strongly recommended that you install this filter, however, because (1) the filter prevents waste material aerosols from being released into the air, and (2) the filter serves as a temporary fluid barrier in the event that the waste bottle is allowed to overfill. (The replacement PN for the filter is 48146.)

BioTek Instruments, Inc.

Setting Up the ELx50

|

29

Figure 13: Waste Tubing Connections (all models except the ELx50™/8FMW)

Note: The washer's top shroud is removed from this illustration for clarity only. While it is acceptable to remove the mist shield, do not

attempt to remove the shroud protecting the washer's components.

ELx50 Operator's Manual

30

| Chapter 2: Installation

Figure 14: Waste Tubing Connections

(ELx50™/8FMW model with Aspiration Carrier installed)

BioTek Instruments, Inc.

Setting Up the ELx50

|

31

Supply System

Refer to

Figure 15

for the following instructions.

1. Two check valves are supplied; note the flow direction arrows. Connect the check valves to the syringe pump top and bottom ports. Compare the flow direction arrows with those shown in

Figure 11

(all models except the

ELx50™/8FMW) or

Figure 12

(ELx50/8FMW model).

2.

3.

4.

Two supply tubes are provided, one short, one long. Connect one end of the short tube to the manifold. Connect the other end to the check valve that is connected to the syringe pump top port.

Washers

without

the multiple valve tubing option: Connect one end of the long tube to the supply bottle and the other end to the check valve connected to the syringe pump bottom port.

Washers

with

the multiple valve tubing option: The long tube is longer than necessary and should be cut to approximately 5 inches when installing to minimize prime volume. Connect one end of the 5-inch tube to the valve output port on the rear panel of the washer and the other end to the check valve connected to the syringe pump bottom port.

Attach the three tubes from supply bottles A, B, and C to valve ports A, B, and C on the back of the washer.

Note: The ELx50/8FMW model does not support the multiple valve tubing option.

ELx50 Operator's Manual

32

| Chapter 2: Installation

Figure 15a: Supply Tubing Connections (front view)

Figure 15b: Supply Tubing Connections (view of connections on rear panel)

Note: The washer's top shroud is removed from this illustration for clarity only. While it is acceptable to remove the mist shield, do not

attempt to remove the shroud protecting the washer's components.

BioTek Instruments, Inc.

Setting Up the ELx50

|

33

Figure 16: Washer with Multiple Valve Tubing Connections

(“V” versions of the ELx50™)

ELx50 Operator's Manual

34

| Chapter 2: Installation

Liquid Level Alert™ Option

If you have not already done so, connect the waste and supply bottles as described on the previous pages.

1. The Liquid Level Alert

select box

has four ports, labeled A, B, and C, for the supply bottle(s), and WASTE.

• For washers with the Valve option, connect the supply bottle cables to the appropriate ports in the select box. If a port is left empty, insert one of the supplied “shorting plugs” (all ports must be filled).

• For washers without the Valve option, connect the supply bottle cable to port A and then insert the supplied shorting plugs into ports B and C.

• For all washers, connect the waste bottle cable to the WASTE port.

2. Connect the select box to the washer via the

LIQUID ALERT

port on the back of the washer.

Connects to washer

Note: After verifying performance (p. 37),

enable Liquid Level

Alert by selecting

UTIL > SETUP >

SENSOR > LIQUID

LEVEL ALERT? > YES.

Shorting plugs

Liquid

Alert port

Supply (A)

Waste

Figure 17: Liquid Level Alert select box (top), and cable connections

BioTek Instruments, Inc.

Setting Up the ELx50

|

35

Final Check

1.

2.

Verify that the tubing was not crimped during installation.

Make sure that the fluid supply and the waste output tubes are attached to the appropriate manifold ports. If the mounted manifold tilts forward or backward, make sure that the tubing to the manifold is not being pulled tight or off to the side.

Attaching the Mist Shield

Refer to

Figure 18

below for the following instructions.

1. Insert the two thumbscrews into the slots on the washer’s left side. Do not tighten completely.

2. Hold the mist shield parallel to the washer’s left side, with the door toward the front of the washer. See below.

3.

4.

5.

Slide the mist shield down so that the two slots fit over the thumbscrew shafts.

Insert the two pins into the slots on the washer’s top left side.

Tighten the thumbscrews.

ELx50 Operator's Manual

Figure 18: Attaching the Mist Shield

36

| Chapter 2: Installation

Connecting the Power Supply and Cord

Warning! Power Rating.

The ELx50™ power supply must be connected to a power receptacle that provides voltage and current within the specified rating for the system. Use of an incompatible power receptacle may produce electrical shock and fire hazards.

Warning! Electrical Grounding.

Never use a twoprong plug adapter to connect primary power to the ELx50 power supply. Use of a two-prong adapter disconnects the utility ground, creating a severe shock hazard. Always connect the power cord directly to a three-prong receptacle with a functional ground.

The ELx50™ Washer uses an external 24-volt power supply. The power supply automatically adjusts for input voltage in the range of 100 - 250 V~.

1.

2.

Plug the power supply’s plug into the washer’s rear panel. See

Figure 11

(all models except the ELx50/8FMW) or

Figure 12

(ELx50/8FMW model).

Insert the power cord into the power supply and into an appropriate wall outlet.

Serial Port for Communication With Other Devices

The ELx50 has a 25-pin serial (RS-232) port located on the rear panel of the instrument

(see

Figure 11

or

Figure 12

). The serial port allows the washer to receive data from a computer for purposes of factory or dealer service.

BioTek Instruments, Inc.

Verifying Performance

|

37

Verifying Performance

Before using the ELx50™ washer for the first time, verify that it is operating properly by turning the washer on. The washer will automatically perform a system Self-Test to check the manifold, microplate carrier, vacuum pump and syringe pump positioning and operation. You can also choose to run this test from the washer’s front panel, by selecting

UTIL > TESTS > SLFCHK

.

Once the self test has completed successfully, the washer is ready for use. For more information regarding verification of the washer’s performance, see

Chapter 4,

Maintenance and Performance Verification (IQ/OQ/PQ)

.

If the test passes

, the Main Menu will be displayed.

If the test fails

, the washer will beep repeatedly and the display will show an error code. If this happens, note the error code and then press the

Stop

key on the keypad to stop the beeping. Refer to

Appendix A

,

Error Codes

, to determine its cause. If the problem is something you can fix, turn off the washer, fix the problem, and then turn the washer back on. If the cause is not something you can fix, contact BioTek’s Technical Assistance Center.

Chapter 4, Maintenance and Performance Verification (IQ/OQ/PQ)

provides recommended Installation and Operational Qualification procedures to be performed after the instrument is installed and set up as described in this chapter, and before the instrument is used in a laboratory environment.

The successful completion of the Installation Qualification confirms that the washer and its components have been supplied as ordered and ensures that they are assembled and configured properly for your lab environment. The successful completion of the

Operational Qualification confirms that the washer is operating according to specification.

Note: An instrument qualification package (PN 4070530) for the

ELx50 is available for purchase from BioTek. The package contains thorough procedures for performing Installation Qualification,

Operational Qualification and Performance Qualification (IQ/OQ/PQ) and preventive maintenance (PM). Extensive Checklists and Logbooks are included for recording results. Contact your local dealer for more information.

Important!

Before operating this washer, review the

General Usage Guidelines

any program, information on performing periodic maintenance, and points to consider when creating or editing washer programs.

in

Chapter 3

. These guidelines include necessary steps to perform before running

ELx50 Operator's Manual

38

| Chapter 2: Installation

Repackaging and Shipping

If you need to ship the ELx50™ Washer to BioTek for service or repair, be sure to use the original packing. Other forms of commercially available packing are not recommended and can void the warranty.

If the original packing materials have been damaged or lost, contact BioTek for replacement packing (see Technical Assistance in

Chapter 1

for contact information).

The instrument’s packaging design is subject to change over time. If the instructions in this section do not appear to apply to the packaging materials you are using, please contact BioTek’s Technical Assistance

Center for guidance.

Before Repackaging the Instrument

Warning!

If the washer has been exposed to potentially hazardous material, decontaminate it to minimize the risk to all who come in contact with the washer during shipping, handling and servicing.

Decontamination prior to shipping is required by the U.S.

Department of Transportation regulations.

Important!

Perform the following steps in the order presented before repackaging the washer for shipment.

1. Decontaminate

the washer and its accessories as necessary. Refer to the

Decontamination

procedure in

Chapter 4

for complete instructions.

2. Obtain a

Return Materials Authorization

(RMA) number from BioTek’s

Technical Assistance Center through BioTek’s website, fax, or e-mail address listed in

Chapter 1

.

3.

4.

When obtaining the RMA, explain whether the washer requires calibration, cleaning, periodic maintenance, warranty work, and/or repair. Make a note of any error messages displayed and their frequency.

Provide BioTek with the name and contact information of a person who may be contacted if questions arise.

5.

6.

Mark the RMA number on the outside of the shipping box.

Insure the instrument for full value before shipping to BioTek.

BioTek Instruments, Inc.

Repackaging and Shipping

|

39

Repackaging the ELx50 and Its Accessories

Refer to

Figures 19

through

29

below and on the following pages when repackaging the ELx50.

Removing the Mist Shield

2.

3.

4.

1. Turn the washer off and disconnect the power supply cord and all waste and supply tubing (and, if applicable, the optional valve tubing) from the washer.

Remove the two mist shield thumbscrews.

Remove the mist shield and place it into a plastic bag (not shown).

Reinstall the two mist shield thumbscrews into their slots on the washer.

ELx50 Operator's Manual

Figure 19: Removing the Mist Shield

40

| Chapter 2: Installation

Removing the Manifold and Microplate Carrier

1. Remove the thumbscrews from the manifold.

2. For all models of the manifold: Lift the manifold carefully from the support pins on the manifold mounting bracket.

3. For the 8-, 8s-, or 12-channel manifolds only: Screw the two thumbscrews back into their support pins on the manifold mounting bracket.

4.

5.

Pack the manifold as instructed on the following two pages.

Remove the microplate carrier and place it into a plastic ziplock bag.

Figure 20: Removing the Manifold and Microplate Carrier

BioTek Instruments, Inc.

Repackaging and Shipping

|

41

Removing the Aspiration Carrier (ELx50™/8FMW)

1.

2.

3.

4.

Disconnect the tubing from the back of the aspiration carrier.

Lift the carrier off the carrier transport rail.

Ensure that the two vent plugs are inserted in the “A” and “B” storage slots in the front panel of the carrier, for shipping.

Place the carrier into a plastic ziplock bag.

Figure 21: Removing the Aspiration Carrier (ELx50/8FMW)

ELx50 Operator's Manual

42

| Chapter 2: Installation

Packing the 8-, 8s-, or 12-Channel Manifold in Its Container

1.

2.

3.

Place the manifold into the foam shipping container as shown.

Wrap a large rubber band around the shipping block.

Place the shipping block into a plastic ziplock bag.

Figure 22: Packing the 8-, 8s-, or 12-Channel Manifold

BioTek Instruments, Inc.

Repackaging and Shipping

|

43

Packing the 16-Channel Manifold in Its Container

1.

2.

3.

Place the three thumbscrews into a small ziplock bag. Insert the bag into the hole in the bottom of the foam shipping container for the 16-channel manifold.

Place the manifold into a plastic bag.

Insert the bag into the container.

Figure 23: Packing the 16-Channel Manifold in Its Container

ELx50 Operator's Manual

44

| Chapter 2: Installation

Rotating the Cam

1.

2.

Ensure that the mist shield, manifold, and carrier have been removed as instructed on the previous pages.

Rotate the cam until it is in the downward position, as shown below.

Figure 24: Rotating the Cam

BioTek Instruments, Inc.

Repackaging and Shipping

|

45

Placing the Washer into the Inner Shipping Container

5.

6.

7.

1.

2.

3.

4.

Ensure that the cam is in the down position as shown on the previous page.

Align the groove in the foam shipping block to the manifold mounting bracket and push the block onto the bracket.

Place the washer into a plastic bag.

Carefully lower the instrument into the inner shipping container, placing it all the way forward into the box.

Place insert #1 into the shipping container, as shown below.

Slide the washer all the way back to insert #1.

Place insert #2 into the front of the shipping container, as shown below.

Figure 25: Placing the Washer into the Inner Shipping Container

ELx50 Operator's Manual

46

| Chapter 2: Installation

Placing the Aspiration Carrier into the Inner Shipping Container

1. ELx50™/8FMW model only: Place the aspiration carrier in its compartment as shown.

2. Close the inner shipping container.

Figure 26: Placing the Aspiration Carrier (ELx50/8FMW) into the Inner Shipping Container

BioTek Instruments, Inc.

Repackaging and Shipping

|

47

Packing the Inner Shipping Container

1.

2.

Place the eight foam corner blocks onto the inner container as shown.

Place the container into the outer shipping container.

ELx50 Operator's Manual

Figure 27: Packing the Inner Shipping Container

48

| Chapter 2: Installation

Packing the 16-Channel Manifold and Mist Shield

1.

2.

3.

Slide the 16-channel manifold underneath the mist shield as shown.

Insert the mist shield and manifold into the top packing container.

Place the container on top of the inner shipping box.

Figure 28: Packing the 16-Channel Manifold and Mist Shield

BioTek Instruments, Inc.

Repackaging and Shipping

|

49

Packing the 8-, 8s-, or 12-Channel Manifold, Microplate Carrier, and Power Supply

1.

2.

3.

Place the 8-, 8s-, or 12-channel manifold, microplate carrier, and power supply into the shipping tray as shown, then place into the outer shipping container.

Close the top of the outer shipping box and tape it shut.

Write “RMA” and the RMA number in large, clear letters on the outside of the shipping container and ship the instrument to the BioTek address provided in the

Product Support & Service

section of

Chapter 1

.

Figure 29: Packing the 8-, 8s-, and 12-Channel Manifold, Microplate Carrier, and Power Supply

ELx50 Operator's Manual

50

| Chapter 2: Installation

BioTek Instruments, Inc.

Chapter 3

Operation

This chapter includes instructions for operating the ELx50™

Microplate Strip Washer and its software.

ELx50 Front Panel....................................................................52

Keypad Description ..................................................................53

Entering Program Names ..........................................................54

System Startup .......................................................................54

Main Menu ..............................................................................55

General Usage Guidelines .........................................................56

Wash .....................................................................................60

Prime.....................................................................................72

Dispense ................................................................................76

Aspiration ...............................................................................79

Soak ......................................................................................83

Linking Programs .....................................................................86

Utility.....................................................................................89

ELx50 Microplate Strip Washer Menu Maps ..................................92

52

| Chapter 3: Operation

ELx50 Front Panel

The front panel on the ELx50™ Microplate Strip Washer features a 25-key keypad and a 2 x 24-character LCD display as shown in

Figure 30

below.

LCD display

SoftKeys (4)

Alphanumeric keys

BIO-TEK ELx50

RUN DEFINE MAINT UTIL

Menu functions

(second line of

LCD display)

Special function keys

Figure 30: ELx50 Front Panel, With LCD and Keypad

BioTek Instruments, Inc.

Keypad Description

|

53

Keypad Description

Main

Menu

Options

ENTER

The keypad has four

SoftKeys,

one below each selectable menu option. Press a SoftKey to choose a menu selection. For example, from the

Main Menu

, press the leftmost SoftKey to select

RUN

, the rightmost to select

UTIL

.

Exit the current screen and return to the

Main Menu.

After defining a washer program, press the

Main Menu

key, then press

YES

to save the program.

To scroll through the different options within a program, press the

Options

ENTER

key or the

Shift

+

Options

key combination. Press the

key to select the current option.

Pressing

ENTER

generally saves the current screen settings and advances to the next screen in a series.

Previous

Screen

To move to a previous menu, press the

Previous Screen

key.

CLEAR

Press

CLEAR

to delete an entry.

Press the ◄ (reverse) arrow to move the cursor to the left in the LCD display.

Press the ► (forward) arrow to move the cursor to the right in the LCD display.

To start running a washer program, press the

START

key.

START

STOP

To stop running (abort) a washer program, press the

STOP

key.

ELx50 Operator's Manual

54

| Chapter 3: Operation

Entering Program Names

Each ELx50™ washer program requires the entry of a unique name, using up to 16 alphanumeric characters. To enter a program name via the washer keypad:

1.

2.

3.

4.

5.

Press

Shift

+ key

A

-

H

, or

Scroll through the alphabet with the Options key for

A

-

Z

.

Press

Shift

+

Options

to reverse direction.

Use the ► (forward) or ◄ (reverse) arrows to move the cursor within the display.

Press the

ENTER

key to store the completed program name and continue.

If the program name already exists, the washer software will require you to enter a unique name. This protects washer programs from inadvertently being overwritten.

The NAME display offers four symbols that can be used in a program name: dash (hyphen), forward slash, colon, and underscore (see below). These symbols appear in the LCD as

SoftKey choices. To include a symbol within a program name, press its corresponding

SoftKey.

NAME: WASH 96

- / : _

System Startup

To turn on the ELx50™, press the on/off switch on the right side of the base. The ELx50 will perform a self-test, displaying the screens shown below until initialization is complete.

During this time, all keys are inactive.

If the self-test fails, the washer will “chirp” and display an error code. Press

STOP

to stop the chirp. Refer to

Chapter 5

,

Troubleshooting

and

Appendix A

,

Error Codes

, to interpret these codes. Contact BioTek Instrument’s Technical Assistance Center (see page

12), for further assistance with troubleshooting.

POWERUP SEQUENCE V1.02

INITIALIZING. . .

(Instrument ID)

SYSTEM SELF-TEST. . .

BioTek Instruments, Inc.

Main Menu

|

55

Main Menu

Following successful power-up of the ELx50™, the

Main Menu

appears:

BIO-TEK ELx50

RUN DEFINE MAINT UTIL

The following options are available through the Main Menu. (See the

Washer Menu

Maps

on pages 92 through 95for a complete listing of menu options.)

BIO-TEK ELx50

RUN DEFINE MAINT UTIL

SELECT UTILITY:

TESTS SETUP AUTOPRIME

SELECT MAINT PROGRAM:

(*See options below.)

SELECT ACTION:

CREATE EDIT COPY DELETE

SELECT PROGRAM TYPE:

WASH PRIME DISP *MORE

SELECT PROGRAM TYPE:

ASPIR SOAK LINK *MORE

*Maintenance options include:

DECONTAMINATION, LONG_SHUTDOWN, OVERNIGHT_LOOP,

RINSE_AND_SOAK, and DAY_RINSE.

ELx50 Operator's Manual

56

| Chapter 3: Operation

General Usage Guidelines

Important!

Please review the before running any programs.

General Usage Guidelines

Vacuum Filter:

It is strongly recommended that you install the hydrophobic in-line vacuum filter before using the ELx50™.

See

Connecting the Tubes and Bottles

in

Chapter 2

for more information.

When operating the washer, follow these guidelines to ensure optimal performance and to prevent the problems described in the troubleshooting charts in

Chapter 5

,

Troubleshooting

.

Before Running Any Program

• Fill the wash/rinse bottles with sufficient fluid. Make sure the supply tube is in the liquid.

• The supply tube should extend to the bottom of the bottle and be cut at the end (see drawing to the right) to ensure free flow of liquid at the bottom of the supply bottle.

• Empty the waste bottle and firmly seat the waste bottle’s stopper. To ensure that fluid does not back up into the vacuum pump during operation, never allow the waste bottle to become more than three-quarters full.

• Check the external tubing connections for kinks and clogs.

• For washers with the aspiration carrier (ELx50/8FMW): when setting up the tubing from the aspiration carrier to the port on the back of the washer, ensure that there is sufficient room for the tubing.

Since the carrier moves in and out during a Dispense program, a clear tubing flex path needs to be maintained so that carrier motion is not impeded.

Pushing the instrument up against a wall or against the vacuum/dispense bottles can result in errors, because the carrier loses steps during movement.

See

Chapter 2

,

Installation

, for illustrations of proper tubing connections.

• Make sure the bottles, solutions, and tubing are clean and do not contain any particles or mold. Water and dye solutions that are recycled over several days will grow algae, bacteria, molds, or other undesirable organisms.

• If there are large air pockets in the tubing, run a Prime program using

200 ml before running another program.

BioTek Instruments, Inc.

General Usage Guidelines

|

57

• To avoid creating air bubbles every time the wash or rinse bottle is filled, make a mark halfway down the bottle and refill when the fluid level has dropped to that point. Unscrew the cap and let it hang over the side just enough to avoid emptying the inside tube and enough to refill the bottle.

• When placing a microplate on the carrier, make sure that well A1 is in the right rear corner as you face the front of the washer, and that the plate is firmly seated in the carrier. The microplate carrier has a microplate spring that holds the microplate against the front edge of the carrier. The aspiration carrier in the

ELx50™/8FMW model has a plate clamp for securing the microplate in place.

(See

Chapter 2

,

Installation

, for illustrations of the microplate and aspiration carriers.)

• Ensure that you are using the correct microplate/microstrip types:

8-, 8s-channel manifold: 96-well plates, 8-well strips.

12-channel manifold: 96-well plates, 12-well strips.

16-channel manifold: 96- or 384-well plates, 8-well strips.

Aspiration carrier (ELx50/8FMW): 96-well filter-bottom microplates only. Microstrips are not supported.

• When placing microstrips on the carrier, always load the strips starting with position one in the holder closest to the rear of the instrument.

Vacuum control for the aspiration carrier:

Make sure that there is enough vacuum for aspiration. Use the vent port (labeled “C”) on the carrier and the two vent plugs (located in storage slots “A” and “B”) to vary the vacuum levels. (See

Figure 10

in

Chapter 2

for an illustration of the vent port and the two vent plugs.)

The vent port is a 0.047” diameter hole, which is equal to approximately

-64 mmHg. The gray vent plug contains a 0.032”diameter hole, which is equal to approximately -116 mmHg. The black vent plug contains a 0.020” diameter hole, which is equal to approximately -240 mmHg.

The approximate vacuum levels listed above (-64, -116, and -240 mmHg) are “typical” only, and may vary, depending upon factors such as relative humidity, barometric pressure, and mechanical tolerances.

For a

low

vacuum level, leave the vent port on the carrier

open

.

For a

medium

vacuum level, insert the

gray

vent plug into the vent port.

For a

high

vacuum level, insert the

black

vent plug into the vent port.

Important!

Ensure that the vent port is not completely sealed. If the vent port is sealed, the washer’s internal vacuum pump may collapse the vacuum reservoir bottle.

ELx50 Operator's Manual

58

| Chapter 3: Operation

Maintenance

Perform Daily , Overnight, and Periodic Maintenance as described in

Chapter 4

,

Maintenance and Performance Verification (IQ/OQ/PQ)

. Periodic Maintenance includes:

• Cleaning the manifold tubes and chambers.

• Cleaning the plate carrier.

• Cleaning the supply, rinse, and waste bottles.

• Rinsing and checking the tubing and the check valves.

Frequent inspection of the washer, tubing, and bottles for the possible causes listed below may help to eliminate many of the problems listed in

Chapter 5

,

Troubleshooting

.

Loose or incorrect connections

of power supply, waste and supply tubing and fittings.

Incomplete

or

incorrect installation

of manifold and carrier,

attachment

of mist shield, or

placement

of microplate/microstrips in carrier.

Accumulation of residue

on carrier or inside aspirate/dispense tubes, waste/supply bottles and tubing, manifold, inlet/outlet fittings, optional in-line vacuum filter, or optional vacuum pump muffler (washers manufactured after January 2008).

Worn

tubing, o-rings, inlet or outlet fittings, manifold gaskets or vacuum seals.

Kinked or bent

aspirate/dispense tubes or waste/supply tubing.

Programming Considerations

• Be sure to enter the correct ‘First Strip’ and ‘Number of Strips’ values at runtime.

• Ensure that the correct manifold type (8-, 8s-, 12-, or 16-) has been specified (by selecting UTIL > SETUP) and is in use.

• Washers with the multiple valve tubing option (“V” version washers): Ensure that the correct reagent bottle is selected (A, B, or C).

• The dispense (overflow) height in the wash protocol should be low enough to ensure that excess fluid is removed by the aspirate tubes when the manifold is dispensing.

Plate type for the 16-channel manifold:

Ensure that the correct plate type (96 or 384) is specified when selecting the Define Wash program options.

• When editing, copying, deleting, running, or linking programs, the list of programs displayed is dependent on the manifold configurations. For example:

BioTek Instruments, Inc.

General Usage Guidelines

|

59

If an 8-channel manifold is configured, then the list of programs available for editing, copying, deleting, running, or linking will not contain 12-channel or 16-channel programs.

The option to perform a bottom aspirate using the aspiration carrier

(ELx50™/8FMW model) is only available when the instrument is configured with either the 8- or 8s-channel manifold. Aspirate-only programs containing a bottom aspiration sequence will not be displayed in the program list, if a 12-channel manifold is configured.

Bottom aspiration time:

When you are using the aspiration carrier in the

ELx50/8FMW washer, ensure that you do not define a bottom aspiration time

(BOT ASPIRATE TIME) that is too long for your assay type. Long aspiration times may cause beads to stick to the bottom of the wells in the filter-bottom plates. Try shorter aspiration times (e.g., 10 seconds) initially, and then increase the aspiration time until suitable aspiration is achieved. You may also wish to vary the vacuum levels by using the vent port on the aspiration carrier and the two vent plugs. Refer to the instructions for using the vent port and plugs in

“Vacuum control for the aspiration carrier” on page 57.

Liquid Level Alert™ Option

ELx50 washers manufactured after October 2007 support the Liquid Level Alert option. With this option, each bottle has a level sensing device inside. When a supply bottle’s fluid level drops to approximately 400-450 mL, or the waste bottle’s fluid level rises to approximately 1400-1450 mL, a sensor will activate.

The washer checks the status of the sensor at the start of all programs and also at the end of Wash and Dispense programs. If a sensor has been activated, an error code

(2800 or 2900) will appear on the display. If this happens, press the STOP key and then select ABORT or CONTINUE. If you choose to continue running a program, it is your responsibility to ensure that the supply bottle contains enough fluid to process the plate.

Sensor detection is disabled on the washer by default. To view or change the sensor detection setting , follow the washer menu path below:

UTIL SETUP SENSOR LIQUID LEVEL ALERT?

YES NO

ELx50 Operator's Manual

60

| Chapter 3: Operation

Wash

This section discusses how to run, define, edit, copy, and delete

Wash

programs.

Run Wash Program

Before running any program, make sure that the microplate or the holder containing the microstrips is properly seated in the plate carrier.

To run a Wash program, follow the menu path shown here. Some of these items are options and some are prompts. The options are described below.

RUN WASH SELECT WASH PROGRAM SELECT

REAGENT BOTTLE* PRIME THE WASHER? SELECT

PRIME PROGRAM FIRST STRIP NUMBER OF STRIPS

PLACE PLATE IN CARRIER AND PRESS <START> KEY

WASH PROGRAM RUNNING.PRESS <STOP> KEY TO QUIT

WASH PROGRAM COMPLETE. PRESS <ENTER> KEY

* SELECT REAGENT BOTTLE only appears on the “V” version washers.

Run Wash Program Options

Options Actions

Wash program

First Strip

Number of Strips

Select a Wash program.

Note: For washers with dual-configuration for switching between the 16- and 8-channel manifolds, only those programs defined under the current configuration will be presented for selection. For example, if you are currently using the 16-channel manifold, only those programs that were defined with the 16-channel configuration will be displayed.

Select Reagent Bottle Select the reagent bottle from which to dispense. This applies only to ″V″ version models.

Prime? Select a prime program to run before starting the wash program.

Specify the number of the first strip to be washed.

Specify the number of microwell strips to be washed (the actual selectable range is based on the First Strip value):

1 to 12 for the 8- or 8s-channel manifold

1 to 8 for the 12-channel manifold

1 to 24 for the 16-channel manifold

BioTek Instruments, Inc.

Wash

|

61

Define Wash Program

To define a Wash program, follow the menu path shown here. Some of these items are options and some are prompts. The options are described below.

DEFINE CREATE WASH NAME PLATE TYPE*

DEFINE WASH COMPONENT Select the Program

Component (Method, Dispense, Aspiration)

Press MAIN MENU KEY to end OK TO SAVE

PROGRAM?

* PLATE TYPE only appears if the washer is configured with the 16-channel

manifold.

Define Wash Program Options

Options

Name

Plate Type

Wash Component

Press Main Menu to end

Actions

Enter a unique name for the Wash program, using up to 16 alphanumeric characters. See Entering Program Names on page 54 for instructions.

Select a Plate Type of 96 if washing 8-well strips or 96-well plates, or 384 if washing 384-well plates (for washers configured with the 16-channel manifold).

METHOD: Wash cycles, plate or strip format, soaking and shaking, soak duration, shake before soak, shake duration, shake intensity, prime after soak, prime volume and prime flow rate. The fields are described in the Method table, page

62.

DISP: Dispense Volume, Dispense Flow Rate, Dispense

Height, Horizontal Dispense Position, Bottom Wash First,

Bottom Dispense Volume, Bottom Flow Rate, Bottom

Dispense Height, Bottom Horizontal Position, Prime Before

Start, Prime Volume, Prime Flow Rate. The fields are described in the Dispense table, page 64.

ASPIR: Aspirate Height, Horizontal Aspirate Position, Aspirate

Rate, Aspirate Delay, Crosswise Aspirate, Crosswise On,

Crosswise Height, Crosswise Horizontal Position, Final

Aspiration, Final Aspirate Delay. The fields are described in the Aspiration table, page 67.

After all wash components have been programmed, press

Main Menu then select YES to save the program.

ELx50 Operator's Manual

62

| Chapter 3: Operation

Define Wash METHOD Program Options

Options Actions

Number of Cycles

Wash Format

Soak/Shake?

Soak Duration

Specify 1 to 10 complete wash cycles. A complete wash cycle represents an aspirate and dispense sequence, which may or may not end with a final aspiration of the liquid from the wells

(see Final Aspiration? in the Define Wash ASPIRATION

Program Options table, page 67.)

Note: If Bottom Wash is enabled, an additional aspirate and dispense sequence is performed using unique bottom washing dispense parameters (see Bottom Wash First? in the Define

Wash DISPENSE Program Options table, page 64).

Choose Plate or Strip wash format.

Plate format applies each cycle to the entire plate (or all of the strips) before it starts the next cycle. Strip format applies all cycles to one strip before it moves to the next strip.

Note: The number of strips to wash is defined at program run-time.

Choose YES to soak and/or shake the wash buffer in the microwells after the buffer has been dispensed.

Soak Duration represents the time to allow the wash buffer to remain in the wells before aspiration. The duration range is from 0 to 600 seconds.

A soak begins after the wash buffer is dispensed to the wells.

In some chemistries it enhances washing by allowing unbound material to diffuse into the wash buffer. When washing strips in a plate format use a soak that lasts as long as it takes to process one wash cycle of all strips.

Shake Before Soak? Choose YES to agitate the microwells back and forth linearly along the carrier track to aid in dispersal.

Before the microwells begin shaking, the manifold raises so that the dispense and aspirate tubes clear the microwells. The carrier returns to its home position so that the tubes are above the carrier trough.

Shake Duration

Shake Intensity

Shake Duration represents the length of time that the microwells are agitated. The duration range is from 1 to 600 seconds.

Specify the intensity of microwell shaking. The options range from 1 to 5, where 1 = Least Intense, 5 = Most Intense. The washer display shows the corresponding cycles/sec value for each option.

Prime After Soak? Choose YES to prime the dispense tubes after a soak cycle.

Prime Volume Specify the volume, in milliliters, of the solution used to prime the dispense tubes. The volume range is from 1 to 200 ml.

BioTek Instruments, Inc.

Wash

|

63

Options

Prime Flow Rate

Actions

Specify the rate at which the priming fluid will be pumped into the dispense tubes. The flow rate options range from 1 to 9 (1 to 7 for the 12-channel manifold), where 1 = Slowest,

9 = Fastest. The washer display shows the corresponding

µl/well/sec value for each option.

Define another wash component. Press ENTER to continue

Press Main Menu to end

Select YES to save the program.

Wash METHOD Defaults and Ranges

Parameter

Number of Cycles

Wash Format

Default

3

Plate

Range

1-10

Plate/Strip

Units

Soak Duration

Shake Before Soak?

Shake Duration

Shake Intensity

30

No

5

3

No

0-600

Yes/No

1-600

1-5

Yes/No seconds seconds

1 = 15 cycles/second

2 = 16

3 = 17

4 = 18

5 = 19

Prime After Soak?

Prime Volume

Prime Flow Rate 5 1-9 (1-7 for the 12-channel manifold)

Selection = Flow rate in ul/well/second

8/8s 12 16

8 = 800 ----- 450

9 = 1000 ----- 500

ELx50 Operator's Manual

64

| Chapter 3: Operation

Define Wash DISPENSE (DISP) Program Options

Note: The shaded options are not available for Dispense-only programs or for washers configured with 16-channel manifolds.

Options

Dispense Volume

Dispense Flow Rate

Dispense Height

(Overflow Position)

Horizontal Disp Pos

Actions

Specify the volume of wash solution to dispense per microwell, per wash cycle. The volume range is 50 to 3000

µl/well for the 8-channel, 8s-channel, 12-channel manifolds, as well as for the 16-channel manifold, but only when it is dispensing to 96-well plates. For the 16-channel manifold dispensing to 384-well plates, the range is 25-3000 μl/well.

Note: During program creation, it is possible to enter a

Dispense Volume/Flow Rate combination that is invalid for the current washer manifold type. The ELx50™ basecode performs validation of these parameters at program run-time only—if the configuration is found to be invalid, the 0F00 error will display. See Dispense Volume Invalid for

Manifold Type (0F00 Error) in Appendix A for a table showing valid Dispense Volume ranges for each Dispense

Flow Rate / Washer Manifold combination. (The same table is also included on page 66.)

Specify the rate at which the dispense fluid is dispensed from the tubes. The flow rate options range from 1 to 9 (1 to 7 for the 12-channel manifold), where 1 = Slowest, 9 = Fastest.

The washer display shows the corresponding µl/well/sec value for each option.

Dispense Height represents the height between the bottom of the aspirate tubes and the carrier surface on which the plate rests, during the dispense part of a wash cycle. See Figure

31 on page 70.

The Dispense Height options range from 12 to 180. The washer display shows the corresponding measurement in millimeters for each option.

12 is the closest to the carrier surface, and 180 is the furthest from this surface. The dispense tubes are shorter than the aspirate tubes.

Note: It is considered good practice to set a dispense height that positions the aspirate tubes at the top of wells to draw off any overflow.

Horizontal Dispense Position represents the position of the manifold tube in relation to the center of the microwell.

The Horizontal Disp Pos options range from: -24 to 24 for the

8- and 12-channel manifolds; -88 to 24 for the 8s-channel;

-24 to 15 for the 16-channel, 96-well plates; and -22 to 10 for the 16-channel, 384-well plates.

BioTek Instruments, Inc.

Wash

|

65

Options Actions

(Horizontal Disp Pos) The washer display shows the corresponding measurement in millimeters for each option.

A negative offset (-24 to -1) moves the dispense tubes away from the front edge of the well. A positive offset (1 to 24) moves the dispense tubes toward the front edge of the well.

A setting of 0 indicates no offset (0.000 mm).

Bottom Wash First?

Bottom Disp Volume Specify the volume of wash solution to dispense per microwell during the bottom wash. The volume range is 50 to 3000

µl/well.

Bottom Flow Rate Specify the rate at which the dispense fluid is dispensed from the tubes during the bottom wash. The flow rate options range from 1 to 9 (1 to 7 for the 12-channel manifold), where

1 = Slowest, 9 = Fastest. The washer display shows the corresponding µl/well/sec value for each option.

Bottom Disp Height

Bottom washing consists of an initial dispense/aspirate sequence, adding another wash cycle to the defined number of cycles. The manifold is positioned at the defined Bottom

Disp Height. The reagent is simultaneously dispensed and aspirated at this height to create cleaning turbulence. The manifold descends to aspirate again and ends with a final dispensing of fluid to fill the well.

Bottom Horiz Pos

Bottom Dispense Height represents the height between the bottom of the aspirate tube and the carrier surface on which the plate rests, during the wash. See Figure 31 on page 70.

The Bottom Disp Height options range from 12 to 180. The washer display shows the corresponding measurement in millimeters for each option.

12 is the closest to the carrier surface; 180 is the furthest.

The dispense tubes are shorter than the aspirate tubes.

See the description for Horizontal Disp Pos in this table. The only difference is that Bottom Horizontal Pos specifically refers to Bottom Washing.

Prime Before Start? Select YES to prime the dispense tubes before starting the wash program.

Prime Volume Specify the volume of the buffer or rinse used to prime the dispense tubes. The volume range is 1 to 200 ml.

Prime Flow Rate

Press Enter to continue

Specify the rate at which the priming fluid is pumped into the dispense tubes. The flow rate options range from 1 to 9 (1 to

7 for the 12-channel manifold), where 1 = Slowest, 9 =

Fastest. The washer display shows the corresponding

µl/well/sec value for each option.

Define another wash component.

Press Main Menu to end

Select YES to save the program.

ELx50 Operator's Manual

66

| Chapter 3: Operation

Wash DISPENSE Defaults and Ranges

Parameter

Dispense Volume*

Default Range

(16-channel manifold, 384-well plates)

50-3000 (8-, 8s-, and

12-channel manifold;

16-channel manifold,

96-well plates)

Units

µl/well

Dispense Height

Horizontal Dispense Position

Bottom Wash First?

Bottom Dispense Volume

Bottom Flow Rate

Bottom Dispense Height

Bottom Horizontal Position

120

0

No

300

5

60

0

12-180

-88 to 24**

Yes/No

50-3000

1-9

12-180

-88 to 24** quarter steps (in mm)

µl/well

µl/well/sec quarter steps (in mm)

Prime Before Start?

Prime Volume

Prime Flow Rate

No Yes/No

5 1-9 µl/well/sec

* Refer to the table below for valid Dispense Volume/Flow Rate combinations.

** Horizontal Dispense Position and Bottom Horizontal Dispense Position are model-specific:

-24 to 24 (8- and 12-channel manifold); -88 to 24 (8s-channel manifold); -24 to 15

(16-channel manifold, 96-well plates); -22 to 10 (16-channel manifold, 384-well plates)

Dispense

Flow Rate

1

2

3

7

8

9

4

5

6

Dispense Volume Range (μl/well) by Manifold Type

8-Channel 12-Channel

16-Channel

96-well plate 384-well plate

50-3000 50-3000 50-3000 25-3000

50-3000 50-3000 50-3000 25-3000

50-3000 50-3000 50-3000 25-3000

50-3000 50-3000 56-3000 28-3000

50-3000 60-3000 60-3000 30-3000

50-3000 70-1100 90-3000 45-3000

55-3000 70-1100 100-3000 50-3000

BioTek Instruments, Inc.

Wash

|

67

Define Wash ASPIRATION (ASPIR) Program Options

Note: The shaded options are not available for Aspirate-only programs.

Note: Bottom aspiration using the aspiration carrier (ELx50™/8FMW washer) is not available for Wash programs.

Options Actions

Aspirate Height Aspirate Height represents the height between the bottom of the aspirate tubes and the carrier surface on which the plate rests when aspirating the microwells. See Figure 31 on page

70.

The Aspirate Height options range from 12 to 180. The washer display shows the corresponding measurement in millimeters for each option.

12 is the closest to the carrier surface, 180 is the furthest from the surface. The aspirate tubes are longer than the dispense tubes.

Horizontal Aspir Pos Horizontal Aspirate Position represents the position of the manifold tube in relation to the center of the microwell. The

Horizontal Aspir Pos options range from: -24 to 24 for the

8- and 12-channel manifolds; -24 to 88 for the 8s-channel;

-15 to 24 for the 16-channel, 96-well plates; and -10 to 22 for the 16-channel, 384-well plates. The washer display shows the corresponding measurement in millimeters for each option.

A negative offset (-24 to -1) moves the aspirate tubes toward the rear edge of the well. A positive offset (1 to 24) moves the aspirate tubes away from the rear edge of the well. A setting of 0 indicates no offset (0.000 mm).

Note: For flat-bottom plates, position the aspirate tubes near the edge of the well for minimal residual volume.

Aspirate Rate

Aspirate Delay

Specify the rate at which the washer manifold travels down into the wells.

The Aspirate Rate options range from 1 to 7, where

1 = Slowest, 7 = Fastest. The washer display shows the corresponding mm/sec value for each option.

Aspirate Delay represents the time delay starting when the tube is at its aspiration height and ending when it next moves. The Aspirate Delay range is 0 to 5000 milliseconds.

The delay applies to the normal (initial) aspiration if

Crosswise Aspirate is not enabled. If Crosswise Aspirate is enabled, the delay applies to the crosswise, not the normal initial, aspiration.

ELx50 Operator's Manual

68

| Chapter 3: Operation

Options Actions

Crosswise Aspir?

Crosswise On

Crosswise Height

A crosswise aspiration is a two-step aspiration. The wells are first aspirated at the horizontal aspirate position. The aspirate tubes raise a fixed number of steps and then do a second

(crosswise) aspiration at the crosswise horizontal position.

The purpose is to eliminate residual sample or reagent from the wall perimeter.

Select YES to designate a second aspiration.

Select ALL to perform a crosswise aspiration on all aspirations, or FINAL to perform the crosswise aspiration on just the final aspiration.

Crosswise Height applies to a second or final aspiration. This is the height between the bottom of the aspirate tube and the carrier surface on which the plate rests. See Figure 31 on page 70.

The Crosswise Height options range from 12 to 180. The washer display shows the corresponding measurement in millimeters for each option.

12 is the closest to the carrier surface, 180 is the furthest from the surface. The aspirate tube is longer than the dispense tube.

Crosswise Horiz Pos Crosswise Horizontal Position applies to a second aspiration.

See Horizontal Aspir Pos, on the previous page, in this table.

The only difference is that Crosswise Horiz Pos specifically refers to crosswise aspirations.

Final Aspiration?

Final Aspir Delay

Select YES to designate a final aspiration, leaving the wells empty.

Final Aspiration Delay represents the time delay starting when the tube is at its final aspiration height and ending when it next moves. The Final Aspir Delay range is 0 to 5000 milliseconds.

Define another wash component. Press Enter to continue

Press Main Menu to end

Select YES to save the program.

Note: The shaded options are not available for Aspirate-only programs.

BioTek Instruments, Inc.

Wash

|

69

Wash ASPIRATE Defaults and Ranges

Parameter Default Range Units

Aspirate Height

Horizontal Aspirate Position

Aspirate Rate

Aspirate Delay

Crosswise Aspirate

Crosswise On

24

0

3

0

No

Final

12-180

-24 to 88* quarter steps (in mm)

1-7**

0-5000 milliseconds

Yes/No

All/Final

Crosswise Height

Crosswise Horizontal Position

Final Aspirate

24

0

Yes

12-180

-24 to 88* quarter steps (in mm)

Yes/No

Final Aspirate Delay 0 0-5000 milliseconds

*Horizontal Aspirate Position and Crosswise

Horizontal Position are model-specific:

-24 to 24 (8- and 12-channel manifold)

-24 to 88 (8s-channel manifold)

-15 to 24 (16-channel manifold, 96-well plates)

-10 to 22 (16-channel manifold, 384-well plates)

**Aspirate Rate

(mm/sec):

1 = 05.0

2 = 05.2

3 = 05.8

4 = 06.7

5 = 08.0

6 = 10.7

7 = 16.0

ELx50 Operator's Manual

70

| Chapter 3: Operation

Microplate

Manifold

Dispense height

Carrier

Aspirate height

Figure 31: Dispense and Aspirate Heights for the 8-, 8s-, and 12-Channel Manifolds

(See

Figure 1

for an illustration of the 16-channel manifold)

Note: The dispense tubes are shorter than the aspirate tubes.

BioTek Instruments, Inc.

Wash

|

71

Edit Wash Program

To change a Wash program, follow the menu path shown here. Some of these items are options and some are prompts.

DEFINE EDIT WASH SELECT WASH PROGRAM

Edit the Program Component (Method, Dispense,

Aspiration) MAIN MENU KEY OK TO SAVE

PROGRAM?

Copy Wash Program

To copy a Wash program, follow the menu path shown here. Some of these items are options and some are prompts. The options are described below.

DEFINE COPY WASH SELECT WASH PROGRAM

NAME OK TO COPY?

Copy Wash Program Options

Options Actions

Select Wash Program Select the Wash program you wish to copy.

Name Enter a unique name for the new Wash program.

OK to Copy? Select YES to copy the program.

Delete Wash Program

To delete a Wash program, follow the menu path shown here. Some of these items are options and some are prompts. The options are described below.

DEFINE DELETE WASH SELECT WASH PROGRAM

OK TO DELETE?

Delete Wash Program Options

Options Actions

Select Wash Program Select the Wash program you wish to delete.

OK to Delete? Select YES to delete the program.

ELx50 Operator's Manual

72

| Chapter 3: Operation

Prime

This section discusses how to define, run, edit, copy, and delete Prime programs.

Define Prime Program

To define a Prime program, follow the menu path shown here. Some of these items are options and some are prompts. The options are described below.

DEFINE CREATE PRIME NAME PRIME

VOLUME PRIME FLOW RATE SOAK AFTER PRIME?

SOAK DURATION OK TO SAVE PROGRAM?

Define Prime Program Options

Options Actions

Name

Prime Volume

Prime Flow Rate

Type a unique name for the Prime program. See Entering

Program Names on page 54 for instructions.

Specify the volume of the buffer or rinse used to prime the dispense tubes. The volume range is 1 to 200 ml.

Specify the rate at which the priming fluid is pumped into the dispense tubes. The flow rate options range from 1 to

9 (1 to 7 for the 12-channel manifold), where 1 = Slowest,

9 = Fastest. The washer display shows the corresponding

µl/well/sec value for each option.

Soak after Prime? Select YES to soak the manifold tubes after they have been primed.

After the dispense tubes have been primed, the manifold raises to a trough clearance height and the carrier returns to its “home” position so that the dispense and aspirate tubes sit above the carrier trough.

The soak solution is then dispensed to the trough. The vacuum pump is turned off and the tubes are lowered into the trough to soak. After the specified Soak Duration (see below), the vacuum pump is turned on and the trough aspirated.

Soak Duration Soak Duration represents the time to allow the dispense and aspirate tubes to be soaked in wash buffer, rinse, or disinfectant. The range is from 1 minute to 18 hours.

OK to Save Program? Select YES to save the program.

BioTek Instruments, Inc.

Prime

|

73

PRIME Program Defaults and Ranges

Parameter Default Range Units

Prime Flow Rate

Soak After Prime

Soak Duration

*Prime Flow Rate (µl/well/sec):

8/8s 12 16

1 = 150 150 225

2 = 200 200 250

3 = 300 300 300

4 = 450 450 325

5 = 500 500 350

6 = 600 600 375

7 = 650 650 400

8 = 800 ----- 450

9 = 1000 ----- 500

5 1-9 (1-7 for

12-channel)*

No Yes/No

1 minute 1 minute to 18 hours

µl/well/sec minutes to hours

ELx50 Operator's Manual

74

| Chapter 3: Operation

Run Prime Program

To run a Prime program, follow the menu path shown here. Some of these items are options and some are prompts. The options are described below.

RUN PRIME SELECT PRIME PROGRAM SELECT

REAGENT BOTTLE* CONNECT REAGENT BOTTLE AND

PRESS <START> KEY PRIME PROGRAM RUNNING.

PRESS <STOP> KEY TO QUIT PRIME PROGRAM

COMPLETE. PRESS <ENTER> KEY

* SELECT REAGENT BOTTLE only appears on the “V” version washers.

Run Prime Program Options

Options Actions

Prime Program Select a Prime program.

Special note for washers with dual-configuration for switching between the 16- and 8-channel manifolds: Only those programs defined under the current configuration will be presented for selection. For example, if you are currently using the 16-channel manifold, only those programs that were defined with the 16-channel configuration will be shown to you.

Select Reagent Bottle Select the reagent bottle from which to dispense. This applies only to “V” version models.

Edit Prime Program

To change a Prime program, follow the menu path shown here. Some of these items are options and some are prompts.

DEFINE EDIT PRIME SELECT PRIME PROGRAM

Edit the Program Parameters MAIN MENU KEY

OK TO SAVE PROGRAM?

BioTek Instruments, Inc.

Prime

|

75

Copy Prime Program

To copy a Prime program, follow the menu path shown here. Some of these items are options and some are prompts. The options are described below.

DEFINE COPY PRIME SELECT PRIME PROGRAM

NAME OK TO COPY?

Copy Prime Program Options

Options Actions

Select Prime Program Select the Prime program you wish to copy.

Name Enter a unique name for the new Prime program.

OK to Copy? Select YES to copy the program.

Delete Prime Program

To delete a Prime program, follow the menu path shown here. Some of these items are options and some are prompts. The options are described below.

DEFINE DELETE PRIME SELECT PRIME

PROGRAM OK TO DELETE?

Delete Prime Program Options

Options Actions

Select Prime Program Select the Prime program you wish to delete.

OK to Delete? Select YES to delete the program.

ELx50 Operator's Manual

76

| Chapter 3: Operation

Dispense

This section discusses how to define, run, edit, copy, and delete Dispense programs.

Define Dispense Program

To define a Dispense program, follow the menu path shown here. Some of these items are options and some are prompts. The options are described below.

DEFINE CREATE DISP NAME PLATE TYPE*

DISPENSE VOLUME DISPENSE FLOW RATE

DISPENSE HEIGHT HORIZONTAL DISP POS PRIME

BEFORE START? PRIME VOLUME PRIME FLOW RATE

OK TO SAVE PROGRAM?

* PLATE TYPE only appears if the washer is configured with the 16-channel manifold.

Define Dispense Program Options

See

Define Wash DISPENSE Program Options

on page 64 for a complete list of dispense program options. Note that the shaded options are not available for Dispense-only programs.

Run Dispense Program

To run a Dispense program, follow the menu path shown here. Some of these items are options and some are prompts. The options are described below.

RUN DISP SELECT DISP PROGRAM SELECT

REAGENT BOTTLE* PRIME THE WASHER? SELECT

PRIME PROGRAM FIRST STRIP NUMBER OF STRIPS

PLACE PLATE IN CARRIER AND PRESS <START> KEY

DISPENSE PROGRAM RUNNING. PRESS <STOP> KEY

TO QUIT DISP PROGRAM COMPLETE. PRESS

<ENTER> KEY

* SELECT REAGENT BOTTLE appears only on the “V” version washers.

BioTek Instruments, Inc.

Dispense

|

77

Run Dispense Program Options

Options

Disp Program

Reagent

First Strip

Number of Strips

Actions

Select a Dispense program.

Special note for washers with dual-configuration for switching between the 16- and 8-channel manifolds: Only those programs defined under the current configuration will be presented for selection. For example, if you are currently using the 16-channel manifold, only those programs that were defined with the 16-channel configuration will be shown to you.

Select which reagent bottle from which to dispense. This applies only to the “V” version models.

Specify the number of the first strip to be washed.

Specify the number of microwell strips to be washed (the actual selectable range is based on the First Strip value):

1 to 12 for the 8- or 8s-channel manifold

1 to 8 for the 12-channel manifold

1 to 24 for the 16-channel manifold

Edit Dispense Program

To change a Dispense program, follow the menu path shown here. Some of these items are options and some are prompts.

DEFINE EDIT DISP SELECT DISP PROGRAM

Edit the Program parameters MAIN MENU KEY

OK TO SAVE PROGRAM?

Copy Dispense Program

To copy a Dispense program, follow the menu path shown here. Some of these items are options and some are prompts. The options are described below.

DEFINE COPY DISPENSE SELECT DISP

PROGRAM NAME OK TO COPY?

ELx50 Operator's Manual

78

| Chapter 3: Operation

Copy Dispense Program Options

Options

Select Dispense

Program

Name

OK to Copy?

Actions

Select the Dispense program you wish to copy.

Enter a unique name for the new Dispense program.

Select YES to copy the program.

Delete Dispense Program

To delete a Dispense program, follow the washer menu path shown here. Some of these items are options and some are prompts. The options are described below.

DEFINE DELETE DISP SELECT DISP PROGRAM

OK TO DELETE?

Delete Dispense Program Options

Options

Select Dispense

Program

OK to Delete?

Actions

Select the Dispense program you wish to delete.

Select YES to delete the program.

BioTek Instruments, Inc.

Aspiration

|

79

Aspiration

Important!

The

ELx50™/8FMW

uses either the manifold

(with the microplate carrier) for top aspiration of solid-bottom plates, or the aspiration carrier for bottom aspiration of filterbottom plates. ASPIRATE TYPE and BOT ASPIRATE TIME only appear if the washer model is the ELx50/8FMW, and the manifold is configured as either 8- or 8s-channel. Because the heights of the microplate and aspiration carriers are different, it is important that you use the correct carrier for the ASPIRATE

TYPE:

- For a TOP ASPIRATE, use the microplate carrier.

- For a BOT ASPIRATE, use the aspiration carrier.

This section discusses how to define, run, edit, copy, and delete Aspiration programs.

Define Aspiration Program

To define an Aspiration program, follow the menu path shown here. Some of these items are options and some are prompts. The options are described below.

DEFINE CREATE MORE ASPIR NAME PLATE

TYPE* ASPIRATE TYPE** BOT ASPIRATE TIME**

ASPIRATE HEIGHT HORIZONTAL ASPR POS

ASPIRATION RATE ASPIRATE DELAY CROSSWISE

ASPIR? CROSSWISE HEIGHT CROSSWISE HORIZ POS

OK TO SAVE PROGRAM?

* PLATE TYPE

only appears if the washer is configured with the 16-channel manifold.

** ASPIRATE TYPE

and

BOT ASPIRATE TIME

only appear if the washer is an

ELx50/8FMW and the manifold is configured as either 8- or 8s-channel.

Define Aspiration Program Options

See

Define Wash ASPIRATION Program Options

on page 67 for a complete listing of Aspiration Program Options that are available for Wash programs and Aspirate-only programs. Note that the shaded options are not available for Aspirate-only programs.

The tables on the following page list options and parameters that are only available for Aspirate-only programs.

ELx50 Operator's Manual

80

| Chapter 3: Operation

Define Aspirate-only Program Options

Options

Aspirate Type

BOT Aspirate Time

Actions

Aspirate Type identifies whether the manifold (with the microplate carrier) or the aspiration carrier is to be used for the aspirate function of an Aspirate-only program. This is only available for ELx50/8FMW models, and the manifold must be either 8- or 8s-channel. The two options are TOP for top aspiration using the manifold with the microplate carrier, and BOT for bottom aspiration using the aspiration carrier. The default is TOP.

The bottom aspirate time indicates how long to aspirate fluid using the aspiration carrier. The bottom aspirate time option ranges from 1 to 180 seconds. The default value is

30 seconds.

Aspirate-only Program Defaults and Ranges

Parameter

Aspirate Type

BOT Aspirate Time

Default Range

TOP

30

TOP/BOT

1-180

Units

seconds

Run Aspiration Program

To run an Aspiration program, follow the menu path shown here. Some of these items are options and some are prompts. The options are described below.

RUN MORE ASPIR SELECT ASPIR PROGRAM

FIRST STRIP* NUMBER OF STRIPS* PLACE PLATE

IN CARRIER AND PRESS <START> KEY ASPIR

PROGRAM RUNNING. PRESS <STOP> KEY TO QUIT

ASPIR PROGRAM COMPLETE. PRESS <ENTER> KEY

* FIRST STRIP

and

NUMBER OF STRIPS will not be displayed for Aspirate programs containing a bottom aspirate.

BioTek Instruments, Inc.

Aspiration

|

81

Run Aspiration Program Options

Options

Aspir Program

First Strip

Number of Strips

Actions

Select an Aspiration program.

Special note for washers with dual-configuration for switching between the 16- and 8-channel manifolds: Only those programs defined under the current configuration will be presented for selection. For example, if you are currently using the 16-channel manifold, only those programs that were defined with the 16-channel configuration will be shown to you.

Specify the number of the first strip to be washed.

Specify the number of microwell strips to be washed (the actual selectable range is based on the First Strip value):

1 to 12 for the 8- or 8s-channel manifold

1 to 8 for the 12-channel manifold

1 to 24 for the 16-channel manifold

Edit Aspiration Program

To change an Aspiration program, follow the menu path shown here. Some of these items are options and some are prompts.

DEFINE EDIT MORE ASPIR SELECT ASPIR

PROGRAM Edit Program parameters MAIN MENU

KEY OK TO SAVE PROGRAM?

Copy Aspiration Program

To copy an Aspiration program, follow the menu path shown here. Some of these items are options and some are prompts. The options are described below.

DEFINE COPY MORE ASPIR SELECT ASPIR

PROGRAM NAME OK TO COPY?

ELx50 Operator's Manual

82

| Chapter 3: Operation

Copy Aspiration Program Options

Options Actions

Select Aspiration Program Select the Aspiration program you wish to copy.

Name

OK to Copy?

Enter a unique name for the new Aspiration program.

Select YES to copy the program.

Delete Aspiration Program

To delete an Aspiration program, follow the washer menu path shown here. Some of these items are options and some are prompts. The options are described below.

DEFINE DELETE MORE ASPIR SELECT

ASPIR PROGRAM OK TO DELETE?

Delete Aspiration Program Options

Options

Select Aspiration

Program

OK to Delete?

Actions

Select the Aspiration program you wish to delete.

Select YES to delete the program.

BioTek Instruments, Inc.

Soak

|

83

Soak

This section discusses how to define, edit, copy, and delete Soak programs.

• A soak begins after the wash buffer is dispensed to the wells. In some chemistries it enhances washing by allowing unbound material to diffuse into the wash buffer.

• A Soak program cannot be run on its own; it must be selected as part of a Link program (see page 86).

• Soak program parameters can also be defined within a Wash program (see page

62).

• Washer basecode PN 4070201 does not support Soak programs. See

Software

Variations

in

Chapter 1

for more information.

Define Soak Program

To define a Soak program, follow the menu path shown here. Some of these items are options and some are prompts. The options are described below.

DEFINE CREATE MORE SOAK NAME SOAK

DURATION SHAKE BEFORE SOAK? SHAKE DURATION

SHAKE INTENSITY OK TO SAVE PROGRAM?

Define Soak Program Options

Options Actions

Name

Soak Duration

Enter the name of the Soak program. See Entering

Program Names on page 54 for instructions.

Soak Duration represents the time to allow the wash buffer to remain in the wells before aspiration. The duration range is from 0 to 600 seconds.

Note: When washing strips in a plate format use a soak that lasts as long as it takes to process one wash cycle of all strips.

Shake Before Soak? Choose YES to agitate the microwells back and forth linearly along the carrier track to aid in dispersal.

Before the microwells begin shaking, the manifold raises so that the dispense and aspirate tubes clear the microwells. The carrier returns to its home position so that the tubes are above the carrier trough.

Shake Duration Shake Duration represents the length of time that the microwells are agitated. The duration range is from 1 to

600 seconds.

ELx50 Operator's Manual

84

| Chapter 3: Operation

Options Actions

Shake Intensity Specify the intensity of microwell shaking. The options range from 1 to 5, where 1 = Least Intense, 5 = Most

Intense. The washer display shows the corresponding cycles/sec value for each option.

OK to save program? Select YES to save the program.

Edit Soak Program

To change a Soak program, follow the menu path shown here. Some of these items are options and some are prompts.

DEFINE EDIT MORE SOAK SELECT SOAK

PROGRAM “Edit the soak parameters” OK TO

SAVE PROGRAM?

Copy Soak Program

To copy a Soak program, follow the menu path shown here. Some of these items are options and some are prompts. The options are described below.

DEFINE COPY MORE SOAK SELECT SOAK

PROGRAM NAME OK TO COPY?

Copy Soak Program Options

Options Actions

Select Soak Program Select the Soak program you wish to copy.

Name Type a unique name for the new Soak program.

OK to Copy? Select YES to copy the program.

BioTek Instruments, Inc.

Soak

|

85

Delete Soak Program

To delete a Soak program, follow the menu path shown here. Some of these items are options and some are prompts. The options are described below.

DEFINE DELETE MORE SOAK SELECT SOAK

PROGRAM OK TO DELETE?

Delete Soak Program Options

Options Actions

Select Soak Program Select the Soak program you wish to delete.

OK to Delete? Select YES to delete the program.

ELx50 Operator's Manual

86

| Chapter 3: Operation

Linking Programs

This section discusses how to define, view, run, and delete Link programs.

• Pre-defined Wash, Prime, Aspirate, Dispense, and/or Soak programs can be

“linked” together and run sequentially and automatically.

• A Link program can contain up to 10 programs, all of which must specify the same plate type (96 or 384).

• For “V” version washers, all of the linked programs must use fluid from the same supply bottle. The supply bottle location (A, B, or C) is specified at program run time.

• Once defined, a Link program can be viewed, run, or deleted. It cannot be edited or copied.

• Washer basecode PN 4070201 does not support Link programs. See

Software

Variations

in

Chapter 1

for more information.

Define Link Program

To define a Link program, follow the washer menu path shown here. Some of these items are options and some are prompts. The options are described below.

DEFINE CREATE MORE LINK NAME PLATE

TYPE* SELECT LINK 01 (02, 03, up to 10)

TYPE: OK TO SAVE PROGRAM?

* PLATE TYPE only appears if the washer is configured with the 16-channel

manifold.

Define Link Program Options

Options Actions

Name Enter a unique name for the Link program. See Entering

Program Names on page 54 for instructions.

Plate Type Select a Plate Type of 96 to link programs defined for

8-well strips or 96-well plates, or 384 to link programs defined for 384-well plates.

Select Link (n) Type Select each Wash, Prime, Dispense, Aspirate, and/or Soak program to be included in the Link program. Programs will run in the order in which they are selected.

Note: If the Plate Type option appeared (see above), only programs with the selected Plate Type will be accessible for inclusion in the Link program.

BioTek Instruments, Inc.

Linking Programs

|

87

Options

Press Main Menu to end

Actions

Select YES to save the program.

Troubleshooting Tip: If your washer has basecode software part number 4070230 with custom programs created before June 1999 and you are unable to create Link programs, try this: for each program you wish to include in a Link program, select DEFINE > EDIT > choose the program > press Main Menu > choose YES to save the program. If this does not solve the problem, please contact BioTek TAC.

View Link Program

A Link program cannot be edited, but its contents can be viewed. To view a Link program, follow the menu path shown below.

DEFINE EDIT MORE LINK SELECT LINK

PROGRAM PLATE TYPE VIEW LINK 01 (02, 03, up to 10) MAIN MENU KEY

Run Link Program

To run a Link program, follow the menu path shown here. Some of these items are options and some are prompts. The options are described below.

RUN MORE LINK SELECT LINK PROGRAM

SELECT REAGENT BOTTLE* PRIME THE WASHER?

FIRST STRIP** NUMBER OF STRIPS** PLACE

PLATE IN CARRIER AND PRESS <START> KEY LINK

PROGRAM RUNNING. PRESS <STOP> KEY TO QUIT

LINK PROGRAM COMPLETE. PRESS <ENTER> KEY

*

SELECT REAGENT BOTTLE

appears only on the “V” version washers.

** If a Link program contains an Aspirate program that uses the aspiration carrier

(ELx50™/8FMW model), the

FIRST STRIP

and

NUMBER OF STRIPS

prompts do not apply to that portion of the program, since the aspiration carrier works with

96-well filter-bottom full plates only.

ELx50 Operator's Manual

88

| Chapter 3: Operation

Run Link Program Options

Options

Link Program

Reagent

Prime?

First Strip

Number of Strips

Actions

Select a Link program.

Only those programs defined under the current configuration will be presented for selection. For example, if you are currently using the 16-channel manifold, only those programs that were defined with the 16-channel configuration will be shown to you.

Select the supply bottle for use with all of the linked programs. This applies only to “V” version models.

Run a prime program before starting the first linked program.

Specify the number of the first strip to be washed.

Specify the number of microwell strips to be washed (the actual selectable range is based on the First Strip value):

1 to 12 for the 8- or 8s-channel manifold

1 to 8 for the 12-channel manifold

1 to 24 for the 16-channel manifold

Delete Link Program

To delete a Link program, follow the menu path shown here. Some of these items are options and some are prompts. The options are described below.

DEFINE DELETE MORE LINK SELECT LINK

PROGRAM OK TO DELETE?

Delete Link Program Options

Options Actions

Select Link Program Select the Link program you wish to delete.

OK to Delete? Select YES to delete the program.

BioTek Instruments, Inc.

Utility

|

89

Utility

The Utility (UTIL) menu options include TESTS, SETUP, and AUTOPRIME. These options are described below and on the following pages.

Tests

The system Self-Test (SLFCHK) and Checksum Test (CHKSUM) are performed automatically whenever the ELx50™ is turned on. The Tests option enables you to manually run both of these tests.

Self-Test (SLFCHK)

The Self-Test verifies system components, such as vacuum, manifold, and manifold to carrier positioning. To run the Self-Test, follow the menu path below:

UTIL TESTS SLFCHK

If the test completes successfully, the washer is ready for use. If the test fails, the washer will “chirp” and display an error code. Press

STOP

to stop the chirp, then look up the code in

Appendix A

,

Error Codes

.

Configuration Information (CHKSUM)

The Checksum Test verifies on-board software with internally recorded checksum values. To run the Checksum Test, follow the menu path below:

UTIL TESTS CHKSUM

The initial checksum test displays the on-board (basecode) software part number, version, and checksum:

4070xxx Verx.xx

CODE CHECKSUM: (DDA7)

After a few moments, the second checksum test displays the customer configuration software part number and version:

4070xxx-FW vx.xx.x

ELx50 Operator's Manual

90

| Chapter 3: Operation

Setup

Manifold (MANIFLD)

Your ELx50™ Microplate Strip Washer was initially configured for and shipped with an 8-, 12-, 8s-, or 16-channel wash manifold. It is important that the washer software configuration accurately represent the manifold installed in the instrument. If you change the hardware, you must also change the software configuration. See

Appendix B

,

ELx50 Manifold Variations

for more information.

To view or change the configuration of the manifold to 8-, 12-, 8s-, or 16channel, follow the washer menu path shown below:

UTIL SETUP MANIFLD SELECT MANIFOLD:

8 12* 8S 16**

* For washers with the aspiration carrier (ELx50/8FMW model), if 12 is selected for the manifold, aspiration using the aspiration carrier is not available.

**16 will not be displayed on ELx50/8FMW washers.

SENSOR

ELx50 washers manufactured after October 2007 support the

Liquid Level

Alert

™ option. With this option, each bottle has a level sensing device inside.

When a supply bottle’s fluid level drops to approximately 400-450 mL, or the waste bottle’s fluid level rises to approximately 1400-1450 mL, a sensor will activate. See page 59 for more information.

Sensor detection is disabled on the washer by default. To view or change the sensor detection setting , follow the washer menu path below:

UTIL SETUP SENSOR LIQUID LEVEL ALERT?

YES NO

BioTek Instruments, Inc.

Utility

|

91

AutoPrime

The AutoPrime option is important for preventive maintenance of the ELx50™.

Frequent priming keeps the washer manifold tubes moist and helps to prevent them from clogging. See

Chapter 4, Maintenance and Performance

Verification (IQ/OQ/PQ)

for more information on the AutoPrime feature.

The AutoPrime option may be enabled, disabled, or edited. If enabled, the program will run automatically after the washer is idle for the specified period of time. To enable or disable the AutoPrime feature, follow the menu paths shown below:

Enable:

UTIL AUTOPRIME ENABLE AUTOPRIME? YES

Disable:

UTIL AUTOPRIME ENABLE AUTOPRIME?

NO

To interrupt the AutoPrime feature, press the

STOP

key.

To edit the AutoPrime program, follow the washer menu path shown below.

DEFINE EDIT PRIME SELECT PRIME PROGRAM

Edit the Program Parameters OK TO SAVE

PROGRAM?

ELx50 Operator's Manual

92

| Chapter 3: Operation

ELx50™ Microplate Strip Washer Menu Maps

Main Menu (DEFINE function path for Create and Edit)

RUN

CREATE

SELECT

PROGRAM TYPE

DEFINE

EDIT

SELECT

PROGRAM TYPE

MAINT

COPY

UTIL

DELETE

WASH

SELECT

WASH PROGRAM

ENTER

PROGRAM NAME

SELECT

PLATE TYPE*

PRIME

SELECT

PRIME PROGRAM

ENTER

PROGRAM NAME

PRIME

VOLUME

RATE

PRIME?

DURATION

DISP

SELECT

DISP PROGRAM

ENTER

PROGRAM NAME

SELECT

PLATE TYPE*

VOLUME

FLOW RATE

HEIGHT

ASPIR

SELECT

ASPIR PROGRAM

ENTER

PROGRAM NAME

SELECT

PLATE TYPE*

ASPIRATE

TYPE**

BOT ASPIRATE

TIME**

ASPIRATE

HEIGHT

DISP POS ASPR POS

BEFORE START?

VOLUME

FLOW RATE

METHOD DISP ASPIR

RATE

DELAY

ASPIR?

CROSSWISE

HEIGHT

CROSSWISE

HORIZ POS

NUMBER

OF CYCLES

WASH FORMAT

PLATE / STRIP

SOAK/SHAKE?

YES / NO

SOAK

DURATION

SHAKE

BEFORE SOAK?

SHAKE DURATION

DISPENSE

VOLUME

DISPENSE

FLOW RATE

DISPENSE

HEIGHT

HORIZONTAL

DISP POS

BOTTOM

WASH FIRST?***

BOTTOM

ASPIRATE

HEIGHT

HORIZONTAL

ASPR POS

ASPIRATION

RATE

ASPIRATE

DELAY

CROSSWISE

ASPIR?

CROSSWISE ON

SHAKE

INTENSITY

PRIME

AFTER SOAK?

PRIME

VOLUME

PRIME

FLOW RATE

DISP VOLUME

BOTTOM

FLOW RATE

BOTTOM

DISP HEIGHT

BOTTOM

HORIZ POS

PRIME

BEFORE START?

ALL / FINAL

CROSSWISE

HEIGHT

CROSSWISE

HORIZ POS

FINAL

ASPIRATION?

FINAL

ASPIR DELAY

PRIME

VOLUME

PRIME

FLOW RATE

SOAK

SELECT

SOAK PROGRAM

ENTER

PROGRAM NAME

SOAK

DURATION

SHAKE

BEFORE SOAK?

SHAKE DURATION

SHAKE

INTENSITY

LINK

SELECT

LINK PROGRAM

ENTER

PROGRAM NAME

SELECT

PLATE TYPE*

SELECT

LINK n TYPE

(up to 10)

* PLATE TYPE only applies to the

16-channel manifold.

**ASPIRATE TYPE and BOT ASPIRATE

TIME only apply to the 8FMW model.

*** The Bottom Wash options do not apply to the 16-channel manifold.

BioTek Instruments, Inc.

ELx50™ Microplate Strip Washer Menu Maps

|

93

Main Menu (DEFINE function path for Copy and Delete)

RUN DEFINE MAINT UTIL

CREATE EDIT COPY DELETE

SELECT

PROGRAM TYPE

WASH

SELECT

WASH PROGRAM

ENTER

PROGRAM NAME

OK TO COPY?

YES / NO

PRIME

SELECT

PRIME PROGRAM

ENTER

PROGRAM NAME

OK TO COPY?

YES / NO

DISP

SELECT

DISP PROGRAM

ENTER

PROGRAM NAME

OK TO COPY?

YES / NO

ASPIR

SELECT

ASPIR PROGRAM

ENTER

PROGRAM NAME

OK TO COPY?

YES / NO

RUN DEFINE MAINT UTIL

CREATE EDIT COPY DELETE

SELECT

PROGRAM TYPE

SOAK

SELECT

SOAK PROGRAM

ENTER

PROGRAM NAME

OK TO COPY?

YES / NO

WASH

SELECT

WASH PROGRAM

OK TO DELETE?

YES / NO

PRIME

SELECT

PRIME PROGRAM

OK TO DELETE?

YES / NO

DISP

SELECT

DISP PROGRAM

OK TO DELETE?

YES / NO

ASPIR

SELECT

ASPIR PROGRAM

OK TO DELETE?

YES / NO

SOAK

SELECT

SOAK PROGRAM

OK TO DELETE?

YES / NO

LINK

SELECT

LINK PROGRAM

ENTER

PROGRAM NAME

OK TO COPY?

YES / NO

LINK

SELECT

LINK PROGRAM

OK TO DELETE?

YES / NO

ELx50 Operator's Manual

94

| Chapter 3: Operation

Main Menu (RUN function path)

RUN DEFINE

WASH

SELECT

WASH PROGRAM

SELECT BUFFER**

A B C

PRIME WASHER?

YES / NO

FIRST STRIP*

NUMBER OF STRIPS*

PLACE PLATE IN

CARRIER AND PRESS

<START> KEY

WASH PROGRAM

RUNNING

WASH PROGRAM

COMPLETE

MAINT

PRIME

SELECT

PRIME PROGRAM

SELECT BUFFER**

A B C

CONNECT REAGENT

BOTTLE AND PRESS

<START> KEY

PRIME PROGRAM

RUNNING

PRIME PROGRAM

COMPLETE

UTIL

DISP

SELECT

DISPENSE PROGRAM

SELECT BUFFER**

A B C

PRIME WASHER?

YES / NO

FIRST STRIP*

NUMBER OF STRIPS*

PLACE PLATE IN

CARRIER AND PRESS

<START> KEY

DISPENSE PROGRAM

RUNNING

DISP PROGRAM

COMPLETE

ASPIR

SELECT

ASPIRATE PROGRAM

FIRST STRIP*

NUMBER OF STRIPS*

PLACE PLATE IN

CARRIER AND PRESS

<START> KEY

ASPIRATE PROGRAM

RUNNING

ASPIR PROGRAM

COMPLETE

SOAK*** LINK

SELECT

LINK PROGRAM

FIRST STRIP*

NUMBER OF STRIPS*

PLACE PLATE IN

CARRIER AND PRESS

<START> KEY

RUNNING

COMPLETE

If the Enter key is pressed at the WASH, PRIME, DISPENSE, ASPIRATE, or LINK PROGRAM

COMPLETE screen, then the user is returned to the corresponding SELECT PROGRAM screen. For example, after a Wash program completes, pressing ENTER returns the user to the SELECT WASH PROGRAM screen.

* FIRST STRIP and NUMBER OF STRIPS are not displayed for Aspirate programs using the aspiration carrier (ELx50/8FMW model).

** The SELECT BUFFER option only appears for ELx50™ washers equipped with the multiple valves hardware.

*** A Soak program cannot be run independently; it must be selected within a Link program. Soak parameters can be defined within a Wash program (see “Main Menu,

DEFINE function path for Create and Edit”).

BioTek Instruments, Inc.

ELx50™ Microplate Strip Washer Menu Maps

|

95

Main Menu (MAINT, UTIL function path)

RUN DEFINE MAINT UTIL

SELECT

MAINT PROGRAM

CONNECT RINSE AND

PRESS <START>

MAINTENANCE

RUNNING PRESS

<STOP> KEY TO QUIT

COMPLETE

RUN DEFINE MAINT UTIL

TESTS

SLFCHK CHKSUM MANIFLD

PERFORM DISPLAY

SYSTEM TEST CHECKSUM 8 12 8s 16*

DISPLAY

SOFTWARE

CONFIGURATION

SELECT MANIFOLD

SETUP

SENSOR

LIQUID LEVEL ALERT?

YES / NO

* “16” is not displayed on ELx50/8FMW washers.

AUTOPRIME

ENABLE

AUTOPRIME?

YES / NO

AUTOPRIME

INTERVAL

ELx50 Operator's Manual

96

| Chapter 3: Operation

BioTek Instruments, Inc.

Chapter 4

Maintenance and

Performance Verification

(IQ/OQ/PQ)

This chapter describes how to maintain, store, decontaminate, and verify performance of the ELx50™.

Overview ................................................................................98

Recommended Maintenance/Performance Verification Schedules ....98

Warnings & Precautions ............................................................101

Maintenance Programs .............................................................102

Prime Programs.......................................................................103

Daily Maintenance....................................................................104

Overnight Maintenance .............................................................106

Monthly Maintenance................................................................108

Long-Term Storage Maintenance................................................115

Storing the ELx50 ....................................................................116

Decontamination .....................................................................117

Performance Verification ...........................................................122

Installation Qualification (IQ).....................................................123

Operational Qualification (OQ) ...................................................123

Performance Qualification (PQ) ..................................................124

Self-Test and Checksum Test.....................................................125

Liquid Tests for the 8-, 8s-, and 12-Channel Manifolds ..................126

Liquid Tests for the 16-Channel Manifold .....................................130

Liquid Test for the Aspiration Carrier (ELx50/8FMW) .....................134

98

| Chapter 4: Maintenance and Performance Verification (IQ/OQ/PQ)

Overview

This chapter describes a set of procedures to be performed regularly to maintain equipment in top condition and also to confirm its operation to specification. For example, during normal operation, salt crystals may build up and clog the valves and tubes.

Adherence to the

Recommended Maintenance

and

Performance Verification

Schedules

on the following pages will reduce this problem, extend the life of the washer, and verify that the washer continues to meet performance specifications.

The

Maintenance

section of this chapter (starting on p. 102) includes instructions for the following:

• Daily, Overnight, and Monthly Maintenance

• Long-Term Storage Maintenance

• Decontamination

The

Performance Verification (IQ/OQ/PQ)

section of this chapter (starting on p. 122) includes:

• System and Checksum Tests

• Evacuation Efficiency Tests

• Dispense Precision Tests

Maintenance/Performance and Operational Qualification

logs are provided at the end of this chapter.

Recommended Maintenance and Performance

Verification Schedules

The schedules on the following pages recommend maintenance and verification tasks, the frequency with which each task should be performed, the pre-defined onboard program that can be run to perform the task, and where to find more information.

BioTek Instruments, Inc.

Recommended Maintenance and Performance Verification Schedules

|

99

Recommended Maintenance Schedule

TASK

Protein-residual removal, p. 102

Daily maintenance program: p. 104

DAY_RINSE

AutoPrime, p. 104

Overnight maintenance program: p. 106

OVERNIGHT_LOOP

Periodic rinse and soak program: p. 107

RINSE_AND_SOAK

Clean bottles, p. 109

Check/empty waste bottle, p. 109

Clean manifold p. 109

Clean mist shield, p. 109

Clean aspirate and dispense tubes, p. 110

Clean microplate carrier and aspiration carrier

(ELx50™/8FMW), p. 111 and p. 112

Clean tubing check valves, p. 114

Remove liquid from washer parts program: p. 115

LONG_SHUTDOWN

Decontaminate program: p. 117

DECONTAMINATION

Frequency

Daily Monthly

Long-

Term

Storage

This schedule summarizes BioTek’s recommended maintenance tasks, and indicates approximately how often each task should be performed.

Daily and periodic routines and minimal guidelines for frequency are listed. Beyond that, it is difficult for BioTek to recommend a fixed frequency for each task to be performed. The risk and performance factors of your assays should determine the frequency with which to conduct these tasks. Therefore, BioTek recommends you develop a maintenance schedule for your washer based on the characteristics of the fluids used and the washer’s activity level.

ELx50 Operator's Manual

100

| Chapter 4: Maintenance and Performance Verification (IQ/OQ/PQ)

Recommended Performance Verification Schedule

TASK

System Self-Test, p. 125

Checksum Test, p. 125

8-, 8s- 12-Channel Manifolds:

Evacuation Efficiency Test, p. 127

Dispense Precision Test, p. 128

16-Channel Manifold:

Evacuation Efficiency Test, p. 131

Dispense Precision Test, p. 132

Aspiration Carrier (ELx50/8FMW):

Evacuation Efficiency Test, p. 134

Installation

Qualification

Operational

Qualification

Initially &

Annually

Performance

Qualification

Monthly

It is important to note that the risk and performance factors associated with your assays may require that some or all of the procedures be performed more frequently than presented in the schedule.

BioTek Instruments, Inc.

Warnings & Precautions

|

101

Warnings & Precautions

Please read the following before performing any Maintenance procedures:

Warning! Internal Voltage.

Turn off and unplug the instrument for all maintenance operations.

Warning!

Wear protective gloves when handling contaminated instruments. Gloved hands should be considered contaminated at all times; keep gloved hands away from eyes, mouth, nose, and ears.

Warning!

Mucous membranes are considered prime entry routes for infectious agents. Wear eye protection and a surgical mask when there is a possibility of aerosol contamination. Intact skin is generally considered an effective barrier against infectious organisms; however, small abrasions and cuts may not always be visible. Wear protective gloves when handling contaminated instruments.

Caution: Chemical Compatibility with Washers.

Some chemicals may cause irreparable damage to washers. The following chemicals have been deemed safe for use in washers: buffer solutions

(such as PBS), saline, surfactants, deionized water, 70% ethyl, isopropyl, or methyl alcohol, 40% formaldehyde, and 20% sodium hydroxide.

Never use acetic acid, DMSO, or organic solvents. Other chemicals may cause severe damage to the instrument. Contact BioTek prior to using any other chemicals. Refer to

Appendix C, Chemical Compatibility

.

Caution: Sodium Hypochlorite.

Do not expose any part of the instrument to the recommended diluted sodium hypochlorite solution

(bleach) for more than 20 minutes. Prolonged contact may damage the instrument surfaces. Be certain to rinse and thoroughly wipe all surfaces.

Important!

Do not immerse the instrument, spray it with liquid, or use a “wet” cloth on it. Do not allow water or other cleaning solution to run into the interior of the instrument. If this happens, contact BioTek’s

Technical Assistance Center.

ELx50 Operator's Manual

102

| Chapter 4: Maintenance and Performance Verification (IQ/OQ/PQ)

Maintenance Programs

The ELx50™ onboard software contains pre-defined maintenance programs, including

DAY_RINSE, RINSE_AND_SOAK, OVERNIGHT_LOOP, DECONTAMINATION, and

LONG_SHUTDOWN.

These programs are accessible from the washer main menu, under the

MAINT

option.

To run a maintenance program:

1. From the Main Menu, select

MAINT

.

2. Select the desired maintenance program.

START

key to begin. To halt the maintenance program in progress, press

STOP

.

4. When the program is complete, press

ENTER

to return to the Main Menu.

Important!

Solutions containing proteins, such as bovine serum albumin (BSA), will compromise the washer's performance over time unless a strict maintenance protocol is adhered to. It is recommended that a daily rinse and overnight soaking with phosphate buffered saline (PBS) be conducted. Rinse well with deionized water to remove the

PBS in the morning. On a weekly or monthly basis, flush with 0.1-0.5N NaOH, followed by neutralization with an equivalent normality (0.1-0.5N) of HCl. Rinse well with deionized water to remove the HCl.

Important!

Keep the manifold in a wetted condition to ensure proper evacuation of fluid. BioTek recommends changing the Soak Duration to 18 hours for overnights and weekends when the washer is not being used. This will decrease flow problems and is a more trouble-free way to operate the washer.

BioTek Instruments, Inc.

Prime Programs

|

103

Prime Programs

Each pre-defined maintenance program runs a pre-defined prime program. The charts below show the default prime program parameters. These parameters can be modified.

To view or edit the parameters for a particular prime program (and thus, the associated

Maintenance program):

1.

2.

From the Main Menu, follow the path below:

DEFINE EDIT PRIME Select the program

Press the

Enter

key to cycle through the parameters.

Maintenance Program: DAY_RINSE

Prime Program:

Prime Volume:

Flow Rate:

Soak After Prime?

P_DAY_RINSE

60 mL

6

No

Maintenance Program: RINSE_AND_SOAK

Prime Program:

Prime Volume:

Flow Rate:

Soak After Prime?

Soak Duration:

P_RINSE_AND_SOAK

60 mL

6

Yes

5 minutes*

Maintenance Program: OVERNIGHT_LOOP

Prime Program:

Prime Volume:

Flow Rate:

Soak After Prime?

Soak Duration:

P_RINSE_LOOP

40 mL

6

Yes

4 hours

ELx50 Operator's Manual

104

| Chapter 4: Maintenance and Performance Verification (IQ/OQ/PQ)

Daily Maintenance

For all daily maintenance programs, ensure that the supply bottle contains at least 200 mL of rinse solution, and that the waste bottle is empty.

DAY_RINSE

Daily maintenance involves flushing the washer with an appropriate reagent at the beginning of the day (BioTek recommends using a buffer solution in the morning).

The

DAY_RINSE program is recommended for the morning. This should be done on

the same day the microplates are washed. This procedure helps prevent the aspirate and dispense tubes from clogging between washes.

AUTOPRIME

Frequent priming keeps the washer manifold tubes moist and helps to prevent them from clogging. BioTek recommends using the

AUTOPRIME

feature.

AUTOPRIME is particularly useful if the washer is used intermittently throughout the day. Once the feature is activated, the AutoPrime prime program runs automatically at specified intervals until the feature is deactivated. If you prefer not to have the

AutoPrime feature on, it would be necessary to run the

DAY_RINSE

program periodically during the day.

To enable or disable the AutoPrime feature, from the Main Menu follow the paths shown below.

Enable:

UTIL AUTOPRIME ENABLE AUTOPRIME? YES

Disable:

UTIL AUTOPRIME ENABLE AUTOPRIME? NO

The AutoPrime program can be edited to change the length of the soak period, eliminate the soak period, or to change the interval between AutoPrime cycles.

To view or edit the AutoPrime program, follow the path below:

DEFINE EDIT PRIME SELECT PRIME PROGRAM

Edit the Program Parameters OK TO SAVE PROGRAM?

AutoPrime Program Options, and a Defaults and Ranges table are on the following page.

BioTek Instruments, Inc.

Daily Maintenance

|

105

AutoPrime Program Options

Options Actions

Prime Volume Specify the volume of the buffer or rinse used to prime the manifold. The volume range is 1 to 200 mL. Do not specify more volume than can be safely contained in the waste bottle during idle periods.

Prime Flow Rate Specify the rate at which the priming fluid is pumped into the dispense tubes. The flow rate options range from 1 to 9, where

1 = Slowest, 9 = Fastest. The washer display shows the corresponding µl/well/sec value for each option.

Soak after Prime? Select YES to soak the manifold tubes after they have been primed.

After the dispense tubes have been primed, the manifold raises to a trough clearance height and the carrier returns to its “home” position so that the dispense and aspirate tubes are above the carrier trough.

The soak solution is then dispensed to the trough. The vacuum pump is turned off and the tubes are lowered into the trough to soak. After soaking, the vacuum pump is turned on and the trough aspirated.

Soak Duration Soak Duration represents the time to allow the dispense and aspirate tubes to be soaked in wash buffer, rinse, or disinfectant.

The range is from 1 minute to 18 hours.

Note: By connecting a rinse bottle to the washer, the AutoPrime feature can be used to automatically rinse the washer.

Parameter

Interval

Prime Volume

Prime Flow Rate

Default

10 minutes

60 mL

6

Soak After Prime? Enabled

Soak Duration 3 hours, 50 minutes

Selectable Range

10 to 600 minutes

1 to 200 mL

1 to 9, or

1 to 7 for 12-channel manifold

Enabled/Disabled

1 minute to 18 hours

ELx50 Operator's Manual

106

| Chapter 4: Maintenance and Performance Verification (IQ/OQ/PQ)

Overnight Maintenance

OVERNIGHT_LOOP

When left overnight or for the weekend, the

OVERNIGHT_LOOP program is recommended. Overnight/multi-day maintenance involves flushing all wash solution out of the instrument, then periodically rinsing and soaking the tubes to keep them moist.

1.

2.

3.

4.

Connect a supply bottle containing deionized water and run

DAY_RINSE

, to flush any wash solution out of the washer.

Run the

OVERNIGHT_LOOP

program to prime and soak the tubes. It is important to keep the manifold in a wetted condition to ensure proper evacuation of fluid. BioTek recommends changing the Soak Duration to 18 hours for overnights and weekends when the washer isn't being used. This will decrease flow problems and is a more trouble-free way to operate the washer.

The

OVERNIGHT_LOOP

does require that the washer be kept plugged in and turned on. As an alternative, run the

RINSE_AND_SOAK

maintenance program and turn off the instrument after the soak begins. This leaves the tubes soaking in the priming trough until the instrument is turned on again.

The

OVERNIGHT_LOOP

program will run continuously until the Stop key is pressed. After pressing the Stop key, choose

YES

to home all axes.

To run the

OVERNIGHT_LOOP program, start at the Main Menu and follow the washer menu paths shown below:

Washers not equipped with automatic valve switching:

Select MAINT Press the Option key until

OVERNIGHT_LOOP appears CONNECT RINSE AND PRESS

<START> KEY MAINTENANCE RUNNING PRESS <STOP> TO

QUIT.

Washers equipped with automatic valve switching:

Select MAINT Press the Option key until

OVERNIGHT_LOOP appears RINSE VALVE? A B C START

MAINTENANCE? YES MAINTENANCE RUNNING PRESS <STOP>

TO QUIT.

BioTek Instruments, Inc.

Overnight Maintenance

|

107

RINSE_AND_SOAK

The

RINSE_AND_SOAK program rinses the wash manifold and leaves the tubes soaking in the trough for the duration of the soak.

To run the

RINSE_AND_SOAK program, start at the Main Menu and follow the washer menu paths shown below.

Washers not equipped with automatic valve switching:

Select MAINT Press the Option key until

RINSE_AND_SOAK appears CONNECT RINSE AND PRESS

<START> KEY MAINTENANCE RUNNING PRESS <STOP> TO

QUIT.

Washers equipped with automatic valve switching:

Select MAINT Press the Option key until

RINSE_AND_SOAK appears RINSE VALVE? A B C START

MAINTENANCE? YES MAINTENANCE RUNNING PRESS <STOP>

TO QUIT.

ELx50 Operator's Manual

108

| Chapter 4: Maintenance and Performance Verification (IQ/OQ/PQ)

Monthly Maintenance

Tools and Supplies

• 70% isopropyl alcohol

• Deionized or distilled water

• Dish soap or other mild cleaner

• Aspirate tube stylus (PN 4072042) and Dispense tube stylus (PN 4072043)

• Phillips head screwdriver

• Lint-free disposable towels

• Soft-bristled brush

• Rubber stopper for waste bottle (PN 4072034)

• Spare vacuum inline filter (PN 48146)

• Spare vacuum pump muffler (PN 4073009) (washers shipped after January 2008)

• Spare check valves (2) (PN 68061)

• Spare manifold gasket (rubber seal) (PN 4072012)

• Spare gasket for aspiration carrier (rubber seal, ELx50™/8FMW) (PN 4072096)

• Spare vent plug o-rings for aspiration carrier (2 per plug, ELx50/8FMW)

(PN 19497)

Part numbers are subject to change over time. Please contact BioTek

Customer Care if you have any questions.

Warning! Internal Voltage.

Always turn off and disconnect the ELx50 from its power supply for all cleaning operations.

Important!

Do not immerse the instrument, spray it with liquid, or use a

“wet” cloth. Do not allow the cleaning solution to run into the interior of the instrument. If this happens, contact the BioTek Service Department.

Do not soak the keypad—

this will cause damage. Moisten a clean cloth with deionized or distilled water and wipe the keypad. Dry immediately with a clean, dry cloth.

Do not apply lubricants

to manifold gasket, manifold channel end seals, fittings, tubing connections, or any surface that is part of the fluid path. The use of any lubricant on the fluid handling components will interfere with the aspirating and dispensing performance, and may cause irreparable damage to these components.

BioTek Instruments, Inc.

Monthly Maintenance

|

109

Cleaning the Bottles

• Clean the supply and waste bottles with deionized water before the first use, before each refill, and if they have been idle for any length of time.

• Accumulated algae, fungi, or mold may require decontamination.

Decontamination is described on page 117.

• Empty the waste bottle. Rinse the cover every time the wash or rinse bottles are filled. Periodic decontamination may be required. Do not allow the waste bottle to overfill!

Cleaning the Manifold

Regular rinsing helps to keep the manifold clean, the aspirate and dispense tubes clear, and it increases the life of the tubing.

If you suspect a particular problem is related to the manifold (for example, clogged tubes can result in poor or uneven aspiration or dispensing), you should perform a thorough cleaning of the manifold. When necessary, follow the

Decontamination

procedure on page 117 to disinfect the manifold and tubing.

To clean the manifold:

1. Run the system “dry”: connect an empty supply bottle, and prime the washer until the tubing is empty.

2. Turn off and unplug the washer.

3.

Moisten a lint-free disposable towel with water, or with water and mild detergent.

Do not soak the cloth.

4. Remove the mist shield if it is attached. See

Figure 19

. Wipe inside and outside surfaces of the mist shield with the towel. Wipe the top surface of the instrument base, and all exposed surfaces of the instrument.

5. If detergent was used, wipe all surfaces with a cloth moistened with water.

6. Use a clean, dry cloth to dry all wet surfaces.

7. Disconnect the tubing from the back of the manifold. See

Figures 11

through

17

.

8. Remove and clean the manifold:

Using a lint-free disposable towel moistened with water, or with water and mild detergent, thoroughly clean the outside of the dispense and aspirate tubes.

Run hot water through both the inlet and outlet fittings. Check to see if water comes out of all of the dispense and aspirate tubes. If not, soak the manifold in hot soapy water and repeat.

ELx50 Operator's Manual

110

| Chapter 4: Maintenance and Performance Verification (IQ/OQ/PQ)

Cleaning the Aspirate and Dispense Tubes

8-, 8s-, or 12-Channel Manifold:

5.

6.

7.

8.

9.

1.

2.

3.

4.

Using a Phillips screwdriver, unscrew the metal plate on top of the manifold (eight screws). Remove the gasket (rubber top seal) of the manifold.

Using hot soapy water and a soft-bristled brush, clean the dispense and aspirate chambers on the manifold top. Thoroughly clean the walls of the chambers.

Clean the insides of all of the aspirate and dispense tubes with the appropriate stylus (aspirate or dispense).

Rinse the manifold with deionized or distilled water. Look through the tubes to ensure tube openings are clear.

When satisfied, replace the gasket and gasket clamp. Do not overtighten.

Remount the manifold and replace the tubing. See

Figures 7

and

11

through

17

.

Replace the mist shield.

Run a Prime program using 200 mL.

Verify Dispenser performance visually or by completing the

Dispense

Precision Test Procedure

, page 128 for the 8-, 8s-, 12-channel manifold, and page 132 for the 16-channel manifold.

16-Channel Manifold:

Important!

Do

not

disassemble the 16-channel manifold. There are five screws on the top metal plate.

Removing any of these screws will cause the manifold to

fall out of alignment!

1.

2.

3.

4.

5.

6.

Remove the three black plugs and the T-shaped plug from the manifold.

Using the appropriate stylus (aspirate/dispense), clean the insides of all of the aspirate and dispense tubes.

Rinse the manifold with deionized or distilled water. Check to see if water comes out of all of the dispense and aspirate tubes.

When satisfied, replace all of the plugs.

Remount the manifold and replace the tubing. See

Figures 8

,

13

,

15a

,

15b

, and

16

.

Replace the mist shield.

BioTek Instruments, Inc.

Monthly Maintenance

|

111

7.

8.

Run a Prime program using 200 mL.

Verify dispenser performance visually or by completing the

Dispense

Precision Test Procedure

, page 128 for the 8-, 8s-, 12-channel manifold; page 132 for the 16-channel manifold.

Cleaning the Carrier

If liquid has overflowed onto the microplate carrier, some crystal buildup may occur, preventing the microplate from seating correctly onto the carrier. The aspiration carrier in the ELx50™/8FMW washer may also have some residual buildup, which can prevent proper seating of filter-bottom plates.

Refer to the instructions below and on the following page for cleaning the microplate and aspiration carriers. See

Figures 7

through

10

in

Chapter 2,

Installation

, for illustrations of the microplate and aspiration carriers.

Microplate Carrier

Important!

Each microplate carrier is specifically aligned to each washer.

Do not

adjust the nylon screws or interchange the carrier with another.

1.

2.

3.

4.

Turn the washer on, wait for the self-test to complete, then turn the washer off. Make sure the manifold and tubes are resting above the priming trough

(“home” position).

Remove the carrier:

Carefully lift the carrier front off the rail, avoiding contact with the manifold tubes.

Carefully lift the carrier rear off the rail, avoiding contact with the manifold tubes.

Clean the carrier using soap and water or 70% isopropyl alcohol.

Replace the carrier:

Hold the carrier over the rail so that the trough is to the rear of the washer.

Line up the pin on the underside of the carrier with the slot on the carrier transport.

Snap the two carrier clips onto the rail. The pin should sit in the slot.

ELx50 Operator's Manual

112

| Chapter 4: Maintenance and Performance Verification (IQ/OQ/PQ)

Aspiration Carrier (ELx50™/8FMW only)

4.

5.

6.

1.

2.

3.

7.

Turn the washer on, wait for the self-test to complete, then turn the washer off. Make sure the manifold and tubes are resting above the priming trough on the aspiration carrier (“home” position).

Disconnect the vacuum tubing between the aspiration carrier and washer.

Remove the aspiration carrier:

Carefully lift the carrier front off the rail, avoiding contact with the manifold tubes.

Carefully lift the carrier rear off the rail, avoiding contact with the manifold tubes.

Unlatch the plate clamp.

Rinse the carrier under tap water and allow to dry.

Inspect the vent port (labeled “C”) on the front of the carrier and the two slots for the vent plugs (labeled “A” and “B”) to ensure they are not plugged. If they are plugged, soak the carrier in hot soapy water first, followed by a rinse, and then allow it to air dry.

Inspect the plate seal gasket and replace it if it is torn. The gasket is held in place with six screws around the perimeter of the top of the carrier. See

Figure 32

on the next page.

If the vacuum grate (located under the gasket) needs to be cleaned, remove the grate, and rinse it under tap water and allow to dry before placing the grate and gasket in the carrier.

Note: The gasket is not symmetrical and only goes in one way.

When replacing the gasket, align the notch in the gasket with the small “bump” on the plate clamp, as shown in Figure 32.

8.

9.

10.

11.

Close the plate clamp.

Inspect the o-rings on the vent plugs (two o-rings per plug) and replace them if they are worn.

Rinse the vacuum tubing under tap water or soak in hot soapy water if extra cleaning is needed.

Place the aspiration carrier back into the washer:

Hold the aspiration carrier over the rail so that the priming/soaking trough is closest to the rear of the washer.

Snap the two carrier rail guides onto the carrier transport rail.

BioTek Instruments, Inc.

Monthly Maintenance

|

113

12.

13.

Reconnect the vacuum tubing between the aspiration carrier and the washer. Ensure that the tubing is connected to the correct port on the back of the washer, and that there is enough flex room for the tubing during carrier movement. Refer to

Connecting the Tubes and Bottles

in

Chapter 2, Installation

.

Verify performance of the aspiration carrier by performing the

Evacuation

Efficiency Test

for the carrier on page 134.

Figure 32: Replacing the Gasket in the Aspiration Carrier (ELx50™/8FMW model)

ELx50 Operator's Manual

114

| Chapter 4: Maintenance and Performance Verification (IQ/OQ/PQ)

Cleaning the Check Valves

In the event that the check valves become stuck or leak, you can either clean or replace them. To replace a check valve, contact BioTek Customer Service (check valve PN 68061). To clean a check valve:

1.

2.

3.

4.

Pull the tubing off the valve and pull the valve off the syringe pump port.

See

Figures 11a

and

12a

.

Insert the stylus into the feed end of the valve to hold it open (observe the arrow on the valve indicating the feed direction).

Flush with hot water. If necessary, the check valve body twists open to allow disassembly and cleaning of internal components. Note proper orientation for re-assembly.

Replace the valve and the tubing.

BioTek Instruments, Inc.

Long-Term Storage Maintenance

|

115

Long-Term Storage Maintenance

Before leaving the washer unused for a long period of time, you should run the

LONG_SHUTDOWN maintenance program below for the manifold and tubes. This program performs three steps:

1.

2.

Flushes and soaks the supply tubing and manifold with disinfectant.

Flushes the system with rinse.

3. Purges the system of fluid.

If your washer has the aspiration carrier (ELx50™/8FMW model) you should also disinfect the carrier as instructed on the following page.

LONG_SHUTDOWN

Prepare an aqueous solution of 0.5% sodium hypochlorite (NaClO, or bleach). Be sure to check the percent NaClO of the bleach you are using; this information is printed on the side of the bottle. Commercial bleach is typically 10% NaClO; if this is the case, prepare a 1:20 dilution. Household bleach is typically 5% NaClO; if this is the case, prepare a 1:10 dilution.

Fill one supply bottle with at least 200 mL of disinfectant (bleach solution) and one with at least 400 mL of rinse (deionized water). You also need one empty supply bottle (this is the “air” bottle). Empty the waste bottle. If the washer is equipped with automatic valve switching, connect all bottles.

To run the

LONG_SHUTDOWN program, follow the washer menu paths shown below.

Washers not equipped with automatic valve switching:

MAINT Press the Option key until LONG_SHUTDOWN appears CONNECT DISINFECTANT AND PRESS <START> KEY

MAINTENANCE RUNNING PRESS <STOP> TO QUIT.

Note:

While this program is running, you will need to periodically check the display panel and follow the instructions for changing the supply bottles.

Washers equipped with automatic valve switching:

MAINT Press the Option key until LONG_SHUTDOWN appears DISINFECTANT VALVE? A B C RINSE VALVE? A

B C AIR VALVE? A B C START MAINTENANCE? YES

MAINTENANCE RUNNING PRESS <STOP> TO QUIT.

ELx50 Operator's Manual

116

| Chapter 4: Maintenance and Performance Verification (IQ/OQ/PQ)

Disinfect the Aspiration Carrier (ELx50™/8FMW)

Do not insert a filter-bottom plate into the aspiration carrier for the following procedure. Disinfectant and rinse solutions will be dispensed into the empty carrier.

2.

3.

4.

5.

1.

6.

7.

8.

Prepare an aqueous solution of 0.5% sodium hypochlorite (NaClO, or bleach).

• Be sure to check the percent NaClO of the bleach you are using; this information is printed on the side of the bottle. Commercial bleach is typically 10% NaClO; if this is the case, prepare a 1:20 dilution. Household bleach is typically 5% NaClO; if this is the case, prepare a 1:10 dilution.

Fill one supply bottle with disinfectant (bleach solution).

Fill another supply bottle with rinse (deionized water).

Empty the waste bottle.

Insert the black vent plug that has the 0.020” diameter hole into the vent port on the front panel of the carrier. (See

Figure 10

in

Chapter 2, Installation

, for a close-up illustration of the vent port and vent plugs.)

Run a 600 μl Dispense program for 12 strips using the disinfectant.

Run an Aspirate program: select BOT for the ASPIRATE TYPE and 20

(seconds) for the BOT ASPIRATE TIME.

Run a 600 μl Dispense for 12 strips again, this time using the rinse, followed by a final 20-second bottom aspiration to purge the system.

Storing the ELx50™

1.

2.

3.

4.

5.

Follow the

Long-Term Storage Maintenance

steps, as described on the previous page and above.

Disconnect the washer from its power source.

Store the washer on a flat surface that is relatively free of vibration, in a dust- and particle-free environment.

Protect the washer from temperature extremes that can cause condensation within the unit and from corrosive fumes and vapors.

Store the washer under the following environmental conditions:

Temperature:

–20° to 50°C (-4° to 122°F)

Relative humidity:

10% to 85% (non-condensing)

BioTek Instruments, Inc.

Decontamination

|

117

Decontamination

Purpose

Any laboratory instrument that has been used for research or clinical analysis is considered a biohazard and requires decontamination prior to handling.

Decontamination minimizes the risk to all who come into contact with the instrument during shipping, handling, and servicing. Decontamination is required by the U.S.

Department of Transportation regulations.

Persons performing the decontamination process must be familiar with the basic setup and operation of the instrument.

Important!

BioTek Instruments, Inc. recommends the use of the following decontamination solutions and methods based on our knowledge of the instrument and recommendations of the Centers for Disease Control and

Prevention (CDC). Neither BioTek nor the CDC assumes any liability for the adequacy of these solutions and methods. Each laboratory must ensure that decontamination procedures are adequate for the Biohazard(s) they handle.

Wear prophylactic gloves when handling contaminated instruments. Gloved hands should be considered contaminated at all times; keep gloved hands away from eyes, mouth, nose and ears. Eating and drinking while decontaminating instruments is not advised.

Mucous membranes are considered prime entry routes for infectious agents. Wear eye protection and a surgical mask when there is a possibility of aerosol contamination. Intact skin is generally considered an effective barrier against infectious organisms; however, small abrasions and cuts may not always be visible. Wear protective gloves when performing the decontamination procedure.

ELx50 Operator's Manual

118

| Chapter 4: Maintenance and Performance Verification (IQ/OQ/PQ)

Tools and Supplies

• Sodium hypochlorite (NaClO or bleach)

• 70% isopropyl alcohol (as an alternative to bleach)

• Deionized or distilled water

• Safety glasses

• Surgical mask

• Protective gloves

• Lab coat

• Biohazard trash bags

• 125 mL beakers

• Clean cotton cloths or paper towels

Cleaning solution/disinfectant:

Prepare an aqueous solution of 0.5% sodium hypochlorite (NaClO, or bleach).

As an alternative, 70% isopropyl alcohol may be used if the effects of bleach are a concern. Be sure to check the percent NaClO of the bleach you are using; this information is printed on the side of the bottle. Commercial bleach is typically 10%

NaClO; if this is the case, prepare a 1:20 dilution. Household bleach is typically 5%

NaClO; if this is the case, prepare a 1:10 dilution.

Note: Do not prepare a stronger bleach solution than described here.

Extended exposure to high concentrations of bleach can deteriorate some waste system components.

Warning! Internal Voltage.

Turn off and disconnect the

ELx50™ from its power supply for all decontamination operations.

The bleach solution is caustic; wear gloves and eye protection when handling this solution.

Do not immerse the instrument, spray it with liquid, or use a “wet” cloth. Do not allow the cleaning solution to run into the interior of the instrument. If this happens, contact the BioTek Service

Department.

Do not soak the keypad—

this will cause damage.

BioTek Instruments, Inc.

Decontamination

|

119

Decontamination Procedure for External Components

7.

8.

9.

1.

2.

3.

4.

5.

6.

Prepare the cleaning solution as described on page 118.

Turn off and unplug the instrument.

Moisten a cloth with the bleach solution or alcohol.

Do not soak the cloth.

Wipe the keypad (do not soak). Wipe again with a clean cloth moistened with deionized or distilled water. Dry immediately with a clean, dry cloth.

Remove the mist shield, if it is attached. Wipe the mist shield’s inside and outside surfaces. Wipe the microplate carrier, top surface of the instrument base, supply bottles and tubing, and all exposed surfaces of the instrument.

Wait 20 minutes. Moisten a cloth with DI or distilled water and wipe all surfaces of the instrument that have been cleaned with the bleach solution or alcohol.

Use a clean, dry cloth to dry all wet surfaces.

Reassemble the instrument as necessary.

Discard the used gloves and cloths using a Biohazard trash bag and an approved Biohazard container.

Decontamination Procedure for Tubing and Manifold

1.

2.

3.

4.

Perform the Decontamination procedure described on the previous page.

Connect the ELx50™ to the power supply, and turn it on.

Prepare to run the

DECONTAMINATION

program, to decontaminate the internal tubing and manifold parts. This program will consume approximately 55 mL of disinfectant and 55 mL of rinse. Start with an empty waste bottle.

To run the

DECONTAMINATION

program, follow one of two washer menu paths shown below.

Note:

While the program is running, you will need to periodically check the display panel and follow the instructions for changing the supply bottles.

Washers not equipped with automatic valve switching:

MAINT Press the Option key DECONTAMINATION

CONNECT DISINFECTANT AND PRESS <START> KEY RUNNING:

DECONTAMINATION/ DISINFECTING CONNECT RINSE BOTTLE

AND PRESS <START> KEY RUNNING: DECONTAMINATION/

RINSING

ELx50 Operator's Manual

120

| Chapter 4: Maintenance and Performance Verification (IQ/OQ/PQ)

Washers equipped with automatic valve switching:

MAINT Press the Option key DECONTAMINATION

DISINFECTANT VALVE? A B C RINSE VALVE? A B C

START MAINTENANCE? YES MAINTENANCE RUNNING PRESS

<STOP> TO QUIT.

5. To ensure that all bleach is removed from the instrument, run the

DAY_RINSE program at least once after decontamination.

Alternate Decontamination Procedure for Tubing and Manifold

If you are unable to run the decontamination program due to a system failure, decontaminate the internal tubing and manifold parts as follows:

1.

2.

3.

4.

5.

6.

7.

8.

Prepare the cleaning solution as described on page 118.

Turn off and unplug the instrument.

Remove the tubing and manifold.

Soak the tubing and manifold in the bleach or alcohol solution.

Wait 20 minutes. Rinse the tubing and manifold with DI water.

Use a clean, dry cloth to dry all wet surfaces of the tubes and manifold.

Reassemble the instrument as necessary.

Discard the used gloves and cloths using a Biohazard trash bag and an approved Biohazard container.

Decontamination Procedure for Aspiration Carrier (ELx50/8FMW)

If your washer is the ELx50™/FMW model, perform the following instructions to decontaminate the aspiration carrier.

1.

2.

3.

4.

Prepare the cleaning solution as described on page 118.

Turn off and unplug the instrument.

Disconnect the tubing from the back of the aspiration carrier.

Remove the aspiration carrier:

• Carefully lift the carrier front off the rail, avoiding contact with the manifold tubes.

• Carefully lift the carrier rear off the rail, avoiding contact with the manifold tubes.

BioTek Instruments, Inc.

Decontamination

|

121

5.

6.

7.

8.

Unlatch the plate clamp.

Remove the vacuum grate, which is located under the plate seal gasket. The vacuum grate and gasket are held in place with six screws around the perimeter of the top of the carrier. See

Figure 32

on page 113.

Soak the carrier, plate clamp, vacuum grate, and gasket in the bleach solution for 20 minutes.

Rinse the carrier, plate clamp, vacuum grate, and gasket thoroughly to remove all bleach.

9.

10.

Use a clean, dry cloth to dry all wet surfaces.

Place the vacuum grate and gasket back into the carrier and replace the six perimeter screws.

Note: The gasket is not symmetrical and only goes in one way. When replacing the gasket, align the notch in the gasket with the small

“bump” on the plate clamp, as shown in

Figure 32

.

11.

12.

Close the plate clamp.

Replace the aspiration carrier:

• Hold the aspiration carrier over the rail so that the priming/soaking trough is closest to the rear of the washer.

• Snap the two carrier rail guides onto the carrier transport rail.

13. Reassemble the instrument, as necessary.

ELx50 Operator's Manual

122

| Chapter 4: Maintenance and Performance Verification (IQ/OQ/PQ)

Performance Verification

Performance verification for the ELx50™ involves three activities: qualification of installation and setup, qualification of routine capability, and qualification of long-term stability. This section refers to these activities as Installation Qualification (IQ), Operational

Qualification (OQ), and Performance Qualification (PQ), respectively.

Note: An Installation-Operational-Performance Qualification

(IQ-OQ-PQ) package (PN 4070530) for the ELx50™ is available for purchase. Contact your local dealer for more information.

The

Recommended Maintenance

and

Performance Verification Schedules

on pages 99 and 99 defines factory-recommended intervals for verification tests for a washer used two to five days a week. Verification tests include the following:

• The

System Self-Test

checks the vacuum pump and syringe pump, manifold, and microplate carrier positioning and operation. The

Checksum Test

verifies the basecode software against internal checksum values.

• The

Evacuation Efficiency Test

for the 8-, 8s-, 12-, and 16-channel manifolds measures the average residual volume per well after a wash program is performed. A solution is dispensed into all the wells of a previously weighed microplate. The wash program is run and the plate is weighed again in order to calculate the total residual, based on the weight difference. The lower the residuals per well, the better the evacuation efficiency of the washer.

The Evacuation Efficiency Test for the

aspiration carrier

verifies that the washer effectively aspirates all fluid down through the wells of a 96-well filter-bottom microplate. A Link program is run for dispensing deionized water into all wells of the plate, followed by bottom-aspiration of the wells. After the program completes, visual verification confirms that there is no residual fluid in any of the wells. If the Millipore filter-bottom plate recommended by BioTek is used, the test also measures the residual volume of the plate after evacuation.

• The

Dispense Precision Test

measures the variability of volumes dispensed from tube to tube across the manifold. A test dye solution is dispensed into a microplate. The optical density of each well is measured at 630 nm and the background at 450 nm is subtracted to account for scratches in the plate or particulates in the well.

The average percent error is calculated, and the amount dispensed for each well is calculated and the %Coefficient of Variation (%CV). It is the %CV, or the ratio expressed in percent, of the standard deviation of the distribution of fluid volumes in the wells, to the mean value of volume per well. The lower the CV, the better the uniformity across the manifold.

BioTek Instruments, Inc.

Installation Qualification (IQ)

|

123

Installation Qualification (IQ)

Installation Qualification

confirms that the washer and its components have been supplied as ordered and ensures that they are assembled and configured properly for your lab environment.

• The recommended IQ procedure consists of setting up the instrument as described in

Chapter 2, Installation

and then performing the System Self-Test and the

Checksum test.

• The IQ procedure should be performed initially (before the washer is used for the first time).

• The successful completion of the IQ procedure verifies that the instrument is installed correctly. The Operational Qualification procedure should be performed immediately following the successful IQ (see below).

Operational Qualification (OQ)

Operational Qualification

confirms that the washer operates according to specification initially and over time.

• The recommended OQ procedure consists of performing the System Self-Test and the Checksum, Evacuation Efficiency, and Dispense Precision tests.

• Your facility’s operating policies may also require that you perform an actual assay prior to accepting the washer for routine use. You should not use the data obtained from the first assay that utilizes the washer until you have confirmed that the package insert criteria have been met.

• The OQ procedure should be performed initially (before first use) and then routinely; the recommended interval is annually. It should also be performed after any major repair or upgrade to the hardware or software.

• Although out-of-tolerance failures will be detected by the OQ tests, results should be compared with those from the monthly Performance Qualification tests and previous OQ tests to monitor for trends.

• The successful completion of the OQ procedure, in combination with results that are comparable to previous PQ and OQ tests, confirms that the washer is performing consistently over time.

ELx50 Operator's Manual

124

| Chapter 4: Maintenance and Performance Verification (IQ/OQ/PQ)

Performance Qualification (PQ)

Performance Qualification

confirms that the washer consistently meets the requirements of the tests performed at your laboratory.

• The recommended PQ procedure consists of performing the System Self-Test and the Checksum, Evacuation Efficiency, and Dispense Precision tests.

• Your facility’s operating policies may also require that you routinely perform an actual assay, to confirm that the washer will consistently give adequate results for the assays to be run with it.

• These tests should be performed routinely; the recommended interval is monthly.

This frequency may be adjusted depending on the trends observed over time.

• The successful completion of the PQ procedure confirms that the washer is performing consistently under normal operating conditions.

BioTek Instruments, Inc.

Self-Test and Checksum Test

|

125

Self-Test and Checksum Test

The Self-Test (SLFCHK) and Checksum Test (CHKSUM) are performed automatically whenever the instrument is turned on. They can also be performed manually from the

ELx50™ Main Menu.

• The

Self-Test

checks the vacuum, manifold, and manifold-to-carrier movement.

• The

Checksum Test

compares the onboard software with internally recorded checksum values to ensure that no corruption has occurred.

Self-Test

To run the Self-Test, select the following from the Main Menu:

UTIL TESTS SLFCHK

If the test passes, the Main Menu is displayed and the washer is ready for use.

If the test fails, the washer will “chirp” and display an error code. Press

STOP

to stop the beep, then look up the code in

Appendix A, Error Codes

.

Configuration Information (CHKSUM)

To run the Checksum Test, select the following from the Main Menu:

UTIL TESTS CHKSUM

The initial checksum test displays the onboard (basecode) software part number, version, and checksum:

4070xxx Verx.xx

CODE CHECKSUM: (DDA7)

After a few moments, the second checksum test displays the customer configuration software part number and version:

4070xxx-FW vx.xx.x

This information is important if you need to contact BioTek TAC.

ELx50 Operator's Manual

126

| Chapter 4: Maintenance and Performance Verification (IQ/OQ/PQ)

Liquid Tests for the 8-, 8s- and 12-Channel

Manifolds

Evacuation Efficiency and Dispense Precision Tests

The

Evacuation Efficiency

test measures the average residual volume per well after evacuation. The lower the residuals per well, the better the evacuation efficiency of the washer. The

Dispense Precision

test measures the variability of the volumes dispensed from tube to tube across the manifold. It is the %CV, or the ratio expressed in percent, of the standard deviation of the distribution of fluid volumes in the wells, to the mean value of volume per well. The lower the CV, the better the uniformity across the manifold.

Materials

• Corning Costar

®

#3590 96-well plates (or equivalent)

• Precision balance with capacity of 100 g minimum and readability of

0.001 g resolution

• Pipettes and graduated beakers

• Microplate absorbance reader capable of dual wavelength reading at 630 and

450 nm

• Test solutions:

Solution #1: 0.1% Tween

®

20 in deionized water

Pipette 1 mL Tween 20 into 1 liter (1000 mL) of dionized water and mix well,

or

Pipette 10 mL of BioTek Solution #1 100X Concentrate Wetting Agent (PN

7773002) into 1 liter of deionized water and mix well.

Solution #2: Residual Test Solution

Mix 100 mL of Solution #1 with 0.0500 grams of FD&C #1 blue dye,

or

Mix 10 mL of BioTek Solution #2 10X Concentrate Blue Test Dye

(PN 7773001) with 90 mL of Solution #1 prepared above

Solution #3: Dispense Precision Solution

Mix 590 mL of deionized water with 10 mL of either Solution #2 prepared above.

Important!

The factory determined the pass/fail specifications for the following tests using the test solutions described here.

BioTek Instruments, Inc.

Liquid Tests for the 8-, 8s- and 12-Channel Manifolds

|

127

Note: PN 7773002 contains 10% Tween

®

20 in deionized water and

0.01% Sodium Azide as a preservative. PN 7773001 contains 5 g per liter of FD&C Blue #1, 0.1% Tween 20 in deionized water, and 0.01%

Sodium Azide as a preservative.

You may use your own buffer solution in place of Solution #1. If any tests fail using your own buffer, however, retry the tests using the suggested solutions.

Using pure DI water in place of Solution #1 is

not

recommended and will likely result in the failure of the unit to meet specifications.

There is a

Liquid Tests Worksheet

on page 129, for recording data reduction results. If your tests are failing, this information will be useful for BioTek TAC to help diagnose any problems.

Evacuation Efficiency Test Procedure

Important!

Specifications for the following Evacuation

Efficiency Test only apply when the correct microplate carrier is in use. Use the carrier that is specific to the manifold type

(see

Carrier Variations

in

Chapter 1, Introduction

).

Do

not

use the aspiration carrier for this test!

1.

2.

3.

4.

5.

Adjust the wash program for optimal residual. You can use your own wash program, or one of the ELx50’s onboard Wash programs, such as

COSTAR_FLAT.

Place a clean, dry 96-well microplate on the balance, then zero the balance.

Prime the washer with the test solution, then run the dispense program a c c u r a c y _ q c _ t e s t

. Run the wash program.

Weigh the microplate containing the residual.

Calculate and report the average residual by dividing the net weight by the number of wells dispensed.

Example:

NW

# wells dispensed

= average residual *

* The average residual must be < 0.002 grams per well.

If this test does not pass on the initial attempt, clean the aspirate tubes with a stylus and adjust the programmed aspiration height and horizontal aspiration position to minimize residual. Add a crosswise aspiration on the final aspiration step if needed, and then retest.

ELx50 Operator's Manual

128

| Chapter 4: Maintenance and Performance Verification (IQ/OQ/PQ)

Dispense Precision Test Procedure

1.

2.

3.

4.

Use a clean, dry 96-well microplate.

Prepare the washer to dispense test solution to the microplate (see

Materials

).

• The Test Solution should contain your buffer solution or 0.1% Tween

®

20 in deionized water. Add 1 mL Tween 20 into 1 liter (1000 mL) of deionized water.

• Add enough colored dye to the solution to produce an absorbance of between 0.70 and 1.30 OD when dispensing 300 μl into the microwell. The formulation will vary, depending on dye purity.

Prime the washer with the test solution, then run the dispense program accuracy_qc_test

. Use the number keys to enter the number of strips to process (6 for the 8- or 8s-channel manifold and 4 for the 12-channel manifold).

Read the plate in an absorbance reader (blank on air) using the dualwavelength method (630-450 nm), to reduce the influence of scratches and foreign particles that could be in the well. Print or export the results.

Note: If you are using one of BioTek’s keypad-based readers, such as the ELx800™ or ELx808™, ensure that the reader is not running in Rapid mode. To check this, select UTIL

R E A D

and cycle through the prompts until you see READ IN RAPID MODE?, then choose NO for an accurate result.

5.

Calculate and report the mean absorbance, standard deviation, and

% Coefficient of Variation (%CV) for the group of wells that were filled.

A %CV greater than 3.0% is a failure. If your result is greater than 3.0%, clean the dispense tubes with a stylus and retest.

Example:

Data:

Mean Absorbance

Standard Deviation

Calculation:

= <OD>

= SD

% Coefficient of Variation = SD

<OD>

= 1.026

= 0.010

= 0.010

1.026 x 100 = 0.98%

If this test does not pass on the initial attempt, use the stylus to clean the dispense tubes that have OD reading results lower than average, and then retest.

BioTek Instruments, Inc.

Liquid Tests for the 8-, 8s- and 12-Channel Manifolds

|

129

Liquid Tests Worksheet (8-, 8s-, and 12-Channel Manifolds)

Washer Serial Number:

Date:

Tested By:

Evacuation Efficiency Test

Total Residual Weight

Verification that no well varies considerably from the others

Mean Residual Weight (Total Residual Weight ÷ number of

wells processed)

Mean Residual Weight must be < 0.002 g

Dispense Precision Test

Standard Deviation

Calculate using a spreadsheet program

Mean Absorbance (sum of all wells ÷ number of wells

processed)

% Coefficient of Variation (%CV)

= (Standard Deviation ÷ Mean Absorbance) * 100

%CV value must be < 3.0

Comments:

grams

Pass Fail grams

Pass Fail

Pass Fail

ELx50 Operator's Manual

130

| Chapter 4: Maintenance and Performance Verification (IQ/OQ/PQ)

Liquid Tests for the 16-Channel Manifold

Evacuation Efficiency and Dispense Precision Tests

The

Evacuation Efficiency

test measures the average residual volume per well after evacuation. The lower the residuals per well, the better the evacuation efficiency of the washer. The

Dispense Precision

test measures the variability of the volumes dispensed from tube to tube across the manifold. It is the %CV, or the ratio expressed in percent, of the standard deviation of the distribution of fluid volumes in the wells, to the mean value of volume per well. The lower the CV, the better the uniformity across the manifold.

Materials

• Corning Costar

®

384-well plates

• Precision balance with capacity of 100 g minimum and readability of

0.001 g resolution

• Pipettes and graduated beakers

• Microplate absorbance reader capable of dual wavelength reading at 630 and

450 nm

• Test solutions:

Solution #1: 0.1% Tween

®

20 in deionized water

Pipette 1 mL Tween 20 into 1 liter (1000 mL) of dionized water and mix well,

or

Pipette 10 mL of BioTek Solution #1 100X Concentrate Wetting Agent* (PN

7773002) into 1 liter of deionized water and mix well.

Solution #2: Residual Test Solution

Mix 100 mL of Solution #1 with 0.0500 grams of FD&C #1 blue dye,

or

Mix 10 mL of BioTek Solution #2 10X Concentrate Blue Test Dye*

(PN 7773001) with 90 mL of Solution #1 prepared above.

Solution #3: Dispense Precision Solution

Mix 590 mL of deionized water with 10 mL of either Solution #2 prepared above.

Important!

The factory determined the pass/fail specifications for the following tests using the test solutions described here.

BioTek Instruments, Inc.

Liquid Tests for the 16-Channel Manifold

|

131

Note: PN 7773002 contains 10% Tween

®

20 in deionized water and

0.01% Sodium Azide as a preservative. PN 7773001 contains 5 g per liter of FD&C Blue #1, 0.1% Tween 20 in deionized water, and 0.01%

Sodium Azide as a preservative.

You may use your own buffer solution in place of Solution #1. If any tests fail using your own buffer, however, retry the tests using the suggested solutions.

Using pure DI water in place of Solution #1 is

not

recommended and will likely result in the failure of the unit to meet specifications.

There is a

Liquid Tests Worksheet

on page 133, for recording data reduction results. If your tests are failing, this information will be useful for BioTek TAC to help diagnose any problems.

Evacuation Efficiency Test Procedure

Important!

Specifications for the following Evacuation

Efficiency Test only apply when the correct microplate carrier is in use. Use the carrier that is specific to the

16-channel manifold.

1.

2.

3.

4.

5.

Adjust the wash program for optimal residual. You can use your own wash program, or one of the ELx50’s onboard Wash programs, such as

COSTAR_FLT16_384.

Place a clean, dry 384-well microplate on the balance, then zero the balance.

Prime the washer with the test solution, then run the dispense program a c c u r _ q c _ t e s t 3 8 4

. Run the wash program.

Weigh the microplate containing the residual.

Calculate and report the average residual by dividing the net weight by the number of wells dispensed.

Example:

NW

# wells dispensed

= average residual *

* The average residual must be < 0.004 grams per well.

If this test does not pass on the initial attempt, clean the aspirate tubes with the stylus and adjust the programmed aspiration height and horizontal aspiration position to minimize residual. Add a crosswise aspiration on the final aspiration step if needed, and then retest.

ELx50 Operator's Manual

132

| Chapter 4: Maintenance and Performance Verification (IQ/OQ/PQ)

Dispense Precision Test Procedure

1.

2.

3.

4.

Use a clean, dry 384-well microplate.

Prepare the washer to dispense test solution to the microplate (see

Materials

).

• The Test Solution should contain your buffer solution or 0.1% Tween

®

20 in deionized water. Add 1 mL Tween 20 into 1 liter (1000 mL) of deionized water.

• Add enough colored dye to the solution to produce an absorbance of between 0.70 and 1.30 OD when dispensing 300 μl into the microwell. The formulation will vary depending on dye purity.

Prime the washer with the test solution, then run the dispense program a c c u r _ q c _ t e s t 3 8 4

. Use the number keys to enter the number of strips to process (6 for the 16-channel manifold).

Read the plate in an absorbance reader (blank on air) using the dualwavelength method (630-450 nm), to reduce the influence of scratches and foreign particles that could be in the well. Print or export the results.

Note: If you are using one of BioTek’s keypad-based readers, such as the ELx800™ or ELx808™, ensure that the reader is not running in Rapid mode. To check this, select UTIL

R E A D

and cycle through the prompts until you see READ IN RAPID MODE?, then choose NO for an accurate result.

5.

Calculate and report the mean absorbance, standard deviation, and

% Coefficient of Variation (%CV) for the group of wells that were filled.

A %CV greater than 4.0% is a failure. If your result is greater than 4.0%, clean the dispense tubes with a stylus and retest.

Example:

Data:

Mean Absorbance

Standard Deviation

Calculation:

= <OD> = 1.200

= SD

% Coefficient of Variation = SD

<OD>

= 0.040

= 0.040

1.200 x 100 = 3.3%

If this test does not pass on the initial attempt, use the stylus to clean the dispense tubes that have OD reading results lower than average, and then retest.

BioTek Instruments, Inc.

Liquid Tests for the 16-Channel Manifold

|

133

Liquid Tests Worksheet (16-Channel Manifold)

Washer Serial Number:

Date:

Tested By:

Evacuation Efficiency Test

Total Residual Weight

Verification that no well varies considerably from the others

Mean Residual Weight (Total Residual Weight ÷ number of

wells processed)

Mean Residual Weight must be < 0.004 g

Dispense Precision Test

Standard Deviation

Calculate using a spreadsheet program

Mean Absorbance (sum of all wells ÷ number of wells

processed)

% Coefficient of Variation (%CV)

= (Standard Deviation ÷ Mean Absorbance) * 100

%CV value must be < 4.0

Comments:

grams

Pass Fail grams

Pass Fail

Pass Fail

ELx50 Operator's Manual

134

| Chapter 4: Maintenance and Performance Verification (IQ/OQ/PQ)

Liquid Test for the Aspiration Carrier

(ELx50/8FMW)

Evacuation Efficiency Test

The

Evacuation Efficiency

test for the aspiration carrier on the ELx50™/8FMW model verifies that the washer effectively aspirates all fluid down through the wells of a

96-well filter-bottom microplate. The test is based on visual verification that there is no residual fluid left in any of the wells. If the Millipore plate recommended by BioTek is used (see PN below), the test also measures the residual volume of the plate after evacuation. The lower the residual weight (< 1 gram), the better the evacuation efficiency of the washer.

Materials

• One new 96-well filter-bottom plate:

0.45 μm filter pore size (recommended: Millipore PN MSHVN4550)

• Precision balance with capacity of 100 g minimum and readability of 0.001 g resolution

• Deionized water

Although BioTek recommends that you use the Millipore filter-bottom plate listed above for this test, you may substitute other manufacturers’ filter-bottom plates that you are most familiar with using. Using a substitute plate, however, may vary your test

results.

There is a Liquid Test Worksheet on page 136, for recording test results. If your tests are failing, this information will be useful for BioTek TAC to help diagnose any problems.

Evacuation Efficiency Test Procedure

Important!

Before performing the Evacuation Efficiency

Test, ensure that the aspiration carrier is installed.

Perform the following test using the 0.45 μm filter pore plate:

1.

2.

Prime the washer using deionized water.

If you are using the Millipore PN MSHVN4550 plate, place the plate on the balance, then zero the balance.

BioTek Instruments, Inc.

Liquid Test for the Aspiration Carrier (ELx50/8FMW)

|

135

3. Ensure that there is proper vacuum. Use the vent port (labeled “C”) on the carrier and the two vent plugs (located in storage slots “A” and “B”) to vary the vacuum levels.

(See

Figure 9

in

Chapter 2

for an illustration of the vent port and the two vent plugs.)

• For a

low

vacuum level, leave the vent port on the carrier open. The vent port is a 0.047” diameter hole, which is equal to approximately

-64 mmHg.

• For a

medium

vacuum level, insert the gray vent plug into the vent port. The gray vent plug contains a 0.032”diameter hole, which is equal to approximately -116 mmHg.

• For a

high

vacuum level, insert the black vent plug into the vent port.

The black vent plug contains a 0.020” diameter hole, which is equal to approximately -240 mmHg.

The approximate vacuum levels listed above (-64, -116, and -240 mmHg) are “typical” only, and may vary, depending upon factors such as relative humidity, barometric pressure, and mechanical tolerances.

Important!

Ensure that the vent port is not completely sealed. If the vent port is sealed, the washer’s internal vacuum pump may collapse the vacuum reservoir bottle.

4.

5.

6.

7.

Run a Link program: select BTST.

Immediately after the program completes, visually verify that all wells have been evacuated.

Pat the bottom of the plate with a paper towel to remove any droplets.

If using the Millipore plate, weigh the plate and verify that the weight is less than 1 gram.

The 1 gram weight specification is specifically for Millipore part number MSHVN4550. An alternate plate type will need to be tested, and a specification determined for that plate.

8.

9.

Enter the weight in the worksheet on the following page.

If the test fails, repeat using another new plate.

ELx50 Operator's Manual

136

| Chapter 4: Maintenance and Performance Verification (IQ/OQ/PQ)

Liquid Test Worksheet (Aspiration Carrier)

Washer Serial Number:

Date:

Tested By:

Evacuation Efficiency Test

0.45 µm filter pore size, 96-well plate

Manufacturer brand and part number for the plate:

Visual verification that all wells have been evacuated:

Residual Weight:

Residual Weight (Millipore MSHVN4550 only) must be < 1 g

Pass Fail grams

Pass Fail

BioTek Instruments, Inc.

Maintenance/Performance and Operational Logs

|

137

Maintenance/Performance and Operational Logs

Maintenance/Performance and Operational logs are provided on the following pages.

ELx50 Operator's Manual

138

| Chapter 4: Maintenance and Performance Verification (IQ/OQ/PQ)

BioTek Instruments, Inc.

Empty waste bottl

Chapter 5

Troubleshooting

This chapter lists problems that you may experience with the

ELx50™ washer, and suggests possible solutions for these problems.

Overview ................................................................................142

Troubleshooting Checklist .........................................................142

Troubleshooting Charts.............................................................143

Washer

Fluid

Start-Up .................................................................143

Aspiration ...................................................................144

Delivery .....................................................................148

Fluid Leakage .....................................................................150

Carrier

Washer Manifold Movement ..................................................151

Scratches............................................................152

Syringe Drive Movement ......................................................152

142

| Chapter 5: Troubleshooting

Overview

Every effort has been made to ensure that the ELx50™ Microplate Strip Washer is extremely reliable and easy to use. Nevertheless, you could experience problems with the washer hardware, software, and/or accessories. This chapter offers information to help resolve these problems.

The table below is a concise summary of possible causes for many of the problems that may occur during operation of the washer.

The troubleshooting charts on the following pages provide detailed lists of problems, their possible causes, and possible solutions.

Troubleshooting Checklist

Periodic inspection of the washer, tubing, and bottles for the possible causes listed below may help to eliminate many of the problems listed on the following pages.

Check for the following:

9

9

9

9

Loose or incorrect connections

of power cord, power supply, waste and supply tubing and fittings

Incomplete or incorrect installation

of manifold and carrier,

attachment

of mist shield, or

placement

of microplate/strips in carrier

Accumulation of residue

on carrier or inside aspirate/dispense tubes, waste/supply bottles and tubing, manifold, inlet/outlet fittings, optional inline vacuum filter, or optional pump muffler (washers manufactured after

January 2008).

Worn

tubing, o-rings, inlet or outlet fittings, manifold gaskets or vacuum seals

9

Kinked or bent

aspirate/dispense tubes or waste/supply tubing

BioTek Instruments, Inc.

Troubleshooting Charts

|

143

Troubleshooting Charts

See

Figures 10

through

15

in

Chapter 2

,

Installation

, and

Figure 30

in

Chapter 4

,

Maintenance and Performance Verification (IQ/OQ/PQ)

for illustrations of the components referenced in the Troubleshooting Charts:

Figure 11

Washer Rear Panel (all models except ELx50™/8FMW)

Figure 12

Washer Rear Panel (ELx50/8FMW model)

Figure 13

Waste Tubing Connections (all models except ELx50/8FMW)

Figure 14

Waste Tubing Connections (ELx50/8FMW model with Aspiration

Carrier installed)

Figure 15a Supply Tubing Connections (front view)

Figure 15b Supply Tubing Connections (view of connections on rear panel)

Figure 16

Washer with Multiple Valve Tubing Connections (“V” versions of the

ELx50)

Figure 32

Replacing the Gasket in the Aspiration Carrier (ELx50/8FMW model)

Washer Start-Up

Problem

Display (LCD) not on.

Syringe, carrier, or manifold position error.

Possible Cause

Power cord not plugged in or loose.

Manifold or carrier is being obstructed.

Motor, sensor, or electrical problem.

Misaligned carrier or manifold.

Possible Solution

Check the power connection.

Remove any obstruction. Also, ensure that the manifold and carrier have been properly installed (see

Ch. 2

,

Installation)

.

Turn the washer off, wait at least 15 seconds, then turn it back on.

If it fails the self-test, the washer will

“chirp” and display an error code.

Press the

STOP

button to stop the beep, and look up the code in

App. A

,

Error Codes

. Contact BioTek TAC.

Contact BioTek TAC.

ELx50 Operator's Manual

144

| Chapter 5: Troubleshooting

Fluid Aspiration

Problem

Poor or uneven aspiration.

Possible Cause

Insufficient or no vacuum.

Insufficient or no vacuum.

Clogged aspirate tubes on the washer manifold.

Microplate aspiration height adjustment too high or too low.

Aspirate tubes not properly positioned horizontally in wells.

Possible Solution

If the optional in-line vacuum filter or optional pump muffler has been installed, the filter or muffler may need to be replaced (see

Connecting the Tubes and Bottles

in

Ch. 2

).

Firmly seat the waste bottle stopper.

Ensure that the tubing is connected properly (see

Fig. 13

:

Waste Tubing

Connections

).

Check all external tubing for kinks or accumulations of residue.

When aspiration begins, you should hear the vacuum pump turn on. If the pump does not turn on, contact

BioTek TAC.

When the vacuum pump turns on, remove the vacuum tubing from the back of the washer while the pump is on (see

Fig. 11: Washer Rear Panel

).

Put your finger over the port; if there is no vacuum, contact BioTek TAC.

Remove and clean the manifold (see

Cleaning the Manifold in Ch. 4

).

Change the programmed aspiration height (see

Define Wash

ASPIRATION Program Options

in

Ch. 3

).

If none of the tubes are bent, try adjusting the programmed horizontal aspirate position (see

Define Wash

ASPIRATION Program Options

in

Ch. 3

).

BioTek Instruments, Inc.

Troubleshooting Charts

|

145

Fluid Aspiration, cont’d

Problem

Poor or uneven aspiration,

cont’d.

Poor or uneven bottom aspiration in aspiration carrier

(ELx50/8FMW only)

Possible Cause Possible Solution

Microplate not level in carrier or strips not level in holder.

Reseat the microplate in the carrier or the strips in their holder.

Ensure that the carrier is clean (see

Cleaning the Carrier

in

Ch. 4

).

Try a different microplate or strip holder. If the problem is unresolved, the carrier may have to be realigned.

Contact BioTek TAC.

Movement rails misaligned. Contact BioTek TAC.

Clogged vacuum filter or pump muffler.

If the optional in-line vacuum filter or optional pump muffler has been installed, the filter or muffler may need to be replaced (see

Connecting the Tubes and Bottles

in

Ch. 2

).

Clogged filters in filter-bottom plates

(ELx50™/8FMW model only).

If the plate is old or has been used more than once, the filter paper may be dirty, resulting in poor aspiration.

Replace the old plate with a new one.

Ensure, also, that the vacuum grate is in place in the aspiration carrier. The grate is located under the plate seal gasket. If this grate is missing, the filter-bottom plate may have drops clinging to the underside of the plate.

A combination of thick solution and high vacuum level may cause clogging in some filter-bottom plates. For instance, assays with beads may become clogged if a high vacuum level is used. Try a lower vacuum level or filter pore size plate.

Bottom aspiration times (BOT

ASPIRATE TIME) that are too long may cause beads to stick in the bottom of the wells. Try shorter aspiration times initially (e.g., 10 seconds), and then increase the aspiration time until suitable aspiration is achieved.

ELx50 Operator's Manual

146

| Chapter 5: Troubleshooting

Fluid Aspiration, cont’d

Problem

Poor or uneven bottom aspiration in aspiration carrier

(ELx50™/8FM

W only), cont’d.

Possible Cause

Vacuum lines not connected or incorrectly connected.

Too much residual left in microwells after aspiration.

Seal did not form.

Incorrect plate size.

Vacuum reservoir bottle is full.

Washer program requires optimization for the plate type used.

Possible Solution

Ensure that the vacuum lines to the manifold and aspiration carrier are connected. Also, if the lines are reversed, the top and bottom functions will be reversed. If this is the case, reconnect the lines as appropriate.

If the filter-bottom plate is not seated correctly in the aspiration carrier, a seal will not form, resulting in no vacuum. Reposition the plate in the carrier. Also, check for residue on the carrier or carrier transport rails, and clean them, if necessary.

Check the rubber gasket on the aspiration carrier. If it is worn, there may be air leaks, resulting in a poor or no seal. If this is the case, replace the gasket.

If all 96 wells do not contain fluid, a seal will not form. Dispense to full plates only.

If the filter-bottom plate is the wrong size, a seal will not form, resulting in no vacuum. Try a plate with a different filter pore size.

If the vacuum reservoir bottle is full, the vacuum pump will draw fluid into the in-line filter or pump muffler and lower the vacuum level. A full bottle will also result in fluid being drawn into the pump, causing the vacuum level to drop. Empty the bottle.

To minimize the residuals, program a second or crosswise aspiration (see

Define Wash ASPIRATION Program

Options

in

Ch. 3

). Other options include decreasing the aspiration rate, increasing or decreasing aspiration height or tube position in the well, or adding a delay on the final aspiration.

BioTek Instruments, Inc.

Troubleshooting Charts

|

147

Fluid Aspiration, cont’d

Problem Possible Cause Possible Solution

Too much residual left in microwells after aspiration,

cont’d.

Waste bottle stopper not properly sealed or fittings not properly connected.

Firmly seat the waste bottle stopper.

Make sure the tubing is connected properly.

Ensure that the fittings are properly connected to the manifold.

Manifold out of alignment or not moving freely.

Check the fit of the gasket (rubber top seal) on the 8-, 8s-, or 12-channel manifold top (see

Cleaning the

Manifold

in

Ch. 4

). If the fit is loose, contact BioTek TAC.

Check for obstructions and ensure that the manifold has been properly installed (see

Ch. 2

,

Installation

). If no obstructions are found and the manifold is correctly installed, contact

BioTek TAC.

Reseat the microplate in the carrier or the strips in their holder.

Microplate not level in carrier or strips not level in holder.

Using 16-channel manifold to wash 8-well strips with round or "V" bottoms.

Aspirate tubes are bent.

Ensure that the carrier is clean (see

Cleaning the Carrier

in

Ch. 4

).

Try a different microplate or strip holder. If the problem is unresolved, the carrier may have to be realigned.

Contact BioTek TAC.

The 16-channel manifold may not be compatible with all 96-well format strips and plates. This manifold uses two aspirating tubes for each well, and it may be difficult for these tubes to reach the bottom of wells that are not flat.

Gently attempt to straighten the tubes using your fingers. If they remain bent, contact BioTek TAC.

Note:

Monthly performance of the Evacuation Efficiency test is recommended to verify that the residual volume per well after evacuation (aspiration) meets the specified criteria. See

Chapter 4

,

Performance Verification, Evacuation Efficiency Test

Procedure

.

ELx50 Operator's Manual

148

| Chapter 5: Troubleshooting

Fluid Delivery

Problem

Unable to dispense fluid.

Plate overfills

(floods).

Possible Cause Possible Solution

Inlet tube not connected at manifold or at bottle.

Supply tube inside the supply bottle is kinked or disconnected.

Clogged dispense tubes on the washer manifold.

Check Valve flow direction is incorrect.

Check Valves are stuck closed.

Dispense tube not plugged into syringe pump.

No wash or rinse fluid.

System not primed.

Faulty syringe pump.

Dispense height too high.

The aspirate tubes are too many steps above the microwells to prevent overflow.

Aspirate tubes hit bottom of trough during Prime or

Maintenance.

Fittings to manifold not connected.

Check all tubing (see

Connecting the

Tubes and Bottles

in

Ch. 2

).

Straighten or connect supply tube.

Make sure the end is cut as shown on page 55.

Remove and clean the manifold (see

Cleaning the Manifold

in

Ch. 4

).

Compare the flow direction of the

Check Valves with

Fig. 11: Washer

Rear Panel.

Switch or replace if necessary.

Clean Check Valves (see

Cleaning the

Check Valves

in

Ch. 4

).

Plug dispense tube into top connector of syringe pump.

Fill bottles with appropriate fluid.

Run a Prime program using 200 ml.

Contact BioTek TAC.

Change the Dispense Height in the program (see

Define Wash DISPENSE

Program Options

in

Ch. 3

).

Contact BioTek TAC.

Reconnect fittings.

Optional in-line vacuum filter or optional pump muffler plugged.

Rubber stopper on waste bottle is loose.

Replace or remove the filter or pump muffler.

Push rubber stopper into waste bottle to create a better seal.

Dispense rate too fast for volume selected.

Use slower dispense rate or lower volume (see

Define Wash DISPENSE

Program Options

in

Ch. 3

).

Insufficient or no vacuum. Firmly seat the waste bottle stopper.

Make sure tubing is connected properly.

BioTek Instruments, Inc.

Troubleshooting Charts

|

149

Fluid Delivery, cont’d

Problem Possible Cause Possible Solution

Plate overfills

(floods), cont’d.

Insufficient or no vacuum,

cont’d.

Check all external tubing for kinks or clogs.

The optional in-line vacuum filter or the optional pump muffler may need to be replaced.

When the program begins, you should be able to hear the vacuum pump turn on. If not, contact BioTek TAC.

If the vacuum pump is turning on, remove the vacuum tubing from the back of the washer while the vacuum pump is on. Put your finger over the port; if there is no vacuum, contact

BioTek TAC.

Contact BioTek TAC. Units with solenoid valves unable to dispense.

Solenoid valve not opening.

Uneven dispensing of fluid; wells not filled.

Solenoid valve tubing disconnected, or wrong solenoid inlet connected.

Manifold or tubing not adequately primed.

Reconnect or redirect tubing.

Run a Prime program using 200 ml.

Dripping dispense tubes.

Dispense flow rate too low. Select a higher flow rate (see

Define

Wash DISPENSE Program Options

in

Ch. 3

).

Clogged dispense tubes on the washer manifold.

Microplate aspiration height adjustment too high or too low.

Dispense tubing routed incorrectly.

Remove and clean the manifold (see

Cleaning the Manifold

in

Ch. 4

).

Change the Aspiration Height in the program (see

Define Wash

ASPIRATION Program Options

in

Ch. 3

).

The supply bottle tube must enter the syringe at the bottom. Also, check for holes in the manifold gasket.

Note:

Monthly performance of the Dispense Precision test is recommended to measure the variability of volumes dispensed from tube to tube across the manifold. See

Chapter

4

,

Performance Verification, Dispense Precision Test Procedure

.

ELx50 Operator's Manual

150

| Chapter 5: Troubleshooting

Fluid Leakage

Problem Possible Cause Possible Solution

Dripping dispense tubes.

Fluid leaking from manifold.

Dispense tubing routed incorrectly.

Defective seals.

The supply bottle tube must enter the syringe at the bottom.

If 8-, 8s-, or 12-channel manifold, replace manifold gasket. If 16-channel manifold, replace plugs and o-ring on

‘T’ plug (see

Cleaning the Manifold in

Ch. 4

).

Replace o-rings on washer manifold inlet fittings.

Teflon

®

sleeves are defective or missing.

Contact BioTek TAC.

Reposition sleeves.

Contact BioTek TAC.

Check valves are leaking. Clean or replace check valves (see

Cleaning the Check Valves

in

Ch. 4

).

Aspirate tubes only: vacuum too low.

Check waste connector tubes, and make sure they are properly connected to the manifold.

Check vacuum filter for clogging, replace if necessary.

Defective tubing connector or inlet tubing.

Push rubber stopper into waste bottle to create a better seal.

Use a slower Aspiration Rate in the program (see

Define Wash

ASPIRATION Program Options

in

Ch.

3

).

Contact BioTek TAC. Fluid leaking from underneath the instrument.

Fluid leaking from external tubing connector.

Leaking syringe seal.

Defective syringe piston.

Defective connector.

Worn tubing.

Worn seal (inlet or vacuum fitting).

Contact BioTek TAC.

Contact BioTek TAC.

Replace connector.

Replace tubing.

Replace filter or seal (see

Connecting the Tubes and Bottles

in

Ch. 2

).

BioTek Instruments, Inc.

Troubleshooting Charts

|

151

Carrier Movement

Problem Possible Cause Possible Solution

Microplate Carrier

Aspirate tubes not entering well correctly.

Aspirate tube(s) bent.

Microplate not properly seated or strips not level.

Push the stylus needle into the tube and then gently attempt to straighten the tube using your fingers. If it remains bent, contact BioTek TAC.

Reseat the microplate in the carrier or the strips in the holder. Make sure the plate is distributed squarely to the front edge of the carrier.

Aspirate

Position is programmed

Change the Horizontal Aspir Pos value in the program (see

Define Wash

too wide for a movement.

ASPIRATION Program Options

3

).

in

Ch.

Manifold tilted.

Make sure the carrier is clean (see

Cleaning the Carrier

in

Ch. 4

).

Check tubing for twists; add manifold screws (PN 12174).

Aspiration Carrier (ELx50/8FMW)

Aspiration carrier loses steps during program.

Insufficient flex room for tubing between carrier and washer.

Pushing the washer up against a wall or against the vacuum/dispense bottles can reduce the flex room needed by the tubing during carrier movement.

Re-position the washer and/or tubing to allow enough room. See

Connecting the Tubes and Bottles

in

Chapter 2

.

Washer Manifold Movement

Problem

Manifold position error.

Possible Cause

Manifold movement is blocked.

Possible Solution

Check orientation of microplate; A1 should be in the right rear corner of the plate carrier as you face the front of the instrument.

Check for and remove any obstructions.

Ensure that the manifold is resting on the manifold support bracket.

Contact BioTek TAC.

ELx50 Operator's Manual

152

| Chapter 5: Troubleshooting

Microplate Scratches

Problem

Scratches on microplate bottom.

Possible Cause

Microplate dispense or aspiration height adjustment too low.

Microplate not properly seated or strips not level.

Possible Solution

Change the Dispense or Aspiration

Height in the program (see

Define

Wash DISPENSE or ASPIRATION

Program Options

in

Ch. 3

).

Reseat microplate in carrier or strips in holder.

Make sure the carrier is clean (see

Cleaning the Carrier

in

Ch. 4

).

Try a different microplate or strip holder. If the problem is unresolved, the carrier may have to be realigned.

Contact BioTek TAC.

Syringe Drive Movement

Problem

Syringe drive gives movement errors.

Possible Cause

Check valves incorrectly mounted.

Obstructions in syringe path.

Possible Solution

Re-mount check valves.

Remove any obstructions.

BioTek Instruments, Inc.

Appendix A

Error Codes

This appendix lists error codes that may appear on the ELx50™ washer. If an error appears, refer to Chapter 5, Troubleshooting.

Contact BioTek’s Technical Assistance Center if further instructions are necessary.

Error Codes ............................................................................154

General Errors.........................................................................155

Fatal Errors.............................................................................157

Dispense Volume Invalid for Manifold Type (0F00 Error) ................158

154

| Appendix A: Error Codes

Error Codes

An error code is displayed on the ELx50™ Microplate Washer as a four-digit identifier.

The first character will be either 0, 1, 2, or A.

• 0 or 1 denote a noncritical (“General”) error, which means that the instrument will still respond to keypad input. See

General Errors

on the following page.

• A denotes a more serious error, which requires that the ELx50™ be turned off and then powered up. Upon restarting the washer, you should be able to enter commands into the keypad. See

Fatal Errors

on page 157.

BioTek Instruments, Inc.

General Errors

|

155

General Errors

General errors indicate non-fatal conditions that require attention. If an error is displayed, refer to

Chapter 5, Troubleshooting

. Contact BioTek’s Technical Assistance Center if further instructions are necessary.

Error Code Cause

ABORT_ERR

NO_SENSOR_ERR

AUTOCAL_JIG_ERROR

0100 Wash function aborted.

0200 Manifold motor couldn’t find optical sensor.

0201 Carrier motor couldn’t find optical sensor.

0202 Syringe motor couldn’t find optical sensor.

0300 Manifold motor couldn’t find edge of jig.

0301 Carrier motor couldn’t find edge of jig.

0302 Syringe motor couldn’t find edge of jig.

MOTOR_VERIFY_ERR 0400 Manifold motor failed positional verify.

0401 Carrier motor failed positional verify.

0402 Syringe motor failed positional verify.

UNDEF_PROG_TYPE_ERR 0500 Undefined program type specified.

PROGRAM_NUM_ERR 060n assay number chosen by user

(n = program number).

PROGRAM_SET_ERR

PROGRAM_DEL_ERR

SYR_VOLUME_TO_LARGE

0700 Invalid write attempt.

0800 Invalid delete attempt.

0900 Syringe volume too large. checksum failure; indicates corrupt assay program (n = program number).

MANIFOLD_AUTOCAL_ERR 0B00 Invalid manifold for calibration.

INVALID_BUFFER_ERR 0C00 Invalid buffer selected.

CAL_CHECKSUM_ERR 0D00 Failed calibration checksum test.

DISPENSE_VOLUME_ERR

CNFG_DATA_ERR

CNFG_CHECKSUM_ERR

0F00 Dispense volume invalid for manifold type

(see page 158).

1000 Necessary configuration data missing.

1100 Failed configuration checksum test.

Note:

Calibration errors (AUTOCAL_JIG_ERROR; MANIFOLD_AUTOCAL_ERR;

CAL_CHECKSUM_ERR) will normally be displayed only during calibration or repair of the instrument by BioTek TAC.

ELx50 Operator's Manual

156

| Appendix A: Error Codes

Error Code Cause

CAL_DATA_ERR

MOTOR_NOT_HOMED_ERR

1200 Necessary calibration data missing.

1300 Manifold motor not homed successfully.

1301 Carrier motor not homed successfully.

1302 Syringe motor not homed successfully.

AUTOCAL _ABORT_ERR

MALLOC_ERR

1400

1900

Autocal of washer was aborted.

Memory allocation failure.

Carrier present manifold.

Carrier present manifold.

SYRINGE_FMEA_ERR* 1C01 Sensor not found at start of syringe aspirate stroke.

*Failure Mode Effects Analysis 1C02 Sensor not open at start of dispense stroke.

1C03 Syringe aspirate stroke failure.

1C04 Sensor verify after last syringe stroke in program failed.

1C05 Sensor clear from home position too many steps.

MANIFOLD_POS_ERR 1D00 Manifold position not in table.

AUTOCAL_CHECKSUM_ERR 1E00 Failed autocal checksum.

INVALID_LINK_ERR 2400 Failed to get program to view/run.

CHECK_FLUID_START_ERR 2800

CHECK_FLUID_END_ERR 2900 the waste bottle’s fluid level has risen above

1400/1450 mL. Or, the Liquid Level Alert option is enabled (via SETUP > SENSORS) but the select box is not properly connected to the washer. See “Liquid Level Alert” in Chapter 3 under

General Usage Guidelines

.

INCOMPATIBLE_HW_ERR 2A00 Both the bottom aspirate and the external valves jumpers are set. This is not a valid hardware configuration.

3100 Error running shake/soak protocol. SOAK_SHAKE_ERR

Notes:

Calibration errors (CAL_DATA_ERR; AUTOCAL_ABORT_ERR) will normally be displayed only during calibration or repair of the instrument by BioTek TAC.

Syringe errors (SYRINGE_FMEA_ERR) indicate movement problems.

BioTek Instruments, Inc.

Fatal Errors

|

157

Fatal Errors

Fatal errors indicate conditions that require immediate attention. If a fatal error is displayed, contact BioTek’s Technical Assistance Center for further instructions.

Error

TCB_NOT_AVAIL_ERR

Code

A100

Cause

Task control block not available.

NOT_AVAIL_ERR

CHECKSUM_ERR

A300

A301

A302

A304

A305

A306

A400

Manifold motor not available.

Carrier motor not available.

Syringe motor not available.

Software timer not available.

Display not available.

Quick flash memory not available.

Failed code checksum test on power-up; memory corruption.

POWER_ERR A500 Power dropped below safe level. flash

Memory corruption; contact TAC.

QFLASH_ERR

HEAP_CORRUPTED_ERR

A700

A900

Quick flash memory read did not match write.

Memory corruption; contact TAC.

Memory manager corruption detected.

Contact TAC.

Note:

Unavailability of a motor, software timer, display or quick flash memory

(NOT_AVAIL_ERR) may be caused by loose cables. Contact TAC.

ELx50 Operator's Manual

158

| Appendix A: Error Codes

Dispense Volume Invalid for Manifold Type

(0F00 Error)

An 0F00 error at run-time indicates a validation error in the washer program, and that the current dispense volume/flow rate combination is invalid for the washer manifold type.

The ELx50™ basecode does not perform this type of validation during program creation.

To eliminate this error, edit the washer program and change the Dispense Volume and/or

Dispense Flow Rate so that they are compatible with your manifold type.

The following table shows the valid Dispense Volume ranges (in µl/well) for each

Dispense Flow Rate/Washer Manifold combination:

BioTek Instruments, Inc.

Appendix B

ELx50 Manifold Variations

This appendix lists the parts you need to purchase and the steps you need to take to reconfigure the ELx50™ Washer to work with a different manifold.

ELx50 Manifold Variations .........................................................160

Reconfiguring the Washer.....................................................160

Steps ..................................................................161

Steps ...................................................................162

160

| Appendix B: ELx50 Manifold Variations

ELx50 Manifold Variations

Your ELx50™ Microplate Strip Washer was initially configured for and shipped with either an 8-, 8s-, 12-, or 16-channel manifold. The 8s-channel manifold has short dispense tubes for special applications. The 16-channel manifold is typically used for 384-well plates (16-well strips), but it can also wash 96-well plates (8-well strips).

Your washer can be configured to work with a different manifold. The level of difficulty for reconfiguration depends on the type of manifold you have now and the type you plan to use.

Each manifold comes with a microplate carrier that is specific to the manifold type

(see

Carrier Variations

in

Chapter 1

). In addition to the microplate carrier,

ELx50/8FMW washers also have an aspiration carrier for bottom aspiration of filter-bottom plates.

Reconfiguring the Washer

Refer to the following table to find out what parts you should purchase and the steps you need to take to reconfigure the washer. The hardware and software steps are described on the following pages.

For ELX50/8, ELX50/12 and ELX50/8FMW

Note:

If you have the ELx50/8FMW model, you can have either an 8-, 8s-, or 12-channel manifold. However, only the 8- and 8s-channel manifolds are compatible with the aspiration carrier.

If your current manifold is:

And you plan to use this manifold: You need to:

12-channel 8- or 8s-channel

8- or 8s-channel 8s- or 8-channel

• Purchase the new manifold.

• Perform the hardware steps.

• Perform the software steps.

• Purchase the new manifold.

• Perform the hardware steps.

• Perform the software steps.

8- or 8s-channel 12-channel • Purchase the new manifold.

• Perform the hardware steps.

• Perform the software steps.

BioTek Instruments, Inc.

ELx50 Manifold Variations

|

161

For ELX50™/16

If your current manifold is:

And you plan to use this manifold: You need to:

16- or 8-channel 8-, 8s-, or 16-channel

• Purchase the new manifold.

• Perform the hardware steps.

• Perform the software steps.

Hardware Steps

Refer to

Figures 7

through

10

in

Chapter 2

for installation of the manifold and microplate/aspiration carriers. Refer to

Figures 20

and

21

in

Chapter 2

for removal of the manifold and microplate/aspiration carriers.

To replace a washer manifold:

1. Remove the washer's mist shield.

2. Detach the tubes from the currently installed manifold.

3. Remove the thumbscrews (three if 16-channel manifold, otherwise two).

4. Lift the manifold up off the manifold support pins.

5. Orient the new manifold so that the tubes are pointed down, and the fittings are pointed toward the back of the washer.

6. Slide the manifold gently down over the support pins.

7. Replace the thumbscrews.

8. Replace the tubing.

9. Replace the washer's mist shield.

To replace the microplate carrier with the aspiration carrier:

1. Remove the microplate carrier.

2. Install the aspiration carrier.

3. Connect the aspiration carrier tubing.

To replace the aspiration carrier with a microplate carrier:

1. Disconnect the aspiration carrier tubing.

2. Remove the aspiration carrier.

3.

Install the microplate carrier.

ELx50 Operator's Manual

162

| Appendix B: ELx50 Manifold Variations

Software Steps

To specify an 8-, 12-, 8s-, or 16-channel manifold, follow the washer menu path shown below.

U T I L

Î SETUP Î MANIFLD Î SELECT MANIFOLD:

8 1 2 8 S 1 6 *

* 16 is not available on ELx50™/8FMW models.

To use the aspiration carrier:

1. Select 8 or 8S from the menu path above for the manifold.

2. Create an aspirate-only program and select BOT for the Aspirate Type.

3.

Enter a Bottom Aspirate Time (BOT ASPIRATE TIME).

See

Chapter 3

for complete washer menu maps.

BioTek Instruments, Inc.

Appendix C

Chemical Compatibility

This appendix contains a name and materials listing of each

ELx50™ component that will come into contact with chemicals dispensed during operation or maintenance. If there are questions about compatibility of chemicals with the ELx50, contact BioTek.

Components and Materials Composition ......................................164

164

| Appendix C: Chemical Compatibility

Components and Materials Composition

Continuous contact with harsh chemicals is not recommended. It is best if the washer is rinsed with deionized water after contact with any strong acid, base, or solvent.

Components Material Composition

Syringe seal, buffer bottle

Fittings in bottles, inline fittings, check valves, waste bottle

Check valve seals

Dispense tubes, syringe piston, check valve spring, carrier spring

Syringe cylinder, carrier

Polyethylene

Polypropylene

Viton

316 SS

Manifold

Manifold mounting bracket

Assay plates

Tubing, manifold bumpers

Manifold seal, waste bottle stopper

Vacuum pump muffler

Noryl (polyphenylene ether + polystyrene)

CPVC (chlorinated polyvinyl chloride)

Anodized aluminum

Polystyrene

Silicone

EP (ethylene propylene)

Body: Nylon; filter: borosilicate glass microfibers with PVDF Fluorocarbon Resin binder

Paint Polyurethane

The optional vacuum pump muffler (PN 4073009) is included with

ELx50™ washers that are equipped with the new vacuum pump

(washers manufactured after January 2008).

BioTek Instruments, Inc.

Appendix D

Default Programs

This appendix lists all default wash, prime, dispense, aspirate, and maintenance programs available onboard the ELx50™.

Default Programs.....................................................................166

166

| Appendix D: Default Programs

Default Programs

WASH

COSTAR_FLAT

COSTAR_FLT16_96

COSTAR_FLT16_384

COSTAR ROUND

NUNC_FLAT

NUNC_FLT16_96

NUNC_FLT16_384

NUNC_ROUND

PRIME

Autoprime

NEW_BUFFER_PRIME

P_DAY_RINSE

ELx50™/8 ELx50/12 ELx50/16 ELx50/8FMW*

X X X X

X X X X

X

X

X

X X X

X X X

X X X X

X

X X X X

X X X X

X X X X

X X X X

X X X X P_DECON1

P_DECON2

P_RINSE_AND_SOAK

X X X X

X X X X

P_RINSE_LOOP X X X X

DISPENSE

accuracy_qc_test X X X X accur_qc_test384 X

ASPIRATE

residual_qc_test X X X X resid_qc_test384 X

MAINTENANCE

DAY_RINSE X X X X

RINSE_AND_SOAK X X X X

OVERNIGHT_LOOP X X X X

LONG_SHUTDOWN X X X X

DECONTAMINATION X X X X

* ELx50/8FMW model: The microplate carrier for the ELx50/8FMW model should be installed when running any of the onboard programs. The aspiration carrier does not support these programs, and is specifically designed for bottom aspiration of filter-bottom plates.

BioTek Instruments, Inc.

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

Download PDF

advertisement

Table of contents