Extech Instruments EX542 12 Function Wireless True RMS Industrial MultiMeter/Datalogger User's Guide
Extech Instruments EX542 is a Wireless TRMS Multimeter that is packed with features to make your work easier. With its ability to measure AC/DC Voltage, AC/DC Current, Resistance, Capacitance, Frequency, Duty Cycle, Diode Test, and Continuity plus Thermocouple Temperature, this meter is perfect for a wide range of applications. It also has the capability to store and recall data, making it easy to keep track of your measurements. Additionally, its rugged design and waterproof construction make it ideal for use in harsh environments.
विज्ञापन
विज्ञापन
User's Guide
Wireless TRMS Multimeter
Model EX542
Introduction
This meter measures AC/DC Voltage, AC/DC Current, Resistance, Capacitance, Frequency
(electrical & electronic), Duty Cycle, Diode Test, and Continuity plus Thermocouple Temperature. It can store and recall data. It features a waterproof, rugged design for heavy duty use. This meter can transmit data wirelessly when linked to a PC. Proper use and care of this meter will provide many years of reliable service.
Safety
WARNING
This symbol adjacent to another symbol, terminal or operating device indicates that the operator must refer to an explanation in the Operating Instructions to avoid personal injury or damage to the meter.
This WARNING symbol indicates a potentially hazardous situation, which if not avoided, could result in death or serious injury.
CAUTION
MAX
1000V
This CAUTION symbol indicates a potentially hazardous situation, which if not avoided, may result damage to the product.
This symbol advises the user that the terminal(s) so marked must not be connected to a circuit point at which the voltage with respect to earth ground exceeds (in this case) 1000 VAC or VDC.
This symbol adjacent to one or more terminals identifies them as being associated with ranges that may, in normal use, be subjected to particularly hazardous voltages. For maximum safety, the meter and its test leads should not be handled when these terminals are energized.
This symbol indicates that a device is protected throughout by double insulation or reinforced insulation.
PER IEC1010 OVERVOLTAGE INSTALLATION CATEGORY
OVERVOLTAGE CATEGORY I
Equipment of OVERVOLTAGE CATEGORY I is equipment for connection to circuits in which measures are taken to limit the transient overvoltages to an appropriate low level.
Note – Examples include protected electronic circuits.
OVERVOLTAGE CATEGORY II
Equipment of OVERVOLTAGE CATEGORY II is energy-consuming equipment to be supplied from the fixed installation.
Note – Examples include household, office, and laboratory appliances.
OVERVOLTAGE CATEGORY III
Equipment of OVERVOLTAGE CATEGORY III is equipment in fixed installations.
Note – Examples include switches in the fixed installation and some equipment for industrial use with permanent connection to the fixed installation.
OVERVOLTAGE CATEGORY IV
Equipment of OVERVOLTAGE CATEGORY IV is for use at the origin of the installation.
Note – Examples include electricity meters and primary over-current protection equipment
2 EX542-en-GB_v1.5 06/15
SAFETY INSTRUCTIONS
This meter has been designed for safe use, but must be operated with caution. The rules listed below must be carefully followed for safe operation.
1. NEVER apply voltage or current to the meter that exceeds the specified maximum:
Input Protection Limits
V DC or V AC mA AC/DC
A AC/DC
Frequency, Resistance, Capacitance,
Duty Cycle, Diode Test, Continuity
1000VDC/AC rms
500mA 1000V fast acting fuse
10A 1000V fast acting fuse (20A for 30 seconds max every 15 minutes)
1000VDC/AC rms
Surge Protection: 8kV peak per IEC 61010
2. USE EXTREME CAUTION when working with high voltages.
3. DO NOT measure voltage if the voltage on the "COM" input jack exceeds 1000V above earth ground.
4. NEVER connect the meter leads across a voltage source while the function switch is in the current, resistance, or diode mode. Doing so can damage the meter.
5. ALWAYS discharge filter capacitors in power supplies and disconnect the power when making resistance or diode tests.
6. ALWAYS turn off the power and disconnect the test leads before opening the covers to replace the fuse or batteries.
7. NEVER operate the meter unless the back cover and the battery and fuse covers are in place and fastened securely.
8. If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
3 EX542-en-GB_v1.5 06/15
Controls and Jacks
1. 40,000 count LCD display
2. STORE(<RECALL) button
3. MAX/MIN (_)button
4. MODE button
5. RANGE(SETUP) button
6. Function switch
7. mA, µA and 10A input jacks
8. COM input jack
9. Positive input jack
1
2
6
10. Backlight button
11. EXIT(AC+DC) button
12. HOLD(PEAKHOLD>) button
13. REL(+) button
Note: Tilt stand and battery compartment are on rear of unit.
Symbols and Annunciators
7
3
4
5
•))) Continuity
Diode test status
-9 ) (capacitance)
(10
-3 ) (volts, amps)
A Amps
) (ohms)
RF transmitter active
%
Ohms
Percent (duty ratio)
PEAK
Hz Hertz V
Peak Hold
Volts
Relative
DC Direct current
ºF Degrees Fahrenheit
MAX Maximum
SET Set up parameter
HOLD
ºC
MIN
Display hold
Degrees Centigrade
Minimum
S second
AC +DC Alternating current + Direct current
STO Store
RCL Recall
Auto Power off enabled Backlight
4 EX542-en-GB_v1.5 06/15
9
8
13
12
11
10
Operating Instructions
WARNING: Risk of electrocution. High-voltage circuits, both AC and DC, are very dangerous and should be measured with great care.
1. ALWAYS turn the function switch to the OFF position when the meter is not in use.
2. If “OL” appears in the display during a measurement, the value exceeds the range you have selected. Change to a higher range.
DC VOLTAGE MEASUREMENTS
CAUTION: Do not measure DC voltages if a motor on the circuit is being switched ON or OFF.
Large voltage surges may occur that can damage the meter.
1. Set the function switch to the VDC position.
2. Insert the black test lead banana plug into the negative COM jack.
Insert the red test lead banana plug into the positive V jack.
3. Touch the black test probe tip to the negative side of the circuit.
Touch the red test probe tip to the positive side of the circuit.
4. Read the voltage in the display.
AC VOLTAGE (FREQUENCY, DUTY CYCLE) MEASUREMENTS
WARNING: Risk of Electrocution. The probe tips may not be long enough to contact the live parts inside some 240V outlets for appliances because the contacts are recessed deep in the outlets. As a result, the reading may show 0 volts when the outlet actually has voltage on it. Make sure the probe tips are touching the metal contacts inside the outlet before assuming that no voltage is present.
CAUTION: Do not measure AC voltages if a motor on the circuit is being switched ON or OFF.
Large voltage surges may occur that can damage the meter.
1. Set the function switch to the VAC/Hz/% position.
2. Insert the black test lead banana plug into the negative COM jack.
Insert red test lead banana plug into the positive V jack.
3. Touch the black test probe tip to the neutral side of the circuit.
Touch the red test probe tip to the “hot” side of the circuit.
4. Read the voltage in the main display and the frequency in the right auxiliary display
6. Read the frequency in the main display.
8. Read the % of duty cycle in the main display.
9. With ACV in the main display, press EXIT for 2 seconds to measure AC+DC.
5 EX542-en-GB_v1.5 06/15
mV VOLTAGE MEASUREMENTS
CAUTION: Do not measure mV voltages if a motor on the circuit is being switched ON or OFF.
Large voltage surges may occur that can damage the meter.
1. Set the function switch to the mV position.
2. Press the MODE button to indicate “DC”.or ““AC ”, or in AC range press EXIT for two seconds and chose ”AC+DC”
3. Insert the black test lead banana plug into the negative COM jack.
Insert the red test lead banana plug into the positive V jack.
4. Touch the black test probe tip to the negative side of the circuit. Touch the red test probe tip to the positive side of the circuit.
5. Read the mV voltage in the main display.
6. With AC mV in the main display, press EXIT for 2 seconds to measure AC+DC.
DC CURRENT MEASUREMENTS
CAUTION: Do not make 20A current measurements for longer than 30 seconds. Exceeding 30 seconds may cause damage to the meter and/or the test leads.
1. Insert the black test lead banana plug into the negative COM jack.
2. For current measurements up to 4000µA DC, set the function switch to the µA position and insert the red test lead banana plug into the µA/mA jack.
3. For current measurements up to 400mA DC, set the function switch to the mA position and insert the red test lead banana plug into the µA/mA jack.
4. For current measurements up to 20A DC, set the function switch to the 10A/HZ/% position and insert the red test lead banana plug into the 10A jack.
6. Remove power from the circuit under test, then open up the circuit at the point where you wish to measure current.
7. Touch the black test probe tip to the negative side of the circuit.
Touch the red test probe tip to the positive side of the circuit.
8. Apply power to the circuit.
9. Read the current in the display.
6 EX542-en-GB_v1.5 06/15
AC CURRENT (FREQUENCY, DUTY CYCLE) MEASUREMENTS
CAUTION: Do not make 20A current measurements for longer than 30 seconds. Exceeding 30 seconds may cause damage to the meter and/or the test leads.
1. Insert the black test lead banana plug into the negative COM jack.
2. For current measurements up to 4000µA AC, set the function switch to the µA position and insert the red test lead banana plug into the µA/mA jack.
3. For current measurements up to 400mA AC, set the function switch to the mA position and insert the red test lead banana plug into the µA/mA jack.
4. For current measurements up to 20A AC, set the function switch to the 10A/HZ/% position and insert the red test lead banana plug into the 10A jack.
6. Remove power from the circuit under test, then open up the circuit at the point where you wish to measure current.
7. Touch the black test probe tip to the neutral side of the circuit.
Touch the red test probe tip to the “hot” side of the circuit.
8. Apply power to the circuit.
9. Read the current in the display. In the 10AAC range, right auxiliary display frequency.
10. Press and hold the MODE button to indicate “Hz”.
11. Read the frequency in the display.
12. Momentarily press the MODE button again to indicate “%”.
13. Read the % duty cycle in the display.
14. Press and hold the MODE button to return to current measurement.
15. With ACA in the main display, press EXIT for 2 seconds to measure AC+DC.
RESISTANCE MEASUREMENTS
WARNING: To avoid electric shock, disconnect power to the unit under test and discharge all capacitors before taking any resistance measurements. Remove the batteries and unplug the line cords.
1. Set the function switch to the Ω CAP position.
2. Insert the black test lead banana plug into the negative COM jack.
Insert the red test lead banana plug into the positive jack. the ”on the display.
4. Touch the test probe tips across the circuit or part under test. It is best to disconnect one side of the part under test so the rest of the circuit will not interfere with the resistance reading.
5. Read the resistance in the display.
7 EX542-en-GB_v1.5 06/15
CONTINUITY CHECK
WARNING: To avoid electric shock, never measure continuity on circuits or wires that have voltage on them.
1. Set the function switch to the Ω CAP
2. Insert the black lead banana plug into the negative COM jack.
Insert the red test lead banana plug into the positive jack.
4. Touch the test probe tips to the circuit or wire you wish to check.
5. If the resistance is less than approximately 35 , the audible signal will sound. If the circuit is open, the display will indicate “OL”.
DIODE TEST
1. Set the function switch to the Ω CAP
2. Insert the black test lead banana plug into the negative COM jack and the red test lead banana plug into the positive V jack. on the display.
4. Touch the test probes to the diode under test. Forward voltage will typically indicate 0.400 to 0.700V. Reverse voltage will indicate “OL”. Shorted devices will indicate near 0V and an open device will indicate “OL” in both polarities.
CAPACITANCE MEASUREMENTS
WARNING: To avoid electric shock, disconnect power to the unit under test and discharge all capacitors before taking any capacitance measurements. Remove the batteries and unplug the line cords.
1. Set the rotary function switch to the Ω CAP
2. Insert the black test lead banana plug into the negative COM jack.
3. Insert the red test lead banana plug into the positive V jack.
5. Touch the test leads to the capacitor to be tested.
6. Read the capacitance value in the display
8 EX542-en-GB_v1.5 06/15
TEMPERATURE MEASUREMENTS
1. Set the function switch to the Temp position.
2. Insert the Temperature Probe into the input jacks, making sure to observe the correct polarity.
4. Touch the Temperature Probe head to the part whose temperature you wish to measure. Keep the probe touching the part under test until the reading stabilizes (about 30 seconds).
5. Read the temperature in the display.
Note: The temperature probe is fitted with a type K mini connector. A mini connector to banana connector adaptor is supplied for connection to the input banana jacks.
FREQUENCY (DUTY CYCLE) MEASUREMENTS (ELECTRONIC)
1. Set the rotary function switch to the Hz/% position.
2. Insert the black lead banana plug into the negative COM jack and the red test lead banana plug into the positive Hz jack.
3. Touch the test probe tips to the circuit under test.
4. Read the frequency on the display.
6. Read the % duty cycle in the display.
9 EX542-en-GB_v1.5 06/15
% 4 – 20mA MEASUREMENTS
1. Set up and connect as described for DC mA measurements.
2. Set the rotary function switch to the 4-20mA% position.
3. The meter will display loop current as a % with 0mA=-25%, 4mA=0%, 20mA=100%, and
24mA=125%.
AUTORANGING/MANUAL RANGE SELECTION
When the meter is first turned on, it automatically goes into AutoRanging. This automatically selects the best range for the measurements being made and is generally the best mode for most measurements. For measurement situations requiring that a range be manually selected, perform the following:
3. To exit the Manual Ranging mode and return to Autoranging, press EXIT
Note: Manual ranging does not apply for the Temperature functions.
MAX/MIN appear. The meterleft auxiliary displaywill display and hold the maximum reading and will update only when a new “max” occurs.The display icon "MIN" will appear. The right auxiliary displaymeter will display and hold the minimum reading and will update only when a new “min” occurs.
2. To exit MAX/MIN mode press EXIT
RELATIVE MODE
The relative measurement feature allows you to make measurements relative to a stored reference value. A reference voltage, current, etc. can be stored and measurements made in comparison to that value. The displayed value is the difference between the reference value and the measured value.
1. Perform the measurement as described in the operating instructions. the display.
3. The Right auxiliary display displays the initial reading (the stored value)
4. The Left auxiliary display displays the actual currently measured value.
5. Main display displays the Relative value (currently measured value minus the stored value).
10 EX542-en-GB_v1.5 06/15
DISPLAY BACKLIGHT
Press the key to turn the backlight on. The backlight will automatically turn off after SET time.
Press the EXIT button to exit the backlight on mode.
HOLD
The hold function freezes the reading in the display. Press the HOLD key momentarily to activate or to exit the HOLD function.
PEAK HOLD
The Peak Hold function captures the peak AC or DC voltage or current. The meter can capture negative or positive peaks as fast as 1 millisecond in duration.
Set the rotary dial to your preferred setting. Press and Hold the HOLD button for 3 seconds and release. Now momentarily press the HOLD button, “PEAK” will appear along with “MAX” that will display in left auxiliary display and MIN” will display in right auxiliary display. Momentarily press the
HOLD button again and “PEAK Hold” will appear. The meter will update the display each time a higher or lower peak occurs. Press the EXIT button to exit the PEAK mode. Auto Power Off feature will be disabled automatically in this mode.
DATA STORAGE
1. Set the function switch to the measurement function desired.
2. Press the STORE button to enter the STORE function. The left upper aux display indicates the current storage location (0000 to 9999). New measurements will begin storing into the next location.
3. Press the PEAKHOLD button to change into the initial storage number 0000. (Press it again to change back). On the right upper aux display shows XXXX which states how many current storage locations are used.
4. Press STORE button again to enter into recording interval time set up function.
5. The left upper aux display indicates 0000 S, which is the recording sample rate; use the + & - buttons to select the sample rate desired (0 to 255 sec)
6. Set the sample rate to 0000 S for manual recording. In this mode, each press of the STORE button will save one measurement reading.
7. Set the sample rate to 1 to 255 S for automatic recording. In this mode, pressing the STORE button will start data recording at the set sample rate.
8. Press the EXIT button to end the recording session.
DATA STORAGE RECALL
1. Press STORE button for two seconds to enter into RECALL function.
2. On the left upper aux display will show XXXX, which is the current storage location. The right aux display will show XXXX, which is the number of storage locations currently used.
3. Press the + or — button to select the storage location. The value in that location will be indicated in the main display.
4. Press the PEAKHOLD button once to continuously scan data from 0000 to XXXX. Press again then scan again.
5. Press the EXIT button to end the recall session.
CLEARING MEMORY
To clear the memory of all stored data, from the OFF position, hold the EXIT button and switch the function switch to any position. Release the EXIT button and the LCD will flash 3 times and a buzzer will beep three times. Memory is now clear.
11 EX542-en-GB_v1.5 06/15
PC WIRELESS COMMUNICATION:
1. Install and launch the pc software
2. Press and Hold the backlight button for two seconds to enter RF wireless transmit mode.
3. The RF icon will appear on the display.
4. When communication is established, the RF icon on the display will blink and the led indicator on the receiver will also blink.
5. Once per second, the data will be displayed on the pc screen, plotted and inserted into a list.
6. Hold the backlight button for two seconds to exit the RF wireless transmit mode
SENDING STORED DATA TO THE PC
1. Launch the pc software
2. Press the STORE button for two seconds to enter into data RECALL function.
3. Press the HOLD button for two seconds. The RF transmit icon will flash while the stored data sent sent to the pc
SETUP
1. Press and Hold the RANGE/SETUP button for two seconds to enter the SET function. The first of five settable functions will appear.
2. Press the RANGE button to step through the functions
A: Upper limit buzzer alarm OFF or Value
B: Lower limit buzzer alarm
C: Auto power off time
OFF or Value
OFF, 10 to 30 sec
D: Button beeper
E: Back light time
ON/OFF
OFF, 10 to 30 sec
Use the +, - ◄, and ◄ buttons to select and change conditions and digits.
3. Press the RANGE/SETUP button until the meter returns to the normal display to exit this mode.
AC+DC
In the measuring modes VAC, mV(AC), 10A(AC), mA(AC) and uA(AC), press the EXIT button for 2 seconds to enter into AC+DC testing. The precision is the same as in the AC measure modes. The
LCD shows AC+DC icon. Press the EXIT button to exit the mode.
LOW BATTERY INDICATION
icon appears in the display, the battery should be replaced When the
12 EX542-en-GB_v1.5 06/15
Maintenance
WARNING: To avoid electric shock, disconnect the test leads from any source of voltage before removing the back cover or the battery or fuse covers.
WARNING: To avoid electric shock, do not operate your meter until the battery and fuse covers are in place and fastened securely.
This MultiMeter is designed to provide years of dependable service, if the following care instructions are performed:
1. KEEP THE METER DRY. If it gets wet, wipe it off.
2. USE AND STORE THE METER IN NORMAL TEMPERATURES. Temperature extremes can shorten the life of the electronic parts and distort or melt plastic parts.
3. HANDLE THE METER GENTLY AND CAREFULLY. Dropping it can damage the electronic parts or the case.
4. KEEP THE METER CLEAN. Wipe the case occasionally with a damp cloth. DO NOT use chemicals, cleaning solvents, or detergents.
5. USE ONLY FRESH BATTERIES OF THE RECOMMENDED SIZE AND TYPE. Remove old or weak batteries so they do not leak and damage the unit.
6. IF THE METER IS TO BE STORED FOR A LONG PERIOD OF TIME, the batteries should be removed to prevent damage to the unit.
BATTERY INSTALLATION
WARNING: To avoid electric shock, disconnect the test leads from any source of voltage before removing the battery cover.
1. Turn power off and disconnect the test leads from the meter.
2. Open the rear battery cover by removing two screws (B) using a Phillips head screwdriver.
3. Insert the battery into battery holder, observing the correct polarity.
4. Put the battery cover back in place. Secure with the screws.
You, as the end user, are legally bound (EU Battery ordinance) to return all used batteries, disposal in the household garbage is prohibited! You can hand over your used batteries / accumulators at collection points in your community or wherever batteries
/ accumulators are sold!
Disposal: Follow the valid legal stipulations in respect of the disposal of the device at the end of its lifecycle
13 EX542-en-GB_v1.5 06/15
WARNING: To avoid electric shock, do not operate the meter until the battery cover is in place and fastened securely.
NOTE: If your meter does not work properly, check the fuses and batteries to make sure that they are still good and that they are properly inserted.
REPLACING THE FUSES
A
B
A
F1 F2
B A
WARNING: To avoid electric shock, disconnect the test leads from any source of voltage before removing the meter cover.
1. Disconnect the test leads from the meter.
2. Remove the protective rubber holster.
3. Remove the battery cover (two “B” screws) and the battery.
4. Remove the six “A” screws securing the rear cover.
5. Gently remove the old fuse and install the new fuse into the holder.
6. Always use a fuse of the proper size and value (0.5A/1000V fast blow for the 400mA range [SIBA
70-172-40], 10A/1000V fast blow for the 20A range [SIBA 50-199-06]).
7. Replace and secure the rear cover, battery and battery cover.
WARNING: To avoid electric shock, do not operate your meter until the fuse cover is in place and fastened securely.
14 EX542-en-GB_v1.5 06/15
Specifications
Function Range
DC Voltage
Resolution
400mV 0.01mV
40V 0.001V
AC Voltage
(AC+DC)
50 to 1000Hz
Accuracy
400V 0.01V
1000V 0.1V (0.1% reading + 2 digits)
400mV 0.01mV
4V 0.0001V
(1.0% reading + 4digits)
40V 0.001V
400V 0.01V
1000V 0.1V
(1.0% reading + 3digits)
DC Current
All AC voltage ranges are specified from 5% of range to 100% of range
400 A 0.01A
4000 A 0.1A
40mA 0.001mA (1.0% reading + 3 digits)
400mA 0.01mA
10A 0.001A
AC Current
(AC+DC)
50 to 1000Hz
(20A: 30 sec max with reduced accuracy)
400 A 0.01A
4000 A 0.1A
40mA 0.001mA
(1.5% reading + 3 digits)
400mA 0.01mA
10A 0.001A
(20A: 30 sec max with reduced accuracy)
All AC voltage ranges are specified from 5% of range to 100% of range
NOTE: Accuracy is stated at 18 o C to 28 o C (65 o F to 83 o F) and less than 75% RH.
15 EX542-en-GB_v1.5 06/15
Function Range Resolution
Resistance 400 0.01
4k 0.0001k
40k 0.001k
400k 0.01k
4M 0.001M
40M 0.001M
Capacitance 40nF 0.001nF
400nF 0.01nF
4 F 0.0001F
40 F 0.001F
400 F 0.01F
4000µF 0.1µF
40mF 0.001mF
Frequency
(electronic)
40Hz 0.001Hz
400Hz 0.01Hz
4kHz 0.0001kHz
40kHz 0.001kHz
400kHz 0.01kHz
4MHz 0.0001MHz
40MHz 0.001MHz
Accuracy
(0.3% reading + 9 digits)
(0.3% reading + 4 digits)
(2.0% reading + 10 digits)
(3.5% reading + 40 digits)
(3.5% reading + 10 digits)
(5% reading + 10 digits)
(0.1% reading + 1 digits)
Frequency
(electrical)
Duty Cycle
Temp
(type-K)
100MHz 0.01MHz
Sensitivity: 0.8V rms min. @ 20% to 80% duty cycle and <100kHz; 5Vrms min @
20% to 80% duty cycle and > 100kHz.
(0.5% reading)
0.001KHz
Sensitivity:5Vrms
0.1 to 99.90% 0.01% (1.2% reading + 2 digits)
Pulse width: 100µs - 100ms, Frequency: 5Hz to 150kHz
-58 to 1382
-50 to 750
F 0.1F
C 0.1C
(1.0% reading + 4.5°F)
(1.0% reading + 2.5°C)
(probe accuracy not included)
4-20mA% -25 to 125% 0.01%
0mA=-25%, 4mA=0%, 20mA=100%, 24mA=125%
Note: Accuracy specifications consist of two elements:
(% reading) – This is the accuracy of the measurement circuit.
(+ digits) – This is the accuracy of the analog to digital converter.
±50 digits
16 EX542-en-GB_v1.5 06/15
Enclosure
Shock (Drop Test)
Double molded, waterproof (IP67)
2 meters (6.5 feet)
Diode Test
Storage capacity
Test current of 0.9mA maximum, open circuit voltage 2.8V DC typical
9999 records
RF transmit distance 10 meters (approx)
Transmitter Frequency 433MHz
Continuity Check
Peak
Audible signal will sound if the resistance is less than 35 (approx.), test current <0.35mA
Captures peaks >1ms
Temperature Sensor Requires type K thermocouple
Input Impedance >10MΩ VDC & >9MΩ VAC
AC Response
ACV Bandwidth
True rms
50Hz to 1000Hz
Crest Factor
Display
≤3 at full scale up to 500V, decreasing linearly to ≤1.5 at 1000V
40,000 count, backlit, liquid crystal with bargraph
Overrange indication “OL” is displayed
Auto Power Off
Polarity
Measurement Rate
15 minutes (approximately) with disable feature
Automatic (no indication for positive); Minus (-) sign for negative
2 times per second, nominal
Low Battery Indication “
Battery
” is displayed if battery voltage drops below operating voltage
One 9 volt (NEDA 1604) battery
Fuses mA, µA ranges; 0.5A/1000V ceramic fast blow
A range; 10A/1000V ceramic fast blow
Operating Temperature 5ºC to 40ºC (41ºF to 104ºF)
Storage Temperature -20 o C to 60 o C (-4 o F to 140 o F)
Operating Humidity
Storage Humidity
Operating Altitude
Weight
Size
Safety
Max 80% up to 31ºC (87ºF) decreasing linearly to 50% at 40ºC (104ºF)
<80%
2000meters (7000ft.) maximum.
342g (0.753lb) (includes holster).
187 x 81 x 50mm (7.36” x 3.2” x 2.0”) (includes holster)
This meter is intended for origin of installation use and protected, against the users, by double insulation per EN61010-1 and IEC61010-1 2 nd Edition
(2001) to Category IV 600V and Category III 1000V; Pollution Degree 2.
The meter also meets UL 61010-1, 2 nd Edition (2004), and CAN/CSA C22.2
No. 61010-1 2 nd Edition (2004)