NuAire Series 9 or Higher NU-8500, NU-8500E, NU-8700, NU-8700E, Series 5 or Higher NU-8500D, Series 4 or Higher NU-8700D CO2 Incubator Operation & Maintenance Manual

Series 9 or Higher NU-8500, Series 5 or Higher NU-8500D, Series 9 or Higher NU-8500E, Series 9 or Higher NU-8700, Series 4 or Higher NU-8700D, Series 9 or Higher NU-8700E are CO2 incubators designed to create a stable in-vitro environment for optimal tissue cell culture growth. These incubators are water-jacketed to ensure temperature uniformity and feature state-of-the-art controls for precise temperature, CO2 and humidity levels. The incubators also have an optional relative humidity display and control.

PDF
Document
NuAire CO2 Incubator Operation & Maintenance Manual | Manualzz

Advertisement

Advertisement

/

Advertisement

Key features

  • Large Capacity Water-Jacket

  • CO2 Display and Control

  • Humidity Display and Control

  • All Stainless Steel Construction

  • Temperature Control System

  • Remote Alarm Output Contacts

Frequently asked questions

The chamber temperature range is 5°C above ambient to +55°C.

Before connecting the CO2 supply, ensure the water-jacket is filled and the temperature set. The CO2 pressure should not exceed 25 PSIG or 1.8 BAR. Use a two-stage CO2 pressure regulator and medical-grade CO2.

The drain plug is located on the bottom left side. The drain uses a ball valve. To drain the jacket, turn the valve stem vertically to open the drain. Remove the white safety plug before draining.

User questions

Z U

Mary Z.

How is a full factory reset performed and what are the consequences?

A full factory reset is performed using the "Reset, Master" diagnosis function, detailed in Section 8.8.5. Executing this function clears the microprocessor’s memory, resets all parameters to their defaults, and results in the loss of all configuration options and calibration offsets.

H S

John H.

What does it mean if the temperature and CO2 level readings are blinking on a Nuaire CO2 Incubator NU5100E after it is turned back on?

Blinking displays after turning the unit back on indicates an interruption of power. To stop the blinking displays, any key should be pressed. If the blinking and alarm occur 30 minutes after the inner glass door is left open, it could also indicate that the inner glass door is not closed or the magnetic switch needs a position adjustment. Close and latch the inner glass door and Adjust the switch position to align it with the disk magnet on the glass door hinge by loosening the acorn nut on the cable clip. Check the door switch; if faulty, replace it.

F A

John F.

The CO2 Inject Status LED was blinking green during the diagnostic test, but is not illuminated after exiting the test. What could be the issue?

This could indicate several possibilities:

  • CO2 control system is off: Ensure that the CO2 control system is enabled in the option configuration parameters. If the CO2 system is disabled, the LED will not illuminate during normal operation.
  • CO2 setpoint: Verify that the CO2 setpoint is not set to 0.0%. If it is, the CO2 control system is turned off, and the LED will not illuminate.
  • CO2 level reached: If the CO2 level in the chamber has already reached the setpoint, the LED will not be illuminated because the system is not actively injecting CO2.
  • Faulty LED: (Less likely) Although the LED blinked during the diagnostic test, there could still be an issue with it not functioning correctly during normal operation.
F A

John F.

How to enable the CO2 control system?

To enable the CO2 control system, access the Option Configuration Parameters:

  1. Press and hold the Hidden key (NuAire logo) for four seconds to enter the diagnostic mode.
  2. Press the Up arrow key to advance to the 'opt' menu (Option Configuration Parameters).
  3. Press the Select key repeatedly until you reach the 'CO2 System Enable' option.
  4. Press the Up arrow key to toggle the setting to 'on'.
  5. Press the Hidden key to exit the diagnostic mode.

After performing these steps, the CO2 control system should be enabled.

H S

John H.

How to check if the CO2 system is enabled in the option configuration parameters on a Nuaire CO2 Incubator NU5100E?

To check if the CO2 system is enabled in the option configuration parameters: a) Press and hold the Hidden key (NuAire logo) for four seconds (in either Run or Setup mode); the temperature display will indicate the first menu “tst”. b) Press the Up arrow key until the temperature display indicates “opt”. c) Press the Select key; this will enter the option configuration menu. d) Press the Select key repeatedly to advance through the function parameters until you reach 'CO2 System Enable'. e) The display will show 'on' or 'off', corresponding to the current condition. If it displays 'off', press the Up arrow key to toggle it to 'on'. f) To exit the diagnostic mode, press the Hidden key several times.

F A

John F.

How to check if the CO2 setpoint is not set to 0.0%?

To check that the CO2 setpoint is not set to 0.0%, follow these steps:

  1. Press and hold the Mode key for three seconds to enter the Setup mode. The green LED above the Mode key should be blinking, and the temperature and CO2 displays should indicate 'SETUP'.
  2. Press the SEL key until the green LED next to the CO2 percent display is lit. The current CO2 setpoint value will be shown in the display.
  3. Verify that the displayed value is not 0.0%. If it is, use the Up arrow key to increase it to the desired setpoint.
  4. Press the Mode key again for three seconds to return to Run mode. The green LED above the Mode key should be on solid.
Z U

Mary Z.

What does the "CO2 percent error 0.5P" alert signify?

This alert signifies an "Auto zero failure," meaning the zero value generated by the auto-zero routine is greater than 0.5%. This is only an alert and does not impact the CO2 system operation.

H S

John H.

The CO2 panel on a Nuaire CO2 Incubator NU5100E shows 5%, but the 'Inject CO2 Status LED' is not green. What could be the reason?

If the CO2 level is at the setpoint (5%) but the 'Inject CO2 Status LED' is not green, it could indicate that the CO2 control system is turned off or there is a potential issue with the LED indicator itself. The CO2 system can be checked and enabled in the option configuration parameters.

O R

Robert O.

How to perform the LED test?

To perform the LED test, enter diagnostic mode by pressing and holding the NuAire logo for four seconds. Navigate to the 'tst' menu using the Up arrow key and press SEL. Repeatedly press SEL until you reach the 'All Lights' or 'Display/LED Test' function parameter. Observe if the CO2 LED lights up during the test. To exit, press the Hidden key (NuAire logo) several times.

F A

John F.

What does '2.tn (off)' mean on the configuration control panel?

The 'CO2 Tank 2 Enable (option)' parameter, when set to 'off', disables the optional CO2 tank 2 system. When this option is disabled, the unit will not use CO2 from tank 2.