GE Healthcare CU-950, CU-960 Control Unit Instructions
The Control Units CU-950 and CU-960 are used as the controlling interface between a UNICORN™ workstation and the components of an ÄKTA™ system. CU-950 is connected to a UNICORN workstation using either a USB port or an Ethernet interface, while CU-960 uses Ethernet interface. CU-960 is the standard control unit delivered with ÄKTAprocess™ and ÄKTAready™ systems. CU-960 is equipped with a 128 MB flash memory card that can buffer run data in the event of a computer failure, communication loss, or power failure. This allows runs to continue and data to be stored until communication is re-established.
Advertisement
Advertisement
CU950-960.book Page 1 Monday, January 12, 2009 2:42 PM
GE Healthcare
Control Units
CU-950 and CU-960
Instructions
CU950-960.book Page 2 Monday, January 12, 2009 2:42 PM
CU950-960.book Page 3 Monday, January 12, 2009 2:42 PM
Important user information
All users must read this instruction to fully understand the safe use of the CU-950 and CU-960 control units.
Safety notices
This manual contains warnings and cautions concerning the safe use of the product. See definitions below.
WARNING! The WARNING symbol and notice highlight instructions that must be followed to avoid personal injury. Do not proceed until all stated conditions are clearly understood and met.
Declaration of conformity
This product meets the requirements of applicable CEdirectives. A copy of the corresponding Declaration of
Conformity is available on request.
The CE symbol and corresponding Declaration of
Conformity is valid for the instrument when it is:
• used as a stand-alone unit, or
• connected to other CE-marked GE Healthcare instruments, or
• connected to other products recommended or described in this instruction, and
• used in the same state as it was delivered from GE
Healthcare, except for alterations described in this instruction.
Note:
The Declaration of conformity is valid only for instruments that are marked with the CE logo.
CAUTION! The CAUTION notice highlights instructions that must be followed to avoid damage to the product or other equipment. Do not proceed until all stated conditions are met and clearly understood.
Note:
A Note is used to indicate information that is important for trouble-free and optimal use of the product.
Recycling
This symbol indicates that the waste of electrical and electronic equipment must not be disposed as unsorted municipal waste and must be collected separately. Please contact an authorized representative of the manufacturer for information concerning the decommissioning of equipment.
WARNING! All repairs should be done by personnel authorized by GE Healthcare. Do not open any covers or replace parts unless specifically stated in the instructions.
CU950-960.book Page 4 Monday, January 12, 2009 2:42 PM
CU950-960.book Page 5 Monday, January 12, 2009 2:42 PM
Contents
1 Introduction ..........................................................................6
2 Description............................................................................7
2.1 Front panel LEDs .........................................................................8
3 CU-950 only: Installation on ÄKTAexplorer and
ÄKTApurifier..........................................................................9
4 CU-950 only: Installation on ÄKTAFPLC ........................ 11
5 CU-950 only: Installation on ÄKTApilot ......................... 13
6 CU-960 only: Installation on ÄKTAprocess ................... 14
7 Flash memory card........................................................... 15
8 Error messages from CU-950/960 displayed in
UNICORN ............................................................................ 16
9 CU-960 only: Installation on ÄKTAready....................... 20
Instructions - 28-9414-37 Edition AA 5
CU950-960.book Page 6 Monday, January 12, 2009 2:42 PM
Control Units CU-950 and CU-960
Control Units CU-950 and CU-960
1 Introduction
The Control Units CU-950 and CU-960 are used as the controlling interface between a UNICORN™ workstation and the components of an ÄKTA™ system.
CU-950
This instruction describes CU-950 and provides information on how to install this control unit with ÄKTAexplorer™, ÄKTApurifier™, ÄKTAFPLC™, and ÄKTApilot™.
CU-960
CU-960 is the standard control unit delivered with ÄKTAprocess™ and
ÄKTAready™ systems.
6 Instructions - 28-9414-37 Edition AA
CU950-960.book Page 7 Monday, January 12, 2009 2:42 PM
Control Units CU-950 and CU-960
2 Description
CU-950
CU-950 unit consists of a back panel with connectors and three indicator lamps
(LEDs) on the front. The unit is supplied with a small power converter and EU and
US mains cables. It is connected to a UNICORN workstation using the:
• USB port. The CU-950 USB is designed as a convenient solution for users running a single systems from one laptop computer. The CU-950 USB is connected via the USB interface and each UNICORN workstation can use one CU-950 USB to control one system.
• Ethernet interface, so called Advance mode. The CU-950 Advanced mode is designed for users that expect robustness in, for example, manufacturing.
The CU-950 Advanced mode is connected to the UNICORN workstation via an Ethernet network interface and operates from a designated IP address.
Up to four CU-950 can be connected enabling simultaneous control of four systems. The CU-950 Advanced mode is offered with an option to use 128
MB internal memory for improved preparedness in case of communication failures between the UNICORN workstation computer and the CU-950.
The control unit is designated its own IP-address and can be connected directly to a network switch or hub instead of a local UNICORN workstation.
USB UniNet-1 Test
Figure 2-1. CU-950 unit, front and back panel.
CU-960
The CU-960 controller card is self-contained with a separate power supply, independent of the controlling workstation. It is connected to a UNICORN workstation using the Ethernet interface. The control unit is designated its own IPaddress and can be connected directly to a network switch or hub instead of a local UNICORN workstation. The CU-960 is equivalent to the CU-950 Advanced mode with the memory option. The CU-960 has communication support for
Profibus components and is used in the ÄKTAprocess systems and other process systems.
Instructions - 28-9414-37 Edition AA
CU-960
POWER
PC
SYSTEM
Figure 2-2. CU-960 unit, front and back panel.
Note: The USB interface is only used by GE Healthcare service personnel for service purposes.
7
CU950-960.book Page 8 Monday, January 12, 2009 2:42 PM
Control Units CU-950 and CU-960
2.1
Front panel LEDs
The functions of the three green LEDs on the front panel labelled Power, PC and
System are described below:
LED
Power
FLASHING
Internal test running or not OK
PC No PC communication
System No system communication
ON
Power on, internal test OK
Contact with PC OK
System communication OK
When the system LED flashes the System Control Run Data Instruments shows
Scanning, the cause can be seen from the table below:
No communication to:
ÄKTA system
Profibus system
ÄKTA and Profibus systems
Approximate flash time
0.5 s
2 s
1 s
When the system LED is On, the System Control Run Data Instruments shows
Ready, if there where any instrument modules with error or that an instrument module is missing. The error will be reported in system control before the
"Instruments" Run Data changes from Scanning to Ready.
2.2
Back panel
The back panel ports are (from left to right):
• Power supply input
• CU-960 only: Profibus port – for system communication
• USB port – for network/controlling workstation communication (UNICORN)
– CU-960: for service use only
• Ethernet port – for network/controlling workstation communication
(UNICORN)
• UniNet-1 port – for system communication
• Test port – for service use only
CAUTION! Ensure that the CU-960 is connected to the proper power supply,
12-24 VDC 1A.
8 Instructions - 28-9414-37 Edition AA
CU950-960.book Page 9 Monday, January 12, 2009 2:42 PM
Control Units CU-950 and CU-960
3
CU-950 only: Installation on ÄKTAexplorer and ÄKTApurifier
1 Hang the CU-950 on the left side of the system by inserting the hooks on the front of CU-950 into the channel on the side of the UV-900 and rotating into position, as shown in the illustration below. ÄKTApurifier system is shown as an example.
Ä
u fier
Figure 3-1. Location of CU-950 on ÄKTApurifier system.
Figures 3-2 and 3-3 show the schematic CU-950 cable connections for
ÄKTAexplorer, and ÄKTApurifier respectively. Remember that the CU-950 unit is actually mounted on the side of UV-900.
Mains supply socket
Fraction collector
(optional)
Instructions - 28-9414-37 Edition AA
CU-950
UniNet-1 cable
UniNet-1 cable
Mains cable Power converter
To computer
USB cable
Figure 3-2. Schematic UniNet-1 and mains cable connections for ÄKTAexplorer.
9
CU950-960.book Page 10 Monday, January 12, 2009 2:42 PM
Control Units CU-950 and CU-960
Fraction collector
(optional)
Mains supply socket
CU-950
UniNet-1 cable
Mains cable
UniNet-1 cable
To computer
USB cable
Power converter
Figure 3-3. Schematic UniNet-1 and mains cable connections for ÄKTApurifier.
How to connect the cables:
1 Connect the UniNet-1 data communication cable between the UniNet-1 port on CU-950 and Pump P-900 in the system.
WARNING! Only use mains cables delivered and approved by GE Healthcare.
Note: If a fraction collector is used, connect the fraction collector UniNet-1 cable between Pump P-900 and the fraction collector, and the 1.5 m
UniNet-1 cable between the fraction collector and CU-950, and then the USB cable between CU-950 and the computer.
2 Connect the power converter cable to the Input port on CU-950.
3 Connect the mains cable between the power converter and the mains supply socket at the rear of the system.
4 If an auto-sampler is used, the mains cable to the autosampler may be connected to an external mains outlet if all mains supply sockets on the system are occupied.
Note: In this case the autosampler must be manually reset by disconnecting the power cable when performing a system reset or restart.
5 Connect the USB cable to the USB port on CU-950. For connection to and installation on the computer, please refer to the UNICORN Administration and technical manual.
10 Instructions - 28-9414-37 Edition AA
CU950-960.book Page 11 Monday, January 12, 2009 2:42 PM
Control Units CU-950 and CU-960
4
CU-950 only: Installation on ÄKTAFPLC
1 Hang the CU-950 on the left side of the system by inserting the hooks on the front of CU-950 into the channel on the side of the UPC-900 and rotating into position, as shown below.
Instructions - 28-9414-37 Edition AA
Figure 4-1. Location of CU-950 on ÄKTAFPLC.
Fig. 4-2 shows the schematic CU-950 cable connections for ÄKTAFPLC.
Remember that the CU-950 unit is actually mounted on the side of UPC-900.
Fraction collector
(optional)
UniNet-1 cable
CU-950
UniNet-1 cable
Conductivity
Flow Cell pH-GroundpH-Probe
Optical
Unit
Analogue out 0-1 V UniNet 1
Lamp
Mains
Voltage Frequency Power, max
100-200 V- 50-60 Hz 30 VA
Remote
Pressure
Analogue out 0-1 V
UniNet 2
UniNet 1
Mains output
Leakage current, max
3,5 mA
Mains
Voltage
100-200 V-
Frequency Power, max Fuse
50-60 Hz 600 VA T 6,3 AL
WARNING! For continued protection against risk of fire, replace only woth fuse of the specified type and current ratings
Mains supply socket
To computer
USB cable
Mains cable
Power converter
Figure 4-2. Schematic UniNet-1 and mains cable connections for ÄKTAFPLC.
11
CU950-960.book Page 12 Monday, January 12, 2009 2:42 PM
Control Units CU-950 and CU-960
How to connect the cables:
1 Connect the UniNet-1 data communication cable between the UniNet-1 port on CU-950 and Pump P-920 in the system.
Note: If a fraction collector is used, connect the fraction collector UniNet-1 cable between P-920 and the fraction collector, and the 1.5 m UniNet-
1 cable between the fraction collector and CU-950, and then the USB cable between CU-950 and the computer.
2 Connect the cable on the CU-950 power converter to the Input port on
CU-950.
3 Connect the mains cable between the power converter and the mains supply socket at the rear of the system.
4 If an autosampler is used, the mains cable to the autosampler may be connected to an external mains outlet if all mains supply sockets on the system are occupied.
Note: In this case the autosampler must be manually reset by disconnecting the power cable when performing a system reset or restart.
5 Connect the USB cable to the USB port on CU-950. For connection to and installation on the computer, please refer to the UNICORN Administration and technical manual.
12 Instructions - 28-9414-37 Edition AA
CU950-960.book Page 13 Monday, January 12, 2009 2:42 PM
Control Units CU-950 and CU-960
5
CU-950 only: Installation on ÄKTApilot
ÄKTApilot uses a CU-950 Advanced, with a 128 MB flash memory card.
The CU-950 unit and the power converter is placed near or inside ÄKTApilot.
Fraction collector
(optional)
Analog IN Analog out UniNet 1 Digital I/O
Internal I/O Press2 Press1 Valves Air 2 Air 1 Fans
Mains
UniNet-1 cable
Mains cable
Power converter
CU-950
UniNet-1 cable
To computer
Ethernet cable
Figure 5-1. Schematic UniNet-1 and mains cable connections for ÄKTApilot.
How to connect the cables:
1 Connect the UniNet-1 data communication cable between the UniNet-1 port on CU-950 and one of the UniNet-1 sockets on the controller rear panel inside ÄKTApilot.
Note: If a fraction collector is used, connect the fraction collector UniNet-1 cable between the controller rear panel inside ÄKTApilot and the fraction collector, and the 1.5 m UniNet-1 cable between the fraction collector and CU-950.
2 Connect the cable on the CU-950 power converter to the Input port on
CU-950.
3 Connect the mains cable between the power converter and the mains supply socket inside the system.
4 Connect the Ethernet cable to the Ethernet port on CU-950. For connection to and installation on the computer, please refer to the UNICORN
Administration and technical manual.
Instructions - 28-9414-37 Edition AA 13
CU950-960.book Page 14 Monday, January 12, 2009 2:42 PM
Control Units CU-950 and CU-960
6
CU-960 only: Installation on ÄKTAprocess
The CU-960 is delivered pre-installed in all ÄKTAprocess systems. All necessary connections are made and tested already when the system is assembled before delivery. The unit is mounted in a rack adaptor which is mounted in the instrument rack inside the electronics cabinet of the ÄKTAprocess system, as illustrated below.
Figure 6-1. CU-960 mounted in an ÄKTAprocess system.
Note: When the CU-960 is mounted in the system cabinet the indicator LEDs may not be visible since the front of the unit is facing inwards.
WARNING! Only personnel authorized by GE Healthcare are allowed to open
the electronics cabinet of the ÄKTAprocess systems.
14 Instructions - 28-9414-37 Edition AA
CU950-960.book Page 15 Monday, January 12, 2009 2:42 PM
Control Units CU-950 and CU-960
7 Flash memory card
The CU-960* is delivered with a 128 MB flash memory card which, depending on the UNICORN system settings, can buffer run data when the control unit is in operation, in the event of a computer failure, broken communication between the unit and workstation or a power failure. The run may proceed and data may be stored in the CU-960* flash memory until communication between the workstation and the CU-960* is re-established or the workstation is functional again. The run will continue only until the flash memory is full.
Run data is saved at intervals by UNICORN. If a power failure occurs immediately before UNICORN has saved the run data several minutes of data may be lost.
However, if selected in UNICORN, run data is saved continuously on the flash memory card and the data will be saved up to the time when the CU-960* lost its power. When power is restored, the run data may be retrieved.
* The CU-950 can also be equipped with a flash memory card and be configured for
Advanced mode.
Instructions - 28-9414-37 Edition AA 15
CU950-960.book Page 16 Monday, January 12, 2009 2:42 PM
Control Units CU-950 and CU-960
8 Error messages from CU-950/960 displayed in UNICORN
1008
1009
1010
1011
1013
1014
1015
Error Id no. Error message
Error codes for Controller card
1001 List does not start with $
Cause
Internal error.
1002
1003
1004
Download not for this system
Unknown list prefix
M/S table index out of range
Internal error.
Internal error.
Internal error.
1005 BASE Internal error.
valid character
1017
1018
1019
1021
1023 type
No decimal point in real
Internal error.
Value table index out of range
Internal error.
No 'V' when expected Internal error.
No type on value Internal error.
Missing parameters Internal error.
Error in local value Internal error.
Unknown opcode Internal error.
real error.
System not in IDLE Internal error.
See Help
No < 1 or >
UCO_MAX_COMMAN
DS
Internal error.
Internal error.
Internal error.
param>UCO_MAX_PA
RAMETERS
Error in terminal field Internal error.
0 or 1 expected, not received
Internal error.
Suggested Action
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
16 Instructions - 28-9414-37 Edition AA
CU950-960.book Page 17 Monday, January 12, 2009 2:42 PM
Instructions - 28-9414-37 Edition AA
Control Units CU-950 and CU-960
Error Id no. Error message
# Method Executor error codes
1030 Missing at least one list
1031 System is already running
Cause
Internal error.
Internal error.
Suggested Action
Create a report.
Create a report.
1033
1034
Unknown opcode
Erroneous value for value table
Internal error.
Internal error.
1036
1037
1038
1039 list
Call and/or loop nest too deep
Internal error.
Missing value table index
Unknown list type
Internal error.
Internal error.
Missing method event list
Internal error.
1040 Erroneous id
Internal error.
1043 list
Internal error.
parameters in instrument
M/S table index out of range
Internal error.
1044 Internal
1045 Value table index out of range
Internal error.
1046 Method stopped error, Internal error.
1047 Wrong value type in write_vt
Internal error.
1061
1070 manual command
Wrong number of pars in command
> 16 data sel blocks active
Internal error.
Internal error.
error
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
17
CU950-960.book Page 18 Monday, January 12, 2009 2:42 PM
Control Units CU-950 and CU-960
Error Id no. Error message
1100
1101
Error in strategy execution
Undef. strat block,
Cause
Internal error.
Suggested Action
Create a report.
The strategy contains an unknown block
Internal error.
Upgrade UNICORN software or reinstall old strategy
1700 >95% CPU Load
# SAP error codes
1303 Tuning forbidden in this state
1305 Wrong test type in strat. block
Internal error.
Internal error.
Create a report.
Create a report.
Create a report.
Create a report.
1309
1310
1311 type in strat.
Parvalue out of range in strat
Internal error.
Internal error.
Missing par in strategy block
Wrong index in strat.
RUNMODE
Internal error.
Create a report.
Create a report.
Create a report.
Create a report.
1315 unknown block
Trm index out of range, RUNMODE
Internal error.
Create a report.
2001 restarted
WARNING - faulty
UniNet comm. controller is restarted - messages are lost.
Check instrument connection.
Faulty UniNet communication- no message lost.
Check instrument connection.
2002 UniNet overrun
Messages are lost. Check instrument connection.
2003 UniNet pckg rec in wrong order
UniNet package received in wrong order.
Make sure that all connected valves have a unique id.
Number.
2010 Too many pars Internal error.
Create a report.
2011
2012
Undef service
No response
Internal error.
System could not find specific unit.
Create a report.
Check instrument connection.
If Error message appears during system start up and specific unit is not included in the system configuration this message can be ignored.
18 Instructions - 28-9414-37 Edition AA
CU950-960.book Page 19 Monday, January 12, 2009 2:42 PM
Control Units CU-950 and CU-960
19
38
51
Error Id no. Error message
2014
2022
2023
2024
52
53
54
55
3000
3001
Cause
instrum. reply Internal error.
No pars in comm Internal error.
instrument Internal error.
responding
No instrms defined in strategy
No relay unit def
Too many instrms def in strat.
Internal error.
Internal error.
Internal error.
See Help
See Help
Internal error.
Internal error.
Internal error.
Unable to allocate
CAN c block
Unable to post CAN cmd block
Internal error.
Unable to allocate
CAN m block
Unable to post CAN msg block
Internal error.
Internal error.
General CAN driver error
Internal error.
Storing index hit max CU-960 will no longer be able to recover data in case of communication loss
Fatal Strategy Error Strategy uses
Profibus card, card not in CU configuration.
Suggested Action
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Create a report.
Please create a report and contact service
Instructions - 28-9414-37 Edition AA 19
CU950-960.book Page 20 Monday, January 12, 2009 2:42 PM
Control Units CU-950 and CU-960
9
CU-960 only: Installation on ÄKTAready
Connection from the system computer to the internal control unit CU-960 is made by the water proof RJ45 connector placed in the bottom of the main cabinet
(VS-08-A-RJ45/MOD-1-IP67 + VS-08-BU-RJ45-5-F/BU). It is absolutely necessary from an EMC perspective to connect this connector shielding to the cabinet using either copper tape or the small sheet-metal shroud developed in the project.
The mating connector cable assembly needs to have the matching water proof plug type provided by Phoenix. (VS-08-RJ45-5-Q/IP67) to maintain IP67. Suitable mating cable assembly is described in assembly drawing 14001358 (part number
28-4022-13) and is supplied with the system in the accessory-kit.
Note: The internal LAN cable is a cross-over type!
If the system is connected to a switch/hub, a cross-over cable must be connected in series with a cable provided by GE Healthcare, part number 28-4022-13.
20 Instructions - 28-9414-37 Edition AA
CU950-960.book Page 21 Monday, January 12, 2009 2:42 PM
CU950-960.book Page 22 Monday, January 12, 2009 2:42 PM
CU950-960.book Page 23 Monday, January 12, 2009 2:42 PM
CU950-960.book Page 24 Monday, January 12, 2009 2:42 PM
For local office contact information, visit www.gelifesciences.com/contact
GE Healthcare Bio-Sciences AB
Björkgatan 30
751 84 Uppsala
Sweden www.gelifesciences.com/protein-purification-bioprocess www.gelifesciences.com/akta
GE, imagination at work and GE monogram are trademarks of General Electric Company.
ÄKTA, ÄKTAexplorer, ÄKTAFPLC, ÄKTApilot, ÄKTAprocess, ÄKTApurifier, ÄKTAready, Drop Design, and UNICORN are trademarks of GE Healthcare companies.
UNICORN: Any use of this software is subject to GE Healthcare Standard Software End-User
License Agreement for Life Sciences Software Products.
All third party trademarks are the property of their respective owners.
© 2009 General Electric Company—All rights reserved.
First published Jan. 2009
All goods and services are sold subject to the terms and conditions of sale of the company within
GE Healthcare which supplies them. A copy of these terms and conditions is available on request. Contact your local GE Healthcare representative for the most current information.
GE Healthcare UK Ltd
Amersham Place, Little Chalfont,
Buckinghamshire, HP7 9NA,
UK
GE Healthcare Bio-Sciences Corp
800 Centennial Avenue,
P.O. Box 1327, Piscataway,
NJ 08855-1327,
USA
GE Healthcare Europe GmbH
Munzinger Strasse 5,
D-79111 Freiburg,
Germany
GE Healthcare Bio-Sciences KK
Sanken Bldg. 3-25-1,
Hyakunincho, Shinjuku-ku,
Tokyo 169-0073,
Japan
28-9414-37 AA 01/2009
Download
Advertisement
Key features
Ethernet communication
USB communication
Flash memory card
Profibus communication
Run data buffering
Frequently asked questions
These control units act as the interface between a UNICORN workstation and the components of an ÄKTA system.
The CU-950 can be connected using either a USB port or an Ethernet interface. The USB port is designed for single system setups, while the Ethernet interface is designed for more robust applications.
The CU-960 is connected to a UNICORN workstation using the Ethernet interface.
The flash memory card provides buffering of run data in case of computer failure, communication loss, or power failure. This ensures that the run can continue and data is not lost.