Welch Allyn SureTemp Plus
Trouble Shooting Guide
SYMPTOM
Oral Temperatures
Inaccurate temperature reading
(low or high)
POSSIBLE CAUSE CORRECTION
Probe is not placed into sublingual pocket Repeat temperature and reposition the probe.
The thermometer is not in the Oral mode Check the mode on the display. Change to the Oral mode using the Mode Selection button
Incorrect probe covers used or no probe cover used
Use only Welch Allyn probe covers. Using non-Welch Allyn probe covers or no probe cover will cause incorrect reading
Patient has ingested something cold or hot Wait at least 20 minutes after eating, drinking, chewing gum or mints, smoking or brushing teeth
Patient was speaking or clenching the probe tip with teeth while taking a temperature
Probe must be completely surrounded by tissue while taking a temperature
Clinician is taking repeated temperatures on the same patient
Wait a minimum of 10 minutes before repeating a patient’s temperature. The site is cooled by repeated temperatures
Axillary Temperatures
Inaccurate temperature reading
(low or high)
Probe is not placed correctly into the axilla Reposition the probe vertically high into the axilla. Avoid skin folds, bandages etc.
The thermometer is not in the axillary mode or is in the incorrect axillary mode
Check the mode on the display. Change to the Adult (>18 yrs) or Peds (<17 yrs) mode using the Mode Selection button
Incorrect probe covers used or no probe cover used
Use only Welch Allyn probe covers.
Using non-Welch Allyn probe covers or no probe cover will cause incorrect reading
Chose opposite axilla or another site.
Patient’s axilla has been exposed to the ambient air for a prolonged time (i.e. patient not wearing gown or shirt and arms are extended over head.
Patient or probe tip was moving in the axilla during temperature taking
Probe must be completely surrounded by tissue while taking a temperature. Arm should be held at snugly at patient’s side
SYMPTOM
Rectal Temperatures
Inaccurate temperature reading
(low or high)
POSSIBLE CAUSE CORRECTION
Probe is not placed correctly into the rectal site
Insert the probe only 5/8 inch into the rectum (less for infants)
The thermometer is not in the Rectal mode Check the mode on the display. Use only probes with red ejection button and red probe well
Incorrect probe covers used or no probe cover used
Use only Welch Allyn probe covers.
Using non-Welch Allyn probe covers or no probe cover will cause incorrect reading
Thermometer
The display is blank when the probe is removed
The batteries are dead
The thermometer is broken
Check the battery icon on the display.
If no segments are visible, replace the batteries.
Call Welch Allyn’s Customer Service
Department
Call Welch Allyn’s Customer Service
Department
The displays shows a Broken
Probe icon or an Instrument
Malfunction icon when the probe is removed from the well
There are no temperature icons and /or words on the display when the probe is removed
The displays shows “SEC” when the probe is removed from the well or the device is removed form the wall holder
The display show a snail icon and the word “Monitor”
The thermometer or probe is broken
The device has been internally changed from the factory default
The device is in the monitor mode.
Contact your Biomedical Department
The security mode has been internally set Contact your Biomedical Department or
Welch Allyn’s Customer Service
Department
One of the following has occurred: The probe was left out of the thermometer for over one minute or your patient’s temperature is too cold or there was poor tissue contact while taking the temperature
© 2005 Welch Allyn MC2747-TSG