Midmark Vantage Panoramic X-ray System Installation manual
Below you will find brief information for Vantage. The Vantage is an easy-to-use and easy-to-install digital panoramic X-ray system. It provides a broad overview of the teeth, jaw, and oral structure of the entire mouth. The panoramic digital receptor is contained in a C-arm that moves around the patient’s head.
Advertisement
Advertisement
Progeny Vantage
Panoramic X-ray System
0120
Installation Guide
For Vantage units manufactured during or after October 2012
Part Number: 00-02-1608
ECN: P3217
REV. H
Contents
2
1 Regulatory
Information
In this Chapter
•
Indications for Use
•
Contraindications
•
Warnings and Precautions
•
Compliance with Applicable Standards
•
Certified Components
•
Device Labeling
•
EC Declaration of Conformity
•
EMC Statement
•
Authorized Representatives
Indications for Use
Statement of Indications for Use
The indications for use of the Progeny Vantage Panoramic X-ray System are to provide dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. When the system is equipped with the Cephalometric option, the system will also provide cephalometric radiographic examinations for the use in orthodontic treatment planning and evaluation.
Guidelines for Patient Selection
The guidelines for use of the Progeny Vantage Panoramic Extraoral X-ray
System are described in the “ADA/FDA Guide to Patient Selection for Dental
Radiographs.” This device is to be operated only for the intended use as indicated by prescription of a qualified dental practitioner.
Contraindications
None known at this time.
3
Progeny Vantage Panoramic X-ray System Installation Guide 1 Regulatory Information
Warnings and Precautions
Radiation Safety
• The operator during X-ray production must remain 2 m (6 ft.) from the focal spot and the X-ray beam for protection.
• Full use must be made of all radiation safety features on the equipment.
• Full use must be made of all radiation protection devices, accessories, and procedures available to protect the patient and operator from X-ray radiation.
Electrical Safety
• Only qualified and authorized service personnel should remove covers on the equipment.
• This equipment must only be used in rooms or areas that comply with all applicable laws and recommendations concerning electrical safety in rooms used for medical purposes, e.g., IEC, US National Electrical Code, or VDE standards concerning provisions of an additional protective earth (ground) terminal for power supply connection.
• Before cleaning or disinfecting, this equipment must always be turned off.
• The Progeny Vantage X-ray System is ordinary medical equipment without protection against ingress of liquids. To protect against short-circuit and corrosion, no water or any other liquid should be allowed to leak inside the equipment.
Explosion Safety
Only qualified and authorized personnel may operate this equipment observing all laws and regulations concerning radiation protection.
This equipment must not be used in the presence of flammable or potentially explosive gases or vapors, which could ignite, causing personal injury and/or damage to the equipment. If flammable disinfectants are used, the vapor must be allowed to disperse before using the equipment.
Damage and Injury
Do not place permanent or non-mobile structures beneath the device. Device movement may result in damage to the device or structure, or in injury to the operator or patient.
Cleanliness
To prevent cross contamination, always clean the patient contact areas and always install a fresh protective sheath over the bite guide before positioning a patient. The sheath recommended for this application is the TIDI Products, part number 21008.
4
Progeny Vantage Panoramic X-ray System Installation Guide
Laser Safety
1 Regulatory Information
Do not stare into the beam. Do not place eyes closer than 100 mm. This equipment contains class 2 lasers of 3 mW output at 650 nm. The beam is a 40° fan line. The lensing on the laser is not removable. Laser on time does not exceed 100 seconds.
CAUTION!
Use of procedures other than those contained within this manual may result in exposure to damaging laser light.
5
Progeny Vantage Panoramic X-ray System Installation Guide 1 Regulatory Information
Compliance with Applicable Standards
Radiation Protection
The certified components of the Progeny Vantage Panoramic Dental X-ray
System comply with Radiation Performance Standards 21 CFR, Subchapter J, at the time of manufacture.
Performance Standards
Standard
IEC 60825-1:2001
IEC 60601-1
IEC 60601-1-1
IEC 60601-2-7
IEC 60601-2-28
IEC 60601-1-3
IEC 60601-1-2
CAN/CSA 22.2 No.
601.1-M90
IEC 60601-2-32:
1994
Content
Safety of Laser Equipment
Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
Medical electrical equipment - Part 1-1:
General requirements for safety - Collateral standard: Safety requirements for medical electrical systems
Medical electrical equipment - Part 2-7:
Particular requirements for the safety of highvoltage generators of diagnostic X-ray generators
Medical electrical equipment - Part 2-28:
Particular requirements for the safety of X-ray source assemblies and X-ray tube assemblies for medical diagnosis
Medical electrical equipment – Part 1-3:
General requirements for radiation protection in diagnostic X-ray equipment
EMI/RFI
Canadian standard for medical electrical equipment
Medical electrical equipment - Part 2-32:
Particular requirements for the safety of associated equipment of X-ray equipment
6
Progeny Vantage Panoramic X-ray System Installation Guide
Certified Components
Component
Pan tubehead assembly
Pan collimator assembly
Pan X-ray control assembly
Pan sensor assembly
Ceph (Option) sensor assembly
Ceph (Option) secondary collimator assembly
1 Regulatory Information
Reference Number
60-A1014
60-A2009
60-A2035
60-A1010
60-A1004
60-F4051
7
Progeny Vantage Panoramic X-ray System Installation Guide
Device Labeling
1 Regulatory Information
8
Progeny Vantage Panoramic X-ray System Installation Guide 1 Regulatory Information
Optional Cephalometric Extension Labeling
9
Progeny Vantage Panoramic X-ray System Installation Guide 1 Regulatory Information
EC Declaration of Conformity
Name and Description of Product
Progeny Vantage Panoramic Dental X-ray System
Catalog: V5000 Progeny Vantage Panoramic X-ray
System, Domestic
Model: 60-A0001
Catalog: V5100 Progeny Vantage Panoramic X-ray
System, Export
Model: 60-A0001
Catalog: V5050 Progeny Vantage Panoramic with Ceph
Sensor, Domestic
Model: 60-A0008
Catalog: V5150 Progeny Vantage Panoramic with Ceph
Sensor, Export
Model : 60-A0008
Catalog: V5000C Progeny Vantage Panoramic with
Cephalometric Extension, 2 Sensor System, Domestic
Model: V5000 + C6000
Catalog: V5100C Progeny Vantage Panoramic with
Cephalometric Extension, 2 Sensor System, Export
Model: V5100 + C6000
Catalog: V5050C Progeny Vantage Panoramic with
Cephalometric Extension, 1 Sensor System, Domestic
Model: V5050 + C4000
Catalog: V5150C Progeny Vantage Panoramic with
Cephalometric Extension, 1 Sensor System, Export
Model: V5150 + C4000
Catalog: C6000 Cephalometric Extension with Sensor
Model: 60-A1009
Catalog: C4000 Cephalometric Extension without Sensor
Model: 60-A1019
Catalog: Ceph Sensor only
Model: 60-A1004
Catalog: Panoramic Sensor
Model: 60-A1010
Class: llb
10
Progeny Vantage Panoramic X-ray System Installation Guide 1 Regulatory Information
Reference Numbers to which Conformity is
Declared
Declaration
Contact
The following regulatory documents apply:
IEC 60601-1
IEC 60601-1-2
IEC 60601-1-3
IEC 60601-2-7
IEC 60601-2-28
IEC 60601-2-32
IEC 60825-1
Medical Device Directive
ISO 13485
Midmark Corporation declares that the products described herein meet all the applicable Essential Requirements of the
EC Medical Device Directive 93/42/EEC in Annex I. For
Class IIb products described herein, the product is manufactured, inspected, tested, and released in accordance with the approved quality assurance system established in accordance with ISO 13485 and Annex II of the EC Medical Device Directive under the Supervision of the SGS United Kingdom Ltd., a Notified Body.
Technical Support
Midmark Corporation
Phone: 888-924-3800 (U.S. and Canada)
+1 847-415-9800 (International)
Fax: 847-415-9810 [email protected]
Hours: 8:00 a.m. – 5:00 p.m. CT
EMC Statement
Information Regarding Potential EMC Interference And Advice For Avoidance
• Magnetic and Electrical fields are capable of interfering with the proper performance of this device. For this reason, make sure that all external devices operated in the vicinity comply with the relevant EMC requirements.
• Main power quality should be that of a typical commercial or hospital environment.
• Power frequency magnetic fields should be at levels characteristic of a typical location in a commercial or hospital environment.
• If the above criteria cannot be verified, precautions shall be taken when using this equipment as the device may inadvertently operate.
11
Progeny Vantage Panoramic X-ray System Installation Guide
Authorized Representatives
North America
MIDMARK CORPORATION
675 Heathrow Dr.
Lincolnshire, Illinois 60069 U.S.A.
Phone: 888-924-3800 (U.S. and Canada)
+1 847-415-9800 (International)
Fax: 847-415-9810
Europe
CE Partner 4U
Esdoornlaah 13
3951DB Maarn
The Netherlands
Phone: +31.343.442.524
Fax: +31.343.442.162
1 Regulatory Information
12
2 Introduction
In this Chapter
•
Product Description
•
Disclaimer about the Manual
•
Symbols and Conventions
•
Obtaining Technical Support
Product Description
The Progeny Vantage Panoramic X-ray System is an easy to use and easy to install digital panoramic X-ray system.
The panoramic X-ray provides a broad overview of the teeth, jaw, and oral structure of the entire mouth. The X-ray image supplies information about the teeth, upper and lower jawbone, sinuses, and other hard and soft tissues of the head and neck. The panoramic digital receptor is contained in a C-arm that moves around the patient’s head.
The Vantage System has many applications that include evaluation of third molars, evaluation of patients with past and present TMJ (temporomandibular joint) problems, patients who require full or partial removable dentures, dental implants, or braces, those who are at risk or suspected of having oral cancer or other tumors of the jaw, those who have impacted teeth, and those who have had any recent trauma to the face or teeth (i.e., can help identify a fractured jaw).
The optional Cephalometric Extension allows for lateral and PA views of oral structures. The ceph digital receptor is mounted on a transit rail which scans horizontally during ceph examinations. Cephalometric images are typically useful in orthodontic evaluation and treatment.
Disclaimer about the Manual
Midmark pursues a policy of continual product development. Although every effort is made to produce up-to-date product documentation, this publication should not be regarded as an infallible guide to current specifications. We reserve the right to make changes without prior notice. The original language of this manual is English.
13
Progeny Vantage Panoramic X-ray System Installation Guide 2 Introduction
Symbols and Conventions
Symbol Explanation
Type B: Protection against electric shock (IEC 60601.1-
1988).
Consult written instructions in the User Guide.
ATTENTION RAYONS-X:
OPERATION SEULEMENT PAR DU PERSONNEL
AUTORISE. VOIR MANUEL DE L’OPERATEUR.
WARNING X-RAY
THIS X-RAY UNIT MAY BE DANGEROUS
TO PATIENT AND OPERATOR UNLESS
SAFE EXPOSURE FACTORS AND
OPERATING INSTRUCTIONS ARE
OBSERVED.
X-RAY EMISSION
Mains HOT WIRE
Mains NEUTRAL WIRE
Earth Ground
LASER RADIATION
DO NOT STARE INTO BEAM
CLASS 2 LASER PRODUCT. 650 nm, 3 mW
Power off (circle)
Power on (line)
14
Progeny Vantage Panoramic X-ray System Installation Guide
Obtaining Technical Support
For Technical Support, contact:
MIDMARK CORPORATION
675 Heathrow Drive
Lincolnshire, Illinois 60069 U.S.A.
Phone: 888-924-3800 (U.S. and Canada)
+1 847-415-9800 (International)
Fax: 847-415-9810 [email protected]
Hours: 8:00 a.m. – 5:00 p.m. CT
2 Introduction
15
3 System Overview
In this Chapter
•
About the Vantage System
•
About the Panoramic X-ray Device
•
About the Optional Cephalometric Extension
About the Vantage System
The Progeny Vantage Panoramic X-ray System consists of the panoramic X-ray device, the exposure button, and a Touch Control Panel.
About the Panoramic X-ray Device
The panoramic X-ray device consists of the telescoping column, the overhead assembly, the patient positioning table, and the touch control panel.
Telescoping Column
The telescoping column has two main parts: the fixed section and the moving or telescoping section. The fixed section contains the actuator to control up and down movement of the panoramic X-ray device. The telescoping section mounts the patient positioning features. Optical sensors in the telescoping section define the maximum and minimum extension of the column.
Overhead Assembly
The overhead assembly consists of an overhead arm and C-arm. The overhead arm supports the C-arm, which rotates. The C-arm includes the tubehead and the removable sensor. The tubehead produces the X-ray beam, and the sensor is a digital image receptor.
Patient Positioning Table
The patient positioning table guides and supports the patient’s head during acquisition of panoramic X-ray images by means of the chin rest, bite guide, and positioning wands. The positioning control on the side of the patient positioning table has 4 buttons for the operator to control the up/down movement of the telescoping column, to apply and release the positioning wands, and to turn on positioning lights. The patient positioning table also contains a storage compartment.
16
Progeny Vantage Panoramic X-ray System Installation Guide 3 System Overview
Touch Control Panel
The touch control panel is mounted on the fixed column section under the patient positioning table. It is the main user interface for taking X-ray images with the
Vantage System and is activated by touch.
Exposure Button
The exposure button is used by the operator to take the X-ray. The basic configuration consists of an exposure button connected to the panoramic X-ray device by a coil cord.
Bite Guide
A bite guide helps the patient keep his or her jaw correctly positioned. Additional bite guides may be obtained from Progeny. Always install a fresh protective sheath over the bite guide before positioning a patient. The sheath for this application is the Progeny part number 60-S0027.
Chin Rest
An easily removable chin rest fits into an opening on the patient positioning table.
Additional chin rests may be obtained from Progeny.
TMJ Positioner
A TMJ positioner for TMJ X-rays fits into the patient positioning table. TMJ positioners are included with the Vantage.
Emergency Stop Switch
The stop switch, mounted under the left side of the patient positioning table near the telescoping column, is for use by the patient. Depressing the button will immediately halt all motor movement. Touching the control panel surface will also abort any movement. The button can be released by turning the knob.
About the Optional Cephalometric
Extension
The Cephalometric Extension consists of the support arm, the cephalometric scanning mechanism and the cephalometric sensor.
Support Arm
The support arm is an aluminum casting that mounts to the panoramic column’s inner (mobile) component. It mounts via a support casting that allows vertical adjustment of the position of the cephalometric mechanism. The arm can be mounted either to the right, or to the left of the column.
17
Progeny Vantage Panoramic X-ray System Installation Guide 3 System Overview
Cephalometric Scanning Mechanism
The cephalometric scanning mechanism is comprised of the rail and the transit assembly. The rail supports the transit mechanism, and houses the motor and bearings that cause the transit mechanism to move. The transit assembly mounts the cephalometric sensor and the secondary collimator, and maintains the alignment of the sensor and secondary collimator to the panoramic tubehead.
The rail also supports the cephalostat, used to position the patient. The cephalostat is comprised of two otic positioning posts which adjust to conform to the width of the patients head, and a nasion locator, which adjusts vertically and laterally to align the patients head. The entire cephalostat rotates in 22.5° increments to support many cephalometric imaging positions.
Cephalometric Sensor
The cephalometric sensor is similar in appearance to the panoramic sensor, but houses within a 21cm long digital detector instead of the 14cm digital detector used for normal panoramic exam. The cephalometric sensor can be used in place of the panoramic sensor, but the panoramic sensor will not operate as a cephalometric sensor.
18
Progeny Vantage Panoramic X-ray System Installation Guide
Vantage System Panoramic X-ray Device
3 System Overview
19
Progeny Vantage Panoramic X-ray System Installation Guide
Vantage System Patient Positioning Table
3 System Overview
20
Progeny Vantage Panoramic X-ray System Installation Guide
Optional Cephalometric Extension
3 System Overview
21
4 Pre-Installation
Planning
In this Chapter
•
Environmental Requirements
•
Support Requirements
•
Power and Cable Requirements
•
Space Requirements
•
Network and System Requirements
Environmental Requirements
Temperature
The Vantage panoramic device is intended for indoor use for normal dental applications at temperatures in the range +10 C to +35 C (+50 F to +95 F).
Storage temperature range should not exceed -35 C to +66 C (-31 F to +150 F).
Humidity
Humidity should not cause condensation to form on the device. When the device is being operated, humidity should not exceed 95% RH non-condensing. When the device is being stored, humidity should not exceed 90% RH non-condensing.
22
Progeny Vantage Panoramic X-ray System Installation Guide 4 Pre-Installation Planning
Support Requirements
About Support
The Vantage panoramic device is wall mounted. As an alternative, the Vantage panoramic device can be installed as a free standing unit. If the free standing installation method is used, the free standing base kit must be attached to the unit for support.
Wall Mounted Installation
Wall fasteners for the Vantage panoramic device must be able to withstand a 68 kg (150 lb.) shear force and a 180 kg (400 lb.) tensile (pull-out) load. The floor must be able to support approximately 90 kg (200 lb./sq. ft.) for wall mounted installation.
Free Standing Installation
The floor must be able to support approximately 158 kg (100 lb./sq. ft.) for free standing installation. Free standing installation cannot be used for the
Cephalometric Extension.
Power and Cable Requirements
Electrical Outlet Requirements
The Vantage System requires a dedicated, 15 amp minimum circuit. If a fixed connection is used, the length of the whip must conform to local codes. If a standard mains receptacle is used, it must be placed within 2 m (6 ft.) of the device and positioned in compliance with local codes.
Ethernet Connections
Cat 5e grade communications cable is required for connection of the Vantage panoramic system.
Optional Remote Exposure Switch
If a remote exposure switch is used, a four conductor cable capable of RJ 11 termination is required for installation.
23
Progeny Vantage Panoramic X-ray System Installation Guide 4 Pre-Installation Planning
Space Requirements
The fully extended column has a maximum height of 236 cm (93 in.) and depth of
121 cm (48 in.). The rotational reach of the overhead arm and C-arm is 108 cm
(48 in.). See the diagram below.
The Vantage panoramic device has an optional right or left entry. The default configuration is left entry. To configure a right entry, see Optional Right Entry
Configuration in Chapter 6.
Panoramic Space Requirements
24
Progeny Vantage Panoramic X-ray System Installation Guide 4 Pre-Installation Planning
25
Network and System Requirements
The Vantage Software Client requires a computer for it to be installed on and a network connection to communicate with the Vantage Panoramic.
Network Requirements:
• An Ethernet based network connection at 100MBps or higher
• A wireless network connection at 802.11n or higher***
***A wireless network connection may introduce speed limitations that can result in a less than desirable image transmission time
System Requirements:
Component
Computer
Hardware
Memory
System
Operating
System
Requirement
PC - compatible Pentium 4 / 1.4 GHz or greater computer
2 GB RAM or higher recommended (minimum 1 GB)
Disk Space
Display
Settings
Microsoft Windows XP Professional with Service Pack 3; Microsoft Vista
(Business or Ultimate editions); Microsoft Windows 7 (Professional or Ultimate editions)
450 MB minimum
NOTE: Additional disk space is needed depending on the size of the practice, the number of images, and other information you plan to store. Each image is approximately 4 MB. For example, approximately 300 GB are needed to store 75 000 images.
1024 x 768 (16 - bit or higher) with 32 MB (or higher) of Video RAM
NOTE: It is possible to increase these settings based on the actual video adapter installed. As a rule, the better your video adapter or capture card the better your images.
26
5 Installation
In this Chapter
•
About Installation
•
Installing the Vantage Panoramic Device on a Wood Stud Wall
•
Installing the Vantage Panoramic Device on a Free Standing Base
•
Optional Right Entry Configuration
About Installation
The Vantage System panoramic device is preassembled and can be installed in a few easy steps. It has been designed to be installed by one person with an assistant in one hour, assuming that all pre-installation requirements have been met. Checking the image alignments can take an additional hour.
Check Pre-installation Requirements
Prior to beginning the installation, be sure that all pre-installation requirements have been completed. This includes confirming that the wall and floor support requirements are adequate for mounting the Vantage panoramic device, that the electrical requirements are met, and that wire locations are proper.
Optional Right Entry Configuration
The Vantage System is packaged with a left entry but can be configured with a right entry. To configure a right entry, see Optional Right Entry Configuration in
this chapter.
27
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
Installing the Vantage Panoramic Device on a Wood Stud Wall
Preparing to Install the Vantage Panoramic Device
Gather Tools
• Level
• 2 mm hex key
• 4 mm hex key
• 6.3 mm (1/4 in.) pilot hole drill
• 9.5 mm (3/8 in.) pilot hole drill (masonry bit for cement floor)
• 14 mm (9/16 in.) socket wrench (for cement or wood floor)
Unpack the Hardware Kit
• Hardware for wall (wood mount): 1 wall bracket, 2 lag bolts, and 2 washers
• Hardware for cement floor: 2 cement anchors, 2 nuts, and 2 washers
• Hardware for wood floor: 2 lag bolts and two washers
• Hardware for attaching wall bracket to column: 2 clamps and 2 screws
Unpack the Removable Parts Kit
• Chin rest
• 2 wands
• Bite guide
28
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
Cautions
When using lag screws as the method of attachment, it is imperative to consider the full scope of the task. Several factors must be considered for safe, permanent installations. Some of the key issues are below:
• Lumber commonly used in construction projects can be different from location to location.
• The grade, age, position, and overall condition of wood can vary greatly.
• The attachment stud may have additional, hidden loads.
• The location of the pilot hole with respect to the center of the stud will affect the load bearing ability.
• The size of the pilot hole required for the lag screw will be different based on the grade, age, and condition of the lumber.
• Never over-tighten the lag screw as this will weaken the mechanical connection (18 ft.-lb. maximum).
• Lumber with splits or cracks should not be used for attachment.
• Plywood, particle board, or similar construction materials should not be
used for attachment.
• Progeny provides fasteners for average installations. Based on specific installation conditions, it may be necessary to choose an alternate fastener or fastening methods.
• Seek the advice of a professional structural engineer to clarify any issues before the installation.
About the Order of Installation
For convenience and access, Progeny recommends the following order of installation described in the following sections:
• Remove the shipping brackets
• Attach the power
• Attach the bracket to the wall
• Snap the Vantage panoramic device in place
• Plumb the Vantage panoramic device and attach the clamps
• Mount the Vantage panoramic device to the floor
29
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
Remove the Shipping Brackets
Before attaching power, you need to remove the shipping brackets.
1. Remove the overhead top cover and remove the three orange shipping bolts from the overhead pivot area.
Overhead Top Cover
2. Remove the 4 bolts holding the orange shipping bracket in place as shown in the following figure.
Remove Aluminum Shipping Bracket
30
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
Attach the Bracket to the Wall
The mounting bracket is a guide for locating where to drill the holes used to mount the Vantage panoramic device to the wall. Carefully placing the mounting bracket on the wall will help to insure correct installation of the Vantage panoramic device.
1. Center the bracket on the studs.
2. Hold the bracket so that it is 121 cm (47 ½ in.) from the floor and mark one of the bracket holes.
3. Drill a 6.3 mm (1/4 in.) hole.
4. Select the 2 wall mount washers and the 2 lag bolts from the hardware kit.
5. Put 1 lag bolt through the washer and bracket mounting hole and loosely tighten the bolt.
6. Lift up the other side of the bracket and level it.
7. Mark the second hole and drill it.
8. Put the second lag bolt through the second washer and second bracket mounting hole and loosely tighten the bolt.
9. Place a level on the top of the bracket and level the bracket.
10. Tighten the bolts to 14-18 ft. lb. maximum.
Test the Bracket
31
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
Snap the Vantage Panoramic Device in Place
The Vantage panoramic device and bracket are designed to snap together when correctly aligned and firm pressure is applied. This snap holds the device in place while you complete the installation.
1. Lift up and move the Vantage panoramic device to the mounting wall with the back positioned towards the wall mounted bracket.
2. Line up the Vantage panoramic device with the wall mounted bracket and snap the device into place, making sure the device snaps solidly.
CAUTION!
The snap partially secures the Vantage panoramic device. The clamps and floor mount steps must be completed to hold the device in place.
Vantage Panoramic Device and Wall Bracket
32
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
Plumb the Vantage Panoramic Device and Attach the Clamps
The Vantage panoramic device and wall bracket must be joined with 2 clamps.
This requires a process of alternately screwing the clamps into place and plumbing the device.
1. Select 2 clamps and 2 screws from the mounting hardware kit.
Clamp and Screw
2. Hold 1 clamp in place, aligning it with one of the small holes in the bracket.
Insert a screw through the wall bracket hole and the Vantage panoramic device hole. Tighten the screw using a 4 mm hex key. Do not tighten completely.
Clamp Placement
33
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
3. Hold a level vertically against the column and move the Vantage panoramic device on the floor until it is plumb.
Column Test
4. Once the column is plumb, repeat the procedure with the second clamp. Do not tighten completely.
5. Do a final level test before completely tightening the clamp screws.
34
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
Mount the Vantage Panoramic Device to the Floor
Use the floor base as a template to drill holes in the floor. You do not need to move the Vantage panoramic device for this procedure. Ensure that the device is positioned vertically before drilling holes.
Note
To install the Vantage panoramic device on a free standing base, see Installing the Vantage Panoramic Device on a Free Standing Base in this chapter.
1. Select 2 cement anchors, 2 washers, and 2 nuts from the hardware kit. For wood floors, select 2 lag bolts and 2 washers.
2. Drill through 1 hole in the floor base of the Vantage panoramic device using an 9.5 mm (3/8 in.) masonry bit. For wood floors, use a 6.3 mm (1/4 in.) bit.
Drilling through Floor Base Hole
35
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
3. Insert 1 cement anchor through the hole in the plate. Assemble 1 nut and 1 washer on top of the cement anchor and tighten. Ensure that the anchor is fully seated in the hole. For wood floors, assemble 1 lag bolt and 1 washer and insert the lag bolt and washer through the hole in the plate. Screw the lag bolt partially into the hole using a socket wrench.
4. Repeat the process for the second anchor, nut, and washer. For wood floors, repeat the process for the second lag bolt and washer.
Inserting Anchor
5. Tighten the anchors using a 14 mm (9/16 in.) socket wrench. For wood floors, tighten the lag bolts using a 14 mm (9/16 in.) socket wrench (same size as used for cement).
Tightening Anchor
36
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
Installing the Vantage Panoramic Device on a Free Standing Base
Preparing to Install the Vantage Panoramic Device on a Free Standing
Base
Gather Tools
• Level
• 2 mm hex key
• 4 mm hex key
• 8 mm hex key
• 17 mm socket wrench
Unpack the Hardware Kit
• 2 kinds of bolts: 4 each
• 8 levelers
• 8 leveler plugs
Unpack the Removable Parts Kit
• Chin rest
• 2 wands
• Bite guide
About the Order of Installation
For convenience and access, Progeny recommends the following order of installation described in the following sections:
• Arrange the free standing base halves and levelers
• Attach the Vantage panoramic device to the free standing base
• Remove the shipping brackets
• Attach the power
• Position and mount the Vantage panoramic device to the floor
37
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
Arrange the Free Standing Base Halves and Levelers
Prepare the free standing base for attachment to the Vantage panoramic device using the following steps:
1. Remove the free standing base halves from the packaging.
2. Install the levelers in the holes as shown below.
3. Arrange the base halves as shown below, with the levelers in contact with the flooring.
Free Standing Base Halves and Levelers
38
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
Attach the Vantage Panoramic Device to the Free Standing Base
To attach the free standing base to the Vantage panoramic device, set the
Vantage panoramic device on top of the base halves and install the bolts and bracket, using the following steps:
1. Move the Vantage panoramic device over the bases with the overhead pointing the same direction as the legs of the bases.
2. Install the bolts and bracket as shown below, back and front views.
Attaching Vantage Panoramic Device to Free Standing Base:
Back View
39
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
Attaching Vantage Panoramic Device to Free Standing Base:
Front View
40
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
Remove the Shipping Brackets
Before attaching power, you need to remove the shipping brackets.
1. Remove the overhead top cover and remove the three orange shipping bolts from the overhead pivot area.
Overhead Top Cover
2. Remove the 4 bolts holding the orange shipping bracket in place as shown in the following figure.
Remove Aluminum Shipping Bracket
41
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
Position and Mount the Vantage Panoramic Device to the Floor
To attach the Vantage panoramic device to the floor, you need move the unit into position, level the unit, and mount it to the floor using the following steps:
1. Move the Vantage panoramic device on its base into position, and use an 8 mm hex wrench to level the device, ensuring that all levelers are in contact with the flooring.
2. Fix the unit to the floor with the flooring appropriate fasteners through the holes in the center of the levelers.
42
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
Installing the Vantage Cephalometric
Extension on a Vantage Panoramic
Device
Preparing to Install the Vantage Panoramic Device on a Free Standing
Base
1. Place the Progeny Vantage at a comfortable working level and then power the unit off.
2. Ensure the box is on its side with the label “This side up” facing up. Do not destroy the box. It will be used to help with the installation of the Ceph Unit.
43
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
3. Open the box on both ends and remove the accessories box and Ceph Senor Box.
Accessories Box Ceph Sensor Box
4. Remove the Cardboard filler and pull out the Ceph unit. The mounting arm is located beneath the Ceph Unit.
Pull out cardboard filler Pull out the ceph Unit
Pull out the Mounting Arm
44
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
RJ45 Ethernet cable
50 Speculum
60-P4009
Daughter
PCB
Head Positioners
5. Open the accessories box and check for the following contents:
Power Jumper
&
RS232 Jumper
Flat Gray RJ50
Cable
Ceph Covers
Ceph Extension mounting casting
Flanged cap screws
M8x1.25
Ceph arm leveling bolt with Nylok M8
Ceph Screws
M6 x 1.0
4 M8 x 1.0 cap screws
Cat 5 cable and
Ceph Cover
Screws adapter
Gap Cover
Calibration
Fixture
45
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
6. Remove the covers, indicated below, from the Vantage a. Upper column cover (with mirror):
Pull the upper Column Cover toward you but do not completely remove it.
Unplug the connector for the Frankfort plane laser. Once the Frankfort plane laser is disconnected, completely remove the cover. Note: you may have to move the overhead all the way to the right to clear the Gap Cover. b. Remove the two lower column covers
Grab the Cover with the
Vantage logo on it and pull it toward you and remove it.
Grab the lower cover at the bottom pull it toward you and remove it.
46
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
7. Mount the Ceph extension mounting casting on the Vantage using 4 M8 x 1.0 cap screws.
Tighten the screws
Four mounting holes
4 M8 x 1.0 cap screws
8. Mount the Ceph arm on the casting using 4 M8 x 1.25 flanged cap screws. Do Not
Tighten.
47
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
9. Mount the ceph arm leveling bolt with Nylok M8 nut between the arm and the mounting casting.
10. Level the arm of the Ceph to the base of the machine.
10A
10B
Place a level on the base of the column. Take
note of the bubble position. You will be transferring this bubble position to the end of the
Ceph Arm.
Place the Level on the flat surface of the Ceph
Arm, as shown above.
You will be adjusting the angle of the Ceph arm to match the level position measured in the previous step.
48
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
10C 10D
Level the arm by turning the Nylon Nut. Ensure that you match the bubble level of the base that was measured in step 10A.
Tighten the 4 M8 x 1.25 flanged cap screws.
11. Mount the RTC Daughter Board. Take note of the Connectors.
Connector
Connector
J4
Connector
J5
J1
RTC Daughter Board
Connector
J2
Connector
Connector
J3
J6
49
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
11A
Remove the ribbon cable from connector J24 of the RTC.
11C
Plug the ribbon cable (removed in step 11A) into connector J5 of the Daughter Board
E
13.
14.
Locate the Power Jumper Cable from the
Accessory Kit and plug it into connector J3 of
17. the RTC the RTC.
11B
Plug the Daughter Board into connector J24 of the RTC
11D
Remove the cable from Connector J21 on the
RTC and move it to connector J3 of the
Daughter board.
11F
Locate the RS232 Jumper Cable from the
Accessory Kit and plug it into connector J2 of the Daughter Board and any of the available
Motor CTLR connections on the RTC (J1, J10, or J12).
50
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
12. Locate the RJ50 Flat Cable and the RJ45 Ethernet cable from the Accessory Kit. Route the cables through the Ceph Arm as shown below.
12A
Route both cables through the arm as shown.
12C
Route the cables through hole in the Arm Mount.
12B
The cables will come out at the other end of the
Arm where the Ceph Unit mounts.
12D
Route the Ethernet Cable through the opening at the end of the patient table as shown.
51
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
12E
Plug the Ethernet Cable into any of the available
3 ports on the left of the POE. Do not use the far left port. This is reserved for the Control Panel.
12F
Plug the Flat RJ50 Cable into connector J1 of the
Daughter board.
12G
Front View Rear View
At the other end of the Ceph Arm, tuck the cables through the hole as shown. Ensure the cables are tucked out of the way to ensure that they are not pinched when the Ceph is mounted.
52
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
13. Install the Ceph onto the Ceph Arm.
13A
The Ceph box can be used to install the
Ceph unit as shown above.
With the top of the box open, place the ceph unit on top of the box.
Turn on the Vantage and wait for the unit to fully boot. Move the column up or down to match the level of the Ceph arm with the Ceph on the box. Turn off the
Vantage.
13B
13C
3 M6 x 1.0
2 M8 x 1.25
13D
Ensure the cables are free and clear and cannot be pinched.
Using 2 M8 x 1.25 flanged screws and 3 M6 x
1.0 flanged screws, secure the Ceph Unit onto the Ceph Arm.
Move Ceph unit into the opening of the Ceph
Arm and prepare to secure it down.
13E
Route the RJ45 Ethernet Cable and the RJ50 Flat cable through the cavity. Plug the RJ45 Ethernet cable into the Silver connector. Plug the RJ50
Flat cable into the Black connector.
53
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
14. Remove the covers from the Secondary Collimator.
Remove the Top Cover from the secondary collimator by removing the two screws.
14A
14C
The Front Cover is pressed into place. Gently pry the front cover away from the Secondary
Collimator. the connector from J2.
Remove the three screws form the secondary collimator rear cover and remove the cover.
14B
Remove the wire guide screws and plate. Unplug
14D
54
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
15. Locate the calibration fixture from the accessory kit and mount it on secondary collimator.
15A end away from the vertical line utilizing two thumb screws.
Front View Rear View
Utilizing 2 thumb screws, Mount the Calibration Fixture onto the secondary collimator with the heavy
15B 15C
Locate the Cat5 cable and adapter in the accessory kit. Plug the Cat5 cable into the connector on the alignment fixture.
At the top of the Ceph, disconnect the Cat 5 cable from the Silver connector. This is the same cable installed in step 13D…see next step
55
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
Connect the Cat5 cable to the adapter.
15D
16. Install the Sensor Calibration File.
16A
Locate the USB drive that is in the same box as the sensor. Remove the USB Port cover on the back of the Operator panel.
16B
Plug the USB Drive into the available
USB port at the back of the Operator
Panel.
16D 16E
E
Select “Options” on the lower right corner of the operator panel.
16F
Select Calibration Files on the upper left.
Press on the Green arrow to load the calibration files onto the system.
This window will display the calibration files loaded on the USB
Drive.
56
15
E
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
The Calibration files will begin transferring.
16G 16H
When the transfer is complete, a green check mark will display. You may now select “Close” and remove the USB Drive from the back of the operator panel.
17. Ceph Calibration Procedure. Power on the Vantage.
Select “Options” on the lower right of the display.
17A
17C
17B
17D
16I
Turn the Vantage off. Wait 10 seconds and then turn it back on.
Select “Service” on the lower left
Enter 77469 and Select OK. Highlight the Ceph option by pressing on it.
57
Progeny Vantage Panoramic X-ray System Installation Guide
Select Axis Service and Diagnostics
17E
17G
Select “L”
The window above will display.
H
5 Installation
Locate the small Zero Alignment pin.
17H
17F
Find the alignment holes on the side closest to the sensor. Continue to move the Ceph arm until the two alignment holes overlap…see next image.
58
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
K
Place the Zero Alignment Pin into the alignment hole to align the two holes. Note: when the pin is inserted properly, you should not be able to move the Ceph Arm…See Next Image
17M
Carefully remove the Zero Position Pin…Try not to disturb the Ceph Arm Position.
17O
Zero Position Pin Correctly placed.
Select Set Zero Alignment.
17L
17N
17P
Select “Yes” in the confirmation window. The Ceph Arm will move slightly and the message “Procedure
Successfully Completed will Display”.
Select “Move to Zero”. The Ceph Arm will move to the home position and then to the Zero
Position.
59
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
17Q
18. Align C-Arm
Select “Align C-Arm”.
Re-insert the Zero Position Pin. It should Fit and lock the unit into place without having to move it. If not, restart at step 17I. If the pin fits properly, select “Finish”.
18A
18C proceeding. If the Pan Sensor is connected, remove it before proceeding.
18B
The message above will appear. This step has already been accomplished in step 15. If this has not been done, please complete step 15 before
18D
Select Calibrate Axes. The Ceph Arm will move, when it is done moving, take an exposure.
You will take a series of exposures. The unit will calibrate itself. Continue to follow the prompts on the display.
60
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
18E
Continue to take exposures until the display indicates that the calibration is complete and the group of numbers are green. Select Finish and Finish again.
19. Align the secondary collimator
19A
Unplug the sensor from the calibration Fixture and plug it into the Ceph Arm…See next step.
19C
Plug the sensor into the Ceph Arm.
19B
19D
Remove the Calibration Fixture and Ethernet
Cable with Coupler.
Plug the Ethernet cable (unplugged in section
15) back into the Silver connector.
61
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
19E
Select “Align Secondary Collimator”
19G
Select “Calibrate Axes”. The unit will move into position and the moving indicator will blink. When the unit stops moving and the moving indicator stops blinking, take an exposure.
Adjustment
Loosen
19I
Loosen the four Allen screws shown above. Turn the adjustment screw in the direction and distance given in the previous step. Tighten the Allen Screws.
19F
The message above will display. Ensure the
PAN Sensor (not the Ceph sensor) is removed before proceeding. Select OK.
19H
The software may indicate that the adjustment screw needs to be turned. Take note of the amount and direction to turn the adjustment screw…See next step.
19J
Once the Adjustment has been made, select “Continue”. When the unit stops moving, take another exposure. If the result is satisfactory, the unit will not ask you to turn the screw again.
62
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
19K
Continue to take exposures until the group of numbers turn green and the display indicates that the procedure has successfully completed. Select Finish and Finish again.
20. Calibrate the Primary Collimator
Select “Calibrate Primary Collimator”
20A 20B
The message above will display. Ensure the
PAN Sensor (not the Ceph sensor) is removed before proceeding. Select OK.
20C
Select “Calibrate Axes”. The unit will move into position and the moving indicator will blink. When the unit stops moving and the moving indicator stops blinking, take an exposure.
20D
Follow the prompts on the display and continue to take exposures until the unit displays “Calibration successfully completed”
63
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
Calibration Successfully Completed. Select Finish
20E 20F
Note: There may be certain installations where the upper or lower calibration numbers never resolve to zero. In these cases, it is acceptable to adjust the level of the ceph arm until the sensor is completely covered by the x-ray field. Adjust the arm up if the bottom is not covered, and down if the top is not covered. Note:
You will need to support the weight of the arm while advancing the adjustment nut.
21. Verify Calibration
Select “Verify Calibration”
21A
When the “Verification Successful” message appears on the screen, select Finish.
21B
Select “Calibrate Axes”. The unit will move into position and the moving indicator will blink. When the unit stops moving and the moving indicator stops blinking, take an exposure.
21
C The calibration of the unit is now complete.
21D
Exit diagnostic mode.
64
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
22. Place the covers on the Vantage and on the Cephalometric unit.
Replace the covers on the Vantage.
These were removed earlier in Step 6.
23A 23B
Replace the secondary collimator cover removed earlier in step 14C.
23C 23D
Reconnect connector J2 and secure. This was removed in step 14B.
Install the top cover on the Ceph unit. The 3
Philips head screws are located in the accessory kit.
65
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
23E
Install the rear cover on Ceph Arm. The cover is located in the accessory kit.
23G
23F
Install the curved cover on the front of the Ceph.
The cover is located in the accessory kit.
Install the cover at the end of the Ceph Arm. The cover is located in the accessory kit.
66
Progeny Vantage Panoramic X-ray System Installation Guide 5 Installation
Optional Right Entry Configuration
The Vantage panoramic device is delivered with left entry. To configure with a right entry, remove and rotate the Frankfort Plane Laser Assembly using these steps:
1. To access the Frankfort Plane Laser Assembly, remove the cover from the telescoping section of the column.
2. Disconnect the laser wires.
3. To access the new screw holes where you will install the assembly, remove the small cover on the left side of the mirror assembly (as viewed from the back of the column). Set the small cover aside for later assembly.
4. Remove the 2 screws from the Frankfort Plane Laser Assembly and lift the assembly out. Set the screws aside for later assembly.
Frankfort Plane Laser Assembly
5. Rotate the Frankfort Plane Laser Assembly 180 degrees so that it is upside down.
6. Install the Frankfort Plane Laser Assembly on the left side of the column (as viewed from the back of the column) using the 2 screws.
7. Place the small cover over the screw holes on the right side (as viewed from the back).
8. Reconnect the laser wires.
9. Replace the column cover.
67
Progeny Vantage Panoramic X-ray System Installation Guide 6 Installing the Cables
6 Installing the
Cables
In this Chapter
•
Connecting the Vantage to Your Network
•
Installing the Exposure Button
Connecting the Vantage to your network
The Vantage System requires a PC that serves as a workstation to receive images from the Vantage panoramic device. The Vantage must be connected to the same network as the designated workstation in order to connect and receive an image.
1. Attach the PC Ethernet cable to the network connection port on the back of the column and the other end into your network wall jack
Cable Connecting Ports
68
Progeny Vantage Panoramic X-ray System Installation Guide 6 Installing the Cables
Installing the Exposure Button
The exposure button is used to take the X-rays. The basic configuration is to connect the button to the Vantage panoramic device by means of a coil cord.
Attach the Exposure Button to the Vantage Panoramic Device
To attach the exposure button, insert the coil cord into one of the exposure switch connections on the column shown in the figure above. The exposure switch connections are the top 2 ports.
Mount the Exposure Button
The exposure button comes with double stick tape and a hole for a screw, if desired, for mounting. It can be mounted to the fixed portion of the column or to a wall.
1. To mount the exposure button using the double stick tape, remove the protective backing covering the adhesive mount.
2. Firmly press the mount to the wall or surface as shown in the figure below.
Exposure Button Mounted
69
7 Starting Up
In this Chapter
•
Turning the Vantage System On
•
Checking Image Quality
•
Progeny Support Information
Turning the Vantage System On
You can turn on the Vantage panoramic X-ray device first or the client software first (see chapter 8 Vantage software client setup). Make sure that all shipping brackets have been removed.
1. To turn on the Vantage panoramic device, press the on/off switch located at the back of the stationary column on top of the box of connections.
Vantage Panoramic Device with On/Off Switch
When you turn on the Vantage panoramic device, the touch control panel displays a “Start Up” screen. The Vantage panoramic device goes through a self diagnostic procedure as part of the startup process.
2. If the Client software is not running, launch it and open the image acquisition software.
70
Progeny Vantage Panoramic X-ray System Installation Guide 7 Starting Up
Checking Image Quality
An image quality phantom is available from Progeny. The phantom simulates the position of average human teeth roots, and markers allow you to check the position of the focal trough. To check for image quality, you need to take an X-ray exposure with the image phantom and then view the image, checking it according to some specific tests. The Cephalometric imaging phantom is included with Cephalometric Extension option.
Install the Panoramic Image Phantom
The Vantage panoramic device is designed so that the chin rest accessory can easily be switched with the image phantom.
1. To install the image phantom, lift the chin rest from the chin rest hole on the patient positioning table.
2. Set the image phantom in the chin rest hole.
Image Checks
The following checks for image quality are performed on the image after exposing the phantom. In general, each individual line should be sharp where the lines cross the center of the image, and they should blur out above the center line.
1. Verify that the semi-projections b = c +/- 3 mm (nominal value with the central vertical line well focused) = 80 mm.
2. Verify that line “a” is vertical +/-3 degrees with respect to the horizontal line
“d”.
3. Verify that line “d” is horizontal and flat within a band of 6 mm.
71
Progeny Vantage Panoramic X-ray System Installation Guide 7 Starting Up
4. Verify that the darkened area is centered in respect to the central vertical line
“a” with a tolerance of +/- 4 mm.
5. Verify that the central vertical line is well focused and adjacent balls are round.
X
√
X
Take an exposure with the Cephalometric Imaging Phantom.
1
Locate the image phantom in the accessory kit.
Hang the image phantom in between the ear posts.
Take an exposure of the image Phantom at
66KV and 8 Ma.
2
72
Progeny Vantage Panoramic X-ray System Installation Guide
Should be centered
7 Starting Up
3
Rectangular boxes
The box within the center squares should be relatively centered in the box.
4
Zoom in on the rectangular boxes. The individual lines should be visible.
73
Progeny Vantage Panoramic X-ray System Installation Guide 8 Vantage Software Client
Setup
8 Vantage Software
Client Setup
In this Chapter
•
About
•
Installation Procedure
•
Connecting to Vantage
About
Vantage acts as a “network appliance” that connects directly to your network. In order to receive x-ray images you must designate at least one workstation on your network to receive them.
Installation Procedure
When installing the software it is assumed that previous versions of the
Progeny Device Suite and Progeny Imaging image management software are not present.
NOTE: Proper operation requires any previous version of Progeny Device Suite and Progeny
Imaging to be removed (uninstalled) prior to the installation process to begin.
Execute the following steps:
• Insert the USB Flash Drive into an available USB port on your computer and allow the computer to recognize the flash drive.
74
Progeny Vantage Panoramic X-ray System Installation Guide 8 Vantage Software Client
Setup
• The main screen of the installation software is shown on Figure 1. If
the software on the USB flash drive does not start automatically, navigate to Windows Explorer™ and select the “Progeny” drive letter. Browse to the content of the flash drive and start “Setup.exe”.
This step begins the installation process.
NOTE: The installation software requires Microsoft
.NET Framework revision 3.5. This software will be installed if it is not yet present to the operating system. Follow all on screen prompts.
NOTE: If the intended configuration is based on
Windows XP, the Service Pack 3 update is required.
This update is included on the USB flash drive and can be installed from folder named ‘Utilities’.
Another option is to use the Windows update tool provided by Microsoft.
Figure 1: Main screen of the Installation software
75
Progeny Vantage Panoramic X-ray System Installation Guide 8 Vantage Software Client
Setup
• Start the installation process by clicking on ‘Install Progeny Device
NOTE: The installed software requires multiple software components that may already be available in your system. These components will be installed if they are not yet present. Follow all on screen prompts.
Figure 2: Starting the Progeny Device Suite installation
• The screen on Figure 3 will be displayed. Choose Vantage Pan and
all other device families that have to be supported by the Imaging
Software.
Figure 3: Selecting the device families to be installed
76
Progeny Vantage Panoramic X-ray System Installation Guide 8 Vantage Software Client
Setup
• A green check mark next to the ‘Install Progeny Device Suite’ button will appear when Progeny Device Suite installation is completed.
Continue by installing Progeny Imaging software by clicking on
‘Install Progeny Imaging’ button (Figure 4) and follow the prompts on
the screen to perform the installation.
Figure 4: Starting the Progeny Imaging installation
• Green check marks next to each of the ‘Install Progeny Device Suite’ and ‘Install Progeny Device Suite’ buttons will appear when both the
Progeny Device Suite and Progeny Imaging are installed (Figure 5).
Figure 5: Progeny Device Suite and Progeny Imaging are installed
77
Progeny Vantage Panoramic X-ray System Installation Guide 8 Vantage Software Client
Setup
Connecting to Vantage
Connecting your computer to the Vantage system can be setup either through
Progeny Imaging image management software or through Progeny’s TWAIN interface.
• If you are using Vantage Pan from a 3 rd
party application you need to open Progeny’s TWAIN interface to connect to the Vantage system.
Some image management programs also support Vantage integration.
See your image management software support for more information on how to access the Vantage.
• Select “Vantage Pan” in the device drop down if it has not already been selected
• To access device configuration: o TWAIN: Click the device configuration icon o Progeny Imaging: In the top menu browse to Tools > Devices >
Device Configuration
• If you have already connected to a device it will be listed in the Server
Connection section. Click the “Connect” button.
78
Progeny Vantage Panoramic X-ray System Installation Guide 8 Vantage Software Client
Setup
• The “Select Pan Server” window will automatically discover the Vantages in your network. Highlight the vantage you want to connect to and click the “Select” button. o Optional User Name: The User Name field is what’s used to identify your Vantage Client in the network. The default is your computer’s name. o Optional Network Adapters: Progeny Device Suite will automatically detect the most suitable network adapter on your computer for connecting to the Vantage. If your Vantage is not automatically discovered click the “Net Config” button to select a specific network adapter.
79
Progeny Vantage Panoramic X-ray System Installation Guide 8 Vantage Software Client
Setup
• Once returned to Device Configuration screen the Vantage you selected will be visible in the “Server Connection” section. Click the “Close” button to complete the connection.
• Once you close Device Configuration and you are back at the client application the light for “Vantage Pan” should now turn green.
• At this point you can take images with the Vantage Panoramic and they will be delivered to the client. You can repeat these steps at any time to connect to another Vantage panoramic device.
Progeny Support Information
For Technical Support, contact:
MIDMARK CORPORATION
675 Heathrow Drive
Lincolnshire, Illinois 60069 U.S.A.
Phone: 888-924-3800 (U.S. and Canada)
+1 847-415-9800 (International)
Fax: 847-415-9810 [email protected]
Hours: 8:00 a.m. – 5:00 p.m. CT
80
Download
Advertisement
Key features
Digital panoramic X-ray system
Easy to install
Easy to use
C-arm that moves around the patient’s head
Provides a broad overview of the teeth, jaw, and oral structure
Frequently asked questions
The Progeny Vantage is an easy-to-use and easy-to-install digital panoramic X-ray system that provides a broad overview of the teeth, jaw, and oral structure of the entire mouth.
The Vantage is preassembled and can be installed in a few easy steps. It has been designed to be installed by one person with an assistant in one hour, assuming that all pre-installation requirements have been met.
The fully extended column has a maximum height of 236 cm (93 in.) and depth of 121 cm (48 in.). The rotational reach of the overhead arm and C-arm is 108 cm (48 in.).