The 6180 AeroDAQ is designed to meet the specifications of the aerospace industry. Automated scheduling of TUS (Temperature Uniformity Survey),
SAT (System Accuracy Test), and instrument calibration is handled by the
AeroDAQ based on furnace class and instrument type as per the relevant tables of AMS2750D. A Service Page clearly displays the number of days until the next activity is due with messages to alert that the due date is approaching. Additionally, the AeroDAQ also monitors control thermocouple life for number of days used, and load thermocouples for both number of days used and number of uses. A pre-batch thermocouple life check will indicate the availability of the thermocouples for the next batch returning either OK or Will Fail with the reason (either days or uses will expire). Data is stored in a tamper –resistant binary format that can be used for secure, long term records of your process. The high accuracy, low drift input boards enable the AeroDAQ to meet the requirements of
AMS2750D for Class 1 Furnace.
Available features*
Display
Channels
Relays
Groups
Auditor features
Virtual channels**
Timers
Alarms
Batch
Bridge- remote viewing software
Screen builder
Modbus Master
Security
6180 AeroDAQ
12.1” XGA
6
3
6 standard
Audit trail
36
Fitted as standard
4 per channel
Standard
Full as standard
Standard
Standard
Unlimited unique user names with configurable access permissions and passwords
Configuration software
Review/Quickchart Lite software
Standard views
Standard
Standard
Vertical and horizontal trending,
Vertical and horizontal bargraphs,
Circular trend and numeric values
* Standard AeroDAQ features shown. Additional options available as per ordering code
** Virtual channels can be configured as maths, totalisers, counters or comms
• E U R O T H E R M
®
F L E X I B L E S O L U T I O N S •
Data logging and archiving
The AeroDAQ has internal Flash memory for secure data storage.
They are also able to accept various removable media types (Compact
Flash or USB memory stick). Data stored within the internal memory can be archived to the removable media on demand or at preset intervals. The AeroDAQ will give indication of how long its internal memory and that of the removable media installed will last according to the configuration of the recorder.
The AeroDAQ has Ethernet capability. It can be configured to archive to the removable media and/or over Ethernet. Archiving files over
Ethernet effectively gives a secure, infinite archiving capacity.
Approximate duration for continuous recording of one group of six channels, high compression:
Archive media
32Mb Internal
Flash (approx. 4 million samples)
0.125s
0.5s
1s
Sample rate
5s
2.83 days 11.3 days 22.6 days 113 days
10s
226 days
30s 60s
1.86 yrs 3.7 yrs
64Mb CF/SD Card 5.66 days 22.6 days 45.3 days 226 days 1.2 yrs or USB memory stick (approx. 8 million samples)
3.7 yrs 7.4 yrs
4.9 yrs 14.8 yrs 20 yrs 256Mb CF/SD Card 22.6 days 90.6 days 181 days 2.4 yrs or USB memory stick (approx, 32 million samples)
9.6 yrs 19 yrs 58 yrs 116 yrs 1Gb CF/SD Card or USB memory stick (approx. 125 million samples)
Ethernet (FTP Server)
88 Days 354 days 1.9 yrs
Infinite
Time synchronisation (SNTP)
The AeroDAQ supports Simple Network Time Protocol which, when enabled, updates the instrument time every 15 minutes from the configured SNTP server. The unit can also act as a Unicast SNTP server on the network, allowing client instruments to synchronise with the
AeroDAQ to a resolution of one millisecond .
Batch recording
Up to ten user-defined fields can be used to enter batch specific data.
Field descriptor
– up to 20 characters
Operator entered batch information
– up to 60 characters
The user can choose to log any number of the given fields on start and / or stop of a batch. The information will appear on the chart as a message and cannot be separated from the process data to which it relates.
Audit trail
Standard on the AeroDAQ is a time stamped audit trail. This allows all actions taken on the recorder including all user logins, batch start and stop, TUS, SAT and calibration reset, and any configuration changes to be shown on the “chart” as well as becoming a permanent part of the history file. The audit trail information can be seen in historic views of the data.
Modbus master
Allows users to view data from multiple instruments connected either by a local Network connection using Modbus TCP, or a Serial connection using Modbus RTU,
ASCII printer output (reports)
Fitted as standard the ASCII text printer option provides the AeroDAQ with the ability to generate up to 10 simple reports that can be directed to a Serial ASCII text printer. Reports, triggered by an event/job can be configured to contain parameters such as time and date, batch names, process values and user defined messages.
Dynamic Host Configuration Protocol (DHCP)
Dynamic Host Configuration Protocol, the successor to BootP, allows an AeroDAQ host to obtain Network parameters, such as IP address,
Subnet Mask, default gateway and DNS server address dynamically.
The implementation of DHCP on the AeroDAQ significantly reduces the overhead for maintaining a network of instrumentation.
SPECIFICATION
Recorder
Environmental performance
Temperature limits Operation: 0 to +50°C
Storage: –20 to 60°C
Humidity limits Operation: 5% to 80% RH
Storage: 5% to 90% RH
Protection Bezel and display: IP66
Sleeve: IP20
Vibration (10 to 150Hz): BSEN60873, Section 9,18
Approvals
Electromagnetic compatibility CE, cUL (EMC)
UL file number: e57766
Emissions and immunity: BS EN61326
Electrical safety
(BS EN61010): Installation cat. II; Pollution degree 2
INSTALLATION CATEGORY II
The rate impulse voltage for equipment on nominal 230V mains is 2500V.
POLLUTION DEGREE 2
Normally, only non-conductive pollution occurs. Occasionally, however, a temporary conductivity caused by condensation shall be expected
Physical
Panel mounting:
Panel mounting angle:
Dimensions
DIN43700
±45°
Bezel size: 292 x 292mm
Panel cutout: 281 x 281mm (both –0/+1mm)
Depth behind bezel rear face: 261mm
Operator interface
Type: Colour TFT LCD with cold cathode backlight, fitted with resistive, analogue,
Touch-Panel
XGA (1024 x 768 pixels) 12.1” Size and resolution:
Power requirements
Supply voltage
Power (Max):
Standard: 85 to 265V ac; 47 to 63Hz or
110 to 370V dc
50W
Fuse type: None
Interrupt protection Standard: Holdup >200msec, at 240V ac, with full load
Back-up battery
Type: Poly-carbonmonofluoride/lithium
(BR2330) Part No. PA261095
Support time (RTC):
Replacement period:
Stored data:
1 year min. with recorder unpowered
3 years
Time; date; values for totalisers, counters and timers; batch data; Fvalue,
Rolling average, Stopwatch etc.
Ethernet communications
Protocols:
Cable
(IEEE802.3)
TCP/IP, FTP, DHCP, BOOTP, SNTP,
MODBUS, SMTP, ICMP
Type: CAT5
Maximum length: 100 metres
Termination: RJ45
Serial communications option
No of ports:
Protocol:
2
ASCII (typical applications: Input of
ASCII string inputs from Barcode readers,
Credit card readers etc.)
ASCII printer support
Modbus RTU Master and Slave
Isolation (dc to 65Hz
BS EN61010):
Terminals to ground:
Transmission standard:
Installation category II;
Pollution degree 2
50V RMS or dc (basic insulation)
EIA232 or EIA485 (software selectable)
Input Board
General
Input types:
Input type mix:
Maximum number of inputs:
A/D conversion method:
Input ranges:
Termination:
Noise rejection (48 to 62Hz): dc Volts, dc millivolts, dc milliamps (with shunt),
Thermocouple, 2/3-wire RTD
Contact closure (not Channels 1, 7, 13, 19,
25, 31, 37, 43) >60 ms
Freely configurable.
6 per board
>16 bits, 2nd order delta sigma
See Table1 and Table 2 below
Edge connector / terminal block
Common mode: >140dB (channel to channel and channel to ground).
Series mode: >60dB
Max common mode voltage: 250V continuous
Max series mode voltage: 45mV at lowest range;
23.74 Volts peak at highest range.
Isolation Channel to channel: 300V RMS or dc (double insulation)
Channel to common electronics: 300V RMS or dc (double insulation)
Channel to ground: 300V RMS or dc (basic insulation)
Dielectric strength
(BS EN61010) (1 minute type tests)
Channel to channel: 2500V ac
Channel to ground: 1500V ac
Insulation resistance:
Input impedance:
Over voltage protection:
>10MΩ at 500V dc
38mV, 150mV, 1V ranges: >10MΩ;
20V range: 65.3kΩ
50V peak (150V with attenuator)
Open circuit detection:
Recognition time:
Minimum break resistance:
± 57nA max.
500msec
10MΩ
Update/archive rates
Input/Relay-output sample rate: 8Hz
Trend update:
Archive sample-value:
Display value:
8Hz maximum
Latest value at archive time
Latest value at display update time (8Hz)
DC Input ranges
Shunt: Externally mounted resistor modules
Additional error due to shunt: 0.1% of input
Additional error due to attenuator:
Performance
Long term drift
0.2% of input
See Table 1
Meets the requirements of AMS2750D for
Class 1 Furnace
Requires an Input Adjust be performed to meet the requirements of AMS2750D
Low High
Range Range Resolution
Typical error
(instrument at 20°C)
Range
Maximum error
(instrument at 20°C)
Range
-38mV 38mV 1.4μV 0.013% I/P + 0.031% 0.030% I/P + 0.052%
-150mV 150mV 5.5μV 0.013% I/P + 0.028% 0.029% I/P + 0.039%
-1V
-20V
1V
20V
37μV
720μV
0.013% I/P + 0.024% 0.029% I/P + 0.029%
0.075% I/P + 0.027% 0.393% I/P + 0.033%
Table 1 Voltage ranges - accuracy and resolution
Worst case temp
Performance
Input per °C
25ppm
25ppm
25ppm
388ppm
Low High
Range Range Resolution
0 Ω
0 Ω
0 Ω
150
600
5K
Ω
Ω
Ω
5m Ω
Typical error
(instrument at 20°C)
Range
Maximum error
(instrument at 20°C)
Range
0.027% I/P +0.034% 0.037% I/P + 0.077%
22m Ω 0.027% I/P +0.035% 0.037% I/P + 0.057%
148m Ω 0.030% I/P + 0.034% 0.040% I/P + 0.041%
Table 2 Resistance ranges - accuracy and resolution
Worst case temp
Performance
Input per °C
30ppm
30ppm
30ppm
Thermocouple data
Temperature scale:
Bias current:
Cold junction types:
ITS 90
0.05nA
Off, internal, external, remote
1°C max with inst. at 25°C CJ error:
CJ rejection ratio:
Upscale/downscale drive
Types and ranges:
50:1 minimum
High, low or none selectable for each thermocouple channel
Additional error: 0.01°C (typ.) if high or low selected
See Table 3
Resistance inputs
Ranges (including lead resistance):
Influence of lead resistance:
0 to 150Ω, 0 to 600Ω, 0 to 6kΩ
Error: Negligible
Temperature scale:
Accuracy and resolution:
RTD types and ranges:
Mismatch: 1Ω/Ω
ITS90
See Table 2
See Table 4
T/C Type
B
Overall range (°C)
0 to +1820
Standard
IEC 584.1
C
D
E
K
L
J
G2
0 to +2300
0 to +2495
–270 to +1000
0 to +2315
–210 to +1200
–270 to +1372
–200 to +900
S
T
N
R
–270 to +1300
–50 to +1768
–50 to +1768
–270 to +400
U –200 to +600
NiMo/NiCo –50 to +1410
Ni/NiMo
Platinel
0 to +1406
0 to +1370
Pt20%Rh/
Pt40%Rh
0 to +1888
Hoskins
Hoskins
IEC 584.1
Hoskins
IEC 584.1
IEC 584.1
DIN43710:1985 (To IPTS68)
IEC 584.1
IEC 584.1
IEC 584.1
IEC 584.1
DIN43710:1985
ASTM E1751-95
Ipsen
Engelhard
ASTM E1751-95
Table 3 Thermocouple types and ranges
Max linearisation error
0 to 400°C = 1.7°C
400 to 1820°C = 0.03°C
0.12°C
0.08°C
0.03°C
0.07°C
0.02°C
0.04°C
0.02°C
0.04°C
0.04°C
0.04°C
0.02°C
0.08°C
0.06°C
0.14°C
0.02°C
0.07°C
RTD Type
Cu10
Cu53
JPT100
Ni100
Ni120
Pt100
Pt100A
Pt1000
Overall range (°C)
–20 to +400
–70 to ± 200
–220 to +630
–60 to +250
–50 to +170
–200 to +850
–200 to +600
–200 to +850
Standard
General Electric Co.
RC21-4-1966
JIS C1604:1989
DIN43760:1987
DIN43760:1987
IEC 751
Eurotherm Recorders SA
IEC 751
Table 4 RTD types and ranges
Max linearisation error
0.02 °C
<0.01°C
0.01 °C
0.01 °C
0.01 °C
0.01 °C
0.09 °C
0.01 °C
Relay Output Board
General
Maximum number of relay boards
Number of relays per board
Estimated mechanical life
Update rate
9 (max no of relay outputs = 36)
3 per C/O
4 per N/C
4 per N/O
30,000,000 operations
See ‘Update rates’ in ‘Recorder
Specification’ above
AC load ratings
Derating
The figures give below are for restive loads. For reactive or inductive loads, de-rate in accordance with Graph 1, in which
F1 = Actually measured results on representative samples
F2 = Typical values according to experience
Contact life = Resistive contact life x reduction factor
Maximum switching power
Maximum contact voltage
Maximum contact current
500VA
250V providing this does not cause the maximum switching power (above) to be exceeded
2 Amps providing this does not cause the maximum switching power (above) to be exceeded
DC load ratings
Maximum switching power
Maximum contact voltage/Current
See Graph 2 for operating volt/amp envelope
See Graph 2 for examples
F
1
0.9
0.8
0.7
0.6
0.5
0.4
Safety isolation
Isolation (dc to 65Hz; BS EN61010) Installation category II; Pollution degree 2
Relay to Relay: 300V RMS or dc (double insulation)
Relay to ground: 300V RMS or dc (basic insulation)
F1
F2
Max. DC load breaking capacity
300
200 resistive load
100
0.3
1 0.8
0.6
0.4
Power factor (cos
φ)
0.2
Graph 1 Derating curves for ac loads
50
40
30
20 inductive load (L/R
= 20msec)
10
0.1 0.2
0.5
1 2 5
DC current (Amps)
2 0
Graph 2 DC load switching curves
Mechanical installation
Front view
292mm
Side Elevation
260.90mm
211mm
Safety
Earth
Access Flap x
Panel cutout
281mm x 281mm
(+1 - 0) or
11.07 x 11.07 inches
) y
Panel thickness = 6 to 25mm
Optimum thickness depends on panel material
Minimum recommended inter-unit spacing
Side clamps Top/bottom clamps x = 25mm (1 inch) y = 12.5 mm (0.5 inch) x = 12.5mm (0.5 inch) y = 25mm (1 inch)
View on underside
Maximum Installed Angle a = b = 45 degrees max
Panel mounting technique
Rear terminal connections
Serial communication port pinout
6
7
4
5
8
9
Pin
1
2
3
EIA232
Not connected
Rx
Tx
DTR
Not Connected
Not Connected
Not Connected
USB 1
1 2 3 4 5
EIA485 (5-wire)
RxA
Not connected
Not connected
Not connected
Signal ground
RxB
TxA
TxB
5.0V via 1500
Ω
EIA485 (3-wire)
Link to pin 7
Not connected
Not connected
Not connected
Link to pin 8
TxA/RxB
TxB/RxA
USB 2
Ethernet connector
(option)
1 2 3 4 5
6 7 8 9
Serial 1
6 7 8 9
Serial 2
Option boards
Input channels
1
1
1
1
1
3
37 - 42
25 - 30
13 - 18
1 - 6
2
4
1
22
1
22
1
22
1
22
5
7
43 - 48
31 - 36
19 - 24
7 - 12
6
8
9
Option boards
22
22
22
22
Input channels a˚ b˚
Input board wiring
Channels
1 2 3 4
1,7,13,19,
25,31,37,43
5 6 7
2,8,14,20,
26,32,38,44
V+ V- I
8 9 10
3,9,15,21,
27,33,39,45
V+ V- I
11 12 13 14 15 16 17 18 19 20 21 22
4,10,16,22,
28,34,40,46
5,11,17,23,
29,35,41,47
6,12,18,24,
30,36,42,48
CJC V+ VI V+ V- I V+ V- I
Relay board wiring
2 Option boards
Relay
Three changeover relays
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 nc c no nc c no nc c no
1 2 3 nc c nc c no nc c no nc c no
1 2 3 no
Maximum number of boards = 9
Input board signal wiring
V+ VI V+ VI
Shunt
V+ VI V+ VI
+ –
–20 to +20V dc
Thermocouples dc mV
+ –
DC millamps
+ –
Attenuator
Assembly
–200 to + 200V dc
RTD
3-wire resistance thermometer
Shunt assembly
100
Ω = LA246779UK10
250 Ω = LA246779UK25
Attenuator assembly
LA244180
V+ V-
RTD
I
2-wire resistance thermometer
V+ VI I
Potentiometer
Potentiometer
V+ VI nc com no
Minimum contact = 60 ms
Contact closure
(Not Chs 1,7,13 etc.)
Ordering code
0 1 2
6180A
14 15 16
3
17
4 5 6 7 8 9 10 11 12 13
18 19 20 21 22 23 24 25 26 27
28 29 30 31 32 33 34
0 Model
6180A 180mm TFT XGA Display . . . . . . . . . . . . . . . . . . . . . . . . . .6180A
1 Number of channels
6 Input Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .U06
12 Input Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .U12
18 Input Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .U18
24 Input Channels (6180A only) . . . . . . . . . . . . . . . . . . . . . . . . . . . .U24
30 Input Channels (6180A only) . . . . . . . . . . . . . . . . . . . . . . . . . . . .U30
36 Input Channels (6180A only) . . . . . . . . . . . . . . . . . . . . . . . . . . . .U36
42 Input Channels (6180A only) . . . . . . . . . . . . . . . . . . . . . . . . . . . .U42
48 Input Channels (6180A only) . . . . . . . . . . . . . . . . . . . . . . . . . . . .U48
2 Industry variant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .AERODAQ
Panel mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .PANEL
Carry handle (Bezel colour silver) . . . . . . . . . . . . . . . . . . . . . . . . . . .CH
4 Lock
Media lock not fitted . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .NOLCK
Electronic lock fitted . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .LOCK
5 Bezel colour
Silver including portable options . . . . . . . . . . . . . . . . . . . . . . . . . . .SLV
Eurotherm Green . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .GRN
Black . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .BLK
6 Power supply
90- 264 Vac (110-370Vdc) 47 –63 Hz . . . . . . . . . . . . . . . . . . . . . . . .VH
20 – 42 Vac (20 – 54Vdc) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .VL
7 Not used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .XXXXX
8 Non standard
Non Standard Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .XXXXXX
9 Internal memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
32M Byte for history – approximately 4 million samples . . . . . . .032M
96M Byte for history – approximately 12 million samples . . . . . .096M
10 Removable media
Compact Flash and Front USB port . . . . . . . . . . . . . . . . . . . . . . . . . .CF
Secure Digital and Front USB port . . . . . . . . . . . . . . . . . . . . . . . . . . .SD
11 Memory card size (CF or SD, dependant on removable media type)
Not fitted . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .NOMC
128M byte Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .128M
256M byte Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .256M
512M byte Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .512M
1Gbyte Card (CF or SD, dependant on removable media type) . .001G
12 USB Memory stick size
Not fitted . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .NOMS
64M byte USB Memory Stick . . . . . . . . . . . . . . . . . . . . . . . . . .064MMS
256M byte USB Memory Stick . . . . . . . . . . . . . . . . . . . . . . . . .256MMS
512M byte USB Memory Stick . . . . . . . . . . . . . . . . . . . . . . . . .512MMS
1G byte USB Memory Stick . . . . . . . . . . . . . . . . . . . . . . . . . . . .001GMS
13 Rear USB
No rear USB ports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0RUSB
Two USB ports at rear of product . . . . . . . . . . . . . . . . . . . . . . . .2RUSB
14 Serial communication ports
Not fitted . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0SRL
Two EIA 232/422/485 Serial Ports . . . . . . . . . . . . . . . . . . . . . . . . . .2SRL
15 Not used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .XXXX
16 Calibration certificates
Not required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .NOCAL
Calibration certificate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .CAL
17 Changeover relays
Not fitted . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .00
3 changeover relays (1 option board) . . . . . . . . . . . . . . . . . . . . . . . .03
6 changeover relays (2 option boards) . . . . . . . . . . . . . . . . . . . . . . . .06
9 changeover relays (3 option boards) . . . . . . . . . . . . . . . . . . . . . . . .09
12 changeover relays (4 option boards) . . . . . . . . . . . . . . . . . . . . . . .12
15 changeover relays (5 option boards, 6180A only) . . . . . . . . . . . .15
18 changeover relays (6 option boards, 6180A only) . . . . . . . . . . . .18
21 changeover relays (7 option boards, 6180A only) . . . . . . . . . . . .21
24 changeover relays (8 option boards, 6180A only) . . . . . . . . . . . .24
27 changeover relays (9 option boards, 6180A only) . . . . . . . . . . . .27
18 Normally closed relays
Not fitted . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .00
4 Normally Closed relays (1 option board) . . . . . . . . . . . . . . . . . . . .04
8 Normally Closed relays (2 option boards) . . . . . . . . . . . . . . . . . . . .08
12 Normally Closed relays (3 option boards) . . . . . . . . . . . . . . . . . . .12
16 Normally Closed relays (4 option boards) . . . . . . . . . . . . . . . . . . .16
20 Normally Closed relays (5 option boards, 6180A only) . . . . . . . .20
24 Normally Closed relays (6 option boards, 6180A only) . . . . . . . .24
28 Normally Closed relays (7 option boards, 6180A only) . . . . . . . .28
32 Normally Closed relays (8 option boards, 6180A only) . . . . . . . .32
36 Normally Closed relays (9 option boards, 6180A only) . . . . . . . .36
19 Normally open relays
Not fitted . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .00
4 Normally Open relays (1 option board) . . . . . . . . . . . . . . . . . . . . .04
8 Normally Open relays (2 option boards) . . . . . . . . . . . . . . . . . . . . .08
12 Normally Open relays (3 option boards) . . . . . . . . . . . . . . . . . . . .12
16 Normally Open relays (4 option boards) . . . . . . . . . . . . . . . . . . . .16
20 Normally Open relays (5 option boards, 6180A only) . . . . . . . . .20
24 Normally Open relays (6 option boards, 6180A only) . . . . . . . . .24
28 Normally Open relays (7 option boards, 6180A only) . . . . . . . . .28
32 Normally Open relays (8 option boards, 6180A only) . . . . . . . . .32
36 Normally Open relays (9 option boards, 6180A only) . . . . . . . . .36
20 Event inputs
Not fitted . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .00
06 Event Inputs (1 board) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .06
12 Event Inputs (2 boards) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
18 Event Inputs (3 boards) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18
24 Event Inputs (4 boards) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
21 Analogue outputs
None . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .00
2 Analogue Outputs (1 option board) . . . . . . . . . . . . . . . . . . . . . . . .02
4 Analogue Outputs (2 option boards) . . . . . . . . . . . . . . . . . . . . . . .04
6 Analogue Outputs (3 option boards) . . . . . . . . . . . . . . . . . . . . . . .06
8 Analogue Outputs (4 option boards) . . . . . . . . . . . . . . . . . . . . . . .08
22 Quantity of shunts
Enter quantity required
23 Shunt value
Not required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .NOS
100 ohm shunts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .100
250 ohm shunts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .250
24 Quantity of attenuators (100:1)
Enter quantity required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ._ _
25 Warranty
Standard warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .XXXXX
Extended warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .WL005
26 Bridge
Bridge Lite (supplied as standard) . . . . . . . . . . . . . . . . . . . . . . . . .BLITE
Bridge Full . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .BFULL
27 Review and quickchart
Review and Quickchart Lite (supplied as standard) . . . . . . . . . . .RLITE
Review and Quickchart Full . . . . . . . . . . . . . . . . . . . . . . . . . . . . .RFULL
28 Auditor
Not required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .NOADT
Audit Trail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ALITE
Auditor Full . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .AFULL
29 Security manager
Not required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .NOSM
Security Manager . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .SECMAN
30 Groups
6 Groups (supplied as standard) . . . . . . . . . . . . . . . . . . . . . .06GROUP
12 Groups . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12GROUP
31 Maths, totalisers and counters
Not required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .MTC00
36 Virtual Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .MTC36
96 Virtual Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .MTC96
128 Virtual Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .MTC128
32 Batch
Not required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .NOBTCH
Batch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .BATCH
33 Screen builder
Not required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .NOSB
Advanced screen builder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ADSB
34 Master communications
Not required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .NOMSTR
Modbus Master Comms for 16 slaves . . . . . . . . . . . . . . . . . . . .MSTR16
Modbus Master Comms for 32 slaves . . . . . . . . . . . . . . . . . .MSTR3223
Standard AeroDAQ Ordering code
6180A/U06/AERODAQ/NOLCK/SLV/VH/XXXXX/XXXXX/032M/CF/NOMC/NOMS/0RUSB/2SRL/XXXX/NOCAL/03/
00/00/00/00/00/NOS/00/XXXXX/BFULL/RLITE/ALITE/NGSM/06GROUP/MTC36/BATCH/ADSB/MSTR16
0 Model 6180A 180mm TFT XGA Display . . . . . . . . . . . . . . . . . . . . . . . . .6180A
1 Number of channels 6 Input Channels . . . . . . . . . . . . . . . . . . . . . . . . . .U06
2 Industry variant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .AERODAQ
4 Lock Media lock not fitted . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .NOLCK
5 Bezel colour Silver including portable options . . . . . . . . . . . . . . . . . . . . .SLV
6 Power supply 90- 264 Vac (110-370Vdc) 47 –63 Hz . . . . . . . . . . . . . . . . .VH
7 Not used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .XXXXX
8 Non standard Non Standard Option . . . . . . . . . . . . . . . . . . . . . . . .XXXXXX
9 Internal memory 32M Byte for history – approx 4 million samples . . .032M
10 Removable media Compact Flash and Front USB port . . . . . . . . . . . . . . .CF
11 Memory card size Not fitted . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .NOMC
12 USB Memory stick size Not fitted . . . . . . . . . . . . . . . . . . . . . . . . . . . .NOMS
13 Rear USB No rear USB ports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0RUSB
14 Serial communication ports Two EIA 232/422/485 Serial Ports . . . . . . .2SRL
15 Not used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .XXXX
16 Calibration certificates Not required . . . . . . . . . . . . . . . . . . . . . . . . .NOCAL
17 Changeover relays 3 changeover relays (1 option board) . . . . . . . . . . . . .03
18 Normally closed relays Not fitted . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .00
19 Normally open relays Not fitted . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .00
20 Event inputs Not fitted . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .00
21 Analogue outputs None . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .00
22 Quantity of shunts Enter quantity required . . . . . . . . . . . . . . . . . . . . . . ._ _
23 Shunt value Not required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .NOS
24 Quantity of attenuators (100:1) Enter quantity required . . . . . . . . . . . ._ _
25 Warranty Standard warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .XXXXX
26 Bridge Bridge Full . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .BFULL
27 Review and quickchart Lite (supplied as standard) . . . . . . . . . . . . . . .RLITE
28 Auditor Audit Trail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ALITE
29 Security manager Not required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .NOSM
30 Groups 6 Groups (supplied as standard) . . . . . . . . . . . . . . . . . . . . .06GROUP
31 Maths, totalisers and counters 36 Virtual Channels . . . . . . . . . . . . .MTC36
32 Batch Batch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .BATCH
33 Screen builder Advanced screen builder . . . . . . . . . . . . . . . . . . . . . . . .ADSB
34 Master communications Modbus Master Comm 16 slaves . . . . . . .MSTR16
Eurotherm: International sales and service
Understanding and providing local support is a key part of Eurotherm's business. Complementing worldwide Eurotherm offices are a whole range of partners and a comprehensive technical support team… to ensure you get a service you will want to go back to.
AUSTRALIA Sydney
Eurotherm Pty. Ltd.
T (+61 2) 9838 0099
F (+61 2) 9838 9288
AUSTRIA Vienna
Eurotherm GmbH
T (+43 1) 7987601
F (+43 1) 7987605
BELGIUM & LUXEMBOURG Moha
Eurotherm S.A/N.V.
T (+32) 85 274080
F (+32) 85 274081
BRAZIL Campinas-SP
Eurotherm Ltda.
T (+5519) 3707 5333
F (+5519) 3707 5345
DENMARK Copenhagen
Eurotherm Danmark AS
T (+45 70) 234670
F (+45 70) 234660
FINLAND Abo
Eurotherm Finland
T (+358) 22506030
F (+358) 22503201
FRANCE Lyon
Eurotherm Automation SA
T (+33 478) 664500
F (+33 478) 352490
E info.[email protected]
GERMANY Limburg
Eurotherm Deutschland GmbH
T (+49 6431) 2980
F (+49 6431) 298119
HONG KONG & CHINA
Eurotherm Limited North Point
T (+85 2) 28733826
F (+85 2) 28700148
Guangzhou Office
T (+86 20) 8755 5099
F (+86 20) 8755 5831
Beijing Office
T (+86 10) 6567 8506
F (+86 10) 6567 8509
Shanghai Office
T (+86 21) 6145 1188
F (+86 21) 6145 1187
INDIA Chennai
Eurotherm India Limited
T (+91 44) 24961129
F (+91 44) 24961831
IRELAND Dublin
Eurotherm Ireland Limited
T (+353 1) 4691800
F (+353 1) 4691300
ITALY Como
Eurotherm S.r.l
T (+39 31) 975111
F (+39 31) 977512
KOREA Seoul
Eurotherm Korea Limited
T (+82 31) 2738507
F (+82 31) 2738508
NETHERLANDS Alphen a/d Rijn
Eurotherm B.V.
T (+31 172) 411752
F (+31 172) 417260
NORWAY Oslo
Eurotherm A/S
T (+47 67) 592170
F (+47 67) 118301
POLAND Katowice
Invensys Eurotherm Sp z o.o.
T (+48 32) 2185100
F (+48 32) 2177171
SPAIN Madrid
Eurotherm España SA
T (+34 91) 6616001
F (+34 91) 6619093
SWEDEN Malmo
Eurotherm AB
T (+46 40) 384500
F (+46 40) 384545
SWITZERLAND Wollerau
Eurotherm Produkte (Schweiz) AG
T (+41 44) 7871040
F (+41 44) 7871044
UNITED KINGDOM Worthing
Eurotherm Limited
T (+44 1903) 268500
F (+44 1903) 265982
E [email protected] www.eurotherm.co.uk
U.S.A. Leesburg VA
Eurotherm Inc.
T (+1 703) 443 0000
F (+1 703) 669 1300
E [email protected] www.eurotherm.com
ED53
© Copyright Eurotherm Limited 2007
Invensys, Eurotherm, the Eurotherm logo, Chessell, EurothermSuite, Mini8, Eycon, Eyris and Wonderware are trademarks of Invensys plc, its subsidiaries and affiliates. All other brands may be trademarks of their respective owners.
All rights are strictly reserved. No part of this document may be reproduced, modified, or transmitted in any form by any means, nor may it be stored in a retrieval system other than for the purpose to act as an aid in operating the equipment to which the document relates, without the prior written permission of Eurotherm limited.
Eurotherm Limited pursues a policy of continuous development and product improvement. The specifications in this document may therefore be changed without notice. The information in this document is given in good faith, but is intended for guidance only.
Eurotherm Limited will accept no responsibility for any losses arising from errors in this document.
Part No. HA029538 Issue 1 6180AeroDAQ Specification Sheet Printed in England 06.07
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement