- No category
advertisement
▼
Scroll to page 2
of 145
Instructions for Use LS2101 Rev C 1 Contents English (EN) 3 German (DE) 18 33 Italian (IT) 49 Spanish (ES) 65 Dutch (NL) 81 Swedish (SV) 97 Danish (DA) 112 Finnish (FI) 127 2 EN EN Instructions for Use Only medical professionals who have experience in performing procedures within the uterine cavity, such as IUD insertions or dilation and curettage, are suitably trained, have adequate knowledge and familiarity with the Thermablate EAS system, should perform endometrial ablation using this device. Read all directions, cautions, and warnings prior to use. These Instructions for Use provide directions for using the Thermablate EAS system. Failure to follow any instructions or to heed any warnings or precautions could result in serious injury to the patient. These Instructions for Use are complemented and reiterated by the “Operator’s Manual” included with the Treatment Control Unit (TCU) kit. Use caution not to perforate the uterine wall during dilation, sounding, or cutterage (if performed). If a perforation is suspected, a hysteroscopy must be performed. Do not initiate the procedure if perforation of the uterine wall or the creation of a false passage is confirmed during the hysteroscopy performed just prior to the insertion of the balloon. Do not perform same day Thermablate EAS procedure and hysteroscopic tubal occlusion/sterilization. Thermablate EAS procedure can be safely and effectively performed with nickel titanium inserts in place, however the procedure should only be performed after the 3 month tubal occlusion confirmation test. The outside temperature measured on the disposable cartridge sheath can reach a maximum of 80°C during treatment cycle. It has been confirmed from clinical studies and historical data, that the heat lacks sufficient energy to cause any heat transfer to surrounding area. To eliminate any potential for patient discomfort in some circumstances Medical Equipment should not to be placed against exposed flesh during treatment 3 PATIENT SELECTION Excessive uterine bleeding can be caused by a variety of underlying problems including but not limited to endometrial cancer, myomas, polyps, and drugs. Patients should always be evaluated to determine if there are underlying causes of their excessive uterine bleeding before any treatment option is initiated. The patient selection criteria are: • • • • • • Documented diagnosis of excessive uterine bleeding with no underlying causes Completed childbearing Pre-menopausal Normal uterine cavity with sounding between 8 cm and 12 cm, inclusive Normal Pap Smear and endometrial biopsy Does not present any of the contraindications below. CONTRAINDICATIONS Thermablate EAS is contraindicated for use in: • A patient with a uterine sounding less than 8 cm or in excess of 12cm (external os to fundus) • A patient with active pelvic inflammatory disease • A patient with known or suspected endometrial carcinoma (uterine cancer) or pre-malignant change of the endometrium, such as unresolved complex (adenomatous) hyperplasia • A patient with history of pelvic malignancy within the past 5 years. • A patient with submucous / intramural myomas greater than 3.0 cm such that the uterine cavity is significantly distorted • A patient with intracavitary lesions (Type 0 or 1 myoma or polyps of any size) as documented by hysteroscopy, saline infusion sonohysterogram (SIS), or MRI performed in the last six months. Ablation can be performed if polyp is removed prior to procedure. • A patient with a septate uterus • • • • • • A patient with an IUD currently in place A patient with any anatomic or pathologic condition in which weakness of the myometrium could exist, such as history of previous classical caesarean section or transmural myomectomy A patient that has had three or more, lower segment C sections and where the linear scar thickness in those patients is less than 8mm. A patient who is pregnant or who wants to become pregnant in the future A patient with active genital or urinary tract infection at the time of procedure (e.g. cervicitis, vaginitis, endometritis, salpingitis, or cystitis) A patient less than 6 months post-partum. 4 REPEAT BALLOON ABLATIONS ARE CONTRAINDICATED As the endometrial cavity following any kind of endometrial ablation is most likely distorted, repeat ablations should not be attempted with Thermablate EAS. Patients requiring further treatment after thermal balloon ablation should be treated medically, by resectoscopic endometrial ablation, or by hysterectomy. Repeat resectoscopic ablation resection should be attempted only by experienced hysteroscopists since the complications may be severe. POTENTIAL ADVERSE EVENTS The following adverse events are potentially associated with endometrial ablation: • Pelvic cramping • Nausea and vomiting • Perforation of the uterus • Rupture of the uterus • Thermal injury to adjacent tissues/organs • Heated liquid escaping into the cervix, vagina, or fallopian tubes • Infection • Post-ablation tubal sterilization syndrome (PATSS) • Hematometra. RECOMMENDED PRE-TREATMENT OF ENDOMETRIUM • Oral Contraceptives (Recommended for minimum 21 days until day of treatment) • NuvaRing® • Timing of the Menstrual Cycle • Suction Curettage (If necessary) • GnRH analogues recommended only for patients with a uterine sounding ≥ 10 cm. PATIENT COUNSELLING As with any procedure, the physician needs to discuss the risks, benefits, and alternatives with the patient prior to performing endometrial ablation. The device is intended for use in women who do not desire to bear children because the likelihood of pregnancy is significantly decreased following this procedure. Patients of childbearing capability should be counselled that endometrial ablation is not a sterilization procedure and should be provided with an appropriate birth control method. These patients should be cautioned of the potential complications that may ensue if they should become pregnant. Vaginal discharge is typically experienced during the first few days following ablation and may last as long as a few weeks. Generally, the discharge is described as: bloody during the first few days; serosanguineous by approximately one week; then profuse and watery thereafter. Excessive pain, heavy bleeding, foul smelling discharge or fever should be reported to the physician. WARNING Endometrial ablation using the Thermablate EAS system is not a sterilization procedure. Pregnancies after ablation can be dangerous for both mother and foetus. 5 Directions for Use SET-UP 1. Check that the following items are present with the Thermablate EAS device: • One (1) Sterile Disposable (Single-Use) Cartridge • One (1) Treatment Control Unit (TCU) • One (1) Universal-Input Power Supply • One (1) Power Cord (to match local AC power receptacles) • One (1) TCU Stand. Frequently Used Functions Operator should make themselves aware of the following functions: On / Off Power switch located on power cord into TCU Emergence Stop Button located on LCD overlay on top of TCU. Blue Treatment Trigger Switch located on front of the TCU handle. 2. Place TCU in Stand. 3. Ensure the “POWER SWITCH” is in the “OFF” position. 4. Connect the Power Supply to the TCU. Plug the Power Cord into both the Power Supply and the wall power outlet. Ensure that all the connections are firmly connected. Ensure that the “POWER SWITCH” and mains plug can be easily accessed to disconnect power to the TCU. 5. Ensure that the device is positioned to allow ease of access to the patient during treatment avoiding any hinderances, stretching or obstructions. Also ensure that no cables or accessories are placed in a dangerous manner or position. OPENING AND INSTALLING THE THERMABLATE EAS STERILE DISPOSABLE CARTRIDGE 1. The Disposable Cartridge is inside a pouch (Pouch made of clear film and Tyvek webs). The pouch is inside a tube with lids. The tube is inside a foil vacuum bag designed to be peeled open. ONLY THE CONTENTS OF THE FOIL BAG ARE DEEMED TO BE IN A STERILE STATE. The foil vacuum bag is inside a cardboard box. 2. Check expiry date. DO NOT USE the Disposable Cartridge if it has expired. 3. Open cardboard box. 4. Remove foil vacuum bag from cardboard box. 5. Ensure vacuum is intact. If it is not, DO NOT USE the Disposable Cartridge. get another Disposable Cartridge instead. 6. Peel open foil vacuum bag. 7. Remove tube from the foil vacuum bag. Do not discard plastic tube but place aside for use in safe disposal of disposable cartridge after use. 8. Remove the lid with label attached from the tube by tearing the blue cap where indictaed by the symbol. 6 9. Remove pouch containing the disposable cartridge.Peel open pouch to remove cartridge using standard aseptic technique. 10. Install Disposable Cartridge in the TCU, by lining up the 2 pins on the cartridge with the 2 slots in the TCU, and rotating it clockwise, as shown in the figure below. Ensure that “THIS SIDE UP” and the guidance markings are visible on the top side of the catheter. Leave the plastic balloon cover on. WARM-UP 1. Turn the “POWER SWITCH” to the “ON” position; the LCD will show the following message: If no display is present then do not continue with procedure. WELCOME TO THERMABLATE Both LED’s should be OFF (see below right-bottom corner of LCD). (If the number of treatments remaining before servicing is less than 50, the TCU will issue a beeping sound and the LCD will show the following message for 10 seconds: X TREATMENT(S) BEFORE SERVICING where “X” is number of remaining uses of the TCU.) 2. The unit automatically performs a self-test, and, if no technical problems are encountered, the initial message is replaced by the following message: Device OK (If the TCU encounters a technical problem, an error number will be displayed on the LCD. Please refer to the Operator’s Manual for troubleshooting.) 3. The following message is then displayed: HEATING – Wait… Fluid Temp XXX °C where “XXX” is the measured temperature of the treatment liquid. The “Heating” LED will turn ON, indicating that the pre-heating of the 7 treatment liquid has begun. The “Ready” LED should be OFF. Before entering the “READY FOR TREATMENT” state the TCU performs a leak test. If this test is not passed (“ERROR #19” will appear on LCD), turn the “POWER SWITCH” to the “OFF” position, and verify that: A) The Disposable Cartridge has been installed properly into the TCU, and B) No fluid is leaking out from the Disposable Cartridge, particularly from the balloon. In case of a leaking Disposable Cartridge that has not spilled treatment fluid into the TCU, replace with a new one. USE EXTRA CAUTION, AS THE METALLIC PART OF THE DISPOSABLE CARTRIDGE MAY BE VERY HOT. Turn the “POWER SWITCH” back to the “ON” position.Wait until the TCU reaches the “READY FOR TREATMENT” state. If treatment fluid has leaked into the TCU, it will need to be repaired. PATIENT PREPARATION 1. Provide patient with adequate analgesia. 2. During and/or prior to the pre-heating of the treatment fluid, patient preparation can be performed. Appropriate sterile technique for vaginal/cervical preparation should be used. 3. Place patient in dorsal lithotomy position. TREATMENT After approximately 12 minutes, the LCD will show ”READY FOR TREATMENT” (see below) and the TCU will beep, indicating that the fluid has been heated up to treatment temperature (approximately 173°C) and the treatment procedure can be initiated. The “Heating” LED turns OFF and the “Ready” LED turns ON. The LCD Display reads: READY FOR TREATMENT NOTICE If the system is left unused, the treatment temperature will be maintained for 35 minutes. After this period, the system will automatically turn OFF. To reinitiate Warm-Up, turn the "POWER SWITCH” to the “OFF” position, and restart the process. The Disposable Cartridge must not be used if it has been heated up and cooled down more than twice or exposed to ambient air for more than 2 hours. WARNING IF THE BALLOON IS NOT COMPLETELY DEFLATED AND LIQUID AND / OR GAS IS SEEN IN BALLOON DURING “READY FOR TREATMENT” STATE, DO NOT PROCEED WITH THE TREATMENT. Instead, turn the “POWER SWITCH” to the “OFF” position, disconnect the power cord, substitute the Disposable Cartridge, and restart the process. 8 WARNING IF THE BALLOON COVER CANNOT BE EASILY REMOVED FROM THE BALLOON, DUE TO PRESSURE IN THE BALLOON, DO NOT DETACH CARTRIDGE FROM TCU. Instead, turn the “POWER SWITCH” to the “OFF” position, disconnect the power cord and cool down both TCU and Cartridge together until Balloon Cover can be easily removed. Remove Cartridge carefully from TCU, verify no fluid leaking out to the TCU, substitute the Disposable Cartridge, and restart the process. CAUTION Patients with either an acutely anteverted or retroverted uterus, or a fixed uterus (e.g. due to significant endometriosis or adhesions), or those that have had previous uterine surgery are at a higher risk. Particular attention should be paid to the angulation of the uterine sound, cervical dilator and Thermablate catheter during insertion. 1. Conduct pelvic examination to confirm position of uterus. 2. Insert Speculum. 3. Apply tenaculum. 4. Measure sounding length of uterus from external os to fundus using uterine sound. Confirm that measurement is between eight (8) and twelve (12) cm. 5. Use dilators to gradually dilate cervix up to seven (7) mm. Dilators should pass easily through cervix with minimal discomfort to the patient. Dilators should not be advanced deeper than the predetermined uterine depth. 6. Measure length of uterus a second time using the uterine sound. Confirm that sounding length of the uterus after dilation is the same as sounding length obtained prior to dilation. If there is a discrepancy of more than 0.5cm between the first and second measurements a false passage or perforation of the uterus may have been created during the dilation. 7. Perform hysteroscopy prior to balloon insertion to ensure that the uterus has not been perforated or that a false passage has not been created during dilatation, sounding or curettage (if performed). CAUTION A PERFORATION OF THE UTERUS OR CREATION OF A FALSE PASSAGE, IF UNDETECTED, CAN LEAD TO THERMAL INJURIES OF ADJACENT ORGANS OR TISSUE Hysteroscopy should reveal both tubal ostia clearly before proceeding with the treatment. If distension of uterus during hysteroscopy cannot be maintained, it is possible that the uterus has been perforated and treatment should not proceed. Should the hysteroscopy reveal an excessively thick endometrial lining, a gentle suction curettage of the uterus may be performed. 9 A second hysteroscopy should be performed immediately following curettage to ensure that the curettage has not created a perforation of the uterus. 8. Alternatively, use ultrasonic surveillance during the treatment to check for correct balloon position inside the uterine cavity. 9. Slide the balloon cover off the balloon. DO NOT DISPOSE OF BALLOON COVER AS IT IS REQUIRED FOR LATER USE. Remove the Thermablate EAS system from the TCU Stand. 10. Slowly insert the Thermablate balloon until balloon tip touches the fundus. Tap the tip of catheter gently against the fundus to confirm placement of the catheter within the uterus. 11. Ensure that the depth marking on the balloon catheter matches the previously obtained sounding measurements. Should there be a discrepancy of more than 0.5 cm between the sounding measurements obtained and depth marking on the catheter, a repeat hysteroscopy should be performed. 12. Activate the treatment cycle by holding Blue Treatment Trigger Switch for 5 seconds. After hearing five (5) short and one (1) long beeping sounds, the treatment will automatically begin. The Finger can be removed from the trigger switch at this point. The LCD will show the following message: Performing System Check Both LEDs will turn OFF. 13. After 15 seconds, if the TCU passes the system check, the actual treatment cycle will start, and the LCD will show the following message: TREATING… Starting… Followed shortly after by: Pressure: XYYY Time Left: Z:ZZ where “X” is the sign (+ or -) for positive or negative pressure. “YYY” is the actual pressure value reached during procedure (mmHg). “Z:ZZ” is the treatment time remaining (min:sec). If the system check fails, the TCU will issue an alarm and stop operating. In this case, turn the “POWER SWITCH” to the “OFF” position, d i s c o n n e c t t h e p o w e r c o r d , a n d remove the Disposable Cartridge from the patient, and verify that: 10 A) The Disposable Cartridge has been installed properly into the TCU, and B) No fluid is leaking out from the Disposable Cartridge, particularly from the balloon. In case of a leaking Disposable Cartridge that has not spilled treatment fluid into the TCU, replace with a new one. USE EXTRA CAUTION, AS THE METALLIC PART OF THE DISPOSABLE CARTRIDGE MAY BE VERY HOT. Turn the “POWER SWITCH” back to the “ON” position and follow the instructions until reaching the system check. If treatment fluid has leaked into the TCU, it will need to be repaired. 14. As the balloon deploys, it may push the catheter slightly backwards (Up to 0.5 cm is normal). Do not push the catheter forward during treatment. CAUTION At no time during the treatment should the catheter advance beyond the pre-determined sounding length. Should this occur, abort the procedure by activating the “Emergency Stop”button, observe the “Treatment Stop” message, while the fluid is actively withdrawn from the balloon and wait for the message: “Finished V: XX ml Withdraw Balloon” to appear on the LCD screen and slowly remove the Thermablate catheter from the uterus. Perform a hysteroscopy to ensure that the uterus has not been perforated. 15. Observe the LCD screen on TCU as it automatically performs system checks and completes the treatment cycle. Automatic operation of the System inflates the balloon, controls the pressure, and pulses the treatment liquid, maintaining uniform temperature in the balloon. During this time, the pump is clearly audible. This is not a malfunction – it is part of the normal operation of the unit. CAUTION IN THE EVENT OF POWER LOSS DURING TREATMENT If the power is lost during treatment, wait for thirty (30) seconds. If, after this 30-second period, the power has not come on again, withdraw the balloon fast but carefully since some liquid may still be contained in the balloon. Turn the “POWER SWITCH” to the “OFF” position, and then go to Section “Post-Treatment” below. If the power has come on again within this 30-second period, DO NOT WITHDRAW THE BALLOON. The TCU will automatically recognize that the “previous” treatment did not finish correctly and will proceed to actively withdraw the liquid from the balloon. REMOVE THE BALLOON FROM THE UTERUS ONLY AFTER THE “FINISHED V: XX ml Withdraw Balloon” MESSAGE APPEARS ON THE LCD SCREEN. 11 EMERGENCY INTERUPTION / TERMINATION OF TREATMENT IF EMERGENCY INTERRUPTION / TERMINATION OF TREATMENT IS REQUIRED OR IF LIQUID IS NOTED LEAKING THROUGH THE CERVIX, DO NOT IMMEDIATELY REMOVE THE BALLOON FROM THE UTERUS. If fitted, first, activate the red “Emergency Stop” button on the LCD overlay which will activate the emergency routine and display “TREATMENT STOP” message on the LCD and proceed to actively withdraw the liquid from the balloon. OR Turn off the power to the TCU. Then proceed to turn power back on to the TCU. In doing so, the TCU will automatically recognise that the “previous” treatment did not finish correctly. The TCU will display the “TREATMENT STOP” message on the LCD and then proceed to actively withdraw the liquid from the balloon. TREATMENT STOP Do NOT Remove REMOVE THE BALLOON FROM THE UTERUS ONLY AFTER “FINISHED V: XX ml Withdraw Balloon” MESSAGE APPEARS ON THE LCD SCREEN. . WARNING If the TCU encounters a technical problem during the procedure, it will issue a “TREATMENT FAILED” message. REMOVE THE BALLOON FROM THE UTERUS ONLY AFTER THE “FINISHED V: XX ml Withdraw Balloon” MESSAGE APPEARS ON THE LCD SCREEN. 16. After completion of the treatment cycle, the LCD will show the following message: FINISHING… Do NOT Remove WARNING If the TCU encounters a technical problem during the deflation procedure, it will issue a “DEFLATION FAILED” message. DO NOT REMOVE THE BALLOON FROM THE UTERUS. Instead, wait until the device withdraws the liquid from the balloon. REMOVE THE BALLOON FROM THE UTERUS ONLY AFTER THE “FINISHED V: XX ml Withdraw Balloon” MESSAGE APPEARS ON THE LCD SCREEN. 17. After 10 seconds of deflation time, the LCD will show the following message FINISHED V: XX ml Withdraw Balloon 12 where “XX” is the estimated uterus volume. This message indicates that the treatment is complete and that the balloon can be withdrawn. Withdraw the balloon carefully from the uterus. After evert treatment the TCU carries out a test on the Filter Quality and if it has deteriorated beyond a set point the following message will be displayed along with an audible indication. Filter Change is Due Right NOW If this message is displayed the TCU MUST be returned for IMMEDIATE Servicing. POST-TREATMENT 1. Place the TCU in the TCU Stand. REPLACE BALLOON COVER ONTO THE BALLOON (To prevent blood splatter during disposal of cartridge). 2. After 50 seconds, the TCU will beep and its LCD will alternate the two following messages, continuously: DISCARD USED CARTRIDGE ***WARNING*** *Metal End *HOT* 3. Turn the “POWER SWITCH” to the “OFF” position. WARNING The Disposable Cartridge IS HOT and should be removed from the TCU and returned into the tube carefully. 4. Remove Disposable Cartridge from TCU. 5. Return the Disposable Cartridge into its original tube by holding the blue connector and inserting ALLUMINIMUM SHIELD END FIRST (so that the hot metal end of the cartridge will be toward the metal reinforced cap) and cover with the capcontaining holes. Allow it to cool, and dispose of it in biohazard-labelled containers or per facility policy. WARNING The Disposable Cartridge is a SINGLE-USE component. Do not reuse it as this could result in serious injuries to the patient and/or the user. 6. Post-treatment hysteroscopy is recommended. 7. Repeat balloon ablations are contraindicated. 13 CLEANING, MAINTENANCE, AND STORAGE CLEANING After each use, the TCU must be cleaned according to the following procedure: 1. D i s c o n n e c t t h e T C U f r o m t h e P o w e r s u p p l y . 2. Wipe down the outside of the TCU housing with warm water using a soft bristle brush until no soilage can be seen.DO NOT SOAK OR IMMERSE. 3. Subsequently, wipe down the outside of the TCU housing with warm water using a clean wipe.DO NOT SOAK OR IMMERSE 4. Disinfect the outside of the TCU housing with a clean wipe and 50%IPA/water solution. DO NOT SOAK OR IMMERSE CAUTION This unit contains electronic components. DO NOT soak, flush, spray or use excessive free liquid on the TCU MAINTENANCE There are no user serviceable parts within the TCU. Opening the device will void the warranty. Return unit to Idoman for service or repair. It is recommended that the TCU and its accompanying power supply are subjected to annual inspection to ensure the safety of the device for both patient and user. This inspection should include the following: • Power connecting cords and power supply for signs of damage or wear. • TCU housing for signs of damage which may leave it in an unsafe or contaminated state. • Presence and legibility of all Safety related markings and Labels. • Copy of Equipment Operators manual present. Testing (If Required) • Protective earth resistance • Equipment leakage current • Patient leakage current • Insulation resistance. • Functional Test of Device WARNING No modification of this equipment is allowed. Unauthorised modifications or access may result in electrical shock or leave the unit in a hazardous condition. All TCU units that have reached their end of service life should be returned to manufacturer for disposable as per WEEE Directive 2002/96/EC STORAGE 1. Store the TCU in the Carrying Case. The storage temperature should never exceed 50°C. 2. Store Disposable Cartridge in dry place at room temperature. 14 SYMBOLS USED ON LABELLING Method of Sterilization – Gamma Irradiation Do not reuse Do not resterilize Read Instruction / Operators Manual Caution – Hot Surface Latex-free Expiry date Classification according to the degree of protection against electric shock: Type BF Date of Manufacture Do Not Use If Package Is Damaged Fragile, Handle with Care Keep Dry Correct Upright Position Humidity Limitation 15 Atmospheric Pressure Limitation On position of power switch Off position of power switch Class II Method of protection against electric shock Product must be disposed of as per WEEE Directive 2002/96/EC Power Supply RoHS Compliant Power Supply Green Energy Level Rating 0120 CE Mark and Identification Number of Notified Body. Product conforms to the essential requirements of the Medical Devices Directive 93/42/EEC. Manufactured by ETL Mark and Control Number. 5000028 16 Idoman Teoranta Killateeaun, Tourmakeady, Co Mayo Ireland Tel No: +353 94 95 44726 Fax No: +353 94 95 44725 E-mail: [email protected] Web Site: www.idoman-med.com 0120 Distributed by: Place Distributor’s Label Here 17 DE DE 18 19 • • • • • • • • 20 2. 3. 4. 5. 6. 7. 21 22 3. B) 23 24 25 B) 26 TREATMENT STOP Do Not Remove ENTFERNEN SIE DEN BALLON AUS DEM UTERUS ERST NACHDEM DIE MELDUNG „FINISHED V: XX ml Withdraw Balloon“ AUF DEM LCD- BILDSCHIRM ERSCHEINT. 27 28 • • • • • • Isolationswiderstand • 29 NOT-AUS Temperaturbegrenzung 30 0120 5000028 31 Idoman Teoranta Killateeaun, Tourmakeady, Co Mayo Ireland Tel No: +353 94 95 44726 Fax No: +353 94 95 44725 E-mail: [email protected] Web Site: www.idoman-med.com 32 FR 33 • • • • 34 35 OUVERTURE ET INSTALLATION DE LA CARTOUCHE JETABLE STERILE THERMABLATE EAS 1. 36 37 38 39 40 41 42 17 43 DISCARD USED CARTRIDGE 44 45 Ouvrir ici 46 0120 47 Idoman Teoranta Killateeaun, Tourmakeady, Co Mayo Ireland Tel No: +353 94 95 44726 Fax No: +353 94 95 44725 Courriel: [email protected] Site Internet: www.idoman-med.com 0120 48 IT 49 • • • • • • • • • • 50 51 52 53 54 55 56 57 58 59 60 61 Aprire Qui ARRESTO DI EMERGENZA 62 0120 63 Idoman Teoranta Killateeaun, Tourmakeady, Co Mayo Ireland Tel No: +353 94 95 44726 Fax No: +353 94 95 44725 E-mail: [email protected] Sito Web: www.idoman-med.com 0120 64 ES 65 • • • • • • • • • • 66 67 68 69 70 71 PRECAUCION LA PERFORACION DEL UTERO O LA CREACION DE UNA LAS VIA, SI NO SE DETECTAN PUEDEN GENERAR LESIONES TERMICAS DE LOS ORGANOS O TEJIDOS AADYCENTES 72 73 74 75 76 • • • • • • • • Testeo functional del dispositivo ADVERTENCIA: 77 Abra aquí APAGADO DE EMERGENCIA Límite de temperatura 78 Límite de humedad 0120 Fabricado por 5000028 79 Idoman Teoranta Killateeaun, Tourmakeady, Co Mayo Ireland Tel No: +353 94 95 44726 Fax No: +353 94 95 44725 Correo electrónico: [email protected] Página Web: www.idomanmed.com 0120 Pegar etiqueta del aqui Distribuido por: 80 NL of 81 • • • • • • • • • • 82 83 84 85 86 87 88 B) 89 VERWIJDER DE BALLON ENKEL UIT DE BAARMOEDER, WANNEER HET BERICHT “FINISHED V: XX ml Withdraw Balloon” VERSCHIJNT OP HET LCD-SCHERM. . 90 ***WARNING*** *Metal End *HOT* 3. Draai de "POWER SWITCH” naar de "UIT"-positie. 91 92 93 de droge op NOODSTOP Temperatuurbegrenzing Juiste rechtop positie 94 Aan positie van schakelaar Stroomtoevoer is ROHS-conform 0120 Vervaardigd door 5000028 95 Idoman Teoranta Killateeaun, Tourmakeady, Co Mayo Ireland Tel No: +353 94 95 44726 Fax No: +353 94 95 44725 E-mail: [email protected] Web Site: www.idomanmed.com 0120 96 SV 97 • • • Pre-menopaus status • • • • • • • • • • • • • • 98 99 finns med 100 101 102 103 104 AVBROTT / AVSLUT AV BEHANDLINGEN OM AVBROTT/AVSLUT AV BEHANDLINGEN ÄR NÖDVÄNDIGT ELLER 105 106 107 108 NÖDSTOPP 109 Produkten måste hanteras enligt WEEE Direktiv 2002/96/EC 0120 5000028 60601,1:08 Medicinsk elektrisk utrustning, del 1 110 Idoman Teoranta Killateeaun, Tourmakeady, Co Mayo Ireland Tel No: +353 94 95 44726 Fax No: +353 94 95 44725 E-mail: [email protected] Web Site: www.idoman-med.com 0120 111 DA 112 113 • • • • • 114 Brugsanvisning OPSÆTNING 1. En (1) strømførende ledning (som passer til lokale AC- stikkontakter) En (1) TCU-holder. 5. 115 116 3 117 118 119 120 121 122 2. 3. 4. Hvis 123 NØDSTOP Opbevares tørt 124 Produktet skal bortskaffes i henhold til WEEE-direktiv 2002/96/EC 0120 125 Idoman Teoranta Killateeaun, Tourmakeady, Co Mayo Ireland Tel No: +353 94 95 44726 Fax No: +353 94 95 44725 E-mail: [email protected] Hjemmeside: www.idoman-med.com 0120 126 FI 127 POTILASVALINTA • • • • • • • • • • • • • 128 129 130 131 132 3. 133 6. 7. 134 135 136 137 138 139 140 141 0120 5000028 142 Idoman Teoranta Killateeaun, Tourmakeady, Co Mayo Irlanti Puhelin: +353 94 95 44726 Faksi: +353 94 95 44725 Sähköposti: [email protected] Verkkosivusto: www.idoman-med.com 0120 143 NOTES 144 For further information please contact Customer Service IDOMAN TEORANTA KILLATEEAUN TOURMAKEADY COUNTY MAYO IRELAND TELEPHONE +353 (0)94 9544726 FAX +353 (0)94 9544725 WWW.IDOMAN-MED.COM For additional enquiries contact: [email protected] Customer Service Canada Tel: + 416 487 8397 Toll Free:1800 768 1836 Toll Free Fax: 1888 763 1356 Thermablate EAS, Endometrial Ablation System Copyright © 2015 by Idoman Teoranta All rights reserved. AND logos are trademark of IDOMAN Teoranta
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Languages
Only pages of the document in English were displayed