LT-10
LT-10-EUR
Users Manual
2/2013, 4265801 A
©2013 Amprobe Test Tools.
All rights reserved. Printed in China
English
Limited Warranty and Limitation of Liability
Your Amprobe product will be free from defects in material and workmanship for one year from
the date of purchase unless local laws require otherwise. This warranty does not cover fuses,
disposable batteries or damage from accident, neglect, misuse, alteration, contamination, or
abnormal conditions of operation or handling. Resellers are not authorized to extend any other
warranty on the behalf of Amprobe. To obtain service during the warranty period, return the
product with proof of purchase to an authorized Amprobe Service Center or to an Amprobe
dealer or distributor. See Repair Section for details. THIS WARRANTY IS YOUR ONLY REMEDY.
ALL OTHER WARRANTIES - WHETHER EXPRESS, IMPLIED OR STATUTORY - INCLUDING IMPLIED
WARRANTIES OF FITNESS FOR A PARTICULAR PURPOSE OR MERCHANTABILITY, ARE HEREBY
DISCLAIMED. MANUFACTURER SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL
OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR THEORY. Since
some states or countries do not allow the exclusion or limitation of an implied warranty or of
incidental or consequential damages, this limitation of liability may not apply to you.
Repair
All Amprobe returned for warranty or non-warranty repair or for calibration should be
accompanied by the following: your name, company’s name, address, telephone number, and
proof of purchase. Additionally, please include a brief description of the problem or the service
requested and include the test leads with the meter. Non-warranty repair or replacement
charges should be remitted in the form of a check, a money order, credit card with expiration
date, or a purchase order made payable to Amprobe.
In-warranty Repairs and Replacement – All Countries
Please read the warranty statement and check your battery before requesting repair. During
the warranty period, any defective test tool can be returned to your Amprobe distributor for
an exchange for the same or like product. Please check the “Where to Buy” section on www.
Amprobe.com for a list of distributors near you. Additionally, in the United States and Canada,
in-warranty repair and replacement units can also be sent to an Amprobe Service Center (see
address below).
Non-warranty Repairs and Replacement – United States and Canada
Non-warranty repairs in the United States and Canada should be sent to an Amprobe Service
Center. Call Amprobe or inquire at your point of purchase for current repair and replacement rates.
USA:
Amprobe
Everett, WA 98203
Tel: 877-AMPROBE (267-7623)
Canada:
Amprobe
Mississauga, ON L4Z 1X9
Tel: 905-890-7600
Non-warranty Repairs and Replacement – Europe
European non-warranty units can be replaced by your Amprobe distributor for a nominal
charge. Please check the “Where to Buy” section on www.Amprobe.eu for a list of
distributors near you.
Amprobe Europe*
Beha-Amprobe
In den Engematten 14
79286 Glottertal, Germany
Tel.: +49 (0) 7684 8009 - 0
www.Amprobe.eu
*(Correspondence only – no repair or replacement available from this address. European customers please
contact your distributor.)
LT-10 / LT-10-EUR Lamp Tester
CONTENTS
SYMBOL.......................................................................................................... 3
SAFETY INFORMATION.................................................................................. 3
UNPACKING AND INSPECTION...................................................................... 4
FEATURE.......................................................................................................... 4
OPERATING THE LAMP TESTER..................................................................... 5
Attaching and removing the antenna..................................................... 6
Attaching and removing the adaptor...................................................... 6
Lamp test................................................................................................... 7
VolTech™ Non-contact voltage detection............................................... 9
Pin test....................................................................................................... 12
DETAILED SPECIFICATIONS............................................................................ 12
MAINTENANCE............................................................................................... 13
TROUBLESHOOTING....................................................................................... 14
BATTERY REPLACEMENT................................................................................ 15
LT-10 / LT-10-EUR Lamp Tester
4
5
1
2
3
1 Test probe
(CAT II 300 V, CAT I 600 V)
2 LED indicator
3 LAMP TEST button
4 Adaptor plate
5 Finger guard
6
6 Locking switch for
antenna connection
7 Antenna connection
to test probe
7
Symbols

W
Caution! Risk of electric shock.
T
The equipment is protected by double insulation or reinforced
insulation.
N
Battery.
)
Caution! Refer to the explanation in this manual.
Canadian Standards Association (NRTL/C)
P
Complies with European Directives.
;
Conforms to relevant Australian standards.
=
Do not dispose this product as unsorted municipal waste.
Contact aqualified recycler.
SAFETY INFORMATION
The meter complies with:
IEC/EN 61010-1 3rd Ed., UL61010-1 2nd Ed. and CAN/CSA C22.2 No. 61010-104 + CSA Update No.1: 2008 to CAT II 300 V, CAT I 600 V, pollution degree 2.
EMC IEC/EN 61326-1
“This product has been tested to the requirements of CAN/CSA-C22.2
No.61010-1, second edition, including Amendment 1, or a later version of
the same standard incorporating the same level of testing requirements.”
Measurement Category II (CAT II) is for measurements
performed on circuits directly connected to low voltage; for example,
measurements on household appliances, portable tools and similar
equipment.
Measurement Category I (CAT I) is for measurements performed on circuits
not directly connected to mains.
CENELEC Directives
The instruments conform to CENELEC low voltage directive 2006/95/EC and
electromagnetic compatibility directive 2004/108/EC
For Use by Competent Persons
Anyone using this instrument should be knowledgeable and trained about
the risks involved with light fixtures and electrical connections. They
must understand the importance of taking safety precautions and testing
the instrument before and after use to ensure that it is in good working
condition.
3
� Warning: Read Before Using
To avoid possible electric shock or personal injury:
• If the tester is used in a manner not specified by the manufacturer,
protection provided by the tester may be impaired.
• For indoor use only. Do not use the tester in rain, snow, damp or wet
locations. Do not use the tester around explosive gas or vapor. Do not
insert or remove the battery in an explosive or flammable environment.
• Comply with local and national safety requirements.
• Use proper protective equipment as required by local or national
authorities.
Unpacking and Inspection
Your shipping carton should include:
1 LT-10 Lamp Tester
1 Detachable antenna
1 Adaptor plate
1 Carrying case
1 9V alkaline battery (installed)
1 Users manual
If any of these items are damaged or missing, return the complete package
to the place of purchase for an exchange.
FEATURES
The Amprobe LT-10 Lamp Tester is a pocket-size tool designed to
troubleshoot florescent lamps and verify presence of voltage in electrical
systems.
Three easy steps to troubleshoot lamps:
1. Check if fluorescent (electroluminescent) light bulb is damaged with
Lamp-test and pin-check functions
2. Verify if voltage is present at the ballast with VolTect™ non-contact
voltage detection.
3. If light bulb passes the test in step 1 and the voltage is present in step 2,
but the lamp is not working, then replace the ballast
4
• Tests fluorescent (electroluminescent) lights
• Built-in VolTect™ non-contact voltage detection
• Lamp and filament test to check fluorescent light bulbs
• Simple one-handed, single button operation
• Ultra-compact design for portability
• 4 foot removable, fully retractable antenna included to test lights and
voltage without a ladder
• Compatible with all fluorescent (electroluminescent) light bulbs:
- T2, T4, T5, T8, T9, T10, T12
- Fluorescent energy saving light bulbs
• Insulated antenna sleeve and tip provided to safely extend and retract
during use
• Does not work with LED and incandescent (standard) light bulbs
• Replacement parts: Antenna LT-10-ANT
OPERATING THE TESTER
The test probe / antenna emits a high frequency voltage (approx. 3 kV) to
ionize the light fixtures in order to diagnose the failure.
� Precautions:
• When LAMP TEST button is pressed, be alert. Do not touch test probe
or antenna. Discharge can cause electric shock or personal injury.
• Do not attempt to overreach. Make sure you have proper footing and
balance at all times.
• Do not allow test probe or antenna to touch energized wires.
• To avoid damages to antenna and personal injury, do not attempt to
bend or use the antenna as a crowbar.
• Do not operate the instrument with the case or the battery door open.
• Do not use if the instrument appears damaged or doesn’t operate
properly. If in doubt, have the instrument serviced.
• The non-contact voltage function is always on. Test on a known live
source within the rated ac voltage range of the product, both before
and after use to ensure the instrument is in good working condition.
5
Test Functions and Indications:
Test
Audio
LED on
None
Pin test
LED on
Solid tone
LED flashing
Modulated beep
Attaching the Antenna
Make sure the switch is securely locked onto the probe before use.
Switch
Figure 1: Attaching the antenna
Removing the Antenna
2
1
Figure 2: Removing the antenna
1 Press and hold down the switch to unlock.
2 Remove the antenna from the probe.
Removing the antenna in a manner not specified in this manual may cause
damages to the instrument and protection provided by the instrument may
be impaired.
6
Attaching and Removing the Adaptor
Attaching adaptor plate
Removing adaptor plate
Figure 3: Attaching and
removing the adaptor
Lamp Test
Non-contact voltage detection is always on. When the probe or the antenna
is close to energized wires, LED flashes and beeps. Non-contact voltage
detection will be disabled when LAMP TEST button is pressed.
For better result, do the following while testing:
• Press and hold down LAMP TEST button for one second and release it
for one second.
Lamp Test with the Probe
1. Attach the antenna to the test probe and make sure the switch is locked
and secured.
2. Pull the antenna to the desired length or fully extended.
3. Touch the surface of the light bulb with the tip of the antenna.
• To boost the test signal, install the adapter plate.

Do not touch the live parts with the antenna.
4. Press and hold down the LAMP TEST button
• The fluorescent lamp is good if lit during lamp test
• The fluorescent lamp is bad if not lit during lamp test
7
• If the fluorescent lamp is lit during the lamp test, but does not
function when installed in the light fixture, the spiral wound
filament, the starter or the ballast may be faulty.

Do not touch the probe during test.
Figure 4: Lamp test against glass
surface in light fixture with the
antenna
Lamp Test with the Antenna
1. Touch the surface of the light bulb or the lamp socket with the probe.

Do not touch energized parts with the probe.
Do not touch the lamp socket during test. This could lead to faulty
test results and cause electric shock.
2. Press and hold down the LAMP TEST button.
• The fluorescent lamp is good if lit during lamp test
• The fluorescent lamp is bad if not lit during lamp test
• If the fluorescent lamp is lit during the lamp test, but does not
function when installed in the light fixture, the spiral wound
filament, the starter or the ballast may be faulty.

Do not touch the live parts with the antenna.
8
Figure 5: Lamp test
against socket with
the probe
Figure 6: Lamp test
against glass surface
with the probe
VolTect™ Non-contact Voltage Detection
� WARNING!
• If the instrument is used in a manner not specified by the manufacturer,
protection provided by the tester may be impaired.
• Test on a known energized source within the rated ac voltage range of
the instrument both before and after use to ensure the instrument is in
good working condition.
• The non-contact voltage detection function is always on. When the
probe or antenna is close to energized wires, LED flashes and beeps.
Non-contact voltage detection will be disabled when LAMP TEST button
is pressed.
• When using the instrument, if LED indicator does not glow or
the instrument does not beep, voltage could still be present. The
instrument indicates active voltage in the presence of electrostatic
fields of sufficient strength generated from the source voltage. If the
field strength is low, the instrument may not provide indication of live
voltage. Lack of an indication occurs if the instrument is unable to sense
the presence of voltage which may be influenced by several factors
including, but not limited to:
- Shielded wire/cables
- Thickness and type of insulation
- Distance from the voltage source
- Fully-isolated users that prevent an effective ground
- Receptacles in recessed sockets or differences in socket design
- Condition of the instrument and batteries
9
• Do not use if the instrument appears damaged or if it doesn’t operate
properly. Closely examine the tip of the probe for cracks or breakage
before use. If in doubt, have the instrument serviced.
• Do not use the instrument to test voltage higher than the rated voltage
as marked on the instrument.
• Use caution with voltages above 30 V ac as a shock hazard may exist.
• Comply with local and national safety requirements.
• Use proper protective equipment as required by local or national
authorities.
The non-contact voltage (NCV) test can be used to check whether the
ballast is functional by verifying the presence of ac voltages at the input
and output of the ballast, or the energized parts of the light fixture. The
light fixture must be turned on before verifying the presence of ac voltage
with the instrument.
Voltage Detection with Antenna
Do not touch the energized parts with the antenna.
1. Turn the light fixture’s power on.
2. Use the antenna to quickly check the presence of ac voltage to the light
fixture.
• If the LED flashes red and beeps (modulated), there is ac voltage
present.
Figure 7: Voltage detection
with antenna
10
Voltage Detection with Probe
Do not touch energized parts with the probe.
1. Turn the light fixture’s power on.
2. Use the probe to detect the presence of ac voltage on the wires at input
of the ballast.
• If the tester’s LED flashes red and beeps, there is ac voltage present.
3. Use the probe to detect the presence of ac voltage on the wires at
output of the ballast.
• If the tester’s LED flashes red and beeps, there is ac voltage present.
The lamp may be faulty.
• If LED and beeper do not activate during voltage detection at the
ballast output wires, the ballast may be faulty.
Figure 8: Voltage detection
at ballast input wires
Figure 9: Voltage detection at ballast
output wires
11
Pin Test
The pin test (filament test) can be used on dual pin fluorescent lamps that
have a filament under the metal cap. The filament is used to excite the gas
inside the tube and turn it on. The lamp will not function correctly if the
filament is broken.
Turn the power off to the light fixture before removing the lamp from
the fixture.
1. PIN test is always on. Plug the lamp pins into the pin test slots on the
back of the tester.
• If the tester does not beep and LED indicator is not on, the filament
is broken. Replace the lamp.
• If the tester beeps and LED indicator is on, the filament is functional.
The pins are good.
2. Repeat step 1 with the other end of the lamp.
LED indicator ON
Solid continuous beep
Figure 10: Plug
the lamp into the
pin test slots
Figure 11: Pin test indications
DETAILED SPECIFICATIONS
Test voltage
(with new battery)
Approximately 3 kV / 280 kHz
Field strength
Approximately 100 μV/m at 260 - 300 kHz
Maximum antenna
length
4-FT (121.92 cm)
T2, T4, T5, T8, T9, T10, T12
AC 90 V to AC 600 V, 50/60 Hz
CAT II 300 V, CAT I 600 V
12
LED illuminates and beeps at approximately 3 mm
Non-contact voltage
(0.12 in) distance from a wire carrying 120 Vac
sensitivity (with probe) LED illuminates and beeps at approximately 5 mm
(0.2 in) distance from a wire carrying 230 Vac
Drop proof
1 meter
Operating time
One second ON, one second OFF for maximum
five cycles and wait for 1 minute before making
another measurement
Operating temperature -10OC to +50OC (14OF to 122OF) ≤85% RH
Storage temperature
-10OC to +50OC (14OF to 122OF) ≤85% RH
Operating altitude
Up to 2000 meters
Battery
1 x 9V battery, 6LR61 / 6LF22 / MN1604 or
equivalent
Battery life
500 tests (alkaline typically)
Approximately 170 x 40 x 24 mm
(6.69 x 1.57 x 0.94 in)
Weight
Approximately 80 g (0.18 lb) with battery
installed
Safety compliance
IEC 61010-1, UL 61010-1
CAN/CSA-C22.2 No. 61010-1-2004
IEC 61326-1
CSA and CE
MAINTENANCE AND REPAIR
If the instrument fails to operate, check the battery and replace as
necessary.
Do the following:
1. Replace the battery if the tester does not work.
2. If antenna is not working, check the antenna connection. Make sure the
antenna is locked and secured.
3. Review the users manual to better understand how the tester operates.
13
Except for the replacement of the battery, repair of the tester should
be performed only by an authorized service center or by other qualified
instrument service personnel.
The front panel and carrying case can be cleaned with a mild solution of
detergent and water. Apply sparingly with a soft cloth and allow to dry
completely before using. Do not use aromatic hydrocarbons, gasoline or
chlorinated solvents for cleaning.
TROUBLESHOOTING
Lamp tester does not work
Problem: Possible low or dead battery
Action: Check and/or replace battery
Low lamp tester response
Problem:
1. Possible low battery
2. Bad contact between the probe/antenna and the lamp tester
Action:
1. Check and/or replace battery
2. Test the area on the lamp where the probe/antenna can make good
contact against lamp glass surface or the lamp socket
Antenna does not work
Problem:
1. Possible low battery
2. Bad contact between the antenna and the lamp under test
3. Bad connection contact between the probe and antenna
Action:
1. Check and/or replace battery
2. Test the area on the lamp where the antenna can make good contact
against lamp glass surface or the lamp socket
3. Check and re-install the antenna. Make sure the antenna is securely
locked onto the probe. If the antenna is still not working, the antenna
may be damaged. Contact a service center for repair or antenna
replacement
14
Probe does not work or work well against lamp socket
Problem:
1. Possible low battery
2. Bad contact between the probe and the lamp socket
NOTE: The insulated probe may not work on socket of
energy saving lamp.
Action:
1. Check and/or replace battery
2. Test the glass tube on the lamp with the probe.
BATTERY REPLACEMENT
Replacing BATTERY follow below steps:
1. On the battery case, turn the screw to “open” position by using a flat
screw driver. Then open the battery cover.
2. Replace battery by removing the battery and replacing it with one 9 V
alkaline battery (6LF22, 6LR61, MN1604) or equivalent. Pay attention to
the polarity signs.
3. Put the battery cover back and re-fasten the screw.
1.
2.
3.
Figure 12: Replacing battery
15
Canada :
Amprobe
Mississauga (Ontario) L4Z 1X9
Tél. : 905-890-7600
5
1
2
3
7
Symboles

W
T
N
Pile.
)
P
;
=
Test
Visuel
Audio
Aucun
Modulation sonore

2
1


6






8

11

Environ 3 kV / 280 kHz
121,92 cm (4 pi)
T2, T4, T5, T8, T9, T10, T12
90 V CA à 600 V CA, 50/60 Hz
CAT II 300 V, CAT I 600 V
12
1 mètre
Pile
Poids
IEC 61010-1, UL 61010-1
CAN/CSA-C22.2 No. 61010-1-2004
IEC 61326-1
CSA et CE
14
2.
3.
15
Reparatur
USA:
Amprobe
Everett, WA 98203
Tel.: 877-AMPROBE (267-7623)
Kanada:
Amprobe
Mississauga, ON L4Z 1X9
Tel.: 905-890-7600
5
1
2
7
Symbole

W
T
N
)
Canadian Standards Association (NRTL/C)
P
;
=
–
Kontaktprüfung
LED blinkt

2
1


6





11

LED indicator ON
Solid continuous beep
Abbildung 11: KontaktprüfungSignalisierung
etwa 3 kV/280 kHz
etwa 100 µV/m bei 260 – 300 kHz
121,92 cm
T2, T4, T5, T8, T9, T10, T12
12
1m
Betriebstemperatur
-10 bis +50 °C, ≤ 85% RH
Lagerungstemperatur
-10 bis +50 °C, ≤ 85% RH
Bis 2000 m
etwa 170 × 40 × 24 mm
(6,69 x 1,57 x 0,94 Zoll)
IEC 61010-1, UL 61010-1
CAN/CSA-C22.2 Nr. 61010-1-2004
IEC 61326-1
2.
3.
15
Italiano
Tester per lampade
USA:
Amprobe
Everett, WA 98203
Tel: 877-AMPROBE (267-7623)
Canada:
Amprobe
Mississauga, ON L4Z 1X9
Tel: 905-890-7600
5
1
2
7
Simboli

W
T
N
)
Canadian Standards Association (NRTL/C)
P
;
=
Test
Visiva
Audio
Test lampada
Test dei pin
Bip modulato

2
1


6






11

Circa 3 kV / 280 kHz
Circa 100 μV/m at 260 - 300 kHz
4-FT (121,92 cm)
Test dei filamenti
T2, T4, T5, T8, T9, T10, T12
Da CA 90 V a CA 600 V, 50/60 Hz
CAT II 300 V, CAT I 600 V
12
1 metro
-10OC - +50OC (14OF - 122OF) ≤85% RH
-10OC - +50OC (14OF - 122OF) ≤85% RH
Durata batteria
Dimensioni
(L x L x H)
Peso
IEC 61010-1, UL 61010-1
CAN/CSA-C22.2 N. 61010-1-2004
IEC 61326-1
CSA e CE
2.
3.
15
Estados Unidos:
Amprobe
Everett, WA 98203
Tfno.: 877-AMPROBE (267-7623)
Canadá:
Amprobe
Mississauga, ON L4Z 1X9
Tfno.: 905-890-7600
5
1
2
6
7
Símbolos

W
T
N
Pila.
)
P
Cumple con la normativa europea.
;
=
Visual
Audio
Sin sonido
Non-contact voltage
(voltaje sin contacto)
LED parpadeando
Pitido modulado

2
1








9
10
11

Aproximadamente 3 kV / 280 kHz
Aproximadamente 100 μV/m a 260 - 300 kHz
4-FT (121,92 cm)
T2, T4, T5, T8, T9, T10, T12
90 V de CA a 600 V de CA, 50/60 Hz
CAT II 300 V, CAT I 600 V
12
1 metro
-10OC a +50OC (14OF a 122OF) ≤85% RH
-10OC a +50OC (14OF a 122OF) ≤85% RH
Hasta 2000 metros
Pila
Aproximadamente 170 x 40 x 24 mm
(6,69 x 1,57 x 0,94 pulgadas)
Peso
IEC 61010-1, UL 61010-1
CAN/CSA-C22.2 N.º 61010-1-2004
IEC 61326-1
2.
3.
Reparationer
USA:
Amprobe®
Everett, WA 98203
Tel: 877-AMPROBE (267-7623)
Kanada:
Amprobe
Mississauga, ON L4Z 1X9
Tel: 905-890-7600
5
1
2
7
Symboler

W
T
N
)
Canadian Standards Association (NRTL/C)
P
;
=
•
•
•
Visuell
Ljud
Ingen
Stifttest
Fast ton
LED blinkar
Modulerat pip
Test

Switch
Figur 1: Sätta fast antennen
Ta bort antennen
2
1

Figur 2: Ta bort antennen


6
Ta bort adapterplatta

4. Tryck och håll inne knappen LAMPTEST
• Den fluorescerande lampan fungerar om den lyser under lamptestet
• Den fluorescerande lampan fungerar inte om den inte lyser under
lamptestet
7
 Rör inte mätsonden under testet.
8

Figur 6: Lamptest
med sond mot
glasyta

LED indicator ON
Solid continuous beep

Figur 11: Stifttestindikationer
DETLAJERADE SPECIFIKATIONER
Testspänning
(med nytt batteri)
Cirka 3 kV / 280 kHz
Fältstyrka
Cirka 100 μV/m vid 260 - 300 kHz
4-FT (121,92 cm)
T2, T4, T5, T8, T9, T10, T12
AC 90 V till AC 600 V, 50/60 Hz
CAT II 300 V, CAT I 600 V
12
1 meter
Arbetstemperatur
Lagringstemperatur
Cirka 170 x 40 x 24 mm
(6,69 x 1,57 x 0,94 in)
Vikt
IEC 61010-1, UL 61010-1
CAN/CSA-C22.2 No. 61010-1-2004
IEC 61326-1
Certifierat
14
2.
3.

Figur 12: Byta batteri
15
Visit www.Amprobe.com for
•
•
•
•
Catalog
Application notes
Product specifications
User manuals
Amprobe®
www.Amprobe.com
[email protected]
Everett, WA 98203
Tel: 877-AMPROBE (267-7623)
Amprobe® Europe
Beha-Amprobe
In den Engematten 14
79286 Glottertal, Germany
Tel.: +49 (0) 7684 8009 - 0
Please
Recycle

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

advertisement

Languages

Only pages of the document in English were displayed