10
Application Programs
Counting
Purpose
With the Counting program you can determine the number of parts that each have approximately equal weight.
Available Features
– Store the current weight value to have it loaded as the preset reference sample quantity next time you initialize the Counting application
– The reference sample quantity can be changed in the operating menu:
See “Configuring the Balance/Scale”
– The average piece weight is automatically output via the optional data interface port after initialization, if the menu code for “Printout with data ID codes” is set
– Press v to toggle the display between piece count and weight
Factory Settings
Reference sample quantity: 10 (menu code 3. 3. 2)
Preparation
§ Configure the Counting application in the operating menu:
See “Configuring the Balance/Scale”
Menu code 2. 1. 4 Counting
§ Reference sample quantity:
Menu code 3. 3. 1
Menu code 3. 3. 2
Menu code 3. 3. 3
Menu code 3. 3. 4
Menu code 3. 3. 5
5 pcs
10 pcs
20 pcs
50 pcs
100 pcs
See also “Configuring the Balance/Scale”
Example
Determine an unknown piece count; weigh the preset reference sample quantity
Settings (changes in the factory settings required for this example):
Menu: Application program: Counting (menu code 2. 1. 4)
Reference sample quantity: 20 pcs (menu code 3. 3. 3)
Key (or instruction) Display/Printout Step
1
1. Turn on the balance/scale e t v >2 sec.
0.0 g ref 20 (briefly)
1 2. Tare the balance/scale
1 3. Display the reference sample quantity (here: 20 pcs)
1 4. Place the reference sample quantity (20 pcs) on the balance/scale (here: 66 g)
1 5. Start the application; if the print format is set to include data ID codes, the following is printed:
1 6. Weigh uncounted parts
(here: 174 pcs)
1 7. Display weight
1 8. Display quantity
1 9. Unload the balance/scale v v v
+
+
66.0 g
20 pcs wRef + 3.300 g
+
+
+
174 pcs
574.2 g
174 pcs
0 pcs c 10. Delete the reference value
11. Repeat the procedure starting from step 6, if desired.
11
12
Weighing in Percent
Purpose
This application program allows you to obtain weight readouts in percent which are in proportion to a reference weight.
Available Features
– Store the current weight value to have it loaded as the preset reference percentage next time you initialize the Weighing in Percent application
– The reference percentage can be changed in the operating menu:
See “Configuring the Balance/Scale”
– The reference percentage is automatically output via the optional data interface port after initialization, if the menu code for “Printout with data ID codes” is set
– Press v to toggle the display between percentage and weight
Factory Settings
Reference percentage: 10 (menu code 3. 3. 2)
Preparation
§ Configure the Weighing in Percent application in the operating menu:
See “Configuring the Balance/Scale”
Menu code 2. 1. 5 Weighing in percent
§ Reference percentage:
Menu code 3. 3. 1
Menu code 3. 3. 2
Menu code 3. 3. 3
Menu code 3. 3. 4
5 %
10 %
20 %
50 %
Menu code 3. 3. 5 100 %
See also “Configuring the Balance/Scale”
Example
Determine an unknown percentage; store the weight on the balance/scale as a reference percentage
Settings (changes in the factory settings required for this example):
Menu: Application program: Weighing in percent (menu code 2. 1. 5)
Menu: Reference percentage 100 % (menu code 3. 3. 5)
Step Key (or instruction) Display/Printout
1 1. Turn on the balance/scale e
1 2. Tare the balance/scale
1 3. Display the reference percentage
1 4. Place the reference weight for 100% on the balance/scale
(here: 222.5 g)
1 5. Start application; if the print format is set to include data
ID codes, the following is printed:
1 6. Place an unknown weight on the balance/scale
(here: 322.5 g)
1 7. Display weight
1 8. Display percentage
1 9. Unload the balance/scale t v > 2 sec.
v v v
+ 222.5 g
+
Wxx% + 222.500 g
+ 144.94 %
+
+
0.0 g ref 100
100.00 %
322.5 g
144.94 %
0.00 % c 10. Delete the reference percentage
11. Repeat the procedure starting from step 6, if desired.
13
14
Weigh Averaging
Purpose
Use this program to determine weights under unstable ambient conditions. In this program, the balance/scale calculates the weight as the average value from a defined number of individual weighing operations. These weighing operations are also known as “subweighing operations” or “subweighs.”
Available Features
– The measured result displayed is the arithmetic mean shown in the selected weight unit; a triangle indicates that this is a calculated value
– You can set the number of subweighing operations performed in the operating menu:
See “Configuring the Balance/Scale”
– Press v for at least 2 sec. to display the preset number of subweighing operations
– Press v to toggle the display between the calculated result and the weight
Factory Settings
Number of subweighs for averaging: 10 (3. 3. 2)
Preparation
§ Configure the Weigh Averaging application in the operating menu:
See “Configuring the Balance/Scale”
Menu code 2. 1.12
Averaging
– Number of subweighs for weigh averaging:
3. 3. 1 5 subweighs
3. 3. 2
3. 3. 3
10 subweighs
20 subweighs
3. 3. 4 50 subweighs
3. 3. 5 100 subweighs
See also “Configuring the Balance/Scale”
Example
Determine the weight of a sample in extremely unstable ambient conditions by calculating the average of 10 subweighing operations.
Settings (changes in the factory settings required for this example):
Menu: Application program: Weigh Averaging (menu code 2. 1.12)
Step Key (or instruction) Display/Printout
1. Turn on the balance/scale e
2. Tare the balance/scale
3. Display the number of subweighs (here: 10)
4. Place sample on the balance/ scale (weight readout fluctuates; here: about 275 g)
5. Start measurement t v >2 sec.
v ref
0.0 g
10 (briefly)
+ 8888
+ 8888
10
9
8
...
1
+ 275.5 g G
After 10 subweighs
If the print format is set to include data ID codes, the following is printed:
6. Unload the balance/scale
RES + 275.5 g
+ 275.5 g G
(stable display) c 7. Delete the result
8. Repeat the procedure starting from step 4, if desired.
15
Net-Total Formulation/Second Tare Memory
With this application program you can weigh in components for formulation of a mixture.
Preparation
Configure the Net-Total Formulation/Second Tare Memory application in the operating menu: See “Configuring the Balance/Scale” Menu code 2.1.3
Example
Step Key (or instruction) Display/Printout
1. Turn on the balance/scale e
2. Place an empty container on the balance/scale
+ 65.0 g
3. Tare the balance/scale
4. Add the first component t
+ 120,5 g
5. Store the first component weight
If the print format is set to include data ID codes, the following is printed:
6. Add the next component
7. Store the 2nd component weight
8. Add further components, if desired
9. Display total weight and fill to desired final weight v 0.0 g
NET
N1 + 120.5 g
+ 70,5 g
0.0 g
NET v
As described for steps 5 and 6 c + 191,0 g
16
Toggle Between Weight Units
With this application program you can toggle the display of a weight value back and forth between two weight units.
Configure the “Toggle Weight Units” application in the operating menu:
See “Configuring the Balance/Scale,” menu code 2.1. 2 Toggle weight units (Factory setting on Gem lite models)
Menu code
1. 7. 1
1. 7. 13
1. 7. 14
1. 7. 15
1. 7. 16
1. 7. 17
1. 7. 18
1. 7. 19
1. 7. 20
1. 7. 2 o
1. 7. 3 1
)
1. 7. 4 2
)
1. 7. 5
1. 7. 6
1. 7. 7
1. 7. 8
1. 7. 9
1. 7. 10
1. 7. 11
1. 7. 12
1. 7. 22
3. 1. 1
3. 1. 13
3. 1. 14
3. 1. 15
3. 1. 16
3. 1. 17
3. 1. 18
3. 1. 19
3. 1. 20
3. 1. 22
3. 1. 2 o
3. 1. 3 1
)
3. 1. 4
3. 1. 5
3. 1. 6
3. 1. 7 3 )
3. 1. 8
3. 1. 9
3. 1. 10
3. 1. 11
3. 1. 12
Unit
Taiwanese taels 4 )
(only selectable on Gem lite scales)
Grams
Kilograms
Carats
Pounds
Ounces
Troy ounces
Hong Kong taels
Singapore taels
Taiwanese taels
Grains
Pennyweights
Milligrams
Parts per pound
Chinese taels
Mommes
Austrian carats
Tola
Baht
Mesghal lb/oz – Gem lite scales only
Conversion factor
0.02666666000
Abbr. on printout o
1.00000000000
0.00100000000
5.00000000000
0.00220462260
0.03527396200
0.03215074700
0.02671725000
0.02645544638
0.02666666000
15.43235835000
0.64301493100
GN dwt
1000.00000000000
mg
1.12876677120
/lb
0.02645547175
0.26670000000
5.00000000000
tlc mom
K
0.08573333810
0.06578947437
0.21700000000
0.03527396200
tol bat
MS o g ozt tlh tls tlt kg ct lb oz
ο = Factory setting
1 ) = not for models with a readability of ≤ 0.2 mg
2 ) = Factory setting only for model GC503
3 ) = Factory setting only for GM models
4 ) = Readability in increments of 1 digit
Function
● To toggle the display between the 1st and 2nd weight units:
Press the v key