SEPARATED SPEECH & NOISE CHANNELS
Tracks 24-35 contain the 12 standard QuickSIN lists recorded with the speech and noise on two separate channels (target speech on channel one and 4-talker babble on channel two). The purpose of these lists is to provide a quick way to verify the effectiveness of hearing aids that have switchable directional microphones. On these tracks, both speech and babble were recorded at constant levels; therefore, the tester must establish and control the signal-to-noise ratios by selecting the presentation levels for both speech and babble channels, and manually change the level of the babble channel for each sentence to adjust the signal-to-noise ratio.
Directional Comparisons
A complete measurement of a directional hearing aid requires extensive laboratory facilities, but a good demonstration of the ability of directional hearing aids to reject sound from the sides and rear can be obtained in a standard test booth with loudspeakers located in the corners (at +45 o and -45 o or 0 o and 180 o azimuth).
It is important to remember that any test conducted in a sound booth will not precisely reflect results in the real world. By design, sound booths have minimal reverberation, and testing is conducting using a limited number of loudspeakers
(usually two) that are in fixed locations. In this setup, it is possible that the location of the speakers may interact with the null of the directional microphone(s). Therefore, these measures should not be used to assess effectiveness of one directional microphone design vs. another (where differences are usually small) but rather as a general measure comparing OMNI to Directional, to verify that the directional microphones are working and providing directivity (rejection of sound from the sides and rear).
Procedure:
If the loudspeakers are located at +45 o and -45 o , test each ear separately. Position the patient in the sound booth so that speech is presented from in front at 45 o and babble from behind at approximately 135 o . Direct the speech (channel one) to the loudspeaker at 45 o and direct the babble (channel two) to the loudspeaker at 135 o .
Note: There are two possible “45 degree” orientations for the patient. The desired orientation places the aided ear between the loudspeakers. See Figure 3. When the other ear is tested, the patient will need to be rotated to face the opposite wall and the speech and babble switched to the opposite speakers.
11
Directional Comparisons Continued
Figure 3
Figure 4
12
If the loudspeakers are located at 0 o and 180 o , you may test each ear separately or both ears together. Position the patient in the sound booth so that speech is presented from in front at 0 o and babble from behind at 180 o
.
See Figure 4.
Since the babble noise on Tracks 24-35 is at a constant level, two types of demonstrations are possible:
1. Subjective. Calibrate both channels. Have the patient set the hearing aid/s to
OMNI. Adjust the speech to 50 dB HL (65 dB SPL conversational speech level) and then adjust the noise to the level where the patient reports that it just prevents understanding the speech. Have the patient switch back and forth between OMNI and Directional positions on the hearing aid/s. The improved intelligibility in the directional mode should be obvious.
2. Objective. Calibrate both channels. Set the dial for channel one (front speaker) to 50 dB HL. Set the dial for channel two to 25 dB HL for the first sentence, and increase the dial setting for channel two (babble) by 5 dB for each succeeding sentence (see Table 2). Score each list as before to obtain SNR Loss. Test in
OMNI and Directional. A minimum of three lists in each condition (six total) is required for a valid comparison to an accuracy of 1.5 dB at the 80% confidence level (see page 20). If you find a difference greater than 4 dB with one list in each condition, you have already reached the 95% confidence level.
Sentence 1
Sentence 2
Sentence 3
Sentence 4
Sentence 5
Sentence 6
Table 2
Channel 1
Dial (dB HL)
Channel 2
Dial (dB HL) SNR (dB)
50
50
50
50
50
50
25
30
35
40
45
50
10
5
0
25
20
15
13