INFCPH and INFCOP INFINITY Microprocessor Based pH Controller

Add to my manuals
56 Pages

advertisement

INFCPH and INFCOP INFINITY Microprocessor Based pH Controller | Manualzz

4

Configuring The Meter

4.5 PERFORMING THREE-POINT CALIBRATION (“CAL.3”)

Note ☞ Only buffer solutions of 4, 7 and 10 are acceptable. IF the buffer value is not accepted, the meter will display a flashing “PH.ER“.

1.

Place your electrode into the pH buffer solution.

2.

Press the MENU button until the meter shows “CAL.3

3.

Press the

T/mV button. The meter shows “IN 1” (Input 1)

4.

Press the 䊳 T/mV button again. The meter shows the previous value of “IN 1”.

5.

Press the 䊳 T/mV button again. The meter shows the actual buffer solution’s pH value.

Note

Please allow enough time for the electrode to settle before processing.

6.

Press the MENU button to store the “IN 1” value. The meter momentarily shows an “OK.4”, followed by “IN 2” (Input 2)

7.

Rinse the electrode with distilled water and place it into the pH 7 buffer solution.

8.

Press the 䊳 T/mV button. The meter shows the previous value of “IN 2

9.

Press the

T/mV button again. The meter shows the actual buffer solution’s pH value.

Note ☞

Please allow enough time for the electrode to settle before processing.

10.

Press the MENU button to store the “IN 2” value. The meter momentarily shows an “OK.7”, followed by “IN 3” (Input 3)

11.

Rinse the electrode with distilled water and place it into the pH 10 buffer solution.

12.

Press the 䊳 T/mV button. The meter shows the previous value of “IN 3

13.

Press the

T/mV button again. The meter shows the actual buffer solution’s pH value.

Note ☞

Please allow enough time for the electrode to settle before processing.

14.

Press the MENU button to store the “IN 3” value. The meter momentarily shows an “OK.10”, then momentarily shows “STRD”, followed by “S1.CF” (Setpoint 1

Configuration).

24

Configuring The Meter

4

4.6 USING SETPOINT 1 CONFIGURATIONS (“S1.CF”)

You may use Setpoint 1 Configuration for the following:

• To set setpoint 1’s active band above or below your chosen value

• To select whether setpoint 1’s operation is latched or unlatched

4.6.1 Setting Setpoint 1's Active Band

1.

Press the MENU button until the meter shows “S1.CF”.

2.

Press the

T/mV button. The meter shows one of the following:

• “S.1=A” (Active above setpoint) (Default)

• “S.1=B” (Active below setpoint)

3.

Press

/pH button to toggle between available choices.

4.

Press the 䊳 T/mV button to select if setpoint 1 is latched or unlatched or press the

MENU button to store your selection and enter “S2.CF” (Setpoint 2 configurations).

4.6.2 Selecting if Setpoint 1 is Latched or Unlatched

1.

Select the 䊳 T/mV button. The meter shows one of the following:

• “S.2=L” Setpoint 1 to be latched

“S.2=U” Setpoint 1 to be unlatched (Default)

2.

Press 䊱 /pH button to toggle between available choices.

3.

Press the MENU button to store your choices. The meter momentarily shows

STRD”, followed by “S2.CF” (Setpoint 2 Configuration).

4.6.3 Selecting Setpoint 1 (for pH/ORP Controllers only)

1.

Select the

T/mV button. The meter shows one of the following:

• “S.3=P” Setpoint 1 assigned to pH

• “S.3=O” Setpoint 1 assigned to ORP

2.

Press 䊱 /pH/ORP button to toggle between available choices.

3.

Press the MENU button to store your choices. The meter momentarily shows

STRD”, followed by “S2.CF” (Setpoint 2 Configuration).

25

4

Configuring The Meter

4.7 USING SETPOINT 2 CONFIGURATION (“S2.CF”)

You may use Setpoint 2 Configuration for the following:

• To set setpoint 2’s active band above or below your chosen value

• To select whether setpoint 2’s operation is latched or unlatched

4.7.1 Setting Setpoint 2's Active Band

1.

Press the MENU button until the meter shows “S2.CF”.

2.

Press the

T/mV button. The meter shows one of the following:

“S.1=A” (Active above setpoint) (Default)

• “S.1=B” (Active below setpoint)

3.

Press the 䊱 /pH button to toggle between available choices.

4.

Press the

T/mV button to select if setpoint 2 is latched or unlatched or press the

MENU button to store your selection and enter “S1.DB” (Setpoint 1 Deadband)

4.7.2 Selecting if Setpoint 2 is Latched or Unlatched

1.

Press the

T/mV button. The meter shows one of the following:

• “S.2=L” setpoint 2 to be latched

“S.2=U” setpoint 2 to be unlatched (Default)

2.

Press the

/pH button to toggle between available choices.

3.

Press the MENU button to store your choices. The meter momentarily shows

STRD”, followed by “S1.DB” (Setpoint 1 Deadband).

4.7.3 Selecting Setpoint 2 (for pH/ORP Controllers only)

1.

Select the 䊳 T/mV button. The meter shows one of the following:

• “S.3=P” Setpoint 2 assigned to pH

• “S.3=O” Setpoint 2 assigned to ORP

2.

Press

/pH/ORP button to toggle between available choices.

3.

Press the MENU button to store your choices. The meter momentarily shows

STRD”, followed by “S1.DB” (Setpoint 1 Deadband).

26

Configuring The Meter

4

4.8 SETTING THE SETPOINT 1 DEADBAND (“S1.DB”)

Factory default deadband is 001.0. The deadband’s resolution value depends upon the decimal point position previously set-up in Section 4.2. The minimum deadband setting should be no less than 0.1 pH. To change the deadband (hysteresis) of setpoint 1, follow these steps:

1.

Press the MENU button until the meter shows “S1.DB”.

2.

Press the

T/mV button. The meter shows the last stored number (0000 through

9999) with flashing 4th digit.

3.

Press the

/pH button to change the value of the flashing digit. If you continue to press the 䊱 /pH button, the flashing digit's value continues to change.

4.

Press the

T/mV button to scroll to the next digit.

5.

Press the MENU button to store your selection. The meter momentarily shows

STRD”, followed by “S2.DB” (Setpoint 2 Deadband).

4.9 SETTING THE SETPOINT 2 DEADBAND (“S2.DB”)

Factory default deadband is 001.0. The deadband’s resolution value depends upon the decimal point position previously set-up in Section 4.2. The minimum deadband setting should be no less than 0.1 pH. To change the deadband (hysteresis) of setpoint 2, follow these steps:

1.

Press the MENU button until the meter shows “S2.DB”.

2.

Press the

T/mV button. The meter shows the last stored number (0000 through

9999) with flashing 4th digit.

3.

Press the 䊱 /pH button to change the value of the flashing digit. If you continue to press the

/pH button, the flashing digit's value continues to change.

4.

Press the 䊳 T/mV button to scroll to the next digit.

5.

Press the MENU button to store your selection. The meter momentarily shows

STRD”, followed by “OT.CF” (Output Configuration).

27

4

Configuring The Meter

Figure 4-1. Relay Output Triggering Example

OFF ON

ACTIVE ABOVE

OFF

SETPOINT

SIGNAL

LEVEL

ON OFF

ACTIVE BELOW

ON

DEADBAND

SIGNAL

LEVEL

ON OFF

ON

ACTIVE BELOW

WITH DEADBAND 3

ON

NOTE: DEADBAND WORKS AS HYSTERISIS

OFF ON

ACTIVE ABOVE

WITH DEADBAND 3

OFF

3

SETPOINT

3

SIGNAL

LEVEL

ON ON ON

ACTIVE BELOW LATCHED

OFF ON ON

ACTIVE ABOVE LATCHED

Note ☞

To reset latched alarms you must:

1.

Input a signal “OUT” of the Relay Triggering zone

2.

Then press SETPTS and then, RESET button

SETPOINT

28

Configuring The Meter

4

4.10 USING OUTPUT CONFIGURATION (OT.CF)

Use Output Configuration to select the following:

• To enable or disable the analog output

• To determine if the analog output is current or voltage

4.10.1 Enabling or Disabling the Analog Output

To enable or disable the analog output, follow these steps:

1.

Press the MENU button until the meter shows “OT.CF”.

2.

Press the

T/mV button. The meter shows one of the following:

“O.1=E” (Analog output enabled) (Default)

• “O.1=D” (Analog output disabled)

3.

Press the 䊱 /pH button to toggle between available choices.

4.

Press the

T/mV button to select the analog output as current/voltage or press the

MENU button to store your selection. The meter momentarily shows “STRD”, followed by “OT.S.O” (Output Scale and Offset) - refer to section 4.12.

4.10.2 Selecting Analog Output as Current or Voltage

1.

Press the

T/mV button. The meter shows one of the following:

“O.2=C" (Analog output = current) (Default)

• “O.2=V” (Analog output = voltage)

2.

Press the

/pH button to toggle between available choices.

3.

Press the 䊳 T/mV button to go back to selecting analog output as enabled or disabled or press the MENU button to store your choices. The meter momentarily shows “STRD”, followed by “OT.S.O” (Output Scale and Offset) - refer to section

4.12.

29

4

Configuring The Meter

4.10.3 Selecting Analog Output as pH or ORP( for pH/ORP Controllers only)

1.

Press the

T/mV button. The meter shows one of the following:

• “O.3=P" (Analog Output assigned to pH)

• “O.3=O” (Analog Output assigned to ORP)

2.

Press the

/pH/ORP button to toggle between available choices.

3.

Press the 䊳 T/mV button to go back to selecting analog output as enabled or disabled or press the MENU button to store your choices. The meter momentarily shows “STRD”, followed by “OT.S.O” (Output Scale and Offset) - refer to section

4.12.

30

Configuring The Meter

4

4.11 CONFIGURING TEMPERATURE (C or F - “TEMP”)

Note

If you have selected “MANU” in “A.T.C.” menu item (Section 4.1), then you have to specify the constant temperature value. Maximum/minimum values are 0 to

199.9 for C or 0 to 398.0 for F.

1.

Press the MENU button until the meter shows “TEMP”.

2.

Press the

T/mV button. The meter shows actual constant temperature value.

3.

Press the 䊱 /pH button until the first digit is correct.

4.

Press the

T/mV button to advance to the next digit.

5.

Repeat steps 3 and 4 until meter shows the desired “TEMP” value.

6.

Press the MENU button to store your selection. The meter momentarily shows

STRD”, followed by “OT.S.O” (Output Scale and Offset) - refer to section 4.12.

7.

Press the MENU button until the meter shows the “CAL.2 or CAL.3” menu.

If the display shows “CAL.2” then proceed to Section 4.4.

If the display shows “CAL.3” then proceed to Section 4.5.

Note ☞

C or F flashes to indicate the unit of measure.

31

4

Configuring The Meter

4.12 USING OUTPUT SCALE AND OFFSET (OT.S.O)

Output Scale and Offset (“OT.S.O”) scales your analog output to be equal to the meter's display and/or any engineering units you require. You may scale the output for direct

(4-20 mA, 0-10 V, etc) or reverse acting (20-4 mA, 10-0 V, etc).

1.

Press the MENU button until the meter shows “OT.S.O”.

2.

Press the

T/mV button. The meter shows “RD 1” (Read 1).

Note ☞

This is your first point of display reading.

3.

Press the 䊳 T/mV button again. The meter shows the last stored Low pH number with flashing 4th digit.

4.

Press the

/pH button to change the value of Read 1.

5.

Press the

T/mV button to scroll to the next digit.

6.

Repeat steps 4 and 5 until the desired value is selected.

7.

Press the MENU button to store your selection. The meter shows “OUT.1” (Output

1) (Low output).

8.

Press the 䊳 T/mV button. The meter shows the selected output.

Note ☞

If you select “O.2=V ” for voltage, the maximum signal you may select is

10.00 for an 0-10 Vdc signal output. If you select “O.2=C” for current, the maximum signal you may select is 20.00 for 0-20 or 4-20 mA dc signal output.

9.

Press the 䊱 /pH button to enter the Output 1 signal selection. If you continue to press the

/pH button, the flashing digit's value continues to change.

10 Press the 䊳 T/mV button to scroll to the next digit.

11.

Repeat steps 9 and 10 until the desired value is selected.

12.

Press the MENU button to store your selection. The meter shows “RD 2” (Read 2).

Note ☞ This is your second point of display reading.

32

Configuring The Meter

4

4.12 USING OUTPUT SCALE AND OFFSET (OT.S.O) (Continued)

13.

Press the

T/mV button. The meter shows the last stored High pH number with flashing 4th digit.

14.

Press the

/pH button to change the value of the flashing digit. If you continue to press the 䊱 /pH button, the flashing digit's value continues to change.

15.

Press the

T/mV button to scroll to the next digit.

16.

Repeat steps 14 and 15 until the desired value is selected.

17.

Press the MENU button to store your selection. The meter shows “OUT.2

(Output 2). (High output)

18.

Press the

T/mV button. The meter shows selected output.

Note ☞

If you select “O.2=V ” for voltage, the maximum signal you may select is

10.00 for an 0-10 Vdc signal output. If you select “O.2=C” for current, the maximum signal you may select is 20.00 for a 0-20 or 4-20 mA dc signal output.

19.

Press the 䊱 /pH button to change the value of the flashing digit. If you continue to press the

/pH button, the flashing digit's value continues to change.

20.

Press the

T/mV button to scroll to the next digit.

21.

Repeat steps 19 and 20 until the desired value is selected.

22.

Press the MENU button to store your selection. The meter momentarily shows

STRD”, then “RST”. The meter will enter into the “RUN” mode (Normal operation).

CAUTION: If the meter displays all flashing values on any item, the value has overflowed. Press the 䊱 /pH button to start new values.

33

4

Configuring The Meter

4.12.1 Example for Output Scale and Offset

You want to send 4-20 mA output for 0 to 14 pH value (default). The meter has 0.01 pH resolution. Complete the following steps:

1.

Press the MENU button until the meter shows “OT.S.O”.

2.

Press the

T/mV button. The meter shows “RD 1” (Read 1).

3.

Press the 䊳 T/mV button again to show the existing value.

4.

Change the value of “RD 1” to 00.00 by pressing the

/pH and

T/mV and buttons.

5.

Press the MENU button to store your selection. The meter shows “OUT.1

(Output 1).

6.

Press the

T/mV button again to show the existing value.

7.

Change the value of “OUT.1” to 04.00 by pressing the 䊱 /pH and 䊳 T/mV and buttons.

8.

Press the MENU button to store your selection. The meter shows “RD 2” (Read 2).

9 Press the

T/mV button to show the existing value.

10.

Change the value of “RD 2” to 14.00 by pressing the

/pH and

T/mV buttons.

11.

Press the MENU button to store your selection. The meter shows “OUT.2

(Output 2).

12.

Press the 䊳 T/mV button to show the existing value.

13.

Change the value of “OUT.2” to 20.00 by pressing the

/pH and

T/mV buttons.

14.

Press the MENU button to store your selection. The meter shows “STRD”.

34

Selecting Setpoint Values

5

SECTION 5. SELECTING SETPOINT VALUES

Follow the steps below to select values for Setpoint 1 and Setpoint 2.

1.

Press the SETPTS button. The meter momentarily shows “SP1” (Setpoint 1), followed by the last stored value with flashing 4th digit. The factory default for

SP1” is 000.0.

2.

Press the

/pH button to change the value of Setpoint 1.

3.

Press the

T/mV button to scroll to the next digit.

4.

Press the SETPTS button to store your selection. The meter momentarily shows

SP2” (Setpoint 2), followed by the last stored value with flashing 4th digit. The factory default for “SP2” is 000.0.

5.

Press the 䊱 /pH button to change the value of Setpoint 2.

6.

Press the

T/mV button to scroll to the next digit.

7.

Press the SETPTS button to store new values. The meter momentarily shows

STRD”, “RUN” and then enters the run mode.

Note ☞

You may press the RESET button anytime during this routine to return to the run mode.

35

6

Display Messages

SECTION 6. DISPLAY MESSAGES

Table 6-1. Display Messages

MESSAGE

PH.ER

±LMT

LOCK

UN.LK

CAL.2

CAL.3

TEMP

OFF

RST

A.T.C.

DEC.P

RD.CF

S1.CF

S2.CF

S1.DB

S2.DB

OT.CF

OT.S.O

±OPN

9999

-1999

ER1

VOLT

METR

MANU

SP.RS

SP1

SP2

ER2

DESCRIPTION

Hard (power on) reset

Temperature Compensation Mode

Decimal point

Reading configuration

Setpoint 1 configuration

Setpoint 2 configuration

Setpoint 1 deadband

Setpoint 2 deadband

Output configuration

Output scale and offset

Sensor break or temperature outside range

Value overflow in setpoint/menu peak deviation routines

Value overflow in setpoint/menu peak routines

2 coordinate format programming error

Value Direct from probe in mV

RTD input for A.T.C.

Manually entered input for A.T.C.

Reset setpoints

Setpoint 1 value

Setpoint 2 value

One or more of these items have overflowed due to decimal point change: setpoint values, setpoint deadbands pH value is not accepted

±pH over limit

Lock for RESET, MENU and SETPTS buttons

Unlock for RESET, MENU and SETPTS buttons

Two-point calibration

Three-point calibration

Temperature of solution (A.T.C. - METR) or Temperature configuration (A.T.C. - MANU)

Constant 25ºC input for A.T.C

36

Menu Configuration Displays

7

SECTION 7. MENU CONFIGURATION DISPLAYS

MENU

A.T.C.

Temperature Compensation

Mode

Table 7-1. Configuration Menu

(Defaults in Bold and Italics)

T/mV

Show A.T.C. choices:

/pH

METR: Metered - default

MANU: Manually

OFF: Constant 25ºC

DEC.P

Decimal Point

RD.CF

Reading Configuration

S1.CF

Setpoint 1 Configurations

S2.CF

Setpoint 2 Configurations

Show decimal point position

R.2

R.1

FFFF.

FFF.F

FF.FF

C: Celsius

F: Fahrenheit

2: Two-Point Calibration

3: Three-Point Calibration

S.1

S.2

S.1

S.2

A: Active above

B: Active below

U: Unlatched

L: Latched

A: Active above

B: Active below

U: Unlatched

L: Latched

S1.DB

Setpoint 1 Configurations

S2.DB

Setpoint 2 Configurations

CAL.2

Two-Point Calibration

IN 1

Input new value and show “IN 1

IN 2

Enter new value and show “IN 2

Scroll right one digit

Scroll right one digit

Show prior and actual value

Show prior and actual value

Change flashing digit’s value

Change flashing digit’s value

37

7

Menu Configuration Displays

SECTION 7. MENU CONFIGURATION DISPLAYS

Table 7-1. Configuration Menu (Continued)

(Defaults in Bold and Italics)

MENU

OT.CF

Output Configuration

T/mV

0.1

D: Disabled

E: Enabled

䊱 /pH

Analog Output Option 0.2

V: Voltage Analog out

C: Current Analog out

OT.S.O

Output Scale & Offset

Enter new value & show

OUT1

Enter new value & show

RD 2

Enter new value & show

OUT2

TEMP

Temperature Configuration

Show “RD 1” & prior value

Scroll right one digit

Show prior value

Scroll right one digit

Show prior value

Scroll right one digit

Show prior entered value

Scroll right one digit

Change flashing digit’s value

Change flashing digit’s value

Change flashing digit’s value

Change flashing digit’s value

CAL.3

Three-Point Calibration

IN 1

Input new value and show “IN 1

IN 2

Input new value and show “IN 2

IN 3

Input new value and show “IN 3

Show prior and actual value

Show prior and actual value

Show prior and actual value

38

Menu Configuration Displays

7

SECTION 7. MENU CONFIGURATION DISPLAYS

Table 7-2. Run Mode Displays

DISPLAY

TEMP

Temperature

Reading

VOLT mV Reading

SP.RS

䊳 T/mV 䊱 /pH

Shows temperature value of input signal. Press again to show mV value of input signal.

Shows mV value of input signal. Press again to show temperature value of input signal.

RESET Description

LATCHED

RESET

Press RESET button to reset your setpoints.

39

8

Setpoint Configuration Displays

SECTION 8. SETPOINT CONFIGURATION DISPLAYS

Table 8-1. Setpoint Configuration Displays

MENU

SP 1

Setpoint 1

䊳 T/mV

Scroll right one digit

䊱 /pH

Change flashing digit’s value

DESCRIPTION

Select from -1999 through 9999

SP 2

Setpoint 2

Scroll right one digit

Change flashing digit’s value

Select from -1999 through 9999

40

Specifications

9

SECTION 9. SPECIFICATIONS

INPUT SIGNAL

pH and Volt Specification

Range:

Resolution:

- 2.00 to 16.00 pH (-620 to +620 mV)

Accuracy:

Calibration:

From: -200 to -620: 1 mV

± 0.01 pH (± 0.1 mV)

Two or Three Point

Temperature Compensation: Automatic: 0ºC to 100ºC (32ºF to 212ºF)

Manual: 0ºC to 100ºC (32ºF to 212º F)

Note

Volt specification is only used to verify proper operation of electrodes.

)

ORP Specification

Range: -620 to +620 mV

Resolution:

Accuracy:

Note

From: -199.9 to 620.0 mV: 0.1 mV

From: -200 to -620: 1 mV

± 0.1 mV

ORP specification is only used to verify proper operation of electrodes.

TEMPERATURE

Range:

Resolution:

Accuracy:

Unit of Measure:

Sensor:

Isolation:

Noise Rejection:

0ºC to 100ºC (32ºF to 212ºF)

0.1ºC

±0.5ºC

ºF or ºC (programmable by software)

100/1000 Ohm Pt RTD (Connection by phone jack or screw terminals)

Dielectric strength to 2500V transient per 3mm spacing based on EN 61010 for 260Vrms or dc working voltage

Normal Mode Rejection (NMR) = 60dB

Common Mode Rejection (CMR) = 120dB

LED 14-segment, 13.8 mm (0.54”) Display:

Symbol:

Input Impedance: >10 12 ohm

41

9

Specifications

SECTION 9. SPECIFICATIONS (Continued)

ANALOG TO DIGITAL

Technique:

Internal Resolution:

Read Rate:

Dual Slope

15 bits

3 per second for pH; 1 per second for Temperature

RELAY OUTPUTS

2 Form "C" on/off relays. Configurable for latched and unlatched by software.

Max current: 5 A, resistive load

Max voltage: 250 Vac or 30 Vd

ANALOG OUTPUT

Signal Type:

Signal Level:

Function:

Linearity:

Step Response Time:

Current or voltage

Current: 10 V max compliance at 20 mA output

Voltage: 20 mA max for 0-10 V output

May be assigned to a display range or proportional control output with setpoint #1 when used as a control output.

0.2%

2 - 3 seconds to 99% of the final value

ISOLATED ANALOG OUTPUT (TB5, if applicable)

Same as non-isolated analog output except isolated to 1000 Vdc.

Signal Type:

Signal Level:

Function:

Linearity:

Step Response Time:

Current or voltage

Current: 10 V max compliance at 20 mA output

Voltage: 20 mA max for 0-10 V output

May be assigned to a display range or proportional control output with setpoint #1 when used as a control output.

0.2%

2 - 3 seconds to 99% of the final value

Note

Only one analog output is available on each unit and it must be factory installed.

42

Specifications

9

SECTION 9. SPECIFICATIONS (Continued)

INPUT POWER INFORMATION

~ AC units 115/230 V~(AC) ±10%, 50/60 Hz

7 W max, power consumption (Non-Isolated Analog Output)

8 W max, power consumption (Isolated Analog Output)

DC units 10-32 Vdc

6 W max, power consumption (Non-Isolated Analog Output)

7 W max, power consumption (Isolated Analog Output)

External Fuse Protection Recommended:

IEC 127-2/III

Power

115 V

Fuse

125 mA @ 250 (T)

230 V 63 mA @ 250 (T)

UL 248-14 (Listed Fuse)

Power Fuse

115 V

230 V

175 mA @ 250 V Slow-Blow

80 mA @ 250 V Slow-Blow

ENVIRONMENT

Operating temperature:

Storage temperature:

Relative humidity:

0° to 50°C (32° to 122°F)

-40° to 85°C (-40° to 185° F)

90% at 40°C (non-condensing)

MECHANICAL

Panel cutout:

Weight:

Case material:

1/8 DIN 3.62 x 1.78" (45 x 92mm)

1.27 lb (575 g)

Polycarbonate, 94 V-O UL rated

43

9

Specifications

SECTION 9. SPECIFICATIONS (Continued)

48,00 (1.890) 96,00 (3.780)

17,78 (.700)

CASE

REAR

COVER

SIDE VIEW TOP VIEW

PANEL THICKNESS

6,4 (.25) MAX

0,8 (.03) MIN

1,5

R(.06)

4 PLCS

45,00 + 0,61/-0,00

(1.772 + .024/–.000)

92,00 + 0,81/–0,00

(3.622 + .032/–.000)

NOTE: Dimensions in Millimeters (Inches)

Figure 9-1. Dimensions/ Panel Cutout

Figure 9-1

44

Factory Preset Values

10

SECTION 10. FACTORY PRESET VALUES

Table 10-1. Factory Preset Values

MENU ITEM FACTORY PRESET VALUES

A.T.C.

DEC.P

RD.S.O

RD.CF

S1.CF

S2.CF

S1.DB

S2.DB

OT.CF

OT.S.O

SP1

SP2

TEMP

Temperature Compensation: METR

Decimal Point Position: FFF.F

Reading Scale and Offset:

4-20 mA dc = 0-1000

Reading Configuration:

R.1=3 (3 Point Calibration)

R.2=C (Degree Celsius)

Setpoint 1 Configuration:

S.1=A (Setpoint is active above)

S.2=U (Setpoint is unlatched)

Setpoint 2 Configuration:

S.1=A (Setpoint is active above)

S.2=U (Setpoint is unlatched)

Setpoint 1 Deadband: 0003

Setpoint 2 Deadband: 0003

Output Configuration:

O.1=E (Analog output is enabled)

O.2=C (Analog output is current)

Output Scale and Offset:

0-14 pH = 4-20 mA

Setpoint 1 Value: 0000

Setpoint 2 Value: 0000

Temperature Compensation value set to 025.0

45

46

CE APPROVAL INFORMATION

1. Electromagnetic Compatibility (EMC)

This device comforms with requirements of EMC Directive 89/336/EEC, amended by 93/68/EEC. This instrument complies with the following EMC

Immunity Standards as tested per EN 50082-2, 1995 (Industrial environment)

Phenomena

Electrostatic

Discharge

Radio Frequency electromagnetic field.

Test Specification

+/- 4 kV contact discharge

+/- 8 kV air discharge

27 - 1000 MHz

10 V/m

80% AM (1 KHz)

Basic Standard

IEC 1000-4-2

Performance

Criteria B

IEC 1000-4-3

Performance

Criteria A

Radio Frequency 900 MHz electromagnetic field.

10 V/m

Pulse modulated.

50% Duty cycle @ 200 Hz

Fast Transients +/- 2 kV (ac mains)

+/- 1 kV (dc, signal I/O)

IEC 1000-4-3

Performance

Criteria A

IEC 1000-4-4

Performance

5/50 ns Tr/Th, 5 KHz rep. freq.

Criteria B

Radio Frequency conducted

0.15 - 80 MHz

10 V/m

80% AM (1 KHz)

IEC 1000-4-6

Performance

Criteria A

This instrument complies with the following EMC Emission Standards as tested per EN 50081-1, 1992 (Residential, Commercial and Light Industrial)

Phenomena

Radiated

Emission

Conducted

Emission

Frequency

Range

30-230 MHz

230-1000 MHz

0.15-0.5 MHz

0.5-5 MHz

5-30 MHz

Limits

30 dB_V/m at 10 m

37 dB_V/m at 10 m quasi peak

66-56 dB_V quasi peak

56 dB_V quasi peak

60 dB_V quasi peak

Basic

Standard

CISPR 22

Class B

CISPR 22

Class B

2.Safety

This device conforms with Low Voltage Directive 73/23/EEC, amended by 93/68/EEC.

The following LVD requirements have been met to comply with EN 61010-1, 1993

(Electrical equipment for measurement, control and laboratory use)

1. Pollution Degree 2

2. Installation Category II

3. Double Insulation

4. Class I Equipment (Units with 115/230 Vac power)

Class III Equipment (Units with 10-32 Vdc power)

Warranty/Disclaimer

NEWPORT Electronics, Inc. warrants this unit to be free of defects in materials and workmanship for a period of one (1) year from the date of purchase. In addition to NEWPORT’s standard warranty period, NEWPORT Electronics will extend the warranty period for one (1) additional year if the warranty card enclosed with each instrument is returned to

NEWPORT.

If the unit should malfunction, it must be returned to the factory for evaluation. NEWPORT’s Customer Service

Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by NEWPORT, if the unit is found to be defective it will be repaired or replaced at no charge. NEWPORT’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of being damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of NEWPORT’s control. Components which wear are not warranted, including but not limited to contact points, fuses, and triacs.

NEWPORT is pleased to offer suggestions on the use of its various products. However, NEWPORT neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by NEWPORT, either verbal or written. NEWPORT warrants only that the parts manufactured by it will be as specified and free of defects. NEWPORT MAKES NO

OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED,

EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF

MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF

LIABILITY: The remedies of purchaser set forth herein are exclusive and the total liability of NEWPORT with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall

NEWPORT be liable for consequential, incidental or special damages.

CONDITIONS: Equipment sold by NEWPORT is not intended to be used, nor shall it be used: (1) as a “Basic

Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, or used on humans, or misused in any way, NEWPORT assumes no responsibility as set forth in our basic WARRANTY /

DISCLAIMER language, and additionally purchaser will indemnify NEWPORT and hold NEWPORT harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.

Return Requests/Inquiries

Direct all warranty and repair requests/inquiries to the NEWPORT Customer Service Department.

BEFORE RETURNING ANY PRODUCT(S) TO NEWPORT, PURCHASER MUST OBTAIN AN

AUTHORIZED RETURN (AR) NUMBER FROM NEWPORT’S CUSTOMER SERVICE DEPARTMENT

(IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence.

The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.

FOR WARRANTY RETURNS, please have the following information available BEFORE contacting NEWPORT:

1.

P.O. number under which the product was

PURCHASED,

2.

Model and serial number of the product under warranty, and

3.

Repair instructions and/or specific problems relative to the product.

FOR NON-WARRANTY REPAIRS, consult

NEWPORT for current repair charges. Have the following information available BEFORE contacting NEWPORT:

1.

P.O. number to cover the COST of the repair,

2.

Model and serial number of product, and

3.

Repair instructions and/or specific problems relative to the product.

NEWPORT’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering.

NEWPORT is a registered trademark of NEWPORT ELECTRONICS, INC.

© Copyright 2001 NEWPORT ELECTRONICS, INC. All rights reserved. This document may not be copied, photocopied, reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without prior written consent of NEWPORT ELECTRONICS, INC.

M1570/N/0801

For immediate technical or application assistance please call:

®

Newport Electronics, Inc.

2229 South Yale Street • Santa Ana, CA • 92704 • U.S.A.

TEL: (714) 540-4914 • FAX: (714) 546-3022

Toll Free: 1-800-639-7678 • http://www.newportUS.com • e-mail:[email protected]

ISO 9001 Certified

Newport Technologies, Inc.

976 Bergar • Laval (Quebec) • H7L 5A1 • Canada

TEL: (514) 335-3183 • FAX: (514) 856-6886

Toll Free: 1-800-639-7678 • http://www.newport.ca • e-mail:[email protected]

Newport Electronics, Ltd.

One Omega Drive • River Bend Technology Centre

Northbank, Irlam • Manchester M44 5BD • United Kingdom

Tel: +44 161 777 6611 • FAX: +44 161 777 6622

Toll Free: 0800 488 488 • http://www.newportuk.co.uk • e-mail:[email protected]

Newport Electronics B.V.

Postbus 8034 • 1180 LA Amstelveen • The Netherlands

TEL: +31 20 3472121 • FAX: +31 20 6434643

Toll Free: 0800 0993344 • http://www.newport.nl • e-mail: [email protected]

Newport Electronics spol s.r.o.

Rudé armády 1868, 733 01 Karviná 8 • Czech Republic

TEL: +420 69 6311899 • FAX: +420 69 6311114

Toll Free: 0800-1-66342 • http://www.newport.cz • e-mail: [email protected]

Newport Electronics GmbH

Daimlerstrasse 26 • D-75392 Deckenpfronn • Germany

TEL: 49 7056 9398-0 • FAX: 49 7056 9398-29

Toll Free: 0800 / 6397678 • http://www.newport.de • e-mail: [email protected]

Newport Electronique S.A.R.L.

9, rue Denis Papin • 78190 Trappes • France

TEL: +33 130 621 400 • FAX: +33 130 699 120

Toll Free: 0800-4-06342 • http://www.newport.fr • e-mail: [email protected]

Mexico and Latin America

TEL: 001-800-826-6342 • FAX: 001 (203) 359-7807

En Español: 001 (203) 359-7803

11799ML-04 Rev. A

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals