Overview. Emerson OCX 8800 O2 / Combustibles Transmitter Hazardous Area, OCX 8800
Add to my manuals
164 Pages
advertisement
Instruction Manual
IM-106-880, Rev 1.0
January 2007
Section 4
OCX 8800
Using HART Communications
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 4-1
HART Communicator Signal Connections . . . . . . . . . . . . page 4-1
HART Communicator PC Connections . . . . . . . . . . . . . . . page 4-4
HART Menu Tree . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 4-5
D/A Trim Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 4-9
OVERVIEW
HART COMMUNICATOR
SIGNAL CONNECTIONS
The HART communicator is a handheld communications interface device. It provides a common communications link to all microprocessor-based instruments that are HART compatible. The handheld communicator contains an 8 x 21 character liquid crystal display (LCD) and 25 keys. A pocket-sized manual, included with the HART communicator, details the specific functions of all the keys.
To interface with the OCX 8800, the HART communicator requires a termination point along the O
2
4-20 mA current loop and a minimum load resistance of 250 ohms between the communicator and the power supply.
The HART communicator accomplishes its task using a Frequency Shift
Keying (FSK) technique. With the use of FSK, high-frequency digital communication signals are superimposed on the 4-20 mA oxygen output signal. The communicator does not disturb the 4-20 mA signal since no net energy is added to the loop. HART information is not available on the COe output signal.
The HART communicator may be interfaced with a personal computer (PC), providing special software has been installed. To connect the HART communicator to a PC, an interface adapter is required. Refer to the proper
HART communicator documentation in regard to the PC interface option.
The HART communicator can connect to the OCX 8800 oxygen analog output signal line at any wiring termination in the O
2
4-20 mA current loop. There are two methods of connecting the HART communicator to the signal line. For applications in which the signal line has a load resistance of 250 ohms or more, refer to method 1. For applications in which the signal line load resistance is less than 250 ohms, refer to method 2.
http://www..raihome.com
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 42 Component Checklist
- 42 System Overview
- 42 Specifications
- 43 Installation
- 43 Electrical Installation
- 43 Pneumatic Installation
- 43 Initial Startup
- 44 Verify Installation
- 44 Initial Power Up
- 44 Set Test Gas Values
- 44 OCX 8800 Reset Procedure
- 45 Overview
- 45 HART Communicator Signal Connections
- 45 HART Communicator PC Connections
- 45 HART Menu Tree
- 45 D/A Trim Procedures
- 46 Overview
- 46 Fully Automatic Calibration
- 46 Operator - Initiated Autocalibration
- 46 Manual Calibration
- 47 Overview
- 47 and Installation
- 47 Remote Electronics
- 47 Repair Sensor Housing
- 47 Sensor Housing Disassembly
- 47 Sensor Housing Assembly
- 47 Repair Electronics Housing
- 47 Electronics Housing Disassembly
- 98 Electronics Housing Assembly
- 98 Replace Tube Fittings
- 98 Remove Tube Fittings
- 98 Install Tube Fittings
- 99 Overview
- 99 Grounding
- 99 Electrical Noise
- 99 Electrostatic Discharge
- 99 Total Power Loss
- 99 Diagnostic Alarms
- 99 Fault Isolation
- 99 Alarm Relay Events
- 100 Sensor Housing
- 100 Electronics Housing
- 100 O2 Cell and Heater Strut Assembly