Gravity Compensation. Cardinal Detecto MARINER WPS10


Add to my manuals
12 Pages

advertisement

Gravity Compensation. Cardinal Detecto MARINER WPS10 | Manualzz

CONFIGURATION, CONT.

GRAVITY COMPENSATION

1. With the display showing :

To Enable the Gravity Compensation mode, press the SET key. The display will change to

On

(use the factory set value of 9.797).

To Disable the Gravity Compensation mode, press the UNIT key. The display will change to OFF (use the value obtained from the factory Tech Support).

2. Press the SET key again. The display will change to show GrAu .

3. Press the UNIT key. The display will change to show

End

.

4. Press the SET key to save the settings and reset the scale.

5. Setup is complete.

GRAVITY COMPENSATION

The Gravity Compensation for your scale has been pre-configured at the factory and should not require changing in most locations. In the event that the factory settings do not meet the requirements of your location, contact the factory Tech Support for the value for your location.

With

GrAu

enabled in the function setting of the Configuration mode, the scale will use the factory set value of 9.797 for the Gravity Compensation.

If the

GrAu

is disabled, the scale will use the value obtained from the factory Tech Support and entered by the operator for the Gravity Compensation.

To change the Gravity Compensation value:

1. Press and hold the UNIT key while pressing the ON/OFF key to turn on the scale and enter Gravity Compensation mode, the display will show 9.797 (or the current value).

2. Press the UNIT key to increase the value or TARE key to decrease the value to achieve the desired value.

3. Press the SET key to confirm the entered gravity and return the scale to the normal operating mode.

CALIBRATION PROCEDURE

1. While the scale is off, press and hold the ON/OFF key.

2. Once all display segments are on, release the ON/OFF key and immediately and press the UNIT key three (3) times.

3. The scale will show “11111”

”55555”, and finally show the counts of zero.

4. Press the SET key. The scale will show “0g” and then go back to the counts.

5. Place a 5 kg test weight on the scale and then press the SET key.

6. The scale will show “5000g”, and change to show “CAL”, and then automatically turn off.

7. Remove the test weight from the scale.

8. The scale is ready for normal operation.