Chapter 3
SPECIAL FUNCTIONS
The infusion is within the range of the maximum soft limit
The infusion is within the range of the minimum soft limit
Violation of the upper soft limit
Violation of the lower soft limit
=
=
=
=
No soft limit is defined
Only a drug name is available
(It is possible to select a drug name only from the drug library)
=
=
The limits of the drug library have to comply with the limits of the pump and the disposable.
Note: An adequate monitoring when infusing highly potent drugs is recommended.
Note: In case a drug from the drug library is selected and the pump is running under dose rate calculation the initial values will be overwritten by the drug library values if selected.
Remote Drug Library update from Upload Manager (Space Online Suite)
The file icon blinks every 2 s. An update is available.
The Drug Library Upload starts as soon as the pump is in Passive mode.
Note: You can cancel the upload by pressing c
.
Please contact your local sales represantative in case you like to use Remote
Drug Library update.
3.4 Patient Controlled Analgesia (PCA)
For PCA a drug list with at least one drug activating the profile PCA is necessary. By this the conditions for an effective and safe therapy are defined.
Switch on pump with o and wait until self-check is finished. Depending on the settings the choice of a drug is offered direcly or the pump is in “Main Menu”.
Select “Special Functions” with t from “Main Menu” and confirm with l .
Select drug list, category and desired drug by using q .
23
24
Chapter 3
SPECIAL FUNCTIONS
After the selection the pump offers additional drug related
Information which are confirmed by l .
Select profile PCA by using and confirm with t .The therapy settings stored in the drug list are displayed *.
The therapy can be started now with s in case all values are defined.
Depending on the pre-defined settings the therapy is started with an initial bolus and a basal rate or not.
Before leaving the patient the pump should be put into DataLock level 3 with q in Menu “Options”. This is necessary especially in case non-authorised access to the settings can be anticipated.
The code is entered with q and confirmed with k .
The pump display now may look like this.
In this state the patient is allowed to demand boli. Depending on the status of the therapy these are either administered or denied. Changing the disposable is also possible by using the code for level 1 or level 2. Altering the settings for
PCA or other therapies however is only possible with the code for level 3.
The status of the therapy can be checked in the menu „Status“ .
Enter the „Main Menu“ with c and select the “Status“ with q .
The A/D-ratio indicates the percentage of administered and demanded boli thus giving an idea about the effectivity of the therapy.
An acoustic confirmation of demanded boli can be activated and modulated by t in Data Lock 3.
Is a demand button connected, the therapy symbol looks like this:
In case there is no demand button connected the therapy symbol looks like this: .
The demand button is connected to the interface P2 at the rear side of the pump.
*Bolus volume is the volume of a single bolus the patient may demand. Max. Limit is the amount of drug or volume a patient may demand within a certain time in total. Lockout is the time in between two boli.