Electrical safety testing
Welch Allyn recommends performing ground continuity, leakage current, and dielectric strength tests
1
when replacing the power supply or primary wiring according to EN/IEC
60601-1 - Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and Essential Performance or EN/IEC 62353 - Medical Electrical Equipment - Recurrent
Test and Test After Repair of Medical Electrical Equipment.
Due to the variability of test equipment in the field, Welch Allyn does not include specific instructions to perform electrical safety tests. When performing electrical safety tests, refer to your test equipment manuals for detailed instruction. The following table provides connections and test limits to assist you in performing these tests.
Test Limits
Ground continuity
Leakage current
Ground continuity from EP stud* (equipotential terminal) to the Gnd pin of the
IEC power connector shall be no greater than 0.1 ohms.
Leakage current shall be less than 300 µA from EP stud* to mains (Line and
Neutral pins of the IEC power connector).
Dielectric strength Dielectric strength shall be 1.8 kVAC EP stud* to IEC mains (Line and Neutral pins of the IEC power connector).
* To locate the equipotential terminal, see “Controls, indicators, and connectors.”
69
1
Perform this test only if there is a reason to doubt the integrity of the electrical insulation (e.g. multiple trips of a residual-current device or liquid ingress of a saline solution). If you determine this test should be performed, return the device to Welch Allyn for service.
70 Electrical safety testing CP 150 12-lead resting electrocardiograph
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 3 Contents
- 5 Symbols
- 5 Documentation symbols
- 5 Power symbols
- 6 Connectivity symbols
- 6 Shipping, storing, and environment symbols
- 7 Miscellaneous symbols
- 7 Electromagnetic compatibility (EMC) information
- 9 Safety
- 9 General warnings
- 9 Warnings related to the environment
- 9 Warnings related to accessories and other equipment
- 10 Warnings related to using the electrocardiograph
- 11 General cautions
- 12 General safety considerations
- 13 Electrostatic discharge (ESD)
- 15 Overview
- 15 Purpose and scope
- 15 Related documents
- 15 Technical support services
- 16 Service loaners
- 16 Service options
- 16 Partners in Care service agreements
- 16 Warranty service
- 16 Non-warranty service
- 17 Repairs
- 18 Configuration options
- 19 The Welch Allyn Service Tool
- 19 Returning products
- 23 Service menu
- 23 Access the Service screen
- 23 Performing the full-functional tests
- 24 Battery test
- 24 Audio test
- 25 USB host test
- 25 USB client test
- 26 Keypad test
- 27 ECG lead test
- 28 Printer pattern test
- 29 Printer speed test
- 30 Ping test
- 30 Test results
- 31 Manage tests and other files
- 33 Troubleshooting
- 33 Symptoms and solutions
- 33 System problems:
- 35 Printer
- 35 Printer problems:
- 37 User interface
- 37 User interface problems:
- 39 Post test error messages
- 39 Post test error messages:
- 41 Disassembly, repair, and reassembly
- 42 CP 150 Disassembly Flow chart
- 42 Required tools and equipment
- 44 Power down the electrocardiograph
- 44 Remove the battery
- 45 Separate the top housing assembly from the bottom housing assembly
- 48 Top housing disassembly
- 48 Remove the Input/Output (I/O) communication board
- 51 Remove the SMCE board
- 53 Remove the LCD frame and display assembly
- 55 Bottom housing disassembly
- 55 Remove the ECG board
- 56 Remove the ECG ferrite
- 57 Remove the battery housing cover and the battery connector board
- 60 Remove the power supply assembly
- 61 Remove and disassemble the printer assembly
- 63 Remove the printer board
- 65 Remove the printer door sensor
- 66 Remove the gear train and motor
- 68 Remove the print head assembly
- 69 Remove the print head
- 73 Electrical safety testing
- 75 Software updates process
- 75 Update process
- 76 Export and import of the device configuration
- 79 Field replaceable units
- 79 Housing top assembly
- 79 Serv Kit, CP150, Hsg Top Assy (material no. CP150-0001)
- 80 Serv Kit, CP150, 7 Inch Display (material no. CP150-0002)
- 80 Serv Kit, Frame LCD (material no. CP150-0003)
- 81 Serv Kit, SMCE PCA (material no. CP150-0004)
- 82 Serv Kit, CP150, I/O PCA (material no. CP150-0005)
- 83 Housing base assembly
- 83 Serv Kit, CP150, ECG PCA (material no. CP150-0006)
- 84 Serv Kit, Battery Case Assy (material no. CP150-0007)
- 85 Serv Kit, Printer PCA (material no. CP150-0008)
- 86 Printer 8 inch assembly tray
- 86 Serv Kit, CP150, Tray paper assy (material no. CP150-0009)
- 87 Serv Kit, Printer Top Assy (material no. CP150-0010)
- 88 Serv Kit, Print Head Assy (material no. CP150-0011)
- 89 Serv Kit, Gear Train Assy (material no. CP150-0012)
- 89 Serv Kit, Motor Assy (material no. CP150-0013)
- 90 Serv Kit, Door Sensor (material no. CP150-0014)
- 90 Bottom assembly
- 91 Serv Kit, Bottom assy (material no. CP150-0015)
- 91 Serv Kit, Power supply (material no. CP150-0016)
- 91 Serv Kit, Software upgrade (material no. CP150-0017)
- 92 Serv Kit, Battery door (material no. CP150-0018)
- 92 Serv Kit, Miscellaneous (material no. CP150-0019)
- 93 Partners in Care service and support agreements
- 93 Service and repair training
- 95 Appendices
- 95 Decontamination and cleaning requirements