- No category
advertisement
Agilent 490 Micro Gas Chromatograph
User Manual
4
GC Channels
Micro Electronic Gas Control (EGC) 47
The instrument contains up to 2 channels in a dual channel cabinet, or up to 4 channels for a quad channel cabinet. A GC channel contains a gas regulator, an injector, a column, and a
TCD detector. See Figure 18 on page 46.
This chapter provides a brief discussion on the major components in the Micro GC and the backflush option.
Agilent Technologies
45
4
GC Channels
Carrier Gas
C A U T I O N
The Micro GC is configured for use with either He and H
2 or
N
2
and Ar.
Agilent recommends you use gases with a minimum purity of
99.995%. Since the injection valve is operated pneumatically, there is a limit of 550 kPa ± 10% (80 psi ± 10%) to the main gas supply.
Your Micro GC is configured either for carrier gas He and H
2 or N
2
and Ar.
Use the carrier gas type for which your instrument is configured, otherwise the detector filaments can be damaged.
Gas Clean unit
(optional)
Microelectronic gas control
(EGC)
Injector
Columns
(analytical and reference)
µTCD
Reference vent
Column vent
Sample Out
Carrier gas
Figure 18 Gas flow diagram
Sample in
46 490 Micro GC User Manual
GC Channels
4
Micro Electronic Gas Control (EGC)
The Micro GCs have built-in regulators that can be adjusted to get a constant or programmed pressure control, which, once constant or programmed pressure control is obtained, results in a constant or programmed flow through the injector, column and detector. The pressure range is from 50 to 350 kPa (7 to
49 psi). This pressure sets a continuous flow of carrier gas of about 0.2 to 4.0 mL/min (depending on column length and type).
A typical pressure rise is 200 kPa/min, which will give a significant pressure increase during the run without excessive baseline disturbance. In most cases baseline subtraction may improve the quality of chromatograms that suffer from baseline drift.
Injector
The injector has a built-in 10-µL sample loop that is filled with the gaseous sample. The pressure of the sample should be between 0 and 100 kPa (0 to 15 psi) and the sample temperature within 5 to 110 °C ± 5 °C of the analyzer.
When the chromatographic data system sends a START command, the vacuum pump draws the gas sample through the loop and the injector injects the gas sample from the sample loop into the gas stream. A typical injection time is 40 milliseconds (ms). This equals an average injection volume of
200 nL. Injection time will be rounded to a multiple of 5 ms. A practical minimum value is 40 ms. A value of 0 to 20 milliseconds might result in no injection.
490 Micro GC User Manual 47
4
GC Channels
Column
A variety of column configurations are possible on the Micro
GC. The columns you require for your specific analyses have been installed at the factory. Other configurations are, of course, possible, but altering the GC channels is a delicate matter that can only be handled by an Agilent service engineer.
shows several standard columns as supplied in the
Micro GCs and selected applications. Other columns are available by contacting Agilent Technologies.
Table 7
Agilent Micro GC columns and applications
Column/Phase type
Molsieve 5Å
Hayesep A
CP-Sil 5 CB
CP-Sil 19 CB
CP-WAX 52 CB
PLOT Al
2
O
3
/KCl
PoraPLOT U
PoraPLOT Q
CP-CO
X
CP-Sil 19CB for THT
CP-Sil 13CB for TBM
MES NGA
Target components
Permanent gases (N
2
/O
2
separation), methane, CO, NO, and so forth. 20 m required for O
2
-Ar baseline separation). Natural gas and biogas analysis. Optional Retention
Time Stability (RTS) configuration.
Hydrocarbons C
1
–C
3
, N
2
, CO
2
, air, volatile solvents, natural gas analysis.
Hydrocarbons C
3
–C
10
, aromatics, organic solvents, natural gas analysis.
Hydrocarbons C
4
–C
10
, high boiling solvents, BTX.
Polar volatile solvents, BTX.
Light hydrocarbons C
1
–C
5
saturated and unsaturated. Refinery gas analysis.
Hydrocarbons C
1
–C
6
, halocarbons/freons, anesthetics, H
2
S, CO solvents. Separation of ethane, ethylene, and acetylene.
2
, SO
2
, volatile
Hydrocarbons C
1
–C
6
, halocarbons/freons, anesthetics, H
2
S, CO
2
, SO
2
, volatile solvents. Separation of propylene and propane, coelution of ethylene and acetylene.
and O
2
), CH
4
.
THT and C
3
–C
6
+
in Natural Gas Matrix.
TBM and C
3
–C
6
+
in Natural Gas Matrix.
Unique column specially tested for MES in natural gas (1 ppm)..
C A U T I O N
All columns except the HayeSep A (160 °C) and MES (110 °C) columns can be used up to 180 °C, the maximum temperature of the column oven.
Exceeding this temperature will cause the column to lose efficiency instantly and the column module will need replacement. All channels have a built-in protection that prevents a setpoint above the maximum temperature.
48 490 Micro GC User Manual
GC Channels
4
Molsieve 5Å columns
The Molsieve 5Å column is designed to separate: hydrogen, carbon monoxide, methane, nitrogen, oxygen, and some noble gases. Higher molecular weight components have much higher retention times on this column.
Table 8
Molsieve 5Å instrument parameters
Parameter
Column temperature
Injector temperature
Column pressure
Sample time
Injection time
Run time
Detector sensitivity
Peak 1
Peak 2
Peak 3
Peak 4
Peak 5
4m Heated
110°C
110°C
100 kPa (15 psi)
30 s
40 ms
25 s
Auto
Hydrogen 1.0%
Argon/Oxygen 0.4%
Nitrogen 0.2%
_________
_________
10m Unheated
40°C
NA
150 kPa (21 psi)
30 s
40 ms
140 s
Auto
Neon 18 ppm
Hydrogen 1.0%
Argon 0.2%
Oxygen 0.2%
Nitrogen 0.2%
20m Unheated
40 °C
NA
200 kPa (28 psi)
30 s
40 ms
210 s
Auto
Neon 18 ppm
Hydrogen 1.0%
Argon 0.2%
Oxygen 0.2%
Nitrogen 0.2%
6
4
2
0 mV
14
Molsieve 5Å 4 m heated
2
12
10
8
3
1
0 5 10 15 20 25
Seconds mV
4.5
4
3.5
Molsieve 5Å 10 m unheated
3
4
mV
450
400
350
3
2.5
2
1.5
5
1
0.5
12
0
0 20 40 60 80 100 120 140 160 seconds
300
250
200
150
100
50
0
0
Molsieve 5Å 20 m unheated
1
50
2
3
4
100 150
5
200 250
Seconds
490 Micro GC User Manual 49
4
GC Channels
CP-Sil 5 CB columns
The natural gas components, mostly hydrocarbons, separate in the same order on the non-polar and medium-polar CP-Sil CB columns. Nitrogen, methane, carbon dioxide, and ethane are not separated on these columns. They produce a composite peak.
For separation of these components, consider a HayeSep A column.
Table 9
CP-Sil 5 CB instrument parameters
Parameters
Column temperature
Injector temperature
Column pressure
Sample time
Injection time
Run time
Detector sensitivity
Peak 1
Peak 2
Peak 3
Peak 4
Peak 5
4m Heated
50 °C
110 °C
150 kPa (21 psi)
30 s
40 ms
30 s
Auto
Composite Balance
Ethane 8.1%
Propane 1.0% i-Butane 0.14% n-Butane 0.2%
6m Unheated
50 °C
NA
150 kPa (21 psi)
30 s
40 ms
30 s
Auto
Composite Balance
Ethane 8.1%
Propane 1.0% i-Butane 0.14% n-Butane 0.2%
25
15
5
-5
0 mV
55
CP Sil 5 CB 4 m heated
1 2
45
3
35
5
4 5
10 15 20 25 30 35
Seconds
24
19
14
9 mV
34
CP Sil 5 CB 6 m unheated
1 2 3
29
4
5
4
-1
0 10 20 30 40 50 60 70
Se conds
50 490 Micro GC User Manual
GC Channels
4
CP-Sil CB columns
Table 10
CP-Sil CB instrument parameters
Parameter
Column temperature
Injector temperature
Column pressure
Sample time
Injection time
Run time
Detector sensitivity
Peak 1
Peak 2
Peak 3
CP-Sil 13 CB 12m Heated (TBM) CP-Sil 19 CB 6m Heated (THT)
40°C 85 °C
50°C
250 kPa (38 psi)
85 °C
200 kPa (25 psi)
30 s
255 ms
80 s
Auto
30 s
255 ms
35 s
Auto
Methane balance
TBM 6.5 ppm
________
Helium balance
THT 4.6 ppm
Noane 4.5 ppm mV
0.6
0.4
0.2
0
-0.2
-0.4
-0.6
-0.8
0
CP Sil 13 CB
12 m heated (TBM)
10 20
1
30 40 50
2
60 70 80 90
Se conds
-0.5
-1
-1.5
mV
0.5
CP Sil 19 CB
6 m unheated (THT)
1
0
-2
0 10 20 30
2
3
40 50 60
Seconds
490 Micro GC User Manual 51
4
GC Channels
PoraPlot 10m column
Table 11
PoraPlot 10m instrument parameters
Parameter
Column temperature
Injector temperature
Column pressure
Sample time
Injection time
Run time
Detector sensitivity
Peak 1
Peak 2
Peak 3
Peak 4
Peak 5
4
5
2
3
Auto
1
PoraPlot u 10m Heated PoraPlot Q 10m Heated
150°C
110°C
150 °C
110 °C
150 kPa (21 psi)
30 s
40 ms
100s
150 kPa (21 psi)
30 s
40 ms
50 s
Auto
Composite Balance
Ethane 8.1%
Propane 1.0% i-Butane 0.14% n-Butane 0.2% mV PoraPlot U 10 m heated
30
25
20
15
50
45
40
35
10
5
0
2 6
1 2
3
4
5
10 14 18 22 26 30 34 38 42 46 50
Seconds mV
680
580
480
380
280
180
80
-20
0
PoraPlot Q 10 m heated
10 20
1 2
30
3
40
4
5
50 60
Seconds
52 490 Micro GC User Manual
GC Channels
4
Hayesep A 40 cm heated column
WA R N I N G
The HayeSep A column separates oxygen, methane, carbon dioxide, ethane, acetylene, ethylene, and selected sulfur gases.
Nitrogen coelutes with oxygen. Components with a higher molecular weight than propane have long retention times on this column.
Maximum allowable column temperature is 160 °C.
Table 12
Hayesep instrument parameters
Parameter
Column temperature
Injector temperature
Column pressure
Sample time
Injection time
Run time
Detector sensitivity
Peak 1
Peak 2
Peak 3
Hayesep A 40 cm Heated
50 °C
110 °C
150 kPa (21 psi)
30 s
40 ms
60 s
Auto
Nitrogen 0.77%
Methane Balance
Ethane 8.1%
245
195
145
95 mV
345
Hayesep A 40 cm heated
2
295
45
-5
0
1
10 20 30
3
40 50 60 70
Seconds
490 Micro GC User Manual 53
4
GC Channels
CO
X
and AL
2
0
3
/KCI columns
Table 13
CO
X
and AL
2
0
3
/KCI instrument parameters
Parameter
Column temperature
Injector temperature
Column pressure
Sample time
Injection time
Run time
Detector sensitivity
Peak 1
Peak 2
Peak 3
Peak 4
Peak 5
CO
X
1m Unheated
80 °C
NA
200 kPa (28 psi)
30 s
40 ms
204 s
Auto
Hydrogen 1.0%
Nitrogen 1.0%
CO 1.0%
Methane 1.0%
CO
2
1.0%
Helium Balance
AL
2
0
3
/KCI 10m Heated
100
°C
110
°C
150 kPa (21 psi)
30 s
40 ms
60 s
Auto
Composite Balance
Ethane 8.1%
Propane 1.0% i-Butane 0.14% n-Butane 0.2%
2.5
2
1.5
1 mV
3.5
3
CO
X
2
1 m unheated
3
4
0.5
0
1
-0.5
0 50 100
5
150 mV
115
Al
2
O
3
/KCl 10 m heated
1 2
95
75
3
55
35
200
15
250
Seconds
-5
0 10 20
4 5
30 40 50 60 70
Seconds
54 490 Micro GC User Manual
GC Channels
4
MES (NGA) and CP-WAX 52 CB columns
Table 14
MES (NGA) and CP-WAX 52 CB instrument parameters
Parameter
Column temperature
Injector temperature
Column pressure
Sample time
Injection time
Run time
Detector sensitivity
Peak 1
Peak 2
Peak 3
Peak 4
MES 10m Heated (NGA) CP-WAX 52 CB 4m Heated
90 °C
110 °C
60 °C
110 °C
70 kPa (10 psi)
30 s
500 ms
120 s
150 kPa (21 psi)
30 s
40 ms
35 s
Auto
Nitrogen Balance
Noane 11.2 ppm
MES 14. 2 ppm
________
Auto
Nitrogen 0.75%
Acetone 750 ppm
Methanol 0.15%
Ethanol 0.30%
Helium Balance mV
4000
3500
3000
2500
2000
1500
1000
500
0
-500
0
MES 10 m heated (NGA)
20 40 60
1
0.9
0.5
2
3
0.1
- 0.3
80 90 100 110 120 130 mV CP-WAX 52 CB 4 m heated
1
13
8
2
3
4
3
80 100 120 140
Seconds
-2
1 6 11 16 21 26 31 36
Seconds
490 Micro GC User Manual 55
4
GC Channels
Column conditioning
Follow this procedure to make sure that any water that might be present inside the analytical column is removed before the TCD is switched on.
Also follow this procedure if the Micro GC module has been stored for a long period.
C A U T I O N
The detector filaments may be damaged by improper conditioning. Follow this procedure to avoid damaging the detector filaments.
Column conditioning procedure
1
2
3
4
Switch off the TCD filaments in the method.
Set the column temperature of the module to the maximum temperature (160 °C or 180 °C depending on the column limit). Leave the filaments off.
Download this method to the Micro GC.
Run the downloaded method to condition the column, preferably overnight.
This will assure you that all the water has been removed from the column and no damage will occur to the TCD filaments.
Nitrogen and oxygen merging in Molsieve columns
On a properly activated column, nitrogen and oxygen will be well separated. However, in time you will find that these two peaks begin to merge together. This is caused by water and carbon dioxide present in the sample or carrier gas, adsorbing to the stationary phase.
To restore the column efficiency, condition the column, Z described above, for about an hour. After reconditioning, you can test the column performance by injecting plain air. If you have a proper separation between nitrogen and oxygen again, the column separation power has been restored. If the Micro GC frequency of use is very high, you might consider routinely leaving the oven temperature at 180 °C overnight. The longer the reconditioning period, the better the column performance.
56 490 Micro GC User Manual
GC Channels
4
Backflush Option
Backflush to vent is an advanced technique used to prevent later-eluting compounds from reaching the analytical column and detector. The main reason for applying this technique is to keep the analytical column clean and reduce analysis time.
The Micro GC is optionally available with GC modules that incorporate backflush capabilities.
1 2 3 4 5
6
8
Natural gas analysis, straight
Figure 19 Natural gas analysis
7
50 s
1 = Methane
2 = Ethane
3 = Propane
4 = iso-Butane
5 = Butane
6 = iso-Pentane
7 = Pentane
1 2
3
4
5
8 50 s
Natural gas analysis, with backflush at 8 seconds
A backflush system always consists of a pre-column and an analytical column. The two columns are coupled at a pressure
point, which makes it possible to invert the carrier gas flow direction through the pre-column at a preset time, called the
backflush time. See Figure 21 on page 58.
The injector, two columns, and detector are in series.
The sample is injected onto the pre-column where a pre-separation takes place; injection takes place in normal mode. See Figure 20 on page 58.
490 Micro GC User Manual 57
4
GC Channels
Restriction
System pressure
Pressure regulator
Pre-column
Analytical column
Pressure point
Detector
Injector Backflush vent
Figure 20 Backflush system normal flows
When all compounds to be quantified are transferred to the analytical column, the backflush valve switches (at the backflush time). On the pre-column, the flow inverts and all compounds left on the pre-column now backflush to the vent. On the analytical column the separation continues because there the flow is not inverted. See Figure 21 .
Restriction
System pressure
Pressure regulator
Pre-column
Analytical column
Pressure point
Detector
Figure 21 Backflush flows
Injector
Backflush vent
58
The standby mode is the backflush configuration (if the instrument is equipped with the optional backflush valve).
Backflushing saves the time required to elute high boiling components that are not of interest and ensures that the pre-column will be in good condition for the next run.
490 Micro GC User Manual
GC Channels
4
Tuning
Use trial and error to tune the backflush time. Reduce the backflush time to transfer fewer compounds to the analytical column.
1
2
3
Obtain a chromatogram in normal mode. Check out the retention times of the compounds you have to quantify.
Set the backflush time at the retention time of the last peak of interest.
Obtain a second chromatogram.
4
Adjust the backflush time (increments of 0.1 seconds are possible) until all components of interest are transferred to the analytical column and all unwanted peaks are backflushed.
Because a small pre-column is used, it is not always possible to cut between two adjacent peaks.
The
Backflush Time range is from 0.5 seconds until the maximum run time.
To disable backflush
To disable backflushing, set the
Backflush Time to 0. This puts the system in normal mode during the entire run.
490 Micro GC User Manual 59
4
GC Channels
TCD Detector
Each GC channel is equipped with a thermal conductivity detector (TCD). This detector responds to the difference in thermal conductivity between a reference cell (carrier gas only) and a measurement cell (carrier gas containing sample components). The construction of a TCD is such that the changing thermal conductivity of the carrier gas stream, due to components present, is compared to the thermal conductivity of a constant reference gas stream.
60 490 Micro GC User Manual
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 3 Contents
- 7 Introduction
- 8 Safety Information
- 8 Important safety warnings
- 8 Hydrogen safety
- 9 Safety symbols
- 9 Safety and regulatory information
- 10 General safety precautions
- 13 Shipping Instructions
- 13 Cleaning
- 13 Instrument Disposal
- 15 Instrument Overview
- 16 Front View
- 17 Back View
- 18 Inside View
- 21 Carrier Gas Connection
- 23 Sample Gas
- 23 Handling a sample
- 23 Using the external filter unit
- 24 Heated sample lines
- 25 Connecting to a heated sample line
- 26 Power
- 26 Power source
- 26 Power Requirements
- 26 Disposal
- 27 Specifications
- 27 Ambient Pressure
- 27 Ambient Temperature
- 27 Maximum Operation Altitude
- 29 Installation and Use
- 30 Pre-Installation Requirements
- 30 Inspect the Shipping Packages
- 31 Unpack the Micro GC
- 32 Review the Packing List
- 33 Install the 490 Micro GC
- 33 Step 1: Install gas regulators and set pressures
- 33 Step 2: Connect carrier gas
- 33 Step 3: Connect to the checkout sample
- 33 Step 4: Connect to power
- 34 Step 5: Connect to the data system computer (or LAN)
- 34 Step 6a: Assign an IP address - for a Micro GC with main board CP740010
- 37 Step 6b: Assign an IP address - for a Micro GC with main board G3581-65000
- 40 Step 7: Note the Micro GC startup cycle
- 40 Step 8: Complete Micro GC configuration in the data system
- 41 Create the Test Method
- 42 Perform a Series of Runs
- 43 Shut Down Procedure
- 43 Long Storage Recovery Procedure
- 45 GC Channels
- 46 Carrier Gas
- 47 Micro Electronic Gas Control (EGC)
- 47 Injector
- 48 Column
- 49 Molsieve 5Å columns
- 50 CP-Sil 5 CB columns
- 51 CP-Sil CB columns
- 52 PoraPlot 10m column
- 53 Hayesep A 40 cm heated column
- 54 COX and AL203/KCI columns
- 55 MES (NGA) and CP-WAX 52 CB columns
- 56 Column conditioning
- 57 Backflush Option
- 59 Tuning
- 59 To disable backflush
- 60 TCD Detector
- 61 Communications
- 62 Access the Connection Ports
- 63 490 Chromatography Data Systems
- 64 Ethernet Networks
- 65 IP Addresses
- 66 Example network configurations
- 69 To restore the factory default IP address
- 70 To change the Micro GC network settings
- 72 Frequently Asked Questions (FAQ)
- 72 Glossary of network terms
- 74 External Digital I/O
- 75 External Analog I/O
- 76 Micro GC Cycle with Constant Pressure
- 77 Micro GC Cycle with Ramped Pressure
- 79 Errors
- 80 Error Handling
- 81 Error List