When the battery does not need charging then be sure to detach the power supply from the power supply socket on the back of the device and remove the charger from the mains supply.
2.5.5 Initial settings
WARNING
The sections of this Manual contained in a frame correspond to the wording shown on the screen of the device. spirolab
III allows for the personalised setting of certain parameters through the
Configuration Menu.
To access the configuration menu, with the device switched on, press contains the following list:
Delete data in memory
Print last calibration
Turbine calibration
Printout header text
Change Date/Time
Choose operating language
Choose predicted values
Setup parameters/printout
Bluetooth settings
Oximetry settings(*)
Incentive
Turbine type
Standard
Date format
Units format
Auto power-off
which
* function available only for spirolab III with colour display
Select the required option using alongside your selection; then press
or , until the symbol on the left of the screen is
to access the option.
Use this key to recall the Configuration Menu, to set-up and/or to change certain main functions of the device.
Delete data in memory
To delete all the data in memory.
The display will show:
DELETE DATA IN MEMORY
WARNING! ALL SPIROMETRY TESTS IN
MEMORY WILL BE DELETED.
Use test data.
to quit without deleting the
The password is
: 122333
TO PROCEED INSERT PASSWORD
( )
spirolab
III – User Manual Code MIR 980067 REV 2.0 Page 29 / 74
After deleting the data in memory, the display shows:
If the password is entered incorrectly, the display shows:
Test data has been DELETED!
Available memory is: 100%
PRESS ANY KEY TO EXIT
PASSWORD INCORRECT
ENTER to Retry
ESC to Quit
Press to repeat the procedure.
Print last calibration
To print the turbine calibration coefficient currently in use plus the date of the last calibration made.
Turbine calibration
WARNING
The turbine flow sensor does not require calibration but needs only regular cleaning. If a calibration must be made then the following guidelines must be carefully noted.
Only the reusable turbine can be calibrated.
NOTE
Each time a calibration is made, the new correction factor is algebraically added to the previous correction factors. Therefore, before making a new calibration, make sure to delete the actual calibration in use as described above.
For an accurate and reliable calibration the syringe volume must be at least 3 L.
spirolab
III – User Manual Code MIR 980067 REV 2.0 Page 30 / 74
To modify the calibration of the turbine based on the
FVC values (for the expiratory phase) and FIVC (for the inspiratory phase), measured during a test made with a calibration syringe.
From configuration menu choosing “turbine calibration”, on the display appears the following informations.
The values “FVC” and “FIVC” are retailed to the last calibration. The values under the “%COR” column indicate the actual correction. They are set on 0 for default.
WARNING
ATP stands for Ambient Temperature and Pressure which indicates that the measurement conditions are ambient temperature and pressure.
BTPS stands for Body Temperature and Pressure Saturated which represents a world standard reference condition for the measurement of spirometric parameters. All spirometric parameters are calculated at body temperature (37
°C) and pressure saturated.
In line with the publication "STANDARDISED LUNG FUNCTION TESTING" of the
European Respiratory Society (Vol 6, Supplement 16, March 1993), the air expired from the mouth is at a temperature of circa 33/34 °C.
The expired flow and volume, to be converted back to BTPS conditions (37 °C), must be increased by 2.6% - this is derived from the BTPS factor of 1.026 at a temperature of 33°C, which represents a correction of 2.6%. In practice the BTPS factor for the expired flow and volumes is therefore constant and equal to 1.026.
For the inspired volumes and flows, the BTPS factor depends instead upon the ambient temperature, as the inspired air is at ambient temperature.
For instance, at an ambient temperature of 20°C with relative humidity at 50%, the
BTPS factor is 1.102, a correction of +10.2%.
The correction of the inspired volumes and flows is made automatically, as the device has an internal temperature sensor; all parameters are thus reported at
BTPS.
To make the calibration:
1 - Insert the volume in cL of the calibration syringe in use (e.g. for a 3L syringe, insert 300 cL).
2 - Insert both the FVC and FIVC values, obtained by the measurement made with the calibration syringe, in the New FVC and New FIVC field. Use vertical scrolls to select the value to be modified (SIRIN, New FVC, New FIVC).
Use numeric keys to modify the parameter value selected.
If the calculated correction factors are acceptable (<20%), they are displayed beside the
New FVC and New FIVC parameters.
spirolab
III – User Manual Code MIR 980067 REV 2.0 Page 31 / 74
Press ESC to return to the configuration Menu without entering the correction.
If the FVC and FIVC values produce a correction factor that is >20%, the FVC and FIVC values will not be accepted. This means that the system cannot correct for such a large calibration error. The following message appears:
WARNING ! ERROR too high
In this case:
- Check the correct functioning of the spirolab III with a new turbine and/or
- Clean the turbine.
To erase the calibration in use and reset to the factory calibration, input 0 in the New
FVC and New FIVC fields.
Then press to return to the Configuration Menu.
WARNING
With the new calibration, the correction factors are algebraically added to the previous correction factor(s).
Before to begin a new calibration set the two values FVC and FIVC at 0, then repeat the calibration procedure.
If a 3L syringe is used to make the calibration and if the spirometer is calibrated correctly then the FVC (syringe) value will be:
3.00 (FVC) x 1.026 (BTPS) = 3.08 L (FVC at BTPS).
If the ambient temperature is 20°C, the FIVC (syringe) value will be:
3.00 (FIVC) x 1.102 (BTPS) = 3.31 L (FIVC at BTPS).
The user must be aware that the volume of the syringe shown by the machine is converted to BTPS conditions, so that the "increase" of the results with respect to the expected values does not constitute an error.
For instance, if the calibration procedure is carried out with measured data:
20 ° C, FVC = 3.08, FIVC = 3.30 the spirometer is perfectly calibrated and the relative correction factors are null.
No calibration is required!
This does not represent an error, but is a logical consequence of the above detailed explanation.
spirolab
III – User Manual Code MIR 980067 REV 2.0 Page 32 / 74
Printer header text
To insert a header that will be printed at the beginning of each spirometry report (see image).
Change DATE/TIME
To change the date and time.
The time is shown in the 24 hour format.
Select language
To change the language used for displaying messages on the screen and the printouts
.
Select predicted values
To select one of the standards available for the calculation of the predicted values. The parameters are organized in the following way:
REFERENCE
1
2
3
ADULTS
ERS
Knudson
Morris/Bass
4
5
ERS
MC-Barcelona
Zapletal
Zapletal
Select with the numeric keys the reference number, automatically the chosen values will set and the display shows the configuration menu.
CHILDREN
Knudson
Knudson
Knudson
Select printout
To enable or disable the printout of: spirometry parameters best values of: FVC, FEV1, FEV1/FVC, PEF, FEF2575 (the last one is always relative to the best test) spirometry interpretation
F/V and V/t curves.
Bluetooth settings
Select this function and the following menu is shown:
Search Device
Options PRINTER
Options PHONE
Remote line Number
Search Device
Select the required option with and confirm with ; spirolab III begins to search for
Bluetooth active devices; when one or more active devices are found then a list is shown, select a device and push to define the device (with vertical scroll) as a printer or as a phone (use PRINTER or PHONE); select one of the two options and push , otherwise
spirolab
III – User Manual Code MIR 980067 REV 2.0 Page 33 / 74
push to return to the Bluetooth options. If the device is already present in the chosen list a warning message will be shown on the display; push a key to turn to the devices list.
Options PRINTER
To control the devices memorized within the “printers list”. Enter the list with having selected a device it is possible:
and
to set the device as default, identified with a yellow point near the device name (so
spirolab
III will automatically connect to this)
to delete the device from the list (spirolab III asks for a confirmation by pressing , otherwise press the list).
to return to the Bluetooth options and not delete the device from
If no devices are shown then a message advises the user to search for active devices; press to start the search or to return to the Bluetooth settings.
Options PHONE
To control the devices memorized in the “printers list”. Enter the list with selected a device it is possible:
and having
to set the device as default, identified with a yellow point near the device name (and
spirolab
III will automatically connect to this)
to delete the device from the list (spirolab III asks for a confirmation by pressing , otherwise press the list).
to return to the Bluetooth options and not delete the device from
Even if the list is empty, a search can be made for active devices as previously described.
Remote line Number
Enter the menu with then it is possible to insert the telephone number that spirolab III will use to connect to a Bluetooth telephone. Insert the required number using the numeric keyboard and then press .
Oximetry settings
This function permits the following kinds of setting up :
reference values for Sp02 and BPM; if, during a test, these values go below the minimum or above the maximum values in the set up an acoustic alarm will ring to alert the user
activation/disactivation of the acoustic alarm
set the sampling time (2 or 4 seconds) of oximetry values during a test
The cursor at the beginning will rest on the minimum Sp02 value, use the numeric keys to select the desired value, then the cursor automatically pass to the next value; if a value
spirolab
III – User Manual Code MIR 980067 REV 2.0 Page 34 / 74
must not be modified press ; once the four values have been set up you can then, if required, activate or disactivate the acoustic alarm by pressing key 1 for OFF and key 2 for
ON. To confirm the selection press . Then is possible to set the sampling time: press key
2 to set 2 sec or press key 4 to set 4 sec. Pressing the configuration menu is displayed.
Incentive
spirolab
III has a pediatric incentive program which can be useful to help a child (or an older person) make a correct spirometry test. The incentive is a hidden image, directly connected to the signal coming from the MiniFlowmeter so the harder the expiration then the further the “curtain” moves away, revealing the image and thus helping the patient to obtain the best possible result.
When the pediatric incentive is selected, a warning message is shown asking for the age of the patient. Set the correct value, the configuration menu is then shown.
Given that the FVC test requires collaboration that is difficult to obtain in children under the age of 4, then if an age of 4 or below is selected then spirolab III automatically shows the incentive during the FVC test.
Turbine setup
Select the type of turbine to be used for testing, either single-patient disposable or reusable. To select the correct option, follow the steps as described previously selecting the required item and press v to change the option.
Standard
To select one of the available standards: ATS/ERS or NANHES III, press .
WARNING
With the standard NAHNES III is not possible to select or to modify the predicted values.
DATE format: dd/mm/yy
To select the required format, toggles between dd/mm/yy or mm/dd/yy or yy/mm/dd or vice versa.
Press to toggle.
UNITS format: cm, kg
To change the units format from cm, kg to in, lb (inches and pounds) or vice versa.
Press to toggle.
Auto power-off: 6 min
To change the wait-time for auto power off to 6 min, 60 min or after 240 min.
spirolab
III – User Manual Code MIR 980067 REV 2.0 Page 35 / 74