advertisement
GasAlertMicro
Users Manual
Maintenance
To keep the detector in good operating condition, perform the following basic maintenance as required:
• Calibrate, test and inspect the detector at regular intervals.
• Keep an Operations Log of all maintenance, calibrations and alarm events.
• Clean the exterior with a soft damp cloth. Do not use solvents, soaps or polishes.
• Do not immerse the detector in liquids.
Replacing the Batteries
a
Warning
To avoid personal injury:
⇒ Replace the batteries as soon as the detector emits
a Low Battery Alarm.
⇒ Use only batteries recommended by
BW Technologies to prevent damage or personal injury.
⇒ Use only approved AA alkaline or charged AA NiMH
batteries, properly installed in the detector case.
See page 49 for approved batteries.
⇒ Charge batteries using only a recommended
charger. Do not use any other charger. Failure to observe this precaution could lead to fire or explosion.
⇒ Do not change or charge batteries in a hazardous
location. Doing so will impair the intrinsic safety of the unit, and may lead to fire or explosion.
40
Table 11 and Figure 4 illustrate how to replace the batteries.
If the detector is on, turn off the detector before replacing the batteries. Press the two release buttons on the detector. The battery tray pulls out and snaps in.
Remove the batteries from the battery drawer and replace according to the polarity markings. Close drawer until you hear both sides click, which locks the battery drawer in place.
Table 11. Replacing the Batteries
Item Description
A
B
C
Detector Front
Release Buttons
Batteries
To preserve battery life, turn off the detector when you are not using it.
GasAlertMicro
Maintenance
Figure 4. Replacing the Batteries
41
GasAlertMicro
Users Manual
Replacing a Sensor or Sensor Filter
a
Warning
To avoid personal injury, use only sensors specifically designed for the detector. See the section
“Replacement Parts and Accessories.”
Each sensor has a high degree of resistance to common vapors and gases. A sensor will most likely clear itself if you remove the detector to a clean environment and wait 10 to
30 minutes. Do not expose a sensor to the vapors of inorganic solvents (such as paint fumes) or organic solvents.
The section “If the Detector Does Not Work” describes problems caused by a sensor in need of calibration or replacement.
Gently rocking the sensor back and forth may help free a tightly held sensor. Insert new sensor on the PCB, ensuring that sensor posts are aligned correctly. Two-gas units require a dummy sensor.
Calibrate the detector after changing any sensor.
Table 12. Replacing a Sensor or Sensor Filter
Item Description
A
B
C
Detector Screws
Sensors
Sensor Filter
Figure 5. Replacing a Sensor or Sensor Filter
42
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Related manuals
advertisement
Table of contents
- 8 GasAlertMicro Multi-Gas Detector
- 8 GasAlertMicro with User Downloadable Datalogger
- 9 Introduction
- 10 Contacting BW Technologies
- 10 Safety Information - Read First
- 14 Getting Started
- 18 Activating the Detector
- 18 Self-Test
- 19 Sensor Test
- 19 Calibration Status
- 20 Battery Test
- 20 Datalogger Operation
- 20 Self-Test Pass
- 20 Self-Test Fail
- 21 Deactivating the Detector
- 21 Confidence Beep
- 22 User Option Menu
- 22 Latched Alarms Function
- 23 “SAFE” Display Enable or Disable
- 24 Combustible Sensor Measuring Selection
- 25 Sensor Enable or Disable
- 25 Disabling/Enabling a Sensor
- 26 Pass Code Protection
- 27 Adjust Datalogger Sampling Rate
- 27 Change the Date and Time
- 28 Alarms
- 32 Gas Exposures Computed
- 33 Viewing Gas Exposures
- 34 Gas Alarm Setpoints
- 35 Resetting Gas Alarm Setpoints
- 35 Stopping a Gas Alarm
- 36 Sensor Alarm
- 36 Low Battery Alarm
- 36 Automatic Shutdown Alarm
- 37 Calibration and Setting Alarm Setpoints
- 37 Guidelines
- 38 Diagnostics Protection
- 38 Applying Gas to the Sensors
- 39 Calibration Procedure
- 39 Start Calibration
- 39 Auto Zero and Oxygen Sensor Calibration
- 40 Pass Code Protect Activated
- 41 Auto Span
- 42 Alarm Setpoints
- 43 Setting the TWA Alarm Setpoint
- 43 Setting the STEL Alarm Setpoint
- 44 Setting the LOW Alarm Setpoint
- 44 Setting the HIGH Alarm Setpoint
- 45 Setting the Remaining Alarm Setpoints
- 45 Setting the Calibration Due Date
- 45 Verification (optional)
- 47 Instrument Maintenance
- 48 Maintenance
- 48 Replacing the Batteries
- 50 Replacing a Sensor or Sensor Filter
- 51 Installing the MultiMediaCard (MMC)
- 51 If the Detector Does Not Work
- 55 Replacement Parts and Accessories
- 56 Specifications
- 59 Appendix A
- 60 Excel Datalog Manager (E.D.M.)
- 60 Excel Datalog Manager (E.D.M.) Software Plug-in
- 60 Sample Data Files
- 61 Installing Excel Datalog Manager (E.D.M.):
- 61 MMC Card Reader Installation
- 61 Types of MultiMediaCard Readers
- 62 Start Excel Datalog Manager
- 62 Import Your Data File.
- 64 Direct Import to Compatible Programs
- 64 Importing the Data File Into Compatible Desktop Applications
- 64 Determining Application Compatibility
- 68 MultiMediaCard Compatibility
- 69 Troubleshooting
- 69 Recovering Data Files