Spitfire

Add to My manuals
116 Pages

advertisement

Spitfire | Manualzz

Maintenance and Troubleshooting

Symptom: Poor contrast ratio

Possible Cause:

Pre-pulse:

Incorrect alignment of the output

Pockels cell

Incorrect alignment of the ¼ wave plate for the input Pockels cell

Post-pulse:

Incorrect alignment of the Input

Pockels cell

Incorrect adjustment of the ¼ wave voltage for Input Pockels cell

Corrective Action:

Contact your Spectra-Physics representative

Contact your Spectra-Physics representative

Contact your Spectra-Physics representative

Contact your Spectra-Physics representative

Symptom: Poor output beam quality

Possible Cause:

Incorrect pump beam alignment

Corrective Action:

Refer to Appendix C for pump beam alignment procedures.

Damage to optical components Check for optical damage; contact your Spectra-Physics representative if present.

Compressor vertical retro-reflector and/or horizontal retro-reflector are incorrectly aligned in the horizontal axis

Contact your Spectra-Physics representative.

Symptom: Optical damage in the amplifier cavity

Possible Cause:

Seed laser not well modelocked (CW breakthrough)

Partial restriction of the stretched spectrum

Failure to remove alignment tools from optical path after checking stretcher or compressor alignment

Incorrect alignment of amplifier cavity

Corrective Action:

Contact your Spectra-Physics representative.

Contact your Spectra-Physics representative.

Contact your Spectra-Physics representative.

Contact your Spectra-Physics representative.

8-5

advertisement

Questions & Answers

2024-05-29

V V M

What are the pulse widths for the different Spitfire amplifier models?
The Spitfire F produces pulses less than 130 fs, the Spitfire P and Spitfire PM produce pulses less than 2 ps, the Spitfire USF produces pulses less than 90 fs, and the Spitfire 50FS produces pulses less than 50 fs.
What is the Spitfire HP?
The Spitfire HP is a high-power option available with most Spitfire configurations, operating at 1 kHz and adding an extra stage of amplification.
How are different wavelength ranges achieved with the Spitfire models?
Different wavelength ranges are achieved by using different optics sets. For example, Optics Set 1 covers the output wavelength range of 750 nm – 840 nm.