Optimizing B-Mode. GE LOGIQ V2, LOGIQ V1
The LOGIQ V2 and LOGIQ V1 ultrasound systems are designed for a variety of clinical applications. These innovative systems combine advanced imaging technologies with intuitive user interface, making them suitable for a diverse range of medical professionals. They offer a comprehensive set of features to perform diagnostic examinations and generate detailed reports.
Advertisement
Advertisement
Optimizing the Image
Optimizing B-Mode
Intended Uses
B-Mode is intended to provide two-dimensional images and measurement capabilities concerning the anatomical structure of soft tissue.
Figure 5-1. B-Mode Display -- Representative Example
B Mode Primary Menu
Different parameters can be assigned to the Primary Menu, the
Primary Menu can be configured in Utility -> Application ->
Imaging Controls.
5-2 LOGIQ V2/LOGIQ V1 – Basic User Manual
Direction 5610735-100 English Rev. 2
Optimizing B-Mode
Typical B-Mode Exam Protocol
A typical examination using B-Mode might proceed
1. Record exam-related patient information. Verify system setup (probes and presets).
2. Position the patient and the console for optimum operator and patient comfort. Perform the scan.
3. Complete the study by collecting all the data.
CAUTION
Ensure that the hands of the patient are away from the system during the exam.
The position of the operator and the patient vary by scan region.
In most case, the operator sits/stands straight in front of the operator console and the patient lies on the bed on the right (or left) side of the system.
CAUTION
Always use the minimum power required to obtain acceptable images in accordance with applicable guidelines and policies.
LOGIQ V2/LOGIQ V1 – Basic User Manual
Direction 5610735-100 English Rev. 2
5-3
Optimizing the Image
B-Mode Scanning Hints
HINTS
These B-Mode controls produce the following results:
Auto Optimize. Automatically improves the contrast resolution
of the image by changing the gray scale to match the image data. Available in B-Mode and Doppler Mode.
Coded Harmonics. Enhances near field resolution for
improved small parts imaging as well as far field penetration.
Frequency. Changes system parameters to best optimize for a particular patient type.
Gray Map. Affects the presentation of B-Mode information.
Choose the gray map prior to making other adjustments. There is an interdependency between gray maps, gain, and dynamic range. If you change a map, revisit gain and dynamic range settings.
Dynamic Range. Changes the amount of gray scale
information displayed. A higher dynamic range shows more gray scale information displayed, while a lower dynamic range displays less gray scale information onto the same display scale. If you increase the gain, you may want to decrease the
Dynamic Range.
Frame Average. Smooths the image by averaging frames.
Affects the amount of speckle reduction.
TGC. Adjust TGC to adjust Gain in specific areas.
Focus Zones. Focal zones should be placed roughly in the
lower half of the display depth, at or below the organ of interest.
Width. Sizes region of interest. Adjust the Width to the smallest
reasonable size to maximize frame rate.
5-4 LOGIQ V2/LOGIQ V1 – Basic User Manual
Direction 5610735-100 English Rev. 2
Optimizing B-Mode
Depth
Description
Adjusting
Preset
Values
Benefits
Affect on other controls
HINTS
Depth controls the distance over which the B-Mode images anatomy, and the field of view. To visualize deeper structures, increase the depth. To visualize flatter structures, decrease the depth.
Each adjustment cycles you to the next Depth setting. Imaging and display parameters adjust automatically.
To increase/decrease, adjust Depth.
You can set the depth by probe and application on the Utility -->
Imaging page.
Depth increments vary by probe and application. Depth displays on the monitor in centimeters.
Depth values are returned to the factory or user preset value when you change the following: Patient/Preset, Exam Calc, or
End Exam.
Depth adjusts your field of view. It increases your field of view to look at larger or deeper structures; it decreases your field of view to look at structures near the skin line.
After adjusting the depth, you may need to adjust the TGC and focus.
Changing the depth may change the TI and/or MI. Observe the output display for possible effects.
Make sure enough space is left below the anatomy of interest to demonstrate shadowing or enhancement.
LOGIQ V2/LOGIQ V1 – Basic User Manual
Direction 5610735-100 English Rev. 2
5-5
Optimizing the Image
Gain
Description
Adjusting
Preset
Values
B-Mode Gain increases or decreases the amount of echo information displayed in an image. It may have the effect of brightening or darkening the image if sufficient echo information is generated.
To decrease/increase, rotate Gain.
Gain values vary depending on the probe; they are not associated with a particular position of the knob.
NOTE: B-Mode gain is independent of M-Mode and Doppler and Color
Flow Gain. Changing the M-Mode Gain while in M-Mode does not affect the B-Mode image gain.
You can set gain by probe and application on the Utility -->
Imaging page.
Gain displays on the monitor in Gn. Maximum gain varies by probe. Gain values vary by probe, application, and frequency setting.
NOTE: Maximum gain is factory preset to an optimum setting to eliminate noise in the display.
Benefits
Gain values are returned to the factory or user preset value when you change the following: Probe, Exam Category/Exam
Calc, or Patient.
Gain allows you to balance echo contrast so that cystic structures appear echo-free and reflecting tissue fills in.
Affect on other controls
Bioeffects
After you adjust the Power Output, you may need to adjust the gain. Generally speaking, if you increase the Power Output, you need to decrease the gain; if you decrease the Power Output, you need to increase the gain. Gain and TGC interact by adding together.
Gain has no affect on Power Output. However, with increased gain, the power output level can usually be reduced to produce an equivalent image quality.
NOTE: Always optimize gain before increasing the Power Output.
5-6 LOGIQ V2/LOGIQ V1 – Basic User Manual
Direction 5610735-100 English Rev. 2
Optimizing B-Mode
Focus
Description
Adjusting
Values
Increases the number of focal zones, moves the focal zone(s) and changes the zone width so that you can tighten up the beam for a specific area. A graphic corresponding to the focal zone position(s) appears on the right edge of the image.
To increase/decrease the number of focal zones, adjust Focus
Number.
To move the focal zone to the near/far field, adjust Focus
Position.
NOTE: You can set the focus (Depth % and Number, and Number
CrossXbeam) by probe and application on the Utility--> Imaging page.
Focus zone number and position vary depending on the depth, zoom, probe, application, and frequency setting selected.
Benefits
Affect on other controls
Bioeffects
Focal zone number values are returned to the factory or user preset value when you change the following: Probe, Exam
Category, Exam Calc, or Patient.
Focus optimizes the image by increasing the resolution for a specific area.
Changing the focal number affects the frame rate. The greater number of focal zones, the slower the frame rate.
Changing the focal zone may change the TI and/or MI. Observe the output display for possible effects.
LOGIQ V2/LOGIQ V1 – Basic User Manual
Direction 5610735-100 English Rev. 2
5-7
Optimizing the Image
Auto Optimize
Description
Benefit
Adjusting
Preset
Values
Affect on other controls
Auto Optimize (Auto) lets you optimize the image based upon the actual B Mode image data (Auto Tissue Optimize, ATO). The preset levels (Low, Medium, and High) allow you to pick a preference for the contrast enhancement in the resulting image.
Low does the least amount of contrast enhancement, high does the most.
Auto is available in single or multi image, on live, frozen or CINE images (in B-Mode only), and while in zoom and in Spectral
Doppler.
Auto in PW Doppler Mode (ASO: Auto Spectral Optimization) optimizes the spectral data. Auto adjusts the Velocity Scale/PRF
(live imaging only), baseline shift, dynamic range, and invert (if preset). Upon deactivation, the spectrum is still optimized.
Auto can be found in reduced optimization time and a more consistent and accurate optimization process.
To activate Auto, press AO key.
Press AO key again to turn off Auto.
Specify the ATO Level: Low, Medium, or High via Utility -->
Imaging --> B-Mode.
Specify the Auto Invert on ASO for Doppler mode: On or Off via
Utility --> System -> System Imaging.
Auto is active until you deactivate it or when you change the following: Probe, Exam Category, Exam Calc, or Patient.
You may need to adjust the Gain.
5-8 LOGIQ V2/LOGIQ V1 – Basic User Manual
Direction 5610735-100 English Rev. 2
Optimizing B-Mode
CrossXBeam
Description
Adjusting
Preset
CrossXBeam is the process of combining three or more frames from different steering angles into a single frame. CrossXBeam is available on Convex and Linear probes.
CrossXBeam combines multiple co-planar images from different view angles into a single image at real-time frame rates, using bi-cubic interpolation.
To activate CrossXBeam, select CrossXBeam.
To adjust the number of frames being compounded, adjust
CrossXBeam.
Select CrossXBeam Type (Off, Low, Medium or High) for
Convex probe. Select CrossXBeam Type (Off, Low, Medium,
High or Max) for Linear probe. Max detects maximum values;
Mean detects averaged values; Hybrid detects a mix of both
average and maximum values.
You can preset B-Mode CrossXBeam:
• Preset the CrossXBeam #
• Preset Focus Number for CrossXBeam
• Preset Line Density CrossXBeam
• Preset Focus Width for CrossXBeam
• Frame Average CrossXBeam
These presets can be made via the Utility --> Imaging page.
LOGIQ V2/LOGIQ V1 – Basic User Manual
Direction 5610735-100 English Rev. 2
5-9
Optimizing the Image
CrossXBeam (continued)
Values
CrossXBeam is available on Convex and Linear probes. Multiple focal zones are supported. B-Mode CrossXBeam is available while in B-Mode, Color Flow, or PW Doppler Mode. Steering is optimized by probe. The displayed compound image depth is equal to the image depth of the non-steered frame.
The following features are supported in CrossXBeam:
Benefits
• CINE Loops acquired using CrossXBeam.
• All Measurement and Analysis packages
• SRI HD, Rejection, Colorize, Revert, Edge Enhance, Gray
Map, Rotation, Biopsy Kit, Dynamic Range, Depth, TGC,
Gain, Acoustic Output, Auto, Virtual Convex, LOGIQ View and Easy 3D.
The combined single image has the benefits of reduced speckle noise, reduced clutter, and continuity of specular reflectors.
Therefore, this technique can improve contrast resolution with increased conspicuity of low contrast lesions, better detection of calcifications, biopsy needle visualization, and cystic boundary definition.
5-10 LOGIQ V2/LOGIQ V1 – Basic User Manual
Direction 5610735-100 English Rev. 2
Optimizing B-Mode
SRI-HD (High Detection Speckle Reduction Imaging)
Description
SRI-HD (High Detection Speckle Reduction Imaging) is an adaptive algorithm to reduce the unwanted effects of speckle in the ultrasound image. Image speckle usually appears as a grainy texture in otherwise uniform areas of tissue. Its appearance is related to image system characteristics, rather than tissue characteristics, so that changes in system settings, such as probe type, frequency, depth, and others, can change the appearance of the speckle. Too much speckle can impair image quality and make it difficult to see the desired detail in the image. Likewise, too much filtering of speckle can mask or obscure desired image detail. Extra care must be taken to select the optimal SRI-HD level.
SRI-HD is available in B-Mode imaging and may be used with any transducer or clinical application when image speckle appears to interfere with the desired image detail.
Figure 5-2. B-Mode Image before SRI-HD (left) and
After SRI-HD (right)
LOGIQ V2/LOGIQ V1 – Basic User Manual
Direction 5610735-100 English Rev. 2
5-11
Optimizing the Image
SRI-HD (High Detection Speckle Reduction Imaging) (continued)
Values
Benefits
We recommend that you select the SRI-HD level by observing the enhanced image in side-by-side dual image comparison with the original, unprocessed image. Dual display mode is activated by pressing the L and R keys simultaneously.
In selecting the level of SRI-HD, you must observe the effects of
SRI-HD in the desired region of interest and should make a real-time comparison with the original image. The optimal level depends on the clinical situation and improves with experience.
Observing the original and SRI-HD-processed images together helps to determine whether too much or too little SRI-HD has been applied.
Dual image mode for SRI-HD can also be activated on a stored
CINE Loop. This allows you to always see the original, unprocessed or enhanced image by going into the Dual display mode and to change the SRI-HD settings when reviewing the
CINE Loop.
SRI-HD is available on 3D.
• You cannot change SRI-HD after the scan starts.
• The effects for the rendered image are less than the
2D-image.
Smooths the image when image speckle interferes with the desired image detail.
5-12 LOGIQ V2/LOGIQ V1 – Basic User Manual
Direction 5610735-100 English Rev. 2
Optimizing B-Mode
Coded Harmonic Imaging (CHI)
Description
Adjusting
Values
Benefits
Bioeffects
Harmonic imaging utilizes Digitally Encoded Ultrasound (DEU).
Coded Harmonics enhances near field resolution for improved small parts imaging as well as far field penetration.
To activate Coded Harmonic imaging, press CHI key on the control panel.
NOTE: You can assign CHI On/Off function to the User Defined key.
See ‘System/User Configurable Key’ on page 16-35 for more information.
NOTE: You can specify the parameters for CHI in Utility -> Imaging ->
HAR.
On/Off. ’CHI’ appears in place of ‘B’ in the imaging parameters.
Coded Harmonics diminishes low frequency high amplitude noise and improves imaging technically difficult patients. Coded
Harmonics may be especially beneficial when imaging isoechoic lesions in shallow-depth anatomy in the breast, liver, and hard-to-visualize fetal anatomy.
Activating CHI mode may change the TI and/or MI. Observe the output display for possible effects.
Frequency
Description
Adjusting
Multi Frequency mode lets you downshift to the probe's next lower frequency or shift up to a higher frequency.
Adjust Frequency until the desired frequency is selected.
The selected frequency appears as “Frq” in the upper, right-hand portion of the monitor display.
NOTE: Frequency change is not active when the image is frozen.
Values
NOTE: Changing frequency resets those parameters which are presettable by frequency to their preset values for the current frequency.
Vary, depending on the probe and application.
Benefits
Bioeffects
Frequency values are returned to the factory or user preset value when you change the following: Probe, Exam Category,
Exam Calcs, or begin a new patient.
This optimizes the probe's wide band imaging capabilities at multiple frequencies to image at greater depths.
Adjusting frequency may change the TI and/or MI. Observe the output display for possible effects.
LOGIQ V2/LOGIQ V1 – Basic User Manual
Direction 5610735-100 English Rev. 2
5-13
Optimizing the Image
B Steer
Description
Adjusting
Values
Preset
Benefits
Bioeffects
You can slant the B-Mode or Color Flow acoustic beam without moving the probe. The steer function only applies to linear probes. The steer function is only available in CrossXBeam Off.
To slant the linear image to the left/right, adjust B Steer.
Linear probes can be steered left, center, or right up to a maximum of 15 degrees, depending on the probe.
Steer values are returned to factory or user preset value when you change: Probe, Exam Category, Exam Calcs, or Patient.
Preset the steer angle via Utility --> Imaging --> B --> B Steer.
Slant the linear image left or right to get more information without moving the probe.
Activating steer may change the TI and/or MI. Observe the output display for possible effects.
M/D Cursor
Description
Adjusting
Benefits
Displays the M/D-Mode cursor on the B-Mode image.
To activate/deactivate the M/D-Mode cursor, press M/D Cursor key. Trackball to position M/D-Mode cursor. Adjust Angle and
SV Length as necessary.
Lets you position the cursor before you go into M-Mode or
Doppler Mode so that you can make optimum use of the larger
B-Mode image.
Virtual Convex
Description
Adjusting
Values
Benefits
Bioeffects
On Linear and Sector probes, Virtual Convex provides a larger field of view in the far field.
To activate/deactivate Virtual Convex, select Virtual Convex.
On/Off
Virtual Convex allows for a wider field of view. Available in
B-Mode, Color Flow Mode, and Doppler Mode. CrossXBeam is available on Virtual Convex with linear probes.
Activating Virtual Convex may change the TI and/or MI. Observe the output display for possible effects.
5-14 LOGIQ V2/LOGIQ V1 – Basic User Manual
Direction 5610735-100 English Rev. 2
Optimizing B-Mode
TGC
Description
Adjusting
Values
Preset
Benefits
CAUTION
TGC amplifies returning signals to correct for the attenuation caused by tissues at increasing depths. TGC slide pots are spaced proportionately to the depth. The area each pot amplifies varies as well. A TGC curve may appear on the display
(if preset), matching the controls that you have set (except during zoom). You can choose to deactivate the TGC curve on the image.
To decrease/increase TGC, move slide pot to the left/right.
When you change the depth, TGC is rescaled across the new depth range. Each pot is proportionately scaled across the depth.
TGC Display On/Off -- preset via Utility --> System --> System
Imaging.
TGC balances the image so that the density of echoes is the same throughout the image.
Do not apply too much power to the TGC slide pots as this could damage the slide pots.
Scan Area
Description
Adjusting
Values
Benefits
Affect on other controls
Bioeffects
You can widen or narrow the size of the sector angle to maximize the image's region of interest (ROI).
To narrow/widen the angle, press Scan Area until the Width is highlighted, then move the Trackball to move left/right to decrease/increase the angle size. Then press Scan Area to set the ROI.
Varies, depending upon the probe (not applicable to linear probes) and application.
Increase the sector angle to see a wide field of view; decrease the sector angle when you need to have a faster frame rate, as in fetal heart.
Changing the sector angle affects the frame rate. The narrower the sector angle, the faster the frame rate.
Changing the sector angle may change the TI and/or MI.
Observe the output display for possible effects.
LOGIQ V2/LOGIQ V1 – Basic User Manual
Direction 5610735-100 English Rev. 2
5-15
Optimizing the Image
Tilt
Description
Adjusting
Values
Benefits
Bioeffects
You can steer the sector angle to get more information without moving the probe while in B-Mode, M-Mode, Doppler Mode, and
Color Flow Mode. Tilt is not available on Linear probes. Tilt disappears when CrossXBeam is activated.
To tilt the angle to the left/right, press Scan Area until Tilt is highlighted, then move the Trackball to the left/right.
Varies, depending on the probe.
Allows you to move a reduced sector angle laterally, without moving the probe. Beneficial in GYN.
Steering the sector angle may change the TI and/or MI. Observe the output display for possible effects.
Dynamic Range
Description
Adjusting
Values
Benefits
Affect on other controls
Dynamic Range controls how echo intensities are converted to shades of gray, thereby increasing the adjustable range of contrast.
To increase/decrease, adjust Dynamic Range.
Dynamic Range values vary by probe, application, and frequency setting.
Dynamic Range levels are returned to the factory or user preset value when you change the following: Patient/Preset, Exam
Calcs, Probe, or Multi Frequency.
Dynamic Range is useful for optimizing tissue texture for different anatomy. Dynamic Range should be adjusted so that the highest amplitude edges appear as white while lowest levels
(such as blood) are just visible.
Dynamic range operates in real-time, Freeze and CINE. It also affects Gain.
5-16 LOGIQ V2/LOGIQ V1 – Basic User Manual
Direction 5610735-100 English Rev. 2
Optimizing B-Mode
Revert
Description
Adjusting
Values
Benefits
Flips the image 180 degrees left/right.
To flip the image 180 degrees, select Revert.
The image rotates in 180 degrees left/right. Revert settings vary by probe and application.
Revert settings are returned to the factory or user preset value when you change the following: Probe, Exam Category, Exam
Calcs, or Patient.
Used for anatomical correctness.
CAUTION
When reading a reverse image, be careful to observe the probe orientation to avoid possible confusion over scan direction or left/right image reversal.
NOTE: The Image can be easily identified as being reversed as the
LOGIQ V logo will be on the opposite side of the image monitor.
LOGIQ V2/LOGIQ V1 – Basic User Manual
Direction 5610735-100 English Rev. 2
5-17
Optimizing the Image
Line Density
Description
Adjusting
Values
Optimizes B-Mode frame rate or spatial resolution for the best possible image.
Adjust Line Density to increase resolution or to increase frame rate.
Select the default value via Utility -> Imaging -> B -> Line
Density and press Save.
Varies by probe.
NOTE: Not available in cine mode.
Benefits
Values vary by probe and application. Line Density values are returned to the factory or user preset value when you change the following: Probe, Exam Category, Exam Calcs, or Patient.
A lower line density is useful in fetal heart beat, adult cardiac applications and in clinical Radiology applications requiring significantly higher frame rates.
A higher line density is useful in obtaining very high resolution, e.g., thyroid, testicles.
Line density changes the vector density and frame rate.
Affect on other controls
Bioeffects
Activating color flow line density may change the TI and/or MI.
Observe the output display for possible effects.
Line Density Zoom
Description
Adjusting
You can set the default value for Line Density in zoom independently.
Select the default value via Utility -> Imaging -> B -> Line
Density Zoom and press Save.
5-18 LOGIQ V2/LOGIQ V1 – Basic User Manual
Direction 5610735-100 English Rev. 2
Optimizing B-Mode
Colorize
Description
Colorize is the colorization of a conventional B-Mode image or
Doppler Spectrum to enhance the user's ability to discern B, M, and Doppler Mode intensity valuations. Colorize is NOT a
Doppler Mode.
NOTE: You can colorize real-time or CINE images.
Colorizes the gray scale image to enhance the eye's discrimination capability.
Adjusting
Spectrum Colorize colorizes the spectrum as a function of power using the inverse of the Colorize map for the signal intensity in each Doppler line.
Colorize enhances the visibility of the spectrum's characteristics and enhances your ability to identify spectral broadening and the edge contours of the spectrum used to define the peak frequency/velocity.
The gray bar displays while Colorize is activated.
To activate Colorize,
2. Trackball to cycle through available maps.
To deselect, select a gray map.
LOGIQ V2/LOGIQ V1 – Basic User Manual
Direction 5610735-100 English Rev. 2
5-19
Optimizing the Image
PRF
Description
Reduces noise artifacts in the image. When you activate PRF, the frame rate decreases and the noise artifacts are filtered.
NOTE: Available only Cardiac application.
Edge Enhance
Description
Adjusting
Values
Benefits
Edge Enhance brings out subtle tissue differences and boundaries by enhancing the gray scale differences corresponding to the edges of structures. Adjustments to
M-Mode's edge enhancement affects the M-Mode only.
To cycle through settings, adjust Edge Enhance.
Values vary by probe, application and multi frequency setting.
Edge Enhance values are returned to the factory or user preset value when you change the following: Probe, Exam Category,
Exam Calcs, or Patient.
Edge Enhance cleans out the B-Mode image/M-Mode time line by subduing some of the gray scale in order to highlight the vessel wall or organ. This is helpful when you cannot differentiate between the chambers of the heart.
Edge Enhance operates in real-time only; not in Freeze or CINE.
Affect on other controls
Frame Average
Description
Adjusting
Values
Benefits
Temporal filter that averages frames together, thereby using more pixels to make up one image. This has the effect of presenting a smoother, softer image.
To adjust frame averaging, adjust Frame Average.
Values vary by probe, application and multi frequency setting.
Frame Average values are returned to the factory or user preset value when you change the following: Probe, Exam Category,
Exam Calcs, or Patient.
Smooths the image.
5-20 LOGIQ V2/LOGIQ V1 – Basic User Manual
Direction 5610735-100 English Rev. 2
Optimizing B-Mode
Gray Map
Description
Adjusting
Values
The system supplies B, M, and PW Mode system maps.
In Doppler mode, the Clear map offers a bluish coloring compared to the standard gray map. Clear Maps provide a more transparent map. Clear maps are displayed under gray maps.
To select a map, select the Gray Map. A map window displays.
The image reflects the map as you go through the selections.
Map values vary by probe, application, and frequency setting.
Map values are returned to the factory or user preset value when you change the following: Probe, Preset, Exam Calcs, or
Patient.
Rejection
Description
Benefits
Selects a level below which echoes will not be amplified (an echo must have a certain minimum amplitude before it will be processed).
Allows for the elimination of lower echoes (caused by noise) from the display.
Rotation
Description
Benefits
CAUTION
Flips the image 180 degrees up/down.
Beneficial in transvaginal scanning.
When reading a rotated image, be careful to observe the probe orientation to avoid possible confusion over scan direction or left/right image reversal.
Suppression
Description
Suppresses the noise in the image.
LOGIQ V2/LOGIQ V1 – Basic User Manual
Direction 5610735-100 English Rev. 2
5-21
Optimizing the Image
Needle Recognition
Description
Needle recognition allows you to obtain better needle imaging in the area surrounded by the yellow dashed line (Figure 5-3.a).
Needle recognition functionality is only available with linear probes (12L-RS, L6-12-RS).
Figure 5-3.b illustrates the position between the needle and probe. Needle recognition can only enhance in-plane needles.
The needle angle is defined as the angle between the needle and probe surface.
a
Figure 5-3. a. Enhanced area b. Needle Angle
2. Needle guide line
b
1. Needle
2. Probe
Adjusting
NOTE: Define a key from U1 - U4 on the control panel for activating/ deactivating Needle via Utility -> System -> User Configurable
Key.
• Activating Needle Recognition:
Press the key assigned for Needle on the control panel to activate Needle Recognition
• Adjusting steer direction:
Press the key assigned for Needle on the control panel to adjust the steer direction.
• Adjusting the needle recognition beam angle:
Adjust Beam Angle by rotating the Primary Menu control.
Max available angles are up to 50 degrees on 12L-RS, 30 degrees on L6-12-RS.
5-22 LOGIQ V2/LOGIQ V1 – Basic User Manual
Direction 5610735-100 English Rev. 2
Optimizing B-Mode
Needle Recognition (continued)
Preset
Benefits
Bioeffects
NOTE: Adjust the needle recognition steering angle to form the needle and the beam angle as perpendicular as possible to get the best needle enhancement.
NOTE: To help verify the location and trajectory of the needle tip, please use needle movement and / or fluid injection.
NOTE: Make sure the needle is always in the ultrasound plane by slightly moving or tilting the probe to get the best needle enhancement during the needle procedure.
Needle Direction - preset via Utility -> Imaging -> B Tab.
Beam Angle - preset via Utility -> Imaging -> B Tab.
Needle Gain - preset via Utility -> Imaging -> B Tab.
Needle Thickness - preset via Utility -> Imaging -> B Tab.
Provides better biopsy needle visualization than normal B-Mode with steered beam angle and post processing.
Activating angle Needle Recognition may change the TI and/or
MI. Observe the output display for possible effect.
LOGIQ V2/LOGIQ V1 – Basic User Manual
Direction 5610735-100 English Rev. 2
5-23
Optimizing the Image
LOGIQ View (Option)
Description
LOGIQ View provides the ability to construct and view a static
2D image which is wider than the field of view of a given transducer. This feature allows viewing and measurements of anatomy that is larger than what would fit in a single image.
Examples include scanning of vascular structures and connective tissues in the arms and legs.
LOGIQ View constructs the extended image from individual image frames as the operator slides the transducer along the surface of the skin in the direction of the scan plane. The quality of the resulting image is somewhat user-dependent and requires some additional skill and practice to develop proper technique and become fully proficient.
LOGIQ View is not available for the following:
• Color Flow Mode
Benefits
Clinical Use
Using LOGIQ View
The user can look at a larger region of interest within one field of view that is wider than any given probe would normally provide.
LOGIQ View is intended for scanning areas too large to fit on one image.
To perform an exam using LOGIQ View,
1. Perform a detailed examination of the anatomy/pathology.
Optimize parameters for tissue texture and visible window
PRIOR TO activating LOGIQ View.
2. Press the user defined key for LOGIQ View.
3. To start acquiring the image, press L key.
HINTS
When you scan, scan slowly and in a uniform motion lengthwise, end-to-end (with or against the probe orientation marker). LOGIQ View acquires images via leading edge vectors (and does not acquire slices, as in CINE). The image is being stored as you perform the scan and you can watch the
LOGIQ View as it is being acquired.
5-24 LOGIQ V2/LOGIQ V1 – Basic User Manual
Direction 5610735-100 English Rev. 2
Optimizing B-Mode
LOGIQ View (Option) (continued)
Uniform Motion
Bioeffects
4. To restart the scan, press L again. You can back up the probe, realign it, then go forward to redo a portion of the scan.
5. To complete the scan, press R or Freeze (or allow the scan to auto complete). The LOGIQ View is then displayed, scaled to fit entirely on the screen.
6. Perform measurements and record images.
The quality and usefulness of LOGIQ View images is affected by transducer motion. Incorrect technique can contribute to image distortion.
Guidance and precautions for uniform motion:
• Continuous contact is required throughout the length of the extended image. DO NOT lift the transducer from the skin surface.
• Always keep the transducer perpendicular to the skin surface. DO NOT rock the transducer.
• Keep the motion within the same scan plane, if possible. DO
NOT slide the transducer laterally.
• Lateral turning (change in direction to follow anatomical structure) can be accommodated with slower motion. DO
NOT make abrupt changes in direction.
• The system accommodates a reasonable range of motion velocity. DO NOT make abrupt changes in speed of motion.
Deeper scans generally require reduced speed.
Activating LOGIQ View has no affect upon Acoustic Output values.
LOGIQ V2/LOGIQ V1 – Basic User Manual
Direction 5610735-100 English Rev. 2
5-25
Download
Advertisement
Key features
Advanced imaging technologies
Intuitive user interface
Comprehensive set of features
Diagnostic examinations
Detailed reports
Frequently asked questions
The LOGIQ V2 and LOGIQ V1 ultrasound systems are intended for a variety of clinical applications, including abdominal, obstetric, gynecologic, cardiovascular, vascular, urologic, and pediatric examinations.
The LOGIQ V2 and LOGIQ V1 systems offer a range of advanced imaging technologies such as B-mode, M-mode, Color Flow, Power Doppler Imaging, and Spectral Doppler.
The user interface is designed to be intuitive and easy to use, with a clear and concise layout for efficient navigation and access to various functions.
Yes, you can customize the systems by creating presets for imaging parameters and adjusting various settings to optimize performance for different applications.