Maintenance

Replacing the Batteries

Refer to the safety information in Table 1 before replacing the batteries.

1. Turn off the thermometer if necessary.

2. Loosen the screw and remove the battery door.

3. Replace the three AA batteries.

4. Replace the battery door and tighten the screw.

Cleaning the Case and Holster

Use soap and water or a mild commercial cleaner.

Wipe with a damp sponge or soft rag.

51 & 52 Series II

Maintenance

Calibration

To ensure that the thermometer performs to its accuracy specifications, Fluke recommends that you calibrate the thermometer annually, starting one year after purchase.

To calibrate the thermometer, contact Fluke for the

Service Center nearest you or follow the calibration procedure in the service manual listed in "Replacement

Parts and Accessories."

Specifications

Environmental

Operating

Temperature

10 o

C to 50

o

C

(14

o

F to 122

o

F)

Storage

Temperature

Humidity

40

o

C to

+

60

o

C

(

40

o

F to

+

140

o

F)

Non condensing <10

o

C (<50

o

F)

95% RH: 10

o

C to 30

o

C (50

o

F to 86

o

F)

75% RH: 30

o

C to 40

o

C (86

o

F to 104

o

F)

45% RH: 40

o

C to 50

o

C (104

o

F to 122

o

F)

13

51 & 52 Series II

Users Manual

General

Weight

Dimensions

(without holster)

Battery

280 g (10 oz)

2.8 cm

×

7.8 cm

×

16.2 cm

(1.1 in

×

3 in

×

6.4 in)

Certification

Safety

3 AA batteries

P

,

) s

CSA C22.2 No. 1010.1 1992

EN 61010 Amendments 1, 2

CAT I

OVERVOLTAGE (Installation)

CATEGORY I, Pollution Degree 2 per IEC1010-1*

*

Refers to the level of Impulse Withstand Voltage protection provided. Equipment of OVERVOLTAGE CATEGORY I is equipment for connection to circuits in which measures are taken to limit the transient over voltages to an appropriate low level. Example include protect electronic circuits.

80 PK-1 Thermocouple (supplied with thermometer)

Type

Temperature

Range

Accuracy

Type K, Chromel Alumel, bead style

40 o

C to

+

260 o

C

(

40 o

F to

+

500 o

F)

±

1.1 o

C (

±

2.0 o

F)

Electrical

Measurement

Range

Display

Resolution

J-type:

210

(

346 o

C to

+

1200 o

F to

+

2192 o

F) o

C

K-type:

200 o

C to

+

1372 o

C

(

328 o

F to

+

2501 o

F)

T-type:

250 o

C to

+

400 o

C

(

418 o

F to

+

752 o

F)

E-type:

150 o

C to

+

1000 o

(

238 o

F to

+

1832 o

F)

C

0.1 o

C / o

1.0 o

C / o

F / K

<

1000 o

F / K

1000 o

14