Table 5 Meter diagnostics (continued)
Option Description
Display test Does a test on the display to make sure that the display is functional.
Event log Lets the user see, delete or export the event log. Export the event log to make the contents available as an accessible file through USB mass storage. This option is used primarily by factory service.
Troubleshoot errors
The meter and sensor contain no user-serviceable parts. For the errors and messages listed, try the corrective action.
If the problem does not go away or a problem occurs that is not in the list, contact the manufacturer.
Message or problem
Sensor is not connected
Value is out of range
Solution
Connect a sensor and try the action again.
Change the measurement parameters or put in a different value, then try the action again.
Clean the sensor and test.
Sensor data is known to be not correct or not accurate
Sensor is not recognized
Display is dim or is not visible
Data is not available or access to the data is not possible
Meter is unresponsive
Check the sensor connection. Make sure that the lock nut on the connection port is tight (finger-tighten only).
Push a key on the keypad.
Make sure that the USB option (Main Menu) is set to Mass Storage.
Push and hold the power button for at least 3 seconds. This deenergizes the meter. Energize the meter again.
Note: Do not use this method to power off while in normal operation or if the file access icon is visible in the display.
Replacement parts and accessories
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to the company website for contact information.
Replacement parts
Table 6 Velocity only sensor
Description
Sensor with 1.5 m (5 ft) cable
Sensor with 6.1 m (20 ft) cable
Sensor with 12.2 m (40 ft) cable
Sensor with 30.5 m (100 ft) cable
Item no.
1040500595-0N
1040500595-1N
1040500595-2N
1040500595-3N
Table 7 Velocity with depth sensor
Description
Sensor with 1.5 m (5 ft) cable
Sensor with 6.1 m (20 ft) cable
Item no.
1040500595-0D
1040500595-1D
English 29
advertisement
Related manuals
advertisement
Table of contents
- 5 Specifications
- 5 Sensor specifications
- 5 Portable meter specifications
- 6 User interface specifications
- 6 General specifications
- 6 General information
- 7 Safety information
- 7 Use of hazard information
- 7 Precautionary labels
- 7 Certification
- 8 Product overview
- 8 System overview
- 9 Sensor overview
- 10 Meter overview
- 10 Product components
- 11 Installation
- 11 Installation with optional accessories
- 11 Install the sensor on the universal sensor mount
- 12 Connect the sensor to the meter
- 12 Attach the lanyard
- 13 Attach the velcro strap
- 13 User interface and navigation
- 13 Keypad and key functions
- 14 Status bar
- 15 Navigation and Main Menu
- 19 Startup and self-test
- 19 Sleep mode
- 20 Operation
- 20 Stream profiles
- 20 Stations and station spacing
- 20 Measure velocity
- 21 Measure velocities in a cross-section
- 23 Insert or delete a station
- 24 Conduit profiles
- 24 0.9 x Vmax measurement method
- 25 0.2/0.4/0.8 method
- 25 Velocity/Level Integration measurement method
- 26 2D measurement method
- 27 Download data
- 27 Delete data files
- 27 Maintenance
- 27 Download the PVM utility
- 27 Update the firmware
- 28 Clean the sensor
- 29 Clean the meter
- 29 Install or replace the battery
- 30 Charge the battery
- 30 Troubleshooting
- 30 Diagnostics
- 31 Troubleshoot errors
- 31 Replacement parts and accessories
- 32 Appendix
- 32 Mean-section and Mid-section methods
- 34 Profiles and measurements
- 35 Site selection
- 35 Do a velocity calibration
- 35 Wet/Dry threshold